ERBE VIO 3: troubleshooting and error codes
The ServiceForge knowledge base holds 14 documented symptoms for the ERBE VIO 3, with 51 ranked causes, 0 diagnostic checks with test points and expected values, and 216 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
N/N-A error displayed patient safety
Error code N or N-A shown on display, detecting module NM (NESSY return electrode monitoring)
Ranked causes
- 1. N1: CRC check alarm (unknown)
- 2. N2: ErrorHandler Buffer Overflow (unknown)
- 3. N5-8: Unspecified error (unknown)
- 4. N10-14: Unspecified error (unknown)
- 5. N17-36: Unspecified error (unknown)
- 6. N38: Faulty circuit, solid state relay not connected or defect (unknown)
- 7. N40: Faulty port pins to EEPROM, I2C short circuit (unknown)
- 8. N42-57: Unspecified error (unknown)
- 9. N70-74: Unspecified error (unknown)
- 10. N80-82: Unspecified error (unknown)
- 11. N128-134: Advisory message (unknown)
- 12. N144: Activation signal not within 100 ms (unknown)
- 13. N145: Second signal not within 1000 ms (unknown)
- 14. N146: Microcontroller temperature rather low (unknown)
- 15. N147: Microcontroller temperature rather high (unknown)
- 16. N247-255: Advisory message (unknown)
16 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
P/P-A error displayed caution
Error code P or P-A shown on display, detecting module PSU (power supply unit)
Ranked causes
- 1. P1: CRC check alarm (unknown)
- 2. P2: ErrorHandler Buffer Overflow (unknown)
- 3. P5-8: Unspecified error (unknown)
- 4. P21-28: Unspecified error (unknown)
- 5. P34: Second signal absent >1.3 seconds, faulty software, timeout (unknown)
- 6. P35: Timeout in status self-check or activation/deactivation, faulty software (unknown)
- 7. P39: 12 V controller (network) voltage less than 8.5 V (unknown)
- 8. P40: 12 V PFC MOSFET driver voltage less than 8.5 V (unknown)
- 9. P51: ADC interrupt renewal before service routine completion, ISR runtime too long (unknown)
- 10. P127: ECB command does not match assembly state specifications (unknown)
- 11. P128-134: Advisory message (unknown)
- 12. P246-255: CAN error in physical layer, CAN STATUS (unknown)
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
S/S-A error displayed caution
Error code S or S-A shown on display, detecting module SCM (system control module)
Ranked causes
- 1. S1: Checksum calculation application check alarm, faulty software (unknown)
- 2. S3: Checksum calculation mode list alarm, faulty software (unknown)
- 3. S4: Checksum calculation version list alarm, faulty software (unknown)
- 4. S10: Mainboard 12 V voltage not within ±10% (unknown)
- 5. S11: Mainboard 5 V voltage not within ±10% (unknown)
- 6. S63-74: Unspecified error (unknown)
- 7. S78-84: Unspecified error (unknown)
- 8. S88-94: Unspecified error (unknown)
- 9. S128: SCM start triggered by watchdog reset (unknown)
- 10. S148: Main fan rotational speed insufficient (occasional)
- 11. S159: Invalid instrument parameters, UI rejection (occasional)
- 12. S-A 18: Maximum average power output reached, heat management (occasional)
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 190: Return electrode symmetry warning patient safety
NESSY detects marginal symmetry of return electrode positioning
Ranked causes
- 1. Return electrode alignment suboptimal (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 191: Return electrode current density warning patient safety
NESSY warns of elevated current density under return electrode, risking thermal injury
Ranked causes
- 1. Return electrode current density elevated (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 192: NESSY HF current exceeds 300 mA patient safety
NESSY detects high-frequency current exceeding 300 mA threshold (standard or neonatal mode)
Ranked causes
- 1. HF current exceeds safe limit, elevated temperature risk at electrode (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 48: No return electrode type defined patient safety
NESSY error - return electrode type not selected in system
Ranked causes
- 1. Return electrode type not configured (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 49: No return electrode contact detected patient safety
NESSY return electrode monitoring has not detected any skin contact
Ranked causes
- 1. Return electrode not making skin contact (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 50: Non-split return electrode connected patient safety
NESSY system detects wrong type of return electrode (non-split when split expected or vice versa)
Ranked causes
- 1. Wrong return electrode type installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 51: Return electrode skin contact insufficient patient safety
NESSY detects inadequate return electrode contact with patient skin (R_NE exceeds dynamic limit)
Ranked causes
- 1. Insufficient return electrode skin contact (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 52: Return electrode symmetry error patient safety
NESSY detects misalignment or poor distribution of return electrode on patient
Ranked causes
- 1. Return electrode not properly aligned or positioned (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 53: Short circuit at split return electrode patient safety
NESSY detects electrical short circuit in split-type return electrode
Ranked causes
- 1. Return electrode internal short circuit or damage (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 54: Resistance at non-split return electrode patient safety
NESSY detects abnormal resistance in non-split return electrode
Ranked causes
- 1. Return electrode contact resistance abnormal (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
N-A 55: Split return electrode connected when non-split expected patient safety
NESSY detects mismatch between configured and physically connected return electrode type
Ranked causes
- 1. Wrong return electrode type connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
APC-1 | CRC check triggers an alarm | CRC checksum mismatch in Gas Control Module | error |
APC-122 | EEPROM read error | Failure to read system EEPROM memory | error |
APC-123 | EEPROM write error | Failure to write to system EEPROM memory | error |
APC 128-134 | Internal error | Various internal faults in GCM | warning |
APC-128 to APC-134 | Advisory message | Advisory condition in GCM | warning |
APC-150 | Internal unit temperature exceeds 80°C | Temperature sensor in unit housing reads >80°C | warning |
APC-151 | Controller temperature exceeds 80°C | Temperature sensor in controller reads >80°C | warning |
APC-2 | ErrorHandler Buffer Overflow | Buffer overflow in error handling system | error |
APC 20-38 | Internal error | Various internal faults in GCM | error |
APC-20 to APC-38 | Unknown error condition | Unspecified error in Gas Control Module | warning |
APC 246-255 | Internal error | Various internal faults in GCM | warning |
APC-246 to APC-255 | Advisory message | Advisory condition in GCM | warning |
APC 40-46 | Internal error | Various internal faults in GCM | error |
APC-40 to APC-46 | Unknown error condition | Unspecified error in Gas Control Module | warning |
APC 50-59 | Internal error | Various internal faults in GCM | error |
APC-50 to APC-59 | Unknown error condition | Unspecified error in Gas Control Module | warning |
APC 5-8 | Internal error | Various internal faults in GCM | error |
APC-5 to APC-8 | Unknown error condition | Unspecified error in Gas Control Module | warning |
APC-A 10 | Input pressure exceeds 7 bar (overpressure in gas supply) | Pressure sensor reading > 7 bar (rel) | warning |
APC-A 11 | Input pressure below 3 bar with bottle pressure above 25 bar (vacuum in gas supply, gas hose not connected) | Pressure < 3 bar (rel) and bottle pressure > 25 bar (rel) | warning |
APC-A 12 | Input pressure below 3 bar with no bottle pressure sensor signal (vacuum in gas supply, sensor not connected) | Pressure < 3 bar (rel) and no bottle pressure sensor signal | warning |
APC-A-13 | Output pressure exceeds 2 bar during activation - possible instrument blockage | Gas output pressure >2 bar (rel) when unit is activated | warning |
APC-A-148 | Bottle pressure low - gas bottle nearly empty | Gas bottle pressure <25 bar (rel) | info |
APC-A 150 | No measurement signal from bottle pressure sensor (sensor not connected) | No bottle pressure sensor signal when gas supply set to gas bottle | warning |
APC-A 151 | Input pressure below 3 bar with bottle pressure above 25 bar (vacuum in gas supply, gas hose not connected) | Pressure < 3 bar (rel) and bottle pressure > 25 bar (rel) | warning |
EIP-A-128 | CRC error in irrigation pump module | CRC checksum mismatch in EIP 2 module | error |
EIP-A 130 | Pump cover opened when switched on | Pump cover detected open during unit operation | error |
EIP-A 132 | Timeout error in irrigation pump | Pump operation timeout | error |
EIP-A-136 | Uetest error | Unspecified error in EIP module test function | error |
EIP-A 136 | UeWS error in irrigation pump | UeWS protocol failure in EIP 2 | error |
EIP-A-144 | No electricity or no consumer detected in pump module | EIP module detects no power supply or load | error |
EIP-A 160 | Current too high when switched on in irrigation pump | Excessive current detected during pump startup | error |
EIP-A-192 | Motor control defective - current too high during activation | Motor control circuit detects excessive current during operation | critical |
G 1 | CRC check triggers an alarm in generator | CRC check failure in UGM | error |
G-10 | D-Flash programming error | Data flash memory programming operation failed | critical |
G-11 | P-Flash programming error | Program flash memory programming operation failed | critical |
G-5 | Oscillator clock fault | Generator main clock circuit malfunction detected | critical |
G-6 | Oscillator fault | Generator oscillator circuit malfunction | critical |
G 75 | Temperature error - generator temperature too high | Generator temperature sensor reading exceeds threshold | warning |
G-A 75 | Temperature error - generator temperature too high | Generator temperature sensor reading exceeds threshold during activation | warning |
I1 | CRC check triggers an alarm on accessory socket module | CRC checksum error detected by IIM module during accessory detection | error |
I 10 | D-Flash programming error in instrument interface | Programming error during D-Flash access | error |
I 11 | P-Flash programming error in instrument interface | Programming error during P-Flash access | error |
I 39 | Undefined or impermissible connector detection or faulty socket | Undefined connector detected or socket malfunction | error |
I45-50 | IIM module error codes 45 through 50 | Unspecified IIM module detection error | error |
I 45-50 | Internal error in instrument interface | Various internal faults in IIM | error |
I 5-7 | Internal error in instrument interface | Various internal faults in IIM | error |
I5-7 | IIM module error codes 5 through 7 | Unspecified IIM module detection error | error |
I57-63 | IIM module error codes 57 through 63 | Unspecified IIM module detection error | error |
I 57-63 | Internal error in instrument interface | Various internal faults in IIM | error |
I-A 30 | 1-wire CRC incorrect | CRC error in 1-wire accessory communication | error |
I-A30 | 1-wire CRC checksum error on accessory identification line | CRC mismatch during 1-wire communication with accessory module | error |
I-A31 | 1-wire accessory type incorrectly detected or identified | Accessory identification data does not match expected type during 1-wire initialization | error |
I-A32 | 1-wire no Erbe instrument detected on accessory line | Accessory module does not respond or not recognized as Erbe instrument via 1-wire | error |
I-A33 | Short circuit detected on instrument recognition (IK) line | IK line short-circuited during accessory initialization or operation | error |
I-A34 | Activation ID malfunction on accessory module | Accessory activation identifier is missing or invalid | error |
I-A35 | SK socket error - undefined: left/center pin present but right pin missing or vice versa | Socket contact pins not fully or correctly inserted in monopolar/bipolar socket | error |
I-A36 | SK socket error - bipolar: 2 pins detected at each plug detection (should be 1 or 3) | Pin count mismatch during monopolar socket insertion detection | error |
I-A37 | SK socket error - monopolar: 3 pins detected at bipolar socket (should be 2+1) | Pin count mismatch during bipolar socket insertion detection | error |
I-A38 | 1-wire protocol version incompatible with accessory module | Accessory protocol version does not match unit protocol version | warning |
I-A40 | One or more buttons detected as pressed during accessory insertion | Button depression detected on accessory at moment of connection | error |
IES-A1 | Smoke evacuator software error - checksum mismatch | Software checksum validation fails during IES module initialization | error |
IES-A128 | CAN bus error on IES module | CAN communication failure or timeout during module operation | error |
IES-A129 | Setting parameter error on IES module | Invalid or out-of-range parameter detected during configuration or operation | error |
IES-A20 | Smoke evacuator overtemperature alarm | Internal temperature of IES module exceeds safe operating limit | warning |
IES-A21 | Smoke evacuator undertemperature alarm | Internal temperature of IES module below safe operating specification | warning |
IES-A22 | Pressure sensor error in smoke evacuator module | Pressure sensor fails self-test or communication error during initialization | error |
IES-A23 | Filter cartridge used up - end of service life reached | Operating hours or filtration capacity threshold exceeded | warning |
IES-A24 | High suction resistance indicating filter blockage | Measured airflow resistance exceeds threshold during operation | warning |
IES-A25 | Filter cartridge not detected during initialization | Filter cartridge presence detection fails at module startup or status check | error |
M1 | CRC check error on program code in MCM boot loader | Boot loader self-test detects program code checksum mismatch | error |
M100-105 | MCM error codes 100 through 105 (unspecified) | Unknown MCM fault during operation or initialization | error |
M 100-105 | Internal error in measurement module | Various internal faults in MCM | error |
M 10-14 | Internal error in measurement module | Various internal faults in MCM | error |
M10-14 | MCM error codes 10 through 14 (unspecified) | Unknown MCM fault during operation or initialization | error |
M110-112 | MCM error codes 110 through 112 (unspecified) | Unknown MCM fault during operation or initialization | error |
M 110-112 | Internal error in measurement module | Various internal faults in MCM | error |
M113 | Spark available in mode that does not permit sparks during activation | Spark detected in non-spark mode during mode activation sequence | error |
M114 | Insufficient HF sensor signal during mode activation | No AC signal from HF current or voltage sensors at expected threshold | error |
M120 | HF power exceeds power supply unit output for extended period during activation | HF output power measured greater than power supply specification during operation | error |
M121 | Power supply nominal voltage reaches maximum with no HF peak voltage generated | Supply voltage saturates while HF output remains absent during activation | error |
M131 | CAN warning - CAN error counter at level 96 (warning threshold) | CAN error counter accumulation reaches 96 during operation | warning |
M132 | CAN error during operation on MCM module | CAN communication failure detected during active unit operation | error |
M133 | MCM temperature measurement malfunction | Temperature sensor on MCM fails self-test or communication error | error |
M134 | PFC timeout - PFC signals not good for 50 ms during operation | Power Factor Correction module fails status check during power supply operation | error |
M30 | CAN timeout of activation message from SCM (120 ms threshold) | CAN bus timeout waiting for mode activation message from SCM module | error |
M31 | Unknown mode identifier - mode not recognized by MCM | Activation request received for unsupported or undefined surgical mode | error |
M32 | Timeout after self-check - no activation message received | MCM self-check completes but no activation message follows within timeout | error |
M33 | Incorrect ECB mode parameter received during activation | Invalid effect setting parameter received via CAN during mode activation | error |
M34-43 | MCM error codes 34 through 43 (unspecified) | Unknown MCM fault during operation or initialization | error |
M 34-43 | Internal error in measurement module | Various internal faults in MCM | error |
M5 | CAN initialization failure on MCM module | CAN bus failed to initialize during MCM startup sequence | error |
M 50-63 | Internal error in measurement module | Various internal faults in MCM | error |
M50-63 | MCM error codes 50 through 63 (unspecified) | Unknown MCM fault during operation or initialization | error |
M 70-88 | Internal error in measurement module | Various internal faults in MCM | error |
M70-88 | MCM error codes 70 through 88 (unspecified) | Unknown MCM fault during operation or initialization | error |
M-A1 | thermoSEAL short circuit detected; no tissue effect in BiCision mode | Short circuit detected in thermoSEAL output during BiCision activation | error |
M-A10 | Initial incision incomplete in highCUT bipolar mode | Initial incision did not complete successfully when highCUT bipolar activated | error |
M-A11 | highCUT bipolar no-load - instrument not in NaCl or poor contact | Expected saline conductivity not detected during highCUT bipolar activation | warning |
M-A2 | thermoSEAL no-load condition - insufficient tissue contact | Expected tissue resistance not detected during thermoSEAL activation | warning |
M-A 20 | Power overdose with PPS/QuickStart | Excessive power detected during PPS or QuickStart activation | warning |
M-A 30 | endoCUT Q idle state (no tissue contact for two initial incision cycles) | No tissue contact detected for duration of two initial incision cycles | warning |
N1 | CRC check alarm | Checksum verification failure in NM module | error |
N10-14 | Unspecified NM error | Various undocumented faults in NM module | error |
N128-134 | Advisory message (NM) | Various warning conditions in NM | warning |
N144 | Activation signal timing error | Activation signal not received within 100 ms window | warning |
N145 | Second signal timing error | Second signal not received within 1000 ms window | warning |
N146 | Microcontroller temperature low | Microcontroller temperature below threshold | warning |
N147 | Microcontroller temperature high | Microcontroller temperature above threshold | warning |
N17-36 | Unspecified NM error | Various undocumented faults in NM module | error |
N2 | ErrorHandler Buffer Overflow | Error handling buffer overflow in NM | error |
N247-255 | Advisory message (NM) | Various warning conditions in NM | warning |
N38 | Faulty circuit, solid state relay issue | Circuit fault or solid state relay not connected | error |
N40 | Faulty port pins to EEPROM, I2C interface error | Short circuit or communication fault at I2C interface | error |
N42-57 | Unspecified NM error | Various undocumented faults in NM module | error |
N5-8 | Unspecified NM error | Various undocumented faults in NM module | error |
N70-74 | Unspecified NM error | Various undocumented faults in NM module | error |
N80-82 | Unspecified NM error | Various undocumented faults in NM module | error |
N-A190 | Return electrode symmetry warning (NESSY) | Electrode contact distribution marginal but functional | warning |
N-A191 | Current density warning at return electrode (NESSY) | HF current density elevated under return electrode | warning |
N-A192 | High-frequency current exceeds 300 mA (NESSY) | NESSY HF current output > 300 mA threshold | warning |
N-A48 | No return electrode type defined (NESSY) | Return electrode type not selected in system setup | warning |
N-A49 | No return electrode contact detected (NESSY) | NESSY monitoring detects no skin contact | error |
N-A50 | Wrong return electrode type connected (NESSY) | Non-split electrode connected when split expected or vice versa | warning |
N-A51 | Return electrode skin contact resistance excessive (NESSY) | R_NE exceeds dynamic limit threshold | error |
N-A52 | Return electrode symmetry error (NESSY) | Electrode contact pressure or position uneven | warning |
N-A53 | Short circuit in split return electrode (NESSY) | Electrical short detected in split-type electrode | error |
N-A54 | Abnormal resistance in non-split return electrode (NESSY) | Contact resistance out of range on non-split electrode | warning |
N-A55 | Split return electrode connected (mismatch) (NESSY) | Split-type electrode detected when non-split configured | warning |
P1 | CRC check alarm (PSU) | Checksum verification failure in power supply unit | error |
P127 | ECB command state mismatch (PSU) | ECB process command does not match assembly state | error |
P128-134 | Advisory message (PSU) | Various warning conditions in PSU | warning |
P2 | ErrorHandler Buffer Overflow (PSU) | Error handling buffer overflow in PSU | error |
P21-28 | Unspecified PSU error | Various undocumented faults in PSU | error |
P246-255 | CAN physical layer error (PSU) | CAN bus communication fault | warning |
P34 | Second signal timeout (PSU) | Second signal absent for >1.3 seconds | error |
P35 | Timeout in status/activation (PSU) | Self-check or activation/deactivation timeout | error |
P39 | Controller 12 V supply low (PSU) | Network controller 12 V < 8.5 V | error |
P40 | PFC MOSFET driver 12 V low (PSU) | PFC power factor correction driver 12 V < 8.5 V | error |
P51 | ADC interrupt service routine overrun (PSU) | New ADC interrupt before previous routine termination | error |
P5-8 | Unspecified PSU error | Various undocumented faults in PSU | error |
P-A | Not available | No P-A error codes documented | unknown |
S1 | Application checksum error (SCM) | Application checksum calculation fails | error |
S10 | Mainboard 12 V supply out of tolerance (SCM) | 12 V supply not within ±10% (10.8–13.2 V) | error |
S11 | Mainboard 5 V supply out of tolerance (SCM) | 5 V supply not within ±10% (4.5–5.5 V) | error |
S128 | Watchdog reset detected (SCM) | SCM restart initiated by watchdog timer | error |
S148 | Main fan speed insufficient (SCM) | Main cooling fan RPM below minimum threshold | warning |
S159 | Invalid instrument parameters (SCM) | UI rejects received instrument parameter values | warning |
S3 | Mode list checksum error (SCM) | Mode list checksum calculation fails | error |
S4 | Version list checksum error (SCM) | Version list checksum calculation fails | error |
S63-74 | Unspecified SCM error | Various undocumented faults in SCM | error |
S78-84 | Unspecified SCM error | Various undocumented faults in SCM | error |
S88-94 | Unspecified SCM error | Various undocumented faults in SCM | error |
S-A 129 | Heat management: High average power output reached | Unit achieves high sustained power output | warning |
S-A 130 | Instrument plugged in is blocked or unsupported | Incompatible or blocked instrument detected during operation | error |
S-A 149 | Double footswitch activation | Two footswitches activated simultaneously | warning |
S-A 150 | Undervoltage flag set in RTC chip | Real-time clock detects low voltage condition | info |
S-A 152 | Rinsing of an APC instrument triggered | APC instrument rinsing cycle initiated | info |
S-A 154 | No footswitch assigned on activation | Activation attempted without assigned footswitch | error |
S-A 155 | No mode for activation source assigned on activation | Activation attempted without assigned mode | error |
S-A 156 | New irregular activation signal applied before previous activation ends regularly | Second activation signal arrives during ongoing activation | warning |
S-A 157 | No instrument connected on activation | Activation attempted without instrument connection | error |
S-A 163 | IOSK instrument connected to the system | IOSK instrument detected and evaluated by sterile interpreter | info |
S-A 164 | Instrument was removed or unplugged from the system | Instrument disconnected during operation or standby | info |
S-A 166 | APC 3 module is not ready for operation | APC 3 module fails initialization or loses readiness | error |
S-A18 | Maximum power output reached (heat management) | Average HF power output at or near system limit | warning |
S-A 19 | Maximum activation time reached (maximum ON time) | Activation time exceeds configured maximum | warning |
S-A 198 | EIP ready for operation | EIP module initialization complete | info |
S-A 199 | EIP disconnected | EIP module disconnection detected | info |
S-A 200 | IES ready for operation | IES module initialization complete | info |
S-A 201 | IES disconnected | IES module disconnection detected | info |
S-A 204 | One pedal ReMode footswitch ready for operation | One pedal ReMode footswitch initialization complete | info |
S-A 205 | One pedal ReMode footswitch disconnected | One pedal ReMode footswitch disconnection detected | info |
S-A 206 | Two pedal ReMode footswitch ready for operation | Two pedal ReMode footswitch initialization complete | info |
S-A 207 | Two pedal ReMode footswitch disconnected | Two pedal ReMode footswitch disconnection detected | info |
S-A 208 | One pedal IES footswitch ready for operation | One pedal IES footswitch initialization complete | info |
S-A 209 | One pedal IES footswitch disconnected | One pedal IES footswitch disconnection detected | info |
S-A 21 | Critical unit temperature has just been reached | Unit temperature reaches critical threshold | warning |
S-A 22 | Second signal detected with higher priority | Incompatible module connected to ECB during operation | error |
S-A 23 | Unsupported unit connected to the ECB | Incompatible module connected to ECB | error |
S-A 24 | Contact monitor assignment when contact detected | Contact detected on instrument during monitoring | warning |
S-A 29 | Activation attempt when two identical footswitches are plugged in | Activation attempted with duplicate footswitch configuration | error |
S-A 30 | Two identical footswitches have been detected | System detects duplicate footswitch types on startup or during operation | error |
S-A 31 | A modified module configuration was detected | Internal module configuration change detected | warning |
S-A 32 | Modified socket configuration detected | Socket configuration change detected (replacement, retrofitting, or detection defect) | warning |
S-A 40 | Two pedal footswitch pedal pressed during first query | Two pedal footswitch reports pressed state during initial query cycle | warning |
S-A 41 | One pedal footswitch pedal pressed during first query | One pedal footswitch reports pressed state during initial query cycle | warning |
U 1 | DCB communication timeout | Data/Device Control Block fails to receive message from SCM during activation | error |
U 48 | Display brightness could not be set | Display brightness adjustment command fails | warning |
U 49 | Volume could not be set | Volume adjustment command fails | warning |
U-A 132 | Program memory full | Attempts to store additional programs exceed available memory | warning |
U-A 133 | Remote control detected | Remote control module detected during initialization | info |
U-A 134 | Remote control disconnected | Remote control module disconnection detected | info |
U-A 135 | Modification not possible | System rejects requested modification | info |
U-A 136 | Data transfer active | System is actively transferring data | info |
U-A 137 | The technical safety check is due | System detects that scheduled safety check interval has elapsed | warning |
U-A 138 | Assign activation type | System requires activation type assignment for operation | warning |
U-A 7 | Line voltage too low | Input mains voltage below acceptable range | error |
W 128–130 | Operating system warning | System error condition | warning |
W 140 | Error during setting of system time | System time configuration failure | warning |
W 17-24 | Write/read error partition | Memory read or write operation fails on partition | error |
W 1-8 | Checksum partition error | Memory partition checksum validation fails | error |
W 28-30 | Memory error | General memory subsystem error detected | error |
W 32 | Update error | Firmware or software update fails | error |
W 33 | Update procedure in progress | System is executing firmware or software update | info |
W 36 | No more RAM available | Available RAM exhausted | error |
W 37 | RAM error | RAM hardware malfunction detected | error |
W 40 | CPU module overheating | CPU temperature exceeds safe operating threshold | critical |
W 41 | Faulty CPU module voltage regulator | Voltage regulator output out of specification | error |
W 44 | Container error | Software container integrity check fails | error |
W 48-49 | Memory or software error | Memory system or software failure | error |
W 52-53 | Memory or software error | Memory system or software failure | error |
W 56 | Unexpected OP system behavior | System malfunction | warning |
W 9-16 | No memory partition found | Memory partition not accessible or missing | error |
W-A 128 | The connection to the system has been interrupted | Loss of system connection | warning |
W-A 129 | The connection to the system has been re-established | System connection restored | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI