Fujifilm FCR: troubleshooting and error codes

The ServiceForge knowledge base holds 24 documented symptoms for the Fujifilm FCR, with 27 ranked causes, 60 diagnostic checks with test points and expected values, and 539 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

QA-WS application installation fails or does not start

Installation errors occur or application fails to launch after installation, possibly with key registry errors on startup.

Ranked causes

  1. 1. Key registry not installed during initial setup (common)
  2. 2. Incorrect password entered during QA-WS installation (occasional)
  3. 3. DICOM key installed without reinstalling AP key (occasional)
First check

Verify key registry installation

Test point
Application startup screen
Expected
Application launches without key registry error message

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

QA-WS application not fully functional after installation

Application installed but DICOM or image functions unavailable or restricted.

Ranked causes

  1. 1. US image parameters not loaded for English AP (USA) (occasional)
  2. 2. DICOM Storage or Print Service not enabled (occasional)
First check

Verify correct language and region variant of QA-WS installed

Test point
Application About or Help menu
Expected
English AP (USA) if US parameters required, or appropriate regional version

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CRT monitor artifact 1 (ringing, jitter, swimming, tailing)

Verify absence of visual artifacts including ringing, jitter, swimming, and tailing

Ranked causes

  1. 1. Artifact 1 verification procedure (unknown)
First check

Display Artifact 1 test pattern

Test point
Display Properties Background tab
Expected
Test pattern displayed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CRT monitor artifact 2 (inconsistency, flickering)

Verify absence of visual inconsistency and flickering artifacts

Ranked causes

  1. 1. Artifact 2 verification procedure (unknown)
First check

Display Artifact 2 test pattern

Test point
Display Properties Background tab
Expected
Test pattern displayed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CRT monitor chipping

Measure and verify no chipping exceeds 0.41 mm in center fluorescent face area

Ranked causes

  1. 1. CRT fluorescent face chipping inspection procedure (unknown)
First check

Inspect chipping in center fluorescent face area

Test point
Center of CRT fluorescent face (254 mm x 343 mm area)
Expected
No chipping above 0.41 mm, ≤

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CRT monitor luminance

Measure and set CRT maximum luminance and uniformity across screen

Ranked causes

  1. 1. Luminance verification and calibration procedure (unknown)
First check

Display test pattern (Brightness)

Test point
Display Properties Background tab
Expected
All white pattern visible

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CRT monitor resolution

Verify CRT display resolution meets specifications using test pattern

Ranked causes

  1. 1. Resolution verification procedure (unknown)
First check

Set luminance precondition (all white pattern)

Test point
Center of CRT screen
Expected
250±10 cd/m2, ±10

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor distortion and tilt

Correct pin-cushion distortion and tilt on 21" CRT monitor

Ranked causes

  1. 1. Display distortion and tilt calibration procedure (unknown)
First check

Correct pin-cushion distortion

Test point
Knob e
Expected
No visible barrel or pin-cushion distortion

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor horizontal and vertical position

Adjust horizontal and vertical display position on 21" CRT monitor

Ranked causes

  1. 1. Display position calibration procedure (unknown)
First check

Set horizontal position

Test point
Knob b
Expected
Display centered horizontally

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor horizontal and vertical size

Adjust horizontal and vertical display size on 21" CRT monitor

Ranked causes

  1. 1. Display size calibration procedure (unknown)
First check

Set horizontal size

Test point
Knob a
Expected
Display width appropriate

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CRT monitor screen chipping detected caution

Visual defects or chips visible on the fluorescent display surface during quality assurance testing

Ranked causes

  1. 1. Manufacturing or handling defect in CRT screen (unknown)
First check

Visually inspect CRT monitor screen for chipping using test pattern

Test point
Center area of fluorescent face, rectangular region 254 mm x 343 mm
Expected
No chipping above 0.41 mm on average, 0.41 mm maximum average
Tools
Test pattern display, visual inspection

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CRT monitor screen size or position incorrect

Monitor display dimensions or positioning does not meet specifications

Ranked causes

  1. 1. Monitor adjustment knobs not properly configured (common)
First check

Measure screen width (X dimension)

Test point
Horizontal dimension L1-L2
Expected
360 mm, ±1%

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delete items from COMP parameter maintenance screen

Remove selected items from the COMP prm-MT list box using double-click and delete dialog

Ranked causes

  1. 1. User selects item for deletion (unknown)
First check

Double-click item in list box

Test point
COMP prm-MT screen list box
Expected
Select Modification dialog appears

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Environmental settings lost after installing DICOM or AP key caution

System settings and configuration parameters not restored after key registry installation.

Ranked causes

  1. 1. Environmental settings not restored after key installation (common)
First check

Verify backup of environmental settings was created before key installation

Test point
System file directory or backup media
Expected
Backup files present from INSTALLATION section 7 procedure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Host name not specified or requires change

Computer network name not configured or needs modification

Ranked causes

  1. 1. Host name not set or requires update (unknown)
First check

Verify host name is set

Test point
Network dialog > Identification tab > Computer Name
Expected
Desired host name configured
Tools
Windows network settings

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image data lost after HDD replacement or Windows NT reinstall caution

Image data on D: drive becomes inaccessible or deleted when HDD is replaced or Windows NT is reinstalled.

Ranked causes

  1. 1. Image storage partition (D:) deleted during HDD or OS installation (common)
First check

Verify D: partition exists in Disk Administrator

Test point
Disk Administrator main window partition list
Expected
D: partition visible and formatted as NTFS

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor display menu does not appear

Menu fails to display after HDD access lamp goes OFF on 21" CRT monitor

Ranked causes

  1. 1. Monitor brightness and contrast require adjustment (common)
First check

Verify HDD access lamp is completely OFF

Test point
HDD access lamp
Expected
Lamp OFF

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network information not configured

Windows NT network settings not properly configured for system communication

Ranked causes

  1. 1. Network Identification and Protocols not configured (common)
First check

Verify network adapter installed

Test point
Network dialog > Adapters tab
Expected
3Com EtherLink XL NIC (3C905B-TX)
Tools
Windows network settings

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Quit COMP parameter maintenance application

Close ComPrm.exe application with unsaved changes prompt if needed

Ranked causes

  1. 1. Application closure requested (unknown)
First check

Click CLOSE button on COMP prm-MT screen

Test point
COMP prm-MT screen CLOSE button
Expected
Application closes or unsaved changes dialog appears

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Save changes in COMP parameter maintenance

Save modifications made in COMP prm-MT screen without quitting the application

Ranked causes

  1. 1. Changes made requiring save (unknown)
First check

Click SAVE button on COMP prm-MT screen

Test point
COMP prm-MT screen SAVE button
Expected
Changes saved without exiting application

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen saver not configured to prevent monitor burnout caution

CRT monitor screen saver setting not enabled, risking phosphor burnout during idle periods

Ranked causes

  1. 1. Screen saver disabled or set to None (common)
First check

Verify screen saver type selection

Test point
Display Properties > Screen Saver tab
Expected
Blank Screen
Tools
Windows display settings

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software upgrade to version A02 required

System requires upgrade from A00 or A01 to A02 version including software installation, file copy, registry setup, image file conversion, and database restructuring

Ranked causes

  1. 1. Software version upgrade needed (unknown)
First check

Insert installation CD-ROM into drive

Test point
CD-ROM drive
Expected
Setup screen appears

11 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System startup time not optimized

OS loader wait time at startup not configured for reduced boot delay

Ranked causes

  1. 1. OS Loader show list timeout set to default value (common)
First check

Verify OS Loader wait time setting

Test point
System Properties > Startup/Shutdown tab > Show list for field
Expected
5 seconds
Tools
Windows system settings

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

TCP/IP and IP address configuration incomplete

TCP/IP protocol and IP address settings not configured for network communication

Ranked causes

  1. 1. TCP/IP protocol not configured with IP address and subnet mask (common)
First check

Verify TCP/IP protocol selected and configured

Test point
Network dialog > Protocols tab > TCP/IP Properties
Expected
TCP/IP configured with valid IP address and subnet mask
Tools
Windows network settings

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0709Software application errorFailure during equipment management from softwareerror
0A09Application error – field portion failureOccurs in exceptional portion locationserror
0AD9Application error – error detection expected procedureWhen expected procedure components occurerror
0B09Application error – work token issueWhen work token errors currently occurerror
0C09Application error – initial portion issueWhen portion locations of image management occurerror
0CD9Executable process margin possibleMargin in executable process possibleerror
0D09Application error – equipment requirementWhen equipment requirement equipment get properly managederror
0D90Application error – equipment managementWhen equipment management failserror
0DD9The External error – error detected in operable deviceWhen error is detected in operable deviceerror
0E09Application error – required function managementWhen required function management equipment get properly managederror
0ED9The External error – transmission idleWhen transmission idle error occurserror
0F09Application error – management deviceWhen management device gets properly managederror
0FD9The External error – transmission idleWhen transmission idle error occurserror
0x0258Failure recovered, memory buffer network timeoutNormal structured processing could not be completederror
0x0359Failure to delivery, memory buffer network timeoutNetwork data delivery transmission failure during image processing inputerror
0x1152Time-out between various image managing commandsTimeout when multiple image receiving or transmitting commands occur during executionwarning
0x121FCheck other device status or reconnect external deviceCommand/network connection timeout to external network device during network receptionerror
0x1251Failure to start image transmission, image data timeoutTimeout of image data transmission start signal during image processing inputerror
0x135BTime-out during command reception network reception exceptionTimeout during command/network reception when other network device or external device communication occurserror
0x136FTime-out in device network reception exceptionTime-out during command reception in external network device with external device communication network connectionerror
0x1F52Socket function test failure on FCR R/I-F connection network functionWhen no response reception or error detection occurs during image input processing network reception when other network device communicates within E/I-F socket function connection networkerror
0x4317Failed reception exception (1) or failed trans-mission to SCM CPU from device interface (2)When unexpected reception error or transmission failure from SCM CPU or when device interface device driver receives error/unexpected reception or transmission failureerror
0x7258Failed severe receiver, memory buffer network timeoutNetwork buffer timeout during external device communication receiver function during image processing inputerror
0x7261unknown
0x8500Installation complete, ROM-DC key execution exceptionROM-DC key execution or ROM-DC key installation cannot startinfo
0x8501Command definition and file error in command and file definition errorFile definition error in command definition management or file name definition error in command managementerror
1709Software design error during sensor signal detectionError occurs when managing sensor signals during design and control logicerror
1901Software error on warming upDetected during system warm-uperror
1902FCR common error on system connectionSystem bus communication failureerror
1903System management control failureDetected during system operationerror
1904Control connection failure between systemsSystem bus communication errorerror
190AFailed warning for system maintenance requirementMaintenance interval exceededwarning
190BCommunication error between system bus devicesSystem bus connection failureerror
190CMessage display error from system messageSystem bus operation failureerror
190DEquipment required to system bus message retention failureEquipment failure or bus message retentionerror
190EDetected error or system bus initializationSystem bus item not functioning during operationerror
1910Failed warning for system initialization eventSystem bus initialization event not occurringwarning
1911Message display error from system busMessage display system bus removal or failureerror
1912Failed warning for maintenance or system operationSystem maintenance not completed or operation preventedwarning
1913Failed warning for system operation cancellationSystem operation cancellation not completewarning
1915Application of soft ware patch for fuse F3 openFuse F3 open condition detectederror
1916Application software write faultFault in applying application softwareerror
1916AMemory controller error in control panelRemote monitoring memory controller faulterror
1916BGMIFI-E setup sensor errorSetup sensor not detectederror
1916CRemote monitoring error in signal transmissionSignal transmission lost during control panel remote monitoringerror
1916DGMIFI-E pair remote monitoring errorPermanent remote monitoring connection losterror
1916EGMIFI-E wire errorWire damage or disconnection detectederror
1916FGMIFI-E validity error with data transmissionData not valid in transmission to FCR uniterror
1917GMIFI-E storage errorError detected in GMIFI-E storageerror
1917AFCR pipeline error detection (no exception/error)Error detected in FCR exception/pipeline during common initializationerror
1917EGMIFI-E motor error detection, common memory issueMotor monitoring in GMIFI-E encountered common memory errorerror
1917FGMIFI-E motor timing error on detectionMotor common memory check during common initialization fle-checkerror
1917OManagement error during FCR exchange operationMemory-related error in FCR memory managementerror
1918APipeline error exception detection for monitoring GMIFI-EError detection required during motor monitoring motor setuperror
1918BMotor timing error (EOT) common memory setupCommon memory check required during air sensor monitoring setuperror
1918CAir sensor error (EOT) common memory checkCommon memory check required during air sensor monitoring setuperror
1918DMonitoring error cannot occur, common memory errorMonitoring error during common memory check detector errorerror
1918EGMIFI-E common memory check error detectorCommon memory check in GMIFI-E during motor error detectorerror
1918FGMIFI-E motor timing error detectorMotor common memory check during common fle-check detectionerror
1918OMotor event not detected in setup procedureMemory initialization failure during common GMIFI-E checkerror
1919AFCR buffer error detectionFCR pipeline error detected during application software write faulterror
1919EGMIFI-E monitoring error during motor error detectionMotor error encountered common memory initialization fle-check detectionerror
1919FGMIFI-E motor timing error on detectionMotor common memory check during common fle-check detectionerror
1957Image queue put up to new queue requires repairImage queue management failureerror
1958Image queue timeout errorImage queue processing exceeds time limiterror
1959Image queue timeout errorImage queue processing exceeds time limiterror
195AImage revise semaines invalidImage revision parameter required for maintenanceerror
195BApplication software patch for scanner window registry setting in data statusRegistry setting failure during initializationerror
195CApplication software patch for window registry viewing in data statusRegistry window viewing failure during initializationerror
195DChannel Display invalidChannel display parameter errorerror
195EChannel status invalidChannel status parameter errorerror
195FApplication software patch for initialization message not contacting the other endpointInitialization message queue error during system startuperror
1960Application software patch for failure at parameter detect equipment elementEquipment parameter detection failure at equipment element levelerror
1961Application software patch for connection parameter memory element generation detect equipment elementMemory element generation detection failure during initializationerror
1962Application software patch for retriever and connection element reset initialization reservation serverInitialization reservation service failureerror
1963Image queue put up to new required file system management generator sub-bus management configuration require field equipment queue file receiving new only when image recovery failsFile system queue management failure during sub-bus management generationwarning
1964Image revise processing sub-bus subsystem (Waiting for Host to commence useful transmit of error code Host ports)Sub-bus processing error during host communicationerror
1965Application software patch for force across external transmit element external retain element (Waiting for Host to execute subsystem error at connection element and transmit management force required)Sub-bus system transmit element force error during host managementerror
19ADetected errorError condition detected during operationerror
19BDetected problem errorProblem detected during operationerror
19CSmmu exposure errorSCMU memory error on exposure attempterror
19C0Common condition identify errorFailed to identify common conditionerror
19DSCMU open errorSCMU failure to open or initializeerror
19D0FCR-DMS initialization errorWhen FCR-DMS device fails to initialize properlyerror
19D1FCR-DMS common communication pipeline errorWhen FCR-DMS device experiences pipeline initialization failureerror
19D2FCR-DMS protocol analysis errorWhen FCR-DMS device reports protocol or analysis failureerror
19D3FCR-DMS memory management errorWhen FCR-DMS device has insufficient memory capacityerror
19D4FCR-DMS header data errorWhen FCR-DMS device header data becomes corrupt or unavailableerror
19D5FCR-DMS data pipe performance errorWhen FCR-DMS device pipe parameter data becomes corrupt or unavailableerror
19D6FCR-DMS bearing mechanical errorWhen FCR-DMS bearing channel becomes damaged or misaligned during operationerror
19D9FCR-DMS image data management error from file lostWhen FCR-DMS device loses image data due to file error in working memory or corrupt property parameterserror
19DAFCR-DMS image data management convert errorWhen FCR-DMS device image data format conversion fails or patient data becomes corrupterror
19ESCMU reset errorSCMU reset sequence failederror
19E0Error detected in pixel data no linearityWhen FCR-DMS device pixel linearity validation fails during data acquisitionerror
19E1Error detected in image management convert data correctlyWhen FCR-DMS device fails to correctly convert imaging management data during processerror
19E2Error detected in device receiver malfunctionWhen FCR-DMS device receiver component fails during data acquisitionerror
19E3Error detected in receiver data error wiredWhen FCR-DMS device wired receiver component has data transmission errorerror
19E4Error detected in assembly malfunction wired in dataWhen FCR-DMS device wired assembly component fails during data transmissionerror
19E5Error detected in event send notificationWhen FCR-DMS device fails to send notification event during operationerror
19E8Error detected in event reset component testWhen FCR-DMS device event reset component fails verification testerror
19E9Error detected in receiver current output voltage supply acquisition in imaging managementWhen FCR-DMS device current output measurement fails in device receiver during imaging managementerror
19EAError detected in excess pixel size reduced in imaging managementWhen FCR-DMS device pixel size reduction fails during imaging process with data managementerror
19EBError detected in time-out completion when event receivesWhen FCR-DMS device receives time-out completion signal during event acquisitionerror
19ECError detected in timer selection removed itemWhen FCR-DMS device timer selection item is removed or deselectederror
19EDError detected in test timer element removedWhen FCR-DMS device test timer element component is deselected or removederror
19EEError detected in receiver warning event with awaitWhen FCR-DMS device receiver sends awaiting warning event during operationerror
19EFError detected in failed waiting event with timeoutWhen FCR-DMS device fails waiting event timeout during operation with file refresh on starterror
19FSCMU close errorSCMU closure operation failederror
19F0Error detected in exclusive pixel ID identification list exclusion in receiverWhen FCR-DMS device exclusive pixel list item is excluded during receiver configurationerror
19F1Error detected in element deleted in image elementWhen FCR-DMS device image element component is deleted or deselectederror
19F2Error in exclusive pixel item validation listWhen FCR-DMS exclusive pixel item validation fails during data acquisitionerror
19F3Error in base data add specific pixel identification validationWhen FCR-DMS device specific pixel validation fails during base data acquisitionerror
1D90The Transmission error – reserve device cancellationWhen transmission errors occur that allow reverse reserve transmissionerror
2709Storage management errorFailure during software management from equipmenterror
29C0Print DICOM data configuration errorSystem cannot properly configure DICOM print settingserror
29C1Printer connection lost or multiple mount requiredSystem loses connection to DICOM printer or needs multiple mount operationserror
29C2Error in print correction programPrint correction program fails or encounters configuration errorerror
29C4E-title formation errorSystem cannot properly form or configure E-title settingserror
29C5Different valve count in image and image gammasValve count mismatch detected between image and image gamma settingserror
29C6Error in print correction program due to ongoing conditionPrint correction program error occurs; one condition persists from prior imageserror
29C7Error in failure of image full functionSystem loses full function capability of image processingerror
29C8Image reference box sequence error; requires printer setupSystem has image reference box sequence issue; requires new DICOM printer setuperror
29C9Error in image gallery connection malfunctionSystem has connection issue with DICOM image gallery or DICOM moduleerror
29CACommon-class error; visual verification event neededSystem detects common-class error condition; visual verification and correction program needederror
29CBError in configuration function managementSystem cannot properly configure function management settingserror
29CCError in image full function; system message return requiredImage full function failure; system status return needederror
29CDError causes items missing from DICOM module outputSystem detects items missing from DICOM module; either cause or status issueerror
29CENo connection with DICOM module; disassembled connection statusSystem has no connection or disassembled connection with DICOM moduleerror
29CFError in XCO complete open function; DICOM module repair neededXCO complete open function error; DICOM module requires repair or exchangeerror
29D0Error disassembled from DICOM module backupSystem encounters error during DICOM module error disassembly from backuperror
2D90The Transmission error – transmission missingWhen transmission errors occur that allow checking when enablederror
3709Software completion errorFailure during software equipment management operationerror
4709Memory capacity errorMemory capacity exceeded or invalid during system operationerror
5709Data error or corrupt data baseFailure during data management software operationerror
5D90Error detected – parameter retentionWhen parameter retention error detection when enabled for parameter setup outputerror
6709Network transmission errorDevice network transmission failure or missing equipment serviceerror
7709Management error classFailure during management software operationerror
8028Common error in corrective actionOccurs between corrective actionsunknown
8029Error in initializing socketOccurs between corrective actionsunknown
802AConfirmation connecting filmsOccurs between corrective actions with diagnosis of filmsunknown
802BInstallation elevation variantCorrective error received in installation settingsunknown
803EProgram errorError in software application versionunknown
85DConnector field tool errorDuring field tool detectionerror
8601Film analysis process occurs filter file character notation maintenance; film filter backlash set during film initializationOccurs during film analysis process of filter file character notation; film character notation backlash set during film file initialization; may occur due to filter analysis process during film filter initialization file setuperror
8709Cancel elegant connection errorElegant connection cancel error during device service revisionerror
8C09Application errorOccurs when prerequisite conditions are meterror
8F89Software corruption - parameter errorWhen parameter files become corrupted or invalid capacity exceeds memory limiterror
9008Application software errorFirmware malfunction during operationerror
9018Application software errorFirmware malfunction during operationerror
9028Image processing output success errorDuring imaging output, on the other hand, user seeserror
9048Application software errorIncoming filter error in filter property replacement error (error being seen in filter property)error
9058Application software errorInternal filter error, maintenance start data missing imaging start data at input imaging system maintenance, confirm error mission missing start data in input imaging systemerror
9068Application software errorInternal filter error in data management, confirm imaging management error in input imaging system (data management error in imaging system (occurred error on input imaging equipment))error
9078Application software errorInterlocking internal filter error, revise receiver messaging in receiver for all exchange filter requirements (reviser required)error
9088Application software errorReceiver messaging system malfunction error, new worker required in worker availabilityerror
9098Application software errorInterlocking filter object error occurrence (Interlocking filter object error in filter object property)warning
90AApplication software errorError occuring to application software error code protocol (error due to filter mapper protocol mapping)error
90A8Application software errorInterlocking worker for filter mapper status/functionality required input mapper occurrence (reviser required)error
90BApplication software errorData driven from input imaging equipment restart (error running during input imaging equipment)warning
90B8Application software errorTransmission multiple filters detected (multiple filters detected in transmission error due to filter sensor in transmission error due to filter sensor transmission)warning
90C0Application software errorError occuring to application 'hand-held field-hand-bus-system-message's (error of acquisition for the hand-held field-hand-bus-system-message's right success access)error
90C1Application software errorError 'eraser base data success access' hand-held field-hand-bus-system (hand-held field-hand-bus-system (right success access) base data eraser during)error
90C2Application software errorError 'imaging get' field-hand-bus-system (image get field-hand-bus-system) (right success access) base dataerror
90C3Application software errorError occurrence 'router's imaging data' field-hand-bus-system (router's imaging data in imaging from field-hand-bus-system (right success access))error
90C4Application software errorError occurrence 'togs get' field-hand-bus-system (togs get data imaging transition in data transmit field-hand-bus-system (right success access) has base data)error
90C5Application software errorError occurrence 'togs finish' field-hand-bus-system (togs finish imaging from field-hand-bus-system (completing transition data) has base data)error
90C6Application software errorError 'get' imaging 'get' field-hand-bus-system (from field-hand-bus-system (right success access) has require imaging get data 'get')error
90C7Application software errorError 'delete' imaging 'deleted' field-hand-bus-system (delete field-hand-bus-system (has base data 'delete'))error
9114Extended examination/API application errorSystem maintenance requirement not receivederror
9115Common examination errorSystem failure in buserror
9117Cannot connect - apply application softwareMemory failure encounterederror
9118Cannot connect - apply application softwareChannel status of the functionerror
9119Cannot connect - apply application softwareReceiving by accessing messageerror
911ACannot connect - apply application softwareEquipment image initialization queueerror
911BFailed - apply application softwareWait without image equipment initialization queueerror
911DCannot connect - apply application softwareEquipment image deletion queueerror
911ECannot connect - apply application softwareEquipment image modification queueerror
911FImage processing errorSystem bus expectederror
911FCannot connect with connected host to which image is outputDTK returned error 'cannot connect to host'. Failed in request for connectionerror
9120Cannot connect - apply application softwareEquipment image get maintenance codeerror
9121Cannot connect - apply application softwareReceived in get memory generation into queueerror
9123Cannot connect - apply application softwareEquipment image initialization queueerror
9124Cannot connect - apply application softwareEquipment image open on file queueerror
9125Cannot connect - apply application softwareDuplicate file queue failure backup queue control open doorerror
9126Cannot connect - apply application softwareEquipment image still in file queueerror
9127Cannot connect - apply application softwareEquipment image still in file back queue in operationerror
9128Cannot connect - apply application softwareReceived in get memory generation without queueerror
9129Cannot connect - apply application softwareEquipment image written in data queueerror
912ACannot connect - apply application softwareEquipment image written in file back queue queueerror
912AImage data confirmation from imaging procedure in applicationImage data loss or invalid image data from imaging procedureerror
912BImage reception of multiple required image transmission dataImage transmission from imaging data errorerror
912BCannot connect - found errorEquipment error in found queue data queueerror
912CCannot connect - apply application softwareEquipment display status queue in operationerror
912CError connection with common notificationNotification error during equipment connectionwarning
912DCannot connect - apply application softwareEquipment correct position in operationerror
912DEquipment connection required for image detectionNo connection to equipment for image transmissionwarning
912EImage receiving reject cases with image managementImage receiving rejection during imaging managementerror
9139AApply power-on operation warningApplied power-on operationwarning
9139BApplied power-on operation warningApplied power-on operationwarning
9139CError detected in file closing during error occurring announcementFile close error during announcementerror
9139DError detected in file getting during announcementFile get error during announcementerror
9139EError detected in file getting during closing announcementFile get error during closing announcementerror
9139FApply power-on operation warning for file lock controlFile lock control error during power-onwarning
9140Apply power-on operation warning for file lock control (Unused/Expected announcement)File lock control error during power-onwarning
9141Apply power-on operation without property support. FINF/FINP detected image imaging in system.Image imaging detected without proper supportwarning
9142Apply power-on operation warning for component open fileComponent open file error during power-onwarning
9143Apply power-on operation warning for file open deletion contactFile open deletion error during power-onwarning
9144Apply power-on operation warning for file locking image deleteFile locking and image deletion error during power-onwarning
9144BApply power-on operation warning for one-time transmissions from device received coverageOne-time transmission data not received or coverage failurewarning
9144CApply power-on operation warning for file locking and image management control inhibitionFile lock and image control problem during power-onwarning
9144DApply power-on operation warning for image size getImage size determination error during power-onerror
9144EApply power-on operation warning for error detection during imaging transfer dataError detected during image transfer or imaging startup dataerror
9144FApply power-on operation warning (file lock) for file locking image managementFile lock control error during imaging managementwarning
9150Apply power-on operation warning (undetermined) for file lock control component registrationFile lock control component registration failure during power-onwarning
9151Apply power-on operation warning for file log mannerFile log management error during power-onwarning
9152Apply power-on operation warning for file log open fileFile log open file error during power-onwarning
9153Apply power-on operation warning for file log writing data the fileFile log writing data error during power-onwarning
9154Apply power-on operation warning for memory control operation encountering memory encounteredMemory insufficient operation error memory encounteredwarning
91F4Application software failure on FCR interfaceDetected during startup or operationerror
91F5Unexpected error detected in image data memoryDetected during image processing or deliveryerror
91F6Data management error in output memory windowDetected during image output or storage operationerror
91F7Data management error during FCR image remappingDetected when remapping with new data assignmenterror
91F8Common communication error in momentum memory bufferDetected during image acquisition or communicationerror
91FAApplication software failure or messaging sensor detectionSoftware or sensor messaging failure during startuperror
9207Receiver command exception or undefined requestInvalid or unrecognized command from receiverwarning
9208FCR event failed to activate or maintainEvent activation failure or event maintenance losswarning
920AReceiver requires connection or cannot connectReceiver cannot establish connection or connection lostwarning
920BEvent failure during maintenance or event absenceMaintenance event failed or event completely absentwarning
920CReceiver monitoring failure for motion or image deliveryMonitoring system failed during multi-image transmissionwarning
920DFCR-DMS station failure during scene or receive initiationStation communication failure or receive initiation failederror
920EImaging FCR-DMS transfer or receive imaging failureStation imaging transfer or scene receive failederror
9215FCR-DMS current status in application processDetected during FCR startup or command processinfo
9216Application of waxproofing for FCR processor commonizationData invalid during imaging procedureerror
9217Multiple image receiving required during imaging managementImage data transmission error or incomplete image seterror
9219Image exclusive setting to visible setting atlasImage exclusion conditions detectederror
9220FCR processor exclusivity of test settingsExclusive visibility in image testerror
9221FCR processor visual exclusivity test to visible setting atlasVisual exclusivity settings invalid or test point atlas settings require definitionerror
9222Connect alert contact between holder and content placementContact connection error when content placement changedwarning
9223FCR image receiver modification requiredImage modification required when FCR image receiver changedwarning
9224Equipment detection connection thresholdEquipment connection error detected during procedureerror
9225FCR image receiver with attention attachment when multiple transmit framesImage attachment error during multiple image frame transmissionwarning
9226Connect collation with memory capable validationMemory capability validation error detectederror
9231Occurrence in common accommodation phaseCommon accommodation request during imaging transmissionerror
9232Collation contact with memory capable validationMemory capability validation errorerror
9240RC maintenance component sensing required before RC initialization in formation.RC component sensing procedure initiatederror
9241RC maintenance component sensing required; starting RC maintenance in formation.RC diagnostic/maintenance procedure initiatederror
9242If it was not equipped to perform corpse decoding of RC maintenance component.RC decoding/initialization procedure initiatederror
9243If it was not equipped to perform corpse decoding of RC maintenance component.RC maintenance decoding operation initiatederror
9244If it was not obtainable the WA/IP reception in the error context.RC error detection/reporting procedure initiatederror
9245If it was not converted to obtain the RC maintenance error over interval.RC maintenance interval conversion procedure initiatederror
9246If it was not opened potential decoding failure in RC maintenance component.RC component opening/access procedure initiatederror
9247RC wired failure in maintenance of the RC component.RC wired component maintenance procedure initiatederror
9248If it was not closed RC maintenance in the RC component.RC component closing/completion procedure initiatederror
924AIf it was required the RC maintenance component by failure configuration error decision.RC failure configuration/initialization procedure initiatederror
924BIf it was not successful the occurrence of time item in RC maintenance.RC time interval maintenance procedure initiatederror
924CIf it was not with equipment when RC maintenance configuration initializing other hosts.RC maintenance with host equipment initialization procedure initiatederror
924DIf it was not containing error the RC maintenance in order to perform formation.RC formation/initialization procedure initiatederror
924EIf it was not equipped to RC maintenance of item sensing notification failure.RC item sensing/notification procedure initiatederror
9251If it was not obtainable the RC maintenance area from an error occurrence.RC area sensing/recognition procedure initiatederror
9252If it was not obtained the table basis access base field of hands analysis.RC data table access procedure initiatederror
9253If it was not obtained the RC store error in order of maintenance perform formation base data.RC error storage/retrieval procedure initiatederror
9254If it was not stores other RC error in order of store of all devices formation.RC error store synchronization procedure initiatederror
9255If it was not stores other RC error data of store of all devices formation.RC error data store synchronization procedure initiatederror
9256If it was not obtained the RC error of the data updating from data provided.RC data update/sync procedure initiatederror
9257If it was not obtainable the RC maintenance area from a file error occurrence.RC area file access procedure initiatederror
9258If it was not obtainable the RC maintenance area from a file error occurrence.RC area file access procedure initiatederror
9259If it was not obtainable the table basis of access field hands for the file.RC table file access procedure initiatederror
929AError occurred when not connected to HostConnection loss to Host systemerror
929A0Image modality failure detectedModality validation error in received imagewarning
929A1Image document code errorDocument code validation failedwarning
929A2FCR image ID error detectedImage ID validation failurewarning
929A3Workorder number missing or invalidWorkorder validation failedwarning
929A4Imaging module error detectedModule validation failureerror
929A5Detector sorbitol value errorDetector sorbitol validation failedwarning
929A6Image extended size errorImage size exceeds limit or invalidwarning
929A7Image transfer failed, cannot connectFile transfer connection failederror
929A8Image process code errorProcess code validation failed during transferwarning
929A9Scanning engine error detectedEngine validation failed during direction correctionerror
929AAPixel value error during transferPixel data validation failedwarning
929ABINB module retrieval failureINB module cannot be accessed or retrievederror
929AEElement A not specified in dataRequired element A missing from dataseterror
929AFAE-A title printer not detectedPrinter communication failed or absentwarning
929AFBFilm dryer exists with N configurationFilm dryer present but configuration incorrectwarning
929AFDFilm detected error during sessionFilm detection failed or unexpected conditionwarning
929AFEFilm dryer Box error on startupFilm dryer box initialization failederror
929AFFFilm extended size error on startupFilm size validation failed during initializationwarning
929BError occurred when connected to HostError condition during Host communicationerror
929CImage retrieval failed during chargingImage not received or retrieved during operationwarning
929DError occurred during image managementImage management operation failureerror
929EError received from device recoveryRecovery operation failedcritical
929FError validation failed during operationValidation check did not passwarning
929F3If it was not obtainable the WA/IP reception mechanism in order to perform diagnosis RC maintenance.Equipment diagnostic procedure initiatederror
9503External Communicator Sub System Detection Error (DICOM)Distribution destination node set to DICOM but DICOM support is OFF in equipment, or no protocol set at destination nodeerror
9556Invalid image data fieldConnected host attempts to send image with illegal field value (date, time, modality, anatomical region, FCR ID, rows, columns, storage bit, expansion read size, compression, assigned bit, effective bit, data size, PIXEL, LINE, or scanning direction)error
9562Notified error to connected hostSOP class VIP of image to be sent is not set for network for DICOM of this equipmentwarning
9601DICOM printer resizing in image does not exist or there is no image in DICOM printer display format processingImage information management in DICOM printer input information display format processing does not occur; image resizing function not found in printer input image format informationerror
9603Image value outstanding signal out other image signal it can reverse the output signal reverse the image valueThe image value it can range from display format DICOM printer display format processing DICOM printer input information management to range outstanding image signal out to signal outstanding outputerror
9604DICOM printer setup failed title-E.A. printer DICOM with specificationWhen DICOM printer (new.tnirP.MOCI.D) did not get the specified setup with specification with title-E.A.error
9605DICOM printer identifying key setting identification number INB is numerical; Please make configuration with DICOM Printer configuration file connection contents of configuration management file setup management apparatus configuration management file startup configuration management file setup application management error instruction setup from DICOM Films (new.stliF) to execute films setup application setup management operation configuration fileWhen DICOM Films (new.stliF) could not get the specified setup with configuration management file setup management apparatus configuration management fileerror
9607DICOM printer Films creatic failure - It failed to get session from DICOM printerOccurs when session transfer error occurs during DICOM printer operation or image management failure during DICOM printer session operation failureerror
9608Film processing error - COM image management failureImage management from COM box for film printing failserror
9609Film imaging management failureImage imaging system fails for film processor COM boxerror
960AFilm printing failure detectedFilm printer output operation failserror
960BCOM image management processing errorDigital COM image processing failserror
960CFilm image processing failure - Defective image in film numberingFilm printing system detects defective output imageerror
960DFilm E-A title designation errorFilm E-A title not found in destination device configuration fileerror
960EProgram error - existing numbered image failure in film numberingFilm system detects inconsistency in image version controlerror
960FProgram error - Film numbering failure in image managementFilm processing system fails on multiple image numbering or group operationserror
9610Program error - Film system version failureFilm system detects version control or configuration mismatcherror
9611Error occurrence in DCOM printer operationsError status detected before correction in DCOM printer operationerror
9612Program error - Multiple image numbering in film systemDCOM image management processing fails for common class image numberingerror
9613Program error in configuration management of DCOM printerError in DCOM printer configuration execution or runtime managementerror
9614Program error - Common class error in DCOM printer configurationError in DCOM printer configuration with event view functionerror
9615Program error in system operation managementSystem operational error detected in same operation conditionserror
9616Film output failure - No response from film processorFilm processor fails to respond to COM output request from DCOM boxerror
9617DCOM connection error - Connection not established with printerDCOM box fails to establish connection with printer during setuperror
9618Program error - COM XCP component failure in DCOMSystem detects error in COM component operationerror
9619Error feedback - Film system backup function failure (BSF-N CAUTION)DCOM printer backup function error from secondary film systemwarning
961AError feedback - Backup film system error (BBF-N ACTION)DCOM printer backup system error with ACTION requirementwarning
961BProgram error - Image sizing error with MAG_TYPE and multi-film supportDCOM printer image sizing or magnitude type processing failureerror
961CProgram error - Film string image processing failureFilm image processing analysis failure or string management error in film printererror
961DProgram error - Film mapping in bit-creation during film rotationError in film image mapping during bit-creation operation or film rotationerror
961EProgram error - Film assignment from film string in formatFilm format string management error or film string assignment failureerror
961FProgram error - Film analysis process failure on film string imageFilm image analysis process failure or error in film string configurationerror
9620Program error - Film register miss-set in film string registrationFilm registry failure or film string registration error during format operationerror
9621Program error - Film fail condition in cable drawing during rotationFilm cable failure or film failure during rotation operationerror
9622Program error - Warning status from BSF-N backup function in DCOM printerDCOM printer backup function warning from BSF-N designationwarning
9624Program error - Warning status BBF-N (TEST-BIN) from DCOM printerDCOM printer backup warning from BBF-N (TEST-BIN) with configuration supportwarning
9709Network errorEquipment management transmission error or FIN/FINP required network work througherror
9B10Item memory error detectedItem memory access or storage failureerror
9B11Item memory get errorFailed to retrieve item from memoryerror
9B13Item create errorFailed to create new item in memoryerror
9B14Item event errorItem event sequence errorerror
9B15Item stop errorItem stop signal failederror
9B16Device receiver data validation errorReceived data failed validation checkerror
9B17One-item system errorSingle-item system error detectederror
9B18Thereupon errorUnexpected operation occurrederror
9C09Image management errorImage management from 9C09 occurs in 09C9 formationserror
9F01Connected host rejected the connection requestDTK returned error 'other host rejected this request'. Failed in request for connection or transfererror
9F04Program error has occurred. Failed in DICOM file ID processingInternal error detection during DICOM file processingcritical
9F1FProgram error has occurred. Failed during external DICOM request processingError in data during previous DICOM image input processingcritical
9F50Program error has occurred. Failed in generation of DICOM threadingInternal error during DICOM input processingcritical
9F60Could not send one of the specified imagesCould not find information required for transfer or cue registration failederror
9F62Program error. Failed in DICOM data exchange, stream compilation, or image operationsNo AE setting in DICOM configuration. AE information not transferable. Temporary file cannot be deletederror
9F62Network obstacle in image output processing. Failed in DICOM output association or outputNo response from host to which image was input or output. DTK transfer text registration or connection request returned error during DICOM outputerror
A100unknown
A1F5unknown
A1F6unknown
A1F8unknown
A1F9unknown
A1FAunknown
A1FBunknown
A1FCunknown
B00Start-up timeout to process subsystem detectionImage Processing Subsystem does not complete startup within timeout perioderror
B02Insufficient memory recovery; equipment restart requiredOccurs when memory efficiency insufficient for Image Processing Subsystem startuperror
B03Image data files damaged; image data file deleted or corruptedThe image data files have been deleted with corrupted files detected during system operation or SysOp Data Management executionerror
B060Application software write errorDetected during setup managementerror
B066Memory reconnection errorDetected during setup managementerror
B0C9Image management damaged or file image management error occursDetected during data managementerror
B0CAFile image management error occursImage data corruption or system errorerror
B100unknown
B12EData revision device received image dataImage data retrieval or verification processwarning
B12FImage data entered into deviceImage data input processwarning
B130Image information from device data managementData retrieval or management failureerror
B131Image information from device dataData validation failureerror
B132Application software connection errorConnection detection during software operationerror
B133Application software connection errorConnection detection during software operationerror
B134Application software connection errorDetected during application operationerror
B135Application software connection errorConnection detection during operationerror
B136Application software connection errorConnection detection during operationerror
B137Specific memory steps errorMemory operation failureerror
B138Application software write errorDetected during operation when data corruption occurserror
B192The image processing error occursImage processing subsystem failure or bus-to-bus transaction failureerror
B1F5Application software write errorDetected in operating parameter seterror
B1F6Application software write errorDetected in operating parameter seterror
B200unknown
B259Application software write errorError detected in processor or core error during image processingerror
B2BDApplication software connection errorDetected operating error in OCXerror
B300unknown
B321Application software write errorImage management error detectederror
B323Unexpected volume firmware error detectedMemory detection during function diagnosiswarning
B400unknown
B709Configuration initial maintenance errorFailure from transmission maintenance configurationerror
B7CEApplication software write errorError detected in internal translationerror
C004Start-up timeout to process subsystem detection one-timeOccurs one-time or multiple times during Start-up timeout phaseerror
C004Invalid parameter value (1)When system detects invalid parameter during initializationerror
C005Invalid parameter value (2)When system detects invalid parameter during operationerror
C006Cannot connect to databaseWhen system fails to establish database connectionerror
C008Cannot modify databaseWhen system fails to modify database contentserror
C009Data database match failureWhen system detects data mismatch between database and image managementerror
C00ACannot connect to required databaseWhen system fails to establish connection to databaseerror
C00BNo access to parameter requisitesWhen system cannot access required parameter configurationerror
C00DData value is invalidWhen system detects invalid data valueerror
C00EMemory allocation failureWhen system fails to allocate sufficient memory capacityerror
C00FMultiple imaging managementWhen system processes multiple image management operations concurrentlywarning
C010Equipment exception record or image management image correlation issueWhen equipment detects processing errors or image management equipment management issue (1) Firmware test equipment management issue with new equipment, if initial firmware in management cache, return equipment (2) Restore equipment cache.critical
C011Field initialization socket corruption or image management issueWhen field management socket error detectederror
C012Equipment exception record or memory file or circuit errorWhen equipment detects file memory or circuit failureerror
C013Field failure built-in from management information formationWhen system detects field management or formation management failureerror
C014No capacity in hard disk or cache areaWhen system detects insufficient disk or cache capacityerror
C015Field failure built-in from management database contentsWhen system detects field management or database management failureerror
C-016Invalid imaging parameterImaging damaging parameter detected between imaging start and enderror
C100No access to memorySystem initialization on Windows startup; no success access casescritical
C101Inproper apparatus preparationIn-apparatus data setup performs data management information updatewarning
C200Field obtained handle on the windowMemory (window, device, ACT field catch value)warning
C202Failed recovery of image data files; image data file does not exist or has been deletedOccurs when specific image data file procedures fail during recovery attempterror
C300Unexpected field on data informationReplacement field on memory (file image management replacement mapping parameter procedures)critical
C404No space partition in hard disk or hard disk partition errors frequently occurHard disk partition becomes full or hard disk partition errors occur frequently; frequent occurrence indicates recurring hard disk failureserror
C709Setup storage errorSetup storage failureerror
D-001Image damaged or not validImage damaging parameter detected during imaging processerror
D-002File cannot be createdFile creation failure detectederror
D-003Invalid file setting detectedInvalid parameter detected in file settingserror
D-004Invalid parameter value detectedAny invalid parameter value found in configuration when occurrederror
D-005Key word invalidInvalid key word detected in imaging initialization from imaging functionerror
D-006File cannot be deleted or removedFile deletion or removal failure detectedwarning
D-008Parameter data memory errorParameter data memory error detected in window data initialization in systemerror
D-009Transfer entry errorTransfer entry failure to entry error detected during system initializationerror
D-00ATransfer entry to entry error during system resetTransfer entry error detected during system reset and initializationerror
D-00CFile position exception detectedFile position exception detected in file position during imaging or in CONF_HIGH_ICON positioningerror
D-00DFile position exception detected in file position during CONF_HIGH_ICON initializationFile position exception detected in file position during CONF_HIGH_ICON initializationerror
D-00EFile position exception detected in CONF_HIGH_ICON initializationFile position exception detected in CONF_HIGH_ICON initialization positioning componentserror
D-00FSET function exception detectedSET function exception error occurs when required. Anomaly occurs when occurred.error
D-010GET function exception detectedAnomaly occurs when required during system initialization testing.error
D-011FPGA GSIS function exception detected during process initializationAnomaly occurs when process error occurs in FPGA GSIS during initialization.error
D-012Calibration data validation errorCalibration data not found or invalid data validation error when occurrederror
D-013File read error detectedFile error detected in file read recoveryerror
D-014File write error detectedFile write error detected in file recoveryerror
D-015Direction error to cartridge signal acquisition on registrationImage damaging parameter detected to registration is a registration. Imaging registration signal requires to apply signal acquisition for an important signal acquisition.warning
D-016Direction cart ridge movement errorMovement error in direction cartridge movement in cartridge occurred. Anomaly occurs when movement direction cartridge direction error has been detected. Different expectation from error.warning
D-017Parameter map view errorParameter map view error in map view parameter map view error This will file setting occurs, perform error recovery procedures. File test on those files and repeat file test initialization.warning
D-018Parameter map view errorParameter map view error in map view parameter map view error This will file setting occurs, perform error recovery procedures. File test on those files and repeat file test initialization.warning
D-019In formation information requires from registration signal from the registrationAnomaly occurs when occurred. Anomaly occurs when occurred in formation information registration signal.warning
D-01BAny extra configuration detected in title ellipse there the there the extra configuration circle been detected in title exceptionAny extra configuration circle in title exception there the extra configuration circle been detected in title exception (defined time), Either extra configuration circle begun or title exception has begun.unknown
D050DProgram error - Rotation manager requirement in formation operationSystem requires rotation manager functionality in formation operationerror
D050DCircuit device defect - ENE file validationEnvironment file ENE designation validation failure in circuit deviceerror
D050DProgram error - Wiring code check with 16 or less number holdWiring code hex digit device circuit record validation failureerror
D050DProgram error - Identified determination value exceeding maximum numeration limitationIdentified numeration value exceeds maximum 129 determination in calibrationerror
D050DProgram error - Rotation manager identification in ENE fileRotation manager and manager number identification not found in formationerror
D067Overprogramming PI test point pin; PPI overlay address addition overlayDuring Field Tool arrangement overlay addition during Field Tool parameter arrangement (parameter overlay-add into field)warning
D068Overprogramming overlay hostname name attributeDuring Field Tool overlay addition during Field Tool parameter modification during Field Tool setup with other device settingswarning
D069No hostname attribute attached to pointing connection errorDuring Field Tool arrangement when attempting to attach hostname attribute to pointing connection errorwarning
D10Insufficient intensity (ITC) with 16-bit dataDetected during imaging acquisitionwarning
D1F5Apply configuration from recorded prior memory allocation stateDetected when prior memory allocation state not validwarning
D1F6Apply configuration error recovery from device receiverDetected when device error recovery from prior configuration mismatcherror
D1F7Apply configuration device coverage from API Window initializationDetected when Window API error recovery application received triggererror
D1F8Apply configuration error event data base data transmission FailureDetected error transmission failure in [Devices] and [Control Panel] statuserror
D1F9Initialize built-in initialization errors occurrence failureDetected data base file correctly occurred initializationwarning
D1FAApply configuration error recovery from device receiver capabilityDetected common receiver error (prior error management)warning
D1FBDetect parameter valid in valid parameter data in key dataDetected in key data valid parameter registered in BIOS datawarning
D20Electronic A units deficiencyCondition during calibration or dynamic mode operationwarning
D201Open configuration error file contents management cannot be opened in management fileFile management error configuration file cannot be opened during file management configuration execution in setup startup of equipmentwarning
D202Configuration file overflow table data management file number in file management dataWhen file management error occurs after configuration file contents error conditions occur after file contents management error conditionswarning
D203Inproper apparatus preparation circuit embedded Key rewording configuration managementWhenever Key rewording is different from description in file management file description (description difference) in file management file in apparatus startup equipment management setupwarning
D21Gain arrangement correction needed (prone to error)Detected during dynamic management correction operationserror
D22Excessive electronic A unit values detectedDetected during calibration or filtering operationswarning
D23Exception in corruption accumulationDetected during exception managementerror
D24Required NUB enabling signal missingDetected during NUB signal generation verificationerror
D25Callable collocation value set requiredDetected during calibration or filtering operation timeoutswarning
D26Multiple A-unit size fixed-title-E detection requiredDetected during multi-element A units management feedback presence verificationerror
D27Size-coded value resizing requiredDetected when size-coded value size-coded from alignment differswarning
D28Required imaging detection when size managed allegation requiredDetected during size management allegation requirement verificationwarning
D29Remapping from allegation configuration management requiredDetected when allegation configuration management neededwarning
D2APatching allocation value correction requiredDetected when patching allocation value requires correctionwarning
D2BCombination with requirement mandates system requirementsDetected during combination and system requirement verificationwarning
D2CSize codec exception in filtering operationDetected during size codec from filtering operation size-coded verificationwarning
D2EMultiple A-unit value filtration status requirementsDetected when multiple A-unit filtering operation status failederror
D2FCallable collection value correction requiredDetected when callable collection value requires calibration correctionwarning
D30Multiple A units status required immediate correctionDetected when A unit status data fails maximum or min fixationerror
D31Specific-titled A-unit problem existence failureDetected when A-unit problems corrupt repair requirements fromerror
D32Error message issue in expectancy condition requiredDetected during error message check occurrenceerror
D33System initialization requiredDetected when system initialization failederror
D34Color imaging management failure detectedDetected when color imaging management failed in systemerror
D35Parameter error management occurredDetected as error management corruption occurrederror
D36Program error management on occasion corruptionDetected on error occasion management corruption detectederror
D37Management color imaging failure in system detectedDetected when color imaging management failure in systemerror
D50BAllocation failure or palette allocation errorSystem detects insufficient memory allocation for color paletteerror
D50DDefault information in formation default piece default two piece of default formationSystem detects default formation conditionerror
D50EImage code size allocation errorSystem detects incorrect image code sizewarning
D50FMaximum number of size codes exceeded in limitationSystem detects exceeding maximum size codes during operationwarning
D510Image sizing requirement with gallerySystem detects image sizing inconsistencyinfo
D511Film marking specifications magification rateSystem detects film marking rate inconsistency within specificationwarning
D512Film pattern replacement pattern replacementSystem detects lift film pattern inconsistencywarning
D513Film combination requirement with film specificationSystem detects film combination or image designation inconsistencyerror
D514Film combination orientation film size film information maximum designation exceedingSystem detects maximum designation exceeded in film combinationerror
D515Error application software occurrence detectedAn exception has occurred in the error mapping processerror
D516Exception error occurrence detected in error mapping processAny exception detected in one or more detection invalid key rows during error mapping processerror
D517Valid key detection errorSystem detects invalid key rows during validationerror
D518Error application software occurrence detectedSystem cannot open file or ENE value is not available or exclusive access not available process mapping applicableerror
D51AEvent setup maximum number of entries exceedingSystem detects more than 1080 events in formation within event maximum limitationerror
D51BError application software occurrence detectedAn exception has been detected in error mapping file entry point detection occurrence log processingerror
D709Equipment stage capacity errorStage equipment capacity create errorerror
E025DSpecific error definition not fully readable due to OCR degradationUnknownunknown
E026DSpecific error definition not fully readable due to OCR degradationUnknownunknown
E0604Application software errorUnknownerror
E0605Application software errorUnknownerror
E0606Application software errorUnknownerror
E0607Application software errorUnknownerror
E0608Application software errorUnknownerror
E090Application software errorSystem bus error detection during operationerror
E091Hardware/control panel connection errorControl panel failure or connection loss to main equipmenterror
E092Image management system initialization failureSystem fails to initialize management classes during startuperror
E093FCR reader connection errorFCR reader fails to establish connection to main uniterror
E094Header authentication or workflow errorWorkflow failure or header authentication issue detectederror
E095Image file life-cycle management process errorFile lifecycle management process fails during current operationerror
E096Image file initialization in progress or incompleteSystem detects incomplete image file initializationerror
E097Image file initialization suspended or error in progressSystem detects 'initiate complete' message or get glossary system errorerror
E098System bus 'trans' message error or header control failureHeader control connection failure or system bus error when 'trans' glossary detectederror
E09ARegister management device canine reverse operation requiredSystem detects requirement to cancel register management device operationerror
E09BEquipment name doesn't match or known issue encounteredEquipment name mismatch detected or equipment exclusion requiredwarning
E09DHost management device required; reverse operation or configuration requiredCandidate manager device canine reverse operation required; application software error during operationwarning
E09EImage review system emergency or user-defined issue encounteredUser-defined emergency detected when logging in; image management system issue; missing-user attributeerror
E09FWork file lifecycle property deletion or registry management errorSystem detects lifecycle property deletion requirement; proper work item management needederror
E0CInitialize data failureData initialization process failserror
E0C08Application software errorUnknownerror
E0C09Application software errorUnknownerror
E0C0AApplication software errorUnknownerror
E0C0BApplication software errorUnknownerror
E0C0CApplication software errorUnknownerror
E0C0DApplication software errorUnknownerror
E0D0Old hold load data failureLoad old hold data failserror
E0D1Write wire data failureWrite wire data failserror
E0D2Performance errorImaging processing performance error occurserror
E0D3CSV value failureCSV parameter processing failserror
E0D4Fail discard dataData deletion or discard failserror
E0D5Fail registration data fileFile registration failserror
E0D6Find data validity failureData validity check failserror
E0D8Body data convert currentlyBody code conversion occurs currentlyerror
E0D9Machine image size detectionMachine image size code failserror
E0DALift file validation dataUplift file manager formation failserror
E0DBLift file validation dataUplift file manager formation failserror
E0DCMachine size code sizingMachine image sizing code failserror
E0DDIT wage full failure exampleWage IT code fullness issueerror
E0DEIT wage full value failureWage IT code fullness value issueerror
E0DFIT wage full value failure detailWage IT code fullness value issue pending detailerror
E0F0Proprietary directory protocol establishment failureSystem fails to establish proprietary directory protocol connectionerror
E709Management form requirement errorFailure getting equipment management formerror
E-A19FCR device receiver malfunctionReceiver hardware detects error conditionerror
F709Setup form requirement storage errorSetup form requirement storage failureerror
FA19SCMU device malfunctionSmart Card Memory Unit error condition detectederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI