FR FieldReady Academy
Mobile C-arm X-ray systems · Series 9600 family

GE Healthcare 9600: troubleshooting and error codes

The ServiceForge knowledge base holds 75 documented symptoms for the GE Healthcare 9600, with 118 ranked causes, 155 diagnostic checks with test points and expected values, and 253 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Beam alignment radiation

Multi-step procedure to align X-ray beam, secondary collimator, variable collimator, and collimator iris with image intensifier electronics center using bore sight tool and beam alignment patterns

Ranked causes

  1. 1. Initial setup and bore sight tool installation (common)
  2. 2. Secondary collimator misalignment (occasional)
  3. 3. Image intensifier electronic center offset (common)
  4. 4. Variable collimator centering misalignment (occasional)
  5. 5. Collimator iris sizing requires calibration after Image Function PCB replacement (common)
  6. 6. Beam alignment verification with film (common)
  7. 7. Camera iris calibration long method baseline (common)
  8. 8. Camera iris calibration short method setup (common)
  9. 9. Adjust iris open stop short method (common)
First check

Bore sight tool centered in image

Test point
Monitor display
Expected
Bore sight tool centered on pattern

18 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stator motor does not start high voltage

Stator fails to initialize during boot sequence; no motor rotation detected; nineteen arrows not displayed on Control Panel

Ranked causes

  1. 1. STAT_RUN signal not going low; transistor Q3 or buffer U45 pins 15/5 failure (occasional)
  2. 2. STATSTRT signal not going low; transistor Q7 or buffer U45 pins 13/7 failure (occasional)
  3. 3. Relay K1 or K2 not energizing on Power Motor Relay PCB (occasional)
  4. 4. Circuit breaker CB2 open; 115VAC not reaching stator motor (occasional)
  5. 5. Stator motor winding open or shorted (rare)
First check

Verify PIO U38 pin 20 output during stator start

Test point
Analog Interface PCB PIO U38-20
Expected
High voltage output (logic high)
Tools
DMM, oscilloscope

7 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stator NOT ON error displayed on Control Panel patient safety

Error message 'STATOR NOT ON' appears on display; stator current not detected; X-rays and motor movement blocked

Ranked causes

  1. 1. Stator motor not drawing current; no current flow through current sense transformer T1 (common)
  2. 2. Current sense transformer T1 secondary open; induced current not detected (occasional)
  3. 3. Operational amplifier U19 or clipping circuit (diodes CR5-7, resistor R15, capacitor C17) failure (occasional)
  4. 4. Inverter U24 pins 5 & 4 failure; signal not inverted to logic low (occasional)
  5. 5. X-ray tube thermal cutoff switch open; stator disabled by overtemperature (occasional)
First check

Verify stator motor receives 115VAC during start phase (first 2 seconds)

Test point
Stator motor terminal leads at X-ray tube assembly
Expected
115VAC applied
Tools
DMM, oscilloscope

8 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PRE-CHARGE FAIL displayed during boot high voltage

Control Panel display shows PRE-CHARGE FAIL message; system boot does not complete; fatal error that prevents system startup

Ranked causes

  1. 1. +200VSNS signal does not reach 4.5 VDC (pre-charge capacitors do not reach 185 VDC) (common)
  2. 2. Fuse F3 on Generator Driver PCB is open (occasional)
  3. 3. Voltage divider R10/R11 malfunction or open circuit (occasional)
  4. 4. Analog Interface PCB A/D Mux U3-8 or signal path failure (occasional)
First check

Measure voltage across pre-charge capacitors C2 & C3

Test point
Generator Controller Assembly capacitors C2 & C3
Expected
approximately 185 VDC during boot sequence, ±10 VDC
Tools
Voltmeter rated for 221 VDC

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charger not receiving AC input high voltage

Battery packs not charging. Yellow/red bargraph indicators DS1 and DS2 inactive.

Ranked causes

  1. 1. Circuit breaker CB2 or CB3 open on Power/Signal Interface PCB (common)
  2. 2. Line filter LF1 or LF3 failed (occasional)
  3. 3. Bridge rectifier CR1 failed (occasional)
First check

Check CB2 and CB3 breaker status

Test point
Power/Signal Interface PCB CB2, CB3
Expected
Both in normal position

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

kV output does not match desired setting radiation

Measured X-ray output kilovoltage does not achieve or maintain desired kV level during fluoroscopy

Ranked causes

  1. 1. kV control voltage out of range or incorrect (common)
  2. 2. kV drive signal duty cycle incorrect (common)
  3. 3. kV sense signal not proportional to output (occasional)
First check

Select 80 kV in MANUAL FLUORO mode without making X-rays

Test point
U20-7 X-ray Regulator PCB
Expected
Approximately -6 VDC (-75 mV/kV) on amp U20-7 of X-ray Regulator PCB, ±1 V
Tools
Multimeter, X-ray Regulator access

6 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Stator motor runs but does not reach 3500 RPM speed high voltage

Stator motor spins during run phase but speed is below nominal 3500 RPM; inadequate voltage supplied after two-second start sequence

Ranked causes

  1. 1. Stator transformer T3 not stepping down 115VAC to 40VAC (common)
  2. 2. Relay K2 contacts stuck open or corroded in run phase (occasional)
  3. 3. Current sense transformer T1 inserting excessive impedance (occasional)
First check

Measure stator transformer T3 primary voltage during run phase

Test point
Stator Transformer T3 primary winding (after start sequence)
Expected
115VAC, ±5%
Tools
DMM, oscilloscope

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power not reaching C-Arm power systems high voltage

Battery charger inactive, power supplies not energizing. AC indicators (DS2 green lamp) not lit.

Ranked causes

  1. 1. Circuit breaker CB2 open or tripped (common)
  2. 2. Interconnect cable disconnected or failed (common)
First check

Verify CB2 breaker status

Test point
CB2 on Power Control PCB
Expected
Reset only breaker in normal position

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charger overvoltage shutdown active patient safety

Battery not charging. Red bargraph indicator (last segment) lit continuously. System cannot boost charge voltage.

Ranked causes

  1. 1. Output voltage exceeded 250 VDC (common)
  2. 2. Voltage feedback potentiometer R32 misadjusted or failed (occasional)
First check

Measure charger output voltage at point before shutdown

Test point
Battery Charger PCB output at fuse F1 location on Generator Controller
Expected
Should not exceed 250 VDC; typical 225–235 VDC in normal mode, ±10 VDC
Tools
DMM DC setting; HIGH IMPEDANCE probe only

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage low during exposure (less than 160V)

Bargraph or DMM measurement shows battery voltage drops below 160V during X-ray exposure, indicating inadequate power delivery.

Ranked causes

  1. 1. Weak or failing battery packs (common)
  2. 2. Battery Charger PCB fault (occasional)
First check

Select FILM mode and perform test exposure

Test point
Exposure result and bargraph display
Expected
No 'X-ray Overtime' message; no kV or mA errors; red LED bargraph shows 160-170 volts
Tools
System controls, DMM (optional)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SRAM Hot-Byte configuration

Hot-Byte is a control byte at address 0000 in EEPROM.DAT file that determines direction of data transfer between EEPROM U58 and SRAM card during boot-up; value 5A (0) is normal, A5 (1) forces SRAM→EEPROM overwrite for recovery.

Ranked causes

  1. 1. Hot-Byte = 0 (5A): normal operation, EEPROM→SRAM or blank EEPROM→SRAM (common)
  2. 2. Hot-Byte = 1 (A5): forced SRAM→EEPROM for EEPROM recovery (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Wig Wag brake fails to lock forward mechanics caution

Brake handle turns without engaging; forward mechanics move when brake should be locked

Ranked causes

  1. 1. Actuator screw insufficiently tightened (common)
  2. 2. Worn or damaged washers or brake pad (occasional)
First check

Attempt brake engagement after actuator screw adjustment

Test point
Wig Wag brake handle
Expected
Brake handle engages and locks; forward mechanics do not move when brake is fully engaged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charger high charge mode not activating patient safety

Battery charging slow when KVP above 30. Yellow bargraph DS2 indicates low charge current even during X-ray operation.

Ranked causes

  1. 1. KVP signal HICHGMOD not received from X-ray Regulator PCB (occasional)
  2. 2. Comparator U5 or feedback divider circuit (R9, R10, R26) failed (occasional)
First check

Measure HICHGMOD signal voltage during high KVP exposure

Test point
Battery Charger PCB U5 input from X-ray Regulator PCB
Expected
Logic high (approx +12V) when KVP >30
Tools
DMM DC or oscilloscope

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator iris motor stuck

Iris motor has not reached selected destination; may be mechanical jam or slow motor operation

Ranked causes

  1. 1. Motor jam or excessive friction (common)
  2. 2. Motor moving too slowly (occasional)

Control Panel diagnostic LED display not showing key codes caution

Diagnostic LED array (DS1) on Control Panel PCB not displaying row/column codes when keys pressed

Ranked causes

  1. 1. Diagnostic jumper not installed or loose on Control Panel PCB E1 pins 7-8 (common)
  2. 2. LED driver circuit failure or LED array defective (occasional)
First check

Verify diagnostic jumper installed on E1 pins 7-8

Test point
Control Panel PCB header E1
Expected
Shorting jumper present on pins 7-8
Tools
Visual inspection, jumper

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image Function PCB reset

Image Function PCB has undergone unexpected reset; watchdog timer or power monitor triggered reset

Ranked causes

  1. 1. Watchdog timer timeout (U8) (occasional)
  2. 2. Power supply glitch or dropout (occasional)

K10 relay not switching AC power to supply rails caution

System in standby; pressing keyswitch does not activate power supplies or motor circuits.

Ranked causes

  1. 1. KEY_PWR +12V signal missing from keyswitch (common)
  2. 2. K10 relay coil open or shorted (occasional)
First check

Verify +12V KEY_PWR signal at K10 relay coil

Test point
Power/Motor Relay PCB K10 relay coil input
Expected
+12V when keyswitch ON, ±0.5 V
Tools
DMM DC setting

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overvoltage Fault displayed on control panel high voltage

System displays OVERVOLTAGE FAULT message and shuts down, preventing X-rays

Ranked causes

  1. 1. High voltage exceeds 140 kVp (occasional)
  2. 2. Disconnected TAP1 or TAP2 signal lines on P4 connector (occasional)
First check

Verify P4 connector TAP1 and TAP2 signal lines are connected

Test point
P4-4 (TAP2), P4-5 (TAP1) on X-ray Regulator PCB
Expected
Signals connected and present
Tools
Visual inspection, oscilloscope

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Peripherals or subsystems not powered caution

Hard drives, floppy drive, control panel, or optional Lenzar camera do not function

Ranked causes

  1. 1. Circuit breaker tripped or open (common)
  2. 2. +12V or +5V output rail too low (occasional)
First check

Inspect circuit breaker status for failed subsystem

Test point
DC Power Distribution PCB (CB1–CB9 per subsystem)
Expected
Breaker lever in up position (not tripped)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply PS1 output voltage out of specification caution

Technique Processor or Control Panel circuits not operating. System may not boot or intermittent failures.

Ranked causes

  1. 1. +5V output drifted above tolerance (occasional)
  2. 2. +15V or -15V output drifted (occasional)
First check

Measure +5V output

Test point
Technique Processor PCB, C9
Expected
+5.05 VDC ±0.05 VDC, ±0.05 VDC
Tools
DMM DC setting, high impedance probe

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System displays 'CHARGER FAILED'

Message indicates battery charger is not functioning properly during normal operation.

Ranked causes

  1. 1. Battery voltage below 225V and charger output current below 1.0 Amp (common)
  2. 2. Weak battery packs (occasional)
First check

Measure charging voltage at Battery Charger PCB

Test point
J1-1 (S/N 69-0001-69-1000) or J5-1 (S/N 69-1001+)
Expected
225 VDC
Tools
Digital Multimeter

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System displays 'RECHARGE NEEDED - ALLOW 24 HOURS'

Message appears during boot-up after 6 months of inactivity without power.

Ranked causes

  1. 1. System not powered on for 6 months (common)
  2. 2. Backup generator software installed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System displays 'WARM-UP NEEDED - ALLOW 4 HOURS'

Message appears during boot-up after 2 months of inactivity without power.

Ranked causes

  1. 1. System not powered on for 2 months (common)
  2. 2. Backup generator software installed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System memory test failure

Memory test stops at error with address of bad memory location displayed on screen

Ranked causes

  1. 1. Base or Extended memory module failure or defective RAM chip (rare)
  2. 2. Memory module unseated or loose connection (common)
First check

Run System Memory Tests from diagnostics menu

Test point
AT Base Memory and Extended Memory
Expected
PASSED for both Base and Extended memory
Tools
Diagnostic software, system memory modules

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Workstation does not power on or boots intermittently high voltage

System fails to start or repeatedly resets during boot sequence; AT Motherboard may lock up

Ranked causes

  1. 1. +5V output from PS1 falls below +4.65V (common)
  2. 2. U1 voltage monitor IC failure (occasional)
First check

Measure +5V output at test point

Test point
TP27 (GND reference TP25)
Expected
+5.0 VDC ±0.75 VDC, ±0.75 VDC
Tools
Digital multimeter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad iris calibration

Potentiometer values at maximum and minimum iris positions do not match EEPROM stored values during boot-up

Ranked causes

  1. 1. Calibration values corrupted or iris mechanical shift (occasional)

Bad iris potentiometer

Erroneous or no potentiometer values sensed during system boot-up

Ranked causes

  1. 1. Potentiometer failure or open circuit (common)

Battery Charger high charge mode evaluation high voltage

Procedure to verify charger operates at high charge voltage during fluoro exposure.

Ranked causes

  1. 1. Battery Charger verification during high-current demand (rare)
First check

Remove covers and connect DMM to batteries

Test point
Battery terminals
Expected
DC voltage reading
Tools
Digital Multimeter set to DC

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Battery Charger normal charge mode evaluation high voltage

Procedure to verify charger is operating at normal charge voltage and current levels.

Ranked causes

  1. 1. Battery Charger verification during routine maintenance (rare)
First check

Remove covers and connect DMM to batteries

Test point
Battery terminals
Expected
DC voltage reading
Tools
Digital Multimeter set to DC

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Battery charger under-voltage protection active

Charger disabled when battery discharged. No trickle charge restoring battery voltage.

Ranked causes

  1. 1. Battery discharged below 160 VDC triggering under-voltage lockout (common)
First check

Measure battery voltage at charger output

Test point
Battery pack terminal voltage at charger output connector
Expected
Should be >160 VDC to enable fast charging; <160 VDC triggers trickle mode
Tools
DMM DC setting, high-impedance probe

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABS Index File parameters

Edit ABS Index File records including VLI Offset and Pulse VLI Offset values for different dose tables and scan rates

Ranked causes

  1. 1. ABS table parameter adjustment for dose optimization (occasional)
First check

Verify VLI Offset values are not modified

Test point
ABS Index Table records (I. Edit VLI Ofst)
Expected
Factory set values: Standard 5R/10R=0, Chest 5R/10R=-100, Extremities 5R/10R=50

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery Charger Output Voltage high voltage

Procedure to adjust charger output voltage to 220-225 VDC with bargraph indicator calibration.

Ranked causes

  1. 1. Charger output voltage calibration required (rare)
First check

Connect DVM to measure output voltage

Test point
J1-1 to chassis ground
Expected
220 Volts initial measurement
Tools
DVM set to DC volts

4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Collimator field sizes radiation

Iris collimator field sizes require calibration and storage in EEPROM on Image Function PCB

Ranked causes

  1. 1. Collimator calibration procedure required (common)
First check

Access Mainframe Menu via Workstation

Test point
Workstation setup
Expected
Mainframe Menu appears on left monitor

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Entrance exposure verification for 5R systems radiation

Verify and adjust entrance exposure dose rate output to ensure FDA compliance (maximum 4.63 R/min for systems S/N 69-0001 to 69-1160 manufactured before May 19, 1995)

Ranked causes

  1. 1. mA Limit File requires adjustment to maintain 4.63 R/min maximum output (common)
First check

Confirm system ABS tables are correct for 5R configuration

Test point
Mainframe control panel ABS SELECT button
Expected
Four ABS tables visible: Standard for Domestic 5R, Low dose for domestic, Chest 5R, Extremities 5R

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Generator parameters and personality file

System Configuration menu creates and updates personality file with generator settings including boost, inhibit, beeper, video levels, and ABS tables

Ranked causes

  1. 1. System Configuration setup during installation or reconfiguration (common)
First check

Verify boost availability setting

Test point
Option A in System Configuration menu
Expected
Y (Yes)

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Line pair resolution radiation

Verify spatial resolution of image on developed film at three field sizes per specification table

Ranked causes

  1. 1. Line pair resolution measurement required after beam alignment (common)
First check

NORMAL field line pair resolution at 60 Hz

Test point
Film
Expected
1.4 lp/mm

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SRAM card backup to floppy after generator calibration

After generator calibration is performed, new calibration data must be copied from SRAM card on Technique Processor PCB back to 3.5-inch floppy backup disk to preserve calibration in case of future SRAM failure.

Ranked causes

  1. 1. Generator calibration completed; backup required (common)
First check

Boot computer with PCMCIA software

Test point
PCMCIA port
Expected
Software recognizes PCMCIA port

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SRAM card restoration from floppy backup

SRAM card on Technique Processor PCB has failed or been corrupted; files must be restored from 3.5-inch floppy backup disk to preserve system-unique calibration and serial number data.

Ranked causes

  1. 1. SRAM card corrupted or failed (occasional)
First check

Boot computer and verify PCMCIA port recognized by software

Test point
PCMCIA port
Expected
Computer recognizes PCMCIA port, software loaded

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator iris jitter

Iris size drifts away from selected position requiring repeated corrections by Image Function PCB

Ranked causes

  1. 1. Unstable iris motor or potentiometer feedback (unknown)

Collimator iris too large

Collimator iris exceeds maximum size for selected field; detected by Image Function PCB comparing potentiometer value with stored calibration

Ranked causes

  1. 1. Iris potentiometer value exceeds stored calibration value (common)

Collimator replacement caution

Procedure to remove and install collimator assembly including cover and alignment bracket

Ranked causes

  1. 1. Collimator failure or misalignment requiring replacement (unknown)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Diagnostic tone not produced when keys pressed

Keys do not produce audible tones during Front Panel Test or Control Panel diagnostics

Ranked causes

  1. 1. Audio amplifier failure or speaker disconnected (occasional)
First check

Press numbered key (e.g., key 1) and listen for tone

Test point
Control Panel speaker output
Expected
Audible tone at frequency matching key (e.g., key 1 = 100 Hz)
Tools
Oscilloscope or audio frequency meter (optional)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament fault error high voltage

System detects no voltage on MA MEASURED line with X-ray switch engaged, indicating little or no tube current; fatal error causing system shutdown

Ranked causes

  1. 1. No tube current present with X-ray switch engaged (occasional)
First check

Measure voltage on MA MEASURED line during X-ray exposure

Test point
MA MEASURED line
Expected
Voltage present indicating tube current flow

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Front Panel keys not responding in diagnostics

Keys pressed on front panel do not produce expected keycode or tone in diagnostic mode

Ranked causes

  1. 1. Front Panel key matrix failure or key contact issue (common)
First check

Press each key on Front Panel and verify keycode output

Test point
Front Panel keys
Expected
Different hex keycode for each key pressed
Tools
Workstation with diagnostic software

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Front Panel LEDs not illuminating in diagnostics

LEDs on front panel do not light up when tested or do not indicate proper key presses

Ranked causes

  1. 1. LED driver failure or LED burnout (occasional)
First check

Activate each LED using cursor keys and verify illumination

Test point
Front Panel and TrackPad Panel LEDs
Expected
Each LED lights sequentially; LED value changes on display
Tools
Workstation diagnostic software

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fuse F1 open or blown on PS1 high voltage

Power supply PS1 does not deliver any output; main fuse is charred or broken

Ranked causes

  1. 1. Internal PS1 short circuit or component failure (occasional)
First check

Inspect main fuse F1 on PS1

Test point
F1 on power supply PS1
Expected
Fuse intact, no charring or breaks

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

HIGH KV AND LOW FLUORO KV error message

Detected kV and selected technique kV do not agree within 20%; system shuts down for 2 seconds after exposure

Ranked causes

  1. 1. KV measurement circuit error or HV regulation fault (unknown)

High mA warning

Actual mA (MA MEASURED) exceeds selected mA by more than 20%; non-fatal error causing 2-second shutdown

Ranked causes

  1. 1. Actual tube current exceeds selected value by >20% (occasional)
First check

Compare actual mA (MA MEASURED) to selected mA value

Test point
Technique Processor software display
Expected
Actual mA within 20% of selected mA, ±20%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image Function PCB timeout

No serial I/O response received from Image Function PCB

Ranked causes

  1. 1. PCB not responding or communication link failure (common)

Image Intensifier removal high voltage

Procedure to remove Image Intensifier tube and power supply from C-arm; includes high-voltage and safety precautions

Ranked causes

  1. 1. Image Intensifier failure requiring replacement (unknown)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image Processor memory test failure

Image Processor test halts with error; specific memory region fails (Left Filter RAM, Right Filter RAM, Bit Planes, or LUT)

Ranked causes

  1. 1. Image Processor memory module defective or unplugged (rare)
First check

Run Image Processor Tests from diagnostics menu

Test point
Image Memory, Filter RAM, Bit Planes, LUT modules
Expected
PASSED for Image Memory, Left/Right Filter RAM, Bit Planes, LUT
Tools
Diagnostic software

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Infrared transmitter key not responding in diagnostics

Infrared remote control keys do not produce keycode output in diagnostic mode

Ranked causes

  1. 1. IR transmitter battery dead or IR receiver module failure (common)
First check

Press each IR transmitter key and verify keycode output

Test point
IR transmitter
Expected
Different keycode displayed for each key pressed
Tools
IR transmitter, diagnostic software

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

KV ON IN ERROR message displayed patient safety

System displays error and shuts down when no X-ray switch is engaged but high voltage is detected on KVP MEASURED line

Ranked causes

  1. 1. Unintended high voltage present without X-ray switch command (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Level or Window value not adjusting in diagnostics

Level and Window control values remain static or do not change when controls adjusted

Ranked causes

  1. 1. Level or Window potentiometer mechanical wear or loose connection (occasional)
First check

Adjust Level control and verify value output

Test point
Level control potentiometer
Expected
Level value changes 0–80
Tools
Diagnostic software

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low fluoroscopy mA

Actual mA (MA MEASURED) is less than selected mA by more than 20%; non-fatal error causing 2-second shutdown

Ranked causes

  1. 1. Actual tube current falls below selected value by >20% (occasional)
First check

Compare actual mA (MA MEASURED) to selected mA value

Test point
Technique Processor software display
Expected
Actual mA within 20% of selected mA, ±20%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

mA on in error high voltage

System detects tube current present on MA MEASURED line without X-ray switch engaged; fatal error causing system shutdown

Ranked causes

  1. 1. Tube current flowing without X-ray switch command (rare)
First check

Verify X-ray switch is disengaged

Test point
X-ray switch input
Expected
No voltage on MA MEASURED line when switch is not engaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitors not receiving power high voltage

Left and right monitors display no power indicator or backlight

Ranked causes

  1. 1. PS4 +24V output below specification (common)
First check

Measure +24V at monitor input

Test point
Monitor circuit board J6 (black/GND to yellow/+24V)
Expected
+24 VDC ±1V, ±1V
Tools
Digital multimeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motion Artifact Reduction control not toggling

MARS (Motion Artifact Reduction) feature cannot be switched on or off

Ranked causes

  1. 1. Averaging mode not available in current X-ray mode (common)
First check

Verify current X-ray mode

Test point
X-ray mode indicator
Expected
System in Continuous X-Ray mode (not Pulse mode)
Tools
System software interface

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No mA limits message

System displays 'NO MA LIMITS' message indicating MALIMIT.DAT file is missing from SRAM disk

Ranked causes

  1. 1. MALIMIT.DAT file not copied to SRAM disk (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overload Fault displayed on control panel high voltage

System displays OVERLOAD FAULT message and shuts down, preventing X-rays

Ranked causes

  1. 1. Primary winding current exceeds 200 Amps (occasional)
First check

Measure integrated current voltage on TP1

Test point
TP1 on X-ray Regulator PCB
Expected
Voltage remains below 1.6 VDC (representing max 200 Amps primary current), less than 1.6 VDC
Tools
Digital multimeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Power supply PS2 (+24V) output out of specification
  • RAM test failure on Control Panel PCB
  • Regulator fail error
  • Saturation Fault displayed on control panel
  • Specific key not producing correct diagnostic code
  • SRAM lithium battery replacement
  • Stator motor shuts down prematurely; power loss after five minutes
  • System displays '50% CHARGE' and film exposures suspended
  • System displays '70% CHARGE'
  • System displays 'BATTERY DISCONNECTED'
  • TrackPad quadrant pressure sensor malfunction in diagnostics
  • TrackPad superkey switch not responding in diagnostics
  • TrackPad X/Y axis values out of range in diagnostics
  • Verify ABS table configuration for 5R systems
  • X-ray exposure time exceeded

15 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01English language selectedDuring boot sequenceinfo
05Footswitch stuckX-ray ON switch, footswitch, or handswitch sensed closed during bootcritical
06Precharge failSystem has not completed precharge cycle; B+ voltage not detected on Generator Driver PCBcritical
07HV inverter onHigh voltage generator sensed on with no X-ray command initiatedcritical
0830 minute time markerTime marker during operationinfo
09Checksum BIOS ROMBoot sequence test on Technique Processor PCBcritical
09Main program loadedDuring boot sequenceinfo
0AInitialization completeAfter hardware initialization completesinfo
0AInitialize Video InterfaceVideo interface initialization during bootcritical
0BImage Function PCB resetImage Function PCB loses communication with Technique Processor and watchdog timer resets IF PCBcritical
0CAD channel 0 failedkVp detected during boot-up without X-ray command initiatedcritical
0DAD channel 1 failedmA detected during boot-up without X-ray command initiatedcritical
0EAD channel 8 failedVoltage sensed on redundant kV sense circuitry in HV tank during boot-upcritical
0ETest CMOS Shutdown ByteCMOS shutdown byte validation during booterror
0FTest Extended CMOS (If Present)Extended CMOS validation during boot if module installederror
0FAD channel 11 failedExcessive current sensed from Battery Charger PCB during boot-upcritical
10MAIN.TXT file read errorError reading MAIN.TXT file from SRAM card during boot-upcritical
100Disk read errorerror
101Disk write errorerror
103File not openerror
104File not open for inputerror
105File not open for outputerror
11Fluoroscopy X-rayLast function performed was fluoroscopy with X-rayinfo
12Press any keyDuring system operationinfo
13Fluoroscopy X-ray offFluoroscopy exposure endedinfo
15Invalid drive numbererror
15Test 1st 64K of System MemoryFirst 64KB system memory test during bootcritical
15Fluoroscopy boost modeFluoroscopy exposure in boost modeinfo
152Drive not readyerror
153Unknown commanderror
154CRC error in dataerror
156Disk seek errorerror
158Sector not founderror
159Printer out of paperwarning
16Set up Interrupt Vector TableInterrupt vector table setup during bootcritical
16Fluoroscopy to film transitionOperator initiated transition from fluoroscopy to film modeinfo
160Device write faulterror
161Device read faulterror
162Hardware failurecritical
17Collimator stuckCollimator iris cannot open or closewarning
17Set Up Video I/O OperationsVideo I/O operations setup during booterror
18Test Video MemoryVideo memory test during boot (when EGA/VGA not found)error
18Iris jitterCollimator cannot maintain specified field sizewarning
19Standby to film transitionOperator initiated transition from standby to film modeinfo
1AFront panel deadCommunication loss between Control Panel Processor PCB and Technique Processor PCBcritical
1BTest CMOS Battery LevelCMOS battery voltage validation during bootwarning
1CTest CMOS ChecksumCMOS checksum validation during booterror
1CBad iris potentiometerCollimator iris potentiometer voltage is incorrect or out of rangecritical
1DBad iris calibrationCollimator iris calibration data in EEPROM on Image Function PCB is not validcritical
1DSet Configuration From CMOSConfiguration load from CMOS during booterror
1ESizing System Memory and Compare with CMOSSystem memory size test and CMOS comparison during booterror
1EBoost exposure overtimeBoost exposure exceeds preset time limitwarning
1FTest Found System MemoryFull system memory test during bootcritical
1FHousing is warmHousing temperature reaches approximately 60% of rated heat capacityinfo
2File not founderror
20Housing is hotHousing temperature reaches approximately 80% of rated heat capacitywarning
201Range check errorerror
202Stack overflow errorerror
203Heap overflow errorerror
204Invalid pointer operationerror
21Anode is warmX-ray tube anode reaches approximately 60% of rated heat capacityinfo
22Anode is hotAnode temperature reaches approximately 80% of rated heat capacitywarning
24Housing overheatedHousing temperature reaches 100% of rated value and thermal cutoff switch activatescritical
25Will overheatTube anode or housing at 80% of rated heat capacity during film modewarning
250-252Communication errorerror
253Communications errorSystem does not boot fully; serial communication losscritical
254-260Communications errorerror
27No boost when hotAnode or housing at 80% of rated heat capacity during boost modewarning
28Instant-on film prearm timeoutPrearm in FILM mode times out after no X-ray command receivedinfo
29Instant-on film front panel cancelPrearm in FILM mode terminated from front panel by operator pressing another keyinfo
2AKeyboard InitializationKeyboard controller initialization during booterror
2BNo mapping fileError opening mapping file during bootcritical
2BDiskette InitializationDiskette drive initialization during bootwarning
2CSerial Port InitializationSerial port controller initialization during bootwarning
2DParallel Port InitializationParallel port controller initialization during bootwarning
2DImage Function PCB timeoutImage Function PCB hasn't responded to Technique Processor within timeout periodcritical
2EFilm to fluoroscopy transitionOperator initiated transition from film to fluoroscopy modeinfo
2EHard Disk InitializationHard disk drive initialization during booterror
2FPlease waitGenerator loading software during bootinfo
3Path not founderror
31Battery disconnectedBatteries not connected or B+ voltage low during early boot-up sequencecritical
31Option ROM InitializationOption ROM modules initialization during bootwarning
32Standby to fluoroscopy transitionOperator initiated transition from standby to fluoroscopy modeinfo
33Standby mode activatedGenerator not used for 30 minutesinfo
34Standby mode endedGenerator activated after standby conditioninfo
35Film exposure endedEnd of film exposureinfo
36X-ray exposure completeExposure cycle completedinfo
37Released earlyX-ray ON switch released before desired mAs achieved during film exposurewarning
38X-ray overtimeProper mA not achieved for allotted time in FILM modewarning
3BNo option fileOPTIONS.DAT file not found on SRAM card during bootcritical
3FWarning high mAmA detected higher than allowed range for desired techniquewarning
4Too many open fileswarning
40Warning low mAmA detected lower than allowed range for desired techniquewarning
42Iris too largeCollimator iris size is incorrect for selected field sizewarning
43mA on in errormA sensed without an X-ray requestcritical
44Regulator failFilament voltage out of range on mA regulator circuitrywarning
45kV on in errorkVp sensed without an X-ray requestcritical
46Stator not onAnode in X-ray tube sensed not rotatingcritical
47Overload faultExcessive current to high voltage tank sensed by tuned circuit L1, C1 on generator controller assemblycritical
48Saturation faultDarlington driver transistors Q1 and Q2 on generator controller assembly not fully saturatedcritical
4AInterlocks open+24V generator interlock circuit has been openedcritical
4CNo range fileFailure to open Calrange.dat file during bootcritical
4DOver-voltage faultExcessive kVp sensed in H.V. tank secondary winding after boot-upcritical
4FNo MUX dataCannot access data in MUX.DAT file on SRAM cardcritical
50Charger failedBattery Charger PCB output detected out of range after boot-upcritical
50% CHARGEBattery charge capacity has been depleted to 50% or less of nominal.Software monitoring of battery state-of-charge calculation.warning
50_PCT_CHARGEBattery charge at 50 percentBattery charge drops to 50% of effective chargewarning
51User endMainframe Menu accessinfo
51Invalid function error during SRAM card software loadTechnique Processor attempts to load software from SRAM card and encounters invalid function requestcritical
52Exit programMainframe Menu exitinfo
52Address Mark Not Found error during SRAM card readTechnique Processor cannot locate address mark on SRAM card during software loaderror
53Write Protect Error during SRAM card accessTechnique Processor attempts to write to write-protected SRAM carderror
54Sector Not Found error during SRAM card accessTechnique Processor cannot locate requested sector on SRAM carderror
54Loading statusStatus Mode is accessedinfo
55Diskette Change Line Active during SRAM card operationTechnique Processor detects active diskette change signal from card interfacewarning
56DMA Overrun error during SRAM card data transferDirect Memory Access controller overruns during high-speed data transfer from SRAM carderror
56No mA limitsmA Limit file missing from SRAM cardcritical
57DMA Boundary Error during SRAM card transferData transfer request crosses invalid memory boundary during DMA operationerror
57Warning high kVkV detected higher than allowed range for desired techniquewarning
58Warning low kVkV detected lower than allowed range for desired techniquewarning
58Media Type Not Available for SRAM card operationTechnique Processor detects SRAM card type is not supported or card type cannot be identifiederror
59Bad CRC during SRAM card data verificationCyclic Redundancy Check fails when verifying software data loaded from SRAM carderror
5ADiskette Controller error during SRAM card operationTechnique Processor detects hardware error in card interface controllercritical
5APercent charge indicatorBattery charge state displayedinfo
5BSeek Failed error during SRAM card sector locationTechnique Processor unable to locate requested sector position on SRAM carderror
5BStandard 5R/min modeABS table displayinfo
5CLow dose modeABS table displayinfo
5CTime-out during SRAM card operationTechnique Processor operation times out waiting for SRAM card responseerror
5DChest 5R/min modeABS table displayinfo
5EExtremities 5R/min modeABS table displayinfo
5FStandard modeABS table displayinfo
6Invalid file handleerror
60Chest modeABS table displayinfo
61Extremities modeABS table displayinfo
63Checksum errorChecksum routine fails on MAIN software during bootcritical
64Boot-up terminatedDisplayed after CHECKSUM ERROR (code 63)critical
70% CHARGEBattery charge capacity has been depleted to 70% or less of nominal.Software monitoring of battery state-of-charge calculation.info
70_PCT_CHARGEBattery charge at 70 percentBattery charge reduced to 70% of effective chargewarning
80Started Execution of Option ROM Codeinfo
81NEAT TEST Fail - Reseterror
82Bad GSP Load - Self Loop with Interrupts Disablederror
83-8FUndefinederror
90Started Disk Boot Procedureinfo
91Data Error on Disk Readerror
92Controller Errorerror
94Seek Failureerror
98Disk Timed Out (Failed to Respond)error
99-9FUndefinederror
A0Undefinedwarning
A1Nobrk Installed Successfullyinfo
A2Undefinedunknown
A3Nobrk Already Installedinfo
A4-AFUndefinedunknown
ANALOG_CIRCUIT_FAULTAnalog interface data acquisition failureMultiple AD channels report failures or invalid valuescritical
ANODE_IS_HOTAnode hot warning in fluoroscopy modeAnode temperature reaches 80% of rated heat capacity during fluoroscopy operationwarning
ANODE_IS_HOT_FILMAnode hot warning in film modeAnode temperature reaches 80% of rated heat capacity during film operationwarning
B0Undefinedwarning
B1Enter TSR Installinfo
B2Undefinedunknown
B3Previous User Interrupt Not Nullwarning
B4Changing User Interrupt vectorinfo
B5Initialize Hardware Vectorinfo
B6Initialize Vectorinfo
B7-BDUndefinedunknown
Bad Iris CalIris calibration mismatch between current and stored valuesBoot-up drive to max/min positions yields different potentiometer values than EEPROMerror
Bad Iris PotIris potentiometer returns erroneous or no valuesDetected during system boot-uperror
BAD_THERMISTORBad thermistor detectedFault detected in housing temperature thermistor or associated circuitry during bootcritical
BATTERY DISCONNECTEDSoftware detects batteries are electrically disconnected from the system.System boot-up when all but one arrows are lit on control panel display; +200V sense line reads 180V or less after charger disabled.error
BATTERY_DISCONNECTED_OPBattery packs disconnected during operationBattery packs not connected when system powered upcritical
BESuccessful Installationinfo
BFTsrser Already Installedinfo
BOOST_OVERTIME_OPBoost exposure timeoutBoost exposure button held over 30 secondswarning
C0-C9Undefinedwarning
CAMicronics Cache Initializationinfo
CBUndefinedunknown
CCNMI Handler Shutdowninfo
CD-CFUndefinedunknown
CHARGER FAILEDBattery charger is not delivering adequate voltage and current to charge batteries.Software detects battery voltage less than 225V and charger output current less than 1.0A during background monitoring (except during X-rays).error
CHARGER_FAILED_OPCharger failure during operationBattery Charger circuit detected as failederror
D0Program Started Executioninfo
D1Error Opening GSP Object Fileerror
D2Undefinedunknown
D3Seek Errorerror
D4Seek Error On Dataerror
D5Virtual Address Errorerror
D6Error Reading File Headererror
D7Error in Test Program Response Prior to Loaderror
D8Undefinedunknown
D9Undefinederror
DALoad Complete and Successfulinfo
DBUndefinedunknown
DCNo Acknowledge from GSPerror
DDError Reading GSP Ready Flagerror
DEError Reading GSP Error Flagerror
DFGSP Program Load Failederror
E0Workstation Program Executinginfo
E1-EBUndefinedunknown
EEUnexpected Processor Exceptionerror
EFUndefinedunknown
F0Workstation Program Exited to DOSinfo
F1-FAUndefinedunknown
FBGSP is not Readyerror
FCGSP is not respondingerror
FDFront Panel is not Functioningerror
FEUndefinedunknown
FFInt 19H Bootinfo
FILAMENT FAULTNo voltage detected on MA MEASURED line with X-ray switch engagedDetected during X-ray exposure attemptcritical
FOOTSWITCH_STUCK_OPFootswitch stuck during operationFault detected in footswitch, X-RAY ON switch, or switch sensing electronics during bootcritical
FRAMING ERRORCommunications error between control modulesSerial communications failure between Control Panel Processor PCB and Technique Processor PCBerror
HIGH KV AND LOW FLUORO KVDetected kV and selected technique kV differ by more than 20%KVP MEASURED does not agree with selected kV technique within 20% toleranceerror
HOUSING_IS_HOTHousing hot warning in fluoroscopy modeHousing temperature reaches 80% of rated heat capacity during fluoroscopy operationwarning
HOUSING_IS_HOT_FILMHousing hot warning in film modeHousing temperature reaches 80% of rated heat capacity during film operationwarning
HV_INTERLOCK_FAULTHigh voltage with interlock openHV generator sensed on when interlock circuit opencritical
HV_INVERTER_ON_OPHigh voltage unexpectedly onHigh voltage detected when it should be offcritical
IFB ResetImage Function PCB has reset unexpectedlyWatchdog timer or power monitor triggered reset conditionwarning
IFB Time-outNo serial communication response from Image Function PCBExpected serial I/O response not received within timeout periodcritical
INTERLOCKS_OPEN_OPInterlock circuit broken during operation+24V generator interlock circuit openedcritical
I/O FAIL HH nnnnI/O software fatal error with hex module code and addressI/O operation failure or timeout in application codecritical
Iris Jitter (Collimator Iris)Iris size drifts and requires repeated correctionsIris position unstable, Image Function PCB making continuous correctionswarning
Iris Too Large (Collimator)Collimator iris exceeds maximum size for selected fieldIris potentiometer value exceeds stored calibration valuewarning
KV_ON_ERROR_OPkV sensed without X-ray request during operationKV sensed without X-ray command requestedcritical
KV ON IN ERRORHigh voltage detected without X-ray switch commandKVP MEASURED voltage present when no X-ray switch is engagedcritical
LOW FLUORO MAActual tube current (MA MEASURED) is less than selected mA by more than 20%During fluoroscopy mode when measured mA is <80% of commanded valuewarning
MA_ON_ERROR_OPmA sensed without X-ray request during operationmA sensed without X-ray command requestedcritical
MA ON IN ERRORTube current detected on MA MEASURED line without X-ray switch commandWhen X-ray switch is not engaged but MA MEASURED voltage is presentcritical
Motor Stuck (Collimator Iris)Iris motor failed to reach selected destinationMotor not moving or moving too slowly during iris driveerror
NO_BOOST_HOTNo boost allowed due to excessive heatHousing or anode temperature at approximately 90% of rated heat capacitywarning
NO MA LIMITSMALIMIT.DAT file is not found in SRAM diskSystem boot or technique loadingerror
OVERFLOW ERRORCommunications buffer overflow between modulesSerial communications overflow between Control Panel Processor PCB and Technique Processor PCBerror
OVERLOAD FAULTPrimary winding current exceeds safe maximum of approximately 200 AmpsCurrent voltage on comparator U1-2 exceeds reference voltage (~1.6 VDC) on U1-3critical
OVERLOAD_FAULT_OPHV tank overload during operationExcessive current to HV tank sensedcritical
OVERVOLTAGE FAULTHigh voltage exceeds safe maximum of approximately 140 kVpTAP1 signal voltage exceeds 5 VDC threshold at comparator U32-3critical
OVERVOLTAGE_FAULT_OPHigh voltage regulation failureExcessive kVp sensed after boot or during operationcritical
PARITY ERRORParity check failure in communicationsParity error in serial communications between Control Panel Processor PCB and Technique Processor PCBerror
PRE-CHARGE FAILPre-charge capacitors did not reach required voltage during boot sequenceTechnique Processor PCB does not detect +200VSNS signal reaching approximately 4.5 VDC during boot (representing 185 VDC across pre-charge capacitors C2 & C3)critical
PRECHARGE_FAIL_OPPrecharge failure during operationPrecharge cycle failscritical
RECHARGE NEEDED - ALLOW 24 HOURS - PRESS ANY KEY TO CONTINUESystem batteries require extended recharge and de-gas cycle after 6 months without operation.System boot-up when power has not been applied for 6 months.warning
REGULATOR FAILFilament regulator output voltage exceeds specified range by ±15%During X-ray exposure when FILSNS voltage monitored through Analog Interface exceeds rangewarning
REGULATOR_FAIL_OPFilament regulator failureFilament regulator circuit failure detectederror
RELEASED_EARLY_OPFilm exposure released earlyX-RAY ON switch released before desired mAs reachedwarning
RT FAIL HH nnnnRun-time software fatal error with hex module code and addressSoftware exception or fatal error condition in application codecritical
SATURATION FAULTDarlington power transistor voltage drop excessive during on-timeSATDET voltage on comparator U14-2 exceeds reference voltage (~8 VDC) on U14-3critical
SATURATION_FAULT_OPDriver saturation failureDarlington transistor saturation faultcritical
STATOR_NOT_ON_OPStator not rotating during operationX-ray tube anode stator detected not rotatingcritical
WAIT_WILL_OVERHEATAll fluoroscopy prevented - tube overheatingHousing or anode at approximately 90% of rated heat capacity in fluoroscopy or film modewarning
WAIT_WILL_OVERHEAT_FILMFilm exposures disabled - predicted overheatingPredicted housing or anode at 90% of rated heat capacity due to selected technique in film modewarning
WARM-UP NEEDED - ALLOW 4 HOURS - PRESS ANY KEY TO CONTINUESystem batteries and image tube require warm-up cycle after 2 months without operation.System boot-up when power has not been applied for 2 months.warning
WARNING_HIGH_KV_OPkV above desired range during operationkV detected higher than allowed rangewarning
WARNING, HIGH MAActual tube current (MA MEASURED) exceeds selected mA by more than 20%During or after X-ray exposure when measured mA is >120% of commanded valuewarning
WARNING_HIGH_MA_OPmA above desired range during operationmA detected higher than allowed rangewarning
WARNING_LOW_KV_OPkV below desired range during operationkV detected lower than allowed rangewarning
WARNING_LOW_MA_OPmA below desired range during operationmA detected lower than allowed rangewarning
X-RAY OVERTIMEExposure time in Film mode has exceeded 10% of expected exposure durationWhen elapsed exposure time exceeds 110% of programmed exposure time in Film modewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.