GE Healthcare 9900: troubleshooting and error codes

The ServiceForge knowledge base holds 1152 documented symptoms for the GE Healthcare 9900, with 1800 ranked causes, 1744 diagnostic checks with test points and expected values, and 473 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Orbital motion fails or runs intermittently patient safety

Motorized orbital motion does not function or operates sporadically

Ranked causes

  1. 1. RUI not properly connected to C-Arm keyswitch panel (common)
  2. 2. Orbital clutch released or damaged (common)
  3. 3. Orbital position potentiometer damaged or drive belt broken (occasional)
  4. 4. System out of calibration (occasional)
  5. 5. Orbital gearbox faulty (occasional)
  6. 6. Orbital motor faulty (occasional)
  7. 7. Orbital servo drive faulty (occasional)
  8. 8. MCU faulty (rare)
  9. 9. Broken or damaged orbital drive belt (occasional)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm completely inoperative; battery charger light is on

Workstation boots but C-arm circuits have no power and display no error messages; battery charger light illuminates normally

Ranked causes

  1. 1. Workstation power problem (common)
  2. 2. Circuit breaker CB1, CB2, or CB3 tripped in Workstation (common)
  3. 3. 100-125V system plugged into 200-250V outlet (occasional)
  4. 4. Power/Signal Interface PCB connector loose or disconnected (occasional)
  5. 5. Power Control PCB jumper E1 set to wrong position (occasional)
  6. 6. DC power distribution problem (occasional)
  7. 7. Defective Relay K10 in Power/Motor Relay PCB (rare)
  8. 8. Loose connectors on Power/Motor Relay PCB (occasional)
First check

Verify Workstation power cord is plugged in

Test point
AC receptacle
Expected
Power cord fully inserted

2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fast Stop activated or interlock circuit broken high voltage

Software displays 'FAST STOP ACTIVATED' error on Vacuum Fluorescent Display; X-ray emission prevented

Ranked causes

  1. 1. Fast Stop button pressed or switch opened (common)
  2. 2. Wire or connection break in interlock loop (occasional)
  3. 3. Relay K9 failed or not latching (occasional)
  4. 4. Relay K8 failed or not energizing (occasional)
  5. 5. CPU Interlock signal not generated or Relay K1 failed (occasional)
  6. 6. Filament Driver PCB transistor failed (occasional)
  7. 7. PS2 power supply not providing +24 VDC (rare)
  8. 8. Relay K10 failed or +12 VDC Key Power not supplied (occasional)
First check

Verify Fast Stop switches are not activated

Test point
Fast Stop button physical state
Expected
Button not pressed, switch contacts closed

4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Workstation boots but nothing on C-arm; battery charger light ON

Workstation completes boot sequence and displays normally, but C-arm circuits are inactive and unresponsive; battery charger indicator light is illuminated

Ranked causes

  1. 1. No Key Power from Workstation (common)
  2. 2. Loose Power/Signal Interface PCB connectors (occasional)
  3. 3. Blown fuse F6 or F10 on Power/Signal Interface PCB (common)
  4. 4. Defective Relay K10 in Power/Motor Relay PCB (rare)
  5. 5. Power Signal Interface AC power problem (occasional)
  6. 6. Fast stop buttons or wiring open (occasional)
  7. 7. Blown F10 fuse on Power/Signal Interface PCB (common)
  8. 8. Interconnect Cable AC power pins or continuity fault (occasional)
First check

Verify Key Power signal from Workstation via interconnect cable

Test point
Interconnect cable at C-Arm
Expected
Key Power signal present when Workstation on

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motorized rotation fails or runs intermittently patient safety

Motorized rotation function does not operate or operates sporadically

Ranked causes

  1. 1. RUI not properly connected to C-Arm keyswitch panel (common)
  2. 2. Rotational position potentiometer damaged or drive belt broken (occasional)
  3. 3. System out of calibration (occasional)
  4. 4. Rotational gearbox faulty (occasional)
  5. 5. Rotational motor faulty (occasional)
  6. 6. MCU faulty (rare)
  7. 7. Rotational servo drive faulty (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No filament voltage observed high voltage

Filament voltage at FIL_VOLT on Generator Driver PCB is zero, preventing X-ray generation

Ranked causes

  1. 1. No FIL_B_PLUS voltage due to open resistors or cable problem (common)
  2. 2. Transistor open or capacitor shorted in supply (occasional)
  3. 3. No filament drive signals present (occasional)
  4. 4. Faulty or shorted filament voltage sense resistors (occasional)
  5. 5. Failure of filament voltage sensor at Generator Driver PCB (occasional)
  6. 6. Broken connection to X-Ray Controller PCB (occasional)
  7. 7. Failure at X-Ray Controller PCB (rare)
First check

Measure FIL_VOLT at FIL_VOLT test point on Generator Driver PCB

Test point
FIL_VOLT
Expected
Non-zero voltage present

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-Arm completely inoperative without error messages high voltage

C-Arm system shows no power, no functionality, and no error messages

Ranked causes

  1. 1. Incorrect isolation transformer tap settings (common)
  2. 2. No Key Power (occasional)
  3. 3. Loose connectors (occasional)
  4. 4. Blown fuse F6 or F10 (occasional)
  5. 5. DC power distribution problem (rare)
  6. 6. Defective Relay K10 on Power/Motor Relay PCB (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High kV warning displayed high voltage

WARNING, HIGH KV - PRESS ANY KEY message appears during kVp servo-loop test after 10 kVp exposure increments from 40-120 kVp

Ranked causes

  1. 1. Low Batteries (common)
  2. 2. HV Supply Regulator PCB failure (occasional)
  3. 3. Generator Driver PCB failure (occasional)
  4. 4. IGBT failure or saturation (occasional)
  5. 5. HV Cables open or corroded (occasional)
  6. 6. X-ray Tube failure or arcing (occasional)
First check

Measure battery voltage and current capacity

Test point
Battery pack terminals
Expected
Full charge voltage and current rating as marked

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

HIGH KV warning displayed radiation

System displays warning that kilovoltage is too high and requests operator response

Ranked causes

  1. 1. Low batteries (common)
  2. 2. HV supply regulator PCB fault (common)
  3. 3. Generator driver failure (occasional)
  4. 4. IGBT failure (occasional)
  5. 5. HV cable damage or poor connection (occasional)
  6. 6. X-ray tube malfunction (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Infrared remote fails to change screen

Pressing the upper right-hand button on the IR remote transmitter does not change the screen

Ranked causes

  1. 1. Failure in IR transmitter (occasional)
  2. 2. Low or dead batteries in the IR transmitter (common)
  3. 3. Loose, faulty or disconnected cable between IR Receiver and EMI box (common)
  4. 4. Loose, faulty or disconnected cable between EMI box and System Interface PCB (common)
  5. 5. IR receiver faulty (occasional)
  6. 6. Faulty or dead power to IR receiver (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low kV warning displayed high voltage

WARNING, LOW KV - PRESS ANY KEY message appears during kVp servo-loop test after 10 kVp exposure increments from 40-120 kVp

Ranked causes

  1. 1. Low Batteries (common)
  2. 2. HV Supply Regulator PCB failure (occasional)
  3. 3. Generator Driver PCB failure (occasional)
  4. 4. IGBT failure or saturation fault (occasional)
  5. 5. HV Cables corroded or disconnected (occasional)
  6. 6. X-ray Tube anode failure (occasional)
First check

Measure battery pack voltage

Test point
Battery terminals
Expected
Full voltage rating per pack specification

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LOW KV warning displayed radiation

System displays warning that kilovoltage is too low and requests operator response

Ranked causes

  1. 1. Low batteries (common)
  2. 2. HV supply regulator PCB fault (common)
  3. 3. Generator driver failure (occasional)
  4. 4. IGBT failure (occasional)
  5. 5. HV cable damage or poor connection (occasional)
  6. 6. X-ray tube malfunction (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Orbital servo checksum error

Logosol orbital servo control module checksum validation failed during diagnostics

Ranked causes

  1. 1. Orbital servo controller defective (occasional)
  2. 2. Orbital motor defective (occasional)
  3. 3. Orbital motor wiring fault (occasional)
  4. 4. Motion power supply PS4 defective (occasional)
  5. 5. Power supply PS4 wiring fault (occasional)
  6. 6. Orbital position reporting potentiometer defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Orbital servo position error

Orbital servo motion failed to reach expected position during diagnostics

Ranked causes

  1. 1. Orbital servo controller defective (occasional)
  2. 2. Orbital motor defective (occasional)
  3. 3. Orbital motor wiring fault (occasional)
  4. 4. Motion power supply PS4 defective (occasional)
  5. 5. Power supply PS4 wiring fault (occasional)
  6. 6. Orbital position reporting potentiometer defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotational servo checksum error

Logosol rotation servo control module checksum validation failed during diagnostics

Ranked causes

  1. 1. Rotation servo controller defective (occasional)
  2. 2. Rotation motor defective (occasional)
  3. 3. Rotation motor wiring fault (occasional)
  4. 4. Motion power supply PS4 defective (occasional)
  5. 5. Power supply PS4 wiring fault (occasional)
  6. 6. Rotation position reporting potentiometer defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotational servo position error

Rotational servo motion failed to reach expected position during diagnostics

Ranked causes

  1. 1. Rotation servo controller defective (occasional)
  2. 2. Rotation motor defective (occasional)
  3. 3. Rotation motor wiring fault (occasional)
  4. 4. Motion power supply PS4 defective (occasional)
  5. 5. Power supply PS4 wiring fault (occasional)
  6. 6. Rotation position reporting potentiometer defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System fails to complete boot-up from hard-drive

Hard-drive boot test fails to load System Software during startup

Ranked causes

  1. 1. AC/DC power problem (common)
  2. 2. Loose cabling connectors or unseated PCBs (common)
  3. 3. Faulty System Interface PCB component (occasional)
  4. 4. Faulty IDE hard-drive (occasional)
  5. 5. Corrupt software (occasional)
  6. 6. Faulty component within generator (occasional)
First check

Verify AC/DC power supply operation

Test point
Power input to Workstation
Expected
Stable AC and DC voltages present

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vertical column does not move when pressing lift switch patient safety

Column remains stationary when operator presses up or down lift switch on control panel; may indicate power, relay, switch, or wiring issue

Ranked causes

  1. 1. Power panel keyswitch in Disable position (common)
  2. 2. Faulty lift switch (common)
  3. 3. Loose or reseated cabling connections (common)
  4. 4. +24 VDC supply failure or fuse open (occasional)
  5. 5. 115 VAC supply failure or fuse open (occasional)
  6. 6. Power Motor Relay PCB failure (rare)
First check

Verify power panel keyswitch position

Test point
Enable/Disable key-switch on C-Arm Power Panel
Expected
Enable position

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vertical column fails to move when pressing lift switch caution

Vertical column does not respond to lift switch input; remains stationary

Ranked causes

  1. 1. Power panel keyswitch in Disable position (common)
  2. 2. Loose Power Signal Interface or Motor relay connector (common)
  3. 3. Faulty lift switch (occasional)
  4. 4. +24 VDC supply issue (occasional)
  5. 5. 115 VAC supply issue or blown fuses (occasional)
  6. 6. Power Motor Relay PCB malfunction (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wrong mA during X-ray shot patient safety

Improper exposure (too high or too low); incorrect patient dose rate

Ranked causes

  1. 1. Incorrect filament selected (common)
  2. 2. Faulty mA sense circuit (common)
  3. 3. Tube aging (occasional)
  4. 4. Wrong kVp (occasional)
  5. 5. Filament regulator gross failure (rare)
  6. 6. kVp sensor failure (rare)
First check

Verify correct filament is selected in system configuration

Test point
Filament selection menu
Expected
Filament type matches tube specification

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

AC power cord or plug damaged or unsafe high voltage

Power cord shows damage, missing parts, broken wires, or frayed insulation; plug missing grounding pin or has loose/corroded conductors

Ranked causes

  1. 1. Power cord or plug has physical damage (occasional)
  2. 2. Wiring insulation broken, frayed or missing (occasional)
  3. 3. Grounding pin missing from AC line plug (occasional)
  4. 4. Plug conductors loose or corroded (occasional)
  5. 5. Power cord strain relief clamp missing or loose (occasional)
First check

Inspect power cord for physical damage

Test point
Full length of AC power cord
Expected
No visible damage, missing parts, cuts or abrasions

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charger light is off with workstation off high voltage

Battery charging indicator is not illuminated when workstation is powered off

Ranked causes

  1. 1. Workstation power or interconnect cable problem (common)
  2. 2. Wrong voltage system or wrong power cord (occasional)
  3. 3. Power/Signal Interface PCB connector problem (occasional)
  4. 4. Power Control PCB jumper E1 set to wrong position (occasional)
  5. 5. Blown fuse F4 on Power/Signal Interface PCB (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament voltage error high voltage

Wrong voltage is applied to the filament, causing incorrect mA generation

Ranked causes

  1. 1. Filament B plus wrong voltage (common)
  2. 2. Duty ratio of filament drive pulses wrong (common)
  3. 3. Wrong filament selected (occasional)
  4. 4. Failure of filament regulation (occasional)
  5. 5. Sensor gain wrong (occasional)
First check

Measure FIL_VOLT relative to FIL_VOLT_REF

Test point
FIL_VOLT, FIL_VOLT_REF
Expected
FIL_VOLT equals FIL_VOLT_REF

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No video on left monitor when X-ray switch pressed radiation

Left monitor displays no video when X-ray switch is activated, no last image hold function

Ranked causes

  1. 1. Missing CCD video or pilot tone signal (occasional)
  2. 2. Loose or damaged video cable (occasional)
  3. 3. Unseated or failed Video Controller or Vortex Image Processor PCB (occasional)
  4. 4. Loose or damaged video cable (duplicate) (occasional)
  5. 5. Unseated or failed Video Controller or Vortex Image Processor PCB (duplicate) (occasional)
First check

Verify X-ray switch activation

Test point
X-ray switch and monitor input
Expected
Video signal present on left monitor

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Room in use light test fails

External interface test indicates Room in Use indicator light not functioning or no signal

Ranked causes

  1. 1. Failed or burnt-out indicator light bulb (common)
  2. 2. No RM-IN-USE signal from System Interface PCB (occasional)
  3. 3. Loose or disconnected wiring to connector (common)
  4. 4. External interface PCB relay failure (occasional)
  5. 5. System interface PCB failure (occasional)
First check

Inspect Room in Use indicator bulb for failure or burn-out

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

X-ray emission blocked by interlock system high voltage

Unit prevents X-ray emission; Fast Stop button may have been pressed or interlock circuit broken; error message 'FAST STOP ACTIVATED' displayed on Vacuum Fluorescent Display

Ranked causes

  1. 1. Fast Stop button pressed (common)
  2. 2. Interlock circuit wire or connection break (occasional)
  3. 3. Software fault in Generator Controller PCB (occasional)
  4. 4. Relay K9 failure on Power/Motor Relay PCB (rare)
  5. 5. Relay K1 or K8 failure on Generator Interface or Power/Motor Relay PCB (rare)
First check

Verify Fast Stop switch position

Test point
Fast Stop switch physical position
Expected
Closed (not opened)

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray ON light test fails

External interface test indicates X-ray ON indicator light not functioning or no signal

Ranked causes

  1. 1. Failed or burnt-out indicator light bulb (common)
  2. 2. No XRAY-ON signal from System Interface PCB (occasional)
  3. 3. Loose or disconnected wiring to connector (common)
  4. 4. External interface PCB relay failure (occasional)
  5. 5. System interface PCB failure (occasional)
First check

Inspect X-ray ON indicator bulb for failure or burn-out

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

AC power cord and plug damage high voltage

AC power cord or plug shows damage, broken/frayed wires, missing grounding pin, or insecure conductors; loose prongs detected.

Ranked causes

  1. 1. Power cord or plug damaged or worn (unknown)
  2. 2. Missing or damaged grounding pin (unknown)
  3. 3. Loose prongs on power plug (unknown)
  4. 4. Missing strain relief clamp on power cord (unknown)
First check

Inspect AC power cord for damage

Test point
Full length of AC power cord
Expected
No damage, wiring insulation intact, no broken or frayed wires

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power cord or plug damaged high voltage

Power cord or AC plug shows damage, missing parts, broken wires, frayed insulation, or missing grounding pin

Ranked causes

  1. 1. Power cord or AC plug insulation damage or broken wires (unknown)
  2. 2. Missing grounding pin on AC plug (unknown)
  3. 3. Loose or insecure plug prongs or conductors (unknown)
  4. 4. Power cord strain relief clamp missing (unknown)
First check

Inspect power cord for damage and wiring insulation

Test point
Entire power cord length
Expected
No damage, missing parts, broken or frayed wires, intact insulation

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm LS1 on ISD Power Control PCB caution

Alarm LS1 sounds, indicating transformer AC voltage is out of normal range (too high or too low)

Ranked causes

  1. 1. Transformer voltage too high (occasional)
  2. 2. Transformer voltage too low (occasional)
  3. 3. UPS Off switch is engaged (occasional)
  4. 4. Broken wire in cable 887763 (rare)
First check

Observe LED status on ISD Power Control PCB

Test point
DS1 (Green), DS2 (Amber), DS3 (Red) LEDs
Expected
DS1 Green LED lit (transformer voltage OK)

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Boot sequence fails at C-Arm battery check (Arrow 18) caution

Boot stops at Arrow 18 when system should check C-Arm battery.

Ranked causes

  1. 1. Generator Events Log indicates battery or sensor fault (common)
  2. 2. C-Arm battery failure or low charge (common)
  3. 3. INV·B·SEN signal fault on Filament Driver PCB (occasional)
  4. 4. INV·B·SEN signal fault on Generator Driver PCB (occasional)
First check

Observe arrow display on Control Panel

Test point
20-character VFD display
Expected
Arrow 18 then progresses to Arrow 19

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot capture navigation image

CAPTURE IMAGE button does not successfully transfer X-ray image to navigation system display

Ranked causes

  1. 1. Image has been modified from Last Image Hold buffer (common)
  2. 2. Excessive motion of phantom or patient during acquisition (common)
  3. 3. Calibration Fixture improperly installed on Image Intensifier (common)
  4. 4. Improper manual fluoro technique during image acquisition (common)
First check

Verify Calibration Fixture attachment

Test point
Image Intensifier mounting point
Expected
Fixture securely fastened, no movement

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication failed between workstation and C-arm

Workstation has not communicated with C-arm for more than 1 second; system finishes in-progress X-ray and disables future X-rays

Ranked causes

  1. 1. Foot pedal toe tapping or repeated control panel button pressing (common)
  2. 2. Workstation software exception or anomalous condition (occasional)
  3. 3. ARCNET communication failure (occasional)
  4. 4. Corrupted persistent object files (rare)
First check

Verify foot pedal and control panel not being pressed rapidly

Test point
C-Arm Control Panel
Expected
No rapid pedal or button activity detected

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Contact bumper sense error caution

System cannot detect contact bumper assembly during boot diagnostics

Ranked causes

  1. 1. Bumper assembly connector defective (common)
  2. 2. Bumper assembly wiring fault (occasional)
  3. 3. Bumper assembly defective (occasional)
  4. 4. MCU defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM print verification fails

DICOM printer not responding or print data transmission fails during verification test

Ranked causes

  1. 1. Printer powered off or disconnected from network (common)
  2. 2. Incorrect IP address in software configuration (common)
  3. 3. Ethernet cable loose or damaged (occasional)
  4. 4. Faulty Ethernet controller on SBC (rare)
First check

Verify printer is powered on and connected to network

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Failed to record Cine to hard-drive radiation

Cine recording does not save to hard-drive when Fluoro and Camera Rotation switches held

Ranked causes

  1. 1. Cine Bridge PCB not seated (common)
  2. 2. FC-AL cable connections loose or disconnected (common)
  3. 3. Cine Bridge PCB LED FC-AL Controller not functioning (occasional)
  4. 4. DC power not supplied to Cine-drive (PS2) (occasional)
First check

Verify Cine Bridge PCB is seated

Test point
Cine Bridge PCB slot
Expected
PCB fully seated

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament current error high voltage

Filament current reading differs from expected value after filament voltage has been verified correct

Ranked causes

  1. 1. Filament aging (rare)
  2. 2. Wrong filament selected (occasional)
  3. 3. Broken connection to filament circuit (occasional)
  4. 4. Current sense gain wrong (occasional)
First check

Verify SM_FIL_ACK* signal state for correct filament relay operation

Test point
SM_FIL_ACK*
Expected
Signal in correct state for selected filament

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament drive pulses absent high voltage

Filament drive pulse missing or both pulses absent; X-ray tube cannot produce X-rays

Ranked causes

  1. 1. Optoisolator failure (common)
  2. 2. Broken connections in filament drive circuit (occasional)
  3. 3. Faulty logic IC (7407 or 74LS14) (occasional)
  4. 4. FIL_CUR sensor fault (rare)
First check

Detect filament drive pulse presence at Generator Driver PCB

Test point
Filament drive pulse input to Generator Driver PCB
Expected
Both filament drive pulses present with normal amplitude and timing
Tools
Oscilloscope

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament supply voltage wrong high voltage

Inability to produce correct filament voltage; system cannot generate proper X-ray exposure

Ranked causes

  1. 1. Bad filament supply regulator (common)
  2. 2. Wrong filament selected (occasional)
  3. 3. Fault in filament voltage sense divider (FIL_B_SN_DIV signal) (occasional)
  4. 4. Output stage fault on Filament Driver PCB (rare)
First check

Verify filament supply voltage regulation

Test point
FIL_B_SEN
Expected
Correct filament voltage as specified

Frame sync signal not detected

TIM frame synchronization signal not generated or signal path is open during diagnostics

Ranked causes

  1. 1. MCU to Backplane wiring fault (occasional)
  2. 2. Camera to Workstation wiring fault (occasional)
  3. 3. Backplane PCB defective (occasional)
  4. 4. MCU PCB defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ground continuity test failure high voltage

Excessive resistance measured between interconnect cable points and AC power plug ground pin during preventive maintenance ground continuity verification.

Ranked causes

  1. 1. Interconnect cable J1-3 (Safety ground) continuity fault (unknown)
  2. 2. Interconnect cable shield (connector shell) continuity fault (unknown)
  3. 3. Interconnect cable J1-9 continuity fault (unknown)
  4. 4. Interconnect cable J1-15 continuity fault (unknown)
First check

Measure continuity at J1-3 safety ground to AC power plug ground pin

Test point
J1-3 to AC Power Plug Ground Pin
Expected
< 0.5 ohm, < 0.5
Tools
Ohmmeter or DMM

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image fails to rotate or invert

Image Rotation and Image Invert functions do not operate; image remains static

Ranked causes

  1. 1. RS-422 communication failure (common)
  2. 2. Cabling to Camera disconnected or damaged (common)
  3. 3. Fluoro Functions PCB not seated (occasional)
  4. 4. Fluoro Functions PCB failure (occasional)
First check

Verify Image Rotation key operation

Test point
Image Intensifier output display
Expected
Image rotates clockwise and counter-clockwise

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image out of focus

Monitor displays video but image is blurry or unfocused

Ranked causes

  1. 1. CCD camera focus error (common)
  2. 2. Camera calibration error (common)
  3. 3. Thermoelectric cooler failure (occasional)
  4. 4. Image intensifier power supply failure (occasional)

Image out of focus radiation

Image appears on monitor but is not sharply focused

Ranked causes

  1. 1. CCD Camera focus error (common)
  2. 2. Camera calibration out of specification (common)
  3. 3. TEC (Thermoelectric Cooler) malfunction (occasional)
  4. 4. II Power Supply output voltage incorrect (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Interconnect cable damaged or connector pins bent caution

Interconnect cable shows abrasions, cuts, excessive wear; Lemo connector pins are bent, missing, or loose; strain relief not secured

Ranked causes

  1. 1. Cable abrasion, cut, or excessive wear (common)
  2. 2. Lemo connector pins bent or missing (occasional)
  3. 3. Lemo connector pins not secured (occasional)
  4. 4. Strain relief not secured to cable or connector (occasional)
First check

Visually inspect interconnect cable for abrasions, cuts, or wear

Test point
Full length of interconnect cable
Expected
No abrasions, cuts, or excessive wear

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard test produces no audible clicks

Letters appear on screen but no click sounds are heard when keys are pressed

Ranked causes

  1. 1. Amplified speaker disconnected from EMI box (common)
  2. 2. Power to amplified speaker disconnected or faulty (common)
  3. 3. No audio output from GPOS SBC to amplified speaker (occasional)
  4. 4. GPOS SBC sound incorrectly adjusted (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lift down not responding caution

C-arm lift down motion fails to activate when requested via RUI or control

Ranked causes

  1. 1. RUI wiring fault (common)
  2. 2. RUI defective (occasional)
  3. 3. MCU wiring fault (occasional)
  4. 4. MCU defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lift up not responding caution

C-arm lift up motion fails to activate when requested via RUI or control

Ranked causes

  1. 1. RUI wiring fault (common)
  2. 2. RUI defective (occasional)
  3. 3. MCU wiring fault (occasional)
  4. 4. MCU defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network error message on DICOM workstation

Workstation displays network functions impaired or network not responding

Ranked causes

  1. 1. Disconnected network cable (common)
  2. 2. Wrong network cable type connected (occasional)
  3. 3. Network is down (occasional)
  4. 4. Network cable or connector damaged (occasional)
First check

Visually inspect and reconnect network cable

Test point
Network port on Workstation and DICOM network
Expected
Cable fully seated with no visible damage

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No +15 volt power to remote user interface caution

Remote control unit not receiving +15V supply voltage

Ranked causes

  1. 1. Faulty Lemo cable or connector connection (common)
  2. 2. Cable or connector defect on assembly 884758 (occasional)
  3. 3. Cable or connector defect on assembly 879450 (occasional)
  4. 4. Cable or connector defect on assembly 884705 (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No +60 volt output from PS4 high voltage

Power supply not supplying required +60V rail

Ranked causes

  1. 1. Power switch on PS4 set to OFF (common)
  2. 2. Open fuse in DC motor power circuit (common)
  3. 3. Loose or faulty ground/connector for cable 902272 (occasional)
  4. 4. Power supply PS4 defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No filament current with filament voltage present high voltage

Filament is not heated despite voltage being applied, preventing X-ray generation

Ranked causes

  1. 1. P3 disconnected from Filament Driver PCB (common)
  2. 2. P1 disconnected from tank A9J1 (common)
  3. 3. Broken connection in cable to tank (occasional)
  4. 4. Filament current sense circuit fault (occasional)
First check

Visually inspect P3 connector at Filament Driver PCB for disconnection or damage

Test point
P3 connector
Expected
Connector firmly seated and undamaged

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No filament voltage observed high voltage

FIL–VOLT signal on Generator Driver PCB reads zero; filament cannot be heated; no X-ray production possible.

Ranked causes

  1. 1. No FIL–B–PLUS voltage; resistors open or cable problem (common)
  2. 2. Transistor open or capacitor short in FIL–B–PLUS circuit (occasional)
  3. 3. No filament drive signals present (occasional)
  4. 4. Faulty or shorted filament voltage sense resistors or sensor failure (occasional)
First check

Measure FIL–B–PLUS voltage supply

Test point
FIL–B–PLUS line at Generator Driver PCB
Expected
Appropriate supply voltage present per schematic

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No image on monitor radiation

Image does not appear on Workstation monitors during X-ray operation

Ranked causes

  1. 1. CCD Camera failure (common)
  2. 2. Image Intensifier tube failure (occasional)
  3. 3. II Power Supply failure (occasional)
  4. 4. Workstation failure (occasional)
First check

Verify Image Control Test generates image

Test point
Workstation monitors
Expected
Image appears on monitor

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No image on monitor

Monitor displays no video signal after CCD camera initialization

Ranked causes

  1. 1. CCD camera failure (common)
  2. 2. Image intensifier failure (common)
  3. 3. Image intensifier power supply failure (occasional)
  4. 4. Workstation failure (occasional)

Power injector test fails high voltage

External interface test indicates no PWR-INJCT signal from System Interface PCB

Ranked causes

  1. 1. No PWR-INJCT signal from System Interface PCB (occasional)
  2. 2. Loose or disconnected wiring to connector (common)
  3. 3. External interface PCB relay failure (occasional)
  4. 4. System interface PCB failure (occasional)
First check

Verify all wiring connections to Room Interface connector are secure

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Rotational servo fault detected caution

Rotational motion controller detects abnormal servo response or collision during diagnostics

Ranked causes

  1. 1. Rotation servo controller wiring fault (occasional)
  2. 2. Rotation motor wiring fault (occasional)
  3. 3. MCU defective (occasional)
  4. 4. Rotation servo controller defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service computer cannot connect to service node

Service computer RUT or UtilitySuite application fails to establish ARCNET connection to service node

Ranked causes

  1. 1. Damaged RS-232 service cable (common)
  2. 2. Damaged ARCNET connector (common)
  3. 3. Loose or damaged cable from J11 to J1 (common)
  4. 4. Failed service ARCNET node (occasional)
First check

Inspect RS-232 service cable for damage and connectors for looseness

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

System fails to complete boot-up from hard-drive

System cannot load software from hard drive after BIOS stage completes

Ranked causes

  1. 1. AC/DC power problem (common)
  2. 2. Loose cabling connectors or unseated PCBs (common)
  3. 3. Faulty System Interface PCB, IDE hard-drive, or corrupt software (occasional)
  4. 4. Faulty component within the Generator (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System heats up too quickly or cooling takes >20 minutes caution

Thermal management problem; excessive heating or slow cooling below 20 minutes

Ranked causes

  1. 1. Fan vent screen blocked (common)
  2. 2. Fan ducting leaks (common)
  3. 3. Heat Pipe thermal contact poor (occasional)
  4. 4. Heat Sink thermal contact poor (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • UPS emergency OFF button functional check
  • Workstation totally inoperative
  • Wrong filament current observed when voltage correct
  • Wrong filament voltage observed
  • X-ray on switch fails to initiate X-rays
  • X-ray On Switch fails to initiate X-rays
  • 12VDC DVD fan B4 not working
  • AC line configuration mismatch
  • AC power cord or plug damaged
  • AC power cord or plug damaged or insulation compromised
  • Antistatic drag wire damaged, loose, or obstructed
  • Auto brightness stabilization failure
  • BAD THERMISTOR DETECTED alarm
  • Bad thermistor detected or thermistor failure
  • Balance wheel does not turn
  • Battery charger failed
  • Battery charger light does not illuminate
  • Boot sequence fails at duty cycle table load (Arrow 13)
  • Boot sequence fails at generator object initialization (Arrow 5)
  • Boot sequence fails at OS loading completion (Arrow 4)
  • Boot sequence fails at pre-charge complete (Arrow 16)
  • Boot sequence fails at stator ramp (Arrow 15)
  • C-Arm hangs failing to boot completely
  • C-Arm hangs failing to boot completely
  • Cable or connection damaged or loose
  • Camera fails to rotate
  • Camera fails to rotate
  • CHARGER FAILED error message displayed
  • Charger light does not illuminate
  • Circuit Breaker trips after inserting power cord into AC receptacle
  • Clutch sensing error
  • Collision detected
  • Collision detection not ready
  • Contact detector fails to detect contact
  • Data error detected
  • Door interlock test fails
  • Failure of one or more ESD-sensitive devices in the Workstation
  • Fans operate but system heats up too quickly or takes longer than 20 minutes to cool
  • GE logo/status screen absent from monitors
  • GE logo/status screen does not appear on LCD monitors during boot
  • High voltage register failure
  • Image fails to rotate or invert
  • Interconnect cable damaged or connector misaligned
  • Interconnect cable damaged or Lemo connector misaligned
  • Interconnect cable or LEMO connector damaged or misaligned
  • Interconnect cable wear or connector damage
  • Interlock system malfunction
  • Intermittent heat messages on LCD monitors
  • Joystick communication failure
  • Joystick initialization error
  • Joystick motion enable error
  • Joystick stuck or not responding
  • Keyboard test produces no audible clicks and no letters on screen
  • Lift column inoperative
  • Lift motor movement test failed
  • Motion Detection - cannot capture navigation image despite no apparent movement
  • Motion disabled or not responding
  • Motion error detected
  • Motion failure detected
  • Network connection not established
  • Network connectivity test failed
  • Network device does not respond to DICOM command
  • No 115 VAC power to PS4 input
  • No bootup
  • No click heard when interconnect cable unplugged from C-Arm
  • No fans operate
  • No power indication on AC Pilot Lamp (DS1)
  • One front caster does not contact floor
  • Orbital servo fault detected
  • Orbital servo overrun error
  • Overload fault displayed
  • Pre-charge voltage error on boot
  • Printer does not power on
  • Printer will not turn on
  • Rear caster does not roll freely on floor
  • Rear caster does not steer freely
  • Remote unit not sensed
  • Rotational servo overrun error
  • Service LED flashing
  • Stator does not start during boot-up
  • Stator does not start during boot-up
  • Stator stops running unexpectedly
  • System does not complete pre-charging, PRE-CHARGE VOLTAGE ERROR displayed
  • System does not complete pre-charging, PRECHARGE VOLTAGE ERROR displayed
  • System fails to boot from BIOS
  • System fails to complete boot-up from BIOS
  • System halted with fatal error
  • UPS error - power supply fault detected
  • UPS internal temperature too high
  • USB mouse cursor not responding
  • Video abnormalities, artifacts, or pixel drop-outs
  • Video abnormalities, artifacts, pixel drop-outs
  • Workstation ARCNET test fails or incomplete boot
  • Workstation boots but C-Arm not responding with charger light on
  • Workstation boots but nothing on C-arm; battery charger light OFF
  • Workstation does not respond to Ping command
  • Wrong filament current observed (voltage correct)
  • Wrong filament voltage observed
  • X-ray exposure switch stuck at startup
  • X-ray ON lamp fails to light when X-rays are generated
  • X-ray switch held in ON position at startup
  • X-ray switch security mismatch
  • AC power cord strain relief clamp loose or missing
  • Actual kVp higher than displayed
  • Actual kVp lower than displayed
  • Actual mA higher than displayed
  • Actual mA lower than displayed
  • Ambient light sensor test screen fails to lighten or darken
  • Antistatic drag wire loose or damaged
  • Audible alarm does not signal exposure start or end
  • Balance wheel is loose on axle bolt or wobbles
  • Bargraph voltage less than 170 VDC during exposure
  • Battery charger evaluation fails
  • Battery Charger Evaluation fails
  • Black image with ABS
  • Black image with technique servo at maximum
  • Boot does not progress past logo/status screen
  • Boot sequence fails at ABS table load (Arrow 14)
  • Boot sequence fails at ARCNET arrow (Arrow 1)
  • Boot sequence fails at filament ramp (Arrow 17)
  • Boot sequence fails at Workstation X-ray enable (Arrow 19)
  • Brake fails to hold cross-arm in lock position
  • Brakes on one or both rear wheels do not engage
  • C-arm anode temperature at 80% capacity
  • C-arm charger failure at startup
  • C-arm charger failure during operation
  • C-arm clutch open during operation
  • C-arm clutch released or disengaged
  • C-arm fails to rotate smoothly through cradle
  • C-Arm fails to rotate smoothly through cradle
  • C-Arm orbital movement not smooth or restricted
  • C-Arm rotational movement not functioning smoothly
  • Calibration: Beam Alignment
  • Calibration: Camera iris stops
  • Calibration: Dose Area Product (DAP) verification
  • Calibration: Enter mA Limit Values
  • Calibration: Laser Aimer beam alignment
  • Calibration: Motorized drive motion and angle positioning
  • Calibration: System software installation
  • Cannot position instrument within 3 mm of anatomical landmark
  • Casters rumble, squeak, or wobble
  • Casters skid
  • CCD Camera rotation binds or exhibits jerky motion
  • CCD camera rotation binds or jerky motion
  • Circuit Breaker trips after engaging power switch
  • Clutch open condition detected
  • Clutch open or not engaged
  • Clutch position sensor error
  • Collimator hesitates, stops, or lugs down
  • Collimator hesitates, stops, or lugs down during movement
  • Collimator hunts for position
  • Collimator hunts for position
  • Collimator iris motor failure
  • Column lift switch stuck
  • Column movement stops before reaching maximum or minimum height
  • Column movement stops before reaching maximum or minimum height
  • Contact detection assembly attachment detected
  • Contact detection assembly removal detected
  • Control panel error
  • Cross-arm scarring, rust or metal splinters present
  • Diagnostic switch configuration error
  • Error message 'FAST STOP ACTIVATED' displayed
  • Evidence of ESD between Workstation and grounded objects
  • Excessive tank temperature
  • Excessive temperature at High Voltage Tank
  • Failed attempt to start UPS on battery power
  • Fan B3 runs too slowly, has bearing noise or does not run at all
  • Fan runs too slowly, has bearing noise or does not run at all (B1/B2)
  • Fast stop activated
  • Filament supply voltage wrong
  • Flip Flop brake does not prevent C-arm rotation
  • Flip Flop brake handle insufficient tension
  • Flip-Flop brake does not prevent C-Arm rotation
  • Fluid ingress into equipment
  • Front caster binds or wobbles on axle bolt
  • Front caster does not pivot freely on vertical axis
  • GE logo/status screen appears on both monitors but boot does not progress
  • GE logo/status screen appears on one LCD monitor but not the other during boot
  • GE logo/status screen on one monitor only
  • Generator communication loss detected
  • Hand Switch indicator lights off
  • Hand switch indicator lights off with VFD OK
  • High voltage generator error
  • High voltage generator software error
  • HLF footswitch fails to initiate X-rays
  • Horizontal Cross-Arm brake fails to lock position
  • HVL below specification range
  • Image resolution inferior to specification
  • Image too dense or too light
  • Incorrect RUI or C-Arm orientation relative to patient
  • Interconnect cable damaged or worn
  • Interlock circuit open
  • Interlock system down preventing X-rays
  • Interlock system down preventing X-rays
  • Invalid port entry
  • Joystick does not respond during initialization
  • Key disable signal error
  • Key switch error
  • L-Arm brake does not secure L-Arm
  • L-Arm brake does not secure L-Arm
  • Laser Aimer beam not visible after power on
  • Laser Aimer misalignment or not securing
  • Laser beam not visible after Laser Aimer powered on
  • Laser beam not visible after power on
  • Laser beam not visible after power on
  • LEMO connector not mating properly
  • Lift button non-functional
  • Lift switch pressed continuously for 30 seconds
  • Lift switch stuck on control panel
  • Low contrast image quality
  • Low contrast images
  • Mode footswitch fails to toggle modes
  • Motion calibration required
  • Motion calibration required
  • Motion stop button not functioning
  • Motorized drive does not respond to control input
  • Motorized motion disabled
  • Motorized orbital movement not functioning smoothly
  • No bootup with VFD and handswitch indicator OK
  • No click heard when workstation plugged into AC power
  • No dynamic image displayed during Cine recording
  • No fans operate
  • No filament current (filament voltage present)
  • No video and no pilot tone
  • No video but pilot tone OK
  • No video on monitor
  • No video output on display
  • No voltage from CAM COOLER PWR to CAM COOLER RTN
  • Only one caster brake works
  • Output film or paper is blank
  • OVERLOAD FAULT
  • Overload fault in HV regulator
  • OVERVOLTAGE ERROR
  • Overvoltage error displayed
  • Overvoltage fault in HV generator
  • Pre-charge timeout error on boot
  • Precharge voltage too high during startup
  • Printer output is blank
  • Red UPS OFF button does not light
  • Red UPS OFF button does not light when pressed
  • Relay not activating during boot test
  • Resolution of captured navigation image is not as good as original
  • Rotation of captured image is different than original
  • RUI button stuck at boot
  • RUI communication failure
  • RUI communication failure
  • RUI control key stuck
  • RUI key stuck
  • RUI key stuck or button stuck
  • Spots on image
  • Spots visible on printed image
  • Stator motor does not start during boot
  • Stator motor stops during operation
  • Stator stops running after boot
  • Stop Motion button does not halt movement or provides no feedback
  • System does not boot - VFD stops at 14th arrow
  • System does not boot - VFD stops at 18th arrow
  • System does not boot – VFD display stops at 14th arrow
  • System does not complete pre-charging, PRE-CHARGE TIME-OUT ERROR displayed
  • System does not complete pre-charging, PRECHARGE TIME-OUT ERROR displayed
  • System does not shut down immediately after red button press
  • System does not shut down immediately when red UPS OFF button pressed
  • System does not shut down immediately when UPS OFF button pressed
  • System won't record navigation snapshot to CD/DVD
  • System won't verify correct registration
  • TEC Cooling test failure
  • UPS communication error
  • UPS does not power off within 10 seconds
  • UPS does not power on
  • UPS does not start
  • Vertical column lift buttons not working
  • Vertical motion commanded but fails
  • Vertical movement not functioning or jerky
  • VFDs inoperative
  • VFDs inoperative with handswitch indicator OK
  • Video present but no pilot tone
  • White image with ABS
  • White image with technique servo at minimum
  • Wig Wag brake fails to prevent motion
  • Wig-Wag brake fails to prevent motion
  • Workstation boots but C-Arm not responding with charger light off
  • Workstation does not boot
  • Workstation keyboard communication failure
  • Workstation UPS communication failure
  • Workstation UPS failure
  • Workstation will not track in a straight line
  • Wrong filament selection
  • Wrong filament selection
  • X-ray disable keyswitch in standby position
  • X-ray exposure switch stuck during boot
  • X-ray key switch in standby position
  • X-ray key switch in standby position
  • X-RAY ON indicator not lit during exposure
  • X-ray overtime; extended exposure
  • X-ray room door open
  • X-ray room door open
  • X-ray switch security line mismatch
  • X-ray switch security mismatch
  • X-ray tube housing hot
  • X-ray tube overheating or thermal runaway
  • X-ray tube temperature sensor failure
  • X-rays temporarily disabled, FFB reboot
  • 100 kVp @ 300 mAs exposure fails to execute
  • 24 hour recharge required — X-rays disabled
  • 24-hour recharge required
  • 4 hour warm-up required
  • 4-hour warm-up required
  • 4-hour warmup required
  • A/D channel failure at startup
  • A/D converter channel failure
  • AC input voltage abnormal
  • AC line configuration does not match measured voltage
  • AC line power loss detected
  • AC line voltage configuration mismatch
  • AC line voltage not configured
  • AC line voltage not configured
  • AC line voltage not configured
  • AC power failure detected by UPS
  • AC power line sensor error
  • AC power line sensor error detected
  • AC power line sensor error detected
  • AD channel fail - power off
  • AD channel fails during startup
  • AD channel failure
  • AD channel failure
  • AD Channel Failure error during boot
  • AD CHANNEL FAILURE message displayed
  • All four indicators flashing; Workstation beeps 3 times every 10 seconds
  • All squares and arrows indicators did not illuminate; error message displayed
  • Ambient light sensor does not adjust monitor brightness
  • Ambient light sensor not responding correctly
  • Anatomical profile corruption detected
  • Anatomical profile not supported
  • Anatomical profiles disabled
  • Anatomical profiles disabled - invalid format
  • Anatomical profiles disabled - serial number mismatch
  • Anatomical profiles unavailable with C-arm
  • Anode is hot — HLF disabled
  • Anode is hot (C-arm display)
  • Anode is warm
  • Anode is warm (C-arm display)
  • Anode rotor not rotating after X-ray shot
  • Anode rotor not spinning
  • Anode temperature at 70-80% capacity
  • Anode temperature at 80% capacity or greater
  • Anode temperature at high capacity threshold
  • Anode temperature at warm threshold
  • Antistatic drag wire damage or detachment
  • Antistatic drag wire damaged or loose
  • Antistatic drag wire loose or damaged
  • Antistatic drag wires damaged or not secured
  • Antistatic drag wires loose, damaged, or obstructed
  • Antistatic drag wires not securely attached or damaged
  • Arcing detected in High Voltage Tank
  • Arcing or carbon trails observed
  • ARCNET transmit pipe temporarily full
  • Armed (film exposure preparation)
  • Artifacts present in fluoro image with Laser Aimer
  • Artifacts present in fluoroscopy image with Laser Aimer attached
  • Artifacts visible in fluoro image with laser aimer
  • Artifacts visible in fluoro image with laser aimer attached
  • Auto Technique Tracking out of range
  • Battery and Overload LEDs on simultaneously
  • Battery charge below 40 percent
  • Battery charge low
  • Battery charge not restored to nominal voltage
  • Battery charge voltage does not reach nominal after exposure
  • Battery charge voltage not at nominal values
  • Battery charger failure
  • Battery charger light does not illuminate
  • Battery charger light is off with workstation powered off
  • Battery charger not operating during boot
  • Battery pack evaluation fails
  • Battery Power LED on and Battery low
  • Battery voltage low during operation
  • Beam Alignment Verification
  • Boot sequence fails at ADC check (Arrow 11)
  • Boot sequence fails at CMOS initialization (Arrow 8)
  • Boot sequence fails at early stage
  • Boot sequence fails at error handler initialization (Arrow 7)
  • Boot sequence fails at generator bridge data initialization (Arrow 12)
  • Boot sequence fails at generator event table load (Arrow 9)
  • Boot sequence fails at generator object relationships (Arrow 6)
  • Boot sequence fails at persistent data CRC (Arrow 2)
  • Boot sequence fails at PIO initialization (Arrow 10)
  • Boot sequence fails at timer initialization (Arrow 3)
  • Brakes do not engage completely
  • Brakes do not release
  • Buzzer sounds when workstation switched on
  • C-arm 24 hour recharge required
  • C-arm 4 hour warm-up required
  • C-arm anode temperature at 70-80% capacity
  • C-Arm batteries critically low
  • C-Arm batteries low charge
  • C-Arm batteries require 24-hour recharge
  • C-arm battery charge at 40-70%
  • C-Arm clutch is open or released
  • C-Arm collision detected
  • C-arm collision detected during motorized motion
  • C-Arm contact with Image Intensifier detection assembly
  • C-arm does not rotate freely when Flip Flop brake released
  • C-Arm does not rotate freely when Flip-Flop brake released
  • C-Arm or Workstation grounding defect
  • C-arm reconnected to workstation
  • C-Arm requires 4-hour warm-up after 60 days inactivity
  • C-Arm rotational movement not smooth or restricted
  • C-arm system armed for film exposure
  • Calibration: Auto Technique Tracking
  • Calibration: Battery charger output voltage indicator
  • Calibration: Battery charger output voltage indicator
  • Calibration: Beam alignment verification
  • Calibration: Beam Alignment Verification
  • Calibration: C-Arm angle position
  • Calibration: C-Arm orbital movement
  • Calibration: Camera cooler
  • Calibration: Clutch switch replacement and verification
  • Calibration: Collimator and filament after software installation
  • Calibration: Collimator Iris Blade Positioning
  • Calibration: Collision detection function
  • Calibration: Contact detection function
  • Calibration: Cumulative Dose Display Accuracy (Skin Dose CAK mode)
  • Calibration: DAP accuracy verification
  • Calibration: DAP Collimator and Iris Calibration (Dual-leaf systems)
  • Calibration: DC power supply PS1 +5.1V output
  • Calibration: DC power supply PS1 9 (C-Arm)
  • Calibration: Dead time between kV drive signals
  • Calibration: Differentiator setup
  • Calibration: Dose Area Product (DAP)
  • Calibration: Dose Area Product (DAP) Calibration Setup
  • Calibration: Dose Area Product (DAP) single-leaf cardiac systems
  • Calibration: Dose Area Product (DAP) Tube Calibration
  • Calibration: Dose Area Product (DAP) verification
  • Calibration: Dose Display Accuracy (DAP mode)
  • Calibration: Dose Display Accuracy (Skin Dose AKR mode)
  • Calibration: End stop position - C-Arm not at 90 degrees horizontal
  • Calibration: Filament and Collimator
  • Calibration: Filament/duty cycle
  • Calibration: Fluorographic pulsed mA accuracy test
  • Calibration: Fluoroscopic kVp accuracy at 5 mA
  • Calibration: Fluoroscopic kVp accuracy at maximum mA
  • Calibration: Generator and functional tests
  • Calibration: High Voltage Supply Regulator
  • Calibration: High voltage supply regulator mA null offset
  • Calibration: Home position - Image Intensifier out of level
  • Calibration: Image intensifier resolution
  • Calibration: Image tube sizing
  • Calibration: ISD PCB installation and verification
  • Calibration: Isolation transformer output
  • Calibration: kVp accuracy at 5 mA
  • Calibration: kVp accuracy at maximum mA
  • Calibration: Lens focus
  • Calibration: Lock-out/Tag-out procedure
  • Calibration: Mag field size
  • Calibration: Monitor arm elbow pivot
  • Calibration: Monitor arm shoulder pivot
  • Calibration: Monitor arm vertical balance
  • Calibration: Monitor arm wrist pivot
  • Calibration: Monitor Latch Cleaning
  • Calibration: Motorized drive angular positioning
  • Calibration: Motorized drive coupling
  • Calibration: Multi-output power supply PS2
  • Calibration: Orbital clutch anti-rotation stop
  • Calibration: Orbital clutch anti-rotation stop
  • Calibration: Orbital clutch anti-rotation stop position
  • Calibration: Orbital potentiometer RP2
  • Calibration: Power supply PS1 DC outputs
  • Calibration: Power supply PS2 DC outputs
  • Calibration: PWM chip synchronization
  • Calibration: Ramp peak voltage
  • Calibration: Remote Control Screen function
  • Calibration: Rotation potentiometer RP1 drive belt replacement
  • Calibration: Rotational potentiometer RP1
  • Calibration: Shoulder Pivot
  • Calibration: Skin dose limit rate measurement and adjustment
  • Calibration: Skin dose rate limit verification
  • Calibration: Skin Dose Rate Limit Verification at Maximum kV/mA (HLF mode)
  • Calibration: Skin dose rate verification
  • Calibration: Stability pot (R83) overshoot adjustment
  • Calibration: Surgical Navigation receiver or transmitter installation
  • Calibration: Surgical Navigation system functional test setup
  • Calibration: System Lock-Out/Tag-Out procedure
  • Calibration: System Software Installation
  • Calibration: Verify mA Limit Values and Maximum HLF
  • Calibration: Vertical Arm Balance
  • Calibration: X-ray On pushbutton replacement
  • Camera rotation symbols not appearing or rotating incorrectly
  • Cannot make network connection
  • Cannot open port on firewall
  • Cannot read image file from CD/DVD
  • Cannot write to removable media
  • CD/DVD drive failure
  • CD/DVD drive not responding
  • CD/DVD drive not responding or not properly installed
  • Charger failed (during operation)
  • Charger failed at boot-up
  • Charging current below specification
  • Charging voltage not reaching specification
  • Cine disk not available
  • Cine file corrupted in Image Directory
  • Cine recording or playback not functioning
  • Clutch fails to engage
  • Clutch fails to release
  • Clutch is open
  • Clutch is open (C-arm)
  • Clutch sense signal not responding
  • Coin cell battery replacement required
  • Collimation control failure
  • Collimator alignment verification
  • Collimator calibration mismatch
  • Collimator calibration required
  • Collimator calibration values mismatch
  • Collimator displays error message
  • Collimator iris larger than indicated
  • Collimator iris malfunction
  • Collimator iris potentiometer error
  • Collimator iris stuck or inoperable
  • Collimator iris too large
  • Collimator iris unstable
  • Collimator leaves malfunction
  • Collimator leaves rotation not functioning
  • Collimator preview icons not indicating correct function
  • Collimator replacement
  • Collimator stuck
  • Collision detected
  • Collision detected on C-arm
  • Collision Detection Assembly status change
  • Collision detection signal anomaly
  • Collision detector fails to detect collision
  • Collision detector reports false collision
  • Column movement is noisy or rough
  • Column movement is noisy or rough
  • Communication error between processors
  • Communication failed between Generator and Workstation
  • Communication failed with generator
  • Communication failure with navigation computer
  • Contact bumper installed and undamaged
  • Contact detected on image intensifier
  • Contact detected on Image Intensifier
  • Contact detected on Image Intensifier assembly
  • Contact detected with detection assembly
  • Contact detection signal not responding
  • Contact detector reports incorrect contact
  • Control panel communication error
  • Control panel does not display technique values
  • Control panel key stuck at startup
  • Control panel key stuck during boot
  • Control panel key stuck during boot
  • Copy failure during disk write
  • Corrupted image cannot be displayed
  • Corrupted images deleted
  • Corrupted images detected at bootup
  • Corrupted software or data detected
  • CPU or OS runtime exception occurred
  • Cross-arm hits rough spots or grinds during extension/retraction
  • Current anatomical profile disabled
  • Current anatomical profile disabled
  • Current anatomical profile disabled
  • DAP display shows incorrect units
  • DATA ERROR message appears during boot
  • Data loss or disk drive damage during operation
  • Delete all complete; reboot required
  • Delete all patient information in progress
  • DICOM device configuration error
  • DICOM error
  • DICOM error
  • DICOM secondary capture cine not supported
  • DICOM server out of resources
  • DICOM server out of resources
  • DICOM server out of resources
  • DICOM server timeout
  • Different C-arm connected to workstation
  • Different C-Arm connected to Workstation
  • Disk full during copy operation
  • Disk not writable for copy operation
  • Display studies and patient information on DICOM server fails
  • Documentation of DAP Calibration
  • Dose Area Product (DAP) reading inaccuracy
  • Dose Area Product calibration required
  • Dose Area Product verification
  • Dose calibration required
  • Elbow pivot movement too tight or too loose
  • Electric shock from high voltage components after power removal
  • Emergency response during energized electrical work
  • Energized electrical work on 0-50V circuits
  • Energized electrical work on 120-240V circuits
  • Energized electrical work on 240-480V circuits
  • Energized electrical work on 50-120V circuits
  • Entity already exists during DICOM retrieve
  • Entrance exposure rate limits verification
  • Equipment instability on slopes or uneven floors
  • Equipment rolls uncontrolled during movement
  • Error processing DICOM file
  • Error reading disk
  • Error reading disk
  • Error reading disk
  • Error retrieving images from PACS
  • Error writing disk
  • Failed to open firewall port
  • Failed to read image file
  • Failed to read image file from media
  • Fast Stop switch does not stop Vertical Column movement
  • Fast Stop switch does not stop Vertical Column movement
  • Field size adjustment not functioning
  • Field size selection failure
  • Filament calibration required
  • Filament current error; wrong filament current reading
  • Filament current out of tolerance
  • Filament drive pulses absent
  • Filament Driver PCB replacement
  • Filament not calibrated
  • Filament regulator failure
  • Filament regulator output out of tolerance
  • Filament select error
  • Filament size mismatch
  • Filament size mismatch with hardware
  • Filament voltage error; wrong voltage applied to filament
  • File system shutdown test via UtilitySuite
  • Film exposure exceeded time limit
  • Film exposure switch released early
  • Film exposure time exceeded
  • Film mode exposure unsuitable or incorrect
  • Film mode functional test
  • Film mode pre-arm fails
  • Fingerprint test for position accuracy
  • Flammable gas leak near equipment
  • Flip-Flop motion restricted or stiff
  • Fluoroscopy image rotation or reversal incorrect
  • Footswitch configuration error detected
  • Footswitch security error
  • Frame sync error disables X-rays
  • Frame sync error during boot
  • Frame Sync Error during operation
  • Generator batteries test - error messages
  • Generator communication lost during exposure
  • Generator Interface PCB installation verification
  • Generator interface PCB requires replacement
  • Gravitational/mechanical energy hazard during component removal
  • Gravitational/mechanical energy hazard mitigation - C-arm with L-arm (Standard C)
  • Gravitational/mechanical energy hazard mitigation - C-arm without L-arm (Super C)
  • Heat management messages do not appear on workstation monitor
  • Heat Management test messages do not appear on Workstation Monitor
  • High current electric shock from battery packs
  • High kVp detected in kV feedback circuit
  • High Level Fluoro exposure time exceeded
  • High level fluoro overtime alarm
  • High mA detected in mA feedback circuit
  • High Voltage and Control Cable Assembly requires replacement
  • High voltage cable damage
  • High voltage cable damaged
  • High voltage cable damaged or worn
  • High voltage cable damaged or worn
  • High voltage generated without X-ray switch activation
  • High voltage generated without X-ray switch activation
  • High voltage on without X-ray switch activation
  • High voltage overvoltage fault
  • High voltage regulator overload fault
  • High voltage regulator saturation fault
  • High-capacity disk drive failure
  • High-capacity disk failure
  • High-voltage regulator overload fault
  • High-voltage regulator saturation fault
  • HLF Footswitch fails to initiate X-rays
  • HLF overtime - exposure time exceeded
  • Horizontal Cross-Arm grinding or rough spots during motion
  • Housing is hot — HLF disabled
  • Housing is warm
  • Housing overheated, X-rays disabled
  • Housing temperature at 70-80% capacity
  • Housing temperature at 80% capacity or greater
  • HV cable damaged or degraded
  • HV cable damaged or worn
  • HV cable damaged or worn
  • HV generator error
  • HV register failure at startup
  • Image annotation and measurements removed during zoom
  • Image brightness adjusts but appears too light or dark
  • Image file exceeds DICOM size limit
  • Image not centered in fluoroscopy
  • Image orientation inversion/mirroring not functioning
  • Image quality unacceptable
  • Image resolution out of specification
  • Image rotation not functioning
  • Image save function not functioning
  • Image swap between monitors not functioning
  • Image too big to copy
  • Image too large to copy
  • Incompatible DICOM format
  • Insufficient disk space for copy completion
  • Insufficient disk space to retrieve DICOM images
  • Insufficient media capacity
  • Insufficient media capacity
  • Insufficient media capacity for copy
  • Insufficient space on current disk for spanning
  • Intelligent Shutdown logic failure
  • Intelligent Shutdown Power Control PCB replacement
  • Intelligent Shutdown test
  • Interconnect cable damaged or worn
  • Interconnect cable defect
  • Interconnect cable strain relief loose or damaged
  • Interlock circuit failure at power-up
  • Interlock circuit failure at startup
  • Invalid IP address
  • Invalid IP address
  • Invalid IP address entered
  • Invalid port number for DICOM local server
  • Invalid port number for LOCAL server
  • Invalid port number for Query/Retrieve
  • Invalid port number for REMOTE server
  • Invalid profile format
  • Invalid year value entered
  • Isolation transformer output out of specification
  • Isolation transformer output out of specification
  • Isolation transformer output voltage out of specification
  • Isolation transformer secondary voltage out of specification
  • Joystick failure
  • Joystick stuck
  • Joystick X position error high
  • Joystick X position error low
  • Joystick X-axis not responding properly
  • Joystick Y position error high
  • Joystick Y position error low
  • Joystick Y-axis not responding properly
  • Key stuck on control panel
  • Key switch in standby; X-rays disabled
  • Key switch signal anomaly
  • Keyboard communication failure
  • kV loop error: actual kVp higher than indicated
  • kV loop error: actual kVp lower than indicated
  • KV on in error
  • kVp accuracy out of specification
  • kVp accuracy out of specification (Keithley meter)
  • kVp accuracy out of specification (RadCal meter)
  • kVp regulation error exceeds high tolerance
  • kVp regulation error exceeds low tolerance
  • kVp values outside Auto-Technique Tracking range
  • L-Arm does not rotate freely
  • L-Arm does not rotate freely
  • Laser Aimer beam not aligned with crosshairs
  • Laser Aimer causes image artifacts
  • Laser aimer does not lock to C-arm
  • Laser Aimer fluoro image artifacts present
  • Laser beam not centered on crosshairs
  • Lemo connector pins loose or unsecured
  • Lift column inoperative
  • Lift Down request signal anomaly
  • Lift switch stuck or continuously pressed
  • Lift Up request signal anomaly
  • Light or dark image with working ABS
  • Line power was lost
  • Loading Low Dose ABS table
  • Loading Standard ABS table
  • Login failed
  • Login failed - incorrect password
  • Login password incorrect
  • Logout confirmation prompt
  • Logout in progress
  • Lost connection to DICOM server
  • Lost connection to DICOM server
  • Lost connection to DICOM server during retrieve
  • Low kVp detected in kV feedback circuit
  • Low mA detected in mA feedback circuit
  • Low mA messages displayed
  • Low mA output
  • Low UPS battery message appears during shutdown
  • mA loop calibration error: actual mA higher than indicated
  • mA loop calibration error: actual mA lower than indicated
  • mA on in error
  • MA ON IN ERROR message after HVSR replacement
  • mA output calibration error exceeds high tolerance
  • mA output calibration error exceeds low tolerance
  • mA sensor failure
  • mA sensor failure detected
  • MCU CMOS requires reconfiguration after battery replacement
  • Media already contains data
  • Media ejected prematurely
  • Media in CD/DVD drive not writable
  • Media in device for copy operation not writable
  • Mode control failure in vascular mode
  • Mode Footswitch fails to toggle modes of operation
  • Monitor arm elbow pivot too tight or too loose
  • Monitor arm shoulder pivot too tight or too loose
  • Monitor arm vertical balance issue
  • Monitor arm wrist pivot stiff or drifts
  • Monitor latch accumulates gray residue
  • Monitor latch accumulation of gray residue
  • Motion disabled
  • Motion disabled by safety stop
  • Motion enabled signal not responding
  • Motion error
  • Motion failure
  • Motion failure
  • Motion stop button will not clear
  • Motion stop function fails
  • Motion stop status signal anomaly
  • Motorized C-Arm rotational motion calibration
  • Motorized drive motion calibration required
  • Motorized drive not calibrated
  • Motorized motion disabled
  • Motorized orbital motion calibration
  • Motorized orbital motion with clutch released
  • Navigation computer not responding
  • Neither caster brake works
  • Network configuration requires reboot after changes
  • Network connection failure
  • Network Verification test
  • New disk needed
  • New password and retyped password do not match
  • New password too short
  • New password too short
  • No +5 volt power to servo drive sensors
  • No alarm sounds when red UPS OFF button pressed
  • No alarm sounds when UPS OFF button pressed
  • No anode motor sound during film mode
  • No backup power after AC loss
  • No beep sound when footswitch pressed
  • No DICOM images match search criteria
  • No disk in CD/DVD drive
  • No disk in drive
  • No Ethernet IP address or subnet mask configured
  • No fluoroscopy image displayed
  • No match found
  • No media in device for copy operation
  • No media in device for copy operation
  • No response, server timed out
  • No static image on left monitor
  • No static image on left monitor (last image hold)
  • Not enough disk space to retrieve images
  • Not enough space on current disk
  • NVRAM CRC test failed
  • Old password incorrect
  • Old password incorrect on change
  • On Battery and Overload indicators both lit; Workstation beeps once per second
  • On Battery indicator flashes, all others off; Workstation beeps every 2 seconds
  • On Battery indicator flashing; Service indicator flashing; Workstation beeps continuously
  • On Battery indicator glows, all others off; Workstation beeps every 4 seconds
  • On Battery indicator lit; Service indicator flashes; Workstation beeps continuously
  • One lift button functional, other non-functional
  • One or two fans do not operate
  • One or two fans do not operate
  • One vertical column lift button works, other does not
  • Operation failed with unexpected error
  • Orbital auxiliary signal not responding
  • Orbital brake fails to lock C-Arm
  • Orbital brake fails to lock C-arm position
  • Orbital fault signal not responding
  • Orbital motion displays incorrect angle
  • Orbital motion displays incorrect direction
  • Orbital movement fault detected
  • Orbital position potentiometer signal anomaly
  • Orbital servo controller not responding
  • Output from UPS abnormal on battery power
  • Output voltage out of limits on battery power
  • Overvoltage condition detected
  • Overvoltage fault in high-voltage regulator
  • Overvoltage message displayed
  • Password entry too short
  • Password mismatch on change
  • Patient data deletion completed
  • Patient data deletion in progress
  • Patient health information deletion
  • Persistent error message during operation
  • Ping response does not return all transmitted packets
  • Plug cover screws loose on prongs
  • Poor image quality: pixel blemishes
  • Poor image quality: rolling or torn image
  • Poor image quality: unstable image
  • Poor image quality: video noise
  • Power cord strain relief clamp missing or loose
  • Power cord strain relief clamp missing or loose
  • Power not lost to workstation on cord removal
  • Power not lost to workstation when cord unplugged
  • Power On and On Battery indicators both lit; Workstation beeps every 2 seconds
  • Power On and Overload indicators lit; Workstation beeps once per second
  • Power On indicator lit; Service indicator flashes; Workstation beeps continuously
  • Power On, Battery, and Overload indicators flashing; Workstation beeps continuously
  • Power On, On Battery, and Service indicators lit; Workstation beeps continuously
  • Power supply cooling fans run too slowly, have bearing noise or do not run
  • Pre-charge voltage does not reach battery voltage
  • Pre-charge voltage exceeds maximum threshold
  • Pre-charge voltage oscillates between 280V and 180V
  • Precharge circuit timeout
  • Precharge circuit timeout during startup
  • Precharge circuit timeout during startup
  • Precharge voltage error
  • Precharge voltage error during startup
  • Precharge voltage too high
  • Printer communication error
  • Printer does not respond to print command
  • Printer error
  • Printer fan maintenance
  • Printer is out of film or paper
  • Printer offline
  • Printer offline
  • Printer out of film or paper
  • PS1 or PS2 replacement
  • PS3 power supply requires replacement
  • Radiographic exposure completes with failure indication
  • Red UPS OFF button alarm does not sound
  • Red UPS OFF button does not light up
  • Remote control battery is low
  • Remote control battery low
  • Remote control cursor or buttons not responding
  • Remote control screen not displaying
  • Remote user interface communication failure
  • Remote user interface key stuck
  • Replace C-Arm power supply PS3
  • Requested operation failed
  • Resolution below specification on image intensifier
  • Restore power after service
  • Restore power after service completion
  • Retrieve image(s) from DICOM server fails
  • Room door open
  • Rotational auxiliary signal not responding
  • Rotational fault signal not responding
  • Rotational motion displays incorrect angle
  • Rotational motion displays incorrect direction
  • Rotational movement fault detected
  • Rotational position potentiometer signal anomaly
  • Rotational servo controller not responding
  • RUI acquisition controls not functioning
  • RUI key press not generating unique code
  • RUI sense signal not responding
  • RUI software version not displaying
  • Running filament calibration
  • Saturation fault in HV regulator
  • Save operation failed
  • Scratch appears on printout
  • Selected image cannot be displayed
  • Serial number mismatch between C-arm and Workstation
  • Shoulder pivot movement too tight or too loose
  • Skin spacer installation verification
  • Software or data corruption detected
  • Stack memory dangerously low
  • Standard operating menus absent from right monitor
  • Standard operating menus do not appear on right LCD monitor
  • Stator not on; power off
  • Stator standby
  • Steering brake ineffective on rear caster
  • Store studies and patient information on DICOM server fails
  • Strain relief not securely attached
  • System communication failure
  • System countdown does not function or display during UPS low battery test
  • System demo test for orbit and rotation movement
  • System does not boot – VFD display stops at 18th arrow
  • System does not boot successfully
  • System does not complete pre-charging with relay clicking
  • System does not complete pre-charging, relays click every few seconds
  • System does not complete pre-charging, relays click every few seconds
  • System does not execute controlled shutdown
  • System error detected
  • System error detected, CPU or OS exception
  • System error, X-rays disabled
  • System error: CPU or OS exception detected
  • System error: excessive or missing frame sync
  • System error: frame sync loss detected
  • System fails to boot after pre-charge completes
  • System fails to compensate for attenuation caused by copper filters
  • System fails to compensate for copper filter attenuation
  • System fails to complete 100 kVp @ 300 mAs exposure
  • System halted with out of memory error
  • System is busy
  • System labels damaged or missing
  • System labels missing or damaged
  • System not shut down properly
  • System performing internal processing
  • System running calibration procedure
  • System running calibration procedure
  • System shuts down unexpectedly without error messages
  • System unexpectedly shuts down during operation
  • System updating or performing internal tests
  • System updating or running internal tests
  • System won't work with C-Arm disconnected
  • TEC failed open
  • Temperature sensor failure
  • Testing and positioning during temporary LOTO removal
  • Testing and positioning with temporary LOTO removal
  • Touchscreen accuracy verification
  • Unintended system shutdown during operation
  • Unique ID field does not match
  • Unique ID field incorrect
  • Unique ID field mismatch
  • Unique ID mismatch between label and Service Screen
  • Unknown error from removable media
  • Unknown error response from removable media
  • Unsupported print format for printer
  • Update system configuration files after calibration
  • UPS battery low warning does not appear
  • UPS does not provide backup power
  • UPS does not provide backup power after AC disconnect
  • UPS does not provide backup power to workstation
  • UPS does not provide expected backup time
  • UPS does not provide expected backup time
  • UPS does not turn off after red button press
  • UPS does not turn off after red button pressed
  • UPS front panel lights do not turn off after shutdown
  • UPS internal fan has failed
  • UPS lights flash or system beeps during self-test
  • UPS operates in battery mode only
  • UPS operates in battery mode only despite normal utility power
  • UPS operates in battery mode only despite utility power present
  • UPS output abnormal on battery
  • UPS power backup system functional verification
  • UPS self test
  • UPS Self Test fails
  • Verification: ADDITIONAL APPLICATIONS function
  • Verification: AUTO CONTRAST/BRIGHTNESS function
  • Verification: BLANK key function
  • Verification: BRIGHTNESS function
  • Verification: Clipping from Secondary Collimator Ring in Normal Mode
  • Verification: Collimator Preview Lines Alignment
  • Verification: CONTRAST function
  • Verification: DAP system accuracy
  • Verification: ENHANCE function
  • Verification: IMAGE ANNOTATION function
  • Verification: Image Clipping at Multiple C-arm Angles and Magnification Modes
  • Verification: IMAGE DIRECTORY function
  • Verification: LOGOUT key function
  • Verification: MAG1 and MAG2 Collimator Field Size
  • Verification: Minimum Collimator Field Size
  • Verification: NEGATE function
  • Verification: NOISE FILTER function
  • Verification: Normal Mode Beam Diameter and Collimator Offset
  • Verification: PATIENT INFORMATION function
  • Verification: SAVE function
  • Verification: SETUP function
  • Verification: SWAP function
  • Verification: ZOOM function
  • Verify operation failed
  • Vertical arm balance incorrect; assembly lifts or sags
  • Vertical column operational with keyswitch disabled
  • Vertical column operational with keyswitch in DISABLE position
  • Vertical white stripe appears on printout
  • White stripes persist after cleaning head and roller
  • Wig Wag brake handle position incorrect
  • Wig Wag motion reduced or sluggish
  • Workstation and generator are not compatible
  • Workstation and Generator software incompatibility
  • Workstation and Generator software incompatible
  • Workstation brake not operating properly
  • Workstation controlled shutdown test
  • Workstation does not perform controlled shutdown
  • Workstation keyboard communication failed
  • Workstation keyboard error
  • Workstation UPS communication failure
  • Workstation UPS failure
  • Workstation won't automatically configure itself
  • Workstation won't communicate correctly after configuration
  • Worn or damaged balance wheel tread
  • Worn or damaged front caster tread
  • Worn or damaged rear caster tread
  • Wrist pivot movement too tight or too loose
  • Wrong mA during exposure
  • X-ray control released early in film mode
  • X-ray exposure exceeded time limit
  • X-ray exposure switch released early
  • X-ray exposure switch stuck during boot
  • X-ray exposure time exceeded
  • X-ray housing overheated and thermal cutoff activated
  • X-ray housing temperature at high capacity threshold
  • X-ray housing temperature at warm threshold
  • X-ray housing temperature sensor failure
  • X-ray housing temperature sensor failure
  • X-ray indicators not functioning
  • X-ray ON indicator light not illuminating
  • X-ray switch released before desired mAs reached
  • X-ray switch security error
  • X-ray switch security error
  • X-ray tube anode is hot
  • X-ray tube anode is warm
  • X-ray tube anode not powered
  • X-ray tube current detected without switch activation
  • X-ray tube current detected without switch activation
  • X-ray tube current on without X-ray switch activation
  • X-ray tube housing overheated
  • X-ray tube housing temperature at 70-80% capacity
  • X-ray tube housing temperature at 80% capacity
  • X-ray tube housing temperature high
  • X-ray tube housing temperature warming
  • X-ray tube housing warm
  • X-ray tube stator not operational
  • X-ray tube stator not rotating
  • X-ray tube stator not rotating
  • X-ray tube temperature sensor failure
  • X-rays disabled
  • X-rays disabled
  • X-rays disabled and system reboot required
  • X-rays disabled by software interlock
  • X-rays disabled during bootup
  • X-rays initiated with keyswitch in disable position
  • X-rays initiated with keyswitch in DISABLE position
  • X-rays temporarily disabled
  • X-rays temporarily disabled during FFB reboot

1092 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01Checking FlashSystem power-up or firmware initialization phaseinfo
01Checking FlashSystem startup; flash memory verification phaseinfo
02Initializing processorSystem startup; processor initialization phaseinfo
02Initializing processorSystem power-up processor initialization phaseinfo
03Clearing Memory (BSS1)System power-up memory clearing phaseinfo
03Clearing Memory (BSS1)System startup; BSS1 memory clearing phaseinfo
04Clearing Memory (BSS2)System startup; BSS2 memory clearing phaseinfo
04Clearing Memory (BSS2)System power-up memory clearing phaseinfo
05OS initializingSystem power-up operating system initializationinfo
05OS initializingSystem startup; operating system initializationinfo
06Application Initialization beginningSystem startup; application initialization phaseinfo
06Application Initialization beginningSystem power-up application initialization phaseinfo
07Architecture utilities initializingSystem power-up architecture utility initializationinfo
07Architecture utilities initializingSystem startup; architecture utility initializationinfo
08Network InitializationSystem startup; network interface initializationinfo
08Network InitializationSystem power-up network interface initializationinfo
09Timers being initializedSystem power-up timer initialization phaseinfo
09Timers being initializedSystem startup; timer initialization phaseinfo
0ATasks InitializingSystem startup; task initialization phaseinfo
0ATasks InitializingSystem power-up task scheduler initializationinfo
0DSystem halted - out of memoryFluoro Functions PCB runs out of available memorycritical
0DSystem halted: Out of MemoryDynamic memory allocation failure; insufficient RAM availablecritical
0x0000Diagnostic test passed successfullyinfo
0x2251MCU Motion Status Register 1: Lift Down request failedDiagnostic test during RUI's Lift Down key testerror
0x2251MCU Err Stat1 Lift Down Request FailedRUI Lift Down key not responding during testerror
0x2252MCU Err Stat1 Lift Up Request FailedRUI Lift Up key not responding during testerror
0x2252MCU Motion Status Register 1: Lift Up request failedDiagnostic test during RUI's Lift Up key testerror
0x2255MCU Motion Status Register 1: Collision Ready failedDiagnostic test; triggered when bumper switch unpluggederror
0x2255MCU Err Stat1 Collision Ready FailedContact detection module or bumper switch not detectederror
0x2256MCU Err Stat1 Rotational Fault FailedRotational servo abnormal response or collisionerror
0x2256MCU Motion Status Register 1/2: Rotational fault or RUI faultCollision or RUI communication issueerror
0x2257MCU Motion Status Register 1: Orbital faultDiagnostic test; occurs with actual or false collisionerror
0x2257MCU Err Stat 1 Orbital Fault FailedOrbital servo abnormal response or collisionerror
0x2258MCU Err Stat 1 Collision FailedBumper contact detected with abnormal statuserror
0x2258MCU Motion Status Register 1: Collision detection failedDiagnostic test; triggered when bumper pressederror
0x225DMCU Motion Status Register 2: Clutch sense failedDiagnostic test when clutch is held openerror
0x225DMCU Err Stat2 Clutch Sense FailedClutch position sensor not detecting open/closed stateerror
0x225EMCU Err Stat2 Key Switch FailedC-arm key switch not responding during testerror
0x225EMCU Motion Status Register 2: Key switch failedDiagnostic test when keyswitch set to Standbyerror
0x2260MCU Motion Status Register 2: RUI sense failedDiagnostic test when RUI unpluggederror
0x2260MCU Err Stat2 RUI Sense FailedRemote unit interface not detectederror
0x2261MCU Err JS Motion Enable Test FailedJoystick motion control signal not at correct levelerror
0x2261MCU Joystick Motion Enable test failedDiagnostic test during joystick grasperror
0x2263MCU Contact Sense test failedDiagnostic test during boot; triggered by bumper presserror
0x2263MCU Err Contact Sense FailedContact bumper assembly not detectederror
0x226AMCU Err RUI Sense Test FailedRemote unit interface not detected during booterror
0x226AMCU RUI sense test failedSystem does not see RUI during bootuperror
0x226BJoystick rest position incorrect in X axisJoystick X axis position too high at resterror
0x226BMCU Err JS X Pos Too HighJoystick X-axis rest position calibration higherror
0x226CMCU Err JS X Pos Too LowJoystick X-axis rest position calibration lowerror
0x226CJoystick rest position incorrect in X axisJoystick X axis position too low at resterror
0x226DJoystick rest position incorrect in Y axisJoystick Y axis position too high at resterror
0x226DMCU Err JS Y Pos Too HighJoystick Y-axis rest position calibration higherror
0x226EMCU Err JS Y Pos Too LowJoystick Y-axis rest position calibration lowerror
0x226EJoystick rest position incorrect in Y axisJoystick Y axis position too low at resterror
0x226FOrbital servo checksum errorProblem with Logosol orbital servo control module during bootuperror
0x226FMCU Err LGS Check Sum ErrorLogosol servo control checksum validation failederror
0x2270MCU Err LGS Position ErrorServo motion failed to reach expected positionerror
0x2270Orbital servo position errorMotion stop button pressed or servo position error detectederror
0x2271Orbital servo overrun errorServo feedback indicates overrun condition detectederror
0x2271MCU Err LGS Servo OverrunServo exceeded travel limits or response timeerror
0x2272MCU Err Motion MS Sw FailedMotion stop switch not detectederror
0x2272Motion stop switch failedRUI motion stop button failure detected during bootuperror
0x2273MCU NVRAM CRC test failedConfiguration data stored in NVRAM is corruptedcritical
0x2273MCU Err NVRAM Test FailedSystem configuration data CRC check failedcritical
0x2274MCU Err Diag Switch FailedDiagnostic switches S2/S3 not at zero during normal booterror
0x2274MCU diagnostic self-test failedDiagnostic switches S2 or S3 not set to 0 at bootuperror
0x2275Frame sync test failedFrame sync signal not generated or signal path openerror
0x2275MCU Err Frame Sync Test FailedFrame sync signal not generated or signal path openerror
0x2277MCU Err Clutch Sense FailedClutch position sensor signal not at correct stateerror
0x2277MCU error: clutch sense failedMCU does not detect clutch position sensor signalerror
0x2280Lift motor movement test failedLift motor does not respond during bootup self-testerror
0x2280MCU Err Lift Motor Movement Test FailedLift motor does not respond to lift commandserror
0x2281MCU Err Key Disable FailedMCU cannot sense key disable line signalerror
0x2281MCU error: key disable failedMCU does not properly sense keyswitch positionerror
24-HOUR RECHARGE REQUIREDC-Arm batteries depleted and require 24-hour recharge cycleBatteries discharged after six months or more of storage without powerwarning
24 HOUR RECHARGE REQUIRED — X-RAYS DISABLEDSystem battery has discharged from storage > 6 months; recharge requiredSystem bootup after extended storage period; battery voltage below operational thresholdwarning
24 HOUR RECHARGE REQUIRED - X-RAYS DISABLEDBattery requires full 24-hour recharge cycleInverter B battery voltage low after 24-hour cycleerror
24 HOUR RECHARGE REQUIRED – X-RAYS DISABLEDSystem batteries require recharging after extended storageSystem in storage for more than six monthswarning
24 HOUR RECHARGE REQUIRED – X-RAYS DISABLEDSystem has been in storage for more than six months, batteries need rechargingSystem storage exceeding six monthswarning
4-HOUR WARM-UP REQUIREDImage Intensifier requires warm-up period for ion removalSystem inactive for more than 60 dayswarning
4 HOUR WARM-UP REQUIRED — PRESS ANY KEYSystem requires 4 hour warm-up period for ion removal from Image IntensifierSystem bootup after extended storage > 60 daysinfo
4 HOUR WARM-UP REQUIRED – PRESS ANY KEYSystem has been stored for more than 60 days, warm-up period necessary for ion removal from Image Intensifier tubeSystem storage exceeding 60 daysinfo
AC LINE CONFIGURATION DOES NOT MATCH MEASURED VOLTAGEIntended AC line operating range does not match measured input voltageSystem detects AC line voltage outside configured range during startup or operationcritical
AC LINE VOLTAGE NOT CONFIGUREDIntended operating AC line range has not been downloaded from UtilitySuiteFresh software installation without AC line parameters loadederror
AC POWER LINE SENSOR ERROR DETECTEDWorkstation detects invalid AC line voltage and/or current readingsAC voltage or current readings outside acceptable range: <50 VAC, <2A/1A @120V/200V, >27A/13.5A @120V/200Vcritical
AD CHANNEL 1s FAIL - POWER OFF - WAIT 5 SECONDSAnalog-to-digital channel failureA/D conversion error on specified channelerror
AD CHANNEL – FAIL-POWER OFF. WAIT 10 SECONDS.Fault detected in analog-to-digital data-conversion process in one or more channelsA/D channel fault detected during system startuperror
AD CHANNEL – FAIL-POWER OFF. WAIT 10 SECONDS.Analog-to-digital channel fault detectedA/D channel checked during system startup; fault in data-conversion process detectederror
AD CHANNEL # FAIL-POWER OFF. WAIT 10 SECONDS.Analog-to-digital conversion fault detected during startup; system inoperableA/D channel self-test at bootup; data conversion error detected in one or more channelscritical
AD CHANNEL FAILUREAnalog-to-digital conversion failure in charger monitoring circuitDetected at system boot when LINE-V-SEN signal reads outside acceptable rangeerror
AD CHANNEL %s FAILAnalog-to-digital converter channel failed; channel number inserted in %sAnalog-to-digital channel failure detected during system startupcritical
A DIFFERENT C-ARM HAS BEEN CONNECTED TO THE WORKSTATIONSerial number mismatch detected between Workstation and C-arm during bootSystem bootup detects C-arm serial number does not match expected C-arm for this Workstationwarning
ALL EXISTING ANNOTATION AND MEASUREMENTS WILL BE REMOVED FROUser attempted to zoom on annotated image; annotations will be lostUser tries to zoom image with text or measurement annotationswarning
ALL EXISTING ANNOTATION AND MEASUREMENTS WILL BE REMOVED FROUser initiated image zoom operation on an image with text annotations or measurement overlaysUser attempts to zoom in on annotated or measured imagewarning
ALL PATIENT HEALTH INFORMATION AND SAVED IMAGES WILL BE DELEUser invoked Delete All Patient Health Information option on Customize Security ScreenUser selects delete all patient data function in security configuration menuwarning
ALL PATIENT HEALTH INFORMATION AND SAVED IMAGES WILL BE DELEConfirmation prompt for deletion of all patient dataUser invokes Delete All Patient Health Information on security screenwarning
ANATOMICAL PROFILES DISABLEDSome anatomical profiles unavailable; serial number mismatch or invalid profile formatSystem detects unavailable profiles (first case) or invalid profiles (second case)warning
ANODE IS HOTX-ray tube anode at 80% or greater of rated thermal capacityAnode temperature reaches 80% (±2%) of rated capacity during operationcritical
ANODE IS HOTX-ray tube anode temperature criticalAnode heat capacity exceeded during operationerror
ANODE IS HOT — HLF DISABLEDX-ray tube anode temperature at 80% or greater of rated heat capacityAnode temperature monitoring; threshold reached during operationwarning
ANODE IS HOT — XX%X-ray tube anode temperature at 80% or greater of rated capacityAnode temperature monitoring during C-arm operation; threshold exceededwarning
ANODE IS HOT – XX°Anode temperature is at 80% of its rated heat capacity or greaterX-ray tube anode reaches 80% of rated heat capacitywarning
ANODE IS WARMX-ray tube anode temperature elevatedAnode approaching thermal limits during operationwarning
ANODE IS WARMX-ray tube anode temperature at 70% or greater of rated heat capacityAnode temperature monitoring; warning threshold reachedinfo
ANODE IS WARM — XX%X-ray tube anode temperature at 70-80% of rated heat capacityAnode temperature monitoring; warning threshold reached during operationinfo
ANODE IS WARM – XX°Anode temperature is at 70–80% of its rated heat capacityX-ray tube anode reaches 70–80% of rated heat capacityinfo
ARMEDSystem is preparing the technique for a film exposureFilm exposure preparation initiatedinfo
ARMEDSystem preparing technique for film exposure; ready for X-ray switch pressFilm exposure mode selected and technique parameters loadedinfo
ARMEDX-ray system prepared and ready for exposure; filament heated and awaiting X-ray triggerFilament preparation completes without X-ray control being pressedinfo
BATTERY CHARGEC-Arm batteries at 40-70% charge level; normal operation continuesBattery charge drops to 40-70% of required level during operation or standbywarning
BATTERY CHARGE - ___Battery charge level critically lowBattery capacity below minimum operating thresholdwarning
Battery charger light off with Workstation offAC power distribution to battery charging circuit is not functioningBattery Charger light fails to illuminate when system is powered but Workstation is offerror
BATTERY CHARGE /WAITC-Arm batteries critically low at below 40% charge; Film and HLF exposures disabledBattery charge drops below 40% of required levelerror
BATTERY CHARGE — XX%Effective battery charge dropped to between 70 and 40 percentBattery monitoring during operation; charge level crossed warning thresholdwarning
BATTERY CHARGE – XX%Effective battery charge has dropped to between 70 and 40 percentBattery charge level drops to 70–40 percentwarning
BUMPER ATTACHEDCollision Detection Assembly has been physically attached to Image IntensifierSystem detects Collision Detection Assembly attachment during operationinfo
BUMPER REMOVEDCollision Detection Assembly has been removed from Image IntensifierSystem detects Collision Detection Assembly removal during operationwarning
CAMERA IRIS FAILUREVideo system iris aperture malfunctionIris position feedback error detectederror
CANCEL COPY OPERATION?Confirmation prompt for canceling copy operation in progressUser presses Cancel button on Copy in Progress pop-upinfo
CANNOT MAKE NETWORK CONNECTIONWorkstation unable to initialize external network interface during bootupNetwork interface initialization fails during system startuperror
CANNOT WRITE TO THE DISK IN THE REMOVABLE MEDIA DEVICEError occurred while writing to a CD or DVD diskWrite operation failed on removable mediaerror
CANNOT WRITE TO THE DISK IN THE REMOVABLE MEDIA DEVICEError occurred while writing to CD or DVD diskWrite operation to removable media failederror
C-Arm completely inoperative with battery charger light onC-Arm unable to operate despite AC power reaching battery charger circuitNo error messages displayed but unit is completely inoperative with battery charger light oncritical
CD/DVD NOT RESPONDINGCD or DVD device is not responding to commands from the systemSystem commands to CD/DVD drive receive no responseerror
CHARGER FAILEDBattery charger failure detected during operationCharger hardware self-test or communication failure during normal operationerror
CHARGER FAILEDBattery charger has failed or is not operatingDetected during system boot if charger is not functioning properlyerror
CHARGER FAILED – POWER OFF. WAIT 10 SECONDS.Battery charger failure detected at system bootup; system inoperableCharger self-test at startup failederror
CHARGER FAILED – POWER OFF. WAIT 10 SECONDS.Battery charger failed at boot-up, system will not operateBattery charger failure detected during system startuperror
CHARGER FAILED - POWER OFF - WAIT 5 SECONDSBattery charging circuit critical failurePersistent charger voltage or current errorerror
Check for error messagesGeneric error message notification during Image Rotation or Invert operationRS-422 communication failure, camera cabling disconnection, or Fluoro Functions PCB not seatedwarning
CLUTCH IS OPENC-arm clutch is released; C-arm free to orbit manuallyClutch engagement switch reads open during operation or startupwarning
CLUTCH IS OPENClutch is released and the C-Arm is free to orbitClutch handle held to the rear or clutch failurewarning
COL IRIS POTENTIOMETER ERROR – PRESS ANY KEYCollimator iris potentiometer is not working, iris position cannot be determinedCollimator iris potentiometer failure detectederror
COL IRIS POTENTIOMETER ERROR – PRESS ANY KEYCollimator iris potentiometer not functional; iris position cannot be determinedSoftware unable to read iris position sense signal during startup/operationerror
COL IRIS TOO LARGECollimator iris diameter larger than control panel software indicatesIris position feedback exceeds maximum expected size during operationerror
COL IRIS TOO LARGECollimator iris is larger than the control panel indicatesIris opening exceeds expected positionwarning
COL IRIS UNSTABLEIris collimator motor control cannot maintain position toleranceIris motor unable to hold position within tolerancewarning
COL IRIS UNSTABLECollimator iris motor control cannot maintain stable position toleranceMotor feedback shows oscillation or inability to hold iris positionwarning
COLLIMATOR CAL REQUIREDCollimator iris and leaf position calibration mismatch; stored reference values do not match current positionsIris and leaf position values detected at startup do not match calibration reference values stored in system memorycritical
COLLIMATOR CAL REQUIRED — PRESS ANY KEYCollimator iris or leaf position values do not match reference values in memorySystem detects mismatch during calibration verificationwarning
COLLIMATOR CAL REQUIRED – PRESS ANY KEYCurrent collimator iris or leaf position values do not match reference values in memoryCollimator calibration drift detected at startupwarning
COLLIMATOR CAL REQUIRED - PRESS ANY KEYCollimator system calibration requiredCollimator leaf or iris position feedback errorerror
COLLIMATOR STUCKCollimator leaves or iris mechanically stuckPotentiometer reading indicates position not changing or out of rangeerror
COLLIMATOR STUCKCollimator iris motor is unable to open or close the collimator irisCollimator iris motor unable to move iriserror
COLLIMATOR STUCKCollimator iris motor unable to open or close iris apertureIris motor command issued but motor cannot move iris through full range or reaches limiterror
COLL IRIS POTENTIOMETER ERRORCollimator iris position potentiometer not functioning or disconnectedPosition potentiometer signal not detected or potentiometer mechanically disconnected during initializationcritical
COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEYCollimator iris potentiometer reading invalidPotentiometer A/D conversion error or mechanical failureerror
COLL IRIS TOO LARGECollimator iris opening exceeds maximumIris potentiometer reading indicates oversized openingerror
COLL IRIS TOO LARGECollimator iris aperture opening larger than position sensor indicates on control panelIris position sensor reading and actual iris position do not correspond; actual opening exceeds indicated sizewarning
COLL IRIS UNSTABLECollimator iris motor exhibits instability or jitter; cannot maintain steady positionIris motor control detects position oscillation or instability during operationwarning
COLL IRIS UNSTABLECollimator iris position unstableIris potentiometer reading erratic or out of rangeerror
COLLISION DETECTEDLoss of C-arm velocity or increased motor current detected during motorized motionMotor current monitoring or velocity feedback during C-arm orbital/rotational motionwarning
COMMUNICATION FAILEDLoss of communication with Generator detected by Workstation during exposureGenerator communication lost during active X-ray exposurecritical
COMMUNICATION FAILEDSystem communication between major components failedLoss of communication on system bus or between boardscritical
COMMUNICATION FAILEDWorkstation and C-arm communication loss for > 1 second; watchdog timeoutCommunication gap exceeds 1 second between workstation and C-armwarning
COMMUNICATION FAILEDSystem communication has failed, C-Arm cannot take X-raysSystem communication loss detectederror
CONTACT DETECTEDObject has contacted Image Intensifier Contact Detection AssemblyContact Detection Assembly physical sensor triggered during C-arm motionwarning
CONTROL PANEL ERRORSystem cannot communicate with C-arm Control Panels; X-rays disabledControl Panel Processor communication timeout during system initializationcritical
CONTROL PANEL ERRORControl panel display or input malfunctionPanel communication or hardware errorerror
CONTROL PANEL ERRORSystem cannot communicate with C-Arm Control Panels, X-rays are disabledControl Panel communication failureerror
CONTROL PANEL KEY STUCKC-arm control panel key stuck on at bootupKey closure detected on C-arm control panel during boot-upwarning
CORRUPTED IMAGES DELETEDCorrupted images detected at bootup and deleted due to inability to repairSystem finds corrupted image files during startup and marks them for deletionwarning
CURRENT ANATOMICAL PROFILE DISABLEDThe active/default anatomical profile has become corrupted or unavailableSystem attempts to load default profile and finds it corrupted or incompatibleerror
DATA ERRORSystem detects invalid data or corruptionInvalid ABS table write/read, corrupted non-volatile memory, software errorcritical
DATA ERRORCorrupted software or data detected; X-rays disabledChecksum failure on firmware or calibration data during startupcritical
DATA ERRORSystem memory or data integrity errorFlash corruption, workstation mismatch, or initialization failureerror
DATA ERRORCorrupted software or data has been detected, X-rays are disabledSoftware or data corruption detectederror
DELETE ALL COMPLETEAll patient health information and saved images have been successfully deletedDeletion process completed successfullyinfo
DELETE ALL IN PROGRESSSystem is erasing all patient health information as user requestedUser confirmed deletion of all patient data and system began erasure processinfo
DICOM ERRORDICOM device returned an error in response to a requestDICOM device reports error; configuration mismatch such as unsupported format layouterror
DICOM SECONDARY CAPTURE CINE NOT SUPPORTEDDICOM store server configuration does not support CINE modality for secondary captureUser attempts to store CINE run to DICOM server configured without CINE supportwarning
DICOM SERVER IS OUT OF RESOURCES — TRY AGAIN LATERDICOM printer or server is busy or out of resourcesDICOM device reports out-of-resources condition during queue submissionwarning
DOSE CAL REQUIREDDose Area Product not calibrated at startupGenerator detects missing DAP calibration during bootupwarning
E0ARCNET transmit pipe temporarily fullNetwork transmission buffer reaches capacitywarning
E0ARCNET transmit pipe temporarily fullARCNET network interface transmit buffer overflow during data transmissionwarning
E4Stack Memory dangerously lowStack memory usage approaching critical limits during operationerror
E4Stack memory dangerously lowStack memory usage exceeds safe thresholdwarning
ENTITY ALREADY EXISTSDICOM retrieve operation encountered a file that already exists on the systemDICOM retrieve attempts to retrieve an entity with same name/ID as existing filewarning
ERROR READING DISKError occurred while reading CD or DVD, or no media is in the driveSystem attempts to read CD/DVD and encounters read error or no media detectederror
ERROR RETRIEVING IMAGESPACS server has incorrect Workstation information for image retrievalPACS server attempts to send images to Workstation but configuration is incorrecterror
ERROR WHILE PROCESSING DICOM FILESystem encountered error accessing necessary data from removable-media DICOM fileSystem attempts to read DICOM data from CD or DVD and failserror
FAILED TO OPEN PORTCould not open firewall port; port number already in useUser attempts to configure firewall port that is already allocated to another serviceerror
FAILED TO READ IMAGE FILESystem attempted to read DICOM image from CD/DVD and failedSystem tries to load DICOM image file from removable mediaerror
FAST STOP ACTIVATEDInterlocks open; typically FAST STOP switch pressedFAST STOP switch activation or 24V interlock loop breakerror
FAST STOP ACTIVATEDFast Stop button has been pressed or interlock circuit fault detectedFast Stop switch opened or interlock circuit failure during operation or bootupwarning
FAST STOP ACTIVATEDInterlock circuit broken; X-ray emission preventedFast Stop switch opened during operation or interlock circuit fault detectederror
FAST STOP ACTIVATED — POWER OFF. WAIT 10 SECONDSInterlocks are open due to FAST STOP switch activation; X-ray generation disabledFAST STOP pushbutton pressed or safety interlock circuit openedwarning
FAST STOP ACTIVATED – POWER OFF. WAIT 10 SECONDS.Interlocks are open, typically from pressing FAST STOP switchFAST STOP switch pressed or interlock loop brokenerror
FAST STOP ACTIVATED - POWER OFF. WAIT 5 SECONDSEmergency stop activated during operationFast stop button pressed or safety condition detectederror
FEFatal Error: System haltedStack overflow, no memory available, or unhandled interrupt during operationcritical
FEFatal Error - system haltedStack overflow, no memory, or unhandled interrupt conditioncritical
FFSystem halted - out of memoryFluoro Functions PCB runs out of available memorycritical
FFSystem halted: Out of MemoryDynamic memory allocation failure; insufficient RAM availablecritical
FILAMENT CAL REQUIREDX-ray tube filament calibration neededFilament performance drift detectederror
FILAMENT CAL REQUIREDFilament calibration not performed at startupSoftware detects missing filament calibration during bootupwarning
FILAMENT CAL REQUIRED — PRESS ANY KEYFilament calibration data missing; generator disabled pending recalibrationFilament calibration flag not set in non-volatile memory during power-upwarning
FILAMENT CAL REQUIRED – PRESS ANY KEYFilaments have not been calibrated, multiple mA errors may occurFilament calibration not performed at startuperror
FILAMENT REGULATOR FAILEDX-ray tube filament voltage regulator malfunctionFilament voltage out of specificationerror
FILAMENT REGULATOR FAILUREX-ray tube filament current out of toleranceFilament current measurement outside specification rangewarning
FILAMENT REGULATOR FAILURE — PRESS ANY KEYFilament current measured out of tolerance; image quality degradation riskmA feedback signal exceeds tolerance window during operationerror
FILAMENT REGULATOR FAILURE – PRESS ANY KEYFilament current is out of tolerance, future image resolution may be degradedFilament current exceeds tolerance limitswarning
FILAMENT SELECT ERRORSoftware filament selection does not match PIO hardware filamentFilament size mismatch between software and hardware detectedwarning
FILAMENT SELECT ERROR — PRESS ANY KEYFilament size mismatch between software and PIO hardware; X-rays disabledFilament select output does not match expected PIO feedback during power-uperror
FILAMENT SELECT ERROR — PRESS ANY KEYSoftware filament selection does not match PIO hardwareMismatch between software configuration and installed filamenterror
FILAMENT SELECT ERROR – PRESS ANY KEYFilament size selected by software does not match current filament in PIO hardwareFilament selection mismatch between software and hardwareerror
FILAMENT SELECT ERROR - PRESS ANY KEYX-ray tube filament selection mismatchDetected filament (small/large) does not match expected selectionerror
FILAMENT STANDBYX-ray filament in standby modeFilament is preheating before exposureinfo
FOOTSWITCH SECURITY ERRORX-ray security line does not match X-ray controlFootswitch security circuit mismatch detectederror
HIGH-CAPACITY DISK FAILURECD/DVD drive has failed or is not properly installed/connectedCD/DVD drive was not detected or connected during bootuperror
HLF OVERTIMEHigh Level Fluoro exposure time exceeded preset limit; exposure terminatedDigital Cine or HLF run duration exceeds preset thresholdwarning
HLF OVERTIMEDigital Cine run or High Level Fluoro has exceeded preset time and been terminatedHLF or Digital Cine run duration exceeds preset time limitwarning
HLF OVERTIMEHigh-level fluoroscopy overtime operationContinuous fluoroscopy mode time exceeded limitsinfo
HLF OVERTIME (digital Cine)High-level fluoroscopy overtime in digital cine modeCine mode operation time limit exceededinfo
HOUSING IS HOT - ___Generator housing temperature criticalHousing temperature capacity approaching 100%error
HOUSING IS HOTX-ray tube thermal load at 80-99% of capacityHousing temperature reaches 80% or more of rated capacitywarning
HOUSING IS HOT — HLF DISABLEDX-ray tube housing temperature at 80% or greater of rated heat capacityHousing temperature monitoring; threshold reached during operationwarning
HOUSING IS HOT — XX%X-ray tube housing temperature at 80% or greater of rated capacityThermistor reading exceeds 80% threshold during normal operationwarning
HOUSING IS HOT – XX°Housing temperature is at 80% of its rated heat capacity or greaterHousing temperature reaches 80% of rated capacitywarning
HOUSING IS WARMX-ray tube housing temperature at 70% or greater of rated heat capacityHousing temperature monitoring; warning threshold reachedinfo
HOUSING OVERHEATEDX-ray tube at 100%+ of rated thermal capacity; thermal cutoff activatedHousing temperature exceeds 100% of rated capacity during or after exposurecritical
HOUSING OVERHEATEDHousing temperature has reached 100% of rated capacity and thermal cutoff switch has activatedHousing temperature reaches 100% of rated capacitycritical
HOUSING OVERHEATEDGenerator housing maximum temperature exceededHousing temperature exceeds safe limitcritical
HOUSING OVERHEATED — X-RAYS DISABLEDHousing temperature reached 100% of rated capacity; thermal cutoff switch activatedHousing temperature threshold exceeded; thermal safety interlock engagedcritical
HOUSING WARMX-ray tube thermal load at 70% (±2%) of capacityHousing temperature reaches 70% of rated capacity during operationinfo
HOUSING WARM - ___Generator housing temperature elevatedHousing temperature at moderate levelwarning
HOUSING WARM – XX°Housing temperature is at 70–80% of its rated heat capacityHousing temperature reaches 70–80% of rated capacityinfo
HOUSING WARM — XX%X-ray tube housing temperature at 70–80% of rated heat capacityThermistor reading in 70–80% range during normal operationinfo
HV GENERATOR ERRORSoftware detected error in high-voltage generatorHV generator status monitoring detects fault conditioncritical
HV GENERATOR ERRORSoftware has detected an error in the high-voltage generatorHigh-voltage generator fault detectederror
HV GENERATOR ERRORHigh-voltage generator software error detected; system shutdown initiatedHV regulation loop error or generator driver fault detection during operationcritical
HV GENERATOR ONHigh-voltage generator remains energizedHigh voltage input register shows active state when not expectederror
HV REGISTER FAILHigh-voltage control register failureRegister address or reading corruptederror
HV REGISTER FAILSoftware has detected a failure in the high-voltage registerHigh-voltage register failure detectederror
HV REGISTER FAILHigh-voltage register failure detected; system shutdown initiatedHV register read verification failure during operation or startupcritical
HV REGISTER FAILSoftware detected failure in high-voltage registerHV register status check fails or communication errorcritical
HV REGISTER FAILHigh voltage generator register error at startupHV register malfunction detected during system power-upcritical
IMAGE TOO BIG TO COPYSize of selected Cine file exceeds DICOM standard limit of 4.29 GBUser attempts to copy Cine file to DICOM device that exceeds 4.29 GB standard limitwarning
INSUFFICIENT MEDIA CAPACITYSelected image or Cine run will not fit on available mediaUser selects file(s) to copy to CD/DVD that exceed remaining free space on mediawarning
INTERLOCK FAILUREInterlock circuit failed during system startup; operation preventedInterlock verification test fails at power-upcritical
INTERLOCK FAILUREThe interlock circuit has failed during system start-upInterlock circuit failure detected at startuperror
INTERLOCK FAILURESystem interlock failed to operate correctlyInterlock failed to close, already closed, or IO task missingerror
INVALID IPIP address entered is not valid or properly formattedUser enters IP address in configuration that fails validationerror
INVALID PORT ENTRYPort number for DICOM Query/Retrieve server is outside valid 1–65536 rangeUser attempts to configure DICOM Query/Retrieve port with invalid numbererror
INVALID PORT ENTRYPort number must be 104 or 1025–49150 for DICOM local serverDICOM local server port configurationerror
JOYSTICK FAILURESystem could not communicate with joystick controllerJoystick fails to respond during system bootup or operationerror
JOYSTICK STUCKJoystick reported initialization error; may have physical obstructionJoystick initialization fails due to detected stuck position or obstructionerror
JOYSTICK STUCKJoystick initialization error detected; reported on monitors and control panelJoystick self-test fails at power-up or during initializationerror
JOYSTICK STUCKJoystick initialization error detectedDuring system boot-up, joystick reports initialization errorwarning
KEY STUCKControl panel key stuck during boot-upSystem detects key press on control panel during startupwarning
KEY STUCKKey closure detected on control panel during bootupControl panel key held closed during system startupwarning
KEY STUCK — RELEASE, and then PRESS ANY KEY TO CONTINUEControl panel key detected stuck during bootup; waiting for key releaseContinuous key signal detected on at least one control panel key during startupwarning
KEY STUCK - RELEASE, THEN PRESS ANY KEY TO CONTINUEOperator control key stuck in pressed positionKey input continuously registered as pressederror
KEY SWITCH IN STANDBYX-ray key switch in Standby positionX-ray key switch detected in Standby rather than On positionwarning
KEY SWITCH IN STANDBYX-ray disable keyswitch in ON position; X-rays and lift disabledX-ray disable keyswitch set to ON (key handle vertical)warning
KEY SWITCH IN STANDBY — X-RAYS AND LIFT DISABLEDX-ray key switch in Standby position; X-rays and column lift motors disabledX-ray key switch scanned in Standby (Off) positioninfo
KV ON IN ERRORX-ray tube current on but no X-ray switch pressedTube voltage/current active without intentional X-ray control activationcritical
KV ON IN ERRORHigh voltage being generated without X-ray switch activationSystem detects kV generation without X-ray switch activatedcritical
KV ON IN ERROR- POWER OFF. WAIT 10 SECONDS.High voltage generated without X-ray switch activationHV detected without X-ray control signalcritical
KV ON IN ERROR— POWER OFF. WAIT 10 SECONDSHigh voltage being generated without X-ray switch activation; generator fault detectedkV present on output without corresponding X-ray enable commandcritical
kV ON IN ERROR - POWER OFF, WAIT 5 SECONDSHigh voltage energized when it should be offUnexpected high-voltage output detectederror
kVp ErrorKilovoltage error; tube voltage out of specificationTriggered during exposure if measured tube voltage deviates from commanded settingwarning
kVp ErrorKilovoltage peak setting does not match commanded valueDuring exposure when actual kVp deviates from set valueerror
LIFT SWITCH STUCKC-arm lift switch held in ON condition for 30+ secondsLift switch activation detected continuously for 30 seconds or morewarning
LIFT SWITCH STUCKLift switch held for 30+ secondsLift switch continuously pressed for 30 secondswarning
LIFT SWITCH STUCK - PRESS ANY KEYTube lift mechanism switch mechanically stuckLift switch input continuously registerederror
LIFT SWITCH STUCK — PRESS ANY KEYColumn lift switch pressed continuously for 30 seconds; motor control failure riskLift switch input signal continuous for 30 seconds during operationwarning
LINE POWER WAS LOSTAC line power was lost with NAV application not in focusAC power disconnected while system running without NAV application activecritical
LOADING LOW DOSE ABSLoading Low Dose ABS table as current ABS tableUser selects Low Dose ABS table in generator configurationinfo
LOADING STANDARD ABSLoading Standard ABS table as current ABS tableUser selects Standard ABS table in generator configurationinfo
LOGIN FAILEDUser entered an invalid password at the LOGON screenUser authentication fails due to incorrect passwordwarning
LOGOUT IN PROGRESSUser confirmed logout and system prompts for final confirmationUser selects logout option; system displays confirmation dialoginfo
LOST CONNECTION TO THE DICOM SERVERDICOM server ended connection unexpectedly during retrieve operationServer closes connection before retrieve operation completeserror
mA ErrorTube current (tube output) out of specificationTriggered during exposure if measured tube current deviates from commanded settingwarning
mA ErrorTube current (milliamperage) does not match commanded valueDuring exposure when actual mA deviates from set valueerror
MA ON IN ERRORmA Sense on but no X-ray switch pressedmA Sense circuit active without X-ray control activationcritical
MA ON IN ERRORX-ray tube current detected without switch activationSystem detects mA without X-ray switch activatedcritical
MA ON IN ERROR — POWER OFF. WAIT 10 SECONDSX-ray tube current detected without X-ray switch activation; generator fault indicatedmA present on tube without corresponding X-ray enable commandcritical
mA ON IN ERROR - POWER OFF, WAIT 5 SECONDSTube current energized when disabledUnexpected tube current output detectederror
MA SENSOR FAILmA sensor malfunction detectedDuring startup or operation, mA sensor test indicates defecterror
MA SENSOR FAILFailure detected in one of the mA sensor circuitsmA sensor measurement failure or circuit malfunction detectedwarning
MA SENSOR FAILmA sensor failure detected; current measurement unreliablemA sensor signal out of range or missing during operationerror
mA SENSOR FAILEDX-ray tube current sensor complete failureCurrent feedback permanently losterror
MEDIA IN DEVICE FOR COPY OPERATION NOT WRITABLEUser attempted to copy to CD or DVD that is not writableUser presses Copy button when write-protected or full media is in drivewarning
MOTION CAL REQUIREDMotorized drive motion calibration missing or invalid; positioning may be inaccurateMotion calibration flag not set or checksum error in motor calibration datawarning
MOTION CAL REQUIREDMotion subsystem has not been calibratedSystem detects missing motion calibration during startuperror
MOTION CAL REQUIREDMotorized drive motion not calibratedSystem startup detects uncalibrated motionwarning
MOTION DISABLEDMotorized C-arm motion disabled by motion stop switch or RUI disconnectionMotion stop engaged or RUI not properly connected during startup/operationwarning
MOTION ERRORTrace of uncommanded motion on orbital and rotational, or no motion detected when motion was commandedMotion control malfunction on orbital or rotational axeserror
MOTION FAILUREOrbital or rotational potentiometer open or short circuitPotentiometer electrical failureerror
NEW DISK NEEDEDCurrent target disk does not have enough space to complete copy operationInsufficient disk space during image copy operationwarning
NEW PASSWORD AND RETYPED PASSWORD DO NOT MATCHUser did not type the same password twice on Change Password screenPassword entry mismatch during password changeinfo
NEW PASSWORD AND RETYPED PASSWORD DO NOT MATCHPassword entries do not match during password changeUser password mismatch on Change Password screenwarning
NEW PASSWORD TOO SHORTUser entered a password of fewer than six charactersPassword length below minimum requirement during password changewarning
NO DISK IN DRIVENo media present in CD/DVD drive when user attempted Copy From or Copy To operationCopy operation attempted without disk in drivewarning
NO MATCH FOUNDNo DICOM images found that match the search criteriaDICOM database search returned no resultsinfo
NO MEDIA IN DEVICE FOR COPY OPERATIONUser attempted to read from or write to disk when no disk in driveCopy operation attempted without disk in CD/DVD drivewarning
NO MEDIA IN DEVICE FOR COPY OPERATIONNo disk in drive during read/write operationCopy operation attempted with empty drivewarning
NO RESPONSE — SERVER TIMED OUTNo response from DICOM server within allotted timeout periodDICOM server communication timeouterror
NOT ENOUGH DISK SPACE TO RETRIEVE IMAGESDICOM retrieve operation failed in the middle due to connection failure or insufficient disk spaceDisk space exhausted during DICOM retrieve operationerror
NOT ENOUGH SPACE ON THE CURRENT DISKCombined size of selected images is greater than space available on target diskImage copy operation size exceeds target disk capacitywarning
OLD PASSWORD INCORRECTUser made an incorrect entry in the Old Password field during password changeIncorrect old password entered during password change operationwarning
OVERLOAD FAULTFault detected in high-voltage regulator circuitHV regulator output current exceeds thresholderror
OVERLOAD FAULTHigh-voltage regulator overload condition detected; possible shutdown on second occurrenceExcessive current sensed in HV circuit; first time message appears briefly post-exposurewarning
OVERLOAD FAULTSystem electrical load exceeds maximum ratingHigh-voltage input register shows overload conditionerror
OVERLOAD FAULTHigh-voltage regulator overload detectedFirst occurrence: briefly after exposure. Second: auto-shutdown.error
OVERVOLTAGE FAULTSystem voltage exceeds safe operating limitHigh-voltage input register shows overvoltage stateerror
OVERVOLTAGE FAULTHigh-voltage regulator overvoltage detectedFirst: briefly after exposure. Second: auto-shutdown.error
OVERVOLTAGE FAULTHigh-voltage overvoltage condition detected; possible shutdown on second occurrenceHV output voltage exceeds regulation limit; first time message appears brieflywarning
OVERVOLTAGE FAULTX-ray generator overvoltage fault detectedHV output exceeds maximum setpointerror
PLEASE WAITSystem updating Generator data or performing internal tests; standby messageGenerator firmware update or self-test routine in progressinfo
PLEASE WAIT...System searching for DICOM-format files on mediaUser initiated search for DICOM-formatted filesinfo
PLEASE WAITSystem updating generator data or performing internal testsDuring normal boot or operation when generator updates or self-test runsinfo
PLEASE WAIT... Searching for DICOM-format files on mediaSystem searching for DICOM files on mediaUser initiated DICOM file searchinfo
PLEASE WAIT... Searching for mediaSystem searching for media during Copy To operationUser initiated Copy To operationinfo
PORT VERIFICATION SUCCESSFUL!Connection made to DICOM serverDICOM server port verification completed successfullyinfo
PRECHARGE CIRCUIT TIMEOUTCONTACT relay failed to close during startupCONTACT relay malfunction during system power-up sequencecritical
PRECHARGE CIRCUIT TIMEOUTContact relay failed to close during startupPrecharge relay K2 fails to close during boot sequencecritical
PRECHARGE CIRCUIT TIMEOUT — POWER OFF. WAIT 10 SECONDS.Contact relay failed to close during start-up prechargePrecharge relay does not close within expected timeframecritical
PRECHARGE CIRCUIT TIMEOUT - POWER OFF. WAIT 5 SECONDSBattery pre-charge initialization failed to completePre-charge timing sequence exceeded limiterror
PRECHARGE VOLTAGE ERRORPrecharge voltage too high during startupPrecharge voltage exceeds safe limit during bootcritical
PRECHARGE VOLTAGE ERRORPrecharge capacitor voltage exceeds 250V at startupCapacitor voltage exceeds 250V during power-up voltage checkerror
PRECHARGE VOLTAGE ERROR — POWER OFF. WAIT 10 SECONDS.Precharge voltage exceeds acceptable range during start-upCapacitor voltage too high before precharge completioncritical
PRE-CHARGE VOLTAGE ERROR - POWER OFF - WAIT 5 SECONDSBattery pre-charge voltage out of specificationInverter B battery voltage abnormal during pre-chargeerror
PRINTER ERRORPrinter reported a hardware error condition while printing an imagePrinter hardware error during print operationerror
PRINTER IS OUT OF FILM OR PAPERPrinter has exhausted film or paper supplyPrinter film or paper supply depleted during print operationwarning
PRINTER OFF-LINEPrinter is offline or not ready to printPrinter not in online/ready state during print operationwarning
RELEASED EARLYX-ray switch released before desired mAs reached; exposure incompleteOperator releases X-ray button before system terminates exposure automaticallyinfo
RELEASED EARLYX-ray switch released before desired mAs reachedOperator releases exposure switch before mAs accumulation completewarning
REMOTE CONTROL BATTERY IS LOWIR remote control battery charge is too low for accurate usabilityRemote control battery voltage drops below usable thresholdwarning
REQUESTED OPERATION FAILEDUnexpected error condition occurred such as DICOM Q/R database lock failure or media manager communication failureDICOM Q/R operation or media manager communication errorerror
ROOM DOOR OPENX-ray room door not fully closedDoor sensor detects X-ray room door in open positionerror
ROOM DOOR OPENX-ray room door is openX-ray door switch contact is openwarning
ROOM DOOR OPEN — CLOSE DOORX-ray room door is open; safety interlock activeDoor switch contact openwarning
RUI FAILURESystem cannot communicate with the remote user interfaceRUI communication error detectederror
RUI FAILURESystem could not communicate with remote user interfaceRUI disconnected or communication losterror
RUI KEY STUCKRemote User Interface key stuck or repeatingRUI sends same key signal multiple times during operationwarning
RUI KEY STUCKSystem receiving repeated signals from single keyPhysical button stuck or in contact with objectwarning
RUI KEY STUCKRUI keypad key stuck; same signal received multiple timesContinuous RUI key signal detectederror
RUI KEY STUCKSystem receiving repeated signals from a single keyRUI button stuck in pressed positionwarning
RUI KEY STUCKRemote User Interface key detected stuck; repeated signal from same RUI inputSame RUI signal received multiple times suggesting stuck keyerror
RUNNING CALIBRATIONRemote Utilities Tool (RUT) performing filament calibrationRUT software initiates filament calibration procedureinfo
RUNNING CALIBRATIONSystem is running calibration routine; informational message onlyAutomatic or user-initiated calibration routine activeinfo
RUNNING CALIBRATIONSystem is executing calibration procedureSystem initiates internal calibration routineinfo
SATURATION FAULTHigh-voltage regulator saturation faultDuring or after exposure, HV regulator saturation detectederror
SATURATION FAULTFault in high voltage regulator circuit; excessive IGBT currentExcessive current detected in IGBT power deviceserror
SATURATION FAULT — POWER OFF. WAIT 10 SECONDS.Fault in high-voltage regulator circuit; IGBT saturationHV regulator IGBT transistor driven into saturationcritical
SATURATION FAULT - POWER OFF. WAIT 5 SECONDSHigh-voltage regulation circuit saturatedHigh-voltage input register shows saturationerror
SAVE OPERATION FAILEDSystem failed to complete requested save operationUser pressed Save key to save image; save did not complete successfullyerror
SAVE OPERATION FAILEDSystem failed to complete the requested Save operationImage save operation completed unsuccessfullyerror
SAVE OPERATION FAILEDImage save to disk failed to completeUser pressed Save key; save operation incompleteerror
SCHEDULED EXAM LIST IS FULLScheduled exam list is full and cannot accept new patient informationNEW EXAM button pressed when scheduled exam list is fullwarning
SCHEDULED EXAM LIST IS FULLNew patient information will not appear in scheduled exam listUser touched NEW EXAM button when scheduled exam list was already fullwarning
Searching for external video source...User selected Endoscope button on Additional Apps screenUser selection of Endoscope functioninfo
Searching for external video source...System searching for external video source during Endoscope mode selectionUser selected Endoscope button on Additional Apps screeninfo
Search in ProgressDICOM server search in progress, results not yet receivedUser initiated search against DICOM serverinfo
Search in Progress - Please WaitDICOM server search in progress; result not immediately availableUser performs DICOM server searchinfo
Search in Progress / Please WaitSearch against DICOM server in progress; result not yet receivedUser performs search against DICOM serverinfo
SHUTDOWN IN PROGRESSSystem is shutting down; shutdown process cannot be stoppedSystem initiated normal or emergency shutdownwarning
SHUTDOWN IN PROGRESSSystem shutting down; shutdown process cannot be stoppedSystem initiated shutdown or UPS low battery shutdownwarning
SHUTDOWN IN PROGRESS (UPS)UPS battery is low; immediate shutdown in progressUPS battery voltage dropped below safe operating thresholdcritical
STATOR NOT ONX-ray tube anode not powered; anode rotation disabledAnode rotation power supply voltage loss or fault detectedcritical
STATOR NOT ONX-ray tube stator not rotatingSystem detects stator not energized during startupcritical
STATOR NOT ON — POWER OFF. WAIT 10 SECONDS.X-ray tube stator motor not running; anode cannot rotateStator motor fails to energize or runcritical
STATOR STANDBY WAITBattery capacity below 40 percentSystem detects low battery during operationwarning
SYSTEM ERRORExcessive or missing frame sync detected on GIB or FFB for more than ten seconds, or FFB has not recovered within 15 seconds following FFB rebootFrame synchronization error on GIB or FFBerror
SYSTEM ERROR DETECTEDCPU exception or runtime OS exception has occurred; X-rays disabledSystem software exception or fault during operationerror
SYSTEM ERROR DETECTED (CPU Exception)CPU exception or runtime OS exception has occurredCPU or operating system fault detectedcritical
SYSTEM ERROR DETECTED - CPU exceptionCPU or runtime OS exception occurred; X-rays disabledCPU or OS exception during operationcritical
SYSTEM ERROR DETECTED (CPU exception)X-rays are disabled; CPU or OS exception occurredCPU exception or runtime OS exceptioncritical
SYSTEM ERROR DETECTED (Frame Sync)Frame sync lost for more than five seconds or too many frame syncs detected within one minuteFrame synchronization loss detectedcritical
SYSTEM ERROR DETECTED - Frame sync lossFrame sync lost for >5 seconds or too many frame syncs in 1 minute; X-rays disabledExcessive frame sync anomalies detectedcritical
SYSTEM ERROR DETECTED (Frame sync lost)X-rays are disabled; frame sync lost or too many frame syncs detectedFrame sync lost for more than five seconds or too many frame syncs within one minutecritical
SYSTEM ERROR - Excessive or missing frame syncExcessive/missing frame sync on GIB/FFB >10s or FFB not recovered within 15s; X-rays disabledFrame sync anomaly sustained beyond recovery timeoutcritical
SYSTEM ERROR (FFB frame sync)Workstation fully functional but X-rays are disabled; excessive or missing frame sync on GIB or FFBExcessive or missing frame sync for more than ten seconds; FFB has not recovered within 15 secondserror
SYSTEM ERROR — X-RAYS DISABLEDGIB or FFB detected no frame syncs or too many frame syncsFrame sync monitoring during operation; sync fault detectederror
SYSTEM FAILED TO CONNECT TO DICOM SERVER (AE Title)System cannot connect to DICOM server due to wrong AE title configurationDICOM server connection attempt with incorrect AE titleerror
SYSTEM FAILED TO CONNECT TO DICOM SERVER (AE title)System could not establish connection to DICOM server due to wrong AE title or Query Information ModelSystem attempted to connect to DICOM server with incorrect AE title configurationerror
SYSTEM FAILED TO CONNECT TO DICOM SERVER (IP/port)System could not establish connection to DICOM server due to wrong IP address or port numberSystem attempted to connect to DICOM server with incorrect network parameterserror
SYSTEM FAILED TO CONNECT TO DICOM SERVER (IP/Port)System cannot connect to DICOM server due to wrong IP address or port number configurationDICOM server connection attempt with incorrect IP/porterror
SYSTEM FAILED TO CONNECT TO DICOM SERVER - Wrong AE titleCannot establish connection to DICOM server; wrong AE title or Query Model configuredDICOM connection attempt with incorrect AE titleerror
SYSTEM FAILED TO CONNECT TO DICOM SERVER - Wrong IPCannot establish connection to DICOM server; wrong IP address or port configuredDICOM connection attempt with incorrect IP/porterror
TEMP SENSOR FAILX-ray tube temperature sensor (thermister) defective; overheating warning disabledThermister out-of-range reading or continuity check fails at startuperror
TEMP SENSOR FAILX-ray tube temperature sensor (thermister) malfunctionThermister self-test during startup or operation indicates failureerror
THE C-Arm HAS BEEN RECONNECTED TO THE WORKSTATIONC-Arm was disconnected and reconnected without proper system shutdownUser disconnected/reconnected C-Arm without shutting down system power firstwarning
THE C-ARM HAS BEEN RECONNECTED TO THE WORKSTATIONC-arm was disconnected and reconnected to workstation without shutting down system firstC-arm physical reconnection detected without proper power shutdownwarning
THE CINE DISK IS NOT AVAILABLECine disk failed to initialize correctly during bootupCine disk initialization failure during system startuperror
THE CINE DISK IS NOT AVAILABLECine subsystem is not available for requested actionCine disk failed to initialize correctly during bootuperror
THE CURRENT CINE RUN WILL FILL THE CINE DISK IN APPROXIMATELThirty seconds of Cine disk storage remain; current cine run will fill disk when completeCine recording continues with minimal disk space remaininginfo
THE CURRENT CINE RUN WILL FILL THE CINE DISK IN APPROXIMATELThirty seconds of storage remain on Cine disk; current CINE RUN will fill the diskCine disk storage capacity approaching exhaustionwarning
THE CURRENT SESSION CINE RUNS WILL BE OVERWRITTEN IN 30 SECOCurrent session Cine runs will be overwritten in 30 seconds or less; oldest runs overwritten firstCine disk space fully consumed; new acquisitions will overwrite existing datawarning
THE CURRENT SESSION CINE RUNS WILL BE OVERWRITTEN IN 30 SECOThirty seconds or less of Cine acquisition time remains before current patient's Cine runs will be overwrittenCine disk capacity approaching overwrite threshold for current sessionwarning
THE LAST IMAGE IS BIGGER THAN THE CURRENT MEDIA CAPACITYUser selected a Cine run bigger than current media's capacityImage or cine run size exceeds media capacity during copy operationwarning
THE LAST IMAGE IS BIGGER THAN THE CURRENT MEDIA CAPACITYCopy operation will fail; selected Cine run is larger than current media capacityUser selected Cine run larger than available media capacitywarning
THE MEDIA HAS BEEN EJECTED PREMATURELYMedia ejected from device before being used by systemMedia physically removed from removable media device during operationwarning
THE MEDIA HAS BEEN EJECTED PREMATURELYMedia has been ejected from the device before being used by the systemMedia ejected during active read/write operationerror
THE SELECTED FUNCTION IS NOT AVAILABLE ON THIS MODELUser selected a Workstation function that is not available on this modelUser attempted to access model-restricted functioninfo
THE SELECTED FUNCTION IS NOT AVAILABLE ON THIS MODELSelected function not supported on this workstation modelUser selected unavailable feature for equipment modelinfo
THE SELECTED FUNCTION IS NOT AVAILABLE ON THIS MODELUser selected Workstation function not available on this modelUser attempted to access model-specific function not supported by this equipment variantinfo
THE SELECTED IMAGE CANNOT BE DISPLAYEDSelected image data is corrupt and could not be repairedUser selected thumbnail of corrupted image fileerror
THE SELECTED IMAGE CANNOT BE DISPLAYEDImage data is corrupt and could not be repairedCorrupted image thumbnail selected for displaywarning
THE SYSTEM WAS NOT SHUT DOWN PROPERLYSystem did not go through the Intelligent Shutdown process when shut downSystem power loss or improper shutdown detected at startuperror
THE SYSTEM WAS NOT SHUT DOWN PROPERLYSystem did not go through Intelligent Shutdown process; software and data may be corruptSystem power loss or forced shutdown without proper Intelligent Shutdown sequencecritical
THIS DICOM SERVER VERIFIED SUCCESSFULLYSystem successfully connected to DICOM server in response to verify requestSystem performed DICOM server verification from Print Definition, Store Definition, or Query Definition screeninfo
THIS DICOM SERVER VERIFIED SUCCESSFULLYSystem successfully connected to DICOM server in response to verify requestDICOM server verification request completed successfullyinfo
This study contains color images. Not all viewer functions wStudy containing color images is loaded in the Study ViewerColor image study loaded into viewerinfo
This study contains color images. Not all viewer functions wStudy containing color images is loaded in Study Viewer; some functions disabledLoading study with color images into Study Viewerinfo
THIS WILL ERASE ALL CAPTURED NAV IMAGESSurgical navigation (NAV) images are being captured and a Workstation operation will erase them (NEW EXAM, SCHEDULED EXAM, or SAVED EXAM initiated).Attempt to start new exam, scheduled exam, or saved exam while NAV is active and capturing images.warning
THIS WILL ERASE ALL CAPTURED NAV IMAGESSurgical navigation (NAV) images are being captured and a Workstation operation caused equivalent of NEW EXAM/SCHEDULED EXAM/SAVED EXAMNAV image erase initiated by exam change operationwarning
TOO MANY MATCHES FOUNDDICOM query produced too many resultsDICOM database search returned excessive number of resultswarning
TOO MANY MATCHES FOUNDA DICOM query produced too many results and cannot display all matches.DICOM query with insufficiently specific search criteria.warning
Transmission completeLog files have been successfully sent to the GE On-Line center.Log file transmission to GE service completed.info
Transmission complete.Displayed log files have been sent to the GE On-Line centerLog file transmission to GE service center completedinfo
Transmission complete. Reboot to restore system functionLog files successfully sent to GE On-Line centerLog file transmission to GE completedinfo
Transmitting files.User selected button to send log files to the GE On-Line centerUser initiated log file transmission to GE service centerinfo
Transmitting filesLog files are currently being transmitted to the GE On-Line center.User initiated log file transmission to remote service.info
Transmitting files. Please waitLog files being sent to GE On-Line centerUser selected button to send log files to GEinfo
UNIQUE ID FIELD DOES NOT MATCHMismatch between first and second Unique ID entry fieldsService screen entries compared; fields do not match exactlywarning
UNIQUE ID FIELD DOES NOT MATCHOne or more fields between the first and second Unique ID entry fields do not match on the Service screen.User entered second Unique ID entry line with mismatched field values.warning
UNIQUE ID FIELD INCORRECTOne or more Unique ID entry fields are incorrect or invalid.User entered Unique ID field with invalid or malformed data on Service screen.warning
UNIQUE ID FIELD INCORRECTOne or more Unique ID entry fields contain incorrect dataService screen Unique ID entry validation failedwarning
UNKNOWN ERROR RESPONSE RECEIVED FROM REMOVABLE MEDIAUnknown error detected in removable media during operationRemovable media operation attempted; media hardware error or data corruption detectedwarning
UNKNOWN ERROR RESPONSE RECEIVED FROM REMOVABLE MEDIAAn unknown error was detected in removable media (USB drive, external storage).Read/write operation to removable media encountered unrecognized error.error
UPS communication errorWorkstation software cannot establish communication with UPS unit at system startupWorkstation power-on sequence attempts to communicate with UPS via USBerror
UPS errorUPS detects internal fault that prevents it from supplying backup power to the systemUPS internal self-test detects power supply malfunctionerror
VERIFY FAILEDVerify Operation did not complete successfullyVerify process initiated; echo verification failederror
VERIFY FAILEDThe Verify Operation failed; the verify process was unsuccessful (echo failed).Attempt to verify system configuration or connectivity resulted in echo failure.error
WAITSystem is busy, not ready for new commandsSystem executing operation or scan in progressinfo
WAITSystem is busy with internal operationsSystem executing internal procedureinfo
WAITSystem is busy processing; standby messageSystem executing time-critical internal task or data processinginfo
WARNING, HIGH KVkV feedback error: actual kVp higher than indicatedkVp measurement exceeds displayed value and tolerancewarning
WARNING - HIGH kV - PRESS ANY KEYX-ray tube voltage exceeds desired settingSensed kV higher than desired kV valuewarning
WARNING, HIGH KV - PRESS ANY KEYMeasured kVp exceeds control panel setting and is out of toleranceActual kVp output higher than displayed setpointerror
WARNING, HIGH KV — PRESS ANY KEYkVp feedback error: actual kVp higher than displayed; out of tolerancekV sense circuit output exceeds expected value during operationwarning
WARNING, HIGH MAmA calibration error: actual mA higher than indicatedmA measurement exceeds displayed value and tolerancewarning
WARNING, HIGH mA - PRESS ANY KEYX-ray tube current exceeds desired settingSensed mA higher than desired mA valuewarning
WARNING, HIGH MA — PRESS ANY KEYmA feedback error: actual mA higher than displayed; calibration errormA sense circuit output exceeds expected tube current thresholdwarning
WARNING, HIGH MA - PRESS ANY KEYMeasured mA exceeds control panel setting and is out of toleranceActual tube current higher than displayed setpointerror
WARNING, LOW KVkV feedback error: actual kVp lower than indicatedkVp measurement below displayed value and tolerancewarning
WARNING - LOW kV - PRESS ANY KEYX-ray tube voltage below desired settingSensed kV lower than desired kV valuewarning
WARNING, LOW KV - PRESS ANY KEYMeasured kVp is below control panel setting and out of tolerance; may degrade imageActual kVp output lower than displayed setpointerror
WARNING, LOW KV — PRESS ANY KEYkVp feedback error: actual kVp lower than displayed; out of tolerance; poor image riskkV sense circuit output below expected value during operationwarning
WARNING, LOW MAmA calibration error: actual mA lower than indicatedmA measurement below displayed value and tolerancewarning
WARNING, LOW mA - PRESS ANY KEYX-ray tube current below desired settingSensed mA lower than desired mA valuewarning
WARNING, LOW MA — PRESS ANY KEYmA feedback error: actual mA lower than displayed; poor image quality riskmA sense circuit output below expected tube current thresholdwarning
WARNING LOW MA - PRESS ANY KEYMeasured mA is below control panel setting and out of tolerance; may degrade imageActual tube current lower than displayed setpointerror
WARNING! X-RAY HOUSING TEMPERATURE SENSOR FAILURETemperature information unavailable because the temperature sensor in the housing had failed.System detects temperature sensor failure at startup or during operationwarning
WARNING! X-RAY HOUSING TEMPERATURE SENSOR FAILUREX-ray tube housing temperature sensor has failed or disconnectedSystem unable to read housing temperature sensor during operationwarning
WORKSTATION AND GENERATOR ARE NOT COMPATIBLEWorkstation software and Generator software versions are incompatibleSystem bootup; software compatibility check failederror
WORKSTATION KEYBOARD ERRORCommunication with Workstation keyboard has failedSystem startup or during operation; keyboard communication losserror
WORKSTATION UPS COMMUNICATION FAILUREConnection between Workstation and UPS disrupted during startupSystem bootup; UPS communication timeout or errorerror
WORKSTATION UPS FAILUREUPS communication established but UPS reported error (not BATTERY LOW)System bootup; UPS self-test or health check failed with non-battery errorerror
X-ray OvertimeX-ray exposure time exceeded maximum permitted durationExposure time exceeds system limit during Film Mode operationwarning
X-ray OvertimeExposure duration exceeded safe limitTriggered during fluoroscopy when system attempts to exceed exposure time limitswarning
X-RAY OVERTIME - PRESS ANY KEYFilm exposure continued beyond required duration; software terminated exposureExposure time exceeded threshold during film shotwarning
X-RAY OVERTIME — PRESS ANY KEYFilm exposure time exceeded preset limit; software terminated exposureFilm shot exposure time exceeds calculated mAs termination pointinfo
X-RAYS DISABLEDX-ray generation is currently disabled in the systemUnknown disable event or safety interlock activationwarning
X-RAYS DISABLEDSystem has disabled X-ray generation capability; informational messageNormal system state when X-ray production is not commandedinfo
X-RAYS DISABLEDX-ray generation disabled by system softwareSafety interlock or system control disables X-raysinfo
X-RAYS DISABLED REBOOT REQUIREDDuring bootup, system detected active X-ray generation input signal and shut downActive X-ray command line, security line, or XRAY_ON, XRAY_EN signal during startuperror
X-RAYS DISABLED — REBOOT REQUIREDX-ray generation input signal active during boot; system shut downX-ray command signal detected at startupcritical
X-RAYS TEMPORARILY DISABLEDX-rays disabled during unexpected FFB reboot sequenceFFB firmware or hardware detected fault requiring restartwarning
X-RAY SWITCH SECURITY ERRORSystem detected X-ray switch security error following X-ray switch pressX-ray switch pressed; security interlock or safety verification failederror
X-RAY SWITCH SECURITY ERRORMismatch between footswitch/Handswitch security line and control connector security lineSecurity signal discrepancy during switch initializationerror
X-RAY SWITCH SECURITY ERRORX-ray switch security mismatch detectedHandswitch or footswitch signal does not match security linecritical
X-RAY SWITCH SECURITY ERROR — PRESS ANY KEYX-ray switch security line mismatch detected; system operation disabledHandswitch or footswitch signal does not match security line stateerror
X-RAY SWITCH SECURITY ERROR — PRESS ANY KEYMismatch between X-ray switch and security line signalsSwitch and security line signals do not correlatecritical
X-RAY SWITCH SECURITY FAILUREX-ray high-voltage switch security interlock failedX-ray switch register shows unexpected stateerror
X-RAY SWITCH STUCKX-ray high-voltage switch mechanically stuckX-ray switch register indicates stuck stateerror
X-RAY SWITCH STUCKX-ray switch enabled during boot-upX-ray switch detects activation during startup sequencecritical
XRAYSWITCH STUCKX-ray switch circuitry malfunction detectedWrong footswitch type (3 Pedal) booted with system when 4 Pedal Neurovascular configurederror
X-RAY SWITCH STUCKX-ray exposure switch detected as enabled during boot-upX-ray switch signal present at startupcritical
X-RAY TUBE TEMPERATURE SENSOR FAILURE - PRESS ANY KEYTube thermistor failed functional test at startup; no temperature monitoring activeThermistor test failure during power-on diagnosticscritical