Dräger Narkomed 2C: troubleshooting and error codes
The ServiceForge knowledge base holds 32 documented symptoms for the Dräger Narkomed 2C, with 58 ranked causes, 85 diagnostic checks with test points and expected values, and 50 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Ventilator Inoperative patient safety
Anesthesia ventilator not functioning, no breathing assistance delivered to patient
Ranked causes
- 1. Incorrect breathing hose or scavenger hose connections to bellows assembly (common)
- 2. Cylinder or pipeline oxygen supply pressure inadequate for ventilator operation (common)
- 3. Ventilator settings incorrect or out of adjustment (common)
- 4. 8V power not present at ventilator controller J2 or power supply and connections defective (occasional)
- 5. Valve and guide assembly or ventilator controller defective or requires replacement (occasional)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Blank Upon System Power-up
Display screen remains blank or shows no information when system is powered on
Ranked causes
- 1. Display cable disconnected from processor board port J4 or display assembly disconnected (common)
- 2. Power supply voltages out of tolerance or defective (occasional)
- 3. Display assembly defective or requires replacement (occasional)
- 4. Processor assembly defective or requires replacement (rare)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Respiratory Volume Monitor patient safety
Respiratory volume monitor reading absent or non-functional on anesthesia machine
Ranked causes
- 1. Volume sensor not installed correctly or sensor cord disconnected from interface panel (common)
- 2. Internal cable from interface panel to J17 on processor board disconnected or failed (occasional)
- 3. Volume sensor defective or requires replacement (occasional)
- 4. Processor assembly defective or requires replacement (rare)
Perform complete PMS on unit
- Test point
- PMS self-test
- Expected
- Unit passes PMS
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Datagrip Response
Datagrip hand control assembly is unresponsive, no input received from Datagrip controls
Ranked causes
- 1. Datagrip cable disconnected from processor port J3 or Datagrip assembly disconnected (common)
- 2. Power supply voltages out of tolerance or defective (occasional)
- 3. Datagrip assembly defective or requires replacement (occasional)
- 4. Processor assembly defective or requires replacement (rare)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No O2 Supply Pressure Alarms patient safety
Oxygen supply pressure alarm system non-functional, no alarm indication when O2 pressure is low or out of range
Ranked causes
- 1. O2 low pressure alarm switch, wire harness, or connections to alarm channel damaged or disconnected (common)
- 2. Cable or connections between alarm channel J1 and processor board J16 disconnected or damaged (occasional)
- 3. Alarm channel defective or requires replacement (occasional)
- 4. Processor assembly defective or requires replacement (rare)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Keypanel Inoperative
Display keypanel buttons or input controls are unresponsive to operator input
Ranked causes
- 1. Display assembly disconnected or requires reconnection (common)
- 2. Display assembly defective or requires replacement (occasional)
- 3. Processor assembly defective or requires replacement (rare)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of O2 Monitor patient safety
O2 monitor display or reading is absent or non-functional, anesthesia machine unable to monitor oxygen concentration
Ranked causes
- 1. Sensor element not installed correctly or sensor cord disconnected from interface panel (common)
- 2. Internal cable from interface panel to J1 on processor board disconnected or failed (occasional)
- 3. Processor assembly defective or requires replacement (occasional)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Audio Alarms patient safety
Audio alarm system is non-functional, no sound output from speaker when alarms are triggered
Ranked causes
- 1. Speaker defective or requires replacement (common)
- 2. Speaker wire harness or connection to processor board J2 disconnected or damaged (occasional)
- 3. Processor assembly defective or requires replacement (rare)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial Port Communication Failure
Serial port unable to communicate with peripheral devices or host computer, data transfer not possible
Ranked causes
- 1. Peripheral cable disconnected from serial interface port or cable defective (common)
- 2. Peripheral or host serial port configuration settings incorrect (occasional)
- 3. Processor assembly defective or requires replacement (rare)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Breathing Pressure Monitor patient safety
Breathing pressure monitor reading absent or non-functional on anesthesia machine display
Ranked causes
- 1. Pilot line disconnected at one or both ends or internal tubing to transducer disconnected (common)
- 2. Processor assembly defective or requires replacement (occasional)
Perform complete PMC on unit
- Test point
- PMC self-test
- Expected
- Unit passes PMC
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not pass power-up diagnostics
Machine powers up and runs diagnostics but displays CONDITIONALLY FUNCTIONAL or NON-FUNCTIONAL message instead of FUNCTIONAL; machine may not proceed to MACHINE MONITOR screen.
Ranked causes
- 1. Minor problem detected in diagnostics (unknown)
- 2. Serious problem detected in diagnostics (unknown)
Review diagnostics display for FUNCTIONAL status
- Test point
- Diagnostics screen
- Expected
- All diagnostic tests show PASS; final message shows FUNCTIONAL
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Absorber maintenance inspection caution
Preventive maintenance checklist for absorber components including valve domes, pins, discs, washers, O-rings, flow sensors, and soda lime canister condition verification
Ranked causes
- 1. Worn or damaged O-rings on sensor or plug assemblies (occasional)
Inspect O-rings on sensor or plug assembly
- Test point
- Oxygen sensor or plug assembly
- Expected
- O-rings in good condition
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery voltage check
Verification and calibration of battery voltage output to ensure full charge and low battery cutoff parameters are within specification
Ranked causes
- 1. Battery voltage out of specification during AC line operation or battery operation mode (unknown)
Measure battery voltage during AC line operation at full charge
- Test point
- TB1 terminals 1 and 2 on power supply
- Expected
- 12.83 to 14.18 VDC, 12.83-14.18
- Tools
- DMM voltmeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display arm hinge tension
Arm does not return to approximately horizontal position when raised or lowered; or arm stays in any position when raised/lowered (over-tightened).
Ranked causes
- 1. Friction adjustment screw incorrectly set (common)
Raise and release display arm
- Test point
- display arm hinge
- Expected
- arm returns to approximately horizontal position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Machine mount joint tension
Excessive free swing at the machine mount joint where display arm connects.
Ranked causes
- 1. Shoulder screw not properly tightened (common)
Tighten shoulder screw at machine mount joint
- Test point
- machine mount joint shoulder screw
- Expected
- torque of 75 in. lbs.
- Tools
- torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Monitor zero calibration patient safety
Perform zero calibration for O2 cells A and B to store baseline current readings for the oxygen monitor sensor
Ranked causes
- 1. Oxygen Monitor zero calibration procedure (common)
Verify current cell A reading before zero calibration
- Test point
- CURRENT CELL A
- Expected
- Current value displayed on screen
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote display assembly mounting screw torque
Remote display assembly or Datagip mounting requires torque adjustment at the end of display arm.
Ranked causes
- 1. Remote display assembly mounting screw not properly tightened (common)
Tighten remote display assembly mounting screw
- Test point
- remote display assembly mounting screw at end of display arm
- Expected
- torque of 3 to 5 foot pounds, 3 to 5 foot pounds
- Tools
- torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canister and soda lime inspection caution
Inspection of canisters, canister gaskets, dust cup, O-ring, and soda lime condition during preventive maintenance
Ranked causes
- 1. Contaminated or degraded soda lime and associated components (common)
Inspect canisters, gaskets, dust cup, O-ring, and soda lime
- Test point
- CO2 absorber components
- Expected
- All canisters, gaskets, dust cup, O-ring, and soda lime in good condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cm H2O gauge zero adjustment caution
Verification and adjustment of pressure gauge to zero calibration using slotted screw adjustment
Ranked causes
- 1. Cm H2O gauge not calibrated to zero (occasional)
Verify gauge reads zero
- Test point
- cm H2O gauge
- Expected
- Gauge reads 0 cm H2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Freshgas Outlet assembly inspection caution
Removal and inspection of 15-mm connector and Freshgas Outlet assembly condition
Ranked causes
- 1. Freshgas Outlet assembly wear or damage (occasional)
Inspect Freshgas Outlet assembly condition
- Test point
- Freshgas Outlet assembly
- Expected
- Assembly in good condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory valve assembly inspection caution
Removal and inspection of inspiratory valve assembly, washer condition, and spring clip verification
Ranked causes
- 1. Inspiratory valve washer wear or damage (occasional)
Inspect washer under inspiratory valve
- Test point
- Inspiratory valve
- Expected
- Washer in good condition
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or incorrect -12VDC processor voltage high voltage
Measured -12VDC at processor board connector J15 pin 13 outside acceptable range
Ranked causes
- 1. Power supply voltage regulation fault or processor board fault (common)
Measure -12VDC at processor board
- Test point
- J15-13 (Orn)
- Expected
- -11.64 to -12.36 VDC, -11.64 to -12.36 VDC
- Tools
- DMM
Low or incorrect +12VDC processor voltage high voltage
Measured +12VDC at processor board connector J15 pin 4 outside acceptable range
Ranked causes
- 1. Power supply voltage regulation fault or processor board fault (common)
Measure +12VDC at processor board
- Test point
- J15-4 (Yel)
- Expected
- 11.76 to 12.24 VDC, 11.76 to 12.24 VDC
- Tools
- DMM
Low or incorrect +5VDC processor voltage high voltage
Measured +5VDC at processor board connector J15 pin 14 outside acceptable range
Ranked causes
- 1. Power supply voltage regulation fault or processor board fault (common)
Measure +5VDC at processor board
- Test point
- J15-14 (Pur)
- Expected
- 4.85 to 5.15 VDC, 4.85 to 5.15 VDC
- Tools
- DMM
Low or incorrect +8VDC alarm channel voltage high voltage
Measured +8VDC at alarm channel connector J3 pin 9 outside acceptable range
Ranked causes
- 1. Power supply voltage regulation fault or alarm channel board fault (common)
Measure +8VDC at alarm channel
- Test point
- J3-9 (Wht)
- Expected
- 7.76 to 8.24 VDC, 7.76 to 8.24 VDC
- Tools
- DMM
MAN/AUTO Selector Valve repacking caution
Removal, cleaning, and lubrication of Manual/Automatic selector valve channel and body with high vacuum grease if applicable
Ranked causes
- 1. Selector valve requires periodic repacking and lubrication for smooth operation (occasional)
Remove stick shift block from selector valve body
- Test point
- Selector valve assembly
- Expected
- Block removed with four screws
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spiromed Sensor lubrication caution
Lubrication of Spiromed sensor lateral holes using approved sensor lubrication kit and pipette application method
Ranked causes
- 1. Spiromed sensor requires periodic lubrication for optimal operation (common)
Inspect washer under expiratory valve
- Test point
- Expiratory valve
- Expected
- Washer in good condition
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrasonic Flow Sensor maintenance caution
Inspection and reassembly of ultrasonic flow sensor connector hose, transducers, and associated components if applicable
Ranked causes
- 1. Damaged or degraded ultrasonic sensor components or O-rings (occasional)
Inspect connector hose, connector, and O-ring
- Test point
- Ultrasonic sensor connector hose
- Expected
- Connector hose, connector, and O-ring in good condition
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage measurement: -12VDC on processor board high voltage
Measurement of -12V power supply voltage at connector J15 pin 13 on NARKOMED 2C processor board
Ranked causes
- 1. Power supply failure if voltage out of range (unknown)
Measure -12VDC at J15-13 using DMM
- Test point
- J15 connector pin 13 (Orange wire) on processor board
- Expected
- -11.64 to -12.36 VDC, +/-0.36 VDC
- Tools
- Digital multimeter (DMM)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage measurement: +12VDC on processor board high voltage
Measurement of +12V power supply voltage at connector J15 pin 4 on NARKOMED 2C processor board
Ranked causes
- 1. Power supply failure if voltage out of range (unknown)
Measure +12VDC at J15-4 using DMM
- Test point
- J15 connector pin 4 (Yellow wire) on processor board
- Expected
- 11.76 to 12.24 VDC, +/-0.24 VDC
- Tools
- Digital multimeter (DMM)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage measurement: +5VDC on processor board high voltage
Measurement of +5V power supply voltage at connector J15 pin 14 on NARKOMED 2C processor board
Ranked causes
- 1. Power supply failure if voltage out of range (unknown)
Measure +5VDC at J15-14 using DMM
- Test point
- J15 connector pin 14 (Purple wire) on processor board
- Expected
- 4.85 to 5.15 VDC, +/-0.15 VDC
- Tools
- Digital multimeter (DMM)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage measurement: +8VDC on alarm channel high voltage
Measurement of +8V power supply voltage at connector J3 pin 9 on NARKOMED 2C alarm channel
Ranked causes
- 1. Power supply failure if voltage out of range (unknown)
Measure +8VDC at J3-9 using DMM
- Test point
- J3 connector pin 9 (White wire) on alarm channel
- Expected
- 7.76 to 8.24 VDC, +/-0.24 VDC
- Tools
- Digital multimeter (DMM)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
C001 | O2 zero calibration A stored | Oxygen sensor zero reference calibration point A recorded | info |
C002 | O2 zero calibration B stored | Oxygen sensor zero reference calibration point B recorded | info |
C011 | Pressure calibration span stored | Pressure sensor span calibration recorded | info |
C012 | Pressure calibration zero stored | Pressure sensor zero calibration recorded | info |
E000 | Task create error | Real-time OS fails to create a new task thread | critical |
E002 | Task ident error | Real-time OS fails to identify task | critical |
E003 | Task start error | Real-time OS cannot start a created task | critical |
E045 | Fatal error | Unrecoverable system exception | critical |
E049 | Device initialize error | Peripheral device fails to initialize | critical |
E050 | Device open error | Device driver cannot open peripheral device | critical |
E052 | Device read error | Data read from peripheral device fails | critical |
E053 | Device write error | Write operation to device fails | critical |
E100 | PDA subsystem error | Patient data acquisition subsystem fails | critical |
E200 | CCM subsystem error | Central control module subsystem fails | critical |
E300 | DSV subsystem error | Dual scavenging valve subsystem fails | critical |
E400 | UDI subsystem error | User display interface subsystem fails | critical |
E500 | EDC subsystem error | Esophageal detector circuit subsystem fails | critical |
E600 | SYS subsystem error | General system subsystem fails | critical |
E700 | IOD subsystem error | Input/output data interface fails | critical |
E800 | SUP subsystem error | Supply/power subsystem fails | critical |
F000 | SRAM test failed | Static RAM memory test fails during self-diagnosis | critical |
F010 | Timer test failed | Timer circuit fails self-diagnostic during startup | critical |
F020 | A/D converter test failed | Analog-to-digital converter self-test fails | critical |
F030 | Primary audio test failed | Speaker or audio output circuit fails self-test | unknown |
F040 | Backup audio test failed | Backup audio circuit fails self-diagnostic | unknown |
F050 | Serial I/O test failed | Serial communication port self-test fails | critical |
F060 | Real-time clock test failed | RTC circuit fails self-diagnostic during boot | critical |
F070 | NVRAM test failed | Non-volatile RAM memory test fails during startup | critical |
HP01 | MECIF communication error | Medical equipment common interface frame/parity error | critical |
HP02 | MECIF communication lost | MECIF connection broken | critical |
HP03 | MECIF communication regained | MECIF link restored | info |
O001 | ORLK communication error | Operating room link serial data error occurs | critical |
O002 | ORLK communication lost | Operating room link connection dropped | critical |
O003 | ORLK communication regained | Operating room link restored after loss | info |
S000 | View mode accessed | Service log view function activated | info |
S001 | Installation mode accessed | Initial system installation/setup procedure begun | info |
S002 | Preventive maintenance service accessed | Preventive maintenance screen entered | info |
S003 | Service mode accessed | General service screen entered | info |
S004 | Secondary service screen entry settings | Secondary service menu accessed | info |
S005 | Service date reset | Service date/time reset by technician | info |
S006 | Service log cleared | Service event log history erased | info |
S010 | Serial number set | Equipment serial number programmed | info |
S011 | Vaporizer settings configured | Anesthesia vaporizer parameters configured | info |
S012 | Vaporizer defaults restored | Vaporizer parameters reset to factory defaults | info |
S013 | Enable protocols configured | Safety protocol settings modified | info |
S014 | Low O2 supply whistle setting changed | Low oxygen alarm parameters configured | info |
S015 | Serial ports updated | Communication port configuration modified | info |
V001 | VTLK communication error | Vaporizer telemetry link frame, overrun, or parity error | critical |
V002 | VTLK communication lost | Vaporizer telemetry link connection lost | critical |
V003 | VTLK communication regained | Vaporizer telemetry link restored after loss | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI