ERBE 10134-000: troubleshooting and error codes
The ServiceForge knowledge base holds 16 documented symptoms for the ERBE 10134-000, with 16 ranked causes, 24 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
APC Coag output power out of specification
Forced APC Coag mode does not deliver expected power output within acceptable range
Ranked causes
- 1. System output power drift or malfunction (unknown)
Measure APC Coag output power in Forced mode
- Test point
- APC output
- Expected
- 96 to 144 Watts, 96 to 144 Watts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow rate out of specification at 0.2 L/min setting
Measured gas flow deviates from acceptable range when flow control set to 0.2 L/min
Ranked causes
- 1. Flow control valve malfunction or blockage (unknown)
Measure gas flow at 0.2 L/min setting
- Test point
- Flow meter or measurement port
- Expected
- 0.16 – 0.24 L/min, 0.16 – 0.24 L/min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow rate out of specification at 0.5 L/min setting
Measured gas flow deviates from acceptable range when flow control set to 0.5 L/min
Ranked causes
- 1. Flow control valve malfunction or blockage (unknown)
Measure gas flow at 0.5 L/min setting
- Test point
- Flow meter or measurement port
- Expected
- 0.4 – 0.6 L/min, 0.4 – 0.6 L/min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow rate out of specification at 1.0 L/min setting
Measured gas flow deviates from acceptable range when flow control set to 1.0 L/min
Ranked causes
- 1. Flow control valve malfunction or blockage (unknown)
Measure gas flow at 1.0 L/min setting
- Test point
- Flow meter or measurement port
- Expected
- 0.8 – 1.2 L/min, 0.8 – 1.2 L/min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow rate out of specification at 2.0 L/min setting
Measured gas flow deviates from acceptable range when flow control set to 2.0 L/min
Ranked causes
- 1. Flow control valve malfunction or blockage (unknown)
Measure gas flow at 2.0 L/min setting
- Test point
- Flow meter or measurement port
- Expected
- 1.6 – 2.4 L/min, 1.6 – 2.4 L/min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow rate out of specification at 5.0 L/min setting
Measured gas flow deviates from acceptable range when flow control set to 5.0 L/min
Ranked causes
- 1. Flow control valve malfunction or blockage (unknown)
Measure gas flow at 5.0 L/min setting
- Test point
- Flow meter or measurement port
- Expected
- 4.0 – 6.0 L/min, 4.0 – 6.0 L/min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow setting inaccurate
Flow meter reading does not match set value across multiple settings from 0.2 to 5.0 L/min
Ranked causes
- 1. Flow calibration out of tolerance (unknown)
Measure gas flow at 0.2 L/min setting
- Test point
- Flow meter
- Expected
- 0.16 – 0.24 L/min., 0.16 – 0.24 L/min.
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Output power below specification
Forced APC Coag mode output power outside acceptable range during verification
Ranked causes
- 1. Output power out of specification (unknown)
Measure output power in Forced APC Coag mode at 500 setting
- Test point
- output
- Expected
- 96 to 144 Watts, 96 to 144 Watts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output pressure error with system activation
Pressure measurement inaccuracy at 30 psi output when system is activated and stopped
Ranked causes
- 1. Pressure measurement drift during system state changes (unknown)
Measure output pressure error at 30 psi with system activation/stop cycles
- Test point
- System output pressure gauge or display
- Expected
- Error ≤ 6 psi, ≤ 6 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure reading error at maximum threshold
System pressure measurement inaccuracy when set to 108 psi maximum pressure condition
Ranked causes
- 1. Pressure transducer or measurement circuit drift (unknown)
Measure pressure error at 108 psi maximum
- Test point
- System pressure gauge or display
- Expected
- Error ≤ 20 psi, ≤ 20 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure reading error at minimum threshold
System pressure measurement inaccuracy when set to 30 psi minimum pressure condition
Ranked causes
- 1. Pressure transducer or measurement circuit drift (unknown)
Measure pressure error at 30 psi minimum
- Test point
- System pressure gauge or display
- Expected
- Error ≤ 6 psi, ≤ 6 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System pressure instability at operating point
System pressure regulation instability when input maintained at 75 psi
Ranked causes
- 1. Pressure regulator malfunction or internal leakage (unknown)
Monitor pressure stability with 75 psi input
- Test point
- System pressure gauge or display
- Expected
- Stable pressure reading without significant drift
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System pressure out of tolerance caution
Pressure readings do not meet specified minimum, maximum or error limits
Ranked causes
- 1. Pressure calibration out of specification (unknown)
Check minimum pressure error at 30 psi input
- Test point
- Pressure gauge
- Expected
- 6 psi maximum error, 6 psi
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Tank pressure display not showing low at threshold
Tank pressure indication fails to display low state when tank pressure drops to 90 psi
Ranked causes
- 1. Low pressure switch malfunction or display circuit failure (unknown)
Check tank display shows low at 90 psi pressure
- Test point
- Tank pressure display or indicator
- Expected
- Display indicates low
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tank pressure display not showing normal at high pressure
Tank pressure indication fails to display normal state when tank pressure is at 750 psi
Ranked causes
- 1. Pressure switch or display circuit malfunction (unknown)
Check tank display shows normal at 750 psi pressure
- Test point
- Tank pressure display or indicator
- Expected
- Display indicates normal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tank pressure indication incorrect caution
Tank display does not show Normal at 750 psi or Low at 90 psi as specified
Ranked causes
- 1. Pressure sensor or display calibration out of range (unknown)
Verify tank displays Normal at 750 psi
- Test point
- Tank display indicator
- Expected
- Normal psi
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI