Dynatronics Solaris 700: troubleshooting and error codes
The ServiceForge knowledge base holds 63 documented symptoms for the Dynatronics Solaris 700, with 135 ranked causes, 151 diagnostic checks with test points and expected values, and 8 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Poor conductance during electrotherapy treatment patient safety
Bar graph shows decreasing green segments during treatment; reduced current delivery to patient; risk of increased current density at small point under electrode causing patient discomfort
Ranked causes
- 1. Electrodes worn or self-adhesive electrodes lost adhesiveness (common)
- 2. Poly adhesive electrode surface not fully adhering (common)
- 3. Self-adhesive electrodes not clean or hydrated (common)
- 4. Snap adapters disconnected or lead wire disconnected (occasional)
- 5. Carbon electrode loose connection with pin ends of leads (occasional)
- 6. Corrosion on lead ends (occasional)
- 7. Carbon electrodes not moistened or have build-up (occasional)
- 8. Poor body area conductance or high skin resistance (occasional)
- 9. Patient skin dry or electrodes loose from patient movement (occasional)
Inspect electrodes for wear or loss of adhesiveness
- Test point
- Self-adhesive or carbon electrodes
- Expected
- Electrodes intact and adhesive present
9 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient experiences pain during ultrasound treatment patient safety
Patient reports pain, tingling, excess heat, aching, or burning sensation during treatment
Ranked causes
- 1. Soundhead size inappropriate for treatment area (common)
- 2. Treatment intensity too high (common)
- 3. Treatment time too long (common)
- 4. Soundhead moved too slowly across skin (common)
- 5. Soundhead held stationary in one location (common)
- 6. Frequency too low for treatment depth (occasional)
- 7. Patient has sensory nerve damage or desensitized skin (occasional)
- 8. Unit not properly calibrated (occasional)
Verify soundhead size matches treatment area
- Test point
- Device display or soundhead specification
- Expected
- 2cm² for small areas, 5cm² for medium areas, 10cm² for large areas
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Burns from improper ultrasound therapy technique patient safety
Risk of thermal burns to patient skin if therapy parameters or technique not administered correctly
Ranked causes
- 1. Intensity (power) too high (common)
- 2. Frequency too low causing excessive penetration and heating (occasional)
- 3. Soundhead held stationary creating hot spots (common)
- 4. Soundhead moved too slowly over treatment area (common)
- 5. Excessive treatment time (common)
- 6. Bony prominences reflecting sound waves increasing intensity (common)
Verify intensity setting before starting treatment
- Test point
- Intensity control display
- Expected
- Appropriate W/cm² for condition stage: acute 0.1-0.5, sub-acute 0.5-1.0, chronic 1.0-2.0
6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient discomfort during treatment patient safety
Patient experiences discomfort and skin reactions during electrotherapy
Ranked causes
- 1. Worn or damaged lead wires (common)
- 2. Poor quality or inappropriate electrodes (common)
- 3. Dry carbon electrodes (common)
- 4. Carbon electrode build-up from body oils or moistening agent (common)
- 5. Self-adhesive electrode loss of adhesion (occasional)
Test lead wires for continuity
- Test point
- Lead pins
- Expected
- Tone stops when pins held together; tone continues if lead is broken
- Tools
- Dynatron Solaris device Lead Test function
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient skin rash, burn, or blister from electrotherapy patient safety
Adverse skin reaction including rash, burn, or blister occurring during or after electrical stimulation treatment
Ranked causes
- 1. High current density from small electrode size (common)
- 2. Excessive treatment intensity (common)
- 3. Worn or damaged electrodes (common)
- 4. Patient skin contamination (dirt, oils, powders) (occasional)
- 5. Patient susceptibility to skin irritation (occasional)
Verify electrode size selected for treatment area
- Test point
- Electrode surface area
- Expected
- Largest practical electrode for application; reference chart for current density
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
General error message in LED display (non-Temperature related)
Device displays alphanumeric error message during setup or treatment delivery, preventing normal operation. May be caused by disconnected soundhead, condensation, or internal issues.
Ranked causes
- 1. Soundhead disconnected during operation (common)
- 2. Soundhead damaged by drop or impact (occasional)
- 3. Moisture condensation inside device (rare)
- 4. Internal device malfunction requiring service (rare)
Stop treatment and power cycle device
- Test point
- Power switch; LED display
- Expected
- Error message clears after 5-10 second power-off interval
- Tools
- Timer
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-adhesive electrode loses adhesion during or after treatment patient safety
Electrode does not adhere to patient skin or falls off during treatment
Ranked causes
- 1. Electrode stored improperly or exposed to air (common)
- 2. Patient skin not clean or has oils/flakiness (common)
- 3. Electrode used for more than 15 treatments (common)
- 4. Electrode secured with straps or weights (rare)
Inspect patient skin condition before electrode application
- Test point
- Electrode contact area on patient skin
- Expected
- Skin clean, dry, free from oils, flakiness, and debris
- Tools
- Visual and tactile inspection
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead crystal or electronic component damage caution
Crystal stress, thermal shock, or mechanical damage to soundhead from improper handling or cooling; may result in loss of coupling or treatment effectiveness
Ranked causes
- 1. Dropping soundhead on hard surfaces (rare)
- 2. Using ice water or ice packs for cooling (occasional)
- 3. Repeated overheating of soundhead (occasional)
- 4. Holding soundhead in air during treatment (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treatment selection based on injury stage and condition patient safety
Selection of appropriate frequency, intensity, duration and treatment frequency based on acute, sub-acute, or chronic condition stage
Ranked causes
- 1. Acute conditions require specific lower intensity settings (common)
- 2. Sub-acute conditions require moderate intensity (common)
- 3. Chronic conditions require higher intensity for thermal effect (common)
- 4. Maximal thermal effect requires specific formula calculation (occasional)
Classify injury or condition stage
- Test point
- Patient history and clinical assessment
- Expected
- Documented as acute, sub-acute, or chronic condition
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequency selection based on lesion depth penetration patient safety
Selection of 1 MHz, 2 MHz, or 3 MHz frequency based on tissue depth and desired penetration distance
Ranked causes
- 1. Deep lesions require 1 MHz frequency (common)
- 2. Moderate depth lesions require 2 MHz frequency (common)
- 3. Superficial lesions require 3 MHz frequency (common)
Determine tissue depth of lesion or treatment area
- Test point
- Clinical assessment and tissue density evaluation
- Expected
- Classified as deep (>5cm), moderate (2-5cm), or superficial (<2cm)
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No current delivery when probe ON/OFF button pressed
Active MultiStim probe fails to deliver current to patient when ON/OFF button is pressed during probe treatment
Ranked causes
- 1. Ground probe or ground electrode not properly connected (common)
- 2. Cotton swab not contacting metal ring at probe tip (common)
- 3. Cotton swab too dry to conduct current (common)
Verify ground probe or ground electrode is connected
- Test point
- Banana plug connection at back of MultiStim probe or adapter
- Expected
- Secure connection to ground probe or ground electrode
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No patient current during Interferential treatment setup patient safety
Device displays LEAD warning with channel number; beeps and prevents intensity increase; open or incomplete circuit detected
Ranked causes
- 1. Lead wire not plugged into device (occasional)
- 2. Electrode disconnected from patient or not making contact (occasional)
- 3. Lead Test sensitivity set too high (rare)
Verify all lead wires are plugged into device channels
- Test point
- Channel 1, 2, 3, 4 jacks
- Expected
- All leads for treatment channels are fully seated in jacks
- Tools
- Visual inspection
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pad connector attachment issues caution
Difficulty inserting or removing keyed pad connectors; improper connector orientation
Ranked causes
- 1. Pad connector not aligned with keyed jack openings (common)
- 2. Connector not released before removal attempt (occasional)
- 3. Excessive force applied during insertion or removal (occasional)
Verify pad connector alignment with keyed jack openings
- Test point
- Connector and keyed openings in jack
- Expected
- Connector aligned with keyed openings; ready for insertion
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ultrasonic coupling to patient patient safety
Insufficient energy delivery from soundhead to patient due to air gaps, inadequate gel, or improper contact; indicated by low coupling bar graph segments and decreased treatment effectiveness
Ranked causes
- 1. Inadequate conductive gel application (common)
- 2. Air gaps or incomplete soundhead-skin contact (common)
- 3. Air in whirlpool or water treatment (occasional)
Verify adequate gel application
- Test point
- Treatment site and soundhead surface
- Expected
- Ample conductive gel visible between soundhead and skin
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin irritation at electrode placement site patient safety
Skin irritation or burns may occur beneath electrodes during or after microcurrent treatment, particularly with long-term application or certain settings.
Ranked causes
- 1. Poor electrode quality combined with high current setting (occasional)
- 2. High current density from small electrode delivery area (occasional)
- 3. Patient hypersensitivity to electrical conductive medium (occasional)
Assess current setting and electrode condition
- Test point
- Device control panel and electrode surfaces
- Expected
- Current setting appropriate for patient tolerance; electrodes clean and undamaged
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin irritation or rash during electrotherapy patient safety
Patient experiences skin irritation, rash, burn, or blister during or after electrical stimulation treatment
Ranked causes
- 1. High current density from worn or dried out electrodes (common)
- 2. Excessive treatment intensity (common)
- 3. Patient skin sensitivity to electrotherapy (occasional)
Inspect electrode condition
- Test point
- Electrode surface
- Expected
- Electrodes clean, not worn or dried out
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treatment ineffective or suboptimal patient safety
Patient reports reduced pain relief, muscle relaxation, or improved blood circulation after treatment completion
Ranked causes
- 1. Probe not maintaining full contact with skin during treatment (common)
- 2. Gels used with light therapy probe (common)
- 3. Treatment area larger than probe aperture not fully covered (occasional)
Verify skin contact area is clean and free of gels
- Test point
- Treatment skin surface and probe aperture
- Expected
- Skin and probe both cleaned; no visible gel or lotion residue
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BATT BAD message displayed and device shuts down
Device displays BATT BAD message, treatment intensity ramps down, and machine shuts off. Indicates either a defective battery or battery is fully depleted.
Ranked causes
- 1. Battery fully depleted (common)
- 2. Battery pack is defective (occasional)
Attempt to recharge battery
- Test point
- Battery charger connection
- Expected
- Battery accepts charge and device resumes normal operation after 24-hour charge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current limit warning triggered during treatment patient safety
Device beeps several times, flashes intensity display, and automatically reduces intensity a few increments when maximum output current limit is reached
Ranked causes
- 1. Patient unable to adequately feel current due to reduced sensation (common)
- 2. Multiple large electrodes dispersing current over larger surface area (occasional)
Confirm patient sensation in treatment area
- Test point
- Treatment electrode site
- Expected
- Patient reports feeling current sensation
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 13 message displayed
Error 13 appears in LED display when soundhead is plugged in while device is powered on, or due to connector detection issues.
Ranked causes
- 1. Soundhead connected while device powered on (common)
- 2. Loose or intermittent soundhead connection (occasional)
Verify proper soundhead connection sequence
- Test point
- Soundhead connector at time of device power-on
- Expected
- Soundhead must be connected BEFORE powering device on
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inability to start combination treatment
Device display flashes and treatment will not start
Ranked causes
- 1. Stim intensity not set (common)
- 2. Ultrasound power not set (common)
Verify stim intensity is set
- Test point
- Stim intensity control and display
- Expected
- Non-zero intensity value displayed; display not flashing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient discomfort during microcurrent treatment patient safety
Microcurrent at high intensity settings or during probe treatment can be felt by patient and may be uncomfortable, result of current concentrated at small area
Ranked causes
- 1. High current density from high intensity settings (common)
- 2. High current density from small electrode area or probe tip (common)
Verify current intensity setting
- Test point
- Intensity control
- Expected
- Moderate setting per published guidelines
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient does not feel stim current through soundhead
No current sensation reported by patient during stim through soundhead treatment
Ranked causes
- 1. Poor soundhead-to-skin coupling (common)
- 2. Insufficient conductive gel (common)
Verify soundhead contact with patient skin
- Test point
- Soundhead surface at treatment site
- Expected
- Soundhead in full contact with skin
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient experiences heat rash or skin irritation patient safety
Patient develops reaction similar to heat rash or skin sensitivity response to ultrasound output
Ranked causes
- 1. Patient has sensitive skin to ultrasound output (occasional)
- 2. Bony prominences reflecting sound waves (common)
Assess patient skin sensitivity and condition
- Test point
- Patient interview and skin inspection
- Expected
- Document baseline skin condition and sensitivity level
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe connector attachment issues caution
Difficulty inserting or removing keyed probe connectors; risk of pin damage
Ranked causes
- 1. Keyed connector not properly aligned with jack notch (common)
- 2. Connector forced during insertion or removal (occasional)
Verify keyed connector alignment before insertion
- Test point
- Raised portion of connector and notched jack opening
- Expected
- Raised portion aligned with notch; connector slides smoothly into jack
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe overheats during treatment caution
Word 'HOT' appears in time display window and treatment automatically pauses until internal probe temperature returns to normal
Ranked causes
- 1. Treatment head vents obstructed (common)
- 2. Gels or lotions clogging internal vents (common)
Inspect probe vents for obstructions
- Test point
- Vents surrounding treatment head
- Expected
- Vents completely clear and unobstructed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin irritation or hypersensitivity from electrical medium patient safety
Patient experiences skin irritation or hypersensitivity reaction to electrical conductive medium or electrical stimulation itself.
Ranked causes
- 1. Sensitivity to conductive medium material (occasional)
- 2. Sensitivity to electrical stimulation at current location (occasional)
Inspect skin for erythema or reaction signs
- Test point
- Area under electrodes
- Expected
- No visible redness, swelling, or rash
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead not detected by device
Device displays HEAD error message during setup or delivery of ultrasound treatment, indicating soundhead is not plugged in or not being detected.
Ranked causes
- 1. Soundhead not firmly plugged into connector (common)
- 2. Soundhead malfunction or damage (occasional)
Verify soundhead is firmly connected
- Test point
- Soundhead connector at ultrasound output jack
- Expected
- Connector fully inserted; no gaps visible
- Tools
- Visual inspection
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead overheats during treatment patient safety
Soundhead temperature approaches or exceeds 103°F, GREEN segments on coupling graph replaced by CLEAR segments, POWER display flashes, treatment automatically paused at 108°F with "HOT" message displayed
Ranked causes
- 1. Poor coupling between soundhead and patient skin (common)
- 2. Insufficient power reduction for difficult coupling areas (occasional)
Monitor coupling bar graph during treatment
- Test point
- COUPLING BAR GRAPH display under POWER/INTENSITY
- Expected
- All twelve segments lighted GREEN
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead temperature too cold at startup
Soundhead has been exposed to cold environment. Keypad does not respond and device will not recognize soundhead until heated to minimum operating temperature (~60°F).
Ranked causes
- 1. Soundhead exposed to cold environment (occasional)
- 2. Keypad temperature lock (occasional)
Check if keypad responds to button presses
- Test point
- Any key on device keypad
- Expected
- Keys respond; display shows normal response
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead temperature too high during treatment patient safety
Soundhead temperature rises due to poor coupling, causing reduced ultrasonic output to patient and risk of crystal overheating. POWER display flashes as warning; treatment auto-pauses when maximum temperature reached.
Ranked causes
- 1. Poor acoustic coupling during treatment (common)
- 2. Whirlpool treatment temperature too high (occasional)
Observe POWER/INTENSITY LED display for flashing
- Test point
- LED display during treatment
- Expected
- Solid illumination
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead uncoupling from patient skin caution
Soundhead loses contact with patient skin during treatment; reduced treatment effectiveness and soundhead overheating
Ranked causes
- 1. Insufficient or poor quality coupling agent (common)
- 2. Movement too fast across treatment area (occasional)
Inspect coupling agent quality and quantity
- Test point
- Between soundhead and patient skin
- Expected
- Sufficient coupling agent present with no visible air bubbles
- Tools
- Visual inspection, coupling gel
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treatment timer not counting as expected
For probe treatments, timer does not count up in seconds or resets unexpectedly; for electrode treatments, timer does not count down
Ranked causes
- 1. Probe current not activated or stopped between treatments (common)
- 2. Electrode treatment timer set to zero (occasional)
Verify probe ON/OFF button is pressed to start current for probe treatments
- Test point
- MultiStim probe ON/OFF button on handle
- Expected
- One beep and timer counting up from zero in seconds after button press
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BATT LO error when battery is not actually low
Device displays BATT LO message indicating low battery when battery charge level is adequate for continued operation.
Ranked causes
- 1. Battery adapter cord is low-quality gauge wire (occasional)
Inspect battery adapter cord gauge and quality
- Test point
- Battery adapter cord connector pins and wire
- Expected
- High-quality gauge wire with 5 amp fast blow fuse installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
BATT LO message displayed
Device displays BATT LO message indicating low battery power during treatment. Treatment can continue but setup of new treatment may not be possible when current treatment ends.
Ranked causes
- 1. Battery charge depleted below operational threshold (common)
Monitor battery charge level during operation
- Test point
- Device display
- Expected
- BATT LO message indicates low battery condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue light exposure hazard with D405 probe patient safety
Operator or patient exposed to blue light wavelength 405nm without proper eye protection
Ranked causes
- 1. Protective eyewear not worn during D405 probe treatment (common)
Verify protective eyewear available and properly fitted
- Test point
- Patient and practitioner eye protection
- Expected
- Protective eyewear in place on both patient and practitioner before D405 treatment begins
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Soundhead parameters entry and verification
Procedure to enter soundhead calibration parameters into device after adding or replacing soundhead
Ranked causes
- 1. New or replaced soundhead requires parameter entry (common)
Obtain Head Calibration Printout for soundhead
- Test point
- Documentation
- Expected
- Printout sheet with soundhead serial number matching plugged-in soundhead
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ultrasound soundhead power output caution
Routine calibration procedure to verify and adjust ultrasound power output at 1, 2, and 3 MHz frequencies using an external power meter.
Ranked causes
- 1. Scheduled maintenance calibration (common)
Verify equipment and water quality
- Test point
- Power meter; degassed water supply
- Expected
- Ultrasound power meter capable of measuring up to 3 MHz; degassed water with oxygen content of 4 ppm or less
- Tools
- Ohmic Instrument UPM-DT1 or UPM-DT-10 power meter; Dynatronics Soundhead Calibration Printout
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac fibrillation from thoracic electrotherapy patient safety
Risk of cardiac fibrillation during electrotherapy treatment applied to thoracic region
Ranked causes
- 1. Electrotherapy applied across thoracic region (common)
Verify treatment location is not thoracic region
- Test point
- Electrode placement area on patient
- Expected
- Electrodes placed on non-thoracic area
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac fibrillation risk from thoracic application patient safety
Potential for cardiac fibrillation if electrical stimulation applied to thoracic region with high output current
Ranked causes
- 1. Thoracic electrotherapy application with high current output (rare)
Confirm treatment area is not thoracic
- Test point
- Patient electrode placement
- Expected
- Electrodes applied to non-thoracic area only
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel 1 unavailable during probe treatment
Channel 1 jack cannot be used for electrode treatment while any microcurrent probe treatment is in progress
Ranked causes
- 1. Probe treatment is active (common)
Verify probe treatment is stopped before selecting Channel 1
- Test point
- CHANNEL 1 LED indicator status and probe ON/OFF button status
- Expected
- CHANNEL 1 LED is GREEN (not YELLOW) indicating no probe treatment in progress
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Combined stimulation device with ultrasound modality requires precautions patient safety
When using simultaneous electrical stimulation output through ultrasound soundhead, additional contraindications and warnings apply
Ranked causes
- 1. Combination therapy requires dual modality contraindication review (common)
Review patient contraindications for ultrasound therapy
- Test point
- Patient history and clinical notes
- Expected
- No ultrasound contraindications documented
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Conductance erratic during probe treatment
Cotton swab dries out during microcurrent probe treatment, causing conductance to become unpredictable or unstable
Ranked causes
- 1. Dry cotton swab at probe tip (common)
Observe if conductance bar fluctuates or becomes unstable
- Test point
- Conductance bar graph on display
- Expected
- Stable conductance reading
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Direct current default parameters setup
Configuration of default treatment settings for frequently used DC therapy protocols to streamline future treatments
Ranked causes
- 1. Need to save custom default settings (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Direct current treatment setup and operation patient safety
Operator needs to configure and deliver DC stimulation therapy using MultiStim probe with adjustable pulse duration, polarity, and intensity up to 20 mA
Ranked causes
- 1. Standard DC treatment procedure (common)
Verify pulse duration setting
- Test point
- TIME display window
- Expected
- 0.1, 0.3, 0.5, 1, 10, 100, or 500 mSec per pulse with 500 mSec interpulse duration
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hydrogel disk method for wound treatment patient safety
Alternative delivery method using hydrogel disk to protect wounds from direct soundhead contact during sonation
Ranked causes
- 1. Need to treat crater wound without direct soundhead contact (occasional)
Verify hydrogel disk is clean and intact
- Test point
- Visual inspection
- Expected
- Hydrogel disk present, clean, and without damage or contamination
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injury from electrode or lead removal during active treatment patient safety
Sudden disconnection of electrodes or leads during treatment causing unexpected stimulus cessation or patient discomfort
Ranked causes
- 1. Electrodes or leads removed while treatment active (common)
Confirm treatment is stopped before electrode/lead modification
- Test point
- Treatment output status
- Expected
- Treatment output is OFF/stopped
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injury from power switch activation while electrodes connected patient safety
Risk of sudden, unexpected electrical stimulation to patient if power is turned on or off while patient is connected to device
Ranked causes
- 1. Power switched while patient connected (common)
Verify all leads and electrodes are disconnected from patient
- Test point
- Lead/electrode connections to patient
- Expected
- No leads or electrodes attached to patient skin
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injury from worn or damaged electrode/lead accessories patient safety
Use of defective or deteriorated electrode leads causing inadequate electrical contact or patient injury
Ranked causes
- 1. Worn or damaged leads or electrodes in use (common)
Inspect leads and electrodes for damage before treatment
- Test point
- Lead insulation, electrode adhesion, connectors
- Expected
- All leads intact with no cracks or exposed wiring; electrodes undamaged with good adhesion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No beep or audio feedback when probe ON/OFF button pressed
Device does not beep or produce audible tone when starting or stopping microcurrent probe treatment
Ranked causes
- 1. Audible tone is turned off (common)
Verify audible tone is enabled
- Test point
- Audible tone status during probe treatment
- Expected
- Beep when ON/OFF button pressed; no audible tone means feature is OFF for electrode treatments or disabled for probe treatments
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient cannot feel microcurrent stimulation
At very low frequencies typical of microcurrent treatment, current is commonly not felt by patient
Ranked causes
- 1. Low frequency and microamp current levels inherent to microcurrent modality (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient injury from damaged leads or electrodes patient safety
Risk of patient skin injury or electrical hazard from worn, damaged, or degraded electrodes and leads
Ranked causes
- 1. Worn or damaged leads or electrodes applied to patient (common)
Inspect all leads and electrodes
- Test point
- Leads and electrode surfaces
- Expected
- No visible wear, cracks, discoloration, or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient injury from power state change during connection patient safety
Risk of patient harm if unit power is turned on or off while electrodes or leads are connected to patient
Ranked causes
- 1. Power switched on or off with patient connected (occasional)
Verify unit power state before electrode attachment
- Test point
- Power switch
- Expected
- Unit powered off
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe or pad not recognized
Error message 'Irhd' appears in time display window when device powers on, indicating probe/pad not properly connected or detected
Ranked causes
- 1. Probe or pad not properly inserted into output jack (common)
Verify probe/pad fully inserted into output jack
- Test point
- IR/RED probe jack or Xp pad jack on Booster Box or base console
- Expected
- Connector fully seated with no gaps
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduced probe output power caution
Probe delivers less therapeutic energy than expected, reduced optical output during treatment
Ranked causes
- 1. Gel or lotion residue on probe lens optical coating (common)
Inspect probe aperture lens for residue or coating degradation
- Test point
- Probe treatment head lens surface
- Expected
- Lens clear and transparent with no visible residue or discoloration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead heat or pause during combination therapy caution
Treatment pauses when soundhead becomes too hot; stim current continues while ultrasound and timer pause
Ranked causes
- 1. Soundhead thermal limit reached (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead not properly connected to device caution
Soundhead is not firmly plugged in before device power-on, resulting in no treatment delivery or intermittent operation
Ranked causes
- 1. Soundhead connection loose or not fully inserted (common)
Verify soundhead connection before powering device
- Test point
- Soundhead connector port on device
- Expected
- Soundhead firmly plugged in and fully seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Soundhead temperature warning at 103°F in whirlpool treatment patient safety
POWER/INTENSITY LED flashes during whirlpool treatment when soundhead reaches 103°F, indicating temperature approaching limit. This is a warning only; treatment may continue at this level.
Ranked causes
- 1. Whirlpool water temperature higher than normal treatment environment (common)
Monitor POWER/INTENSITY LED during whirlpool treatment
- Test point
- LED display flashing behavior
- Expected
- LED steady; no flashing at 103°F or below
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treatment area size guideline for ultrasound dose management
Area covered by ultrasound treatment affects total dosage received; larger areas receiving same intensity and time distribute dose over more tissue
Ranked causes
- 1. Treatment area size affects ultrasound dosage distribution (common)
Measure treatment area size
- Test point
- Area assessment (cm²)
- Expected
- Area size documented and appropriate for selected soundhead
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrasound unit requires external calibration verification
Device requires calibration every six months to one year to maintain accuracy and ensure proper operation of all soundheads
Ranked causes
- 1. Scheduled calibration maintenance due (common)
Verify last calibration date
- Test point
- Device maintenance log or records
- Expected
- Calibration performed within last 12 months for optimal performance
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Unsafe treatment without remote stop cable provided to patient
- Water immersion method for ultrasound treatment requires proper coupling
- Xp Pad structural damage during use
3 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
BATT BAD | Battery is either defective or fully depleted | Battery power drops below minimum threshold to operate device | error |
BATT LO | Low battery power indication | Available battery power reduced to operational threshold | warning |
ERROR 13 | Soundhead connection error during power-on | Occurs when soundhead is plugged in while device is powered on, or due to intermittent connection detection | error |
HEAD | Soundhead not detected by device | Device cannot detect soundhead during setup or delivery of ultrasound treatment | error |
HEAD 3 ERROR | Soundhead parameters not saved during calibration | Occurs at end of calibration procedure when one or more head parameters fail to save to device memory | error |
HOT | Soundhead temperature exceeds safe operating limit | Soundhead temperature approaches or exceeds 108 degrees Fahrenheit during ultrasound treatment | critical |
Irhd | Probe or pad hardware detection error | Probe or Xp Pad not properly inserted into output jack when device powers on | error |
LEAD | Open or incomplete circuit detected during Interferential treatment setup or high-frequency Premodulated treatment | Device detects open circuit; lead not plugged in, electrode fallen off, electrode not making skin contact, or any condition causing incomplete circuit | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.