Fresenius Medical Care 2008T: troubleshooting and error codes
The ServiceForge knowledge base holds 33 documented symptoms for the Fresenius Medical Care 2008T, with 54 ranked causes, 65 diagnostic checks with test points and expected values, and 34 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: CPRE sensor stability
Monitor CPRE (conductivity/pressure reading) value stability on debug screen #5 over 1 minute to verify sensor performance
Ranked causes
- 1. Sensor board malfunction (unknown)
- 2. Motor noise interference (unknown)
- 3. -12V power supply problems (unknown)
Monitor CPRE value stability
- Test point
- CPRE on debug screen #5
- Expected
- Difference between highest and lowest values ≤ 8 units, ≤ 8
Calibration: TPRE sensor stability
Monitor TPRE (temperature/pressure reading) value stability on debug screen #5 over 1 minute to verify sensor performance
Ranked causes
- 1. Sensor board malfunction (unknown)
- 2. Motor noise interference (unknown)
- 3. -12V power supply problems (unknown)
Access debug mode
- Test point
- keyboard
- Expected
- Main display changes to indicate debug mode
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
CPOS sensor value variation exceeds tolerance
Debug screen #5 CPOS value fluctuates more than 8 units over 1-minute observation period
Ranked causes
- 1. Sensor board malfunction (unknown)
- 2. Motor noise coupling into sensor (unknown)
- 3. -12V power supply problems (unknown)
Observe CPOS value on debug screen #5 for 1 minute
- Test point
- Debug screen #5, CPOS parameter
- Expected
- Difference between highest and lowest ≤ 8
CPRE sensor variability out of specification
Debug screen #5 shows CPRE value fluctuating; difference between highest and lowest readings exceeds 8 units over 1 minute observation
Ranked causes
- 1. Sensor board malfunction (unknown)
- 2. Motor noise interference (unknown)
- 3. -12V power supply problems (unknown)
Measure CPRE value stability
- Test point
- Debug screen #5 CPRE display
- Expected
- Difference between highest and lowest ≤ 8 units, ≤ 8
Power failure battery alarm does not sound or Battery Test fails patient safety
Audible alarm does not activate when mains power removed, or internal battery test fails during alarm test
Ranked causes
- 1. 9-volt battery depleted or missing (common)
- 2. Power Logic Board IC9 failure (rare)
- 3. Power Logic Board failure due to swollen battery (occasional)
Remove 9-volt battery and perform Alarm Test
- Test point
- Battery Test function
- Expected
- Battery Test fails with battery removed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TPOS sensor value variation exceeds tolerance
Debug screen #5 TPOS value fluctuates more than 2 units over 1-minute observation period
Ranked causes
- 1. Sensor board malfunction (unknown)
- 2. Motor noise coupling into sensor (unknown)
- 3. -12V power supply problems (unknown)
Observe TPOS value on debug screen #5 for 1 minute
- Test point
- Debug screen #5, TPOS parameter
- Expected
- Difference between highest and lowest ≤ 2
TPRE sensor variability out of specification
Debug screen #5 shows TPRE value fluctuating; difference between highest and lowest readings exceeds 2 units over 1 minute observation
Ranked causes
- 1. Sensor board malfunction (unknown)
- 2. Motor noise interference (unknown)
- 3. -12V power supply problems (unknown)
Access debug mode
- Test point
- Keyboard CTRL + [▲] and [▼] keys
- Expected
- Main display changes to indicate debug mode
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Alarm Test fails to trigger expected system response patient safety
Test board stimulus applied but functional board and blood systems do not respond as expected; alarms not triggered or systems not shut down
Ranked causes
- 1. Test board malfunction or communication failure with Functional board (common)
- 2. Functional board detector or response circuit failure (occasional)
Verify Alarm Test prerequisites before start
- Test point
- Machine state
- Expected
- alarm free, TX Clock stopped, dialysate lines in shunt box, no blood sensed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arterial Pressure Hard Limit alarm not triggered patient safety
Test board stimulus to arterial pressure transducer does not cause Functional board to detect over/under limit condition and shut blood pump
Ranked causes
- 1. Arterial pressure transducer offset adjustment circuit malfunction (common)
- 2. Functional board ADC conversion failure or threshold comparison error (occasional)
Art High Hard: Measure transducer offset voltage with DAC detuning at 0V
- Test point
- Arterial pressure transducer output (Vout)
- Expected
- considerably above 5V (>350 mmHg equivalent)
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure module pressure calibration error
Display pressure and external pressure meter reading differ by more than 3 mmHg
Ranked causes
- 1. Calibration drift or sensor error in module (unknown)
- 2. Over-pressure relief not functioning (unknown)
Compare display pressure to calibrated external meter
- Test point
- M400/TM-2910/TM-2915/TM-2917 use external pressure meter; M2000/M2600/M3600/SunTech use T-fitting setup
- Expected
- Display and meter pressures agree within 3 mmHg at stabilization
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diasafe Test fails or does not meet pressure rise requirement patient safety
Diasafe valve test does not achieve 60 mmHg pressure rise at 2000ml/min UF rate or final pressure remains below 300 mmHg
Ranked causes
- 1. Diasafe valve V28 or check valves stuck or leaking (common)
- 2. UF pump membrane or rotor worn; flow delivery degraded (occasional)
Record baseline dialysate pressure at 4000ml/min UF rate after 100 strokes
- Test point
- Dialysate pressure reading (DIAL-PRESSURE channel 4)
- Expected
- stable reading after 100 strokes
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure Test fails during stabilization phase patient safety
Pressure reading during 20-30 second stabilization period falls outside expected range, test terminates as failed
Ranked causes
- 1. Pressure transducer drift or calibration error (common)
- 2. Air in pressure lines during test (occasional)
Measure positive pressure (Get Pos TMP) at start of test
- Test point
- Dialysate pressure sensor
- Expected
- at least +350 mmHg
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure Test fails negative pressure phase patient safety
Negative pressure reading during test falls outside range -250 to -450 mmHg or varies more than 21 mmHg
Ranked causes
- 1. Dialysate pump or UF pump mechanical wear or failure (common)
- 2. Pressure transducer malfunction or calibration drift (occasional)
Achieve negative pressure (Get Neg TMP) before flow test
- Test point
- Dialysate pressure sensor
- Expected
- at least -270 mmHg
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Relay contact test LEDs do not light caution
D6 and/or D19 LED on LP450 board fail to light when level down switch is pressed
Ranked causes
- 1. Relay contacts are bad or LP450 board malfunction (unknown)
- 2. Relay contacts fail when blood alarm active (unknown)
Press level down switch and verify LED illumination
- Test point
- D6 and D19 LEDs on LP450 board
- Expected
- both LEDs D6 and D19 light
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5V supply voltage out of tolerance high voltage
Internal 5-volt supply measured on debug screen #4 not within 0.2V specification
Ranked causes
- 1. 12-volt supply out of tolerance (common)
Access debug screen #4
- Test point
- Machine display
- Expected
- Debug screen #4 displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arterial pressure transducer calibration out of specification patient safety
Arterial Pressure bargraph does not match external pressure meter at 0 and 200mmHg calibration points
Ranked causes
- 1. Pressure transducer contamination or malfunction (occasional)
Inspect internal pressure transducer protector
- Test point
- Arterial transducer PART port protector
- Expected
- No contamination visible
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood dimness detector voltage out of range patient safety
Blood dimness level measured on debug screen #4 outside 5.0±1.0V specification
Ranked causes
- 1. Contamination in glass detector tube (occasional)
Access debug screen #4
- Test point
- Machine display
- Expected
- Debug screen #4 displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood leak detector voltage out of range patient safety
Blood leak level measured on debug screen #4 outside 4.5-5.2V specification
Ranked causes
- 1. Contamination in glass detector tube (occasional)
Access debug screen #4
- Test point
- Machine display
- Expected
- Debug screen #4 displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure module air leakage exceeds limit
Cuff pressure drops more than specification during air leakage test
Ranked causes
- 1. Tubing leak, connection leak, or internal valve leakage (unknown)
Measure air leakage rate
- Test point
- Module-specific Dummy Cuff port (M400/M2000/M2600: 220cc; M3600: 220cc; TM-2910/TM-2915/TM-2917: 700cc; SunTech: 700cc)
- Expected
- M400/M2000/M2600: ≤15 mmHg; M3600: ≤12 mmHg; TM-2910/TM-2915/TM-2917: ≤30 mmHg; SunTech: ≤22 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure module deflation speed out of range
Module deflates at rates outside specification at various pressure points during deflation test
Ranked causes
- 1. Degraded or faulty deflation control valve (unknown)
Measure deflation rate at 240 mmHg
- Test point
- Dummy Cuff 220cc port
- Expected
- M400: 4.3-6.2 mmHg/sec; M2000/M2600: 4.4-6.7 mmHg/sec; M3600: 3.8-6.6 mmHg/sec; TM-2910/TM-2915/TM-2917: 3.5-7.5 mmHg/sec
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure module inflation speed out of range
Module takes longer than specified to pressurize dummy cuff to 250 mmHg during inflation speed test
Ranked causes
- 1. Module type mismatch or degraded inflation performance (unknown)
Measure inflation time to 250 mmHg
- Test point
- Dummy Cuff 700cc port
- Expected
- M400: ≤7.3 seconds; M2000/M2600/M3600: ≤10.0 seconds; TM-2910/TM-2915/TM-2917: ≤10.0 seconds; SunTech: ≤11.0 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pump rotor occlusion pressure out of range patient safety
Peak pressure measured at pump outlet during occlusion test falls outside 25-35 psi range
Ranked causes
- 1. Blood pump rotor springs worn or damaged (common)
Measure peak pressure at blood pump outlet during occlusion
- Test point
- pressure gauge between pump output and clamp
- Expected
- between 25 and 35psi, 25-35 psi
- Tools
- pressure gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deaeration pressure out of specification caution
Deaeration pump inlet pressure not within acceptable range when dialysate flow is ON
Ranked causes
- 1. Deaeration pump adjustment needed (unknown)
Install pressure gauge on deaeration pump inlet
- Test point
- Inlet (suction) side of deaeration pump
- Expected
- Gauge installed with T-fitting
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Dialysate line safety interlock malfunction patient safety
Safety switches for dialysate lines and bypass shunt cover do not function properly
Ranked causes
- 1. Dialysate line safety interlock switches defective or misaligned (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dialysate tubing damaged or cracked caution
Visible damage, cracks, or deterioration on dialysate lines at machine connections
Ranked causes
- 1. Aging, UV exposure, or mechanical damage to tubing (unknown)
Visually inspect upper dialysate line for damage
- Test point
- Upper dialysate line connection point at back of machine
- Expected
- No cracks, tears, discoloration, or visible deterioration
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow relief pressure out of specification caution
Flow pump outlet pressure not within specified range during service mode test
Ranked causes
- 1. Flow pump relief valve adjustment needed (unknown)
Enter Service Mode and access Calibrate Hydraulics screen
- Test point
- Machine main display
- Expected
- Calibrate Hydraulics screen displayed
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Heat disinfect mode dialysate temperature below specification patient safety
Dialysate temperature fails to reach 80-90°C during heat disinfect mode
Ranked causes
- 1. Heater malfunction or insufficient heating (common)
Monitor dialysate temperature during heat disinfect mode
- Test point
- external temperature meter on dialysate lines
- Expected
- between 80°C and 90°C, 80-90°C
- Tools
- external temperature meter capable of reading at least 85°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Level detector alarm channel LEDs do not light in sequence patient safety
Channel 1 and/or Channel 2 LEDs on level detector circuit board do not light when venous chamber is removed
Ranked causes
- 1. Level detector circuit board malfunction (unknown)
Watch Channel 1 and Channel 2 LEDs when venous chamber is removed
- Test point
- Channel 1 and Channel 2 LEDs on circuit board
- Expected
- Channel 1 LED lights first followed quickly by Channel 2
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion clamp does not fully seal venous line patient safety
Air escapes from venous line when pressure applied, indicating incomplete occlusion
Ranked causes
- 1. Occlusion clamp worn or damaged (common)
Apply pressure to venous line and observe for air leakage
- Test point
- end of venous line in water container
- Expected
- no air escapes from the venous line psi
- Tools
- syringe, pressure gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OLC test resistor plug replacement
Replace test connectors on distribution board and install calibration resistor plugs for OLC testing
Ranked causes
- 1. OLC sensor testing required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient exposure to improper dialyzing fluid patient safety
Patient exposed to dialyzing fluid with incorrect temperature or conductivity during treatment
Ranked causes
- 1. Dialyzer valves V24, V25, V26 fail to bypass dialyzer during alarm condition (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rinse flow rate below minimum caution
Less than 310 ml collected in 30 seconds during rinse mode flow test
Ranked causes
- 1. Rinse circuit blockage or pump flow reduction (common)
Measure rinse drain flow volume in 30 seconds
- Test point
- drain line collected in graduated cylinder
- Expected
- minimum of 310ml in 30 seconds
- Tools
- 1000ml graduated cylinder, timer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrafiltration pump does not deliver output
Green pump light is on but pump produces no flow output
Ranked causes
- 1. Check valve installed backwards (occasional)
Verify check valve orientation in UF pump
- Test point
- Check valve inlet/outlet markings on valve body
- Expected
- Flow direction arrow pointing toward outlet; valve installed with correct polarity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
A.11 | Pump is not reaching speed. | Blood pump fails to achieve target speed | warning |
A.13 | Pump rotor is turning in the wrong direction. | Blood pump rotor rotates in reverse or incorrect direction | error |
A.16 | Key stuck or held in too long. | User holds a key down or key is stuck mechanically | warning |
A.20 | Set speed read-back analog voltage at X348/14 is out of limits. | Set speed feedback voltage exceeds allowable range | error |
A.21 | Actual speed read-back analog voltage at X348/10 is out of limits. | Actual speed feedback voltage exceeds allowable range | error |
A.22 | Arterial pressure read-back analog voltage at X348/7 is out of limits. | Arterial pressure feedback voltage exceeds allowable range | error |
A.24 | Optical sensor frequency (tach) not in range. | Tachometer signal frequency outside expected range | warning |
A.25 | Pressure increases too quickly when Level Up key is pressed. | Pressure rises at excessive rate during level adjustment | warning |
A.26 | Pressure was adjusted too much in calibration mode. | Pressure adjustment exceeds safe limits during calibration | critical |
A.27 | Timeout when receiving Intel-Hex-line or overflowed receive buffer. | Communication timeout or buffer overflow during firmware update | error |
A.28 | Error in received Intel-Hex-line. | Corrupted or invalid data received during firmware transfer | error |
A.29 | Pump rotor turning when it should not be (first revolution). | Rotor rotates during initialization or standby | error |
E.01 | EPROM CRC error. | EPROM checksum validation fails at startup | critical |
E.02 | Flash ROM CRC error. | Flash ROM checksum validation fails | critical |
E.03 | RAM check error. | RAM self-test fails during initialization | critical |
E.04 | Reference Voltage error. | Reference voltage circuit failure detected | critical |
E.05 | Serial EEPROM error. | EEPROM communication or checksum failure | critical |
E.06 | Watchdog timeout. | CPU watchdog timer expires indicating system hang | critical |
E.07 | +12 volts is outside the allowable range of 10.8 to 13.2 volts. | +12V supply voltage falls outside specified limits | critical |
E.08 | +24 volts is outside the allowable range of 22.8 to 28.0 volts. | +24V supply voltage falls outside specified limits | critical |
E.09 | -12 volts is outside the allowable range of -9.6 to -13.2 volts. | -12V supply voltage falls outside specified limits | critical |
E.10 | +5 volts is outside the allowable range of 4.75 to 5.25 volts. | +5V supply voltage falls outside specified limits | critical |
E.14 | 50 ms time period exceeded. | Software task execution exceeds 50 millisecond deadline | error |
E.15 | Software task was not completed correctly. | Task completion check fails during execution | error |
E.23 | Pump rotor turning when it should not be (second revolution). | Rotor continues rotating during initialization standby phase | error |
E.97 | Error copying data into Flash ROM. | Flash write operation fails during firmware update | critical |
E.98 | Error erasing Flash ROM. | Flash erase operation fails during firmware update | critical |
E.99 | Transmit error during Flash update. | Communication error occurs while sending firmware data | critical |
Fill Program Alarm | Filling program runs for more than 60 seconds with blood sensed | Filling program continues beyond 60 seconds while blood is detected | critical |
Flow Error | Chamber full switch does not switch balance chamber | CFS (chamber full switch) does not switch the balance chamber for 10 seconds | error |
Flow Inlet Error | Float switch does not detect low water level after balance chamber cycle | Float switch does not detect low level within 2.5 seconds after balance chamber cycles | error |
High Flow Error | Balance chamber cycles too rapidly | CFS (chamber full switch) switches the balance chamber faster than 1.8 seconds | error |
No Water | Water level not detected after float switch opens valve 41 | Float switch opens valve 41 but water level not up within approximately 6 seconds | error |
RED LED | Pump door is open longer than preset time (default 30 seconds). | Door remains open beyond time limit set by dip-switch 4 on LP955 | warning |