Dräger Narkomed Mobile: troubleshooting and error codes
The ServiceForge knowledge base holds 16 documented symptoms for the Dräger Narkomed Mobile, with 49 ranked causes, 62 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Ventilator Inoperative patient safety
Ventilator system not functioning; no breath delivery or control unresponsive.
Ranked causes
- 1. Ventilator settings incorrect or not verified (common)
- 2. Cylinder or pipeline oxygen supply inadequate pressure (common)
- 3. Breathing or scavenger hose connections incorrect (occasional)
- 4. Ventilator controller power supply failure (occasional)
- 5. Ventilator controller wire harness or connections faulty (occasional)
- 6. Valve and guide assembly failure (occasional)
- 7. Ventilator controller assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Blank Upon System Power-up
Display does not show any information or graphics when system is powered on.
Ranked causes
- 1. Display cable not connected to display port (common)
- 2. Internal cable to J204 on processor disconnected or damaged (occasional)
- 3. Power supply voltages out of tolerance (occasional)
- 4. Display assembly failure (occasional)
- 5. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad Inoperative
Keypad buttons not responding to user input; system cannot accept command inputs.
Ranked causes
- 1. Display cable not connected to display port (common)
- 2. Internal cable to J4 on processor disconnected or damaged (occasional)
- 3. Power supply voltages out of tolerance (occasional)
- 4. Display and keypad assembly failure (occasional)
- 5. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Respiratory Volume Monitor patient safety
Respiratory volume monitor not displaying; volume readings unavailable.
Ranked causes
- 1. Volume sensor not installed correctly (common)
- 2. Sensor cord disconnected from interface panel (common)
- 3. Internal cable from interface panel to J102 faulty (occasional)
- 4. Volume sensor failure (occasional)
- 5. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial Port Communication Failure
Serial port not communicating with external devices or host system.
Ranked causes
- 1. Peripheral cable not connected to serial interface port (common)
- 2. Internal cable from serial port to processor faulty (occasional)
- 3. Peripheral or host serial port configuration incorrect (occasional)
- 4. Serial cable defective (occasional)
- 5. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Breathing Pressure Monitor patient safety
Breathing pressure monitor not displaying; readings unavailable; unit may operate otherwise.
Ranked causes
- 1. Pilot line disconnected from pressure monitoring system (common)
- 2. Pressure hose connections loose or disconnected (common)
- 3. Internal tubing from interface panel to transducer faulty (occasional)
- 4. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of O2 Monitor patient safety
O2 monitor not displaying or responding; unit may be operational otherwise.
Ranked causes
- 1. Sensor element not installed correctly (common)
- 2. Sensor cord disconnected from interface panel (common)
- 3. Internal cable from interface panel to J103 on processor board is faulty (occasional)
- 4. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Audio Alarms patient safety
Audio alarm function not working; no sound output when alarms are triggered.
Ranked causes
- 1. Speaker wire harness disconnected or damaged (common)
- 2. Speaker failure (common)
- 3. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No O2 Supply Pressure Alarms patient safety
Oxygen supply pressure alarm function not operational; low O2 pressure not detected.
Ranked causes
- 1. O2 low pressure alarm switch or wire harness faulty (common)
- 2. Speaker or speaker wire harness failure (common)
- 3. Processor assembly failure (occasional)
Perform complete PMS on unit
- Test point
- Unit self-test
- Expected
- Unit passes all PMS tests
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scavenger system pressure out of specification patient safety
Manometer reading exceeds safe limits during negative or positive pressure tests, indicating improper scavenger function or blockage.
Ranked causes
- 1. Scavenger needle valve not properly adjusted (common)
- 2. Scavenger system blockage or obstruction (occasional)
Negative pressure test - verify scavenger system vacuum
- Test point
- Scavenger adapter hose barb
- Expected
- pressure less than or equal to 0.5 cm H2O, ≤0.5 cm H2O
- Tools
- Manometer
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of specification
Battery voltage at full charge not within 13.50 to 14.80 VDC range during AC line operation, or low battery cutoff not within 10.5 to 10.0 VDC during battery operation.
Ranked causes
- 1. Battery voltage out of specification (unknown)
Measure battery voltage at terminals while on AC power
- Test point
- Battery terminals
- Expected
- 13.50 to 14.80 VDC, 13.50 to 14.80 VDC
- Tools
- Voltmeter, multimeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cylinder pressure regulator output caution
Adjust O2, N2O, Air, He, He/O2, or N2 regulator output pressure to specification using compensated settings based on cylinder supply pressure
Ranked causes
- 1. Regulator output requires calibration per cylinder pressure (common)
Set System Power switch position
- Test point
- System Power switch
- Expected
- STANDBY
19 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure Monitor span
Current airway pressure reading requires span calibration at known test pressure
Ranked causes
- 1. Pressure sensor span calibration drift (unknown)
Disconnect pressure sample line from absorber
- Test point
- Pressure sample line connection
- Expected
- Line removed and absorber isolated
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure Monitor zero
Current airway pressure reading requires zero calibration reset to atmospheric baseline
Ranked causes
- 1. Pressure sensor zero offset drift (unknown)
Disconnect pressure sample line from absorber
- Test point
- Pressure sample line connection
- Expected
- Line removed and absorber isolated
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 SUPPLY alarm set point out of range patient safety
Alarm activates outside the specified 40 to 34 psi range during system depressurization.
Ranked causes
- 1. Adjustment set screw out of calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor replacement patient safety
Oxygen sensor housing and capsule replacement procedure for Narkomed Mobile anesthesia machine. Sensor located in pressure sensor adapter on inspiratory valve dome.
Ranked causes
- 1. Oxygen sensor capsule degradation or failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.