Dräger Narkomed M: troubleshooting and error codes
The ServiceForge knowledge base holds 34 documented symptoms for the Dräger Narkomed M, with 70 ranked causes, 139 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Ventilator inoperative patient safety
Ventilator is not functioning or not working properly during operation
Ranked causes
- 1. Incorrect ventilator settings (common)
- 2. Valve and guide assembly failure (occasional)
- 3. Inadequate gas supply pressure (occasional)
- 4. Incorrect breathing hose or scavenger connections (occasional)
- 5. No 8V power at ventilator controller J2 (occasional)
- 6. Wire harness or connector failure (occasional)
- 7. Ventilator controller assembly failure (rare)
Verify ventilator switch and all settings are correct
- Test point
- Ventilator switch, flow, frequency, I:E ratio, tidal volume, pressure limit
- Expected
- Switch ON; all parameters set to prescribed values
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Blank Upon System Power-up patient safety
Display does not show any information or image when system is powered on
Ranked causes
- 1. Display cable not connected to display port (common)
- 2. Internal cable to J204 on processor disconnected or damaged (occasional)
- 3. Display assembly defective (occasional)
- 4. Power supply voltages out of tolerance (occasional)
- 5. Processor assembly failure (rare)
Verify display cable connected to display port
- Test point
- Display port connector
- Expected
- Display cable properly connected
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad Inoperative patient safety
Keypad keys do not respond to presses; user cannot input commands via keypad
Ranked causes
- 1. Display cable not connected to display port (common)
- 2. Internal cable to J4 on processor disconnected or damaged (occasional)
- 3. Display and keypad assembly defective (occasional)
- 4. Power supply voltages out of tolerance (occasional)
- 5. Processor assembly failure (rare)
Verify display cable connected to display port
- Test point
- Display port connector
- Expected
- Display cable properly connected
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Respiratory Volume Monitor patient safety
Respiratory volume monitor not working or displaying properly
Ranked causes
- 1. Sensor cord disconnected from interface panel (common)
- 2. Sensor installed improperly (common)
- 3. Sensor valve mounting or gaskets degraded (occasional)
- 4. Internal cable from interface panel to processor board disconnected (occasional)
- 5. Processor assembly board failure (rare)
Is respiratory volume monitor working properly
- Test point
- Respiratory volume monitor display
- Expected
- Monitor displays respiratory volume readings
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial port communication failure
Peripheral device cannot communicate with anesthesia machine via serial interface port
Ranked causes
- 1. Peripheral cable not connected to serial interface port (common)
- 2. Peripheral or host serial port configuration mismatch (occasional)
- 3. Defective peripheral cable (occasional)
- 4. Internal cable from serial port to processor damaged (occasional)
- 6. Processor assembly still defective after replacement (rare)
Verify peripheral cable is physically connected
- Test point
- Serial interface port
- Expected
- Cable connected
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Breathing Pressure Monitor patient safety
Breathing pressure monitor not reading or not displaying pressure on Narkomed M
Ranked causes
- 1. Pressure hose connections loose or disconnected (common)
- 2. Pilot line not connected (common)
- 3. Internal tubing from interface panel to transducer disconnected or damaged (occasional)
- 4. Processor assembly failure (rare)
Check all pressure hose connections
- Test point
- Absorber, O2 sensor adapter, interface panel hose ports
- Expected
- All hoses seated and secure
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of O2 Monitor patient safety
Oxygen monitor not reading or displaying oxygen level on Narkomed M anesthesia machine
Ranked causes
- 1. Sensor cord not connected to interface panel (common)
- 2. Sensor element not installed correctly (common)
- 3. Internal cable from interface panel to J103 damaged or disconnected (occasional)
- 4. Processor assembly failure (rare)
Verify sensor cord connected to interface panel
- Test point
- Interface panel sensor port
- Expected
- Sensor cord properly seated
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSP O2 LOW alarm condition patient safety
Central alarm display shows low inspired oxygen concentration message with audible alarm
Ranked causes
- 1. Oxygen analyzer sensor not calibrated or out of range (common)
- 2. Oxygen supply interruption or low pressure (occasional)
- 3. Oxygen sensor blockage or malfunction (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Audio Alarms patient safety
System does not produce audio alarm sounds
Ranked causes
- 1. Speaker wire harness loose or connector not seated (common)
- 2. Speaker defective (occasional)
- 3. Processor assembly failure (rare)
Check speaker wire harness and J12 connection
- Test point
- J12 connector on processor board
- Expected
- Wire harness securely connected to J12
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No O2 Supply Pressure Alarms patient safety
Oxygen supply low pressure alarms not functioning or sounding
Ranked causes
- 1. Speaker wire harness loose or connection to J12 not seated (common)
- 2. O2 low pressure alarm switch malfunction or wire harness loose (common)
- 3. Processor assembly failure (rare)
Verify O2 supply pressure alarms functional
- Test point
- Low O2 supply pressure alarm
- Expected
- Alarms sound when O2 low
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen supply pressure alarms patient safety
Low oxygen pressure alarm fails to sound when supply pressure is low
Ranked causes
- 1. Speaker circuit or speaker defective preventing alarm output (common)
- 2. Processor power supply voltage insufficient for alarm function (occasional)
Test speaker alarm output function
- Test point
- Speaker output
- Expected
- Audible alarm sound when O2 pressure alarm triggered
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMC failure after sensor troubleshooting patient safety
Unit does not pass performance and maintenance check
Ranked causes
- 1. Original processor assembly requires reinstallation (occasional)
- 2. Original volume sensor requires reinstallation (occasional)
Perform complete PMC on the unit
- Test point
- All performance and maintenance check points
- Expected
- Unit passes all PMC criteria
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Absorber system maintenance required
Routine inspection and maintenance of CO2 absorber system including valve components, sensors, and soda lime canister.
Ranked causes
- 1. Scheduled maintenance procedure (common)
Examine O-rings on sensor housing, oxygen sensor plug, and breathing pressure hose assemblies
- Test point
- O2 Sensor Housing, Oxygen Sensor Plug, Breathing Pressure Hose Assembly
- Expected
- O-rings intact and undamaged
12 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of specification
Battery voltage during AC operation or battery mode outside specified range
Ranked causes
- 1. Battery not fully charged or charge circuit defective (common)
Measure battery voltage during AC operation at full charge
- Test point
- Battery terminals
- Expected
- 13.50 to 14.80 VDC, +/-0.65 VDC
- Tools
- Digital voltmeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing Pressure Monitor zero not reading correctly
After disconnection and stabilization, breathing pressure monitor zero point is inaccurate
Ranked causes
- 1. Zero calibration point requires reset (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cylinder Pressure Regulator high voltage
Adjust oxygen cylinder pressure regulator output to correct compensated setting based on cylinder pressure; verify tolerance after adjustment and test oxygen supply alarm function.
Ranked causes
- 1. Regulator output pressure requires adjustment or verification (common)
Disconnect all pipeline hoses
- Test point
- System
- Expected
- Hoses disconnected
24 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen monitor patient safety
One-year service interval oxygen monitor calibration procedure for the Narkomed M anesthesia machine
Ranked causes
- 1. Scheduled one-year oxygen monitor calibration (common)
Access Oxygen Monitor Service Screen
- Test point
- Mon Cal key
- Expected
- Oxygen Monitor Service Screen appears
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Monitor zero calibration
Zero calibration procedure for O2 cells A and B displayed on service screen with current and stored values
Ranked causes
- 1. O2 cells require zero calibration (common)
Access Oxygen Monitor Service Screen
- Test point
- Service Screen menu
- Expected
- Oxygen Monitor Service Screen displayed
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Ratio Controller (ORC) adjustment patient safety
Adjust ORC to maintain proper oxygen concentration across flow range using calibrated oxygen monitor
Ranked causes
- 1. ORC requires calibration adjustment (unknown)
Set O2 flow
- Test point
- O2 flow control
- Expected
- 8 l/min
- Tools
- O2 flowmeter
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMC Procedure verification patient safety
Perform PMC Procedure as part of ORC calibration completion
Ranked causes
- 1. PMC Procedure required after ORC adjustment (unknown)
Compensated Regulator Output Setting Verification caution
Verification that the regulator output pressure matches the compensated setting based on actual cylinder pressure, using a lookup table for pressures from 1000 to 2000 psi.
Ranked causes
- 1. Output pressure varies with cylinder supply pressure (common)
Determine current cylinder pressure
- Test point
- Cylinder pressure indicator
- Expected
- 1000, 1200, 1400, 1600, 1800, or 2000 psi
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Narkomed M battery voltage low during battery operation caution
Battery voltage drops below cutoff threshold during battery-powered operation
Ranked causes
- 1. Battery depleted or aging (common)
Measure battery voltage during battery-powered operation
- Test point
- Battery terminals during battery mode
- Expected
- Cutoff voltage 10.5 to 10.0 VDC, 10.0 to 10.5 VDC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Narkomed M battery voltage out of range during AC operation caution
Battery voltage measured at terminals during AC line operation falls outside specified range
Ranked causes
- 1. Charging circuit malfunction (unknown)
Measure battery voltage during AC line operation
- Test point
- Battery terminals
- Expected
- 13.50 to 14.80 VDC at full charge, +0.0 to +1.3 VDC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 sensor will not calibrate
Oxygen sensor fails to calibrate properly despite following calibration procedure steps.
Ranked causes
- 1. Sensor not stabilizing or improper sensor condition (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ORC adjusting screw out of spec patient safety
Oxygen concentration does not fall within specification when adjusting oxygen and nitrous oxide flows
Ranked causes
- 1. ORC adjusting screw requires repositioning (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor replacement patient safety
Oxygen sensor capsule requires replacement; sensor is located in pressure sensor adapter on top of inspiratory valve.
Ranked causes
- 1. Sensor capsule degradation or failure (unknown)
Turn system power to STANDBY
- Test point
- System Power switch
- Expected
- STANDBY position
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary power supply output voltage out of range high voltage
Voltage measurement at Condor power supply J3 connector (+15VDC) does not fall within acceptable limits; all downstream voltages may be affected.
Ranked causes
- 1. Condor power supply failure or fault (unknown)
Measure +15VDC at Condor power supply J3
- Test point
- J3-1 (Brn)
- Expected
- 14.0 to 16.0 VDC, 14.0 to 16.0 VDC
Processor board power supply voltages out of range high voltage
Voltage measurements at J14 connector (±5V, ±12V rails) do not fall within acceptable limits; display or monitoring functions may be impaired or non-functional.
Ranked causes
- 1. Primary power supply fault or voltage regulation failure (unknown)
Measure +5VDC at processor board J14
- Test point
- J14-12,14 (Red, Orn)
- Expected
- 4.80 to 5.25 VDC, 4.80 to 5.25 VDC
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Replace Spiromed respiratory volume sensor caution
Sensor installed between lower absorber pipe and PEEP valve; requires disconnection from monitor, removal from pipe, and reinstallation with proper gasket seating.
Ranked causes
- 1. Spiromed sensor requires replacement (unknown)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spiromed Sensor lubrication caution
Lubrication of Spiromed Sensor lateral holes using specified lubrication kit during maintenance.
Ranked causes
- 1. Scheduled Spiromed Sensor lubrication (common)
Locate four lateral holes on Spiromed Sensor
- Test point
- Spiromed Sensor lateral holes marked by arrows
- Expected
- Four holes visible and accessible
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vaporizer internal leak or exclusion system test failure patient safety
Vaporizer fails internal leak test or exclusion system test during vapor concentration verification
Ranked causes
- 1. Internal leak in vaporizer system (occasional)
Vaporizer internal leak or exclusion system test
- Test point
- Vaporizer system integrity
- Expected
- PASS - no internal leaks detected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator controller assembly failure confirmation patient safety
Ventilator still inoperative after all prior checks
Ranked causes
- 1. Ventilator controller assembly defective (rare)
Check all cables and tubing connections at ventilator controller
- Test point
- Ventilator controller
- Expected
- All connections secure and properly routed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator controller power supply voltage out of range high voltage
Voltage measurement at J2 connector (+8VDC rail) does not fall within acceptable limits; ventilator functions may be impaired or non-functional.
Ranked causes
- 1. Primary power supply fault or voltage regulation failure (unknown)
Measure +8VDC at ventilator controller J2
- Test point
- J2-3 (Brn)
- Expected
- 7.70 to 8.30 VDC, 7.70 to 8.30 VDC
Verify processor board power supply voltages high voltage
Technician must measure DC voltages at processor board connector J14 to verify proper power distribution.
Ranked causes
- 1. Power supply output voltage out of tolerance (occasional)
Measure +5VDC at processor board
- Test point
- J14 pins 12, 14 (Red, Orange wires)
- Expected
- 4.80 to 5.25 VDC, 4.80 to 5.25 VDC
- Tools
- DMM
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.