Encision AEM Monitor: troubleshooting and error codes
The ServiceForge knowledge base holds 16 documented symptoms for the Encision AEM Monitor, with 55 ranked causes, 37 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on (Bipolar End Point Monitor)
No power, no indicators, no beeps when powered on. POST function not executed.
Ranked causes
- 1. Disconnected or faulty power cord (common)
- 2. Blown fuse or open fuse drawer (common)
- 3. Malfunctioning power entry module (occasional)
- 4. Malfunctioning power switch (occasional)
- 5. Malfunctioning AC Power Cable (internal) (occasional)
- 6. Malfunctioning power supply (occasional)
- 7. Internal cables loose or faulty (common)
- 8. Malfunctioning ESM Display Board (occasional)
Verify power cord connections
- Test point
- power cord at monitor and wall
- Expected
- firmly connected
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
No power, no indicators, no beeps. POST function did not execute at startup.
Ranked causes
- 1. Disconnected or faulty power cord (common)
- 2. Blown fuse or open fuse drawer (common)
- 3. Malfunctioning power entry module (occasional)
- 4. Malfunctioning power switch (occasional)
- 5. Malfunctioning AC power cable (internal) (occasional)
- 6. Malfunctioning power supply (occasional)
- 7. Internal cables loose or disconnected (common)
Verify power cord connections
- Test point
- power cord at monitor and wall receptacle
- Expected
- firmly connected
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AEM Cord indicator remains on patient safety
Electrosurgical generator locked in alarm state. AEM (Active Electrode Monitor) cord indicator continuously lit.
Ranked causes
- 1. Poor cord connection to instrument or monitor (common)
- 2. Malfunctioning or damaged AEM cord (occasional)
- 3. Malfunctioning connection to ESM Main Board (occasional)
- 4. Malfunctioning ESM Main Board (rare)
- 5. Malfunctioning ESM Display Board (occasional)
Verify cord connections
- Test point
- AEM cord connectors at instrument and monitor
- Expected
- firmly seated
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel display does not turn off when remote plugged in
Front panel bar graph display continues showing indication when End Point Monitor Remote is plugged into rear connector.
Ranked causes
- 1. Malfunctioning EPM Main Board (occasional)
- 2. Internal cables loose or malfunctioning (occasional)
- 3. Malfunctioning End Point Monitor Remote unit (occasional)
- 4. Defective EPM Main Board (rare)
- 5. Remote cable or connection problem (occasional)
Test remote resistance
- Test point
- remote display plug pins 1 & 2
- Expected
- <=2 ohms, <=2
- Tools
- Ohmmeter
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Return Electrode indicator remains on patient safety
Electrosurgical generator locked in alarm state. Return Electrode indicator continuously lit.
Ranked causes
- 1. Improper return electrode installed (common)
- 2. Dual Electrode pin not being detected (common)
- 3. Malfunctioning connection to ESM Main Board (occasional)
- 4. Defective ESM Main Board (rare)
- 5. Defective ESM Display Board (occasional)
Verify dual-pad return electrode
- Test point
- return electrode connector
- Expected
- dual-pad with projecting pin
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrosurgical generator in alert state patient safety
Contact quality monitor activated. Generator not in full alarm but alert condition is triggered.
Ranked causes
- 1. Dual pad open or short circuited (common)
- 2. Poor connection in return electrode circuit (common)
- 3. Malfunctioning electrosurgical generator (occasional)
- 4. Malfunctioning connection to ESM Main Board (occasional)
Verify dual pad resistance
- Test point
- dual pad electrodes
- Expected
- correct resistance per ESG
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audible indication (no clicking)
Speaker does not click when ESG activated, but bar graph shows correct indication.
Ranked causes
- 1. Volume Control switch in wrong position (common)
- 2. Malfunctioning speaker, switch, or cable (occasional)
- 3. Malfunctioning EPM Main Board (occasional)
- 4. Faulty power supply (occasional)
Check Volume Control switch position
- Test point
- Volume Control switch
- Expected
- switch in High (up) position
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ready indicator remains on with no fault indicators patient safety
Generator locked in alarm state but Ready indicator lights and other fault indicators are off.
Ranked causes
- 1. Inappropriate internal state of AEM Monitor (common)
- 2. Malfunctioning electrosurgical generator (occasional)
- 3. Malfunctioning ESM Display Board (occasional)
- 4. Malfunctioning connection to ESM Main Board (occasional)
Power cycle the unit
- Test point
- power switch
- Expected
- unit resets, Ready indicator off
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuzzy indication on bar graph at high end
Bar graph indication is unstable at high end of current scale. Bars at right end neither fully on nor off.
Ranked causes
- 1. Internal cables loose or malfunctioning (common)
- 2. Malfunctioning EPM Display Board (occasional)
- 3. Malfunctioning EPM Main Board (occasional)
Reseat internal connections
- Test point
- all internal connectors
- Expected
- firmly seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insulation indicator does not light during fault patient safety
AEM Monitor fails to detect operative fault. Insulation light remains off when fault occurs.
Ranked causes
- 1. Arc detector failure (rare)
- 2. Short detector failure (rare)
- 3. Blown fuse(s) on power supply (unknown)
Test arc detection
- Test point
- AEM signal line
- Expected
- Insulation light on, ESG deactivated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No bar graph indication
Bipolar current measurement display shows no bar graph indication when ESG is activated.
Ranked causes
- 1. Malfunctioning EPM Display Board (occasional)
- 2. Malfunctioning EPM Main Board (occasional)
Test with replacement display board
- Test point
- J2 connector, bipolar output setup
- Expected
- bar graph indicates at ~3W
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
End Point Monitor Remote Display does not indicate when plugged in
Remote display connector shows no indication when plugged into rear connector, but front panel display works when remote is removed.
Ranked causes
- 1. Defective End Point Monitor Remote unit (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
End Point Monitor Remote plugged into rear connector
Issue with remote display connection to End Point Monitor.
Ranked causes
- 1. End Point Monitor Remote plugged in (common)
Inspect remote connector
- Test point
- Remote Display connector on rear
- Expected
- empty/no remote plugged in
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel not functioning
POST begins but front panel indicators do not light or function properly.
Ranked causes
- 1. Front panel malfunction (occasional)
Test with replacement display board
- Test point
- J2 connector on ESM Main Board
- Expected
- indicators function with replacement board
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator not in alert state during fault patient safety
Ready indicator is off and any of the following are on: AEM Cord, Return Electrode, or Insulation indicators.
Ranked causes
- 1. Defective ESM Main Board (rare)
Measure ESM Main Board resistance
- Test point
- J1 pins 5 & 6
- Expected
- >2 ohms, >2
- Tools
- Ohmmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply output voltages incorrect high voltage
After Three Output Power Supply replacement, one or more output voltages at J2 are out of specification
Ranked causes
- 1. Three Output Power Supply board failure or misadjustment (common)
Measure output voltages at power supply J2
- Test point
- J2 pins 1, 2, 3, 4, 5
- Expected
- Pin 1: +15.0 ±.1V; Pin 2: ground; Pin 3: +5.27 ±.05V; Pin 4: ground; Pin 5: -15.0 ±.1V, ±0.1V (±15V), ±0.05V (+5.27V), ±0.1V (-15V)
- Tools
- Digital multimeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI