GE Healthcare ADVANTEX: troubleshooting and error codes

The ServiceForge knowledge base holds 261 documented symptoms for the GE Healthcare ADVANTEX, with 555 ranked causes, 510 diagnostic checks with test points and expected values, and 986 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Cone servo fails to reach requested position

Error 4B52 displayed on console and in-room display shows 'CHECK FOR OBSTRUCTION'. The Cone servo did not reach the requested position within the allotted timeout.

Ranked causes

  1. 1. Servo controller timeout - servo in position bit not asserted (common)
  2. 2. Feedback potentiometer failure (occasional)
  3. 3. Servo amplifier board malfunction (occasional)
  4. 4. Motor not moving (occasional)
  5. 5. Servo belt improper tension (occasional)
  6. 6. Servo damping or positioning calibration error (occasional)
  7. 7. R+F Positioner module firmware or memory issue (rare)
  8. 8. 80188 CPU timer malfunction (rare)
First check

Run Cone Axis Test diagnostics

Test point
R+F Servo Controller Board in R+F menu
Expected
Test passes without error

7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 420B displayed

System displays error code 420B during fluoroscopy operation

Ranked causes

  1. 1. Temporary grid short condition (occasional)
  2. 2. Intermittent grid shorts (occasional)
  3. 3. Tube spitting causing filament correction drift (occasional)
  4. 4. Invalid calibration database (rare)
  5. 5. ECM or Power Unit module fault (rare)
  6. 6. Power Unit system level fault (rare)
  7. 7. ECM system level fault (rare)
First check

Release expose switch and attempt exposure

Test point
Expose switch
Expected
Error 420B does not reoccur

2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Variable mask servo position timeout

Console displays 'Retry selection, cycle console power or call service'. In-room display shows 'RETRY OR CALL SERVICE'. System fails to move Variable Mask to requested position within timeout.

Ranked causes

  1. 1. Feedback potentiometer malfunction (common)
  2. 2. Servo amplifier board failure (common)
  3. 3. Motor not responding to command (occasional)
  4. 4. Servo belt slack or misaligned (occasional)
  5. 5. Servo calibration drift (occasional)
  6. 6. R+F Positioner module hardware fault (rare)
  7. 7. 80188 CPU timer malfunction (rare)
First check

Run Variable Mask Axis Test

Test point
R+F Servo Controller Board
Expected
Test passes without error

9 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament drive error during exposure radiation

Filament drive error detected; system initiates auto-reset; exposure unable to complete

Ranked causes

  1. 1. Power unit board or component failure (common)
  2. 2. Filament control voltage or demand issue (occasional)
  3. 3. Filament calibration data error (occasional)
  4. 4. Filament demand exceeds limits (occasional)
  5. 5. Filament inverter board malfunction (occasional)
  6. 6. Closed loop system or cable issue (occasional)
First check

Monitor filament demand voltage at J1-a13

Test point
J1-a13 on i/o board 46-264024
Expected
5V when error code generated
Tools
Voltmeter

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optical distributor not positioned high voltage

Console displays 'Optical distributor not positioned. Try again or call service.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' Gate rotate failed to position during initialization or normal move.

Ranked causes

  1. 1. Gate rotate electrical in position bit not received within 2 seconds at initialization (common)
  2. 2. Both electrical and mechanical in position bits not lost within 100 milliseconds during move (occasional)
  3. 3. Gate & NDF Control Board failure (rare)
  4. 4. Gate & NDF Interface Board failure (rare)
  5. 5. XRII Interface Board failure (rare)
  6. 6. Gate Sensor failure (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optical distributor positioning error

Console displays 'Optical distributor not positioned. Try again or call service.' In-room display shows 'TURN CONSOLE POWER OFF/ON.'

Ranked causes

  1. 1. ND Filter amp indicated stall at bad time (unknown)
  2. 2. Gate & NDF Control Board failure (unknown)
  3. 3. Gate & NDF Interface Board failure (unknown)
  4. 4. XRII Interface Board failure (unknown)
  5. 5. Gate Sensor failure (unknown)
  6. 6. Gate Mechanics failure (unknown)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Positioner not in valid Cerebral Contact position patient safety

Interlock message displayed at Prep; CC button flashing; film changer not meeting valid position requirements for CC mode

Ranked causes

  1. 1. Film changer distance exceeds 30 cm ISO distance requirement (common)
  2. 2. Film changer not at proper end of travel position for L/U angle combination (common)
  3. 3. CC mode not properly enabled or CC switch signal not detected (occasional)
  4. 4. L or U angle display incorrect for actual positioner position (occasional)
  5. 5. Hardware failure in angle sensors or switches (rare)
  6. 6. Firmware or power state issue (rare)
First check

Measure ISO distance of film changer from reference

Test point
ISO distance measurement point
Expected
<= 30 cm

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power unit tube spit recovery error high voltage

Console displays 'Power unit tube spit recovery. Continue'; in-room display shows 'POWER UNIT ERROR. CALL SERVICE.' kV is automatically cut off, automatic retry performed. After 7 spits during same exposure, exposure aborts with error 42CA.

Ranked causes

  1. 1. Tube not formed (new tube) (common)
  2. 2. Excessive kV or mA settings (common)
  3. 3. Tube degradation (old tube) (occasional)
  4. 4. HV tank problem (occasional)
  5. 5. Measurement/switching board issue (occasional)
  6. 6. Divider board calibration problem (rare)
First check

Determine if tube is new and not formed

Test point
Tube status
Expected
New tube not yet formed

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System communications problem

Console displays 'System communications problem. Please try again.' In-room display shows 'PLEASE TRY AGAIN.' Module attempted transmission but did not receive acknowledge before timeout.

Ranked causes

  1. 1. Destination ID invalid or node IDs set incorrectly (common)
  2. 2. Destination node is off or not receiving (common)
  3. 3. GE bus cables damaged or disconnected (common)
  4. 4. CPU RAM corrupted (occasional)
  5. 5. BIU board failure (occasional)
  6. 6. Applications floppy corrupted (occasional)
First check

Verify destination ID is valid for system

Test point
System configuration
Expected
Destination ID matches valid system node

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Camera did not respond high voltage

Console shows 'Camera did not respond. Try again or call service.' In-room display shows 'CAMERA DID NOT RESPOND.' System times out waiting for expose to generator signal.

Ranked causes

  1. 1. Expose to generator signal not reaching Power Amp board (common)
  2. 2. PREP/EXPOSE signal not reaching camera (common)
  3. 3. PCM CPU board timer or PIC failure (occasional)
  4. 4. PHOTOSPOT module RAM/Arcnet failure (occasional)
  5. 5. System hardware failure requiring restart (rare)
First check

Trace expose to generator signal path

Test point
Camera to Power Amp board
Expected
Signal present throughout path

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Camera shutter does not open radiation

Photospot prep fails; console displays 'Camera shutter did not open. Try again or call service'; in-room display shows 'TRY AGAIN. CAMERA FAULT.'

Ranked causes

  1. 1. Shutter mechanism stuck or not moving freely (common)
  2. 2. SHTR OPEN* signal not reaching shutter properly (occasional)
  3. 3. SHTR OPEN CMD* signal not being sent or shutter motor voltage insufficient (occasional)
  4. 4. CPU or timer hardware failure (rare)
  5. 5. Camera Shutter test failure (occasional)
First check

Observe shutter movement at camera

Test point
105 camera shutter
Expected
Shutter opens smoothly and moves freely
Tools
Visual inspection, listening

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Console receives bad data from ASC

Console displays 'Bad data received. Turn console power OFF/ON.' message; in-room display shows 'TURN CONSOLE POWER OFF/ON.'

Ranked causes

  1. 1. Invalid application command in GE bus packet from ASC (unknown)
  2. 2. Console CPU board RAM failure (unknown)
  3. 3. ASC CPU board RAM failure (unknown)
  4. 4. ASC GE bus interface (Arcnet) failure (unknown)
  5. 5. Corrupted system boot disk (unknown)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 4A5F displayed on console and in-room

Console displays 'Bad data received. Turn console power OFF/ON.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' CPU detected math division by zero error in Positioner module.

Ranked causes

  1. 1. Division by zero in Positioner CPU (unknown)
  2. 2. Hardware motion or potentiometer calibration issue (occasional)
  3. 3. Positioner module firmware or RAM failure (occasional)
  4. 4. S188 CPU Board defect (rare)
  5. 5. Application firmware corruption (rare)
First check

Power cycle system

Test point
Main power switch
Expected
System restarts and reloads firmware without error

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid addresses sent to AMB utilities

Error 6601 indicates AUP software is calculating invalid memory addresses, typically due to corrupted AUP database values or AUP memory board failures

Ranked causes

  1. 1. AUP database corruption (common)
  2. 2. AUP memory board not securely attached or low battery voltage (occasional)
  3. 3. AUP memory board failure or AMB memory corruption (occasional)
  4. 4. ARCNET communication failure (biplane systems) (occasional)
  5. 5. ASC module hardware failure (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

OCCLUSION or UPSTREAM OCCLUSION alarm with door closed patient safety

OCCLUSION or UPSTREAM OCCLUSION alarm triggered when pump door is closed and START key pressed.

Ranked causes

  1. 1. Tubing mispositioned or not seated in sensor (common)
  2. 2. Obstruction upstream or downstream of pump (common)
  3. 3. Wrong administration set type or code (occasional)
  4. 4. Ambient or solution temperature below 60°F (occasional)
  5. 5. Pressure sensor contaminated with residue (common)
First check

Verify tubing position and seating

Test point
Tubing in sensor and safety clamp
Expected
Tubing properly positioned and seated

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Photospot camera not loaded properly caution

Console displays 'Photospot camera not loaded properly. Reload to continue.' In-room display shows 'FILM NOT LOADED PROPERLY.' Camera unable to transport film.

Ranked causes

  1. 1. Film path loops incorrect or film pan misaligned (common)
  2. 2. Camera transport motor failure (occasional)
  3. 3. Camera ready signal line fault (occasional)
  4. 4. PHOTOSPOT module PRD test failure (rare)
  5. 5. System power or reset issue (occasional)
First check

Visually inspect film path and pan alignment

Test point
Film path, film pan, drive sprocket
Expected
Loops proper, film pan in correct position, sprocket dot close to film pan dot

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Unit communication error high voltage

Console displays 'Power Unit communication error. Try again or cycle power.' In-room display shows 'TRY AGAIN OR CYCLE POWER.' LFX type power units only.

Ranked causes

  1. 1. LFX I/O or Bias Inverter board faulty (common)
  2. 2. Bias Interrupt circuit failure (common)
  3. 3. Bad Bias Inverter Power Supply (occasional)
  4. 4. Bad Bias Inverter board (occasional)
  5. 5. Cable from LFX to bias tank faulty (occasional)
First check

Run Bias Interrupt diagnostic

Test point
LFX CPU
Expected
Bias Interrupt test passes

8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power unit inverter 1 failure high voltage

Console displays 'Power unit inverter 1 failure. Call service'. In-room display shows 'POWER UNIT ERROR. CALL SERVICE.' Only reported by MPPU and VAMP type power units.

Ranked causes

  1. 1. Fuses F11 or F12 blown on power supply (common)
  2. 2. Flat cable between Control board and Inverter 1 disconnected or damaged (occasional)
  3. 3. Inverter 1 failure (occasional)
  4. 4. Security board failure (rare)
  5. 5. Control board failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power unit KV balance fault detected high voltage

X-ray system power unit has detected an imbalance in high voltage (kV) generation or regulation

Ranked causes

  1. 1. Newly formed X-ray tube not yet stabilized (common)
  2. 2. Excessive kV or mA demand beyond current tube capability (common)
  3. 3. Aged X-ray tube failing to regulate kV properly (occasional)
  4. 4. High voltage tank failure or regulation problem (occasional)
  5. 5. Measurement or power switching board failure (occasional)
First check

Check tube age and formation status in system log

Expected
If new tube without formation, error is expected

4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Rear door lock fails to lock or unlock caution

Error 4B34 displayed on console and in-room display. System reports rear door lock solenoid signal failure after repeated lock/unlock attempts.

Ranked causes

  1. 1. Rear door not closed properly (common)
  2. 2. Rear door lock solenoid malfunction (occasional)
  3. 3. Fuse open on RF interface board (occasional)
  4. 4. Rear door lock not receiving voltage (occasional)
  5. 5. R+F Positioner module firmware or CPU error (rare)
First check

Verify rear door is fully closed

Test point
Rear door
Expected
Door fully and securely closed

9 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Supply magazine not ready

Console displays 'Reload or insert supply magazine to continue', in-room display shows 'CHECK SUPPLY MAGAZINE'. Camera indicates supply magazine needs reloading.

Ranked causes

  1. 1. Supply magazine empty or not in place (common)
  2. 2. CAMERA RDY signal failure (occasional)
  3. 3. FILM REMAINING STATUS signal incorrect (occasional)
  4. 4. Photospot module fault (rare)
  5. 5. Camera hardware failure (rare)
First check

Verify supply magazine is filled and installed

Test point
Supply magazine
Expected
Magazine full and seated in camera

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table angulation non-functional

Operator requests table angulation but no motion occurs. Console displays 'Table Angulation non-functional. Call Service.' In-room display shows 'ANGULATION NON-FUNCTIONAL.'

Ranked causes

  1. 1. Table position feedback pot not indicating movement (common)
  2. 2. Angulation and Table Top I/F board malfunction (occasional)
  3. 3. External angulation deadman enable switch preventing motion (occasional)
  4. 4. R+F S188 CPU Board timer fault (rare)
  5. 5. R+F Positioner module hardware failure (rare)
First check

Run Table Top Angulation Test diagnostic

Test point
R+F Interface Board entry in R+F menu
Expected
Test completes without error

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table at angulation limit patient safety

Console displays 'Table at angulation limit' error; In-Room Display shows 'ANGULATION LIMIT'. Occurs when angulating the table.

Ranked causes

  1. 1. Foot step not fully retracted (common)
  2. 2. Angulation limit switch malfunction (occasional)
  3. 3. Servo controller interlock signal fault (occasional)
  4. 4. Software/firmware fault (rare)
  5. 5. R+F Positioner module fault (rare)
First check

Verify foot step position

Test point
Table foot step
Expected
Fully retracted/inserted

7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table locks non-functional caution

Console displays 'Table locks non-functional. Call Service'. In-room display shows 'LOCKS NON-FUNCTIONAL'. Lock did not respond as expected: either failed to set when requested, or set when not requested. Determined by comparing expected state to actual state reported by Power Assist board. Error reported after timeout with lock in wrong state.

Ranked causes

  1. 1. Lock mechanism not operating correctly (common)
  2. 2. Power Assist board timer error (occasional)
  3. 3. Power Assist board failure (occasional)
  4. 4. Cable connection issue (occasional)
  5. 5. Lock mechanism broken (rare)
First check

Run Lock Test from diagnostics

Test point
R+F Power Assist/Lock Control menu
Expected
Test completes without error

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AIR alarm occurs with no air in tubing or door closed patient safety

AIR alarm triggered when pump door is closed and START key pressed, or when there is no visible air in the tubing.

Ranked causes

  1. 1. Tubing mispositioned in air sensor (common)
  2. 2. Tubing surface scratches or deformation (common)
  3. 3. Wrong administration set type or code (occasional)
  4. 4. Air sensor contaminated with residue (common)
First check

Check tubing position in air sensor

Test point
Air sensor - tubing insertion
Expected
Tubing fully seated in air sensor

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Camera rotator stalled

Console displays 'WARNING, Camera rotator stalled. Please try again.' TV Rotator failed to track command.

Ranked causes

  1. 1. Mechanical jam in the head (common)
  2. 2. TV Rotator Board failure (occasional)
  3. 3. Iris and TV Rotator Interface failure (occasional)
  4. 4. Rotator Motor failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cine film jammed in camera

Camera stalled; console displays 'Cine film jammed in camera. Reload magazine.' In-room display shows 'CINE FILM JAMMED. RELOAD.' Camera fails to reach 98% of expected speed within timeout period.

Ranked causes

  1. 1. Film jammed in transport mechanism (common)
  2. 2. Camera RPM limiter switch tripped (occasional)
  3. 3. 80188 timer or PIC circuit fault (occasional)
  4. 4. Camera tach or shutter circuit fault (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 420E: Auto Brightness Control problem high voltage

ABC Algorithm attempted to reload pulse width timer during ongoing exposure

Ranked causes

  1. 1. ECM timer configuration allows ABC update timeout during exposure (unknown)
  2. 2. EXT SYNC or EXP GATE signal module problem (unknown)
  3. 3. ECM servicing higher priority interrupts delays ABC interrupt (unknown)
  4. 4. Hardware or Real Time Bus timing problem (unknown)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 4247 displayed

ECM and power unit communication error or power unit sending unexpected error value

Ranked causes

  1. 1. ECM or power unit requires reinitialization (common)
  2. 2. Incorrect DIP switch settings on ECM or power unit (occasional)
  3. 3. Serial data link communication fault (occasional)
  4. 4. ECM or power unit module failure (rare)
First check

Power cycle the system

Test point
Console power
Expected
System reinitializes ECM and power unit

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 42CB displayed

Console shows 'Instruction not understood. Please try again.' In-room display shows 'RELEASE SWITCH, TRY AGAIN.'

Ranked causes

  1. 1. LFX/MPPU/VAMP power unit reporting error (common)
  2. 2. SCPU power unit with VRTX error detected (occasional)
  3. 3. Floating point calculation error in SCPU (occasional)
  4. 4. Corrupted SCPU error code table (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 4634 - ARCNET transmission failed

Console displays 'Instruction not understood. Please try again'; in-room display shows 'RELEASE SWITCH, TRY AGAIN'

Ranked causes

  1. 1. Invalid destination node ID (common)
  2. 2. CINE module CPU RAM error (occasional)
  3. 3. System power supply unstable (occasional)
  4. 4. Corrupted application software (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exposure time below minimum radiation

Console displays 'Min. expose time attempted. Select new technique to continue.' In-room display shows 'MIN. EXPOSE TIME ATTEMPTED.'

Ranked causes

  1. 1. AEC LUT time plus tail time minus anticipation time results in negative value (common)
  2. 2. High technique with very little filtration causing excessively short exposure time (occasional)
  3. 3. AEC LUT is defective (occasional)
  4. 4. AEC LUT addressing hardware is defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

FastBoot auto return failed, default parameters restored

Console displays 'Parameters not as previously selected. Check console settings.' Mode and parameters differ from previous settings after reboot.

Ranked causes

  1. 1. NVRAM failed (occasional)
  2. 2. System halted during runtime snapshots (occasional)
  3. 3. LAT plane snapshot mismatch in biplane systems (rare)
  4. 4. LAT plane brought online after receptor change (rare)
First check

Verify console displays correct application mode and parameters

Test point
Console screen
Expected
Application mode and parameters match previously used settings

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament correction value increasing rapidly (Error 4204H)

System continuously increases filament correction value in large steps; indicates calibration or feedback problem; mA meter saturation or tube spit errors may accompany

Ranked causes

  1. 1. Filament control parameters not properly calibrated (common)
  2. 2. mA meter malfunction (ABC modes) (occasional)
  3. 3. mA integrator fault (rad modes) (occasional)
  4. 4. Incorrect mA feedback from power unit (occasional)
First check

Review error log for related errors

Test point
System error log
Expected
Check for 42EAH (mA meter saturation) or 42CAH (tube spit) errors occurring when 4204H observed

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Film changer programmer not ready

Console displays 'Film Changer Programmer not ready'; in-room display shows 'CHECK FC PROGRAMMER'. Occurs at prep time when film changer signals indicate film magazine or receiver out of device or software cannot determine film position.

Ranked causes

  1. 1. Film magazine or receiver out of device (common)
  2. 2. FC RDY signal indicates not ready or internal flag shows film not in place (common)
  3. 3. Failure in FC RDY signal path from programmer to I/O board (occasional)
  4. 4. Positioner module failure or misconfiguration (rare)
First check

Verify film changer programmer indicates ready status

Test point
SEP film changer programmer display
Expected
No error or fault indication

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluoroscopy not operable patient safety

Console displays 'Cannot do Fluoro. Exposure control inoperable. Cycle Console.' In-room display shows 'FLUORO NOT OPERABLE.'

Ranked causes

  1. 1. ECM not configured online during room setup (common)
  2. 2. No reply from ECM to load command (occasional)
  3. 3. No reply from ECM to reset fluoro timer (occasional)
  4. 4. No reply from ECM to prep sequence enable command (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Node response timeout during test sequence

Console displays timeout error; test aborted; GE bus communication failure with system node

Ranked causes

  1. 1. ASC BIU communication failure (common)
  2. 2. Node not responding to code download request (common)
  3. 3. Node hardware malfunction (occasional)
  4. 4. BIU board defect on receiving node (occasional)
First check

Verify ASC can communicate with system nodes

Test point
ASC BIU to any available node
Expected
Communication established without timeout

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Obstruction detected at front patient safety

Console displays 'Obstruction detected, clear using Front/Rear load buttons.' In-room display shows 'OBSTRUCTION DETECTED.' System enters operator-initiated recovery mode.

Ranked causes

  1. 1. Front sensor beam (IN FRONT CASS signal) falsely broken or cassette already in system (common)
  2. 2. IN FRONT CASS signal indicating falsely (occasional)
  3. 3. Front sensor adjustment out of tolerance (occasional)
  4. 4. R+F Positioner module or RAM/Arcnet failure (rare)
First check

Perform automatic cassette unload

Test point
Front load button
Expected
Cassette unloads via front, recovery completes

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Positioner software initialization error

Error 4A6A displayed on console and in-room display; console message instructs to cycle power or call service

Ranked causes

  1. 1. Positioner to LP controls link failure (common)
  2. 2. Device I/O board malfunction (occasional)
  3. 3. Positioner module RAM or Arcnet failure (occasional)
  4. 4. LP control system failure (occasional)
First check

Verify link status between Positioner and LP controls

Test point
Physical connection Positioner-LP
Expected
Connected and secure

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power plug not lit or battery icon lit when plugged in caution

Plug icon fails to illuminate when pump plugged in, or battery icon incorrectly illuminates with AC power connected.

Ranked causes

  1. 1. Power plug not fully inserted into AC outlet (common)
  2. 2. Rear power fuse failed (common)
  3. 3. AC outlet voltage incorrect or below specification (occasional)
  4. 4. Power cord discontinuity (occasional)
First check

Verify power plug connection

Test point
AC power plug in grounded outlet
Expected
Plug fully seated in outlet

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power unit tube select error radiation

Power unit cannot select correct X-ray tube; system displays error and requires power cycle

Ranked causes

  1. 1. Power unit transient fault (occasional)
  2. 2. Power unit hardware or communication fault (occasional)
  3. 3. TIRC circuit malfunction (occasional)
  4. 4. Interface 1 board fault (occasional)
First check

Power cycle the system

Expected
System restarts successfully

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump door will not open or close smoothly caution

Door movement is stiff or binding when attempting to open or close.

Ranked causes

  1. 1. Tubing or slide clamp mispositioned or not seated (common)
  2. 2. Wrong administration set type or code (occasional)
  3. 3. Solution spills, liquids or residues present (common)
  4. 4. Door latch roller pin sticking or damaged (occasional)
First check

Verify tubing and slide clamp positioning

Test point
Tubing guides and slide clamp slot
Expected
Tubing and clamp properly seated in guides

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table top in illegal region patient safety

Table top is extended in angulated position when it should be retracted, or extended beyond calibrated boundary. Error message on console: 'Press OFF button to stop table motion. Call Service.' In-room display: 'PRESS EMERGENCY OFF BUTTON.'

Ranked causes

  1. 1. Table feedback pot reading outside calibrated positions (common)
  2. 2. Angulation switches wired backwards (occasional)
  3. 3. Table motors not stopping fast enough (occasional)
  4. 4. Calibration data corrupted (rare)
First check

Run Table Top Lat/Long Motion Test

Test point
R+F Interface Board in R+F menu
Expected
Test passes without error

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table top motion non-functional patient safety

Console displays 'Table top motion non-functional. Call Service.' In-room display shows 'TABLE TOP NON-FUNCTIONAL.' Table top longitudinal motion requested but no motion detected after timeout. Position feedback pot shows no movement.

Ranked causes

  1. 1. Table top longitudinal feedback pot failure or path failure (common)
  2. 2. Angulation and Table Top I/F board failure (occasional)
  3. 3. R+F S188 CPU board timer failure (occasional)
  4. 4. R+F Positioner module RAM/Arcnet failure (occasional)
First check

Run Table Top Lat/Long Motion Test

Test point
R+F Interface Board, R+F menu
Expected
Test completes without error

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Video aperture did not respond

Console displays 'Video aperture did not respond. Try again or call service.' In-room display shows 'APERTURE DID NOT RESPOND.' Iris electronics failed to generate interrupt within 1 second timeout.

Ranked causes

  1. 1. Iris electronics database parameter ap_pos_cal not set (common)
  2. 2. XRII & Iris Control Board failure (occasional)
  3. 3. Iris and TV Rotator Interface failure (occasional)
  4. 4. Iris assembly failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC exposure time exceeds four seconds

After AEC exposure terminates, LUT address exceeds 1299 (4 sec limit), error 4267 reported, LUT defaults to 1299

Ranked causes

  1. 1. Backup timer loaded with value less than four seconds but not functioning (common)
  2. 2. Backup timer hardware fault on ETL1 board (occasional)
  3. 3. AEC hardware fault on ETL2 board (occasional)
First check

Verify backup timer is loaded with value less than four seconds

Test point
ETL1 board backup timer setting
Expected
value less than 4 seconds

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auto-positioner control problem

Serial link interface between Vascular positioner and tableside control has failed. Console displays 'Auto-positioner control problem. Use manual positioning or try again.' In-room display shows 'AUTO-POSITIONER PROBLEM.'

Ranked causes

  1. 1. Serial link interface failure due to loose connections, noise, or power supply issues (unknown)
  2. 2. CPU board or device I/O board failure (unknown)
  3. 3. Tableside auto-positioner board failure (unknown)
First check

Check cabling for loose connections

Test point
Serial link connectors
Expected
Connections secure and tight

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Autoprotocols unavailable at power-up

Console displays 'Autoprotocols unavailable. Continue and call service.' PROCEDURE MENU button not displayed. AUP software removes itself from active applications.

Ranked causes

  1. 1. AUP memory board not securely attached or battery voltage low (common)
  2. 2. AUP memory board defective or inaccessible (common)
  3. 3. ASC module or system-level hardware failure (occasional)
First check

Reboot system to clear transient errors

Test point
System startup
Expected
Error clears and PROCEDURE MENU button appears

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad data received from power unit high voltage

Console displays 'Bad data received. Turn console power OFF/ON.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' ECM resets power unit.

Ranked causes

  1. 1. DMA controller malfunction (occasional)
  2. 2. Software configuration problem (occasional)
  3. 3. Associated hardware failures (linked errors) (rare)
First check

Verify DMA controller operation

Test point
DMA controller
Expected
Packet length >= 5 bytes

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad data received from XRII interface radiation

Console displays 'Bad data received. Turn console power OFF/ON.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' System halts X-ray operation.

Ranked causes

  1. 1. DUART framing error on XRII & Iris Control Board channel A (common)
  2. 2. XRII I/F Board failure (occasional)
  3. 3. XRII & Iris Control Board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette not detected in tray radiation

Console error message 'Load cassette in tray to continue' appears when attempting RAD exposure on selected receptor

Ranked causes

  1. 1. Cassette not loaded in tray (common)
  2. 2. Tray not fully inserted (common)
  3. 3. Cassette-in-place signal not reaching positioner (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cine camera not line locked

Console warning message displayed; hardware phase lock loop cannot lock to incoming signal during camera startup; LOCK signal on port 0 not asserted when stall timer expires.

Ranked causes

  1. 1. LOCK signal not asserted from power amp board (common)
  2. 2. Incoming signals to phase comparator not in phase (common)
  3. 3. CINE CPU board defective (occasional)
First check

Run diagnostics on CINE CPU board

Test point
CINE CPU Board menu
Expected
8253 Diagnostic Timer, 8253 Timers 0/1/2, 80188 PIC, Phase Locked Loop tests pass

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cine camera out of film

Console displays 'Cine camera out of film. Reload magazine.' and in-room display shows 'CAMERA OUT OF FILM.' System detects INTLOK bit as off on port 0 and film remaining A/D value is below calibrated no-film threshold.

Ranked causes

  1. 1. Magazine empty or film depleted (common)
  2. 2. Film sensor or motor malfunction (occasional)
  3. 3. CINE module circuitry fault (rare)
First check

Verify film presence in magazine

Test point
Cine camera magazine
Expected
Film loaded in magazine

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cine camera overspeed or jam fault

Console displays 'Reset Cine camera overspeed switch at camera or check for film jam'. In-room display shows 'CINE CAMERA FAULT'. Camera will not run.

Ranked causes

  1. 1. Film not loaded correctly (common)
  2. 2. Camera RPM limiter switch tripped (common)
  3. 3. INTLOK signal failure or defective camera (occasional)
First check

Verify film is loaded correctly in camera

Test point
Camera film compartment
Expected
Film properly seated and threaded

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Database error in ECM with invalid cal/config data high voltage

ECM received calibration or configuration data out of range. ECM CPU LEDs toggle between 00FH and 0F0H in endless loop until cold reset. Module not marked online; not ready signal not sent to ASC.

Ranked causes

  1. 1. Invalid go-to-cold time parameter in configuration data (occasional)
  2. 2. Other out-of-range parameters in calibration/config data (occasional)
  3. 3. Multiple invalid parameters detected by HW status field (rare)
First check

Observe ECM CPU board LED state

Test point
ECM CPU board LEDs
Expected
LEDs toggling between 00FH and 0F0H hex

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DF module not ready for acquisitions

Console displays 'Enter DF acquisition ready screen to continue'; in-room display shows 'SET DF CONSOLE FOR EXPOSURE'

Ranked causes

  1. 1. DSMON not entered at DF keyboard (common)
  2. 2. DSMON in REPRO or XFER mode instead of acquisition mode (common)
  3. 3. Console power cycle required (occasional)
First check

Verify DSMON has been entered at DF keyboard

Test point
DF keyboard display
Expected
DSMON mode active

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 4246

ECM and power unit communications error; system displays error code and may not operate normally

Ranked causes

  1. 1. ECM or power unit requires re-initialization or DIP switch misconfiguration (common)
  2. 2. ECM or power unit module failure detected by PRD diagnostics (occasional)
  3. 3. System level failure in ECM or power unit (rare)
First check

Cycle console power to re-initialize

Test point
Console power switch
Expected
System fully reboots and initializes

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 4264 displayed on console and in-room caution

Console shows 'Press CPU RESET at control cabinet. System requires reboot.' In-room display shows 'PRESS CPU RESET.' ECM cannot bring power unit online.

Ranked causes

  1. 1. Power unit did not set RESET DONE bit within ten seconds (common)
  2. 2. Power unit CPU board failure (occasional)
  3. 3. ECM power unit interface problem (occasional)
First check

Inspect power unit CPU board LED indicators

Test point
Power unit CPU board LEDs
Expected
No PRD error indicated on LEDs

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 42B9: Power unit tube select error radiation

Error message displayed when tube selection fails during power unit operation. Console shows 'Power unit tube select error. Cycle power, call service & continue.' In-room display shows 'TRY AGAIN OR CYCLE POWER.'

Ranked causes

  1. 1. Compressed pins in high voltage connectors (common)
  2. 2. HV tank failure (low filament current with kV present) (occasional)
  3. 3. HV switch or measurement commutation board failure (occasional)
First check

Cycle power to console

Test point
Power switch
Expected
Error clears on restart

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 4A09 Bad data received

Console displays 'Bad data received. Turn console power OFF/ON.' with in-room display showing 'TURN CONSOLE POWER OFF/ON.'

Ranked causes

  1. 1. Software task scheduler overflow; no more room in timer list (common)
  2. 2. Positioner module task scheduler or memory issue (occasional)
  3. 3. Corrupted Applications software on floppy (occasional)
First check

Power cycle system

Test point
System power and reset
Expected
Console and in-room displays boot normally; no error messages

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Error 4B33 displayed on console
  • Error 4B3D diagnostic procedure
  • Error 6602 occurs at warm start
  • Exposure run terminated by STARTWHEN timeout
  • Exposure time calculated as negative
  • Film supply depleted or camera not ready
  • INSERT SLIDECLAMP alert or alarm when door closed
  • kVp control error
  • Magazine lever not closed
  • Obstruction prevents table angulation
  • Power unit tube spit recovery
  • Rear door open interlock
  • Receptor ready interrupt lost during X-ray exposure
  • Safety clamp will not latch open
  • STARTWHEN timeout, overlap detector
  • Supply magazine empty indication
  • System timeout during exposure startup
  • Table not positioned at horizontal
  • Tube anode not rotating
  • Tube spit during exposure
  • Tube spit recovery error during exposure
  • Vascular positioner serial link communication failure
  • X-ray tube and table receptor misalignment
  • Camera A/D converter timeout
  • CHECK SET/LOADING alarm when door closed
  • Cine magazine missing
  • Circle adjust failure during calibration
  • Collimator non-functional
  • Digital Record inoperative in fluoroscopy mode
  • Error 420A
  • Error 421A: Power unit inoperative
  • Error 4245 displayed
  • Error 4274 displayed
  • Error 4277 occurs during power unit operation
  • Error 42C5
  • Error 4500 displayed
  • Error 4903 displayed during RAD exposure
  • Error 4B57: Remove all cassettes to continue
  • Error 6405: Exposure control inoperable
  • Error 6707 displayed
  • Filament control adjusting message displayed
  • Filament drive error
  • Filament drive error with auto-reset in progress
  • Film changer not correctly locked
  • Film changer rotation interlock persists after repositioning
  • Floppy diskette read error
  • Image Intensifier not at required SID position
  • Improper exposure error 4214
  • Jog not allowed during Cine recording
  • Light bias inoperative
  • Maximum exposure time reached
  • Maximum tube housing heat units reached
  • Obstruction in fluoro image
  • Optical filter not in position
  • Procedure database version incompatible
  • Procedure menu blocked during auto-protocol mode
  • RAD exposure blocked by room door
  • RATE, VTBI, or START key not accepted
  • Selected filament open
  • Serial link communication failure
  • Servo axis not maintaining position
  • SID error
  • System error persists after diagnostics
  • Table height error during RAD exposure
  • Table motion disabled with film changer engaged
  • TV video signal out of range
  • Unrecognized command error
  • VTR not in remote mode
  • X-ray tube overheating
  • ABC brightness instability
  • ASC sends invalid video gain to ECM
  • Audible alarm volume is not loud enough
  • Backlight is off during internal battery operation
  • Bias fault
  • Bias fault detected
  • Bias fault indicated
  • Bias inverter supply too low
  • Biplane system communication failure
  • Camera still hot error on startup
  • Cannot take exposures with LP parked
  • Cassette carriage jammed
  • Cassette not detected in Spot Film Device
  • Cassette not positioning
  • Cine camera film running low
  • Cine mode unavailable
  • Collimator blades not closed during tube warmup
  • Cone button fails to operate
  • Console displays bad data error
  • Console power cycle required
  • Console receives invalid video format command
  • Database Error: Incompatible Database Version
  • DF service mode exposure zero milliseconds error
  • Diagnostic test failed on startup
  • Digital error - service required
  • Disk space exhausted during acquisition
  • Disk space remaining less than 15 frames
  • Duty cycle limit exceeded
  • DX module not ready for acquisitions
  • E4156 Filament drive error
  • ECM loses power to CPU
  • ECM module not coming online after error
  • Error 4202 power unit reset for unknown reason
  • Error 4217: No exposure, power unit inoperative
  • Error 4245 with POWER UNIT issues
  • Error 4260 displayed
  • Error 4264 persists after initial troubleshooting
  • Error 4266 displayed on console and in-room display
  • Error 4273 displayed
  • Error 4282 detection and recipe abort
  • Error 4297: Start when timeout, boost done
  • Error 4299 displayed on console
  • Error 429D displayed
  • Error 42C9: Filament drive error detected by power unit
  • Error 42F9 detected
  • Error 4405 Digital system maximum run length reached
  • Error 4706 divide by zero fault
  • Error 4913 diagnostic procedure
  • Error 4A59 film changer position
  • Error 4A5A diagnostic procedure
  • Error 4A5F - Positioner module division by zero
  • Error 4A81
  • Error 4B19 displayed
  • Error 4B22 displayed
  • Error 4B2C
  • Error 4B33 persists after initial troubleshooting
  • Error 4B34 displayed
  • Error 4B3F displayed
  • Error 4C04 diagnostic procedure
  • Error 4C05 displayed
  • Error 4C08 verification
  • Error 6417: Bad data received from Digital Record
  • Error 6600 displayed
  • Error 6601 occurs on lateral biplane system
  • Error 6602 during SAVE/RETRIEVE operation
  • Error code 42A1 detected
  • Error code 4A02 displayed
  • Error code 4A8E
  • Error code 4B36 diagnostic procedure
  • Error code 4C11 triggered
  • Exit sequence not executed
  • Exposure run terminated with timeout
  • Exposure start signal timeout (Error 4282)
  • Filament command failure with auto-reset
  • Filament warmup delay message displayed
  • Film changer rotated unexpectedly
  • Floppy disk error
  • FLOW RATE or CHECK VTBI alarm occurs on START
  • Frontal trial image too bright, lateral trial image too dark
  • HI or LO displayed during Volume-Time programming
  • II on LP collided with object
  • Image Hold Fluoro mode only available
  • Injector not ready on prep switch depression
  • Insufficient disk space for cardiac exposure
  • Insufficient disk space for image acquisition
  • Interval between audible alarm tones too long
  • Last exposure may be bad
  • mA exceeds 125% of maximum (Error 420AH)
  • mA meter saturation (Error 42EAH)
  • No exposure, power unit inoperative
  • Object blocking front load port sensor
  • Object in front load port
  • Photo cathode current read during AGC
  • Photospot camera inoperable
  • Photospot prep done timeout during exposure
  • Positioner module task scheduling failure
  • Positioner position not selected within time limit
  • Positioners not orthogonal
  • Power unit control not operable
  • Power unit fault with high voltage unavailable
  • Power unit recovery sequence initiated
  • Power unit rotor control error
  • Pump fails to run on internal battery
  • Pump stops with BATTERY LOW alarm
  • Rear load door forced open
  • Rotor braking sequence not completing
  • Secondary errors during power unit recovery
  • SID zero value calibration drift or error
  • Signal transmission error to ECM
  • Spotfilmer locked in forward position
  • Spotfilmer servo non-functional
  • Standby cooling not performed due to invalid time parameter
  • Stepper selected at table but not at console
  • System aborts with Error 4011
  • System communication problem
  • System communications problem between ASC and DR module
  • Table motion non-functional
  • Table not ready during exposure prep
  • Table top not centered laterally
  • Table top position data incorrect
  • Tomo procedure cannot start
  • Tomo table not properly prepared for acquisitions
  • Tube casing water flow not detected
  • Tube spit error (Error 42CAH)
  • TV gain curve data bad
  • Unable to establish tube selection during operation
  • Video camera system inoperative
  • Video setup corrupted
  • Video shading correction fault
  • VTR communication error
  • VTR left idle in ready state over 30 minutes
  • Watchdog timer timeout (error indeterminate)

201 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
30Analog output different from expected value; possible shorted/over-loaded analog output or AD/DA converter failureWatchdog detects analog output voltage out of range or converter circuit malfunctionerror
4000System error - RMX error detectedAPOS code running in ASC module detects RMX error conditionerror
4000-40FFAPOS (Applications Operating System) error rangeError encountered during Applications mode operationunknown
4001System error - error code not found in error tableAPOS code running in ASC module encounters unknown conditionerror
4011System error: no free buffers in APOS running in ASC moduleTOMO function selected on Advantx Console prior to turning on Ordograph Control, especially with long interconnect cable (80 feet or longer)critical
4012System error - no free mailboxesAPOS code running in ASC module detects mailbox allocation failurecritical
4013APOS kernel semaphore exhaustionOperating system runs out of available semaphoreserror
4020System communications problemLoss of communication with nodeerror
4022System communications problem in APOS running on ASC moduleASC module node fails to respond within timeout perioderror
4023System communications corruptedARCNET initialization failurecritical
4024Wrong node address errorASC module detects wrong node address configurationerror
4025System communications problemAPOS code in ASC moduleerror
4027System communication problemApplications Operating System (APOS) code running in the ASC modulecritical
4030System communications problemAPOS code running in ASC module generates errorerror
4031System communications problemAPOS code running in ASC module detects bad data type in descriptor blockerror
4032System communications problem; APOS stack/data structure errorNo data pointer present in ASC module during operationcritical
4033System communications problemASC module communication failureerror
4035System communications problemUnknownerror
4036System communications problemAPOS code running in ASC moduleerror
4037System communications problemAPOS (Applications Operating System) code in ASC module detects communication issue or invalid character in input fielderror
4038System communications problemAPOS code in ASC moduleerror
4039System communications problemApplications Operating System (APOS) code running in ASC module detected bad data type in MSG$DECODE functionerror
4040System communications problemASC module call table index errorerror
4041System communications problemModule code filename not found in filename tableerror
4042System communications problemSystem communications error detected in ASC moduleerror
4043Text input field size exceededApplications Operating System (APOS) receives text input exceeding field size limiterror
4044APOS stack limit exceeded in ASC moduleApplications Operating System code running in ASC module generates stack overflowcritical
4050System error message queue is fullAPOS applications operating system error in ASC moduleerror
4060System error: no default task address in tableAPOS operating system unable to locate default task address during initializationerror
4061Bad default task number in APOSAPOS code error during startup or operationcritical
4065ASC Dynamic RAM Memory Parity Error; problem with ASC main memorySystem automatic self-check detecting memory parity failureerror
4067System power down event marker in Event LogSystem was powered downinfo
406FSystem requires rebootCPU error conditionerror
4070System communications problemSystem error during operationerror
4073System communications problem in APOS (Applications Operating System) running on ASC moduleMailbox released with data unexpectedly during system operationerror
4074System communications problemAPOS code running in ASC module generates errorerror
4075System error - GE bus reconfiguration occurredAPOS code in ASC module detects bus reconfigurationerror
4076unknown
4077System error in Applications Operating System (APOS) moduleBad translate table detected in ASC moduleerror
4078Other ASC did not respond to biplane communication requestLateral plane did not respond to Operating Mode Request commanderror
4082unknown
4083unknown
4084unknown
4086Database is unusable; restore requiredApplication detects corrupted or incompatible database at startuperror
4090unknown
4091unknown
4092System error with module and description not fully documentedUnknown; messages and recommended actions not yet availableunknown
4093unknown
40A0System communications errorASC software attempts to reallocate a software timer that is already in use; occurs when bouncing on and off prep and fluoro switcheserror
40FFSystem communications problemAPOS code in ASC module detected communication failurecritical
4111Both frontal and lateral trial images are too brightBiplane trial image acquisition with excessive brightness on both planeswarning
4112Frontal trial too bright, lateral trial too darkBiplane trial image quality assessment fails due to exposure mismatch between frontal and lateral projectionswarning
4113Frontal trial image is too dark and lateral trial image is too bright.Biplane trial execution with suboptimal technique values for frontal and lateral planes.warning
4114Both frontal and lateral trial images are too darkTrial images acquired with insufficient exposure or techniquewarning
4115DX store function in progress on the in-room monitorDX action initiated; takes 2 seconds to completeinfo
4116DX recall function in progress on the in-room monitorDX action recall initiated; takes approximately 2 seconds to completeinfo
4150Power Unit communication errorPower unit sends request for service and receives no response from ECM after timeout of 130serror
4151Power Unit communication errorError detected in the format of the packet received from ECM (checksum, overrun...)error
4152Instructions not understood. Please try again.Bad command from ECM concerning tube number, kV demand, mA demand or mAs demand.error
4153ECM maintains boost heater command for more than 400 msBoost heater command duration exceeds 400 millisecondserror
4154Instructions not understood. Please try again.ECM sends a rotor command while the tube is switching.error
4155Instructions not understoodConsole command or in-room control input not recognizedwarning
4156Filament drive errorFilament current sense system detects filament current above threshold, tripping latcherror
4157Filament drive error in heater systemFilament current sense system detects current above thresholderror
4158Filament drive errorFilament current sense system detects current above threshold on heater boarderror
4159Filament drive errorFilament current sense system detects filament current above threshold, trips latcherror
415AFilament drive error detected in heater board filament current sense systemFilament current exceeds threshold on heater board, latch trips and reports error to CPUerror
415BFilament drive error - overcurrent condition detected on heater board filament current senseFilament current sense system detects filament current above threshold while in acquisition modeerror
415CFilament drive error in small focus heater circuitNo current detected in small focus heater circuit with non-zero heating demanderror
415DFilament drive errorNo current detected in the small focus heater circuit with a non-zero heating demanderror
415EFilament drive error in small focus heater circuitNo current detected in small focus heater circuit with non-zero heating demand during acquisitionerror
415FFilament drive errorNo current detected in large focus heater circuit when heating demand is non-zeroerror
4160Filament drive errorNo current detected in large focus heater circuit with non-zero heating demanderror
4161Filament drive error in large focus heater circuitNo current detected in large focus heater circuit with non-zero heating demand during acquisitionerror
4162Filament drive errorOver-current in small focus inverter circuit or open circuit detected during heater pre-light stateerror
4163Filament drive errorOver-current in small focus inverter circuit or open circuit detectederror
4164Filament drive error in SCPU type power unitOver-current in small focus inverter or open circuit detected during heater acquisition stateerror
4165Filament drive errorOver-current in large focus inverter circuit or open circuit detected during heater state (pre-light)error
4166Filament drive error in X-ray tube heater circuitOver-current in large focus inverter circuit or open circuit detected during heater operationerror
4167Filament drive errorOver-current detected in large focus inverter circuit or open circuit conditionerror
4168Power unit rotor control errorCPU detects no current in rotor controller circuit while commanded onerror
4169Power unit rotor control errorCPU detects no current in rotor controller circuit while commanded onerror
416APower unit rotor control errorCPU detects no current in rotor controller circuit while commanded onerror
416BPower unit rotor control errorCPU detects no current in rotor controller circuit while commanded onerror
416CPower unit rotor control errorCPU detects no current in rotor controller circuit while commanded onerror
416DPower unit rotor control errorCPU detects no current in rotor controller circuit while commanded onerror
416EPower unit rotor control errorCPU detects overcurrent in rotor controller circuit while commanded onerror
416FPower unit rotor control errorCPU detects over current in rotor controller circuit while commanded on during High Speed Runerror
4170Bias fault detectedSignal grid_func_ok tested at bad level during mode change in secondary timed modes or when bias is selectedcritical
4171Bias FaultSignal grid_pulse_ok detected false during CINE or PULSED FLUORO initialization (EXP_CMND rise)error
4172Bias FaultGrid monitoring measures mA during grid off time (>1.1 mA fluoro or >11 mA cine >0.6 after RAD pulse)error
4173Bias Fault in VAMP type power unitNo RAD_PULSE interrupt received by CPU board within 2 s after Exp Cmderror
41B0Power unit tube spit recoverykV drop detected on anode side during tube exposurewarning
41B1Power unit tube spit recoverykV drop on anode side during tube spit eventwarning
41B2Power unit tube spit recoveryTube spit condition detected during X-ray exposure with automatic KV cutoffwarning
41B3Power unit kV regulation errorDetected during tube spit; kV automatically cut offerror
41B4Tube casing water flow not detectedChiller circuitry detects water not flowing or temperature too high with MX150 tubeerror
41C0Duty cycle limit exceeded during cine operationThree consecutive pulses shot with KV75% signal not releasederror
41C1Bias fault in VAMP type power unitBias inverter feedback signal from micro Grid/Bias tank is at an incorrect levelerror
41C2Bias inverter supply voltage exceeds acceptable rangeBias feedback voltage more than 10% above demand valueerror
41C3Bias inverter supply voltage too lowBias inverter supply measured more than 10% below demand valueerror
41D0Power unit rotor control errorCPU detects overcurrent in rotor controller circuit while commanded onerror
41D1Power unit rotor control errorCPU detects over current in rotor controller circuit while commanded onerror
41D2Power unit rotor control errorCPU detects over current in rotor controller circuit while commanded onerror
41D3Power unit rotor control errorCPU detects over current in rotor controller circuit while commanded onerror
41D4Power unit rotor control errorLow speed relay status not compatible with demanderror
41D5Power unit tube select errorSCPU receives a tube switching command while the rotor is still running (RUN, ACCELERATION or BRAKE state)error
41D6Power unit tube select errorSCPU receives tube switching command from ECM while heater on large or small focus is ONerror
41D7Power unit tube select errorSCPU receives tube switching command from ECM while kV measured exceeds 9 kVwarning
41D8Power unit tube select errorHigh voltage tube switch status is different from the tube requestederror
41D9Power unit tube select errorRotor tube switch status is different from the tube requestederror
41DAPower unit tube select errorHeater tube switch status differs from tube requestederror
41DBPower unit control not operableStatus of mA scale relays or kV/mA test relays not compatible with demand or error in mA measurement circuiterror
41DCPower unit control not operableBad connection between SCPU front panel boards detectederror
41DDFilament command failureADC result does not match DAC reference value for heater and kV commanderror
41DEPower unit control not operableSafety latch from inverter is set out of exposureerror
41DFPower unit control not operableSafety from heater on large focus function is set while heater is not activatederror
41E0Power unit control not operableSafety rotor function set while rotor is not activatederror
41E1Power unit control not operableDC BUS voltage out of range: less than 400V or greater than 762Vcritical
41E2Power unit control not operableError detected in the inverter floating power supplyerror
41E3Power unit control not operableError detected in low voltage power supply (5V, +15V, -15V or 24V)error
41E4Power unit control not operableSafety interlock activated when small focus function enabled but heater not activatedcritical
4200Software error: ECM received undefined ARCNET command outside valid range (00H–0FH or 80H–FFH) or corrupted packet from hardware failureARCNET hardware failure in ECM or sending node; undefined command sent to ECMerror
4200-4277ECM module Applications software error rangeError in ECM module during Applications modeunknown
4201Bad data received from console communicationsECM recipe pre-checker detects error with recipe during initializationerror
4202Power unit reset for unknown reasonPower unit sent reset done bit in packet without prior ECM reset commandwarning
4203Bus timeout NMI (Non-Maskable Interrupt)CPU attempts to access undefined or unreachable memory location; Multibus decode logic failure on ECM circuit boardserror
4204Filament correction value increased by 4 or more counts during debug modeHIGH word of filament correction value increased by 4+ counts since last 4204H or 4205H messageinfo
4204HFilament correction value increasing rapidly in large stepsSystem repeatedly increases filament correction value; occurs frequently with large step increaseswarning
4205Filament correction value decreased by 4 or more counts; informational message for Filament Control Algorithm monitoringHIGH word of filament correction value decreased by 4+ counts since last 4204H/4205H message, or lower limit of filament correction adjustment encounteredinfo
4205Filament correction factor adjusted to lower limit due to mA feedback error or other transient conditionmA feedback from power unit lower than actual mA, or prior mA meter saturation (42EAH) or tube spit (42CAH) errors clearedwarning
4206Instruction not understood. Please try again.Recipe processor interrupt received when ECM did not think valid recipe was executing, or ASC sent Exit/Prep/Expose/kick start before recipe downloaded successfullycritical
4207Received recipe poke command with no RIP downloadedA patch recipe command (ARCNET command 0A1H) was received from the ASC when no valid recipe resides in the ECMerror
4208Loss of receptor ready interrupt during X-ray exposureReceptor ready signal (RTB-1) goes low while X-rays are actively oncritical
4209Software error; ECM could not find EXITMARK opcode in recipe or detected DONE before EXITMARKExit Sequence Enable command (ARCNET 085H) received without valid EXITMARK opcodeerror
420AFilament control adjustingmA returned from power unit reaches 1.25 times expected mA during Fluoro exposure for at least 3 secondswarning
420AmA meter error or related power delivery faultIncorrect mA meter calibration, hardware fault in ECM or Power Unit, or related grid/tube errorserror
420AHmA exceeds 125% of maximum rated mAMeasured mA value exceeds 125% of specified maximum mA limiterror
420BmA below specification during fluoroscopy operationActual fluoroscopy mA output falls below programmed minimum thresholderror
420BFilament control limitmA feedback from power unit is less than half of calculated maximum mA while ECM expects near maximum, during ABC Algorithm operation at KV override or above 90% max pulse width for pulsed systems, or temporary grid short condition clears during exposureerror
420CFilament control threshold exceededDuring pulsed Fluoro exposures, mA feedback is more than twice expected value, indicating possible grid short not detected by power unit, or mA metering circuit problemerror
420DMinimum kV temporarily limited for thermal protectionGenerator detects thermal overload condition approaching limitswarning
420ESuspected Auto Brightness Control problemABC Algorithm attempted to reload pulse width timer during active exposurewarning
4210No brightness samples availableASC requested brightness samples before buffer completely full or ECM not receiving brightness ready interruptserror
4213Power Unit communication errorAll 20 ECM power unit packet buffers full with valid packet information; packets queued faster than sent over SDLC linkerror
4214Improper exposure - power unit HV ON timeout errorPower unit fails to set HV ON within 50-100 milliseconds of exposure command from ECMcritical
4215Power unit did not set HV ON in response to EXP CMND from ECM within 50-100 msecPower unit never received EXP ENBL signal or ECM's EXP ENBL signal is low when EXP CMND is seterror
4216Power unit did not set HV ON in response to exposure command from ECM within 50-100 msec due to SCR ON or precontacting faultPower unit fails to respond to EXP CMND from ECM within 50-100 msecerror
4217Power unit failed to respond to exposure commandPower unit did not set HV ON within 50-100 msec of exposure command from ECMcritical
4218Power unit inoperative - safety contactor did not closePower unit does not set HV ON within 50-100 msec of exposure command because safety contactor failed to closecritical
4219Power unit did not set HV ON in response to EXP CMND from ECMPower unit failed to respond to exposure command within 50–100 msecerror
421APower unit inoperative; no exposure possibleCharger SCR fails to turn on within 50-100 msec of exposure commandcritical
421BNo exposure. Power unit inoperative.Power unit did not set HV ON in response to EXP CMND from ECM within 50-100 msec because commutating capacitors never chargedcritical
421CPower unit inoperative; ECM did not receive feedback #1 from power unit after watchdog timeout requestECM requests watchdog timeout cause from power unit but next packet lacks feedback #1; power unit may die between error detection and ECM information requestcritical
421FSuspected Auto Brightness Control problemABC configuration data sent by ASC was incorrecterror
4220Maximum tube housing heat units reachedTube housing thermal load reaches maximum limitwarning
4221X-ray tube heat wait condition; ECM detected that X-ray tube requires cooling before continued operationX-ray tube heat level exceeds safe operating threshold; ECM detects heat wait condition and notifies ASC via ARCNET command 35Hwarning
4222Bad data received; ECM received invalid database identifier in cal/config packetARCNET command 0CH received with database number other than 2 or 3error
4222ECM or ASC module software errorSystem error detected during operation or initializationerror
4223Bad data received from software commandECM received cal/config command (ARCNET 00CH) with invalid part numbererror
4224Bad data received from ECM packetECM receives cal/config packet with invalid number of packets parametererror
4225Bad data received from ECM cal/config packetECM received ARCNET command 0CH with incorrect data byte count for part numbererror
4226Parameter out of range error in ASC or ECM moduleSoftware reload or database inconsistency detected during system initializationerror
4227Power unit inverter thermal wait condition detectedECM detects inverter heat wait condition or predictive thermal threshold exceededwarning
4228Bad data received from ECMECM received a cal/config packet with invalid packet number parametererror
4229Tube bulk temperature exceeds 1600 degrees C or database error in cooling table indexBulk temperature sent with non-volatile run-time parameters is too large; ECM calculates index >199 when dividing bulk temp by 8error
422ASelected focal spot out of rangeASC sends ARCNET command 0B1H with focal spot value outside valid range 0-5error
422BSoftware error: ECM did not receive all part numbers from ASC during calibration/config data transmissionECM thinks ASC finished sending cal/config data but one or more part numbers were not receivederror
422CSoftware error in automatic video gain control routineASC sends Fluoro detail value greater than two to ARCNET command 0B4Herror
422DPower unit in reset state – prep/expose switch ignoredECM resets power unit due to prior error; prep/exposure switch transitions occur during resetwarning
4230Instruction not understood. Please try again.ASC tried to download a new recipe when ECM thought it was still executing current recipeerror
4231Recipe downloaded from ASC contains odd number of bytesRecipe download from ASC with incorrect byte counterror
4244Bad data received from ASC; X-ray tube distance downloaded is less than SID zeroECM receives x-ray tube distance (command 0A0H) less than SID zero parameterwarning
4244SID zero calibration error or sensor driftSID (source-to-image distance) sensor or calibration data out of specificationerror
4245Communication failure between ECM and POWER UNITPacket transmission or clock signal loss between moduleserror
4245Power Unit communication errorECM did not receive a reply packet from the power unit within one second of transmissionerror
4246ECM received PRD error from power unit after PU Reset line assertedPower unit PRD error during warm start or after fatal error reseterror
4247Invalid power unit error code received by ECM from power unitPower unit sends unrecognized error code or ECM misinterprets received error codewarning
4248Bad data received from power unit during communications testECM detected error during PRD communications tests between ECM and power unitcritical
4255ETL-1 board or system level errorUnknown from provided texterror
4255Spurious interrupt received by ECMSignal noise or interrupt not held long enough; vector number often 03FHerror
4256ASC sent video gain information to ECM that was out of range; ECM overrides with valid value between 1.0 and 40.0ASC module sends video gain value outside valid range to ECMwarning
4256Suspected Auto Brightness Control problemAuto Brightness Control malfunction detectedwarning
4257Video gain control not workingTimeout waiting for Transmitter Ready bit in 2681 DUART status register after writing to TX holding registererror
4259A RAM high parity error interrupt occurredRAM high parity error in ECM modulecritical
425ARAM low parity error interrupt occurred on ECM moduleSystem boot or operation when ECM RAM parity check failserror
425CPower Unit communication errorChecksum mismatch on power unit packet received by ECMerror
425DPower Unit communication errorCRC error bit set during 8530 receive interrupt when reading from power uniterror
425EPower Unit communication errorECM detects Receive Overrun Error bit set when 8530 receive interrupt occurs and End of Frame bit is not seterror
425FBad data received from communicationConsole receives corrupted or invalid data during operationerror
4260Power unit reset in progress or recent reset occurredECM detects power unit in reset state during operator command or packet transmissionwarning
4261ECM detected a loss of power to its CPUCPU power rail dropped or disconnectedcritical
4262Software error: data byte count mismatch during power unit database download to ECMPower unit database download command with mismatched byte counterror
4263Divide by zero operation or uninitialized NV parameterECM attempted divide by zero operation or database errorerror
4264Power unit failed to complete reset and initialization within ten secondsECM sends reset command to power unit; RESET DONE bit not set within ten secondserror
4265Maximum exposure time reached; exposure backup timer timed outExposure reaches maximum allowed time for technique and is terminated by backup timer rather than AEC LUT sequencer hardwarewarning
4266ECM software error due to invalid AEC gain commandECM receives load AEC gain command (097H) with one or both data bytes equal to zeroerror
4267AEC LUT address exceeds maximum value (1299) after exposure terminationUpon termination of AEC exposure, if LUT address is greater than 1299 (4 sec exposure)error
4268Prep/expose switch transition interrupt with redundancy bit mismatchECM receives prep/expose switch transition interrupt; switch status bits from two separate storage locations do not matcherror
4270Software error: ASC told ECM that latch through was set but 2.1 Second Trail Delay Enable bits were not set in ETL configuration latchesRecord mode acquisitions attemptederror
4271Last exposure may be bad. Proceed with caution/call service.Offset on brightness signal exceeds 1.5 volts (153 when scaled by 102)error
4272Exposure time for AEC type exposure was less than zeroCalculated AEC LUT time + tail time minus anticipation time results in negative or zero valueerror
4273Exposure time calculated for fixed time exposure was less than zero (expose time + tail time – anticipation time)ECM module calculates negative exposure time during fixed time exposure setup or executionerror
4274Backup timer mode not configured correctly for primary timed exposure; prevents proper tube heating exposure time calculationTimer mode programmed as value other than F1–F5 in ECM backup timer registererror
4275Total prep/expose switch state is out of range (valid range zero to fifteen)Invalid value received for prep/expose switch state on ECM moduleerror
4276Recipe poke attempted past end of recipeASC tried to poke a value into the recipe at a RIP index greater than the size of the RIPerror
4277Power unit communication error requiring system recoveryDetected during power unit operation or reset sequenceerror
4277Bad data received from power unit; DMA packet less than 5 bytesECM receives incomplete DMA packet from power unit during communicationerror
4278Selected exposure time is greater than device will allowExposure terminated by loss of expose gate rather than exp/rad timer, such as film changer expose time incompatible with SEP programmer settingswarning
4279ASC sent initial AEC expose time outside acceptable range (less than 5 msec or greater than 2 sec); ECM AEC backup time algorithm expects initial expose time within this rangeSoftware error during AEC initializationerror
427ADatabase or software error in AEC backup time table calculation resulted in negative divisorOne of the calculations to compute the AEC backup time table resulted in a negative numbererror
427BTable top position data incorrectRT bus 8, 9, 10 report unrecognized table position during film changer stepper acquisitionserror
427CFilament warmup in progress; exposure not yet permittedPrep or expose switch pressed within 5 seconds of new filament activationinfo
427DBad data received from ASC to ECMFocal spot change or ABC mode change (Cine/Fluoro/progressive) after Prep Sequence Enable Command (Arcnet 88H)error
427EABC Algorithm calculated filament current exceeding maximum rating of active filamentFilament current calculation exceeds filament maximum (5.5 or 6.5 Amps)error
4280STARTWHEN timeout - receptor ready signal not achieved within specified recipe timeoutParticular signal or state not achieved after given amount of time as specified in currently active recipeerror
4281STARTWHEN timeout error indicating a signal or state not achieved within specified timeA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
4282STARTWHEN timeout: required signal or state not achieved within recipe-specified timeSignal or state does not reach expected condition within timeout period specified in active recipeerror
4282-42FFPower unit module Applications software error rangeError in Power unit module during Applications modeunknown
4283Start when timeout, external synchA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
4284Start when timeout, expose gateA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
4285Start when timeout, expose rateA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
4286STARTWHEN timeout: a signal or state not achieved within specified time from active recipeSignal or state fails to reach expected condition within recipe timeout windowerror
4287STARTWHEN timeout on control '2'; signal or state not achieved within recipe-specified timeParticular signal or state has not been achieved after given amount of time as specified in active recipeerror
4288STARTWHEN timeout on RT bus 9Particular signal or state not achieved after given timeout in active recipeerror
428ASTARTWHEN timeout on RT bus 11A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
428BStart when timeout on RT bus 12A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
428CSTARTWHEN timeout on real-time bus 13A required signal or state not achieved within recipe-specified timeout on RTB 13error
428DTimeout waiting for a particular signal or state from RT bus 14Signal or state not achieved within time specified in active recipeerror
428ESTARTWHEN timeout on RT bus 15Particular signal or state not achieved after given time specified in active recipeerror
428FSTARTWHEN timeout on RT bus 16Signal or state not achieved within timeout specified in active recipeerror
4290Start when timeout, timer 8A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
4291STARTWHEN timeout on timer 11 - a required signal or state not achieved within recipe timeoutA particular signal or state has not been achieved after specified time in active recipeerror
4292Start when timeout, timer 12A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
4294Start when timeout, timer 14Particular signal or state not achieved after specified time in active recipeerror
4295STARTWHEN timeout, overlap detectorA particular signal or state has not been achieved after a given amount of time specified in the currently active recipe.error
4296Start when timeout, bit 22 - signal or state not achieved within recipe timeoutParticular signal or state specified in active recipe not achieved within allotted time during exposure startuperror
4297Start when timeout, boost doneParticular signal or state not achieved after timeout specified in active recipeerror
4298STARTWHEN timeout errorSignal or state not achieved within recipe timeout period; causes exposure run terminationerror
4299STARTWHEN timeout error; recipe aborted and exposure terminatedRequired signal or state not achieved within timeout period specified in active recipeerror
429ASTARTWHEN timeout; rotor rotatingParticular signal or state not achieved after specified time in active recipeerror
429BStart when timeout, rotor at speedA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
429CSTARTWHEN timeout: a required signal or state was not achieved within the timeout specified in the active recipeA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
429DSTARTWHEN timeout error; signal or state not achieved within recipe timeout periodExposure run initiated when required signal does not reach ECM within specified timeouterror
429EStart when timeout, expose sequence enableA particular signal or state not achieved after timeout specified in active recipeerror
429FStart when timeout, prep seq. enable.A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipecritical
42A0kVp control errorVoltpac brush does not reach position commanded by ECM during servo lock sequenceerror
42A1Selected filament openFilament current during primary timed exposure not greater than 1.5 A with no inverter faulterror
42A2Filament control faultSTARTWHEN signal not achieved within recipe timeout periodcritical
42A3Filament control faultA particular signal or state has not been achieved after a given amount of time as specified in the currently active recipecritical
42B1Power unit communication errorCommunication failure between ECM and MPPU or VAMP power uniterror
42B2Power Unit communication errorLFX, VAMP or MPPU type power units experience checksum problems or receive packet with undefined bits/dataerror
42B3Power Unit communication errorSDLC chip detects CRC-CCITT polynomial mismatch in packeterror
42B4Power Unit communication errorPower Unit (LFX, MPPU, or VAMP) unable to communicate with ECMerror
42B5Power Unit communication errorLoss of communication between LFX/MPPU/VAMP power unit and ECM (main controller)error
42B6Grid-Cathode short errorGrid-cathode circuit short detected during high voltage operation in Cine modecritical
42B7Power unit soft start errorMPPU or VAMP type power units experience soft start failureerror
42B8Power unit sequence errorReported by MPPU or VAMP type power units during power sequencingerror
42B9Power unit tube select errorHV switch fails to commutate or compressed pins prevent proper cable contact during tube selectionerror
42BAPower unit KV range error in X-ray systemOnly reported by MPPU or VAMP type power units when KV range error is detectederror
42BBPower unit not operableReported by MPPU or VAMP type power units onlyerror
42BCkVp control not operableLFX, MPPU, or VAMP type power unit malfunctionerror
42BDPower unit regulation selection errorPower unit selection or configuration mismatch in MPPU or VAMP type power unitserror
42BEPower unit inverter 1 failureInverter 1 fault detected in power unit (MPPU or VAMP type)critical
42BFPower unit inverter 2 failureDetected during self-test in MPPU and VAMP type power unitserror
42C0Power Unit communication errorGenerator does not measure mA during exposure, or mA jumper not re-installed after calibration on Model Period 11 MP phase 2error
42C1Power Unit communication errorLFX, MPPU, or VAMP power unit communication failureerror
42C2Power Unit communication errorLFX type power unit loses communication with main consoleerror
42C3Overload detected in power unit mA (tube current) limiting circuitTube current exceeds safe limits during exposure; reported by LFX, MPPU, or VAMP power unitscritical
42C4Tube anode not rotatingAnode rotation control failure detected by LFX or MPPU or VAMP power unitscritical
42C5Maximum tube housing heat units reachedTube housing thermal limit exceeded; rotor controller thermal limit (42DE) or inverter thermal limit (42DF) may present similarlyerror
42C5Tube pressure sensor voltage out of specificationVoltage at tube pressure point not 24V during system operationerror
42C6Filament drive error - auto-reset in progressFilament drive system detects error condition; system attempts automatic resetcritical
42C7Filament drive errorFilament drive circuit malfunction detected in power unit (LFX, MPPU, or VAMP type)error
42C8Filament drive error; auto-reset in progressFilament heating problem during LFX, MPPU, or VAMP power unit operation; only serviced when inverter A controls radiationerror
42C9Filament drive errorPower unit detects filament drive circuit failure or DC offset error during operation when inverter B is controlling radiationerror
42CAmA overload occurredPower unit detects current exceeding safe limits during prep/expose switch transitionserror
42CAHTube spit error - X-ray tube arcing or sputteringX-ray tube arcing detected causing excessive mA feedbackerror
42CBInstruction not understood or power unit communication failureLFX/MPPU/VAMP power unit error; SCPU VRTX error, floating point error, or corrupted error tableerror
42CCPower unit memory corruptedReported by LFX and MPPU and VAMP type power units when power unit memory is corruptedcritical
42CDPower unit communication errorCommunication failure between console and LFX type power uniterror
42CEBad data received from power unitLFX type power units only; data communication errorerror
42CFIncoming line voltage lowSystem detects low incoming AC line voltage; only reported by LFX type power unitswarning
42D0Power unit TIRC errorPower unit communication or control error in MPPU or VAMP unitserror
42D1Power unit tirc errorPower unit communication or control failureerror
42D2Power unit tube select errorReported by MPPU and VAMP type power units; tube selection circuit fails or communication losterror
42D3Power unit tube select errorPower unit selection failure during operationerror
42D4Invalid power unit error codePower unit malfunction detected by MPPU or VAMP type power unitserror
42D6Input power lossOne phase of the network has dropped to zerocritical
42D7Power Unit Fault. High voltage unavailablePower unit failure detected, high voltage supply not availablecritical
42D8Power Unit Fault - High Voltage UnavailableMini Bridge switching feedback incompatible with selection or fuse/switch failurecritical
42DEMaximum rotor controller temperature reachedRotor controller temperature exceeds limitwarning
42DFMaximum power unit temperature reached. Allow time to cool.Inverter thermal limit reached in SCPU type power unitserror
42E0Power unit self test error occurredSelf-test failure in MPPU or VAMP type power unitserror
42E1Power unit tube spit recoveryX-ray tube arcing or spit eventwarning
42E2Power unit missing current on filament AOnly reported by MPPU and VAMP type power units when filament A current missingerror
42E3Power unit missing current on filament BOnly reported by MPPU and VAMP type power units when filament B current missingerror
42E4Power unit tube spit recoveryTube spit detected during exposure; kV automatically cut off with automatic retrywarning
42E5Power unit tube spit recoveryTube spit detected during X-ray exposure; kV automatically cut off with automatic retrywarning
42E6Power unit tube spit recovery during exposureDuring X-ray exposure when tube spit event detectederror
42E7Power Unit Tube Spit RecoverykV automatically cut off during tube spit eventerror
42E8Power unit tube spit recoverykV is cut off during tube spit; automatic retry performed; after 7 spits during same exposurewarning
42E9Power unit tube spit recoveryOccurs during tube spit event when kV automatically cut off; detected after automatic retry attemptwarning
42EAmA meter limitPower unit (LFX, MPPU, VAMP, SCPU) mA meter saturates during exposurewarning
42EAHmA meter saturation detectedmA measurement exceeds meter saturation threshold during operationwarning
42EBmA meter error in Power UnitFailure of mA metering circuit test performed after power up or tube selectionerror
42ECClosed loop mA errorReception of error from LFX, MPPU, or VAMP power unitserror
42F0Excessive tube current attemptedAttempt to operate tube at current exceeding safe limitserror
42F1Power unit power supply errorPower unit voltage regulation failure detected by MPPU or VAMP power unitcritical
42F2Power unit exposure without commandPower unit receives exposure signal without a prior commanderror
42F3Power unit backup mAs errorPower unit backup milliampere-second (mAs) measurement out of specification or unavailableerror
42F4Power unit exposure time exceededPower unit exposure exceeds 10 secondscritical
42F5KVP (kilovolt peak) power failure in SCPU type power unitVoltage regulation or measurement board circuit error in power uniterror
42F6MA (tube current) failurePower unit detects tube current supply malfunctionerror
42F7Power unit internal errorReported only by SCPU type power units during operationerror
42F8Overload detectedPower unit detects overload conditionerror
42F9Power unit tube spit recoveryX-ray tube unable to complete recovery phase after exposurewarning
42FAPower unit TIRC errorReported by MPPU and VAMP type power unitserror
42FBTube casing water flow not detectedWater flow sensor in tube casing circuit fails to detect circulationerror
42FCPower unit no KV or KV ripplekV does not reach 12 kV within 500 microseconds after start of exposureerror
42FDPower unit inverter overheatInverter temperature exceeds safe operating limitwarning
42FEPower unit KV balance faultDetected by MPPU or VAMP type power units when kV regulation failserror
4300Bad data receivedConsole receives invalid application command in GE bus packet from ASCerror
4300-43FFConsole module Applications software error rangeError in Console module during Applications modeunknown
4301Bad data received from ASC in GE bus packetConsole receives invalid video format commanderror
4302Bad data receivedConsole receives a video format command of invalid length in a GE bus packet from the ASCerror
4303Bad data received from ASC to console via GE bus packetConsole receives invalid display command character in GE bus packet from ASCerror
4304Bad data received from ASC; invalid video display character in GE bus packetConsole receives malformed data packet from the Acquisition and Storage Controller (ASC)error
4305Bad data received from ASC to ConsoleConsole receives invalid GE bus command from ASC to position video cursorerror
4306Bad data received from ASC to ConsoleConsole receives invalid GE bus command from ASC to selectively erase video displayerror
4307Bad data received from ASC to ConsoleConsole receives invalid GE bus command from ASC to erase video displayerror
4308Console received invalid GE bus command from ASC to disable touch screen or trackball reportsInvalid data transmission from ASC to Console on GE buserror
4309Bad data received from ASC to ConsoleConsole receives invalid GE bus command from ASC to enable touch screen or trackball reportserror
430ABad data received from ASC to ConsoleConsole receives invalid GE bus command from ASC to configure handswitch signalserror
430BBad data received from ASC to Console; invalid GE bus command for programmable touch screen row templateConsole receives invalid GE bus command from ASC during touch screen row template loaderror
430CInvalid GE bus command received by ConsoleConsole receives invalid GE bus command from ASC to sound audible beep toneerror
430DBad data received from ASC to ConsoleConsole receives invalid GE bus command from ASC to sound audible error toneerror
430FInvalid box erase coordinatesAttempt to erase a box with invalid coordinate parameterserror
4380Touch screen may not be workingConsole detects a failure in communications with, or performance of the optical touch frameerror
4400Communication problem between console and DX/DLX moduleCommunication failure detectederror
4400-44FFAdvantx DX Applications software error rangeError in DX module during Applications modeunknown
4401Communication problem between console and DX/DLX moduleLoss of communication between Advantx console and DX/DLX moduleerror
4402Communication problem between console and DX/DLX moduleDX/DLX module communication failureerror
4404Digital system is out of disk spaceStorage capacity exhausted on DX/DLX moduleerror
4405Digital system maximum run length reachedMaximum allowable continuous digital acquisition time exceedederror
4406Digital system image disk is write protectedSystem detects write protection on image diskerror
4407Digital system image disk faultDX/DLX subsystem errorerror
4408Max number of sequences for this patient reachedAttempt to create additional sequences for a patient that has reached the system limitinfo
4409Max number of categories for this patient reachedAttempting to add more categories than system allows for current patientinfo
440AMax sequences on digital system reachedMaximum number of sequences stored on digital systemwarning
440BDigital system offsets failedCommunication error between DLX and VIC modules or VIC module A15 circuit board failureerror
440CDigital system losing too many images due to insufficient compression ratioImage compression cannot maintain required ratio; digital storage system cannot keep pace with acquisitionerror
440DTrial too brightX-ray image overexposure during trial exposurewarning
440ETrial image too darkImage acquisition returns signal below acceptable brightness levelwarning
440FImages being recorded to Cine film onlyCine film recording mode active without alternative recording mediawarning
4410Communication problem between console and DX/DLX moduleDX/DLX error that propagates to Advantx consoleerror
4411Front and lateral trial images too brightExposure technique or filtration generates overexposed trial imageswarning
4412Front trial image too bright, lateral trial image too darkExposure technique or filtration settings produce inconsistent image brightness between frontal and lateral projectionswarning
4413Front trial image too dark, lateral trial image too brightRadiographic trial images expose incorrectlywarning
4414Front trial too dark, lateral trial too brightImage acquisition results in improper exposure levelswarning
4415DX store function in progress on the in-room monitorStore function is active during DX/DLX operationinfo
4416DX recall function in progress on the in-room monitorRecall operation initiated on DX/DLX moduleinfo
4417Digital System ErrorDigital system malfunction in DX/DLX moduleerror
4500DR busy/fault existsDR Module failed to respond during power up or DR module reported an internal error during command execution or status reportingerror
4500-45FFAdvantx DR Applications software error rangeError in DR module during Applications modeunknown
4502System communications problem between ASC and DR BIUFluoro switch depression; DR Module fails to set DR-RDY signal high within 50 msec after FLUORO SWITCH signal is set lowerror
4503System communications problemDR BIU to DR Module communication failure, ASC to DR BIU data validation failure, or ASC pass-through command errorerror
4504System communications problemSerial communication packet transmission validation failure; STX code not found in first position of DR Module BIU packeterror
4505System communications problem between DR BIU and DR ModuleSerial communication packet validation failure; calculated last packet location does not contain ETX code (3H)error
4506System communications problem - message queue overflowDR BIU processor detects 20 stacked messages waiting to transmit via serial link to DR Moduleerror
4507System communications problemSerial link Duart (U133) transmitter fails to become ready after specified perioderror
4508System communications problem between ASC and DR moduleSerial link Duart detects parity error in character reception from DR Moduleerror
450ASystem communications problem between DR BIU processor and DR ModuleDR Module fails to acknowledge command (ACK character) within 20 msec after ETX transmission; retransmit attempts fail after 3 trieserror
450BSystem communications problemDR Module fails to set DR-RDY signal low within 10 msec after PREP SWITCH signalerror
450CConsole displays: System communications problem. Please try again.DR-RDY signal (Pin 7 of U133) not set low within 10 msec of FLUORO SWITCH signal going lowerror
450EDR function inoperativeDR Module fails to respond to INT-SYNC signal within approx. 0.5 seconds by setting SYNC-ACK signal appropriatelycritical
4510Enter new study information and continueDR module rejected New Study Command from consoleerror
4511System communications problem. Please try again.DR Module fails to respond to EXPOSE SWITCH signal from DR BIU within approximately 1 milliseconderror
4512System communications problem between DR BIU and DR ModuleDR Module fails to respond within specified time frame after ETX command transmissionerror
4513System communications problemSerial link Duart (U133) detects overrun of data in receiving Fifoerror
4514System communications problemDR Module fails to acknowledge command from DR BIU processor within 20 msec or sends NAK after 3 retransmit attemptserror
4515System communications problemDR BIU processor detects bad values for certain parameters in its database during warm start initializationerror
4517XY shift failureSystem in service (Cal) mode; DR Module response indicates problem with downloading x-y shift parameters for blanking circleerror
4518Circle adjust failureCalibration attempted during service (Cal) mode when downloading blanking circle size parameter to DR Modulewarning
4519Smooth filter failureDR Module's response to downloading coefficients for filter number 2 indicates a problem with transmitted commanderror
451ASharp filter failureSystem in service (Cal) mode; DR Module response to coefficient download for filter 1 indicates problem with transmitted commanderror
451BMaximum number of studies (50) defined on the DRS (Dig Rec module)ASC sends DRS command to define 51st new studyerror
451CMaximum number of images have been stored in this study99th image reached during acquisitionwarning
451DDig Rec temperature too highMHR temperature rises above operational levelwarning
451ENo storage of images on the Digital Recording disk is availablePower-up test failure; disk storage unavailable for image acquisitionerror
451FDRS (Digital Radiography System) Power-Up Test FailureDRS power-on self-test failure; system enters degraded modecritical
4520Digital imaging system failure in DRS modulePower-up test failure in Digital Radiography Systemerror
4521Filters and post-processing functions not availablePower-up test failureerror
4522Power-up test failure in DRS moduleSystem power-up self-testerror
4523Digital Filming is not availablePower-up test failure in DRS moduleerror
4524Communication problem encounteredCommunication failure between system moduleserror
4525Communication problem encounteredCommunication failure between system componentserror
4526Communication problem encounteredCommunication failure between system componentserror
4600-46FFCine module Applications software error rangeError in Cine module during Applications modeunknown
4602Invalid frame rate received by CINE module from ASCModule receives invalid frame rate from ASC in set frame rate commanderror
4603Invalid phase received or camera running during set phase commandModule receives invalid phase from ASC in set phase command or camera is running when command receivederror
4604CINE module software thinks camera is already running when commanded to operateCINE module receives expose command while software believes camera is runningerror
4605Software state mismatch on CINE module during film transport requestRequest for film transport received from ASC while software thinks camera already runningerror
4606Jog not allowed during Cine recordingEnable cine trial command received while software running cameraerror
4608System communications corruptedCommunications failure between console and in-room display or modulescritical
4609Instruction not understoodUser command not recognized by systemwarning
460ACine camera out of filmINTLOK bit is off on port 0 and film remaining A/D value is less than calibrated no-film valueinfo
460BCine magazine missing or not lockedINTLOK bit off on port 0 and film remaining A/D below 8 countserror
460CCine camera fault; camera interlock error indicating camera cannot runINTLOK bit read as zero by CPU on port 0 during prep or camera starterror
460DCine camera overheated. Heat protection timeout timer expired.Heat protection timeout timer (80188 timer 1) expires, generating interrupt to stop camera.warning
460ECine mode faultWhen attempting to run the camera and Power-amp board indicates fault via PAFAULT* signal on port 0error
460FCine film jammed in camera; camera stalled and not reaching speed within timeoutCamera fails to reach 98% expected speed within timeout; heat timer interrupt fires while camera below speederror
4610Camera still hot - thermal protection activeAttempting to start camera when heat protection software detects excessive camera temperaturewarning
4611Cine mode fault; wrong command sequence from ASC or software malfunction during prepImproper exposure setup command sequence; camera motor on commands not receivederror
4620Cine camera not line lockedHardware phase lock loop fails to lock during camera startup; LOCK signal remains unasserted after stall timer expireswarning
4621System communications corruptedMAST-OSC signal unstable or not close to final speed after five seconds pre-run checkerror
4630System communications problemModule tried to transmit packet to node and did not receive transmit acknowledge before timeouterror
4631Improper updating of packet indexer; indexer went past end of packet after processing command due to insufficient packet dataSystem communications problem during packet processingerror
4632Software attempted to transmit more data than would fit in a single packetData transmission exceeds packet capacityerror
4633GE Bus transmit queue full; module software lost or ARCNET not transmittingSoftware internal GE Bus transmit queue was full when module attempted to send packeterror
4634ARCNET hardware could not find designated node on system to transmit packet toARCNET hardware indicates transmitter available but no transmit acknowledge given before timeouterror
4635Instruction not understood. Please try again.GE bus transmission received with command not found in command tableerror
4636Software task scheduling failure on system timerSystem attempted to schedule a task on the timer with no available slots in the listerror
4637Instruction not understood. Timer ID outside scheduled task list.Attempt to update or flush scheduled task with timer ID outside list of scheduled taskserror
4640Instruction not understoodUser input not recognized by systemwarning
4641Instruction not understoodUser input not recognized by systemerror
4642Instruction not understoodUser input not recognized by consolewarning
4700Instruction not understoodCommand received across GE Bus (ARCNET) is not a defined command for VIC moduleerror
4700-47FFVIC module Applications software error rangeError in VIC module during Applications modeunknown
4701ARCNET transmission attempted to non-existent nodeTA was set and TMA was not seterror
4702System communications problemARCNET transmission failed; destination module did not acknowledge reception within 2 secondserror
4703GE Bus command interpreter packet index overrun; number of bytes transmitted did not match expected numberVIC module receives command packet with incorrect data byte counterror
4704Packet transmission exceeds maximum transmittable sizeSoftware attempts to transmit a packet larger than the maximum allowederror
4705Integer math overflow occurredMath overflow in VIC moduleerror
4706Divide by zero error in VIC moduleComputational fault in VIC circuit boarderror
4707Bad data received from consoleData integrity or communication failure between console and VIC moduleerror
4708Bad data received from GE Bus command with invalid parameter outside valid rangeGE Bus command parameter outside valid range for the commanderror
4709TV gain curve data badDuring system operation or initialization when gain linearization data becomes invaliderror
470AThe GE Bus output queue is full.GE Bus output queue becomes fullerror
470BBad data received from consoleAttempt to schedule a time-delayed task using SCHEDULE_SOFT_TIMER_TASK with a timer already allocated to another taskerror
470CTV camera power supply failedInterrupt received indicating warm power supply has failedcritical
470DTV camera power supply failedInterrupt received indicating cold power supply has failedcritical
470EPhoto cathode current read attempted during AGCAttempt to sample photo cathode current with AGC enablederror
470FAGC attempted during photo cathode current readAttempt to enable Automatic Gain Control Function while Photo Cathode Sample function was activeerror
4710Photo cathode current read already activeAttempt to sample photo cathode current while previous sample command still executingerror
4711Instruction not understood. Please try again.Attempt to flush or set a soft timer greater than the number of timers configurederror
4712Instruction not understood. Please try again.Attempt to schedule a task when no more timers availableerror
4713Video setup corruptedSystem warm state data retrieval failure with corrupted parameterswarning
4720Video camera power supply has failedInterrupt received indicating camera power supply failureerror
4721Video camera system inoperativeCamera sweep fail occurrederror
4722Video camera system inoperativeCamera hardware generates interrupt indicating camera sync failerror
4723Video camera system inoperativeCamera phase lock loop failure occurrederror
4724Camera A/D converter did not signal completed conversion within timeout period (40+ milliseconds)A/D conversion timeout on camera interfaceerror
4725TV Video signal out of rangeSetup adjust task unable to achieve readable video signal rangeerror
4726Video shading correction faultAmount of shading compensation required exceeds hardware capabilityerror
4727Light bias inoperativeVIC initialization detects signal current change within 10 A/D bits when DAC output is varied, or when non-zero light bias is requestederror
4728Light bias inoperativeRequested signal current from light bias cannot be achieved; DAC at limiterror
4740Video aperture did not respondIris electronics did not generate interrupt indicating iris reached requested position within 1 second allowable timeerror
4741Preset value requested exceeds maximum aperture valueAperture preset command sent to VIC when requested value is larger than Max Aperture Command limiterror
4742Video Gain Command (VGC) received over RT Bus serial link from ECM was outside allowable rangeVGC received with value outside 66H through 0FFFH rangeerror
4743Aperture in Position interrupt received when none was expectedUnexpected aperture position interrupt signal from VIC modulewarning
4744IRIS A/D conversion interrupt not received within timeoutIRIS AD RDY interrupt not received within 10 milliseconds of conversion starterror
4780Optical distributor not positionedGate rotate failed to position in time. At initialization, electrical in position bit not received within 2 seconds, or during move bits not lost within 100 ms or do not reappear within 1 second.error
4781Optical distributor not positionedIn position bits go true then false within 100 milliseconds during normal gate movementerror
4782Optical distributor not positionedNeutral Density Filter shows both position bits true (close to in bit and close to out bit from XRII I/F board both logic low)error
4783Optical distributor not positionedOptical distributor positioning failure during initialization or operationerror
4785Optical distributor not positionedSystem detects optical distributor is not in correct position during operationerror
4786Optical distributor not positionedGate pivot failed to position in timeerror
4787Optical distributor not positionedGate pivot amp indicated stall at bad timeerror
4788Optical distributor not positionedSystem detects optical distributor is not in correct positionerror
4789Optical distributor not positionedGate clock watchdog timer timeout after 10 milliseconds without gate clock reseterror
478AGate and/or filter motion not allowed due to XRII interface link failureSerial link to XRII I/F board is downerror
478BGate pivot or ndf did not recoverGate and/or ND filter attempting to recover from a prior error and could not successfully complete the recoveryerror
478CGate stall hardware latch kicked inStall watchdog detectederror
4790Output buffer containing messages to XRII I/F micro-controller is fullCommunications with XRII I/F not proceeding normallyerror
4791XRII I/F micro-controller has reported an errorBad data received by console from XRII interfaceerror
4792Bad data received from consoleData transmission error between console and systemerror
4793Requested field of view in XRII command not in valid rangeInvalid XRII field of view command sent to systemerror
4794XRII I/F micro-controller did not respond appropriately to VIC within prescribed timeXRII I/F micro-controller communication timeouterror
4795DUART framing error on XRII interface communication channelFraming error detected on DUART channel A communicating with XRII I/F Board microcontrollererror
4796Bad data received from XRII interfaceDUART on aperture board reports parity error on channel A communicating with XRII I/F micro-controllererror
4797DUART overrun error on XRII & Iris Control Board channel ACharacters received from micro-controller faster than removed from input queueerror
4798XRII I/F board 8031 microcontroller power-up diagnostics failedPower-up self-test of microcontroller on XRII I/F board detects errorerror
4799Bad data received from 8031 microcontroller on XRII I/F boardResponse byte from 8031 does not match command sent to iterror
47B0TV Rotator failed to track commandRotator unable to respond to positioning commandswarning
47D0Video Tape Recorder did not respond to commandVTR did not respond to a command from the systemwarning
47D1VTR Communication errorVTR sent a NAK to the VICerror
47D2VTR not in remote modeVTR front panel used or tape becomes unthreadedwarning
47D3VTR Communication errorCommunication failure between console and VTR moduleerror
47D4Search not confirmed; tape search did not completeAttempted search for mark or run too close to tape beginning or end, or cue/record attempted before valid runs existwarning
47D5Freeze deselected due to inactivityVTR left in freeze frame state for 30 minutesinfo
47D7VTR communication errorVTR module communication failureerror
47D8VTR communication errorVTR software attempted to stimulate Action-State driver with invalid stimuluserror
47D9VTR communication errorVTR responded with invalid transport modeerror
47DAHPR–50 VTR left idle in ready state for more than 30 minutesVTR remains in ready state without use for over 30 minuteswarning
47DBVTR communication errorVTR DUART detected a framing errorerror
47DCVTR communication errorVTR DUART detected parity error during communicationerror
47DDVTR communications errorVTR DUART detected an overrun errorerror
47E0UnknownUnknownunknown
47FEInvalid data base item ID numberDatabase access error on VIC moduleerror
47FFInvalid data base item sizeDatabase item size validation failureunknown
4800Communication problem between console and display moduleUndefined command received by DF moduleerror
4800-48FFDigital system Applications software error rangeError in Digital system during Applications modeunknown
4801Communication problem between console and in-room componentsArcnet packet received with legal command but incorrect byte counterror
4802Communication problem between console and DF (Digital Fluorography) moduleApplication attempts exposure sequence when DF unable to acquire imageserror
4803Communication problem between DF module and ASCDF module receives packet with invalid application mode (neither Static nor Dynamic)error
4804Communication problem between DF module and consoleDF module receives packet indicating acquisition mode change during active acquisitionerror
4805Fatal error in DF (Digital Fluoroscopy) module detectedDF module detects fatal error conditioncritical
4806Insufficient disk space available for acquisitionAvailable disk space drops below 2 seconds of run time during acquisitionwarning
4807DF maximum run length of 999 images was reachedCurrent run sequence attempted to store more than 999 imagesinfo
4900X-ray tube and table receptor alignment mismatchTable selected as RAD receptor but tube and table receptor not in proper alignmenterror
4900-4BFFPositioner module Applications software error rangeError in Positioner module during Applications modeunknown
4901Room door not closed during RAD exposure attemptRAD exposure attempted with room door open or signal missingwarning
4902Cassette not in place or cassette-in-place signal not presented to positionerRAD exposure attempted for selected receptor without cassette detectederror
4903Bucky rotation mismatch errorRAD exposure attempt when receptor rotation does not match X-Ray fielderror
4904X-ray tube is in an improper orientationTube and/or receptor are not in proper alignment for RAD exposure or operator intention cannot be determinederror
4905Wall receptor alignment failure. Tube and wall receptor not in proper alignment for exposure.Wall unit selected as RAD receptor at console but tube and wall receptor misaligned.error
4906Table height not at proper exposure positionRAD exposure attempted with variable-height table not at proper heighterror
4907Exposure terminated early during RAD exposure with reciprocating grid receptorRAD exposure attempted with reciprocating grid receptor when grid did not reach speed or speed signal bouncingerror
4908Film not in place, film jam, or patient ID card not in place during RAD exposure with film changer wall deviceRAD exposure attempted with Changex or ACS-1000 film changer when film is absent, jammed, or patient ID card missingerror
4909Collimator motion problem detectedCollimator fails to move or respond to commandserror
490FInvalid receptor; software corrupted because internal receptor indicator is not acceptable valueSystem detects unacceptable internal receptor indicator value during operationerror
4911Film changer programmer not readyFilm magazine and/or film receiver not attached to film changer, or FC READY signal indicates not ready after required delayserror
4912Film changer is out of filmFILM IN PL signal at prep and during run indicates no film when expected, film changer indicates not ready on FC READY signalwarning
4913Table not positioned at horizontal for film changer operationSoftware detects HPOS0 signal not asserted when prep mode enterederror
4914Stepper switch in wrong position on film changer programmerStepper selection switch on PDQ-2 film changer does not match console selectionwarning
4915Injector selection switch mismatch between PDQ-2 film changer and consoleInjector selected on console but not at changer, or vice versawarning
4916Film changer programmer not readyInjector is selected for use and injector is not armed when used with PDQ-2 film changerwarning
4917Remove all cassettes to continueUnknown from excerptunknown
4918Collimator failure detectedOperator attempts to manually move Rad Ultranet Collimator blades using collimator knobs immediately after tube warm-up or within 30 seconds of warm-up completionwarning
4919Collimator failure detectedCollimator mechanism malfunction or positioning errorerror
4A00System communications problemGE bus transmission received but command not found in command tableerror
4A01Bad data received from Positioner module via ARCNETSoftware internal GE Bus transmit queue became full when attempting to send packet, indicating module software failure or ARCNET communication failureerror
4A02System communications problem - transmit timeout on Positioner modulePositioner module attempted to transmit packet and did not receive acknowledge before timeout expirederror
4A03Bad data received from consoleSoftware attempted to schedule a task to run and an internal timer indicator was allocated to a different scheduled taskerror
4A04System communications problemSoftware attempted to transmit more data than expected to fit in a single packeterror
4A05System communications problemCommunication failure between console and Positioner moduleerror
4A06Improper updating of packet indexer; indexer went past end of packet after processing commandPacket had less data than expected during command processingerror
4A07ARCNET transmission failed; destination node not found on networkARCNET transmitter available but no transmit acknowledge received before timeouterror
4A08Bad data received from Positioner module during scheduled task update/flushAttempt to update or flush a scheduled task with a timer id outside the list of scheduled taskserror
4A09Bad data received from Positioner module; system timer task scheduler overflowSoftware attempts to schedule task on system timer when list is fullerror
4A0AA/D conversion did not complete in reasonable time on Positioner moduleA/D conversion complete bit not found set within expected time period when reading A/D on interface board or collimator boarderror
4A0BSystem communications problemPart or all of data from GE bus packet is not valid for the commanderror
4A10Bad data received from consoleProblem transmitting request for download of calibration and configuration dataerror
4A31System communications problemCommunication failure between console and in-room display or connected moduleserror
4A32System communications problemUnknown communications error in Positioner moduleerror
4A34System communications problemPositioner module communication failurewarning
4A42Bad data received from Positioner moduleData communication error between console and Positioner during operationerror
4A51Tableside control communication failure after reset attemptsSoftware detects problem communicating with tableside controls; reset diagnostic shows continued problem or many unsuccessful reset attemptserror
4A52Tableside control communication error detected by DUART status registerError in reception from tableside control detected during system operationwarning
4A53Invalid state transition in Positioner moduleSoftware detects an invalid transition state based on CC select state, auto format select state, and II or Film changer configurationerror
4A54Software could not determine arc sine, cosine or tangent of L & U anglesPositioner module software failure during angle calculation from trig tableerror
4A55Room door needs to be closed to continue operationRoom door signal indicated open at Prep timewarning
4A57Image Intensifier not at or below 30cm SIDII position not at or below 30cm from ISO-center during Interchange operationwarning
4A58Film changer rotation interlockFilm changer is in rotated position when system enters prep modeerror
4A59Film changer not correctly locked at end-of-travel positionAt Prep time when film changer is not at End-of-travel (EOT) or AUX EOT as indicated by signalserror
4A5APositioner not in valid Cerebral Contact mode positionAttempt to use Cerebral Contact mode when positioner does not meet valid position requirements; one of three position criteria not satisfiedwarning
4A5BCerebral Contact mode setup failed Case I/II checks; film changer not properly positionedAttempted Cerebral Contact mode prep when L/U arm angles or film changer position do not meet valid conditionserror
4A5CFilm changer position interlock error in Cerebral Contact modePositioner does not meet conditions: L angle III–<+45° OR U angle >+45° with film changer fully extended and at auxiliary end-of-travelerror
4A5DNon-maskable interrupt occurred due to invalid I/O port access, unacknowledged hardware, or non-existent memory accessCPU Timer 0 expired without acknowledgmenterror
4A5ECPU detected a math overflow during calculationMath overflow in positioner CPU during some calculationerror
4A5FCPU detected a math attempt to divide by zero in Positioner moduleOccurs during system operation or hardware motion that triggers undefined math operationerror
4A60Film Changer Programmer not readyAt prep time when film changer signals indicate programmer not ready or film not in placeerror
4A61Film Changer out of filmFC RDY signal not ready and FC FILM IN PLACE SW signal not indicating film present during PREPwarning
4A63Collimator has stalledLONG STALL and/or TRANS STALL signals active during routine status checkerror
4A64Cannot select CC (Continuous Cinefluoroscopy) mode when table stepper is selectedCC button pressed while table stepper is selected or STPR ON signal activeerror
4A65CC (Cine Continuous) mode was deselected because table stepper unit power was detectedInternal flag indicates CC is selected while STPR ON signal shows stepper is activatederror
4A66Injector did not inject as programmedInjector does not appear to inject after being commanded; delayed injection timer expires without INJECTING signal detectederror
4A67FC program is not compatible with console selectionPositioner finds plane selections on console and SEP programmer do not agree at PREP timeerror
4A68SEP programmer not indicating any plane selectionFilm changer selected as SEP, rad acquisition requested, SEP power-on delay expired, neither CHANGER 1 SEL nor CHANGER 2 SEL signals activeerror
4A69Film changer error reportedFC RDY signal went away or FC FILM IN PLACE signal not presenterror
4A6APositioner software initialization failurePositioner module software fails to initialize during system startup or warm booterror
4A6BCollimator problemDuring X-ray exposure in CFM/AFM vascular systemserror
4A81Serial link between Vascular positioner module and host communication failed5 packets transmitted to Host with no response after successful link initializationerror
4A82Receiver error on host link - Serial Link interface between Vascular positioner module and host has failed; DUART status for received byte was badSerial communication error detected between Positioner and host system during normal operation or after initializationerror
4A83Serial link between Vascular positioner module and host not working correctlyUnsequence packet received; high-order bits of data bytes do not toggle in predictable patternerror
4A84Host not in application modeHost executed internal mode change during applications mode and reported to Vascular positionererror
4A85Auto-positioner control problemSerial link interface between Vascular positioner and autopositioner tableside control has failederror
4A86Auto-positioner control problemSerial link interface between Vascular positioner and autopositioner tableside control has failed. Link must have successfully initialized for this error to be reported, implying actual equipment failure.error
4A87Auto-positioner control problemSerial link interface between Vascular positioner and autopositioner tableside control fails; invalid byte sent from tableside controlerror
4A88Cannot memorize current position; L-arm positioned outside allowable limitsUser attempts to store L-arm position while outside permitted rangewarning
4A89Timeout: pre-programmed position not activated within 5 secondsUser fails to use selected pre-programmed position within 5 second window after selectionwarning
4A8ATable top is not centered laterallyTable top detected in non-centered lateral position during system initialization or operationwarning
4A8BLP (positioning arm) is in an invalid state for storing pre-programmed positionsUser attempts to store a position while LP is not at park or iso-centerwarning
4A8CI/O driver failedI/O-driver call within the software failederror
4A8DUnexpected start byte on serial link between Vascular positioner module and hostSerial communication protocol violation: more than one byte missed in packet causes receiver to not detect packet end and misidentify subsequent start byteerror
4A8EMissing start byte in serial packet from Vascular positioner moduleNext received byte after complete packet is not encoded as valid start byteerror
4A8FModule Utilities and Driver package (M.U.D.) reported an internal failureBad status received from MUD moduleerror
4A90Vascular positioner calibration and configuration data transfer failedTransfer of calibration & configuration data from Vascular positioner module to host across serial link did not completeerror
4A91Gantry operation degradedVascular positioner module commanded host to change operating modes but host did not change to required modewarning
4A92Gantry problem detected by positioner moduleVascular positioner protocol software driver detects internal error conditionerror
4A93Gantry link restart in processGantry link restart eventinfo
4AA1Cannot take exposures with LP parkedLateral Positioner in parked position at PREP stagewarning
4AA2LP control errorSerial link to LP appears down; no response after 20 tries to communicate at 50ms intervalserror
4AA3LP control error detected during data byte receptionParity, framing, or overrun error in LP data byte receptionerror
4AA4Communication problem with LP (Positioner)Packet parity error from Positioner transmission to consoleerror
4AA5Communication problem with LP (Positioner)Unrecognized data byte in last Positioner transmission; transmission start byte indicates errorerror
4AA6Communication problem with LP (Positioner)Data byte received from LP is out of sequence; MSB does not correspond to data byte counterror
4AA7Lateral system not operating. Cannot perform task.Lateral ASC appears down; Positioner requested cal/config data with two minute timeout and no responseerror
4AA8Table not rotationally centered to allow LP motionTT ROTATNL CNTR signal indicates not centered during LP motion or smart handle requestserror
4AA9Positioners not orthogonal - L angle not at zero degreesAttempt to move LP to or from vertical position when L angle is not zeroerror
4AAAPositioner cannot drive to next auto position due to interferenceTransmission start byte from LP electronics indicates positioner interferencewarning
4AABAttempted auto position operation with LP at isocenter and Frontal plane selectedAuto position operation initiated with LP at isocenter; Frontal plane selected on Smart Handleerror
4AACExceed travel limitLP (linear positioner) reaches its travel limit in current directionerror
4AADDrive not respondingLP indicating axis drive failure to positioner in transmission start byteerror
4AAEX-ray tube on the LP has collided with something and motion is disabledPhysical collision detected via third data byte from LPerror
4AAFII on lateral LP has collided with object; motion disabledPhysical collision detected during II motion on lateral positionererror
4AB0X-ray tube collision with L/U; motion disabledX-ray tube on L/U collides with object; collision status indicated in third data byte from LPerror
4AB1Imaging Intensifier (II) collision on L/U; motion disabledII on L/U collides with object; collision status in third data byte from LPerror
4AB2Type 1 bad interlock. Safety mask installed when it should not be, inhibiting Rad Acquisition.Safety mask installed on collimator when system expects noneerror
4AB3SID ErrorType 1 bad interlock; SID value too small or too large or SID change during fluoroscopyerror
4AB4Collimator error; collimator blades are open too farType 1 bad interlock during X-ray exposure or fluoro while changing SIDerror
4AB5Collision detection problemOperating fault detected on collision detection system switcherror
4AD9L-arm outside safety cone during spin examination learning phaseAttempt to record pivot arm spin trajectory while L-arm is outside safety conewarning
4ADAReady. Clear rotation area.Between end of trajectory check and beginning of image recording in spin examinationinfo
4ADBPosition not allowedRequested start or stop position refused during Spin and Bolus mode examinationwarning
4ADCTable ReadyAll Bolus Technique learning phases have ended correctlyinfo
4ADDReady for spin test message displayed to userA spin trajectory has been memorizedinfo
4ADESpin test in progressDuring spin test procedure executioninfo
4ADFSpin set-up abortedSoftware detects a condition that is not part of an allowable spin sequenceerror
4AE0Hardware error detected on the Stepping Board from the PositionerHardware malfunction on Stepping Boarderror
4AE0Table problem. Operation degraded, continue with caution.Table operation fault detectederror
4AE1LC or Tilt-C gantry subsystem board initialization failureGantry board not responding to initialization command at power-up or reseterror
4AE1Gantry problemGantry system error during operationerror
4AE2Gantry problem. Operation degraded, continue with caution.LC or Tilt-C CPU software detected an internal error conditionwarning
4AE3Gantry problem with degraded operationLC or Tilt-C gantry internal EEPROM storage failurewarning
4AE4Gantry problem; LC or Tilt-C CPU watchdog timer failed during runtimeLC or Tilt-C CPU watchdog timer failure during system operationerror
4AE5Gantry problem. Operation degraded, continue with caution.LC or Tilt-C gantry reports error detected at host end of Vascular positioner module serial linkwarning
4AE6Gantry collision occurredSecond-level collision or limit switch was hiterror
4AE7Increase SID for full displaySID reduced to point where collimator cannot track with 40cm image intensifierwarning
4AE8Stepper problem in positionerStepper motor or related control circuit failureerror
4AE9Position not allowed. Table too close to electrical and/or software limits.In Auto Positioner mode when table position prevents safe trajectory completionwarning
4AEAConsole/In-Room Display: Table or gantry not in authorized position.System detects table or gantry position outside allowed rangeerror
4AEBTable is moving to the start positionPositioning operation initiatedinfo
4AECPosition attainedProgrammed auto-positions have all been reachedinfo
4AF0Gantry problem. Operation degraded, continue with caution.LC or Tilt-C gantry subsystem boards report communications problemwarning
4AF1One of the LC, LCA or Tilt-C gantry subsystem boards has reported one of several possible error conditionsError conditions on gantry subsystem boards; multiple errors logically OR-ed togethererror
4AF1Gantry problem with operation degradedGantry positioning system detects fault or limits operator capabilitywarning
4AF2Positioner move command incomplete or failedPositioner cannot complete 90-degree (3-degree tolerance) lateral moveunknown
4AF3Positioner cannot obtain required LAO/RAO angle of 15 degreesPositioner fails to reach 15° LAO or RAO positionunknown
4AF4Obtain CRA/CAU 3 deg.Positioning request to Cine Radiography Angle / C-Arm Utilization at 3 degreesunknown
4AF5Bolus filters problemPositioner module detects bolus filter malfunctionerror
4AF6General positioner module software errorPositioner module software detects an internal error conditionerror
4AF7Emergency Stop depressedEmergency Stop Switch actuatedwarning
4AF8Gantry power or enable failureGantry problem detectederror
4AF9Gantry DMT (Lifetest) failureDMT cycle failed during gantry operationerror
4AFAPositioner Console FaultCommunications failure with Remote or Inroom Consoleerror
4AFBGantry Link FaultCommunications failure with Gantryerror
4AFCHardstop Bypass ErrorEmergency Stop is actuatederror
4AFDCollision recovery failedFirst Level Collision backout did not clear collisionerror
4AFFFilm changer in working positionFilm changer placed in working positioninfo
4B01Film changer collision detected on PUCK film changerCollision detected during film changer positioningerror
4B07Smart handle problemPermanent command detected on one axis of smart handle A or B for one minuteerror
4B10DUART status register error for character received on channel ACharacter received on DUART channel A with error statuserror
4B11Spot Film Device function unavailable due to DUART character reception errorDUART status register indicates error for character received on channel Berror
4B12Spot Film Device function unavailableSerial link to SFD panel is not active even after reset attempt; polling fails within timeouterror
4B13Spot Film Device function unavailableSerial link to Remote panel not active after reset attempt; polling timeout detectederror
4B14Unsolicited input received on serial linkCharacter received that looks like a command but is unexpected; can occur on power-up in pre-MP6 systemswarning
4B15Spot Film Device function unavailableSoftware corruption; invalid serial link identifier valueerror
4B16Room door interlock error - room door does not appear to be closedExposure request made with room door open or door signal indicating open statewarning
4B17Spot Film Device (SFD) not in forward positionPrep request initiated while SFD POS SW signal indicates SFD is not forwardwarning
4B18Object in front load portObject detected blocking front sensor(s) during fluoroscopy exposure requesterror
4B19Rear door open on Spot Film DeviceREAR DOOR SW signal reads low when exposure is requestedcritical
4B1ACassette not detected in Spot Film DevicePrep request made without cassette loadederror
4B1BCassette being loaded. One moment please.Exposure request while cassette is in middle of loading or unloadinginfo
4B1CInvalid format selected for current cassettePrep request with invalid format selectionwarning
4B1ESystem communications problemCommand received out of expected sequenceerror
4B1FSystem communications problemCommand received out of expected sequenceerror
4B20Obstruction may be in way of fluoro imageCarriage is stuck when fluoroscopy is requestedwarning
4B21Cone not parked at startupSystem startup when cone is not in valid position (Front or Park)error
4B2280188 Internal Timer faultDetected during system diagnosticsunknown
4B22Servo axis not maintaining proper positionOne of the servo axes jumps out of the position window and does not return within timeout period, or repeatedly jumps in and outerror
4B23Table locks non-functionalLock did not respond as expected: requested to set but did not, or set when not requestederror
4B24Power Assist non-functional due to thermal overloadPower Assist board generates OVERTEMP signal over Multibuserror
4B25Power Assist clutch has failed or is non-functionalPower Assist clutch either did not set when requested, set when not requested, or set but then failederror
4B26Cone button operation attempted with Horizontal Stop switch in OUT positionUser attempts to use Cone button while Horizontal Stop switch is OUTerror
4B27Invalid format selection for cassetteAttempting to select a format that is not currently valid for the cassetteerror
4B28Wrong size cassette detectedSFD unable to recognize loaded cassette; cassette may be jammed or invalid sizeerror
4B29Cassette currently in system is thought to be fully exposed and cannot be usedPrep request when internal flags indicate cassette is fully exposedwarning
4B2ASelected format has become invalid due to film depletionDuring Prep when currently selected format becomes invalid based on internal flags indicating film usagewarning
4B2BPatient Step or Overhead Obstruction Preventing AngulationAngulation requested while table step is extendederror
4B2DTable top motion non-functionalTable top longitudinal motion requested but no movement detected after timeout perioderror
4B2ECassette carriage jammedExpected servo motion did not complete in time; servo appears stallederror
4B2FTable at angulation limitAngulation limits reached or foot step extended; INTERLOCK signal true on Servo controller boarderror
4B30Collimator non-functionalOne or both collimator blades not at requested position for extended time perioderror
4B31Rear door on Spot Film Device is not closedRear door switch signal indicates door is open when carriage motion is to occurerror
4B32Object detected in front load portFront sensors detect cassette or object in front cassette opening; internal software confirms unload operation but object still presentwarning
4B33Invalid cassette type or positioner calibration errorRear load button pressed on 8835 positioner with 14x14 cassette loaded or carriage calibration mismatcherror
4B34Rear door lock failed to lock or unlock after repeated retriesSolenoid POS SW signal did not give proper lock indication within timeout perioderror
4B35Table stepper engaged; manual motion not permittedAngulation or lateral table motion requested while film changer not parked, or Film Changer mode selected with table horizontalerror
4B36Obstruction detected at frontFront sensor beam (IN FRONT CASS signal) detected as broken when carriage not at front or cassette already in systemerror
4B37CPU requires reset and system rebootPositioning module error conditionerror
4B3ATable top in illegal region (extended while angulated or beyond calibrated boundary)Table top feedback pot showing value outside calibrated positions for current table positionerror
4B3BTable motion non-functionalLongitudinal or angulation pot reads zero or full A/D converter valueerror
4B3CSpot Film Device non-functionalHandshake with servo controller board did not occur within ~100ms; servo controller did not toggle bit 0 of servo board status porterror
4B3DRear load door forced openRear door opened when not expected; REAR DR SW signal indicates opening without prior load button requestwarning
4B3ESpot Film Device non-functionalS188 unable to pass setup parameters to servo controller during initialization; read back data did not match passed dataerror
4B3FCollimator non-functionalSID or collimator blade feedback pot A/D value registers at zero or full scaleerror
4B40Spotfilmer servo non-functionalServo feedback A/D values register at zero or full scaleerror
4B41Cone not parked during table motionTable top motion requested while cone not parked in U.K. cone operation modeerror
4B48Step command requested when in external modeASC sends step command to table top while positioner is in film changer control modeerror
4B4BInjector did not inject as requestedInjector did not give an injecting indication within specified time period after INJ START DR signal assertederror
4B50Fixed mask servo failed to reach requested position within timeoutServo in position bit from servo controller not asserted within allotted timeerror
4B51Variable mask servo position timeout errorVariable Mask servo did not reach requested position within allotted timeout period; servo in position bit not asserted by servo controllererror
4B53Grid servo failed to reach requested position within timeoutServo in position bit from servo controller not asserted within timeouterror
4B54Lateral cassette servo failed to reach requested position within timeoutServo in position bit from servo controller not asserted within allotted timeerror
4B55Longitudinal cassette servo failed to reach requested position within timeoutServo in position bit from servo controller not asserted within timeouterror
4B56SFD locked in forward position interlock errorSFD in forward position at Prep time when not appropriate for selected functionwarning
4B57Cassettes detected in table/wall buckets during auxiliary or warmup mode exposure attemptRAD exposure attempted in AUX or Tube Warmup Mode with cassettes in table and/or wall bucky trayserror
4C00-4CFFPhoto Spot module Applications software error rangeError in Photo Spot module during Applications modeunknown
4C02Film Advance in processPhotospot module during film advance operationinfo
4C03Film Advance in processRemote film advance on console attempted while camera already doing transport due to manual pushbutton film advanceerror
4C04Camera shutter did not openPhotospot prep initiated; shutter commanded open by ASC but remains closed after timeout perioderror
4C05Math error in Photospot software, division by zero attemptedSoftware encountered division by zero condition in Photospot moduleerror
4C06Photospot camera inoperablePower amp board indicates a fault condition during pre-transport check with PAFAULT signal activeerror
4C07Magazine lever not closed; knife switch closed when it should be openCamera pre-check during transport or Prep operation detects knife switch signal indicating closed positionerror
4C08Photospot camera film supply depleted or camera not readyFilm remaining (FILM REM) A/D indicates low film AND camera signals not-ready during Prep phasewarning
4C09Photospot camera not loaded properlyCamera indicates not ready during pre-transport ready check; CAM RDY* signal indicates not ready statuserror
4C0AInstruction not understood. Please try again.GE bus transmission received and software cannot find command in its command tableerror
4C0BTakeup magazine is full and must be emptied to continueCamera indicates takeup magazine full; CAMERA RDY signal not ready; TAKE UP MAG OK and SUPPLY MAG OK in place; CARD OK present; FILM REMAINING STATUS > 2.5V; TAKE UP STATUS > 5Vwarning
4C0CTakeup magazine full and supply magazine empty condition detectedCAMERA RDY* signal shows not ready; takeup magazine A/D (TAKE UP STATUS) > 5 volts; supply magazine A/D (FILM REMAINING STATUS) < 2 volts; both magazines in placewarning
4C0DNo patient ID card inserted in cameraCamera indicates missing patient ID card detected via CARD OK signal at Prep timeerror
4C0ESupply magazine not in placeCamera detects SUPPLY MAG OK signal absent at Prep timewarning
4C0FTakeup magazine not in placeTAKEUP MAG OK signal on camera status port indicates magazine absent at prep timeerror
4C10Camera not ready: missing takeup magazine, full magazine, or missing patient ID cardCamera not ready at Prep when film supply A/D > 2 voltserror
4C11Supply magazine needs to be reloaded or magazine not in placeCAMERA RDY* signal not ready AND FILM REMAINING STATUS A/D <2 volts (<30 frames)error
4C13Supply magazine emptyCAMERA RDY* signal indicates not ready AND FILM REMAINING STATUS A/D reads less than 2.5 volts AND TAKEUP STATUS A/D is less than 5 volts AND patient ID card is in placeerror
4C14Camera did not respond before timeout expiredCamera did not generate expose to generator signal before timeout, EXP TO GEN signal not true on camera status port despite CAMERA RDY* signal indicating readyerror
4C15Camera did not respond to expose command before timeout expiredCamera did not release expose to generator signal or EXP TO GEN signal did not go low or stay low after close shutter commanderror
4C16Film transport faultCamera did not complete transporting film within one second from start of transporterror
4C30System communication problem. Photospot module failed to receive transmit acknowledge within timeout.Module attempted to transmit packet to node; no acknowledge before 80188 timer 2 timeout.error
4C31Improper updating of the packet indexer; indexer went past end of packet after processing command because packet had less data than expectedPacket indexer processing error in Photospot moduleerror
4C32Software attempted to transmit more data than would fit in a single packetData transmission exceeds packet size capacityerror
4C33GE Bus transmit queue full, module software lost or ARCNET not transmittingSoftware's internal GE Bus transmit queue was full when module attempted to send packeterror
4C34ARCNET hardware could not find designated node on system to transmit packetTransmitter available before transmission timeout but no transmit acknowledge givenerror
4C35Internal timer indicator allocated to different scheduled task during software task schedulingSoftware attempted to schedule a task and internal timer was already in use by another taskerror
4C40System timer task queue full; instruction not understoodSoftware attempted to schedule a task on system timer with no available queue spaceerror
4C41Instruction not understood. Please try again.Attempt to update or flush a scheduled task with timer ID outside list of scheduled taskserror
4D00System communications problemARCNET transmission failed; destination module did not acknowledge reception within 250 millisecondserror
4D00-4DFFPulscan module Applications software error rangeError in Pulscan module during Applications modeunknown
4D01GE Bus command interpreter packet index overrunNumber of bytes transmitted in a packet does not match expected numbererror
4D02System communications problemPacket transmission attempted exceeds maximum transmittable sizeerror
4D03GE Bus output queue is fullSystem attempting to send data on GE Bus when queue at capacityerror
4D04System communications problemARCNET transmission attempted to non-existent node (TA set, TMA not set)error
4D05System communications problemCommand received across GE Bus (ARCNET) that is not a defined command for Pulscan moduleerror
4D06Bad data received from consoleAttempt to schedule a task when no timers availableerror
4D07Bad data received from Pulscan moduleAttempt to flush or set a soft timer greater than configured system timerserror
4D08Bad data received from consoleAttempt to schedule a time-delayed task through SCHEDULE_SOFT_TIMER_TASK using a timer already allocated to another taskerror
4D0FSystem communications problemCommunications failure in Pulscan moduleerror
4E00Bolus filter errorBolus filter malfunction or misalignmenterror
4E00-4EFFAFM/CFM positioners Applications software error rangeError in AFM/CFM positioners during Applications modeunknown
4E01General PVC console errorPVC circuit malfunction detected by Positioner moduleerror
4E02Gantry problemGeneral error with the gantry system detectederror
4E03M200 Table is disengaged. Remote control not allowed.Table top is not engagedwarning
4E04General M200 Table errorPositioner table malfunction detectederror
4E05II collision occurredPositioner mechanical collision detectederror
4E06Collimator collision occurredCollimator arm strikes patient or table during positioningerror
4E07General error with the tableTable malfunction detectederror
4E08Emergency stop occurredEmergency stop detection in Positioner modulewarning
4E09Gantry not in anteroposterior (AP) positionGantry position sensor detected gantry not positioned for AP imagingerror
4E0ACollimator errorCollimator mask not installed or collimator system malfunctionerror
4E0BCollision CHNGRPositioner collision detected during operationerror
4E0CCollimator mask installed when FLUORO mode requires it to be removedUser attempts FLUORO operation with collimator mask installederror
4E0DFilm changer direction to X-rayOnboard film changer moves from parking to working positioninfo
4E0EFilm changer direction to head positionOnboard film changer moves from Working to Parking head positionunknown
4E0FFilm changer moves from Working to Parking Foot positionOnboard film changer direction command to Foot positioninfo
4E10Request cannot be executed with film changer in this positionFilm changer positioned incorrectly for requested operationerror
4E11Gantry Angulation not allowed for Film Changer motion.Attempt to perform film changer interchange at current gantry angulationwarning
4E12Interchange not allowed in AP angulation if table is not on topAttempted AP angulation with table not in top positionwarning
4E13Gantry is movingGantry movement detectedinfo
4E14Film Changer in middle of travel; no gantry motionFilm changer positioned at middle of travel during operationerror
4E15Film Changer Collision RiskDuring film changer interchange operationwarning
4E16Film Changer Interchange CompleteFilm changer completes interchange cycleinfo
4E17Lift collision detectedCollision between lift and mechanical structureerror
4E18Film changer not correctly lockedFilm changer position incorrect or locking mechanism failederror
4E19Film changer not in work positionFilm changer is not positioned correctly for operationerror
4E1AGantry not centeredGantry position monitoring system detects misalignmentwarning
4E1BTable positioning error in AP modeSystem detects table is not fully backward before AP interchangeerror
4E1CFilm changer is not in park positionType II bad interlock detected; film changer not at park positionerror
4E1DTable is moving to the Start PositionDuring DA / DSA stepping acquisition when table transitions to start positioninfo
6000Floppy Disk errorDisk hardware failureerror
6002Floppy disk error - volume granularityFloppy disk read/write error; volume allocation unit errorerror
6003Floppy Disk errorDevice granularity errorerror
6004Floppy Disk errorFloppy disk read command failureerror
6005Floppy Disk error - software module errorASC (Floppy Disk Code Module) fails to process file - no TEND on filenameerror
6007Floppy disk error: filename too longSoftware attempted to access file with name exceeding length limiterror
6008File not foundFloppy disk code module cannot locate required fileerror
6009Floppy disk error - pointer pair overflowFloppy disk operation or self-test failurecritical
600AFloppy disk error; VOL block not foundFloppy disk communication failureerror
600CFloppy Disk errorEntry point not found in floppy diskerror
600DRuntime database error in NV RAM fil correction valuesWarm start when NV_G_fil_correction_X does not equal NOT (NV_G_fil_correction_not_X)warning
600ERecommended X-ray tube Preventive MaintenanceFilament correction values increased more than 3B$0000H from calibration baseline at warm startwarning
6010Fastboot functionality not operationalFastboot ROM or boot process failurewarning
6011Floppy disk error - incorrect data countFloppy disk read/write data mismatcherror
6020SPROM access error in ASC moduleINSITE Electronic Device ID software failure to read/access ASC i2c devices or bad device programminginfo
6021Board ID: Elapsed Hours On Checksum ErrorINSITE Electronic Device ID software failure to read/access ASC i2c devices or Advantx-E software finds bad programmingwarning
6100-67FFASC module Applications software error rangeError in ASC module during Applications modeunknown
6110Room deselect attempted during fluoro exposuresA room deselect is attempted during fluoro exposureswarning
6111Room control module inoperable; fluoroscopy cannot proceedAttempt to perform fluoroscopy exposure when Positioner module is offline, misconfigured, or not respondingerror
6112Exposure control module (ECM) not responding or not configured.Operator attempts fluoroscopy exposure when ECM is offline or unresponsive.error
6113Video control module inoperable; fluoroscopy cannot be performedVIC not configured online or no reply received from VIC to RT Bus driver command during fluoro exposure attempterror
6116Video Tape Recorder communication failureNo reply received from VIC to VTR status commanderror
6117Optical distributor did not respondAperture in place reply not received from VIC module when fluoro exposures attemptederror
6118Optical filter did not respondImage gate positioned reply not received from VIC module during fluoro exposureserror
6119VTR being used for playback, unable to recordVTR is not in stopped state during fluoro or record exposurewarning
6120Performing camera auto zero. Please wait.Operator presses horizontal reverse button at consoleinfo
6121Pulscan inoperative. Reverted to next available FL mode. Call service.Pulscan not configured, initialization incomplete, or no fluoro exposures due to Pulscan offline or bad reply to status/RT Bus commanderror
6122Selection cannot be made during acquisitionFluoro operation mode change is attempted while taking exposureswarning
6123Digital Record inoperative in fluoroscopy modeDigital Record not marked online, not configured, or bad status reply receivederror
6124Digital Record module busy; no last image hold will occurReply received from Digital Record when attempting Digital Record fluoro exposures indicates Digital Record is currently busywarning
6125Cannot do Progressive Fluoro, DX not availableAttempting DX fluoro exposures when no fluoro configure reply received from DXerror
6126Cannot do Progressive Fluoro; DX offline or not ready for acquisitionsAttempting DX fluoro exposures when DX is offline or unavailablewarning
6127Horizontal sweep reversal already in progressOperator presses horizontal reverse button while previous reversal still in progresswarning
6128Lateral FLUORO VIEW not allowed with DIGITAL FLUORO selectedFluoro operation mode changed to DX fluoro with lateral fluoro view, or fluoro view changed when DX fluoro is activewarning
61F9VTR did not start recordingNo reply received to VTR record command sent for HPR-50warning
61FANo run data saved, acquisition less than 3 secondsLess than 3 seconds of tape has elapsed since the beginning of recordinginfo
61FBMark frame outside current run rangeOperator attempts to mark frame on tape section outside current run selectionerror
61FCNo run data to search onAttempt to mark a frame, select a run number, or select a mark when VTR current run directory is emptyinfo
61FDNo mark saved for this runOperator attempts to select a mark for a VTR run with no stored mark positioninfo
61FEInformational message displayed when VTR Reset button pressedOperator presses VTR Reset button on consoleinfo
6202Injector not readyPrep switch depressed when injector is not in ready statewarning
6203Table not readyPrep switch depressed when positioner/table not in ready statewarning
6212Stepper selection mismatch between console and tablePrep switch depressed with stepper selected at table but not at consolewarning
62CAmA overload occurredmA overload condition during X-ray exposurewarning
6304Optical distributor did not respond during Cine exposure attemptImage Gate positioned reply not received from VIC module when Cine exposures attemptederror
6306Optical filter did not respondNeutral Density Filter positioned reply not received from VIC module during Cine exposureserror
630ACine camera not line lockedError status reply from Cine module in response to RT Bus receivers configure commandwarning
630EAuto stabilize sequence timeout20 second timeout on start of auto stabilize sequence reachederror
630FCine module offlineCine mode select performed or record exposures attempted in Cine, DF Dynamic, or DX Cardiacerror
6310Cine mode unavailable due to video control module inoperableVIC module found offline when cine mode selected or record exposures attempted in Cine, DF Dynamic, or DX Cardiacerror
6311Positioner module offline when cine mode selected or cine exposures attemptedCine mode select performed or cine exposures attempted with positioner module offlineerror
6312Cine mode unavailable due to ECM module offlineCine mode select performed or cine exposures attempted while ECM offlineerror
6314Injector not readyPrep switch depressed when injector status indicates not readywarning
6315VIC Pedestal Servo did not respondNo response received from imaging module to black level servo command when cine record acquisition attemptederror
6316Pulscan module offline during cine exposurePulscan module found offline when cine exposures using Pulscan fill-in attemptedwarning
6318Opposite plane acquisition sequence timeoutTimeout on opposite plane response reached when cine exposures attemptederror
632AMaximum single plane run time for a biplane acquisition exceededBiplane acquisition run duration exceeds system limitswarning
63FFCine camera has less than 20 feet of film remainingFilm update packet received from Cine module at end of exposures when remaining film drops below 20 feetwarning
6401Bad data received from Positioner during prep done timeout10 second timeout on receiving prep done when attempting record exposures (except Cine) because Configure for Acquisition command reply from Positioner not received or was baderror
6402Bad data received from VIC module during exposure preparation10 second timeout reached while waiting for prep done signal when attempting record exposures (DF Static, DF Dynamic, DX Vascular or DX Cardiac)error
6403Bad data received from Positioner; timeout on interlock status10 minute timeout reached when attempting record exposures while waiting for Positioner Interlocks command replyerror
6404Timeout on receiving exit done at end of record exposuresReply to expose sequence disable command from ECM not received within 1.5 secondserror
6405Exposure control inoperableASC waits 20 seconds for ECM compliance to select tube and room; no response receivedcritical
6406Bad data received from module nodeModule node found offline during communications (record acquisition, dx offsets, film speed change, temporal value change)error
6407Wait for Trial Done TimeoutTimeout during trial executionerror
6408Mode mismatch between console and table positionerAttempting to record exposures when selected mode at console does not match table type from positioner auxiliary mode packeterror
6409Expose time packet not received from Positioner during Tomo record setupAttempting to initiate Tomo record exposures without exposure length parameterswarning
640ABad data received from variable exceeding mathematical function boundariesVariable exceeds boundaries in application generic software set during console operationerror
640BBad data received from application softwareDividend of zero passed to divide routine in generic software seterror
640CBad data received from action-state driver routineInvalid value passed into action-state driver in generic software seterror
640DCannot use Stepper while CC selectedOperator attempts to use stepper function while CC mode is selectederror
640ENo auxiliary mode packet received from PositionerAttempting to record exposures for Rad and Tomo modes without mode packet from Positionererror
640FProper auxiliary mode packet not received from Positioner for Tomo acquisitionsAttempting Tomo record exposures without proper table preparationerror
6411Invalid action-state driver routine value for current application stateA value was passed into the action-state driver routine which was invalid for the current state in the application generic software seterror
6412Cannot switch filaments during acquisitionAttempt to change focal spot during fluoro or record exposureswarning
6413Cannot switch patient size during acquisitionAttempt to change patient size during fluoro or record exposureswarning
6414Cannot switch AEC mode during acquisitionAttempted change of record mode between AEC and fixed during fluoro or record exposureswarning
6415Bad data received from temperature to energy conversion routineBulk temperature greater than 2000 degree maximum passed into conversion routine used for sequence prediction calculationserror
6416Bad data received during sequence prediction calculationBulk energy value exceeding maximum passed to energy-to-temperature conversion routineerror
6417Bad data received from Digital Record module during exposure attempt10 second timeout reached when waiting for Digital Record prep done confirmation during Digital Record exposureerror
6418Bad data received during Photospot prep sequence10-second timeout when Photospot module fails to send prep done signalerror
6501Film remaining value from Photospot module outside valid range for configured camera typeInvalid photospot module data received by consoleerror
6601Invalid addresses calculated by AUP software for AMB utilities accessAUP database corruption, AUP memory board failure, communication loss, or AMB memory faulterror
6601Communication failure with auto protocol memory boardAUP software attempted to access AUP memory boarderror
6602Memory or procedure backup initialization/restore errorOccurs at warm start or during SAVE/RETRIEVE operationerror
6602Procedure database version is not compatibleWarm start with mismatched AUP database version or restore from incompatible backup floppyerror
6603No procedure present, cannot perform requested operation.No procedure has been selected.error
6604Procedure menu is full with no space to add new dataAttempt to add procedure when memory capacity reachederror
6605Procedure data base version is not compatibleAUP data base version mismatch at warm start during PED modecritical
6606Cast editing requested while another room is editingSoftware attempts to edit cast/auto-protocols when another room is performing cast editinginfo
6607Procedure edit must be run; procedures cannot be accessed until NV RAM compatibility check completesOperator presses PROCEDURE MENU during auto-protocol (AUP) mode when ped_required flag is seterror
6608No AEC data, cannot select it as defaultAttempt to select AEC as default when no AEC data has been programmedwarning
6609No fixed data has been programmed.Attempt to select a default protocol when no fixed data existserror
660ACategory full, cannot perform requested operationAttempt made to add procedure to full category in Procedure Edit moduleerror
660BMaximum receptors reached, cannot perform requested operationAttempt to add another receptor when procedure already has maximum number assignedwarning
660COther room currently using SAVE/RESTOREAnother room is in the save/restore screeninfo
660DSystem floppy in drive, cannot perform requested operationAttempted save/restore operation with Advantx software floppy in driveerror
6610Frontal/Lateral database mismatchProtocol database inconsistency between frontal and lateral imaging mode protocolserror
6611No unique receptors, cannot perform requested operationAttempt to copy receptors to a procedure that already contains all requested receptorswarning
6612Autoprotocols unavailable. Continue and call service.At power-up, AUP software fails to write and read back test patterns to reserved test location in AUP database, or data read back does not match data written.error
6613Autoprotocols unavailable; AUP software task creation failed on frontal biplane system at power upPower-up initialization; error encountered during RMX task creation by AUP software on frontal half of biplane system for lateral database read/write requestserror
6614A protection key floppy is required to edit proceduresAttempt to edit procedures without protection key floppy installedinfo
6680Media error, insert new disketteFailure to properly read floppy diskerror
6681Disk error - floppy drive failureFailure occurs with floppy drive during operationerror
6682Floppy disk is write protectedAttempt to write to a write-protected floppy diskwarning
66C0Communication failure while communicating with the other plane in a biplane systemASC (Procedure Edit Auto-protocols) module loses communication with opposite planeerror
6700Cannot switch receptors during an exposureAttempt to switch receptors at console or spot film device during fluoroscopy or record exposureswarning
6701Tube conditioning in progressAttempt to switch receptors at console or spot film device during tube conditioningwarning
6702Instruction not understoodSoftware detected execution pathway designed to be unreachableerror
6703Application not loadedApplication file configured for system did not upload properly from floppy diskerror
6704Receptor not available due to floppy disk read errorApplication file not properly uploaded from floppy disk during operator selectionerror
6705Lateral application not found. Reconfiguring as single plane.Application file configured as biplane did not upload properly to lateral ASC from floppy diskwarning
6706Application not loadedTube warmup application file did not upload properly from floppy diskerror
6707FastBoot auto return failed; default application mode and parameters restoredNVRAM failure, system halt during runtime snapshots, or biplane plane mismatchwarning
6708Biplane communication error: frontal failed to send packet containing biplane-common values to lateralIn biplane systems, communication error occurs during packet transmission of common values between frontal and lateral unitserror
6800Communication problem between console and DF (digitalfluoroscopy) modulePacket from DF not received within allotted timeerror
6801Communication problem between console and DF modulePacket from DF not received within allotted timeerror
6802DF module status indicates a fault conditionLast DF module status received indicates a faulterror
6803Communication problem between DF module and Advantx softwareDF module software and Advantx software not in same acquisition mode (e.g. DF in Static, Advantx in Dynamic)error
6804DF module not ready for acquisitionsDF module initialization fails or acquisition prerequisites not metwarning
6805Insufficient disk space for image acquisitionStatic sequence requires more image disk space than availableerror
6806Disk space available is less than 2 seconds of run timeDynamic acquisition attempted with insufficient disk spacewarning
6807Image disk directory is full (contains 100 images)Attempting to acquire or store images when disk reaches capacityerror
6808DF service mode exposure timing errorAttempted zero millisecond exposure in DF service modeerror
6902Collimator blades open during tube warmupTube warmup sequence entered with manually controlled collimator blades openwarning
6903Non-volatile database checksum mismatchChecksum validation fails at system startupcritical
6904Non-volatile database checksum is wrongSystem startup when NV database checksum does not match stored valueerror
6905Warmup inactive while two soak cycles in progressTube change attempted while on tube warmup screen with two soak timers activewarning
6906Cannot perform tube conditioningDuring tube warmup startup, selected focal spot value set using initialized MA value found in errorerror
6A00Digital Record New Study annotation not enteredDigital Record exposures attempted without New Study annotationwarning
6A01Disk space remaining is less than 15 framesDigital Record exposures attempted when available disk space drops below 15 framesinfo
6A02Digital Recorder module busy; no last image hold will occurDigital Record exposures attempted while Digital Recorder in busy statewarning
6A03Disk full. Remove images to continue.No more frames of storage available on Digital Disk during Digital Record exposures.warning
6A04Maximum number of images have been stored in this studyAcquisition attempted to append images to a study that contains 99 images (maximum allowable)info
6A05Digital Recorder temperature too highTemperature sensor on MHR image processor detects overheating conditionerror
6A06Images cannot be appended to this study at this timeAcquisition attempted on a study not selected for APPEND modeerror
6A07No current study exists on DRSAcquisition attempted when no study selected or new study information not enterederror
6A08Communication problem encounteredAcquisition attempted when XDIU was not functioningerror
6B00Number of images exceeds disk spaceVascular sequence requires more disk space than currently availablewarning
6B01Maximum number of patients on digital system has been reachedDX image disk cannot store additional imageswarning
6B02The maximum number of sequences on digital system has been reached.DX image disk cannot store additional images.warning
6B03Digital system offsets failedDX video adjustment process did not complete successfullyerror
6B04Digital system offset increment is enabledDX service mode left noise measurement functionality enabledwarning
6B05Digital system offset adjustment in process (console is locked)DX video adjustment process initiatedinfo
6B06Digital system offset adjustment has completed successfullyDX video adjustment process completioninfo
6B07System communications problemPacket from DX module not received within allotted timeerror
6B08Communication problem between console and DX code modulePacket from DX not received within allotted timeerror
6B09Digital system is not in READY stateDX module fault indicated by last DX status receivederror
6B0AOffsets failedDigital offset values could not be applied or re-calibratederror
6B11DX module not ready for acquisitionsDX system fails to enter ready state for image acquisitionerror
6B12Disk space insufficient for cardiac exposure sequenceLess than 2 seconds of image storage availablewarning
6B13Communication problem between console and DX Code ModuleDX packet not received within allotted timeerror
6B14Digital system offset failed to complete within the allowed timeDX video adjustment process does not finish within allotted time windowerror
6B15AEC exposure time is less than the suggested minimumTrial or mask AEC exposure resulted in exposure time less than five millisecondswarning
6B16Film record not allowed at this frame rate. Record Medium set to Disk.Attempt to record film at two lowest frame rates in Cardiac modeinfo
6B17Selection of DENS QUANT with Record Medium Film is not allowedOperator selects Dens Quant with film-only record medium in Cardiac when DX software version is prior to A4warning
6B18Trial expose time too longTrial exposure results in expose time greater than half of trial backup timewarning
6B19Frame rate too slow to use cine cameraCine camera operation with insufficient frame ratewarning
6B1ABiplane acquisitions not possible in TECH MENUAttempted biplane acquisition while in TECH MENU modeinfo
6B1BFrontal AEC expose time is less than the suggested minimumFrontal trial or mask AEC exposure resulted in expose time less than five milliseconds during biplane acquisitionwarning
6B1CLateral AEC expose time is less than the suggested minimumLateral trial or mask AEC exposure resulted in expose time less than five milliseconds during biplane acquisitionwarning
6B1DBoth AEC expose times are less than the suggested minimumBoth frontal and lateral trial or mask AEC exposures result in expose times less than five milliseconds during biplane acquisitionwarning
8100An MTE error has occurredMTE (Module Test Error) condition detectederror
8100Hardware Status 0900—1000MTE error occurrenceerror
8100MTE (Module Test Error) has occurred with bad data receivedHardware Status 0000–0800 conditionerror
8100Hardware Status error in range 0000-0800, MTE (Multi-Tiling Engine or similar) error with bad data receivedCorrupted or invalid data received by systemerror
8101RMX system error occurredRMX error detected in systemerror
8101-81FFSLOS (System Level Operating System) error rangeError in Diagnostics mode by SLOSunknown
8102Non-Volatile Database CorruptedInvalid database checksum detectederror
8103System programming error: invalid parameter passed to Clear Call Table utilityProcedure using Clear Call Table utility called with out-of-range parameterserror
8104Database Error: Incompatible Database VersionNon-Volatile Database major revision does not match software floppy release numbererror
8105Database Error: Default Override Switch ActiveNon-Volatile Database Checksum Override Switch is set on ASC BIUinfo
8180Transmitter Not Ready. Test AbortedGE bus transmit procedure cannot obtain ARCNET transmit token; SLD command not senterror
8181Node response timeout during GE bus transmissionASC sends SLD command to node; bus transmit times out before node respondserror
8182Node Not Responding. Test AbortedGE bus transmit procedure cannot communicate to the receiving node; SLD command not senterror
8183Node error during communication with receiving nodeTest Sequencer receives node error while communicating with receiving nodeerror
8184Invalid packet received during inter-node communication; test abortedTest Sequencer receives malformed or corrupt data packet from receiving node during test executionerror
8185Node is off-line. Test aborted. Node communication problem.Test Sequencer received a node off-line error while attempting to communicate with a nodeerror
81A0Floppy read error, boot information not foundError occurred while attempting to read Third Party data file from floppyerror
81A1Floppy not initialized properlyThird Party data file was not initialized properlyerror
81A2Floppy incompatible with systemThird Party data file key checksum does not match non-volatile checksumwarning
81A3Invalid system identificationThird Party data file ID checksum does not match non-volatile checksumerror
A000-A0FFSMOS (Service Mode Operating System) error rangeError in Service Mode by SMOSunknown
E-6001Floppy Disk errorUnknown disk format in floppy driveerror
Error 4026System error from Applications Operating System (APOS) codeGE bus transmitter is not availablecritical
Error 4051System communications problemASC module communication failure detectedcritical
Error 4066System was reset from the consoleinfo
Error 4071System communications problem in ASC module running APOS codeBad node ID received over communications linkerror
Error 4072System communications problemAPOS code in ASC module detects node errorerror
Error 4202Power unit interface errorPower unit error condition detected during system operation or recovery sequenceerror
Error 4226Database error: ECM received calibration or configuration data out of range.ECM receives cal/config data with one or more parameters exceeding valid range during initialization or operation.error
Error 4246ECM and power unit communications error or module initialization failurePower-up, re-initialization failure, or DIP switch misconfiguration causing diagnostic interferenceerror
Error 4247ECM receiving unused error value from power unit or misinterpreting normal error valuePower unit sending unexpected error value or ECM firmware interpreting normal value incorrectlywarning
Error 4289STARTWHEN timeout on RT bus 10Particular signal or state not achieved within recipe-specified time; recipe index displayed in HW STATUS fielderror
Error 4293Start when timeout, timer 13A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipeerror
Error 4403Digital system error, acquisition terminatedDigital system fault detectionerror
Error 4509Serial link framing error detected on DR BIU processorSerial link Duart (U133) detects framing error in character reception from DR Moduleerror
Error 4784Optical distributor not positionedND Filter amp indicated stall at bad timeerror
Error 4A56Film changer not parked at system prep time; interlock prevents continuationFC PARKED signal indicates film changer not in park position during preperror
Error 4B23Table locks non-functionalLock feedback indicates incorrect state after timeouterror
Error 6006Floppy disk error in ASC moduleFilename not specified during floppy disk operationerror
Error 6108Neutral density filter positioning reply not received from VIC moduleFluoro exposure attempted while VIC module offline or filter reply not receivedwarning
Error 6600Unrecognized commandAUP software receives command packet from other applications that does not match recognized command formatinfo
Error 660EFloppy drive door left open during save operationFile save attempt with drive door openwarning
Error 660FFloppy drive door was left openAttempted to retrieve data with floppy drive door openerror
Indeterminate (watchdog error)Watchdog timer timeout indicating serious software or electronics errorWatchdog circuit detects failure during normal operation, at power-up, or during shockcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.