Dräger Narkomed 4: troubleshooting and error codes

The ServiceForge knowledge base holds 59 documented symptoms for the Dräger Narkomed 4, with 127 ranked causes, 237 diagnostic checks with test points and expected values, and 155 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

O.R. DATA MANAGER display blank

O.R. DATA MANAGER remote display screen shows no output; user cannot access data manager functions

Ranked causes

  1. 1. Remote display not plugged in (occasional)
  2. 2. Interface panel cable disconnected (occasional)
  3. 3. Remote display defective (common)
  4. 4. Backplane assembly defective (common)
  5. 5. Processor board defective (NM4 mode only) (occasional)
  6. 6. Backup processor board defective (NM4 mode only) (occasional)
  7. 7. O.R. DATA MANAGER CPU assembly defective (common)
  8. 8. O.R. DATA MANAGER video cable defective or disconnected (occasional)
First check

Test if display blank in both NM4 and ORDM modes

Test point
Display output
Expected
Display functional in both modes

14 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence function failure patient safety

Silence button or key not functioning at main or remote alarm panels

Ranked causes

  1. 1. Backplane assembly defect (common)
  2. 2. Remote display malfunction (occasional)
  3. 3. Front bezel assembly failure (occasional)
  4. 4. Processor board failure (occasional)
  5. 5. Internal wiring harness disconnection (rare)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DataScan function not logging data

DataScan readings not present on remote display; data logging not functioning

Ranked causes

  1. 1. Processor board defect (common)
  2. 2. Backplane assembly failure (occasional)
  3. 3. Front bezel assembly failure (occasional)
  4. 4. Remote display malfunction (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disk error messages on remote display

System displays disk error messages including disk full, invalid, not inserted, or general error during operation.

Ranked causes

  1. 1. Disk not inserted or not detected (common)
  2. 2. Disk invalid or corrupted (common)
  3. 3. Disk full or insufficient space (common)
  4. 4. CPU assembly defective or disk drive failure (occasional)
First check

Is disk not inserted error on remote display

Test point
Remote display messages
Expected
No disk not inserted message

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External printer not functioning

External printer does not function or does not respond to print commands from system.

Ranked causes

  1. 1. External printer power not connected (common)
  2. 2. External printer cable not connected to system (common)
  3. 3. Internal printer cable disconnected from CPU or backplane (common)
  4. 4. ORDM CPU assembly defective (occasional)
First check

Is external printer power connected

Test point
Printer power cord
Expected
Power cord connected and printer powered on

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of blood pressure monitor patient safety

NIBP (non-invasive blood pressure) monitor error, cuff error, or cuff disconnect displayed

Ranked causes

  1. 1. Backplane assembly failure (common)
  2. 2. NIBP pump and sensor circuit assembly failure (occasional)
  3. 3. Processor board defect (occasional)
  4. 4. NIBP cuff leak or disconnection (common)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of GAS monitor patient safety

CO2/AGT error or CO2 line blockage indicated on central alarm panel

Ranked causes

  1. 1. Processor board failure (common)
  2. 2. Backplane assembly defect (occasional)
  3. 3. MULTISPEC analyzer failure (occasional)
  4. 4. Occluded sample line or water trap (common)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of pulse oximetry monitor patient safety

SpO2 monitor display panel blank or oximeter error indicated

Ranked causes

  1. 1. Processor board defect (common)
  2. 2. Backplane assembly failure (occasional)
  3. 3. SpO2 sensor circuit assembly failure (occasional)
  4. 4. SpO2 cable disconnection (common)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ORDM communication error on remote display

Remote display shows ORDM communication error advisory. Communications between main and backup processors may be affected.

Ranked causes

  1. 1. Data cable disconnected or loose between CPU and backplane (common)
  2. 2. ORDM CPU communication failure (occasional)
  3. 3. Backup processor board failure (occasional)
  4. 4. Main processor board failure (occasional)
First check

Is ORDM COM error present on remote display

Test point
Remote display advisory
Expected
No ORDM COM error

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vitalink communication failure

Host fails to communicate with peripheral device; unit may be non-functional or display monitoring data

Ranked causes

  1. 1. Processor board defective (common)
  2. 2. Backplane assembly defective (common)
  3. 3. Serial interface assembly defective (common)
  4. 4. Peripheral cable disconnected or misconfigured (occasional)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display arm tension adjustment

Display arm does not maintain position when raised or lowered, or excessive free swing at machine mount joint

Ranked causes

  1. 1. Friction adjustment screw too loose or over-tightened (common)
  2. 2. Excessive free swing at machine mount joint (occasional)
  3. 3. Remote display assembly mounting screw loose (occasional)
First check

Verify display arm returns to approximately horizontal position after raising and releasing

Test point
Display arm
Expected
Returns to approximately horizontal position

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve dome pins or discs damaged patient safety

Expiratory valve does not seal or pass flow; may cause gas leakage or backpressure in expiratory breathing circuit.

Ranked causes

  1. 1. Mechanical damage to expiratory valve crater pins (occasional)
  2. 2. Mechanical damage to expiratory valve dome pins (occasional)
  3. 3. Expiratory valve disc worn or deteriorated (occasional)
First check

Inspect expiratory valve crater pins

Test point
Expiratory valve crater
Expected
No damage to pins

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory valve dome pins or discs damaged patient safety

Valve does not seal or pass flow properly; may cause gas leakage or resistance in inspiratory breathing circuit.

Ranked causes

  1. 1. Mechanical damage to inspiratory valve crater pins (occasional)
  2. 2. Mechanical damage to inspiratory valve dome pins (occasional)
  3. 3. Inspiratory valve disc worn or deteriorated (occasional)
First check

Inspect inspiratory valve crater pins

Test point
Inspiratory valve crater
Expected
No damage to pins

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal printer not working

Printer does not respond or produce output; recording function unavailable

Ranked causes

  1. 1. Printer cable disconnected (occasional)
  2. 2. Backplane assembly defective (common)
  3. 3. Internal printer defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard error or keyboard not present during power-up

System detects keyboard error or keyboard not found during power-up. Keyboard may be disconnected or defective.

Ranked causes

  1. 1. Keyboard cable disconnected from CPU board J5 (common)
  2. 2. Keyboard assembly defective (occasional)
  3. 3. CPU assembly defective (occasional)
First check

Is keyboard cable connected to CPU

Test point
CPU board J5 and monitor box floating connector
Expected
Connected

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of breathing pressure monitor patient safety

Barometric pressure monitor (BAROMED) error displayed on central alarm panel

Ranked causes

  1. 1. Processor board defect (common)
  2. 2. VPO board failure (occasional)
  3. 3. Pilot line disconnection (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of O2 monitor patient safety

Oxygen monitor display absent or error indicated on central alarm panel

Ranked causes

  1. 1. Processor board failure (common)
  2. 2. VPO board malfunction (occasional)
  3. 3. Oxygen sensor failure (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of respiratory volume monitor patient safety

Spirometric error (SPIROMED ERR) or volume sensor disconnect (VOL SEN DISC) displayed

Ranked causes

  1. 1. Backplane assembly failure (common)
  2. 2. VPO board defect (occasional)
  3. 3. Respiratory volume sensor failure (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

New Patient Alarm not displayed patient safety

New patient alarms absent from central alarm panel display

Ranked causes

  1. 1. Remote display malfunction (common)
  2. 2. Backplane assembly failure (occasional)
  3. 3. Processor board defect (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No audio alarms patient safety

Warnings and cautions not audibly announced; speaker output absent

Ranked causes

  1. 1. Backplane PCB failure (common)
  2. 2. Processor board defect (occasional)
  3. 3. Speaker failure or wiring issue (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen supply pressure alarm not functional patient safety

Oxygen supply pressure alarm does not trigger when required; critical safety monitoring lost

Ranked causes

  1. 1. Alarm channel assembly defective (common)
  2. 2. Oxygen supply pressure alarm switch defective (common)
  3. 3. Wire harness damaged or disconnected (occasional)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen/nitrous oxide ratio alarm not functional patient safety

O2/N2O ratio monitoring alarm does not trigger when required; patient safety monitoring compromised

Ranked causes

  1. 1. Backplane assembly defective (common)
  2. 2. ORMC adjustment out of specification (occasional)
  3. 3. Alarm channel assembly defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen failure

Touch screen not responsive to user input or touch commands not registering

Ranked causes

  1. 1. Processor board defect (common)
  2. 2. Backplane assembly failure (occasional)
  3. 3. Front bezel assembly failure (occasional)
First check

Unit functional and passes PMS

Test point
System startup
Expected
Unit operational, PMS completed successfully

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

B drive error messages on remote display

System displays B drive error messages during operation or at power-up.

Ranked causes

  1. 1. CPU assembly defective (common)
  2. 2. Memory battery defective or depleted (common)
First check

Are any B drive errors shown on remote display

Test point
Remote display messages
Expected
No B drive errors

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display screens blank on power-up

Both display panels remain blank when system powered on; no display output visible

Ranked causes

  1. 1. Front bezel assembly defective (common)
  2. 2. Backplane assembly defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement data incorrect patient safety

Analyzer displays or reports incorrect measurement values; patient monitoring data unreliable

Ranked causes

  1. 1. Cooling fans not operating (common)
  2. 2. Multispec analyzer assembly defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No keypanel or selection dial response

Keypanel buttons and/or selection dial do not respond to user input; unit not controllable

Ranked causes

  1. 1. Backplane assembly defective (common)
  2. 2. Front bezel assembly defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply voltage out of tolerance or backup battery mismatch high voltage

Red LED DS2 on power supply board lighted indicating voltage fault or battery level difference

Ranked causes

  1. 1. One or more DC-DC converter output voltages out of range (common)
  2. 2. Backup battery voltage levels differ by more than 1.1 volt (occasional)
First check

Verify all voltages present at backplane connectors J28, J29, J30

Test point
Backplane connectors J28, J29, J30
Expected
All voltages within acceptable ranges per Table 3-1
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Processor test failure detected

Main or Back-Up processor board reports failure during self-diagnostics on power-up; green lamp on processor board indicates power status

Ranked causes

  1. 1. Main processor board defective (unknown)
  2. 2. Back-Up processor board defective (unknown)
First check

Observe display during power-up self-diagnostics

Test point
Display
Expected
PASS or FAIL result shown for each test

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2/Pulse monitor not reading patient safety

Monitor fails to display oxygen saturation or pulse rate values after sensor connection

Ranked causes

  1. 1. Nellcor sensor or pre-amp cable disconnected or faulty (common)
  2. 2. Novametrix sensor extension cable disconnected or faulty (common)
First check

Verify Nellcor MP-202 module oxygen saturation reading with pocket simulator

Test point
Monitor display
Expected
80 to 82 for oxygen saturation, 60 to 62 for pulse rate %, 80-82 SpO2, 60-62 bpm

10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gas Analyzer Monitor patient safety

Span calibration procedure for multispec monitor accessible via Service/Test Hardware screen

Ranked causes

  1. 1. Gas analyzer (multispec) monitor requires span calibration (common)
First check

Access Gas Analyzer Monitor calibration screen

Test point
Service/Test Hardware screen
Expected
GAS ANALYZER MONITOR option accessible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Machine mount joint shoulder screw torque

Display arm has excessive free swing at the machine mount joint

Ranked causes

  1. 1. Shoulder screw at machine mount joint under-torqued (common)
First check

Tighten shoulder screw at machine mount joint to correct torque

Test point
Machine mount joint shoulder screw
Expected
75 in. lbs.
Tools
Torque wrench

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP inflation pressure patient safety

Verify that the NIBP module inflates the blood pressure cuff to the correct pressure during operation.

Ranked causes

  1. 1. NIBP module pressure calibration out of specification (unknown)
First check

Verify external pressure gauge reading during cuff inflation

Test point
External pressure gauge in test fixture TEE
Expected
180 mm Hg
Tools
External pressure gauge, test fixture TEE assembly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP Monitor patient safety

Cuff inflation pressure verification procedure accessible via Service/Test Hardware screen

Ranked causes

  1. 1. NIBP monitor cuff inflation pressure requires verification (common)
First check

Access NIBP Monitor calibration verification screen

Test point
Service/Test Hardware screen
Expected
NIBP MONITOR option accessible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Oxygen sensor zero patient safety

Oxygen sensor requires zero calibration adjustment to establish baseline cell current values before clinical use.

Ranked causes

  1. 1. Oxygen sensor zero calibration required (common)
First check

Remove oxygen sensor assembly from inspiratory valve dome

Test point
Inspiratory valve dome
Expected
Sensor assembly removed

8 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure Monitor patient safety

Zero and span calibration procedures for pressure monitor accessible via Service/Test Hardware screen

Ranked causes

  1. 1. Pressure monitor requires zero or span calibration (common)
First check

Access Pressure Monitor calibration screen

Test point
Service/Test Hardware screen
Expected
PRESSURE MONITOR option accessible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remote display assembly mounting torque

Remote display assembly or Datagriп mounting needs adjustment for proper tension

Ranked causes

  1. 1. Remote display assembly mounting screw under-torqued (common)
First check

Remove plastic cap at end of display arm

Test point
Display arm end cap
Expected
Cap removed to expose mounting screw

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve washer under flow sensor degraded caution

Washer does not seal properly under expiratory valve; may cause gas leakage or improper flow sensor function.

Ranked causes

  1. 1. Washer under expiratory valve worn or deteriorated (occasional)
First check

Inspect washer under expiratory valve

Test point
Washer under expiratory valve
Expected
Good condition, not worn or cracked

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect display screen shown

Wrong monitor screen displayed; unit may not show expected interface or data

Ranked causes

  1. 1. Backplane assembly defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose interconnecting cables caution

Ribbon cables or wire harnesses between suspected assemblies not properly seated

Ranked causes

  1. 1. Improper cable connection during assembly or service (occasional)
First check

Visual inspection of cable connections

Test point
Between processor boards and peripheral devices
Expected
All cables fully seated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

MAN/AUTO selector valve sticking or not responding caution

Selector valve does not switch smoothly between manual and automatic modes; may require excessive force or fail to engage.

Ranked causes

  1. 1. Excessive lubricant residue in valve channel or body (common)
First check

Inspect valve channel and body for residual lubricant

Test point
Valve channel and valve body interior
Expected
Clean, free of lubricant residue

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing backplane voltages high voltage

Voltage measurements missing at backplane connectors J28, J29, or J30 during startup or operation

Ranked causes

  1. 1. Power supply failure or malfunction (unknown)
First check

Verify all voltages present at backplane

Test point
J28, J29, J30 or J15
Expected
All voltages per Table 3-1
Tools
DMM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Multispec analyzer not in full accuracy mode

CO2 and Agent warm-up messages appear on advisory display, preventing valid accuracy testing

Ranked causes

  1. 1. Multispec warm-up cycle in progress (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O-rings on oxygen sensor housing or plug degraded patient safety

O-rings do not seal; oxygen sensor assembly leaks or does not mount properly on inspiratory valve dome.

Ranked causes

  1. 1. O-rings on oxygen sensor housing or plug cracked, dried, or deteriorated (occasional)
First check

Inspect O-rings on oxygen sensor housing and plug

Test point
O-rings on O2 sensor housing and oxygen sensor plug
Expected
O-rings intact, pliable, no cracks or drying

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 supply alarm activates outside setpoint range patient safety

Alarm triggers below 34 psi or above 40 psi instead of within the correct 40–34 psi band

Ranked causes

  1. 1. Oxygen supply pressure alarm switch setpoint out of adjustment (common)
First check

Verify test gauge is connected to TEE fitting in O2 regulator output line

Test point
TEE fitting, 1/4 in. diameter O2 regulator output line
Expected
Test gauge properly connected and reading pressure
Tools
O2 test gauge

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen sensor cell current values

Verify recorded oxygen sensor fuel cell A and cell B zero calibration current values for documentation.

Ranked causes

  1. 1. Zero calibration values recorded (common)
First check

Verify CURRENT CELL A and CELL B values displayed before STORE ZERO

Test point
Display screen current cell readout
Expected
CURRENT CELL A: 3, CURRENT CELL B: 6 (or equivalent current values for the sensor)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen sensor replacement patient safety

Sensor located on inspiratory valve requires replacement or maintenance

Ranked causes

  1. 1. Oxygen sensor capsule degradation or failure (unknown)
First check

Set System Power switch to STANDBY

Test point
System Power switch
Expected
STANDBY position

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen supply alarm activates above setpoint range patient safety

Alarm activates when oxygen pressure is above 40 psi instead of between 40 and 34 psi.

Ranked causes

  1. 1. Alarm switch setpoint too high (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen supply alarm activates below setpoint range patient safety

Alarm activates when oxygen pressure falls below 34 psi instead of between 40 and 34 psi.

Ranked causes

  1. 1. Alarm switch setpoint too low (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply self-test failure high voltage

Red LED DS4 remains lighted after power-up indicating self-test failure of monitoring and communication circuitry

Ranked causes

  1. 1. Power supply board monitoring or communication circuit fault (common)
First check

Observe LED DS4 status during power-up self-test

Test point
Power supply board LED DS4 (red)
Expected
LED DS4 lighted during power up, off after self-test completion

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply voltages out of tolerance caution

Measured power supply voltages do not meet specifications; power distribution compromised

Ranked causes

  1. 1. Power supply circuit board defective (common)
First check

Complete preventive maintenance

Test point
Unit functional status
Expected
Unit passes PMS

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power-up diagnostic test failure or setup error

One or more power-up diagnostic tests fail or setup errors occur during system initialization.

Ranked causes

  1. 1. CPU assembly defective or misconfigured (common)
First check

Do any power-up diagnostic tests fail or setup errors occur

Test point
System power-up initialization
Expected
All diagnostic tests pass, no setup errors

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace SPIROMED respiratory volume sensor caution

Sensor replacement procedure for the respiratory volume sensor installed between absorber assembly and expiratory valve

Ranked causes

  1. 1. Sensor requires replacement (unknown)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Spiromed sensor lubrication required caution

Sensor operation stiff, sluggish, or making noise; lubrication depletion or contamination.

Ranked causes

  1. 1. Spiromed sensor insufficient or depleted lubricant (common)
First check

Locate lubrication access holes on spiromed sensor

Test point
Four lateral holes at sides of spiromed sensor
Expected
Holes visible and accessible, marked by arrows

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Spiromed sensor washer degraded caution

Washer under sensor does not seal properly; may cause gas leakage or sensor function issues.

Ranked causes

  1. 1. Washer under spiromed sensor worn or deteriorated (occasional)
First check

Inspect washer under spiromed sensor

Test point
Washer under spiromed sensor
Expected
Good condition, not worn or cracked

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 monitor battery depleted

Novametrix simulator not providing consistent readings or readings unstable

Ranked causes

  1. 1. Novametrix simulator 9V battery voltage below 8.5 VDC (occasional)
First check

Measure Novametrix simulator battery voltage

Test point
Battery terminals
Expected
9.0 VDC minimum for reliable operation, >8.5 VDC for operation, <8.5 VDC requires replacement
Tools
Digital multimeter (DMM)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ultrasonic flow sensor connector hose or O-ring degraded caution

Connector hose leaks, connector does not seal, or O-rings deteriorated; flow sensor does not function or provides inaccurate readings.

Ranked causes

  1. 1. Connector hose, connector, or O-ring components degraded or damaged (occasional)
First check

Inspect connector hose, connector, and O-ring condition

Test point
Ultrasonic flow sensor connector hose and assembly
Expected
Good condition, no cracks or degradation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ultrasonic flow sensor transducers or O-rings degraded caution

Flow sensor provides inaccurate readings or does not function; may indicate volume/flow delivery problems.

Ranked causes

  1. 1. Transducer O-rings or components degraded or damaged (occasional)
First check

Inspect transducer O-rings and components

Test point
Ultrasonic flow sensor transducers
Expected
Good condition, no cracks or degradation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Valve dome washers worn or degraded caution

Washers do not provide proper seal between valve dome and valve body; may cause gas leakage or improper valve function.

Ranked causes

  1. 1. Valve dome washers worn, cracked, or deteriorated (occasional)
First check

Inspect valve dome washers condition

Test point
Valve dome washers
Expected
Good condition, not worn or cracked

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0000Power onSystem power initializationinfo
A001Pattern AdjustedPattern adjustment performedinfo
B Drive Error MessagesBackup drive (B:) reports various errorsDetected during disk access or power-upwarning
C001O2 Store Zero AO2 zero calibration performedinfo
C002O2 Store Zero BO2 zero calibration B performedinfo
C003Line Block Store ReferenceLine blockage reference storedinfo
C004Gas Analyzer - StartGas analyzer calibration startedinfo
C005Gas Analyzer - ContinueGas analyzer calibration continuedinfo
C006NIBP InflateNIBP inflation calibrationinfo
C007NIBP StopNIBP calibration stoppedinfo
C008NIBP Take ReadingNIBP reading takeninfo
C009Respiratory Flow Store SpanRespiratory flow span calibrationinfo
C010Respiratory Flow Store ZeroRespiratory flow zero calibrationinfo
C011Pressure Store SpanPressure span calibrationinfo
C012Pressure Store ZeroPressure zero calibrationinfo
C013NIBP Inflation CheckNIBP inflation check performedinfo
C014NIBP Deflation CheckNIBP deflation check performedinfo
C015NIBP Leak CheckNIBP leak check performedinfo
CMOS Checksum FailCMOS memory checksum error detectedDetected during power-up self testerror
Communications Error with HostO.R. DATA MANAGER cannot communicate with main NM4 hostDetected during power-up or operationerror
DG00Datagrip Internal ErrorDatagrip module internal errorerror
DG01Datagrip Internal ErrorDatagrip module internal errorerror
DG02Datagrip Internal ErrorDatagrip module internal errorerror
Disk ErrorHard disk or floppy disk error detectedDetected during power-up or disk accesserror
Disk Full ErrorDisk storage capacity exhaustedAttempted to write data when disk fullwarning
Disk Invalid ErrorDisk file system invalid or corruptedDetected during disk access or power-uperror
Disk Not Inserted ErrorFloppy disk not present in driveAttempted to access floppy disk without disk insertedinfo
F011QPDM Address Register ErrorQPDM address register test failureerror
F012QPDM Data Register ErrorQPDM data register test failureerror
F013QPDM Vblank Int Timeout ErrorQPDM vertical blank interrupt timeouterror
F014QPDM VRAM Memory ErrorQPDM video memory test failureerror
F021PROM CRC Test ErrorPROM checksum validation failureerror
F031SRAM Memory Test ErrorSRAM memory test failureerror
F051MFP Bus ErrorMFP bus communication failureerror
F052MFP TimerA ErrorMFP Timer A test failureerror
F053MFP TimerB ErrorMFP Timer B test failureerror
F054MFP TimerC ErrorMFP Timer C test failureerror
F055MFP TimerD ErrorMFP Timer D test failureerror
F056System Clock ErrorSystem clock test failureerror
F057FIFO Empty Interrupt ErrorFIFO empty interrupt test failureerror
F058ADC Interrupt ErrorADC interrupt test failureerror
F059CNTOVF Interrupt ErrorCounter overflow interrupt test failureerror
F061ADC Interrupt ErrorADC interrupt test failureerror
F06212V to -12V Out of Tolerance ErrorPower supply voltage out of rangeerror
F0635V to GND Out of Tolerance Error5V power supply out of rangeerror
F071Primary Speaker Static Test ErrorSpeaker static test failureerror
F072Primary Speaker Dynamic Test ErrorSpeaker dynamic test failureerror
F073Primary Speaker FIFO Test ErrorSpeaker FIFO test failureerror
F074Primary Speaker No Volume Test ErrorSpeaker volume test failureerror
F081Backup Speaker Dynamic Test ErrorBackup speaker dynamic test failureerror
F082Backup Speaker No Volume Test ErrorBackup speaker volume test failureerror
F091SAO2 Speaker Static Test ErrorSAO2 speaker static test failureerror
F092SAO2 Speaker Dynamic Test ErrorSAO2 speaker dynamic test failureerror
F093SAO2 Speaker No Volume Test ErrorSAO2 speaker volume test failureerror
F111Real Time Clock Test ErrorReal time clock test failureerror
F121Bad File WarningFile corruption detectedwarning
F122Bad Files ErrorFile corruption detectederror
F123Event Memory ErrorEvent memory test failureerror
F130Alternate Processor FailAlternate processor failureerror
F200UART1 Internal Init FailUART1 initialization failureerror
F201UART1 Register Readback FailUART1 register read test failureerror
F202UART1 Transmit Interrupt FailUART1 transmit interrupt failureerror
F203UART1 Receive Interrupt FailUART1 receive interrupt failureerror
F204UART1 Receive Error Interrupt FailUART1 receive error interrupt failureerror
F205UART1 Baud Rate Test 1 FailUART1 baud rate test 1 failureerror
F206UART1 Baud Rate Test 2 FailUART1 baud rate test 2 failureerror
F207UART1 Baud Rate Test 3 FailUART1 baud rate test 3 failureerror
F210UART1 Channel 0 FailUART1 channel 0 test failureerror
F211UART1 Channel 1 FailUART1 channel 1 test failureerror
F212UART1 Channel 2 FailUART1 channel 2 test failureerror
F213UART1 Channel 3 FailUART1 channel 3 test failureerror
F214UART1 Channel 4 FailUART1 channel 4 test failureerror
F215UART1 Channel 5 FailUART1 channel 5 test failureerror
F216UART1 Channel 6 FailUART1 channel 6 test failureerror
F217UART1 Channel 7 FailUART1 channel 7 test failureerror
F300UART2 Internal Init FailUART2 initialization failureerror
F301UART2 Register Readback FailUART2 register read test failureerror
F302UART2 Transmit Interrupt FailUART2 transmit interrupt failureerror
F303UART2 Receive Interrupt FailUART2 receive interrupt failureerror
F304UART2 Receive Error Interrupt FailUART2 receive error interrupt failureerror
F305UART2 Baud Rate Test 1 FailUART2 baud rate test 1 failureerror
F306UART2 Baud Rate Test 2 FailUART2 baud rate test 2 failureerror
F307UART2 Baud Rate Test 3 FailUART2 baud rate test 3 failureerror
F310UART2 Channel 0 FailUART2 channel 0 test failureerror
F311UART2 Channel 1 FailUART2 channel 1 test failureerror
F312UART2 Channel 2 FailUART2 channel 2 test failureerror
F313UART2 Channel 3 FailUART2 channel 3 test failureerror
F314UART2 Channel 4 FailUART2 channel 4 test failureerror
F315UART2 Channel 5 FailUART2 channel 5 test failureerror
F316UART2 Channel 6 FailUART2 channel 6 test failureerror
F317UART2 Channel 7 FailUART2 channel 7 test failureerror
H001Head Temperature ErrorHead assembly temperature out of rangeerror
Internal Diagnostic Failure upon System Power-UpO.R. DATA MANAGER self-test detected internal failureExecuted during power-up initializationcritical
Interrupt Controller #1 FailFirst interrupt controller hardware failedDetected during power-up self testcritical
Interrupt Controller #2 FailSecond interrupt controller hardware failedDetected during power-up self testcritical
Keyboard Error or Keyboard Not PresentO.R. DATA MANAGER keyboard not detected or malfunctioningDetected during power-up self testerror
M001Channel Error Active CPUActive CPU channel errorerror
M002Channel Error Inactive CPUInactive CPU channel errorerror
M003Command RetryCommand retry requiredwarning
M004Communication ErrorMSPEC communication failureerror
M005System Fault MaxMaximum system fault threshold reachederror
M006Version ID RetryVersion identification retrywarning
M007Span FailSpan calibration failureerror
M008Solenoid StuckSolenoid valve stuckerror
M009Zero TimeoutZero calibration timeouterror
M010Zero Fail MaxMaximum zero calibration failureserror
M011RX Buffer OverflowReceive buffer overflow conditionerror
M012No Continuous DataContinuous data stream failureerror
M013Line Block DifferentialGas line blockage detectedwarning
M014Line Block Less Than MinimumGas line blockage below minimum thresholdwarning
M016Multispec ShutdownMSPEC module shutdownerror
M101CO2 Channel ErrorCO2 measurement channel errorerror
M102Channel Error O2 Info P1O2 measurement channel error packet 1error
M103Channel Error O2 Info P2O2 measurement channel error packet 2error
Memory Battery FailBattery for memory backup has failed or lost chargeDetected during power-up self testwarning
Memory in Protected Mode FailProtected mode memory access failedDetected during power-up self testcritical
N001NIBP Monitor ErrorNIBP module monitoring failureerror
N002NIBP Communication ErrorNIBP module communication failureerror
N003NIBP Bad DataNIBP data validation failureerror
N004Inflation Pressure Not Greater Than 30NIBP inflation pressure lowwarning
N005Error Received from NIBPNIBP module error indicationerror
N006NIBP Mode ConflictNIBP operating mode conflicterror
NO OXI PULSENovametrix SpO2 sensor signal lost or disconnectedSensor disconnected from finger or cable unplugged for 10 secondswarning
NO SpO2 PULSENellcor SpO2 sensor signal lost or disconnectedSensor disconnected from finger or cable unplugged for 10 secondswarning
OXI ALARMS DISABLEOxygen saturation alarms disabledDepressing both Open Test keys together during module testinfo
OXI SEN DISCOxygen sensor disconnected or not detectedDepressing both Open Test keys or sensor physically disconnectedwarning
P001AC FailAC mains power losterror
P002AC GoodAC mains power restoredinfo
P003Battery LowBattery voltage below thresholdwarning
P004Battery GoodBattery voltage normalinfo
P005Brick BadPower supply module failureerror
Printer InoperativeO.R. DATA MANAGER printer not respondingAttempted to print or detected during self-testwarning
Processor Exception Interrupts FailProcessor exception interrupt handling failedDetected during power-up self testcritical
Protected Mode FailProtected mode operation failedDetected during power-up self testcritical
PULSE LONovametrix pulse signal weak or absentDepressing Open Test RED key during simulator test or loss of sensor signalwarning
S001InstallationInstallation eventinfo
S002PMSPreventive maintenance performedinfo
S003ServiceService performedinfo
S004Secondary Service Screen EntrySecondary service screen accessedinfo
S005Reset Service DateService date resetinfo
S006Service Log ClearService log clearedinfo
Sizing Extended Memory 00384K Not FoundSystem did not detect expected extended memoryDetected during power-up self testerror
Sizing System Memory 640K Not FoundSystem did not detect expected 640K base memoryDetected during power-up self testcritical
System Memory 640K FailBase memory 640K failed self-testDetected during power-up self testcritical
T001Bad Device RecalibrationTouch panel calibration errorerror
TEM0Template Defaults LoadedTemplate defaults loadedinfo
TEM1Recreating Template DatabaseTemplate database recreationinfo
TEM2Templates ReinitializedTemplate reinitializationinfo
TEM3Templates ReinitializedTemplate reinitializationinfo
TEM4Factory Defaults InstalledFactory defaults appliedinfo
V001VTLK Communication ErrorData Manager communication errorerror
V002VTLK Communication LostData Manager link losterror
V003VTLK Communication RegainedData Manager link restoredinfo
V010ORDM Startup FailureOR Data Manager startup failederror
V011ORDM Inactive Startup FailureOR Data Manager inactive startup failederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.