Essex Cryogenics of Missouri CRU-87/U: troubleshooting and error codes
The ServiceForge knowledge base holds 44 documented symptoms for the Essex Cryogenics of Missouri CRU-87/U, with 68 ranked causes, 74 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Converter will not fill patient safety
Filling operation cannot be initiated or completed
Ranked causes
- 1. Ice in filler valve or filler line obstructs LOX flow (common)
- 2. Filler line not properly purged prior to filling (common)
- 3. Converter not sub-cooled before filling (occasional)
- 4. Filling pressure too low (occasional)
- 5. Defective converter (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Container assembly not filling with liquid oxygen high voltage
Liquid oxygen does not flow from vent in solid stream during fill procedure; gaseous spurts interrupt flow or no LOX flows
Ranked causes
- 1. Servicing hose not fully purged before filling (common)
- 2. Servicing pressure below forty (40) PSIG (common)
- 3. Container assembly already 75% or more full (false fill condition) (occasional)
Verify servicing hose is purged
- Test point
- oxygen tank trailer servicing hose outlet
- Expected
- liquid oxygen flows from hose
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter does not fill in required time patient safety
Filling operation proceeds but takes longer than specified
Ranked causes
- 1. Filling pressure too low (common)
- 2. Converter not sub-cooled before filling (occasional)
- 3. Defective converter (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter will fill only partially with gas only emitted from vent patient safety
Liquid oxygen does not accumulate; only gas exits vent valve
Ranked causes
- 1. Converter not sub-cooled prior to filling (common)
- 2. Buildup, vent and filler valve defective or partially open (occasional)
- 3. Pressure relief valve open (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter will not build up pressure patient safety
Converter fails to maintain or develop required pressure after filling
Ranked causes
- 1. System leakage (common)
- 2. Fill, buildup and vent valve partially open venting gas (occasional)
- 3. Pressure relief valve open (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insufficient oxygen delivery rate patient safety
Converter not delivering required oxygen flow to aircraft systems.
Ranked causes
- 1. LOX level too low in converter (common)
- 2. Heat exchanger blockage or malfunction (occasional)
- 3. Pressure regulator failure (occasional)
Measure LOX level in converter
- Test point
- Converter sight glass or level indicator
- Expected
- Adequate LOX visible; converter should be near full capacity liters
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or no converter pressure patient safety
Pressure gauge indicates below normal or zero pressure
Ranked causes
- 1. Converter empty, defective probe or quantity indicator (common)
- 2. System leakage (occasional)
- 3. Pressure relief valve open or out of adjustment (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Quantity indicator using TTU-62/E Tester
Calibrate oxygen quantity indicator using TTU-62/E Tester capacitance measurement at empty and full states
Ranked causes
- 1. Quantity indicator requires calibration via capacitance tester (occasional)
- 2. Battery power or indicator failure (rare)
Connect tester to converter probes and verify reading within limits
- Test point
- Converter probes
- Expected
- Within limits in T.O. 15X-1-1 Section VIII Figure 8-4 (10 Liter Capacitance Chart)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter fails to reach operating pressure after filling
Pressure closing valve not building pressure to nominal 75 psig after refill operation.
Ranked causes
- 1. Pressure closing valve bellows leak (occasional)
- 2. Buildup coil obstruction or leak (rare)
Measure buildup time after filling
- Test point
- Converter outlet pressure gauge
- Expected
- Pressure reaches 75 psig within 5 minutes of fill completion, ±5 psig
- Tools
- Pressure gauge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter pressure exceeds safe limits high voltage
Relief valve does not function; internal pressure could exceed 12,000 psig if demand not placed on converter, creating hazardous overpressure situation.
Ranked causes
- 1. Relief valve failed or not set correctly (unknown)
- 2. Pressure closing valve malfunction (unknown)
Verify relief valve setting
- Test point
- Relief valve outlet
- Expected
- Pressure relief occurs at approximately 110 psig, ±10 psig
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive pressure in servicing hose during fill high voltage
Pressure buildup in oxygen tank trailer servicing hose during filling operation
Ranked causes
- 1. Servicing hose not fully depressurized before disconnecting (common)
- 2. Pressure relief valve not opened or not equipped (occasional)
Close servicing line shutoff valve
- Test point
- oxygen tank trailer servicing line shutoff valve (V-3)
- Expected
- valve fully closed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive time to drain liquid oxygen from container assembly caution
Drain cycle takes longer than expected 3-hour baseline for 10-liter capacity with three valves at maximum flow
Ranked causes
- 1. Flow control valves not set to maximum flow rate during drain (common)
- 2. One or more flow control valves delivering below rated capacity (occasional)
Verify each flow control valve set to maximum position
- Test point
- Flow control valve handle position
- Expected
- All three valves at maximum 15 LPM position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid oxygen leakage from container assembly patient safety
LOX escaping from outlet receptacles, fill port, or supply hoses during operation or storage
Ranked causes
- 1. Supply hose not properly connected to container assembly outlet receptacle (common)
- 2. Flow control valve not set to zero position before disconnecting supply hoses (occasional)
Inspect LOX fill port for condition and function
- Test point
- LOX fill port
- Expected
- General good condition, smooth and positive action, freedom from all contamination, dust cover in place
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quantity Indicator indicates empty patient safety
Indicator displays empty reading when converter contains liquid
Ranked causes
- 1. Dead batteries (common)
- 2. Converter empty, defective probe or quantity indicator (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quantity Indicator no reading patient safety
Indicator display is blank or unresponsive
Ranked causes
- 1. Dead batteries (common)
- 2. Converter empty, defective probe or quantity indicator (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen pressure relief valve high voltage
Test and calibrate Type V oxygen pressure relief valve with operating range 120-140 PSIG
Ranked causes
- 1. Pressure relief valve requires calibration per manual procedure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure regulator high voltage
Set master pressure gauge to 50 PSIG with prescribed oxygen flow (45 LPM total) before operation
Ranked causes
- 1. Pressure regulator requires calibration per procedure (common)
Verify pressure indication has stabilized on master pressure gauge
- Test point
- Master pressure gauge
- Expected
- Stabilized reading
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Quantity indicator using weight method
Calibrate oxygen quantity indicator by measuring indicator reading against actual liquid oxygen weight in five 2-liter increments
Ranked causes
- 1. Quantity indicator requires calibration verification (common)
Ensure all LOX drained and indicator registers zero
- Test point
- Quantity indicator with OPERATE switch
- Expected
- Zero reading
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Container assembly lacks positive nitrogen pressure
Pressure gauge does not register a reading after shipment
Ranked causes
- 1. Nitrogen pressure lost during shipment (occasional)
Verify pressure gauge reading on arrival
- Test point
- Pressure gauge
- Expected
- Positive pressure of gaseous nitrogen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Container assembly leakage test high voltage
Pneumatic pressure test to verify all fittings and connections are leak-free
Ranked causes
- 1. Leakage test required for container integrity verification (common)
Remove front panel assembly
- Test point
- Container assembly front
- Expected
- Panel fully removed
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Container assembly pressure below 45 PSIG
Pressure gauge reading indicates pressure lower than minimum safe operating pressure during operation or standby
Ranked causes
- 1. Insufficient pressure in container assembly (unknown)
Check container assembly pressure gauge for proper operating pressure
- Test point
- Pressure gauge on top surface of container assembly
- Expected
- 50+5 psi, +/-5 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Container assembly temperature unstable after filling caution
Container assembly temperature has not stabilized after liquid oxygen fill operation before transport or loading
Ranked causes
- 1. Insufficient stabilization time after filling (common)
Wait for stabilization period
- Test point
- container assembly
- Expected
- thirty (30) minutes elapsed since fill completion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter purging procedure high voltage
Procedure to purge converter assembly using heated nitrogen/air followed by cold oxygen purge to remove contaminants and moisture before use
Ranked causes
- 1. Converter requires purging prior to operation (common)
Ascertain converter assembly is completely drained and empty
- Test point
- Converter assembly
- Expected
- Empty with no residual liquid oxygen
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filler line cannot be disconnected from filler valve caution
Filler line assembly stuck or frozen to valve connection
Ranked causes
- 1. Filler nozzle frozen to filler valve (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow control valve delivers incorrect flow rate patient safety
Rotameter reading does not match expected flow tolerance at a given valve position setting during verification test
Ranked causes
- 1. Flow control valve out of specification (unknown)
Verify flow control valve flow rate at each position
- Test point
- Rotameter outlet of flow control valve
- Expected
- Position 0: No leakage; Position 2: 2.0±0.35 SLPM; Position 4: 4.0±0.59 SLPM; Position 6: 6.0±0.88 SLPM; Position 8: 8.0±1.17 SLPM; Position 10: 10.0±1.47 SLPM; Position 12: 12.0±1.77 SLPM; Position 15: 15.0±2.21 SLPM PSIG, ±5 PSIG
- Tools
- Rotameter (0-20 LPM), master pressure gauge (0-100 psig), tee fitting, tubing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frost collects on entire outer jacket of converter caution
Visible frost or ice accumulation on external converter surface
Ranked causes
- 1. Heat loss due to annular space leakage (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect flow control valve connection patient safety
Reverse connection of inlet/outlet fittings or non-compliant gas source may cause inaccurate oxygen flow rates and patient injury
Ranked causes
- 1. Incorrect fitting connection or non-standard gas pressure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid oxygen proximity to combustible materials caution
Equipment operated near absorbent materials, loose clothing, rags, gasoline, kerosene, oil, grease or other hydrocarbons, creating fire or spontaneous ignition hazard.
Ranked causes
- 1. Equipment located near combustible materials or hydrocarbons (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid oxygen quantity indicator battery test reading outside acceptable range
Activation of test switch on quantity indicator does not display reading between 10.00 and 19.99 liters
Ranked causes
- 1. Indicator storage battery depleted or faulty (unknown)
Activate indicator test switch and verify battery power
- Test point
- Liquid crystal digital display on quantity indicator
- Expected
- Display shows reading between 10.00 and 19.99 liters, 10.00 to 19.99
Liquid oxygen skin contact patient safety
Skin exposed to liquid oxygen drops or splashes freezes immediately, causing severe frostbite and frozen tissue.
Ranked causes
- 1. Operator handling error or equipment malfunction causing spillage (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loud or continuous hissing sound with vapor visible patient safety
Audible hissing from container assembly vent line that is loud and continuous, or vapor is observed venting
Ranked causes
- 1. Abnormal venting or pressure release from container assembly (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LOX converter contaminated patient safety
Converter contaminated with undesirable odors, moisture or foreign substances
Ranked causes
- 1. Contamination from non-compatible materials during servicing (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen delivery pressure falls below 45 PSIG at 15 LPM
During flow check at 15 LPM setting, measured pressure at outlet drops below minimum of 45 PSIG
Ranked causes
- 1. Insufficient pressure regulation or supply pressure low (unknown)
Check oxygen delivery pressure at 15 LPM flow setting
- Test point
- Supply outlet after flow control valve set to 15 LPM
- Expected
- Pressure does not fall below 45 PSIG
Oxygen equipment near combustible materials patient safety
Risk of spontaneous ignition or rapid combustion when liquid oxygen equipment is near hydrocarbons or absorbent materials
Ranked causes
- 1. LOX equipment proximity to incompatible substances (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen odor detected other than normal hose smell patient safety
Emitted oxygen smells of substances other than the normal hose odor during flow check
Ranked causes
- 1. Contamination in oxygen system (unknown)
Smell emitted oxygen for any odors
- Test point
- Oxygen outlet at 15 LPM
- Expected
- Normal hose smell only
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen supply consumed too quickly patient safety
Pressure or quantity decreases faster than normal during use
Ranked causes
- 1. System leakage (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure gauge calibration check high voltage
Verify converter pressure gauge accuracy against calibrated master gauge
Ranked causes
- 1. Pressure gauge requires calibration verification (common)
Fill container assembly with liquid oxygen
- Test point
- Filler valve
- Expected
- Container filled with liquid oxygen
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure regulator calibration check high voltage
Verify pressure regulator output against master gauge during operation
Ranked causes
- 1. Regulator requires pressure testing and adjustment (common)
Remove front panel screws and panel
- Test point
- Container assembly front
- Expected
- Panel removed to access regulator
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quantity indicator battery power depleted
Indicator does not display reading when membrane test switch activated
Ranked causes
- 1. BA-3090/U batteries discharged or missing (common)
Activate membrane test switch and observe reading
- Test point
- Quantity indicator display
- Expected
- Reading between 10.00 and 19.99
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quantity indicator display not responding
Quantity indicator does not show reading when TEST switch is actuated; battery power may be insufficient
Ranked causes
- 1. Quantity indicator storage battery depleted (common)
Verify quantity indicator battery power by activating TEST switch
- Test point
- Quantity indicator display
- Expected
- Reading between 10.00 and 19.99
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse connection of flow control valve inlet and outlet patient safety
Flow control valve connected with inlet to humidifier and outlet to container assembly instead of correct orientation
Ranked causes
- 1. Operator connected valve fittings in reverse order (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse inlet/outlet connection patient safety
Flow control valve fittings connected in reverse order or supplied with incorrect gas source pressure, causing inaccurate oxygen flow and potential patient injury.
Ranked causes
- 1. Inlet/outlet fitting reversed or incorrect gas source (not 50±5 psi oxygen) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unobstructed vent line required high voltage
Vent line or fitting becomes obstructed, preventing safe pressure release from liquid oxygen expansion.
Ranked causes
- 1. Vent line/fitting obstruction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent line obstruction high voltage
Pressure buildup when liquid oxygen expands to gas with obstructed vent line or fitting
Ranked causes
- 1. Vent line or fitting blocked or obstructed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI