Ethicon Total Energy: troubleshooting and error codes

The ServiceForge knowledge base holds 13 documented symptoms for the Ethicon Total Energy, with 13 ranked causes, 40 diagnostic checks with test points and expected values, and 62 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Alert notification displayed

Grey pop-up message shown on User Screen with designated alert; two quick audio tones emitted

Ranked causes

  1. 1. Non-hazardous condition affecting patient, equipment, or workflow (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Advanced Bipolar (ENSEAL) energy output high voltage

Verify ENSEAL Test 1, Test 2, and Test 3 output power at 10, 100, and 500 ohm loads through Advanced port; results recorded on Output Verification Check Form

Ranked causes

  1. 1. Energy module output out of specification (unknown)
First check

Insert ENSEAL test leads into Advanced port

Test point
Advanced port
Expected
Test leads fully inserted and connected

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Advanced Monopolar energy output high voltage

Verify output power at 120 watts in Coag 1 mode through Advanced port with 500 ohm load; results recorded on Output Verification Check Form

Ranked causes

  1. 1. Energy module output out of specification (unknown)
First check

Select Advanced Energy in Output Verification mode

Test point
ESU display
Expected
Advanced mode selected

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Contact Quality Monitoring (CQM) patient safety

Verify patient return electrode port light response at 50, 150, and threshold ohm loads using CQM Safety Leads; confirm threshold is between 63 and 77 ohms

Ranked causes

  1. 1. CQM system out of specification (unknown)
First check

Power on system and prepare for CQM test

Test point
Communications Module on/off button
Expected
System powered on; no patient return electrode attached to Energy Module

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Electrical safety – Earth Leakage Current high voltage

Verify earth leakage current under normal condition, reversed polarity, and single fault (neutral open) conditions does not exceed specification limits

Ranked causes

  1. 1. Leakage current out of specification (unknown)
First check

Measure earth leakage current at 264 V, 60 Hz normal condition

Test point
System leakage to earth/chassis
Expected
≤0.5 mA, ≤0.5 mA
Tools
Electrical safety analyzer, 264 V 60 Hz source

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Electrical safety – Ground Bond test high voltage

Verify total system ground resistance without and with power cord attached meets specification of ≤0.100 ohms or ≤0.200 ohms respectively at 25 A, 60 Hz

Ranked causes

  1. 1. Ground resistance out of specification (unknown)
First check

Measure total system ground resistance without power cord

Test point
System frame to safety ground
Expected
≤0.100 ohms at 25 A, 60 Hz, ≤0.100
Tools
Ground bond analyzer, 25 A 60 Hz source

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Electrical safety – Patient Leakage Current (optional) patient safety

Optional test: Verify patient leakage current under normal condition, reversed polarity, and single fault conditions does not exceed specification limits

Ranked causes

  1. 1. Patient leakage current out of specification (unknown)
First check

Measure patient leakage current at 264 V, 60 Hz normal condition

Test point
Patient circuit leakage to earth
Expected
≤50 uA, ≤50 uA
Tools
Electrical safety analyzer, 264 V 60 Hz source

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Harmonic energy output high voltage

Verify Harmonic Test 1, Test 2, and Test 3 output power at 50, 200, and 800 ohm loads through Advanced port; results recorded on Output Verification Check Form

Ranked causes

  1. 1. Energy module output out of specification (unknown)
First check

Insert HARMONIC test leads into Advanced port

Test point
Advanced port
Expected
Test leads fully inserted and connected

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch Screen test (optional)

Optional test to verify User Screen touch functionality; each cell should change to lighter shade when touched

Ranked causes

  1. 1. User Screen touch sensor defective (rare)
First check

Navigate to Touch Screen Test utility

Test point
User Screen display
Expected
Maintenance menu displayed; Touch Screen Test option visible

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Contact Quality Monitoring alarm triggered patient safety

Flashing red pop-up message on User Screen; Patient Return Port light turns red; three-tone audio alarm sounds; Monopolar output disabled

Ranked causes

  1. 1. Patient return electrode resistance increased by 40% within 10-135 ohms range (unknown)
First check

Measure patient return electrode resistance

Test point
Patient Return Port
Expected
10 to 135 ohms, within range; alarm at +40% increase

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low priority alarm triggered

Yellow pop-up message displayed on User Screen with low priority alarm symbol; two-tone audio alarm sound emitted

Ranked causes

  1. 1. Issue requiring surgical staff attention for procedural progress (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Medium priority alarm triggered patient safety

Flashing yellow pop-up message displayed on User Screen with medium priority alarm symbol; three-tone audio alarm sound emitted

Ranked causes

  1. 1. Hazardous condition requiring prompt attention (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Notification not resolved after on-screen guidance

Alarm, alert, or other event persists despite following displayed prompts and recommended actions

Ranked causes

  1. 1. Unresolved system condition requiring technical support (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
ACCSY-FSW001Footswitch Port ErrorError detected on one or more Footswitch Portserror
ACCSY-IO001-AUSB Port ErrorSystem detected an error on one or more of the back panel USB portserror
ACCSY-IO001-BEthernet Port ErrorSystem detected an error on the back panel Ethernet porterror
ACCSY-IO001-CDigital Accessory Port ErrorSystem detected an error on the back panel Digital Accessory Porterror
Factory reset errorError occurred during factory reset processFailure during factory reset executionerror
HEAD-CFG002Invalid Module(s) DetectedSystem detected an invalid configuration upon startup due to extra, invalid, or unidentified moduleerror
HEAD-CFG003No Module(s) DetectedSystem detected an invalid configuration upon startup due to no modules connectederror
HEAD-CFG007Software/Hardware Conflict DetectedDuring initial setup, system received user input indicating an issue with hardware configurationerror
HEAD-CON001Software Update ErrorSystem could not complete the software update processerror
HEAD-CON002Data Upload FailureSystem could not complete upload of software update confirmationwarning
HEAD-TBL001System Restart RequiredSystem detects error requiring restarterror
HEAD-TBL002System Detected Module Error – System Unavailable for UsePresence of faulty module detectedcritical
HEAD-TBL003System Power LossPower supply disruptioncritical
HEAD-TBL005System ErrorSystem detected a fault that requires further troubleshootingerror
HEAD-TBL006Maintenance Required Soon - Clock Battery LowUpon startup, system detected a low clock batterywarning
HEAD-TBL007Clock Battery Maintenance RequiredClock battery critically lowwarning
HEAD-TBL008Communication Module ErrorError detected in Communications Moduleerror
INST-AE001Replace InstrumentMultiple instances detected requiring instrument replacementwarning
INST-AE001-AReplace Instrument – Continuous Instrument Jaw ErrorOpen circuit condition detected on Advanced Energy instrument after several occurrenceswarning
INST-AE001-BReplace Instrument – Continuous Contact with Metal ErrorAdvanced Energy Instrument activated on metal (staples, clips, retractors, clamps)warning
INST-AE001-CReplace Instrument – Unable to Pass HARMONIC™ Instrument Attachment TestConnected HARMONIC™ instrument fails attachment testerror
INST-AE001-DReplace Instrument – HARMONIC™ Blade Error DetectedError detected with blade of connected HARMONIC™ instrumenterror
INST-AE002Check Tissue Thickness and ReactivateOpen circuit condition detected on Advanced Energy instrument preventing energy deliveryinfo
INST-AE003Reposition Instrument Jaws and ReactivateAdvanced Energy instrument activated on metal (staples, clips, retractors, clamps)info
INST-AE004Connect Compatible Instrument to HARMONIC™ Hand PieceIncompatible HARMONIC™ instrument detectedwarning
INST-AE005Software Upgrade Required to Use This InstrumentInstrument not compatible with installed software versionwarning
INST-AE007Simultaneous Activation Not Available Over 150WSystem detected two Energy Modules each with a Monopolar Instrument with power settings above 150 W allowable thresholdwarning
INST-AE008Simultaneous Activation Not AvailableAdvanced Energy instrument cannot be used simultaneously as power level exceeds allowable system limit or combination with second Advanced Energy instrument exceeds thresholdwarning
INST-AE009Connect Compatible Instrument to the HARMONIC Hand PieceSystem detected a HARMONIC handpiece without an instrument connectedwarning
INST-AE010Remaining Uses on HARMONIC Hand PieceSystem detected the number of uses remaining on a connected hand piece is between 1 and 10info
INST-AE011Replace HARMONIC™ Hand PieceIssue detected with connected hand pieceerror
INST-AE012Replace HARMONIC™ Hand PieceHand piece has reached end of lifeerror
INST-AE014No Uses RemainingInstrument exceeded allowable usage or single-use reusedwarning
INST-AE015HARMONIC Hand Piece TestUser initiated HARMONIC hand piece diagnostic testinfo
INST-AE016Advanced Features Are Not Available in This InstrumentSystem detected an issue with the instrument that prevents use of advanced featureswarning
INST-AIT002Tighten HARMONIC™ Hand Piece AssemblyIssue with attachment between instrument and hand piece detectedwarning
INST-AIT003HARMONIC Instrument Attachment TestUser initiated HARMONIC instrument attachment diagnostic testinfo
INST-BASIC001Set Power Settings to Enable Energy OutputUser attempted to activate an instrument without selecting power settingswarning
MODU-CQM001Check Patient Return ElectrodeIssue with quality of contact or connection with patient return electrode detectedwarning
MODU-GEN001Software Update Required to Use the ModuleSystem detected one or more modules upon startup with module software version that does not match currently installed system software versionwarning
MODU-GEN004Energy Module ErrorError detected with connected Energy Modulecritical
MODU-GEN007Energy Module ErrorError detected with connected energy moduleerror
Software update errorError occurred during software update processFailure during Ethernet or USB software update executionerror
SYST-CLNCL001Return to Main Screen to Activate InstrumentUser attempting to activate instrument while viewing non-clinical screensinfo
SYST-LSE001System ErrorSystem detects fault and stops activationwarning
SYST-OVH001Cooling Fan ErrorLoss of cooling fan operation detectederror
SYST-OVH002Module OverheatingOne or more modules overheatingwarning
USER-ACT002Footswitch Not Assigned to InstrumentSystem detected user depression of a footswitch assigned to a port without an instrument or where instrument is not properly connectedwarning
USER-ACT003Instrument Jaws OpenSystem detected HARMONIC instrument jaws are in an open state during activation mode that requires closed jawswarning
USER-ACT004Release Secondary InputSystem detected activation request for second energy modality while another energy modality on same Energy Module is already activewarning
USER-ACT005Reactivate to Continue SealingSystem detected an ENSEAL activation lasting longer than 2 minutesinfo
USER-ACT006-ARelease Hand SwitchSystem detected a depressed activation button upon instrument connection or system startupwarning
USER-ACT006-BRelease FootswitchSystem detected a depressed footswitch pedal upon instrument connection or system startupwarning
USER-ACT007Simultaneous Activation Not AllowedSystem detected an attempted activation on instrument(s) where simultaneous activation of two instruments has been disabledwarning
USER-ACT008Current Threshold ExceededAccumulated current during monopolar activation exceeds allowable thresholdwarning
USER-ACT012-ARelease All Activation SwitchesSystem detected a depressed switch on an instrument or footswitch pedal when energy is not being deliveredwarning
USER-ACT013Open Jaws to Enable Auto BipolarPrior to delivering therapeutic output, system detected clamped tissue in jaws on a bipolar instrument while Auto Bipolar is enabledwarning
USER-BIT001HARMONIC Blade Approaching Maximum ImpedanceSystem detected excessive force against the blade on a HARMONIC instrumentwarning
USER-BIT002Maximum Impedance on HARMONIC™ Blade DetectedInstrument loaded to point where output stopswarning
USER-BIT003HARMONIC™ Blade Error DetectedSystem detects need to diagnose bladeerror
USER-OV001Disconnect Instruments to Enter Energy Output Verification ModeSystem detected the Output Verification Key while instruments or leads are connected to one or more Energy Module portswarning
USER-OV003Connect Energy Output Verification Key to Test This Energy ModuleEnergy Output Verification attempted without Output Verification Key connectedwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.