GE Healthcare ADVANTEX: troubleshooting and error codes
The ServiceForge knowledge base holds 261 documented symptoms for the GE Healthcare ADVANTEX, with 555 ranked causes, 510 diagnostic checks with test points and expected values, and 986 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Cone servo fails to reach requested position
Error 4B52 displayed on console and in-room display shows 'CHECK FOR OBSTRUCTION'. The Cone servo did not reach the requested position within the allotted timeout.
Ranked causes
- 1. Servo controller timeout - servo in position bit not asserted (common)
- 2. Feedback potentiometer failure (occasional)
- 3. Servo amplifier board malfunction (occasional)
- 4. Motor not moving (occasional)
- 5. Servo belt improper tension (occasional)
- 6. Servo damping or positioning calibration error (occasional)
- 7. R+F Positioner module firmware or memory issue (rare)
- 8. 80188 CPU timer malfunction (rare)
Run Cone Axis Test diagnostics
- Test point
- R+F Servo Controller Board in R+F menu
- Expected
- Test passes without error
7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 420B displayed
System displays error code 420B during fluoroscopy operation
Ranked causes
- 1. Temporary grid short condition (occasional)
- 2. Intermittent grid shorts (occasional)
- 3. Tube spitting causing filament correction drift (occasional)
- 4. Invalid calibration database (rare)
- 5. ECM or Power Unit module fault (rare)
- 6. Power Unit system level fault (rare)
- 7. ECM system level fault (rare)
Release expose switch and attempt exposure
- Test point
- Expose switch
- Expected
- Error 420B does not reoccur
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Variable mask servo position timeout
Console displays 'Retry selection, cycle console power or call service'. In-room display shows 'RETRY OR CALL SERVICE'. System fails to move Variable Mask to requested position within timeout.
Ranked causes
- 1. Feedback potentiometer malfunction (common)
- 2. Servo amplifier board failure (common)
- 3. Motor not responding to command (occasional)
- 4. Servo belt slack or misaligned (occasional)
- 5. Servo calibration drift (occasional)
- 6. R+F Positioner module hardware fault (rare)
- 7. 80188 CPU timer malfunction (rare)
Run Variable Mask Axis Test
- Test point
- R+F Servo Controller Board
- Expected
- Test passes without error
9 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament drive error during exposure radiation
Filament drive error detected; system initiates auto-reset; exposure unable to complete
Ranked causes
- 1. Power unit board or component failure (common)
- 2. Filament control voltage or demand issue (occasional)
- 3. Filament calibration data error (occasional)
- 4. Filament demand exceeds limits (occasional)
- 5. Filament inverter board malfunction (occasional)
- 6. Closed loop system or cable issue (occasional)
Monitor filament demand voltage at J1-a13
- Test point
- J1-a13 on i/o board 46-264024
- Expected
- 5V when error code generated
- Tools
- Voltmeter
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optical distributor not positioned high voltage
Console displays 'Optical distributor not positioned. Try again or call service.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' Gate rotate failed to position during initialization or normal move.
Ranked causes
- 1. Gate rotate electrical in position bit not received within 2 seconds at initialization (common)
- 2. Both electrical and mechanical in position bits not lost within 100 milliseconds during move (occasional)
- 3. Gate & NDF Control Board failure (rare)
- 4. Gate & NDF Interface Board failure (rare)
- 5. XRII Interface Board failure (rare)
- 6. Gate Sensor failure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optical distributor positioning error
Console displays 'Optical distributor not positioned. Try again or call service.' In-room display shows 'TURN CONSOLE POWER OFF/ON.'
Ranked causes
- 1. ND Filter amp indicated stall at bad time (unknown)
- 2. Gate & NDF Control Board failure (unknown)
- 3. Gate & NDF Interface Board failure (unknown)
- 4. XRII Interface Board failure (unknown)
- 5. Gate Sensor failure (unknown)
- 6. Gate Mechanics failure (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Positioner not in valid Cerebral Contact position patient safety
Interlock message displayed at Prep; CC button flashing; film changer not meeting valid position requirements for CC mode
Ranked causes
- 1. Film changer distance exceeds 30 cm ISO distance requirement (common)
- 2. Film changer not at proper end of travel position for L/U angle combination (common)
- 3. CC mode not properly enabled or CC switch signal not detected (occasional)
- 4. L or U angle display incorrect for actual positioner position (occasional)
- 5. Hardware failure in angle sensors or switches (rare)
- 6. Firmware or power state issue (rare)
Measure ISO distance of film changer from reference
- Test point
- ISO distance measurement point
- Expected
- <= 30 cm
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power unit tube spit recovery error high voltage
Console displays 'Power unit tube spit recovery. Continue'; in-room display shows 'POWER UNIT ERROR. CALL SERVICE.' kV is automatically cut off, automatic retry performed. After 7 spits during same exposure, exposure aborts with error 42CA.
Ranked causes
- 1. Tube not formed (new tube) (common)
- 2. Excessive kV or mA settings (common)
- 3. Tube degradation (old tube) (occasional)
- 4. HV tank problem (occasional)
- 5. Measurement/switching board issue (occasional)
- 6. Divider board calibration problem (rare)
Determine if tube is new and not formed
- Test point
- Tube status
- Expected
- New tube not yet formed
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System communications problem
Console displays 'System communications problem. Please try again.' In-room display shows 'PLEASE TRY AGAIN.' Module attempted transmission but did not receive acknowledge before timeout.
Ranked causes
- 1. Destination ID invalid or node IDs set incorrectly (common)
- 2. Destination node is off or not receiving (common)
- 3. GE bus cables damaged or disconnected (common)
- 4. CPU RAM corrupted (occasional)
- 5. BIU board failure (occasional)
- 6. Applications floppy corrupted (occasional)
Verify destination ID is valid for system
- Test point
- System configuration
- Expected
- Destination ID matches valid system node
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Camera did not respond high voltage
Console shows 'Camera did not respond. Try again or call service.' In-room display shows 'CAMERA DID NOT RESPOND.' System times out waiting for expose to generator signal.
Ranked causes
- 1. Expose to generator signal not reaching Power Amp board (common)
- 2. PREP/EXPOSE signal not reaching camera (common)
- 3. PCM CPU board timer or PIC failure (occasional)
- 4. PHOTOSPOT module RAM/Arcnet failure (occasional)
- 5. System hardware failure requiring restart (rare)
Trace expose to generator signal path
- Test point
- Camera to Power Amp board
- Expected
- Signal present throughout path
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Camera shutter does not open radiation
Photospot prep fails; console displays 'Camera shutter did not open. Try again or call service'; in-room display shows 'TRY AGAIN. CAMERA FAULT.'
Ranked causes
- 1. Shutter mechanism stuck or not moving freely (common)
- 2. SHTR OPEN* signal not reaching shutter properly (occasional)
- 3. SHTR OPEN CMD* signal not being sent or shutter motor voltage insufficient (occasional)
- 4. CPU or timer hardware failure (rare)
- 5. Camera Shutter test failure (occasional)
Observe shutter movement at camera
- Test point
- 105 camera shutter
- Expected
- Shutter opens smoothly and moves freely
- Tools
- Visual inspection, listening
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Console receives bad data from ASC
Console displays 'Bad data received. Turn console power OFF/ON.' message; in-room display shows 'TURN CONSOLE POWER OFF/ON.'
Ranked causes
- 1. Invalid application command in GE bus packet from ASC (unknown)
- 2. Console CPU board RAM failure (unknown)
- 3. ASC CPU board RAM failure (unknown)
- 4. ASC GE bus interface (Arcnet) failure (unknown)
- 5. Corrupted system boot disk (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 4A5F displayed on console and in-room
Console displays 'Bad data received. Turn console power OFF/ON.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' CPU detected math division by zero error in Positioner module.
Ranked causes
- 1. Division by zero in Positioner CPU (unknown)
- 2. Hardware motion or potentiometer calibration issue (occasional)
- 3. Positioner module firmware or RAM failure (occasional)
- 4. S188 CPU Board defect (rare)
- 5. Application firmware corruption (rare)
Power cycle system
- Test point
- Main power switch
- Expected
- System restarts and reloads firmware without error
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid addresses sent to AMB utilities
Error 6601 indicates AUP software is calculating invalid memory addresses, typically due to corrupted AUP database values or AUP memory board failures
Ranked causes
- 1. AUP database corruption (common)
- 2. AUP memory board not securely attached or low battery voltage (occasional)
- 3. AUP memory board failure or AMB memory corruption (occasional)
- 4. ARCNET communication failure (biplane systems) (occasional)
- 5. ASC module hardware failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OCCLUSION or UPSTREAM OCCLUSION alarm with door closed patient safety
OCCLUSION or UPSTREAM OCCLUSION alarm triggered when pump door is closed and START key pressed.
Ranked causes
- 1. Tubing mispositioned or not seated in sensor (common)
- 2. Obstruction upstream or downstream of pump (common)
- 3. Wrong administration set type or code (occasional)
- 4. Ambient or solution temperature below 60°F (occasional)
- 5. Pressure sensor contaminated with residue (common)
Verify tubing position and seating
- Test point
- Tubing in sensor and safety clamp
- Expected
- Tubing properly positioned and seated
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Photospot camera not loaded properly caution
Console displays 'Photospot camera not loaded properly. Reload to continue.' In-room display shows 'FILM NOT LOADED PROPERLY.' Camera unable to transport film.
Ranked causes
- 1. Film path loops incorrect or film pan misaligned (common)
- 2. Camera transport motor failure (occasional)
- 3. Camera ready signal line fault (occasional)
- 4. PHOTOSPOT module PRD test failure (rare)
- 5. System power or reset issue (occasional)
Visually inspect film path and pan alignment
- Test point
- Film path, film pan, drive sprocket
- Expected
- Loops proper, film pan in correct position, sprocket dot close to film pan dot
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Unit communication error high voltage
Console displays 'Power Unit communication error. Try again or cycle power.' In-room display shows 'TRY AGAIN OR CYCLE POWER.' LFX type power units only.
Ranked causes
- 1. LFX I/O or Bias Inverter board faulty (common)
- 2. Bias Interrupt circuit failure (common)
- 3. Bad Bias Inverter Power Supply (occasional)
- 4. Bad Bias Inverter board (occasional)
- 5. Cable from LFX to bias tank faulty (occasional)
Run Bias Interrupt diagnostic
- Test point
- LFX CPU
- Expected
- Bias Interrupt test passes
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power unit inverter 1 failure high voltage
Console displays 'Power unit inverter 1 failure. Call service'. In-room display shows 'POWER UNIT ERROR. CALL SERVICE.' Only reported by MPPU and VAMP type power units.
Ranked causes
- 1. Fuses F11 or F12 blown on power supply (common)
- 2. Flat cable between Control board and Inverter 1 disconnected or damaged (occasional)
- 3. Inverter 1 failure (occasional)
- 4. Security board failure (rare)
- 5. Control board failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power unit KV balance fault detected high voltage
X-ray system power unit has detected an imbalance in high voltage (kV) generation or regulation
Ranked causes
- 1. Newly formed X-ray tube not yet stabilized (common)
- 2. Excessive kV or mA demand beyond current tube capability (common)
- 3. Aged X-ray tube failing to regulate kV properly (occasional)
- 4. High voltage tank failure or regulation problem (occasional)
- 5. Measurement or power switching board failure (occasional)
Check tube age and formation status in system log
- Expected
- If new tube without formation, error is expected
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Rear door lock fails to lock or unlock caution
Error 4B34 displayed on console and in-room display. System reports rear door lock solenoid signal failure after repeated lock/unlock attempts.
Ranked causes
- 1. Rear door not closed properly (common)
- 2. Rear door lock solenoid malfunction (occasional)
- 3. Fuse open on RF interface board (occasional)
- 4. Rear door lock not receiving voltage (occasional)
- 5. R+F Positioner module firmware or CPU error (rare)
Verify rear door is fully closed
- Test point
- Rear door
- Expected
- Door fully and securely closed
9 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply magazine not ready
Console displays 'Reload or insert supply magazine to continue', in-room display shows 'CHECK SUPPLY MAGAZINE'. Camera indicates supply magazine needs reloading.
Ranked causes
- 1. Supply magazine empty or not in place (common)
- 2. CAMERA RDY signal failure (occasional)
- 3. FILM REMAINING STATUS signal incorrect (occasional)
- 4. Photospot module fault (rare)
- 5. Camera hardware failure (rare)
Verify supply magazine is filled and installed
- Test point
- Supply magazine
- Expected
- Magazine full and seated in camera
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table angulation non-functional
Operator requests table angulation but no motion occurs. Console displays 'Table Angulation non-functional. Call Service.' In-room display shows 'ANGULATION NON-FUNCTIONAL.'
Ranked causes
- 1. Table position feedback pot not indicating movement (common)
- 2. Angulation and Table Top I/F board malfunction (occasional)
- 3. External angulation deadman enable switch preventing motion (occasional)
- 4. R+F S188 CPU Board timer fault (rare)
- 5. R+F Positioner module hardware failure (rare)
Run Table Top Angulation Test diagnostic
- Test point
- R+F Interface Board entry in R+F menu
- Expected
- Test completes without error
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table at angulation limit patient safety
Console displays 'Table at angulation limit' error; In-Room Display shows 'ANGULATION LIMIT'. Occurs when angulating the table.
Ranked causes
- 1. Foot step not fully retracted (common)
- 2. Angulation limit switch malfunction (occasional)
- 3. Servo controller interlock signal fault (occasional)
- 4. Software/firmware fault (rare)
- 5. R+F Positioner module fault (rare)
Verify foot step position
- Test point
- Table foot step
- Expected
- Fully retracted/inserted
7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table locks non-functional caution
Console displays 'Table locks non-functional. Call Service'. In-room display shows 'LOCKS NON-FUNCTIONAL'. Lock did not respond as expected: either failed to set when requested, or set when not requested. Determined by comparing expected state to actual state reported by Power Assist board. Error reported after timeout with lock in wrong state.
Ranked causes
- 1. Lock mechanism not operating correctly (common)
- 2. Power Assist board timer error (occasional)
- 3. Power Assist board failure (occasional)
- 4. Cable connection issue (occasional)
- 5. Lock mechanism broken (rare)
Run Lock Test from diagnostics
- Test point
- R+F Power Assist/Lock Control menu
- Expected
- Test completes without error
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AIR alarm occurs with no air in tubing or door closed patient safety
AIR alarm triggered when pump door is closed and START key pressed, or when there is no visible air in the tubing.
Ranked causes
- 1. Tubing mispositioned in air sensor (common)
- 2. Tubing surface scratches or deformation (common)
- 3. Wrong administration set type or code (occasional)
- 4. Air sensor contaminated with residue (common)
Check tubing position in air sensor
- Test point
- Air sensor - tubing insertion
- Expected
- Tubing fully seated in air sensor
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Camera rotator stalled
Console displays 'WARNING, Camera rotator stalled. Please try again.' TV Rotator failed to track command.
Ranked causes
- 1. Mechanical jam in the head (common)
- 2. TV Rotator Board failure (occasional)
- 3. Iris and TV Rotator Interface failure (occasional)
- 4. Rotator Motor failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cine film jammed in camera
Camera stalled; console displays 'Cine film jammed in camera. Reload magazine.' In-room display shows 'CINE FILM JAMMED. RELOAD.' Camera fails to reach 98% of expected speed within timeout period.
Ranked causes
- 1. Film jammed in transport mechanism (common)
- 2. Camera RPM limiter switch tripped (occasional)
- 3. 80188 timer or PIC circuit fault (occasional)
- 4. Camera tach or shutter circuit fault (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 420E: Auto Brightness Control problem high voltage
ABC Algorithm attempted to reload pulse width timer during ongoing exposure
Ranked causes
- 1. ECM timer configuration allows ABC update timeout during exposure (unknown)
- 2. EXT SYNC or EXP GATE signal module problem (unknown)
- 3. ECM servicing higher priority interrupts delays ABC interrupt (unknown)
- 4. Hardware or Real Time Bus timing problem (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 4247 displayed
ECM and power unit communication error or power unit sending unexpected error value
Ranked causes
- 1. ECM or power unit requires reinitialization (common)
- 2. Incorrect DIP switch settings on ECM or power unit (occasional)
- 3. Serial data link communication fault (occasional)
- 4. ECM or power unit module failure (rare)
Power cycle the system
- Test point
- Console power
- Expected
- System reinitializes ECM and power unit
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 42CB displayed
Console shows 'Instruction not understood. Please try again.' In-room display shows 'RELEASE SWITCH, TRY AGAIN.'
Ranked causes
- 1. LFX/MPPU/VAMP power unit reporting error (common)
- 2. SCPU power unit with VRTX error detected (occasional)
- 3. Floating point calculation error in SCPU (occasional)
- 4. Corrupted SCPU error code table (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 4634 - ARCNET transmission failed
Console displays 'Instruction not understood. Please try again'; in-room display shows 'RELEASE SWITCH, TRY AGAIN'
Ranked causes
- 1. Invalid destination node ID (common)
- 2. CINE module CPU RAM error (occasional)
- 3. System power supply unstable (occasional)
- 4. Corrupted application software (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposure time below minimum radiation
Console displays 'Min. expose time attempted. Select new technique to continue.' In-room display shows 'MIN. EXPOSE TIME ATTEMPTED.'
Ranked causes
- 1. AEC LUT time plus tail time minus anticipation time results in negative value (common)
- 2. High technique with very little filtration causing excessively short exposure time (occasional)
- 3. AEC LUT is defective (occasional)
- 4. AEC LUT addressing hardware is defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
FastBoot auto return failed, default parameters restored
Console displays 'Parameters not as previously selected. Check console settings.' Mode and parameters differ from previous settings after reboot.
Ranked causes
- 1. NVRAM failed (occasional)
- 2. System halted during runtime snapshots (occasional)
- 3. LAT plane snapshot mismatch in biplane systems (rare)
- 4. LAT plane brought online after receptor change (rare)
Verify console displays correct application mode and parameters
- Test point
- Console screen
- Expected
- Application mode and parameters match previously used settings
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament correction value increasing rapidly (Error 4204H)
System continuously increases filament correction value in large steps; indicates calibration or feedback problem; mA meter saturation or tube spit errors may accompany
Ranked causes
- 1. Filament control parameters not properly calibrated (common)
- 2. mA meter malfunction (ABC modes) (occasional)
- 3. mA integrator fault (rad modes) (occasional)
- 4. Incorrect mA feedback from power unit (occasional)
Review error log for related errors
- Test point
- System error log
- Expected
- Check for 42EAH (mA meter saturation) or 42CAH (tube spit) errors occurring when 4204H observed
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film changer programmer not ready
Console displays 'Film Changer Programmer not ready'; in-room display shows 'CHECK FC PROGRAMMER'. Occurs at prep time when film changer signals indicate film magazine or receiver out of device or software cannot determine film position.
Ranked causes
- 1. Film magazine or receiver out of device (common)
- 2. FC RDY signal indicates not ready or internal flag shows film not in place (common)
- 3. Failure in FC RDY signal path from programmer to I/O board (occasional)
- 4. Positioner module failure or misconfiguration (rare)
Verify film changer programmer indicates ready status
- Test point
- SEP film changer programmer display
- Expected
- No error or fault indication
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluoroscopy not operable patient safety
Console displays 'Cannot do Fluoro. Exposure control inoperable. Cycle Console.' In-room display shows 'FLUORO NOT OPERABLE.'
Ranked causes
- 1. ECM not configured online during room setup (common)
- 2. No reply from ECM to load command (occasional)
- 3. No reply from ECM to reset fluoro timer (occasional)
- 4. No reply from ECM to prep sequence enable command (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Node response timeout during test sequence
Console displays timeout error; test aborted; GE bus communication failure with system node
Ranked causes
- 1. ASC BIU communication failure (common)
- 2. Node not responding to code download request (common)
- 3. Node hardware malfunction (occasional)
- 4. BIU board defect on receiving node (occasional)
Verify ASC can communicate with system nodes
- Test point
- ASC BIU to any available node
- Expected
- Communication established without timeout
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Obstruction detected at front patient safety
Console displays 'Obstruction detected, clear using Front/Rear load buttons.' In-room display shows 'OBSTRUCTION DETECTED.' System enters operator-initiated recovery mode.
Ranked causes
- 1. Front sensor beam (IN FRONT CASS signal) falsely broken or cassette already in system (common)
- 2. IN FRONT CASS signal indicating falsely (occasional)
- 3. Front sensor adjustment out of tolerance (occasional)
- 4. R+F Positioner module or RAM/Arcnet failure (rare)
Perform automatic cassette unload
- Test point
- Front load button
- Expected
- Cassette unloads via front, recovery completes
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Positioner software initialization error
Error 4A6A displayed on console and in-room display; console message instructs to cycle power or call service
Ranked causes
- 1. Positioner to LP controls link failure (common)
- 2. Device I/O board malfunction (occasional)
- 3. Positioner module RAM or Arcnet failure (occasional)
- 4. LP control system failure (occasional)
Verify link status between Positioner and LP controls
- Test point
- Physical connection Positioner-LP
- Expected
- Connected and secure
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power plug not lit or battery icon lit when plugged in caution
Plug icon fails to illuminate when pump plugged in, or battery icon incorrectly illuminates with AC power connected.
Ranked causes
- 1. Power plug not fully inserted into AC outlet (common)
- 2. Rear power fuse failed (common)
- 3. AC outlet voltage incorrect or below specification (occasional)
- 4. Power cord discontinuity (occasional)
Verify power plug connection
- Test point
- AC power plug in grounded outlet
- Expected
- Plug fully seated in outlet
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power unit tube select error radiation
Power unit cannot select correct X-ray tube; system displays error and requires power cycle
Ranked causes
- 1. Power unit transient fault (occasional)
- 2. Power unit hardware or communication fault (occasional)
- 3. TIRC circuit malfunction (occasional)
- 4. Interface 1 board fault (occasional)
Power cycle the system
- Expected
- System restarts successfully
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump door will not open or close smoothly caution
Door movement is stiff or binding when attempting to open or close.
Ranked causes
- 1. Tubing or slide clamp mispositioned or not seated (common)
- 2. Wrong administration set type or code (occasional)
- 3. Solution spills, liquids or residues present (common)
- 4. Door latch roller pin sticking or damaged (occasional)
Verify tubing and slide clamp positioning
- Test point
- Tubing guides and slide clamp slot
- Expected
- Tubing and clamp properly seated in guides
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table top in illegal region patient safety
Table top is extended in angulated position when it should be retracted, or extended beyond calibrated boundary. Error message on console: 'Press OFF button to stop table motion. Call Service.' In-room display: 'PRESS EMERGENCY OFF BUTTON.'
Ranked causes
- 1. Table feedback pot reading outside calibrated positions (common)
- 2. Angulation switches wired backwards (occasional)
- 3. Table motors not stopping fast enough (occasional)
- 4. Calibration data corrupted (rare)
Run Table Top Lat/Long Motion Test
- Test point
- R+F Interface Board in R+F menu
- Expected
- Test passes without error
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table top motion non-functional patient safety
Console displays 'Table top motion non-functional. Call Service.' In-room display shows 'TABLE TOP NON-FUNCTIONAL.' Table top longitudinal motion requested but no motion detected after timeout. Position feedback pot shows no movement.
Ranked causes
- 1. Table top longitudinal feedback pot failure or path failure (common)
- 2. Angulation and Table Top I/F board failure (occasional)
- 3. R+F S188 CPU board timer failure (occasional)
- 4. R+F Positioner module RAM/Arcnet failure (occasional)
Run Table Top Lat/Long Motion Test
- Test point
- R+F Interface Board, R+F menu
- Expected
- Test completes without error
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video aperture did not respond
Console displays 'Video aperture did not respond. Try again or call service.' In-room display shows 'APERTURE DID NOT RESPOND.' Iris electronics failed to generate interrupt within 1 second timeout.
Ranked causes
- 1. Iris electronics database parameter ap_pos_cal not set (common)
- 2. XRII & Iris Control Board failure (occasional)
- 3. Iris and TV Rotator Interface failure (occasional)
- 4. Iris assembly failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AEC exposure time exceeds four seconds
After AEC exposure terminates, LUT address exceeds 1299 (4 sec limit), error 4267 reported, LUT defaults to 1299
Ranked causes
- 1. Backup timer loaded with value less than four seconds but not functioning (common)
- 2. Backup timer hardware fault on ETL1 board (occasional)
- 3. AEC hardware fault on ETL2 board (occasional)
Verify backup timer is loaded with value less than four seconds
- Test point
- ETL1 board backup timer setting
- Expected
- value less than 4 seconds
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto-positioner control problem
Serial link interface between Vascular positioner and tableside control has failed. Console displays 'Auto-positioner control problem. Use manual positioning or try again.' In-room display shows 'AUTO-POSITIONER PROBLEM.'
Ranked causes
- 1. Serial link interface failure due to loose connections, noise, or power supply issues (unknown)
- 2. CPU board or device I/O board failure (unknown)
- 3. Tableside auto-positioner board failure (unknown)
Check cabling for loose connections
- Test point
- Serial link connectors
- Expected
- Connections secure and tight
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autoprotocols unavailable at power-up
Console displays 'Autoprotocols unavailable. Continue and call service.' PROCEDURE MENU button not displayed. AUP software removes itself from active applications.
Ranked causes
- 1. AUP memory board not securely attached or battery voltage low (common)
- 2. AUP memory board defective or inaccessible (common)
- 3. ASC module or system-level hardware failure (occasional)
Reboot system to clear transient errors
- Test point
- System startup
- Expected
- Error clears and PROCEDURE MENU button appears
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad data received from power unit high voltage
Console displays 'Bad data received. Turn console power OFF/ON.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' ECM resets power unit.
Ranked causes
- 1. DMA controller malfunction (occasional)
- 2. Software configuration problem (occasional)
- 3. Associated hardware failures (linked errors) (rare)
Verify DMA controller operation
- Test point
- DMA controller
- Expected
- Packet length >= 5 bytes
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad data received from XRII interface radiation
Console displays 'Bad data received. Turn console power OFF/ON.' In-room display shows 'TURN CONSOLE POWER OFF/ON.' System halts X-ray operation.
Ranked causes
- 1. DUART framing error on XRII & Iris Control Board channel A (common)
- 2. XRII I/F Board failure (occasional)
- 3. XRII & Iris Control Board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette not detected in tray radiation
Console error message 'Load cassette in tray to continue' appears when attempting RAD exposure on selected receptor
Ranked causes
- 1. Cassette not loaded in tray (common)
- 2. Tray not fully inserted (common)
- 3. Cassette-in-place signal not reaching positioner (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cine camera not line locked
Console warning message displayed; hardware phase lock loop cannot lock to incoming signal during camera startup; LOCK signal on port 0 not asserted when stall timer expires.
Ranked causes
- 1. LOCK signal not asserted from power amp board (common)
- 2. Incoming signals to phase comparator not in phase (common)
- 3. CINE CPU board defective (occasional)
Run diagnostics on CINE CPU board
- Test point
- CINE CPU Board menu
- Expected
- 8253 Diagnostic Timer, 8253 Timers 0/1/2, 80188 PIC, Phase Locked Loop tests pass
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cine camera out of film
Console displays 'Cine camera out of film. Reload magazine.' and in-room display shows 'CAMERA OUT OF FILM.' System detects INTLOK bit as off on port 0 and film remaining A/D value is below calibrated no-film threshold.
Ranked causes
- 1. Magazine empty or film depleted (common)
- 2. Film sensor or motor malfunction (occasional)
- 3. CINE module circuitry fault (rare)
Verify film presence in magazine
- Test point
- Cine camera magazine
- Expected
- Film loaded in magazine
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cine camera overspeed or jam fault
Console displays 'Reset Cine camera overspeed switch at camera or check for film jam'. In-room display shows 'CINE CAMERA FAULT'. Camera will not run.
Ranked causes
- 1. Film not loaded correctly (common)
- 2. Camera RPM limiter switch tripped (common)
- 3. INTLOK signal failure or defective camera (occasional)
Verify film is loaded correctly in camera
- Test point
- Camera film compartment
- Expected
- Film properly seated and threaded
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Database error in ECM with invalid cal/config data high voltage
ECM received calibration or configuration data out of range. ECM CPU LEDs toggle between 00FH and 0F0H in endless loop until cold reset. Module not marked online; not ready signal not sent to ASC.
Ranked causes
- 1. Invalid go-to-cold time parameter in configuration data (occasional)
- 2. Other out-of-range parameters in calibration/config data (occasional)
- 3. Multiple invalid parameters detected by HW status field (rare)
Observe ECM CPU board LED state
- Test point
- ECM CPU board LEDs
- Expected
- LEDs toggling between 00FH and 0F0H hex
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DF module not ready for acquisitions
Console displays 'Enter DF acquisition ready screen to continue'; in-room display shows 'SET DF CONSOLE FOR EXPOSURE'
Ranked causes
- 1. DSMON not entered at DF keyboard (common)
- 2. DSMON in REPRO or XFER mode instead of acquisition mode (common)
- 3. Console power cycle required (occasional)
Verify DSMON has been entered at DF keyboard
- Test point
- DF keyboard display
- Expected
- DSMON mode active
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 4246
ECM and power unit communications error; system displays error code and may not operate normally
Ranked causes
- 1. ECM or power unit requires re-initialization or DIP switch misconfiguration (common)
- 2. ECM or power unit module failure detected by PRD diagnostics (occasional)
- 3. System level failure in ECM or power unit (rare)
Cycle console power to re-initialize
- Test point
- Console power switch
- Expected
- System fully reboots and initializes
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 4264 displayed on console and in-room caution
Console shows 'Press CPU RESET at control cabinet. System requires reboot.' In-room display shows 'PRESS CPU RESET.' ECM cannot bring power unit online.
Ranked causes
- 1. Power unit did not set RESET DONE bit within ten seconds (common)
- 2. Power unit CPU board failure (occasional)
- 3. ECM power unit interface problem (occasional)
Inspect power unit CPU board LED indicators
- Test point
- Power unit CPU board LEDs
- Expected
- No PRD error indicated on LEDs
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 42B9: Power unit tube select error radiation
Error message displayed when tube selection fails during power unit operation. Console shows 'Power unit tube select error. Cycle power, call service & continue.' In-room display shows 'TRY AGAIN OR CYCLE POWER.'
Ranked causes
- 1. Compressed pins in high voltage connectors (common)
- 2. HV tank failure (low filament current with kV present) (occasional)
- 3. HV switch or measurement commutation board failure (occasional)
Cycle power to console
- Test point
- Power switch
- Expected
- Error clears on restart
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 4A09 Bad data received
Console displays 'Bad data received. Turn console power OFF/ON.' with in-room display showing 'TURN CONSOLE POWER OFF/ON.'
Ranked causes
- 1. Software task scheduler overflow; no more room in timer list (common)
- 2. Positioner module task scheduler or memory issue (occasional)
- 3. Corrupted Applications software on floppy (occasional)
Power cycle system
- Test point
- System power and reset
- Expected
- Console and in-room displays boot normally; no error messages
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Error 4B33 displayed on console
- Error 4B3D diagnostic procedure
- Error 6602 occurs at warm start
- Exposure run terminated by STARTWHEN timeout
- Exposure time calculated as negative
- Film supply depleted or camera not ready
- INSERT SLIDECLAMP alert or alarm when door closed
- kVp control error
- Magazine lever not closed
- Obstruction prevents table angulation
- Power unit tube spit recovery
- Rear door open interlock
- Receptor ready interrupt lost during X-ray exposure
- Safety clamp will not latch open
- STARTWHEN timeout, overlap detector
- Supply magazine empty indication
- System timeout during exposure startup
- Table not positioned at horizontal
- Tube anode not rotating
- Tube spit during exposure
- Tube spit recovery error during exposure
- Vascular positioner serial link communication failure
- X-ray tube and table receptor misalignment
- Camera A/D converter timeout
- CHECK SET/LOADING alarm when door closed
- Cine magazine missing
- Circle adjust failure during calibration
- Collimator non-functional
- Digital Record inoperative in fluoroscopy mode
- Error 420A
- Error 421A: Power unit inoperative
- Error 4245 displayed
- Error 4274 displayed
- Error 4277 occurs during power unit operation
- Error 42C5
- Error 4500 displayed
- Error 4903 displayed during RAD exposure
- Error 4B57: Remove all cassettes to continue
- Error 6405: Exposure control inoperable
- Error 6707 displayed
- Filament control adjusting message displayed
- Filament drive error
- Filament drive error with auto-reset in progress
- Film changer not correctly locked
- Film changer rotation interlock persists after repositioning
- Floppy diskette read error
- Image Intensifier not at required SID position
- Improper exposure error 4214
- Jog not allowed during Cine recording
- Light bias inoperative
- Maximum exposure time reached
- Maximum tube housing heat units reached
- Obstruction in fluoro image
- Optical filter not in position
- Procedure database version incompatible
- Procedure menu blocked during auto-protocol mode
- RAD exposure blocked by room door
- RATE, VTBI, or START key not accepted
- Selected filament open
- Serial link communication failure
- Servo axis not maintaining position
- SID error
- System error persists after diagnostics
- Table height error during RAD exposure
- Table motion disabled with film changer engaged
- TV video signal out of range
- Unrecognized command error
- VTR not in remote mode
- X-ray tube overheating
- ABC brightness instability
- ASC sends invalid video gain to ECM
- Audible alarm volume is not loud enough
- Backlight is off during internal battery operation
- Bias fault
- Bias fault detected
- Bias fault indicated
- Bias inverter supply too low
- Biplane system communication failure
- Camera still hot error on startup
- Cannot take exposures with LP parked
- Cassette carriage jammed
- Cassette not detected in Spot Film Device
- Cassette not positioning
- Cine camera film running low
- Cine mode unavailable
- Collimator blades not closed during tube warmup
- Cone button fails to operate
- Console displays bad data error
- Console power cycle required
- Console receives invalid video format command
- Database Error: Incompatible Database Version
- DF service mode exposure zero milliseconds error
- Diagnostic test failed on startup
- Digital error - service required
- Disk space exhausted during acquisition
- Disk space remaining less than 15 frames
- Duty cycle limit exceeded
- DX module not ready for acquisitions
- E4156 Filament drive error
- ECM loses power to CPU
- ECM module not coming online after error
- Error 4202 power unit reset for unknown reason
- Error 4217: No exposure, power unit inoperative
- Error 4245 with POWER UNIT issues
- Error 4260 displayed
- Error 4264 persists after initial troubleshooting
- Error 4266 displayed on console and in-room display
- Error 4273 displayed
- Error 4282 detection and recipe abort
- Error 4297: Start when timeout, boost done
- Error 4299 displayed on console
- Error 429D displayed
- Error 42C9: Filament drive error detected by power unit
- Error 42F9 detected
- Error 4405 Digital system maximum run length reached
- Error 4706 divide by zero fault
- Error 4913 diagnostic procedure
- Error 4A59 film changer position
- Error 4A5A diagnostic procedure
- Error 4A5F - Positioner module division by zero
- Error 4A81
- Error 4B19 displayed
- Error 4B22 displayed
- Error 4B2C
- Error 4B33 persists after initial troubleshooting
- Error 4B34 displayed
- Error 4B3F displayed
- Error 4C04 diagnostic procedure
- Error 4C05 displayed
- Error 4C08 verification
- Error 6417: Bad data received from Digital Record
- Error 6600 displayed
- Error 6601 occurs on lateral biplane system
- Error 6602 during SAVE/RETRIEVE operation
- Error code 42A1 detected
- Error code 4A02 displayed
- Error code 4A8E
- Error code 4B36 diagnostic procedure
- Error code 4C11 triggered
- Exit sequence not executed
- Exposure run terminated with timeout
- Exposure start signal timeout (Error 4282)
- Filament command failure with auto-reset
- Filament warmup delay message displayed
- Film changer rotated unexpectedly
- Floppy disk error
- FLOW RATE or CHECK VTBI alarm occurs on START
- Frontal trial image too bright, lateral trial image too dark
- HI or LO displayed during Volume-Time programming
- II on LP collided with object
- Image Hold Fluoro mode only available
- Injector not ready on prep switch depression
- Insufficient disk space for cardiac exposure
- Insufficient disk space for image acquisition
- Interval between audible alarm tones too long
- Last exposure may be bad
- mA exceeds 125% of maximum (Error 420AH)
- mA meter saturation (Error 42EAH)
- No exposure, power unit inoperative
- Object blocking front load port sensor
- Object in front load port
- Photo cathode current read during AGC
- Photospot camera inoperable
- Photospot prep done timeout during exposure
- Positioner module task scheduling failure
- Positioner position not selected within time limit
- Positioners not orthogonal
- Power unit control not operable
- Power unit fault with high voltage unavailable
- Power unit recovery sequence initiated
- Power unit rotor control error
- Pump fails to run on internal battery
- Pump stops with BATTERY LOW alarm
- Rear load door forced open
- Rotor braking sequence not completing
- Secondary errors during power unit recovery
- SID zero value calibration drift or error
- Signal transmission error to ECM
- Spotfilmer locked in forward position
- Spotfilmer servo non-functional
- Standby cooling not performed due to invalid time parameter
- Stepper selected at table but not at console
- System aborts with Error 4011
- System communication problem
- System communications problem between ASC and DR module
- Table motion non-functional
- Table not ready during exposure prep
- Table top not centered laterally
- Table top position data incorrect
- Tomo procedure cannot start
- Tomo table not properly prepared for acquisitions
- Tube casing water flow not detected
- Tube spit error (Error 42CAH)
- TV gain curve data bad
- Unable to establish tube selection during operation
- Video camera system inoperative
- Video setup corrupted
- Video shading correction fault
- VTR communication error
- VTR left idle in ready state over 30 minutes
- Watchdog timer timeout (error indeterminate)
201 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
30 | Analog output different from expected value; possible shorted/over-loaded analog output or AD/DA converter failure | Watchdog detects analog output voltage out of range or converter circuit malfunction | error |
4000 | System error - RMX error detected | APOS code running in ASC module detects RMX error condition | error |
4000-40FF | APOS (Applications Operating System) error range | Error encountered during Applications mode operation | unknown |
4001 | System error - error code not found in error table | APOS code running in ASC module encounters unknown condition | error |
4011 | System error: no free buffers in APOS running in ASC module | TOMO function selected on Advantx Console prior to turning on Ordograph Control, especially with long interconnect cable (80 feet or longer) | critical |
4012 | System error - no free mailboxes | APOS code running in ASC module detects mailbox allocation failure | critical |
4013 | APOS kernel semaphore exhaustion | Operating system runs out of available semaphores | error |
4020 | System communications problem | Loss of communication with node | error |
4022 | System communications problem in APOS running on ASC module | ASC module node fails to respond within timeout period | error |
4023 | System communications corrupted | ARCNET initialization failure | critical |
4024 | Wrong node address error | ASC module detects wrong node address configuration | error |
4025 | System communications problem | APOS code in ASC module | error |
4027 | System communication problem | Applications Operating System (APOS) code running in the ASC module | critical |
4030 | System communications problem | APOS code running in ASC module generates error | error |
4031 | System communications problem | APOS code running in ASC module detects bad data type in descriptor block | error |
4032 | System communications problem; APOS stack/data structure error | No data pointer present in ASC module during operation | critical |
4033 | System communications problem | ASC module communication failure | error |
4035 | System communications problem | Unknown | error |
4036 | System communications problem | APOS code running in ASC module | error |
4037 | System communications problem | APOS (Applications Operating System) code in ASC module detects communication issue or invalid character in input field | error |
4038 | System communications problem | APOS code in ASC module | error |
4039 | System communications problem | Applications Operating System (APOS) code running in ASC module detected bad data type in MSG$DECODE function | error |
4040 | System communications problem | ASC module call table index error | error |
4041 | System communications problem | Module code filename not found in filename table | error |
4042 | System communications problem | System communications error detected in ASC module | error |
4043 | Text input field size exceeded | Applications Operating System (APOS) receives text input exceeding field size limit | error |
4044 | APOS stack limit exceeded in ASC module | Applications Operating System code running in ASC module generates stack overflow | critical |
4050 | System error message queue is full | APOS applications operating system error in ASC module | error |
4060 | System error: no default task address in table | APOS operating system unable to locate default task address during initialization | error |
4061 | Bad default task number in APOS | APOS code error during startup or operation | critical |
4065 | ASC Dynamic RAM Memory Parity Error; problem with ASC main memory | System automatic self-check detecting memory parity failure | error |
4067 | System power down event marker in Event Log | System was powered down | info |
406F | System requires reboot | CPU error condition | error |
4070 | System communications problem | System error during operation | error |
4073 | System communications problem in APOS (Applications Operating System) running on ASC module | Mailbox released with data unexpectedly during system operation | error |
4074 | System communications problem | APOS code running in ASC module generates error | error |
4075 | System error - GE bus reconfiguration occurred | APOS code in ASC module detects bus reconfiguration | error |
4076 | unknown | ||
4077 | System error in Applications Operating System (APOS) module | Bad translate table detected in ASC module | error |
4078 | Other ASC did not respond to biplane communication request | Lateral plane did not respond to Operating Mode Request command | error |
4082 | unknown | ||
4083 | unknown | ||
4084 | unknown | ||
4086 | Database is unusable; restore required | Application detects corrupted or incompatible database at startup | error |
4090 | unknown | ||
4091 | unknown | ||
4092 | System error with module and description not fully documented | Unknown; messages and recommended actions not yet available | unknown |
4093 | unknown | ||
40A0 | System communications error | ASC software attempts to reallocate a software timer that is already in use; occurs when bouncing on and off prep and fluoro switches | error |
40FF | System communications problem | APOS code in ASC module detected communication failure | critical |
4111 | Both frontal and lateral trial images are too bright | Biplane trial image acquisition with excessive brightness on both planes | warning |
4112 | Frontal trial too bright, lateral trial too dark | Biplane trial image quality assessment fails due to exposure mismatch between frontal and lateral projections | warning |
4113 | Frontal trial image is too dark and lateral trial image is too bright. | Biplane trial execution with suboptimal technique values for frontal and lateral planes. | warning |
4114 | Both frontal and lateral trial images are too dark | Trial images acquired with insufficient exposure or technique | warning |
4115 | DX store function in progress on the in-room monitor | DX action initiated; takes 2 seconds to complete | info |
4116 | DX recall function in progress on the in-room monitor | DX action recall initiated; takes approximately 2 seconds to complete | info |
4150 | Power Unit communication error | Power unit sends request for service and receives no response from ECM after timeout of 130s | error |
4151 | Power Unit communication error | Error detected in the format of the packet received from ECM (checksum, overrun...) | error |
4152 | Instructions not understood. Please try again. | Bad command from ECM concerning tube number, kV demand, mA demand or mAs demand. | error |
4153 | ECM maintains boost heater command for more than 400 ms | Boost heater command duration exceeds 400 milliseconds | error |
4154 | Instructions not understood. Please try again. | ECM sends a rotor command while the tube is switching. | error |
4155 | Instructions not understood | Console command or in-room control input not recognized | warning |
4156 | Filament drive error | Filament current sense system detects filament current above threshold, tripping latch | error |
4157 | Filament drive error in heater system | Filament current sense system detects current above threshold | error |
4158 | Filament drive error | Filament current sense system detects current above threshold on heater board | error |
4159 | Filament drive error | Filament current sense system detects filament current above threshold, trips latch | error |
415A | Filament drive error detected in heater board filament current sense system | Filament current exceeds threshold on heater board, latch trips and reports error to CPU | error |
415B | Filament drive error - overcurrent condition detected on heater board filament current sense | Filament current sense system detects filament current above threshold while in acquisition mode | error |
415C | Filament drive error in small focus heater circuit | No current detected in small focus heater circuit with non-zero heating demand | error |
415D | Filament drive error | No current detected in the small focus heater circuit with a non-zero heating demand | error |
415E | Filament drive error in small focus heater circuit | No current detected in small focus heater circuit with non-zero heating demand during acquisition | error |
415F | Filament drive error | No current detected in large focus heater circuit when heating demand is non-zero | error |
4160 | Filament drive error | No current detected in large focus heater circuit with non-zero heating demand | error |
4161 | Filament drive error in large focus heater circuit | No current detected in large focus heater circuit with non-zero heating demand during acquisition | error |
4162 | Filament drive error | Over-current in small focus inverter circuit or open circuit detected during heater pre-light state | error |
4163 | Filament drive error | Over-current in small focus inverter circuit or open circuit detected | error |
4164 | Filament drive error in SCPU type power unit | Over-current in small focus inverter or open circuit detected during heater acquisition state | error |
4165 | Filament drive error | Over-current in large focus inverter circuit or open circuit detected during heater state (pre-light) | error |
4166 | Filament drive error in X-ray tube heater circuit | Over-current in large focus inverter circuit or open circuit detected during heater operation | error |
4167 | Filament drive error | Over-current detected in large focus inverter circuit or open circuit condition | error |
4168 | Power unit rotor control error | CPU detects no current in rotor controller circuit while commanded on | error |
4169 | Power unit rotor control error | CPU detects no current in rotor controller circuit while commanded on | error |
416A | Power unit rotor control error | CPU detects no current in rotor controller circuit while commanded on | error |
416B | Power unit rotor control error | CPU detects no current in rotor controller circuit while commanded on | error |
416C | Power unit rotor control error | CPU detects no current in rotor controller circuit while commanded on | error |
416D | Power unit rotor control error | CPU detects no current in rotor controller circuit while commanded on | error |
416E | Power unit rotor control error | CPU detects overcurrent in rotor controller circuit while commanded on | error |
416F | Power unit rotor control error | CPU detects over current in rotor controller circuit while commanded on during High Speed Run | error |
4170 | Bias fault detected | Signal grid_func_ok tested at bad level during mode change in secondary timed modes or when bias is selected | critical |
4171 | Bias Fault | Signal grid_pulse_ok detected false during CINE or PULSED FLUORO initialization (EXP_CMND rise) | error |
4172 | Bias Fault | Grid monitoring measures mA during grid off time (>1.1 mA fluoro or >11 mA cine >0.6 after RAD pulse) | error |
4173 | Bias Fault in VAMP type power unit | No RAD_PULSE interrupt received by CPU board within 2 s after Exp Cmd | error |
41B0 | Power unit tube spit recovery | kV drop detected on anode side during tube exposure | warning |
41B1 | Power unit tube spit recovery | kV drop on anode side during tube spit event | warning |
41B2 | Power unit tube spit recovery | Tube spit condition detected during X-ray exposure with automatic KV cutoff | warning |
41B3 | Power unit kV regulation error | Detected during tube spit; kV automatically cut off | error |
41B4 | Tube casing water flow not detected | Chiller circuitry detects water not flowing or temperature too high with MX150 tube | error |
41C0 | Duty cycle limit exceeded during cine operation | Three consecutive pulses shot with KV75% signal not released | error |
41C1 | Bias fault in VAMP type power unit | Bias inverter feedback signal from micro Grid/Bias tank is at an incorrect level | error |
41C2 | Bias inverter supply voltage exceeds acceptable range | Bias feedback voltage more than 10% above demand value | error |
41C3 | Bias inverter supply voltage too low | Bias inverter supply measured more than 10% below demand value | error |
41D0 | Power unit rotor control error | CPU detects overcurrent in rotor controller circuit while commanded on | error |
41D1 | Power unit rotor control error | CPU detects over current in rotor controller circuit while commanded on | error |
41D2 | Power unit rotor control error | CPU detects over current in rotor controller circuit while commanded on | error |
41D3 | Power unit rotor control error | CPU detects over current in rotor controller circuit while commanded on | error |
41D4 | Power unit rotor control error | Low speed relay status not compatible with demand | error |
41D5 | Power unit tube select error | SCPU receives a tube switching command while the rotor is still running (RUN, ACCELERATION or BRAKE state) | error |
41D6 | Power unit tube select error | SCPU receives tube switching command from ECM while heater on large or small focus is ON | error |
41D7 | Power unit tube select error | SCPU receives tube switching command from ECM while kV measured exceeds 9 kV | warning |
41D8 | Power unit tube select error | High voltage tube switch status is different from the tube requested | error |
41D9 | Power unit tube select error | Rotor tube switch status is different from the tube requested | error |
41DA | Power unit tube select error | Heater tube switch status differs from tube requested | error |
41DB | Power unit control not operable | Status of mA scale relays or kV/mA test relays not compatible with demand or error in mA measurement circuit | error |
41DC | Power unit control not operable | Bad connection between SCPU front panel boards detected | error |
41DD | Filament command failure | ADC result does not match DAC reference value for heater and kV command | error |
41DE | Power unit control not operable | Safety latch from inverter is set out of exposure | error |
41DF | Power unit control not operable | Safety from heater on large focus function is set while heater is not activated | error |
41E0 | Power unit control not operable | Safety rotor function set while rotor is not activated | error |
41E1 | Power unit control not operable | DC BUS voltage out of range: less than 400V or greater than 762V | critical |
41E2 | Power unit control not operable | Error detected in the inverter floating power supply | error |
41E3 | Power unit control not operable | Error detected in low voltage power supply (5V, +15V, -15V or 24V) | error |
41E4 | Power unit control not operable | Safety interlock activated when small focus function enabled but heater not activated | critical |
4200 | Software error: ECM received undefined ARCNET command outside valid range (00H–0FH or 80H–FFH) or corrupted packet from hardware failure | ARCNET hardware failure in ECM or sending node; undefined command sent to ECM | error |
4200-4277 | ECM module Applications software error range | Error in ECM module during Applications mode | unknown |
4201 | Bad data received from console communications | ECM recipe pre-checker detects error with recipe during initialization | error |
4202 | Power unit reset for unknown reason | Power unit sent reset done bit in packet without prior ECM reset command | warning |
4203 | Bus timeout NMI (Non-Maskable Interrupt) | CPU attempts to access undefined or unreachable memory location; Multibus decode logic failure on ECM circuit boards | error |
4204 | Filament correction value increased by 4 or more counts during debug mode | HIGH word of filament correction value increased by 4+ counts since last 4204H or 4205H message | info |
4204H | Filament correction value increasing rapidly in large steps | System repeatedly increases filament correction value; occurs frequently with large step increases | warning |
4205 | Filament correction value decreased by 4 or more counts; informational message for Filament Control Algorithm monitoring | HIGH word of filament correction value decreased by 4+ counts since last 4204H/4205H message, or lower limit of filament correction adjustment encountered | info |
4205 | Filament correction factor adjusted to lower limit due to mA feedback error or other transient condition | mA feedback from power unit lower than actual mA, or prior mA meter saturation (42EAH) or tube spit (42CAH) errors cleared | warning |
4206 | Instruction not understood. Please try again. | Recipe processor interrupt received when ECM did not think valid recipe was executing, or ASC sent Exit/Prep/Expose/kick start before recipe downloaded successfully | critical |
4207 | Received recipe poke command with no RIP downloaded | A patch recipe command (ARCNET command 0A1H) was received from the ASC when no valid recipe resides in the ECM | error |
4208 | Loss of receptor ready interrupt during X-ray exposure | Receptor ready signal (RTB-1) goes low while X-rays are actively on | critical |
4209 | Software error; ECM could not find EXITMARK opcode in recipe or detected DONE before EXITMARK | Exit Sequence Enable command (ARCNET 085H) received without valid EXITMARK opcode | error |
420A | Filament control adjusting | mA returned from power unit reaches 1.25 times expected mA during Fluoro exposure for at least 3 seconds | warning |
420A | mA meter error or related power delivery fault | Incorrect mA meter calibration, hardware fault in ECM or Power Unit, or related grid/tube errors | error |
420AH | mA exceeds 125% of maximum rated mA | Measured mA value exceeds 125% of specified maximum mA limit | error |
420B | mA below specification during fluoroscopy operation | Actual fluoroscopy mA output falls below programmed minimum threshold | error |
420B | Filament control limit | mA feedback from power unit is less than half of calculated maximum mA while ECM expects near maximum, during ABC Algorithm operation at KV override or above 90% max pulse width for pulsed systems, or temporary grid short condition clears during exposure | error |
420C | Filament control threshold exceeded | During pulsed Fluoro exposures, mA feedback is more than twice expected value, indicating possible grid short not detected by power unit, or mA metering circuit problem | error |
420D | Minimum kV temporarily limited for thermal protection | Generator detects thermal overload condition approaching limits | warning |
420E | Suspected Auto Brightness Control problem | ABC Algorithm attempted to reload pulse width timer during active exposure | warning |
4210 | No brightness samples available | ASC requested brightness samples before buffer completely full or ECM not receiving brightness ready interrupts | error |
4213 | Power Unit communication error | All 20 ECM power unit packet buffers full with valid packet information; packets queued faster than sent over SDLC link | error |
4214 | Improper exposure - power unit HV ON timeout error | Power unit fails to set HV ON within 50-100 milliseconds of exposure command from ECM | critical |
4215 | Power unit did not set HV ON in response to EXP CMND from ECM within 50-100 msec | Power unit never received EXP ENBL signal or ECM's EXP ENBL signal is low when EXP CMND is set | error |
4216 | Power unit did not set HV ON in response to exposure command from ECM within 50-100 msec due to SCR ON or precontacting fault | Power unit fails to respond to EXP CMND from ECM within 50-100 msec | error |
4217 | Power unit failed to respond to exposure command | Power unit did not set HV ON within 50-100 msec of exposure command from ECM | critical |
4218 | Power unit inoperative - safety contactor did not close | Power unit does not set HV ON within 50-100 msec of exposure command because safety contactor failed to close | critical |
4219 | Power unit did not set HV ON in response to EXP CMND from ECM | Power unit failed to respond to exposure command within 50–100 msec | error |
421A | Power unit inoperative; no exposure possible | Charger SCR fails to turn on within 50-100 msec of exposure command | critical |
421B | No exposure. Power unit inoperative. | Power unit did not set HV ON in response to EXP CMND from ECM within 50-100 msec because commutating capacitors never charged | critical |
421C | Power unit inoperative; ECM did not receive feedback #1 from power unit after watchdog timeout request | ECM requests watchdog timeout cause from power unit but next packet lacks feedback #1; power unit may die between error detection and ECM information request | critical |
421F | Suspected Auto Brightness Control problem | ABC configuration data sent by ASC was incorrect | error |
4220 | Maximum tube housing heat units reached | Tube housing thermal load reaches maximum limit | warning |
4221 | X-ray tube heat wait condition; ECM detected that X-ray tube requires cooling before continued operation | X-ray tube heat level exceeds safe operating threshold; ECM detects heat wait condition and notifies ASC via ARCNET command 35H | warning |
4222 | Bad data received; ECM received invalid database identifier in cal/config packet | ARCNET command 0CH received with database number other than 2 or 3 | error |
4222 | ECM or ASC module software error | System error detected during operation or initialization | error |
4223 | Bad data received from software command | ECM received cal/config command (ARCNET 00CH) with invalid part number | error |
4224 | Bad data received from ECM packet | ECM receives cal/config packet with invalid number of packets parameter | error |
4225 | Bad data received from ECM cal/config packet | ECM received ARCNET command 0CH with incorrect data byte count for part number | error |
4226 | Parameter out of range error in ASC or ECM module | Software reload or database inconsistency detected during system initialization | error |
4227 | Power unit inverter thermal wait condition detected | ECM detects inverter heat wait condition or predictive thermal threshold exceeded | warning |
4228 | Bad data received from ECM | ECM received a cal/config packet with invalid packet number parameter | error |
4229 | Tube bulk temperature exceeds 1600 degrees C or database error in cooling table index | Bulk temperature sent with non-volatile run-time parameters is too large; ECM calculates index >199 when dividing bulk temp by 8 | error |
422A | Selected focal spot out of range | ASC sends ARCNET command 0B1H with focal spot value outside valid range 0-5 | error |
422B | Software error: ECM did not receive all part numbers from ASC during calibration/config data transmission | ECM thinks ASC finished sending cal/config data but one or more part numbers were not received | error |
422C | Software error in automatic video gain control routine | ASC sends Fluoro detail value greater than two to ARCNET command 0B4H | error |
422D | Power unit in reset state – prep/expose switch ignored | ECM resets power unit due to prior error; prep/exposure switch transitions occur during reset | warning |
4230 | Instruction not understood. Please try again. | ASC tried to download a new recipe when ECM thought it was still executing current recipe | error |
4231 | Recipe downloaded from ASC contains odd number of bytes | Recipe download from ASC with incorrect byte count | error |
4244 | Bad data received from ASC; X-ray tube distance downloaded is less than SID zero | ECM receives x-ray tube distance (command 0A0H) less than SID zero parameter | warning |
4244 | SID zero calibration error or sensor drift | SID (source-to-image distance) sensor or calibration data out of specification | error |
4245 | Communication failure between ECM and POWER UNIT | Packet transmission or clock signal loss between modules | error |
4245 | Power Unit communication error | ECM did not receive a reply packet from the power unit within one second of transmission | error |
4246 | ECM received PRD error from power unit after PU Reset line asserted | Power unit PRD error during warm start or after fatal error reset | error |
4247 | Invalid power unit error code received by ECM from power unit | Power unit sends unrecognized error code or ECM misinterprets received error code | warning |
4248 | Bad data received from power unit during communications test | ECM detected error during PRD communications tests between ECM and power unit | critical |
4255 | ETL-1 board or system level error | Unknown from provided text | error |
4255 | Spurious interrupt received by ECM | Signal noise or interrupt not held long enough; vector number often 03FH | error |
4256 | ASC sent video gain information to ECM that was out of range; ECM overrides with valid value between 1.0 and 40.0 | ASC module sends video gain value outside valid range to ECM | warning |
4256 | Suspected Auto Brightness Control problem | Auto Brightness Control malfunction detected | warning |
4257 | Video gain control not working | Timeout waiting for Transmitter Ready bit in 2681 DUART status register after writing to TX holding register | error |
4259 | A RAM high parity error interrupt occurred | RAM high parity error in ECM module | critical |
425A | RAM low parity error interrupt occurred on ECM module | System boot or operation when ECM RAM parity check fails | error |
425C | Power Unit communication error | Checksum mismatch on power unit packet received by ECM | error |
425D | Power Unit communication error | CRC error bit set during 8530 receive interrupt when reading from power unit | error |
425E | Power Unit communication error | ECM detects Receive Overrun Error bit set when 8530 receive interrupt occurs and End of Frame bit is not set | error |
425F | Bad data received from communication | Console receives corrupted or invalid data during operation | error |
4260 | Power unit reset in progress or recent reset occurred | ECM detects power unit in reset state during operator command or packet transmission | warning |
4261 | ECM detected a loss of power to its CPU | CPU power rail dropped or disconnected | critical |
4262 | Software error: data byte count mismatch during power unit database download to ECM | Power unit database download command with mismatched byte count | error |
4263 | Divide by zero operation or uninitialized NV parameter | ECM attempted divide by zero operation or database error | error |
4264 | Power unit failed to complete reset and initialization within ten seconds | ECM sends reset command to power unit; RESET DONE bit not set within ten seconds | error |
4265 | Maximum exposure time reached; exposure backup timer timed out | Exposure reaches maximum allowed time for technique and is terminated by backup timer rather than AEC LUT sequencer hardware | warning |
4266 | ECM software error due to invalid AEC gain command | ECM receives load AEC gain command (097H) with one or both data bytes equal to zero | error |
4267 | AEC LUT address exceeds maximum value (1299) after exposure termination | Upon termination of AEC exposure, if LUT address is greater than 1299 (4 sec exposure) | error |
4268 | Prep/expose switch transition interrupt with redundancy bit mismatch | ECM receives prep/expose switch transition interrupt; switch status bits from two separate storage locations do not match | error |
4270 | Software error: ASC told ECM that latch through was set but 2.1 Second Trail Delay Enable bits were not set in ETL configuration latches | Record mode acquisitions attempted | error |
4271 | Last exposure may be bad. Proceed with caution/call service. | Offset on brightness signal exceeds 1.5 volts (153 when scaled by 102) | error |
4272 | Exposure time for AEC type exposure was less than zero | Calculated AEC LUT time + tail time minus anticipation time results in negative or zero value | error |
4273 | Exposure time calculated for fixed time exposure was less than zero (expose time + tail time – anticipation time) | ECM module calculates negative exposure time during fixed time exposure setup or execution | error |
4274 | Backup timer mode not configured correctly for primary timed exposure; prevents proper tube heating exposure time calculation | Timer mode programmed as value other than F1–F5 in ECM backup timer register | error |
4275 | Total prep/expose switch state is out of range (valid range zero to fifteen) | Invalid value received for prep/expose switch state on ECM module | error |
4276 | Recipe poke attempted past end of recipe | ASC tried to poke a value into the recipe at a RIP index greater than the size of the RIP | error |
4277 | Power unit communication error requiring system recovery | Detected during power unit operation or reset sequence | error |
4277 | Bad data received from power unit; DMA packet less than 5 bytes | ECM receives incomplete DMA packet from power unit during communication | error |
4278 | Selected exposure time is greater than device will allow | Exposure terminated by loss of expose gate rather than exp/rad timer, such as film changer expose time incompatible with SEP programmer settings | warning |
4279 | ASC sent initial AEC expose time outside acceptable range (less than 5 msec or greater than 2 sec); ECM AEC backup time algorithm expects initial expose time within this range | Software error during AEC initialization | error |
427A | Database or software error in AEC backup time table calculation resulted in negative divisor | One of the calculations to compute the AEC backup time table resulted in a negative number | error |
427B | Table top position data incorrect | RT bus 8, 9, 10 report unrecognized table position during film changer stepper acquisitions | error |
427C | Filament warmup in progress; exposure not yet permitted | Prep or expose switch pressed within 5 seconds of new filament activation | info |
427D | Bad data received from ASC to ECM | Focal spot change or ABC mode change (Cine/Fluoro/progressive) after Prep Sequence Enable Command (Arcnet 88H) | error |
427E | ABC Algorithm calculated filament current exceeding maximum rating of active filament | Filament current calculation exceeds filament maximum (5.5 or 6.5 Amps) | error |
4280 | STARTWHEN timeout - receptor ready signal not achieved within specified recipe timeout | Particular signal or state not achieved after given amount of time as specified in currently active recipe | error |
4281 | STARTWHEN timeout error indicating a signal or state not achieved within specified time | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
4282 | STARTWHEN timeout: required signal or state not achieved within recipe-specified time | Signal or state does not reach expected condition within timeout period specified in active recipe | error |
4282-42FF | Power unit module Applications software error range | Error in Power unit module during Applications mode | unknown |
4283 | Start when timeout, external synch | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
4284 | Start when timeout, expose gate | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
4285 | Start when timeout, expose rate | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
4286 | STARTWHEN timeout: a signal or state not achieved within specified time from active recipe | Signal or state fails to reach expected condition within recipe timeout window | error |
4287 | STARTWHEN timeout on control '2'; signal or state not achieved within recipe-specified time | Particular signal or state has not been achieved after given amount of time as specified in active recipe | error |
4288 | STARTWHEN timeout on RT bus 9 | Particular signal or state not achieved after given timeout in active recipe | error |
428A | STARTWHEN timeout on RT bus 11 | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
428B | Start when timeout on RT bus 12 | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
428C | STARTWHEN timeout on real-time bus 13 | A required signal or state not achieved within recipe-specified timeout on RTB 13 | error |
428D | Timeout waiting for a particular signal or state from RT bus 14 | Signal or state not achieved within time specified in active recipe | error |
428E | STARTWHEN timeout on RT bus 15 | Particular signal or state not achieved after given time specified in active recipe | error |
428F | STARTWHEN timeout on RT bus 16 | Signal or state not achieved within timeout specified in active recipe | error |
4290 | Start when timeout, timer 8 | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
4291 | STARTWHEN timeout on timer 11 - a required signal or state not achieved within recipe timeout | A particular signal or state has not been achieved after specified time in active recipe | error |
4292 | Start when timeout, timer 12 | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
4294 | Start when timeout, timer 14 | Particular signal or state not achieved after specified time in active recipe | error |
4295 | STARTWHEN timeout, overlap detector | A particular signal or state has not been achieved after a given amount of time specified in the currently active recipe. | error |
4296 | Start when timeout, bit 22 - signal or state not achieved within recipe timeout | Particular signal or state specified in active recipe not achieved within allotted time during exposure startup | error |
4297 | Start when timeout, boost done | Particular signal or state not achieved after timeout specified in active recipe | error |
4298 | STARTWHEN timeout error | Signal or state not achieved within recipe timeout period; causes exposure run termination | error |
4299 | STARTWHEN timeout error; recipe aborted and exposure terminated | Required signal or state not achieved within timeout period specified in active recipe | error |
429A | STARTWHEN timeout; rotor rotating | Particular signal or state not achieved after specified time in active recipe | error |
429B | Start when timeout, rotor at speed | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
429C | STARTWHEN timeout: a required signal or state was not achieved within the timeout specified in the active recipe | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
429D | STARTWHEN timeout error; signal or state not achieved within recipe timeout period | Exposure run initiated when required signal does not reach ECM within specified timeout | error |
429E | Start when timeout, expose sequence enable | A particular signal or state not achieved after timeout specified in active recipe | error |
429F | Start when timeout, prep seq. enable. | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | critical |
42A0 | kVp control error | Voltpac brush does not reach position commanded by ECM during servo lock sequence | error |
42A1 | Selected filament open | Filament current during primary timed exposure not greater than 1.5 A with no inverter fault | error |
42A2 | Filament control fault | STARTWHEN signal not achieved within recipe timeout period | critical |
42A3 | Filament control fault | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | critical |
42B1 | Power unit communication error | Communication failure between ECM and MPPU or VAMP power unit | error |
42B2 | Power Unit communication error | LFX, VAMP or MPPU type power units experience checksum problems or receive packet with undefined bits/data | error |
42B3 | Power Unit communication error | SDLC chip detects CRC-CCITT polynomial mismatch in packet | error |
42B4 | Power Unit communication error | Power Unit (LFX, MPPU, or VAMP) unable to communicate with ECM | error |
42B5 | Power Unit communication error | Loss of communication between LFX/MPPU/VAMP power unit and ECM (main controller) | error |
42B6 | Grid-Cathode short error | Grid-cathode circuit short detected during high voltage operation in Cine mode | critical |
42B7 | Power unit soft start error | MPPU or VAMP type power units experience soft start failure | error |
42B8 | Power unit sequence error | Reported by MPPU or VAMP type power units during power sequencing | error |
42B9 | Power unit tube select error | HV switch fails to commutate or compressed pins prevent proper cable contact during tube selection | error |
42BA | Power unit KV range error in X-ray system | Only reported by MPPU or VAMP type power units when KV range error is detected | error |
42BB | Power unit not operable | Reported by MPPU or VAMP type power units only | error |
42BC | kVp control not operable | LFX, MPPU, or VAMP type power unit malfunction | error |
42BD | Power unit regulation selection error | Power unit selection or configuration mismatch in MPPU or VAMP type power units | error |
42BE | Power unit inverter 1 failure | Inverter 1 fault detected in power unit (MPPU or VAMP type) | critical |
42BF | Power unit inverter 2 failure | Detected during self-test in MPPU and VAMP type power units | error |
42C0 | Power Unit communication error | Generator does not measure mA during exposure, or mA jumper not re-installed after calibration on Model Period 11 MP phase 2 | error |
42C1 | Power Unit communication error | LFX, MPPU, or VAMP power unit communication failure | error |
42C2 | Power Unit communication error | LFX type power unit loses communication with main console | error |
42C3 | Overload detected in power unit mA (tube current) limiting circuit | Tube current exceeds safe limits during exposure; reported by LFX, MPPU, or VAMP power units | critical |
42C4 | Tube anode not rotating | Anode rotation control failure detected by LFX or MPPU or VAMP power units | critical |
42C5 | Maximum tube housing heat units reached | Tube housing thermal limit exceeded; rotor controller thermal limit (42DE) or inverter thermal limit (42DF) may present similarly | error |
42C5 | Tube pressure sensor voltage out of specification | Voltage at tube pressure point not 24V during system operation | error |
42C6 | Filament drive error - auto-reset in progress | Filament drive system detects error condition; system attempts automatic reset | critical |
42C7 | Filament drive error | Filament drive circuit malfunction detected in power unit (LFX, MPPU, or VAMP type) | error |
42C8 | Filament drive error; auto-reset in progress | Filament heating problem during LFX, MPPU, or VAMP power unit operation; only serviced when inverter A controls radiation | error |
42C9 | Filament drive error | Power unit detects filament drive circuit failure or DC offset error during operation when inverter B is controlling radiation | error |
42CA | mA overload occurred | Power unit detects current exceeding safe limits during prep/expose switch transitions | error |
42CAH | Tube spit error - X-ray tube arcing or sputtering | X-ray tube arcing detected causing excessive mA feedback | error |
42CB | Instruction not understood or power unit communication failure | LFX/MPPU/VAMP power unit error; SCPU VRTX error, floating point error, or corrupted error table | error |
42CC | Power unit memory corrupted | Reported by LFX and MPPU and VAMP type power units when power unit memory is corrupted | critical |
42CD | Power unit communication error | Communication failure between console and LFX type power unit | error |
42CE | Bad data received from power unit | LFX type power units only; data communication error | error |
42CF | Incoming line voltage low | System detects low incoming AC line voltage; only reported by LFX type power units | warning |
42D0 | Power unit TIRC error | Power unit communication or control error in MPPU or VAMP units | error |
42D1 | Power unit tirc error | Power unit communication or control failure | error |
42D2 | Power unit tube select error | Reported by MPPU and VAMP type power units; tube selection circuit fails or communication lost | error |
42D3 | Power unit tube select error | Power unit selection failure during operation | error |
42D4 | Invalid power unit error code | Power unit malfunction detected by MPPU or VAMP type power units | error |
42D6 | Input power loss | One phase of the network has dropped to zero | critical |
42D7 | Power Unit Fault. High voltage unavailable | Power unit failure detected, high voltage supply not available | critical |
42D8 | Power Unit Fault - High Voltage Unavailable | Mini Bridge switching feedback incompatible with selection or fuse/switch failure | critical |
42DE | Maximum rotor controller temperature reached | Rotor controller temperature exceeds limit | warning |
42DF | Maximum power unit temperature reached. Allow time to cool. | Inverter thermal limit reached in SCPU type power units | error |
42E0 | Power unit self test error occurred | Self-test failure in MPPU or VAMP type power units | error |
42E1 | Power unit tube spit recovery | X-ray tube arcing or spit event | warning |
42E2 | Power unit missing current on filament A | Only reported by MPPU and VAMP type power units when filament A current missing | error |
42E3 | Power unit missing current on filament B | Only reported by MPPU and VAMP type power units when filament B current missing | error |
42E4 | Power unit tube spit recovery | Tube spit detected during exposure; kV automatically cut off with automatic retry | warning |
42E5 | Power unit tube spit recovery | Tube spit detected during X-ray exposure; kV automatically cut off with automatic retry | warning |
42E6 | Power unit tube spit recovery during exposure | During X-ray exposure when tube spit event detected | error |
42E7 | Power Unit Tube Spit Recovery | kV automatically cut off during tube spit event | error |
42E8 | Power unit tube spit recovery | kV is cut off during tube spit; automatic retry performed; after 7 spits during same exposure | warning |
42E9 | Power unit tube spit recovery | Occurs during tube spit event when kV automatically cut off; detected after automatic retry attempt | warning |
42EA | mA meter limit | Power unit (LFX, MPPU, VAMP, SCPU) mA meter saturates during exposure | warning |
42EAH | mA meter saturation detected | mA measurement exceeds meter saturation threshold during operation | warning |
42EB | mA meter error in Power Unit | Failure of mA metering circuit test performed after power up or tube selection | error |
42EC | Closed loop mA error | Reception of error from LFX, MPPU, or VAMP power units | error |
42F0 | Excessive tube current attempted | Attempt to operate tube at current exceeding safe limits | error |
42F1 | Power unit power supply error | Power unit voltage regulation failure detected by MPPU or VAMP power unit | critical |
42F2 | Power unit exposure without command | Power unit receives exposure signal without a prior command | error |
42F3 | Power unit backup mAs error | Power unit backup milliampere-second (mAs) measurement out of specification or unavailable | error |
42F4 | Power unit exposure time exceeded | Power unit exposure exceeds 10 seconds | critical |
42F5 | KVP (kilovolt peak) power failure in SCPU type power unit | Voltage regulation or measurement board circuit error in power unit | error |
42F6 | MA (tube current) failure | Power unit detects tube current supply malfunction | error |
42F7 | Power unit internal error | Reported only by SCPU type power units during operation | error |
42F8 | Overload detected | Power unit detects overload condition | error |
42F9 | Power unit tube spit recovery | X-ray tube unable to complete recovery phase after exposure | warning |
42FA | Power unit TIRC error | Reported by MPPU and VAMP type power units | error |
42FB | Tube casing water flow not detected | Water flow sensor in tube casing circuit fails to detect circulation | error |
42FC | Power unit no KV or KV ripple | kV does not reach 12 kV within 500 microseconds after start of exposure | error |
42FD | Power unit inverter overheat | Inverter temperature exceeds safe operating limit | warning |
42FE | Power unit KV balance fault | Detected by MPPU or VAMP type power units when kV regulation fails | error |
4300 | Bad data received | Console receives invalid application command in GE bus packet from ASC | error |
4300-43FF | Console module Applications software error range | Error in Console module during Applications mode | unknown |
4301 | Bad data received from ASC in GE bus packet | Console receives invalid video format command | error |
4302 | Bad data received | Console receives a video format command of invalid length in a GE bus packet from the ASC | error |
4303 | Bad data received from ASC to console via GE bus packet | Console receives invalid display command character in GE bus packet from ASC | error |
4304 | Bad data received from ASC; invalid video display character in GE bus packet | Console receives malformed data packet from the Acquisition and Storage Controller (ASC) | error |
4305 | Bad data received from ASC to Console | Console receives invalid GE bus command from ASC to position video cursor | error |
4306 | Bad data received from ASC to Console | Console receives invalid GE bus command from ASC to selectively erase video display | error |
4307 | Bad data received from ASC to Console | Console receives invalid GE bus command from ASC to erase video display | error |
4308 | Console received invalid GE bus command from ASC to disable touch screen or trackball reports | Invalid data transmission from ASC to Console on GE bus | error |
4309 | Bad data received from ASC to Console | Console receives invalid GE bus command from ASC to enable touch screen or trackball reports | error |
430A | Bad data received from ASC to Console | Console receives invalid GE bus command from ASC to configure handswitch signals | error |
430B | Bad data received from ASC to Console; invalid GE bus command for programmable touch screen row template | Console receives invalid GE bus command from ASC during touch screen row template load | error |
430C | Invalid GE bus command received by Console | Console receives invalid GE bus command from ASC to sound audible beep tone | error |
430D | Bad data received from ASC to Console | Console receives invalid GE bus command from ASC to sound audible error tone | error |
430F | Invalid box erase coordinates | Attempt to erase a box with invalid coordinate parameters | error |
4380 | Touch screen may not be working | Console detects a failure in communications with, or performance of the optical touch frame | error |
4400 | Communication problem between console and DX/DLX module | Communication failure detected | error |
4400-44FF | Advantx DX Applications software error range | Error in DX module during Applications mode | unknown |
4401 | Communication problem between console and DX/DLX module | Loss of communication between Advantx console and DX/DLX module | error |
4402 | Communication problem between console and DX/DLX module | DX/DLX module communication failure | error |
4404 | Digital system is out of disk space | Storage capacity exhausted on DX/DLX module | error |
4405 | Digital system maximum run length reached | Maximum allowable continuous digital acquisition time exceeded | error |
4406 | Digital system image disk is write protected | System detects write protection on image disk | error |
4407 | Digital system image disk fault | DX/DLX subsystem error | error |
4408 | Max number of sequences for this patient reached | Attempt to create additional sequences for a patient that has reached the system limit | info |
4409 | Max number of categories for this patient reached | Attempting to add more categories than system allows for current patient | info |
440A | Max sequences on digital system reached | Maximum number of sequences stored on digital system | warning |
440B | Digital system offsets failed | Communication error between DLX and VIC modules or VIC module A15 circuit board failure | error |
440C | Digital system losing too many images due to insufficient compression ratio | Image compression cannot maintain required ratio; digital storage system cannot keep pace with acquisition | error |
440D | Trial too bright | X-ray image overexposure during trial exposure | warning |
440E | Trial image too dark | Image acquisition returns signal below acceptable brightness level | warning |
440F | Images being recorded to Cine film only | Cine film recording mode active without alternative recording media | warning |
4410 | Communication problem between console and DX/DLX module | DX/DLX error that propagates to Advantx console | error |
4411 | Front and lateral trial images too bright | Exposure technique or filtration generates overexposed trial images | warning |
4412 | Front trial image too bright, lateral trial image too dark | Exposure technique or filtration settings produce inconsistent image brightness between frontal and lateral projections | warning |
4413 | Front trial image too dark, lateral trial image too bright | Radiographic trial images expose incorrectly | warning |
4414 | Front trial too dark, lateral trial too bright | Image acquisition results in improper exposure levels | warning |
4415 | DX store function in progress on the in-room monitor | Store function is active during DX/DLX operation | info |
4416 | DX recall function in progress on the in-room monitor | Recall operation initiated on DX/DLX module | info |
4417 | Digital System Error | Digital system malfunction in DX/DLX module | error |
4500 | DR busy/fault exists | DR Module failed to respond during power up or DR module reported an internal error during command execution or status reporting | error |
4500-45FF | Advantx DR Applications software error range | Error in DR module during Applications mode | unknown |
4502 | System communications problem between ASC and DR BIU | Fluoro switch depression; DR Module fails to set DR-RDY signal high within 50 msec after FLUORO SWITCH signal is set low | error |
4503 | System communications problem | DR BIU to DR Module communication failure, ASC to DR BIU data validation failure, or ASC pass-through command error | error |
4504 | System communications problem | Serial communication packet transmission validation failure; STX code not found in first position of DR Module BIU packet | error |
4505 | System communications problem between DR BIU and DR Module | Serial communication packet validation failure; calculated last packet location does not contain ETX code (3H) | error |
4506 | System communications problem - message queue overflow | DR BIU processor detects 20 stacked messages waiting to transmit via serial link to DR Module | error |
4507 | System communications problem | Serial link Duart (U133) transmitter fails to become ready after specified period | error |
4508 | System communications problem between ASC and DR module | Serial link Duart detects parity error in character reception from DR Module | error |
450A | System communications problem between DR BIU processor and DR Module | DR Module fails to acknowledge command (ACK character) within 20 msec after ETX transmission; retransmit attempts fail after 3 tries | error |
450B | System communications problem | DR Module fails to set DR-RDY signal low within 10 msec after PREP SWITCH signal | error |
450C | Console displays: System communications problem. Please try again. | DR-RDY signal (Pin 7 of U133) not set low within 10 msec of FLUORO SWITCH signal going low | error |
450E | DR function inoperative | DR Module fails to respond to INT-SYNC signal within approx. 0.5 seconds by setting SYNC-ACK signal appropriately | critical |
4510 | Enter new study information and continue | DR module rejected New Study Command from console | error |
4511 | System communications problem. Please try again. | DR Module fails to respond to EXPOSE SWITCH signal from DR BIU within approximately 1 millisecond | error |
4512 | System communications problem between DR BIU and DR Module | DR Module fails to respond within specified time frame after ETX command transmission | error |
4513 | System communications problem | Serial link Duart (U133) detects overrun of data in receiving Fifo | error |
4514 | System communications problem | DR Module fails to acknowledge command from DR BIU processor within 20 msec or sends NAK after 3 retransmit attempts | error |
4515 | System communications problem | DR BIU processor detects bad values for certain parameters in its database during warm start initialization | error |
4517 | XY shift failure | System in service (Cal) mode; DR Module response indicates problem with downloading x-y shift parameters for blanking circle | error |
4518 | Circle adjust failure | Calibration attempted during service (Cal) mode when downloading blanking circle size parameter to DR Module | warning |
4519 | Smooth filter failure | DR Module's response to downloading coefficients for filter number 2 indicates a problem with transmitted command | error |
451A | Sharp filter failure | System in service (Cal) mode; DR Module response to coefficient download for filter 1 indicates problem with transmitted command | error |
451B | Maximum number of studies (50) defined on the DRS (Dig Rec module) | ASC sends DRS command to define 51st new study | error |
451C | Maximum number of images have been stored in this study | 99th image reached during acquisition | warning |
451D | Dig Rec temperature too high | MHR temperature rises above operational level | warning |
451E | No storage of images on the Digital Recording disk is available | Power-up test failure; disk storage unavailable for image acquisition | error |
451F | DRS (Digital Radiography System) Power-Up Test Failure | DRS power-on self-test failure; system enters degraded mode | critical |
4520 | Digital imaging system failure in DRS module | Power-up test failure in Digital Radiography System | error |
4521 | Filters and post-processing functions not available | Power-up test failure | error |
4522 | Power-up test failure in DRS module | System power-up self-test | error |
4523 | Digital Filming is not available | Power-up test failure in DRS module | error |
4524 | Communication problem encountered | Communication failure between system modules | error |
4525 | Communication problem encountered | Communication failure between system components | error |
4526 | Communication problem encountered | Communication failure between system components | error |
4600-46FF | Cine module Applications software error range | Error in Cine module during Applications mode | unknown |
4602 | Invalid frame rate received by CINE module from ASC | Module receives invalid frame rate from ASC in set frame rate command | error |
4603 | Invalid phase received or camera running during set phase command | Module receives invalid phase from ASC in set phase command or camera is running when command received | error |
4604 | CINE module software thinks camera is already running when commanded to operate | CINE module receives expose command while software believes camera is running | error |
4605 | Software state mismatch on CINE module during film transport request | Request for film transport received from ASC while software thinks camera already running | error |
4606 | Jog not allowed during Cine recording | Enable cine trial command received while software running camera | error |
4608 | System communications corrupted | Communications failure between console and in-room display or modules | critical |
4609 | Instruction not understood | User command not recognized by system | warning |
460A | Cine camera out of film | INTLOK bit is off on port 0 and film remaining A/D value is less than calibrated no-film value | info |
460B | Cine magazine missing or not locked | INTLOK bit off on port 0 and film remaining A/D below 8 counts | error |
460C | Cine camera fault; camera interlock error indicating camera cannot run | INTLOK bit read as zero by CPU on port 0 during prep or camera start | error |
460D | Cine camera overheated. Heat protection timeout timer expired. | Heat protection timeout timer (80188 timer 1) expires, generating interrupt to stop camera. | warning |
460E | Cine mode fault | When attempting to run the camera and Power-amp board indicates fault via PAFAULT* signal on port 0 | error |
460F | Cine film jammed in camera; camera stalled and not reaching speed within timeout | Camera fails to reach 98% expected speed within timeout; heat timer interrupt fires while camera below speed | error |
4610 | Camera still hot - thermal protection active | Attempting to start camera when heat protection software detects excessive camera temperature | warning |
4611 | Cine mode fault; wrong command sequence from ASC or software malfunction during prep | Improper exposure setup command sequence; camera motor on commands not received | error |
4620 | Cine camera not line locked | Hardware phase lock loop fails to lock during camera startup; LOCK signal remains unasserted after stall timer expires | warning |
4621 | System communications corrupted | MAST-OSC signal unstable or not close to final speed after five seconds pre-run check | error |
4630 | System communications problem | Module tried to transmit packet to node and did not receive transmit acknowledge before timeout | error |
4631 | Improper updating of packet indexer; indexer went past end of packet after processing command due to insufficient packet data | System communications problem during packet processing | error |
4632 | Software attempted to transmit more data than would fit in a single packet | Data transmission exceeds packet capacity | error |
4633 | GE Bus transmit queue full; module software lost or ARCNET not transmitting | Software internal GE Bus transmit queue was full when module attempted to send packet | error |
4634 | ARCNET hardware could not find designated node on system to transmit packet to | ARCNET hardware indicates transmitter available but no transmit acknowledge given before timeout | error |
4635 | Instruction not understood. Please try again. | GE bus transmission received with command not found in command table | error |
4636 | Software task scheduling failure on system timer | System attempted to schedule a task on the timer with no available slots in the list | error |
4637 | Instruction not understood. Timer ID outside scheduled task list. | Attempt to update or flush scheduled task with timer ID outside list of scheduled tasks | error |
4640 | Instruction not understood | User input not recognized by system | warning |
4641 | Instruction not understood | User input not recognized by system | error |
4642 | Instruction not understood | User input not recognized by console | warning |
4700 | Instruction not understood | Command received across GE Bus (ARCNET) is not a defined command for VIC module | error |
4700-47FF | VIC module Applications software error range | Error in VIC module during Applications mode | unknown |
4701 | ARCNET transmission attempted to non-existent node | TA was set and TMA was not set | error |
4702 | System communications problem | ARCNET transmission failed; destination module did not acknowledge reception within 2 seconds | error |
4703 | GE Bus command interpreter packet index overrun; number of bytes transmitted did not match expected number | VIC module receives command packet with incorrect data byte count | error |
4704 | Packet transmission exceeds maximum transmittable size | Software attempts to transmit a packet larger than the maximum allowed | error |
4705 | Integer math overflow occurred | Math overflow in VIC module | error |
4706 | Divide by zero error in VIC module | Computational fault in VIC circuit board | error |
4707 | Bad data received from console | Data integrity or communication failure between console and VIC module | error |
4708 | Bad data received from GE Bus command with invalid parameter outside valid range | GE Bus command parameter outside valid range for the command | error |
4709 | TV gain curve data bad | During system operation or initialization when gain linearization data becomes invalid | error |
470A | The GE Bus output queue is full. | GE Bus output queue becomes full | error |
470B | Bad data received from console | Attempt to schedule a time-delayed task using SCHEDULE_SOFT_TIMER_TASK with a timer already allocated to another task | error |
470C | TV camera power supply failed | Interrupt received indicating warm power supply has failed | critical |
470D | TV camera power supply failed | Interrupt received indicating cold power supply has failed | critical |
470E | Photo cathode current read attempted during AGC | Attempt to sample photo cathode current with AGC enabled | error |
470F | AGC attempted during photo cathode current read | Attempt to enable Automatic Gain Control Function while Photo Cathode Sample function was active | error |
4710 | Photo cathode current read already active | Attempt to sample photo cathode current while previous sample command still executing | error |
4711 | Instruction not understood. Please try again. | Attempt to flush or set a soft timer greater than the number of timers configured | error |
4712 | Instruction not understood. Please try again. | Attempt to schedule a task when no more timers available | error |
4713 | Video setup corrupted | System warm state data retrieval failure with corrupted parameters | warning |
4720 | Video camera power supply has failed | Interrupt received indicating camera power supply failure | error |
4721 | Video camera system inoperative | Camera sweep fail occurred | error |
4722 | Video camera system inoperative | Camera hardware generates interrupt indicating camera sync fail | error |
4723 | Video camera system inoperative | Camera phase lock loop failure occurred | error |
4724 | Camera A/D converter did not signal completed conversion within timeout period (40+ milliseconds) | A/D conversion timeout on camera interface | error |
4725 | TV Video signal out of range | Setup adjust task unable to achieve readable video signal range | error |
4726 | Video shading correction fault | Amount of shading compensation required exceeds hardware capability | error |
4727 | Light bias inoperative | VIC initialization detects signal current change within 10 A/D bits when DAC output is varied, or when non-zero light bias is requested | error |
4728 | Light bias inoperative | Requested signal current from light bias cannot be achieved; DAC at limit | error |
4740 | Video aperture did not respond | Iris electronics did not generate interrupt indicating iris reached requested position within 1 second allowable time | error |
4741 | Preset value requested exceeds maximum aperture value | Aperture preset command sent to VIC when requested value is larger than Max Aperture Command limit | error |
4742 | Video Gain Command (VGC) received over RT Bus serial link from ECM was outside allowable range | VGC received with value outside 66H through 0FFFH range | error |
4743 | Aperture in Position interrupt received when none was expected | Unexpected aperture position interrupt signal from VIC module | warning |
4744 | IRIS A/D conversion interrupt not received within timeout | IRIS AD RDY interrupt not received within 10 milliseconds of conversion start | error |
4780 | Optical distributor not positioned | Gate rotate failed to position in time. At initialization, electrical in position bit not received within 2 seconds, or during move bits not lost within 100 ms or do not reappear within 1 second. | error |
4781 | Optical distributor not positioned | In position bits go true then false within 100 milliseconds during normal gate movement | error |
4782 | Optical distributor not positioned | Neutral Density Filter shows both position bits true (close to in bit and close to out bit from XRII I/F board both logic low) | error |
4783 | Optical distributor not positioned | Optical distributor positioning failure during initialization or operation | error |
4785 | Optical distributor not positioned | System detects optical distributor is not in correct position during operation | error |
4786 | Optical distributor not positioned | Gate pivot failed to position in time | error |
4787 | Optical distributor not positioned | Gate pivot amp indicated stall at bad time | error |
4788 | Optical distributor not positioned | System detects optical distributor is not in correct position | error |
4789 | Optical distributor not positioned | Gate clock watchdog timer timeout after 10 milliseconds without gate clock reset | error |
478A | Gate and/or filter motion not allowed due to XRII interface link failure | Serial link to XRII I/F board is down | error |
478B | Gate pivot or ndf did not recover | Gate and/or ND filter attempting to recover from a prior error and could not successfully complete the recovery | error |
478C | Gate stall hardware latch kicked in | Stall watchdog detected | error |
4790 | Output buffer containing messages to XRII I/F micro-controller is full | Communications with XRII I/F not proceeding normally | error |
4791 | XRII I/F micro-controller has reported an error | Bad data received by console from XRII interface | error |
4792 | Bad data received from console | Data transmission error between console and system | error |
4793 | Requested field of view in XRII command not in valid range | Invalid XRII field of view command sent to system | error |
4794 | XRII I/F micro-controller did not respond appropriately to VIC within prescribed time | XRII I/F micro-controller communication timeout | error |
4795 | DUART framing error on XRII interface communication channel | Framing error detected on DUART channel A communicating with XRII I/F Board microcontroller | error |
4796 | Bad data received from XRII interface | DUART on aperture board reports parity error on channel A communicating with XRII I/F micro-controller | error |
4797 | DUART overrun error on XRII & Iris Control Board channel A | Characters received from micro-controller faster than removed from input queue | error |
4798 | XRII I/F board 8031 microcontroller power-up diagnostics failed | Power-up self-test of microcontroller on XRII I/F board detects error | error |
4799 | Bad data received from 8031 microcontroller on XRII I/F board | Response byte from 8031 does not match command sent to it | error |
47B0 | TV Rotator failed to track command | Rotator unable to respond to positioning commands | warning |
47D0 | Video Tape Recorder did not respond to command | VTR did not respond to a command from the system | warning |
47D1 | VTR Communication error | VTR sent a NAK to the VIC | error |
47D2 | VTR not in remote mode | VTR front panel used or tape becomes unthreaded | warning |
47D3 | VTR Communication error | Communication failure between console and VTR module | error |
47D4 | Search not confirmed; tape search did not complete | Attempted search for mark or run too close to tape beginning or end, or cue/record attempted before valid runs exist | warning |
47D5 | Freeze deselected due to inactivity | VTR left in freeze frame state for 30 minutes | info |
47D7 | VTR communication error | VTR module communication failure | error |
47D8 | VTR communication error | VTR software attempted to stimulate Action-State driver with invalid stimulus | error |
47D9 | VTR communication error | VTR responded with invalid transport mode | error |
47DA | HPR–50 VTR left idle in ready state for more than 30 minutes | VTR remains in ready state without use for over 30 minutes | warning |
47DB | VTR communication error | VTR DUART detected a framing error | error |
47DC | VTR communication error | VTR DUART detected parity error during communication | error |
47DD | VTR communications error | VTR DUART detected an overrun error | error |
47E0 | Unknown | Unknown | unknown |
47FE | Invalid data base item ID number | Database access error on VIC module | error |
47FF | Invalid data base item size | Database item size validation failure | unknown |
4800 | Communication problem between console and display module | Undefined command received by DF module | error |
4800-48FF | Digital system Applications software error range | Error in Digital system during Applications mode | unknown |
4801 | Communication problem between console and in-room components | Arcnet packet received with legal command but incorrect byte count | error |
4802 | Communication problem between console and DF (Digital Fluorography) module | Application attempts exposure sequence when DF unable to acquire images | error |
4803 | Communication problem between DF module and ASC | DF module receives packet with invalid application mode (neither Static nor Dynamic) | error |
4804 | Communication problem between DF module and console | DF module receives packet indicating acquisition mode change during active acquisition | error |
4805 | Fatal error in DF (Digital Fluoroscopy) module detected | DF module detects fatal error condition | critical |
4806 | Insufficient disk space available for acquisition | Available disk space drops below 2 seconds of run time during acquisition | warning |
4807 | DF maximum run length of 999 images was reached | Current run sequence attempted to store more than 999 images | info |
4900 | X-ray tube and table receptor alignment mismatch | Table selected as RAD receptor but tube and table receptor not in proper alignment | error |
4900-4BFF | Positioner module Applications software error range | Error in Positioner module during Applications mode | unknown |
4901 | Room door not closed during RAD exposure attempt | RAD exposure attempted with room door open or signal missing | warning |
4902 | Cassette not in place or cassette-in-place signal not presented to positioner | RAD exposure attempted for selected receptor without cassette detected | error |
4903 | Bucky rotation mismatch error | RAD exposure attempt when receptor rotation does not match X-Ray field | error |
4904 | X-ray tube is in an improper orientation | Tube and/or receptor are not in proper alignment for RAD exposure or operator intention cannot be determined | error |
4905 | Wall receptor alignment failure. Tube and wall receptor not in proper alignment for exposure. | Wall unit selected as RAD receptor at console but tube and wall receptor misaligned. | error |
4906 | Table height not at proper exposure position | RAD exposure attempted with variable-height table not at proper height | error |
4907 | Exposure terminated early during RAD exposure with reciprocating grid receptor | RAD exposure attempted with reciprocating grid receptor when grid did not reach speed or speed signal bouncing | error |
4908 | Film not in place, film jam, or patient ID card not in place during RAD exposure with film changer wall device | RAD exposure attempted with Changex or ACS-1000 film changer when film is absent, jammed, or patient ID card missing | error |
4909 | Collimator motion problem detected | Collimator fails to move or respond to commands | error |
490F | Invalid receptor; software corrupted because internal receptor indicator is not acceptable value | System detects unacceptable internal receptor indicator value during operation | error |
4911 | Film changer programmer not ready | Film magazine and/or film receiver not attached to film changer, or FC READY signal indicates not ready after required delays | error |
4912 | Film changer is out of film | FILM IN PL signal at prep and during run indicates no film when expected, film changer indicates not ready on FC READY signal | warning |
4913 | Table not positioned at horizontal for film changer operation | Software detects HPOS0 signal not asserted when prep mode entered | error |
4914 | Stepper switch in wrong position on film changer programmer | Stepper selection switch on PDQ-2 film changer does not match console selection | warning |
4915 | Injector selection switch mismatch between PDQ-2 film changer and console | Injector selected on console but not at changer, or vice versa | warning |
4916 | Film changer programmer not ready | Injector is selected for use and injector is not armed when used with PDQ-2 film changer | warning |
4917 | Remove all cassettes to continue | Unknown from excerpt | unknown |
4918 | Collimator failure detected | Operator attempts to manually move Rad Ultranet Collimator blades using collimator knobs immediately after tube warm-up or within 30 seconds of warm-up completion | warning |
4919 | Collimator failure detected | Collimator mechanism malfunction or positioning error | error |
4A00 | System communications problem | GE bus transmission received but command not found in command table | error |
4A01 | Bad data received from Positioner module via ARCNET | Software internal GE Bus transmit queue became full when attempting to send packet, indicating module software failure or ARCNET communication failure | error |
4A02 | System communications problem - transmit timeout on Positioner module | Positioner module attempted to transmit packet and did not receive acknowledge before timeout expired | error |
4A03 | Bad data received from console | Software attempted to schedule a task to run and an internal timer indicator was allocated to a different scheduled task | error |
4A04 | System communications problem | Software attempted to transmit more data than expected to fit in a single packet | error |
4A05 | System communications problem | Communication failure between console and Positioner module | error |
4A06 | Improper updating of packet indexer; indexer went past end of packet after processing command | Packet had less data than expected during command processing | error |
4A07 | ARCNET transmission failed; destination node not found on network | ARCNET transmitter available but no transmit acknowledge received before timeout | error |
4A08 | Bad data received from Positioner module during scheduled task update/flush | Attempt to update or flush a scheduled task with a timer id outside the list of scheduled tasks | error |
4A09 | Bad data received from Positioner module; system timer task scheduler overflow | Software attempts to schedule task on system timer when list is full | error |
4A0A | A/D conversion did not complete in reasonable time on Positioner module | A/D conversion complete bit not found set within expected time period when reading A/D on interface board or collimator board | error |
4A0B | System communications problem | Part or all of data from GE bus packet is not valid for the command | error |
4A10 | Bad data received from console | Problem transmitting request for download of calibration and configuration data | error |
4A31 | System communications problem | Communication failure between console and in-room display or connected modules | error |
4A32 | System communications problem | Unknown communications error in Positioner module | error |
4A34 | System communications problem | Positioner module communication failure | warning |
4A42 | Bad data received from Positioner module | Data communication error between console and Positioner during operation | error |
4A51 | Tableside control communication failure after reset attempts | Software detects problem communicating with tableside controls; reset diagnostic shows continued problem or many unsuccessful reset attempts | error |
4A52 | Tableside control communication error detected by DUART status register | Error in reception from tableside control detected during system operation | warning |
4A53 | Invalid state transition in Positioner module | Software detects an invalid transition state based on CC select state, auto format select state, and II or Film changer configuration | error |
4A54 | Software could not determine arc sine, cosine or tangent of L & U angles | Positioner module software failure during angle calculation from trig table | error |
4A55 | Room door needs to be closed to continue operation | Room door signal indicated open at Prep time | warning |
4A57 | Image Intensifier not at or below 30cm SID | II position not at or below 30cm from ISO-center during Interchange operation | warning |
4A58 | Film changer rotation interlock | Film changer is in rotated position when system enters prep mode | error |
4A59 | Film changer not correctly locked at end-of-travel position | At Prep time when film changer is not at End-of-travel (EOT) or AUX EOT as indicated by signals | error |
4A5A | Positioner not in valid Cerebral Contact mode position | Attempt to use Cerebral Contact mode when positioner does not meet valid position requirements; one of three position criteria not satisfied | warning |
4A5B | Cerebral Contact mode setup failed Case I/II checks; film changer not properly positioned | Attempted Cerebral Contact mode prep when L/U arm angles or film changer position do not meet valid conditions | error |
4A5C | Film changer position interlock error in Cerebral Contact mode | Positioner does not meet conditions: L angle III–<+45° OR U angle >+45° with film changer fully extended and at auxiliary end-of-travel | error |
4A5D | Non-maskable interrupt occurred due to invalid I/O port access, unacknowledged hardware, or non-existent memory access | CPU Timer 0 expired without acknowledgment | error |
4A5E | CPU detected a math overflow during calculation | Math overflow in positioner CPU during some calculation | error |
4A5F | CPU detected a math attempt to divide by zero in Positioner module | Occurs during system operation or hardware motion that triggers undefined math operation | error |
4A60 | Film Changer Programmer not ready | At prep time when film changer signals indicate programmer not ready or film not in place | error |
4A61 | Film Changer out of film | FC RDY signal not ready and FC FILM IN PLACE SW signal not indicating film present during PREP | warning |
4A63 | Collimator has stalled | LONG STALL and/or TRANS STALL signals active during routine status check | error |
4A64 | Cannot select CC (Continuous Cinefluoroscopy) mode when table stepper is selected | CC button pressed while table stepper is selected or STPR ON signal active | error |
4A65 | CC (Cine Continuous) mode was deselected because table stepper unit power was detected | Internal flag indicates CC is selected while STPR ON signal shows stepper is activated | error |
4A66 | Injector did not inject as programmed | Injector does not appear to inject after being commanded; delayed injection timer expires without INJECTING signal detected | error |
4A67 | FC program is not compatible with console selection | Positioner finds plane selections on console and SEP programmer do not agree at PREP time | error |
4A68 | SEP programmer not indicating any plane selection | Film changer selected as SEP, rad acquisition requested, SEP power-on delay expired, neither CHANGER 1 SEL nor CHANGER 2 SEL signals active | error |
4A69 | Film changer error reported | FC RDY signal went away or FC FILM IN PLACE signal not present | error |
4A6A | Positioner software initialization failure | Positioner module software fails to initialize during system startup or warm boot | error |
4A6B | Collimator problem | During X-ray exposure in CFM/AFM vascular systems | error |
4A81 | Serial link between Vascular positioner module and host communication failed | 5 packets transmitted to Host with no response after successful link initialization | error |
4A82 | Receiver error on host link - Serial Link interface between Vascular positioner module and host has failed; DUART status for received byte was bad | Serial communication error detected between Positioner and host system during normal operation or after initialization | error |
4A83 | Serial link between Vascular positioner module and host not working correctly | Unsequence packet received; high-order bits of data bytes do not toggle in predictable pattern | error |
4A84 | Host not in application mode | Host executed internal mode change during applications mode and reported to Vascular positioner | error |
4A85 | Auto-positioner control problem | Serial link interface between Vascular positioner and autopositioner tableside control has failed | error |
4A86 | Auto-positioner control problem | Serial link interface between Vascular positioner and autopositioner tableside control has failed. Link must have successfully initialized for this error to be reported, implying actual equipment failure. | error |
4A87 | Auto-positioner control problem | Serial link interface between Vascular positioner and autopositioner tableside control fails; invalid byte sent from tableside control | error |
4A88 | Cannot memorize current position; L-arm positioned outside allowable limits | User attempts to store L-arm position while outside permitted range | warning |
4A89 | Timeout: pre-programmed position not activated within 5 seconds | User fails to use selected pre-programmed position within 5 second window after selection | warning |
4A8A | Table top is not centered laterally | Table top detected in non-centered lateral position during system initialization or operation | warning |
4A8B | LP (positioning arm) is in an invalid state for storing pre-programmed positions | User attempts to store a position while LP is not at park or iso-center | warning |
4A8C | I/O driver failed | I/O-driver call within the software failed | error |
4A8D | Unexpected start byte on serial link between Vascular positioner module and host | Serial communication protocol violation: more than one byte missed in packet causes receiver to not detect packet end and misidentify subsequent start byte | error |
4A8E | Missing start byte in serial packet from Vascular positioner module | Next received byte after complete packet is not encoded as valid start byte | error |
4A8F | Module Utilities and Driver package (M.U.D.) reported an internal failure | Bad status received from MUD module | error |
4A90 | Vascular positioner calibration and configuration data transfer failed | Transfer of calibration & configuration data from Vascular positioner module to host across serial link did not complete | error |
4A91 | Gantry operation degraded | Vascular positioner module commanded host to change operating modes but host did not change to required mode | warning |
4A92 | Gantry problem detected by positioner module | Vascular positioner protocol software driver detects internal error condition | error |
4A93 | Gantry link restart in process | Gantry link restart event | info |
4AA1 | Cannot take exposures with LP parked | Lateral Positioner in parked position at PREP stage | warning |
4AA2 | LP control error | Serial link to LP appears down; no response after 20 tries to communicate at 50ms intervals | error |
4AA3 | LP control error detected during data byte reception | Parity, framing, or overrun error in LP data byte reception | error |
4AA4 | Communication problem with LP (Positioner) | Packet parity error from Positioner transmission to console | error |
4AA5 | Communication problem with LP (Positioner) | Unrecognized data byte in last Positioner transmission; transmission start byte indicates error | error |
4AA6 | Communication problem with LP (Positioner) | Data byte received from LP is out of sequence; MSB does not correspond to data byte count | error |
4AA7 | Lateral system not operating. Cannot perform task. | Lateral ASC appears down; Positioner requested cal/config data with two minute timeout and no response | error |
4AA8 | Table not rotationally centered to allow LP motion | TT ROTATNL CNTR signal indicates not centered during LP motion or smart handle requests | error |
4AA9 | Positioners not orthogonal - L angle not at zero degrees | Attempt to move LP to or from vertical position when L angle is not zero | error |
4AAA | Positioner cannot drive to next auto position due to interference | Transmission start byte from LP electronics indicates positioner interference | warning |
4AAB | Attempted auto position operation with LP at isocenter and Frontal plane selected | Auto position operation initiated with LP at isocenter; Frontal plane selected on Smart Handle | error |
4AAC | Exceed travel limit | LP (linear positioner) reaches its travel limit in current direction | error |
4AAD | Drive not responding | LP indicating axis drive failure to positioner in transmission start byte | error |
4AAE | X-ray tube on the LP has collided with something and motion is disabled | Physical collision detected via third data byte from LP | error |
4AAF | II on lateral LP has collided with object; motion disabled | Physical collision detected during II motion on lateral positioner | error |
4AB0 | X-ray tube collision with L/U; motion disabled | X-ray tube on L/U collides with object; collision status indicated in third data byte from LP | error |
4AB1 | Imaging Intensifier (II) collision on L/U; motion disabled | II on L/U collides with object; collision status in third data byte from LP | error |
4AB2 | Type 1 bad interlock. Safety mask installed when it should not be, inhibiting Rad Acquisition. | Safety mask installed on collimator when system expects none | error |
4AB3 | SID Error | Type 1 bad interlock; SID value too small or too large or SID change during fluoroscopy | error |
4AB4 | Collimator error; collimator blades are open too far | Type 1 bad interlock during X-ray exposure or fluoro while changing SID | error |
4AB5 | Collision detection problem | Operating fault detected on collision detection system switch | error |
4AD9 | L-arm outside safety cone during spin examination learning phase | Attempt to record pivot arm spin trajectory while L-arm is outside safety cone | warning |
4ADA | Ready. Clear rotation area. | Between end of trajectory check and beginning of image recording in spin examination | info |
4ADB | Position not allowed | Requested start or stop position refused during Spin and Bolus mode examination | warning |
4ADC | Table Ready | All Bolus Technique learning phases have ended correctly | info |
4ADD | Ready for spin test message displayed to user | A spin trajectory has been memorized | info |
4ADE | Spin test in progress | During spin test procedure execution | info |
4ADF | Spin set-up aborted | Software detects a condition that is not part of an allowable spin sequence | error |
4AE0 | Hardware error detected on the Stepping Board from the Positioner | Hardware malfunction on Stepping Board | error |
4AE0 | Table problem. Operation degraded, continue with caution. | Table operation fault detected | error |
4AE1 | LC or Tilt-C gantry subsystem board initialization failure | Gantry board not responding to initialization command at power-up or reset | error |
4AE1 | Gantry problem | Gantry system error during operation | error |
4AE2 | Gantry problem. Operation degraded, continue with caution. | LC or Tilt-C CPU software detected an internal error condition | warning |
4AE3 | Gantry problem with degraded operation | LC or Tilt-C gantry internal EEPROM storage failure | warning |
4AE4 | Gantry problem; LC or Tilt-C CPU watchdog timer failed during runtime | LC or Tilt-C CPU watchdog timer failure during system operation | error |
4AE5 | Gantry problem. Operation degraded, continue with caution. | LC or Tilt-C gantry reports error detected at host end of Vascular positioner module serial link | warning |
4AE6 | Gantry collision occurred | Second-level collision or limit switch was hit | error |
4AE7 | Increase SID for full display | SID reduced to point where collimator cannot track with 40cm image intensifier | warning |
4AE8 | Stepper problem in positioner | Stepper motor or related control circuit failure | error |
4AE9 | Position not allowed. Table too close to electrical and/or software limits. | In Auto Positioner mode when table position prevents safe trajectory completion | warning |
4AEA | Console/In-Room Display: Table or gantry not in authorized position. | System detects table or gantry position outside allowed range | error |
4AEB | Table is moving to the start position | Positioning operation initiated | info |
4AEC | Position attained | Programmed auto-positions have all been reached | info |
4AF0 | Gantry problem. Operation degraded, continue with caution. | LC or Tilt-C gantry subsystem boards report communications problem | warning |
4AF1 | One of the LC, LCA or Tilt-C gantry subsystem boards has reported one of several possible error conditions | Error conditions on gantry subsystem boards; multiple errors logically OR-ed together | error |
4AF1 | Gantry problem with operation degraded | Gantry positioning system detects fault or limits operator capability | warning |
4AF2 | Positioner move command incomplete or failed | Positioner cannot complete 90-degree (3-degree tolerance) lateral move | unknown |
4AF3 | Positioner cannot obtain required LAO/RAO angle of 15 degrees | Positioner fails to reach 15° LAO or RAO position | unknown |
4AF4 | Obtain CRA/CAU 3 deg. | Positioning request to Cine Radiography Angle / C-Arm Utilization at 3 degrees | unknown |
4AF5 | Bolus filters problem | Positioner module detects bolus filter malfunction | error |
4AF6 | General positioner module software error | Positioner module software detects an internal error condition | error |
4AF7 | Emergency Stop depressed | Emergency Stop Switch actuated | warning |
4AF8 | Gantry power or enable failure | Gantry problem detected | error |
4AF9 | Gantry DMT (Lifetest) failure | DMT cycle failed during gantry operation | error |
4AFA | Positioner Console Fault | Communications failure with Remote or Inroom Console | error |
4AFB | Gantry Link Fault | Communications failure with Gantry | error |
4AFC | Hardstop Bypass Error | Emergency Stop is actuated | error |
4AFD | Collision recovery failed | First Level Collision backout did not clear collision | error |
4AFF | Film changer in working position | Film changer placed in working position | info |
4B01 | Film changer collision detected on PUCK film changer | Collision detected during film changer positioning | error |
4B07 | Smart handle problem | Permanent command detected on one axis of smart handle A or B for one minute | error |
4B10 | DUART status register error for character received on channel A | Character received on DUART channel A with error status | error |
4B11 | Spot Film Device function unavailable due to DUART character reception error | DUART status register indicates error for character received on channel B | error |
4B12 | Spot Film Device function unavailable | Serial link to SFD panel is not active even after reset attempt; polling fails within timeout | error |
4B13 | Spot Film Device function unavailable | Serial link to Remote panel not active after reset attempt; polling timeout detected | error |
4B14 | Unsolicited input received on serial link | Character received that looks like a command but is unexpected; can occur on power-up in pre-MP6 systems | warning |
4B15 | Spot Film Device function unavailable | Software corruption; invalid serial link identifier value | error |
4B16 | Room door interlock error - room door does not appear to be closed | Exposure request made with room door open or door signal indicating open state | warning |
4B17 | Spot Film Device (SFD) not in forward position | Prep request initiated while SFD POS SW signal indicates SFD is not forward | warning |
4B18 | Object in front load port | Object detected blocking front sensor(s) during fluoroscopy exposure request | error |
4B19 | Rear door open on Spot Film Device | REAR DOOR SW signal reads low when exposure is requested | critical |
4B1A | Cassette not detected in Spot Film Device | Prep request made without cassette loaded | error |
4B1B | Cassette being loaded. One moment please. | Exposure request while cassette is in middle of loading or unloading | info |
4B1C | Invalid format selected for current cassette | Prep request with invalid format selection | warning |
4B1E | System communications problem | Command received out of expected sequence | error |
4B1F | System communications problem | Command received out of expected sequence | error |
4B20 | Obstruction may be in way of fluoro image | Carriage is stuck when fluoroscopy is requested | warning |
4B21 | Cone not parked at startup | System startup when cone is not in valid position (Front or Park) | error |
4B22 | 80188 Internal Timer fault | Detected during system diagnostics | unknown |
4B22 | Servo axis not maintaining proper position | One of the servo axes jumps out of the position window and does not return within timeout period, or repeatedly jumps in and out | error |
4B23 | Table locks non-functional | Lock did not respond as expected: requested to set but did not, or set when not requested | error |
4B24 | Power Assist non-functional due to thermal overload | Power Assist board generates OVERTEMP signal over Multibus | error |
4B25 | Power Assist clutch has failed or is non-functional | Power Assist clutch either did not set when requested, set when not requested, or set but then failed | error |
4B26 | Cone button operation attempted with Horizontal Stop switch in OUT position | User attempts to use Cone button while Horizontal Stop switch is OUT | error |
4B27 | Invalid format selection for cassette | Attempting to select a format that is not currently valid for the cassette | error |
4B28 | Wrong size cassette detected | SFD unable to recognize loaded cassette; cassette may be jammed or invalid size | error |
4B29 | Cassette currently in system is thought to be fully exposed and cannot be used | Prep request when internal flags indicate cassette is fully exposed | warning |
4B2A | Selected format has become invalid due to film depletion | During Prep when currently selected format becomes invalid based on internal flags indicating film usage | warning |
4B2B | Patient Step or Overhead Obstruction Preventing Angulation | Angulation requested while table step is extended | error |
4B2D | Table top motion non-functional | Table top longitudinal motion requested but no movement detected after timeout period | error |
4B2E | Cassette carriage jammed | Expected servo motion did not complete in time; servo appears stalled | error |
4B2F | Table at angulation limit | Angulation limits reached or foot step extended; INTERLOCK signal true on Servo controller board | error |
4B30 | Collimator non-functional | One or both collimator blades not at requested position for extended time period | error |
4B31 | Rear door on Spot Film Device is not closed | Rear door switch signal indicates door is open when carriage motion is to occur | error |
4B32 | Object detected in front load port | Front sensors detect cassette or object in front cassette opening; internal software confirms unload operation but object still present | warning |
4B33 | Invalid cassette type or positioner calibration error | Rear load button pressed on 8835 positioner with 14x14 cassette loaded or carriage calibration mismatch | error |
4B34 | Rear door lock failed to lock or unlock after repeated retries | Solenoid POS SW signal did not give proper lock indication within timeout period | error |
4B35 | Table stepper engaged; manual motion not permitted | Angulation or lateral table motion requested while film changer not parked, or Film Changer mode selected with table horizontal | error |
4B36 | Obstruction detected at front | Front sensor beam (IN FRONT CASS signal) detected as broken when carriage not at front or cassette already in system | error |
4B37 | CPU requires reset and system reboot | Positioning module error condition | error |
4B3A | Table top in illegal region (extended while angulated or beyond calibrated boundary) | Table top feedback pot showing value outside calibrated positions for current table position | error |
4B3B | Table motion non-functional | Longitudinal or angulation pot reads zero or full A/D converter value | error |
4B3C | Spot Film Device non-functional | Handshake with servo controller board did not occur within ~100ms; servo controller did not toggle bit 0 of servo board status port | error |
4B3D | Rear load door forced open | Rear door opened when not expected; REAR DR SW signal indicates opening without prior load button request | warning |
4B3E | Spot Film Device non-functional | S188 unable to pass setup parameters to servo controller during initialization; read back data did not match passed data | error |
4B3F | Collimator non-functional | SID or collimator blade feedback pot A/D value registers at zero or full scale | error |
4B40 | Spotfilmer servo non-functional | Servo feedback A/D values register at zero or full scale | error |
4B41 | Cone not parked during table motion | Table top motion requested while cone not parked in U.K. cone operation mode | error |
4B48 | Step command requested when in external mode | ASC sends step command to table top while positioner is in film changer control mode | error |
4B4B | Injector did not inject as requested | Injector did not give an injecting indication within specified time period after INJ START DR signal asserted | error |
4B50 | Fixed mask servo failed to reach requested position within timeout | Servo in position bit from servo controller not asserted within allotted time | error |
4B51 | Variable mask servo position timeout error | Variable Mask servo did not reach requested position within allotted timeout period; servo in position bit not asserted by servo controller | error |
4B53 | Grid servo failed to reach requested position within timeout | Servo in position bit from servo controller not asserted within timeout | error |
4B54 | Lateral cassette servo failed to reach requested position within timeout | Servo in position bit from servo controller not asserted within allotted time | error |
4B55 | Longitudinal cassette servo failed to reach requested position within timeout | Servo in position bit from servo controller not asserted within timeout | error |
4B56 | SFD locked in forward position interlock error | SFD in forward position at Prep time when not appropriate for selected function | warning |
4B57 | Cassettes detected in table/wall buckets during auxiliary or warmup mode exposure attempt | RAD exposure attempted in AUX or Tube Warmup Mode with cassettes in table and/or wall bucky trays | error |
4C00-4CFF | Photo Spot module Applications software error range | Error in Photo Spot module during Applications mode | unknown |
4C02 | Film Advance in process | Photospot module during film advance operation | info |
4C03 | Film Advance in process | Remote film advance on console attempted while camera already doing transport due to manual pushbutton film advance | error |
4C04 | Camera shutter did not open | Photospot prep initiated; shutter commanded open by ASC but remains closed after timeout period | error |
4C05 | Math error in Photospot software, division by zero attempted | Software encountered division by zero condition in Photospot module | error |
4C06 | Photospot camera inoperable | Power amp board indicates a fault condition during pre-transport check with PAFAULT signal active | error |
4C07 | Magazine lever not closed; knife switch closed when it should be open | Camera pre-check during transport or Prep operation detects knife switch signal indicating closed position | error |
4C08 | Photospot camera film supply depleted or camera not ready | Film remaining (FILM REM) A/D indicates low film AND camera signals not-ready during Prep phase | warning |
4C09 | Photospot camera not loaded properly | Camera indicates not ready during pre-transport ready check; CAM RDY* signal indicates not ready status | error |
4C0A | Instruction not understood. Please try again. | GE bus transmission received and software cannot find command in its command table | error |
4C0B | Takeup magazine is full and must be emptied to continue | Camera indicates takeup magazine full; CAMERA RDY signal not ready; TAKE UP MAG OK and SUPPLY MAG OK in place; CARD OK present; FILM REMAINING STATUS > 2.5V; TAKE UP STATUS > 5V | warning |
4C0C | Takeup magazine full and supply magazine empty condition detected | CAMERA RDY* signal shows not ready; takeup magazine A/D (TAKE UP STATUS) > 5 volts; supply magazine A/D (FILM REMAINING STATUS) < 2 volts; both magazines in place | warning |
4C0D | No patient ID card inserted in camera | Camera indicates missing patient ID card detected via CARD OK signal at Prep time | error |
4C0E | Supply magazine not in place | Camera detects SUPPLY MAG OK signal absent at Prep time | warning |
4C0F | Takeup magazine not in place | TAKEUP MAG OK signal on camera status port indicates magazine absent at prep time | error |
4C10 | Camera not ready: missing takeup magazine, full magazine, or missing patient ID card | Camera not ready at Prep when film supply A/D > 2 volts | error |
4C11 | Supply magazine needs to be reloaded or magazine not in place | CAMERA RDY* signal not ready AND FILM REMAINING STATUS A/D <2 volts (<30 frames) | error |
4C13 | Supply magazine empty | CAMERA RDY* signal indicates not ready AND FILM REMAINING STATUS A/D reads less than 2.5 volts AND TAKEUP STATUS A/D is less than 5 volts AND patient ID card is in place | error |
4C14 | Camera did not respond before timeout expired | Camera did not generate expose to generator signal before timeout, EXP TO GEN signal not true on camera status port despite CAMERA RDY* signal indicating ready | error |
4C15 | Camera did not respond to expose command before timeout expired | Camera did not release expose to generator signal or EXP TO GEN signal did not go low or stay low after close shutter command | error |
4C16 | Film transport fault | Camera did not complete transporting film within one second from start of transport | error |
4C30 | System communication problem. Photospot module failed to receive transmit acknowledge within timeout. | Module attempted to transmit packet to node; no acknowledge before 80188 timer 2 timeout. | error |
4C31 | Improper updating of the packet indexer; indexer went past end of packet after processing command because packet had less data than expected | Packet indexer processing error in Photospot module | error |
4C32 | Software attempted to transmit more data than would fit in a single packet | Data transmission exceeds packet size capacity | error |
4C33 | GE Bus transmit queue full, module software lost or ARCNET not transmitting | Software's internal GE Bus transmit queue was full when module attempted to send packet | error |
4C34 | ARCNET hardware could not find designated node on system to transmit packet | Transmitter available before transmission timeout but no transmit acknowledge given | error |
4C35 | Internal timer indicator allocated to different scheduled task during software task scheduling | Software attempted to schedule a task and internal timer was already in use by another task | error |
4C40 | System timer task queue full; instruction not understood | Software attempted to schedule a task on system timer with no available queue space | error |
4C41 | Instruction not understood. Please try again. | Attempt to update or flush a scheduled task with timer ID outside list of scheduled tasks | error |
4D00 | System communications problem | ARCNET transmission failed; destination module did not acknowledge reception within 250 milliseconds | error |
4D00-4DFF | Pulscan module Applications software error range | Error in Pulscan module during Applications mode | unknown |
4D01 | GE Bus command interpreter packet index overrun | Number of bytes transmitted in a packet does not match expected number | error |
4D02 | System communications problem | Packet transmission attempted exceeds maximum transmittable size | error |
4D03 | GE Bus output queue is full | System attempting to send data on GE Bus when queue at capacity | error |
4D04 | System communications problem | ARCNET transmission attempted to non-existent node (TA set, TMA not set) | error |
4D05 | System communications problem | Command received across GE Bus (ARCNET) that is not a defined command for Pulscan module | error |
4D06 | Bad data received from console | Attempt to schedule a task when no timers available | error |
4D07 | Bad data received from Pulscan module | Attempt to flush or set a soft timer greater than configured system timers | error |
4D08 | Bad data received from console | Attempt to schedule a time-delayed task through SCHEDULE_SOFT_TIMER_TASK using a timer already allocated to another task | error |
4D0F | System communications problem | Communications failure in Pulscan module | error |
4E00 | Bolus filter error | Bolus filter malfunction or misalignment | error |
4E00-4EFF | AFM/CFM positioners Applications software error range | Error in AFM/CFM positioners during Applications mode | unknown |
4E01 | General PVC console error | PVC circuit malfunction detected by Positioner module | error |
4E02 | Gantry problem | General error with the gantry system detected | error |
4E03 | M200 Table is disengaged. Remote control not allowed. | Table top is not engaged | warning |
4E04 | General M200 Table error | Positioner table malfunction detected | error |
4E05 | II collision occurred | Positioner mechanical collision detected | error |
4E06 | Collimator collision occurred | Collimator arm strikes patient or table during positioning | error |
4E07 | General error with the table | Table malfunction detected | error |
4E08 | Emergency stop occurred | Emergency stop detection in Positioner module | warning |
4E09 | Gantry not in anteroposterior (AP) position | Gantry position sensor detected gantry not positioned for AP imaging | error |
4E0A | Collimator error | Collimator mask not installed or collimator system malfunction | error |
4E0B | Collision CHNGR | Positioner collision detected during operation | error |
4E0C | Collimator mask installed when FLUORO mode requires it to be removed | User attempts FLUORO operation with collimator mask installed | error |
4E0D | Film changer direction to X-ray | Onboard film changer moves from parking to working position | info |
4E0E | Film changer direction to head position | Onboard film changer moves from Working to Parking head position | unknown |
4E0F | Film changer moves from Working to Parking Foot position | Onboard film changer direction command to Foot position | info |
4E10 | Request cannot be executed with film changer in this position | Film changer positioned incorrectly for requested operation | error |
4E11 | Gantry Angulation not allowed for Film Changer motion. | Attempt to perform film changer interchange at current gantry angulation | warning |
4E12 | Interchange not allowed in AP angulation if table is not on top | Attempted AP angulation with table not in top position | warning |
4E13 | Gantry is moving | Gantry movement detected | info |
4E14 | Film Changer in middle of travel; no gantry motion | Film changer positioned at middle of travel during operation | error |
4E15 | Film Changer Collision Risk | During film changer interchange operation | warning |
4E16 | Film Changer Interchange Complete | Film changer completes interchange cycle | info |
4E17 | Lift collision detected | Collision between lift and mechanical structure | error |
4E18 | Film changer not correctly locked | Film changer position incorrect or locking mechanism failed | error |
4E19 | Film changer not in work position | Film changer is not positioned correctly for operation | error |
4E1A | Gantry not centered | Gantry position monitoring system detects misalignment | warning |
4E1B | Table positioning error in AP mode | System detects table is not fully backward before AP interchange | error |
4E1C | Film changer is not in park position | Type II bad interlock detected; film changer not at park position | error |
4E1D | Table is moving to the Start Position | During DA / DSA stepping acquisition when table transitions to start position | info |
6000 | Floppy Disk error | Disk hardware failure | error |
6002 | Floppy disk error - volume granularity | Floppy disk read/write error; volume allocation unit error | error |
6003 | Floppy Disk error | Device granularity error | error |
6004 | Floppy Disk error | Floppy disk read command failure | error |
6005 | Floppy Disk error - software module error | ASC (Floppy Disk Code Module) fails to process file - no TEND on filename | error |
6007 | Floppy disk error: filename too long | Software attempted to access file with name exceeding length limit | error |
6008 | File not found | Floppy disk code module cannot locate required file | error |
6009 | Floppy disk error - pointer pair overflow | Floppy disk operation or self-test failure | critical |
600A | Floppy disk error; VOL block not found | Floppy disk communication failure | error |
600C | Floppy Disk error | Entry point not found in floppy disk | error |
600D | Runtime database error in NV RAM fil correction values | Warm start when NV_G_fil_correction_X does not equal NOT (NV_G_fil_correction_not_X) | warning |
600E | Recommended X-ray tube Preventive Maintenance | Filament correction values increased more than 3B$0000H from calibration baseline at warm start | warning |
6010 | Fastboot functionality not operational | Fastboot ROM or boot process failure | warning |
6011 | Floppy disk error - incorrect data count | Floppy disk read/write data mismatch | error |
6020 | SPROM access error in ASC module | INSITE Electronic Device ID software failure to read/access ASC i2c devices or bad device programming | info |
6021 | Board ID: Elapsed Hours On Checksum Error | INSITE Electronic Device ID software failure to read/access ASC i2c devices or Advantx-E software finds bad programming | warning |
6100-67FF | ASC module Applications software error range | Error in ASC module during Applications mode | unknown |
6110 | Room deselect attempted during fluoro exposures | A room deselect is attempted during fluoro exposures | warning |
6111 | Room control module inoperable; fluoroscopy cannot proceed | Attempt to perform fluoroscopy exposure when Positioner module is offline, misconfigured, or not responding | error |
6112 | Exposure control module (ECM) not responding or not configured. | Operator attempts fluoroscopy exposure when ECM is offline or unresponsive. | error |
6113 | Video control module inoperable; fluoroscopy cannot be performed | VIC not configured online or no reply received from VIC to RT Bus driver command during fluoro exposure attempt | error |
6116 | Video Tape Recorder communication failure | No reply received from VIC to VTR status command | error |
6117 | Optical distributor did not respond | Aperture in place reply not received from VIC module when fluoro exposures attempted | error |
6118 | Optical filter did not respond | Image gate positioned reply not received from VIC module during fluoro exposures | error |
6119 | VTR being used for playback, unable to record | VTR is not in stopped state during fluoro or record exposure | warning |
6120 | Performing camera auto zero. Please wait. | Operator presses horizontal reverse button at console | info |
6121 | Pulscan inoperative. Reverted to next available FL mode. Call service. | Pulscan not configured, initialization incomplete, or no fluoro exposures due to Pulscan offline or bad reply to status/RT Bus command | error |
6122 | Selection cannot be made during acquisition | Fluoro operation mode change is attempted while taking exposures | warning |
6123 | Digital Record inoperative in fluoroscopy mode | Digital Record not marked online, not configured, or bad status reply received | error |
6124 | Digital Record module busy; no last image hold will occur | Reply received from Digital Record when attempting Digital Record fluoro exposures indicates Digital Record is currently busy | warning |
6125 | Cannot do Progressive Fluoro, DX not available | Attempting DX fluoro exposures when no fluoro configure reply received from DX | error |
6126 | Cannot do Progressive Fluoro; DX offline or not ready for acquisitions | Attempting DX fluoro exposures when DX is offline or unavailable | warning |
6127 | Horizontal sweep reversal already in progress | Operator presses horizontal reverse button while previous reversal still in progress | warning |
6128 | Lateral FLUORO VIEW not allowed with DIGITAL FLUORO selected | Fluoro operation mode changed to DX fluoro with lateral fluoro view, or fluoro view changed when DX fluoro is active | warning |
61F9 | VTR did not start recording | No reply received to VTR record command sent for HPR-50 | warning |
61FA | No run data saved, acquisition less than 3 seconds | Less than 3 seconds of tape has elapsed since the beginning of recording | info |
61FB | Mark frame outside current run range | Operator attempts to mark frame on tape section outside current run selection | error |
61FC | No run data to search on | Attempt to mark a frame, select a run number, or select a mark when VTR current run directory is empty | info |
61FD | No mark saved for this run | Operator attempts to select a mark for a VTR run with no stored mark position | info |
61FE | Informational message displayed when VTR Reset button pressed | Operator presses VTR Reset button on console | info |
6202 | Injector not ready | Prep switch depressed when injector is not in ready state | warning |
6203 | Table not ready | Prep switch depressed when positioner/table not in ready state | warning |
6212 | Stepper selection mismatch between console and table | Prep switch depressed with stepper selected at table but not at console | warning |
62CA | mA overload occurred | mA overload condition during X-ray exposure | warning |
6304 | Optical distributor did not respond during Cine exposure attempt | Image Gate positioned reply not received from VIC module when Cine exposures attempted | error |
6306 | Optical filter did not respond | Neutral Density Filter positioned reply not received from VIC module during Cine exposures | error |
630A | Cine camera not line locked | Error status reply from Cine module in response to RT Bus receivers configure command | warning |
630E | Auto stabilize sequence timeout | 20 second timeout on start of auto stabilize sequence reached | error |
630F | Cine module offline | Cine mode select performed or record exposures attempted in Cine, DF Dynamic, or DX Cardiac | error |
6310 | Cine mode unavailable due to video control module inoperable | VIC module found offline when cine mode selected or record exposures attempted in Cine, DF Dynamic, or DX Cardiac | error |
6311 | Positioner module offline when cine mode selected or cine exposures attempted | Cine mode select performed or cine exposures attempted with positioner module offline | error |
6312 | Cine mode unavailable due to ECM module offline | Cine mode select performed or cine exposures attempted while ECM offline | error |
6314 | Injector not ready | Prep switch depressed when injector status indicates not ready | warning |
6315 | VIC Pedestal Servo did not respond | No response received from imaging module to black level servo command when cine record acquisition attempted | error |
6316 | Pulscan module offline during cine exposure | Pulscan module found offline when cine exposures using Pulscan fill-in attempted | warning |
6318 | Opposite plane acquisition sequence timeout | Timeout on opposite plane response reached when cine exposures attempted | error |
632A | Maximum single plane run time for a biplane acquisition exceeded | Biplane acquisition run duration exceeds system limits | warning |
63FF | Cine camera has less than 20 feet of film remaining | Film update packet received from Cine module at end of exposures when remaining film drops below 20 feet | warning |
6401 | Bad data received from Positioner during prep done timeout | 10 second timeout on receiving prep done when attempting record exposures (except Cine) because Configure for Acquisition command reply from Positioner not received or was bad | error |
6402 | Bad data received from VIC module during exposure preparation | 10 second timeout reached while waiting for prep done signal when attempting record exposures (DF Static, DF Dynamic, DX Vascular or DX Cardiac) | error |
6403 | Bad data received from Positioner; timeout on interlock status | 10 minute timeout reached when attempting record exposures while waiting for Positioner Interlocks command reply | error |
6404 | Timeout on receiving exit done at end of record exposures | Reply to expose sequence disable command from ECM not received within 1.5 seconds | error |
6405 | Exposure control inoperable | ASC waits 20 seconds for ECM compliance to select tube and room; no response received | critical |
6406 | Bad data received from module node | Module node found offline during communications (record acquisition, dx offsets, film speed change, temporal value change) | error |
6407 | Wait for Trial Done Timeout | Timeout during trial execution | error |
6408 | Mode mismatch between console and table positioner | Attempting to record exposures when selected mode at console does not match table type from positioner auxiliary mode packet | error |
6409 | Expose time packet not received from Positioner during Tomo record setup | Attempting to initiate Tomo record exposures without exposure length parameters | warning |
640A | Bad data received from variable exceeding mathematical function boundaries | Variable exceeds boundaries in application generic software set during console operation | error |
640B | Bad data received from application software | Dividend of zero passed to divide routine in generic software set | error |
640C | Bad data received from action-state driver routine | Invalid value passed into action-state driver in generic software set | error |
640D | Cannot use Stepper while CC selected | Operator attempts to use stepper function while CC mode is selected | error |
640E | No auxiliary mode packet received from Positioner | Attempting to record exposures for Rad and Tomo modes without mode packet from Positioner | error |
640F | Proper auxiliary mode packet not received from Positioner for Tomo acquisitions | Attempting Tomo record exposures without proper table preparation | error |
6411 | Invalid action-state driver routine value for current application state | A value was passed into the action-state driver routine which was invalid for the current state in the application generic software set | error |
6412 | Cannot switch filaments during acquisition | Attempt to change focal spot during fluoro or record exposures | warning |
6413 | Cannot switch patient size during acquisition | Attempt to change patient size during fluoro or record exposures | warning |
6414 | Cannot switch AEC mode during acquisition | Attempted change of record mode between AEC and fixed during fluoro or record exposures | warning |
6415 | Bad data received from temperature to energy conversion routine | Bulk temperature greater than 2000 degree maximum passed into conversion routine used for sequence prediction calculations | error |
6416 | Bad data received during sequence prediction calculation | Bulk energy value exceeding maximum passed to energy-to-temperature conversion routine | error |
6417 | Bad data received from Digital Record module during exposure attempt | 10 second timeout reached when waiting for Digital Record prep done confirmation during Digital Record exposure | error |
6418 | Bad data received during Photospot prep sequence | 10-second timeout when Photospot module fails to send prep done signal | error |
6501 | Film remaining value from Photospot module outside valid range for configured camera type | Invalid photospot module data received by console | error |
6601 | Invalid addresses calculated by AUP software for AMB utilities access | AUP database corruption, AUP memory board failure, communication loss, or AMB memory fault | error |
6601 | Communication failure with auto protocol memory board | AUP software attempted to access AUP memory board | error |
6602 | Memory or procedure backup initialization/restore error | Occurs at warm start or during SAVE/RETRIEVE operation | error |
6602 | Procedure database version is not compatible | Warm start with mismatched AUP database version or restore from incompatible backup floppy | error |
6603 | No procedure present, cannot perform requested operation. | No procedure has been selected. | error |
6604 | Procedure menu is full with no space to add new data | Attempt to add procedure when memory capacity reached | error |
6605 | Procedure data base version is not compatible | AUP data base version mismatch at warm start during PED mode | critical |
6606 | Cast editing requested while another room is editing | Software attempts to edit cast/auto-protocols when another room is performing cast editing | info |
6607 | Procedure edit must be run; procedures cannot be accessed until NV RAM compatibility check completes | Operator presses PROCEDURE MENU during auto-protocol (AUP) mode when ped_required flag is set | error |
6608 | No AEC data, cannot select it as default | Attempt to select AEC as default when no AEC data has been programmed | warning |
6609 | No fixed data has been programmed. | Attempt to select a default protocol when no fixed data exists | error |
660A | Category full, cannot perform requested operation | Attempt made to add procedure to full category in Procedure Edit module | error |
660B | Maximum receptors reached, cannot perform requested operation | Attempt to add another receptor when procedure already has maximum number assigned | warning |
660C | Other room currently using SAVE/RESTORE | Another room is in the save/restore screen | info |
660D | System floppy in drive, cannot perform requested operation | Attempted save/restore operation with Advantx software floppy in drive | error |
6610 | Frontal/Lateral database mismatch | Protocol database inconsistency between frontal and lateral imaging mode protocols | error |
6611 | No unique receptors, cannot perform requested operation | Attempt to copy receptors to a procedure that already contains all requested receptors | warning |
6612 | Autoprotocols unavailable. Continue and call service. | At power-up, AUP software fails to write and read back test patterns to reserved test location in AUP database, or data read back does not match data written. | error |
6613 | Autoprotocols unavailable; AUP software task creation failed on frontal biplane system at power up | Power-up initialization; error encountered during RMX task creation by AUP software on frontal half of biplane system for lateral database read/write requests | error |
6614 | A protection key floppy is required to edit procedures | Attempt to edit procedures without protection key floppy installed | info |
6680 | Media error, insert new diskette | Failure to properly read floppy disk | error |
6681 | Disk error - floppy drive failure | Failure occurs with floppy drive during operation | error |
6682 | Floppy disk is write protected | Attempt to write to a write-protected floppy disk | warning |
66C0 | Communication failure while communicating with the other plane in a biplane system | ASC (Procedure Edit Auto-protocols) module loses communication with opposite plane | error |
6700 | Cannot switch receptors during an exposure | Attempt to switch receptors at console or spot film device during fluoroscopy or record exposures | warning |
6701 | Tube conditioning in progress | Attempt to switch receptors at console or spot film device during tube conditioning | warning |
6702 | Instruction not understood | Software detected execution pathway designed to be unreachable | error |
6703 | Application not loaded | Application file configured for system did not upload properly from floppy disk | error |
6704 | Receptor not available due to floppy disk read error | Application file not properly uploaded from floppy disk during operator selection | error |
6705 | Lateral application not found. Reconfiguring as single plane. | Application file configured as biplane did not upload properly to lateral ASC from floppy disk | warning |
6706 | Application not loaded | Tube warmup application file did not upload properly from floppy disk | error |
6707 | FastBoot auto return failed; default application mode and parameters restored | NVRAM failure, system halt during runtime snapshots, or biplane plane mismatch | warning |
6708 | Biplane communication error: frontal failed to send packet containing biplane-common values to lateral | In biplane systems, communication error occurs during packet transmission of common values between frontal and lateral units | error |
6800 | Communication problem between console and DF (digitalfluoroscopy) module | Packet from DF not received within allotted time | error |
6801 | Communication problem between console and DF module | Packet from DF not received within allotted time | error |
6802 | DF module status indicates a fault condition | Last DF module status received indicates a fault | error |
6803 | Communication problem between DF module and Advantx software | DF module software and Advantx software not in same acquisition mode (e.g. DF in Static, Advantx in Dynamic) | error |
6804 | DF module not ready for acquisitions | DF module initialization fails or acquisition prerequisites not met | warning |
6805 | Insufficient disk space for image acquisition | Static sequence requires more image disk space than available | error |
6806 | Disk space available is less than 2 seconds of run time | Dynamic acquisition attempted with insufficient disk space | warning |
6807 | Image disk directory is full (contains 100 images) | Attempting to acquire or store images when disk reaches capacity | error |
6808 | DF service mode exposure timing error | Attempted zero millisecond exposure in DF service mode | error |
6902 | Collimator blades open during tube warmup | Tube warmup sequence entered with manually controlled collimator blades open | warning |
6903 | Non-volatile database checksum mismatch | Checksum validation fails at system startup | critical |
6904 | Non-volatile database checksum is wrong | System startup when NV database checksum does not match stored value | error |
6905 | Warmup inactive while two soak cycles in progress | Tube change attempted while on tube warmup screen with two soak timers active | warning |
6906 | Cannot perform tube conditioning | During tube warmup startup, selected focal spot value set using initialized MA value found in error | error |
6A00 | Digital Record New Study annotation not entered | Digital Record exposures attempted without New Study annotation | warning |
6A01 | Disk space remaining is less than 15 frames | Digital Record exposures attempted when available disk space drops below 15 frames | info |
6A02 | Digital Recorder module busy; no last image hold will occur | Digital Record exposures attempted while Digital Recorder in busy state | warning |
6A03 | Disk full. Remove images to continue. | No more frames of storage available on Digital Disk during Digital Record exposures. | warning |
6A04 | Maximum number of images have been stored in this study | Acquisition attempted to append images to a study that contains 99 images (maximum allowable) | info |
6A05 | Digital Recorder temperature too high | Temperature sensor on MHR image processor detects overheating condition | error |
6A06 | Images cannot be appended to this study at this time | Acquisition attempted on a study not selected for APPEND mode | error |
6A07 | No current study exists on DRS | Acquisition attempted when no study selected or new study information not entered | error |
6A08 | Communication problem encountered | Acquisition attempted when XDIU was not functioning | error |
6B00 | Number of images exceeds disk space | Vascular sequence requires more disk space than currently available | warning |
6B01 | Maximum number of patients on digital system has been reached | DX image disk cannot store additional images | warning |
6B02 | The maximum number of sequences on digital system has been reached. | DX image disk cannot store additional images. | warning |
6B03 | Digital system offsets failed | DX video adjustment process did not complete successfully | error |
6B04 | Digital system offset increment is enabled | DX service mode left noise measurement functionality enabled | warning |
6B05 | Digital system offset adjustment in process (console is locked) | DX video adjustment process initiated | info |
6B06 | Digital system offset adjustment has completed successfully | DX video adjustment process completion | info |
6B07 | System communications problem | Packet from DX module not received within allotted time | error |
6B08 | Communication problem between console and DX code module | Packet from DX not received within allotted time | error |
6B09 | Digital system is not in READY state | DX module fault indicated by last DX status received | error |
6B0A | Offsets failed | Digital offset values could not be applied or re-calibrated | error |
6B11 | DX module not ready for acquisitions | DX system fails to enter ready state for image acquisition | error |
6B12 | Disk space insufficient for cardiac exposure sequence | Less than 2 seconds of image storage available | warning |
6B13 | Communication problem between console and DX Code Module | DX packet not received within allotted time | error |
6B14 | Digital system offset failed to complete within the allowed time | DX video adjustment process does not finish within allotted time window | error |
6B15 | AEC exposure time is less than the suggested minimum | Trial or mask AEC exposure resulted in exposure time less than five milliseconds | warning |
6B16 | Film record not allowed at this frame rate. Record Medium set to Disk. | Attempt to record film at two lowest frame rates in Cardiac mode | info |
6B17 | Selection of DENS QUANT with Record Medium Film is not allowed | Operator selects Dens Quant with film-only record medium in Cardiac when DX software version is prior to A4 | warning |
6B18 | Trial expose time too long | Trial exposure results in expose time greater than half of trial backup time | warning |
6B19 | Frame rate too slow to use cine camera | Cine camera operation with insufficient frame rate | warning |
6B1A | Biplane acquisitions not possible in TECH MENU | Attempted biplane acquisition while in TECH MENU mode | info |
6B1B | Frontal AEC expose time is less than the suggested minimum | Frontal trial or mask AEC exposure resulted in expose time less than five milliseconds during biplane acquisition | warning |
6B1C | Lateral AEC expose time is less than the suggested minimum | Lateral trial or mask AEC exposure resulted in expose time less than five milliseconds during biplane acquisition | warning |
6B1D | Both AEC expose times are less than the suggested minimum | Both frontal and lateral trial or mask AEC exposures result in expose times less than five milliseconds during biplane acquisition | warning |
8100 | An MTE error has occurred | MTE (Module Test Error) condition detected | error |
8100 | Hardware Status 0900—1000 | MTE error occurrence | error |
8100 | MTE (Module Test Error) has occurred with bad data received | Hardware Status 0000–0800 condition | error |
8100 | Hardware Status error in range 0000-0800, MTE (Multi-Tiling Engine or similar) error with bad data received | Corrupted or invalid data received by system | error |
8101 | RMX system error occurred | RMX error detected in system | error |
8101-81FF | SLOS (System Level Operating System) error range | Error in Diagnostics mode by SLOS | unknown |
8102 | Non-Volatile Database Corrupted | Invalid database checksum detected | error |
8103 | System programming error: invalid parameter passed to Clear Call Table utility | Procedure using Clear Call Table utility called with out-of-range parameters | error |
8104 | Database Error: Incompatible Database Version | Non-Volatile Database major revision does not match software floppy release number | error |
8105 | Database Error: Default Override Switch Active | Non-Volatile Database Checksum Override Switch is set on ASC BIU | info |
8180 | Transmitter Not Ready. Test Aborted | GE bus transmit procedure cannot obtain ARCNET transmit token; SLD command not sent | error |
8181 | Node response timeout during GE bus transmission | ASC sends SLD command to node; bus transmit times out before node responds | error |
8182 | Node Not Responding. Test Aborted | GE bus transmit procedure cannot communicate to the receiving node; SLD command not sent | error |
8183 | Node error during communication with receiving node | Test Sequencer receives node error while communicating with receiving node | error |
8184 | Invalid packet received during inter-node communication; test aborted | Test Sequencer receives malformed or corrupt data packet from receiving node during test execution | error |
8185 | Node is off-line. Test aborted. Node communication problem. | Test Sequencer received a node off-line error while attempting to communicate with a node | error |
81A0 | Floppy read error, boot information not found | Error occurred while attempting to read Third Party data file from floppy | error |
81A1 | Floppy not initialized properly | Third Party data file was not initialized properly | error |
81A2 | Floppy incompatible with system | Third Party data file key checksum does not match non-volatile checksum | warning |
81A3 | Invalid system identification | Third Party data file ID checksum does not match non-volatile checksum | error |
A000-A0FF | SMOS (Service Mode Operating System) error range | Error in Service Mode by SMOS | unknown |
E-6001 | Floppy Disk error | Unknown disk format in floppy drive | error |
Error 4026 | System error from Applications Operating System (APOS) code | GE bus transmitter is not available | critical |
Error 4051 | System communications problem | ASC module communication failure detected | critical |
Error 4066 | System was reset from the console | info | |
Error 4071 | System communications problem in ASC module running APOS code | Bad node ID received over communications link | error |
Error 4072 | System communications problem | APOS code in ASC module detects node error | error |
Error 4202 | Power unit interface error | Power unit error condition detected during system operation or recovery sequence | error |
Error 4226 | Database error: ECM received calibration or configuration data out of range. | ECM receives cal/config data with one or more parameters exceeding valid range during initialization or operation. | error |
Error 4246 | ECM and power unit communications error or module initialization failure | Power-up, re-initialization failure, or DIP switch misconfiguration causing diagnostic interference | error |
Error 4247 | ECM receiving unused error value from power unit or misinterpreting normal error value | Power unit sending unexpected error value or ECM firmware interpreting normal value incorrectly | warning |
Error 4289 | STARTWHEN timeout on RT bus 10 | Particular signal or state not achieved within recipe-specified time; recipe index displayed in HW STATUS field | error |
Error 4293 | Start when timeout, timer 13 | A particular signal or state has not been achieved after a given amount of time as specified in the currently active recipe | error |
Error 4403 | Digital system error, acquisition terminated | Digital system fault detection | error |
Error 4509 | Serial link framing error detected on DR BIU processor | Serial link Duart (U133) detects framing error in character reception from DR Module | error |
Error 4784 | Optical distributor not positioned | ND Filter amp indicated stall at bad time | error |
Error 4A56 | Film changer not parked at system prep time; interlock prevents continuation | FC PARKED signal indicates film changer not in park position during prep | error |
Error 4B23 | Table locks non-functional | Lock feedback indicates incorrect state after timeout | error |
Error 6006 | Floppy disk error in ASC module | Filename not specified during floppy disk operation | error |
Error 6108 | Neutral density filter positioning reply not received from VIC module | Fluoro exposure attempted while VIC module offline or filter reply not received | warning |
Error 6600 | Unrecognized command | AUP software receives command packet from other applications that does not match recognized command format | info |
Error 660E | Floppy drive door left open during save operation | File save attempt with drive door open | warning |
Error 660F | Floppy drive door was left open | Attempted to retrieve data with floppy drive door open | error |
Indeterminate (watchdog error) | Watchdog timer timeout indicating serious software or electronics error | Watchdog circuit detects failure during normal operation, at power-up, or during shock | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI