Electrosurgery units

ERBE VIO 300D: troubleshooting and error codes

The ServiceForge knowledge base holds 19 documented symptoms for the ERBE VIO 300D, with 21 ranked causes, 42 diagnostic checks with test points and expected values, and 210 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Neutral electrode critical resistance monitoring dual surface failure patient safety

Neutral electrode lamp does not light red or green; unit activates without warning when should inhibit

Ranked causes

  1. 1. NE monitoring circuit malfunction for dual surface electrode (unknown)
  2. 2. Asymmetrical dual surface electrode monitoring malfunction (unknown)
First check

Dual surface neutral electrode lamp color

Test point
NE socket connected to VIO Testbox 20100-101
Expected
Red lamp indication

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neutral electrode critical resistance monitoring single surface failure patient safety

Neutral electrode lamp does not light red or green; unit activates without warning when should inhibit

Ranked causes

  1. 1. NE monitoring circuit malfunction for single surface electrode (unknown)
  2. 2. NE circuit passes but monitoring disabled or second stage malfunction (unknown)
First check

Neutral electrode lamp color on unit display

Test point
NE socket connected to VIO Testbox 20100-101
Expected
Red lamp indication

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acknowledgement tones not emitted

Unit fails to produce acoustic signals during system start, button presses, or pedal activation

Ranked causes

  1. 1. Acoustic signaling system malfunction (unknown)
First check

Check acknowledgement tone during system start

Test point
Speaker/audio output
Expected
Acoustic signal emitted during self-test

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Automatic start mode delay not functioning patient safety

Unit does not start after set delay in AUTO START 1 mode or fails to terminate on command

Ranked causes

  1. 1. Auto start timing circuit or command processing failure (unknown)
First check

Verify auto start delay timing

Test point
VIO Testbox T1 button activation
Expected
Unit starts after set AUTO START 1 delay (record delay value from Setup) watts

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bipolar resection adapter decoupling capacitor insulation failure high voltage

Decoupling capacitor insulation resistance falls below specification; potential for electrical leakage

Ranked causes

  1. 1. Decoupling capacitor insulation degradation (rare)
First check

Decoupling capacitor DC insulation resistance

Test point
Bipolar resection adapter MF socket to safety tester
Expected
Resistance > 200 MOhm, > 200
Tools
Safety tester with insulation testing > 500 V DC

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bipolar resection adapter HF power output low high voltage

HF power delivered into 50 ohm load falls below specification; output stage malfunction suspected

Ranked causes

  1. 1. RF power amplifier or output coupling malfunction (occasional)
First check

HF power meter impedance configuration

Test point
HF power meter selector
Expected
RL = 50 ohms

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bipolar resection adapter transfer ratio out of specification high voltage

HF output voltage transfer ratio does not meet specification; indicates RF amplifier or transformer issue

Ranked causes

  1. 1. RF transformer or HF amplifier output stage malfunction (occasional)
First check

Setup level 2 access and test program selection

Test point
VIO 300 D menu system
Expected
Navigate to Test programs > Hardware TP > TP generator

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control buttons or pushbuttons not responsive

Selection buttons, focus buttons, or enter button fail to respond or function intermittently

Ranked causes

  1. 1. Button contact or switch failure (unknown)
First check

Test control button functionality

Test point
Control panel buttons
Expected
Smooth operation, acknowledgement tone emitted with each press

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display backlighting not functional

Display lacks illumination or backlighting is dim or faulty

Ranked causes

  1. 1. Backlighting circuit or LED failure (unknown)
First check

Verify display backlighting operation

Test point
Display screen
Expected
Backlighting illuminates properly, image visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dual-pedal footswitch CUT or COAG activation not registered

Dual-pedal footswitch CUT or COAG pedal does not register in test program or MF icon does not illuminate

Ranked causes

  1. 1. Dual-pedal footswitch contact or cable malfunction (unknown)
First check

Test CUT pedal on dual-pedal footswitch

Test point
CUT pedal on dual-pedal footswitch art. no. 20189-303
Expected
MF icon field lights yellow in test program

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fingerswitch CUT or COAG buttons not responsive

Electrode handle CUT or COAG buttons fail to activate or do not show visual confirmation in test program

Ranked causes

  1. 1. Electrode handle button contact failure (unknown)
First check

Test CUT button on electrode handle (Monopolar)

Test point
CUT button on handle art. no. 20190-045
Expected
Handpiece icon field lights in color in test program

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Footswitch CUT or COAG pedal not responsive

Single or dual-pedal footswitch pedals fail to activate or produce no acknowledgement tone

Ranked causes

  1. 1. Footswitch pedal contact or cable failure (unknown)
First check

Test CUT pedal activation

Test point
CUT pedal on dual-pedal footswitch
Expected
Acknowledgement tone emitted each time pedal pressed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Instrument detection via EEPROM not working

Multifunction socket does not detect connected instrument or COAG mode does not switch to BiClamp automatically

Ranked causes

  1. 1. EEPROM detection circuit or sensor failure (unknown)
First check

Test instrument detection on MF socket

Test point
MF socket with measuring cable art. no. 20100-152
Expected
System displays instrument detection advisory message, COAG mode switches to BiClamp

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LEDs do not light up at startup

Status LEDs fail to illuminate or light up at inconsistent intensity during system startup

Ranked causes

  1. 1. LED circuit or individual LED failure (unknown)
First check

Observe LED illumination at startup

Test point
All status LEDs
Expected
All LEDs light briefly at same intensity during system start

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power switch does not operate smoothly

Power switch sticks, scrapes, or is difficult to operate

Ranked causes

  1. 1. Power switch mechanical defect (unknown)
First check

Operate power switch manually

Test point
Power switch
Expected
Smooth operation, snaps into ON position without sticking or scraping

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ReMode button not switching modes

ReMode button on footswitch or electrode handle fails to switch between modes or does not show visual confirmation

Ranked causes

  1. 1. ReMode button contact failure (unknown)
First check

Test ReMode button on dual-pedal footswitch

Test point
ReMode button on dual-pedal footswitch art. no. 20189-303
Expected
MF icon field lights in color in test program confirming mode switchover

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Single-pedal footswitch activation not registered

Single-pedal footswitch COAG pedal does not register activation in test program

Ranked causes

  1. 1. Single-pedal footswitch contact or cable malfunction (unknown)
First check

Test single-pedal footswitch COAG activation

Test point
COAG pedal on single-pedal footswitch art. no. 20188-300
Expected
Single-pedal icon lights in color in test program

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not start or displays error messages

Unit fails to perform system start or shows error messages during startup

Ranked causes

  1. 1. System initialization failure (unknown)
First check

Verify system start without errors

Test point
Display
Expected
System performs start routine without error messages

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Warning message C-84 received

BiClamp short-circuit protection activated and interrupted cutting operation.

Ranked causes

  1. 1. Short-circuit detected in BiClamp circuit (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
2,3,5,6-1-7Socket IIF errorInstrument Interface error in socket 2/3/5/6error
2,3,5,6-70-72Socket IIF errorInstrument Interface error in socket 2/3/5/6error
2,3,5,6-8Overcurrent activation recognitionMeasurement current too high during activation recognitionerror
2,3,5,6-81Short circuit instrument recognition RInstrument recognition resistor value indicates short circuiterror
2,3,5,6-82Short circuit activation recognition R_NEUActivation recognition resistor value indicates short circuiterror
2,3,5,6-9Socket IIF errorInstrument Interface error in socket 2/3/5/6error
2,3,5,6-AOvercurrent instrument recognitionMeasurement current too high during instrument recognitionerror
2,3,5,6-C-ESocket IIF errorInstrument Interface error in socket 2/3/5/6error
2,3,5,6-FU Coag error: +12V voltage too low+12V voltage during Coag activation drops below thresholderror
4(NE)1Nessy2 generic errorUnspecified error condition in Nessy2 moduleerror
4(NE)10Nessy2 generic errorUnspecified error condition in Nessy2 moduleerror
4(NE)13Nessy2 generic errorUnspecified error condition in Nessy2 moduleerror
4(NE)CNessy2 generic errorUnspecified error condition in Nessy2 moduleerror
83No-load activation recognition R_NEU: resistor value determined for activation recognition corresponds to no-loadInstrument interface socket detects no-load condition during activation recognitionerror
84Instrument recognition R-window violated: resistor value cannot be reliably assigned to setpointInstrument interface socket cannot match resistor value to valid setpointerror
85Short circuit activation recognition R_ALT: resistor value corresponds to short circuitInstrument interface socket detects short circuit condition during activation recognitionerror
86Activation error: activation not consistent with instrument typeActivation attempt inconsistent with connected instrument typeerror
87Generic module error (IIF socket)Unspecified error condition in instrument interface moduleerror
88Generic module error (IIF socket)Unspecified error condition in instrument interface moduleerror
89Generic module error (IIF socket)Unspecified error condition in instrument interface moduleerror
8AGeneric module error (IIF socket)Unspecified error condition in instrument interface moduleerror
8BGeneric module error (IIF socket)Unspecified error condition in instrument interface moduleerror
8CGeneric module error (IIF socket)Unspecified error condition in instrument interface moduleerror
8DGeneric module error (IIF socket)Unspecified error condition in instrument interface moduleerror
8EGeneric module error (IIF socket)Unspecified error condition in instrument interface moduleerror
8FGeneric module error (IIF socket)Unspecified error condition in instrument interface moduleerror
90Generic module error (IIF socket or Nessy2 or EIP 2)Unspecified error condition in moduleerror
980Erbe Irrigation Pump EIP 2 generic errorUnspecified error condition in EIP 2 pump moduleerror
982Pump lid open at ON: EIP 2 pump lid not closed at start or opened during activationPump lid detected open during activationwarning
984Time Out Error: EIP 2 activation maintained longer than maximum 20 secondsEIP 2 activation continuous for more than 20 secondswarning
988Erbe Irrigation Pump EIP 2 generic errorUnspecified error condition in EIP 2 pump moduleerror
990Erbe Irrigation Pump EIP 2 generic errorUnspecified error condition in EIP 2 pump moduleerror
9A0Erbe Irrigation Pump EIP 2 generic errorUnspecified error condition in EIP 2 pump moduleerror
9C0Erbe Irrigation Pump EIP 2 generic errorUnspecified error condition in EIP 2 pump moduleerror
A1APC 2 module error code 1Unknown fault condition in APC 2error
A10Underpressure at selected gas inputGas pressure below minimum at selected inputwarning
A11Overpressure at selected gas inputGas pressure above maximum at selected inputwarning
A12Caloric and differential pressure sensor mismatchGas type sensors disagree on pressure differentialerror
A13APC 2 module error code 13Unknown fault condition in APC 2error
A21-23APC 2 module error codes 21-23Unknown fault conditions in APC 2 or VEM 2error
A2-4APC 2 module error codes 2-4Unknown fault conditions in APC 2 or VEM 2error
A30APC 2 module error code 30Unknown fault condition in APC 2error
A38Type detection mismatch on socket 1Socket 1 instrument type does not match stored configurationerror
A39Type detection mismatch on socket 2Socket 2 instrument type does not match stored configurationerror
A40Flow specification not attainedGas or fluid flow below specificationerror
A41APC 2 module error code 41Unknown fault condition in APC 2error
A5Button error socket 1Keyboard malfunction on socket 1warning
A6Button error socket 2Keyboard malfunction on socket 2warning
A7-9APC 2 module error codes 7-9Unknown fault conditions in APC 2error
A7D+7EAPC 2 or VEM 2 error codes 7D and 7EUnknown fault conditionserror
A7FOperating system errorOperating system exception detectedinfo
A80Internal state incorrectSystem state machine inconsistency detectedinfo
A81-83APC 2 or VEM 2 error codes 81-83Unknown fault conditionserror
A85Invalid resistance instrument numberInstrument resistance measurement out of expected rangeerror
A86Gas underdose (hose blocked)Insufficient gas flow delivered to patient circuitcritical
A90Low pressure at cylinder 1Gas tank 1 pressure below minimum operating levelwarning
A91Low pressure at cylinder 2Gas tank 2 pressure below minimum operating levelwarning
AAAPC 2 module error AUnknown fault condition in APC 2error
AA0APC 2 module error A0Unknown fault condition in APC 2error
ABError in test modeError detected during test mode operationinfo
AC+DTest mode or diagnostic error codes C and DError during test mode or diagnosticserror
B10Activation signal not terminatedActivation signal (footswitch) held longer than 5 seconds after auto-stopwarning
B11Control panel error code 11Unknown fault condition in control panelerror
B1+2Control panel error codes 1 and 2Unknown fault conditions in control panelerror
B12Activation signal present during system initializationActivation signal detected (footswitch pressed) during boot sequencewarning
B13Error or event list flash file corruptionFlash file corrupted when managing error or event listerror
B14+15Control panel error codes 14 and 15Unknown fault conditions in control panelerror
B16Insufficient EEPROM memory for user programsUser program storage fullwarning
B17Double activation detected simultaneouslyTwo activation signals (both footswitch pedals) detected within 100 mswarning
B18No deactivation signal from HF or APC moduleModule fails to react to deactivation request for longer than 110 mserror
B19PowerFail signal without actual power failurePowerFail signal received but power does not fail within 2 serror
B1ACRC check not yet completedUser attempts to operate unit after software download before successful CRC checkwarning
B1BMaximum ON time exceededActivation lasts longer than ON time selected in Setuperror
B1CIncorrect instrument parametersInstrument detected via recognition with parameters outside permissible toleranceserror
B1DKeyboard errorButton pressed is recognized during initializationerror
B1ENESSY symmetry monitoring errorError signaled by Nessy symmetry monitoring during activationerror
B1FUnknown NESSY errorControl panel detects undefined conditionerror
B21Activation terminated with tissue still touchedActivation was terminated automatically (e.g. by AutoStop) and tissue contact remainswarning
B22Incorrect status message from IES 2Invalid status message received from IES moduleerror
B23Modified socket configuration detectedSocket configuration changed (e.g. replacement or retrofitting)warning
B24Unknown socket configuration errorControl panel detects undefined conditionerror
B25Maximum PRECISE APC application time exceededPRECISE APC activation on longer than permittedinfo
B26ERBECRYO 2 error receivedCryogenic module signals error conditionerror
B27ERBECRYO 2 status message timeoutTimeout when receiving ERBECRYO 2 status messageerror
B28Footswitch status message timeoutTimeout when receiving footswitch status message during activationerror
B29MF connector disconnected or incompatible instrumentNo MF connector or different instrument to HYBRIDtherm probe connected to MF socketerror
B2AERBECRYO 2 input pressure too highInput pressure sensor on ERBECRYO 2 signals value too high for HYBRIDthermwarning
B2BERBECRYO 2 waste gas tube obstructionPressure or flow sensor indicates blockagewarning
B2CERBECRYO 2 unknown instrument detectedERBECRYO 2 signals unrecognized instrument typeerror
B2DUnknown control panel errorControl panel detects undefined conditionerror
B5+6Control panel error codes 5 and 6Unknown fault conditions in control panelerror
B81Dual-pedal footswitch recognizedDual-pedal footswitch is connectedinfo
B82Dual-pedal footswitch disconnectedDual-pedal footswitch is disconnected from the systeminfo
B83APC socket 1 recognizedAPC socket 1 is detectedinfo
B84Unknown control panel eventControl panel detects undefined conditionerror
B85Single-pedal footswitch recognizedSingle-pedal footswitch is connectedinfo
B86Single-pedal footswitch disconnectedSingle-pedal footswitch is disconnected from the systeminfo
B87APC socket 2 recognizedAPC socket 2 is detectedinfo
B88Unknown control panel eventControl panel detects undefined conditionerror
B89APC module recognizedAPC module is detectedinfo
B8ANo status message from APC moduleAPC module is disconnected from the systemerror
B8BUnknown control panel eventControl panel detects undefined conditionerror
B8EUnknown control panel eventControl panel detects undefined conditionerror
B9Control panel error code 9Unknown fault condition in control panelerror
B90Unknown control panel eventControl panel detects undefined conditionerror
B91No status message from IES moduleIES module is disconnected from the systemerror
B92IES module recognizedIES module is detectedinfo
B93Multifunction footswitch recognizedMultifunction footswitch is connectedinfo
B94Multifunction footswitch disconnectedMultifunction footswitch is disconnected from the systeminfo
B95New instrument recognized by systemInstrument with instrument recognition is connectedinfo
B96Unknown control panel eventControl panel detects undefined conditionerror
B97Unknown control panel eventControl panel detects undefined conditionerror
B98Incorrect activation type assignedUser assigns activation signal not currently available (e.g. unconnected footswitch)error
B99Built-in realtime clock battery depletedVIO switched off for extended period; clock backup power insufficientwarning
B9AMaster remote control recognizedMaster remote control is recognized by the VIOinfo
B9BMaster remote control disconnectedMaster remote control is disconnected from the systeminfo
B9CRemote control recognizedRemote control is recognized by the VIOinfo
B9DRemote control disconnectedRemote control is disconnected from the systeminfo
B9FInstrument has been disconnected from the systemInstrument with instrument recognition disconnectedinfo
BAControl panel error AUnknown fault condition in control panelerror
BA0No other mode can be set for this instrumentUser attempts to assign different mode to fixed-mode instrumentinfo
BA1Unknown errorUnknown conditionerror
BA3Footswitch not assignedFootswitch without assignment pressedinfo
BA4Two footswitches are connectedTwo footswitches of same type connected simultaneouslywarning
BA6EEPROM is updatedUnit programmed using connected PCinfo
BA8Unknown errorUnknown conditionerror
BA9Activation not possible until settings are confirmedUser attempts activation after power-on without confirming settingsinfo
BAANo valid mode is assignedOutput channel activated without assigned modeinfo
BABActivation is only possible with a valid instrumentMF socket activated without recognized instrumentwarning
BACAutoStart is not possible if contact is made during assignmentAutoStart function assigned while contact already recognizedwarning
BAD-BAFUnknown errorUnknown conditionerror
BBNeutral electrode contact resistance out of rangeNE contact resistance measurement outside valid rangewarning
BB0NESSY symmetry warningNESSY symmetry monitoring signals value between 20-50%warning
BB1NESSY current density warningNESSY current density above limit curvewarning
BB2IES footswitch detectedIES footswitch connectedinfo
BB3Unknown errorUnknown conditionerror
BB4Rinsing of an APC instrumentRinsing of APC instrument initiatedinfo
BB5-BB6Unknown errorUnknown conditionerror
BB7Permissible limit value of AUTO START exceededAttempt to assign non-permissible power limitation to AUTO STARTinfo
BB8New interface module recognizedInterface module (e.g. VIO PORTAL) logs oninfo
BB9Interface module has been disconnected from the systemInterface module (e.g. VIO PORTAL) disconnectedinfo
BBAWarning from interface moduleInterface module signals an errorwarning
BBBReminder that a safety check is dueSafety check period elapsedinfo
BBCVEM 2 module recognizedSocket extension module VEM 2 logs into systeminfo
BBDVEM 2 module disconnectedSocket extension module VEM 2 disconnectedinfo
BBEVEM 2 module socket 1 no longer ready for operation1st socket of VEM 2 no longer respondsinfo
BBFVEM 2 module socket 2 no longer ready for operation2nd socket of VEM 2 no longer respondsinfo
BCFunction not available on moduleModule unable to implement required functionerror
BC0Neither footswitch nor AUTO START assignedInstrument without fingerswitch recognized with no footswitch/AUTO START assignedinfo
BC1EIP 2 was recognized by the systemEIP 2 connected to systeminfo
BC2EIP 2 was disconnected from the systemEIP 2 disconnectedinfo
BC3IES 2 footswitch was disconnected from the systemIES 2 footswitch disconnectedinfo
BC4Purge function not assignedPurge button on APC pressed with no APC socket selectedinfo
BC5Configuration download from STORZ OR-1 system has failedConfiguration sync with STORZ OR-1 failserror
BC6NESSY temperature monitoring alertNESSY current exceeds 300 mA during electrode monitoringinfo
BC7VIO received valid status and version from ERBECRYO 2ERBECRYO 2 communicates status and versioninfo
BC8Time out occurred when receiving ERBECRYO 2 status messageERBECRYO 2 status message reception timeoutinfo
BCAHYBRIDtherm probe connected with no ERBECRYO 2 status receivedHYBRIDtherm probe connected but ERBECRYO 2 not respondinginfo
BCBUser operated VIO footswitch during activationFootswitch pressed while activation in progressinfo
BCCImpedance limit has been exceededLoad impedance exceeds maximum allowable valueinfo
BCDThe set ablation duration has been reachedAblation time limit expiresinfo
BCEUser operated VIO STOP button during activationSTOP button pressed during activationinfo
BCFHYBRIDtherm instrument not connected to ERBECRYO 2HYBRIDtherm connected to VIO but not to ERBECRYO 2info
BDI²C bus error - EEPROM write/read failureControl panel CPU cannot read/write control panel EEPROMerror
BD0ERBECRYO 2 footswitch has been activatedERBECRYO 2 footswitch pressedinfo
BD1HYBRIDtherm connected to VIO without required software versionHYBRIDtherm device connected to VIO version < 2.8.xerror
BD2More than one HYBRIDtherm instrument connectedMultiple HYBRIDtherm instruments connected simultaneouslyinfo
BD3-BD4ERBECRYO 2 gas bottle pressure too lowInput pressure below minimum during or after operationerror
BD5ERBECRYO 2 input pressure too highInput pressure sensor signals very high value for HYBRIDthermwarning
BD6ERBECRYO 2 waste gas nozzle pressure highWaste gas nozzle pressure high outside of activationwarning
BD8ERBECRYO 2 input pressure too high outside activationInput pressure high when not activatingwarning
BD9ERBECRYO 2 input pressure too low during activationInput pressure low while activating HYBRIDthermerror
BDAERBECRYO 2 input pressure too low during activationInput pressure below minimum during HYBRIDtherm activationerror
BDBERBECRYO 2 input pressure too low outside activationInput pressure low when not activatingerror
BDCERBECRYO 2 input pressure too low during activationInput pressure below safe level during HYBRIDtherm activationerror
BDDHYBRIDtherm minimum power not reachedAblation power below minimum thresholdinfo
BECAN bus error - connection interruptedControl panel detects CAN bus communication failureerror
BFControl panel error FUnknown fault condition in control panelerror
BFCPower DownUnit switched offinfo
BFDSystem ResetSystem reset performed (e.g. on power-up)info
BFEPowerFailPowerFail signaled (e.g. on power-off)info
BFFCAN MESSAGECAN message transmitted for logginginfo
C1Unknown errorCPU/Sensors errorerror
C21Unknown errorCPU/Sensors errorerror
C26Unknown errorCPU/Sensors errorerror
C30-C38Unknown errorCPU/Sensors errorerror
C41-C48Unknown errorCPU/Sensors errorerror
C4-C6Unknown errorCPU/Sensors errorerror
C51-C58Unknown errorCPU/Sensors errorerror
C5A-C5BUnknown errorCPU/Sensors errorerror
C61-C66Unknown errorCPU/Sensors errorerror
C70-C72Unknown errorCPU/Sensors errorerror
C75Unknown errorCPU/Sensors errorerror
C7D-C7FUnknown errorCPU/Sensors errorerror
C81-C84Unknown errorCPU/Sensors errorerror
C85BiCision short circuit detectionElectrical connection between instrument branches during activationinfo
C91-C92Unknown errorCPU/Sensors errorerror
CA0-CA1Unknown errorCPU/Sensors errorerror
CC0Unknown errorCPU/Sensors errorerror
CDUnknown errorCPU/Sensors errorerror
CD0Unknown errorCPU/Sensors errorerror
CF0Trial activation with Pmax==0Test activation with zero maximum powerinfo
D80ECB faultIES 2 smoke evacuation controller malfunctionerror
D81Excess temperature IES2IES 2 temperature exceeds 60°Cwarning
D82Undertemperature IES2IES 2 temperature below 0°Cwarning
D83Unknown errorIES 2 unknown conditionerror
D86Unknown errorIES 2 unknown conditionerror
D87Suction tube obstructionBlockage in IES 2 suction linewarning
D88Filter missing or obstructedIES 2 filter absent or cloggedwarning
F1CRC error in footswitchCRC check fails during footswitch start-uperror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI