Gaymar Medi-Therm II: troubleshooting and error codes
The ServiceForge knowledge base holds 37 documented symptoms for the Gaymar Medi-Therm II, with 113 ranked causes, 138 diagnostic checks with test points and expected values, and 13 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Blanket won't heat in AUTO or MANUAL mode
Blanket temperature does not reach setpoint when heating is activated; temperature remains below target
Ranked causes
- 1. Pump not running (common)
- 2. Inadequate water flow (common)
- 3. Cold solenoid SV1 stuck open or hot solenoid SV2 stuck closed (occasional)
- 4. Hot solenoid SV2 not energized (occasional)
- 5. Hot solenoid SV2 coil shorted (occasional)
- 6. Heater element open or shorted (occasional)
- 7. Heater not energized (occasional)
- 8. Pin 7 of U51 not at 0 VDC relative to pin 10 (rare)
- 9. Pin 11 of U51 not at 0 VDC relative to pin 10 (rare)
- 10. Optocoupler U1 input not pulled low (rare)
- 11. Optocoupler U7 input not pulled low (rare)
- 12. Optocoupler U8 input not pulled low (rare)
- 13. No line voltage at Q3 output (rare)
- 14. No line voltage at Q8 output (rare)
- 15. No line voltage at Q9 output (rare)
- 16. No line voltage at TB1 terminals 1 and 3 (rare)
- 17. No line voltage at TB1 terminals 2 and 3 (rare)
- 18. No line voltage at TB1 terminals 11 and 8 (rare)
- 19. Temperature sensor RT2 inaccurate (rare)
Verify pump is running
- Test point
- Pump motor
- Expected
- Pump running
11 more checks with pass and fail actions, 21 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Water Flow alert is on
Water flow alarm activated; blanket temperature does not reach setpoint
Ranked causes
- 1. Hoses not connected properly (common)
- 2. Blankets, hoses or connectors pinched or occluded (common)
- 3. Blankets or hoses excessively dirty (occasional)
- 4. Pump motor not running (common)
- 5. Flow switch S5 failure (occasional)
- 6. Cold solenoid SV1 shorted (occasional)
- 7. Hot solenoid SV2 shorted (occasional)
- 8. Impeller broken or cracked (occasional)
- 9. Control/display board failure (rare)
- 10. Board-to-board connection failure (rare)
- 11. Base-to-power supply board connection failure (rare)
- 12. Q8 or Q9 triac failure (rare)
- 13. Solenoid coil not energized (rare)
Verify hose connections
- Test point
- SUPPLY and RETURN ports
- Expected
- One hose connected to SUPPLY, one to RETURN
11 more checks with pass and fail actions, 16 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blanket will not cool high voltage
Refrigeration system faulty; blanket display temperature not within 1.8°F (1°C) of setpoint or reservoir cannot reach 42.8°F (6°C)
Ranked causes
- 1. Cold solenoid SV1 not energized or stuck closed (common)
- 2. Dirty or restricted refrigeration condenser (common)
- 3. Refrigeration unit may have lost its charge (occasional)
- 4. Hot solenoid SV2 stuck open or control/display board failure (occasional)
- 5. Pump not running or inadequate flow (common)
- 6. Compressor not running or thermal overload tripped (occasional)
- 7. Power supply board or relay failure (occasional)
- 8. Reservoir temperature probe RT1 failure or control/display board failure (occasional)
Check if steel object is attracted to center top of valve (magnetic test)
- Test point
- Valve center top
- Expected
- Yes = solenoid working; No = solenoid may be faulty
26 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump motor not running
Pump does not start or run; no fluid circulation detected
Ranked causes
- 1. Unit not in AUTO or MANUAL mode (common)
- 2. Pump winding shorted (occasional)
- 3. Optocoupler U2 input not pulled low (occasional)
- 4. No line voltage at Q4 output (occasional)
- 5. Pin 8 of U51 not at 0 VDC relative to pin 10 (rare)
- 6. Board-to-board connection failure (rare)
- 7. Base-to-power supply board connection failure (rare)
- 8. No line voltage at TB1 terminals 6 and 8 (rare)
Verify operating mode setting
- Test point
- Unit mode selector
- Expected
- Unit in AUTO or MANUAL mode
5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 1: Temperature sensor reading out of range
Machine displays RFU Code 1 in Service Mode 1, indicating measured values from RT1 or RT2 are not within 32°F to 122°F (0°C to 50°C) range.
Ranked causes
- 1. RT1 sensor faulty (common)
- 2. RT2 sensor faulty (common)
- 3. Faulty signal lines between control/display board and RT1 (occasional)
- 4. Faulty signal lines between control/display board and RT2 (occasional)
- 5. Control/display board temperature measurement circuitry failure (occasional)
- 6. Controller failure (rare)
Measure RT1 actual temperature
- Test point
- RT1 sensor
- Expected
- Between 32°F and 122°F (0°C and 50°C)
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 8: Blanket probe reading faulty
Machine displays RFU Code 8 in Service Mode 1, indicating the blanket probe (RT2) reading is faulty.
Ranked causes
- 1. RT2 sensor faulty (common)
- 2. Faulty signal lines between control/display board and RT2 (occasional)
- 3. Control/display board temperature measurement circuitry failure (occasional)
- 4. Controller failure caused RT2 reading outside acceptable range (rare)
Disconnect RT2 connector and measure signal line resistance
- Test point
- Pins 3 and 5 of connector P6
- Expected
- Between 811 and 7355 ohms
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 9: Reservoir probe reading faulty
Machine displays RFU Code 9 in Service Mode 1, indicating the reservoir probe (RT1) reading is faulty.
Ranked causes
- 1. RT1 sensor faulty (common)
- 2. Faulty signal lines between control/display board and RT1 (occasional)
- 3. Control/display board temperature measurement circuitry failure (occasional)
- 4. Controller failure caused RT1 reading outside acceptable range (rare)
Disconnect RT1 connector and measure signal line resistance
- Test point
- Pins 2 and 9 of connector P6
- Expected
- Between 811 and 7355 ohms
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Algae growth in fluid system caution
Cloudiness, discoloration, or biological growth visible in water circulation system or blankets; reduced flow rate or foul odor from system.
Ranked causes
- 1. Tap water use or infrequent water changes (common)
- 2. Excessive algaecide concentration (occasional)
- 3. Use of bleach in cleaning (rare)
Inspect water clarity and system appearance
- Test point
- Fluid reservoir and blanket tubes
- Expected
- Clear distilled water, no visible growth or discoloration
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chassis condition degraded or corroded caution
Rust, corrosion, or discoloration visible on chassis exterior; exterior screws loose or corroded; casters not functioning smoothly; visible dirt or debris in ventilation areas.
Ranked causes
- 1. Moisture exposure or environmental corrosion over time (occasional)
- 2. Caster wheels damaged or locked (occasional)
- 3. Accumulated dirt in ventilation vents and cooling fins (common)
Visually inspect chassis for corrosion and discoloration
- Test point
- External chassis surfaces and frame
- Expected
- Clean, minimal rust or corrosion; no paint peeling or heavy discoloration
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate below minimum specification patient safety
Flowmeter reading lower than specified minimum; machine unable to deliver adequate circulating flow during heating or cooling mode; slow temperature response.
Ranked causes
- 1. Air trapped in pump system after complete drain (occasional)
- 2. Blocked or restricted flow path in blanket connector hose (occasional)
- 3. Pump cavitation or bearing wear (rare)
Fill machine water reservoir to proper level
- Test point
- Reservoir and float stem green band indicator
- Expected
- Green band on float stem fully visible; machine filled to operating level
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code H: Microprocessor received shutdown signal from base
Machine displays RFU Code H, indicating a shutdown signal was received from the base thermostats (S1, S2, S3).
