GE Healthcare Avance CS2: troubleshooting and error codes
The ServiceForge knowledge base holds 479 documented symptoms for the GE Healthcare Avance CS2, with 659 ranked causes, 850 diagnostic checks with test points and expected values, and 298 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
System not ready for operation patient safety
Equipment damaged, components incorrectly attached, breathing circuit damaged or disconnected, gas supplies not connected, cylinder valves open when pipeline in use, scavenging not properly connected, casters loose, brakes not set, power cord not connected
Ranked causes
- 1. Equipment is damaged (unknown)
- 2. Components incorrectly attached (unknown)
- 3. Breathing circuit damaged or incorrectly connected (unknown)
- 4. Pipeline gas supplies not connected (unknown)
- 5. Cylinder valves left open during pipeline use (unknown)
- 6. Scavenging not connected or adjusted (unknown)
- 7. Cylinder wrench missing (unknown)
- 8. Reserve O2 supply not connected (unknown)
- 9. Casters loose or brakes not set (unknown)
- 10. Power cord not connected to wall outlet (unknown)
Inspect equipment for damage
- Test point
- Visual inspection
- Expected
- No visible damage to equipment structure or components
9 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Inspiratory Flow Valve
Two-stage flow valve calibration to establish pressure-to-flow mapping across the full operating range using calibrated orifice reference
Ranked causes
- 1. Flow valve DAC counts >= 1000 during Stage 1 points 2-6 (unknown)
- 2. Leaks around test plugs or manifold (unknown)
- 3. Inspiratory Flow Valve not closing completely (leaky) (occasional)
- 4. Drive gas regulator not adjusted or unstable (occasional)
- 5. Inspiratory Flow Valve not linear (occasional)
- 6. Flow valve DAC reaches 4095 before Lift-Off Point (Stage 2) (unknown)
- 7. Inadequate drive gas supply (cannot deliver > 100 l/min) (occasional)
Leave Drive Port (A) and Manifold Port (B) on interface valve plugged
- Test point
- Vent Engine interface valve ports A and B
- Expected
- Ports remain sealed
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure leak caution
Gas escaping from high-pressure system components (pipeline, O2 flush valve, system switch, cylinder installation, transducer, gaskets, relief valves).
Ranked causes
- 1. Pipeline leak (unknown)
- 2. O2 flush valve defective (unknown)
- 3. System switch defective (unknown)
- 4. Cylinder not installed properly (unknown)
- 5. Cylinder transducer defective (unknown)
- 6. Cylinder gaskets defective (unknown)
- 7. Relief valves defective (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure leak detected patient safety
Gas escaping from high-pressure system components (pipeline, O2 flush valve, system switch, cylinder connection, transducer, gaskets, or relief valves)
Ranked causes
- 1. Pipeline leak (unknown)
- 2. O2 flush valve defect (unknown)
- 3. System switch defect (unknown)
- 4. Cylinder not installed properly (unknown)
- 5. Cylinder transducer defect (unknown)
- 6. Cylinder gaskets defective (unknown)
- 7. Relief valves defective (unknown)
Locate source of high pressure leak
- Test point
- All high-pressure system connections
- Expected
- No gas escaping when checked with leak detector or Snoop
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system circuit leak suspected patient safety
Low airway pressure during gas delivery setup; inability to maintain adequate breathing circuit pressure
Ranked causes
- 1. Flow sensor module defective or disconnected (common)
- 2. Circuit module internal leak or blocked ports (common)
- 3. Soda lime canister defective or improperly sealed (common)
- 4. Inspiratory side circuit disconnection or leak (occasional)
- 5. Negative pressure relief valve stuck open (occasional)
- 6. Flow sensor connector o-ring or internal tubing leak (occasional)
Verify flow sensor integrity without circuit module
- Test point
- Flow Sensor connector and housing
- Expected
- Bulb remains deflated or inflates in more than 30 seconds
- Tools
- Low-pressure leak test device (bulb)
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bleed Resistor
Bleed resistor calibration to establish pressure-to-flow table for system bleed path characterization
Ranked causes
- 1. Flow exceeds 50 l/min before 105 cmH2O pressure reached (unknown)
- 2. Leaks around test plugs (unknown)
- 3. Drive gas regulator not adjusted or unstable (occasional)
- 4. Inspiratory Valve Calibration not complete (occasional)
- 5. Inspiratory Flow Valve not closing completely (leaky) (occasional)
- 6. Inspiratory Flow Valve not linear (occasional)
Plug Drive Port (A) and Manifold Port (B) on interface valve
- Test point
- Vent Engine interface valve ports A and B
- Expected
- Ports sealed with plugs
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in Bag Mode patient safety
Full Test - Circuit Leak fails during system checkout. Bellows do not inflate properly or airway pressure cannot be maintained.
Ranked causes
- 1. Bag/Vent Switch lower seal leaking (common)
- 2. APL diaphragm leaking (occasional)
- 3. APL/BTV Manifold O-Ring seal compromised (occasional)
- 4. Bag port cover or APL valve defective (occasional)
- 5. Negative Pressure Relief Valve leaking (occasional)
Perform Test 4 - Testing bag port, APL valve, Bag/Vent switch, and negative pressure relief
- Test point
- Bag port connection
- Expected
- No pressure loss; bellows inflate when pressurized
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in both Bag and Vent Mode patient safety
Full Test fails in both Bag and Vent Mode. Common areas: Flow Sensor Module, Circuit Module, absorbent canister, negative pressure relief, or optional EZchange/Condenser Module.
Ranked causes
- 1. Flow Sensor Module or bulkhead connector O-ring defective (common)
- 2. Circuit Module O-ring or seal leaking (common)
- 3. Absorbent canister leak or improper installation (occasional)
- 4. Negative Pressure Relief Valve defective (occasional)
- 5. EZchange/Condenser Module or condenser assembly leaking (occasional)
Perform Test 9 - Testing flow sensor module, circuit module, and absorbent canister
- Test point
- Circuit interface ports
- Expected
- No pressure loss; bulb remains collapsed
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure leak without vaporizer caution
Gas escaping from low-pressure system without vaporizer or with vaporizer mounted (leaking port valve, mixer, relief valve, flush valve, system switch).
Ranked causes
- 1. Leaking port valve on vaporizer manifold (unknown)
- 2. Leak in mixer (unknown)
- 3. Leaking relief valve on vaporizer manifold (unknown)
- 4. Leaking flush valve (unknown)
- 5. Leaking system switch (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alternate O2 flowmeter not indicating correct flow patient safety
Flowmeter ball does not rise smoothly, does not maintain steady position, or does not indicate expected flow rate during Alternate O2 flow control operation.
Ranked causes
- 1. Flow control mechanism stiff or not rotating smoothly (occasional)
- 2. Flowmeter ball stuck or obstructed (occasional)
- 3. Alternate O2 flow not activating (occasional)
- 4. O2 supply pressure insufficient or not connected (occasional)
O2 supply connected and pressure available
- Test point
- O2 cylinder valve or pipeline inlet
- Expected
- Cylinder valve fully open or pipeline connected
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows collapsed patient safety
Unable to drive bellows, manifold pressure exceeds calculated limit
Ranked causes
- 1. Breathing circuit leak or hose occlusion (common)
- 2. Drive gas check valve failure (occasional)
- 3. cVIB cabling issue (occasional)
- 4. cVIB module failure (occasional)
Inspect breathing circuit
- Test point
- Breathing circuit hoses and connections
- Expected
- No leaks, no obstructions
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows falls during vertical seal test patient safety
Bellows assembly fails to remain inverted when ports are sealed with test plugs, indicating loss of seal integrity or component malfunction
Ranked causes
- 1. Ports not tightly sealed (common)
- 2. Bellows incorrectly installed (occasional)
- 3. Internal bellows seal incorrectly installed (occasional)
- 4. Parts are damaged (occasional)
Verify test plug seating in all ports
- Test point
- Bellows assembly ports
- Expected
- Test plugs fully seated and secure
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows falls past guide line during upright position test patient safety
Bellows assembly fails to remain extended to guide line when single port is sealed, indicating seal or pressure-relief valve malfunction
Ranked causes
- 1. Port not tightly sealed (common)
- 2. Bellows incorrectly installed (occasional)
- 3. Pressure-relief valve incorrectly installed (occasional)
- 4. Parts are damaged (occasional)
Verify test plug seating in upright position port
- Test point
- Bellows assembly port
- Expected
- Test plug fully seated and secure
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows fill time abnormal
Bellows takes longer or shorter than expected to fill when O2 Flush button is depressed continuously
Ranked causes
- 1. Large leak in breathing system (common)
- 2. Flush regulator setting incorrect (common)
- 3. Flush regulator cross-connection (occasional)
- 4. SCGO/ACGO selector valve inlet cross-connection (occasional)
Set up test apparatus
- Test point
- Bag/Vent switch and patient circuit
- Expected
- Bag/Vent switch in Vent position, patient circuit attached with patient port plugged, bellows completely collapsed
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system leak suspected patient safety
Airway pressure fails to rise or rises below expected threshold during breathing system leak test procedures.
Ranked causes
- 1. Leak in Bag/Vent Switch (common)
- 2. APL valve malfunction (common)
- 3. Bellows module leak or pop-off valve failure (common)
- 4. Flow sensor module or circuit module leak (common)
Verify airway pressure during controlled test
- Test point
- Airway Pressure on Gas Delivery Schematic
- Expected
- ≥ 30 cm H2O, ≥
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check flow sensor patient safety
Measured flow for 6 consecutive mechanical breaths does not meet criteria, or no/negative flow on inspiration/expiration
Ranked causes
- 1. Flow sensor connection issue (common)
- 2. Flow sensor calibration needed (common)
- 3. Inspiratory or expiratory check valve failure (occasional)
- 4. Flow sensor failure (occasional)
Check flow sensor connection
- Test point
- Flow sensor connectors and cables
- Expected
- All connections secure
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive fan noise
Fan generates excessive or abnormal noise during operation
Ranked causes
- 1. Display case fan inlet filter clogged (common)
- 2. Obstructions within case fan (common)
- 3. Com Express fan noise (occasional)
- 4. Defective fan assembly (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas module related errors in logs patient safety
Error log entries indicate issues with the Airway Module gas sampling or sensor functionality
Ranked causes
- 1. Sample gas line blocked or residue in inlet filter (common)
- 2. Sample line not connected (occasional)
- 3. Sensor inoperative (occasional)
- 4. Module not compatible (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High intrinsic PEEP during operation patient safety
Elevated positive end-expiratory pressure measured continuously during ventilation in bag or vent modes
Ranked causes
- 1. APL (adjustable pressure limit) valve sticking or obstruction downstream (common)
- 2. Exhalation valve or pop-off valve sticking or misassembly (common)
- 3. Scavenging system problem or breathing circuit obstruction (occasional)
- 4. Flow control valve malfunction or improper regulator setting (occasional)
Perform flow and pressure sensor calibration
- Test point
- Calibration software in service application
- Expected
- Calibration passes
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate volume ventilation measured patient safety
Set tidal volume does not match measured tidal volume; volume delivery discrepancy
Ranked causes
- 1. Flow sensor calibration out of range or sensor failure (common)
- 2. Inspiratory or expiratory check valve leak (occasional)
- 3. VIB board failure or pneumatic sense tubing leak (occasional)
- 4. Flow sensor failure or blockage (occasional)
Perform flow and pressure calibration
- Test point
- PC Service Application calibration module
- Expected
- Calibration passes
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in Vent Mode patient safety
Full Test - Vent and Gas fails during system checkout. Bellows do not fill or exhaust gas cannot be vented properly.
Ranked causes
- 1. Bellows assembly compromised (common)
- 2. Bellows pop-off valve leaking (common)
- 3. Bag/Vent Switch upper seal, disk, or O-ring defective (occasional)
- 4. Exhalation valve or drive circuit malfunction (rare)
Perform Test 6 - Testing bellows assembly and Bag/Vent switch
- Test point
- Bellows outlet port
- Expected
- Bellows fill and empty without pressure loss
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System will not boot from external USB Flash Drive during software installation
System fails to initiate boot process from USB Flash Drive when attempting software installation
Ranked causes
- 1. USB Flash Drive is incorrect part number or improperly inserted (common)
- 2. Primary USB port malfunction (occasional)
- 3. Defective USB Flash Drive (occasional)
- 4. Carrier Board or Com Express Board Assembly failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen fails to work
Touchscreen does not respond to touch input or is non-functional
Ranked causes
- 1. Screen lock is enabled (common)
- 2. Internal cables or connections disconnected or loose (occasional)
- 3. Touch Controller Board failure (rare)
- 4. Touch Bezel Assembly failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible beep not heard during system startup
Expected beep(s) from annunciators (speakers) during power-on sequence are missing or not audible
Ranked causes
- 1. Power Controller Board annunciator failure (unknown)
- 2. Anesthesia Control Board annunciator failure (unknown)
- 3. Display Unit annunciator failure (unknown)
Verify first audible beep heard immediately after System switch turned to On
- Test point
- System switch
- Expected
- Audible beep from Power Controller Board
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging
Battery status LEDs (CHG, FLOAT, OVER) not illuminating or batteries remain uncharged
Ranked causes
- 1. 28 VDC not reaching battery charging circuit (common)
- 2. Battery charging fuse FH2 or FH3 open (occasional)
- 3. U-Frame Power Supply not delivering 28 VDC (occasional)
Measure 28 VDC at J6 input
- Test point
- J6 connector
- Expected
- 28 VDC
- Tools
- DMM
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage outside acceptable range caution
Battery test shows voltage below minimum threshold or battery not holding charge
Ranked causes
- 1. Battery 1 (right) voltage below 12.10 Vdc (occasional)
- 2. Battery 2 (left) voltage below 12.10 Vdc (occasional)
- 3. Battery current abnormal or excessive discharge (rare)
Execute Battery test
- Test point
- AC mains power
- Expected
- AC mains shut off, wait 2 minutes for results
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows does not move patient safety
Ventilator bellows assembly does not rise and fall during ventilation attempt
Ranked causes
- 1. Drive gas regulator output pressure low or absent (common)
- 2. Gas inlet valve or solenoid failure (occasional)
- 3. Hospital pressure supply failure or inlet filter blockage (occasional)
Check drive gas regulator output pressure
- Test point
- Regulator output
- Expected
- 25 psi
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows leak caution
Gas escaping from bellows assembly due to pop-off valve diaphragm, mounting rim, or bellows damage.
Ranked causes
- 1. Pop-off valve diaphragm not sealing properly (unknown)
- 2. Bellows mounting rim loose (unknown)
- 3. Bellows improperly mounted or has hole or tear (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Primary regulators high voltage
Verify output pressure of O2, Air, or N2O primary regulators under low and high flow conditions; regulators must be tested after setup and adjusted if out of specification.
Ranked causes
- 1. Primary regulator output pressure out of specification at low flow (common)
- 2. Cylinder supply pressure decays too quickly or does not stabilize (occasional)
- 3. Primary regulator output pressure drops excessively at high flow (occasional)
Set system to Service mode and access Gas Delivery Schematic
- Test point
- Service Application (Section 12.3.2)
- Expected
- System in Service mode; schematic accessible
17 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
cVIB pneumatic sense tubing leak caution
Flow measurements jump or fail to stabilize near zero during pneumatic sense tubing leak test; sensor readings indicate flow when circuit should be sealed
Ranked causes
- 1. Leak in flow sensor pneumatic circuit (common)
- 2. Defective flow sensor (occasional)
- 3. Leak in sense line tubing (occasional)
Set fresh gas flow and vent configuration
- Test point
- Gas Delivery Schematic, PC Service Application; APL valve; Bag-to-Vent switch
- Expected
- Fresh gas flow 2 l/min; APL valve 0 cm H2O; Bag-to-Vent switch in Bag/APL position; short tube connected between inspiratory and expiratory flow sensors
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display arm positioning requires adjustment
Display arm movement, side-to-side drift, or tilt position needs calibration to maintain proper positioning and stability
Ranked causes
- 1. Arm attachment mounting nut loose (common)
- 2. Display mount attachment nut loose (common)
- 3. Display clamp assembly clamping screws loose (common)
Position display parallel to arm and apply finger pressure along left side
- Test point
- Left side of display
- Expected
- Display does not move
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure leak with vaporizer caution
Gas escaping from low-pressure system with vaporizer mounted (vaporizer installation, o-ring, fill port, port valve, mixer, relief valve, flush valve, system switch, bellows, pop-off valve, bellows mounting, breathing system).
Ranked causes
- 1. Vaporizer not installed properly (unknown)
- 2. Missing or damaged o-ring on vaporizer manifold (unknown)
- 3. Loose fill port (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mixer ACB communication failure
Anesthesia control board lost communication with gas mixer module or mixer not responding to commands
Ranked causes
- 1. Loose or damaged communication cable between ACB and mixer (common)
- 2. Mixer module not powered or internal failure (occasional)
- 3. ACB board communication port malfunction (rare)
Check mixer XMT/RCV LED activity
- Test point
- Mixer board transmission/reception LEDs
- Expected
- XMT or RCV LED indicates activity; not both absent
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ventilation delivery patient safety
Ventilator unable to deliver breath; bellows not moving or no fresh gas flow; respiratory gas delivery failure
Ranked causes
- 1. Drive gas regulator output pressure insufficient or zero (common)
- 2. No fresh gas flow or blender malfunction (common)
- 3. Bellows drive system failure or exhalation valve stuck (occasional)
Perform flow/pressure sensor calibration
- Test point
- PC Service Application sensor calibration
- Expected
- Calibration passes
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltage out of range high voltage
Diagnostic menu shows voltage readings outside normal ranges on power boards
Ranked causes
- 1. Power Controller Board supply fault (common)
- 2. Anesthesia Control Board power regulation failure (occasional)
- 3. Display Unit power distribution failure (occasional)
Verify Power Controller Board 12Vdc supply
- Test point
- Power Diagnostics menu - Power Board section
- Expected
- 11.70 to 12.30 Vdc, +0.30 / -0.30
24 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotary encoder fails to work
Rotary encoder does not respond to rotation or does not function
Ranked causes
- 1. Rotary encoder cable disconnected or improperly seated (common)
- 2. Rotary encoder failure (occasional)
- 3. Carrier Board or Com Express Board Assembly failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System application screen does not appear
Display does not show GE or Datex Ohmeda logo or system interface after power-on
Ranked causes
- 1. +12V LED not illuminated or power supply failure (common)
- 2. LCD touch bezel assembly failure (occasional)
- 3. Compact flash card or memory module failure (occasional)
Check green power indicator LED on dashboard
- Test point
- Dashboard green LED
- Expected
- Illuminated
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Checkout Stage 2 Step 8: Battery and Electrical status patient safety
Check AC/Mains power connection and battery charge status during system checkout
Ranked causes
- 1. AC/Mains disconnected (unknown)
- 2. Battery failure (unknown)
- 3. Battery still charging (unknown)
Verify AC/Mains power connection
- Test point
- AC/Mains power cord
- Expected
- Connected and active
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to begin mechanical ventilation patient safety
ABS not fully engaged or no O2 supply or defective Bag/Vent switch prevents mechanical ventilation.
Ranked causes
- 1. ABS not fully engaged (unknown)
- 2. No O2 supply (unknown)
- 3. Defective Bag/Vent switch (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alternate O2 flowmeter not indicating correctly patient safety
Float does not rise smoothly, does not maintain steady position, or indicates outside expected range during functional or accuracy testing
Ranked causes
- 1. Flowmeter mechanical obstruction or float sticking (common)
- 2. Flowmeter accuracy drift or calibration error (occasional)
Functional flowmeter response at minimum setting
- Test point
- Alt O2 flowmeter display
- Expected
- 0.5 to 0.7 l/min with flow control fully clockwise
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
App Not Loaded message is displayed
Display shows App Not Loaded message
Ranked causes
- 1. Software not loaded (common)
- 2. Defective CF card (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary O2 flowmeter accuracy out of specification
Flow test device reading does not fall within acceptable limits for set flowmeter values
Ranked causes
- 1. Flowmeter calibration drift (common)
- 2. Test device reading error or incorrect setup (occasional)
Measure flow at 1 L/min setting
- Test point
- Flow test device outlet
- Expected
- 0.5 to 1.5 L/min, +0.5 -0.5 L/min
- Tools
- Flow test device (0–15 L/min, ±2% accuracy)
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Balance gas selector valve leaks
Balance Gas Valve Leak test fails indicating selector valve or proportional valve not sealing
Ranked causes
- 1. Balance gas selector valve seal compromised (occasional)
- 2. Balance gas proportional valve seal degraded (occasional)
Balance Gas Valve Leak test
- Test point
- Balance gas selector valve outlet
- Expected
- Pass (valve holds pressure)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery status abnormal caution
Battery voltage out of range, charge state or discharge time unusual
Ranked causes
- 1. Battery voltage low or high (common)
- 2. Battery discharge capacity reduced (occasional)
Check Battery 1 voltage reading
- Test point
- Power Diagnostics menu - Power Board section
- Expected
- 10.00 to 16.50 Vdc (black display), <10.00 or >16.50 (red display), +3.50 / -3.00
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows does not fill within time specification caution
Flush Flow test shows bellows fill time outside 1.8 to 2.3 second range
Ranked causes
- 1. Gas supply pressure insufficient (occasional)
- 2. Flush valve obstruction or malfunction (occasional)
Perform Flush Flow test and time bellows fill
- Test point
- Bellows fill time
- Expected
- The bellows fills in 1.8 to 2.3 seconds, 1.8 to 2.3 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Board temperature excessive caution
Power Controller Board temperature exceeds safe operating limit
Ranked causes
- 1. Inadequate cooling or fan failure (common)
- 2. Fan 1 or Fan 2 voltage out of specification (occasional)
Monitor board temperature
- Test point
- Power Diagnostics menu - Power Board section
- Expected
- ≤75°C (black); >75°C (red fail), max 75
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system leak caution
Gas escaping from breathing system due to absorber canister or canister o-ring issues.
Ranked causes
- 1. Absorber canister open or missing (unknown)
- 2. Damaged or missing canister o-ring (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system leak detected patient safety
System or circuit leak alarm triggered; bellows falls or pressure not maintained in breathing circuit
Ranked causes
- 1. Loose or disconnected breathing circuit connection (common)
- 2. Leak in bellows base or exhalation valve seal (occasional)
Perform pneumatic sense tubing leak test on VIB
- Test point
- VIB pneumatic sense tubing
- Expected
- No leaks; both channels pass
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Manifold P Span
Manifold pressure transducer calibration to set 75 cmH2O reference point using drive gas pressure through inspiratory flow valve adjustment
Ranked causes
- 1. ADC value outside acceptable range (15740-20250 counts) (unknown)
- 2. Pressure transducer out of range limits (unknown)
Remove ABS breathing system, exhalation valve, and vent engine cover
- Test point
- Vent Engine interface valve
- Expected
- System disassembled and accessible
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Paw Span
Patient airway pressure (Paw) transducer calibration to set 75 cmH2O reference point using inspiratory flow valve adjustment with full circuit assembly
Ranked causes
- 1. ADC value outside acceptable range (21000-27000 counts) (unknown)
- 2. Pressure transducer outside range limits (unknown)
Leave Drive port (A) plugged
- Test point
- Vent Engine interface valve Drive port (A)
- Expected
- Port remains sealed
12 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU LED not illuminated
CPU LED on Power Controller board is off or not flashing, indicates no power or missing Boot Code
Ranked causes
- 1. No 28 VDC power from U-Frame Power Supply or Battery (common)
- 2. No Boot Code in Power Controller board (rare)
Measure 28 VDC input voltage
- Test point
- J6 connector
- Expected
- 28 VDC
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cylinder high pressure leak detected caution
Pressure decay in cylinder exceeds acceptable limit during one-minute hold test
Ranked causes
- 1. Cylinder valve seal leaking (occasional)
- 2. Connection fitting leak (occasional)
Test cylinder #1 for high pressure leaks
- Test point
- Cylinder #1 pressure gauge
- Expected
- Pressure decay after one minute is less than 690 kPa (100 psig)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cylinder pressure gauge does not indicate patient safety
Pressure gauge fails to show cylinder pressure when cylinder valve is opened
Ranked causes
- 1. Cylinder valve closed or not fully opened (common)
- 2. Cylinder empty or low pressure (occasional)
Open each cylinder valve one at a time
- Test point
- Gas Supply window pressure gauge
- Expected
- Corresponding gauge indicates cylinder pressure
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cylinder pressure not displayed
Gas Supply window shows no pressure reading when cylinder valve opened
Ranked causes
- 1. Cylinder valve not fully open (common)
- 2. Cylinder empty or pressure insufficient (occasional)
Cylinder valve open
- Test point
- Gas Supply window
- Expected
- Corresponding display indicates appropriate cylinder pressure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cylinder pressure not indicated on gauge patient safety
Cylinder pressure gauge does not show expected cylinder pressure when valve is opened
Ranked causes
- 1. Cylinder valve not opened or connection incomplete (common)
- 2. Cylinder empty or gauge malfunction (occasional)
Open each cylinder valve one at a time
- Test point
- Cylinder outlet
- Expected
- Corresponding pressure gauge indicates cylinder pressure
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display controller board ACB communication failure
Display unit unable to communicate with anesthesia control board; serial link failure
Ranked causes
- 1. ACB TXD/RXD LED not indicating activity or communication cable damage (occasional)
- 2. Display memory module or carrier board failure (occasional)
Observe ACB TXD/RXD LED activity
- Test point
- ACB board transmission/reception LEDs
- Expected
- Both TXD and RXD LEDs showing activity or one active at a time
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display unit requires service caution
Fan filter clogged, cable cover loose, or internal components need replacement
Ranked causes
- 1. Fan filter fouled (common)
- 2. Cable cover mounting loose (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive gas lost patient safety
Ventilator has no drive gas, O2 or Air supply low
Ranked causes
- 1. O2 supply disconnected or low pressure (common)
- 2. Air supply disconnected or low pressure (common)
Verify drive gas supply connection and pressure
- Test point
- O2 or Air supply inlet
- Expected
- Supply connected, pressure in range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fans not running in Standby mode caution
Fans do not turn on when system enters Standby position with AC Mains connected
Ranked causes
- 1. Power Controller board not receiving 28 VDC from U-Frame Power Supply (common)
- 2. Fan connector disconnected or failed fan motor (occasional)
Measure 28 VDC at J6 input
- Test point
- J6 connector on Power Controller board
- Expected
- 28 VDC from U-Frame Power Supply
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FiO2 High alarm does not occur patient safety
FiO2 High alarm fails to trigger when oxygen concentration rises above alarm threshold
Ranked causes
- 1. FiO2 sensor malfunction or disconnection (unknown)
- 2. Alarm limit not set to approved clinical values (unknown)
Trigger FiO2 High alarm condition
- Test point
- Ventilator oxygen delivery system
- Expected
- FiO2 High alarm occurs
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- FiO2 Low alarm does not occur
- Free breathing valve flapper damaged or contaminated
- Full Test - Circuit Leak check fails
- Full Test - Circuit O2 Cell reading unstable or incorrect
- Full Test - Vent and Gas check fails
- Gas supply system pressure reading incorrect
- High current to mixer 10VA
- High current to pressure transducers
- Machine Check - Circuit: Manual circuit leak test
- Machine Check - Circuit: O2 availability and flow test
- Mains indicator does not illuminate with power cord connected
- Mains power indicator does not illuminate
- Mixer command lost during transmission
- Mixer gas channel flow failure
- MVexp Low alarm does not occur
- O2 supply pressure low alarm not triggered
- Over Pressure Circuit test failure
- PEEP High alarm does not occur
- Peripheral circuit supply failure
- Pipeline supply not indicating facility pressure
- Power failure alarm does not cancel when power cord is plugged back in
- Power failure alarm does not display on battery
- Power failure alarm does not display when power cord is unplugged
- Power failure alarm not displayed or power indicator not active
- Ppeak High alarm does not occur
- Ppeak Low alarm does not occur
- RCV LED constantly on (no communication with Display Unit)
- Scavenger receiver replacement or cleaning
- Suction gauge does not respond to regulator adjustment
- System Checkout Stage 1 Step 2: Low O2 supply pressure
- System Checkout Stage 2 Step 2: Ventilator circuit leak detected
- System Checkout Stage 2 Step 5: O2 Flow mixer test
- System Checkout Stage 2 Step 6: AIR Flow mixer test
- System Checkout Stage 2 Step 7: N2O Flow mixer test
- Tec 7 vaporizer filler cap leaking or damaged
- Zero pressure transducers
- +12.0V H AMPS MGAS power issue
- +12.0V high current draw on mixer power
- +12.0V high current on Alt O2 system
- +12.0V high current on gas selection valves
- +12.5V supply to ACB failure
- AA, CO2 monitoring not connected
- ABS inspection and maintenance
- AC power inlet filter contamination
- AC power inlet filter dirty
- ACB clock speed error
- ACB CPU test failure
- ACB EEPROM failure
- ACB flash memory failure
- ACB hardware watchdog reset
- ACB microprocessor error
- ACB RAM error
- ACB redundant memory failure
- ACB software error
- ACB software watchdog failure
- ACB unexpected reset
- ACGO mode selected - volume and apnea monitoring off
- ACGO port flapper valve damaged or worn
- ACGO port flapper valve obstruction or debris
- ADC reference voltage out of range
- ADC timeout failure
- AGSS system not in working order
- Air pressure low during 21% O2
- Air supply pressure low
- Airway overpressure signal failure
- Airway sensor calibration error
- Alarm tests
- Alt O2 mode activated without operator action
- Alt O2 selection valve failure
- Alternate O2 button pressed
- Alternate O2 flowmeter accuracy out of range
- Alternate O2 flowmeter ball does not maintain steady position
- Alternate O2 flowmeter ball does not rise smoothly
- Alternate O2 flowmeter calibration and functional verification
- Alternate O2 flowmeter does not respond at minimum setting
- Alternate O2 flowmeter does not rise smoothly
- Alternate O2 flowmeter float does not respond to control rotation
- Alternate O2 flowmeter float does not rise
- Alternate O2 flowmeter float does not stop at bottom
- Alternate O2 flowmeter minimum flow outside spec
- Alternate O2 flowmeter not responding to control rotation
- Alternate O2 initial flow reading out of range
- Alternate O2 outlet blockage not detected
- Alternate O2 switch failure
- Alternate O2 switch non-responsive
- Anesthetic agent delivery during software download
- Annunciator beep missing after power-up delay
- Annunciator beep missing at System On
- Annunciators not functioning at startup
- Audible annunciator beep absent or incorrect after power-up
- Auxiliary O2 flowmeter accuracy out of range
- Auxiliary O2 flowmeter ball does not maintain steady position
- Auxiliary O2 flowmeter ball does not maintain steady position at set flow
- Auxiliary O2 flowmeter ball does not move at minimum setting
- Auxiliary O2 flowmeter ball does not rest at bottom when flow off
- Auxiliary O2 flowmeter ball does not rise smoothly
- Auxiliary O2 flowmeter ball does not rise smoothly on opening
- Auxiliary O2 flowmeter does not shut off
- Auxiliary O2 flowmeter float does not maintain steady position
- Auxiliary O2 flowmeter float does not move
- Auxiliary O2 flowmeter float does not respond after shutdown
- Auxiliary O2 flowmeter float does not respond to control
- Auxiliary O2 flowmeter float does not return to bottom when fully closed
- Auxiliary O2 flowmeter float stuck or does not rise
- Auxiliary O2 flowmeter flow does not increase smoothly
- Auxiliary O2 flowmeter not responding to control
- Auxiliary O2 flowmeter reading outside tolerance
- Auxiliary O2 outlet leak detected
- Auxiliary O2 outlet occlusion does not stop float
- Backup mode activated - apnea detected
- Balance channel flow control failure
- Balance channel proportional valve leak
- Balance gas flow does not meet specification at 10 LPM
- Balance gas flow does not meet specification at 3 LPM
- Balance proportional valve failure
- Balanced gas leak test skipped
- Battery charge failure
- Battery empty
- Battery low
- Battery missing
- Battery power unavailable during mains failure
- Battery reversed connections
- Battery voltage failure
- Battery voltage low
- Battery voltage very low
- Bellows assembly does not hold position
- Bellows assembly does not hold pressure
- Bellows assembly fails inspection
- Bellows falls more than 100 ml/min after initial drop
- Bellows fill rate test
- Bellows fills slowly when O2 Flush button pressed
- Bellows fills slowly with O2 Flush button
- Bellows leaks excessively
- Bellows module and bag vent function test
- Breathing system not latched
- Bulb inflation time excessive
- Cabinet, drawer, or frame parts loose or damaged
- Cabinet, frame, or drawer parts loose or damaged
- Calibration data failure in EEPROM
- Calibration error during airway module calibration
- Calibration: Airway module
- Calibration: Airway Pressure (Paw) Span
- Calibration: breathing circuit pressure test
- Calibration: Integrated Suction Regulator
- Calibration: Manifold Pressure Span
- Calibration: Mixer P Zero
- Calibration: O2 Flush Regulator
- Calibration: Positive pressure relief valve functional check
- Calibration: Scavenger Positive Relief Valve
- Calibration: User calibrations and system diagnostics
- Calibration: Zero Gas Transducers
- Carrier board and Com Express assembly replacement
- Carrier board overheat
- Case fan speed failure
- Caster malfunction or misalignment
- Caster operation or brake malfunction
- Casters not rolling or locking properly
- Change drive gas connection
- Check AC Power Inlet filter
- Check casters for proper operation and mounting
- Check external electrical cabling
- Check external tubing, rubber goods, hoses, and pipeline hoses
- Check labels for legibility
- Check mechanical parts condition and tightness
- Circuit Leak test fails
- Circuit O2 Cell 100% test fails
- Circuit O2 Cell 21% test fails
- Circuit O2 Cell test reading instability
- Communication error ventilator to ACB
- Compatibility failure no version info
- Compatibility failure software error
- Compatibility incomplete no versions from Vent SIB
- Compatibility incomplete no versions received from Vent SIB
- Configuration: Drive Gas and Altitude verification
- Consolidated Ventilator Interface board replacement
- Cooling fan current low failure
- CPU fan speed failure
- CPU overheat
- Cylinder pressure drops excessively
- Cylinder pressure test
- Cylinder supply high pressure leak
- DC commanded AC to failure
- DC-DC converter failure with battery in use
- DCB procedure communication failure
- Display fan filter clogged
- Display fan filter contamination
- Display Power fuse FH4 open
- Display Unit annunciator beeps missing during power-up
- Display Unit annunciator beeps missing during startup
- Display unit disassembly required
- Drive gas configuration mismatch
- Drive gas regulator pressure out of specification
- Electrical safety testing
- Electronic Gas Mixer assembly requires replacement
- Expiratory check valve leaking
- External electrical cabling damaged or disconnected
- External electrical cabling damaged or improperly connected
- External tubing or hoses deteriorated or loose
- External tubing, hoses, or pipeline deteriorated
- Fan filter clogged or missing
- Flow and Pressure test fails
- Flow and pressure test failure
- Flow sensor readings exceed ±10% of reference device
- Flow valve current feedback failure
- Flows do not drop to zero when O2 supply fails
- Full system test required
- Gas flows do not drop to zero when O2 supply fails
- Gas inlet valve feedback closed state
- Gas Inlet Valve Feedback does not read Closed
- Gas Inlet Valve Feedback test fails
- Gas supply not visible on display
- Ground resistance exceeds specification
- Ground resistance out of specification
- Ground resistance out of tolerance
- Hard keys not responding
- High airway overpressure signal detected
- High current to alternate O2 valve
- High current to consolidated ventilator interface board
- High current to gas select valves
- High current to MGAS 10VA
- High current to vent and outlet valves
- High pressure cylinder leak test
- Incompatible software detected during download
- Inspiration flow valve 10VA overcurrent
- Inspiratory check valve leaking
- Inspiratory port occlusion
- Integrated suction regulator calibration
- Integrated suction regulator does not reduce to minimum vacuum
- Integrated suction regulator gauge does not rise smoothly
- Integrated Suction Regulator gauge does not rise with adjustment
- Integrated suction regulator gauge exceeds 200 mmHg during test
- Integrated suction regulator gauge out of calibration
- Integrated Suction Regulator gauge rises when inlet occluded at minimum
- Integrated suction regulator pressure exceeds 200 mmHg at minimum setting
- Integrated Suction Regulator vacuum exceeds minimum setting
- Intermittent breathing system leak
- Invalid air pipeline measurement
- Keypad assembly replacement
- Labels illegible or missing
- Labels missing or illegible
- Labels not legible
- LCD display not functioning
- LCD/Touch Bezel assembly replacement
- Leakage current exceeds limit
- Leakage current exceeds specification
- LEDs not illuminating
- Loss of gas delivery user settings
- Loss of ventilation parameter settings
- Low drive gas pressure alarm
- Low O2 supply alarm not triggered on disconnect
- Low O2 supply alarm occurs when pipelines disconnected
- Low P Leak check fails (ACGO)
- Low P Leak check fails (SCGO)
- Low voltage to alternate O2 valve
- Low voltage to cVIB
- Low voltage to vent and outlet valves
- Low-pressure leak check (ACGO)
- Low-pressure leak check (SCGO)
- Machine Check - Circuit: Airway pressure too high
- Machine Check - Circuit: Alt O2 Flow control test
- Machine Check - Circuit: Bag/Vent and ACGO switch verification
- Main power failure - system on battery
- Mains indicator does not illuminate
- Mechanical parts loose or damaged
- Memory failure - EEPROM
- Minimum alternate O2 flow setting incorrect
- Missing second audible beep during power-up
- Missing third or fourth audible beep during power-up
- Mixed fresh gas selection valve overcurrent
- Mixer balance gas check valve leak rate exceeds specification
- Mixer balance gas selection failure
- Mixer channel zero calibration failure
- Mixer channel zero calibration failure due to leak
- Mixer CPU test failure
- Mixer CRC EEPROM failure
- Mixer CRC flash failure
- Mixer flow attainment failure balance gas channel
- Mixer flow attainment failure channel 1
- Mixer flow check failure
- Mixer flow inaccurate on air
- Mixer flow inaccurate on N2O
- Mixer flow inaccurate on O2
- Mixer flow sensor failure balance gas channel
- Mixer flow sensor failure channel 1
- Mixer flow verification failure balance gas channel
- Mixer flow verification out of limits
- Mixer gas channel flow control failure
- Mixer gas channel proportional valve failure
- Mixer gas channel selection valve failure
- Mixer gas channel temperature limit exceeded
- Mixer gas channel temperature sensor malfunction
- Mixer outlet check valve leaking
- Mixer outlet check valve leaks
- Mixer outlet check valve leaks excessively
- Mixer outlet flow inaccurate
- Mixer POST failure during power-up
- Mixer pressure sensor P1/P3 failure
- Mixer pressure sensor P2 failure
- Mixer RAM failure
- Mixer software watchdog failure
- Mixer temperature sensor failure balance gas channel
- Mixer temperature sensor failure channel 1
- Mixer voltage supply failure
- MOPV pressure relief valve malfunction
- MOPV pressure relief valve not functioning
- MOPV pressure relief valve test fails
- No AA data
- No audible beep after system switch on
- No CO2 data or waveform
- No flow waveform displayed
- No fresh gas flow - aux outlet circuit feedback fault
- No measured Vte
- No O2 data
- No PEEP pressure waveform displayed
- Non-Circle SCGO selected
- O2 alarm verification with Airway Module
- O2 bypass valve leaks
- O2 flow does not meet specification at 10 LPM
- O2 flow does not meet specification at 3 LPM
- O2 Flush button friction or wear
- O2 Flush Regulator pressure out of specification
- O2 proportional valve leaks
- O2 selector valve leaks
- O2 sensor not connected
- O2 supply alarm does not occur
- Over-pressure circuit does not function
- Overdue 12-hour power cycle self test
- Oxygen cell calibration failure
- Oxygen leak test skipped
- Pan enclosure fan requires replacement
- PEEP high blockage alarm does not occur
- PEEP/PCV not available
- Pipeline and cylinder pressure test
- Pipeline pressure gauge not indicating correctly
- Pipeline pressure not displayed
- Power Controller communication failure
- Power Controller communication failure with internal failure
- Power Controller self-test failure
- Power failure alarm does not activate
- Power failure alarm does not display
- Power failure alarm response
- Power supply temperature high
- Ppeak low leak alarm does not occur
- Pressure limit circuit test fails
- Pressure relief valve test
- Pressure transducer out of specification
- Reverse expiratory flow detected
- Scavenger positive relief functional check
- Scavenger positive relief valve not functional
- Second annunciator beep missing during startup
- Severe sustained airway pressure
- Speaker not producing sound
- Standby patient detection failure
- Suction control valve not delivering adequate flow
- Suction flow rate below minimum
- Suction gauge does not bleed to zero on OFF
- Suction gauge needle moves with knob in OFF position
- Suction gauge needle not at zero when idle
- Suction gauge needle not at zero when off
- Suction regulation unstable
- Suction regulator does not respond to adjustment
- Suction regulator pressure exceeds limit during operation
- Suction regulator pressure not holding setpoint
- Suction regulator vacuum accuracy out of range
- Suction regulator vacuum gauge not zeroing
- Suction system has vacuum leak with regulator off
- Suction system vacuum does not return to zero after shutdown
- Suction vacuum gauge reading inaccurate
- Sustained high airway pressure
- System Checkout Stage 1 Step 1: Bag/Vent Switch incorrect
- System Checkout Stage 1 Step 3: Ventilator has no drive gas
- System Checkout Stage 2 Step 1: Cannot empty bellows
- System Checkout Stage 2 Step 3: Mechanical Ventilation test
- System Checkout Stage 2 Step 4: Circuit Compliance test
- System date and time incorrect
- System flushing with oxygen after leak test
- System malfunction - ACB clock speed incorrect
- System malfunction - ACB communication failure
- System malfunction - ACB CPU test failure
- System malfunction - ACB Display Controller communication failure
- System malfunction - ACB flash memory CRC failure
- System malfunction - ACB hardware watchdog failure
- System malfunction - ACB microprocessor exception
- System malfunction - ACB procedure communication failure
- System malfunction - ACB RAM error
- System malfunction - ACB redundant memory failure
- System malfunction - ACB software error
- System malfunction - ACB software watchdog failure
- System malfunction - ACB unexpected reset
- System power-on tests failed
- Temperature measurement channel 1 high
- Temperature measurement channel 2 high
- Touch points are misaligned
- Touch screen calibration misaligned
- Touchscreen failure
- Unit does not power on
- Vaporizer alarms triggered during back pressure test
- Vaporizer back pressure and interlock malfunction
- Vaporizer system issues during concentration adjustment
- Vent and Gas test fails
- Ventilation -6V supply failure
- Ventilation +12.5V supply failure
- Ventilation +6V supply failure
- Ventilation 1.22V supply failure
- Ventilation ADC reference voltage failure
- Ventilation check stage 1 failed
- Ventilation check stage 2 failed
- Ventilation flow valve DAC feedback failure
- Ventilation valve 10VA power supply failure
- Ventilator free breathing valve requires maintenance
- Ventilator interface board 10VA overcurrent
- Ventilator interface board communication failure
- Verify and reset Altitude setting
- Verify and reset Drive Gas setting
- Verify audible beeps during power-up
- Visual inspection of display fan filter
- Voltage out of range to ACB
- Zero calibration: Gas transducers
- Zero calibration: Mixer Pressure
419 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
+12.0V H AMPS ALT O2 | High current to alternate O2 valve detected | Alternate O2 valve +10VA turned on | warning |
+12.0V H AMPS ALT O2 | Alternate oxygen system +12V overcurrent; excessive current drawn | +12V supply to Alt O2 circuit drops due to overcurrent | error |
+12.0V H AMPS GAS SEL VALVES | High current to gas select valves detected | Fresh gas select valves +10VA is turned on | warning |
+12.0V H AMPS GAS SEL VALVES | Gas selection valve circuits drawing excessive current; overcurrent condition | +12V supply to gas selection solenoids drops due to overcurrent | error |
+12.0V H AMPS MGAS | High current to MGAS 10VA detected; gas monitoring not available | MGAS 10VA turned on after 3 consecutive ACB MGAS error log messages | warning |
+12.0V H AMPS MGAS | High current condition on +12V Airway Module power rail | Power supply detects excessive current draw from Airway Module | critical |
+12.0V H AMPS MIXER | Mixer +12V power supply overcurrent condition; excessive current detected | +12V supply voltage drops or overcurrent protection activates | critical |
+12.0V H AMPS MIXER | High current to mixer 10VA supply detected | Mixer +10VA turned on | warning |
+12.0V H AMPS VENT&OUTLET VALVES | High current to vent and outlet valves 10VA detected | Ventilator valves +10VA is on | warning |
+12.0V H AMPS VENTSIB | High current to consolidated ventilator interface board 10VA detected | cVIB 10VA is turned on | error |
+12.0V L VOLT ALT O2 | Low voltage to mixer alternate O2 valve detected | ACB detects low voltage condition | warning |
+12.0V L VOLT GAS SEL VALVES | Gas selection valve circuits receiving low +12V supply; voltage below acceptable range | Gas selection solenoid supply voltage below minimum threshold | error |
+12.0V L VOLT VENT&OUTLET VALVES | Low voltage to vent and outlet valves detected | GIV or inspiratory flow valve drawing excess power | warning |
+12.0V L VOLT VENTSIB | Low voltage to cVIB detected | ACB detects low voltage condition to cVIB | warning |
+12.5V TO ACB | Power controller board unable to supply +12.5V to anesthesia control board | ACB power supply voltage below specification; power delivery failure | error |
+12.5V TO ACB | Voltage to ACB out of specification range | Power supply voltage less than 11.9 or greater than 12.9 Vdc | error |
12 HR TEST | System power-up self test overdue | System has been operating longer than 12 hours without power cycle | info |
+5V H AMP GAS SUPPLY XDUCERS | High current to pressure transducers detected; cannot read gas supply pressures | Pressure transducer +10VA turned on | warning |
AA, CO2 MONITORING NOT CONNECTED | Anesthetic agent or CO2 monitoring module not detected | System unable to communicate with monitoring module at startup or during operation | error |
ACB 4.096V ADC REF | ADC reference voltage out of specification; cannot monitor gas supplies | Reference voltage less than 4.018 or greater than 4.176 Vdc | info |
ACB ADC FAIL | ADC timeout detected; cannot monitor gas supplies | ADC timeout on any MUX channel | info |
ACB CLOCK SPEED | AC clock frequency incorrect; system malfunction | Clock frequency greater than 1.1× expected or less than 0.9× expected | error |
ACB CLOCK SPEED | Anesthesia control board clock speed out of specification; timing error detected | CPU clock frequency measurement outside normal range during self-test | error |
ACB COM FAIL | System malfunction; display controller not receiving messages from anesthesia computer | DC does not receive messages from AC for 10 seconds after initial communication established | error |
ACB COM FAIL | ACB communication failure | Loss of communication with ACB | critical |
ACB CPU TEST FAIL | Anesthesia control board CPU self-test failed; processor malfunction | CPU built-in self-test (BIST) detects processor fault during startup | error |
ACB DCB COM FAIL | System malfunction; anesthesia computer not receiving system state messages from display controller | AC receives no system state messages from DC for 10 seconds after initial communication established | error |
ACB DCB COM FAIL | ACB-DCB communication failure | Loss of communication between ACB and DCB | critical |
ACB EEPROM FAIL | Anesthesia control board EEPROM memory failure; data storage error | EEPROM read/write test failed; memory chip unresponsive | error |
ACB EEPROM FAIL | Memory failure; EEPROM read/write or CRC failure on anesthesia control board | EEPROM read/write failure or CRC failure | info |
ACB FLASH FAIL | System malfunction; CRC failure in ACB code space | CRC failure detected in flash memory code space | error |
ACB FLASH FAIL | Anesthesia control board flash memory failure; program storage error | Flash memory read test failed; program code unreadable | error |
ACB HW WATCHDOG | ACB hardware watchdog timeout | Watchdog timer expired, system hung | critical |
ACB HW WATCHDOG | ACB hardware watchdog circuit triggered reset; system lockup or hang occurred | ACB processor did not reset watchdog timer before timeout | error |
ACB MICROPROC ERROR | System malfunction; unexpected microcontroller exception | Bus error, address error, or other microcontroller exception | error |
ACB MICROPROC ERROR | Anesthesia control board microprocessor error; CPU fault detected | Processor exception or fault condition during operation | error |
ACB PROCEDURE COM FAIL | System malfunction; anesthesia computer not receiving procedure messages from display controller | DC receives no procedure messages from AC for 15±1 seconds, or procedure message mismatch for 15±1 seconds | error |
ACB RAM ERROR | System malfunction; memory test failure with multiple bit errors | Multiple bit errors detected during RAM test | error |
ACB RAM ERROR | ACB RAM memory error | RAM test failure or memory parity error | critical |
ACB RAM ERROR | Anesthesia control board RAM memory failure; read/write error detected | RAM memory self-test failed; memory chip unresponsive | error |
ACB REDUNDANT MEMORY FAIL | ACB redundant memory verification failed; data mismatch detected | Redundant memory copies do not match; data integrity check failed | error |
ACB REDUNDANT MEMORY FAIL | System malfunction; redundantly stored parameter corrupted or not stored | Redundant parameter storage or retrieval failure | error |
ACB SW ERROR | System malfunction; unexpected software error | Software error detected during execution | error |
ACB SW ERROR | Anesthesia control board software execution error; firmware fault | Software exception or illegal instruction execution detected | error |
ACB SW WATCHDOG | ACB software watchdog timer triggered; processor hang or deadlock detected | Software watchdog timer expired; ACB firmware did not reset watchdog | error |
ACB UNEXPECTED RESET | System malfunction; unexpected reset of anesthesia control board | Unexpected reset event detected | error |
ACB UNEXPECTED RESET | Anesthesia control board experienced unexpected reset; no watchdog trigger | Reset detected without hardware or software watchdog being triggered | warning |
ACGO | Volume and apnea monitoring off; non-circle mode selected | Non-circle (ACGO) mode selected in system configuration | info |
ACMains POWER FAIL | Plug in power cable; system on battery | ACMains_GOOD goes low and stays low for at least 300 msec | warning |
AIR PIPE INVALID | Cannot monitor air pipeline; invalid air pipeline pressure reading | Air pipeline pressure measurement invalid | warning |
AIR PRESS LOW | Air supply pressure low | Air pipeline pressure less than 252 kPa and air cylinder pressure less than 2633 kPa for 1 second | warning |
AIR PRESS LOW DURING 21% O2 | Air pressure low; increase O2 percentage | Air pipeline pressure less than 252 kPa and air cylinder pressure less than 2633 kPa during 21% O2 mode | error |
AIRWAY SENSOR CAL ERROR | Calibrate flow sensors; airway pressure sensor zero offset out of range | Airway pressure sensor zero offset out of specified range | info |
ALT O2 BUTTON PRESSED | Alternate O2 button is selected; alternate O2 screen displayed | Alternate O2 button selected by operator | info |
ALT O2 SEL VALVE FAILURE | Alternate oxygen selection valve malfunction; unable to select Alt O2 source | Alt O2 selector solenoid unresponsive; valve does not switch | error |
ALT O2 SWITCH FAIL | Alternate O2 switch status indicates fault condition | Alternate O2 switch fault condition persists for 1 second | warning |
Anesthesia Control Board | Communications status of Anesthesia Control Board | System checks communication link during Tools — Communication Status | unknown |
APNEA | No breathing detected for specified time interval; patient not breathing | Respiratory rate drops below configured apnea threshold or time exceeds limit | critical |
APNEA > 120s | Apnea condition detected for longer than 120 seconds; prolonged no breathing | No breathing detected for more than 120 seconds during operation | critical |
App Not Loaded | Application failed to load on display | Software not loaded or CF card defect | warning |
AUX OUTLET FAIL | No fresh gas flow; auxiliary outlet circuit feedback fault | Measured SCGO position does not match commanded position | error |
BACKUP MODE ENTERED | Backup mode active; apnea detected | No spontaneous breaths for set apnea time and 30 seconds elapsed in PSVPro mode | info |
BAL CHANNEL PROP VALVE LEAK FAIL | Balance proportional valve leak detected; alternate O2 screen displayed | Mixer detects proportional valve leak on balance channel | warning |
BAL FLOW CTRL FAIL | Balance channel flow control failure | Mixer status bit STS_FLOW_CTRL_CH2_FAIL indicates flow attainment failure | warning |
BAL PROP VALVE FAIL | Balance channel proportional valve failure (overcurrent, etc.) | Mixer status bit STS_CH2_PROPN VALVE FAIL indicates valve failure | warning |
BATTERY CHARGE FAIL | No battery backup, battery charge current excessive | System in standby with charge current >4.0 A or powered on with battery current >1.7 A | warning |
BATTERY EMPTY | Battery empty, system shutdown in less than 5 minutes | Available battery power between 1 and 5 minutes | error |
BATTERY FAIL | No battery backup, battery voltage or charging anomaly detected | Battery voltage <10.5 V or voltage >16.5 V during bulk/over charging, or excessive bulk charging time | warning |
BATTERY LOW | Battery low, plug in power cable | Available battery power between 20 and 30 minutes | warning |
BATTERY MISSING | No battery backup, battery not connected | Battery voltage reading between ±1.0 VDC | warning |
BATTERY REVERSED CONNECTIONS | No battery backup, battery connected with reversed polarity | Battery voltage reading less than –1.0 VDC | warning |
Battery test FAILED | One or both battery voltages are below minimum threshold or battery test did not complete | Battery 1 (right) <12.10 Vdc OR Battery 2 (left) <12.10 Vdc | error |
BATTERY V LOW | Battery voltage low, plug in power cable | Available battery power between 10 and 20 minutes | warning |
BATT V VERY LOW | Battery voltage very low, plug in power cable | Available battery power decreases to between 10 and 5 minutes | warning |
BELLOWS COLLAPSED | Unable to drive bellows | Manifold pressure exceeds Paw + 10 + [0.25*(Insp valve flow)] | info |
Breathing system loose | Breathing circuit connection loose or partially disconnected during operation | Leak detected in breathing circuit connections; pressure drop in circuit | error |
BREATHING SYSTEM NOT LATCHED | Breathing system loose or not properly latched | Breathing system detection switch indicates system not latched | info |
CAL DATA FAILURE IN EEPROM | Service calibration advised due to corrupt calibration data | Default calibration data in use due to corrupt data in calibration region | info |
Calibrate flow sensors | Flow sensor calibration is due or required; sensor accuracy uncertain | Calibration interval exceeded or sensor reading inconsistent during checkout | warning |
CARRIER BOARD OVERHEAT | Cooling fans may overheat due to high Display Unit temperature | Display Unit thermistors exceed 60 °C | warning |
CASE FAN SPEED FAIL | Cooling fans may overheat due to case fan speed failure | Case fan speed less than 800 RPM (50% of normal speed) | warning |
CHECK FLOW SENSOR | Flow sensor reading out of range during mechanical ventilation | Measured flow for 6 consecutive breaths does not meet criteria, or no/negative flow on inspiration/expiration | warning |
Check flow sensors | Flow sensor diagnostic test failed; sensor not responding within specification | Flow sensor malfunction detected during system checkout or operation | warning |
Circuit check failed | System checkout failure detected | Message or failure displayed during system checkout | unknown |
Circuit Leak | Air leak detected in anesthesia breathing circuit; loss of pressurization | Pressure or flow imbalance detected in inspiration vs. expiration | error |
Circuit O2 check skipped | O2 circuit check not performed during system checkout | Message displayed during system checkout | unknown |
COM ERROR VENT TO ACB | System malfunction, communication lost between ventilator and ACB | Ventilator boundary object receives no messages from cVIB CPU for 35 milliseconds | error |
Compatibility failure: No version info in file for subsystem | Subsystem did not report compatibility information | Subsystem version information missing from file | unknown |
Compatibility failure: Software Error | Subsystem did not report compatibility information | Software error preventing subsystem from reporting version | unknown |
Compatibility incomplete: No versions from Vent SIB | Consolidated Ventilator Interface Board compatibility information incomplete | cVIB compatibility info does not match or GIV test failed | unknown |
Compatibility incomplete: No versions received from Vent SIB | Consolidated Ventilator Interface Board compatibility information not received | cVIB compatibility info does not match or GIV test failed | unknown |
Controls (Front Panel) Board | Communications status of Controls (Front Panel) Board | System checks communication link during Tools — Communication Status | unknown |
COOLING FAN CURRENT LOW FAILURE | Mixer Pan Fan drawing too little current | Mixer Pan Fan current below normal threshold | warning |
CPU FAN SPEED FAIL | Cooling fans failed, CPU may overheat | CPU fan speed less than 50% of nominal speed | warning |
CPU OVERHEAT | Cooling fans may overheat due to high CPU temperature | Display Unit thermistor temperature reading >70 °C | warning |
DCB PROCEDURE COM FAIL | System malfunction, procedure communication failure between computers | No procedure messages from Display Computer for >15 seconds or message mismatch for >15 seconds | error |
DC COMMANDED AC TO FAILURE | Display Controller detected issues and commanded safe state | Display Controller initiated safe state transition | unknown |
DRIVE GAS LOST | Ventilator has no drive gas | O2 or Air supply low or disconnected during mechanical ventilation | error |
EC 06 | System Malfunction | Self test failure or multi bit error detected | critical |
EC 06 | Error code 06 | Condition triggering EC 06 | error |
ERROR!! INCOMPATIBLE SOFTWARE. | Software version on USB Media is not compatible with installed subsystem | Software Download detects incompatible subsystem during download process | error |
***ERROR: No bootable device available. | No valid bootable device found during startup | System startup cannot locate bootable operating system | critical |
***ERROR: Program load failed - CRC. | Software file checksum error detected during load | System startup detects CRC mismatch in program file | error |
***ERROR: RAM memory error. | Random access memory hardware failure detected | System startup POST detects RAM failure | critical |
***ERROR: System reset: ECxx xx xx . . . | System reset due to software or boot failure with error code | System detects software fault or boot failure during startup | error |
***ERROR: System reset: FFFF FF FF | System reset due to critical battery or power failure | System startup detects Carrier board battery contact loss or below voltage | critical |
***ERROR: watchdog circuit failed. | Watchdog timer circuit hardware failure | System startup POST detects watchdog circuit malfunction | critical |
EXP FLOW SENSOR CAL ERROR | Calibrate flow sensors | Exp Flow Sensor zero offset out of range | warning |
EXP FLOW SENSOR EEPROM FAILURE | Replace exp flow sensor | EEPROM cal data read failure | warning |
FAN FAIL | Cooling fan needs service. System OK | Fan Power Status Bit is Low (FAN1_GOOD) | warning |
FiO2 High | Oxygen fraction above acceptable level | Oxygen delivery or measurement exceeds maximum threshold | warning |
FiO2 Low | Oxygen fraction below acceptable level | Oxygen delivery or measurement falls below minimum threshold | warning |
Flow Attain Fail Bal Gas Channel | Unable to achieve commanded flow on balance gas channel; flow control failed | Balance gas flow unable to reach setpoint despite proportional valve movement | error |
Flow Attain Fail Bal Gas Channel | Flow attainment failure on balance gas channel | Unable to attain set flow on balance gas | error |
Flow Attainment Failure Ch1 | Flow attainment failure on channel 1 | Channel 1 unable to attain set flow | error |
Flow Attainment Failure Ch1 | O2 flow mismatch | Commanded flow through O2 Gas channel does not match measured flow via sensor and transducers | error |
Flow Attainment Failure Ch1 | Unable to achieve commanded flow on channel 1; flow control failed | Channel 1 flow unable to reach setpoint despite proportional valve adjustment | error |
FLOW SENSOR CAL ERROR | Calibrate flow sensors | Insp or Exp flow sensor or airway or manifold pressure sensor zero offset out of range | warning |
Flow Sensor Fail Balance Gas Ch | Flow sensor for balance gas channel failed; sensor not responding | Balance gas flow sensor malfunction detected during system check | error |
Flow Sensor Failure Ch1 | Flow sensor for channel 1 (primary gas) failed; sensor not responding | Channel 1 flow sensor malfunction detected during system check | error |
Flow Sensor Failure Ch1 | Flow sensor failure on channel 1 | Channel 1 flow sensor malfunction | error |
FLOW VALVE CURRENT FAILURE | Insp Flow Valve current feedback incorrect | Current feedback from Insp Flow Valve incorrect for seven consecutive readings | error |
FLOW VALVE DAC FAILURE | Insp Flow Valve DAC failure | Current feedback from Insp Flow Valve incorrect for seven consecutive readings | error |
Flow Verif Fail (dP) Bal Gas Ch | Balance gas flow verification failed; differential pressure reading incorrect | Mixer unable to verify balance gas flow using differential pressure method | error |
Flow Verification Failure (dP) Ch1 | O2 pressure transducer mismatch | Commanded O2 flow and Hot-wire anemometer agree but pressure transducers do not agree | error |
FRONT PANEL COM FAIL | Display panel controls failure | Key pad controller fails to send 'life tick' for greater than 10 seconds | warning |
FRONT PANEL KEY STUCK | Stuck Keypad or encoder on Display | Stuck Keypad or encoder detected | warning |
GAS INLET VALVE BOOT UP TEST FAIL | Vent Fail. Monitoring Only | Boot-up test failed | critical |
Gas Module | Communications status of Gas Module | System checks communication link during Tools — Communication Status | unknown |
INSP FLOW SENSOR CAL ERROR | Calibrate flow sensors | Insp Flow Sensor zero offset out of range | warning |
INSP FLOW SENSOR EEPROM FAILURE | Replace insp flow sensor | EEPROM cal data read failure | warning |
Inspiration Stopped | Inspiratory phase unexpectedly terminated; ventilation cycle incomplete | Flow or pressure signal lost during inspiration; control system error | error |
Link Status | Overall communication link status | System checks communication during Tools — Communication Status | unknown |
LOSS OF GAS DELIVERY USER SETTINGS | Loss of gas delivery user settings | User settings lost or corrupted | warning |
LOSS OF GAS DELIVERY USER SETTINGS | Gas delivery user settings lost; saved parameters not recovered after reset | System restart or error recovery; settings not found in ACB memory | warning |
LOSS OF GAS DELIVERY USER SETTINGS / LOSS OF VENT PARAMETER | Vent Fail. Monitoring Only | Gas Delivery User Setting or Ventilator Parameter Settings from Display Computer arrived more than 10 seconds ago | critical |
LOSS OF VENT DELIVERY USER SETTINGS | Ventilation delivery user settings lost; saved parameters not recovered after reset | System restart or error recovery; ventilation settings not found in ACB memory | warning |
LOSS OF VENT PARAMETER SETTINGS | Loss of ventilation parameter settings | Vent parameters lost or corrupted | warning |
Low O2 supply alarm | Oxygen supply pressure below operational threshold | Pipeline supplies disconnected and all cylinder valves closed | warning |
Low Pressure Leak check failed | System Checkout Failed | Low Pressure Leak section of System Checkout Failed | warning |
Low Pressure Leak check fails | Leak measured greater than 50 ml/min | Leak measured is greater than 50 ml/min during Low Pressure Leak Check with SCGO | warning |
MANIFOLD PAW SENSOR FAIL | Vent Fail. Monitoring Only | Calibration failure at bootup | warning |
MANIFOLD PRESSURE SENSR FAILURE | Calibration failure at bootup | Manifold pressure sensor calibration failure at bootup | warning |
MANIFOLD SENSOR CAL ERROR | Manifold Pressure zero failed | Manifold Pressure zero calibration failure | warning |
MGAS CHECK SAMPLE GAS OUT >20 SEC | Sample gas check line taking longer than 20 seconds to respond | Airway Module unable to verify sample gas delivery within expected time | warning |
MGAS CHECK SAMPLE GAS OUT >20 SEC | Check sample gas out | MGAS continuous occlusion bit set for 20 seconds | warning |
MGAS INLET FILTER RESIDUE >40 SEC | Replace D-Fend | MGAS communicates Replace Trap alarm bit for 40 seconds | warning |
MGAS INLET FILTER RESIDUE >40 SEC | Inlet filter residue check taking longer than 40 seconds | Excessive residue detected in Airway Module inlet filter | warning |
MGAS LINE BLOCKED >20 SEC | Gas sample line is blocked | Airway Module detects blockage in gas sampling line | error |
MGas module removed | Airway Module has been physically removed from system | System detects Airway Module is no longer present | warning |
MGAS SAMPLE FLOW DEVIATION | Low gas sample flow | Sample flow less than 80% of nominal for 20 consecutive seconds | warning |
MGAS SAMPLE LINE NOT CONNECTED >40 S | Sample line not connected or connection lost | Airway Module unable to detect sample line connection after 40 seconds | warning |
MGAS SAMPLE LINE NOT CONNECTED >40 S | Check D-Fend | MGAS communicates OpenGasCircuit alarm for 40 seconds | warning |
MGAS SENSOR INOP > XX | Airway Module sensor is inoperative | Sensor malfunction detected during system check | error |
MGAS SENSOR INOP > XX | Module fail. No CO2, AA, O2 data | MGAS communicates hardware failure or UPI does not initialize | warning |
MIXER ACB COM FAIL | Mixer unable to communicate with anesthesia control board; serial link failure | ACB sends command to mixer but receives no response; timeout occurs | error |
MIXER ACB COM FAIL | Alternate O2 Screen | Five seconds pass without measured flow data from mixer | warning |
MIXER ACB COM FAILLLOST CMD | Alternate O2 Screen | Mixer lost AC flow commands for 5 seconds or received illegal commands | warning |
MIXER ACB COM FAIL LOST CMD | Mixer communication lost during command transmission; packet loss or corruption | Mixer command transmission started but not completed; link interruption | error |
MIXER AIR SELECTION VLV FAIL | Alternate O2 Screen | Air selector valve status does not match commanded state | warning |
MIXER BAL GAS CHANGE FAIL | Balance gas selection failure; mixer unable to switch balance gas channel | Mixer selection valve unable to change balance gas source | error |
MIXER BAL GAS CHANGE FAIL | Alternate O2 Screen | After balance gas change command, selector valve shows old balance gas still connected | warning |
MIXER BAL GAS CHANGE FAIL | Mixer balance gas change failure | Unable to change balance gas setting | error |
MIXER BAL GAS FLOW CHECK FAIL | Mixer status bit | Balance gas proportional valve fails flow check | warning |
MIXER BALGAS Flow FAIL | Mixer error bit | Pressure difference between P3 and P2 differs from expected at measured flow for channel 2 | warning |
MIXER BAL GAS F SENSOR FAIL | Alternate O2 Screen | Balance gas flow sensor failure | warning |
MIXER BAL GAS TEMP LIMIT | Mixer error bit | Balance gas temperature exceeds 50 °C | warning |
MIXER BAL GAS TSENSOR FAIL | Mixer error bit | Balance gas temperature sensor failure | warning |
MIXER BAL GAS ZERO FAILLLEAK | Mixer status bit | Balance gas proportional valve fails zero flow check, shows flow while closed | warning |
Mixer Board | Communications status of Mixer Board | System checks communication link during Tools — Communication Status | unknown |
MIXER CPU TEST FAIL | Mixer CPU self-test failed during startup; processor malfunction | Mixer processor self-test detects CPU defect during POST | error |
MIXER CPU TEST FAIL | Alternate O2 Screen | Mixer CPU test failure | warning |
MIXER CRC EEPROM FAIL | Mixer EEPROM checksum error; data integrity check failed | EEPROM CRC calculation does not match stored checksum | error |
MIXER CRC FLASH FAIL | Mixer flash memory checksum error; program code integrity check failed | Flash memory CRC calculation does not match stored checksum | error |
MIXER N2O SELECTION VLV FAIL | N2O selector valve status does not match commanded state | Mixer Status Bit STS_SELV_VN2O_NOTIFY_FAIL | error |
MIXER O2 FLOW CHECK FAIL | O2 proportional valve fails flow check | Mixer status bit STS_FLOW_TEST_CH1_FAIL1 LPM_FLOW_TEST_FAIL | error |
MIXER O2 Flow FAIL | Alternate O2 Screen | Pressure difference differs from expected at measured flow for Channel 1 (O2) | warning |
MIXER O2 F SENSOR FAIL | O2 flow sensor failure | Mixer error bit STS_F2F1_SENSOR_FAIL (O2) | error |
MIXER O2 SELECTION VLV FAIL | O2 selector valve status does not match commanded state | Mixer Status Bit STS_SELV_VOXY_NOTIFY_FAIL | error |
MIXER O2 TEMP LIMIT | O2 temperature exceeds 50 °C | Mixer error bit STS_CH1_TEMP_LIMIT (O2) | error |
MIXER O2 TSENSOR FAIL | O2 temperature sensor failure | Mixer error bit STS_T1_SENSOR_FAIL | error |
MIXER O2 ZERO LEAK FAIL | O2 proportional valve fails zero flow checks for leaks | Mixer status bit STS_CH1_ZERO_FLOW_TESP ROPN_V_LEAK_FAILT_FAIL | error |
MIXER POST FAIL | Mixer power-on self-test (POST) failed during system startup | Mixer initialization test failed during system boot sequence | error |
MIXER PRES SENSOR P1P3 FAIL | Mixer pressure sensor P1/P3 malfunction; one or both sensors not responding | Pressure sensor P1 or P3 returns invalid or out-of-range reading | error |
MIXER P SENSOR P2 FAIL | Mixer pressure sensor P2 malfunction; sensor not responding or reading invalid | Pressure sensor P2 returns invalid or out-of-range reading during diagnostics | error |
MIXER RAM FAIL | Runtime walking pattern check on RAM failed | Mixer Status Bit STS_RAMPATTERN_FAIL | error |
MIXER RAM FAIL | Mixer RAM memory failure; self-test detected defective memory | Mixer RAM self-test failed during startup or operation | error |
MIXER SW WDOG | Mixer Watchdog has been activated | Watchdog timer expiration | warning |
MIXER SW WDOG FAIL | Mixer software watchdog timer triggered; processor hang or firmware deadlock | Mixer watchdog timer expired; firmware did not reset watchdog counter | error |
MIXER VOLTS FAIL | Mixer power supply (on board) is out of tolerance | Status bit STS_VOLT_REF_FAIL. +12.5 V (10 VA) to mixer not OK. | error |
MIXER XX FLOW CHECK FAIL | Flow verification failed on specified gas channel; measured flow not as expected | Mixer flow verification test shows actual flow does not match commanded value | error |
MIXER XX FLOW CHECK FAIL | Mixer gas-specific flow check failure | Flow check on specified gas line failed | error |
MIXER XX Flow FAIL | Flow failure on specified gas channel; unable to deliver requested flow | Proportional valve unable to achieve commanded flow on specified channel | error |
MIXER XX Flow FAIL | Mixer gas-specific flow failure | Unable to deliver specified gas flow | error |
MIXER XX F SENSOR FAIL | Flow sensor failure on specified gas channel (XX = BAL GAS, O2, AIR, or N2O) | Flow sensor malfunction detected on specified channel during diagnostics | error |
MIXER XX F SENSOR FAIL | Mixer gas-specific flow sensor failure | Flow sensor on specified gas line malfunction | error |
MIXER XX SELECTION VLV FAIL | Selection valve failure on specified gas channel; valve unable to switch | Mixer selection valve malfunction; solenoid unresponsive to control signal | error |
MIXER XX SELECTION VLV FAIL | Mixer gas-specific selection valve failure | Selection valve for specified gas malfunction | error |
MIXER XX TEMP LIMIT | Mixer gas-specific temperature limit exceeded | Temperature on specified gas line exceeded limit | warning |
MIXER XX TEMP LIMIT | Gas temperature on specified channel exceeded safe operating limit | Temperature sensor reading above maximum safe temperature threshold | critical |
MIXER XX TSENSOR FAIL | Temperature sensor failure on specified gas channel (XX = BAL GAS, O2, etc) | Temperature sensor malfunction on specified channel during startup | error |
MIXER XX TSENSOR FAIL | Mixer gas-specific temperature sensor failure | Temperature sensor on specified gas line malfunction | error |
MIXER XX ZERO FAIL LEAK | Zero calibration failed on specified channel due to detected internal leak | Mixer zero calibration unable to pass; internal leak detected on channel | error |
MIXER XX ZERO FAIL LEAK | Mixer gas-specific zero calibration failure, possible leak | Zero calibration failed or gas leak detected | warning |
MIX FRSH GAS SLCT VLV 10VA OVR CUR | Fresh gas selection valve overcurrent; solenoid drawing excessive current | +12V supply to fresh gas selection valve shows overcurrent or voltage drop | error |
MIX O2 BYPASS VLV 10VA OVER CURR | Current feedback from O2 Bypass Selector incorrect for 7+ readings | Excessive current readings from O2 Bypass Selector Valve | error |
MODULE NOT COMPATIBLE | Installed Airway Module is not compatible with system | System detects incompatible module firmware or hardware revision | error |
MVexp low | Expiratory minute volume below acceptable level | Measured expiratory minute volume falls below minimum threshold | warning |
N2O PRESS LOW | N2O supply pressure low | N2O pipeline <252 kPa and cylinder <2633 kPa | warning |
Negative airway pressure | Negative pressure detected in airway; potential PEEP valve failure or obstruction | Pressure transducer reads below zero during exhalation phase | error |
NO EXPIRATORY FLOW SENSOR | No Expiratory sensor connected and not calibrating | Expiratory sensor not detected during startup | error |
No Fresh Gas Flow! | Fresh gas flow not detected during ventilation; gas delivery failure | No flow measured from gas outlet during ventilation attempt | critical |
NO INSP/EXP FLOW SENSOR | No inspiratory/expiratory flow sensor detected or functional during ventilator circuit test | Full Test - Vent and Gas execution | warning |
NO INSP/EXP FLOW SENSOR | Flow sensors have been removed from circuit | When instructed to calibrate flow sensors; message appears after sensor removal | info |
NO INSPIRATORY FLOW SENSOR | No Inspiratory sensor connected and not calibrating | Inspiratory sensor not detected during startup | error |
***NOTE: Alarm speaker not detected. Check connection. | Speaker connection fault detected at startup | System startup POST detects no speaker or failed connection | warning |
***NOTE: CMOS battery is weak. Please replace. | Carrier board CMOS battery voltage insufficient | System startup POST detects weak CMOS battery | warning |
***NOTE: CPU Board Supply Voltage Out of Range. | CPU power supply voltage outside specification | System startup POST detects 1.5V_CB, 5V_CB, or 3.3V_CB out of range | warning |
***NOTE: RTC date/time error. Battery may be weak. | Real-time clock date/time not set or invalid | System startup POST detects RTC error | warning |
O2 CAL ERROR | O2 sensor calibration offset, slope, cell voltage not in range or O2 >110% | Galvanic O2 sensor calibration failure | warning |
O2 FLOW CTRL FAIL | O2 flow control failure | Mixer status bit STS_FLOW_CTRL_CH1_FAIL with O2 gas supply pressure OK | error |
O2 FLUSH FAILURE | O2 flush switch detected on continuously >30 sec | Switch reads on for >30 sec without cycling | warning |
O2 PIPE INVALID | Cannot monitor O2 pipeline; O2 Pipeline pressure is invalid | O2 pressure transducer malfunction | error |
O2 PRESS LOW | O2 supply pressure low | O2 pipeline <252 kPa and O2 cylinder <2633 kPa for 1 sec | critical |
O2 PROP VALVE FAIL | O2 proportional valve failure | Mixer status bit STS_CH1_PROPN VALVE FAIL | error |
O2 SENSOR FAILURE | O2 sensor failure; O2 <5% | Galvanic O2 sensor connected and O2 reading low | warning |
O2 SENSOR NOT CONNECTED | Oxygen sensor not connected to Airway Module | Airway Module unable to detect O2 sensor connection | warning |
ON/STANDBY SWITCH TO STANDBY | On/Standby switch transitions from On to Standby | Switch moves to Standby position during therapy | critical |
PATIENT VOLUME MISMATCH OCCURRED | Patient Volume Mismatch alarm occurred; calibrate, dry, or replace flow sensors | Inspiration/expiration volume discrepancy during checkout | warning |
PCB Alarm Off: 1 min time left | Battery alarm displayed during system checkout | Low battery time detection during checkout | info |
PCB Alarm Off: AC Mains failure | AC Power Removed | AC mains voltage loss | info |
PCB Alarm Off: DC-DC power module failure | System on battery despite AC connected (battery status LOW) | AC GOOD HIGH but BATT STAT 1 and 2 LOW | info |
PCB Alarm On: AC mains failure | AC Power Removed | AC mains voltage loss | info |
PCB Alarm On: Bulk charge time exceeded in stdby | Batteries in bulk charge >12 hours while in Standby | Extended bulk charge duration in Standby mode | error |
PCB Saved Alarm Off: 1 min time left | Battery alarm displayed during system checkout | Low battery time detection during checkout | info |
PCB Saved Alarm Off: AC Mains failure | AC Power Removed | AC mains voltage loss | info |
PCB Saved Alarm Off: Blk chrg time exceed in stdby | Batteries in bulk charge >12 hours while in Standby | Extended bulk charge duration in Standby mode | error |
PCB Saved Alarm Off: DC-DC power module failure | Current draw from batteries >0.25A for 2 min while AC connected | Excessive battery current drain with AC mains present | error |
PCB Saved Alarm On: 1 min time left | Battery alarm displayed during system checkout | Low battery time detection during checkout | info |
PCB Saved Alarm On: AC Mains failure | AC Power Removed | AC mains voltage loss | info |
PCB Saved Alarm On: DC-DC power module failure | Current draw from batteries >0.25A for 2 min while AC connected | Excessive battery current drain with AC mains present | error |
PCB Saved Alarm On: DU to PSC Comm Error | Power Controller to Display Unit communication was lost then restored | Communication loss and subsequent recovery | info |
PCB Saved Error: POWER CNTRL COM FAIL | Communication lost between Display Unit and ACB; error stored until re-established | Communication loss between Display Unit computer and ACB | error |
PEEP high. Blockage? | Positive end-expiratory pressure elevated; possible obstruction detected | PEEP rises above set value, indicating possible blockage | error |
PEEP High. Blockage? | Measured PEEP higher than set value; possible breathing circuit obstruction | Positive end-expiratory pressure exceeds commanded setpoint | error |
PEEP PCV NOT AVAILABLE | Volume ventilation only; no PEEP or PSV available | Paw data in range but Pmanifold <= -15 cmH2O | warning |
Plug in power cable. On battery. | Power failure; system operating on battery power | Mains power disconnected while system is in operation | critical |
PORT PLUG POST FAIL | Ventilation fail; monitoring only | Paw > 100 cmH2O for 1 second in POST state | warning |
Power Controller Board | Communications status of Power Controller Board | System checks communication link during Tools — Communication Status | unknown |
POWER CONTROLLER COM FAIL | Internal failure; system may shut down | Communications with PC and DC cannot be established for 10 seconds | error |
POWER CONTROLLER SELF TEST | Internal failure; Power Controller fails self-tests | Power Controller fails self-tests | critical |
POWER SUPPLY 75C | Circuitry >75°C; shutdown possible | Power supply temperature exceeds 75°C | warning |
Ppeak high | Peak inspiratory pressure higher than expected; possible resistance or obstruction | Measured peak pressure exceeds acceptable range for ventilation parameters | error |
Ppeak low. Leak? | Peak inspiratory pressure below expected level; possible leak detected | Peak pressure falls below set value, indicating possible circuit leak | warning |
Pressure Sensor 1/2/3 Failure | Pressure transducer (P1/P2/P3) on Mixer out of specification | Pressure transducer out of specification | error |
PWR CNTRL DC-DC FAIL | Using battery; PC fail. AC supply OK but system reports using battery | AC GOOD HIGH but BATT STAT 1 and 2 LOW | warning |
Quick check fails | Message or failure displayed during system checkout | Failure or message during system checkout | error |
Replace insp / exp flow sensor | Inspiratory or expiratory flow sensor has failed and requires replacement | Flow sensor failure confirmed during diagnostic testing | error |
REVERSE EXPIRATORY FLOW | Reverse expiration flow detected; check valves may be faulty | Flow towards patient on both expiratory and inspiratory sensors during inspiration for 6 consecutive mechanical breaths | warning |
Reverse insp / exp Flow | Inspiratory and expiratory flow sensor channels may be reversed or misconfigured | Flow sensor readings inconsistent with breathing pattern during checkout | warning |
SCGO | Non-Circle SCGO selected; volume and apnea monitoring off | Non-Circle SCGO mode selected | info |
SEVERE SUSTAINED PAW | Measured airway pressure >100 cmH2O for 10 seconds | Airway pressure exceeds 100 cmH2O for 10 seconds | error |
STANDBY PATIENT DETECTION | No fresh gas flow; patient detected during standby | 3 volume breaths or 3 CO2 breaths detected within 30 seconds during system checkout | info |
System Leak | Leak detected in ventilator or breathing system pneumatics; gas escape detected | Bellows pressure or flow imbalance; breathing system leak detection | error |
System Self-tests failed | Power-on tests failed | Power-on tests fail | error |
Temp Measurement Ch1 High | Channel 1 temperature measurement above safe operating range; sensor reading high | Temperature sensor on channel 1 reading above acceptable threshold | error |
Temp Measurement Ch2 High | Channel 2 temperature measurement above safe operating range; sensor reading high | Temperature sensor on channel 2 reading above acceptable threshold | error |
Temp Sensor Fail Balance Gas Ch | Temperature sensor failure on balance gas channel; sensor malfunction | Balance gas temperature sensor unresponsive during system startup | error |
Temp Sensor Failure Ch1 | Temperature sensor failure on channel 1; sensor malfunction detected | Channel 1 temperature sensor unresponsive or returns invalid reading | error |
TOUCHSCREEN FAILURE | Touchscreen controller fails communications POST test | Touch panel controller fails POST communications test | warning |
Tv not achieved | Tidal volume not reached; measured volume less than set volume | Ventilator unable to deliver programmed tidal volume during breathing cycle | error |
Unable to Drive Bellows | Ventilator bellows cannot be driven; no bellows movement during ventilation attempt | Drive gas pressure absent or insufficient; bellows mechanical failure | critical |
Valve Ch1 Leak Tests Not Done | O2 leak test skipped | O2 leak test skipped (no O2 supply at power-up) | info |
VENT 1.22V FAIL | Ventilation fail; monitoring only | Voltage <1.074 Vdc or >1.367 Vdc | warning |
VENT +12.5V FAIL | Ventilation fail; monitoring only | Nominal 12.5V <11.3 Vdc or >13.13 Vdc | warning |
VENT -6V FAIL | Ventilation fail; monitoring only | VSIB -6V out of range <-6.72 Vdc or >-5.28 Vdc | error |
VENT +6V FAIL | Ventilation fail; monitoring only | VSIB +6V out of range <5.51 Vdc or >6.5 Vdc | error |
VENT ADC VREF FAIL | Ventilation fail; monitoring only | VSIB ADC 3.200V ref voltage <3.179 or >3.221 Vdc | error |
VENT AIRWAY OVERPRESS SIGNAL | Inspiration stopped; high airway overpressure signal set | High airway overpressure signal set during mechanical ventilation | warning |
VENT AIRWAY OVERPRESS SIGNAL FAIL | Ventilation fail; monitoring only | High airway overpressure signal set with Paw <90 cmH2O and Pmanifold <80 cmH2O | warning |
Vent check stage 1 failed | Failure displayed during system checkout stage 1 | Failure during checkout stage 1 | error |
Vent check stage 2 failed | Failure displayed during system checkout stage 2 | Failure during checkout stage 2 | error |
VENT FLOW VALVE FAIL CURRENT | Ventilation fail; monitoring only | Incorrect current feedback for 7 consecutive readings | warning |
VENT FLOW VALVE FAIL DAC | Ventilation fail; monitoring only | Incorrect DAC feedback for 3 consecutive readings | warning |
Ventilate Manually | Automatic ventilation failed; manual ventilation mode recommended or required | Ventilator subsystem failure; unable to provide mechanical ventilation | error |
Ventilator has no Drive Gas | Drive gas supply to ventilator bellows is absent or insufficient for operation | Drive gas pressure below minimum threshold during ventilation attempt | critical |
Ventilator Interface Board | Communications status of Ventilator Interface Board | System checks communication link during Tools — Communication Status | unknown |
VENT SIB 10VA OVERCURRENT | Current feedback from cVIB incorrect | Incorrect current feedback from cVIB | error |
VENT SIB COMMUNICATION FAILURE | Loss of communication between ACB and cVIB | Loss of communication between ACB and cVIB after initial establishment | error |
VENT SUSTAINED PAW SDOWN | Ventilation fail; monitoring only | Paw >100 cmH2O for 10 seconds | warning |
VENT VALVE 10VA OVERCURRENT | Current feedback from Insp Flow Valve incorrect for 7 consecutive readings | Incorrect current feedback from Insp Flow Valve | error |
VENT VALVE POWER FAIL | Ventilation fail; monitoring only | Nominal 12.5V <11.3 V or >13.13 Vdc | warning |
Vlv Bal Gas Ch Leak Tests Not Done | Balanced Gas leak test skipped | Balanced Gas leak test skipped (no Balanced Gas at power-up) | info |
Vol vent only. No PEEP or PSV | Volume control ventilation available but no PEEP or PSV modes working; feature unavailable | Drive gas or control system failure; mode selection error | error |
XX FLOW CTRL FAIL | Flow control failure on specified gas channel (XX = BAL GAS, O2, AIR, N2O) | Proportional valve unable to reach commanded flow setpoint | error |
XX FLOW CTRL FAIL | Gas-specific flow control failure | Flow control on specified gas line malfunction | error |
XX PROP VALVE FAIL | Proportional valve failure on specified gas channel; valve not responding to control | Proportional valve solenoid or electronics malfunction; no valve response | error |
XX PROP VALVE FAIL | Gas-specific proportional valve failure | Proportional valve on specified gas line malfunction | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community