FR FieldReady Academy
Anesthesia machines · Avance family

GE Healthcare Avance CS2: troubleshooting and error codes

The ServiceForge knowledge base holds 479 documented symptoms for the GE Healthcare Avance CS2, with 659 ranked causes, 850 diagnostic checks with test points and expected values, and 298 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

System not ready for operation patient safety

Equipment damaged, components incorrectly attached, breathing circuit damaged or disconnected, gas supplies not connected, cylinder valves open when pipeline in use, scavenging not properly connected, casters loose, brakes not set, power cord not connected

Ranked causes

  1. 1. Equipment is damaged (unknown)
  2. 2. Components incorrectly attached (unknown)
  3. 3. Breathing circuit damaged or incorrectly connected (unknown)
  4. 4. Pipeline gas supplies not connected (unknown)
  5. 5. Cylinder valves left open during pipeline use (unknown)
  6. 6. Scavenging not connected or adjusted (unknown)
  7. 7. Cylinder wrench missing (unknown)
  8. 8. Reserve O2 supply not connected (unknown)
  9. 9. Casters loose or brakes not set (unknown)
  10. 10. Power cord not connected to wall outlet (unknown)
First check

Inspect equipment for damage

Test point
Visual inspection
Expected
No visible damage to equipment structure or components

9 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Inspiratory Flow Valve

Two-stage flow valve calibration to establish pressure-to-flow mapping across the full operating range using calibrated orifice reference

Ranked causes

  1. 1. Flow valve DAC counts >= 1000 during Stage 1 points 2-6 (unknown)
  2. 2. Leaks around test plugs or manifold (unknown)
  3. 3. Inspiratory Flow Valve not closing completely (leaky) (occasional)
  4. 4. Drive gas regulator not adjusted or unstable (occasional)
  5. 5. Inspiratory Flow Valve not linear (occasional)
  6. 6. Flow valve DAC reaches 4095 before Lift-Off Point (Stage 2) (unknown)
  7. 7. Inadequate drive gas supply (cannot deliver > 100 l/min) (occasional)
First check

Leave Drive Port (A) and Manifold Port (B) on interface valve plugged

Test point
Vent Engine interface valve ports A and B
Expected
Ports remain sealed

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure leak caution

Gas escaping from high-pressure system components (pipeline, O2 flush valve, system switch, cylinder installation, transducer, gaskets, relief valves).

Ranked causes

  1. 1. Pipeline leak (unknown)
  2. 2. O2 flush valve defective (unknown)
  3. 3. System switch defective (unknown)
  4. 4. Cylinder not installed properly (unknown)
  5. 5. Cylinder transducer defective (unknown)
  6. 6. Cylinder gaskets defective (unknown)
  7. 7. Relief valves defective (unknown)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure leak detected patient safety

Gas escaping from high-pressure system components (pipeline, O2 flush valve, system switch, cylinder connection, transducer, gaskets, or relief valves)

Ranked causes

  1. 1. Pipeline leak (unknown)
  2. 2. O2 flush valve defect (unknown)
  3. 3. System switch defect (unknown)
  4. 4. Cylinder not installed properly (unknown)
  5. 5. Cylinder transducer defect (unknown)
  6. 6. Cylinder gaskets defective (unknown)
  7. 7. Relief valves defective (unknown)
First check

Locate source of high pressure leak

Test point
All high-pressure system connections
Expected
No gas escaping when checked with leak detector or Snoop

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system circuit leak suspected patient safety

Low airway pressure during gas delivery setup; inability to maintain adequate breathing circuit pressure

Ranked causes

  1. 1. Flow sensor module defective or disconnected (common)
  2. 2. Circuit module internal leak or blocked ports (common)
  3. 3. Soda lime canister defective or improperly sealed (common)
  4. 4. Inspiratory side circuit disconnection or leak (occasional)
  5. 5. Negative pressure relief valve stuck open (occasional)
  6. 6. Flow sensor connector o-ring or internal tubing leak (occasional)
First check

Verify flow sensor integrity without circuit module

Test point
Flow Sensor connector and housing
Expected
Bulb remains deflated or inflates in more than 30 seconds
Tools
Low-pressure leak test device (bulb)

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bleed Resistor

Bleed resistor calibration to establish pressure-to-flow table for system bleed path characterization

Ranked causes

  1. 1. Flow exceeds 50 l/min before 105 cmH2O pressure reached (unknown)
  2. 2. Leaks around test plugs (unknown)
  3. 3. Drive gas regulator not adjusted or unstable (occasional)
  4. 4. Inspiratory Valve Calibration not complete (occasional)
  5. 5. Inspiratory Flow Valve not closing completely (leaky) (occasional)
  6. 6. Inspiratory Flow Valve not linear (occasional)
First check

Plug Drive Port (A) and Manifold Port (B) on interface valve

Test point
Vent Engine interface valve ports A and B
Expected
Ports sealed with plugs

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in Bag Mode patient safety

Full Test - Circuit Leak fails during system checkout. Bellows do not inflate properly or airway pressure cannot be maintained.

Ranked causes

  1. 1. Bag/Vent Switch lower seal leaking (common)
  2. 2. APL diaphragm leaking (occasional)
  3. 3. APL/BTV Manifold O-Ring seal compromised (occasional)
  4. 4. Bag port cover or APL valve defective (occasional)
  5. 5. Negative Pressure Relief Valve leaking (occasional)
First check

Perform Test 4 - Testing bag port, APL valve, Bag/Vent switch, and negative pressure relief

Test point
Bag port connection
Expected
No pressure loss; bellows inflate when pressurized

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in both Bag and Vent Mode patient safety

Full Test fails in both Bag and Vent Mode. Common areas: Flow Sensor Module, Circuit Module, absorbent canister, negative pressure relief, or optional EZchange/Condenser Module.

Ranked causes

  1. 1. Flow Sensor Module or bulkhead connector O-ring defective (common)
  2. 2. Circuit Module O-ring or seal leaking (common)
  3. 3. Absorbent canister leak or improper installation (occasional)
  4. 4. Negative Pressure Relief Valve defective (occasional)
  5. 5. EZchange/Condenser Module or condenser assembly leaking (occasional)
First check

Perform Test 9 - Testing flow sensor module, circuit module, and absorbent canister

Test point
Circuit interface ports
Expected
No pressure loss; bulb remains collapsed

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak without vaporizer caution

Gas escaping from low-pressure system without vaporizer or with vaporizer mounted (leaking port valve, mixer, relief valve, flush valve, system switch).

Ranked causes

  1. 1. Leaking port valve on vaporizer manifold (unknown)
  2. 2. Leak in mixer (unknown)
  3. 3. Leaking relief valve on vaporizer manifold (unknown)
  4. 4. Leaking flush valve (unknown)
  5. 5. Leaking system switch (unknown)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alternate O2 flowmeter not indicating correct flow patient safety

Flowmeter ball does not rise smoothly, does not maintain steady position, or does not indicate expected flow rate during Alternate O2 flow control operation.

Ranked causes

  1. 1. Flow control mechanism stiff or not rotating smoothly (occasional)
  2. 2. Flowmeter ball stuck or obstructed (occasional)
  3. 3. Alternate O2 flow not activating (occasional)
  4. 4. O2 supply pressure insufficient or not connected (occasional)
First check

O2 supply connected and pressure available

Test point
O2 cylinder valve or pipeline inlet
Expected
Cylinder valve fully open or pipeline connected

8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows collapsed patient safety

Unable to drive bellows, manifold pressure exceeds calculated limit

Ranked causes

  1. 1. Breathing circuit leak or hose occlusion (common)
  2. 2. Drive gas check valve failure (occasional)
  3. 3. cVIB cabling issue (occasional)
  4. 4. cVIB module failure (occasional)
First check

Inspect breathing circuit

Test point
Breathing circuit hoses and connections
Expected
No leaks, no obstructions

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows falls during vertical seal test patient safety

Bellows assembly fails to remain inverted when ports are sealed with test plugs, indicating loss of seal integrity or component malfunction

Ranked causes

  1. 1. Ports not tightly sealed (common)
  2. 2. Bellows incorrectly installed (occasional)
  3. 3. Internal bellows seal incorrectly installed (occasional)
  4. 4. Parts are damaged (occasional)
First check

Verify test plug seating in all ports

Test point
Bellows assembly ports
Expected
Test plugs fully seated and secure

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows falls past guide line during upright position test patient safety

Bellows assembly fails to remain extended to guide line when single port is sealed, indicating seal or pressure-relief valve malfunction

Ranked causes

  1. 1. Port not tightly sealed (common)
  2. 2. Bellows incorrectly installed (occasional)
  3. 3. Pressure-relief valve incorrectly installed (occasional)
  4. 4. Parts are damaged (occasional)
First check

Verify test plug seating in upright position port

Test point
Bellows assembly port
Expected
Test plug fully seated and secure

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows fill time abnormal

Bellows takes longer or shorter than expected to fill when O2 Flush button is depressed continuously

Ranked causes

  1. 1. Large leak in breathing system (common)
  2. 2. Flush regulator setting incorrect (common)
  3. 3. Flush regulator cross-connection (occasional)
  4. 4. SCGO/ACGO selector valve inlet cross-connection (occasional)
First check

Set up test apparatus

Test point
Bag/Vent switch and patient circuit
Expected
Bag/Vent switch in Vent position, patient circuit attached with patient port plugged, bellows completely collapsed

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system leak suspected patient safety

Airway pressure fails to rise or rises below expected threshold during breathing system leak test procedures.

Ranked causes

  1. 1. Leak in Bag/Vent Switch (common)
  2. 2. APL valve malfunction (common)
  3. 3. Bellows module leak or pop-off valve failure (common)
  4. 4. Flow sensor module or circuit module leak (common)
First check

Verify airway pressure during controlled test

Test point
Airway Pressure on Gas Delivery Schematic
Expected
≥ 30 cm H2O, ≥

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check flow sensor patient safety

Measured flow for 6 consecutive mechanical breaths does not meet criteria, or no/negative flow on inspiration/expiration

Ranked causes

  1. 1. Flow sensor connection issue (common)
  2. 2. Flow sensor calibration needed (common)
  3. 3. Inspiratory or expiratory check valve failure (occasional)
  4. 4. Flow sensor failure (occasional)
First check

Check flow sensor connection

Test point
Flow sensor connectors and cables
Expected
All connections secure

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive fan noise

Fan generates excessive or abnormal noise during operation

Ranked causes

  1. 1. Display case fan inlet filter clogged (common)
  2. 2. Obstructions within case fan (common)
  3. 3. Com Express fan noise (occasional)
  4. 4. Defective fan assembly (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas module related errors in logs patient safety

Error log entries indicate issues with the Airway Module gas sampling or sensor functionality

Ranked causes

  1. 1. Sample gas line blocked or residue in inlet filter (common)
  2. 2. Sample line not connected (occasional)
  3. 3. Sensor inoperative (occasional)
  4. 4. Module not compatible (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High intrinsic PEEP during operation patient safety

Elevated positive end-expiratory pressure measured continuously during ventilation in bag or vent modes

Ranked causes

  1. 1. APL (adjustable pressure limit) valve sticking or obstruction downstream (common)
  2. 2. Exhalation valve or pop-off valve sticking or misassembly (common)
  3. 3. Scavenging system problem or breathing circuit obstruction (occasional)
  4. 4. Flow control valve malfunction or improper regulator setting (occasional)
First check

Perform flow and pressure sensor calibration

Test point
Calibration software in service application
Expected
Calibration passes

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate volume ventilation measured patient safety

Set tidal volume does not match measured tidal volume; volume delivery discrepancy

Ranked causes

  1. 1. Flow sensor calibration out of range or sensor failure (common)
  2. 2. Inspiratory or expiratory check valve leak (occasional)
  3. 3. VIB board failure or pneumatic sense tubing leak (occasional)
  4. 4. Flow sensor failure or blockage (occasional)
First check

Perform flow and pressure calibration

Test point
PC Service Application calibration module
Expected
Calibration passes

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in Vent Mode patient safety

Full Test - Vent and Gas fails during system checkout. Bellows do not fill or exhaust gas cannot be vented properly.

Ranked causes

  1. 1. Bellows assembly compromised (common)
  2. 2. Bellows pop-off valve leaking (common)
  3. 3. Bag/Vent Switch upper seal, disk, or O-ring defective (occasional)
  4. 4. Exhalation valve or drive circuit malfunction (rare)
First check

Perform Test 6 - Testing bellows assembly and Bag/Vent switch

Test point
Bellows outlet port
Expected
Bellows fill and empty without pressure loss

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System will not boot from external USB Flash Drive during software installation

System fails to initiate boot process from USB Flash Drive when attempting software installation

Ranked causes

  1. 1. USB Flash Drive is incorrect part number or improperly inserted (common)
  2. 2. Primary USB port malfunction (occasional)
  3. 3. Defective USB Flash Drive (occasional)
  4. 4. Carrier Board or Com Express Board Assembly failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen fails to work

Touchscreen does not respond to touch input or is non-functional

Ranked causes

  1. 1. Screen lock is enabled (common)
  2. 2. Internal cables or connections disconnected or loose (occasional)
  3. 3. Touch Controller Board failure (rare)
  4. 4. Touch Bezel Assembly failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible beep not heard during system startup

Expected beep(s) from annunciators (speakers) during power-on sequence are missing or not audible

Ranked causes

  1. 1. Power Controller Board annunciator failure (unknown)
  2. 2. Anesthesia Control Board annunciator failure (unknown)
  3. 3. Display Unit annunciator failure (unknown)
First check

Verify first audible beep heard immediately after System switch turned to On

Test point
System switch
Expected
Audible beep from Power Controller Board

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not charging

Battery status LEDs (CHG, FLOAT, OVER) not illuminating or batteries remain uncharged

Ranked causes

  1. 1. 28 VDC not reaching battery charging circuit (common)
  2. 2. Battery charging fuse FH2 or FH3 open (occasional)
  3. 3. U-Frame Power Supply not delivering 28 VDC (occasional)
First check

Measure 28 VDC at J6 input

Test point
J6 connector
Expected
28 VDC
Tools
DMM

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage outside acceptable range caution

Battery test shows voltage below minimum threshold or battery not holding charge

Ranked causes

  1. 1. Battery 1 (right) voltage below 12.10 Vdc (occasional)
  2. 2. Battery 2 (left) voltage below 12.10 Vdc (occasional)
  3. 3. Battery current abnormal or excessive discharge (rare)
First check

Execute Battery test

Test point
AC mains power
Expected
AC mains shut off, wait 2 minutes for results

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows does not move patient safety

Ventilator bellows assembly does not rise and fall during ventilation attempt

Ranked causes

  1. 1. Drive gas regulator output pressure low or absent (common)
  2. 2. Gas inlet valve or solenoid failure (occasional)
  3. 3. Hospital pressure supply failure or inlet filter blockage (occasional)
First check

Check drive gas regulator output pressure

Test point
Regulator output
Expected
25 psi

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows leak caution

Gas escaping from bellows assembly due to pop-off valve diaphragm, mounting rim, or bellows damage.

Ranked causes

  1. 1. Pop-off valve diaphragm not sealing properly (unknown)
  2. 2. Bellows mounting rim loose (unknown)
  3. 3. Bellows improperly mounted or has hole or tear (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Primary regulators high voltage

Verify output pressure of O2, Air, or N2O primary regulators under low and high flow conditions; regulators must be tested after setup and adjusted if out of specification.

Ranked causes

  1. 1. Primary regulator output pressure out of specification at low flow (common)
  2. 2. Cylinder supply pressure decays too quickly or does not stabilize (occasional)
  3. 3. Primary regulator output pressure drops excessively at high flow (occasional)
First check

Set system to Service mode and access Gas Delivery Schematic

Test point
Service Application (Section 12.3.2)
Expected
System in Service mode; schematic accessible

17 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

cVIB pneumatic sense tubing leak caution

Flow measurements jump or fail to stabilize near zero during pneumatic sense tubing leak test; sensor readings indicate flow when circuit should be sealed

Ranked causes

  1. 1. Leak in flow sensor pneumatic circuit (common)
  2. 2. Defective flow sensor (occasional)
  3. 3. Leak in sense line tubing (occasional)
First check

Set fresh gas flow and vent configuration

Test point
Gas Delivery Schematic, PC Service Application; APL valve; Bag-to-Vent switch
Expected
Fresh gas flow 2 l/min; APL valve 0 cm H2O; Bag-to-Vent switch in Bag/APL position; short tube connected between inspiratory and expiratory flow sensors

6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display arm positioning requires adjustment

Display arm movement, side-to-side drift, or tilt position needs calibration to maintain proper positioning and stability

Ranked causes

  1. 1. Arm attachment mounting nut loose (common)
  2. 2. Display mount attachment nut loose (common)
  3. 3. Display clamp assembly clamping screws loose (common)
First check

Position display parallel to arm and apply finger pressure along left side

Test point
Left side of display
Expected
Display does not move

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak with vaporizer caution

Gas escaping from low-pressure system with vaporizer mounted (vaporizer installation, o-ring, fill port, port valve, mixer, relief valve, flush valve, system switch, bellows, pop-off valve, bellows mounting, breathing system).

Ranked causes

  1. 1. Vaporizer not installed properly (unknown)
  2. 2. Missing or damaged o-ring on vaporizer manifold (unknown)
  3. 3. Loose fill port (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mixer ACB communication failure

Anesthesia control board lost communication with gas mixer module or mixer not responding to commands

Ranked causes

  1. 1. Loose or damaged communication cable between ACB and mixer (common)
  2. 2. Mixer module not powered or internal failure (occasional)
  3. 3. ACB board communication port malfunction (rare)
First check

Check mixer XMT/RCV LED activity

Test point
Mixer board transmission/reception LEDs
Expected
XMT or RCV LED indicates activity; not both absent

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ventilation delivery patient safety

Ventilator unable to deliver breath; bellows not moving or no fresh gas flow; respiratory gas delivery failure

Ranked causes

  1. 1. Drive gas regulator output pressure insufficient or zero (common)
  2. 2. No fresh gas flow or blender malfunction (common)
  3. 3. Bellows drive system failure or exhalation valve stuck (occasional)
First check

Perform flow/pressure sensor calibration

Test point
PC Service Application sensor calibration
Expected
Calibration passes

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply voltage out of range high voltage

Diagnostic menu shows voltage readings outside normal ranges on power boards

Ranked causes

  1. 1. Power Controller Board supply fault (common)
  2. 2. Anesthesia Control Board power regulation failure (occasional)
  3. 3. Display Unit power distribution failure (occasional)
First check

Verify Power Controller Board 12Vdc supply

Test point
Power Diagnostics menu - Power Board section
Expected
11.70 to 12.30 Vdc, +0.30 / -0.30

24 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotary encoder fails to work

Rotary encoder does not respond to rotation or does not function

Ranked causes

  1. 1. Rotary encoder cable disconnected or improperly seated (common)
  2. 2. Rotary encoder failure (occasional)
  3. 3. Carrier Board or Com Express Board Assembly failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System application screen does not appear

Display does not show GE or Datex Ohmeda logo or system interface after power-on

Ranked causes

  1. 1. +12V LED not illuminated or power supply failure (common)
  2. 2. LCD touch bezel assembly failure (occasional)
  3. 3. Compact flash card or memory module failure (occasional)
First check

Check green power indicator LED on dashboard

Test point
Dashboard green LED
Expected
Illuminated

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Checkout Stage 2 Step 8: Battery and Electrical status patient safety

Check AC/Mains power connection and battery charge status during system checkout

Ranked causes

  1. 1. AC/Mains disconnected (unknown)
  2. 2. Battery failure (unknown)
  3. 3. Battery still charging (unknown)
First check

Verify AC/Mains power connection

Test point
AC/Mains power cord
Expected
Connected and active

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to begin mechanical ventilation patient safety

ABS not fully engaged or no O2 supply or defective Bag/Vent switch prevents mechanical ventilation.

Ranked causes

  1. 1. ABS not fully engaged (unknown)
  2. 2. No O2 supply (unknown)
  3. 3. Defective Bag/Vent switch (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alternate O2 flowmeter not indicating correctly patient safety

Float does not rise smoothly, does not maintain steady position, or indicates outside expected range during functional or accuracy testing

Ranked causes

  1. 1. Flowmeter mechanical obstruction or float sticking (common)
  2. 2. Flowmeter accuracy drift or calibration error (occasional)
First check

Functional flowmeter response at minimum setting

Test point
Alt O2 flowmeter display
Expected
0.5 to 0.7 l/min with flow control fully clockwise

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

App Not Loaded message is displayed

Display shows App Not Loaded message

Ranked causes

  1. 1. Software not loaded (common)
  2. 2. Defective CF card (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary O2 flowmeter accuracy out of specification

Flow test device reading does not fall within acceptable limits for set flowmeter values

Ranked causes

  1. 1. Flowmeter calibration drift (common)
  2. 2. Test device reading error or incorrect setup (occasional)
First check

Measure flow at 1 L/min setting

Test point
Flow test device outlet
Expected
0.5 to 1.5 L/min, +0.5 -0.5 L/min
Tools
Flow test device (0–15 L/min, ±2% accuracy)

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Balance gas selector valve leaks

Balance Gas Valve Leak test fails indicating selector valve or proportional valve not sealing

Ranked causes

  1. 1. Balance gas selector valve seal compromised (occasional)
  2. 2. Balance gas proportional valve seal degraded (occasional)
First check

Balance Gas Valve Leak test

Test point
Balance gas selector valve outlet
Expected
Pass (valve holds pressure)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery status abnormal caution

Battery voltage out of range, charge state or discharge time unusual

Ranked causes

  1. 1. Battery voltage low or high (common)
  2. 2. Battery discharge capacity reduced (occasional)
First check

Check Battery 1 voltage reading

Test point
Power Diagnostics menu - Power Board section
Expected
10.00 to 16.50 Vdc (black display), <10.00 or >16.50 (red display), +3.50 / -3.00

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows does not fill within time specification caution

Flush Flow test shows bellows fill time outside 1.8 to 2.3 second range

Ranked causes

  1. 1. Gas supply pressure insufficient (occasional)
  2. 2. Flush valve obstruction or malfunction (occasional)
First check

Perform Flush Flow test and time bellows fill

Test point
Bellows fill time
Expected
The bellows fills in 1.8 to 2.3 seconds, 1.8 to 2.3 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Board temperature excessive caution

Power Controller Board temperature exceeds safe operating limit

Ranked causes

  1. 1. Inadequate cooling or fan failure (common)
  2. 2. Fan 1 or Fan 2 voltage out of specification (occasional)
First check

Monitor board temperature

Test point
Power Diagnostics menu - Power Board section
Expected
≤75°C (black); >75°C (red fail), max 75

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system leak caution

Gas escaping from breathing system due to absorber canister or canister o-ring issues.

Ranked causes

  1. 1. Absorber canister open or missing (unknown)
  2. 2. Damaged or missing canister o-ring (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system leak detected patient safety

System or circuit leak alarm triggered; bellows falls or pressure not maintained in breathing circuit

Ranked causes

  1. 1. Loose or disconnected breathing circuit connection (common)
  2. 2. Leak in bellows base or exhalation valve seal (occasional)
First check

Perform pneumatic sense tubing leak test on VIB

Test point
VIB pneumatic sense tubing
Expected
No leaks; both channels pass

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Manifold P Span

Manifold pressure transducer calibration to set 75 cmH2O reference point using drive gas pressure through inspiratory flow valve adjustment

Ranked causes

  1. 1. ADC value outside acceptable range (15740-20250 counts) (unknown)
  2. 2. Pressure transducer out of range limits (unknown)
First check

Remove ABS breathing system, exhalation valve, and vent engine cover

Test point
Vent Engine interface valve
Expected
System disassembled and accessible

7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Paw Span

Patient airway pressure (Paw) transducer calibration to set 75 cmH2O reference point using inspiratory flow valve adjustment with full circuit assembly

Ranked causes

  1. 1. ADC value outside acceptable range (21000-27000 counts) (unknown)
  2. 2. Pressure transducer outside range limits (unknown)
First check

Leave Drive port (A) plugged

Test point
Vent Engine interface valve Drive port (A)
Expected
Port remains sealed

12 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPU LED not illuminated

CPU LED on Power Controller board is off or not flashing, indicates no power or missing Boot Code

Ranked causes

  1. 1. No 28 VDC power from U-Frame Power Supply or Battery (common)
  2. 2. No Boot Code in Power Controller board (rare)
First check

Measure 28 VDC input voltage

Test point
J6 connector
Expected
28 VDC
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cylinder high pressure leak detected caution

Pressure decay in cylinder exceeds acceptable limit during one-minute hold test

Ranked causes

  1. 1. Cylinder valve seal leaking (occasional)
  2. 2. Connection fitting leak (occasional)
First check

Test cylinder #1 for high pressure leaks

Test point
Cylinder #1 pressure gauge
Expected
Pressure decay after one minute is less than 690 kPa (100 psig)

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cylinder pressure gauge does not indicate patient safety

Pressure gauge fails to show cylinder pressure when cylinder valve is opened

Ranked causes

  1. 1. Cylinder valve closed or not fully opened (common)
  2. 2. Cylinder empty or low pressure (occasional)
First check

Open each cylinder valve one at a time

Test point
Gas Supply window pressure gauge
Expected
Corresponding gauge indicates cylinder pressure

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cylinder pressure not displayed

Gas Supply window shows no pressure reading when cylinder valve opened

Ranked causes

  1. 1. Cylinder valve not fully open (common)
  2. 2. Cylinder empty or pressure insufficient (occasional)
First check

Cylinder valve open

Test point
Gas Supply window
Expected
Corresponding display indicates appropriate cylinder pressure

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cylinder pressure not indicated on gauge patient safety

Cylinder pressure gauge does not show expected cylinder pressure when valve is opened

Ranked causes

  1. 1. Cylinder valve not opened or connection incomplete (common)
  2. 2. Cylinder empty or gauge malfunction (occasional)
First check

Open each cylinder valve one at a time

Test point
Cylinder outlet
Expected
Corresponding pressure gauge indicates cylinder pressure

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display controller board ACB communication failure

Display unit unable to communicate with anesthesia control board; serial link failure

Ranked causes

  1. 1. ACB TXD/RXD LED not indicating activity or communication cable damage (occasional)
  2. 2. Display memory module or carrier board failure (occasional)
First check

Observe ACB TXD/RXD LED activity

Test point
ACB board transmission/reception LEDs
Expected
Both TXD and RXD LEDs showing activity or one active at a time

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display unit requires service caution

Fan filter clogged, cable cover loose, or internal components need replacement

Ranked causes

  1. 1. Fan filter fouled (common)
  2. 2. Cable cover mounting loose (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive gas lost patient safety

Ventilator has no drive gas, O2 or Air supply low

Ranked causes

  1. 1. O2 supply disconnected or low pressure (common)
  2. 2. Air supply disconnected or low pressure (common)
First check

Verify drive gas supply connection and pressure

Test point
O2 or Air supply inlet
Expected
Supply connected, pressure in range

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fans not running in Standby mode caution

Fans do not turn on when system enters Standby position with AC Mains connected

Ranked causes

  1. 1. Power Controller board not receiving 28 VDC from U-Frame Power Supply (common)
  2. 2. Fan connector disconnected or failed fan motor (occasional)
First check

Measure 28 VDC at J6 input

Test point
J6 connector on Power Controller board
Expected
28 VDC from U-Frame Power Supply
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

FiO2 High alarm does not occur patient safety

FiO2 High alarm fails to trigger when oxygen concentration rises above alarm threshold

Ranked causes

  1. 1. FiO2 sensor malfunction or disconnection (unknown)
  2. 2. Alarm limit not set to approved clinical values (unknown)
First check

Trigger FiO2 High alarm condition

Test point
Ventilator oxygen delivery system
Expected
FiO2 High alarm occurs

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • FiO2 Low alarm does not occur
  • Free breathing valve flapper damaged or contaminated
  • Full Test - Circuit Leak check fails
  • Full Test - Circuit O2 Cell reading unstable or incorrect
  • Full Test - Vent and Gas check fails
  • Gas supply system pressure reading incorrect
  • High current to mixer 10VA
  • High current to pressure transducers
  • Machine Check - Circuit: Manual circuit leak test
  • Machine Check - Circuit: O2 availability and flow test
  • Mains indicator does not illuminate with power cord connected
  • Mains power indicator does not illuminate
  • Mixer command lost during transmission
  • Mixer gas channel flow failure
  • MVexp Low alarm does not occur
  • O2 supply pressure low alarm not triggered
  • Over Pressure Circuit test failure
  • PEEP High alarm does not occur
  • Peripheral circuit supply failure
  • Pipeline supply not indicating facility pressure
  • Power failure alarm does not cancel when power cord is plugged back in
  • Power failure alarm does not display on battery
  • Power failure alarm does not display when power cord is unplugged
  • Power failure alarm not displayed or power indicator not active
  • Ppeak High alarm does not occur
  • Ppeak Low alarm does not occur
  • RCV LED constantly on (no communication with Display Unit)
  • Scavenger receiver replacement or cleaning
  • Suction gauge does not respond to regulator adjustment
  • System Checkout Stage 1 Step 2: Low O2 supply pressure
  • System Checkout Stage 2 Step 2: Ventilator circuit leak detected
  • System Checkout Stage 2 Step 5: O2 Flow mixer test
  • System Checkout Stage 2 Step 6: AIR Flow mixer test
  • System Checkout Stage 2 Step 7: N2O Flow mixer test
  • Tec 7 vaporizer filler cap leaking or damaged
  • Zero pressure transducers
  • +12.0V H AMPS MGAS power issue
  • +12.0V high current draw on mixer power
  • +12.0V high current on Alt O2 system
  • +12.0V high current on gas selection valves
  • +12.5V supply to ACB failure
  • AA, CO2 monitoring not connected
  • ABS inspection and maintenance
  • AC power inlet filter contamination
  • AC power inlet filter dirty
  • ACB clock speed error
  • ACB CPU test failure
  • ACB EEPROM failure
  • ACB flash memory failure
  • ACB hardware watchdog reset
  • ACB microprocessor error
  • ACB RAM error
  • ACB redundant memory failure
  • ACB software error
  • ACB software watchdog failure
  • ACB unexpected reset
  • ACGO mode selected - volume and apnea monitoring off
  • ACGO port flapper valve damaged or worn
  • ACGO port flapper valve obstruction or debris
  • ADC reference voltage out of range
  • ADC timeout failure
  • AGSS system not in working order
  • Air pressure low during 21% O2
  • Air supply pressure low
  • Airway overpressure signal failure
  • Airway sensor calibration error
  • Alarm tests
  • Alt O2 mode activated without operator action
  • Alt O2 selection valve failure
  • Alternate O2 button pressed
  • Alternate O2 flowmeter accuracy out of range
  • Alternate O2 flowmeter ball does not maintain steady position
  • Alternate O2 flowmeter ball does not rise smoothly
  • Alternate O2 flowmeter calibration and functional verification
  • Alternate O2 flowmeter does not respond at minimum setting
  • Alternate O2 flowmeter does not rise smoothly
  • Alternate O2 flowmeter float does not respond to control rotation
  • Alternate O2 flowmeter float does not rise
  • Alternate O2 flowmeter float does not stop at bottom
  • Alternate O2 flowmeter minimum flow outside spec
  • Alternate O2 flowmeter not responding to control rotation
  • Alternate O2 initial flow reading out of range
  • Alternate O2 outlet blockage not detected
  • Alternate O2 switch failure
  • Alternate O2 switch non-responsive
  • Anesthetic agent delivery during software download
  • Annunciator beep missing after power-up delay
  • Annunciator beep missing at System On
  • Annunciators not functioning at startup
  • Audible annunciator beep absent or incorrect after power-up
  • Auxiliary O2 flowmeter accuracy out of range
  • Auxiliary O2 flowmeter ball does not maintain steady position
  • Auxiliary O2 flowmeter ball does not maintain steady position at set flow
  • Auxiliary O2 flowmeter ball does not move at minimum setting
  • Auxiliary O2 flowmeter ball does not rest at bottom when flow off
  • Auxiliary O2 flowmeter ball does not rise smoothly
  • Auxiliary O2 flowmeter ball does not rise smoothly on opening
  • Auxiliary O2 flowmeter does not shut off
  • Auxiliary O2 flowmeter float does not maintain steady position
  • Auxiliary O2 flowmeter float does not move
  • Auxiliary O2 flowmeter float does not respond after shutdown
  • Auxiliary O2 flowmeter float does not respond to control
  • Auxiliary O2 flowmeter float does not return to bottom when fully closed
  • Auxiliary O2 flowmeter float stuck or does not rise
  • Auxiliary O2 flowmeter flow does not increase smoothly
  • Auxiliary O2 flowmeter not responding to control
  • Auxiliary O2 flowmeter reading outside tolerance
  • Auxiliary O2 outlet leak detected
  • Auxiliary O2 outlet occlusion does not stop float
  • Backup mode activated - apnea detected
  • Balance channel flow control failure
  • Balance channel proportional valve leak
  • Balance gas flow does not meet specification at 10 LPM
  • Balance gas flow does not meet specification at 3 LPM
  • Balance proportional valve failure
  • Balanced gas leak test skipped
  • Battery charge failure
  • Battery empty
  • Battery low
  • Battery missing
  • Battery power unavailable during mains failure
  • Battery reversed connections
  • Battery voltage failure
  • Battery voltage low
  • Battery voltage very low
  • Bellows assembly does not hold position
  • Bellows assembly does not hold pressure
  • Bellows assembly fails inspection
  • Bellows falls more than 100 ml/min after initial drop
  • Bellows fill rate test
  • Bellows fills slowly when O2 Flush button pressed
  • Bellows fills slowly with O2 Flush button
  • Bellows leaks excessively
  • Bellows module and bag vent function test
  • Breathing system not latched
  • Bulb inflation time excessive
  • Cabinet, drawer, or frame parts loose or damaged
  • Cabinet, frame, or drawer parts loose or damaged
  • Calibration data failure in EEPROM
  • Calibration error during airway module calibration
  • Calibration: Airway module
  • Calibration: Airway Pressure (Paw) Span
  • Calibration: breathing circuit pressure test
  • Calibration: Integrated Suction Regulator
  • Calibration: Manifold Pressure Span
  • Calibration: Mixer P Zero
  • Calibration: O2 Flush Regulator
  • Calibration: Positive pressure relief valve functional check
  • Calibration: Scavenger Positive Relief Valve
  • Calibration: User calibrations and system diagnostics
  • Calibration: Zero Gas Transducers
  • Carrier board and Com Express assembly replacement
  • Carrier board overheat
  • Case fan speed failure
  • Caster malfunction or misalignment
  • Caster operation or brake malfunction
  • Casters not rolling or locking properly
  • Change drive gas connection
  • Check AC Power Inlet filter
  • Check casters for proper operation and mounting
  • Check external electrical cabling
  • Check external tubing, rubber goods, hoses, and pipeline hoses
  • Check labels for legibility
  • Check mechanical parts condition and tightness
  • Circuit Leak test fails
  • Circuit O2 Cell 100% test fails
  • Circuit O2 Cell 21% test fails
  • Circuit O2 Cell test reading instability
  • Communication error ventilator to ACB
  • Compatibility failure no version info
  • Compatibility failure software error
  • Compatibility incomplete no versions from Vent SIB
  • Compatibility incomplete no versions received from Vent SIB
  • Configuration: Drive Gas and Altitude verification
  • Consolidated Ventilator Interface board replacement
  • Cooling fan current low failure
  • CPU fan speed failure
  • CPU overheat
  • Cylinder pressure drops excessively
  • Cylinder pressure test
  • Cylinder supply high pressure leak
  • DC commanded AC to failure
  • DC-DC converter failure with battery in use
  • DCB procedure communication failure
  • Display fan filter clogged
  • Display fan filter contamination
  • Display Power fuse FH4 open
  • Display Unit annunciator beeps missing during power-up
  • Display Unit annunciator beeps missing during startup
  • Display unit disassembly required
  • Drive gas configuration mismatch
  • Drive gas regulator pressure out of specification
  • Electrical safety testing
  • Electronic Gas Mixer assembly requires replacement
  • Expiratory check valve leaking
  • External electrical cabling damaged or disconnected
  • External electrical cabling damaged or improperly connected
  • External tubing or hoses deteriorated or loose
  • External tubing, hoses, or pipeline deteriorated
  • Fan filter clogged or missing
  • Flow and Pressure test fails
  • Flow and pressure test failure
  • Flow sensor readings exceed ±10% of reference device
  • Flow valve current feedback failure
  • Flows do not drop to zero when O2 supply fails
  • Full system test required
  • Gas flows do not drop to zero when O2 supply fails
  • Gas inlet valve feedback closed state
  • Gas Inlet Valve Feedback does not read Closed
  • Gas Inlet Valve Feedback test fails
  • Gas supply not visible on display
  • Ground resistance exceeds specification
  • Ground resistance out of specification
  • Ground resistance out of tolerance
  • Hard keys not responding
  • High airway overpressure signal detected
  • High current to alternate O2 valve
  • High current to consolidated ventilator interface board
  • High current to gas select valves
  • High current to MGAS 10VA
  • High current to vent and outlet valves
  • High pressure cylinder leak test
  • Incompatible software detected during download
  • Inspiration flow valve 10VA overcurrent
  • Inspiratory check valve leaking
  • Inspiratory port occlusion
  • Integrated suction regulator calibration
  • Integrated suction regulator does not reduce to minimum vacuum
  • Integrated suction regulator gauge does not rise smoothly
  • Integrated Suction Regulator gauge does not rise with adjustment
  • Integrated suction regulator gauge exceeds 200 mmHg during test
  • Integrated suction regulator gauge out of calibration
  • Integrated Suction Regulator gauge rises when inlet occluded at minimum
  • Integrated suction regulator pressure exceeds 200 mmHg at minimum setting
  • Integrated Suction Regulator vacuum exceeds minimum setting
  • Intermittent breathing system leak
  • Invalid air pipeline measurement
  • Keypad assembly replacement
  • Labels illegible or missing
  • Labels missing or illegible
  • Labels not legible
  • LCD display not functioning
  • LCD/Touch Bezel assembly replacement
  • Leakage current exceeds limit
  • Leakage current exceeds specification
  • LEDs not illuminating
  • Loss of gas delivery user settings
  • Loss of ventilation parameter settings
  • Low drive gas pressure alarm
  • Low O2 supply alarm not triggered on disconnect
  • Low O2 supply alarm occurs when pipelines disconnected
  • Low P Leak check fails (ACGO)
  • Low P Leak check fails (SCGO)
  • Low voltage to alternate O2 valve
  • Low voltage to cVIB
  • Low voltage to vent and outlet valves
  • Low-pressure leak check (ACGO)
  • Low-pressure leak check (SCGO)
  • Machine Check - Circuit: Airway pressure too high
  • Machine Check - Circuit: Alt O2 Flow control test
  • Machine Check - Circuit: Bag/Vent and ACGO switch verification
  • Main power failure - system on battery
  • Mains indicator does not illuminate
  • Mechanical parts loose or damaged
  • Memory failure - EEPROM
  • Minimum alternate O2 flow setting incorrect
  • Missing second audible beep during power-up
  • Missing third or fourth audible beep during power-up
  • Mixed fresh gas selection valve overcurrent
  • Mixer balance gas check valve leak rate exceeds specification
  • Mixer balance gas selection failure
  • Mixer channel zero calibration failure
  • Mixer channel zero calibration failure due to leak
  • Mixer CPU test failure
  • Mixer CRC EEPROM failure
  • Mixer CRC flash failure
  • Mixer flow attainment failure balance gas channel
  • Mixer flow attainment failure channel 1
  • Mixer flow check failure
  • Mixer flow inaccurate on air
  • Mixer flow inaccurate on N2O
  • Mixer flow inaccurate on O2
  • Mixer flow sensor failure balance gas channel
  • Mixer flow sensor failure channel 1
  • Mixer flow verification failure balance gas channel
  • Mixer flow verification out of limits
  • Mixer gas channel flow control failure
  • Mixer gas channel proportional valve failure
  • Mixer gas channel selection valve failure
  • Mixer gas channel temperature limit exceeded
  • Mixer gas channel temperature sensor malfunction
  • Mixer outlet check valve leaking
  • Mixer outlet check valve leaks
  • Mixer outlet check valve leaks excessively
  • Mixer outlet flow inaccurate
  • Mixer POST failure during power-up
  • Mixer pressure sensor P1/P3 failure
  • Mixer pressure sensor P2 failure
  • Mixer RAM failure
  • Mixer software watchdog failure
  • Mixer temperature sensor failure balance gas channel
  • Mixer temperature sensor failure channel 1
  • Mixer voltage supply failure
  • MOPV pressure relief valve malfunction
  • MOPV pressure relief valve not functioning
  • MOPV pressure relief valve test fails
  • No AA data
  • No audible beep after system switch on
  • No CO2 data or waveform
  • No flow waveform displayed
  • No fresh gas flow - aux outlet circuit feedback fault
  • No measured Vte
  • No O2 data
  • No PEEP pressure waveform displayed
  • Non-Circle SCGO selected
  • O2 alarm verification with Airway Module
  • O2 bypass valve leaks
  • O2 flow does not meet specification at 10 LPM
  • O2 flow does not meet specification at 3 LPM
  • O2 Flush button friction or wear
  • O2 Flush Regulator pressure out of specification
  • O2 proportional valve leaks
  • O2 selector valve leaks
  • O2 sensor not connected
  • O2 supply alarm does not occur
  • Over-pressure circuit does not function
  • Overdue 12-hour power cycle self test
  • Oxygen cell calibration failure
  • Oxygen leak test skipped
  • Pan enclosure fan requires replacement
  • PEEP high blockage alarm does not occur
  • PEEP/PCV not available
  • Pipeline and cylinder pressure test
  • Pipeline pressure gauge not indicating correctly
  • Pipeline pressure not displayed
  • Power Controller communication failure
  • Power Controller communication failure with internal failure
  • Power Controller self-test failure
  • Power failure alarm does not activate
  • Power failure alarm does not display
  • Power failure alarm response
  • Power supply temperature high
  • Ppeak low leak alarm does not occur
  • Pressure limit circuit test fails
  • Pressure relief valve test
  • Pressure transducer out of specification
  • Reverse expiratory flow detected
  • Scavenger positive relief functional check
  • Scavenger positive relief valve not functional
  • Second annunciator beep missing during startup
  • Severe sustained airway pressure
  • Speaker not producing sound
  • Standby patient detection failure
  • Suction control valve not delivering adequate flow
  • Suction flow rate below minimum
  • Suction gauge does not bleed to zero on OFF
  • Suction gauge needle moves with knob in OFF position
  • Suction gauge needle not at zero when idle
  • Suction gauge needle not at zero when off
  • Suction regulation unstable
  • Suction regulator does not respond to adjustment
  • Suction regulator pressure exceeds limit during operation
  • Suction regulator pressure not holding setpoint
  • Suction regulator vacuum accuracy out of range
  • Suction regulator vacuum gauge not zeroing
  • Suction system has vacuum leak with regulator off
  • Suction system vacuum does not return to zero after shutdown
  • Suction vacuum gauge reading inaccurate
  • Sustained high airway pressure
  • System Checkout Stage 1 Step 1: Bag/Vent Switch incorrect
  • System Checkout Stage 1 Step 3: Ventilator has no drive gas
  • System Checkout Stage 2 Step 1: Cannot empty bellows
  • System Checkout Stage 2 Step 3: Mechanical Ventilation test
  • System Checkout Stage 2 Step 4: Circuit Compliance test
  • System date and time incorrect
  • System flushing with oxygen after leak test
  • System malfunction - ACB clock speed incorrect
  • System malfunction - ACB communication failure
  • System malfunction - ACB CPU test failure
  • System malfunction - ACB Display Controller communication failure
  • System malfunction - ACB flash memory CRC failure
  • System malfunction - ACB hardware watchdog failure
  • System malfunction - ACB microprocessor exception
  • System malfunction - ACB procedure communication failure
  • System malfunction - ACB RAM error
  • System malfunction - ACB redundant memory failure
  • System malfunction - ACB software error
  • System malfunction - ACB software watchdog failure
  • System malfunction - ACB unexpected reset
  • System power-on tests failed
  • Temperature measurement channel 1 high
  • Temperature measurement channel 2 high
  • Touch points are misaligned
  • Touch screen calibration misaligned
  • Touchscreen failure
  • Unit does not power on
  • Vaporizer alarms triggered during back pressure test
  • Vaporizer back pressure and interlock malfunction
  • Vaporizer system issues during concentration adjustment
  • Vent and Gas test fails
  • Ventilation -6V supply failure
  • Ventilation +12.5V supply failure
  • Ventilation +6V supply failure
  • Ventilation 1.22V supply failure
  • Ventilation ADC reference voltage failure
  • Ventilation check stage 1 failed
  • Ventilation check stage 2 failed
  • Ventilation flow valve DAC feedback failure
  • Ventilation valve 10VA power supply failure
  • Ventilator free breathing valve requires maintenance
  • Ventilator interface board 10VA overcurrent
  • Ventilator interface board communication failure
  • Verify and reset Altitude setting
  • Verify and reset Drive Gas setting
  • Verify audible beeps during power-up
  • Visual inspection of display fan filter
  • Voltage out of range to ACB
  • Zero calibration: Gas transducers
  • Zero calibration: Mixer Pressure

419 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
+12.0V H AMPS ALT O2High current to alternate O2 valve detectedAlternate O2 valve +10VA turned onwarning
+12.0V H AMPS ALT O2Alternate oxygen system +12V overcurrent; excessive current drawn+12V supply to Alt O2 circuit drops due to overcurrenterror
+12.0V H AMPS GAS SEL VALVESHigh current to gas select valves detectedFresh gas select valves +10VA is turned onwarning
+12.0V H AMPS GAS SEL VALVESGas selection valve circuits drawing excessive current; overcurrent condition+12V supply to gas selection solenoids drops due to overcurrenterror
+12.0V H AMPS MGASHigh current to MGAS 10VA detected; gas monitoring not availableMGAS 10VA turned on after 3 consecutive ACB MGAS error log messageswarning
+12.0V H AMPS MGASHigh current condition on +12V Airway Module power railPower supply detects excessive current draw from Airway Modulecritical
+12.0V H AMPS MIXERMixer +12V power supply overcurrent condition; excessive current detected+12V supply voltage drops or overcurrent protection activatescritical
+12.0V H AMPS MIXERHigh current to mixer 10VA supply detectedMixer +10VA turned onwarning
+12.0V H AMPS VENT&OUTLET VALVESHigh current to vent and outlet valves 10VA detectedVentilator valves +10VA is onwarning
+12.0V H AMPS VENTSIBHigh current to consolidated ventilator interface board 10VA detectedcVIB 10VA is turned onerror
+12.0V L VOLT ALT O2Low voltage to mixer alternate O2 valve detectedACB detects low voltage conditionwarning
+12.0V L VOLT GAS SEL VALVESGas selection valve circuits receiving low +12V supply; voltage below acceptable rangeGas selection solenoid supply voltage below minimum thresholderror
+12.0V L VOLT VENT&OUTLET VALVESLow voltage to vent and outlet valves detectedGIV or inspiratory flow valve drawing excess powerwarning
+12.0V L VOLT VENTSIBLow voltage to cVIB detectedACB detects low voltage condition to cVIBwarning
+12.5V TO ACBPower controller board unable to supply +12.5V to anesthesia control boardACB power supply voltage below specification; power delivery failureerror
+12.5V TO ACBVoltage to ACB out of specification rangePower supply voltage less than 11.9 or greater than 12.9 Vdcerror
12 HR TESTSystem power-up self test overdueSystem has been operating longer than 12 hours without power cycleinfo
+5V H AMP GAS SUPPLY XDUCERSHigh current to pressure transducers detected; cannot read gas supply pressuresPressure transducer +10VA turned onwarning
AA, CO2 MONITORING NOT CONNECTEDAnesthetic agent or CO2 monitoring module not detectedSystem unable to communicate with monitoring module at startup or during operationerror
ACB 4.096V ADC REFADC reference voltage out of specification; cannot monitor gas suppliesReference voltage less than 4.018 or greater than 4.176 Vdcinfo
ACB ADC FAILADC timeout detected; cannot monitor gas suppliesADC timeout on any MUX channelinfo
ACB CLOCK SPEEDAC clock frequency incorrect; system malfunctionClock frequency greater than 1.1× expected or less than 0.9× expectederror
ACB CLOCK SPEEDAnesthesia control board clock speed out of specification; timing error detectedCPU clock frequency measurement outside normal range during self-testerror
ACB COM FAILSystem malfunction; display controller not receiving messages from anesthesia computerDC does not receive messages from AC for 10 seconds after initial communication establishederror
ACB COM FAILACB communication failureLoss of communication with ACBcritical
ACB CPU TEST FAILAnesthesia control board CPU self-test failed; processor malfunctionCPU built-in self-test (BIST) detects processor fault during startuperror
ACB DCB COM FAILSystem malfunction; anesthesia computer not receiving system state messages from display controllerAC receives no system state messages from DC for 10 seconds after initial communication establishederror
ACB DCB COM FAILACB-DCB communication failureLoss of communication between ACB and DCBcritical
ACB EEPROM FAILAnesthesia control board EEPROM memory failure; data storage errorEEPROM read/write test failed; memory chip unresponsiveerror
ACB EEPROM FAILMemory failure; EEPROM read/write or CRC failure on anesthesia control boardEEPROM read/write failure or CRC failureinfo
ACB FLASH FAILSystem malfunction; CRC failure in ACB code spaceCRC failure detected in flash memory code spaceerror
ACB FLASH FAILAnesthesia control board flash memory failure; program storage errorFlash memory read test failed; program code unreadableerror
ACB HW WATCHDOGACB hardware watchdog timeoutWatchdog timer expired, system hungcritical
ACB HW WATCHDOGACB hardware watchdog circuit triggered reset; system lockup or hang occurredACB processor did not reset watchdog timer before timeouterror
ACB MICROPROC ERRORSystem malfunction; unexpected microcontroller exceptionBus error, address error, or other microcontroller exceptionerror
ACB MICROPROC ERRORAnesthesia control board microprocessor error; CPU fault detectedProcessor exception or fault condition during operationerror
ACB PROCEDURE COM FAILSystem malfunction; anesthesia computer not receiving procedure messages from display controllerDC receives no procedure messages from AC for 15±1 seconds, or procedure message mismatch for 15±1 secondserror
ACB RAM ERRORSystem malfunction; memory test failure with multiple bit errorsMultiple bit errors detected during RAM testerror
ACB RAM ERRORACB RAM memory errorRAM test failure or memory parity errorcritical
ACB RAM ERRORAnesthesia control board RAM memory failure; read/write error detectedRAM memory self-test failed; memory chip unresponsiveerror
ACB REDUNDANT MEMORY FAILACB redundant memory verification failed; data mismatch detectedRedundant memory copies do not match; data integrity check failederror
ACB REDUNDANT MEMORY FAILSystem malfunction; redundantly stored parameter corrupted or not storedRedundant parameter storage or retrieval failureerror
ACB SW ERRORSystem malfunction; unexpected software errorSoftware error detected during executionerror
ACB SW ERRORAnesthesia control board software execution error; firmware faultSoftware exception or illegal instruction execution detectederror
ACB SW WATCHDOGACB software watchdog timer triggered; processor hang or deadlock detectedSoftware watchdog timer expired; ACB firmware did not reset watchdogerror
ACB UNEXPECTED RESETSystem malfunction; unexpected reset of anesthesia control boardUnexpected reset event detectederror
ACB UNEXPECTED RESETAnesthesia control board experienced unexpected reset; no watchdog triggerReset detected without hardware or software watchdog being triggeredwarning
ACGOVolume and apnea monitoring off; non-circle mode selectedNon-circle (ACGO) mode selected in system configurationinfo
ACMains POWER FAILPlug in power cable; system on batteryACMains_GOOD goes low and stays low for at least 300 msecwarning
AIR PIPE INVALIDCannot monitor air pipeline; invalid air pipeline pressure readingAir pipeline pressure measurement invalidwarning
AIR PRESS LOWAir supply pressure lowAir pipeline pressure less than 252 kPa and air cylinder pressure less than 2633 kPa for 1 secondwarning
AIR PRESS LOW DURING 21% O2Air pressure low; increase O2 percentageAir pipeline pressure less than 252 kPa and air cylinder pressure less than 2633 kPa during 21% O2 modeerror
AIRWAY SENSOR CAL ERRORCalibrate flow sensors; airway pressure sensor zero offset out of rangeAirway pressure sensor zero offset out of specified rangeinfo
ALT O2 BUTTON PRESSEDAlternate O2 button is selected; alternate O2 screen displayedAlternate O2 button selected by operatorinfo
ALT O2 SEL VALVE FAILUREAlternate oxygen selection valve malfunction; unable to select Alt O2 sourceAlt O2 selector solenoid unresponsive; valve does not switcherror
ALT O2 SWITCH FAILAlternate O2 switch status indicates fault conditionAlternate O2 switch fault condition persists for 1 secondwarning
Anesthesia Control BoardCommunications status of Anesthesia Control BoardSystem checks communication link during Tools — Communication Statusunknown
APNEANo breathing detected for specified time interval; patient not breathingRespiratory rate drops below configured apnea threshold or time exceeds limitcritical
APNEA > 120sApnea condition detected for longer than 120 seconds; prolonged no breathingNo breathing detected for more than 120 seconds during operationcritical
App Not LoadedApplication failed to load on displaySoftware not loaded or CF card defectwarning
AUX OUTLET FAILNo fresh gas flow; auxiliary outlet circuit feedback faultMeasured SCGO position does not match commanded positionerror
BACKUP MODE ENTEREDBackup mode active; apnea detectedNo spontaneous breaths for set apnea time and 30 seconds elapsed in PSVPro modeinfo
BAL CHANNEL PROP VALVE LEAK FAILBalance proportional valve leak detected; alternate O2 screen displayedMixer detects proportional valve leak on balance channelwarning
BAL FLOW CTRL FAILBalance channel flow control failureMixer status bit STS_FLOW_CTRL_CH2_FAIL indicates flow attainment failurewarning
BAL PROP VALVE FAILBalance channel proportional valve failure (overcurrent, etc.)Mixer status bit STS_CH2_PROPN VALVE FAIL indicates valve failurewarning
BATTERY CHARGE FAILNo battery backup, battery charge current excessiveSystem in standby with charge current >4.0 A or powered on with battery current >1.7 Awarning
BATTERY EMPTYBattery empty, system shutdown in less than 5 minutesAvailable battery power between 1 and 5 minuteserror
BATTERY FAILNo battery backup, battery voltage or charging anomaly detectedBattery voltage <10.5 V or voltage >16.5 V during bulk/over charging, or excessive bulk charging timewarning
BATTERY LOWBattery low, plug in power cableAvailable battery power between 20 and 30 minuteswarning
BATTERY MISSINGNo battery backup, battery not connectedBattery voltage reading between ±1.0 VDCwarning
BATTERY REVERSED CONNECTIONSNo battery backup, battery connected with reversed polarityBattery voltage reading less than –1.0 VDCwarning
Battery test FAILEDOne or both battery voltages are below minimum threshold or battery test did not completeBattery 1 (right) <12.10 Vdc OR Battery 2 (left) <12.10 Vdcerror
BATTERY V LOWBattery voltage low, plug in power cableAvailable battery power between 10 and 20 minuteswarning
BATT V VERY LOWBattery voltage very low, plug in power cableAvailable battery power decreases to between 10 and 5 minuteswarning
BELLOWS COLLAPSEDUnable to drive bellowsManifold pressure exceeds Paw + 10 + [0.25*(Insp valve flow)]info
Breathing system looseBreathing circuit connection loose or partially disconnected during operationLeak detected in breathing circuit connections; pressure drop in circuiterror
BREATHING SYSTEM NOT LATCHEDBreathing system loose or not properly latchedBreathing system detection switch indicates system not latchedinfo
CAL DATA FAILURE IN EEPROMService calibration advised due to corrupt calibration dataDefault calibration data in use due to corrupt data in calibration regioninfo
Calibrate flow sensorsFlow sensor calibration is due or required; sensor accuracy uncertainCalibration interval exceeded or sensor reading inconsistent during checkoutwarning
CARRIER BOARD OVERHEATCooling fans may overheat due to high Display Unit temperatureDisplay Unit thermistors exceed 60 °Cwarning
CASE FAN SPEED FAILCooling fans may overheat due to case fan speed failureCase fan speed less than 800 RPM (50% of normal speed)warning
CHECK FLOW SENSORFlow sensor reading out of range during mechanical ventilationMeasured flow for 6 consecutive breaths does not meet criteria, or no/negative flow on inspiration/expirationwarning
Check flow sensorsFlow sensor diagnostic test failed; sensor not responding within specificationFlow sensor malfunction detected during system checkout or operationwarning
Circuit check failedSystem checkout failure detectedMessage or failure displayed during system checkoutunknown
Circuit LeakAir leak detected in anesthesia breathing circuit; loss of pressurizationPressure or flow imbalance detected in inspiration vs. expirationerror
Circuit O2 check skippedO2 circuit check not performed during system checkoutMessage displayed during system checkoutunknown
COM ERROR VENT TO ACBSystem malfunction, communication lost between ventilator and ACBVentilator boundary object receives no messages from cVIB CPU for 35 millisecondserror
Compatibility failure: No version info in file for subsystemSubsystem did not report compatibility informationSubsystem version information missing from fileunknown
Compatibility failure: Software ErrorSubsystem did not report compatibility informationSoftware error preventing subsystem from reporting versionunknown
Compatibility incomplete: No versions from Vent SIBConsolidated Ventilator Interface Board compatibility information incompletecVIB compatibility info does not match or GIV test failedunknown
Compatibility incomplete: No versions received from Vent SIBConsolidated Ventilator Interface Board compatibility information not receivedcVIB compatibility info does not match or GIV test failedunknown
Controls (Front Panel) BoardCommunications status of Controls (Front Panel) BoardSystem checks communication link during Tools — Communication Statusunknown
COOLING FAN CURRENT LOW FAILUREMixer Pan Fan drawing too little currentMixer Pan Fan current below normal thresholdwarning
CPU FAN SPEED FAILCooling fans failed, CPU may overheatCPU fan speed less than 50% of nominal speedwarning
CPU OVERHEATCooling fans may overheat due to high CPU temperatureDisplay Unit thermistor temperature reading >70 °Cwarning
DCB PROCEDURE COM FAILSystem malfunction, procedure communication failure between computersNo procedure messages from Display Computer for >15 seconds or message mismatch for >15 secondserror
DC COMMANDED AC TO FAILUREDisplay Controller detected issues and commanded safe stateDisplay Controller initiated safe state transitionunknown
DRIVE GAS LOSTVentilator has no drive gasO2 or Air supply low or disconnected during mechanical ventilationerror
EC 06System MalfunctionSelf test failure or multi bit error detectedcritical
EC 06Error code 06Condition triggering EC 06error
ERROR!! INCOMPATIBLE SOFTWARE.Software version on USB Media is not compatible with installed subsystemSoftware Download detects incompatible subsystem during download processerror
***ERROR: No bootable device available.No valid bootable device found during startupSystem startup cannot locate bootable operating systemcritical
***ERROR: Program load failed - CRC.Software file checksum error detected during loadSystem startup detects CRC mismatch in program fileerror
***ERROR: RAM memory error.Random access memory hardware failure detectedSystem startup POST detects RAM failurecritical
***ERROR: System reset: ECxx xx xx . . .System reset due to software or boot failure with error codeSystem detects software fault or boot failure during startuperror
***ERROR: System reset: FFFF FF FFSystem reset due to critical battery or power failureSystem startup detects Carrier board battery contact loss or below voltagecritical
***ERROR: watchdog circuit failed.Watchdog timer circuit hardware failureSystem startup POST detects watchdog circuit malfunctioncritical
EXP FLOW SENSOR CAL ERRORCalibrate flow sensorsExp Flow Sensor zero offset out of rangewarning
EXP FLOW SENSOR EEPROM FAILUREReplace exp flow sensorEEPROM cal data read failurewarning
FAN FAILCooling fan needs service. System OKFan Power Status Bit is Low (FAN1_GOOD)warning
FiO2 HighOxygen fraction above acceptable levelOxygen delivery or measurement exceeds maximum thresholdwarning
FiO2 LowOxygen fraction below acceptable levelOxygen delivery or measurement falls below minimum thresholdwarning
Flow Attain Fail Bal Gas ChannelUnable to achieve commanded flow on balance gas channel; flow control failedBalance gas flow unable to reach setpoint despite proportional valve movementerror
Flow Attain Fail Bal Gas ChannelFlow attainment failure on balance gas channelUnable to attain set flow on balance gaserror
Flow Attainment Failure Ch1Flow attainment failure on channel 1Channel 1 unable to attain set flowerror
Flow Attainment Failure Ch1O2 flow mismatchCommanded flow through O2 Gas channel does not match measured flow via sensor and transducerserror
Flow Attainment Failure Ch1Unable to achieve commanded flow on channel 1; flow control failedChannel 1 flow unable to reach setpoint despite proportional valve adjustmenterror
FLOW SENSOR CAL ERRORCalibrate flow sensorsInsp or Exp flow sensor or airway or manifold pressure sensor zero offset out of rangewarning
Flow Sensor Fail Balance Gas ChFlow sensor for balance gas channel failed; sensor not respondingBalance gas flow sensor malfunction detected during system checkerror
Flow Sensor Failure Ch1Flow sensor for channel 1 (primary gas) failed; sensor not respondingChannel 1 flow sensor malfunction detected during system checkerror
Flow Sensor Failure Ch1Flow sensor failure on channel 1Channel 1 flow sensor malfunctionerror
FLOW VALVE CURRENT FAILUREInsp Flow Valve current feedback incorrectCurrent feedback from Insp Flow Valve incorrect for seven consecutive readingserror
FLOW VALVE DAC FAILUREInsp Flow Valve DAC failureCurrent feedback from Insp Flow Valve incorrect for seven consecutive readingserror
Flow Verif Fail (dP) Bal Gas ChBalance gas flow verification failed; differential pressure reading incorrectMixer unable to verify balance gas flow using differential pressure methoderror
Flow Verification Failure (dP) Ch1O2 pressure transducer mismatchCommanded O2 flow and Hot-wire anemometer agree but pressure transducers do not agreeerror
FRONT PANEL COM FAILDisplay panel controls failureKey pad controller fails to send 'life tick' for greater than 10 secondswarning
FRONT PANEL KEY STUCKStuck Keypad or encoder on DisplayStuck Keypad or encoder detectedwarning
GAS INLET VALVE BOOT UP TEST FAILVent Fail. Monitoring OnlyBoot-up test failedcritical
Gas ModuleCommunications status of Gas ModuleSystem checks communication link during Tools — Communication Statusunknown
INSP FLOW SENSOR CAL ERRORCalibrate flow sensorsInsp Flow Sensor zero offset out of rangewarning
INSP FLOW SENSOR EEPROM FAILUREReplace insp flow sensorEEPROM cal data read failurewarning
Inspiration StoppedInspiratory phase unexpectedly terminated; ventilation cycle incompleteFlow or pressure signal lost during inspiration; control system errorerror
Link StatusOverall communication link statusSystem checks communication during Tools — Communication Statusunknown
LOSS OF GAS DELIVERY USER SETTINGSLoss of gas delivery user settingsUser settings lost or corruptedwarning
LOSS OF GAS DELIVERY USER SETTINGSGas delivery user settings lost; saved parameters not recovered after resetSystem restart or error recovery; settings not found in ACB memorywarning
LOSS OF GAS DELIVERY USER SETTINGS / LOSS OF VENT PARAMETERVent Fail. Monitoring OnlyGas Delivery User Setting or Ventilator Parameter Settings from Display Computer arrived more than 10 seconds agocritical
LOSS OF VENT DELIVERY USER SETTINGSVentilation delivery user settings lost; saved parameters not recovered after resetSystem restart or error recovery; ventilation settings not found in ACB memorywarning
LOSS OF VENT PARAMETER SETTINGSLoss of ventilation parameter settingsVent parameters lost or corruptedwarning
Low O2 supply alarmOxygen supply pressure below operational thresholdPipeline supplies disconnected and all cylinder valves closedwarning
Low Pressure Leak check failedSystem Checkout FailedLow Pressure Leak section of System Checkout Failedwarning
Low Pressure Leak check failsLeak measured greater than 50 ml/minLeak measured is greater than 50 ml/min during Low Pressure Leak Check with SCGOwarning
MANIFOLD PAW SENSOR FAILVent Fail. Monitoring OnlyCalibration failure at bootupwarning
MANIFOLD PRESSURE SENSR FAILURECalibration failure at bootupManifold pressure sensor calibration failure at bootupwarning
MANIFOLD SENSOR CAL ERRORManifold Pressure zero failedManifold Pressure zero calibration failurewarning
MGAS CHECK SAMPLE GAS OUT >20 SECSample gas check line taking longer than 20 seconds to respondAirway Module unable to verify sample gas delivery within expected timewarning
MGAS CHECK SAMPLE GAS OUT >20 SECCheck sample gas outMGAS continuous occlusion bit set for 20 secondswarning
MGAS INLET FILTER RESIDUE >40 SECReplace D-FendMGAS communicates Replace Trap alarm bit for 40 secondswarning
MGAS INLET FILTER RESIDUE >40 SECInlet filter residue check taking longer than 40 secondsExcessive residue detected in Airway Module inlet filterwarning
MGAS LINE BLOCKED >20 SECGas sample line is blockedAirway Module detects blockage in gas sampling lineerror
MGas module removedAirway Module has been physically removed from systemSystem detects Airway Module is no longer presentwarning
MGAS SAMPLE FLOW DEVIATIONLow gas sample flowSample flow less than 80% of nominal for 20 consecutive secondswarning
MGAS SAMPLE LINE NOT CONNECTED >40 SSample line not connected or connection lostAirway Module unable to detect sample line connection after 40 secondswarning
MGAS SAMPLE LINE NOT CONNECTED >40 SCheck D-FendMGAS communicates OpenGasCircuit alarm for 40 secondswarning
MGAS SENSOR INOP > XXAirway Module sensor is inoperativeSensor malfunction detected during system checkerror
MGAS SENSOR INOP > XXModule fail. No CO2, AA, O2 dataMGAS communicates hardware failure or UPI does not initializewarning
MIXER ACB COM FAILMixer unable to communicate with anesthesia control board; serial link failureACB sends command to mixer but receives no response; timeout occurserror
MIXER ACB COM FAILAlternate O2 ScreenFive seconds pass without measured flow data from mixerwarning
MIXER ACB COM FAILLLOST CMDAlternate O2 ScreenMixer lost AC flow commands for 5 seconds or received illegal commandswarning
MIXER ACB COM FAIL LOST CMDMixer communication lost during command transmission; packet loss or corruptionMixer command transmission started but not completed; link interruptionerror
MIXER AIR SELECTION VLV FAILAlternate O2 ScreenAir selector valve status does not match commanded statewarning
MIXER BAL GAS CHANGE FAILBalance gas selection failure; mixer unable to switch balance gas channelMixer selection valve unable to change balance gas sourceerror
MIXER BAL GAS CHANGE FAILAlternate O2 ScreenAfter balance gas change command, selector valve shows old balance gas still connectedwarning
MIXER BAL GAS CHANGE FAILMixer balance gas change failureUnable to change balance gas settingerror
MIXER BAL GAS FLOW CHECK FAILMixer status bitBalance gas proportional valve fails flow checkwarning
MIXER BALGAS Flow FAILMixer error bitPressure difference between P3 and P2 differs from expected at measured flow for channel 2warning
MIXER BAL GAS F SENSOR FAILAlternate O2 ScreenBalance gas flow sensor failurewarning
MIXER BAL GAS TEMP LIMITMixer error bitBalance gas temperature exceeds 50 °Cwarning
MIXER BAL GAS TSENSOR FAILMixer error bitBalance gas temperature sensor failurewarning
MIXER BAL GAS ZERO FAILLLEAKMixer status bitBalance gas proportional valve fails zero flow check, shows flow while closedwarning
Mixer BoardCommunications status of Mixer BoardSystem checks communication link during Tools — Communication Statusunknown
MIXER CPU TEST FAILMixer CPU self-test failed during startup; processor malfunctionMixer processor self-test detects CPU defect during POSTerror
MIXER CPU TEST FAILAlternate O2 ScreenMixer CPU test failurewarning
MIXER CRC EEPROM FAILMixer EEPROM checksum error; data integrity check failedEEPROM CRC calculation does not match stored checksumerror
MIXER CRC FLASH FAILMixer flash memory checksum error; program code integrity check failedFlash memory CRC calculation does not match stored checksumerror
MIXER N2O SELECTION VLV FAILN2O selector valve status does not match commanded stateMixer Status Bit STS_SELV_VN2O_NOTIFY_FAILerror
MIXER O2 FLOW CHECK FAILO2 proportional valve fails flow checkMixer status bit STS_FLOW_TEST_CH1_FAIL1 LPM_FLOW_TEST_FAILerror
MIXER O2 Flow FAILAlternate O2 ScreenPressure difference differs from expected at measured flow for Channel 1 (O2)warning
MIXER O2 F SENSOR FAILO2 flow sensor failureMixer error bit STS_F2F1_SENSOR_FAIL (O2)error
MIXER O2 SELECTION VLV FAILO2 selector valve status does not match commanded stateMixer Status Bit STS_SELV_VOXY_NOTIFY_FAILerror
MIXER O2 TEMP LIMITO2 temperature exceeds 50 °CMixer error bit STS_CH1_TEMP_LIMIT (O2)error
MIXER O2 TSENSOR FAILO2 temperature sensor failureMixer error bit STS_T1_SENSOR_FAILerror
MIXER O2 ZERO LEAK FAILO2 proportional valve fails zero flow checks for leaksMixer status bit STS_CH1_ZERO_FLOW_TESP ROPN_V_LEAK_FAILT_FAILerror
MIXER POST FAILMixer power-on self-test (POST) failed during system startupMixer initialization test failed during system boot sequenceerror
MIXER PRES SENSOR P1P3 FAILMixer pressure sensor P1/P3 malfunction; one or both sensors not respondingPressure sensor P1 or P3 returns invalid or out-of-range readingerror
MIXER P SENSOR P2 FAILMixer pressure sensor P2 malfunction; sensor not responding or reading invalidPressure sensor P2 returns invalid or out-of-range reading during diagnosticserror
MIXER RAM FAILRuntime walking pattern check on RAM failedMixer Status Bit STS_RAMPATTERN_FAILerror
MIXER RAM FAILMixer RAM memory failure; self-test detected defective memoryMixer RAM self-test failed during startup or operationerror
MIXER SW WDOGMixer Watchdog has been activatedWatchdog timer expirationwarning
MIXER SW WDOG FAILMixer software watchdog timer triggered; processor hang or firmware deadlockMixer watchdog timer expired; firmware did not reset watchdog countererror
MIXER VOLTS FAILMixer power supply (on board) is out of toleranceStatus bit STS_VOLT_REF_FAIL. +12.5 V (10 VA) to mixer not OK.error
MIXER XX FLOW CHECK FAILFlow verification failed on specified gas channel; measured flow not as expectedMixer flow verification test shows actual flow does not match commanded valueerror
MIXER XX FLOW CHECK FAILMixer gas-specific flow check failureFlow check on specified gas line failederror
MIXER XX Flow FAILFlow failure on specified gas channel; unable to deliver requested flowProportional valve unable to achieve commanded flow on specified channelerror
MIXER XX Flow FAILMixer gas-specific flow failureUnable to deliver specified gas flowerror
MIXER XX F SENSOR FAILFlow sensor failure on specified gas channel (XX = BAL GAS, O2, AIR, or N2O)Flow sensor malfunction detected on specified channel during diagnosticserror
MIXER XX F SENSOR FAILMixer gas-specific flow sensor failureFlow sensor on specified gas line malfunctionerror
MIXER XX SELECTION VLV FAILSelection valve failure on specified gas channel; valve unable to switchMixer selection valve malfunction; solenoid unresponsive to control signalerror
MIXER XX SELECTION VLV FAILMixer gas-specific selection valve failureSelection valve for specified gas malfunctionerror
MIXER XX TEMP LIMITMixer gas-specific temperature limit exceededTemperature on specified gas line exceeded limitwarning
MIXER XX TEMP LIMITGas temperature on specified channel exceeded safe operating limitTemperature sensor reading above maximum safe temperature thresholdcritical
MIXER XX TSENSOR FAILTemperature sensor failure on specified gas channel (XX = BAL GAS, O2, etc)Temperature sensor malfunction on specified channel during startuperror
MIXER XX TSENSOR FAILMixer gas-specific temperature sensor failureTemperature sensor on specified gas line malfunctionerror
MIXER XX ZERO FAIL LEAKZero calibration failed on specified channel due to detected internal leakMixer zero calibration unable to pass; internal leak detected on channelerror
MIXER XX ZERO FAIL LEAKMixer gas-specific zero calibration failure, possible leakZero calibration failed or gas leak detectedwarning
MIX FRSH GAS SLCT VLV 10VA OVR CURFresh gas selection valve overcurrent; solenoid drawing excessive current+12V supply to fresh gas selection valve shows overcurrent or voltage droperror
MIX O2 BYPASS VLV 10VA OVER CURRCurrent feedback from O2 Bypass Selector incorrect for 7+ readingsExcessive current readings from O2 Bypass Selector Valveerror
MODULE NOT COMPATIBLEInstalled Airway Module is not compatible with systemSystem detects incompatible module firmware or hardware revisionerror
MVexp lowExpiratory minute volume below acceptable levelMeasured expiratory minute volume falls below minimum thresholdwarning
N2O PRESS LOWN2O supply pressure lowN2O pipeline <252 kPa and cylinder <2633 kPawarning
Negative airway pressureNegative pressure detected in airway; potential PEEP valve failure or obstructionPressure transducer reads below zero during exhalation phaseerror
NO EXPIRATORY FLOW SENSORNo Expiratory sensor connected and not calibratingExpiratory sensor not detected during startuperror
No Fresh Gas Flow!Fresh gas flow not detected during ventilation; gas delivery failureNo flow measured from gas outlet during ventilation attemptcritical
NO INSP/EXP FLOW SENSORNo inspiratory/expiratory flow sensor detected or functional during ventilator circuit testFull Test - Vent and Gas executionwarning
NO INSP/EXP FLOW SENSORFlow sensors have been removed from circuitWhen instructed to calibrate flow sensors; message appears after sensor removalinfo
NO INSPIRATORY FLOW SENSORNo Inspiratory sensor connected and not calibratingInspiratory sensor not detected during startuperror
***NOTE: Alarm speaker not detected. Check connection.Speaker connection fault detected at startupSystem startup POST detects no speaker or failed connectionwarning
***NOTE: CMOS battery is weak. Please replace.Carrier board CMOS battery voltage insufficientSystem startup POST detects weak CMOS batterywarning
***NOTE: CPU Board Supply Voltage Out of Range.CPU power supply voltage outside specificationSystem startup POST detects 1.5V_CB, 5V_CB, or 3.3V_CB out of rangewarning
***NOTE: RTC date/time error. Battery may be weak.Real-time clock date/time not set or invalidSystem startup POST detects RTC errorwarning
O2 CAL ERRORO2 sensor calibration offset, slope, cell voltage not in range or O2 >110%Galvanic O2 sensor calibration failurewarning
O2 FLOW CTRL FAILO2 flow control failureMixer status bit STS_FLOW_CTRL_CH1_FAIL with O2 gas supply pressure OKerror
O2 FLUSH FAILUREO2 flush switch detected on continuously >30 secSwitch reads on for >30 sec without cyclingwarning
O2 PIPE INVALIDCannot monitor O2 pipeline; O2 Pipeline pressure is invalidO2 pressure transducer malfunctionerror
O2 PRESS LOWO2 supply pressure lowO2 pipeline <252 kPa and O2 cylinder <2633 kPa for 1 seccritical
O2 PROP VALVE FAILO2 proportional valve failureMixer status bit STS_CH1_PROPN VALVE FAILerror
O2 SENSOR FAILUREO2 sensor failure; O2 <5%Galvanic O2 sensor connected and O2 reading lowwarning
O2 SENSOR NOT CONNECTEDOxygen sensor not connected to Airway ModuleAirway Module unable to detect O2 sensor connectionwarning
ON/STANDBY SWITCH TO STANDBYOn/Standby switch transitions from On to StandbySwitch moves to Standby position during therapycritical
PATIENT VOLUME MISMATCH OCCURREDPatient Volume Mismatch alarm occurred; calibrate, dry, or replace flow sensorsInspiration/expiration volume discrepancy during checkoutwarning
PCB Alarm Off: 1 min time leftBattery alarm displayed during system checkoutLow battery time detection during checkoutinfo
PCB Alarm Off: AC Mains failureAC Power RemovedAC mains voltage lossinfo
PCB Alarm Off: DC-DC power module failureSystem on battery despite AC connected (battery status LOW)AC GOOD HIGH but BATT STAT 1 and 2 LOWinfo
PCB Alarm On: AC mains failureAC Power RemovedAC mains voltage lossinfo
PCB Alarm On: Bulk charge time exceeded in stdbyBatteries in bulk charge >12 hours while in StandbyExtended bulk charge duration in Standby modeerror
PCB Saved Alarm Off: 1 min time leftBattery alarm displayed during system checkoutLow battery time detection during checkoutinfo
PCB Saved Alarm Off: AC Mains failureAC Power RemovedAC mains voltage lossinfo
PCB Saved Alarm Off: Blk chrg time exceed in stdbyBatteries in bulk charge >12 hours while in StandbyExtended bulk charge duration in Standby modeerror
PCB Saved Alarm Off: DC-DC power module failureCurrent draw from batteries >0.25A for 2 min while AC connectedExcessive battery current drain with AC mains presenterror
PCB Saved Alarm On: 1 min time leftBattery alarm displayed during system checkoutLow battery time detection during checkoutinfo
PCB Saved Alarm On: AC Mains failureAC Power RemovedAC mains voltage lossinfo
PCB Saved Alarm On: DC-DC power module failureCurrent draw from batteries >0.25A for 2 min while AC connectedExcessive battery current drain with AC mains presenterror
PCB Saved Alarm On: DU to PSC Comm ErrorPower Controller to Display Unit communication was lost then restoredCommunication loss and subsequent recoveryinfo
PCB Saved Error: POWER CNTRL COM FAILCommunication lost between Display Unit and ACB; error stored until re-establishedCommunication loss between Display Unit computer and ACBerror
PEEP high. Blockage?Positive end-expiratory pressure elevated; possible obstruction detectedPEEP rises above set value, indicating possible blockageerror
PEEP High. Blockage?Measured PEEP higher than set value; possible breathing circuit obstructionPositive end-expiratory pressure exceeds commanded setpointerror
PEEP PCV NOT AVAILABLEVolume ventilation only; no PEEP or PSV availablePaw data in range but Pmanifold <= -15 cmH2Owarning
Plug in power cable. On battery.Power failure; system operating on battery powerMains power disconnected while system is in operationcritical
PORT PLUG POST FAILVentilation fail; monitoring onlyPaw > 100 cmH2O for 1 second in POST statewarning
Power Controller BoardCommunications status of Power Controller BoardSystem checks communication link during Tools — Communication Statusunknown
POWER CONTROLLER COM FAILInternal failure; system may shut downCommunications with PC and DC cannot be established for 10 secondserror
POWER CONTROLLER SELF TESTInternal failure; Power Controller fails self-testsPower Controller fails self-testscritical
POWER SUPPLY 75CCircuitry >75°C; shutdown possiblePower supply temperature exceeds 75°Cwarning
Ppeak highPeak inspiratory pressure higher than expected; possible resistance or obstructionMeasured peak pressure exceeds acceptable range for ventilation parameterserror
Ppeak low. Leak?Peak inspiratory pressure below expected level; possible leak detectedPeak pressure falls below set value, indicating possible circuit leakwarning
Pressure Sensor 1/2/3 FailurePressure transducer (P1/P2/P3) on Mixer out of specificationPressure transducer out of specificationerror
PWR CNTRL DC-DC FAILUsing battery; PC fail. AC supply OK but system reports using batteryAC GOOD HIGH but BATT STAT 1 and 2 LOWwarning
Quick check failsMessage or failure displayed during system checkoutFailure or message during system checkouterror
Replace insp / exp flow sensorInspiratory or expiratory flow sensor has failed and requires replacementFlow sensor failure confirmed during diagnostic testingerror
REVERSE EXPIRATORY FLOWReverse expiration flow detected; check valves may be faultyFlow towards patient on both expiratory and inspiratory sensors during inspiration for 6 consecutive mechanical breathswarning
Reverse insp / exp FlowInspiratory and expiratory flow sensor channels may be reversed or misconfiguredFlow sensor readings inconsistent with breathing pattern during checkoutwarning
SCGONon-Circle SCGO selected; volume and apnea monitoring offNon-Circle SCGO mode selectedinfo
SEVERE SUSTAINED PAWMeasured airway pressure >100 cmH2O for 10 secondsAirway pressure exceeds 100 cmH2O for 10 secondserror
STANDBY PATIENT DETECTIONNo fresh gas flow; patient detected during standby3 volume breaths or 3 CO2 breaths detected within 30 seconds during system checkoutinfo
System LeakLeak detected in ventilator or breathing system pneumatics; gas escape detectedBellows pressure or flow imbalance; breathing system leak detectionerror
System Self-tests failedPower-on tests failedPower-on tests failerror
Temp Measurement Ch1 HighChannel 1 temperature measurement above safe operating range; sensor reading highTemperature sensor on channel 1 reading above acceptable thresholderror
Temp Measurement Ch2 HighChannel 2 temperature measurement above safe operating range; sensor reading highTemperature sensor on channel 2 reading above acceptable thresholderror
Temp Sensor Fail Balance Gas ChTemperature sensor failure on balance gas channel; sensor malfunctionBalance gas temperature sensor unresponsive during system startuperror
Temp Sensor Failure Ch1Temperature sensor failure on channel 1; sensor malfunction detectedChannel 1 temperature sensor unresponsive or returns invalid readingerror
TOUCHSCREEN FAILURETouchscreen controller fails communications POST testTouch panel controller fails POST communications testwarning
Tv not achievedTidal volume not reached; measured volume less than set volumeVentilator unable to deliver programmed tidal volume during breathing cycleerror
Unable to Drive BellowsVentilator bellows cannot be driven; no bellows movement during ventilation attemptDrive gas pressure absent or insufficient; bellows mechanical failurecritical
Valve Ch1 Leak Tests Not DoneO2 leak test skippedO2 leak test skipped (no O2 supply at power-up)info
VENT 1.22V FAILVentilation fail; monitoring onlyVoltage <1.074 Vdc or >1.367 Vdcwarning
VENT +12.5V FAILVentilation fail; monitoring onlyNominal 12.5V <11.3 Vdc or >13.13 Vdcwarning
VENT -6V FAILVentilation fail; monitoring onlyVSIB -6V out of range <-6.72 Vdc or >-5.28 Vdcerror
VENT +6V FAILVentilation fail; monitoring onlyVSIB +6V out of range <5.51 Vdc or >6.5 Vdcerror
VENT ADC VREF FAILVentilation fail; monitoring onlyVSIB ADC 3.200V ref voltage <3.179 or >3.221 Vdcerror
VENT AIRWAY OVERPRESS SIGNALInspiration stopped; high airway overpressure signal setHigh airway overpressure signal set during mechanical ventilationwarning
VENT AIRWAY OVERPRESS SIGNAL FAILVentilation fail; monitoring onlyHigh airway overpressure signal set with Paw <90 cmH2O and Pmanifold <80 cmH2Owarning
Vent check stage 1 failedFailure displayed during system checkout stage 1Failure during checkout stage 1error
Vent check stage 2 failedFailure displayed during system checkout stage 2Failure during checkout stage 2error
VENT FLOW VALVE FAIL CURRENTVentilation fail; monitoring onlyIncorrect current feedback for 7 consecutive readingswarning
VENT FLOW VALVE FAIL DACVentilation fail; monitoring onlyIncorrect DAC feedback for 3 consecutive readingswarning
Ventilate ManuallyAutomatic ventilation failed; manual ventilation mode recommended or requiredVentilator subsystem failure; unable to provide mechanical ventilationerror
Ventilator has no Drive GasDrive gas supply to ventilator bellows is absent or insufficient for operationDrive gas pressure below minimum threshold during ventilation attemptcritical
Ventilator Interface BoardCommunications status of Ventilator Interface BoardSystem checks communication link during Tools — Communication Statusunknown
VENT SIB 10VA OVERCURRENTCurrent feedback from cVIB incorrectIncorrect current feedback from cVIBerror
VENT SIB COMMUNICATION FAILURELoss of communication between ACB and cVIBLoss of communication between ACB and cVIB after initial establishmenterror
VENT SUSTAINED PAW SDOWNVentilation fail; monitoring onlyPaw >100 cmH2O for 10 secondswarning
VENT VALVE 10VA OVERCURRENTCurrent feedback from Insp Flow Valve incorrect for 7 consecutive readingsIncorrect current feedback from Insp Flow Valveerror
VENT VALVE POWER FAILVentilation fail; monitoring onlyNominal 12.5V <11.3 V or >13.13 Vdcwarning
Vlv Bal Gas Ch Leak Tests Not DoneBalanced Gas leak test skippedBalanced Gas leak test skipped (no Balanced Gas at power-up)info
Vol vent only. No PEEP or PSVVolume control ventilation available but no PEEP or PSV modes working; feature unavailableDrive gas or control system failure; mode selection errorerror
XX FLOW CTRL FAILFlow control failure on specified gas channel (XX = BAL GAS, O2, AIR, N2O)Proportional valve unable to reach commanded flow setpointerror
XX FLOW CTRL FAILGas-specific flow control failureFlow control on specified gas line malfunctionerror
XX PROP VALVE FAILProportional valve failure on specified gas channel; valve not responding to controlProportional valve solenoid or electronics malfunction; no valve responseerror
XX PROP VALVE FAILGas-specific proportional valve failureProportional valve on specified gas line malfunctionerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.