Ethicon Total Energy: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Ethicon Total Energy, with 13 ranked causes, 40 diagnostic checks with test points and expected values, and 62 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Alert notification displayed
Grey pop-up message shown on User Screen with designated alert; two quick audio tones emitted
Ranked causes
- 1. Non-hazardous condition affecting patient, equipment, or workflow (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Advanced Bipolar (ENSEAL) energy output high voltage
Verify ENSEAL Test 1, Test 2, and Test 3 output power at 10, 100, and 500 ohm loads through Advanced port; results recorded on Output Verification Check Form
Ranked causes
- 1. Energy module output out of specification (unknown)
Insert ENSEAL test leads into Advanced port
- Test point
- Advanced port
- Expected
- Test leads fully inserted and connected
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Advanced Monopolar energy output high voltage
Verify output power at 120 watts in Coag 1 mode through Advanced port with 500 ohm load; results recorded on Output Verification Check Form
Ranked causes
- 1. Energy module output out of specification (unknown)
Select Advanced Energy in Output Verification mode
- Test point
- ESU display
- Expected
- Advanced mode selected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Contact Quality Monitoring (CQM) patient safety
Verify patient return electrode port light response at 50, 150, and threshold ohm loads using CQM Safety Leads; confirm threshold is between 63 and 77 ohms
Ranked causes
- 1. CQM system out of specification (unknown)
Power on system and prepare for CQM test
- Test point
- Communications Module on/off button
- Expected
- System powered on; no patient return electrode attached to Energy Module
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electrical safety – Earth Leakage Current high voltage
Verify earth leakage current under normal condition, reversed polarity, and single fault (neutral open) conditions does not exceed specification limits
Ranked causes
- 1. Leakage current out of specification (unknown)
Measure earth leakage current at 264 V, 60 Hz normal condition
- Test point
- System leakage to earth/chassis
- Expected
- ≤0.5 mA, ≤0.5 mA
- Tools
- Electrical safety analyzer, 264 V 60 Hz source
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electrical safety – Ground Bond test high voltage
Verify total system ground resistance without and with power cord attached meets specification of ≤0.100 ohms or ≤0.200 ohms respectively at 25 A, 60 Hz
Ranked causes
- 1. Ground resistance out of specification (unknown)
Measure total system ground resistance without power cord
- Test point
- System frame to safety ground
- Expected
- ≤0.100 ohms at 25 A, 60 Hz, ≤0.100
- Tools
- Ground bond analyzer, 25 A 60 Hz source
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electrical safety – Patient Leakage Current (optional) patient safety
Optional test: Verify patient leakage current under normal condition, reversed polarity, and single fault conditions does not exceed specification limits
Ranked causes
- 1. Patient leakage current out of specification (unknown)
Measure patient leakage current at 264 V, 60 Hz normal condition
- Test point
- Patient circuit leakage to earth
- Expected
- ≤50 uA, ≤50 uA
- Tools
- Electrical safety analyzer, 264 V 60 Hz source
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Harmonic energy output high voltage
Verify Harmonic Test 1, Test 2, and Test 3 output power at 50, 200, and 800 ohm loads through Advanced port; results recorded on Output Verification Check Form
Ranked causes
- 1. Energy module output out of specification (unknown)
Insert HARMONIC test leads into Advanced port
- Test point
- Advanced port
- Expected
- Test leads fully inserted and connected
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch Screen test (optional)
Optional test to verify User Screen touch functionality; each cell should change to lighter shade when touched
Ranked causes
- 1. User Screen touch sensor defective (rare)
Navigate to Touch Screen Test utility
- Test point
- User Screen display
- Expected
- Maintenance menu displayed; Touch Screen Test option visible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contact Quality Monitoring alarm triggered patient safety
Flashing red pop-up message on User Screen; Patient Return Port light turns red; three-tone audio alarm sounds; Monopolar output disabled
Ranked causes
- 1. Patient return electrode resistance increased by 40% within 10-135 ohms range (unknown)
Measure patient return electrode resistance
- Test point
- Patient Return Port
- Expected
- 10 to 135 ohms, within range; alarm at +40% increase
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low priority alarm triggered
Yellow pop-up message displayed on User Screen with low priority alarm symbol; two-tone audio alarm sound emitted
Ranked causes
- 1. Issue requiring surgical staff attention for procedural progress (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Medium priority alarm triggered patient safety
Flashing yellow pop-up message displayed on User Screen with medium priority alarm symbol; three-tone audio alarm sound emitted
Ranked causes
- 1. Hazardous condition requiring prompt attention (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Notification not resolved after on-screen guidance
Alarm, alert, or other event persists despite following displayed prompts and recommended actions
Ranked causes
- 1. Unresolved system condition requiring technical support (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ACCSY-FSW001 | Footswitch Port Error | Error detected on one or more Footswitch Ports | error |
ACCSY-IO001-A | USB Port Error | System detected an error on one or more of the back panel USB ports | error |
ACCSY-IO001-B | Ethernet Port Error | System detected an error on the back panel Ethernet port | error |
ACCSY-IO001-C | Digital Accessory Port Error | System detected an error on the back panel Digital Accessory Port | error |
Factory reset error | Error occurred during factory reset process | Failure during factory reset execution | error |
HEAD-CFG002 | Invalid Module(s) Detected | System detected an invalid configuration upon startup due to extra, invalid, or unidentified module | error |
HEAD-CFG003 | No Module(s) Detected | System detected an invalid configuration upon startup due to no modules connected | error |
HEAD-CFG007 | Software/Hardware Conflict Detected | During initial setup, system received user input indicating an issue with hardware configuration | error |
HEAD-CON001 | Software Update Error | System could not complete the software update process | error |
HEAD-CON002 | Data Upload Failure | System could not complete upload of software update confirmation | warning |
HEAD-TBL001 | System Restart Required | System detects error requiring restart | error |
HEAD-TBL002 | System Detected Module Error – System Unavailable for Use | Presence of faulty module detected | critical |
HEAD-TBL003 | System Power Loss | Power supply disruption | critical |
HEAD-TBL005 | System Error | System detected a fault that requires further troubleshooting | error |
HEAD-TBL006 | Maintenance Required Soon - Clock Battery Low | Upon startup, system detected a low clock battery | warning |
HEAD-TBL007 | Clock Battery Maintenance Required | Clock battery critically low | warning |
HEAD-TBL008 | Communication Module Error | Error detected in Communications Module | error |
INST-AE001 | Replace Instrument | Multiple instances detected requiring instrument replacement | warning |
INST-AE001-A | Replace Instrument – Continuous Instrument Jaw Error | Open circuit condition detected on Advanced Energy instrument after several occurrences | warning |
INST-AE001-B | Replace Instrument – Continuous Contact with Metal Error | Advanced Energy Instrument activated on metal (staples, clips, retractors, clamps) | warning |
INST-AE001-C | Replace Instrument – Unable to Pass HARMONIC™ Instrument Attachment Test | Connected HARMONIC™ instrument fails attachment test | error |
INST-AE001-D | Replace Instrument – HARMONIC™ Blade Error Detected | Error detected with blade of connected HARMONIC™ instrument | error |
INST-AE002 | Check Tissue Thickness and Reactivate | Open circuit condition detected on Advanced Energy instrument preventing energy delivery | info |
INST-AE003 | Reposition Instrument Jaws and Reactivate | Advanced Energy instrument activated on metal (staples, clips, retractors, clamps) | info |
INST-AE004 | Connect Compatible Instrument to HARMONIC™ Hand Piece | Incompatible HARMONIC™ instrument detected | warning |
INST-AE005 | Software Upgrade Required to Use This Instrument | Instrument not compatible with installed software version | warning |
INST-AE007 | Simultaneous Activation Not Available Over 150W | System detected two Energy Modules each with a Monopolar Instrument with power settings above 150 W allowable threshold | warning |
INST-AE008 | Simultaneous Activation Not Available | Advanced Energy instrument cannot be used simultaneously as power level exceeds allowable system limit or combination with second Advanced Energy instrument exceeds threshold | warning |
INST-AE009 | Connect Compatible Instrument to the HARMONIC Hand Piece | System detected a HARMONIC handpiece without an instrument connected | warning |
INST-AE010 | Remaining Uses on HARMONIC Hand Piece | System detected the number of uses remaining on a connected hand piece is between 1 and 10 | info |
INST-AE011 | Replace HARMONIC™ Hand Piece | Issue detected with connected hand piece | error |
INST-AE012 | Replace HARMONIC™ Hand Piece | Hand piece has reached end of life | error |
INST-AE014 | No Uses Remaining | Instrument exceeded allowable usage or single-use reused | warning |
INST-AE015 | HARMONIC Hand Piece Test | User initiated HARMONIC hand piece diagnostic test | info |
INST-AE016 | Advanced Features Are Not Available in This Instrument | System detected an issue with the instrument that prevents use of advanced features | warning |
INST-AIT002 | Tighten HARMONIC™ Hand Piece Assembly | Issue with attachment between instrument and hand piece detected | warning |
INST-AIT003 | HARMONIC Instrument Attachment Test | User initiated HARMONIC instrument attachment diagnostic test | info |
INST-BASIC001 | Set Power Settings to Enable Energy Output | User attempted to activate an instrument without selecting power settings | warning |
MODU-CQM001 | Check Patient Return Electrode | Issue with quality of contact or connection with patient return electrode detected | warning |
MODU-GEN001 | Software Update Required to Use the Module | System detected one or more modules upon startup with module software version that does not match currently installed system software version | warning |
MODU-GEN004 | Energy Module Error | Error detected with connected Energy Module | critical |
MODU-GEN007 | Energy Module Error | Error detected with connected energy module | error |
Software update error | Error occurred during software update process | Failure during Ethernet or USB software update execution | error |
SYST-CLNCL001 | Return to Main Screen to Activate Instrument | User attempting to activate instrument while viewing non-clinical screens | info |
SYST-LSE001 | System Error | System detects fault and stops activation | warning |
SYST-OVH001 | Cooling Fan Error | Loss of cooling fan operation detected | error |
SYST-OVH002 | Module Overheating | One or more modules overheating | warning |
USER-ACT002 | Footswitch Not Assigned to Instrument | System detected user depression of a footswitch assigned to a port without an instrument or where instrument is not properly connected | warning |
USER-ACT003 | Instrument Jaws Open | System detected HARMONIC instrument jaws are in an open state during activation mode that requires closed jaws | warning |
USER-ACT004 | Release Secondary Input | System detected activation request for second energy modality while another energy modality on same Energy Module is already active | warning |
USER-ACT005 | Reactivate to Continue Sealing | System detected an ENSEAL activation lasting longer than 2 minutes | info |
USER-ACT006-A | Release Hand Switch | System detected a depressed activation button upon instrument connection or system startup | warning |
USER-ACT006-B | Release Footswitch | System detected a depressed footswitch pedal upon instrument connection or system startup | warning |
USER-ACT007 | Simultaneous Activation Not Allowed | System detected an attempted activation on instrument(s) where simultaneous activation of two instruments has been disabled | warning |
USER-ACT008 | Current Threshold Exceeded | Accumulated current during monopolar activation exceeds allowable threshold | warning |
USER-ACT012-A | Release All Activation Switches | System detected a depressed switch on an instrument or footswitch pedal when energy is not being delivered | warning |
USER-ACT013 | Open Jaws to Enable Auto Bipolar | Prior to delivering therapeutic output, system detected clamped tissue in jaws on a bipolar instrument while Auto Bipolar is enabled | warning |
USER-BIT001 | HARMONIC Blade Approaching Maximum Impedance | System detected excessive force against the blade on a HARMONIC instrument | warning |
USER-BIT002 | Maximum Impedance on HARMONIC™ Blade Detected | Instrument loaded to point where output stops | warning |
USER-BIT003 | HARMONIC™ Blade Error Detected | System detects need to diagnose blade | error |
USER-OV001 | Disconnect Instruments to Enter Energy Output Verification Mode | System detected the Output Verification Key while instruments or leads are connected to one or more Energy Module ports | warning |
USER-OV003 | Connect Energy Output Verification Key to Test This Energy Module | Energy Output Verification attempted without Output Verification Key connected | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.