DRE Medical Waveline Pro: troubleshooting and error codes
The ServiceForge knowledge base holds 33 documented symptoms for the DRE Medical Waveline Pro, with 47 ranked causes, 44 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Keyboard abnormal message displayed patient safety
Error message 'Keyboard abnormal' appears on schedule bar during monitoring software operation after power-up.
Ranked causes
- 1. Main board PC104 is bad (occasional)
- 2. Keyboard is bad (occasional)
- 3. Control board is bad (occasional)
Check keyboard response and connections
- Test point
- Keyboard connector and keys
- Expected
- Keyboard responds to key presses without error message
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 and pulse rate values displayed patient safety
SpO2 probe connected but no oxygen saturation or pulse rate values are shown on display.
Ranked causes
- 1. SpO2 probe not connected properly (common)
- 2. Red LED light not illuminating in SpO2 probe (occasional)
- 3. SpO2 probe is bad (occasional)
Verify SpO2 probe connection to monitor
- Test point
- SpO2 probe connector
- Expected
- Probe fully seated with good contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC indicator illuminated but display is black
Power indicator lights but TFT screen remains black with no output. Fan may or may not operate depending on power source.
Ranked causes
- 1. Display converter module is bad (common)
- 2. Power supply module is bad (occasional)
Ensure fan operates when AC power is connected
- Test point
- Rear panel fan outlet
- Expected
- Fan running with audible airflow when AC powered
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery capacity exhausted
Battery voltage drops below 7V, causing permanent battery damage. Occurs due to power supply module energy waste on other circuits.
Ranked causes
- 1. Extended non-use without AC power connection (common)
- 2. Power supply module not connected to AC for extended period (common)
Verify battery voltage
- Test point
- Battery terminals
- Expected
- Above 7V
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lost message displayed during monitoring patient safety
Error message 'ECG lost' displayed on message area during patient monitoring operation.
Ranked causes
- 1. ECG electrodes not attached properly (common)
- 2. ECG lead selection is incorrect (occasional)
Visually inspect ECG electrode placement on patient
- Test point
- Patient chest electrodes
- Expected
- Electrodes adhered well with good skin contact
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead off message displayed during monitoring patient safety
Error message 'lead off' displayed on message area during patient monitoring.
Ranked causes
- 1. ECG leads not attached properly (common)
- 2. ECG cable is bad (occasional)
Check ECG lead connections to monitor
- Test point
- ECG connector on monitor
- Expected
- All leads firmly seated in monitor connector
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC/104 Module cannot boot system
System fails to boot; PC/104 controller does not initialize
Ranked causes
- 1. Button battery depleted or faulty (common)
- 2. Incorrect resistor R74 installed (occasional)
Verify button battery condition
- Test point
- PC/104 front side PCB battery location
- Expected
- Battery present and functional
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform displays as baseline patient safety
Respiration waveform shows flat line (baseline) instead of normal undulating waveform during patient monitoring.
Ranked causes
- 1. RA and LL electrodes not attached properly (common)
- 2. ECG cable is bad (occasional)
Visually inspect RA and LL electrode placement
- Test point
- Patient RA and LL electrodes
- Expected
- Electrodes well adhered with no lifting
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal weak message during NBP measure patient safety
Error message 'Signal weak' displayed during non-invasive blood pressure measurement indicating inadequate signal.
Ranked causes
- 1. Cuff not attached tightly (common)
- 2. Patient arm position suboptimal (occasional)
Verify cuff is attached securely to patient arm
- Test point
- NBP cuff around arm
- Expected
- Cuff snug but not causing discomfort or cutting off circulation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature parameter displays lost patient safety
Temperature parameter shows 'lost' or invalid reading during patient monitoring.
Ranked causes
- 1. Temperature probe not connected properly (common)
- 2. Temperature probe is bad (occasional)
Check temperature probe connector
- Test point
- Temperature probe connector on monitor
- Expected
- Probe connector fully seated and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TFT screen displays all white
Display shows white or blank screen after power-up, unable to display any graphics or text.
Ranked causes
- 1. Display cable disconnection (common)
- 2. Main board PC104 is bad (occasional)
Ensure display cable is connected well
- Test point
- Display connector on PC104 and display module
- Expected
- Cable fully seated and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up
Monitor does not respond to power switch. AC indicator may or may not be illuminated. No display or fan operation.
Ranked causes
- 1. Power supply module is bad (common)
- 2. Battery is bad (occasional)
Take out the battery
- Test point
- Battery compartment
- Expected
- Monitor powers on with external AC only
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air pressure error message during NBP measure patient safety
Error message 'Air pressure error' displayed during non-invasive blood pressure measurement.
Ranked causes
- 1. Blocked hoses in pneumatic system (common)
Inspect NBP tubing for blockage
- Test point
- Pneumatic hoses from pump to cuff
- Expected
- All hoses clear and properly routed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Airleak message displayed during NBP measure patient safety
Error message 'Airleak' displayed on screen during non-invasive blood pressure measurement indicating air loss in system.
Ranked causes
- 1. Blocked hoses in pneumatic system (common)
Inspect all NBP pneumatic hoses
- Test point
- NBP tubing from cuff to pump
- Expected
- Hoses open and unobstructed, not pinched or kinked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Beyond range message during NBP measure patient safety
Error message 'Beyond range' displayed during non-invasive blood pressure measurement indicating reading outside normal limits.
Ranked causes
- 1. Patient cuff or arm is in abnormal state (common)
Examine patient arm and cuff condition
- Test point
- Patient arm and NBP cuff
- Expected
- Arm appears normal, no swelling, injury, or deformity. Cuff intact.
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PC/104 Module boot configuration
Procedure to restore PC/104 Module boot functionality after component replacement
Ranked causes
- 1. Configuration requires button battery and resistor R73 (common)
Disassemble button battery
- Test point
- PC/104 front side PCB battery location (Figure 7-1)
- Expected
- Battery accessible and removable
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot enter monitoring interface after power-up patient safety
Display shows content but system does not respond to user input or fail to load monitoring software interface.
Ranked causes
- 1. Main board PC104 is bad (common)
Power-up monitor and wait for boot sequence
- Test point
- Monitoring interface display
- Expected
- Monitoring software interface appears and responds to input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff error message displayed during NBP measure patient safety
Error message 'Cuff Error' displayed on screen when attempting non-invasive blood pressure measurement.
Ranked causes
- 1. Cuff type is incorrect (common)
Verify NBP cuff type in setup menu
- Test point
- NBP cuff type setting
- Expected
- Cuff type matches physical cuff applied to patient
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERT communication error message displayed patient safety
Error messages 'ERT Com Error', 'RESP Com Error', or both displayed on message area in upper middle of screen.
Ranked causes
- 1. ERT communication trouble (common)
Verify ERT and RESP module cable connections
- Test point
- ERT/RESP module connectors
- Expected
- Cables fully seated and latched
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate does not match ECG waveform patient safety
Displayed heart rate value or pulse rate does not correspond to ECG or pulse waveform on display.
Ranked causes
- 1. HR/Pulse display setting is incorrect (common)
Access HR-Display setup menu
- Test point
- Monitor setup menu
- Expected
- 'Heart' or 'Pulse' setting matches displayed parameter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing R20 resistance on rear PC/104 controller PCB
Chip resistance R20 (4.7K ohm, 0603 package) is not assembled on the rear side PC/104 controller PCB.
Ranked causes
- 1. Manufacturing variance in pre-June 2005 units with non-EmCore-i313 PC104 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing R20 resistor on PC/104 controller PCB
Chip resistance R20 (4.7K ohm, 0603 package) is missing or not assembled on rear side PC/104 controller PCB at R20 location (Figure 7-3).
Ranked causes
- 1. R20 resistor not assembled during manufacturing (pre-June 2005 models) (common)
Inspect R20 location on rear PC/104 controller PCB
- Test point
- R20 location, Figure 7-3
- Expected
- 4.7K ohm resistor (0603 package) present and soldered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing resistor R20 on PC/104 controller PCB
Monitors manufactured before June 2005 without EmCore-i313 type PC104 may have missing chip resistance at R20 location
Ranked causes
- 1. Missing or defective R20 resistor on rear PC/104 controller PCB (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multiple monitoring module communication errors patient safety
Multiple error messages 'SpO2 Com Error', 'ERT Com Error', 'RESP Com Error', and 'TEMP Com Error' all displayed.
Ranked causes
- 1. Control module trouble (common)
Verify all module connections to control module
- Test point
- All module connectors
- Expected
- All cables connected and latched
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over motion message during NBP measure patient safety
Error message 'Over motion' displayed during non-invasive blood pressure measurement due to excessive patient arm movement.
Ranked causes
- 1. Patient arm is moving during measurement (common)
Observe patient arm movement during measurement
- Test point
- Patient arm position
- Expected
- Patient arm remains still without motion or muscle tension
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over pressure message during NBP measure patient safety
Error message 'Over pressure' displayed during non-invasive blood pressure measurement indicating excessive pressure buildup.
Ranked causes
- 1. Patient cuff or arm is in abnormal state (common)
Verify cuff size and fit on patient arm
- Test point
- NBP cuff and patient arm
- Expected
- Cuff appropriate size, not compressing excessively
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient type error message during NBP measure patient safety
Error message 'Patient type error' displayed on screen during non-invasive blood pressure measurement attempt.
Ranked causes
- 1. Cuff type is incorrect for patient (common)
Verify patient type and cuff match
- Test point
- NBP cuff type and patient type setting
- Expected
- Cuff type matches patient age/size category
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC/104 module will not boot system
System fails to boot from PC/104 controller module
Ranked causes
- 1. Battery failure or improper resistor configuration on PC/104 controller PCB (unknown)
Verify button battery is present and seated
- Test point
- Front side PC/104 controller PCB
- Expected
- Battery present and fully seated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal over message during NBP measure patient safety
Error message 'Signal over' displayed during non-invasive blood pressure measurement.
Ranked causes
- 1. Patient arm doing unwinding or winding action (common)
Observe if patient is rotating/winding arm
- Test point
- Patient arm motion during measurement
- Expected
- Patient arm stationary without rotational motion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 communication error message displayed patient safety
Error message 'SpO2 Com Error' displayed on message area in upper middle of screen after entering monitoring interface.
Ranked causes
- 1. SpO2 communication trouble (common)
Verify SpO2 module cable connection
- Test point
- SpO2 module connector
- Expected
- Cable fully seated with good contact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TFT displays NBP message continuously
Screen shows 'NBP' word continuously on display and cannot advance to normal monitoring interface.
Ranked causes
- 1. NBP communication trouble (common)
Verify NBP module cable connection
- Test point
- NBP module connector
- Expected
- Cable fully seated and latched
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TFT screen displays skewbald pattern
Display shows scrambled, garbled, or discolored pattern instead of normal graphics or text.
Ranked causes
- 1. Main board PC104 is bad (common)
Visual inspection of display pattern
- Test point
- TFT screen output
- Expected
- Corrupted or scrambled display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wasting energy on circuits when power switch off caution
Power supply module drains energy from battery when switched off due to parasitic current draw on other circuits
Ranked causes
- 1. Parasitic current draw in power supply circuits when switch off (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.