Dräger Narkomed 6000: troubleshooting and error codes
The ServiceForge knowledge base holds 78 documented symptoms for the Dräger Narkomed 6000, with 167 ranked causes, 374 diagnostic checks with test points and expected values, and 64 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Loss of Oxygen Monitor patient safety
O2 monitoring display not functioning or not showing readings
Ranked causes
- 1. O2 Sensor Cord disconnected from interface (common)
- 2. O2 Sensor Capsule not installed correctly (common)
- 3. O2 Sensor Capsule failed (occasional)
- 4. O2 Sensor Cord failed (occasional)
- 5. VPO (Vital Parameters Module) failed (occasional)
- 6. WPU (Work Processing Unit) failed (rare)
- 7. Cable or Vitalbus connection failure (rare)
Is O2 Sensor Cord connected to interface?
- Test point
- Interface panel connection
- Expected
- O2 Sensor Cord connected
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Respiratory Volume Monitor patient safety
Respiratory volume monitoring display not functioning or not showing readings
Ranked causes
- 1. Ultrasonic Flow Sensor Cord not connected to interface (common)
- 2. Ultrasonic Flow Sensor transducers not fully seated (common)
- 3. Ultrasonic Flow Sensor or adapter faulty (occasional)
- 4. Ultrasonic Flow Sensor transducers failed (occasional)
- 5. VPO (Vital Parameters Module) failed (occasional)
- 6. WPU (Work Processing Unit) failed (rare)
- 7. Cable or Vitalbus HUB connection failure (rare)
Is Ultrasonic Flow Sensor Cord connected to interface panel?
- Test point
- Interface panel connection
- Expected
- Cord fully connected
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Event log transfer to Windows XP laptop caution
Download Narkomed 6000 Series Event, Alarm and Data logs to service laptop for analysis
Ranked causes
- 1. Need to configure laptop hardware and network (common)
- 2. Need to power on and initialize system (common)
- 3. Need to verify or configure user access (occasional)
- 4. Need to verify computer name and network TCP/IP settings (occasional)
- 5. Need to create and configure NAD Software folder structure (occasional)
- 6. Need to share NAD Software folder with correct permissions (occasional)
Verify Vitalan cable connection to Vitalbus
- Test point
- Vitalbus port on Hub
- Expected
- Etherlink cable from laptop connected to Vitalbus port
7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display failure
CRT or Flat Panel display not showing image or is blank
Ranked causes
- 1. Video cable disconnected or faulty (common)
- 2. WPU Video PCB not properly seated (common)
- 3. Incorrect voltage at CRT or Flat Panel (occasional)
- 4. CRT or Flat Panel display failed (occasional)
- 5. WPU battery needs replacement and BIOS reset (occasional)
- 6. WPU (Work Processing Unit) failed (rare)
Turn power to Standby, remove AC power, wait 3 minutes, disable breakers
- Test point
- Power cycle procedure
- Expected
- Equipment fully powered down
6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audible alarms or speaker failure patient safety
Audible alarm signals not heard or speaker fail message displayed
Ranked causes
- 1. Master Volume set below 50% (common)
- 2. Speaker has poor resistance or connection (common)
- 3. Speaker failed (occasional)
- 4. Speaker connections faulty (occasional)
- 5. Vitalbus PCB failed (rare)
- 6. WPU (Work Processing Unit) failed (rare)
Is Windows NT Master Volume at least 50% or higher?
- Test point
- Volume setting
- Expected
- Master Volume at 50% or higher
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communications port failure
Peripheral device cannot communicate with NM6K via communications port
Ranked causes
- 1. Interface cable not connected to peripheral device (common)
- 2. Peripheral device or NM6K serial port configuration incorrect (common)
- 3. NM6K serial port internal cabling faulty (occasional)
- 4. ISO Port PCB failed (occasional)
- 5. WPU (Work Processing Unit) failed (rare)
Is peripheral interface cable connected to Comm port?
- Test point
- Comm port connection
- Expected
- Cable fully connected between peripheral and NM6K
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Event log transfer to laptop fails
NM6000 Series unable to transfer event logs to service laptop via Vitalbus/Etherlink connection during remote save operation
Ranked causes
- 1. Laptop computer name is not SRVLAP028 (common)
- 2. Target folder is not NAD Software (common)
- 3. TCP/IP network configuration incorrect (occasional)
- 4. Network workgroup is not configured to Goldisk (occasional)
- 5. Etherlink card not properly configured (occasional)
Verify network workgroup name
- Test point
- Windows NT Network Neighborhood Properties, Identification tab
- Expected
- Goldisk
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GAP/GAP2 fails self-diagnostic test
GAP or GAP2 analyzer fails self diagnostic test while all other PODs pass
Ranked causes
- 1. System power supply voltage issue (common)
- 2. GAP/GAP2 needs sample flow calibration (common)
- 3. Common Processor failed (occasional)
- 4. GAP/GAP2 Personality PCB voltage issue (occasional)
- 5. GAP/GAP2 Personality PCB failed (rare)
Perform GAP/GAP2 flow, line block and accuracy tests
- Test point
- GAP/GAP2 analyzer
- Expected
- All tests pass
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak test failure at ventilator start-up patient safety
Divan reports leak during power-on self-test; system fails leak integrity test
Ranked causes
- 1. Sub-System 1 leak (breathing circuit components) (common)
- 2. Sub-System 3 leak (breathing bag and related components) (common)
- 3. Sub-System 1/2 or Sub-System 2 leak (CO2 canister and exhaust path) (common)
- 4. Breasy assembly faulty or misaligned (occasional)
- 5. Piston-cylinder unit faulty or misaligned (occasional)
Run Divan power-on selftest
- Test point
- Divan control system
- Expected
- Selftest passes without leak alerts
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Breathing Pressure Monitor patient safety
Pressure monitoring display not functioning or not showing readings
Ranked causes
- 1. Pilot line or mechanical gauge not connected (common)
- 2. Pressure Monitor needs calibration (common)
- 3. VPO (Vital Parameters Module) failed (occasional)
- 4. WPU (Work Processing Unit) failed (rare)
- 5. Cable or Vitalbus HUB connection failure (rare)
Are pilot line and mechanical gauge connected at both ends?
- Test point
- Pilot line and gauge connections
- Expected
- Both connections fully seated
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve test manual patient safety
Manual solenoid valve operation test to check pneumatic control system function
Ranked causes
- 1. Fresh gas valve malfunction (unknown)
- 2. APL valve malfunction (unknown)
- 3. Air inlet valve malfunction (unknown)
- 4. Piston seal malfunction (unknown)
- 5. O2 supply valve malfunction (unknown)
Test fresh gas valve MY1
- Test point
- Breathing system interface V1
- Expected
- Control pressure reading changes when Pmax key pressed
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faulty vapor exclusion system caution
Exclusion system does not operate properly; vaporizer access or isolation fails
Ranked causes
- 1. Slider bars require lubrication or are binding (common)
- 2. Vaporizer height misaligned with exclusion bar (common)
- 3. Exclusion knob has free play or does not reach far right position (occasional)
- 4. Vapor Mount assembly faulty or worn (occasional)
Verify slider bar lubrication and free movement
- Test point
- Exclusion slider bars
- Expected
- Bars slide freely without binding or resistance
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GAP/GAP2 displays line block
LINE BLOCK message displayed on GAP or GAP2 analyzer screen indicating sample line obstruction
Ranked causes
- 1. External component obstruction (waterlok, filter, nafion, sample line) (common)
- 2. Internal component restriction or occlusion (common)
- 3. GAP pump flow restriction (occasional)
- 4. Pump failed (GAP2) (rare)
Perform GAP or GAP2 flow calibration
- Test point
- GAP/GAP2 analyzer
- Expected
- Flow calibration completes successfully
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GAP/GAP2 displays SERVICE GAP message
SERVICE GAP message displayed on GAP or GAP2 analyzer screen
Ranked causes
- 1. GAP/GAP2 needs sample flow calibration (common)
- 2. System power supply voltage issue (common)
- 3. IRIA error code logged (occasional)
- 4. GAP/GAP2 Personality PCB failed (occasional)
Perform GAP or GAP2 sample flow, line block and accuracy tests
- Test point
- GAP/GAP2 analyzer
- Expected
- All tests pass
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen not responding
Touch screen input not responding to operator touches or screen press actions not registering
Ranked causes
- 1. Touch/Video cable connection loose or faulty (common)
- 2. Touch screen needs calibration (common)
- 3. CRT or Flat Panel display failed (occasional)
- 4. WPU (Work Processing Unit) failed (rare)
Check Touch or Touch/Video cable connections from WPU to monitor
- Test point
- Cable connectors at WPU and monitor
- Expected
- Cables fully seated and secured
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator piston-cylinder unit not operating caution
Piston-cylinder unit fails to function or generates excessive audible hum during operation
Ranked causes
- 1. Shaft lubrication insufficient (common)
- 2. Loose or detached canister, piston, faulty diaphragm, or loss of seal pressure (common)
- 3. Piston removal position out of reset (occasional)
- 4. Piston-cylinder unit faulty or worn (occasional)
Lubricate piston-cylinder shaft
- Test point
- Shaft
- Expected
- Proper lubrication applied
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Event log transfer to service laptop fails
NM6000 Series unable to transfer event log and POD logs to service laptop using Ethernet connection
Ranked causes
- 1. Laptop not properly configured for network file sharing (common)
- 2. Virus protection not enabled or definitions outdated (occasional)
- 3. Ethernet connection not established (occasional)
Verify laptop computer name setting
- Test point
- Network configuration, Identification tab
- Expected
- srvlap
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak test manual
Service diagnostic to measure subsystem leak rate and locate leaks
Ranked causes
- 1. Leak in bag hose or bag (unknown)
- 2. Leak in fresh gas hose or flow meter circuit (unknown)
- 3. Leak in vaporizer mount or ORC (unknown)
Execute manual leak test for subsystems 1, 2, 3
- Test point
- Full breathing system and vaporizer circuit
- Expected
- Leak rate displayed in mL/min; zero or minimal acceptable leakage
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen supply pressure alarms patient safety
Low O2 supply pressure alarm does not trigger when oxygen supply is low
Ranked causes
- 1. O2 supply pressure switch faulty or disconnected (common)
- 2. Switch wiring or System I/O cable connection faulty (occasional)
- 3. WPU (Work Processing Unit) failed (rare)
Disconnect switch and test continuity
- Test point
- Switch continuity
- Expected
- Continuity present, switch operates properly
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature 1 reading incorrect or not displayed
T1 TEMP parameter window does not appear on monitor display or reading is not 37.0°C ±0.2°C when simulator is set to 37°C output
Ranked causes
- 1. Incorrect simulator configuration for temperature output (common)
- 2. Wrong temperature adapter cable connected (occasional)
- 3. Temperature connector not properly connected to T1 (occasional)
Verify simulator is configured for 37°C temperature output
- Test point
- Simulator Code entry
- Expected
- Code 191 entered and accepted
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unusual Divan noise during operation caution
Machine produces abnormal creaking, grinding, or hum sounds during breathing cycles
Ranked causes
- 1. Motor drive belt requires lubrication (common)
- 2. Piston-cylinder valve rod requires lubrication (common)
- 3. Non-return valve in Breasy defective or worn (occasional)
Determine if noise is reproducible
- Test point
- Divan operation with standard settings
- Expected
- Noise reproduces consistently with same FGF, Pmax, Vt, %IP, PEEP, Rate
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Double Vapor Exclusion System alignment
Right mounted vaporizer difficult to engage or exclusion bar does not properly slide into vaporizer cap due to excessive sagging
Ranked causes
- 1. Misaligned chassis bracket causing vaporizer sagging (common)
- 2. Obstruction preventing proper shim installation (occasional)
Remove vaporizer from right exclusion mount and bottom cover from cockpit area
- Test point
- Right mounted vaporizer and cockpit area
- Expected
- Bottom cover removed, vaporizer accessible
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff inflation cycle malfunction
Cuff does not inflate repetitively or inflation-to-inflation interval exceeds three seconds when STAT measurement initiated.
Ranked causes
- 1. Cuff connection or inflation valve malfunction (unknown)
- 2. STAT cycle timing too slow (unknown)
Touch STAT key and observe cuff inflation and deflation cycle
- Test point
- NIBP window, STAT control, cuff
- Expected
- Cuff inflates and deflates repetitively
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP diastolic alarm function verification
Verify diastolic pressure alarm triggers and ceases correctly at low and high setpoints during BP measurement cycle.
Ranked causes
- 1. Diastolic low alarm setpoint incorrectly configured (unknown)
- 2. Diastolic high alarm setpoint incorrectly configured (unknown)
Touch PRESSURE key and toggle to Dias
- Test point
- Alarms tab, PRESSURE control
- Expected
- Diastolic pressure parameters displayed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP mean alarm function verification
Verify mean pressure alarm triggers and ceases correctly at low and high setpoints during BP measurement cycle.
Ranked causes
- 1. Mean low alarm setpoint incorrectly configured (unknown)
- 2. Mean high alarm setpoint incorrectly configured (unknown)
Touch PRESSURE key and toggle to Mean
- Test point
- Alarms tab, PRESSURE control
- Expected
- Mean pressure parameters displayed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP systolic alarm function verification
Verify systolic pressure alarm triggers and ceases correctly at low and high setpoints during BP measurement cycle.
Ranked causes
- 1. Systolic low alarm setpoint incorrectly configured (unknown)
- 2. Systolic high alarm setpoint incorrectly configured (unknown)
Install BP cuff on arm and open NIBP window
- Test point
- NIBP interface
- Expected
- NIBP window accessible, cuff installed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Piston movement test patient safety
Manual piston operation test to verify motor and position sensor function
Ranked causes
- 1. Motor malfunction or disconnection (unknown)
- 2. Piston position encoder malfunction (unknown)
Activate piston movement in both directions
- Test point
- Piston mechanism and motor
- Expected
- Piston moves smoothly in both directions when knob turned
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System electronics do not energize at switch ON high voltage
When the main power switch is turned to ON position, the system electronics fail to energize.
Ranked causes
- 1. Main switch not properly adjusted or mechanically binding (common)
- 2. Circuit breakers disabled or power not restored (common)
Verify AC power is connected to rear panel
- Test point
- Rear power supply connector
- Expected
- AC power cord firmly connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Non-Functional
Serious diagnostic problem detected; system will not proceed
Ranked causes
- 1. VPO Pod Firmware failure (unknown)
- 2. Ventilator subsystem failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature 1 high alarm not functioning
TEMP 1 HIGH caution message and intermittent audible alarm do not appear when alarm limit is decreased below actual temperature
Ranked causes
- 1. Temperature 1 alarm not enabled (common)
- 2. Incorrect T1 high alarm limit setting (occasional)
Verify Temperature 1 alarm is enabled and T1 channel is selected
- Test point
- Alarm control and Alarm Limits channel selection key
- Expected
- Alarm control ON, Alarm Limits set to T1 channel
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature 1 low alarm not functioning
TEMP 1 LOW caution message and intermittent audible alarm do not appear when alarm limit is raised above actual temperature
Ranked causes
- 1. Temperature 1 alarm not enabled (common)
- 2. Incorrect T1 low alarm limit setting (occasional)
Verify Temperature 1 alarm is enabled and T1 channel is selected
- Test point
- Alarm control and Alarm Limits channel selection key
- Expected
- Alarm control ON, Alarm Limits set to T1 channel
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature 2 high alarm not functioning
TEMP 2 HIGH caution message and intermittent audible alarm do not appear when alarm limit is decreased below actual temperature
Ranked causes
- 1. Temperature 2 alarm not enabled or T2 not selected (common)
- 2. Incorrect T2 high alarm limit setting (occasional)
Toggle Alarm Limits channel selection to T2 and note settings
- Test point
- Alarm Limits channel selection key and alarm settings display
- Expected
- T2 channel selected; T2 high and low alarm settings visible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature 2 low alarm not functioning
TEMP 2 LOW caution message and intermittent audible alarm do not appear when alarm limit is raised above actual temperature
Ranked causes
- 1. Temperature 2 alarm not enabled or T2 not selected (common)
- 2. Incorrect T2 low alarm limit setting (occasional)
Toggle Alarm Limits channel selection to T2 and note settings
- Test point
- Alarm Limits channel selection key and alarm settings display
- Expected
- T2 channel selected; T2 high and low alarm settings visible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature 2 reading incorrect or not displayed
T2 TEMP parameter window does not appear on monitor display or reading is not 37.0°C ±0.2°C when connected to T2 connector
Ranked causes
- 1. Temperature simulator cable not connected to T2 connector (common)
- 2. Temperature adapter cable damaged or loose connection at T2 (occasional)
Move temperature simulator cable to T2 connector
- Test point
- T2 connector of Integrated Patient Monitoring Module
- Expected
- Cable moved from T1 to T2 connector
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood temperature alarm not triggering patient safety
Caution alarm T BLOOD LO or T BLOOD HI does not appear on central alarm display or audible alarm does not sound when limits are adjusted
Ranked causes
- 1. Alarm Control not enabled or alarm limits not properly adjusted (common)
Toggle Alarm Control key to ON and verify original settings noted
- Test point
- Alarm Control key
- Expected
- Alarm Control enabled
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration too loose with free play remaining at lever caution
After completing calibration procedure, free play detected at top of lever when exclusion knob is at far right position, indicating insufficient tightening.
Ranked causes
- 1. Insufficient upper calibration screw tightening in initial pass (occasional)
Loosen both calibration screws
- Test point
- Upper and lower calibration screws
- Expected
- Several rotations counterclockwise
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarm sound volume adjustment
Reduce maximum alarm annunciation sound volume level from 100% to 50% by adjusting Windows NT master volume via keyboard and control panel commands
Ranked causes
- 1. Customer request to reduce alarm volume (common)
Power on unit and reach startup completion
- Test point
- Divan Standby key
- Expected
- Unit completes start-up
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarm speaker functionality verification
Verify speaker operation and audio quality after volume adjustment at both maximum and minimum levels
Ranked causes
- 1. Post-calibration speaker verification required (common)
Listen for tone at maximum volume setting
- Test point
- Unit speaker output, Volume slider maximum
- Expected
- Audible tone present
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Breathing Pressure Monitor
Zero and span calibration procedure for breathing pressure monitor to ensure accurate pressure measurement
Ranked causes
- 1. Calibration required for accurate pressure readings (unknown)
Turn System Power switch to ON
- Test point
- System Power switch
- Expected
- ON position
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Breathing system leveling adjustment patient safety
Breathing system top surface not parallel to back edge of ventilator chassis; requires leveling via eccentric mounts on left side pins
Ranked causes
- 1. Breathing system not level during installation or after maintenance (common)
Turn System Power Switch to STANDBY
- Test point
- Power switch
- Expected
- STANDBY position
13 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CRT alignment high voltage
CRT picture size, centering, and rotation require adjustment
Ranked causes
- 1. CRT alignment adjustment needed (unknown)
Remove upper back cover
- Test point
- Machine upper back
- Expected
- Cover removed safely
7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cylinder Pressure Regulator Adjustment high voltage
Set cylinder pressure regulators to correct output pressure for O2, N2O, and Air cylinders using test pressure gauge at regulator test ports.
Ranked causes
- 1. Regulator pressure requires setting or verification (common)
Disconnect all pipeline hoses
- Test point
- Pipeline connections
- Expected
- All hoses disconnected
13 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas Analyzer Sample Flow Rate Adjustment
Set gas analyzer sample flow rate using test flowmeter.
Ranked causes
- 1. Gas analyzer sample flow rate requires setting (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive Blood Pressure (BP1/ART)
Verify invasive blood pressure sensitivity calibration at 5µV/V1 mmHg and test ART1 connector with simulator cable
Ranked causes
- 1. Calibration procedure for BP1 invasive pressure monitoring (common)
Filter frequency setting
- Test point
- BP1 Setup tab
- Expected
- 12
13 more checks with pass and fail actions, 16 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive Blood Pressure (BP2/CVP2)
Verify CVP2 central venous pressure sensitivity calibration and test CVP2 connector with simulator cable
Ranked causes
- 1. Calibration procedure for BP2/CVP2 invasive pressure monitoring (common)
Filter frequency setting
- Test point
- BP2/CVP2 Setup tab
- Expected
- 12
9 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive Blood Pressure (BP3/PA3)
Verify PA3 pulmonary artery pressure sensitivity calibration and test PA3 connector with simulator cable
Ranked causes
- 1. Calibration procedure for BP3/PA3 invasive pressure monitoring (common)
Filter frequency setting
- Test point
- BP3/PA3 Setup tab
- Expected
- 12
13 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive Blood Pressure (BP4/FEM4)
Verify FEM4 femoral artery pressure sensitivity calibration if applicable; test BP4 connector with simulator cable
Ranked causes
- 1. Calibration procedure for BP4/FEM4 invasive pressure monitoring (if applicable) (common)
Filter frequency setting
- Test point
- BP4/FEM4 Setup tab
- Expected
- 12
9 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Ratio Controller Adjustment
Adjust Oxygen Ratio Controller using oxygen monitor.
Ranked causes
- 1. Oxygen Ratio Controller requires calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Right vaporizer vertical adjustment
Difficulty engaging right vaporizer; exclusion system operation needs verification or correction
Ranked causes
- 1. Right vaporizer misalignment requiring set screw adjustment (unknown)
Power system to STANDBY
- Test point
- System Power switch
- Expected
- STANDBY position
11 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System electrical switch adjustment high voltage
Adjustment of the main power switch, lever switch, and paddle to ensure correct actuation and overtravel on the Narkomed 6000 anesthesia machine.
Ranked causes
- 1. Switch requires mechanical alignment and paddle adjustment (common)
Turn System Power switch to STANDBY
- Test point
- Main power switch
- Expected
- STANDBY position
15 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Test all vaporizers on right exclusion mount
Verify exclusion system operation for all vaporizers installed on NM6000 Series
Ranked causes
- 1. Multiple vaporizers may require individual alignment adjustment (occasional)
Install each additional vaporizer in right exclusion mount
- Test point
- Right exclusion mount
- Expected
- Vaporizer engages properly with exclusion system
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch Screen alignment
Touch Screen calibration required after CRT alignment
Ranked causes
- 1. Touch Screen calibration required following CRT adjustment (unknown)
Perform Touch Screen calibration
- Test point
- Touch Screen display
- Expected
- Calibration completed per Section 5.10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch Screen and Alarms Adjustment
Perform advanced touch screen calibration, adjust alarm volume, and set Time Zone using keyboard.
Ranked causes
- 1. Touch screen, alarm volume or time zone requires adjustment (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Triple Vapor Mount patient safety
Selector knob exclusion mechanism requires calibration to ensure proper relationship between selector knob position and upper exclusion pin engagement, preventing inadvertent vaporizer selection and ensuring smooth knob travel.
Ranked causes
- 1. Tolerance stack-up in exclusion mechanism assembly (common)
Remove back cover of Triple Vapor Mount
- Test point
- Triple Vapor Mount assembly
- Expected
- Back cover removed per Section 4.8.10
12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Triple Vapor Mount upper and left position rest height caution
Verify and adjust vaporizer mount height to ensure proper slider bar alignment and clearance with vaporizer sockets or D-Tec pin alignment.
Ranked causes
- 1. Vaporizer height misalignment (unknown)
Verify slider bar clearance with vaporizer socket
- Test point
- Upper slider bar position against Vapor 19.n or Vapor 2000 socket
- Expected
- Equal clearance at top and bottom
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator control pressure adjustment patient safety
87 mbar control pressure regulator requires adjustment to ensure proper ventilator operation; procedure involves accessing the control pressure assembly and setting pneumatic pressure to specification.
Ranked causes
- 1. Control pressure regulator out of specification (unknown)
Verify breathing system diaphragm integrity at location Y2
- Test point
- Y2
- Expected
- No leakage
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator Oxygen Supply Flow Rate Adjustment
Set ventilator oxygen supply flow rate using test flowmeter.
Ranked causes
- 1. Ventilator oxygen supply flow rate requires setting (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator Supply Pressure - Verification Step
Verify ventilator supply control pressure is within specification 1.4 to 1.6 bar after adjustment.
Ranked causes
- 1. Control pressure verification required after regulator adjustment (common)
Read pressure meter after locking regulator adjustment knob
- Test point
- Pressure meter display
- Expected
- 1.4 to 1.6 bar (20.3 to 23 psi), 1.4 to 1.6 bar or 20.3 to 23 psi
- Tools
- Meriam 350 Smart Manometer or equivalent
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator Supply Pressure Adjustment
Adjust ventilator O2 supply regulator pressure to 1.5 bar (21.8 psi) on Narkomed 6000 anesthesia machine.
Ranked causes
- 1. Ventilator supply pressure requires calibration adjustment (common)
Verify breathing system diaphragm integrity at location Y2
- Test point
- Location Y2
- Expected
- Diaphragm intact
14 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot remove ventilator piston-cylinder unit caution
Piston unit handle cannot be lifted to full up position or unit is stuck in chassis
Ranked causes
- 1. U-shaped channel on piston coupling assembly in wrong position (common)
Verify system power in STANDBY mode
- Test point
- System Power switch
- Expected
- Switch set to STANDBY
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Cardiac output monitor reading out of range
- CO injectate temperature reading out of acceptable range
- Disable P-SET NOT MET ALARM (optional)
- Error: Breathing system does not seat properly on pins after adjustment
- Event log transfer completes but files incomplete
- Exclusion knob binds during travel in T-slot
- Exclusion knob cannot reach far right end of T-slot
- Insufficient transferred file size
- Lever free play not eliminated after upper screw tightening
- NIBP and SpO2 alarm interaction malfunction
- Oxygen sensor will not calibrate properly
- PEEP valve operation test
- Pressure value display
- System Conditionally Functional
- Temperature test accuracy drift
- Touch screen cursor does not appear or does not move properly
- Upper calibration screw loosens when knob moved to vertical position
- Verify power supply voltages
18 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0000 | Operating system error | System startup or operation | critical |
0001 | Unknown | Unknown | unknown |
0003 | Unknown | Unknown | unknown |
0Axx | Unknown | Unknown | unknown |
1001 to 1005 | Unknown | Unknown | unknown |
1007 | Reset command received. Module restarted. | Reset command received | info |
1008 | Internal Software error | Software execution failure | error |
1009 | No data from Oxigraf sensor. Oxigraf restarted. | Oxigraf sensor communication loss | error |
1020 to 1030 | Checksum error | Data integrity check failure | warning |
1040 | The two sensors on MOPS PCB measure the same parameter (e.g. O2) | Duplicate parameter detection | info |
1111 | Module restarted | Module restart event | info |
1112 | Module reset itself | Automatic module reset | warning |
1113 | The temperature of the filter wheel or sample cell exceeds 65 degrees C | Sensor head temperature exceeds threshold | error |
1114 | Sensor is in ISO or F mode and light-source voltage is incorrect | Incorrect light-source voltage detected | error |
1115 | IRIA Sensor Head error | Sensor head malfunction | error |
1116 | Unknown | Unknown | unknown |
1117 | IRIA Sensor Head error (GAP-2 specific) | Sensor head malfunction in GAP-2 | error |
1118 | IRIA Sensor Head error | Sensor head malfunction | error |
1119 | Unknown | Unknown | unknown |
111A | 12 V supply voltage lies outside of valid range | Supply voltage out of specification | warning |
111B | Unknown | Unknown | unknown |
111C | 15 V supply voltage outside of valid range | Supply voltage out of specification | warning |
111D | Unknown | Unknown | unknown |
111E | 6.4 V supply voltage outside of valid range | Supply voltage out of specification | warning |
111F | Unknown | Unknown | unknown |
1200 | Incorrect checksum | Checksum verification failure | critical |
1201 | Incorrect ROM checksum | ROM checksum verification failure | critical |
1202 | Incorrect calibration data | Calibration data verification failure | error |
2111 to 2115 | Unknown | Unknown | unknown |
2116 | Unknown | Unknown | unknown |
2117 | Unknown error code | Internal failure | error |
2118 | Internal software error | Software execution failure | error |
2180 | One correction factor during spectral correction is zero. Correction cannot be carried out. | Spectral correction factor is zero | error |
2200 | Water-trap self-test showed an error | Water trap self-test failure | error |
3111 | The absolute pressure value is zero | Pressure sensor returns zero value | error |
3112 | Pressure sensor conversion table not calculated | System conversion table missing during pressure conversion | error |
3113 | Pressure sensor calibration data could not be read | Calibration data read failure | error |
3114 | EEPROM error | EEPROM read/write failure | error |
3115 | Unknown | Unknown | unknown |
3116 | Unknown | Unknown | unknown |
3117 | Pressure sensor calibration error | Pressure sensor calibration failure | error |
3200 | No Pressure-sensor calibration available | Pressure sensor calibration data missing | error |
3300 | NOVRAM error | NVRAM read/write failure | error |
3400 | No Factory Zero Data available on EEPROM | Factory calibration data missing | error |
3401 | No zero values available on NOVRAM | Zero calibration data missing | warning |
3402 | Real-time clock error | RTC malfunction or synchronization failure | warning |
3403 to 3408 | Incorrect data from O2 Sensor | O2 sensor data validation failure | warning |
3420 | Module startup failure or incorrect configuration | Initialization failure or configuration error | error |
3501 | Factory Zero was incorrect | Factory zero calibration invalid | error |
3502 | Zeroing failed because sufficient valid zero values could not be detected | Zero calibration procedure failure | error |
3503 to 3405 | Zeroing failed | Zero calibration procedure failure | error |
3506 to 3509 | Factory zeroing failed because valid zero values could not be detected | Factory calibration failure during startup | error |
350A to 350D | Zeroing failed because values of one reference filter were outside permissible range | Reference filter calibration out of range | error |
350E to 3511 | Zeroing failed because zero values deviate considerably from factory zeroing | Zero calibration drift detected | error |
3512 to 3515 | Zeroing failed because measured values differ considerably | Zero calibration instability | error |
3516 | Zeroing failed due to excessive negative pressure in pneumatic system during purging | Excessive negative pressure during purge cycle | error |
3517 | Zeroing failed due to insufficient negative pressure in pneumatic system or positive pressure during purging | Inadequate purge pressure during calibration | error |
3518 to 3520 | Incorrect signal from O2 Sensor | O2 sensor signal validation failure | info |
3A11 | Status of pump was not recognized | Pump status detection failure | error |
4000 | Motor is blocked and will be switched off | Motor stalled or blocked condition detected | critical |
5000 | Internal software error | Software execution failure | error |
7000 to 700C | Errors with sensor specific monitoring | Sensor monitoring system failure | error |
8000 | Error with numerical filter 1 | Digital filter processing failure | error |
Fxxx | Internal software error | Software execution failure | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.