Fujifilm FN-PS551: troubleshooting and error codes
The ServiceForge knowledge base holds 3 documented symptoms for the Fujifilm FN-PS551, with 3 ranked causes, 8 diagnostic checks with test points and expected values, and 123 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: TCP/IP network configuration for GX150 control unit
Setting up IP address, subnet mask, and default gateway during Windows NT software installation on FN-PS551 control unit
Ranked causes
- 1. Initial network configuration required during software installation (common)
Verify [Specify an IP address] option is selected
- Test point
- [IP Address] tab in Microsoft TCP/IP Properties
- Expected
- [Specify an IP address] option selected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network driver installation and Service Pack update
Restart procedure after network driver installation and reapplication of Service Pack 5
Ranked causes
- 1. Service Control Manager window appears after network driver installation (common)
Observe Service Control Manager window during restart
- Test point
- System startup display
- Expected
- Service Control Manager window appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Utilities installation for OptiPlex GX150
Installing Dell system utilities including AutoShutdown Service from OptiPlex ResourceCD
Ranked causes
- 1. System Utilities installation required for GX150 Series control unit (common)
Launch Dell OptiPlex ResourceCD menu
- Test point
- CD-ROM drive
- Expected
- ResourceCD menu screen displayed
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
40 | WSA control error occurred | Network error during WSA initialization | warning |
61 | Received image lacks printing information | Received image does not contain enough information | warning |
62 | Invalid exposure conditions | Exposure conditions for received image are incorrect | warning |
63 | Invalid film size | Film size not designated correctly for received image | warning |
64 | Image size (rows) out of limitation | Image size rows for received image is incorrect | warning |
65 | Image size (columns) out of limitation | Image size columns for received image is incorrect | warning |
66 | Invalid allocated bit | Allocated bit for received image is incorrect | warning |
67 | Invalid stored bit | Stored bit for received image is incorrect | warning |
68 | Invalid high bit | High bit for received image is incorrect | warning |
69 | Overlay size (rows) out of limitation | Overlay size rows of received image is incorrect | warning |
70 | Overlay size (columns) out of limitation | Overlay size columns of received image is incorrect | warning |
71 | Overlay origin position out of limitation | Overlay origin position of received image is incorrect | warning |
72 | Invalid window information | Window processing of received image is incorrect | warning |
73 | Film base density invalid | Film base colour specification error | warning |
74 | Overlay data size error | Size of overlay data is invalid | warning |
75 | Image data size error | Size of image data is invalid | warning |
76 | Image information invalid | Image information is incorrect | warning |
77 | Window level processing error | Error occurred with image processing (window) | warning |
78 | TCP/IP communication fault | Error occurred with TCP/IP communication | warning |
79 | Memory allocation failed | Allocating memory failed | warning |
80 | Data transmission error | Error occurred in sending data | warning |
81 | Data reception error | Error occurred in receiving data | warning |
82 | Send data timeout | Timeout occurred in sending data | warning |
83 | Receive data timeout | Timeout occurred in receiving data | warning |
84 | Data reception aborted | Receiving data was aborted | warning |
85 | Data transmission aborted | Sending data was aborted | warning |
86 | Thread creation failed (TPS) | Error occurred with creating thread (TPS) | warning |
87 | Thread creation failed (TPR) | Error occurred with creating thread (TPR) | warning |
88 | Image data too long | Data size was too long and overflow buffer | warning |
89 | ABOR command received | ABOR command was sent | warning |
91 | Inter-frame transmission timeout | Timeout occurred with sending each images | warning |
92 | FTP timeout | Timeout occurred with FTP | warning |
93 | Image information file access error | Error occurred with accessing image information file | warning |
94 | Image data file access error | Error occurred with accessing image file | warning |
95 | Overlay file access error | Error occurred with accessing overlay file | warning |
99 | Unknown error occurred | Unknown error occurred | warning |
DICOM 1 | DICOM initialization failed - DPM invocation | Unable to launch the DICOM main thread | error |
DICOM 2 | DICOM initialization failed - invalid settings | The DICOM protocol settings are invalid | error |
DICOM 3 | DICOM initialization failed - DPC invocation | Unable to launch DICOM receive main thread | error |
DICOM 4 | DICOM initialization failed - insufficient memory | Memory insufficient during DICOM initialization | error |
EifPrDrv 1 | System call failed | A system call error occurred | error |
EifPrDrv 2 | SMCU driver error | An error occurred in SMCU driver operation | error |
EifPrDrv 3 | EIFIMG driver error | An error occurred in EIFIMG driver operation | error |
EifPrDrv 4 | Unexpected command received | An unsupported command was received from printer | error |
EifPrDrv 5 | Timer setup failed | An attempt to set timer settings failed | error |
ImagePr 1 | System call failed | A system call error occurred | error |
ImagePr 2 | System memory acquisition failed | An attempt to acquire system memory failed | error |
ImagePr 3 | Image processing object creation failed | An attempt to create image processing object failed | error |
ImagePr 4 | Image processing parameter retrieval failed | Attempt to obtain print job, film object, or parameters failed | error |
ImagePr 5 | File operation error | An attempt to create or open file failed | error |
ImagePr 6 | Image processing error | All attempts at image processing result in error | error |
PCPSMain-1 | PC-PS system information file access failed | System attempts to access system information file at startup | error |
PCPSMain-10 | Calling 'ExecPrint' method failed | Raw data transmission request fails | error |
PCPSMain-11 | Calling 'OnEndPrint' method failed | Job completed notification fails | error |
PCPSMain-12 | Calling 'AnsOnPrintRequest' method failed | Print request completed notification fails | error |
PCPSMain-13 | Calling 'SetJOBPriority' method failed | Change job priority request fails | error |
PCPSMain-14 | Calling 'DoSpool' method failed | Job spool request fails | error |
PCPSMain-15 | Calling 'AnsPrinterInfo' method failed | Printer information request answer fails | error |
PCPSMain-16 | Calling 'GetImgPrcJOB' method failed | Get image processing job request fails | error |
PCPSMain-17 | Calling 'GetRawDataJOB' method failed | Get raw data transmission job request fails | error |
PCPSMain-18 | Calling 'DeleteJOB' method failed | Delete request fails | error |
PCPSMain-19 | Calling 'PrinterStatusChange' method failed | Change in printer status notification fails | error |
PCPSMain-2 | Calling 'Run' method failed | Execute start request fails | error |
PCPSMain-20 | Calling 'ExecPrint' method failed | Raw data transmission request fails | error |
PCPSMain-21 | Calling 'AnsSetPrintPriority' method failed | Change in job priority completed fails | error |
PCPSMain-22 | Calling 'GetPrintJOBCount' method failed | Number of PrintJob acquisition request fails | error |
PCPSMain-23 | System setting is invalid | System information file not set correctly | error |
PCPSMain-24 | Printer information is not found | Printer information cannot be found for designated printer | error |
PCPSMain-25 | Calling 'ExeImageProcess' method failed | Image processing request fails | error |
PCPSMain-26 | Hard disk is full | No open space on disk | warning |
PCPSMain-27 | Calling 'PrintJOBMake' method failed | PrintJob creation request fails | error |
PCPSMain-28 | File copy error | File copying fails | error |
PCPSMain-3 | Calling 'RestorePrintJOB' method failed | Restore unprocessed job request fails | error |
PCPSMain-4 | Calling 'SetFilmEndStatus' method failed | Film output complete setting request fails | error |
PCPSMain-5 | Calling 'DoPrinterStatusChanged' method failed | Change printer status method call fails | error |
PCPSMain-6 | Calling 'SetRawDataEndStatus' method failed | Raw data transmission completed set request fails | error |
PCPSMain-7 | Calling 'SetImgPrcEndStatus' method failed | Image processing completed setting request fails | error |
PCPSMain-8 | Error return on 'EndRestorePrintJOB' event | Completion of unprocessed job restoration returns error | error |
PCPSMain-9 | Calling 'AnsPrintCheck' method failed | Print request availability check answer fails | error |
ServiceUty 1 | Service Utility startup | Service Utility started up | info |
Spooler 1 | Thread creation failed | Thread creation failed (internal error) | error |
Spooler 10 | Raw file deletion failed | An attempt to delete raw data file failed | warning |
Spooler 11 | Invalid FN-PS status | The status of FN-PS itself is invalid | warning |
Spooler 12 | PrintJob access failed | PrintJob dispatch pointer invalid | warning |
Spooler 13 | Film Object Number invalid | The designated film number is invalid | warning |
Spooler 14 | PrintJob not found for deletion | PrintJob nominated in queue delete request not found | warning |
Spooler 15 | Printer information not found | Printer information for nominated printer not found | warning |
Spooler 16 | Change priority invalid | The designated priority change is not valid | warning |
Spooler 2 | Message transmission failed | Message transmission failed (internal error) | warning |
Spooler 3 | Insufficient memory | An attempt to obtain memory failed (internal error) | warning |
Spooler 4 | PrintJob dispatch creation failed | PrintJob object creation when unprocessed job restored failed | warning |
Spooler 5 | PrintJob file read failed | An attempt to read print job file failed | warning |
Spooler 6 | Item size invalid | The data size for print job or film object is invalid | warning |
Spooler 7 | Item data acquisition failed | Print job or film object data acquisition failed | warning |
Spooler 8 | Spooler queue registration failed | An attempt to register in spooler queue failed | warning |
Spooler 9 | Film Object invalid | Film object invalid or insufficient data | warning |
TPrc-1 | Hard disk is full | Not enough space on HDD | warning |
TPrc-10 | Image set ID is invalid | Message set ID in received image is incorrect | warning |
TPrc-11 | Image position ID is invalid | Page set ID for received image is incorrect | warning |
TPrc-12 | Deleting image was failed | Attempt to delete temporarily stored image failed | warning |
TPrc-14 | Allocating memory was failed | Attempt to allocate memory failed | warning |
TPrc-15 | Error occurred with creating thread(TTx) | Attempt to launch a sub-process failed | warning |
TPrc-16 | Error occurred with creating Print JOB | Attempt to create a print request failed | warning |
TPrc-17 | Error occurred with saving Print JOB in Hard Disk | Unable to save print request to HDD | warning |
TPrc-18 | Error occurred with requesting to print | Unable to generate print command | warning |
TPrc-19 | Error occurred with creating Film Object | Unable to create film information | warning |
TPrc-2 | Hard disk is full | HDD is full | warning |
TPrc-20 | Error occurred with saving Film Object to Hard Disk | Unable to save film information to HDD | warning |
TPrc-21 | No entry in configuration file | No settings file present | warning |
TPrc-3 | Print request could not accepted (shutdown process in progress) | Data transmitted from other device during shutdown | warning |
TPrc-31 | My host name could not be resolved | Unable to acquire host name | warning |
TPrc-32 | Network information could not be resolved | Unable to acquire network information | warning |
TPrc-33 | Error occurred in socket level | A socket error occurred | warning |
TPrc-34 | Error occurred in socket option | A socket option error occurred | warning |
TPrc-35 | Error occurred in Bind state | A bind error occurred | warning |
TPrc-36 | Error occurred in Listen state | A listen error occurred | warning |
TPrc-37 | Error occurred in Accept state | An accept error occurred | warning |
TPrc-38 | Error occurred with creating thread | A fork error occurred in a sub-process | warning |
TPrc-39 | Error occurred in socket control | A socket control error occurred | warning |
TPrc-4 | There is no film | Data received but no film in FL-IM | warning |
TPrc-5 | There is no chemical | Data received but no chemical in FL-IM | warning |
TPrc-6 | Laser Printer is not available | Data received but FL-IM is temporarily disabled | warning |
TPrc-7 | Laser Printer is not available | Data received but FL-IM is unusable due to fatal error | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI