Dräger Narkomed 4: troubleshooting and error codes
The ServiceForge knowledge base holds 59 documented symptoms for the Dräger Narkomed 4, with 127 ranked causes, 237 diagnostic checks with test points and expected values, and 155 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
O.R. DATA MANAGER display blank
O.R. DATA MANAGER remote display screen shows no output; user cannot access data manager functions
Ranked causes
- 1. Remote display not plugged in (occasional)
- 2. Interface panel cable disconnected (occasional)
- 3. Remote display defective (common)
- 4. Backplane assembly defective (common)
- 5. Processor board defective (NM4 mode only) (occasional)
- 6. Backup processor board defective (NM4 mode only) (occasional)
- 7. O.R. DATA MANAGER CPU assembly defective (common)
- 8. O.R. DATA MANAGER video cable defective or disconnected (occasional)
Test if display blank in both NM4 and ORDM modes
- Test point
- Display output
- Expected
- Display functional in both modes
14 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence function failure patient safety
Silence button or key not functioning at main or remote alarm panels
Ranked causes
- 1. Backplane assembly defect (common)
- 2. Remote display malfunction (occasional)
- 3. Front bezel assembly failure (occasional)
- 4. Processor board failure (occasional)
- 5. Internal wiring harness disconnection (rare)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DataScan function not logging data
DataScan readings not present on remote display; data logging not functioning
Ranked causes
- 1. Processor board defect (common)
- 2. Backplane assembly failure (occasional)
- 3. Front bezel assembly failure (occasional)
- 4. Remote display malfunction (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disk error messages on remote display
System displays disk error messages including disk full, invalid, not inserted, or general error during operation.
Ranked causes
- 1. Disk not inserted or not detected (common)
- 2. Disk invalid or corrupted (common)
- 3. Disk full or insufficient space (common)
- 4. CPU assembly defective or disk drive failure (occasional)
Is disk not inserted error on remote display
- Test point
- Remote display messages
- Expected
- No disk not inserted message
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External printer not functioning
External printer does not function or does not respond to print commands from system.
Ranked causes
- 1. External printer power not connected (common)
- 2. External printer cable not connected to system (common)
- 3. Internal printer cable disconnected from CPU or backplane (common)
- 4. ORDM CPU assembly defective (occasional)
Is external printer power connected
- Test point
- Printer power cord
- Expected
- Power cord connected and printer powered on
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of blood pressure monitor patient safety
NIBP (non-invasive blood pressure) monitor error, cuff error, or cuff disconnect displayed
Ranked causes
- 1. Backplane assembly failure (common)
- 2. NIBP pump and sensor circuit assembly failure (occasional)
- 3. Processor board defect (occasional)
- 4. NIBP cuff leak or disconnection (common)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of GAS monitor patient safety
CO2/AGT error or CO2 line blockage indicated on central alarm panel
Ranked causes
- 1. Processor board failure (common)
- 2. Backplane assembly defect (occasional)
- 3. MULTISPEC analyzer failure (occasional)
- 4. Occluded sample line or water trap (common)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of pulse oximetry monitor patient safety
SpO2 monitor display panel blank or oximeter error indicated
Ranked causes
- 1. Processor board defect (common)
- 2. Backplane assembly failure (occasional)
- 3. SpO2 sensor circuit assembly failure (occasional)
- 4. SpO2 cable disconnection (common)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ORDM communication error on remote display
Remote display shows ORDM communication error advisory. Communications between main and backup processors may be affected.
Ranked causes
- 1. Data cable disconnected or loose between CPU and backplane (common)
- 2. ORDM CPU communication failure (occasional)
- 3. Backup processor board failure (occasional)
- 4. Main processor board failure (occasional)
Is ORDM COM error present on remote display
- Test point
- Remote display advisory
- Expected
- No ORDM COM error
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vitalink communication failure
Host fails to communicate with peripheral device; unit may be non-functional or display monitoring data
Ranked causes
- 1. Processor board defective (common)
- 2. Backplane assembly defective (common)
- 3. Serial interface assembly defective (common)
- 4. Peripheral cable disconnected or misconfigured (occasional)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display arm tension adjustment
Display arm does not maintain position when raised or lowered, or excessive free swing at machine mount joint
Ranked causes
- 1. Friction adjustment screw too loose or over-tightened (common)
- 2. Excessive free swing at machine mount joint (occasional)
- 3. Remote display assembly mounting screw loose (occasional)
Verify display arm returns to approximately horizontal position after raising and releasing
- Test point
- Display arm
- Expected
- Returns to approximately horizontal position
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve dome pins or discs damaged patient safety
Expiratory valve does not seal or pass flow; may cause gas leakage or backpressure in expiratory breathing circuit.
Ranked causes
- 1. Mechanical damage to expiratory valve crater pins (occasional)
- 2. Mechanical damage to expiratory valve dome pins (occasional)
- 3. Expiratory valve disc worn or deteriorated (occasional)
Inspect expiratory valve crater pins
- Test point
- Expiratory valve crater
- Expected
- No damage to pins
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory valve dome pins or discs damaged patient safety
Valve does not seal or pass flow properly; may cause gas leakage or resistance in inspiratory breathing circuit.
Ranked causes
- 1. Mechanical damage to inspiratory valve crater pins (occasional)
- 2. Mechanical damage to inspiratory valve dome pins (occasional)
- 3. Inspiratory valve disc worn or deteriorated (occasional)
Inspect inspiratory valve crater pins
- Test point
- Inspiratory valve crater
- Expected
- No damage to pins
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal printer not working
Printer does not respond or produce output; recording function unavailable
Ranked causes
- 1. Printer cable disconnected (occasional)
- 2. Backplane assembly defective (common)
- 3. Internal printer defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard error or keyboard not present during power-up
System detects keyboard error or keyboard not found during power-up. Keyboard may be disconnected or defective.
Ranked causes
- 1. Keyboard cable disconnected from CPU board J5 (common)
- 2. Keyboard assembly defective (occasional)
- 3. CPU assembly defective (occasional)
Is keyboard cable connected to CPU
- Test point
- CPU board J5 and monitor box floating connector
- Expected
- Connected
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of breathing pressure monitor patient safety
Barometric pressure monitor (BAROMED) error displayed on central alarm panel
Ranked causes
- 1. Processor board defect (common)
- 2. VPO board failure (occasional)
- 3. Pilot line disconnection (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of O2 monitor patient safety
Oxygen monitor display absent or error indicated on central alarm panel
Ranked causes
- 1. Processor board failure (common)
- 2. VPO board malfunction (occasional)
- 3. Oxygen sensor failure (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of respiratory volume monitor patient safety
Spirometric error (SPIROMED ERR) or volume sensor disconnect (VOL SEN DISC) displayed
Ranked causes
- 1. Backplane assembly failure (common)
- 2. VPO board defect (occasional)
- 3. Respiratory volume sensor failure (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
New Patient Alarm not displayed patient safety
New patient alarms absent from central alarm panel display
Ranked causes
- 1. Remote display malfunction (common)
- 2. Backplane assembly failure (occasional)
- 3. Processor board defect (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio alarms patient safety
Warnings and cautions not audibly announced; speaker output absent
Ranked causes
- 1. Backplane PCB failure (common)
- 2. Processor board defect (occasional)
- 3. Speaker failure or wiring issue (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen supply pressure alarm not functional patient safety
Oxygen supply pressure alarm does not trigger when required; critical safety monitoring lost
Ranked causes
- 1. Alarm channel assembly defective (common)
- 2. Oxygen supply pressure alarm switch defective (common)
- 3. Wire harness damaged or disconnected (occasional)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen/nitrous oxide ratio alarm not functional patient safety
O2/N2O ratio monitoring alarm does not trigger when required; patient safety monitoring compromised
Ranked causes
- 1. Backplane assembly defective (common)
- 2. ORMC adjustment out of specification (occasional)
- 3. Alarm channel assembly defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen failure
Touch screen not responsive to user input or touch commands not registering
Ranked causes
- 1. Processor board defect (common)
- 2. Backplane assembly failure (occasional)
- 3. Front bezel assembly failure (occasional)
Unit functional and passes PMS
- Test point
- System startup
- Expected
- Unit operational, PMS completed successfully
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
B drive error messages on remote display
System displays B drive error messages during operation or at power-up.
Ranked causes
- 1. CPU assembly defective (common)
- 2. Memory battery defective or depleted (common)
Are any B drive errors shown on remote display
- Test point
- Remote display messages
- Expected
- No B drive errors
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display screens blank on power-up
Both display panels remain blank when system powered on; no display output visible
Ranked causes
- 1. Front bezel assembly defective (common)
- 2. Backplane assembly defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement data incorrect patient safety
Analyzer displays or reports incorrect measurement values; patient monitoring data unreliable
Ranked causes
- 1. Cooling fans not operating (common)
- 2. Multispec analyzer assembly defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No keypanel or selection dial response
Keypanel buttons and/or selection dial do not respond to user input; unit not controllable
Ranked causes
- 1. Backplane assembly defective (common)
- 2. Front bezel assembly defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltage out of tolerance or backup battery mismatch high voltage
Red LED DS2 on power supply board lighted indicating voltage fault or battery level difference
Ranked causes
- 1. One or more DC-DC converter output voltages out of range (common)
- 2. Backup battery voltage levels differ by more than 1.1 volt (occasional)
Verify all voltages present at backplane connectors J28, J29, J30
- Test point
- Backplane connectors J28, J29, J30
- Expected
- All voltages within acceptable ranges per Table 3-1
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor test failure detected
Main or Back-Up processor board reports failure during self-diagnostics on power-up; green lamp on processor board indicates power status
Ranked causes
- 1. Main processor board defective (unknown)
- 2. Back-Up processor board defective (unknown)
Observe display during power-up self-diagnostics
- Test point
- Display
- Expected
- PASS or FAIL result shown for each test
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2/Pulse monitor not reading patient safety
Monitor fails to display oxygen saturation or pulse rate values after sensor connection
Ranked causes
- 1. Nellcor sensor or pre-amp cable disconnected or faulty (common)
- 2. Novametrix sensor extension cable disconnected or faulty (common)
Verify Nellcor MP-202 module oxygen saturation reading with pocket simulator
- Test point
- Monitor display
- Expected
- 80 to 82 for oxygen saturation, 60 to 62 for pulse rate %, 80-82 SpO2, 60-62 bpm
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas Analyzer Monitor patient safety
Span calibration procedure for multispec monitor accessible via Service/Test Hardware screen
Ranked causes
- 1. Gas analyzer (multispec) monitor requires span calibration (common)
Access Gas Analyzer Monitor calibration screen
- Test point
- Service/Test Hardware screen
- Expected
- GAS ANALYZER MONITOR option accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Machine mount joint shoulder screw torque
Display arm has excessive free swing at the machine mount joint
Ranked causes
- 1. Shoulder screw at machine mount joint under-torqued (common)
Tighten shoulder screw at machine mount joint to correct torque
- Test point
- Machine mount joint shoulder screw
- Expected
- 75 in. lbs.
- Tools
- Torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP inflation pressure patient safety
Verify that the NIBP module inflates the blood pressure cuff to the correct pressure during operation.
Ranked causes
- 1. NIBP module pressure calibration out of specification (unknown)
Verify external pressure gauge reading during cuff inflation
- Test point
- External pressure gauge in test fixture TEE
- Expected
- 180 mm Hg
- Tools
- External pressure gauge, test fixture TEE assembly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP Monitor patient safety
Cuff inflation pressure verification procedure accessible via Service/Test Hardware screen
Ranked causes
- 1. NIBP monitor cuff inflation pressure requires verification (common)
Access NIBP Monitor calibration verification screen
- Test point
- Service/Test Hardware screen
- Expected
- NIBP MONITOR option accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen sensor zero patient safety
Oxygen sensor requires zero calibration adjustment to establish baseline cell current values before clinical use.
Ranked causes
- 1. Oxygen sensor zero calibration required (common)
Remove oxygen sensor assembly from inspiratory valve dome
- Test point
- Inspiratory valve dome
- Expected
- Sensor assembly removed
8 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure Monitor patient safety
Zero and span calibration procedures for pressure monitor accessible via Service/Test Hardware screen
Ranked causes
- 1. Pressure monitor requires zero or span calibration (common)
Access Pressure Monitor calibration screen
- Test point
- Service/Test Hardware screen
- Expected
- PRESSURE MONITOR option accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote display assembly mounting torque
Remote display assembly or Datagriп mounting needs adjustment for proper tension
Ranked causes
- 1. Remote display assembly mounting screw under-torqued (common)
Remove plastic cap at end of display arm
- Test point
- Display arm end cap
- Expected
- Cap removed to expose mounting screw
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve washer under flow sensor degraded caution
Washer does not seal properly under expiratory valve; may cause gas leakage or improper flow sensor function.
Ranked causes
- 1. Washer under expiratory valve worn or deteriorated (occasional)
Inspect washer under expiratory valve
- Test point
- Washer under expiratory valve
- Expected
- Good condition, not worn or cracked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect display screen shown
Wrong monitor screen displayed; unit may not show expected interface or data
Ranked causes
- 1. Backplane assembly defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose interconnecting cables caution
Ribbon cables or wire harnesses between suspected assemblies not properly seated
Ranked causes
- 1. Improper cable connection during assembly or service (occasional)
Visual inspection of cable connections
- Test point
- Between processor boards and peripheral devices
- Expected
- All cables fully seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MAN/AUTO selector valve sticking or not responding caution
Selector valve does not switch smoothly between manual and automatic modes; may require excessive force or fail to engage.
Ranked causes
- 1. Excessive lubricant residue in valve channel or body (common)
Inspect valve channel and body for residual lubricant
- Test point
- Valve channel and valve body interior
- Expected
- Clean, free of lubricant residue
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing backplane voltages high voltage
Voltage measurements missing at backplane connectors J28, J29, or J30 during startup or operation
Ranked causes
- 1. Power supply failure or malfunction (unknown)
Verify all voltages present at backplane
- Test point
- J28, J29, J30 or J15
- Expected
- All voltages per Table 3-1
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multispec analyzer not in full accuracy mode
CO2 and Agent warm-up messages appear on advisory display, preventing valid accuracy testing
Ranked causes
- 1. Multispec warm-up cycle in progress (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O-rings on oxygen sensor housing or plug degraded patient safety
O-rings do not seal; oxygen sensor assembly leaks or does not mount properly on inspiratory valve dome.
Ranked causes
- 1. O-rings on oxygen sensor housing or plug cracked, dried, or deteriorated (occasional)
Inspect O-rings on oxygen sensor housing and plug
- Test point
- O-rings on O2 sensor housing and oxygen sensor plug
- Expected
- O-rings intact, pliable, no cracks or drying
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 supply alarm activates outside setpoint range patient safety
Alarm triggers below 34 psi or above 40 psi instead of within the correct 40–34 psi band
Ranked causes
- 1. Oxygen supply pressure alarm switch setpoint out of adjustment (common)
Verify test gauge is connected to TEE fitting in O2 regulator output line
- Test point
- TEE fitting, 1/4 in. diameter O2 regulator output line
- Expected
- Test gauge properly connected and reading pressure
- Tools
- O2 test gauge
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor cell current values
Verify recorded oxygen sensor fuel cell A and cell B zero calibration current values for documentation.
Ranked causes
- 1. Zero calibration values recorded (common)
Verify CURRENT CELL A and CELL B values displayed before STORE ZERO
- Test point
- Display screen current cell readout
- Expected
- CURRENT CELL A: 3, CURRENT CELL B: 6 (or equivalent current values for the sensor)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor replacement patient safety
Sensor located on inspiratory valve requires replacement or maintenance
Ranked causes
- 1. Oxygen sensor capsule degradation or failure (unknown)
Set System Power switch to STANDBY
- Test point
- System Power switch
- Expected
- STANDBY position
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen supply alarm activates above setpoint range patient safety
Alarm activates when oxygen pressure is above 40 psi instead of between 40 and 34 psi.
Ranked causes
- 1. Alarm switch setpoint too high (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen supply alarm activates below setpoint range patient safety
Alarm activates when oxygen pressure falls below 34 psi instead of between 40 and 34 psi.
Ranked causes
- 1. Alarm switch setpoint too low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply self-test failure high voltage
Red LED DS4 remains lighted after power-up indicating self-test failure of monitoring and communication circuitry
Ranked causes
- 1. Power supply board monitoring or communication circuit fault (common)
Observe LED DS4 status during power-up self-test
- Test point
- Power supply board LED DS4 (red)
- Expected
- LED DS4 lighted during power up, off after self-test completion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltages out of tolerance caution
Measured power supply voltages do not meet specifications; power distribution compromised
Ranked causes
- 1. Power supply circuit board defective (common)
Complete preventive maintenance
- Test point
- Unit functional status
- Expected
- Unit passes PMS
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-up diagnostic test failure or setup error
One or more power-up diagnostic tests fail or setup errors occur during system initialization.
Ranked causes
- 1. CPU assembly defective or misconfigured (common)
Do any power-up diagnostic tests fail or setup errors occur
- Test point
- System power-up initialization
- Expected
- All diagnostic tests pass, no setup errors
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace SPIROMED respiratory volume sensor caution
Sensor replacement procedure for the respiratory volume sensor installed between absorber assembly and expiratory valve
Ranked causes
- 1. Sensor requires replacement (unknown)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spiromed sensor lubrication required caution
Sensor operation stiff, sluggish, or making noise; lubrication depletion or contamination.
Ranked causes
- 1. Spiromed sensor insufficient or depleted lubricant (common)
Locate lubrication access holes on spiromed sensor
- Test point
- Four lateral holes at sides of spiromed sensor
- Expected
- Holes visible and accessible, marked by arrows
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spiromed sensor washer degraded caution
Washer under sensor does not seal properly; may cause gas leakage or sensor function issues.
Ranked causes
- 1. Washer under spiromed sensor worn or deteriorated (occasional)
Inspect washer under spiromed sensor
- Test point
- Washer under spiromed sensor
- Expected
- Good condition, not worn or cracked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 monitor battery depleted
Novametrix simulator not providing consistent readings or readings unstable
Ranked causes
- 1. Novametrix simulator 9V battery voltage below 8.5 VDC (occasional)
Measure Novametrix simulator battery voltage
- Test point
- Battery terminals
- Expected
- 9.0 VDC minimum for reliable operation, >8.5 VDC for operation, <8.5 VDC requires replacement
- Tools
- Digital multimeter (DMM)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrasonic flow sensor connector hose or O-ring degraded caution
Connector hose leaks, connector does not seal, or O-rings deteriorated; flow sensor does not function or provides inaccurate readings.
Ranked causes
- 1. Connector hose, connector, or O-ring components degraded or damaged (occasional)
Inspect connector hose, connector, and O-ring condition
- Test point
- Ultrasonic flow sensor connector hose and assembly
- Expected
- Good condition, no cracks or degradation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrasonic flow sensor transducers or O-rings degraded caution
Flow sensor provides inaccurate readings or does not function; may indicate volume/flow delivery problems.
Ranked causes
- 1. Transducer O-rings or components degraded or damaged (occasional)
Inspect transducer O-rings and components
- Test point
- Ultrasonic flow sensor transducers
- Expected
- Good condition, no cracks or degradation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve dome washers worn or degraded caution
Washers do not provide proper seal between valve dome and valve body; may cause gas leakage or improper valve function.
Ranked causes
- 1. Valve dome washers worn, cracked, or deteriorated (occasional)
Inspect valve dome washers condition
- Test point
- Valve dome washers
- Expected
- Good condition, not worn or cracked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0000 | Power on | System power initialization | info |
A001 | Pattern Adjusted | Pattern adjustment performed | info |
B Drive Error Messages | Backup drive (B:) reports various errors | Detected during disk access or power-up | warning |
C001 | O2 Store Zero A | O2 zero calibration performed | info |
C002 | O2 Store Zero B | O2 zero calibration B performed | info |
C003 | Line Block Store Reference | Line blockage reference stored | info |
C004 | Gas Analyzer - Start | Gas analyzer calibration started | info |
C005 | Gas Analyzer - Continue | Gas analyzer calibration continued | info |
C006 | NIBP Inflate | NIBP inflation calibration | info |
C007 | NIBP Stop | NIBP calibration stopped | info |
C008 | NIBP Take Reading | NIBP reading taken | info |
C009 | Respiratory Flow Store Span | Respiratory flow span calibration | info |
C010 | Respiratory Flow Store Zero | Respiratory flow zero calibration | info |
C011 | Pressure Store Span | Pressure span calibration | info |
C012 | Pressure Store Zero | Pressure zero calibration | info |
C013 | NIBP Inflation Check | NIBP inflation check performed | info |
C014 | NIBP Deflation Check | NIBP deflation check performed | info |
C015 | NIBP Leak Check | NIBP leak check performed | info |
CMOS Checksum Fail | CMOS memory checksum error detected | Detected during power-up self test | error |
Communications Error with Host | O.R. DATA MANAGER cannot communicate with main NM4 host | Detected during power-up or operation | error |
DG00 | Datagrip Internal Error | Datagrip module internal error | error |
DG01 | Datagrip Internal Error | Datagrip module internal error | error |
DG02 | Datagrip Internal Error | Datagrip module internal error | error |
Disk Error | Hard disk or floppy disk error detected | Detected during power-up or disk access | error |
Disk Full Error | Disk storage capacity exhausted | Attempted to write data when disk full | warning |
Disk Invalid Error | Disk file system invalid or corrupted | Detected during disk access or power-up | error |
Disk Not Inserted Error | Floppy disk not present in drive | Attempted to access floppy disk without disk inserted | info |
F011 | QPDM Address Register Error | QPDM address register test failure | error |
F012 | QPDM Data Register Error | QPDM data register test failure | error |
F013 | QPDM Vblank Int Timeout Error | QPDM vertical blank interrupt timeout | error |
F014 | QPDM VRAM Memory Error | QPDM video memory test failure | error |
F021 | PROM CRC Test Error | PROM checksum validation failure | error |
F031 | SRAM Memory Test Error | SRAM memory test failure | error |
F051 | MFP Bus Error | MFP bus communication failure | error |
F052 | MFP TimerA Error | MFP Timer A test failure | error |
F053 | MFP TimerB Error | MFP Timer B test failure | error |
F054 | MFP TimerC Error | MFP Timer C test failure | error |
F055 | MFP TimerD Error | MFP Timer D test failure | error |
F056 | System Clock Error | System clock test failure | error |
F057 | FIFO Empty Interrupt Error | FIFO empty interrupt test failure | error |
F058 | ADC Interrupt Error | ADC interrupt test failure | error |
F059 | CNTOVF Interrupt Error | Counter overflow interrupt test failure | error |
F061 | ADC Interrupt Error | ADC interrupt test failure | error |
F062 | 12V to -12V Out of Tolerance Error | Power supply voltage out of range | error |
F063 | 5V to GND Out of Tolerance Error | 5V power supply out of range | error |
F071 | Primary Speaker Static Test Error | Speaker static test failure | error |
F072 | Primary Speaker Dynamic Test Error | Speaker dynamic test failure | error |
F073 | Primary Speaker FIFO Test Error | Speaker FIFO test failure | error |
F074 | Primary Speaker No Volume Test Error | Speaker volume test failure | error |
F081 | Backup Speaker Dynamic Test Error | Backup speaker dynamic test failure | error |
F082 | Backup Speaker No Volume Test Error | Backup speaker volume test failure | error |
F091 | SAO2 Speaker Static Test Error | SAO2 speaker static test failure | error |
F092 | SAO2 Speaker Dynamic Test Error | SAO2 speaker dynamic test failure | error |
F093 | SAO2 Speaker No Volume Test Error | SAO2 speaker volume test failure | error |
F111 | Real Time Clock Test Error | Real time clock test failure | error |
F121 | Bad File Warning | File corruption detected | warning |
F122 | Bad Files Error | File corruption detected | error |
F123 | Event Memory Error | Event memory test failure | error |
F130 | Alternate Processor Fail | Alternate processor failure | error |
F200 | UART1 Internal Init Fail | UART1 initialization failure | error |
F201 | UART1 Register Readback Fail | UART1 register read test failure | error |
F202 | UART1 Transmit Interrupt Fail | UART1 transmit interrupt failure | error |
F203 | UART1 Receive Interrupt Fail | UART1 receive interrupt failure | error |
F204 | UART1 Receive Error Interrupt Fail | UART1 receive error interrupt failure | error |
F205 | UART1 Baud Rate Test 1 Fail | UART1 baud rate test 1 failure | error |
F206 | UART1 Baud Rate Test 2 Fail | UART1 baud rate test 2 failure | error |
F207 | UART1 Baud Rate Test 3 Fail | UART1 baud rate test 3 failure | error |
F210 | UART1 Channel 0 Fail | UART1 channel 0 test failure | error |
F211 | UART1 Channel 1 Fail | UART1 channel 1 test failure | error |
F212 | UART1 Channel 2 Fail | UART1 channel 2 test failure | error |
F213 | UART1 Channel 3 Fail | UART1 channel 3 test failure | error |
F214 | UART1 Channel 4 Fail | UART1 channel 4 test failure | error |
F215 | UART1 Channel 5 Fail | UART1 channel 5 test failure | error |
F216 | UART1 Channel 6 Fail | UART1 channel 6 test failure | error |
F217 | UART1 Channel 7 Fail | UART1 channel 7 test failure | error |
F300 | UART2 Internal Init Fail | UART2 initialization failure | error |
F301 | UART2 Register Readback Fail | UART2 register read test failure | error |
F302 | UART2 Transmit Interrupt Fail | UART2 transmit interrupt failure | error |
F303 | UART2 Receive Interrupt Fail | UART2 receive interrupt failure | error |
F304 | UART2 Receive Error Interrupt Fail | UART2 receive error interrupt failure | error |
F305 | UART2 Baud Rate Test 1 Fail | UART2 baud rate test 1 failure | error |
F306 | UART2 Baud Rate Test 2 Fail | UART2 baud rate test 2 failure | error |
F307 | UART2 Baud Rate Test 3 Fail | UART2 baud rate test 3 failure | error |
F310 | UART2 Channel 0 Fail | UART2 channel 0 test failure | error |
F311 | UART2 Channel 1 Fail | UART2 channel 1 test failure | error |
F312 | UART2 Channel 2 Fail | UART2 channel 2 test failure | error |
F313 | UART2 Channel 3 Fail | UART2 channel 3 test failure | error |
F314 | UART2 Channel 4 Fail | UART2 channel 4 test failure | error |
F315 | UART2 Channel 5 Fail | UART2 channel 5 test failure | error |
F316 | UART2 Channel 6 Fail | UART2 channel 6 test failure | error |
F317 | UART2 Channel 7 Fail | UART2 channel 7 test failure | error |
H001 | Head Temperature Error | Head assembly temperature out of range | error |
Internal Diagnostic Failure upon System Power-Up | O.R. DATA MANAGER self-test detected internal failure | Executed during power-up initialization | critical |
Interrupt Controller #1 Fail | First interrupt controller hardware failed | Detected during power-up self test | critical |
Interrupt Controller #2 Fail | Second interrupt controller hardware failed | Detected during power-up self test | critical |
Keyboard Error or Keyboard Not Present | O.R. DATA MANAGER keyboard not detected or malfunctioning | Detected during power-up self test | error |
M001 | Channel Error Active CPU | Active CPU channel error | error |
M002 | Channel Error Inactive CPU | Inactive CPU channel error | error |
M003 | Command Retry | Command retry required | warning |
M004 | Communication Error | MSPEC communication failure | error |
M005 | System Fault Max | Maximum system fault threshold reached | error |
M006 | Version ID Retry | Version identification retry | warning |
M007 | Span Fail | Span calibration failure | error |
M008 | Solenoid Stuck | Solenoid valve stuck | error |
M009 | Zero Timeout | Zero calibration timeout | error |
M010 | Zero Fail Max | Maximum zero calibration failures | error |
M011 | RX Buffer Overflow | Receive buffer overflow condition | error |
M012 | No Continuous Data | Continuous data stream failure | error |
M013 | Line Block Differential | Gas line blockage detected | warning |
M014 | Line Block Less Than Minimum | Gas line blockage below minimum threshold | warning |
M016 | Multispec Shutdown | MSPEC module shutdown | error |
M101 | CO2 Channel Error | CO2 measurement channel error | error |
M102 | Channel Error O2 Info P1 | O2 measurement channel error packet 1 | error |
M103 | Channel Error O2 Info P2 | O2 measurement channel error packet 2 | error |
Memory Battery Fail | Battery for memory backup has failed or lost charge | Detected during power-up self test | warning |
Memory in Protected Mode Fail | Protected mode memory access failed | Detected during power-up self test | critical |
N001 | NIBP Monitor Error | NIBP module monitoring failure | error |
N002 | NIBP Communication Error | NIBP module communication failure | error |
N003 | NIBP Bad Data | NIBP data validation failure | error |
N004 | Inflation Pressure Not Greater Than 30 | NIBP inflation pressure low | warning |
N005 | Error Received from NIBP | NIBP module error indication | error |
N006 | NIBP Mode Conflict | NIBP operating mode conflict | error |
NO OXI PULSE | Novametrix SpO2 sensor signal lost or disconnected | Sensor disconnected from finger or cable unplugged for 10 seconds | warning |
NO SpO2 PULSE | Nellcor SpO2 sensor signal lost or disconnected | Sensor disconnected from finger or cable unplugged for 10 seconds | warning |
OXI ALARMS DISABLE | Oxygen saturation alarms disabled | Depressing both Open Test keys together during module test | info |
OXI SEN DISC | Oxygen sensor disconnected or not detected | Depressing both Open Test keys or sensor physically disconnected | warning |
P001 | AC Fail | AC mains power lost | error |
P002 | AC Good | AC mains power restored | info |
P003 | Battery Low | Battery voltage below threshold | warning |
P004 | Battery Good | Battery voltage normal | info |
P005 | Brick Bad | Power supply module failure | error |
Printer Inoperative | O.R. DATA MANAGER printer not responding | Attempted to print or detected during self-test | warning |
Processor Exception Interrupts Fail | Processor exception interrupt handling failed | Detected during power-up self test | critical |
Protected Mode Fail | Protected mode operation failed | Detected during power-up self test | critical |
PULSE LO | Novametrix pulse signal weak or absent | Depressing Open Test RED key during simulator test or loss of sensor signal | warning |
S001 | Installation | Installation event | info |
S002 | PMS | Preventive maintenance performed | info |
S003 | Service | Service performed | info |
S004 | Secondary Service Screen Entry | Secondary service screen accessed | info |
S005 | Reset Service Date | Service date reset | info |
S006 | Service Log Clear | Service log cleared | info |
Sizing Extended Memory 00384K Not Found | System did not detect expected extended memory | Detected during power-up self test | error |
Sizing System Memory 640K Not Found | System did not detect expected 640K base memory | Detected during power-up self test | critical |
System Memory 640K Fail | Base memory 640K failed self-test | Detected during power-up self test | critical |
T001 | Bad Device Recalibration | Touch panel calibration error | error |
TEM0 | Template Defaults Loaded | Template defaults loaded | info |
TEM1 | Recreating Template Database | Template database recreation | info |
TEM2 | Templates Reinitialized | Template reinitialization | info |
TEM3 | Templates Reinitialized | Template reinitialization | info |
TEM4 | Factory Defaults Installed | Factory defaults applied | info |
V001 | VTLK Communication Error | Data Manager communication error | error |
V002 | VTLK Communication Lost | Data Manager link lost | error |
V003 | VTLK Communication Regained | Data Manager link restored | info |
V010 | ORDM Startup Failure | OR Data Manager startup failed | error |
V011 | ORDM Inactive Startup Failure | OR Data Manager inactive startup failed | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.