GE Healthcare ApexPro: troubleshooting and error codes
The ServiceForge knowledge base holds 59 documented symptoms for the GE Healthcare ApexPro, with 83 ranked causes, 114 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Apex Oximeter SpO2 parameter not displayed
SpO2 parameter box does not appear on Clinical Information Center after connecting oximeter to transmitter and admitting to system
Ranked causes
- 1. Oximeter not properly connected to transmitter (common)
- 2. Transmitter not admitted to ApexPro telemetry system (occasional)
- 3. Oximeter hardware or firmware fault (rare)
Verify oximeter serial number recorded
- Test point
- Oximeter label
- Expected
- Serial number documented
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequent lead fail events
Lead fail alarms occurring regularly during transmitter operation
Ranked causes
- 1. Leadwire damage from wrapping or handling (common)
- 2. Leadwire internal damage or brittleness (common)
- 3. Improper patient electrode preparation (occasional)
Inspect leadwires for visible damage, kinks or brittle areas
- Test point
- Entire leadwire set
- Expected
- No cracks, breaks or sharp bends visible
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 Data message
Clinical Information Center displays 'No SpO2 Data' message
Ranked causes
- 1. ApexPro transmitter not used or transmitter not configured for SpO2 (common)
- 2. Oximeter disconnected from transmitter (common)
- 3. Patient out of antenna range (occasional)
Verify oximeter connection to transmitter
- Test point
- Oximeter connector
- Expected
- Oximeter firmly connected to transmitter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nonin Xpod Oximeter not transmitting data patient safety
Oximeter connected but SpO2 data not appearing at Clinical Information Center
Ranked causes
- 1. SpO2 probe not connected to oximeter (common)
- 2. Oximeter not connected to transmitter (common)
- 3. Wrong probe type installed (occasional)
Verify SpO2 probe connection to oximeter
- Test point
- 8-pin connector on oximeter
- Expected
- Nonin SpO2 probe connected to 8-pin connector
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Short battery life in Apex Oximeter
Oximeter battery depletes faster than expected; low battery indicator appears
Ranked causes
- 1. Used or expired batteries installed (common)
- 2. Rechargeable batteries used instead of alkaline (occasional)
- 3. Battery beyond expiration date (occasional)
Verify battery type and expiration date
- Test point
- Battery compartment
- Expected
- New alkaline AA batteries within expiration date installed
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Short battery life in transmitter
Transmitter operates for less than expected duration on batteries
Ranked causes
- 1. Used or expired batteries installed (common)
- 2. Rechargeable batteries used instead of alkaline (occasional)
- 3. Battery beyond expiration date (occasional)
Verify battery type and status
- Test point
- Battery compartment
- Expected
- New alkaline AA batteries within expiration date
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform dropout
Periodic loss of waveform signal on the display
Ranked causes
- 1. Transmitter out of antenna range (common)
- 2. External RF interference (occasional)
- 3. Defective transmitter (occasional)
Determine type of dropout by enabling flags
- Test point
- RF Signal Integrity flags on page 4-16
- Expected
- Flags indicate type of dropout
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accutracker does not transmit or receive data correctly
Accutracker display numbers do not match display numbers on Clinical Information Center screen during communication test
Ranked causes
- 1. Transmitter-receiver communication link failure (unknown)
- 2. ECG leads or shorting cable not attached to transmitter (occasional)
Verify ECG leads or shorting cable attached to transmitter
- Test point
- ApexPro Telemetry system ECG input
- Expected
- ECG leads with shorting cable or simulator connected
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accutracker self-test does not complete or fails
Unit does not display TEST PASSED after diagnostic self-test, or unable to abort reading when OFFICE SELF TEST displays
Ranked causes
- 1. Pump or valve circuit malfunction (unknown)
- 2. Reading abort function not responding (occasional)
Turn off Accutracker and wait 5 seconds
- Test point
- Accutracker power
- Expected
- Unit powered off, wait 5 seconds to prevent lockup
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter does not power on
Unit does not respond to power button press; no LED or display indication
Ranked causes
- 1. Batteries not installed or depleted (common)
- 2. Battery polarity reversed (occasional)
Verify batteries present and fresh
- Test point
- Battery compartment
- Expected
- 2 new AA alkaline batteries installed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter weak or poor SpO2 signal patient safety
Perfusion LED indicates weak or very poor quality SpO2 signal; red LED indication
Ranked causes
- 1. Poor SpO2 probe contact or signal quality (common)
- 2. Marginal signal quality (occasional)
Observe perfusion LED color
- Test point
- Perfusion LED indicator
- Expected
- Green indicates acceptable signal strength
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Probe message on oximeter
Clinical Information Center displays 'Check Probe' message from oximeter
Ranked causes
- 1. Probe disconnected from oximeter (common)
- 2. Probe sensor defective (occasional)
Verify probe connection to oximeter
- Test point
- Probe connector
- Expected
- Probe firmly connected to oximeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal gain incorrect
Graph output signal level does not measure 1 cm for 1 mV input when gain set to 1X
Ranked causes
- 1. ECG gain setting not at 1X (common)
- 2. Transmitter/receiver system signal attenuation (occasional)
Verify ECG simulator connected
- Test point
- Transmitter input
- Expected
- ECG simulator connected
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Quality Signal message on oximeter
Clinical Information Center displays 'Low Quality Signal' message from oximeter
Ranked causes
- 1. Poor patient connection or probe placement (common)
- 2. Probe sensor defective (occasional)
Verify probe attachment to patient
- Test point
- Probe placement on patient
- Expected
- Probe properly secured to patient
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker not detected or transmitted
Pace pulse triggered at ECG simulator but pace mark does not appear on Clinical Information Center display
Ranked causes
- 1. PACE function not enabled on Clinical Information Center (common)
- 2. ECG simulator not connected or not generating pace pulse (occasional)
Verify ECG waveforms displaying on Clinical Information Center
- Test point
- Clinical Information Center display
- Expected
- ECG waveforms visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe Off message on oximeter
Clinical Information Center displays 'Probe Off' message from oximeter
Ranked causes
- 1. Probe not attached to patient (common)
- 2. Sensor defective (occasional)
Verify probe attachment to patient
- Test point
- Probe placement
- Expected
- Probe attached to patient skin
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Saturation value is 'x' message on oximeter
Clinical Information Center displays 'The Saturation Value is x' message lasting up to 3 minutes
Ranked causes
- 1. Patient out of antenna range (common)
- 2. Oximeter disconnected from transmitter (common)
Verify oximeter connection to transmitter
- Test point
- Oximeter connector
- Expected
- Oximeter firmly connected to transmitter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accutracker hardware time-out or system leak exceeds specification
Hardware time-out does not occur at 3 minutes ±45 seconds, or system leaks more than 2 mmHg per minute
Ranked causes
- 1. Hardware time-out circuit malfunction or system leak in cuff/tubing (unknown)
Re-enter technical calibrate mode
- Test point
- Accutracker power button
- Expected
- Unit in Technical Calibrate mode; no external pressure applied
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accutracker LCD display does not show version information on startup
Hardware version, software version, gain setting, or pressure setting not displayed during startup when power button held while powering on
Ranked causes
- 1. Display or microprocessor fault (unknown)
Power on Accutracker with button held
- Test point
- Accutracker power button and LCD display
- Expected
- Startup display shows hardware version, software version, gain setting, and pressure setting in mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accutracker over-pressure circuit does not release caution
Valve #1 does not open and pressure does not release when 285 mmHg is applied to pressure transducer
Ranked causes
- 1. Over-pressure release circuit failure (unknown)
Re-enter technical calibrate mode
- Test point
- Accutracker power button
- Expected
- Unit in Technical Calibrate mode; no external pressure applied
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accutracker pressure reading exceeds tolerance
Pressure display differs by more than +/– 2 mmHg from mercury column or PDM200 reference at any test point (0–250 mmHg in 50 mmHg increments)
Ranked causes
- 1. Pressure transducer calibration drift or fault (unknown)
Connect PDM200 or Mercury manometer to patient cable/cuff
- Test point
- Accutracker patient cable/cuff connection
- Expected
- Manometer connected; no external pressure applied to Accutracker
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter CHECK PROBE message not displayed after disconnect
CHECK PROBE message does not appear on Clinical Information Center when oximeter is disconnected from simulator
Ranked causes
- 1. Communication fault between oximeter and Clinical Information Center (unknown)
Disconnect oximeter from simulator
- Test point
- Oximeter simulator cable
- Expected
- Cable disconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter display does not turn on
Digital display remains off after powering up unit or pressing Display On/Off button
Ranked causes
- 1. Display in power-save mode after one minute timeout (common)
Observe power LED status
- Test point
- Power LED in upper left corner of display area
- Expected
- Flashing power LED indicates unit powered on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter display flashing
Digital display on Apex Oximeter starts flashing, indicating low battery reserve power remaining
Ranked causes
- 1. Batteries depleted (common)
Observe digital display flashing pattern
- Test point
- Digital display
- Expected
- Digital display starts flashing when battery reserve depleted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter perfusion indication incorrect
Perfusion LED does not change to yellow when selector switch changed to Ohmeda mode
Ranked causes
- 1. Perfusion LED fault or sensor malfunction (unknown)
Set simulator selector switch to Ohmeda
- Test point
- Simulator selector switch
- Expected
- Switch in Ohmeda position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apex Oximeter powerup sequence abnormal
Display and perfusion LED do not cycle through specified sequence on startup; may indicate software or hardware failure
Ranked causes
- 1. Software or display controller malfunction (unknown)
Observe display shows 888 888 with red perfusion LED
- Test point
- Display and perfusion LED
- Expected
- 888 888 display and red perfusion LED for approximately 1 second
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery contacts dirty or corroded caution
Battery contacts show dirt, corrosion, or oxidation in battery compartment
Ranked causes
- 1. Corrosion or dirt accumulation on contacts (occasional)
Open battery compartment and inspect contacts
- Test point
- battery compartment contacts
- Expected
- contacts clean and free of corrosion or dirt
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carrier frequency out of tolerance
Transmitter center frequency deviates more than ±650Hz from assigned frequency when measured with manufacturing code active
Ranked causes
- 1. Frequency synthesizer malfunction or calibration drift (rare)
Manufacturing code activation
- Test point
- Interface connector pins 2 and 4
- Expected
- Pins shorted; transmitter outputs test pattern for frequency measurement
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Change Battery LED flashing
LED indicator on transmitter flashes to indicate low battery power; approximately one hour of reserve power remains before unit shuts down
Ranked causes
- 1. Battery power depleted (common)
Observe Change Battery LED indicator status
- Test point
- Change Battery LED on transmitter
- Expected
- LED flashing indicates low battery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dust covers loose or missing caution
Dust covers on transmitter ports not securely attached or missing
Ranked causes
- 1. Dust covers not properly attached (occasional)
Inspect dust covers for secure attachment
- Test point
- all dust cover locations
- Expected
- all dust covers securely attached
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Environmental seals compromised patient safety
Case seals opened or damaged, potentially from impact or drop
Ranked causes
- 1. Unit dropped or impacted (unknown)
Inspect case seals for opening or damage
- Test point
- all case seals
- Expected
- seals intact and properly sealed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Graph button not functioning
Pressing Graph button does not trigger graph run at configured printer device; LEDs do not acknowledge button press
Ranked causes
- 1. Graph button circuit malfunction or CIC printer not configured (common)
LED acknowledgment of Graph button press
- Test point
- LEAD STATUS and CHANGE BATTERY LEDs
- Expected
- Both LEDs flash twice when Graph button pressed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LA LED flashes rapidly after power-up
Only the LA LED flashes rapidly when transmitter is powered on
Ranked causes
- 1. Manufacturing/service code in EEPROM erased or corrupted (common)
Power cycle transmitter and observe LED status
- Test point
- LA LED indicator
- Expected
- LA LED flashes rapidly only
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead fail detection not functioning patient safety
Transmitter does not detect open or disconnected leadwires; lead status LEDs do not indicate failure; CIC does not show lead fail status
Ranked causes
- 1. Lead sensing circuit malfunction or software error (common)
Initial ECG signal reception
- Test point
- Clinical Information Center waveform display
- Expected
- ECG signals display at CIC after transmitter admitted to receiver
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leadwire connector pins corroded or bent caution
Leadwire connection pins show corrosion, oxidation, or physical bending
Ranked causes
- 1. Corrosion or oxidation on connector pins (occasional)
Inspect leadwire connector pins for corrosion or bending
- Test point
- leadwire connectors
- Expected
- pins clean, straight, and free of corrosion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leadwires cracked or damaged caution
Leadwires have visible cracks, damage, or reduced flexibility
Ranked causes
- 1. Physical damage or material degradation (occasional)
Inspect leadwires for cracks, damage, or loss of flexibility
- Test point
- entire length of leadwires
- Expected
- leadwires intact, flexible, and undamaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED Displays not operating
LEAD STATUS, CHANGE BATTERY, or PAUSE ALARM LEDs do not flash or illuminate as expected during button press or power-up
Ranked causes
- 1. LED circuit malfunction or loose connections (unknown)
LED response to button press
- Test point
- LEAD STATUS and CHANGE BATTERY LEDs
- Expected
- Flash twice when VERIFY LEADS and GRAPH buttons pressed simultaneously, then PAUSE ALARM LED flashes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery condition not detected caution
CHANGE BATTERY LED does not flash when battery voltage drops below 1.9V; low battery message not displayed at Clinical Information Center
Ranked causes
- 1. Low battery detection circuit malfunction or battery voltage not reaching threshold (common)
Transmitter startup voltage
- Test point
- Battery voltage input
- Expected
- 2.3V minimum to start transmitter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery message displays on Accutracker
Low Batt message appears on the Accutracker DX monitor display
Ranked causes
- 1. AA batteries depleted (common)
Observe Low Batt message on Accutracker display
- Test point
- Accutracker monitor display
- Expected
- Low Batt message visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Membrane switch cracked or damaged caution
GRAPH and VERIFY LEADS control membrane visibly cracked or damaged
Ranked causes
- 1. Physical damage to membrane (unknown)
Inspect membrane switch for cracks or damage
- Test point
- GRAPH and VERIFY LEADS membrane controls
- Expected
- membrane intact without cracks or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Membrane switch label loose or missing caution
Membrane switch label not securely attached to transmitter case
Ranked causes
- 1. Label not properly attached (occasional)
Verify membrane switch label secure attachment
- Test point
- membrane switch label
- Expected
- label securely attached to case
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Moisture detected in battery compartment patient safety
Visible signs of moisture or water inside the battery compartment
Ranked causes
- 1. Water ingress into unsealed battery compartment (rare)
Inspect battery compartment for visible moisture
- Test point
- inside battery compartment
- Expected
- no visible moisture or water
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nonin Xpod Oximeter CHECK PROBE message not displayed after disconnect
CHECK PROBE message does not appear on Clinical Information Center when Nonin Xpod oximeter is disconnected from simulator
Ranked causes
- 1. Communication fault between Nonin Xpod and Clinical Information Center (unknown)
Disconnect Nonin Xpod oximeter from simulator
- Test point
- Oximeter simulator cable
- Expected
- Cable disconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nonin Xpod Oximeter short battery life in transmitter
ApexPro transmitter battery depletes faster when Nonin Xpod Oximeter connected
Ranked causes
- 1. Nonin Xpod draws power from transmitter battery (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oximeter heart rate accuracy out of tolerance
Heart rate measurement at simulator does not meet specified accuracy at test rates (70, 100, 160 BPM)
Ranked causes
- 1. Oximeter sensor or signal processing fault (unknown)
Place Apex Oximeter in continuous display mode
- Test point
- Oximeter Display On/Off switch
- Expected
- Continuous display mode active
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pause Alarm LED does not flash
LED should flash for programmed period (typically 5 min) after pressing Graph and Verify Leads buttons simultaneously
Ranked causes
- 1. Transmitter not entering PAUSE ALARMS condition (unknown)
Verify Graph and Verify Leads buttons respond
- Test point
- Transmitter buttons
- Expected
- LEAD STATUS and Change Battery LEDs flash twice
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power shuts off during leads fail
RF output amplifier power automatically shuts down when all leads are disconnected or fail for more than approximately six seconds. Power resumes about one second after leads are reconnected.
Ranked causes
- 1. All leads disconnected or failed for >6 seconds (common)
Verify all leads are connected
- Test point
- Patient lead connectors
- Expected
- All leads securely seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Powerup selftests not completing
Transmitter does not perform memory, frequency synthesizer, or LED startup sequence tests after battery installation
Ranked causes
- 1. Battery installation issue or selftest failure (common)
Memory and frequency synthesizer selftest
- Test point
- Transmitter power-up sequence
- Expected
- Transmitter performs memory test, frequency synthesizer test, and LED startup sequence; begins normal operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RA LED flashes rapidly after power-up
Only the RA LED flashes rapidly when transmitter is powered on
Ranked causes
- 1. Application code in EEPROM erased or corrupted (common)
Power cycle transmitter and observe LED status
- Test point
- RA LED indicator
- Expected
- RA LED flashes rapidly only
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF power not reducing on lead failure
Transmitter does not reduce RF output when all leads fail for more than six seconds; NO TELEM signal does not display at Clinical Information Center
Ranked causes
- 1. RF power shutdown circuit malfunction (unknown)
NO TELEM signal display after lead failure
- Test point
- Clinical Information Center display
- Expected
- NO TELEM signal appears after six seconds of all-leads-fail condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF power output out of specification
Transmitter RF power output measured at RF monitor does not meet minimum -20dBm specification
Ranked causes
- 1. RF power amplifier malfunction or transmitter in leads-fail condition (common)
RF monitor temperature stabilization
- Test point
- RF service monitor equipment
- Expected
- Operate for at least 15 minutes before measurement
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF signal integrity poor or dropout observed patient safety
ECG waveform signal breakup or dropout observed on receiver when transmitter is connected to ECG simulator; waveform data missing on one or more admitted beds
Ranked causes
- 1. RF signal integrity issue caused by transmitter malfunction (occasional)
Diagnostic color mark interpretation
- Test point
- Clinical Information Center waveform display bottom
- Expected
- No ECG signal breakup on any waveforms; if dropout present, identify color (Yellow=transmitter data missing, Dark Green=receiver data missing, Magenta=receiver system data missing, Orange=hardware manager data missing, Light Gray=invalid telemetry data)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Single lead displayed with multi-lead cable connected
Multi-lead cable connected but only one lead appears on Clinical Information Center
Ranked causes
- 1. Transmitter configured for single lead use (common)
Review transmitter feature configuration
- Test point
- Transmitter settings
- Expected
- Multi-lead configuration enabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmitter case cracked or damaged patient safety
Visual cracks or physical damage visible on the transmitter case
Ranked causes
- 1. Physical damage or impact to case (unknown)
Inspect transmitter case for cracks or damage
- Test point
- entire case exterior
- Expected
- no cracks or visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmitter configuration does not match appearance patient safety
Transmitter labels and interface connector port dust cover colors do not match enabled features
Ranked causes
- 1. Configuration mismatch or incorrect labeling (rare)
Observe startup LED sequence
- Test point
- LEDs during power-up (second flash)
- Expected
- LED pattern matches transmitter configuration
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmitter EEPROM test fails at powerup
EEPROM test loops continuously on failure; transmitter does not proceed to frequency synthesizer test
Ranked causes
- 1. EEPROM failure in transmitter circuit (unknown)
Observe LED behavior at powerup after RAM test passes
- Test point
- Transmitter case LEDs
- Expected
- LED sequence should progress past EEPROM test phase
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmitter frequency synthesizer lock test fails at powerup
Frequency synthesizer fails to lock on specified customer frequency; transmitter resets itself
Ranked causes
- 1. Frequency synthesizer circuit failure or lock circuit malfunction (unknown)
Observe LED behavior at powerup after EEPROM test passes
- Test point
- Transmitter case LEDs
- Expected
- LED sequence should complete synthesizer lock and proceed to normal operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmitter RAM test fails at powerup
RAM test in digital signal processor loops continuously on failure; transmitter does not proceed to EEPROM test
Ranked causes
- 1. RAM failure in digital signal processor circuit (unknown)
Observe LED behavior at powerup after battery installation
- Test point
- Transmitter case LEDs
- Expected
- LED sequence should progress past RAM test phase
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TTX label missing or inaccurate patient safety
TTX label not present on transmitter or recorded data does not match programmed TTX number
Ranked causes
- 1. Label missing or data mismatch with programmed value (unknown)
Verify TTX label is present
- Test point
- transmitter case label area
- Expected
- TTX label present and readable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.