ERBE ICC 200: troubleshooting and error codes
The ServiceForge knowledge base holds 20 documented symptoms for the ERBE ICC 200, with 22 ranked causes, 56 diagnostic checks with test points and expected values, and 62 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: HF generator output parameters high voltage
Adjustment of HF output voltage and current limitation on ICC electrosurgical generator to ensure correct power output and phase angle measurement.
Ranked causes
- 1. HF output voltage requires adjustment via TP3 (common)
- 2. HF current limitation requires adjustment via TP4 (common)
Measure HF effective voltage in AUTO CUT field
- Test point
- AUTO CUT display, TP3 adjustment point
- Expected
- 223 volts HF effective voltage V, 220…223 V
- Tools
- APM 600 HF power meter set to 500 ohms
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: HF generator output parameters amplified high voltage
Fine adjustment of amplified HF voltage and current measurement circuits for low-power output measurement (5 watts).
Ranked causes
- 1. Amplified HF voltage measurement requires adjustment via TP5 (common)
- 2. Amplified HF current measurement requires adjustment via TP6 (common)
Verify 5 watts output at HF power meter before adjustment
- Test point
- APM 600 HF power meter display
- Expected
- 5 watts output W, 4.5…5.5 W
- Tools
- APM 600 HF power meter at 500 ohms
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Activation detection (MIC board) high voltage
Setting activation detection using trimming potentiometer TP2 in program 12 MIC
Ranked causes
- 1. Activation detection requires initial adjustment after instrument detection calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AUTO START delay time
Setting delay time between tissue contact detection and HF generator activation in AUTO START function
Ranked causes
- 1. AUTO START delay time requires adjustment for specific operation modes or per user program (common)
Verify AUTO START 2 delay time setting (ICC 350)
- Test point
- Test program 23 display
- Expected
- 2 ≤ t ≤ 10 seconds in 0.5 second steps
- Tools
- Key 14, Up/Down keys
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bipolar CUT function (MIEN probe) high voltage
Testing bipolar CUT function activation using MIEN probe with 15W Effect 3 setting
Ranked causes
- 1. MIEN bipolar CUT function requires calibration and verification (common)
Measure bipolar CUT output power
- Test point
- MIC socket NE1 CUT, Effect 3 setting
- Expected
- 15 ± 3 W, ± 3 W
- Tools
- APM, 500 Ohm load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bipolar SOFT coagulation (MIEN probe) high voltage
Testing bipolar SOFT coagulation with no-load voltage measurement at 15W setting
Ranked causes
- 1. MIEN bipolar SOFT coagulation requires verification after CUT testing (common)
Measure bipolar SOFT output power
- Test point
- MIC socket NE1 CUT, SOFT setting
- Expected
- 7 ± 2 W, ± 2 W
- Tools
- APM, 500 Ohm load
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CUT function (MIC program) high voltage
Testing CUT cutting function activation on MIC instrument with footswitch and power measurement
Ranked causes
- 1. CUT function requires testing after instrument and activation detection adjustment (common)
Measure CUT function output power
- Test point
- MIC socket NE1 and CUT pin
- Expected
- 40 ± 15 % W, ± 15 %
- Tools
- APM (Automatic Power Meter), 500 Ohm load
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display brightness (7-segment)
Adjusting brightness intensity of 7-segment display indicators in 10 stages via test program 17
Ranked causes
- 1. Display brightness requires adjustment for optimal visibility in different room lighting conditions (common)
Observe 7-segment display intensity at current setting
- Test point
- 4-line display showing AUTO CUT/AUTO COAG 1/AUTO COAG 2/AUTO BIPOLAR
- Expected
- Intensity from 1–10, adjustable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Instrument detection (MIC board) high voltage
Setting instrument detection resistor value to proper range using trimming potentiometer TP1 in program 12 MIC
Ranked causes
- 1. Instrument detection requires calibration during setup or after component replacement (common)
Measure instrument detection I value with 51 Ohm resistor connected
- Test point
- Measurement display in AUTO CUT/AUTO COAG 1 view
- Expected
- 52, at set value
- Tools
- Program 12 MIC display
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monopolar coagulation FORCED function high voltage
Testing FORCED coagulation activation at specified power setting with high load resistance
Ranked causes
- 1. FORCED coagulation function requires verification after SOFT testing (common)
Measure FORCED coagulation output power
- Test point
- MIC socket NE1
- Expected
- 60 ± 15 % W, ± 15 %
- Tools
- APM, 350 Ohm load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monopolar coagulation SOFT function high voltage
Testing SOFT coagulation activation with blue footswitch at specified power setting
Ranked causes
- 1. SOFT coagulation function requires verification after CUT testing (common)
Measure SOFT coagulation output power
- Test point
- MIC socket NE1
- Expected
- 120 ± 15 % W, ± 15 %
- Tools
- APM, 75 Ohm load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Neurotest board feedback frequency high voltage
Set feedback frequency of output current measurement via trimming potentiometer TP1 on Neurotest board EE 30128-070. Requires 1 kOhm 0.5W 1% test resistor at output.
Ranked causes
- 1. Neurotest board EE 30128-070 requires feedback frequency adjustment (common)
Measure feedback frequency of output current at MP7
- Test point
- MP7 (reference ground MP6)
- Expected
- 73 Hz at SPH intensity 3, +/-10 Hz
- Tools
- Frequency meter
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Neurotest constant-current circuit output high voltage
Performance test of constant-current output circuit. Connect 1 kOhm test resistor and 50 kOhm potentiometer in series at Neurotest output. Measure square-wave voltage on 1 kOhm resistor (1 mA = 1 V).
Ranked causes
- 1. Neurotest constant-current source requires output current verification (common)
Measure square-wave peak output current at SPH 1
- Test point
- 1 kOhm test resistor
- Expected
- 5 mA, 3.75–6.25 mA (±25%)
- Tools
- Oscilloscope, DMM
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Neurotest electronic fuse protection high voltage
Performance test of electronic fuse protection. Excess dose triggers fuse via transistor T3. Test by activating Neurotest and briefly connecting MP9 to MP2. Fuse should respond without blowing microfuse.
Ranked causes
- 1. Electronic fuse protection circuit requires functional verification (common)
Activate Neurotest and briefly connect MP9 to MP2
- Test point
- MP9, MP2
- Expected
- Electronic fuse trips; error code appears
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Neurotest LED activation control
Test HF and NT LED activation. HF LED lights only at intensity 0. At intensities 1–5, NT LED controlled by CPU; HF LED must not light. On TUR version, HF inhibited at 1–5. On ZMK version, HF still possible.
Ranked causes
- 1. LED activation control requires verification per device version (common)
Set intensity to 0 and verify HF LED on
- Test point
- HF LED on remote control
- Expected
- HF LED illuminated at intensity 0
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Neurotest output error monitoring high voltage
Performance test of output error monitoring function. Connect 1 kOhm 0.5W 1% test resistor at Neurotest output, set SPH 1, activate unit and verify error message on circuit interruption.
Ranked causes
- 1. Output error monitoring circuit requires verification (common)
Set intensity to SPH 1 and activate Neurotest
- Test point
- Neurotest output
- Expected
- Unit activated; output current flowing
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Phase angle (cos Phi) high voltage
Adjustment of phase angle measurement between HF voltage and current using capacitive test load to verify analog/digital converter accuracy.
Ranked causes
- 1. Phase angle measurement via TP7 requires calibration (common)
Connect APM 600 to ICC output with 1 nF capacitive test load
- Test point
- Active electrode (CUT/COAG 2) and NE sockets
- Expected
- Capacitor creates 45° phase shift at 340 kHz frequency
- Tools
- APM 600, ERBE original cable, 1 nF capacitor EE 51103-026 with short lines
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Protective circuit voltage limitation (MIEN) high voltage
Testing protective circuit voltage limiting function at different HF output voltage settings on MIEN program
Ranked causes
- 1. Protective circuit voltage limiting requires verification to ensure proper fuse operation and diode protection (common)
Measure highest HF peak voltage at Effect 3 (15W) setting
- Test point
- MIC output socket NE1 and CUT with oscilloscope
- Expected
- 340 Vp max without limitation
- Tools
- Oscilloscope
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote control feedback frequency for Neurotest
Set feedback frequency of remote control intensity setting via trimming potentiometer TP1 on board EE 30121-442. Measure frequency at Pin 1 of AD654 IC1. Setting at intensity 5: 38.0 Hz (–0.5, +0.0).
Ranked causes
- 1. Remote control intensity frequency requires adjustment (common)
Measure frequency at Pin 1 AD654 IC1 for intensity 1
- Test point
- Pin 1 AD654 IC1 (reference ground MP6)
- Expected
- 7 Hz, 5–10 Hz
- Tools
- Frequency meter
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote control LED display for Neurotest
Test HF LED and NT LED activation on remote control board EE 30121-442. HF LED at intensity 0 via rotary switch. NT LED at intensities 1–5 via CPU; lights only when feedback frequency in valid range.
Ranked causes
- 1. Remote control LED display requires functional verification (common)
Set intensity 0 and verify HF LED on
- Test point
- HF LED on remote control
- Expected
- HF LED illuminated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | No HF voltage present at the HF voltage sensor | Error in the HF generator or short circuit in accessories | error |
10 | Error during activation of AUTO CUT, no valid front panel setting | Invalid control panel setting | error |
11 | Error during activation of AUTO COAG 1 or AUTO COAG 2, no valid front panel setting | Invalid control panel setting | error |
12 | Error during activation of AUTO BIPOLAR, no valid front panel setting | Invalid control panel setting | error |
13 | Error during activation of AUTO CUT; NE impedance out of range | Neutral electrode impedance measurement error | error |
17 | HF leakage current > 300 mA during activation | Excessive HF leakage current detected | error |
18 | HF leakage current between 150 and 300 mA during activation | Elevated HF leakage current detected | warning |
19 | NESSY current density error during AUTO CUT or AUTO COAG activation | Current density monitoring failure | error |
2 | HF output voltage too high | Error in the HF generator | error |
20 | NESSY current symmetry error; unequal current in NE joint phases | Current symmetry monitoring failure | error |
21 | Current symmetry error | Asymmetric current detected | error |
22 | AUTO CUT switched on when switching on power | Fingerswitch or footswitch activated during power-on | error |
23 | AUTO COAG 1 switched on when switching on power | Fingerswitch or footswitch activated during power-on | error |
24 | AUTO COAG 2 switched on when switching on power | Footswitch activated during power-on | error |
25 | AUTO BIPOLAR switched on by footswitch when switching on power | Footswitch activated during power-on | error |
26 | Contact detected when switching on power; contact monitor is active | Short circuit detected during power-on | error |
27 | Overheating; internal unit temperature > 100°C | Excessive internal temperature | warning |
28 | Activation signal for blue footswitch button already on | Blue footswitch button detected as pressed during activation | error |
29 | Double activation detected (Spanish version) | Further activation signal detected during activation | error |
3 | Switched-mode power supply unit supplies no voltage during activation | Error in the switched-mode power supply unit | error |
30 | Mismatch; load resistance < 50 ohms for > 5 seconds | Low load resistance detected during operation | error |
31 | Mismatch; output power > 100 watts and load resistance remains < 300 ohms | Excessive output power with low load resistance | error |
32 | Multiple activation detected | Simultaneous activation from multiple sources | error |
33 | New activation signal immediately appears at end of activation | Immediate re-activation detected | error |
34 | Multiple activation detected during AUTO BIPOLAR | Simultaneous activation during AUTO BIPOLAR | error |
35 | New activation signal during AUTO BIPOLAR with AUTO STOP or AUTO START | Simultaneous activation detected during AUTO BIPOLAR mode | error |
36 | AUTO START signal already present when pressing BIPOLAR AUTO START key | Duplicate AUTO START signal | error |
37 | Power supply unit voltage has not dissipated at activation time | Power supply discharge failure | error |
38 | Contact resistance < 6 ohms during AUTO START activation | Low contact resistance detected | error |
4 | Switched-mode power supply unit supplies too high a voltage upon ST generator activation | Error in the switched-mode power supply unit | error |
40 | ST generator time control output error; actual value > set value plus 20% | Time control monitoring deviation detected | error |
41 | ST generator time control output error; actual value < set value minus 20% | Time control monitoring deviation detected | error |
42 | ST generator time control output error; actual value > set value minus 40% | Time control monitoring deviation detected | error |
43 | ST generator time control monitor error; no frequency feedback | Frequency feedback loss | error |
44 | ST generator output voltage monitor error | Output voltage monitoring failure | error |
45 | HF generator error during self-check; HF output voltage too low | Self-check failure | error |
46 | Self-check error within permissible limits; power supply versus HF voltage mismatch | Self-check comparison failure | error |
47 | HF generator self-check error outside permissible limits | Self-check comparison failure | error |
48 | No signal from output current monitor | Current monitoring failure | error |
5 | LF leakage current > 50 µA flowing into unit via neutral electrode | Patient positioning issue or defective equipment connected to patient | warning |
50-70 | Key on front panel is depressed during power start-up phase | Front panel key activation during startup | error |
6 | Safety shutoff feature output reports ON during activation phase | Malfunction during activation | error |
7 | Safety shutoff feature output reports OFF during activation phase | Malfunction during activation | error |
71 | Interface error for ICC 350 DOKU; start/stop bit or number of bits incorrect | Serial communication error | error |
72 | Interface error for ICC 350 DOKU; signal in receiving register not read out | Data reception failure | error |
73 | Interface error for ICC 350 DOKU; parity sign placed incorrectly | Parity check failure | error |
74 | Interface error for ICC 350 DOKU; no connection to PC available | PC connection loss | error |
75 | Interface error for ICC 350 DOKU; sign stays in transmitter buffer; SIO module not transmitting | Data transmission failure | error |
76 | Interface error for ICC 350 DOKU; storage takes too long; hard disk too slow | Data storage latency | error |
80 | ICC 350 NEUROTEST; no valid frequency for 5-stage intensity setting | Frequency configuration error | error |
81 | ICC 350 NEUROTEST; intensity setting set to 0 during activation | Invalid intensity configuration | error |
82 | ICC 350 NEUROTEST; NESSY measurement NE impedance > 120 ohms | Neutral electrode impedance measurement failure | error |
83 | ICC 350 NEUROTEST; output current too high but within permissible limits | Current level monitoring | warning |
84 | ICC 350 NEUROTEST; output current too high outside permissible limits | Excessive output current | error |
85 | ICC 350 NEUROTEST; no output current; circuit interruption or frequency feedback failure | Output current loss | error |
86 | ICC 350 NEUROTEST; output current too low but within permissible limits | Insufficient output current | warning |
87 | ICC 350 NEUROTEST; output current too low below permissible limit | Excessive current reduction | error |
88 | ICC 350 NEUROTEST; maximum time limit has run out | Continuous operation exceeds time limit | warning |
89 | ICC 350 NEUROTEST; activation signal already present when switching on power | Premature activation signal on startup | error |
9 | Maximum continuous time limit exceeded | Unit activated for too long continuously | warning |
90 | ICC 350 NEUROTEST; electronic fuse triggered or frequency feedback system failed | Protection activation or feedback loss | error |
91 | ICC 350 NEUROTEST; electronic fuse not triggered or frequency feedback system failed after intensity adjustment | Protection circuit or feedback failure during setup | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI