GE Healthcare 9800: troubleshooting and error codes
The ServiceForge knowledge base holds 511 documented symptoms for the GE Healthcare 9800, with 886 ranked causes, 860 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Orbital motion fails or runs intermittently
C-arm orbital motion does not operate or operates sporadically
Ranked causes
- 1. RUI not properly connected to C-Arm keyswitch panel, or damaged RUI connector (common)
- 2. Orbital clutch released or damaged (occasional)
- 3. Orbital position-reporting potentiometer damaged or drive belt broken (occasional)
- 4. System out of calibration (occasional)
- 5. Faulty orbital gearbox (occasional)
- 6. Faulty orbital motor (occasional)
- 7. Faulty orbital servo drive (occasional)
- 8. Faulty MCU (rare)
- 9. Broken or damaged orbital drive belt (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motorized rotation fails or runs intermittently
C-arm rotational motion does not operate or operates sporadically
Ranked causes
- 1. RUI not properly connected to C-Arm keyswitch panel, or damaged RUI connector (common)
- 2. Rotational position-reporting potentiometer damaged or drive belt broken (occasional)
- 3. System out of calibration (occasional)
- 4. Faulty rotational gearbox (occasional)
- 5. Faulty rotational motor (occasional)
- 6. Faulty MCU (rare)
- 7. Faulty rotational servo drive (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No +60-Volt output from PS4 power supply
Power supply PS4 does not provide required +60V output
Ranked causes
- 1. Motor Power Relay PCB failure (common)
- 2. Cable failure between Power Motor Relay PCB and PS4 (occasional)
- 3. No AC power to PS4 input (occasional)
- 4. PS4 power switch set to OFF (occasional)
- 5. Motor power fuse open (common)
- 6. Ground or connector cable 902272 failure (occasional)
- 7. PS4 power supply failure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger inoperative high voltage
Charger does not operate or maintain battery voltage
Ranked causes
- 1. Workstation disconnected from AC line or AC line down (common)
- 2. Workstation Interconnect Cable disconnected from C-Arm (common)
- 3. Fuse blown or Battery Charger failed (occasional)
- 4. Instantaneous peak charging current exceeds 15.5 amps causing Overcurrent shutdown (occasional)
- 5. Output voltage exceeds 250V causing Overvoltage shutdown (occasional)
- 6. CHG_DISAB signal asserted by X-Ray Controller (common)
Verify AC power to Workstation
- Test point
- AC input to Workstation
- Expected
- 115VAC present
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
C-arm completely inoperative with battery charger light ON
C-Arm does not respond to commands, no error messages displayed, but battery charger indicator lamp is illuminated
Ranked causes
- 1. Buzzer sounds when Workstation switched on due to incorrect Isolation Transformer tap settings (occasional)
- 2. No Key Power (unknown)
- 3. Loose connectors on Power/Signal Interface PCB (common)
- 4. Blown fuse F6 or F10 on Power/Signal Interface PCB (common)
- 5. DC power distribution problem (occasional)
- 6. Defective Relay K10 in Power/Motor Relay PCB (rare)
Listen for buzzer sound when Workstation is switched on
- Test point
- Workstation power switch
- Expected
- No buzzer sound or buzzer indicates alarm condition only
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flat Panel Display screen is dark, distorted, or shows solid color
Video display is completely dark, image is distorted, or entire screen shows single color
Ranked causes
- 1. Display monitor power switch off or switches in wrong state (common)
- 2. Incorrect monitor control settings (common)
- 3. Defective video cable or connections (occasional)
- 4. Faulty power input to Flat Panel Display (occasional)
- 5. Faulty graphics card or motherboard in Navigation Computer (rare)
- 6. Failure of Flat Panel Display (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Navigation Computer has no power input high voltage
Navigation Computer not receiving AC or DC power. Monitor and system not operational.
Ranked causes
- 1. Workstation power failure (common)
- 2. Faulty power cable to Navigation Computer (common)
- 3. Faulty PS4 power supply (occasional)
- 4. UPS battery faulty or disconnected (occasional)
- 5. AC Fail cable disconnected (occasional)
- 6. Auxiliary Power Control PCB failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No image displayed on Color FPD caution
Color Flat Panel Display shows no video output; screen remains blank or off
Ranked causes
- 1. Display power off or video source inactive (common)
- 2. Video source inactive or not transmitting (common)
- 3. Loose video interface connector or pin contact failure (common)
- 4. Faulty or damaged video cabling (occasional)
- 5. AC power not reaching DC power supply (occasional)
- 6. DC power supply output failure (occasional)
Verify FPD display power is on
- Test point
- FPD power switch and indicator
- Expected
- Display powered ON with visible power indicator
- Tools
- Visual inspection
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Orbital servo checksum error
MCU detects checksum error in Logosol orbital servo control module during bootup
Ranked causes
- 1. Faulty orbital servo controller or related circuitry (unknown)
- 2. Faulty orbital motor (unknown)
- 3. Faulty wiring to orbital motor (unknown)
- 4. Faulty motion power supply (PS4) (unknown)
- 5. Faulty wiring to PS4 (unknown)
- 6. Faulty orbital position reporting potentiometer (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Orbital servo position error
MCU detects position error in Logosol orbital servo control during operation
Ranked causes
- 1. Faulty orbital servo controller or related circuitry (unknown)
- 2. Faulty orbital motor (unknown)
- 3. Faulty wiring to orbital motor (unknown)
- 4. Faulty motion power supply (PS4) (unknown)
- 5. Faulty wiring to PS4 (unknown)
- 6. Faulty orbital position reporting potentiometer (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotation servo checksum error
MCU detects checksum error in Logosol rotation servo control module during bootup
Ranked causes
- 1. Faulty rotation servo controller or related circuitry (unknown)
- 2. Faulty rotation motor (unknown)
- 3. Faulty wiring to rotation motor (unknown)
- 4. Faulty motion power supply (PS4) (unknown)
- 5. Faulty wiring to PS4 (unknown)
- 6. Faulty rotation position reporting potentiometer (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotation servo position error
MCU detects position error in Logosol rotation servo control during operation
Ranked causes
- 1. Faulty rotation servo controller or related circuitry (unknown)
- 2. Faulty rotation motor (unknown)
- 3. Faulty wiring to rotation motor (unknown)
- 4. Faulty motion power supply (PS4) (unknown)
- 5. Faulty wiring to PS4 (unknown)
- 6. Faulty rotation position reporting potentiometer (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Workstation boots but nothing on C-arm. Battery Charger Light ON high voltage
Workstation operating but C-arm non-responsive with charger light illuminated
Ranked causes
- 1. Blown fuse F6 or F10 on Power/Signal Interface PCB (occasional)
- 2. DC power distribution problem (occasional)
- 3. Defective Relay K10 in Power/Motor Relay PCB (occasional)
- 4. Power Signal Interface problem (occasional)
- 5. Fast stop buttons or wiring open (occasional)
- 6. F10 Fuse on Power/Signal Interface PCB (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger inoperative high voltage
Workstation unable to charge battery; charger not functioning
Ranked causes
- 1. Workstation disconnected from AC line or AC line down (common)
- 2. Workstation Interconnect Cable disconnected from C-Arm (common)
- 3. Fuse blown or Battery Charger failed (occasional)
- 4. Instantaneous peak battery charging current exceeds 15.5 amps (occasional)
- 6. X-Ray Controller asserts CHG_DISAB signal during normal operation (common)
Verify AC power is present at Workstation
- Test point
- AC inlet
- Expected
- AC line present and stable
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger light is OFF with Workstation off
Battery charger indicator lamp on C-Arm Interface Panel does not illuminate when Workstation is powered off
Ranked causes
- 1. Workstation power problem (common)
- 2. Wrong power cord voltage (occasional)
- 3. Loose connectors (common)
- 4. Power Control PCB jumper E1 set to wrong position (occasional)
- 5. Blown fuse F4 on Power/Signal Interface PCB (common)
Verify Workstation power and interconnect cables are plugged in
- Test point
- Workstation power inlet, interconnect cable connector
- Expected
- Cables fully inserted and seated
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger light OFF when workstation powered off
With Workstation in OFF state, battery charger indicator light is not illuminated.
Ranked causes
- 1. Workstation power problem (common)
- 2. Wrong voltage input for system (occasional)
- 3. Loose connectors on PCB (occasional)
- 4. Power Control PCB jumper E1 incorrect setting (occasional)
- 5. Blown fuse F4 on Power/Signal Interface PCB (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cine disk communication failure
Cine prepare fails or does not complete successfully; fibre channel communications not established
Ranked causes
- 1. Fibre Channel cables reversed at backplane connection (common)
- 2. Fibre Channel cables failed (occasional)
- 3. Power not connected to disk backplane (common)
- 4. Cine Bridge PCB failed (occasional)
- 5. One cine disk failure (occasional)
Check LED indicators on Cine Bridge PCB at boot-up
- Test point
- green LED indicators on Cine Bridge PCB
- Expected
- LEDs flicker during boot-up
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament voltage error patient safety
Wrong filament voltage applied to tube; mA output will also be incorrect
Ranked causes
- 1. Filament B plus wrong (common)
- 2. Wrong duty ratio of filament drive pulses (common)
- 3. Wrong filament selected (occasional)
- 4. Failure of filament regulation (occasional)
- 5. Sensor gain wrong (rare)
Measure filament voltage at FIL_VOLT relative to FIL_VOLT_REF
- Test point
- FIL_VOLT and FIL_VOLT_REF on Generator Driver PCB
- Expected
- FIL_VOLT equals FIL_VOLT_REF or expected voltage for selected filament
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flat Panel Display screen is blank high voltage
Display shows no image or backlight. Screen completely dark or inactive.
Ranked causes
- 1. Display control switches not correctly set (common)
- 2. No AC power to workstation (common)
- 3. Faulty power cable to FluoroTrak display power supply (common)
- 4. FluoroTrak power supply PS3 failure (occasional)
- 5. Faulty cable in A-Arm display connection (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No filament voltage observed high voltage
Filament voltage reading is zero at FIL_VOLT on Generator Driver PCB
Ranked causes
- 1. No FIL_B_PLUS voltage (common)
- 2. No filament drive signals present (common)
- 3. Faulty or shorted filament voltage sense resistors (occasional)
- 4. Failure of filament voltage sensor (occasional)
- 5. Broken connection to X-Ray Controller PCB or failure at X-Ray Controller PCB (occasional)
Measure FIL_VOLT at Generator Driver PCB
- Test point
- FIL_VOLT on Generator Driver PCB
- Expected
- Non-zero voltage present for selected filament
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Workstation off, battery charger light is OFF high voltage
Battery charger light not illuminated when workstation powered off
Ranked causes
- 1. Workstation power problem (common)
- 2. 100-125V system plugged into 200-250V (occasional)
- 3. Loose connectors (common)
- 4. Power Control PCB jumper E1 set to 2800 position (rare)
- 5. Blown fuse F4 on Power/Signal Interface PCB (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wrong mA during shot patient safety
mA output during X-ray exposure is not at expected level for selected technique
Ranked causes
- 1. Incorrect filament selected (common)
- 2. Faulty mA sense circuit (common)
- 3. Tube aging (occasional)
- 4. kVp wrong (occasional)
- 5. Improper exposure setting (occasional)
Verify correct filament selected via SM_FIL_ACK*
- Test point
- SM_FIL_ACK* signal state
- Expected
- Signal matches selected filament
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Collimator Alignment Verification radiation
Verify collimator iris alignment, field size diameters, and preview circle/line accuracy across normal and magnified modes using beam alignment tool and film exposure
Ranked causes
- 1. Collimator iris or leaves visible in fluoroscopy image (common)
- 2. Field diameter outside specification (common)
- 3. Collimator preview circles/lines misaligned (occasional)
- 4. Camera rotation preview symbols not functioning (occasional)
Fluoro in Normal field and verify iris/leaves visibility
- Test point
- Image intensifier display
- Expected
- No collimator iris or leaves visible in image
15 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motorized movement power-on check patient safety
Boot test to verify all power is produced and distributed correctly to motorized movement system. System should remain stable with no uncommanded motions.
Ranked causes
- 1. No +5 Volt power to Servo (common)
- 2. Faulty cable or connector on cable assembly 879450 (common)
- 3. Faulty cable or connector on cable assembly 884705 (common)
- 4. Drive sensors faulty (occasional)
Verify Workstation is plugged into active AC outlet
- Test point
- Power cord assembly
- Expected
- Pilot light on power cord assembly glows
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication failed between Workstation and C-Arm
Workstation not communicating with C-Arm for >1 second; system finishes X-ray in progress then disables future X-rays
Ranked causes
- 1. Foot pedal toe-tapping causing watchdog timeout (occasional)
- 2. Repeated random control panel button pressing (occasional)
- 3. Workstation software exception or processing anomaly (occasional)
- 4. Corrupted persistent object files (rare)
Observe foot pedal and button operation technique
- Test point
- Operator control
- Expected
- Deliberate, intentional control input
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fans operate, but system heats up too quickly or takes longer than 20 minutes to cool off caution
Heat dissipation performance degraded; either excessive heating or slow cooling
Ranked causes
- 1. Screen on fan vents blocked (common)
- 2. Fan ducting leaks (occasional)
- 3. Heat Pipe mounting loose or insufficient thermal grease (occasional)
- 4. Heat Sink mounting loose or insufficient thermal grease (occasional)
Visually inspect fan vent screens
- Test point
- All fan vent locations
- Expected
- Screens clear of blockage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament current error patient safety
Filament current reading does not match expected value for applied voltage and selected filament
Ranked causes
- 1. Filament aging (common)
- 2. Wrong filament selected (occasional)
- 3. Broken connection (occasional)
- 4. Current sense gain wrong (rare)
Verify filament voltage is correct
- Test point
- FIL_VOLT from Generator Driver PCB
- Expected
- Correct voltage for selected filament and technique
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament supply voltage wrong patient safety
Filament supply voltage (FIL_B_PLUS) is not at expected level for selected filament
Ranked causes
- 1. Bad filament supply regulator (common)
- 2. Filament voltage sense divider fault (occasional)
- 3. Wrong filament selected (occasional)
- 4. Wrong duty ratio of drive pulses (rare)
Measure FIL_B_SEN voltage at X-Ray Controller PCB
- Test point
- FIL_B_SEN analog input
- Expected
- Either about 160V or about 140V depending on selected filament, approximately ±20 V
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frame sync signal test failed
MCU detects missing or open frame sync signal path, TIM SystemTime.Fsync not incrementing
Ranked causes
- 1. Faulty wiring between MCU and backplane (unknown)
- 2. Faulty wiring between camera and workstation (unknown)
- 3. Faulty backplane PCB (unknown)
- 4. Faulty MCU PCB (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front caster binds or wobbles on axle bolt caution
Front caster wheel exhibits binding, stiction or lateral play on its mounting bolt
Ranked causes
- 1. Worn caster bearing (common)
- 2. Worn caster axle bolt (common)
- 3. Caster contaminated by debris (occasional)
- 4. Caster's bearing is dry (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No filament current (filament voltage present) high voltage
Filament voltage is normal but no filament heating current flowing to tube
Ranked causes
- 1. Connector disconnected at Filament Driver PCB or tank (common)
- 2. Broken connection in cable to tank (common)
- 3. Filament drive wire broken (occasional)
- 4. Filament current sense circuit fault (occasional)
Verify P3 and P1 connectors are properly seated
- Test point
- P3 at Filament Driver PCB; P1 at tank A9J1
- Expected
- Connectors fully inserted and locked
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tracking inactive error at boot patient safety
SNS boots to Visualization screen then displays 'WARNING: Tracking Inactive. System can only be used to perform surgical planning.'
Ranked causes
- 1. Incorrect boot process (unknown)
- 2. Incorrect power input to FluoroTrak Unit (unknown)
- 3. Serial cable connection fault between FluoroTrak Unit and Navigation computer (unknown)
- 4. Calibration and verification functional test failure (unknown)
Boot process function
- Test point
- SNS startup sequence
- Expected
- System boots without errors to Visualization screen
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wrong filament current observed (voltage correct) patient safety
Filament current reading is incorrect for applied filament voltage
Ranked causes
- 1. Incorrect filament selected (common)
- 2. Faulty filament current sense resistor (common)
- 3. Filament regulator faults (occasional)
- 4. Tube aging (occasional)
Measure filament current via true RMS sensing circuit
- Test point
- FIL_CUR signal from Generator Driver PCB
- Expected
- Current in expected range for selected filament and technique
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Wrong filament voltage observed patient safety
Filament voltage measurement is not in expected range for selected filament and technique
Ranked causes
- 1. Incorrect FIL_B_PLUS voltage (common)
- 2. INV_B_PLUS voltage low (common)
- 3. Faulty or disconnected filament voltage sense resistor (occasional)
- 4. Incorrect filament selected (occasional)
Measure filament voltage via true RMS sensing circuit
- Test point
- Filament voltage output from true RMS circuit on Generator Driver PCB
- Expected
- Expected range for selected filament and technique
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Boot-up fails with no error message
System fails to complete boot sequence with no error display
Ranked causes
- 1. Unseated PCBs (common)
- 2. Loose cable connectors (common)
- 3. Incorrect CMOS settings (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
C-arm completely inoperative with battery charger light ON high voltage
C-arm system is not operational and does not respond. Battery charger indicator light is illuminated.
Ranked causes
- 1. Buzzer sounds - incorrect isolation transformer tap setting (occasional)
- 2. No Key Power (occasional)
- 3. Loose connectors on Power/Signal Interface PCB (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
C-arm completely inoperative; no error messages; battery charger light is ON high voltage
C-arm non-responsive with charger light illuminated
Ranked causes
- 1. Incorrect tap settings on Isolation Transformer (occasional)
- 2. No Key Power (rare)
- 3. Loose connectors (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
C-Arm wheels do not roll smoothly caution
Wheels exhibit binding, vibration, or rough movement when pushed
Ranked causes
- 1. Wheel damage or debris (occasional)
- 2. Caster bearing failure (occasional)
- 3. Wheel insecurely attached to caster (occasional)
Visually inspect C-Arm wheels for damage
- Test point
- C-Arm wheels
- Expected
- No visible damage, cracks or flat spots
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCD Camera cooling not functioning
Cooling system fails to maintain proper temperature control for CCD camera
Ranked causes
- 1. No power at TEC (Thermoelectric Cooler) (common)
- 2. No +5VDC output from Fluoro Functions PCB (common)
- 3. No +12VDC at Power/Motor Relay PCB output (occasional)
Measure power at TEC
- Test point
- TEC power input
- Expected
- Power present
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cine Bridge PCB green LED indicators do not flicker at boot-up
Green LED indicators on Cine Bridge PCB remain static during system boot; indicates possible cable or PCB failure
Ranked causes
- 1. Fibre Channel cables installed backwards (common)
- 2. Fibre Channel cables damaged or failed (occasional)
- 3. Cine Bridge PCB failure (occasional)
Inspect and verify Fibre Channel cable orientation
- Test point
- Fibre Channel connections at 2-Disk or 4-Disk Backplanes
- Expected
- Cables in correct orientation
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data error in system memory
System detects data error (invalid ABS table write/read, corrupted non-volatile memory, or software problem); system shuts down immediately
Ranked causes
- 1. Non-volatile memory corruption (occasional)
- 2. ABS table corruption or invalid data write/read (occasional)
- 3. Software corruption or bug (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drag wires damaged or improperly positioned patient safety
Static drag wires not securely attached, not touching floor, or dirty
Ranked causes
- 1. Drag wire loose attachment or wear (occasional)
- 2. Drag wire not contacting floor (occasional)
- 3. Drag wire contamination (occasional)
Visually inspect drag wire attachment points
- Test point
- Drag wire fasteners (Mobile C-Arm and WorkStation)
- Expected
- Wires are securely attached
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive temperature at tank caution
High Voltage Tank or IGBT heat sink reaches abnormally high temperature during operation
Ranked causes
- 1. Oil expansion beyond set limits (occasional)
- 2. IGBT failure (occasional)
- 3. Cooling fan failure or disconnection (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External network (DICOM) not active
DICOM external network communication not functioning; boot message shows network link down
Ranked causes
- 1. Local network not operational (common)
- 2. External network cable not connected to motherboard (occasional)
- 3. Back-panel Ethernet cable or connector defective (occasional)
Verify Local Network is operational
- Test point
- Local network components
- Expected
- Local network functioning correctly
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament drive pulses absent high voltage
One or both filament drive signals not present; Filament Driver PCB produces no output
Ranked causes
- 1. Optoisolators faulty, broken connections, or no FIL_DRV_RTN connection (common)
- 2. Fault in 7407 or 74LS14 components (occasional)
- 3. Fault in FIL_CUR sensor (occasional)
Verify filament drive pulses present at Generator Driver PCB
- Test point
- Generator Driver PCB filament drive output
- Expected
- Pulses present on both filament drive lines
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament drive pulses absent or incorrect high voltage
No filament current or voltage; filament driver produces no drive; X-ray tube cannot generate X-rays
Ranked causes
- 1. Optoisolators faulty or broken connections to Filament Driver PCB (common)
- 2. Missing FIL_DRV_RTN connection or 7407 fault (occasional)
- 3. Faulty 74LS14 or FIL_CUR sensor (occasional)
Verify filament drive pulse presence at Generator Driver PCB
- Test point
- FILDRA and FILDRB outputs on Generator Driver PCB
- Expected
- Drive pulses present on both lines during X-ray request
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament supply voltage wrong high voltage
Filament B+ supply voltage is not at the correct level for the selected filament, preventing proper filament heating
Ranked causes
- 1. Bad filament supply regulator on Filament Driver PCB (common)
- 2. Filament voltage sense divider or sensor fault (occasional)
- 3. High filament supply voltage due to output stage fault (rare)
Measure FIL_B_PLUS voltage at Filament Driver PCB output
- Test point
- FIL_B_PLUS
- Expected
- About 160V or 140V per filament selection, ±10V
- Tools
- DMM
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament voltage error (wrong voltage applied) high voltage
Filament voltage at FIL_VOLT does not match FIL_VOLT_REF, resulting in incorrect mA generation
Ranked causes
- 1. Filament B plus wrong or duty ratio of filament drive pulses wrong (common)
- 2. Failure of filament regulation or sensor gain wrong (occasional)
- 3. Wrong filament relay selection or relay problem on Filament Driver PCB (occasional)
Measure filament voltage at FIL_VOLT relative to FIL_VOLT_REF
- Test point
- FIL_VOLT
- Expected
- Within specified range for selected filament and technique
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front caster does not pivot freely on vertical axis caution
Front caster wheel cannot rotate or turn smoothly around its vertical mounting point
Ranked causes
- 1. Damaged axle parts (occasional)
- 2. Axle parts are contaminated by debris (common)
- 3. Axle parts require lubrication (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interconnect cable or connectors damaged high voltage
Interconnect cable, Mobile C-Arm connector, or pins damaged or strain relief not secure
Ranked causes
- 1. Interconnect cable mechanical damage (occasional)
- 2. Connector or pins damaged (occasional)
- 3. LEMO connector strain relief loose (occasional)
Visually inspect interconnect cable for damage
- Test point
- Interconnect cable entire length
- Expected
- No cuts, pinches, or damaged insulation
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Joystick failure
System cannot communicate with joystick or Remote User Interface; joystick control inoperative.
Ranked causes
- 1. Joystick mechanically obstructed or at incorrect position (common)
- 2. RUI not properly connected or Motion PCB fault (common)
- 3. Joystick assembly or RUI hardware failure (occasional)
Clear joystick area and verify home position
- Test point
- Joystick physical area and center position
- Expected
- No obstruction; joystick centered; enable line inactive
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Joystick stuck
Joystick reported error during initialization; joystick control may be inoperative.
Ranked causes
- 1. Joystick physically obstructed or stuck (common)
- 2. Joystick not at home position or enable line active (occasional)
- 3. Joystick assembly or RUI failure (occasional)
Check for joystick obstruction
- Test point
- Joystick physical area
- Expected
- No objects, cables, or hands in contact with joystick
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift column inoperative caution
Vertical lift column does not respond
Ranked causes
- 1. Loose connectors (common)
- 2. No MOTOR-115-AC to PS3 caused by defective K15 or K16 relay (occasional)
- 3. Lift column control problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift column inoperative
C-Arm lift column does not respond to control inputs or fails to raise/lower
Ranked causes
- 1. Loose connectors on Power/Motor Relay PCB or PS3 inputs (common)
- 2. No MOTOR-115-AC to PS3 caused by defective K15 or K16 relay (rare)
- 3. Lift column control problem (unknown)
Inspect connectors on Power/Motor Relay PCB
- Test point
- Power/Motor Relay PCB connectors
- Expected
- All connectors fully seated with no corrosion or bent pins
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift motor movement test failed
MCU detects failure in lift motor up/down motion during test
Ranked causes
- 1. Faulty power motor relay PCB (unknown)
- 2. Faulty lift power supply (unknown)
- 3. Faulty lift column assembly (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift switch held on for 30 seconds or more caution
Lift switch remains in ON condition for extended period; all generator functions inhibited
Ranked causes
- 1. Object pressing on lift switch (common)
- 2. Faulty lift switch (occasional)
- 3. Faulty lift circuit (occasional)
Inspect lift switch for obstruction
- Test point
- C-Arm lift switch
- Expected
- No objects pressing on switch
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
mA sensor failure detected high voltage
Message alternates with technique display; system detects failure in one of mA sensors during operation
Ranked causes
- 1. mA sensor malfunction (common)
- 2. HV tank sense point fault (occasional)
- 3. HV Supply Regulator PCB failure (occasional)
Measure HV tank sense point voltages
- Test point
- HV tank sense points
- Expected
- Nominal operating range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MCU contact sense test failed patient safety
MCU diagnostics indicate bumper contact detection is not functioning; system cannot sense bumper position.
Ranked causes
- 1. Bumper assembly not properly attached or faulty (common)
- 2. Bumper assembly connector fault (common)
- 3. Bumper assembly wiring or MCU fault (occasional)
Verify bumper assembly attachment to Image Intensifier
- Test point
- Bumper mechanical mounting and connectors
- Expected
- Bumper physically attached; all connectors fully seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MCU joystick motion enable test failed
MCU diagnostics indicate joystick motion enable signal is not responding correctly or stuck.
Ranked causes
- 1. Joystick physically grasped or obstructed (common)
- 2. Joystick assembly failure (occasional)
- 3. Joystick wiring or MCU failure (occasional)
Release joystick and center position
- Test point
- Joystick mechanical position
- Expected
- Joystick at center (0,0) position; enable line signal responding
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MCU RUI sense test failed
MCU diagnostics indicate Remote User Interface is not detected during bootup; RUI communication unavailable.
Ranked causes
- 1. RUI disconnected or improperly connected (common)
- 2. RUI connector or wiring fault (common)
- 3. RUI hardware failure (occasional)
Disconnect and reconnect RUI connector
- Test point
- RUI connector physical interface
- Expected
- Connector fully seated with audible click; no loose contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Motion disabled
- Motion error
- Motion failure
- Motion Status Register 1 error: Collision fault
- Motion Status Register 1 error: Collision Ready failed
- Motion Status Register 1 error: Lift Down request failed
- Motion Status Register 1 error: Lift Up request failed
- Motion Status Register 1 error: Orbital fault
- Motion Status Register 1 error: Rotational fault
- Motion Status Register 2 error: Clutch sense failed
- Motion Status Register 2 error: Key switch failed
- Motion Status Register 2 error: RUI sense failed
- Navigation Computer not responding
- No +5 Volt power to Servo
- No exposure from Handswitch or Footswitch
- No filament current (filament voltage present)
- No filament voltage observed
- No response to touch controls after Visualization screen appears
- Orbital potentiometer status error
- Orbital servo overrun error
- Overload fault in high voltage regulator
- Overvoltage fault in X-ray generator
- Pre-charge voltage fails to reach battery voltage
- Rear caster does not roll freely on the floor
- Rear caster does not steer freely
- Rotation servo overrun error
- Rotational potentiometer status error
- RS422 serial communication failure
- System bootup fails at 13 Arrows stage
- System bootup fails at 14 Arrows stage
- System bootup fails at 18 Arrows stage
- System fatal error or memory failure
- Workstation boots but nothing on C-arm with battery charger light ON
- Wrong filament voltage observed
- Wrong mA during shot
- X-ray control switch stuck during startup
- X-ray switch security mismatch error
- X-ray tube overheats
- X-ray tube overheats; thermal sensor high or thermal cutout activated
- A-Arm excessive movement
- A-Arm movement
- A-Arm moves unintentionally
- AC power cord damaged
- Actual kVp higher than displayed value
- Actual kVp lower than displayed value
- Actual mA higher than displayed value
- Actual mA lower than displayed value
- Arcing observed
- Balance wheel does not turn
- Balance wheel is loose or wobbles
- Battery voltage below 160 volts
- Battery voltage drops below 163VDC during operation
- Black image – Technique servos to maximum
- Brakes on one or both rear wheels do not engage
- Bumper attached
- Bumper removed
- Calibration: A-Arm removal and installation
- Calibration: Camera Decentration Verification
- Calibration: Collimator Mechanical Centering
- Calibration: Image Tube Sizing Verification
- Calibration: mechanical pivot function and resistance
- CCD camera cooling connections defective
- Charger indicator light not illuminated
- Clutch is open
- Clutch position sensor signal failed
- Collision detected
- Collision with motorized vertical column during operation
- Contact detector fails to detect contact
- Contact detector incorrectly reports a contact
- Cooling fans not operating
- Display monitor excessive movement
- Display monitor movement
- Display monitor moves unintentionally
- Excessive temperature rise at High Voltage Tank
- External Interface PCB troubleshooting
- Fast stop activated
- Filament current error
- Flip flop brake does not prevent movement
- FluoroTrak display power supply no output
- Generator battery voltage below 160 Volts
- Ground continuity failure
- Hand switch indicator lights off
- Hand Switch indicator lights off
- Hard disk drive LED not illuminated
- Image Processor PCB or power strip output voltage out of range
- Image tube input dose out of specification
- Interlock system down
- Interlock system down
- Isolation transformer secondary output voltage out of specification
- Key disable signal sensing failed
- Light or dark image with working ABS – Technique adjusts and stabilizes
- Light output out of specification
- Low contrast images
- MCU diagnostic switch failed
- MCU joystick X position too high
- MCU joystick X position too low
- MCU joystick Y position too high
- Monitor support arm excessive movement
- Monitor Support arm movement
- Monitor support arm moves unintentionally
- Motion stop switch failed
- Motion Stop will not clear
- Motorized C-Arm rotation not responding or misaligned
- Motorized orbital motion not responding or misaligned
- Navigation Computer boot fails with 'Can't Open Boot Device' message
- No +15 Volt power to Remote User Interface
- No bootup
- No bootup
- No fans operate
- No image and kV tracks to 120
- No video but pilot tone OK
- No video output
- One front caster does not contact the floor
- Power supply voltages out of specification
- Pre-charge failure indicated by system
- Pre-charge voltage exceeds 250V maximum
- Printed image quality degraded or media jammed
- PS1 output incorrect
- RUI failure
- RUI key stuck
- Saturation fault in high voltage regulator circuit
- Stator does not start during boot-up
- Stator stops running
- Steering test failure
- System bootup fails at 15 Arrows stage
- System bootup fails at 16 Arrows stage
- System bootup fails at 19 Arrows stage
- System bootup fails at Five Arrows stage
- System does not complete pre-charging; PRECHARGE TIMEOUT ERROR displayed
- System does not complete pre-charging; PRECHARGE VOLTAGE ERROR displayed
- Various errors from IGBT peak voltage exceedance
- Vertical motion commanded from RUI fails or runs intermittently
- VFDs inoperative
- VFDs inoperative
- Video present but no pilot tone
- White image – Technique servos to minimum
- Workstation boots but nothing on C-arm with battery charger light OFF
- Workstation boots but nothing on C-arm. Battery Charger Light OFF
- WorkStation wheels do not roll smoothly
- Wrong filament current observed (when voltage correct)
- Wrong filament selection
- X-ray disable keyswitch in standby position
- +14VDC or -14VDC section of PS2 failed
- 24-hour recharge required
- 4-hour warm-up required
- A/D channel failure during startup
- Analog-to-digital channel failure
- Anode overheated warning
- Anode warm warning
- Anode/housing hot - HLF disabled
- Anode/housing warm warning
- Arcing detected in High Voltage Tank
- Arcing in High Voltage Tank
- Arcing in X-ray tube
- Auxiliary DAP control using DOS keyboard X-key toggle
- Balance wheel tread is worn or damaged
- Battery capacity below 40%
- Battery charge low warning
- Battery charge state warning
- Battery charge very low warning
- Battery charging disabled by X-Ray Controller
- BIOS not displayed during boot-up
- Brakes do not release
- Buzzer sounds when Workstation switched on
- C-Arm angle calibration verification
- C-Arm collision detection during motorized movement
- C-arm does not rotate freely when Flip Flop brake is released
- C-arm does not rotate freely with brake released
- C-Arm does not turn when steering handle turned with brake in neutral
- C-Arm Orbital brake ineffective
- C-arm PCB battery replacement
- C-Arm PS1 output voltage out of specification
- Cable pusher blades damaged
- Calibration: A-Arm lower end adjustment
- Calibration: Battery charger output voltage
- Calibration: Battery charger output voltage adjustment
- Calibration: Battery Charging System
- Calibration: Beam alignment tool to image tube centering
- Calibration: Beam Alignment Tool to Image Tube Centering
- Calibration: Beam alignment verification
- Calibration: Brake adjustment for ineffective braking
- Calibration: Camera decentration
- Calibration: Collimator Mechanical Centering
- Calibration: Collimator rotation and dose rate
- Calibration: DAP rotation constants (1-leaf collimator)
- Calibration: Dead time between kV drive signals
- Calibration: Differentiator setup
- Calibration: display monitor stowed position and retention
- Calibration: Dose Area Product using UtilitySuite
- Calibration: Filament/duty cycle calibration
- Calibration: Flat panel display (touch screen)
- Calibration: Fluorographic Pulsed mA Accuracy
- Calibration: Fluoroscopic kVp Accuracy at Maximum mA
- Calibration: Generator, Beam and Camera alignment
- Calibration: Horizontal Cross Arm motion and locking
- Calibration: Image Tube Electronic Centering Verification
- Calibration: Image tube sizing verification
- Calibration: Iris Stop
- Calibration: kVp accuracy
- Calibration: Lens Focus
- Calibration: Light Sensor
- Calibration: Light Source Fostec
- Calibration: mA accuracy verification
- Calibration: mA null adjustment
- Calibration: MAG2 field verification
- Calibration: monitor pivot mount adjustment
- Calibration: motorized C-arm angular positioning accuracy verification
- Calibration: motorized C-arm orbital and rotational positioning
- Calibration: Motorized movement system
- Calibration: motorized orbital motion
- Calibration: motorized rotational motion
- Calibration: Orbital clutch operation
- Calibration: Power Supply PS1
- Calibration: PWM chip sync adjustment
- Calibration: Ramp peak voltage adjustment
- Calibration: Rotation Pin adjustment
- Calibration: Secondary Arm removal and installation
- Calibration: Stability pot overshoot adjustment
- Calibration: Tracking system
- Camera decentration offset detected
- Charger failed
- Charger output voltage out of specification
- Cine Enclosure power supply voltages out of specification
- Clutch fails to engage
- Clutch fails to release
- Clutch Sense signal not responding to open clutch
- Collimator calibration required
- Collimator hesitates or stops during movement
- Collimator hunts for position
- Collimator iris not fully closing
- Collimator iris position potentiometer error
- Collimator iris potentiometer error
- Collimator iris too large
- Collimator iris unstable
- Collimator iris visibility in fluoro mode - assessment for misalignment
- Collimator leaf dose accuracy verification
- Collimator stuck
- Collision bumper contact sensing
- Collision detector fails to detect collision
- Collision Ready signal incorrect or bumper not detected
- Communication failed
- Configuration for single-leaf and dual-leaf collimator systems
- Contact detected
- CONTACT relay failed during startup
- Contact signal not responding to collision
- Control panel communication error
- Control panel error
- Control panel key stuck at bootup
- Control panel key stuck at operation
- Cross Arm cannot be moved
- Cross Arm does not lock in place
- Cross Arm does not remain locked
- Data error in memory
- Dead time adjustment for HV generation
- DICOM interface unable to archive or print images
- Display does not respond to touch input
- Drag wire damaged or improperly mounted
- Electrical hazard from liquid ingress during cleaning
- Equipment requires safety inspection before operation
- Error message displayed during boot-up
- External network inactive; boot message shows Link down
- Filament calibration required
- Filament driver fault indicated
- Filament regulator failed
- Filament regulator failure
- Filament select error
- Film exposure continues beyond required time
- Flat Panel Display assembly or arm positioning difficulties
- Flat Panel Display assembly positioning difficulties
- Flat Panel Display does not shut off immediately
- Flip Flop brake ineffective
- Flip Flop rotation restricted or rough
- Fluoro exposure technique fluctuates or exhibits abnormalities
- Fluoroscopic kVp Accuracy at 1 mA
- Fluoroscopic kVp Accuracy at Maximum mA
- FluoroTrak 9800 Plus navigation subsystem malfunction
- FluoroTrak Unit power not connected
- Footswitch security error
- Generator driver fault indicated
- High Voltage and Control Cable Assembly replacement required
- High Voltage Supply Regulator PCB replacement
- HLF overtime exceeded
- Horizontal Cross Arm cannot be moved from locked position
- Horizontal Cross Arm does not lock in position
- Horizontal Cross-Arm brake ineffective
- Horizontal Cross-Arm does not slide smoothly
- Housing is hot warning
- Housing overheated
- Housing overheated - X-ray disabled
- Housing warm or hot
- Housing warm warning
- HV cable arcing suspected
- HV generator error
- HV generator on unexpectedly
- HV register failure at startup
- HV register read failure
- Image tube focus adjustment
- Incorrect ARCNet node count display
- Interlock failure
- Interlock stage not progressing
- Joystick X-axis raw counts out of range
- Joystick Y-axis position incorrect
- Joystick Y-axis raw counts out of range
- Key stuck on panel
- Key Switch signal incorrect or not changing
- Keyboard keys unreliable or unresponsive
- kV on in error
- kV out of range (high)
- kV out of range (low)
- KVP drive timing problem indicated
- L-Arm brake ineffective
- Large filament non-functional
- Lift Down key signal not registering
- Lift switch stuck
- Lift Up key signal not registering
- Line voltage regulation outside acceptable range
- Low mA messages
- Low mA messages displayed
- Low mA output
- mA on in error
- mA out of range (high)
- mA out of range (low)
- mA sense active with no X-ray switch pressed
- mA sensor failure
- MCU NVRAM test failed
- MCU watchdog timer not incrementing
- Microcontroller not running
- Mobile C-Arm wheel brake ineffective
- Motion calibration required
- Motion enable signal not responding to joystick grasp
- Motion Stop button not engaging or disengaging
- Motion Stop fails to stop
- Motorized rotation does not display correct angle on workstation monitor
- Motorized rotation does not display correct direction on workstation monitor
- Navigation Computer fans do not shut down properly
- Navigation Computer fans not running
- Navigation Computer power supply voltage missing
- Neither vertical lift column button works
- No image and kV tracks to 120
- No image on Flat Panel Display
- No image on Flat Panel Display, screen dark or distorted
- No response to touch controls
- No video, no pilot tone
- Non-compliant external device connected to system
- One lift button works, but the other does not
- One lift button works, other does not
- One or two fans do not operate
- Orbital auxiliary status not responding
- Orbital clutch does not engage
- Orbital Clutch does not engage or disengage
- Orbital Fault signal not responding to collision
- Orbital motion does not display correct angle on workstation monitor
- Orbital motion does not display correct direction of movement on monitor
- Orbital rotation restricted or rough
- Orbital servo controller status not responding
- Overcurrent shutdown triggered
- Overload fault detected
- Overvoltage condition indicated
- Overvoltage fault detected
- Overvoltage messages
- Overvoltage messages displayed
- Overvoltage shutdown triggered
- Precharge capacitor voltage exceeds 250 Volts
- Precharge circuit timeout
- PRECHARGE VOLTAGE ERROR message
- PRECHARGE VOLTAGE ERROR message displayed
- Printer does not power on after installation
- Private Network is down
- PS1 output voltage incorrect or dropped
- PS2 output voltage incorrect
- PS2 test point verification
- PS3 output voltage incorrect
- Radial rotation restricted or rough
- Ramp adjustment for pulse shaping
- Room door open
- Rotational auxiliary status not changing
- Rotational Fault signal not responding to collision
- Rotational servo controller status not changing
- RUI key does not register unique code on press
- RUI key release code does not register consistently
- RUI Sense signal not detecting RUI connection
- RUI software revision not displaying
- Safety interlock bypassed or disabled
- Saturation fault detected
- Small filament non-functional
- Software fails to boot after pre-charge voltage cycling
- Stability adjustment for HV supply
- Stator not energized
- Stator runs at 115V, tube overheats
- Stator runs normally but error messages seen
- Status LEDs not illuminated on power distribution boards
- Steering handle does not align with rear wheels
- Steering handle misaligned with rear wheels
- Stuck control panel key
- Sync adjustment for PWM oscillator
- System boots with errors after reboot
- System bootup fails at 11 Arrows stage
- System bootup fails at 12 Arrows stage
- System bootup fails at 17 Arrows stage
- System bootup fails at Eight Arrows stage
- System bootup fails at Nine Arrows stage
- System bootup fails at Seven Arrows stage
- System bootup fails at Six Arrows stage
- System bootup fails at Ten Arrows stage
- System does not boot – stops at 14th arrow
- System does not boot – stops at 18th arrow on Control Panel
- System does not boot at 14th arrow on Control Panel
- System does not boot at 18th arrow on Control Panel
- System does not boot completely
- System does not complete boot sequence
- System does not complete pre-charging but relays click every few seconds
- System incorrectly reports a collision
- System performs motorized orbital motion with clutch released
- System shuts down with no error messages
- System shuts down with no error messages
- Temperature sensor failure
- Touch screen cursor does not align with touch point
- Touch screen does not respond to finger input
- Uninterruptible Power Supply (UPS) malfunction
- UPS battery voltage low
- UPS battery voltage low during discharge
- Various errors - Snubber components damaged or defective
- Verification: DAP constant calculation and system validation
- Vertical column operation not disabled when Fast Stop switch pressed
- Vertical Lift Column lift buttons non-functional
- Visualization screen did not appear during initialization
- Warm-up required
- Water or conductive fluid ingress causes short circuit or electric shock
- Water or conductive fluid ingress hazard
- Wheels do not move 90° when steering handle turned
- Wig-Wag brake ineffective
- Wig-Wag movement restricted
- Workstation casters do not lock when wheel pedal fully up
- Workstation displays Navigation error message
- WorkStation wheel brake ineffective
- Workstation wheels and casters do not turn freely in center pedal position
- Worn or damaged front caster tread
- Worn or damaged rear caster tread
- Wrong filament selection
- X-ray control released before film properly exposed
- X-ray generation not disabled when Fast Stop switch released
- X-ray housing temperature sensor failure
- X-ray room door open
- X-ray switch security failure
- X-ray switch stuck
- X-ray tube anode not powered
- X-ray tube current on without X-ray switch pressed
- X-ray tube fan not working
- X-ray tube overtime
- X-ray tube temperature sensor failure at startup
- X-rays disabled
451 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0D | System halted - out of memory | Memory allocation failure or exhausted memory pool during board initialization or operation | critical |
0x2251 | MCU Err Stat1 Lift Down Request Failed | Lift Down key pressed and held on RUI; MCU detects no response | error |
0x2252 | MCU Err Stat1 Lift Up Request Failed | Lift Up key pressed and held on RUI; MCU detects no response | error |
0x2255 | MCU Err Stat1 Collision Ready Failed | Bumper switch assembly unplugged or disconnected; collision detection unavailable | error |
0x2256 | MCU Err Stat1 Rotational Fault Failed | Actual collision detected during C-arm rotational movement | error |
0x2257 | MCU Err Stat 1 Orbital Fault Failed | Actual collision detected during C-arm orbital movement | error |
0x2258 | MCU Err Stat 1 Collision Failed | Bumper pressed and held during diagnostic; collision signal stuck active | error |
0x225D | MCU Err Stat2 Clutch Sense Failed | Clutch held open during diagnostic; sense circuit cannot detect clutch state | error |
0x225E | MCU Err Stat2 Key Switch Failed | C-Arm key switch turned to Standby position or not responding | error |
0x2260 | MCU Err Stat2 RUI Sense Failed | RUI unplugged or disconnected from system; MCU cannot detect RUI | error |
0x2261 | MCU Err JS Motion Enable Test Failed | Joystick grasped or held during bootup; motion enable signal not responding | error |
0x2263 | MCU Err Contact Sense Failed | Contact bumper pressed and held during bootup; system does not recognize bumper | error |
0x226A | MCU Err RUI Sense Test Failed | RUI disconnected or unplugged from system during bootup | error |
0x226B | MCU Err JS X Pos Too High | Joystick X axis rest position incorrect; joystick off-center at bootup | warning |
0x226C | MCU Err JS X Pos Too Low | Joystick X axis rest position too low; off-center from true neutral | warning |
0x226D | MCU Err JS Y Pos Too High | Joystick Y axis rest position incorrect; joystick off-center at bootup | warning |
0x226E | MCU Err JS Y Pos Too Low | Joystick Y-axis rest position reading below expected threshold | warning |
0x226F | MCU Err LGS Checksum Error | Checksum validation fails for orbital or rotational servo control module | error |
0x2270 | MCU Err LGS Position Error | Servo position error detected during orbital or rotational movement | error |
0x2271 | MCU Err LGS Servo Overrun | Orbital or rotational servo command overrun condition detected | error |
0x2272 | MCU Err Motion MS Sw Failed | Motion stop switch signal not detected or incorrect state on RUI | warning |
0x2273 | MCU Err NVRAM Test Failed | Binary configuration file CRC test fails during bootup | critical |
0x2274 | MCU Err Diag Switch Failed | Diagnostic switches S2 and S3 not set to 0,0 during boot or MCU hardware failure | warning |
0x2275 | MCU Err Frame Sync Test Failed | TIM SystemTime.Fsync signal not incrementing within 60 ms | error |
0x2277 | MCU Err Clutch Sense Failed | Clutch position sensor signal not detected or incorrect state | warning |
0x2280 | MCU Err Lift Motor Movement Test Failed | Lift up/down motor motion not detected during MCU self-test | warning |
Anode overtemp (implied from ANODE IS HOT message) | X-ray tube anode heat capacity critical | Anode heat level reaches or exceeds 80% of rated capacity | critical |
Battery low (BATTERY CHARGE message) | C-Arm battery charge below safe operating level | Battery voltage/charge below 40% of required capacity | warning |
Collimator fault (COLL IRIS POTENTIOMETER ERROR) | Collimator iris position feedback failure | COLL_IRIS_POSN line on Fluoro Functions PCB disconnected or not transitioning | error |
Communication timeout (COMMUNICATION FAILED message) | Workstation-to-C-Arm communication interrupted >1 second | No valid communication received for >1 second; Watchdog timer expires | error |
E0 | ARCNET transmit pipe temporarily full | Network transmission buffer overflow; excessive ARCNET traffic or transmission queue backlog | warning |
E-14 or equivalent A/D Channel failure code | Analog-to-digital converter channel failed | A/D channel malfunction detected during bootup self-test | critical |
E4 | Stack memory dangerously low | Available stack memory falls below critical threshold during operation | error |
Fan failure | One or more cooling fans not operating | Fan motor not running, no airflow detected | warning |
FAST STOP ACTIVATED | Fast Stop switch has been activated; X-ray generation disabled and vertical column operation suspended | Fast Stop switch pressed and held | warning |
FE | Fatal error - system halted | Stack overflow, no memory available, or unhandled interrupt detected during operation or initialization | critical |
FF | System halted - out of memory or fatal error | Memory exhaustion or Fatal Error condition | critical |
Filament regulation fault (FILAMENT REGULATOR FAILURE) | X-ray tube filament current out of tolerance | Filament current measurement outside acceptable range during operation | error |
Footswitch security fault (FOOTSWITCH SECURITY ERROR) | X-ray control security verification failed | X-ray security line state does not match X-ray control state | critical |
Housing overtemp (implied from HOUSING OVERHEATED message) | X-ray tube housing heat capacity exceeded maximum limit | Housing heat level reaches 100% or greater of rated capacity; thermal cutoff switch activates | critical |
HV generator fault (HV GENERATOR ERROR) | High voltage generator register error | HV generator register contents invalid or corrupted | critical |
KEY SWITCH IN STANDBY - X-RAYS AND LIFT DISABLED | X-ray disable keyswitch on generator is in vertical (ON) position | X-ray disable keyswitch activated | warning |
KV ON IN ERROR | X-ray tube current on but no X-ray switch pressed | System detects X-ray tube current without switch activation | critical |
LIFT SWITCH STUCK - PRESS ANY KEY | Lift switch held in ON condition for 30 seconds or more | Lift switch remains ON for extended period | warning |
LOADING LOW DOSE ABS | Low Dose ABS table selected as current ABS table | User selects Low Dose ABS table on generator | info |
LOADING STANDARD ABS | Standard ABS table selected as current ABS table | User selects Standard ABS table on generator | info |
MA ON IN ERROR | mA Sense on but no X-ray switch activated | System detects mA sense active without X-ray switch | critical |
MA SENSOR FAIL | System detects failure in one of the mA sensors | mA sensor functional test fails during operation | error |
Navigation Computer not responding | Navigation Computer unresponsive | Power or boot issue on Navigation Computer | warning |
No video, but pilot tone OK | Video signal missing while pilot tone (LIVE message) is present | Blown fuses F15 or F16 or loose connector P7 | error |
No video, no pilot tone | Both video signal and pilot tone are missing | Blown fuses F13 or F14 or loose connector P7 | error |
OVERLOAD FAULT | Possible fault in high voltage regulator circuit | Excessive current or overload condition detected | error |
OVERVOLTAGE FAULT | Possible X-ray generator failure indicated by overvoltage condition | Overvoltage detected in X-ray system | error |
PLEASE WAIT | System performing internal processing; X-rays disabled during this period | System internal processing active | info |
PRECHARGE CIRCUIT TIMEOUT | CONTACT relay failed to close during startup process | CONTACT relay does not close within timeout period during precharge initialization | critical |
PRECHARGE RELAY CLICK | System does not complete pre-charging but relays click every few seconds | Software failure with overvoltage protection activated during pre-charge | warning |
PRECHARGE TIMEOUT ERROR | Pre-charging did not complete within expected time duration | Pre-charging process exceeds maximum allowed duration | error |
PRECHARGE VOLTAGE ERROR | Voltage target not reached during pre-charging phase | Pre-charging cannot establish required voltage level | error |
PRECHARGE VOLTAGE ERROR | Precharge capacitor voltage exceeds 250 Volts during bootup | Precharge voltage check during system startup | critical |
PRECHARGE VOLTAGE ERROR | Battery charger output voltage is too high during normal charging operation | Battery charger output voltage exceeds specification when charging battery in normal mode | error |
pre-OPIS | System is pre-OPIS software version | Connection error when launching Utility Suite DAP Calibration | warning |
RELEASED EARLY | X-ray control released before film properly exposed in Film mode | X-ray control released prematurely during film exposure | warning |
ROOM DOOR OPEN – CLOSE DOOR | Door of X-ray room is open | X-ray room door not fully closed and latched | warning |
RUNNING CALIBRATION | Remote Utilities Tool (RUT) software calibrating generator filament | RUT software filament calibration procedure initiated | info |
SATURATION FAULT | Fault in high voltage regulator circuit; system will not operate | Saturation condition detected in HV regulator | critical |
STATOR NOT ON | X-ray tube anode not powered; stator not receiving voltage | System determines X-ray tube anode is not powered | critical |
STATOR STANDBY WAIT | System detects batteries below 40% capacity; charging required | Battery capacity drops below 40% during operation or standby | warning |
SUNW hme0 - Link down, Cable problems? | SUNW hme0 network link down | DCOM network connectivity issue | warning |
WARNING, HIGH KV - PRESS ANY KEY | Actual X-ray output kVp is higher than displayed value and not within tolerance | kVp sense circuit detects output exceeds displayed value | warning |
WARNING, HIGH MA - PRESS ANY KEY | Actual X-ray output mA is higher than displayed value and not within tolerance | mA sense circuit detects output exceeds displayed value | warning |
WARNING, LOW KV - PRESS ANY KEY | Actual X-ray output kVp is lower than displayed value and not within tolerance; poor image quality may result | kVp sense circuit detects output below displayed value | warning |
WARNING LOW MA - PRESS ANY KEY | Actual X-ray output mA is lower than displayed value and not within tolerance; poor image quality may result | mA sense circuit detects output below displayed value | warning |
WARNING: Tracking Inactive | Tracking system is not operational; system can only be used for surgical planning | System boots to Visualization screen with tracking inactive condition | warning |
WARNING: Tracking Inactive | FluoroTrak tracking function is not operational | System boots to Visualization screen and displays warning message | error |
X-RAY OVERTIME - PRESS ANY KEY | Film shot continues beyond required exposure time; software terminates exposure | Exposure duration exceeds programmed time limit during film shot | warning |
X-RAYS DISABLED | X-ray system disabled for safety or operational reason | Multiple possible triggers: keyswitch, room door open, safety interlocks | warning |
X-RAY SWITCH SECURITY ERROR | Mismatch between Security line on footswitch/handswitch and matching connector | Security line wiring mismatch detected during system startup or operation | error |
X-RAY SWITCH STUCK | X-ray control in ON state detected during system startup; applies to Left or Right controls | X-ray control switch sensed in ON position during boot sequence | critical |
X-RAY TUBE TEMPERATURE SENSOR FAILURE - PRESS ANY KEY | X-ray tube temperature sensing thermistor fails functional testing at start-up | Temperature sensor functional test fails during system bootup | error |
ZNB (0): 100 Mbps Full-Duplex | Network initialization message | Boot process initialization | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.