Fisher & Paykel Healthcare AIRVO 2: troubleshooting and error codes
The ServiceForge knowledge base holds 21 documented symptoms for the Fisher & Paykel Healthcare AIRVO 2, with 65 ranked causes, 52 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Check for blockages (E121)
Unit displays blockage alert. Flow may be restricted in tube, chamber, filter, or patient interface.
Ranked causes
- 1. Non-return valve flaps displaced in water chamber (common)
- 2. Water chamber overfilled above black line (occasional)
- 3. Heated breathing tube blocked or kinked (common)
- 4. Patient interface blocked or kinked (common)
- 5. Wrong tube mode for interface (Default mode with Junior tube) (occasional)
- 6. Unsuitable cannula in use (occasional)
- 7. Air filter dirty or significantly discolored (common)
- 8. Foreign object blocking air filter or holder (occasional)
- 9. Device operating above 2,000 meters altitude (rare)
Inspect non-return valve position
- Test point
- Left-hand chamber port interior silicone flaps
- Expected
- Flaps in correct vertical position
5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 too low patient safety
Oxygen level has fallen below minimum allowed threshold (typically 25%). Displayed on screen or as alert.
Ranked causes
- 1. Oxygen flow rate set too low or flow meter adjustment needed (common)
- 2. Oxygen source not turned on (common)
- 3. Oxygen source empty or faulty (occasional)
- 4. Oxygen inlet kit not installed correctly or tubing kinked (common)
- 5. Oxygen tubing not securely connected to device (common)
- 6. Insufficient device warm-up time (occasional)
- 7. Minimum oxygen limit set to 25% threshold reached (occasional)
Verify oxygen flow rate
- Test point
- Oxygen flow meter setting on source
- Expected
- Flow rate sufficient to maintain >25% (or set limit)
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check for leaks (E122) patient safety
Unit displays leak alert. Water or humidified air escaping from chamber, tube, or interface connection.
Ranked causes
- 1. Water chamber not fitted correctly or not fully inserted (common)
- 2. Black O-ring missing or damaged on heated tube connection (common)
- 3. Heated breathing tube not attached correctly (common)
- 4. Patient interface not properly connected to tube (common)
- 5. Wrong tube mode for interface (Junior mode with Default tube) (occasional)
- 6. Air filter or filter cover not correctly fitted (occasional)
Verify water chamber insertion
- Test point
- Water chamber and finger guard at connector
- Expected
- Finger guard fully clicked into place
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check water
Water supply exhausted or not reaching chamber. Device displays alert prompting refill or chamber replacement.
Ranked causes
- 1. Water bag empty (common)
- 2. Water chamber empty or damaged (occasional)
- 3. Kink in fluid line preventing water flow (common)
- 4. Vent cap on water bag closed (common)
Check water bag status
- Test point
- Water bag indicator or by weight
- Expected
- Water bag contains sufficient water
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: disinfection tube error range 34-37, 42-43, 70-73, 75-78
Unit displays error code related to disinfection tube or cycle failure
Ranked causes
- 1. Disinfection cycle interrupted or components changed mid-cycle (common)
- 2. Disinfection tube connection issue (occasional)
- 3. Air filter or breathing tube connection issue (occasional)
- 4. Faulty disinfection tube (occasional)
Verify disinfection cycle status before error occurred
- Test point
- User observation of cycle progress
- Expected
- Clear information on cycle stage when error appeared
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot reach target flow
Unit cannot achieve the desired flow rate setting. Alert prompts user to confirm operation at lower achievable rate.
Ranked causes
- 1. Target flow setting exceeds patient interface capability (common)
- 2. System blockage or restricted flow (common)
- 3. Device operating above 2,000 meters altitude (rare)
Confirm user decision to proceed
- Test point
- Mode button on display
- Expected
- User presses mode button to accept lower flow rate
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: brownout error 161
Unit displays error code 161 indicating low mains supply voltage condition
Ranked causes
- 1. Low mains supply voltage (brownout condition) (common)
- 2. Electrical interference from nearby equipment (occasional)
- 3. Unit power supply internal failure (rare)
Check for nearby high-power electrical equipment
- Test point
- Power outlet surroundings
- Expected
- No heavy loads on same circuit
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: chamber error range 61-69, 74, 91-106
Unit displays error code related to chamber malfunction or temperature measurement
Ranked causes
- 1. Environmental temperature extremes or chamber overfill (common)
- 2. Chamber connection issue (occasional)
- 3. Temperature sensor malfunction (occasional)
Check environmental temperature
- Test point
- Room environment
- Expected
- Room temperature within normal operating range
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: flow error range 121-122 patient safety
Unit displays error code indicating flow measurement or delivery failure
Ranked causes
- 1. Breathing tube blockage or leak (common)
- 2. Air filter blockage (occasional)
- 3. Unit internal flow sensor or motor failure (occasional)
Verify prior blockage or leak alarm display
- Test point
- Unit display history
- Expected
- User recalls 'Check leaks' or 'Check blockages' message
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: heated breathing tube error range 11-33, 38-41, 44-50, 141-157 patient safety
Unit displays error code related to heated breathing tube malfunction
Ranked causes
- 1. Breathing tube connection issue (common)
- 2. Faulty breathing tube (occasional)
- 3. Unit hardware failure (rare)
Verify breathing tube connection
- Test point
- Tube connection port
- Expected
- Tube fully seated and secure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on
Device does not power up when ON/OFF button is pressed. May have broken display but be operating.
Ranked causes
- 1. Power cord not plugged into mains or device (common)
- 2. Power outlet not working (common)
- 3. Broken display (occasional)
Press and hold ON/OFF button
- Test point
- ON/OFF button
- Expected
- Device turns on or audible alarm sounds
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot reach target temperature
Device cannot achieve the desired temperature setting. Humidity level may be compromised. Most common at high flow in cold environment.
Ranked causes
- 1. Operating at high flow rate in cold ambient room (common)
- 2. Ambient room temperature below 18°C (64°F) (common)
Check ambient room temperature
- Test point
- Ambient environment
- Expected
- Room temperature 18°C (64°F) or above
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check tube (E38)
Alert displayed on screen or auditory signal indicating heated breathing tube fault. Tube may be disconnected, damaged, or improperly attached.
Ranked causes
- 1. Heated breathing tube not attached correctly (common)
- 2. Heated breathing tube visibly damaged (common)
Verify tube attachment
- Test point
- Heated breathing tube connection port
- Expected
- Tube seated firmly and alarm clears
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: internal error range 81-85, 111, 171
Unit displays error code indicating internal processing or logic failure
Ranked causes
- 1. Environmental shock or temporary malfunction (occasional)
- 2. Internal circuit or microprocessor failure (rare)
Document error details and triggering event
- Test point
- Unit display and user observation
- Expected
- Clear record of error code and preceding action
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: motor error range 1-10
Unit displays error code between 1-10 indicating motor-related failure
Ranked causes
- 1. Environmental shock or moisture exposure (occasional)
- 2. Motor component failure (occasional)
Check for physical damage from dropping or shaking
- Test point
- Unit exterior and internal components
- Expected
- No visible damage or moisture
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: oxygen sensor error range 181-192 patient safety
Unit displays error code related to oxygen sensor malfunction or signal failure
Ranked causes
- 1. Environmental temperature extremes affecting sensor (common)
- 2. Oxygen sensor degradation or failure (occasional)
Check operating environment temperature
- Test point
- Room where unit is used or stored
- Expected
- Temperature within normal range, not cold or hot
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code displayed: temperature error range 131-135 patient safety
Unit displays error code indicating temperature measurement or control failure
Ranked causes
- 1. Environmental temperature extremes (common)
- 2. Chamber or tube temperature sensor malfunction (occasional)
Check operating environment temperature
- Test point
- Room where unit is in use
- Expected
- Temperature within normal operating range (not cold or hot)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 too high patient safety
Oxygen concentration has risen above maximum allowed threshold. Displayed on screen or as alert.
Ranked causes
- 1. Oxygen flow rate set too high (common)
- 2. Oxygen limit threshold set too high (occasional)
Reduce oxygen flow rate
- Test point
- Oxygen flow meter on source
- Expected
- Oxygen level reduced below set maximum
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check operating conditions
Device displays alert indicating abnormal operating environment. Often triggered by sudden ambient temperature changes.
Ranked causes
- 1. Ambient room temperature below 10°C or above 30°C (common)
Verify ambient temperature within range
- Test point
- Ambient environment
- Expected
- Room temperature 10–30°C (50–86°F)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Condensation prevention and management
Condensation build-up in heated breathing tube during operation. Risk higher in cold, humid environments or with cooling sources.
Ranked causes
- 1. Ambient room temperature below 18°C or local cooling source (common)
Verify ambient temperature in optimal range
- Test point
- Ambient room environment
- Expected
- Temperature 18–28°C (64–82°F)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power out (black screen)
Display is completely black and auditory alarm sounds for at least 120 seconds.
Ranked causes
- 1. Dislodged or disconnected power cord (common)
Inspect and reseat power cord
- Test point
- Power cord connections at outlet and device
- Expected
- Cord fully inserted and secured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI