Fujifilm X-Porte: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the Fujifilm X-Porte, with 41 ranked causes, 11 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No audio output
System produces no sound; audio controls accessible via Touch Control Panel or system mouse
Ranked causes
- 1. Audio muted or volume too low (common)
- 2. HDMI cable connection issue (common)
- 3. Defective HDMI cable (occasional)
- 4. Stalk Cable Assembly failure (occasional)
- 5. Clinical Display Monitor failure (occasional)
- 6. Dock failure (occasional)
- 7. X-Porte Engine failure (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ultrasound image but screen-saver icon present
Clinical display shows screen-saver icon only; no ultrasound image visible; stand version system
Ranked causes
- 1. Stalk Cable Assembly connection issue (common)
- 2. Dock failure (occasional)
- 3. Clinical Display Monitor failure (occasional)
- 4. X-Porte Engine failure (occasional)
Check Stalk Cable connections
- Test point
- Dock to Clinical Display Monitor
- Expected
- Connections secure
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Assert Code 19325 with istouchinfopresent() error
Touch Control Panel not detected or communicating with system
Ranked causes
- 1. Touch Control Panel cable disconnected from Dock port (common)
- 2. Touch Control Panel Assy defective (occasional)
- 3. Dock Assy defective (rare)
Verify Touch Control Panel cable connection to Dock port
- Test point
- Dock port connector
- Expected
- Cable firmly seated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blooming, jumpy, or jittery graphic display
Display image instability or distortion affecting video quality
Ranked causes
- 1. TTC Assy not properly installed in Stand channel (common)
- 2. Touch Control Panel cable disconnected from Dock Assy (occasional)
- 3. Touch Control Panel Assy defective (occasional)
Verify TTC Assy all four roller guides are in Stand channel holes
- Test point
- TTC roller guides in Stand channels
- Expected
- All four roller guides seated in channel holes
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image background is purple, red, or green
Clinical display shows abnormal color cast in image background
Ranked causes
- 1. HDMI cable connection issue (common)
- 2. Stalk Cable Assembly failure (occasional)
- 3. Clinical Display Monitor failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ultrasound image on desktop system screen-saver icon present
Desktop system shows screen-saver icon only; no ultrasound image; other video sources work
Ranked causes
- 1. HDMI cable connection issue (common)
- 2. X-Porte Engine failure (occasional)
- 3. Clinical Display Monitor failure (occasional)
Check HDMI cable and test with other video source
- Test point
- X-Porte Engine HDMI port to Clinical Monitor Display
- Expected
- Monitor functions with other HDMI cable and video source
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Assert screen displayed
System displays assert screen instructing user to contact SonoSite Technical Support for hardware or software issues; assert code output to clinical display and touch panel
Ranked causes
- 1. Software assert fault (occasional)
- 2. Hardware assert fault (occasional)
Record assert code and attempt power cycle
- Test point
- Assert screen
- Expected
- System powers on normally without assert condition
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM socket communication failed
Error message indicates socket communication failure with Printer/Archiver; Tiller error code TSOCKET_CONNECT_FAILURE
Ranked causes
- 1. Invalid network configuration or wrong port number (common)
- 2. Printer offline or application not running (occasional)
Use Ping to verify connectivity
- Test point
- Printer/Archiver network address
- Expected
- Ping successful
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No displayed video image on clinical monitor
Clinical display monitor shows blank, black screen or no screen saver icon; stand version system
Ranked causes
- 1. Missing +24V supply from Dock to Clinical Display Monitor (common)
- 2. Clinical Display Monitor failure (occasional)
Check +24V supply voltage
- Test point
- Dock to Clinical Display Monitor connection
- Expected
- +24V present
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No displayed video image on desktop system
Desktop system clinical display shows blank or black screen with no screen-saver icon
Ranked causes
- 1. Missing +24V from external Power Supply (common)
- 2. Clinical Display Monitor failure (occasional)
Check +24V supply voltage
- Test point
- External Power Supply to Clinical Display Monitor
- Expected
- +24V present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPMU bottom Blue LED not illuminated
Solid Blue LED under stand base not visible or not illuminated; may indicate no +24V DC output
Ranked causes
- 1. Batteries not charged or no AC power (common)
- 2. SPMU failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch Control Panel is black or not functional
Display panel is dark or unresponsive to user input
Ranked causes
- 1. Touch Control Panel Assy defective (common)
- 2. Dock Assy defective (occasional)
Replace Touch Control Panel Assy and test functionality
- Test point
- Touch Control Panel display and response
- Expected
- Display illuminates and responds to touch input
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brighter background shading at image corners
Blooming effect visible at any corners of the ultrasound image
Ranked causes
- 1. Clinical Monitor Display defect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM Archiver transaction failed
Archiver transaction failed; Tiller error code TDICARCH_OPEN_FAILURE
Ranked causes
- 1. Wrong Capture Type selected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM internal failure detected
Network read failure; Tiller error code TDNETWORK_READ_FAILURE
Ranked causes
- 1. Invalid DICOM Attribute (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM network communication failed
Network communication failure; Tiller error code TDNETWORK_OPEN_FAILURE
Ranked causes
- 1. Device does not recognize X-Porte; rejects association (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM Printer transaction failed
Printer transaction failed; Tiller error code TDICPRNT_OPEN_FAILURE
Ranked causes
- 1. Wrong Image settings (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPMU Green LED flashing
Top-rear Green LED on stand base flashing at 1 Hz repetitive rate; system batteries are charging
Ranked causes
- 1. System batteries charging (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
16203 | PS1 PCBA in Engine | PS1 PCBA failure detected | critical |
17193 | PS1 PCBA in Engine | PS1 PCBA failure detected | critical |
18706 | BE PCBA in Engine | BE PCBA failure detected | critical |
19325 | istouchinfopresent() function failure | Touch Control Panel not detected | error |
19361 | BE PCBA in Engine | BE PCBA failure detected | critical |
19466 | Front-end not detected | Front-end not responding or not detected | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.