Edwards Lifesciences HemoSphere: troubleshooting and error codes
The ServiceForge knowledge base holds 32 documented symptoms for the Edwards Lifesciences HemoSphere, with 42 ranked causes, 49 diagnostic checks with test points and expected values, and 56 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Cannot export clinical data to USB device
Export Data screen accessible but data download fails or does not complete
Ranked causes
- 1. USB device out of space (occasional)
- 2. USB device not approved by Edwards (occasional)
- 3. USB device contains virus or malware (rare)
Verify USB device has sufficient free space
- Test point
- USB device
- Expected
- Adequate space available
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Finger cuff status light amber flashing
Pressure controller LED shows amber flashing ON/OFF pattern for connected finger cuff, indicating defective, expired, or incompatible cuff
Ranked causes
- 1. Defective finger cuff (occasional)
- 2. Expired finger cuff (occasional)
- 3. Non-compatible Edwards finger cuff connected (occasional)
Verify cuff compatibility and condition
- Test point
- Pressure controller cuff status light
- Expected
- Amber flashing ON/OFF
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot access diagnostic export function
Unable to reach Diagnostics Export button or complete diagnostic data download
Ranked causes
- 1. Incorrect Super User password entered (common)
- 2. USB ports unavailable or non-functional (occasional)
Enter Super User password
- Test point
- Password entry field
- Expected
- Password accepted; access to Diagnostics Export granted
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fault: StO2 FORE-SIGHT ELITE module communication error patient safety
Module disconnected or communication lost between tissue oximetry module and FORE-SIGHT ELITE oximeter module (port A or B)
Ranked causes
- 1. FORE-SIGHT ELITE oximeter module (FSM) physically disconnected from tissue oximetry module port (common)
- 2. Communication loss between tissue oximetry module and FORE-SIGHT ELITE module due to connection fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart reference sensor status light amber flashing
Pressure controller LED shows amber flashing ON/OFF pattern for heart reference sensor, indicating defective or non-Edwards sensor connected
Ranked causes
- 1. Defective heart reference sensor (occasional)
- 2. Non-Edwards heart reference sensor detected (occasional)
Verify heart reference sensor is from Edwards
- Test point
- Pressure controller heart reference sensor status light
- Expected
- Amber flashing ON/OFF
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module cable connections damaged or misaligned caution
Visibly damaged cable ports on module; cables cannot insert properly; CCO cable insertion difficult or impossible
Ranked causes
- 1. Bent or damaged connector pins in cable port (unknown)
- 2. Cable incompatibility or contamination in port (occasional)
Inspect all module cable ports visually
- Test point
- Cable connection ports on module
- Expected
- Ports intact with no bent pins or visible damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module housing damaged or labeling unreadable caution
Cracks, scratches, dents in module housing; labeling missing, unreadable, or damaged
Ranked causes
- 1. Physical damage during handling or installation (unknown)
- 2. Labeling degraded or removed (occasional)
Inspect module housing condition
- Test point
- Module exterior housing
- Expected
- Intact housing with no cracks, scratches, or dents; housing clean
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module slots obstructed or damaged caution
Module slots contain debris or have visible damage to slots or cable connection points
Ranked causes
- 1. Accumulation of dust or debris in slots (occasional)
- 2. Physical damage to slot connectors or cable ports (unknown)
Inspect module slots for debris
- Test point
- L-Tech expansion module slot and two expansion module slots
- Expected
- Slots clear of debris and foreign material
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Applied part leakage current out of specification high voltage
Individual patient-connected components (CCO cable, oximetry cable, sensors, transducers) exceed leakage limits.
Ranked causes
- 1. Insulation failure in specific patient cable or sensor (occasional)
Wrap each applied part (CCO cable, oximetry cable, sensors, transducers) individually in aluminum foil
- Test point
- Each component separately
- Expected
- Foil completely covering component with minimum 30 cm cable coverage
- Tools
- Aluminum foil, alligator clip leads
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable damage or degradation caution
Exposed wire, splits, cracks, or stress marks visible on cables; loss of strain relief
Ranked causes
- 1. Physical wear or strain on cables from handling or routing (unknown)
Inspect cable housing and strain relief
- Test point
- Cable exterior and connection points
- Expected
- Intact housing with visible strain relief; no exposed wires or shielding
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable optical door malfunction caution
Oximetry cable optical connection door does not move freely or fails to latch properly
Ranked causes
- 1. Door mechanism stuck or damaged; hinges misaligned (rare)
Test oximetry cable optical door movement
- Test point
- Optical connection enclosure door on HEMOXSC100 cable
- Expected
- Door moves freely without sticking; door latches securely when closed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment leakage current out of specification high voltage
Leakage current from applied parts (catheters, cables, sensors) exceeds safe limits, posing electrical hazard to patient.
Ranked causes
- 1. Insulation degradation or breakdown in patient cables or connectors (occasional)
Measure equipment leakage current with all modules and patient cables connected
- Test point
- EQP terminal to combined foil-wrapped patient connection points
- Expected
- Not to exceed 500 µA, 500 µA maximum
- Tools
- Electrical Safety Analyzer (ESA) per IEC 62353:2014
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Finger cuff status light green flashing
Pressure controller LED shows green flashing ON/OFF, indicating finger cuff is actively monitoring
Ranked causes
- 1. Normal operation: finger cuff actively monitoring (common)
Finger cuff status light solid green
Pressure controller LED shows green solid ON, indicating finger cuff is connected, authenticated, and not expired
Ranked causes
- 1. Normal operation: compatible finger cuff connected (common)
Finger cuff status light solid off
Pressure controller LED shows no light (solid OFF) for finger cuff, indicating no cuff is connected or detected
Ranked causes
- 1. No finger cuff connected (common)
FORE-SIGHT ELITE oximeter module LED not responding patient safety
One or more LEDs on the FORE-SIGHT ELITE oximeter module fail to turn on or respond
Ranked causes
- 1. Module damaged or faulty (unknown)
Observe FORE-SIGHT ELITE oximeter module LEDs
- Test point
- Channel 1 status, Channel 2 status, Module status LEDs
- Expected
- LEDs illuminate according to sensor connection and port status
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart reference sensor status light solid green
Pressure controller LED shows green solid ON for heart reference sensor, indicating sensor is connected and system is ready to start measurement
Ranked causes
- 1. Normal operation: heart reference sensor connected (common)
Heart reference sensor status light solid off
Pressure controller LED shows no light (solid OFF) for heart reference sensor, indicating no sensor is connected or detected
Ranked causes
- 1. No heart reference sensor connected (common)
Housing damage visible caution
Cracks, dents, scratches, signs of falling or abuse on monitor housing or touchscreen
Ranked causes
- 1. Physical damage from impact or mishandling (unknown)
Visually inspect housing for damage
- Test point
- Monitor exterior housing
- Expected
- Intact housing with no scratches, dents, or cracks
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LIVE DEMO banner displayed with tissue oximetry simulator patient safety
Monitor screen shows LIVE DEMO banner when StO2 simulator connected to tissue oximetry module, indicating simulated values not from patient
Ranked causes
- 1. Tissue oximetry simulator connected for functional testing (common)
Observe LIVE DEMO banner presence on screen during simulator testing
- Test point
- monitor display top banner area
- Expected
- LIVE DEMO banner visible across top of screen
Loose parts or foreign bodies present caution
After service or repair, loose components or debris remain inside or on the device
Ranked causes
- 1. Incomplete final assembly or cleanup after maintenance (occasional)
Check for loose parts or foreign bodies
- Test point
- Interior and exterior of monitor
- Expected
- No loose components, debris, or foreign material present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor rear labeling missing or unreadable
Rear center label or serial number label not present or illegible on monitor
Ranked causes
- 1. Label removed, damaged, or degraded from wear (occasional)
Verify rear center label present
- Test point
- Rear center of monitor housing
- Expected
- Label present and clearly readable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Port damage or obstruction caution
Visible damage to USB, HDMI, Ethernet, COM1, Analog input, ECG input, or pressure output ports
Ranked causes
- 1. Physical damage to port connectors from impact or insertion (unknown)
Inspect rear panel ports visually
- Test point
- HDMI, Ethernet, COM1, Analog inputs 1 and 2, ECG input, pressure output, equipotential stud
- Expected
- All ports intact with no damage, bent pins, or obstructions
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure cable LED flashing green
Pressure cable LED flashing green when sensor connected
Ranked causes
- 1. Pressure sensor connected but not zeroed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure cable LED off
No light on pressure cable LED indicator
Ranked causes
- 1. No pressure sensor or transducer connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure cable or sensor connection defective caution
Pressure sensor or transducer cannot insert into pressure cable port; connection damaged or misaligned
Ranked causes
- 1. Cable port connector damaged or misaligned; sensor incompatible (unknown)
Test pressure sensor or transducer insertion
- Test point
- Pressure sensor/transducer connection port on HEMPSC100 cable
- Expected
- Compatible pressure sensor inserts smoothly and seats fully; connection is secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Protective earth resistance out of specification high voltage
Earth conductor resistance between grounding pin and EQP terminal exceeds safe limits, indicating potential electrical safety hazard.
Ranked causes
- 1. Damaged or degraded earth conductor in mains power cord (common)
Measure protective earth resistance using ESA with test current 200-1000 mA
- Test point
- Grounding pin of power cord to EQP terminal on back panel
- Expected
- Not to exceed 300 mOhms, 300 mOhms maximum
- Tools
- Electrical Safety Analyzer (ESA) per IEC 62353:2014
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings cannot be exported to USB
Export Settings button does not function or settings fail to save to USB stick
Ranked causes
- 1. USB stick is not approved Edwards brand or is not inserted (common)
Verify USB stick is inserted and approved Edwards brand
- Test point
- Monitor USB port
- Expected
- Approved Edwards USB stick physically inserted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings import fails or pattern does not match
Imported settings do not apply or on-screen pattern verification fails during import to another monitor
Ranked causes
- 1. Pattern image does not match between export and import (occasional)
Verify on-screen pattern matches exported image
- Test point
- Monitor display during import
- Expected
- Pattern on-screen matches photo taken during export
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Error Screen displayed during startup
After power-on, System Error Screen replaces the normal Startup Screen instead of showing POST status and system information
Ranked causes
- 1. POST detected error condition in critical hardware components (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System leakage current with LAN connection out of specification high voltage
When HemoSphere is networked via Ethernet, system leakage current exceeds limits, indicating electrical hazard.
Ranked causes
- 1. Grounding issue introduced by LAN network connection (rare)
Measure equipment leakage current with LAN cable connected to hardwired Ethernet port
- Test point
- EQP terminal to combined foil-wrapped patient connection points with LAN attached
- Expected
- Not to exceed 500 µA, 500 µA maximum
- Tools
- Electrical Safety Analyzer (ESA) per IEC 62353:2014
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB stick contains virus or malware caution
USB stick is contaminated and may infect monitor when inserted
Ranked causes
- 1. USB stick not scanned for virus or malware before insertion (occasional)
Scan USB stick with antivirus software
- Test point
- Computer antivirus tool
- Expected
- No viruses or malware detected on USB stick
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Alert: Battery Disconnected | Previously inserted battery not detected | Battery not detected or poor connection | warning |
Alert: Low Battery | Battery has less than 20% charge remaining or will be depleted within 8 minutes | Battery charge drops below 20% or depletion predicted within 8 minutes | warning |
Alert: Service Battery | Internal battery fault occurred, battery cannot sustain system adequately | Internal battery fault detected | warning |
Alert: System Buzzer Inoperable | Speaker hardware or software communication error | Speaker or buzzer malfunction detected | warning |
Alert: System LED Indicators Inoperable | Visual alarm indicator hardware or communication error | Visual alarm indicator malfunction or communication error | warning |
Alert: System Temperature Too High | Internal temperature reaching critically high level | Internal temperature reaches elevated threshold | warning |
Alert: Transmit Pressure Not Active | Connection of new patient monitor pressure channel detected | New patient monitor pressure channel connected | info |
Alert: Wireless Module Failure | Internal hardware failure in wireless module | Wireless module hardware failure detected | warning |
Battery Needs Conditioning | Gas gauge not synced to actual battery capacity status | Battery gauge synchronization failure | warning |
Channel 1 Green LED | Sensor connected and active on Channel 1 | Valid sensor connected to Channel 1 | info |
Channel 1 White LED | No sensor connected to Channel 1 | Sensor disconnected or not plugged in to Channel 1 | info |
Channel 2 Green LED | Sensor connected and active on Channel 2 | Valid sensor connected to Channel 2 | info |
Channel 2 White LED | No sensor connected to Channel 2 | Sensor disconnected or not plugged in to Channel 2 | info |
Fault: Battery Depleted | Battery is depleted and system will shut down in 1 minute if not plugged in | Battery depletion detected | critical |
Fault: Cable Port 1 – Communication Error | Cable not inserted properly or connection points damaged | Cable port 1 communication failure | error |
Fault: Cable Port 1 – Hardware Failure | Cable is not inserted properly or connection points are damaged | Cable 1 insertion failure or damaged connectors | error |
Fault: Cable Port 1 - Incompatible Software Version | Unsuccessful software upgrade or incompatible software version in cable port 1 | Software upgrade failure or version incompatibility in cable port 1 | error |
Fault: Cable Port 1 – Software Failure | Software error with cable in port 1 | Software error detected in cable port 1 | error |
Fault: Cable Port 2 – Communication Error | Cable not inserted properly or connection points damaged | Cable port 2 communication failure | error |
Fault: Cable Port 2 – Hardware Failure | Cable is not inserted properly or connection points are damaged | Cable 2 insertion failure or damaged connectors | error |
Fault: Cable Port 2 - Incompatible Software Version | Unsuccessful software upgrade or incompatible software version in cable port 2 | Software upgrade failure or version incompatibility in cable port 2 | error |
Fault: Cable Port 2 – Software Failure | Software error with cable in port 2 | Software error detected in cable port 2 | error |
Fault: Cable Port <#> – Pressure Cable Disconnected | Pressure cable disconnected or not detected during monitoring | Pressure cable disconnected or not detected | error |
Fault: HemoSphere ClearSight Module | Defective HemoSphere ClearSight module | ClearSight module malfunction detected | error |
Fault: HemoSphere ClearSight Module Disconnected | HemoSphere ClearSight module removed or not detected | ClearSight module disconnected or not detected during monitoring | error |
Fault: HIS Connectivity Loss | Loss of HL7 communication | HL7 communication loss detected | warning |
Fault: Internal System Failure | Internal system malfunction | Internal system malfunction detected | critical |
Fault: L-Tech Module Slot – Communication Error | Large technology module not inserted properly or connection points damaged | L-Tech module communication failure | error |
Fault: L-Tech Module Slot – Hardware Failure | Large technology module is not inserted properly or connection points are damaged | L-Tech module insertion failure or damaged connectors | error |
Fault: L-Tech Module Slot – Incompatible Software Version | Unsuccessful software upgrade or incompatible software version in L-Tech slot | Software upgrade failure or version incompatibility in L-Tech module slot | error |
Fault: L-Tech Module Slot – Software Failure | Software error with large technology module | Software error detected in L-Tech module slot | error |
Fault: Module Slot 1 – Communication Error | Module 1 not inserted properly or connection points damaged | Module 1 communication failure | error |
Fault: Module Slot 1 – Hardware Failure | Module 1 is not inserted properly or connection points are damaged | Module 1 insertion failure or damaged connectors | error |
Fault: Module Slot 1 - Incompatible Software Version | Unsuccessful software upgrade or incompatible software version in slot 1 | Software upgrade failure or version incompatibility in module slot 1 | error |
Fault: Module Slot 1 – Software Failure | Software error with module in slot 1 | Software error detected in module slot 1 | error |
Fault: Module Slot 2 – Communication Error | Module 2 not inserted properly or connection points damaged | Module 2 communication failure | error |
Fault: Module Slot 2 – Hardware Failure | Module 2 is not inserted properly or connection points are damaged | Module 2 insertion failure or damaged connectors | error |
Fault: Module Slot 2 - Incompatible Software Version | Unsuccessful software upgrade or incompatible software version in slot 2 | Software upgrade failure or version incompatibility in module slot 2 | error |
Fault: Module Slot 2 – Software Failure | Software error with module in slot 2 | Software error detected in module slot 2 | error |
Fault: Monitor – Incompatible Software Version | Unsuccessful software upgrade or incompatible software version detected | Software upgrade failure or version incompatibility at monitor level | error |
Fault: Oximetry Cable Disconnected | Oximetry cable connection not detected at monitor | Oximetry cable connection not detected | error |
Fault: Pressure-Out – Hardware Failure | Pressure-out cable is not properly connected or damaged | Pressure-out cable connection failure or damage | error |
Fault: Second CO Pressure Sensor Detected | Multiple pressure cables with CO sensor connections detected | Detection of more than one CO pressure sensor connection | error |
Fault: Second Oximetry Cable Detected | Multiple oximetry cable connections detected | Detection of more than one oximetry cable connection | error |
Fault: Second Swan-Ganz Module Detected | Multiple Swan-Ganz module connections detected | Detection of more than one Swan-Ganz module connection | error |
Fault: Swan-Ganz Module Disconnected | HemoSphere Swan-Ganz module removed or not detected | Swan-Ganz module disconnected during monitoring or not detected | error |
Fault: System Temperature Too High - Shutdown Imminent | Internal temperature at critically high level, shutdown imminent | Internal temperature reaches critical level | critical |
Incorrect password entered | Password entered is incorrect | User enters incorrect password | warning |
Module Status Blue LED | Channels associated with port B on HemoSphere tissue oximetry module | Module configured for port B | info |
Module Status Green LED | Channels associated with port A on HemoSphere tissue oximetry module | Module configured for port A | info |
Please enter valid date | Date entered is invalid | User enters invalid date format | info |
Please enter valid time | Time entered is invalid | User enters invalid time format | info |
Service Battery | Internal battery fault occurred | Battery fault detected | warning |
Value must be ≥ xx | Entered value is lower than the low value setting | User enters value below minimum setting | info |
Value must be ≤ xx | Entered value is higher than the high value setting | User enters value above maximum setting | info |
Value out of range (xx-yy) | Entered value is outside the allowed range | User enters value higher or lower than allowed range | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI