ERBE ICC 200: troubleshooting and error codes

The ServiceForge knowledge base holds 20 documented symptoms for the ERBE ICC 200, with 22 ranked causes, 56 diagnostic checks with test points and expected values, and 62 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: HF generator output parameters high voltage

Adjustment of HF output voltage and current limitation on ICC electrosurgical generator to ensure correct power output and phase angle measurement.

Ranked causes

  1. 1. HF output voltage requires adjustment via TP3 (common)
  2. 2. HF current limitation requires adjustment via TP4 (common)
First check

Measure HF effective voltage in AUTO CUT field

Test point
AUTO CUT display, TP3 adjustment point
Expected
223 volts HF effective voltage V, 220…223 V
Tools
APM 600 HF power meter set to 500 ohms

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: HF generator output parameters amplified high voltage

Fine adjustment of amplified HF voltage and current measurement circuits for low-power output measurement (5 watts).

Ranked causes

  1. 1. Amplified HF voltage measurement requires adjustment via TP5 (common)
  2. 2. Amplified HF current measurement requires adjustment via TP6 (common)
First check

Verify 5 watts output at HF power meter before adjustment

Test point
APM 600 HF power meter display
Expected
5 watts output W, 4.5…5.5 W
Tools
APM 600 HF power meter at 500 ohms

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Activation detection (MIC board) high voltage

Setting activation detection using trimming potentiometer TP2 in program 12 MIC

Ranked causes

  1. 1. Activation detection requires initial adjustment after instrument detection calibration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AUTO START delay time

Setting delay time between tissue contact detection and HF generator activation in AUTO START function

Ranked causes

  1. 1. AUTO START delay time requires adjustment for specific operation modes or per user program (common)
First check

Verify AUTO START 2 delay time setting (ICC 350)

Test point
Test program 23 display
Expected
2 ≤ t ≤ 10 seconds in 0.5 second steps
Tools
Key 14, Up/Down keys

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bipolar CUT function (MIEN probe) high voltage

Testing bipolar CUT function activation using MIEN probe with 15W Effect 3 setting

Ranked causes

  1. 1. MIEN bipolar CUT function requires calibration and verification (common)
First check

Measure bipolar CUT output power

Test point
MIC socket NE1 CUT, Effect 3 setting
Expected
15 ± 3 W, ± 3 W
Tools
APM, 500 Ohm load

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bipolar SOFT coagulation (MIEN probe) high voltage

Testing bipolar SOFT coagulation with no-load voltage measurement at 15W setting

Ranked causes

  1. 1. MIEN bipolar SOFT coagulation requires verification after CUT testing (common)
First check

Measure bipolar SOFT output power

Test point
MIC socket NE1 CUT, SOFT setting
Expected
7 ± 2 W, ± 2 W
Tools
APM, 500 Ohm load

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CUT function (MIC program) high voltage

Testing CUT cutting function activation on MIC instrument with footswitch and power measurement

Ranked causes

  1. 1. CUT function requires testing after instrument and activation detection adjustment (common)
First check

Measure CUT function output power

Test point
MIC socket NE1 and CUT pin
Expected
40 ± 15 % W, ± 15 %
Tools
APM (Automatic Power Meter), 500 Ohm load

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display brightness (7-segment)

Adjusting brightness intensity of 7-segment display indicators in 10 stages via test program 17

Ranked causes

  1. 1. Display brightness requires adjustment for optimal visibility in different room lighting conditions (common)
First check

Observe 7-segment display intensity at current setting

Test point
4-line display showing AUTO CUT/AUTO COAG 1/AUTO COAG 2/AUTO BIPOLAR
Expected
Intensity from 1–10, adjustable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Instrument detection (MIC board) high voltage

Setting instrument detection resistor value to proper range using trimming potentiometer TP1 in program 12 MIC

Ranked causes

  1. 1. Instrument detection requires calibration during setup or after component replacement (common)
First check

Measure instrument detection I value with 51 Ohm resistor connected

Test point
Measurement display in AUTO CUT/AUTO COAG 1 view
Expected
52, at set value
Tools
Program 12 MIC display

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monopolar coagulation FORCED function high voltage

Testing FORCED coagulation activation at specified power setting with high load resistance

Ranked causes

  1. 1. FORCED coagulation function requires verification after SOFT testing (common)
First check

Measure FORCED coagulation output power

Test point
MIC socket NE1
Expected
60 ± 15 % W, ± 15 %
Tools
APM, 350 Ohm load

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monopolar coagulation SOFT function high voltage

Testing SOFT coagulation activation with blue footswitch at specified power setting

Ranked causes

  1. 1. SOFT coagulation function requires verification after CUT testing (common)
First check

Measure SOFT coagulation output power

Test point
MIC socket NE1
Expected
120 ± 15 % W, ± 15 %
Tools
APM, 75 Ohm load

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Neurotest board feedback frequency high voltage

Set feedback frequency of output current measurement via trimming potentiometer TP1 on Neurotest board EE 30128-070. Requires 1 kOhm 0.5W 1% test resistor at output.

Ranked causes

  1. 1. Neurotest board EE 30128-070 requires feedback frequency adjustment (common)
First check

Measure feedback frequency of output current at MP7

Test point
MP7 (reference ground MP6)
Expected
73 Hz at SPH intensity 3, +/-10 Hz
Tools
Frequency meter

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Neurotest constant-current circuit output high voltage

Performance test of constant-current output circuit. Connect 1 kOhm test resistor and 50 kOhm potentiometer in series at Neurotest output. Measure square-wave voltage on 1 kOhm resistor (1 mA = 1 V).

Ranked causes

  1. 1. Neurotest constant-current source requires output current verification (common)
First check

Measure square-wave peak output current at SPH 1

Test point
1 kOhm test resistor
Expected
5 mA, 3.75–6.25 mA (±25%)
Tools
Oscilloscope, DMM

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Neurotest electronic fuse protection high voltage

Performance test of electronic fuse protection. Excess dose triggers fuse via transistor T3. Test by activating Neurotest and briefly connecting MP9 to MP2. Fuse should respond without blowing microfuse.

Ranked causes

  1. 1. Electronic fuse protection circuit requires functional verification (common)
First check

Activate Neurotest and briefly connect MP9 to MP2

Test point
MP9, MP2
Expected
Electronic fuse trips; error code appears

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Neurotest LED activation control

Test HF and NT LED activation. HF LED lights only at intensity 0. At intensities 1–5, NT LED controlled by CPU; HF LED must not light. On TUR version, HF inhibited at 1–5. On ZMK version, HF still possible.

Ranked causes

  1. 1. LED activation control requires verification per device version (common)
First check

Set intensity to 0 and verify HF LED on

Test point
HF LED on remote control
Expected
HF LED illuminated at intensity 0

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Neurotest output error monitoring high voltage

Performance test of output error monitoring function. Connect 1 kOhm 0.5W 1% test resistor at Neurotest output, set SPH 1, activate unit and verify error message on circuit interruption.

Ranked causes

  1. 1. Output error monitoring circuit requires verification (common)
First check

Set intensity to SPH 1 and activate Neurotest

Test point
Neurotest output
Expected
Unit activated; output current flowing

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Phase angle (cos Phi) high voltage

Adjustment of phase angle measurement between HF voltage and current using capacitive test load to verify analog/digital converter accuracy.

Ranked causes

  1. 1. Phase angle measurement via TP7 requires calibration (common)
First check

Connect APM 600 to ICC output with 1 nF capacitive test load

Test point
Active electrode (CUT/COAG 2) and NE sockets
Expected
Capacitor creates 45° phase shift at 340 kHz frequency
Tools
APM 600, ERBE original cable, 1 nF capacitor EE 51103-026 with short lines

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Protective circuit voltage limitation (MIEN) high voltage

Testing protective circuit voltage limiting function at different HF output voltage settings on MIEN program

Ranked causes

  1. 1. Protective circuit voltage limiting requires verification to ensure proper fuse operation and diode protection (common)
First check

Measure highest HF peak voltage at Effect 3 (15W) setting

Test point
MIC output socket NE1 and CUT with oscilloscope
Expected
340 Vp max without limitation
Tools
Oscilloscope

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remote control feedback frequency for Neurotest

Set feedback frequency of remote control intensity setting via trimming potentiometer TP1 on board EE 30121-442. Measure frequency at Pin 1 of AD654 IC1. Setting at intensity 5: 38.0 Hz (–0.5, +0.0).

Ranked causes

  1. 1. Remote control intensity frequency requires adjustment (common)
First check

Measure frequency at Pin 1 AD654 IC1 for intensity 1

Test point
Pin 1 AD654 IC1 (reference ground MP6)
Expected
7 Hz, 5–10 Hz
Tools
Frequency meter

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remote control LED display for Neurotest

Test HF LED and NT LED activation on remote control board EE 30121-442. HF LED at intensity 0 via rotary switch. NT LED at intensities 1–5 via CPU; lights only when feedback frequency in valid range.

Ranked causes

  1. 1. Remote control LED display requires functional verification (common)
First check

Set intensity 0 and verify HF LED on

Test point
HF LED on remote control
Expected
HF LED illuminated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1No HF voltage present at the HF voltage sensorError in the HF generator or short circuit in accessorieserror
10Error during activation of AUTO CUT, no valid front panel settingInvalid control panel settingerror
11Error during activation of AUTO COAG 1 or AUTO COAG 2, no valid front panel settingInvalid control panel settingerror
12Error during activation of AUTO BIPOLAR, no valid front panel settingInvalid control panel settingerror
13Error during activation of AUTO CUT; NE impedance out of rangeNeutral electrode impedance measurement errorerror
17HF leakage current > 300 mA during activationExcessive HF leakage current detectederror
18HF leakage current between 150 and 300 mA during activationElevated HF leakage current detectedwarning
19NESSY current density error during AUTO CUT or AUTO COAG activationCurrent density monitoring failureerror
2HF output voltage too highError in the HF generatorerror
20NESSY current symmetry error; unequal current in NE joint phasesCurrent symmetry monitoring failureerror
21Current symmetry errorAsymmetric current detectederror
22AUTO CUT switched on when switching on powerFingerswitch or footswitch activated during power-onerror
23AUTO COAG 1 switched on when switching on powerFingerswitch or footswitch activated during power-onerror
24AUTO COAG 2 switched on when switching on powerFootswitch activated during power-onerror
25AUTO BIPOLAR switched on by footswitch when switching on powerFootswitch activated during power-onerror
26Contact detected when switching on power; contact monitor is activeShort circuit detected during power-onerror
27Overheating; internal unit temperature > 100°CExcessive internal temperaturewarning
28Activation signal for blue footswitch button already onBlue footswitch button detected as pressed during activationerror
29Double activation detected (Spanish version)Further activation signal detected during activationerror
3Switched-mode power supply unit supplies no voltage during activationError in the switched-mode power supply uniterror
30Mismatch; load resistance < 50 ohms for > 5 secondsLow load resistance detected during operationerror
31Mismatch; output power > 100 watts and load resistance remains < 300 ohmsExcessive output power with low load resistanceerror
32Multiple activation detectedSimultaneous activation from multiple sourceserror
33New activation signal immediately appears at end of activationImmediate re-activation detectederror
34Multiple activation detected during AUTO BIPOLARSimultaneous activation during AUTO BIPOLARerror
35New activation signal during AUTO BIPOLAR with AUTO STOP or AUTO STARTSimultaneous activation detected during AUTO BIPOLAR modeerror
36AUTO START signal already present when pressing BIPOLAR AUTO START keyDuplicate AUTO START signalerror
37Power supply unit voltage has not dissipated at activation timePower supply discharge failureerror
38Contact resistance < 6 ohms during AUTO START activationLow contact resistance detectederror
4Switched-mode power supply unit supplies too high a voltage upon ST generator activationError in the switched-mode power supply uniterror
40ST generator time control output error; actual value > set value plus 20%Time control monitoring deviation detectederror
41ST generator time control output error; actual value < set value minus 20%Time control monitoring deviation detectederror
42ST generator time control output error; actual value > set value minus 40%Time control monitoring deviation detectederror
43ST generator time control monitor error; no frequency feedbackFrequency feedback losserror
44ST generator output voltage monitor errorOutput voltage monitoring failureerror
45HF generator error during self-check; HF output voltage too lowSelf-check failureerror
46Self-check error within permissible limits; power supply versus HF voltage mismatchSelf-check comparison failureerror
47HF generator self-check error outside permissible limitsSelf-check comparison failureerror
48No signal from output current monitorCurrent monitoring failureerror
5LF leakage current > 50 µA flowing into unit via neutral electrodePatient positioning issue or defective equipment connected to patientwarning
50-70Key on front panel is depressed during power start-up phaseFront panel key activation during startuperror
6Safety shutoff feature output reports ON during activation phaseMalfunction during activationerror
7Safety shutoff feature output reports OFF during activation phaseMalfunction during activationerror
71Interface error for ICC 350 DOKU; start/stop bit or number of bits incorrectSerial communication errorerror
72Interface error for ICC 350 DOKU; signal in receiving register not read outData reception failureerror
73Interface error for ICC 350 DOKU; parity sign placed incorrectlyParity check failureerror
74Interface error for ICC 350 DOKU; no connection to PC availablePC connection losserror
75Interface error for ICC 350 DOKU; sign stays in transmitter buffer; SIO module not transmittingData transmission failureerror
76Interface error for ICC 350 DOKU; storage takes too long; hard disk too slowData storage latencyerror
80ICC 350 NEUROTEST; no valid frequency for 5-stage intensity settingFrequency configuration errorerror
81ICC 350 NEUROTEST; intensity setting set to 0 during activationInvalid intensity configurationerror
82ICC 350 NEUROTEST; NESSY measurement NE impedance > 120 ohmsNeutral electrode impedance measurement failureerror
83ICC 350 NEUROTEST; output current too high but within permissible limitsCurrent level monitoringwarning
84ICC 350 NEUROTEST; output current too high outside permissible limitsExcessive output currenterror
85ICC 350 NEUROTEST; no output current; circuit interruption or frequency feedback failureOutput current losserror
86ICC 350 NEUROTEST; output current too low but within permissible limitsInsufficient output currentwarning
87ICC 350 NEUROTEST; output current too low below permissible limitExcessive current reductionerror
88ICC 350 NEUROTEST; maximum time limit has run outContinuous operation exceeds time limitwarning
89ICC 350 NEUROTEST; activation signal already present when switching on powerPremature activation signal on startuperror
9Maximum continuous time limit exceededUnit activated for too long continuouslywarning
90ICC 350 NEUROTEST; electronic fuse triggered or frequency feedback system failedProtection activation or feedback losserror
91ICC 350 NEUROTEST; electronic fuse not triggered or frequency feedback system failed after intensity adjustmentProtection circuit or feedback failure during setuperror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI