Gaymar TP600 & TP700: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Gaymar TP600 & TP700, with 29 ranked causes, 11 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Warning indicator flashing and audible alarm beeping caution
LED warning indicator flashes and unit beeps audibly, signaling alarm or test condition
Ranked causes
- 1. High heat setpoint (107°F / 42°C) or cooling setpoint (50°F / 10°C) selected (common)
- 2. Loss of power during therapy mode (power failure) (occasional)
- 3. Unit timing 20- or 30-minute on/off cycle (TP700 Series) (common)
- 4. Unit entering or leaving Lock mode (TP700 Series) (common)
- 5. Safety Circuit Function Test started (occasional)
- 6. Setpoint outside body temperature range selected (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow indicator light on
Flow indicator illuminated, signaling water flow restriction to pad or hose
Ranked causes
- 1. Water flow restricted to pad or hose (common)
- 2. Clamp is closed on line (common)
- 3. Water level low or reservoir empty (common)
- 4. Water temperature too hot (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
T/Pump will not cool caution
Unit fails to cool fluid to therapy temperature
Ranked causes
- 1. Reservoir empty (common)
- 2. Flow blocked (common)
- 3. Ice depleted in cooling water line (common)
Check reservoir water level
- Test point
- Reservoir
- Expected
- Filled to proper level
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
T/Pump will not heat caution
Unit fails to heat fluid to therapy temperature
Ranked causes
- 1. Reservoir empty (common)
- 2. Flow blocked (common)
- 3. Therapy cycle off (common)
Check reservoir water level
- Test point
- Reservoir
- Expected
- Filled to proper level
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water leaks from hose connectors caution
Water escapes from connections between pump and pad or hose joints
Ranked causes
- 1. Damaged O-ring in connector (common)
- 2. Clik-Tite connector locking ring not snapped into place (common)
- 3. Quick disconnect not seated properly (occasional)
Visually inspect connector O-ring
- Test point
- Clik-Tite connector O-ring
- Expected
- O-ring intact and seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature Setpoint light blinking
Temperature setpoint indicator light is blinking
Ranked causes
- 1. Unit is warming up to the selected setpoint (common)
- 2. Unit is in Cooling mode for longer than 40 minutes (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning indicator on with unit in Standby mode caution
Warning light illuminated while unit is in Standby mode
Ranked causes
- 1. Unit shut down in over-temperature condition (common)
- 2. Safety Circuit Function Test completed successfully (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both temperature and therapy cycle setpoint lights blinking
Temperature and therapy cycle indicator lights flash continuously on TP700 series
Ranked causes
- 1. Unit in off therapy cycle time (common)
Verify therapy cycle status
- Test point
- Therapy cycle indicator
- Expected
- Blinking stops when cycle turns on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow indicator and Standby indicator on with T/Pump not pumping patient safety
Flow and Standby lights illuminated but pump is not delivering fluid
Ranked causes
- 1. Unit detected Flow warning for more than 5 minutes and went to standby (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
T/Pump running with Standby light on
Pump operates while Standby indicator is illuminated
Ranked causes
- 1. Safety Circuit Function Test has started (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature or therapy time buttons do not work
Control buttons for temperature or therapy time setting are non-responsive on TP700 series
Ranked causes
- 1. Buttons locked (common)
Attempt to unlock buttons
- Test point
- Lock button
- Expected
- Buttons become responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not pump
T/Pump runs but no fluid circulates through pads or hoses
Ranked causes
- 1. Water level low or reservoir empty (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on
T/Pump fails to power up when switch activated
Ranked causes
- 1. Power cord not plugged into grounded hospital receptacle (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI