ERBE VIO 300D: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the ERBE VIO 300D, with 21 ranked causes, 42 diagnostic checks with test points and expected values, and 210 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Neutral electrode critical resistance monitoring dual surface failure patient safety
Neutral electrode lamp does not light red or green; unit activates without warning when should inhibit
Ranked causes
- 1. NE monitoring circuit malfunction for dual surface electrode (unknown)
- 2. Asymmetrical dual surface electrode monitoring malfunction (unknown)
Dual surface neutral electrode lamp color
- Test point
- NE socket connected to VIO Testbox 20100-101
- Expected
- Red lamp indication
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neutral electrode critical resistance monitoring single surface failure patient safety
Neutral electrode lamp does not light red or green; unit activates without warning when should inhibit
Ranked causes
- 1. NE monitoring circuit malfunction for single surface electrode (unknown)
- 2. NE circuit passes but monitoring disabled or second stage malfunction (unknown)
Neutral electrode lamp color on unit display
- Test point
- NE socket connected to VIO Testbox 20100-101
- Expected
- Red lamp indication
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acknowledgement tones not emitted
Unit fails to produce acoustic signals during system start, button presses, or pedal activation
Ranked causes
- 1. Acoustic signaling system malfunction (unknown)
Check acknowledgement tone during system start
- Test point
- Speaker/audio output
- Expected
- Acoustic signal emitted during self-test
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic start mode delay not functioning patient safety
Unit does not start after set delay in AUTO START 1 mode or fails to terminate on command
Ranked causes
- 1. Auto start timing circuit or command processing failure (unknown)
Verify auto start delay timing
- Test point
- VIO Testbox T1 button activation
- Expected
- Unit starts after set AUTO START 1 delay (record delay value from Setup) watts
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bipolar resection adapter decoupling capacitor insulation failure high voltage
Decoupling capacitor insulation resistance falls below specification; potential for electrical leakage
Ranked causes
- 1. Decoupling capacitor insulation degradation (rare)
Decoupling capacitor DC insulation resistance
- Test point
- Bipolar resection adapter MF socket to safety tester
- Expected
- Resistance > 200 MOhm, > 200
- Tools
- Safety tester with insulation testing > 500 V DC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bipolar resection adapter HF power output low high voltage
HF power delivered into 50 ohm load falls below specification; output stage malfunction suspected
Ranked causes
- 1. RF power amplifier or output coupling malfunction (occasional)
HF power meter impedance configuration
- Test point
- HF power meter selector
- Expected
- RL = 50 ohms
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bipolar resection adapter transfer ratio out of specification high voltage
HF output voltage transfer ratio does not meet specification; indicates RF amplifier or transformer issue
Ranked causes
- 1. RF transformer or HF amplifier output stage malfunction (occasional)
Setup level 2 access and test program selection
- Test point
- VIO 300 D menu system
- Expected
- Navigate to Test programs > Hardware TP > TP generator
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control buttons or pushbuttons not responsive
Selection buttons, focus buttons, or enter button fail to respond or function intermittently
Ranked causes
- 1. Button contact or switch failure (unknown)
Test control button functionality
- Test point
- Control panel buttons
- Expected
- Smooth operation, acknowledgement tone emitted with each press
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display backlighting not functional
Display lacks illumination or backlighting is dim or faulty
Ranked causes
- 1. Backlighting circuit or LED failure (unknown)
Verify display backlighting operation
- Test point
- Display screen
- Expected
- Backlighting illuminates properly, image visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dual-pedal footswitch CUT or COAG activation not registered
Dual-pedal footswitch CUT or COAG pedal does not register in test program or MF icon does not illuminate
Ranked causes
- 1. Dual-pedal footswitch contact or cable malfunction (unknown)
Test CUT pedal on dual-pedal footswitch
- Test point
- CUT pedal on dual-pedal footswitch art. no. 20189-303
- Expected
- MF icon field lights yellow in test program
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fingerswitch CUT or COAG buttons not responsive
Electrode handle CUT or COAG buttons fail to activate or do not show visual confirmation in test program
Ranked causes
- 1. Electrode handle button contact failure (unknown)
Test CUT button on electrode handle (Monopolar)
- Test point
- CUT button on handle art. no. 20190-045
- Expected
- Handpiece icon field lights in color in test program
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch CUT or COAG pedal not responsive
Single or dual-pedal footswitch pedals fail to activate or produce no acknowledgement tone
Ranked causes
- 1. Footswitch pedal contact or cable failure (unknown)
Test CUT pedal activation
- Test point
- CUT pedal on dual-pedal footswitch
- Expected
- Acknowledgement tone emitted each time pedal pressed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument detection via EEPROM not working
Multifunction socket does not detect connected instrument or COAG mode does not switch to BiClamp automatically
Ranked causes
- 1. EEPROM detection circuit or sensor failure (unknown)
Test instrument detection on MF socket
- Test point
- MF socket with measuring cable art. no. 20100-152
- Expected
- System displays instrument detection advisory message, COAG mode switches to BiClamp
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEDs do not light up at startup
Status LEDs fail to illuminate or light up at inconsistent intensity during system startup
Ranked causes
- 1. LED circuit or individual LED failure (unknown)
Observe LED illumination at startup
- Test point
- All status LEDs
- Expected
- All LEDs light briefly at same intensity during system start
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power switch does not operate smoothly
Power switch sticks, scrapes, or is difficult to operate
Ranked causes
- 1. Power switch mechanical defect (unknown)
Operate power switch manually
- Test point
- Power switch
- Expected
- Smooth operation, snaps into ON position without sticking or scraping
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ReMode button not switching modes
ReMode button on footswitch or electrode handle fails to switch between modes or does not show visual confirmation
Ranked causes
- 1. ReMode button contact failure (unknown)
Test ReMode button on dual-pedal footswitch
- Test point
- ReMode button on dual-pedal footswitch art. no. 20189-303
- Expected
- MF icon field lights in color in test program confirming mode switchover
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Single-pedal footswitch activation not registered
Single-pedal footswitch COAG pedal does not register activation in test program
Ranked causes
- 1. Single-pedal footswitch contact or cable malfunction (unknown)
Test single-pedal footswitch COAG activation
- Test point
- COAG pedal on single-pedal footswitch art. no. 20188-300
- Expected
- Single-pedal icon lights in color in test program
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not start or displays error messages
Unit fails to perform system start or shows error messages during startup
Ranked causes
- 1. System initialization failure (unknown)
Verify system start without errors
- Test point
- Display
- Expected
- System performs start routine without error messages
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning message C-84 received
BiClamp short-circuit protection activated and interrupted cutting operation.
Ranked causes
- 1. Short-circuit detected in BiClamp circuit (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
2,3,5,6-1-7 | Socket IIF error | Instrument Interface error in socket 2/3/5/6 | error |
2,3,5,6-70-72 | Socket IIF error | Instrument Interface error in socket 2/3/5/6 | error |
2,3,5,6-8 | Overcurrent activation recognition | Measurement current too high during activation recognition | error |
2,3,5,6-81 | Short circuit instrument recognition R | Instrument recognition resistor value indicates short circuit | error |
2,3,5,6-82 | Short circuit activation recognition R_NEU | Activation recognition resistor value indicates short circuit | error |
2,3,5,6-9 | Socket IIF error | Instrument Interface error in socket 2/3/5/6 | error |
2,3,5,6-A | Overcurrent instrument recognition | Measurement current too high during instrument recognition | error |
2,3,5,6-C-E | Socket IIF error | Instrument Interface error in socket 2/3/5/6 | error |
2,3,5,6-F | U Coag error: +12V voltage too low | +12V voltage during Coag activation drops below threshold | error |
4(NE)1 | Nessy2 generic error | Unspecified error condition in Nessy2 module | error |
4(NE)10 | Nessy2 generic error | Unspecified error condition in Nessy2 module | error |
4(NE)13 | Nessy2 generic error | Unspecified error condition in Nessy2 module | error |
4(NE)C | Nessy2 generic error | Unspecified error condition in Nessy2 module | error |
83 | No-load activation recognition R_NEU: resistor value determined for activation recognition corresponds to no-load | Instrument interface socket detects no-load condition during activation recognition | error |
84 | Instrument recognition R-window violated: resistor value cannot be reliably assigned to setpoint | Instrument interface socket cannot match resistor value to valid setpoint | error |
85 | Short circuit activation recognition R_ALT: resistor value corresponds to short circuit | Instrument interface socket detects short circuit condition during activation recognition | error |
86 | Activation error: activation not consistent with instrument type | Activation attempt inconsistent with connected instrument type | error |
87 | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
88 | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
89 | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
8A | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
8B | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
8C | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
8D | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
8E | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
8F | Generic module error (IIF socket) | Unspecified error condition in instrument interface module | error |
90 | Generic module error (IIF socket or Nessy2 or EIP 2) | Unspecified error condition in module | error |
980 | Erbe Irrigation Pump EIP 2 generic error | Unspecified error condition in EIP 2 pump module | error |
982 | Pump lid open at ON: EIP 2 pump lid not closed at start or opened during activation | Pump lid detected open during activation | warning |
984 | Time Out Error: EIP 2 activation maintained longer than maximum 20 seconds | EIP 2 activation continuous for more than 20 seconds | warning |
988 | Erbe Irrigation Pump EIP 2 generic error | Unspecified error condition in EIP 2 pump module | error |
990 | Erbe Irrigation Pump EIP 2 generic error | Unspecified error condition in EIP 2 pump module | error |
9A0 | Erbe Irrigation Pump EIP 2 generic error | Unspecified error condition in EIP 2 pump module | error |
9C0 | Erbe Irrigation Pump EIP 2 generic error | Unspecified error condition in EIP 2 pump module | error |
A1 | APC 2 module error code 1 | Unknown fault condition in APC 2 | error |
A10 | Underpressure at selected gas input | Gas pressure below minimum at selected input | warning |
A11 | Overpressure at selected gas input | Gas pressure above maximum at selected input | warning |
A12 | Caloric and differential pressure sensor mismatch | Gas type sensors disagree on pressure differential | error |
A13 | APC 2 module error code 13 | Unknown fault condition in APC 2 | error |
A21-23 | APC 2 module error codes 21-23 | Unknown fault conditions in APC 2 or VEM 2 | error |
A2-4 | APC 2 module error codes 2-4 | Unknown fault conditions in APC 2 or VEM 2 | error |
A30 | APC 2 module error code 30 | Unknown fault condition in APC 2 | error |
A38 | Type detection mismatch on socket 1 | Socket 1 instrument type does not match stored configuration | error |
A39 | Type detection mismatch on socket 2 | Socket 2 instrument type does not match stored configuration | error |
A40 | Flow specification not attained | Gas or fluid flow below specification | error |
A41 | APC 2 module error code 41 | Unknown fault condition in APC 2 | error |
A5 | Button error socket 1 | Keyboard malfunction on socket 1 | warning |
A6 | Button error socket 2 | Keyboard malfunction on socket 2 | warning |
A7-9 | APC 2 module error codes 7-9 | Unknown fault conditions in APC 2 | error |
A7D+7E | APC 2 or VEM 2 error codes 7D and 7E | Unknown fault conditions | error |
A7F | Operating system error | Operating system exception detected | info |
A80 | Internal state incorrect | System state machine inconsistency detected | info |
A81-83 | APC 2 or VEM 2 error codes 81-83 | Unknown fault conditions | error |
A85 | Invalid resistance instrument number | Instrument resistance measurement out of expected range | error |
A86 | Gas underdose (hose blocked) | Insufficient gas flow delivered to patient circuit | critical |
A90 | Low pressure at cylinder 1 | Gas tank 1 pressure below minimum operating level | warning |
A91 | Low pressure at cylinder 2 | Gas tank 2 pressure below minimum operating level | warning |
AA | APC 2 module error A | Unknown fault condition in APC 2 | error |
AA0 | APC 2 module error A0 | Unknown fault condition in APC 2 | error |
AB | Error in test mode | Error detected during test mode operation | info |
AC+D | Test mode or diagnostic error codes C and D | Error during test mode or diagnostics | error |
B10 | Activation signal not terminated | Activation signal (footswitch) held longer than 5 seconds after auto-stop | warning |
B11 | Control panel error code 11 | Unknown fault condition in control panel | error |
B1+2 | Control panel error codes 1 and 2 | Unknown fault conditions in control panel | error |
B12 | Activation signal present during system initialization | Activation signal detected (footswitch pressed) during boot sequence | warning |
B13 | Error or event list flash file corruption | Flash file corrupted when managing error or event list | error |
B14+15 | Control panel error codes 14 and 15 | Unknown fault conditions in control panel | error |
B16 | Insufficient EEPROM memory for user programs | User program storage full | warning |
B17 | Double activation detected simultaneously | Two activation signals (both footswitch pedals) detected within 100 ms | warning |
B18 | No deactivation signal from HF or APC module | Module fails to react to deactivation request for longer than 110 ms | error |
B19 | PowerFail signal without actual power failure | PowerFail signal received but power does not fail within 2 s | error |
B1A | CRC check not yet completed | User attempts to operate unit after software download before successful CRC check | warning |
B1B | Maximum ON time exceeded | Activation lasts longer than ON time selected in Setup | error |
B1C | Incorrect instrument parameters | Instrument detected via recognition with parameters outside permissible tolerances | error |
B1D | Keyboard error | Button pressed is recognized during initialization | error |
B1E | NESSY symmetry monitoring error | Error signaled by Nessy symmetry monitoring during activation | error |
B1F | Unknown NESSY error | Control panel detects undefined condition | error |
B21 | Activation terminated with tissue still touched | Activation was terminated automatically (e.g. by AutoStop) and tissue contact remains | warning |
B22 | Incorrect status message from IES 2 | Invalid status message received from IES module | error |
B23 | Modified socket configuration detected | Socket configuration changed (e.g. replacement or retrofitting) | warning |
B24 | Unknown socket configuration error | Control panel detects undefined condition | error |
B25 | Maximum PRECISE APC application time exceeded | PRECISE APC activation on longer than permitted | info |
B26 | ERBECRYO 2 error received | Cryogenic module signals error condition | error |
B27 | ERBECRYO 2 status message timeout | Timeout when receiving ERBECRYO 2 status message | error |
B28 | Footswitch status message timeout | Timeout when receiving footswitch status message during activation | error |
B29 | MF connector disconnected or incompatible instrument | No MF connector or different instrument to HYBRIDtherm probe connected to MF socket | error |
B2A | ERBECRYO 2 input pressure too high | Input pressure sensor on ERBECRYO 2 signals value too high for HYBRIDtherm | warning |
B2B | ERBECRYO 2 waste gas tube obstruction | Pressure or flow sensor indicates blockage | warning |
B2C | ERBECRYO 2 unknown instrument detected | ERBECRYO 2 signals unrecognized instrument type | error |
B2D | Unknown control panel error | Control panel detects undefined condition | error |
B5+6 | Control panel error codes 5 and 6 | Unknown fault conditions in control panel | error |
B81 | Dual-pedal footswitch recognized | Dual-pedal footswitch is connected | info |
B82 | Dual-pedal footswitch disconnected | Dual-pedal footswitch is disconnected from the system | info |
B83 | APC socket 1 recognized | APC socket 1 is detected | info |
B84 | Unknown control panel event | Control panel detects undefined condition | error |
B85 | Single-pedal footswitch recognized | Single-pedal footswitch is connected | info |
B86 | Single-pedal footswitch disconnected | Single-pedal footswitch is disconnected from the system | info |
B87 | APC socket 2 recognized | APC socket 2 is detected | info |
B88 | Unknown control panel event | Control panel detects undefined condition | error |
B89 | APC module recognized | APC module is detected | info |
B8A | No status message from APC module | APC module is disconnected from the system | error |
B8B | Unknown control panel event | Control panel detects undefined condition | error |
B8E | Unknown control panel event | Control panel detects undefined condition | error |
B9 | Control panel error code 9 | Unknown fault condition in control panel | error |
B90 | Unknown control panel event | Control panel detects undefined condition | error |
B91 | No status message from IES module | IES module is disconnected from the system | error |
B92 | IES module recognized | IES module is detected | info |
B93 | Multifunction footswitch recognized | Multifunction footswitch is connected | info |
B94 | Multifunction footswitch disconnected | Multifunction footswitch is disconnected from the system | info |
B95 | New instrument recognized by system | Instrument with instrument recognition is connected | info |
B96 | Unknown control panel event | Control panel detects undefined condition | error |
B97 | Unknown control panel event | Control panel detects undefined condition | error |
B98 | Incorrect activation type assigned | User assigns activation signal not currently available (e.g. unconnected footswitch) | error |
B99 | Built-in realtime clock battery depleted | VIO switched off for extended period; clock backup power insufficient | warning |
B9A | Master remote control recognized | Master remote control is recognized by the VIO | info |
B9B | Master remote control disconnected | Master remote control is disconnected from the system | info |
B9C | Remote control recognized | Remote control is recognized by the VIO | info |
B9D | Remote control disconnected | Remote control is disconnected from the system | info |
B9F | Instrument has been disconnected from the system | Instrument with instrument recognition disconnected | info |
BA | Control panel error A | Unknown fault condition in control panel | error |
BA0 | No other mode can be set for this instrument | User attempts to assign different mode to fixed-mode instrument | info |
BA1 | Unknown error | Unknown condition | error |
BA3 | Footswitch not assigned | Footswitch without assignment pressed | info |
BA4 | Two footswitches are connected | Two footswitches of same type connected simultaneously | warning |
BA6 | EEPROM is updated | Unit programmed using connected PC | info |
BA8 | Unknown error | Unknown condition | error |
BA9 | Activation not possible until settings are confirmed | User attempts activation after power-on without confirming settings | info |
BAA | No valid mode is assigned | Output channel activated without assigned mode | info |
BAB | Activation is only possible with a valid instrument | MF socket activated without recognized instrument | warning |
BAC | AutoStart is not possible if contact is made during assignment | AutoStart function assigned while contact already recognized | warning |
BAD-BAF | Unknown error | Unknown condition | error |
BB | Neutral electrode contact resistance out of range | NE contact resistance measurement outside valid range | warning |
BB0 | NESSY symmetry warning | NESSY symmetry monitoring signals value between 20-50% | warning |
BB1 | NESSY current density warning | NESSY current density above limit curve | warning |
BB2 | IES footswitch detected | IES footswitch connected | info |
BB3 | Unknown error | Unknown condition | error |
BB4 | Rinsing of an APC instrument | Rinsing of APC instrument initiated | info |
BB5-BB6 | Unknown error | Unknown condition | error |
BB7 | Permissible limit value of AUTO START exceeded | Attempt to assign non-permissible power limitation to AUTO START | info |
BB8 | New interface module recognized | Interface module (e.g. VIO PORTAL) logs on | info |
BB9 | Interface module has been disconnected from the system | Interface module (e.g. VIO PORTAL) disconnected | info |
BBA | Warning from interface module | Interface module signals an error | warning |
BBB | Reminder that a safety check is due | Safety check period elapsed | info |
BBC | VEM 2 module recognized | Socket extension module VEM 2 logs into system | info |
BBD | VEM 2 module disconnected | Socket extension module VEM 2 disconnected | info |
BBE | VEM 2 module socket 1 no longer ready for operation | 1st socket of VEM 2 no longer responds | info |
BBF | VEM 2 module socket 2 no longer ready for operation | 2nd socket of VEM 2 no longer responds | info |
BC | Function not available on module | Module unable to implement required function | error |
BC0 | Neither footswitch nor AUTO START assigned | Instrument without fingerswitch recognized with no footswitch/AUTO START assigned | info |
BC1 | EIP 2 was recognized by the system | EIP 2 connected to system | info |
BC2 | EIP 2 was disconnected from the system | EIP 2 disconnected | info |
BC3 | IES 2 footswitch was disconnected from the system | IES 2 footswitch disconnected | info |
BC4 | Purge function not assigned | Purge button on APC pressed with no APC socket selected | info |
BC5 | Configuration download from STORZ OR-1 system has failed | Configuration sync with STORZ OR-1 fails | error |
BC6 | NESSY temperature monitoring alert | NESSY current exceeds 300 mA during electrode monitoring | info |
BC7 | VIO received valid status and version from ERBECRYO 2 | ERBECRYO 2 communicates status and version | info |
BC8 | Time out occurred when receiving ERBECRYO 2 status message | ERBECRYO 2 status message reception timeout | info |
BCA | HYBRIDtherm probe connected with no ERBECRYO 2 status received | HYBRIDtherm probe connected but ERBECRYO 2 not responding | info |
BCB | User operated VIO footswitch during activation | Footswitch pressed while activation in progress | info |
BCC | Impedance limit has been exceeded | Load impedance exceeds maximum allowable value | info |
BCD | The set ablation duration has been reached | Ablation time limit expires | info |
BCE | User operated VIO STOP button during activation | STOP button pressed during activation | info |
BCF | HYBRIDtherm instrument not connected to ERBECRYO 2 | HYBRIDtherm connected to VIO but not to ERBECRYO 2 | info |
BD | I²C bus error - EEPROM write/read failure | Control panel CPU cannot read/write control panel EEPROM | error |
BD0 | ERBECRYO 2 footswitch has been activated | ERBECRYO 2 footswitch pressed | info |
BD1 | HYBRIDtherm connected to VIO without required software version | HYBRIDtherm device connected to VIO version < 2.8.x | error |
BD2 | More than one HYBRIDtherm instrument connected | Multiple HYBRIDtherm instruments connected simultaneously | info |
BD3-BD4 | ERBECRYO 2 gas bottle pressure too low | Input pressure below minimum during or after operation | error |
BD5 | ERBECRYO 2 input pressure too high | Input pressure sensor signals very high value for HYBRIDtherm | warning |
BD6 | ERBECRYO 2 waste gas nozzle pressure high | Waste gas nozzle pressure high outside of activation | warning |
BD8 | ERBECRYO 2 input pressure too high outside activation | Input pressure high when not activating | warning |
BD9 | ERBECRYO 2 input pressure too low during activation | Input pressure low while activating HYBRIDtherm | error |
BDA | ERBECRYO 2 input pressure too low during activation | Input pressure below minimum during HYBRIDtherm activation | error |
BDB | ERBECRYO 2 input pressure too low outside activation | Input pressure low when not activating | error |
BDC | ERBECRYO 2 input pressure too low during activation | Input pressure below safe level during HYBRIDtherm activation | error |
BDD | HYBRIDtherm minimum power not reached | Ablation power below minimum threshold | info |
BE | CAN bus error - connection interrupted | Control panel detects CAN bus communication failure | error |
BF | Control panel error F | Unknown fault condition in control panel | error |
BFC | Power Down | Unit switched off | info |
BFD | System Reset | System reset performed (e.g. on power-up) | info |
BFE | PowerFail | PowerFail signaled (e.g. on power-off) | info |
BFF | CAN MESSAGE | CAN message transmitted for logging | info |
C1 | Unknown error | CPU/Sensors error | error |
C21 | Unknown error | CPU/Sensors error | error |
C26 | Unknown error | CPU/Sensors error | error |
C30-C38 | Unknown error | CPU/Sensors error | error |
C41-C48 | Unknown error | CPU/Sensors error | error |
C4-C6 | Unknown error | CPU/Sensors error | error |
C51-C58 | Unknown error | CPU/Sensors error | error |
C5A-C5B | Unknown error | CPU/Sensors error | error |
C61-C66 | Unknown error | CPU/Sensors error | error |
C70-C72 | Unknown error | CPU/Sensors error | error |
C75 | Unknown error | CPU/Sensors error | error |
C7D-C7F | Unknown error | CPU/Sensors error | error |
C81-C84 | Unknown error | CPU/Sensors error | error |
C85 | BiCision short circuit detection | Electrical connection between instrument branches during activation | info |
C91-C92 | Unknown error | CPU/Sensors error | error |
CA0-CA1 | Unknown error | CPU/Sensors error | error |
CC0 | Unknown error | CPU/Sensors error | error |
CD | Unknown error | CPU/Sensors error | error |
CD0 | Unknown error | CPU/Sensors error | error |
CF0 | Trial activation with Pmax==0 | Test activation with zero maximum power | info |
D80 | ECB fault | IES 2 smoke evacuation controller malfunction | error |
D81 | Excess temperature IES2 | IES 2 temperature exceeds 60°C | warning |
D82 | Undertemperature IES2 | IES 2 temperature below 0°C | warning |
D83 | Unknown error | IES 2 unknown condition | error |
D86 | Unknown error | IES 2 unknown condition | error |
D87 | Suction tube obstruction | Blockage in IES 2 suction line | warning |
D88 | Filter missing or obstructed | IES 2 filter absent or clogged | warning |
F1 | CRC error in footswitch | CRC check fails during footswitch start-up | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.