Fisher & Paykel Healthcare MR700: troubleshooting and error codes
The ServiceForge knowledge base holds 36 documented symptoms for the Fisher & Paykel Healthcare MR700, with 84 ranked causes, 81 diagnostic checks with test points and expected values, and 16 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Excessive water in circuit patient safety
Rainout or water accumulation in breathing circuit tubing
Ranked causes
- 1. Water trap not at lowest point (non-heated mode) (common)
- 2. Heater wire too close to temperature probe (occasional)
- 3. Circuit cooled by external air conditioner (occasional)
- 4. Chamber control set too high (occasional)
- 5. Maximum permissible peak flow rate of chamber exceeded (occasional)
- 6. Humidifier chamber tilted (common)
- 7. Faulty temperature probe (low resistance) (rare)
- 8. Inspiratory heater wire not connected in parallel dual circuit (occasional)
Verify chamber level
- Test point
- Humidification chamber
- Expected
- Chamber level and stable
1 more check with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low temperature alarm patient safety
Airway or chamber temperature drops below set threshold; alarm activates
Ranked causes
- 1. Gas flow stopped or interrupted (common)
- 2. Set temperature increased by more than 2°C (occasional)
- 3. Temperature probe not in probe housing (common)
- 4. Gas flow rate too low (occasional)
- 5. Excessive distance between heater wire and temperature probe (occasional)
- 6. Poor thermal contact between chamber base and heater plate (occasional)
- 7. Temperature probe not inserted at chamber outlet (occasional)
- 8. Heater element needs replacement (rare)
Verify gas flow to humidifier
- Test point
- Gas inlet
- Expected
- Gas flowing continuously
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature probe disconnected or faulty patient safety
Display shows '- - -' with flashing red Temperature Probe LED and audio alarm. Various fault patterns shown by decimal point sequences indicate open/short circuits.
Ranked causes
- 1. Temperature probe not plugged in or open circuit chamber outlet probe (common)
- 2. Open circuit in common lead in probe cable (occasional)
- 3. Open circuit patient airway probe (occasional)
- 4. Short circuit chamber outlet probe (occasional)
- 5. Short circuit patient airway probe (occasional)
- 6. Short in probe cable between airway and chamber probes (rare)
Observe display pattern for fault type identification
- Test point
- Airway temperature display
- Expected
- '- - -' with no decimal points = probe not plugged in or open chamber probe
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High temperature alarm patient safety
Airway or chamber temperature rises above set threshold; alarm activates
Ranked causes
- 1. Sudden spurious increase in gas flow (occasional)
- 2. Heater wire too close to temperature probe (common)
- 3. Faulty EOH sensor on temperature probe (low resistance) (rare)
- 4. No water in chamber (common)
- 5. Faulty TOP sensor on temperature probe (low resistance) (rare)
Verify water level in chamber
- Test point
- Humidification chamber
- Expected
- Water present at proper level
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing circuit leak patient safety
Loss of pressure or gas leakage from breathing circuit connections
Ranked causes
- 1. Faulty circuit connections (common)
- 2. No water bag connected (common)
- 3. Temperature probes incorrectly fitted (occasional)
- 4. Chamber filler stopper missing (occasional)
Inspect all breathing circuit connections
- Test point
- Circuit connectors
- Expected
- All connections seated and secure
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater wire alarm while temperature displayed caution
Heater wire alarm activates but temperature is still shown on display
Ranked causes
- 1. Heater wire unplugged (common)
- 2. Heater wire broken (occasional)
- 3. Heater wire switch turned off but wire connected (common)
- 4. Heater wire switch on but not connected (occasional)
Verify heater wire connection
- Test point
- Heater wire connector
- Expected
- Connector fully seated and secure
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low humidity patient safety
Humidifier output humidity insufficient for patient needs
Ranked causes
- 1. Chamber control set too low (common)
- 2. No water in humidification chamber (common)
- 3. Chamber temperature too low (occasional)
- 4. Gas flow rate too high for chamber capacity (occasional)
Verify water level in chamber
- Test point
- Humidification chamber
- Expected
- Water present at appropriate level
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature fluctuating patient safety
High and low temperature alarms alternate; temperature control unstable
Ranked causes
- 1. Ventilation rate too low or too erratic (common)
- 2. External heating or cooling influence (occasional)
- 3. Temperature probes incorrectly positioned (common)
- 4. Faulty temperature probe (intermittent open or short) (occasional)
Verify temperature probe position
- Test point
- Breathing circuit
- Expected
- Probe positioned correctly at outlet or airway
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature display, no alarm patient safety
Temperature is not displayed and no alarm lights are active
Ranked causes
- 1. Humidifier voltage rating incorrect (common)
- 2. Heaterplate overheat protector operated (occasional)
- 3. Transformer overheat protector operated (rare)
Verify mains voltage rating
- Test point
- Unit label / mains supply
- Expected
- Unit voltage rating matches supply voltage
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature probe alarm patient safety
Display shows '---' indicating temperature probe disconnection or fault
Ranked causes
- 1. Temperature probe unplugged (common)
- 2. Temperature probe plug dirty (common)
- 3. Temperature probe faulty (occasional)
Verify temperature probe connection
- Test point
- Probe connector
- Expected
- Probe connector fully seated and secure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Airway temperature too high (heater wire mode) patient safety
Red Airway Temp High LED flashes and audio alarm sounds when gas temperature is 2 °C higher than Temperature Control setting. Both heater circuits immediately disabled.
Ranked causes
- 1. Temperature control setting too high (common)
- 2. Heater malfunction causing overheating (occasional)
Monitor airway temperature reading against Temperature Control setting
- Test point
- Airway temperature display
- Expected
- Airway temperature within 2 °C of Temperature Control setting, ≤ 2 °C
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Airway temperature too low (heater wire mode) patient safety
Red Airway Temp Low LED flashes and audio alarm sounds when gas temperature is 2 °C lower (4 °C for MR720) than Temperature Control setting. If condition persists >100 seconds, heater circuits shut off. Mute button silences alarm and resumes heat for 100 seconds.
Ranked causes
- 1. Insufficient heater output (common)
- 2. Heater wire or heaterplate malfunction (occasional)
Monitor airway temperature reading against Temperature Control setting
- Test point
- Airway temperature display
- Expected
- Airway temperature within 2 °C (MR700) or 4 °C (MR720) of Temperature Control setting, ≥ -2 °C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Airway temperature too low (non-heater wire mode) patient safety
Red Airway Temp Low LED flashes. Audio alarm sounds after 30 second delay. Pressing mute silences alarm for 10 minutes and restores heaterplate power during that time. If temperature drops below 5 °C, display shows 'LO' and audio alarm sounds.
Ranked causes
- 1. Insufficient heaterplate output (common)
- 2. Heaterplate malfunction (occasional)
Monitor airway temperature reading
- Test point
- Airway temperature display
- Expected
- Airway temperature ≥ 29.5 °C, > 29.5 °C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature measurement
Calibration of VR4 (41 °C comparator) and VR5 (A-to-D converter span) after PCB service or E0x-E7x errors
Ranked causes
- 1. PCB repair or replacement affecting ±5V rails or A-to-D converter (common)
- 2. E0x to E7x error messages displayed (common)
Verify unit warm-up time
- Test point
- Unit operation time
- Expected
- at least twenty minutes
9 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber temperature too high patient safety
Chamber Temp High amber LED flashes and audio alarm sounds (can be muted). Heater circuits switch off. Triggered when chamber outlet gas exceeds chamber set temperature by >4 °C for >20 minutes, or by ≥10 °C immediately.
Ranked causes
- 1. Chamber set temperature too high (common)
- 2. Heater malfunction causing chamber overheating (occasional)
Monitor chamber outlet temperature against chamber set temperature
- Test point
- Chamber temperature sensor reading
- Expected
- Chamber outlet temp within 4 °C of chamber set temperature, ≤ 4 °C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber temperature too low patient safety
Chamber Temp Low amber LED flashes and audio alarm sounds. Triggered by two criteria depending on set temperature: <4 °C below set temp for >20 min (if set <35 °C), or <31 °C for >20 min (if set ≥35 °C).
Ranked causes
- 1. Insufficient heater output or set temperature too low (common)
- 2. Heater malfunction or insufficient heating (occasional)
Monitor chamber outlet temperature relative to chamber set temperature
- Test point
- Chamber temperature sensor reading
- Expected
- If set <35 °C: outlet ≥ (set - 4 °C); if set ≥35 °C: outlet ≥ 31 °C
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater wire disconnected or faulty (MR700/720/730) caution
Heater Wire alarm LED flashes, audio alarm sounds. Both heater wire and heaterplate circuits shut down. Display may show 'F-2' if fuse is open.
Ranked causes
- 1. Heater wire disconnected (common)
- 2. F2 fuse open circuit (occasional)
Check heater wire physical connection
- Test point
- Heater wire connector
- Expected
- Heater wire firmly seated and undamaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature display with 'see manual' indicator patient safety
No temperature display shown and 'see manual' indicator light is on
Ranked causes
- 1. Control PCB fault from water damage (occasional)
- 2. Failed self test with control PCB issue (occasional)
Inspect control PCB for water or contamination
- Test point
- Control PCB
- Expected
- PCB clean and dry, no corrosion or damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Set low warning active caution
Set Low amber LED enabled continuously (no audio alarm). Version 1.0/2.0: triggered if chamber set <31.0 °C. Version 3.0: triggered if chamber set <34.0 °C or Temperature Control setting <34.0 °C.
Ranked causes
- 1. Chamber set temperature below minimum threshold (common)
- 2. Temperature Control setting below minimum (Version 3.0 software) (occasional)
Check software version and chamber set temperature
- Test point
- Software version and chamber set temperature display
- Expected
- v1.0/2.0: chamber set ≥31.0 °C; v3.0: chamber set ≥34.0 °C and Temperature Control ≥34.0 °C
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
41°C overheat calibration fault caution
41°C overheat protection circuit calibration out of limits
Ranked causes
- 1. 41°C comparator calibration out of limits (rare)
Verify 41°C comparator calibration
- Test point
- Overheat protection circuit
- Expected
- Calibration within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Airway temperature too high (non-heater wire mode) patient safety
Red Airway Temp High LED flashes and audio alarm sounds when gas temperature is 41 °C or higher. Heaterplate power immediately shut off.
Ranked causes
- 1. Heaterplate set temperature too high (common)
Monitor airway temperature
- Test point
- Airway temperature display
- Expected
- Airway temperature < 41 °C, < 41 °C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED flashing without audible sound patient safety
Alarm LED is blinking but no audible alarm tone is heard
Ranked causes
- 1. Mute switch has been pressed (common)
Check mute switch status
- Test point
- Mute button / display
- Expected
- Mute mode inactive or timer expired
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Airway temperature control
Set airway temperature from 31 to 40 °C using front panel control knob. Resolution 0.1 °C; display flashes for 5 seconds after adjustment.
Ranked causes
- 1. Airway temperature adjustment required (unknown)
Set airway temperature
- Test point
- Front panel Temperature Control knob
- Expected
- Display shows selected temperature (31 to 40 °C) flashing for 5 seconds °C minimum, 31 to 40 °C range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Chamber control (heater wire mode)
Adjust chamber outlet temperature offset from airway temperature. Range -5 to +2 °C in 0.5 °C steps (heater wire mode only). Fixed knob option available to prevent inadvertent adjustment.
Ranked causes
- 1. Chamber outlet temperature offset adjustment required (unknown)
Set chamber control offset
- Test point
- Front panel Chamber Control knob (heater wire mode only)
- Expected
- Display shows offset value (-5 to +2 °C) flashing for 5 seconds °C minimum, -5 to +2 °C range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Heater wire mode selection caution
Select heater wire mode (MR720, MR730 only) using latching push switch. Green LED indicates heater wire mode active. Audible beep confirms mode change. Alarm triggers if mode switched with heater wire disconnected.
Ranked causes
- 1. Heater wire mode enabled/disabled (unknown)
Verify heater wire mode selection
- Test point
- Front panel Heater Wire Switch and adjacent green LED
- Expected
- Green LED on for heater wire mode; LED off for non-heater wire mode; audible beep confirms change
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power-up sequence verification (Version 1.0)
Verify complete power-up sequence for Version 1.0 software: watchdog test, display test, temperature and chamber settings, software version, heater plate setting, and audio beep. Each displays for 2 seconds.
Ranked causes
- 1. Power-up sequence check required (unknown)
Watchdog circuit test
- Test point
- Power-on sequence display
- Expected
- SEE MANUAL LED lights briefly
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power-up sequence verification (Version 2.0/3.0)
Verify complete power-up sequence for Version 2.0 and 3.0 software: watchdog test, display test, model number, software version, temperature setting, chamber setting, heater plate setting, and audio beep. Each displays for 2 seconds with momentary blanking.
Ranked causes
- 1. Power-up sequence check required (unknown)
Watchdog circuit test
- Test point
- Power-on sequence display
- Expected
- SEE MANUAL LED lights briefly
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Warm-up mode (heater wire mode)
Verify warm-up mode exiting conditions for heater wire mode (MR700, MR730 all versions, MR720 v1.0). Low temperature alarm disabled during warm-up; enabled when exiting. Mode exits when: airway temp within 0.8 °C of setpoint, OR no increase in 4 min (2 min v3.0), OR not within 2 °C after 10 min.
Ranked causes
- 1. Warm-up mode operation and exit conditions (unknown)
Monitor airway temperature during warm-up
- Test point
- Front panel airway temperature display
- Expected
- Temperature rises from initial value toward setpoint
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Warm-up mode (non-heater wire mode)
Verify warm-up mode exiting conditions for non-heater wire mode (MR730/MR720 v1.0, MR480). Low temperature alarm disabled during warm-up; enabled when exiting. Mode exits when: airway temp above 30 °C, OR no increase in 15 min, OR 2 °C increase in 15 min but not 30 °C after further 15 min.
Ranked causes
- 1. Warm-up mode operation and exit conditions (non-heater wire) (unknown)
Monitor airway temperature during warm-up
- Test point
- Front panel airway temperature display
- Expected
- Temperature above 30 °C
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External water leak - disposable chamber caution
Water leaking from disposable humidification chamber exterior
Ranked causes
- 1. Chamber used beyond lifespan (common)
Check chamber age and condition
- Test point
- Disposable chamber
- Expected
- Chamber within service life, no cracks or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External water leak - reusable chamber caution
Water leaking from reusable humidification chamber exterior
Ranked causes
- 1. Chamber O-ring faulty (occasional)
Inspect chamber O-ring
- Test point
- Chamber seal
- Expected
- O-ring clean, undamaged and properly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater wire alarm with F2 display caution
Heater wire alarm activates with 'F2' fault code displayed
Ranked causes
- 1. Internal fuse blown (rare)
Visual fuse inspection if accessible
- Test point
- Internal fuses F1-F4
- Expected
- Fuses intact and properly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater wire connected in non-heater wire mode caution
Heater Wire alarm LED flashes and audio alarm sounds. Indicates a heater wire has been connected when non-heater-wire mode is selected (Heater Wire 'On' LED is off). Heaterplate is switched off.
Ranked causes
- 1. Heater wire incorrectly connected in non-heater wire mode (common)
Verify operating mode and heater wire connection status
- Test point
- Heater Wire 'On' LED and physical heater wire connection
- Expected
- If non-heater-wire mode: no heater wire connected; if heater-wire mode: heater wire connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MR480 jumper cable connection (Revision D or earlier)
Special cable connection required for MR480 model during calibration
Ranked causes
- 1. MR480 calibration jumper requirement (common)
Install MR480 jumper cable
- Test point
- Pin 7 of U10 to cathode of D7
- Expected
- Jumper cable connected with 22 kΩ resistor in series
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Set Low LED on caution
'Set Low' LED is illuminated on the control display
Ranked causes
- 1. Chamber or airway temperature control setting too low (common)
Review temperature control setting
- Test point
- Control display setting
- Expected
- Setting appropriate for clinical requirement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby LED flashing patient safety
Standby indicator LED is blinking on the humidifier display
Ranked causes
- 1. Humidifier in standby mode (common)
Verify standby mode status
- Test point
- Standby LED / mode button
- Expected
- Standby mode inactive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
- - - | Temperature probe not plugged in or low temperature (~0 °C) | Temperature probe disconnected at connector or open circuit chamber outlet probe | error |
- - -. | Open circuit in common lead of probe cable | Open circuit detected in common lead of probe cable | error |
-.- - | Short in probe cable between patient airway and chamber outlet probes | Short circuit detected between patient airway probe and chamber outlet probe in cable | error |
--- | Temperature sensor probe fault | Probe signal absent, shorted, or out of range | error |
b-1 | Mute button stuck on at turn on or held > 5 minutes | Mute button remains pressed at power-on or continuously for over 5 minutes | warning |
b-2 | Standby button is on | Standby button pressed or stuck | warning |
Ec | Checksum error; possible failure of U1 EPROM | Microprocessor detects checksum mismatch during EPROM read | critical |
En | Memory test error; possible failure of U4 RAM | Microprocessor detects RAM test failure on U4 | critical |
EP1 | Hardware fault; Q13 short not effective | Software-driven shorted probe cable test detects Q13 short circuit failure | error |
EP2 | Low probe temperature (~0 °C) | Probe temperature reading abnormally low during shorted probe cable test | warning |
EP3 | High probe temperature (~70 °C) | Probe temperature reading abnormally high during shorted probe cable test | warning |
EP4 | Partial short between active probe cable conductors | Shorted probe cable test detects capacitive coupling or partial short | warning |
EXx | ADC overflow or underflow error at multiplexer stage X (0-7); x=o for overflow, x=u for underflow | ADC counts exceed expected range at indicated multiplexer stage | error |
F-2 | Open circuit heater wire fuse (MR700/720/730 only) | Heater wire fuse F2 is open circuit or defective | error |
F-2 | Heater Wire fuse F2 is open circuited | Fuse F2 fails during heater wire operation | error |
LO | Airway temperature below 5 °C | Airway temperature drops below 5 °C in either heater wire or non-heater wire mode | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.