ERBE VIO 200 D: troubleshooting and error codes

The ServiceForge knowledge base holds 46 documented symptoms for the ERBE VIO 200 D, with 46 ranked causes, 46 diagnostic checks with test points and expected values, and 247 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

APC module timeout or activation signal failure

System fails to receive activation signal from APC module or signal transmission is interrupted

Ranked causes

  1. 1. CAN transmission interrupted or cable positioning issue (common)
First check

Verify CAN transmission integrity

Test point
FS cable connection
Expected
Uninterrupted signal, no interference from HF cables

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

APC protocol error in status message (Control panel)

Control panel receives incorrect status message from APC module with invalid CAN ID or format

Ranked causes

  1. 1. APC module protocol error or software version mismatch (occasional)
First check

Verify APC status message format and CAN ID

Test point
APC module CAN message output
Expected
Status message has correct CAN ID and valid format per protocol specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

APC setup parameters invalid

System detects invalid APC setup parameters stored in configuration

Ranked causes

  1. 1. Invalid sector error in CAN data (occasional)
First check

Analyze CAN data for sector errors

Test point
CAN bus
Expected
Valid sector data in APC parameters

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

APC status error (Control panel) patient safety

APC module reports status 'active' when it should be switched off; status signal conflict

Ranked causes

  1. 1. APC module malfunction or EMC interference (occasional)
First check

Verify APC status signal accuracy

Test point
APC module status output on CAN bus
Expected
Status 'inactive' when APC is switched off; status 'active' only during activation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Button error on receptacle 1

Short circuit or interruption detected on receptacle board 1 button circuit

Ranked causes

  1. 1. Short circuit or interruption on receptacle board 1 connector (common)
First check

Inspect receptacle 1 button circuit

Test point
Receptacle board 1 connector
Expected
No short circuits or interruptions; clean connector contact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Button error on receptacle 2

Short circuit or interruption detected on receptacle board 2 button circuit

Ranked causes

  1. 1. Short circuit or interruption on receptacle board 2 connector (common)
First check

Inspect receptacle 2 button circuit

Test point
Receptacle board 2 connector
Expected
No short circuits or interruptions; clean connector contact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Caloric and differential pressure sensor mismatch caution

Caloric and differential pressure sensors disagree on gas type; indicates wrong gas supply

Ranked causes

  1. 1. Wrong gas supplied or sensor calibration/hardware fault (common)
First check

Verify correct gas type at input

Test point
Gas input connector
Expected
Gas type matches system requirement (e.g., O2, N2O, CO2)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CAN bus error

CAN (Controller Area Network) bus detects length or data error in transmitted packet

Ranked causes

  1. 1. CAN data length or format error (occasional)
First check

Verify CAN bus packet length

Test point
CAN bus interface
Expected
All CAN packets conform to length and format specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CAN protocol violation (length)

CAN bus detects protocol violation in packet length or format

Ranked causes

  1. 1. CAN packet format error or device malfunction (occasional)
First check

Verify CAN protocol compliance

Test point
CAN bus interface
Expected
All CAN packets conform to protocol specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Checkstop interrupt invalid memory access (Control panel)

Control panel detects memory access violation; program crash due to software error or malfunction

Ranked causes

  1. 1. Software error or EMC interference (occasional)
First check

Monitor for memory access errors

Test point
System console output
Expected
No checkstop interrupt messages; normal memory access

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Differential pressure sensor reading low

Differential pressure sensor measures less than half the expected setting; sensor appears defective

Ranked causes

  1. 1. Sensor calibration error or sensor hardware fault (occasional)
First check

Measure differential pressure sensor output

Test point
Differential pressure sensor connector
Expected
Reading >= 50% of expected setting for current gas supply pressure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

EEPROM not ready to read

System cannot access EEPROM memory; read operation fails intermittently

Ranked causes

  1. 1. EEPROM access conflict or secondary fault condition (occasional)
First check

Verify EEPROM access is available

Test point
EEPROM interface
Expected
EEPROM read operation completes successfully without timeout

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error in test mode or calibration

Error occurs during calibration or diagnosis procedure, typically software-related and often uncritical

Ranked causes

  1. 1. PC program error during calibration or diagnosis (occasional)
First check

Repeat calibration or diagnostic test

Test point
PC diagnostic interface
Expected
Test completes without software error

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow specification exceeded caution

Gas flow rate exceeds maximum target flow specification

Ranked causes

  1. 1. Calibration error or faulty flow sensor (occasional)
First check

Measure actual gas flow rate

Test point
Gas outlet measurement point
Expected
Flow rate <= target specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow specification not attained

Gas flow rate does not reach target flow specification

Ranked causes

  1. 1. User error in flow rate setting (common)
First check

Measure actual gas flow rate

Test point
Gas outlet measurement point
Expected
Flow rate >= target specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Function not available (Control panel)

Control panel detects that a required function cannot be implemented by the available modules

Ranked causes

  1. 1. Module software version mismatch or missing feature (occasional)
First check

Verify module software version and enabled features

Test point
Module firmware version and feature configuration
Expected
All modules have compatible software version; required function is enabled

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas input pressure above maximum caution

Overpressure detected at selected gas input; gas supply pressure too high

Ranked causes

  1. 1. User error or faulty input pressure measuring circuit (common)
First check

Verify gas input pressure at selected input

Test point
Gas input connector
Expected
Pressure <= maximum operating pressure per screen instructions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas input pressure below minimum caution

Underpressure detected at selected gas input; gas supply pressure too low

Ranked causes

  1. 1. User error or faulty input pressure measuring circuit (common)
First check

Verify gas input pressure at selected input

Test point
Gas input connector
Expected
Pressure >= minimum operating pressure per screen instructions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas underdose from instrument caution

Gas delivery from instrument is insufficient; hose may be blocked or instrument defective

Ranked causes

  1. 1. Instrument defective or hose blocked (common)
First check

Verify gas flow from instrument

Test point
Gas outlet hose of instrument
Expected
Adequate gas flow with no blockage or restriction

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

I²C bus error on EEPROM interface (Control panel) high voltage

Serial I²C communication error on control panel EEPROM interface; cannot read or write EEPROM data

Ranked causes

  1. 1. EEPROM or I²C bus component fault (occasional)
First check

Verify I²C bus voltage and EEPROM communication

Test point
I²C SDA/SCL lines on control panel PCB
Expected
I²C bus voltage normal (3-5 V); EEPROM responds to read/write commands

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

IIF module serial communication failure

Serial communication with Instrument Interface (IIF) module not possible; connection lost

Ranked causes

  1. 1. IIF module connection faulty or disconnected (common)
First check

Verify IIF module serial communication

Test point
IIF module connector
Expected
Serial communication established and responsive

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Input 2 selected with partial complement

Input 2 selection error; received configuration does not match expected

Ranked causes

  1. 1. CAN sector error in input configuration data (occasional)
First check

Verify Input 2 configuration in CAN data

Test point
CAN bus Input 2 configuration sector
Expected
Complete configuration data without sector errors

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal state incorrect

System internal state machine or data structure is in unexpected state

Ranked causes

  1. 1. Software error or data corruption (rare)
First check

Verify system internal state via diagnostics

Test point
System diagnostic menu
Expected
System state consistent with current operational mode

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid EEPROM calibration data

EEPROM contains invalid or corrupted calibration data; cannot load calibration on startup

Ranked causes

  1. 1. EEPROM calibration data corrupted or not programmed (occasional)
First check

Verify EEPROM calibration data integrity

Test point
EEPROM calibration sector
Expected
Valid calibration data present and checksum correct

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid instrument resistance number

Instrument resistance measurement reports invalid or undefined instrument number

Ranked causes

  1. 1. Instrument not properly registered or detected (common)
First check

Verify instrument resistance measurement value

Test point
Instrument resistance measurement circuit
Expected
Measured resistance corresponds to valid instrument type per database

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low gas pressure at cylinder 1

Gas bottle 1 pressure is low; bottle needs replacement

Ranked causes

  1. 1. Gas bottle 1 depleted or low pressure (common)
First check

Verify gas bottle 1 pressure

Test point
Cylinder 1 pressure gauge or sensor readout
Expected
Pressure >= minimum operating pressure (typically > 500 psi for full bottles)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low gas pressure at cylinder 2

Gas bottle 2 pressure is low; bottle needs replacement

Ranked causes

  1. 1. Gas bottle 2 depleted or low pressure (common)
First check

Verify gas bottle 2 pressure

Test point
Cylinder 2 pressure gauge or sensor readout
Expected
Pressure >= minimum operating pressure (typically > 500 psi for full bottles)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement circuit overvoltage high voltage

Voltage on measurement circuit exceeds maximum operating threshold

Ranked causes

  1. 1. Power supply regulator fault or wiring damage (occasional)
First check

Measure voltage on measurement circuit

Test point
Measurement circuit power supply node
Expected
Voltage <= maximum specified (typically 5.5 V for sensor circuits)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement circuit undervoltage high voltage

Voltage on measurement circuit drops below minimum operating threshold

Ranked causes

  1. 1. Power supply or wiring fault in measurement circuit (occasional)
First check

Measure voltage on measurement circuit

Test point
Measurement circuit power supply node
Expected
Voltage >= minimum specified (typically 5 V for sensor circuits)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement circuit voltage error combined high voltage

Both undervoltage and overvoltage conditions detected simultaneously; reference voltage fault

Ranked causes

  1. 1. Reference voltage circuit fault or power supply failure (occasional)
First check

Verify reference voltage stability

Test point
Reference voltage node on measurement circuit
Expected
Stable voltage output, no oscillation or dropout

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module timeout during activation (Control panel)

Control panel detects timeout; module failed to transmit status message via CAN within 110 ms timeout

Ranked causes

  1. 1. Faulty module or EMC interference on CAN bus (common)
First check

Verify module CAN status message transmission

Test point
CAN bus interface for identified module
Expected
Module transmits valid status message every < 110 ms

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

NE (Nessy2) module not available patient safety

NE module missing or not logged on when monopolar activation is initiated

Ranked causes

  1. 1. NE module failure or disconnection (common)
First check

Verify NE module is logged on and available

Test point
NE (Nessy2) module interface
Expected
NE module logged on and ready for monopolar operations

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

NE contact resistance out of range patient safety

NESSY message indicates NE (neutral electrode) contact resistance is not within valid range on activation

Ranked causes

  1. 1. Patient neutral electrode not properly applied or NE module faulty (common)
First check

Measure NE contact resistance on activation

Test point
NE electrode contact resistance measurement
Expected
Resistance within valid operating range (typically < 50 ohms)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Operating system error

Operating system encounters fault during execution or after software update

Ranked causes

  1. 1. Software error during operation or after update; EMC interference possible (occasional)
First check

Monitor system operation after power cycle

Test point
System console output
Expected
No operating system error message; normal system operation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Program CRC test failure

Cyclic Redundancy Check (CRC) of program memory detects corruption or mismatch

Ranked causes

  1. 1. Program memory corruption or version mismatch (occasional)
First check

Perform firmware integrity check

Test point
System program memory
Expected
Program CRC matches expected value; no memory corruption

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proportional valve control transistor fault

Control transistor for proportional valve is faulty or damaged

Ranked causes

  1. 1. Control transistor faulty (occasional)
First check

Test proportional valve control transistor

Test point
Transistor junction on APC PCB
Expected
Transistor forward/reverse biased normally; no short or open junction

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proportional valve interruption

Open circuit detected in proportional valve coil circuit

Ranked causes

  1. 1. Proportional valve coil not inserted or faulty (common)
First check

Verify proportional valve coil insertion and continuity

Test point
Proportional valve coil connector
Expected
Coil properly inserted with secure connection and continuity

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proportional valve short circuit

Short circuit detected in proportional valve coil circuit

Ranked causes

  1. 1. Short circuit in proportional valve circuit (occasional)
First check

Measure proportional valve coil resistance

Test point
Proportional valve coil connector
Expected
Normal coil resistance (no short), typically 10-50 ohms depending on design

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Receptacle 1 type detection mismatch

Type detection of receptacle 1 does not match stored configuration value

Ranked causes

  1. 1. Receptacle configuration mismatch or incorrect version list (common)
First check

Verify receptacle 1 type against version list

Test point
Receptacle 1 physical slot and system configuration database
Expected
Type detection matches stored configuration value per version list

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Receptacle 2 type detection mismatch

Type detection of receptacle 2 does not match stored configuration value

Ranked causes

  1. 1. Receptacle configuration mismatch or incorrect version list (common)
First check

Verify receptacle 2 type against version list

Test point
Receptacle 2 physical slot and system configuration database
Expected
Type detection matches stored configuration value per version list

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety relay IIF1 position fault patient safety

Relay on IIF module 1 or control on APC controller is faulty or misaligned

Ranked causes

  1. 1. Relay on IIF module 1 or APC controller faulty (common)
First check

Verify safety relay IIF1 position

Test point
IIF module 1 relay contact
Expected
Relay in correct position with secure contact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety relay IIF2 addressing error patient safety

Safety relay IIF2 is incorrectly addressed via relay 1+i instead of direct addressing

Ranked causes

  1. 1. Relay on IIF module 2 or APC controller faulty (common)
First check

Verify safety relay IIF2 addressing

Test point
IIF module 2 relay control
Expected
Direct relay addressing, not via relay 1+i

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SIO to IIF protocol violation

Serial I/O to Instrument Interface communication protocol error detected

Ranked causes

  1. 1. IIF module communication fault or timeout (occasional)
First check

Monitor SIO to IIF communication errors

Test point
IIF module serial interface
Expected
Communication completes without protocol violation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SIO to IIF timeout

Serial I/O to Instrument Interface communication timeout; no response from IIF

Ranked causes

  1. 1. IIF module not responding or cable fault (occasional)
First check

Verify SIO to IIF communication response time

Test point
IIF module serial interface
Expected
IIF responds to SIO query within timeout period (typically < 500 ms)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software watchdog triggered (Control panel)

Control panel CPU watchdog timer detects program crash or hang due to software error or malfunction

Ranked causes

  1. 1. Software error or EMC interference (occasional)
First check

Monitor for software watchdog occurrences

Test point
System console output
Expected
No watchdog timeout messages; stable program execution

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Valve error on APC module

Input valve coil is not inserted or is faulty; no coil connection detected

Ranked causes

  1. 1. Input valve coil not inserted or faulty (common)
First check

Verify input valve coil insertion and continuity

Test point
Input valve coil connector
Expected
Coil properly inserted with secure connection and continuity

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1Interrupt error: undefined IR vector invokedInvalid interrupt vector accessed during operationcritical
10Current sensor 1 defectiveNo signal provided by current sensor UE13error
11Current sensor 2 defectiveNo signal provided by current sensor UE12error
2CRC error: defective program memory discovered during self-checkProgram memory CRC check failed after resetcritical
2CRC errorDefective program memory discovered during self-check after reseterror
3Undervoltage 5V5V voltage discovered to be too low during self-check after reseterror
3Undervoltage 5 V: voltage too low during self-check5 V supply measured below acceptable range after reseterror
4Overvoltage 5V5V voltage discovered to be too high during self-check after reseterror
4Overvoltage 5 V: voltage too high during self-check5 V supply measured above acceptable range after reseterror
5Undervoltage ±12 V: +12 V too low or –12 V too high+12 V or –12 V supply out of range during self-checkerror
6Overvoltage ±12 V: +12 V too high or –12 V too low+12 V or –12 V supply out of range during self-checkerror
6Overvoltage ±12V+12V voltage discovered to be too high or -12V voltage too low during self-check after reseterror
7Undercurrent activation recognition: measurement current too lowActivation recognition current below acceptable thresholderror
7I-source defectiveNo signal provided by power source for NE contact resistance measurementerror
70Relay errorRelay configuration or state error detectederror
71Relay errorRelay state validation failederror
72Relay errorRelay state validation failederror
8Overcurrent activation recognition: measurement current too highActivation recognition current above acceptable thresholderror
8MOSI_errorSPI receive register contains $FF after Interrupt call-uperror
80CRC ErrorEIP 2 CRC monitoring procedure at unit startup discovers an errorerror
81Short circuit instrument recognition R: resistor value indicates shortInstrument recognition resistor measurement detects short circuit conditionerror
82Pump lid open at ONPump lid of EIP 2 not closed at start or opened during activation; Hall sensor circuit board not connected or faultywarning
82Short circuit activation recognition R_NEU: resistor indicates shortActivation recognition resistor measurement detects short circuiterror
83No-load activation recognition R_NEU: resistor indicates open circuitActivation recognition resistor measurement detects open circuiterror
84Instrument recognition R-window violated: resistor cannot be assignedResistor value outside defined recognition window for instrument typeerror
84Time Out ErrorActivation of EIP 2 maintained for more than 20 secondswarning
85Short circuit activation recognition R_ALT: resistor indicates shortAlternate activation recognition resistor detects short circuiterror
86Activation error: activation not consistent with instrument typeActivation mode selected does not match recognized instrument typeerror
87Framing error: UART did not detect stop bit at end of byteSerial receive failed to detect valid stop biterror
88UeWS ErrorMotor control monitoring circuit is faultyerror
88Parity error: check for even parity was unsuccessfulSerial data received with incorrect parity biterror
89Unknown identifier: interrupt routine detected non-agreed identifierReceived message has unrecognized identifier or formaterror
8ADS2430 CRC error: received data record from instrument defectiveCRC check failed on one-wire device data recorderror
8BNon-permissible instrument recognition R_S3Non-permissible value determined for resistor-coded instrument recognitionerror
8CNon-permissible sockets and relay statusNon-permissible value determined for socket recognitionerror
8DNon-permissible instrument recognition R_S4Non-permissible value determined for resistor-coded instrument recognitionerror
8ENon-permissible instrument recognition R_S8Non-permissible value determined for resistor-coded instrument recognitionerror
8FUART timeoutUPR Receive_UART has not received anything 351us after call-uperror
9Undercurrent instrument recognition: measurement current too lowInstrument recognition current below acceptable thresholderror
90No current or no loadNo measurable current input to motor during activationerror
90DS2430_Write errorAttempt to write to EEPROM for instrument recognition unsuccessfulerror
AOvercurrent instrument recognition: measurement current too highInstrument recognition current above acceptable thresholderror
A0Current too high at ONCurrent input to motor is too higherror
A1Timeout of activation signalCAN transmission interrupted or FS cable improperly positioned relative to HF cableerror
A10Gas input underpressurePressure sensor reads below minimum threshold at selected gas inputwarning
A11Gas input overpressurePressure sensor reads above maximum threshold at selected gas inputwarning
A12Gas type mismatch between sensorsCaloric and differential pressure sensors disagree on gas typewarning
A13Differential pressure sensor reading lowDifferential pressure sensor output less than 50% of expected settingwarning
A2APC setup parameters invalidInvalid sector error in CAN configuration dataerror
A21Measurement circuit undervoltageVoltage on measurement circuit drops below minimum operating thresholderror
A22Measurement circuit overvoltageVoltage on measurement circuit exceeds maximum operating thresholderror
A23Measurement circuit voltage error combinedBoth undervoltage (A21) and overvoltage (A22) conditions detected simultaneouslyerror
A3Safety relay IIF1 position faultRelay on IIF module 1 or APC controller malfunctioncritical
A30Input 2 selected with partial complementInput 2 configuration sector error in CAN datawarning
A38Receptacle 1 type detection mismatchType detection does not match stored configuration value in system databasewarning
A39Receptacle 2 type detection mismatchType detection does not match stored configuration value in system databasewarning
A4Safety relay IIF2 addressing errorRelay on IIF module 2 or APC controller malfunctioncritical
A40Flow specification not attainedGas flow rate does not reach target flow specificationwarning
A41Flow specification exceededGas flow rate exceeds maximum target flow specificationwarning
A5Button error receptacle 1Short circuit or interruption on receptacle board 1 connectorerror
A6Button error receptacle 2Short circuit or interruption on receptacle board 2 connectorerror
A7Input valve errorInput valve coil not inserted or faulty; no continuity detectedwarning
A7DProgram CRC test failureCyclic Redundancy Check detects program memory corruption or mismatcherror
A7EEEPROM not ready to readSystem cannot access EEPROM memory; read operation failswarning
A7FOperating system errorOperating system fault during execution or after software updateerror
A8Proportional valve interruptionOpen circuit in proportional valve coil circuitwarning
A80Internal state incorrectSystem internal state machine in unexpected statewarning
A81CAN protocol violationCAN bus detects packet length or format violationerror
A82SIO to IIF protocol violationSerial I/O to IIF communication protocol errorwarning
A83SIO to IIF timeoutSerial I/O to IIF communication timeout; no response from IIFwarning
A85Invalid instrument resistance numberInstrument resistance measurement reports undefined instrument numberwarning
A86Gas underdose from instrumentGas delivery from instrument insufficient; underdose detectedwarning
A9Proportional valve short circuitShort circuit detected in proportional valve coil circuitwarning
A90Low pressure at cylinder 1Gas bottle 1 pressure drops below minimum thresholdwarning
A91Low pressure at cylinder 2Gas bottle 2 pressure drops below minimum thresholdwarning
AAProportional valve control transistor faultControl transistor malfunction or component failurewarning
AA0Invalid EEPROM calibration dataEEPROM contains invalid or corrupted calibration data on startuperror
ABError in test mode or calibrationSoftware error during calibration or diagnostic test executionwarning
ACIIF module serial communication failureSerial communication with IIF module lost or unavailableerror
ADCAN bus errorCAN packet length or data format violation detectederror
B1Software watchdog triggeredProgram crash due to software error or malfunctionerror
B10Please terminate activationActivation automatically terminated but activation signal remains longer than 5 swarning
B11Error during activationActivation not enabled for over 110 ms during activationerror
B12Activation signal during switch-onActivation signal present during initialization (e.g. footswitch pedal pressed)warning
B13Errors in list managementError identified in flash file during event or error list managementerror
B14CRC errorContinuous CRC monitoring detects program errorcritical
B15FLASH RAM verification errorContinuous monitoring detects error in DRAM memoryerror
B16Insufficient EEPROM memoryMemory for user programs is fullwarning
B17Double activationTwo activation signals present simultaneously within 100 mswarning
B18No deactivation signalHF or APC module fails to react to deactivation request for longer than 110 mserror
B19PowerFailPowerFail signal received but power failure does not occur within 2 serror
B1AError in parameter memoryChecksum monitoring of program detects errorerror
B1BCRC check not yet completedUnit operated before CRC check completed after software downloadwarning
B1CMaximum ON time exceededActivation lasts longer than ON time selected in SETUPwarning
B1DIncorrect instrument parametersInstrument recognized via instrument recognition but parameter outside toleranceserror
B1EKeyboard errorButton pressed recognized during initializationwarning
B1FNESSY symmetry monitoringError signaled by Nessy symmetry monitoring during activationwarning
B2Module timeout during activationModule fails to transmit CAN status message within 110 ms timeouterror
B21Incorrect or missing bitmap fileIncompatible bitmap file founderror
B22Tissue still touched after activation terminatedActivation automatically terminated (e.g. by AutoStop) but tissue still in contactwarning
B23Configuration error of HF CPUHF CPU transmits invalid status message (0x27)error
B24Incorrect status message from IES 2Invalid status message received from IES 2 moduleerror
B25Modified receptacle configuration detectedReceptacle configuration changed (e.g. replacement or retrofitting)error
B5APC status errorAPC module signals 'active' status when it should be offcritical
B6APC protocol error in status messageInvalid status message received from APC module with incorrect CAN IDerror
B81Indication of incorrect CAN protocolNon-ECB-specified CAN message receivederror
B82Indication of receptacle arrangement changeChanges in receptacle assignment signaled during operationwarning
B84Dual-pedal footswitch recognizedDual-pedal footswitch connected to systeminfo
B85Dual-pedal footswitch disconnectedDual-pedal footswitch disconnected from systeminfo
B86APC receptacle 1 recognizedAPC receptacle 1 detected and connectedinfo
B87No status message from APC receptacle 1APC receptacle 1 disconnected or not respondingerror
B88Single-pedal footswitch recognizedSingle-pedal footswitch connected to systeminfo
B89Single-pedal footswitch disconnectedSingle-pedal footswitch disconnected from systeminfo
B8AAPC receptacle 2 recognizedAPC receptacle 2 detected and connectedinfo
B8BNo status message from APC receptacle 2APC receptacle 2 disconnected or not respondingerror
B8CAPC module recognizedAPC module detected and connectedinfo
B8DNo status message from APC moduleAPC module disconnected or not respondingerror
B8ENo status message from HF receptacle 1HF receptacle 1 disconnected or not respondingerror
B8FNo status message from HF receptacle 2HF receptacle 2 disconnected or not respondingerror
B9Checkstop interrupt invalid memory accessProgram crash due to memory access violationerror
B90No status message from HF receptacle 3HF receptacle 3 disconnected or not respondingerror
B91No status message from IES moduleIES module is disconnected from the systemwarning
B92IES module recognizedAn IES module is detectedinfo
B93Multifunctional footswitch recognizedA multifunctional footswitch is connectedinfo
B94Multifunctional footswitch has been disconnected from the systemA multifunctional footswitch is disconnectedinfo
B95New instrument recognized by systemAn instrument with instrument recognition is connectedinfo
B97Error in the program modification listChecksum of a stored setting is not correct, e.g. due to storage cell flippingerror
B98Error in the program listChecksum of a stored program is not correct, e.g. due to storage cell flippingerror
B99Activation type not availableUser assigns an activation signal that is not currently available, e.g. unconnected footswitchwarning
B9APlease check the clockVIO switched off for long time, supply current to real-time clock insufficientwarning
B9BMaster remote control recognizedA master remote control is recognized by the VIOinfo
B9CMaster remote control has been disconnected from the systemMaster remote control is disconnectedinfo
B9DRemote control recognizedA remote control is recognized by the VIOinfo
B9ERemote control has been disconnected from the systemRemote control is disconnectedinfo
B9FInstrument has been disconnected from the systemAn instrument with instrument recognition is disconnectedinfo
BANE module not availableNE (Nessy2) module missing or not logged on during monopolar activationcritical
BA0No other mode can be set for this instrumentUser wants to assign a different mode to an instrument with a fixed modeinfo
BA1No status message from HF moduleControl panel does not receive status message from HF module for longer than 1 seconderror
BA3Footswitch not assignedA footswitch which has not been assigned to an output socket is pressedwarning
BA4Two footswitches are connectedTwo footswitches of the same type are connectedwarning
BA6EEPROM is updatedUnit is programmed using a connected PCinfo
BA8Invalid ECB versionA module with an invalid ECB version logs onerror
BA9Activation not possible until settings are confirmedUser wants to perform activation after switching on unit without confirming settingswarning
BAANo valid mode is assignedOutput channel is activated without a mode being assigned to itwarning
BABActivation is only possible with a valid instrumentMF receptacle is activated at which no instrument is recognizedwarning
BACAutoStart is not possible if contact is made during assignmentAutoStart function is assigned while contact has already been recognizedwarning
BADError with version checkA module with an invalid version tries to log onerror
BAENo modules logged onNo modules log on during initialization phasecritical
BAFNE module ceases to respondNE module stops respondingerror
BBNE contact resistance out of rangeNeutral electrode contact resistance outside valid range on activationcritical
BB0NESSY symmetry warningNessy symmetry monitoring signals a value between 20% and 50%warning
BB1NESSY current density warningNessy current density monitoring signals a value above limit curvewarning
BB2IES footswitch detectedAn IES footswitch is connectedinfo
BB3Calibration of keyboardCapacitive keyboard is recalibratedinfo
BB4Rinsing of an APC instrumentRinsing of an APC instrument is initiatedinfo
BB5CAN warningError counter of CAN controller reaches critical valuewarning
BB6Incorrect receptacle version signalIncorrect receptacle version signal is receivedwarning
BB7Permissible limit value of AUTO START activation has been exceededAttempt is made to assign AutoStart activation a non-permissible power limitation valueinfo
BB8New interface module recognizedInterface module (e.g. VIO PORTAL) logs on to systeminfo
BB9Interface module has been disconnected from the systemInterface module (e.g. VIO PORTAL) is disconnectedinfo
BBAWarning from interface moduleInterface module (e.g. VIO PORTAL) signals an errorwarning
BBBReminder that a safety check is dueSafety check is dueinfo
BBCVEM 2 module recognizedReceptacle extension module VEM 2 logs into systeminfo
BBDVEM 2 module disconnectedReceptacle extension module VEM 2 is disconnectedinfo
BBEVEM 2 module receptacle 1 no longer ready for operationFirst receptacle of VEM 2 module no longer respondserror
BBFVEM 2 module receptacle 2 no longer ready for operationSecond receptacle of VEM 2 module no longer respondserror
BCFunction not availableRequired function cannot be implemented by available moduleswarning
BC0Neither footswitch nor AUTO START assignedInstrument without fingerswitch recognized and no footswitch or AUTO START assignedinfo
BC1EIP 2 was recognized by the systemEIP 2 connected to systeminfo
BC2EIP 2 was disconnected from the systemEIP 2 disconnected from systeminfo
BC3IES 2 footswitch was disconnected from the systemIES 2 footswitch disconnected from systeminfo
BC4Purge function not assignedPurge button on APC pressed without APC receptacle selectedwarning
BDI²C bus error on EEPROM interfaceSerial I²C write or read function fails on EEPROMerror
BEECB bus errorControl panel CPU detects error at CAN buserror
BFNo signal from capacitive keyboardControl panel CPU does not receive signal from capacitive keyboarderror
BFCPower DownUnit switched offinfo
BFDSystem ResetSystem reset performed, e.g. when switching on unitinfo
BFEPowerFailPowerFail signaled, e.g. when switching off unitinfo
BFFCAN MESSAGECAN message transmitted by control panel to be logged in protocolinfo
CStack error: stack pointer not at Top of Stack in main loopStack pointer validation failed during executioncritical
CStack errorStack pointer does not point to Top of Stack in main program looperror
C0Motor control faulty (voltage too high at ON)Measured voltage is too high during motor control operationerror
C1System error: CPU has not received necessary CAN message from control panel at proper timeCAN message not received by CPU at expected timeerror
C21Overdose: Output 20% too highHF output exceeds target by 20%critical
C26Measurement values of P_HF and P_NT do not agreeHF power output greater than NT power output or values divergeerror
C30Redundant measurement value for HF voltage too great during activationHF voltage redundancy check exceeds limit during activationerror
C31Measurement value for HF voltage too great with OFFHF voltage exceeds limit when unit is OFFerror
C32Measurement value for HF voltage too small with OFFHF voltage below limit when unit is OFFerror
C33Measurement value for HF voltage too great with ONHF voltage exceeds limit when unit is ONerror
C34Measurement value for HF voltage too small with ONHF voltage below limit when unit is ONerror
C35Measurement value for HF current too great with OFFHF current exceeds limit when unit is OFFerror
C36Measurement value for HF current too small with OFFHF current below limit when unit is OFFerror
C37Measurement value for HF current too great with ONHF current exceeds limit when unit is ONerror
C38Measurement value for HF current too small with ONHF current below limit when unit is ONerror
C4System error: Internal statusInternal CPU status errorerror
C41Measurement value for spark too great with OFFSpark measurement exceeds limit when unit is OFFerror
C42Measurement value for spark too small with OFFSpark measurement below limit when unit is OFFerror
C43Measurement value for spark too great with ONSpark measurement exceeds limit when unit is ONerror
C44Measurement value for spark too small with ONSpark measurement below limit when unit is ONerror
C45Measurement value for phase too great with OFFPhase measurement exceeds limit when unit is OFFerror
C46Measurement value for phase too small with OFFPhase measurement below limit when unit is OFFerror
C47Measurement value for phase too great with ONPhase measurement exceeds limit when unit is ONerror
C48Measurement value for phase too small with ONPhase measurement below limit when unit is ONerror
C5System error: CAN message with wrong lengthCAN message received with incorrect lengtherror
C51Measurement value for power supply unit voltage too great with OFFPower supply unit voltage exceeds limit when unit is OFFerror
C52Measurement value for power supply unit voltage too small with OFFPower supply unit voltage below limit when unit is OFFerror
C53Measurement value for power supply unit voltage too great with ONPower supply unit voltage exceeds limit when unit is ONerror
C54Measurement value for power supply unit voltage too small with ONPower supply unit voltage below limit when unit is ONerror
C55Measurement value for power supply unit current too great with OFFPower supply unit current exceeds limit when unit is OFFerror
C56Measurement value for power supply unit current too small with OFFPower supply unit current below limit when unit is OFFerror
C57Measurement value for power supply unit current too great with ONPower supply unit current exceeds limit when unit is ONerror
C58Measurement value for power supply unit current too small with ONPower supply unit current below limit when unit is ONerror
C5APower supply unit actuating voltage too greatPSU actuating voltage exceeds limiterror
C5BPower supply unit actuating voltage too smallPSU actuating voltage below limiterror
C6System error: Activation request without parameter messageActivation request received without required parameter messageerror
C61Operating voltage +15 V incorrectSupply voltage +15 V out of toleranceerror
C62Operating voltage –15 V incorrectSupply voltage –15 V out of toleranceerror
C63Operating voltage +24 V incorrectVoltage monitoring detected deviation from specificationerror
C64Operating voltage –5 V incorrectVoltage monitoring detected deviation from specificationerror
C70Relay configuration insufficientSelf-check of relay configuration during startup or operationerror
C71At least 1 relay signaled as closed although all should be openRelay self-check during operationerror
C72At least 1 relay signaled as open although all should be closedRelay self-check during operationerror
C75IIF signals invalid instrument numberIIF communication detected invalid instrument identification codeerror
C7DSystem error: CRC error in program memoryCRC check of program memory failed during self-testcritical
C7ESystem error: EEPROM not ready to readEEPROM access failed during self-checkcritical
C7FSystem error: MultitaskingMultitasking execution error detectedcritical
C81System error: CAN protocol length incorrectCAN bus message received with incorrect lengtherror
C82System error: SIO protocol length incorrectSerial I/O message received with incorrect lengtherror
C83System error: IIF has not responded at proper timeTimeout waiting for IIF communication responseerror
C84Short circuit between AE and NEShort circuit detection between active and neutral electrodeswarning
C91Redundancy error: Current at NESSY greater than at CPU + sensorsCurrent measurement mismatch between sensor moduleserror
C92Redundancy error: Power supply current available but no HF currentCurrent monitoring detects mismatch between supply and outputerror
CA0Temperature sensor at CPU + sensors defectiveTemperature sensor self-check failedwarning
CA1EEPROM calibration values incorrectCalibration data verification failedwarning
CC0Relay error in APC during self-checkRelay self-check failed in Active Power Control moduleerror
CDSystem error: CAN length of messageCAN message length error detectederror
CF0Trial activation with Pmax==0 (USA) -> no HFActivation attempted when maximum power setting is zeroinfo
DState error: state variable contains non-permissible valueState machine entered invalid statecritical
DState errorState variable contains a non-permissible valueerror
D80ECB faultCommunication error with IES 2 smoke evacuation control boarderror
D81Excess temperature (> 60 °C)Temperature sensor reading exceeds 60 °C in IES 2 unitwarning
D82Insufficient temperature (< 0 °C)Temperature sensor reading drops below 0 °C in IES 2 unitwarning
D83Software error (check total)Software checksum or integrity verification failedcritical
D86Pressure sensor errorPressure sensor self-test failederror
ECRC errorDefective program memory discovered by system check during operationerror
ECRC error: defective program memory during operationProgram memory CRC check failed during system operationcritical
FU Coag error: +12 V voltage too low during Coag activation+12 V supply drops below minimum during coagulation modeerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.