ERBE VIO 200 D: troubleshooting and error codes
The ServiceForge knowledge base holds 46 documented symptoms for the ERBE VIO 200 D, with 46 ranked causes, 46 diagnostic checks with test points and expected values, and 247 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
APC module timeout or activation signal failure
System fails to receive activation signal from APC module or signal transmission is interrupted
Ranked causes
- 1. CAN transmission interrupted or cable positioning issue (common)
Verify CAN transmission integrity
- Test point
- FS cable connection
- Expected
- Uninterrupted signal, no interference from HF cables
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APC protocol error in status message (Control panel)
Control panel receives incorrect status message from APC module with invalid CAN ID or format
Ranked causes
- 1. APC module protocol error or software version mismatch (occasional)
Verify APC status message format and CAN ID
- Test point
- APC module CAN message output
- Expected
- Status message has correct CAN ID and valid format per protocol specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APC setup parameters invalid
System detects invalid APC setup parameters stored in configuration
Ranked causes
- 1. Invalid sector error in CAN data (occasional)
Analyze CAN data for sector errors
- Test point
- CAN bus
- Expected
- Valid sector data in APC parameters
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APC status error (Control panel) patient safety
APC module reports status 'active' when it should be switched off; status signal conflict
Ranked causes
- 1. APC module malfunction or EMC interference (occasional)
Verify APC status signal accuracy
- Test point
- APC module status output on CAN bus
- Expected
- Status 'inactive' when APC is switched off; status 'active' only during activation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Button error on receptacle 1
Short circuit or interruption detected on receptacle board 1 button circuit
Ranked causes
- 1. Short circuit or interruption on receptacle board 1 connector (common)
Inspect receptacle 1 button circuit
- Test point
- Receptacle board 1 connector
- Expected
- No short circuits or interruptions; clean connector contact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Button error on receptacle 2
Short circuit or interruption detected on receptacle board 2 button circuit
Ranked causes
- 1. Short circuit or interruption on receptacle board 2 connector (common)
Inspect receptacle 2 button circuit
- Test point
- Receptacle board 2 connector
- Expected
- No short circuits or interruptions; clean connector contact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Caloric and differential pressure sensor mismatch caution
Caloric and differential pressure sensors disagree on gas type; indicates wrong gas supply
Ranked causes
- 1. Wrong gas supplied or sensor calibration/hardware fault (common)
Verify correct gas type at input
- Test point
- Gas input connector
- Expected
- Gas type matches system requirement (e.g., O2, N2O, CO2)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CAN bus error
CAN (Controller Area Network) bus detects length or data error in transmitted packet
Ranked causes
- 1. CAN data length or format error (occasional)
Verify CAN bus packet length
- Test point
- CAN bus interface
- Expected
- All CAN packets conform to length and format specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CAN protocol violation (length)
CAN bus detects protocol violation in packet length or format
Ranked causes
- 1. CAN packet format error or device malfunction (occasional)
Verify CAN protocol compliance
- Test point
- CAN bus interface
- Expected
- All CAN packets conform to protocol specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Checkstop interrupt invalid memory access (Control panel)
Control panel detects memory access violation; program crash due to software error or malfunction
Ranked causes
- 1. Software error or EMC interference (occasional)
Monitor for memory access errors
- Test point
- System console output
- Expected
- No checkstop interrupt messages; normal memory access
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Differential pressure sensor reading low
Differential pressure sensor measures less than half the expected setting; sensor appears defective
Ranked causes
- 1. Sensor calibration error or sensor hardware fault (occasional)
Measure differential pressure sensor output
- Test point
- Differential pressure sensor connector
- Expected
- Reading >= 50% of expected setting for current gas supply pressure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEPROM not ready to read
System cannot access EEPROM memory; read operation fails intermittently
Ranked causes
- 1. EEPROM access conflict or secondary fault condition (occasional)
Verify EEPROM access is available
- Test point
- EEPROM interface
- Expected
- EEPROM read operation completes successfully without timeout
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error in test mode or calibration
Error occurs during calibration or diagnosis procedure, typically software-related and often uncritical
Ranked causes
- 1. PC program error during calibration or diagnosis (occasional)
Repeat calibration or diagnostic test
- Test point
- PC diagnostic interface
- Expected
- Test completes without software error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow specification exceeded caution
Gas flow rate exceeds maximum target flow specification
Ranked causes
- 1. Calibration error or faulty flow sensor (occasional)
Measure actual gas flow rate
- Test point
- Gas outlet measurement point
- Expected
- Flow rate <= target specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow specification not attained
Gas flow rate does not reach target flow specification
Ranked causes
- 1. User error in flow rate setting (common)
Measure actual gas flow rate
- Test point
- Gas outlet measurement point
- Expected
- Flow rate >= target specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Function not available (Control panel)
Control panel detects that a required function cannot be implemented by the available modules
Ranked causes
- 1. Module software version mismatch or missing feature (occasional)
Verify module software version and enabled features
- Test point
- Module firmware version and feature configuration
- Expected
- All modules have compatible software version; required function is enabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas input pressure above maximum caution
Overpressure detected at selected gas input; gas supply pressure too high
Ranked causes
- 1. User error or faulty input pressure measuring circuit (common)
Verify gas input pressure at selected input
- Test point
- Gas input connector
- Expected
- Pressure <= maximum operating pressure per screen instructions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas input pressure below minimum caution
Underpressure detected at selected gas input; gas supply pressure too low
Ranked causes
- 1. User error or faulty input pressure measuring circuit (common)
Verify gas input pressure at selected input
- Test point
- Gas input connector
- Expected
- Pressure >= minimum operating pressure per screen instructions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas underdose from instrument caution
Gas delivery from instrument is insufficient; hose may be blocked or instrument defective
Ranked causes
- 1. Instrument defective or hose blocked (common)
Verify gas flow from instrument
- Test point
- Gas outlet hose of instrument
- Expected
- Adequate gas flow with no blockage or restriction
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
I²C bus error on EEPROM interface (Control panel) high voltage
Serial I²C communication error on control panel EEPROM interface; cannot read or write EEPROM data
Ranked causes
- 1. EEPROM or I²C bus component fault (occasional)
Verify I²C bus voltage and EEPROM communication
- Test point
- I²C SDA/SCL lines on control panel PCB
- Expected
- I²C bus voltage normal (3-5 V); EEPROM responds to read/write commands
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IIF module serial communication failure
Serial communication with Instrument Interface (IIF) module not possible; connection lost
Ranked causes
- 1. IIF module connection faulty or disconnected (common)
Verify IIF module serial communication
- Test point
- IIF module connector
- Expected
- Serial communication established and responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input 2 selected with partial complement
Input 2 selection error; received configuration does not match expected
Ranked causes
- 1. CAN sector error in input configuration data (occasional)
Verify Input 2 configuration in CAN data
- Test point
- CAN bus Input 2 configuration sector
- Expected
- Complete configuration data without sector errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal state incorrect
System internal state machine or data structure is in unexpected state
Ranked causes
- 1. Software error or data corruption (rare)
Verify system internal state via diagnostics
- Test point
- System diagnostic menu
- Expected
- System state consistent with current operational mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid EEPROM calibration data
EEPROM contains invalid or corrupted calibration data; cannot load calibration on startup
Ranked causes
- 1. EEPROM calibration data corrupted or not programmed (occasional)
Verify EEPROM calibration data integrity
- Test point
- EEPROM calibration sector
- Expected
- Valid calibration data present and checksum correct
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid instrument resistance number
Instrument resistance measurement reports invalid or undefined instrument number
Ranked causes
- 1. Instrument not properly registered or detected (common)
Verify instrument resistance measurement value
- Test point
- Instrument resistance measurement circuit
- Expected
- Measured resistance corresponds to valid instrument type per database
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low gas pressure at cylinder 1
Gas bottle 1 pressure is low; bottle needs replacement
Ranked causes
- 1. Gas bottle 1 depleted or low pressure (common)
Verify gas bottle 1 pressure
- Test point
- Cylinder 1 pressure gauge or sensor readout
- Expected
- Pressure >= minimum operating pressure (typically > 500 psi for full bottles)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low gas pressure at cylinder 2
Gas bottle 2 pressure is low; bottle needs replacement
Ranked causes
- 1. Gas bottle 2 depleted or low pressure (common)
Verify gas bottle 2 pressure
- Test point
- Cylinder 2 pressure gauge or sensor readout
- Expected
- Pressure >= minimum operating pressure (typically > 500 psi for full bottles)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement circuit overvoltage high voltage
Voltage on measurement circuit exceeds maximum operating threshold
Ranked causes
- 1. Power supply regulator fault or wiring damage (occasional)
Measure voltage on measurement circuit
- Test point
- Measurement circuit power supply node
- Expected
- Voltage <= maximum specified (typically 5.5 V for sensor circuits)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement circuit undervoltage high voltage
Voltage on measurement circuit drops below minimum operating threshold
Ranked causes
- 1. Power supply or wiring fault in measurement circuit (occasional)
Measure voltage on measurement circuit
- Test point
- Measurement circuit power supply node
- Expected
- Voltage >= minimum specified (typically 5 V for sensor circuits)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement circuit voltage error combined high voltage
Both undervoltage and overvoltage conditions detected simultaneously; reference voltage fault
Ranked causes
- 1. Reference voltage circuit fault or power supply failure (occasional)
Verify reference voltage stability
- Test point
- Reference voltage node on measurement circuit
- Expected
- Stable voltage output, no oscillation or dropout
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module timeout during activation (Control panel)
Control panel detects timeout; module failed to transmit status message via CAN within 110 ms timeout
Ranked causes
- 1. Faulty module or EMC interference on CAN bus (common)
Verify module CAN status message transmission
- Test point
- CAN bus interface for identified module
- Expected
- Module transmits valid status message every < 110 ms
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NE (Nessy2) module not available patient safety
NE module missing or not logged on when monopolar activation is initiated
Ranked causes
- 1. NE module failure or disconnection (common)
Verify NE module is logged on and available
- Test point
- NE (Nessy2) module interface
- Expected
- NE module logged on and ready for monopolar operations
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NE contact resistance out of range patient safety
NESSY message indicates NE (neutral electrode) contact resistance is not within valid range on activation
Ranked causes
- 1. Patient neutral electrode not properly applied or NE module faulty (common)
Measure NE contact resistance on activation
- Test point
- NE electrode contact resistance measurement
- Expected
- Resistance within valid operating range (typically < 50 ohms)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating system error
Operating system encounters fault during execution or after software update
Ranked causes
- 1. Software error during operation or after update; EMC interference possible (occasional)
Monitor system operation after power cycle
- Test point
- System console output
- Expected
- No operating system error message; normal system operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Program CRC test failure
Cyclic Redundancy Check (CRC) of program memory detects corruption or mismatch
Ranked causes
- 1. Program memory corruption or version mismatch (occasional)
Perform firmware integrity check
- Test point
- System program memory
- Expected
- Program CRC matches expected value; no memory corruption
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proportional valve control transistor fault
Control transistor for proportional valve is faulty or damaged
Ranked causes
- 1. Control transistor faulty (occasional)
Test proportional valve control transistor
- Test point
- Transistor junction on APC PCB
- Expected
- Transistor forward/reverse biased normally; no short or open junction
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proportional valve interruption
Open circuit detected in proportional valve coil circuit
Ranked causes
- 1. Proportional valve coil not inserted or faulty (common)
Verify proportional valve coil insertion and continuity
- Test point
- Proportional valve coil connector
- Expected
- Coil properly inserted with secure connection and continuity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proportional valve short circuit
Short circuit detected in proportional valve coil circuit
Ranked causes
- 1. Short circuit in proportional valve circuit (occasional)
Measure proportional valve coil resistance
- Test point
- Proportional valve coil connector
- Expected
- Normal coil resistance (no short), typically 10-50 ohms depending on design
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Receptacle 1 type detection mismatch
Type detection of receptacle 1 does not match stored configuration value
Ranked causes
- 1. Receptacle configuration mismatch or incorrect version list (common)
Verify receptacle 1 type against version list
- Test point
- Receptacle 1 physical slot and system configuration database
- Expected
- Type detection matches stored configuration value per version list
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Receptacle 2 type detection mismatch
Type detection of receptacle 2 does not match stored configuration value
Ranked causes
- 1. Receptacle configuration mismatch or incorrect version list (common)
Verify receptacle 2 type against version list
- Test point
- Receptacle 2 physical slot and system configuration database
- Expected
- Type detection matches stored configuration value per version list
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety relay IIF1 position fault patient safety
Relay on IIF module 1 or control on APC controller is faulty or misaligned
Ranked causes
- 1. Relay on IIF module 1 or APC controller faulty (common)
Verify safety relay IIF1 position
- Test point
- IIF module 1 relay contact
- Expected
- Relay in correct position with secure contact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety relay IIF2 addressing error patient safety
Safety relay IIF2 is incorrectly addressed via relay 1+i instead of direct addressing
Ranked causes
- 1. Relay on IIF module 2 or APC controller faulty (common)
Verify safety relay IIF2 addressing
- Test point
- IIF module 2 relay control
- Expected
- Direct relay addressing, not via relay 1+i
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIO to IIF protocol violation
Serial I/O to Instrument Interface communication protocol error detected
Ranked causes
- 1. IIF module communication fault or timeout (occasional)
Monitor SIO to IIF communication errors
- Test point
- IIF module serial interface
- Expected
- Communication completes without protocol violation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIO to IIF timeout
Serial I/O to Instrument Interface communication timeout; no response from IIF
Ranked causes
- 1. IIF module not responding or cable fault (occasional)
Verify SIO to IIF communication response time
- Test point
- IIF module serial interface
- Expected
- IIF responds to SIO query within timeout period (typically < 500 ms)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software watchdog triggered (Control panel)
Control panel CPU watchdog timer detects program crash or hang due to software error or malfunction
Ranked causes
- 1. Software error or EMC interference (occasional)
Monitor for software watchdog occurrences
- Test point
- System console output
- Expected
- No watchdog timeout messages; stable program execution
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve error on APC module
Input valve coil is not inserted or is faulty; no coil connection detected
Ranked causes
- 1. Input valve coil not inserted or faulty (common)
Verify input valve coil insertion and continuity
- Test point
- Input valve coil connector
- Expected
- Coil properly inserted with secure connection and continuity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Interrupt error: undefined IR vector invoked | Invalid interrupt vector accessed during operation | critical |
10 | Current sensor 1 defective | No signal provided by current sensor UE13 | error |
11 | Current sensor 2 defective | No signal provided by current sensor UE12 | error |
2 | CRC error: defective program memory discovered during self-check | Program memory CRC check failed after reset | critical |
2 | CRC error | Defective program memory discovered during self-check after reset | error |
3 | Undervoltage 5V | 5V voltage discovered to be too low during self-check after reset | error |
3 | Undervoltage 5 V: voltage too low during self-check | 5 V supply measured below acceptable range after reset | error |
4 | Overvoltage 5V | 5V voltage discovered to be too high during self-check after reset | error |
4 | Overvoltage 5 V: voltage too high during self-check | 5 V supply measured above acceptable range after reset | error |
5 | Undervoltage ±12 V: +12 V too low or –12 V too high | +12 V or –12 V supply out of range during self-check | error |
6 | Overvoltage ±12 V: +12 V too high or –12 V too low | +12 V or –12 V supply out of range during self-check | error |
6 | Overvoltage ±12V | +12V voltage discovered to be too high or -12V voltage too low during self-check after reset | error |
7 | Undercurrent activation recognition: measurement current too low | Activation recognition current below acceptable threshold | error |
7 | I-source defective | No signal provided by power source for NE contact resistance measurement | error |
70 | Relay error | Relay configuration or state error detected | error |
71 | Relay error | Relay state validation failed | error |
72 | Relay error | Relay state validation failed | error |
8 | Overcurrent activation recognition: measurement current too high | Activation recognition current above acceptable threshold | error |
8 | MOSI_error | SPI receive register contains $FF after Interrupt call-up | error |
80 | CRC Error | EIP 2 CRC monitoring procedure at unit startup discovers an error | error |
81 | Short circuit instrument recognition R: resistor value indicates short | Instrument recognition resistor measurement detects short circuit condition | error |
82 | Pump lid open at ON | Pump lid of EIP 2 not closed at start or opened during activation; Hall sensor circuit board not connected or faulty | warning |
82 | Short circuit activation recognition R_NEU: resistor indicates short | Activation recognition resistor measurement detects short circuit | error |
83 | No-load activation recognition R_NEU: resistor indicates open circuit | Activation recognition resistor measurement detects open circuit | error |
84 | Instrument recognition R-window violated: resistor cannot be assigned | Resistor value outside defined recognition window for instrument type | error |
84 | Time Out Error | Activation of EIP 2 maintained for more than 20 seconds | warning |
85 | Short circuit activation recognition R_ALT: resistor indicates short | Alternate activation recognition resistor detects short circuit | error |
86 | Activation error: activation not consistent with instrument type | Activation mode selected does not match recognized instrument type | error |
87 | Framing error: UART did not detect stop bit at end of byte | Serial receive failed to detect valid stop bit | error |
88 | UeWS Error | Motor control monitoring circuit is faulty | error |
88 | Parity error: check for even parity was unsuccessful | Serial data received with incorrect parity bit | error |
89 | Unknown identifier: interrupt routine detected non-agreed identifier | Received message has unrecognized identifier or format | error |
8A | DS2430 CRC error: received data record from instrument defective | CRC check failed on one-wire device data record | error |
8B | Non-permissible instrument recognition R_S3 | Non-permissible value determined for resistor-coded instrument recognition | error |
8C | Non-permissible sockets and relay status | Non-permissible value determined for socket recognition | error |
8D | Non-permissible instrument recognition R_S4 | Non-permissible value determined for resistor-coded instrument recognition | error |
8E | Non-permissible instrument recognition R_S8 | Non-permissible value determined for resistor-coded instrument recognition | error |
8F | UART timeout | UPR Receive_UART has not received anything 351us after call-up | error |
9 | Undercurrent instrument recognition: measurement current too low | Instrument recognition current below acceptable threshold | error |
90 | No current or no load | No measurable current input to motor during activation | error |
90 | DS2430_Write error | Attempt to write to EEPROM for instrument recognition unsuccessful | error |
A | Overcurrent instrument recognition: measurement current too high | Instrument recognition current above acceptable threshold | error |
A0 | Current too high at ON | Current input to motor is too high | error |
A1 | Timeout of activation signal | CAN transmission interrupted or FS cable improperly positioned relative to HF cable | error |
A10 | Gas input underpressure | Pressure sensor reads below minimum threshold at selected gas input | warning |
A11 | Gas input overpressure | Pressure sensor reads above maximum threshold at selected gas input | warning |
A12 | Gas type mismatch between sensors | Caloric and differential pressure sensors disagree on gas type | warning |
A13 | Differential pressure sensor reading low | Differential pressure sensor output less than 50% of expected setting | warning |
A2 | APC setup parameters invalid | Invalid sector error in CAN configuration data | error |
A21 | Measurement circuit undervoltage | Voltage on measurement circuit drops below minimum operating threshold | error |
A22 | Measurement circuit overvoltage | Voltage on measurement circuit exceeds maximum operating threshold | error |
A23 | Measurement circuit voltage error combined | Both undervoltage (A21) and overvoltage (A22) conditions detected simultaneously | error |
A3 | Safety relay IIF1 position fault | Relay on IIF module 1 or APC controller malfunction | critical |
A30 | Input 2 selected with partial complement | Input 2 configuration sector error in CAN data | warning |
A38 | Receptacle 1 type detection mismatch | Type detection does not match stored configuration value in system database | warning |
A39 | Receptacle 2 type detection mismatch | Type detection does not match stored configuration value in system database | warning |
A4 | Safety relay IIF2 addressing error | Relay on IIF module 2 or APC controller malfunction | critical |
A40 | Flow specification not attained | Gas flow rate does not reach target flow specification | warning |
A41 | Flow specification exceeded | Gas flow rate exceeds maximum target flow specification | warning |
A5 | Button error receptacle 1 | Short circuit or interruption on receptacle board 1 connector | error |
A6 | Button error receptacle 2 | Short circuit or interruption on receptacle board 2 connector | error |
A7 | Input valve error | Input valve coil not inserted or faulty; no continuity detected | warning |
A7D | Program CRC test failure | Cyclic Redundancy Check detects program memory corruption or mismatch | error |
A7E | EEPROM not ready to read | System cannot access EEPROM memory; read operation fails | warning |
A7F | Operating system error | Operating system fault during execution or after software update | error |
A8 | Proportional valve interruption | Open circuit in proportional valve coil circuit | warning |
A80 | Internal state incorrect | System internal state machine in unexpected state | warning |
A81 | CAN protocol violation | CAN bus detects packet length or format violation | error |
A82 | SIO to IIF protocol violation | Serial I/O to IIF communication protocol error | warning |
A83 | SIO to IIF timeout | Serial I/O to IIF communication timeout; no response from IIF | warning |
A85 | Invalid instrument resistance number | Instrument resistance measurement reports undefined instrument number | warning |
A86 | Gas underdose from instrument | Gas delivery from instrument insufficient; underdose detected | warning |
A9 | Proportional valve short circuit | Short circuit detected in proportional valve coil circuit | warning |
A90 | Low pressure at cylinder 1 | Gas bottle 1 pressure drops below minimum threshold | warning |
A91 | Low pressure at cylinder 2 | Gas bottle 2 pressure drops below minimum threshold | warning |
AA | Proportional valve control transistor fault | Control transistor malfunction or component failure | warning |
AA0 | Invalid EEPROM calibration data | EEPROM contains invalid or corrupted calibration data on startup | error |
AB | Error in test mode or calibration | Software error during calibration or diagnostic test execution | warning |
AC | IIF module serial communication failure | Serial communication with IIF module lost or unavailable | error |
AD | CAN bus error | CAN packet length or data format violation detected | error |
B1 | Software watchdog triggered | Program crash due to software error or malfunction | error |
B10 | Please terminate activation | Activation automatically terminated but activation signal remains longer than 5 s | warning |
B11 | Error during activation | Activation not enabled for over 110 ms during activation | error |
B12 | Activation signal during switch-on | Activation signal present during initialization (e.g. footswitch pedal pressed) | warning |
B13 | Errors in list management | Error identified in flash file during event or error list management | error |
B14 | CRC error | Continuous CRC monitoring detects program error | critical |
B15 | FLASH RAM verification error | Continuous monitoring detects error in DRAM memory | error |
B16 | Insufficient EEPROM memory | Memory for user programs is full | warning |
B17 | Double activation | Two activation signals present simultaneously within 100 ms | warning |
B18 | No deactivation signal | HF or APC module fails to react to deactivation request for longer than 110 ms | error |
B19 | PowerFail | PowerFail signal received but power failure does not occur within 2 s | error |
B1A | Error in parameter memory | Checksum monitoring of program detects error | error |
B1B | CRC check not yet completed | Unit operated before CRC check completed after software download | warning |
B1C | Maximum ON time exceeded | Activation lasts longer than ON time selected in SETUP | warning |
B1D | Incorrect instrument parameters | Instrument recognized via instrument recognition but parameter outside tolerances | error |
B1E | Keyboard error | Button pressed recognized during initialization | warning |
B1F | NESSY symmetry monitoring | Error signaled by Nessy symmetry monitoring during activation | warning |
B2 | Module timeout during activation | Module fails to transmit CAN status message within 110 ms timeout | error |
B21 | Incorrect or missing bitmap file | Incompatible bitmap file found | error |
B22 | Tissue still touched after activation terminated | Activation automatically terminated (e.g. by AutoStop) but tissue still in contact | warning |
B23 | Configuration error of HF CPU | HF CPU transmits invalid status message (0x27) | error |
B24 | Incorrect status message from IES 2 | Invalid status message received from IES 2 module | error |
B25 | Modified receptacle configuration detected | Receptacle configuration changed (e.g. replacement or retrofitting) | error |
B5 | APC status error | APC module signals 'active' status when it should be off | critical |
B6 | APC protocol error in status message | Invalid status message received from APC module with incorrect CAN ID | error |
B81 | Indication of incorrect CAN protocol | Non-ECB-specified CAN message received | error |
B82 | Indication of receptacle arrangement change | Changes in receptacle assignment signaled during operation | warning |
B84 | Dual-pedal footswitch recognized | Dual-pedal footswitch connected to system | info |
B85 | Dual-pedal footswitch disconnected | Dual-pedal footswitch disconnected from system | info |
B86 | APC receptacle 1 recognized | APC receptacle 1 detected and connected | info |
B87 | No status message from APC receptacle 1 | APC receptacle 1 disconnected or not responding | error |
B88 | Single-pedal footswitch recognized | Single-pedal footswitch connected to system | info |
B89 | Single-pedal footswitch disconnected | Single-pedal footswitch disconnected from system | info |
B8A | APC receptacle 2 recognized | APC receptacle 2 detected and connected | info |
B8B | No status message from APC receptacle 2 | APC receptacle 2 disconnected or not responding | error |
B8C | APC module recognized | APC module detected and connected | info |
B8D | No status message from APC module | APC module disconnected or not responding | error |
B8E | No status message from HF receptacle 1 | HF receptacle 1 disconnected or not responding | error |
B8F | No status message from HF receptacle 2 | HF receptacle 2 disconnected or not responding | error |
B9 | Checkstop interrupt invalid memory access | Program crash due to memory access violation | error |
B90 | No status message from HF receptacle 3 | HF receptacle 3 disconnected or not responding | error |
B91 | No status message from IES module | IES module is disconnected from the system | warning |
B92 | IES module recognized | An IES module is detected | info |
B93 | Multifunctional footswitch recognized | A multifunctional footswitch is connected | info |
B94 | Multifunctional footswitch has been disconnected from the system | A multifunctional footswitch is disconnected | info |
B95 | New instrument recognized by system | An instrument with instrument recognition is connected | info |
B97 | Error in the program modification list | Checksum of a stored setting is not correct, e.g. due to storage cell flipping | error |
B98 | Error in the program list | Checksum of a stored program is not correct, e.g. due to storage cell flipping | error |
B99 | Activation type not available | User assigns an activation signal that is not currently available, e.g. unconnected footswitch | warning |
B9A | Please check the clock | VIO switched off for long time, supply current to real-time clock insufficient | warning |
B9B | Master remote control recognized | A master remote control is recognized by the VIO | info |
B9C | Master remote control has been disconnected from the system | Master remote control is disconnected | info |
B9D | Remote control recognized | A remote control is recognized by the VIO | info |
B9E | Remote control has been disconnected from the system | Remote control is disconnected | info |
B9F | Instrument has been disconnected from the system | An instrument with instrument recognition is disconnected | info |
BA | NE module not available | NE (Nessy2) module missing or not logged on during monopolar activation | critical |
BA0 | No other mode can be set for this instrument | User wants to assign a different mode to an instrument with a fixed mode | info |
BA1 | No status message from HF module | Control panel does not receive status message from HF module for longer than 1 second | error |
BA3 | Footswitch not assigned | A footswitch which has not been assigned to an output socket is pressed | warning |
BA4 | Two footswitches are connected | Two footswitches of the same type are connected | warning |
BA6 | EEPROM is updated | Unit is programmed using a connected PC | info |
BA8 | Invalid ECB version | A module with an invalid ECB version logs on | error |
BA9 | Activation not possible until settings are confirmed | User wants to perform activation after switching on unit without confirming settings | warning |
BAA | No valid mode is assigned | Output channel is activated without a mode being assigned to it | warning |
BAB | Activation is only possible with a valid instrument | MF receptacle is activated at which no instrument is recognized | warning |
BAC | AutoStart is not possible if contact is made during assignment | AutoStart function is assigned while contact has already been recognized | warning |
BAD | Error with version check | A module with an invalid version tries to log on | error |
BAE | No modules logged on | No modules log on during initialization phase | critical |
BAF | NE module ceases to respond | NE module stops responding | error |
BB | NE contact resistance out of range | Neutral electrode contact resistance outside valid range on activation | critical |
BB0 | NESSY symmetry warning | Nessy symmetry monitoring signals a value between 20% and 50% | warning |
BB1 | NESSY current density warning | Nessy current density monitoring signals a value above limit curve | warning |
BB2 | IES footswitch detected | An IES footswitch is connected | info |
BB3 | Calibration of keyboard | Capacitive keyboard is recalibrated | info |
BB4 | Rinsing of an APC instrument | Rinsing of an APC instrument is initiated | info |
BB5 | CAN warning | Error counter of CAN controller reaches critical value | warning |
BB6 | Incorrect receptacle version signal | Incorrect receptacle version signal is received | warning |
BB7 | Permissible limit value of AUTO START activation has been exceeded | Attempt is made to assign AutoStart activation a non-permissible power limitation value | info |
BB8 | New interface module recognized | Interface module (e.g. VIO PORTAL) logs on to system | info |
BB9 | Interface module has been disconnected from the system | Interface module (e.g. VIO PORTAL) is disconnected | info |
BBA | Warning from interface module | Interface module (e.g. VIO PORTAL) signals an error | warning |
BBB | Reminder that a safety check is due | Safety check is due | info |
BBC | VEM 2 module recognized | Receptacle extension module VEM 2 logs into system | info |
BBD | VEM 2 module disconnected | Receptacle extension module VEM 2 is disconnected | info |
BBE | VEM 2 module receptacle 1 no longer ready for operation | First receptacle of VEM 2 module no longer responds | error |
BBF | VEM 2 module receptacle 2 no longer ready for operation | Second receptacle of VEM 2 module no longer responds | error |
BC | Function not available | Required function cannot be implemented by available modules | warning |
BC0 | Neither footswitch nor AUTO START assigned | Instrument without fingerswitch recognized and no footswitch or AUTO START assigned | info |
BC1 | EIP 2 was recognized by the system | EIP 2 connected to system | info |
BC2 | EIP 2 was disconnected from the system | EIP 2 disconnected from system | info |
BC3 | IES 2 footswitch was disconnected from the system | IES 2 footswitch disconnected from system | info |
BC4 | Purge function not assigned | Purge button on APC pressed without APC receptacle selected | warning |
BD | I²C bus error on EEPROM interface | Serial I²C write or read function fails on EEPROM | error |
BE | ECB bus error | Control panel CPU detects error at CAN bus | error |
BF | No signal from capacitive keyboard | Control panel CPU does not receive signal from capacitive keyboard | error |
BFC | Power Down | Unit switched off | info |
BFD | System Reset | System reset performed, e.g. when switching on unit | info |
BFE | PowerFail | PowerFail signaled, e.g. when switching off unit | info |
BFF | CAN MESSAGE | CAN message transmitted by control panel to be logged in protocol | info |
C | Stack error: stack pointer not at Top of Stack in main loop | Stack pointer validation failed during execution | critical |
C | Stack error | Stack pointer does not point to Top of Stack in main program loop | error |
C0 | Motor control faulty (voltage too high at ON) | Measured voltage is too high during motor control operation | error |
C1 | System error: CPU has not received necessary CAN message from control panel at proper time | CAN message not received by CPU at expected time | error |
C21 | Overdose: Output 20% too high | HF output exceeds target by 20% | critical |
C26 | Measurement values of P_HF and P_NT do not agree | HF power output greater than NT power output or values diverge | error |
C30 | Redundant measurement value for HF voltage too great during activation | HF voltage redundancy check exceeds limit during activation | error |
C31 | Measurement value for HF voltage too great with OFF | HF voltage exceeds limit when unit is OFF | error |
C32 | Measurement value for HF voltage too small with OFF | HF voltage below limit when unit is OFF | error |
C33 | Measurement value for HF voltage too great with ON | HF voltage exceeds limit when unit is ON | error |
C34 | Measurement value for HF voltage too small with ON | HF voltage below limit when unit is ON | error |
C35 | Measurement value for HF current too great with OFF | HF current exceeds limit when unit is OFF | error |
C36 | Measurement value for HF current too small with OFF | HF current below limit when unit is OFF | error |
C37 | Measurement value for HF current too great with ON | HF current exceeds limit when unit is ON | error |
C38 | Measurement value for HF current too small with ON | HF current below limit when unit is ON | error |
C4 | System error: Internal status | Internal CPU status error | error |
C41 | Measurement value for spark too great with OFF | Spark measurement exceeds limit when unit is OFF | error |
C42 | Measurement value for spark too small with OFF | Spark measurement below limit when unit is OFF | error |
C43 | Measurement value for spark too great with ON | Spark measurement exceeds limit when unit is ON | error |
C44 | Measurement value for spark too small with ON | Spark measurement below limit when unit is ON | error |
C45 | Measurement value for phase too great with OFF | Phase measurement exceeds limit when unit is OFF | error |
C46 | Measurement value for phase too small with OFF | Phase measurement below limit when unit is OFF | error |
C47 | Measurement value for phase too great with ON | Phase measurement exceeds limit when unit is ON | error |
C48 | Measurement value for phase too small with ON | Phase measurement below limit when unit is ON | error |
C5 | System error: CAN message with wrong length | CAN message received with incorrect length | error |
C51 | Measurement value for power supply unit voltage too great with OFF | Power supply unit voltage exceeds limit when unit is OFF | error |
C52 | Measurement value for power supply unit voltage too small with OFF | Power supply unit voltage below limit when unit is OFF | error |
C53 | Measurement value for power supply unit voltage too great with ON | Power supply unit voltage exceeds limit when unit is ON | error |
C54 | Measurement value for power supply unit voltage too small with ON | Power supply unit voltage below limit when unit is ON | error |
C55 | Measurement value for power supply unit current too great with OFF | Power supply unit current exceeds limit when unit is OFF | error |
C56 | Measurement value for power supply unit current too small with OFF | Power supply unit current below limit when unit is OFF | error |
C57 | Measurement value for power supply unit current too great with ON | Power supply unit current exceeds limit when unit is ON | error |
C58 | Measurement value for power supply unit current too small with ON | Power supply unit current below limit when unit is ON | error |
C5A | Power supply unit actuating voltage too great | PSU actuating voltage exceeds limit | error |
C5B | Power supply unit actuating voltage too small | PSU actuating voltage below limit | error |
C6 | System error: Activation request without parameter message | Activation request received without required parameter message | error |
C61 | Operating voltage +15 V incorrect | Supply voltage +15 V out of tolerance | error |
C62 | Operating voltage –15 V incorrect | Supply voltage –15 V out of tolerance | error |
C63 | Operating voltage +24 V incorrect | Voltage monitoring detected deviation from specification | error |
C64 | Operating voltage –5 V incorrect | Voltage monitoring detected deviation from specification | error |
C70 | Relay configuration insufficient | Self-check of relay configuration during startup or operation | error |
C71 | At least 1 relay signaled as closed although all should be open | Relay self-check during operation | error |
C72 | At least 1 relay signaled as open although all should be closed | Relay self-check during operation | error |
C75 | IIF signals invalid instrument number | IIF communication detected invalid instrument identification code | error |
C7D | System error: CRC error in program memory | CRC check of program memory failed during self-test | critical |
C7E | System error: EEPROM not ready to read | EEPROM access failed during self-check | critical |
C7F | System error: Multitasking | Multitasking execution error detected | critical |
C81 | System error: CAN protocol length incorrect | CAN bus message received with incorrect length | error |
C82 | System error: SIO protocol length incorrect | Serial I/O message received with incorrect length | error |
C83 | System error: IIF has not responded at proper time | Timeout waiting for IIF communication response | error |
C84 | Short circuit between AE and NE | Short circuit detection between active and neutral electrodes | warning |
C91 | Redundancy error: Current at NESSY greater than at CPU + sensors | Current measurement mismatch between sensor modules | error |
C92 | Redundancy error: Power supply current available but no HF current | Current monitoring detects mismatch between supply and output | error |
CA0 | Temperature sensor at CPU + sensors defective | Temperature sensor self-check failed | warning |
CA1 | EEPROM calibration values incorrect | Calibration data verification failed | warning |
CC0 | Relay error in APC during self-check | Relay self-check failed in Active Power Control module | error |
CD | System error: CAN length of message | CAN message length error detected | error |
CF0 | Trial activation with Pmax==0 (USA) -> no HF | Activation attempted when maximum power setting is zero | info |
D | State error: state variable contains non-permissible value | State machine entered invalid state | critical |
D | State error | State variable contains a non-permissible value | error |
D80 | ECB fault | Communication error with IES 2 smoke evacuation control board | error |
D81 | Excess temperature (> 60 °C) | Temperature sensor reading exceeds 60 °C in IES 2 unit | warning |
D82 | Insufficient temperature (< 0 °C) | Temperature sensor reading drops below 0 °C in IES 2 unit | warning |
D83 | Software error (check total) | Software checksum or integrity verification failed | critical |
D86 | Pressure sensor error | Pressure sensor self-test failed | error |
E | CRC error | Defective program memory discovered by system check during operation | error |
E | CRC error: defective program memory during operation | Program memory CRC check failed during system operation | critical |
F | U Coag error: +12 V voltage too low during Coag activation | +12 V supply drops below minimum during coagulation mode | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.