Ranked causes
- 1. S1 (low temperature thermostat) open or faulty (common)
- 2. S2 or S3 (high temperature thermostat) open or faulty (common)
- 3. Power supply board REMOVE FROM USE NOW/MACHINE SHUTDOWN circuitry failure (occasional)
Check thermostat continuity
- Test point
- TB1 terminals 10 and 8
- Expected
- 0 ohms (closed)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Attachment plug damaged or deteriorated high voltage
Power plug shows visible wear, cracks, or discoloration; pins corroded or pitted; difficult to insert into outlet; loose connection with cord.
Ranked causes
- 1. Age-related deterioration or prior damage to plug (occasional)
- 2. Corrosion on plug pins or contacts (occasional)
Inspect attachment plug for physical damage and corrosion
- Test point
- Power cord plug and pins
- Expected
- No cracks, discoloration, or corrosion; pins intact and shiny; plug seats firmly in outlet without force
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AUTO mode blanket water exceeds 42°C limit with simulated patient temperature patient safety
During AUTO mode testing with simulated 37°C patient probe resistance, machine allows blanket water to exceed 42°C setpoint limit; machine does not properly limit blanket water temperature in AUTO mode.
Ranked causes
- 1. Blanket water limit controller malfunction in AUTO mode (occasional)
- 2. Patient probe simulator resistance incorrect or connection unstable (occasional)
Prepare machine for AUTO mode test
- Test point
- Machine water fill level and mode selector
- Expected
- Machine filled to green band level; ready to enter AUTO mode
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blanket water temperature not matching display or out of tolerance patient safety
Display shows incorrect temperature reading compared to actual water temperature measured with external thermometer; temperature reading on display is unstable or fluctuates excessively.
Ranked causes
- 1. Temperature sensor out of calibration or malfunction (occasional)
- 2. Loose or corroded sensor wiring connection (occasional)
Connect test equipment and set initial temperature setpoint
- Test point
- Test setup per figure 13 with TPT9 tester and mercury thermometer
- Expected
- Setup completed; machine in MANUAL mode; setpoint set per INSPECTION FORM item i
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: High temperature backup thermostat trip points patient safety
Measure the actuation temperatures at which the machine's redundant safety thermostats S2 and S3 shut down the system to verify they operate within specification for the model.
Ranked causes
- 1. Thermostat trip point verification required (common)
- 2. Thermostat malfunction or independent shutdown failure (rare)
Fill machine with distilled water
- Test point
- float stem green band
- Expected
- green band fully visible
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit breaker worn or intermittent caution
Machine shuts off unexpectedly or repeatedly during operation; no error code displayed; cycling circuit breaker on and off does not restore operation reliably; no other apparent failure symptoms.
Ranked causes
- 1. Worn circuit breaker contacts or loose terminals (common)
- 2. Loose circuit breaker terminals (occasional)
Inspect circuit breaker physical condition and terminals
- Test point
- Circuit breaker and push-on terminals at rear
- Expected
- No discoloration, no heating marks, terminals fully seated and snug
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cold water reservoir controller not maintaining setpoint temperature patient safety
Machine unable to cool water to setpoint temperature; compressor does not cycle off as expected; water temperature drifts during COOL mode operation.
Ranked causes
- 1. Compressor not cycling off at target setpoint (occasional)
- 2. Thermostat sensor malfunction or loose connection (rare)
Fill machine to operating level and connect test equipment
- Test point
- Water reservoir and TPT9 test equipment connections
- Expected
- Green band on float stem fully visible; test setup per figure 13 with return line connected to TPT9 tester (blanket optional for speed)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Level switch not actuating at proper water level patient safety
ADD WATER indicator remains off when fluid level is low; indicator does not turn off when water is refilled; bobber float mechanism appears stuck or unresponsive.
Ranked causes
- 1. Bobber float stuck or not seating properly on drain screen (occasional)
- 2. ADD WATER switch contact corrosion or loose wiring (occasional)
Fill machine to proper operating level
- Test point
- Water reservoir and float stem green band
- Expected
- Green band on float stem fully visible; ADD WATER indicator off
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line cord or strain relief damaged high voltage
Visible cuts, cracks, or insulation damage on power cord; burn marks or melting at strain relief ends; intermittent power loss or arcing sounds.
Ranked causes
- 1. Physical damage to line cord or strain relief during use or storage (occasional)
- 2. Poor strain relief condition or improper seating (occasional)
Visually inspect entire line cord for damage
- Test point
- Line cord from plug to machine connector
- Expected
- No cuts, cracks, or discoloration; insulation intact; no exposed wires
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 4: Compensation resistor 1 reading too high
Machine displays RFU Code 4, indicating compensation resistor R12 on control/display board reading is above specification.
Ranked causes
- 1. R12 compensation resistor out of specification (common)
- 2. Control/display board temperature measurement circuitry failure (occasional)
Measure R12 resistance
- Test point
- R12 on control/display board
- Expected
- 816 ohms ±0.1%, ±0.1%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 5: Compensation resistor 1 reading too low
Machine displays RFU Code 5, indicating compensation resistor R12 on control/display board reading is below specification.
Ranked causes
- 1. R12 compensation resistor out of specification (common)
- 2. Control/display board temperature measurement circuitry failure (occasional)
Measure R12 resistance
- Test point
- R12 on control/display board
- Expected
- 816 ohms ±0.1%, ±0.1%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 6: Compensation resistor 2 reading too high
Machine displays RFU Code 6, indicating compensation resistor R13 on control/display board reading is above specification.
Ranked causes
- 1. R13 compensation resistor out of specification (common)
- 2. Control/display board temperature measurement circuitry failure (occasional)
Measure R13 resistance
- Test point
- R13 on control/display board
- Expected
- 7320 ohms ±0.1%, ±0.1%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFU Code 7: Compensation resistor 2 reading too low
Machine displays RFU Code 7, indicating compensation resistor R13 on control/display board reading is below specification.
Ranked causes
- 1. R13 compensation resistor out of specification (common)
- 2. Control/display board temperature measurement circuitry failure (occasional)
Measure R13 resistance
- Test point
- R13 on control/display board
- Expected
- 7320 ohms ±0.1%, ±0.1%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test lights or alarm indicators not functioning patient safety
LEDs on display panel do not illuminate; audible alarm does not sound during test; status legends or temperature displays remain dark when TEST LIGHTS button pressed.
Ranked causes
- 1. Burned-out LED or defective lamp (common)
- 2. Loose or corroded wiring connection to indicator lamps or speaker (occasional)
Perform TEST LIGHTS button function test
- Test point
- Display panel LEDs and audible alarm
- Expected
- Right and left halves light and blank alternately; FLOW OK, HEAT, IN-TEMP, COOL, ALERT with four triangular indicators, three mode legends and SELECT all lit; displays show '888', '888.8', '888.8'; temperature scale and setpoint lines lit; audible alarm sounds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current leakage chassis high voltage
Verify chassis current leakage does not exceed safety limits in all operating modes
Ranked causes
- 1. Current leakage exceeds limit (rare)
Measure leakage with machine grounded
- Test point
- Chassis to ground
- Expected
- Not to exceed 10 microamps, ≤ 10 microamps
- Tools
- Leakage current meter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current leakage patient probe patient safety
Verify patient probe current leakage does not exceed safety limits with foil wrapping simulation
Ranked causes
- 1. Patient probe current leakage exceeds limit (rare)
Wrap patient probe with metal foil
- Test point
- Patient probe entire length
- Expected
- Fully wrapped
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Grounding resistance high voltage
Verify chassis grounding resistance is within safety specification
Ranked causes
- 1. Grounding resistance exceeds specification (rare)
Measure grounding resistance with ohmmeter
- Test point
- Grounding pin on line cord attachment plug to exposed chassis metal
- Expected
- Less than 0.5 ohms, < 0.5 ohms
- Tools
- Ohmmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Low temperature backup thermostat (S1) caution
Verify trip point of low temperature backup thermostat to protect against freezing
Ranked causes
- 1. Thermostat trip point out of specification (unknown)
Drain machine completely
- Test point
- Reservoir and circuit
- Expected
- Empty
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Patient temperature display
Verify patient temperature display accuracy using resistive simulation of probe values
Ranked causes
- 1. Display not within tolerance (unknown)
Fill reservoir with distilled water
- Test point
- Float stem green band
- Expected
- Fully visible
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reset RFU code
Reset stored REMOVE FROM USE code to zero before returning machine to service
Ranked causes
- 1. Previous RFU code retained in memory (occasional)
Enter service mode 1
- Test point
- Control panel service button
- Expected
- Service mode 1 active; displays RFU code and software version
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Thermostat verification patient safety
Verify that each backup thermostat (S2 and S3) independently causes machine shutdown to confirm redundant safety circuit operation.
Ranked causes
- 1. Safety thermostat functional verification required (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor shipping brace not removed before operation patient safety
Machine has not been placed into service after delivery; compressor contains restraint brace to prevent damage during shipping.
Ranked causes
- 1. Lack of awareness of shipping brace requirement or procedure (common)
Verify compressor shipping brace has been removed
- Test point
- Compressor area per figure B on page 86
- Expected
- No shipping brace present; compressor mounting secure and in normal operating position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor unit may have lost refrigerant charge high voltage
Low pressure on refrigeration circuit; compressor cannot cool adequately.
Ranked causes
- 1. Refrigerant leak in system (common)
Visually inspect all refrigeration lines, fittings, service valves, compressor, and condenser for oil residue, corrosion, or damage
- Test point
- Entire refrigerant circuit
- Expected
- No visible signs of leakage or damage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Concealed damage upon receipt caution
Physical damage to machine visible after unpacking; cracks, dents, or broken components observable during initial inspection.
Ranked causes
- 1. Shipping damage during transport (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump motor bearing wear or noise caution
Squealing, grinding, or rattling sounds from pump; motor does not rotate smoothly or pump fails to circulate water.
Ranked causes
- 1. Dry pump motor bearings due to insufficient lubrication (common)
Inspect pump motor for oil ports and lubrication level
- Test point
- Pump motor label locations
- Expected
- Visible oil ports per label; no signs of drying or rust
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quick-disconnect fittings stiff or difficult to engage caution
Connector hose will not attach smoothly to machine fittings; excessive force required to engage quick-disconnects; potential leak risk at connection points.
Ranked causes
- 1. Lack of lubrication on quick-disconnect fittings (occasional)
Test quick-disconnect engagement resistance
- Test point
- Machine supply and return fittings and blanket connectors
- Expected
- Smooth engagement with moderate hand pressure, no grinding or resistance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduced compressor efficiency caution
Compressor runs longer than normal, machine takes extended time to reach setpoint temperature, or insufficient cooling/heating capability.
Ranked causes
- 1. Dirt accumulation on condenser coils and cooling fins (common)
Inspect condenser coils and cooling fins for dirt accumulation
- Test point
- Rear interior of machine near compressor
- Expected
- Clean coils and fins free of visible dirt and debris
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
RFU [–] | CHECKSUM failure | Software checksum validation failure | error |
RFU 1 | Temperature sensor reading out of range | RT1 or RT2 measured values outside 32°F to 122°F (0°C to 50°C) range | error |
RFU 2 | Conversion failure | A/D converter or temperature conversion circuit malfunction | error |
RFU 3 | Conversion failure | Temperature conversion circuit malfunction | error |
RFU 4 | Compensation resistor 1 reading too high | R12 on control/display board resistance above specification | error |
RFU 5 | Compensation resistor 1 reading too low | R12 on control/display board resistance below specification | error |
RFU 6 | Compensation resistor 2 reading too high | R13 on control/display board resistance above specification | error |
RFU 7 | Compensation resistor 2 reading too low | R13 on control/display board resistance below specification | error |
RFU 8 | Blanket probe reading faulty | RT2 sensor or signal integrity failure | error |
RFU 9 | Reservoir probe reading faulty | RT1 sensor or signal integrity failure | error |
RFU E | RAMTEST failure | Random access memory self-test failure | error |
RFU H | Microprocessor received shutdown signal from base | S1, S2, or S3 thermostat opened or power supply board shutdown circuit activated | error |
RFU L | RAMTEST failure | Random access memory self-test failure | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.