Dynatronics Solaris Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 7 documented symptoms for the Dynatronics Solaris Plus, with 11 ranked causes, 13 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Treatment cannot be set up or delivered with error message displayed
Machine displays error message and will not allow treatment setup or delivery
Ranked causes
- 1. Soundhead disconnected from machine (common)
- 2. Soundhead damaged (occasional)
- 3. Soundhead temperature too high (occasional)
- 4. Moisture condensation in device (occasional)
Verify soundhead connection
- Test point
- soundhead port
- Expected
- soundhead firmly connected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead not detected during treatment setup or delivery
Error message 'SOUNDHEAD IS NOT CONNECTED, HEAD WARMING WILL BE DISABLED!' appears in Treatment Display Screen
Ranked causes
- 1. Soundhead not firmly plugged into connector (common)
- 2. Soundhead disconnected or damaged (occasional)
Verify soundhead connector is fully seated
- Test point
- soundhead connector port
- Expected
- soundhead firmly plugged in
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carbon pad does not conduct electricity
Carbon pad shows excessive resistance; electrical contact through pad is poor or absent
Ranked causes
- 1. Carbon pad is degraded or worn (common)
Set DMM to Ohms resistance setting
- Test point
- DMM selector
- Expected
- Ohms mode selected
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Combination therapy underwater patient safety
Patient safety hazard when performing combination therapy underwater with active electrodes
Ranked causes
- 1. Active electrodes placed in water during combination therapy (unknown)
Verify treatment type and electrode placement
- Test point
- treatment setup
- Expected
- combination therapy not used underwater; electrodes not submerged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead wire does not conduct electricity properly
Lead wire shows poor or no electrical conductance; no stimulation felt when testing
Ranked causes
- 1. Lead wire has intermittent or continuous open circuit (common)
Insert lead wire into CH 1 output jack
- Test point
- CH 1 output jack
- Expected
- Lead wire fully inserted
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead temperature too high during whirlpool treatment caution
Overheated soundhead caution appears in Treatment Display Screen when treating in hot whirlpool at approximately 103°F
Ranked causes
- 1. Whirlpool water temperature too high (common)
Verify whirlpool water temperature
- Test point
- whirlpool
- Expected
- below 103°F
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead too cold to operate
Keypad does not respond to key presses and device cannot be used when soundhead has been in cold room or vehicle
Ranked causes
- 1. Soundhead temperature below 60°F (common)
Measure or verify soundhead temperature
- Test point
- soundhead
- Expected
- approximately 60°F or higher
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
SOUNDHEAD IS NOT CONNECTED, HEAD WARMING WILL BE DISABLED! | Device cannot detect soundhead during treatment setup or delivery | Soundhead not properly connected to console or missing | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI