Dräger Resuscitaire Radiant Warmer: troubleshooting and error codes
The ServiceForge knowledge base holds 90 documented symptoms for the Dräger Resuscitaire Radiant Warmer, with 131 ranked causes, 171 diagnostic checks with test points and expected values, and 23 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Cannot obtain desired suction pressure caution
Suction/vacuum system pressure is lower than expected or cannot reach target value
Ranked causes
- 1. Suction jar rim gasket is worn (common)
- 2. Suction jar is loose or chipped (common)
- 3. Connecting hose is bent or tangled (common)
- 4. Hose is leaking between regulator and bottle (occasional)
- 5. Pressure or suction gauge not measuring correctly (rare)
Perform RAP 2.23 diagnostic and repair procedure
- Test point
- Suction system
- Expected
- Desired pressure obtained
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Breath rate timing patient safety
Breath rate timing does not match specification; expiratory or inspiratory time out of tolerance at 18 BPM or 60 BPM
Ranked causes
- 1. Breath Rate control knob out of adjustment (common)
- 2. Timing variable resistor out of adjustment (common)
- 3. Expiratory resistor out of adjustment (common)
- 4. Inspiratory check valve resistor out of adjustment (common)
Connect exhalation valve to transducer and transducer to digital counter
- Test point
- Exhalation valve port to transducer to counter
- Expected
- Circuit connected and operational
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe Fail alarm activates patient safety
Skin temperature probe alarm is triggered indicating probe contact or connection issue
Ranked causes
- 1. Skin temperature probe detached from infant (common)
- 2. Skin temperature probe is damaged (occasional)
- 3. Skin temperature probe disconnected from receptacle (common)
- 4. Skin temperature probe connector damaged (occasional)
Verify probe is in contact with infant skin
- Test point
- Skin temperature probe tip
- Expected
- Firm contact with infant skin
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suction does not obtain desired pressure caution
Suction system fails to achieve or maintain target vacuum pressure; low or no suction output
Ranked causes
- 1. Suction jar rim gasket or jar damaged or worn (common)
- 2. Suction hose kinked, tangled, or leaking (common)
- 3. Suction filter clogged or dirty (common)
- 4. Pressure or suction gauge measuring incorrectly; Suction Module failure (occasional)
Inspect suction jar and rim gasket
- Test point
- Suction jar (A) and rim gasket
- Expected
- No visible wear, cracks, or damage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature probe reading incorrect or missing patient safety
Diagnostic Test #10 shows incorrect ambient temperature probe values or display shows rail value (17.67±0.03) indicating no probe or circuit failure
Ranked causes
- 1. Ambient temperature probe disconnected or damaged (common)
- 2. Display P.C. Board (PCB1) failure (occasional)
- 3. Ambient temperature probe temperature difference excessive (occasional)
Access off-line Diagnostic Mode and proceed to Diagnostic Test #10; observe ambient probe readings
- Test point
- Baby Temperature and Apgar Timer displays
- Expected
- Ambient Probe 1 and 2 temperature values displayed (or 17.67±0.03 if no probe)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender Module alarm activates caution
Blender Module alarm is triggered indicating oxygen/air supply issue
Ranked causes
- 1. Oxygen and/or air is disconnected (common)
- 2. Oxygen and/or air tank not opened (common)
- 3. Pressure differential exceeds 30 psi ± 2 psi (occasional)
Verify oxygen and air supply connections
- Test point
- Oxygen and air inlet connectors
- Expected
- Both supplies connected and secured
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Suction Relief Valve Maximum Pressure caution
Verify and adjust suction relief valve spring to maintain maximum suction pressure within specification; calibrate gauge accuracy across range.
Ranked causes
- 1. Suction gauge accuracy not within 3% of full scale (unknown)
- 2. Suction relief valve maximum pressure out of specification (unknown)
- 3. Suction gauge zero point drift (unknown)
Verify suction gauge accuracy across operating range
- Test point
- Suction hose connected to RT200 tester; gauge readings compared to test device
- Expected
- Readings within 3% of full scale
- Tools
- RT200 tester or equivalent, pressure range 185 mm Hg (99"H2O)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic Test #6 Heater/Relays AC Line Voltage Display patient safety
Tests safety relay status and AC line voltage via Solid State Relays K3 and K4; quartz heater runs ~40 seconds per relay
Ranked causes
- 1. Power and Control P.C. Board (PCB2) failure (common)
- 2. Quartz heater element failure (occasional)
- 3. AC line voltage out of specification (occasional)
Access off-line Diagnostic Mode; at Test #6, verify safety relay (K1) status is initially Off on Apgar Timer display
- Test point
- Apgar Timer display
- Expected
- Safety relay status Off
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater does not turn on caution
Heater fails to activate; no radiant heat output detected
Ranked causes
- 1. Quartz heating element open or damaged (common)
- 2. Connection at J4 on Power and Control P.C. Board (PCB2) loose or damaged (occasional)
- 3. J4 or K1 connector damaged on Power and Control P.C. Board (PCB2) (occasional)
Remove and inspect quartz heating element
- Test point
- Quartz heating element (A)
- Expected
- No signs of breakage, discoloration, or damage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Fail alarm activates high voltage
Power Fail alarm is triggered on unit startup or during operation
Ranked causes
- 1. Circuit breaker on power input is tripped (common)
- 2. Power cord is unplugged (common)
- 3. Power cord is damaged (occasional)
Verify circuit breaker status
- Test point
- Power input circuit breaker
- Expected
- Not tripped
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin temperature probe reading incorrect or missing patient safety
Diagnostic Test #11 shows incorrect skin probe temperature or displays rail value (17.67±0.03) indicating probe disconnected or circuit failure
Ranked causes
- 1. Skin temperature probe disconnected or damaged (common)
- 2. Display P.C. Board (PCB1) failure (occasional)
- 3. Skin probe damaged or skin probe receptacle damaged (occasional)
Access off-line Diagnostic Mode and proceed to Diagnostic Test #11; attach skin probe to body surface; observe readings
- Test point
- Baby Temperature and Apgar Timer displays
- Expected
- Skin Probe 1 and 2 temperature values displayed (or 17.67±0.03 if no probe/problem)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby skin temperature probe not displaying temperature patient safety
Baby Temperature display does not activate when probe connected to probe connector
Ranked causes
- 1. Temperature probe disconnected or defective (common)
- 2. Probe connector circuit fault (occasional)
Verify probe connection at connector
- Test point
- Skin Temperature Probe connector
- Expected
- Probe physically connected and seated properly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender Module O2 concentration setting does not match display caution
Blender Module display setting does not correspond to the actual oxygen concentration knob position
Ranked causes
- 1. Blender Module knob is loose (common)
- 2. Blender Module has failed (occasional)
Perform RAP 2.24 diagnostic and repair procedure
- Test point
- Blender Module O2 setting and display
- Expected
- Setting matches display
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender module output mixture out of specified range patient safety
Blender module output oxygen concentration is outside the specified range of 20.8% to 100%
Ranked causes
- 1. Water contamination in air inlet (common)
- 2. Blender module damaged (occasional)
Verify oxygen concentration output
- Test point
- Blender Module output reading
- Expected
- Between 20.8% and 100%, 20.8-100%
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender Module output mixture outside specified range caution
Blender Module gas mixture is not within 20.8% to 100% oxygen specification
Ranked causes
- 1. Air inlet H2O filter is missing or clogged (common)
- 2. Blender Module has failed (occasional)
Perform RAP 2.26 diagnostic and repair procedure
- Test point
- Blender Module output O2 percentage
- Expected
- 20.8% to 100%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender module oxygen concentration reading does not match setting patient safety
Oxygen concentration display on blender module does not match the knob setting
Ranked causes
- 1. Blender module knob not secure or misaligned (common)
- 2. Defective blender module (occasional)
Verify blender knob is secure and properly aligned
- Test point
- Blender Module knob (A)
- Expected
- Knob fully seated and aligned
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic Test #5 Heater Duty Cycle patient safety
Displays AC heater duty cycle percentage applied to achieve 600 Watts; verifies AC line voltage correction
Ranked causes
- 1. Display P.C. Board (PCB1) failure (occasional)
- 2. Incorrect AC line voltage (occasional)
Verify AC line voltage per RAP 2.7; consult table 2-3 for correction percentage values
- Test point
- AC mains
- Expected
- Voltage within normal range for model
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heating element does not turn ON high voltage
Radiant heater element fails to energize during normal operation
Ranked causes
- 1. Heating element is open circuit (occasional)
- 2. Internal malfunction (unknown)
Verify heating element connectors are seated
- Test point
- Heating element connector
- Expected
- Firmly attached and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Perform back-up audio annunciation test
Diagnostic Test #17 activates piezo back-up audio annunciator at 1 second intervals
Ranked causes
- 1. Display PCB (PCB1) malfunction (common)
- 2. Power and Control PCB (PCB2) malfunction (occasional)
Access Diagnostic Test #17 and verify OFF appears on display
- Test point
- Baby Temperature display
- Expected
- OFF appears
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Perform speaker test across volume ranges
Diagnostic Test #16 activates speaker test at low, mid-range, and high volume levels
Ranked causes
- 1. Display PCB (PCB1) malfunction (common)
- 2. Power and Control PCB (PCB2) malfunction (occasional)
Access Diagnostic Test #16 and verify OFF appears on display
- Test point
- Baby Temperature display
- Expected
- OFF appears
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pre-Warm Mode not activated on system reset caution
Unit does not enter Pre-Warm Mode when powered on or reset
Ranked causes
- 1. System is in Diagnostic Mode (common)
- 2. Internal malfunction with PCB2 (occasional)
Verify Diagnostic Mode is not active
- Test point
- Front panel display
- Expected
- Normal display, not Diagnostic Mode pattern
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pre-warm mode not activated upon system reset caution
System does not enter pre-warm mode when powered on after reset; heater does not activate in pre-warm phase
Ranked causes
- 1. System in Diagnostic Mode (common)
- 2. ROM test failure on Power and Control P.C. Board (PCB2) (occasional)
Verify system is out of Diagnostic Mode
- Test point
- Display and system operation
- Expected
- Normal operation mode displayed; system responsive to standard controls
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin probe temperature accuracy out of specification patient safety
Diagnostic Test #12 shows skin probe accuracy temperature readings not within 36.00°C ±0.05°C tolerance
Ranked causes
- 1. Skin temperature sensor out of calibration (common)
- 2. Display P.C. Board (PCB1) failure (occasional)
Access off-line Diagnostic Mode and proceed to Diagnostic Test #12; observe skin probe accuracy temperature readings
- Test point
- Baby Temperature and Apgar Timer displays
- Expected
- Accuracy temperature values displayed for Skin Probes 1 and 2
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify logic voltage display reads out of range
CPU voltage on Baby Temperature display does not show 5.100V DC ± 0.1V DC during Diagnostic Test #13
Ranked causes
- 1. Display PCB (PCB1) malfunction (common)
- 2. DC power supply voltage out of specification (occasional)
Access Diagnostic Test #13 and read CPU voltage
- Test point
- Baby Temperature display
- Expected
- 5.100V DC ± 0.1V DC, ±0.1 V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify supercap voltage display reads out of range
Supercap voltage on Baby Temperature display during Diagnostic Test #14 not between 3.750V DC to 5.300V DC
Ranked causes
- 1. Display PCB (PCB1) malfunction (common)
- 2. Power and Control PCB (PCB2) malfunction (occasional)
Access Diagnostic Test #14 and read supercap voltage
- Test point
- Baby Temperature display
- Expected
- 3.750V DC to 5.300V DC, 3.750 to 5.300 V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADC gain or offset out of specification
Diagnostic Test #9 displays ADC gain below 4085 or offset above 7, showing FAIL on displays instead of numeric values
Ranked causes
- 1. Display P.C. Board (PCB1) failure or ADC calibration error (occasional)
Access off-line Diagnostic Mode and proceed to Diagnostic Test #9; measure ADC gain
- Test point
- Apgar Timer display
- Expected
- Gain between 4085 and 4095 counts, 4085-4095
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound at 30-second intervals during 5-minute wait patient safety
During the 5-minute wait after initial 10-minute Chk Patient alarm, alarm does not sound at 30-second intervals as expected
Ranked causes
- 1. Alarm circuit timer or buzzer malfunction (occasional)
Monitor alarm sound interval during extended wait
- Test point
- Alarm buzzer; timer
- Expected
- Alarm sounds at 30-second intervals for 5 minutes
Apgar Timer alarm does not sound at 1 minute caution
After Apgar Timer reaches 1:00 minute, alarm does not beep once as expected
Ranked causes
- 1. Timer interval alarm circuit malfunction (occasional)
Monitor timer for 1-minute alarm
- Test point
- Apgar Timer display; Alarm buzzer
- Expected
- Alarm beeps once when timer reaches 1:00 minute
Apgar Timer does not pause or reset caution
Pressing Start/Stop key does not pause Apgar Timer count, or pressing Reset key does not turn off timer display
Ranked causes
- 1. Start/Stop or Reset key circuit malfunction (occasional)
Press Start/Stop to pause timer
- Test point
- Start/Stop key; Apgar Timer display
- Expected
- Timer count stops
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Apgar Timer does not start or count up caution
After reconnecting probe and pressing Start/Stop key, Apgar Timer display does not come on or does not begin counting up from 0 seconds
Ranked causes
- 1. Apgar Timer circuit or Start/Stop key malfunction (occasional)
Reconnect probe and start Apgar Timer
- Test point
- Skin Temperature Probe connector; Start/Stop key; Apgar Timer display
- Expected
- Apgar Timer display comes on and counts up from 0 seconds
AutoBreath Infant Resuscitator BPM outside specified range caution
AutoBreath module measured breaths-per-minute rate is outside normal operating specification
Ranked causes
- 1. Internal malfunction with AutoBreath module (unknown)
Perform RAP 2.30 diagnostic procedure
- Test point
- AutoBreath module BPM measurement
- Expected
- Within specified BPM range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AutoBreath infant resuscitator breaths-per-minute out of range patient safety
AutoBreath infant resuscitator measured breaths-per-minute (BPM) is outside the specified range of 10-60 BPM
Ranked causes
- 1. Resuscitation module defective (common)
Measure AutoBreath measured breaths-per-minute
- Test point
- AutoBreath module
- Expected
- Between 10 BPM and 60 BPM, 10-60 BPM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AutoBreath infant resuscitator I/E ratio out of specification patient safety
Inhalation/Exhalation (I/E) ratio measured on AutoBreath infant resuscitator is outside specification
Ranked causes
- 1. Resuscitation module defective or miscalibrated (common)
Measure AutoBreath I/E ratio
- Test point
- AutoBreath module output
- Expected
- Within specified I/E ratio range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Mode does not activate or set temperature defaults incorrectly patient safety
Pressing Mode Select key to enter Baby Mode does not light Baby indicator, or Set Temperature display does not activate and default to 97.7°F (36.5°C), or Baby Temperature indicator does not flash
Ranked causes
- 1. Mode select circuit or temperature sensor initialization fault (occasional)
Select Baby Mode
- Test point
- Mode Select key; Baby indicator; Set Temperature display; Baby Temperature indicator
- Expected
- Baby indicator lights; Set Temperature display activates and defaults to 97.7°F (36.5°C); Baby Temperature indicator flashes; alarm sounds
Bassinet front panel not properly engaged or cannot position patient safety
Front panel cannot slide under mattress, does not return to upright position, or does not stay positively engaged to confine infant
Ranked causes
- 1. Front panel hinge, engagement mechanism, or corner bracket defect or wear (common)
Test front panel sliding under mattress
- Test point
- Front panel (D) and mattress (E) sliding interface
- Expected
- Front panel raises and slides smoothly under mattress
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bassinet side panels not properly engaged to confine infant patient safety
Side panels cannot be raised and pivoted down, or do not stay positively engaged in upright position to confine infant
Ranked causes
- 1. Side panel hinge or engagement mechanism defect or wear (common)
Test side panel raising and lowering
- Test point
- Side panel (C) hinges and pivots
- Expected
- Side panels raise and pivot to hang straight down; return smoothly to upright position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bassinet tilt mechanism does not function patient safety
Bassinet assembly cannot be placed in 5° and 10° tilt positions or does not return to level position smoothly
Ranked causes
- 1. Tilt mechanism wear, mechanical binding, or defect (common)
Test standard tilt mechanism (if applicable)
- Test point
- Mattress tilt control lever; Bassinet assembly rear edge
- Expected
- Bassinet places smoothly into 5° and 10° tilt positions; returns smoothly to level position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender module sets off alarms patient safety
Alarms are triggered by the blender module during operation
Ranked causes
- 1. Gas source pressure or flow decreased or increased (common)
Verify pipeline and cylinder gas flow
- Test point
- Gas inlet connections
- Expected
- Proper flow from pipeline and/or cylinder
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Auxiliary Flow Control caution
Adjust auxiliary flow control knob and valve to ensure readings match test device across full range (1.0 to 15.0 LPM); verify regulator pressure stability.
Ranked causes
- 1. Auxiliary flow control knob reading does not match RT200 tester reading (unknown)
Set and verify auxiliary flow at 1.0 LPM
- Test point
- Auxiliary Outlet connected to RT200 tester high flow range (180 LPM)
- Expected
- 1.0 LPM (2.1 scfh) on RT200 tester
- Tools
- RT200 tester or equivalent
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Auxiliary Flow Relief Valve caution
Verify and adjust relief valve spring to maintain specified pressure limit based on module type; ensure correct pressure relief at maximum flow.
Ranked causes
- 1. Auxiliary relief valve pressure setting out of specification (unknown)
Verify relief valve pressure for module without AutoBreath (PN 81 400 72 or 81 400 73)
- Test point
- Auxiliary Outlet connected to RT200 tester low pressure range (250 cm H2O)
- Expected
- 160 cm H2O (16 kPa) ± 10%, ±10%
- Tools
- RT200 tester or equivalent
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Auxiliary Pressure Regulator caution
Verify and adjust auxiliary pressure regulator to specified setpoint; incorrect pressure setting affects auxiliary flow control accuracy.
Ranked causes
- 1. Auxiliary pressure regulator not set to 20 psi (138 kPa) (unknown)
Measure auxiliary pressure regulator setting
- Test point
- Auxiliary pressure regulator test port
- Expected
- 20 psi (138 kPa), Exact
- Tools
- Pressure gauge assembly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Heater head electrical verification high voltage
Verify heater element resistance and illumination at specified AC input voltage
Ranked causes
- 1. Heater element resistance out of specification (unknown)
Measure heater element resistance at 100V AC input
- Test point
- Heater element
- Expected
- 12Ω to 15Ω, 12 to 15Ω
- Tools
- Digital voltmeter (Fluke 1 8080A or equivalent)
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen and Air Regulators high voltage
Verify regulator output pressure matches specification for tank pressure; regulator output pressure outside tolerance indicates miscalibration.
Ranked causes
- 1. Regulator output pressure out of specification (unknown)
Measure oxygen regulator output pressure with tank at specified pressure
- Test point
- Oxygen Tank Regulator white hose output
- Expected
- For U.S.: 65 psi ± 3 psi (448 kPa ± 21 kPa) with 1200 psi tank pressure, ±3 psi
- Tools
- RT200 Tester or equivalent
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Patient Gas Supply Pressure Regulator patient safety
Verify and adjust patient gas supply regulator to 20 psi (138 kPa) setpoint; pressure adjustment affects patient airway pressure control.
Ranked causes
- 1. Patient gas supply regulator pressure out of specification (unknown)
Measure patient gas supply regulator pressure
- Test point
- TP7A test port (orange plug line location)
- Expected
- 20 psi (138 kPa)
- Tools
- Pressure gauge assembly
Calibration: PEEP pressure adjustment patient safety
PEEP pressure does not read 21 cm H2O (2 kPa) when controls set to maximum
Ranked causes
- 1. PEEP variable resistor out of adjustment (common)
Connect patient circuit and set all controls to maximum
- Test point
- Patient outlet to exhalation valve port
- Expected
- Flow Rate, Airway Pressure Relief, and PEEP knobs at maximum
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PEEP valve zero adjustment patient safety
PEEP valve does not null flow when control knob turned counterclockwise
Ranked causes
- 1. PEEP control valve setscrews loose or PEEP knob misaligned (common)
Connect exhalation valve to test device low flow connector
- Test point
- Exhalation valve port to RT200 connector
- Expected
- Connected for low flow measurement 5.0 LPM
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pilot pressure adjustment patient safety
Pilot pressure circuit pressure does not read 60 cm H2O (5.9 kPa) on test device
Ranked causes
- 1. Pilot pressure variable resistor out of adjustment (common)
Connect patient circuit and test device
- Test point
- Exhalation valve port to RT200 connector
- Expected
- Circuit connected with low pressure range set to 250 cm H2O
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Resuscitation Module pressure switch point patient safety
Highest pressure point does not correspond to 22 psi (152 kPa) switch point on AutoBreath Infant Resuscitator
Ranked causes
- 1. Timing variable resistor out of adjustment (common)
Verify highest pressure point setting
- Test point
- Pressure measurement point
- Expected
- 22 psi (152 kPa), exact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Suction Valve Zero Flow caution
Adjust suction control valve to eliminate leakage at zero position; suction flow should read zero when control knob is at minimum.
Ranked causes
- 1. Suction control knob zero position does not correspond to zero flow reading (unknown)
Measure suction flow at minimum control knob position
- Test point
- Suction hose connected to RT200 tester
- Expected
- 0 LPM
- Tools
- RT200 tester or equivalent, clamping devices
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cart/bassinet docking mechanism or caster locks do not function patient safety
Cart/bassinet cannot be docked or undocked properly, or caster locks do not keep cart from moving when depressed (Birthing Room model)
Ranked causes
- 1. Docking latch mechanism or caster lock defect or wear (occasional)
Test undocking and caster lock engagement
- Test point
- Cart/bassinet docking mechanism (L); Caster locks (M)
- Expected
- Cart/bassinet undocks smoothly; caster locks keep cart from moving when depressed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chk Patient alarm does not sound at 10 minutes with 100% heater power patient safety
After 10 minutes at 100% heater power with probe connected, Chk Patient indicator does not light and alarm does not sound once
Ranked causes
- 1. Alarm circuit or timer malfunction (occasional)
Set Heater Power to 100%
- Test point
- Heater Power display; up arrow key
- Expected
- All segments light showing 100%
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Controller fails to enter Pre-Warm Mode caution
Self-test completes but controller does not begin Pre-Warm Mode operation or heater power sequence not as expected (should show 10 segments for 3 min, 6 segments for 12 min, then 3 segments)
Ranked causes
- 1. Controller firmware or mode logic fault (unknown)
Verify Pre-Warm Mode activation and heater power sequence
- Test point
- Pre-Warm indicator; Heater Power indicator display
- Expected
- Pre-Warm indicator on; Heater Power indicator displays ten segments (100%) for 3 min, reduces to six segments (60%) for 12 min, then reduces to three segments (30%)
Controller Module self-test fails to complete high voltage
Power switch turned on but self-test does not engage or display expected test pattern (all 8s, alarm indicators, mode indicators, heater power segments, alarm tones)
Ranked causes
- 1. Power supply or controller circuit fault (common)
Verify AC power supply connection
- Test point
- Power connector on rear panel
- Expected
- AC line cord plugged into properly grounded hospital-use receptacle of correct voltage
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic Test #4 Key Check
Tests keypad functionality; seven zeroes appear in displays; pressing any key changes nearest 0 to 1
Ranked causes
- 1. Display P.C. Board (PCB1) failure (common)
Press Apgar Start/Stop key; verify display segments toggle between 0 and 1
- Test point
- Apgar Timer and Baby Temperature displays
- Expected
- Corresponding segments toggle between 0 and 1
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows "Lo" in baby temperature display patient safety
Baby temperature display shows "Lo" indicating very low or invalid temperature reading
Ranked causes
- 1. Skin temperature probe reads below 18°C (occasional)
Measure skin temperature probe temperature
- Test point
- Skin temperature probe reading
- Expected
- > 18°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEPROM status flag incorrect
Diagnostic Test #7 displays EEPROM status flag on Baby Temperature display showing values other than 001 (normal) or 008 (acceptable)
Ranked causes
- 1. EEPROM data block never initialized or corrupted (occasional)
Access off-line Diagnostic Mode and proceed to Diagnostic Test #7
- Test point
- Baby Temperature display
- Expected
- Status flag 001 or 008
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error log clear function not working
Error log does not clear when pressing Mode Up arrow key; Baby Temperature display does not go blank after attempting to clear error log during Diagnostic Test #18
Ranked causes
- 1. Power and Control P.C. Board (PCB2) failure (common)
Press Up arrow Mode Select key during error log display
- Test point
- Baby Temperature display
- Expected
- Error log clears and Baby Temperature display goes blank; reference number returns to 0:01 on Apgar Timer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error log not displaying
Error Log Display test fails; error reference numbers 0:01-0:04 do not appear on Apgar Timer display or error codes do not appear on Baby Temperature display during Diagnostic Test #18
Ranked causes
- 1. Display P.C. Board (PCB1) failure (common)
Access Diagnostic Test #18 and verify error log display
- Test point
- Apgar Timer display and Baby Temperature display
- Expected
- Error reference numbers 0:01-0:04 appear on Apgar Timer; pressing °C/°F key cycles through errors showing on both displays
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Examination Light does not turn on or off caution
Pressing Examination Light key does not turn on light, or pressing again does not turn it off
Ranked causes
- 1. Examination Light key circuit or lamp malfunction (occasional)
Press Examination Light key to turn on
- Test point
- Examination Light key; Light fixture
- Expected
- Examination light comes on
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Heater does not shut down after 15 minutes total runtime patient safety
After 15 minutes total (10+5 min), heater does not shut down, Heater Power indicators do not go off, or alarm does not sound continuously with ramping volume
Ranked causes
- 1. Heater shutdown circuit or safety timer fault (common)
Verify heater shutdown after 15 minutes total
- Test point
- Heater Power indicators; Heater on/off state; Alarm buzzer
- Expected
- Heater shuts down; Heater Power indicators go off; Alarm sounds continuously and ramps up in volume
More symptoms
- Heater ON but indicators are OFF
- Heater Power controls do not function in Manual Mode
- High Temperature alarm does not activate at ≥39°C
- High Temperature alarm does not stop when temperature lowered
- Inspiratory/Expiratory ratio greater than 1:2.4 and less than 1.6:1 or 1:1
- Instrument tray does not swing or function
- Keypad Lock function not operating
- No power and Power Fail alarm does not activate
- Pass-through drawer does not slide or function
- Patient gas supply measured flow out of specified range
- Patient Gas Supply measured flow outside specification
- Patient gas supply pressure relief gauge exceeds 60 cm H2O
- Patient Gas Supply pressure relief gauge exceeds limit
- Perform watchdog timer test
- Power Fail alarm does not activate on loss of AC power
- Probe Disconnect alarm does not trigger
- Silence/Reset key does not restart heater
- Silence/Reset key does not stop alarm in Baby Mode
- Skin probe heating too rapid or too close to heater element
- System Fail alarm activates with error code
- System Fail alarm activates without error code
- Temperature below set point does not trigger alarm
- Temperature override above set point does not trigger alarm
- Temperature override alarm does not silence for 10 minutes
- Temperature Override Mode (>37°C) does not activate
- Temperature Override Mode does not deactivate at lower threshold
- Unit configuration mismatch with jumper settings
- Warmer head assembly does not swivel or is not locked in center
- Writing surface does not slide or function
- X-ray cassette tray does not slide or function
30 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
10 | Power fail circuit failure | Power failure detection circuit malfunctions | error |
11 | Central processing unit (CPU) voltage is out of specification | CPU supply voltage exceeds tolerance | error |
12 | Keypad failed | Keypad not responding or false inputs | error |
13 | ROM test failure | ROM checksum or read test fails | error |
14 | RAM test failure | RAM read/write test fails | error |
15 | EEPROM test failure | EEPROM checksum or access test fails | error |
16 | Analog/digital (A/D) offset is out of specification | A/D converter offset calibration exceeds tolerance | error |
17 | A/D gain is out of specification | A/D converter gain calibration exceeds tolerance | error |
18 | DELETED by software version 1.06 in 1998. Older units may still display this code | Unit running software version prior to 1.06 | info |
2 | AC line voltage is too high | AC mains voltage exceeds system specification | warning |
20 | External watchdog timer failure | Watchdog timer circuit failure or timeout | error |
21 | Difference between ambient temperature probe 1 and 2. Displayed temperatures exceed 0.4°C in diagnostic test #10 | Ambient temperature probes reading difference >0.4°C | warning |
22 | 36°C probe calibration error exceeds 0.2°C | Skin probe calibration accuracy exceeds ±0.2°C error | warning |
23 | Ambient temperature is above 32°C (90°F) | Ambient temperature sensor reads above 32°C threshold | warning |
24 | Internal malfunction | Unspecified internal system failure detected | critical |
25 | Internal malfunction | Unspecified internal system failure detected | critical |
3 | Solid state relay (SSR) path short circuits | SSR develops short circuit condition | error |
4 | Heater circuit failed (short circuit) | Heater circuit develops short circuit | error |
5 | SSR #1 failed (open circuits or heater open circuits) | SSR #1 develops open circuit or heater opens | error |
6 | SSR #2 failed (open circuits or heater open circuits) | SSR #2 develops open circuit or heater opens | error |
7 | Safety Relay failed (short circuits) | Safety relay develops short circuit | error |
8 | AC heater current is high | Heater current exceeds specification | warning |
9 | Hardware detects a high temperature >39.5°C (103.1°F) | Temperature sensor reads above 39.5°C threshold | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.