Dräger Narkomed 2C: troubleshooting and error codes

The ServiceForge knowledge base holds 32 documented symptoms for the Dräger Narkomed 2C, with 58 ranked causes, 85 diagnostic checks with test points and expected values, and 50 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Ventilator Inoperative patient safety

Anesthesia ventilator not functioning, no breathing assistance delivered to patient

Ranked causes

  1. 1. Incorrect breathing hose or scavenger hose connections to bellows assembly (common)
  2. 2. Cylinder or pipeline oxygen supply pressure inadequate for ventilator operation (common)
  3. 3. Ventilator settings incorrect or out of adjustment (common)
  4. 4. 8V power not present at ventilator controller J2 or power supply and connections defective (occasional)
  5. 5. Valve and guide assembly or ventilator controller defective or requires replacement (occasional)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display Blank Upon System Power-up

Display screen remains blank or shows no information when system is powered on

Ranked causes

  1. 1. Display cable disconnected from processor board port J4 or display assembly disconnected (common)
  2. 2. Power supply voltages out of tolerance or defective (occasional)
  3. 3. Display assembly defective or requires replacement (occasional)
  4. 4. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of Respiratory Volume Monitor patient safety

Respiratory volume monitor reading absent or non-functional on anesthesia machine

Ranked causes

  1. 1. Volume sensor not installed correctly or sensor cord disconnected from interface panel (common)
  2. 2. Internal cable from interface panel to J17 on processor board disconnected or failed (occasional)
  3. 3. Volume sensor defective or requires replacement (occasional)
  4. 4. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMS on unit

Test point
PMS self-test
Expected
Unit passes PMS

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Datagrip Response

Datagrip hand control assembly is unresponsive, no input received from Datagrip controls

Ranked causes

  1. 1. Datagrip cable disconnected from processor port J3 or Datagrip assembly disconnected (common)
  2. 2. Power supply voltages out of tolerance or defective (occasional)
  3. 3. Datagrip assembly defective or requires replacement (occasional)
  4. 4. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No O2 Supply Pressure Alarms patient safety

Oxygen supply pressure alarm system non-functional, no alarm indication when O2 pressure is low or out of range

Ranked causes

  1. 1. O2 low pressure alarm switch, wire harness, or connections to alarm channel damaged or disconnected (common)
  2. 2. Cable or connections between alarm channel J1 and processor board J16 disconnected or damaged (occasional)
  3. 3. Alarm channel defective or requires replacement (occasional)
  4. 4. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display Keypanel Inoperative

Display keypanel buttons or input controls are unresponsive to operator input

Ranked causes

  1. 1. Display assembly disconnected or requires reconnection (common)
  2. 2. Display assembly defective or requires replacement (occasional)
  3. 3. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of O2 Monitor patient safety

O2 monitor display or reading is absent or non-functional, anesthesia machine unable to monitor oxygen concentration

Ranked causes

  1. 1. Sensor element not installed correctly or sensor cord disconnected from interface panel (common)
  2. 2. Internal cable from interface panel to J1 on processor board disconnected or failed (occasional)
  3. 3. Processor assembly defective or requires replacement (occasional)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Audio Alarms patient safety

Audio alarm system is non-functional, no sound output from speaker when alarms are triggered

Ranked causes

  1. 1. Speaker defective or requires replacement (common)
  2. 2. Speaker wire harness or connection to processor board J2 disconnected or damaged (occasional)
  3. 3. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial Port Communication Failure

Serial port unable to communicate with peripheral devices or host computer, data transfer not possible

Ranked causes

  1. 1. Peripheral cable disconnected from serial interface port or cable defective (common)
  2. 2. Peripheral or host serial port configuration settings incorrect (occasional)
  3. 3. Processor assembly defective or requires replacement (rare)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of Breathing Pressure Monitor patient safety

Breathing pressure monitor reading absent or non-functional on anesthesia machine display

Ranked causes

  1. 1. Pilot line disconnected at one or both ends or internal tubing to transducer disconnected (common)
  2. 2. Processor assembly defective or requires replacement (occasional)
First check

Perform complete PMC on unit

Test point
PMC self-test
Expected
Unit passes PMC

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not pass power-up diagnostics

Machine powers up and runs diagnostics but displays CONDITIONALLY FUNCTIONAL or NON-FUNCTIONAL message instead of FUNCTIONAL; machine may not proceed to MACHINE MONITOR screen.

Ranked causes

  1. 1. Minor problem detected in diagnostics (unknown)
  2. 2. Serious problem detected in diagnostics (unknown)
First check

Review diagnostics display for FUNCTIONAL status

Test point
Diagnostics screen
Expected
All diagnostic tests show PASS; final message shows FUNCTIONAL

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Absorber maintenance inspection caution

Preventive maintenance checklist for absorber components including valve domes, pins, discs, washers, O-rings, flow sensors, and soda lime canister condition verification

Ranked causes

  1. 1. Worn or damaged O-rings on sensor or plug assemblies (occasional)
First check

Inspect O-rings on sensor or plug assembly

Test point
Oxygen sensor or plug assembly
Expected
O-rings in good condition

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery voltage check

Verification and calibration of battery voltage output to ensure full charge and low battery cutoff parameters are within specification

Ranked causes

  1. 1. Battery voltage out of specification during AC line operation or battery operation mode (unknown)
First check

Measure battery voltage during AC line operation at full charge

Test point
TB1 terminals 1 and 2 on power supply
Expected
12.83 to 14.18 VDC, 12.83-14.18
Tools
DMM voltmeter

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display arm hinge tension

Arm does not return to approximately horizontal position when raised or lowered; or arm stays in any position when raised/lowered (over-tightened).

Ranked causes

  1. 1. Friction adjustment screw incorrectly set (common)
First check

Raise and release display arm

Test point
display arm hinge
Expected
arm returns to approximately horizontal position

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Machine mount joint tension

Excessive free swing at the machine mount joint where display arm connects.

Ranked causes

  1. 1. Shoulder screw not properly tightened (common)
First check

Tighten shoulder screw at machine mount joint

Test point
machine mount joint shoulder screw
Expected
torque of 75 in. lbs.
Tools
torque wrench

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Oxygen Monitor zero calibration patient safety

Perform zero calibration for O2 cells A and B to store baseline current readings for the oxygen monitor sensor

Ranked causes

  1. 1. Oxygen Monitor zero calibration procedure (common)
First check

Verify current cell A reading before zero calibration

Test point
CURRENT CELL A
Expected
Current value displayed on screen

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remote display assembly mounting screw torque

Remote display assembly or Datagip mounting requires torque adjustment at the end of display arm.

Ranked causes

  1. 1. Remote display assembly mounting screw not properly tightened (common)
First check

Tighten remote display assembly mounting screw

Test point
remote display assembly mounting screw at end of display arm
Expected
torque of 3 to 5 foot pounds, 3 to 5 foot pounds
Tools
torque wrench

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canister and soda lime inspection caution

Inspection of canisters, canister gaskets, dust cup, O-ring, and soda lime condition during preventive maintenance

Ranked causes

  1. 1. Contaminated or degraded soda lime and associated components (common)
First check

Inspect canisters, gaskets, dust cup, O-ring, and soda lime

Test point
CO2 absorber components
Expected
All canisters, gaskets, dust cup, O-ring, and soda lime in good condition

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cm H2O gauge zero adjustment caution

Verification and adjustment of pressure gauge to zero calibration using slotted screw adjustment

Ranked causes

  1. 1. Cm H2O gauge not calibrated to zero (occasional)
First check

Verify gauge reads zero

Test point
cm H2O gauge
Expected
Gauge reads 0 cm H2O

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Freshgas Outlet assembly inspection caution

Removal and inspection of 15-mm connector and Freshgas Outlet assembly condition

Ranked causes

  1. 1. Freshgas Outlet assembly wear or damage (occasional)
First check

Inspect Freshgas Outlet assembly condition

Test point
Freshgas Outlet assembly
Expected
Assembly in good condition

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory valve assembly inspection caution

Removal and inspection of inspiratory valve assembly, washer condition, and spring clip verification

Ranked causes

  1. 1. Inspiratory valve washer wear or damage (occasional)
First check

Inspect washer under inspiratory valve

Test point
Inspiratory valve
Expected
Washer in good condition

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low or incorrect -12VDC processor voltage high voltage

Measured -12VDC at processor board connector J15 pin 13 outside acceptable range

Ranked causes

  1. 1. Power supply voltage regulation fault or processor board fault (common)
First check

Measure -12VDC at processor board

Test point
J15-13 (Orn)
Expected
-11.64 to -12.36 VDC, -11.64 to -12.36 VDC
Tools
DMM

Low or incorrect +12VDC processor voltage high voltage

Measured +12VDC at processor board connector J15 pin 4 outside acceptable range

Ranked causes

  1. 1. Power supply voltage regulation fault or processor board fault (common)
First check

Measure +12VDC at processor board

Test point
J15-4 (Yel)
Expected
11.76 to 12.24 VDC, 11.76 to 12.24 VDC
Tools
DMM

Low or incorrect +5VDC processor voltage high voltage

Measured +5VDC at processor board connector J15 pin 14 outside acceptable range

Ranked causes

  1. 1. Power supply voltage regulation fault or processor board fault (common)
First check

Measure +5VDC at processor board

Test point
J15-14 (Pur)
Expected
4.85 to 5.15 VDC, 4.85 to 5.15 VDC
Tools
DMM

Low or incorrect +8VDC alarm channel voltage high voltage

Measured +8VDC at alarm channel connector J3 pin 9 outside acceptable range

Ranked causes

  1. 1. Power supply voltage regulation fault or alarm channel board fault (common)
First check

Measure +8VDC at alarm channel

Test point
J3-9 (Wht)
Expected
7.76 to 8.24 VDC, 7.76 to 8.24 VDC
Tools
DMM

MAN/AUTO Selector Valve repacking caution

Removal, cleaning, and lubrication of Manual/Automatic selector valve channel and body with high vacuum grease if applicable

Ranked causes

  1. 1. Selector valve requires periodic repacking and lubrication for smooth operation (occasional)
First check

Remove stick shift block from selector valve body

Test point
Selector valve assembly
Expected
Block removed with four screws

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Spiromed Sensor lubrication caution

Lubrication of Spiromed sensor lateral holes using approved sensor lubrication kit and pipette application method

Ranked causes

  1. 1. Spiromed sensor requires periodic lubrication for optimal operation (common)
First check

Inspect washer under expiratory valve

Test point
Expiratory valve
Expected
Washer in good condition

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ultrasonic Flow Sensor maintenance caution

Inspection and reassembly of ultrasonic flow sensor connector hose, transducers, and associated components if applicable

Ranked causes

  1. 1. Damaged or degraded ultrasonic sensor components or O-rings (occasional)
First check

Inspect connector hose, connector, and O-ring

Test point
Ultrasonic sensor connector hose
Expected
Connector hose, connector, and O-ring in good condition

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement: -12VDC on processor board high voltage

Measurement of -12V power supply voltage at connector J15 pin 13 on NARKOMED 2C processor board

Ranked causes

  1. 1. Power supply failure if voltage out of range (unknown)
First check

Measure -12VDC at J15-13 using DMM

Test point
J15 connector pin 13 (Orange wire) on processor board
Expected
-11.64 to -12.36 VDC, +/-0.36 VDC
Tools
Digital multimeter (DMM)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement: +12VDC on processor board high voltage

Measurement of +12V power supply voltage at connector J15 pin 4 on NARKOMED 2C processor board

Ranked causes

  1. 1. Power supply failure if voltage out of range (unknown)
First check

Measure +12VDC at J15-4 using DMM

Test point
J15 connector pin 4 (Yellow wire) on processor board
Expected
11.76 to 12.24 VDC, +/-0.24 VDC
Tools
Digital multimeter (DMM)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement: +5VDC on processor board high voltage

Measurement of +5V power supply voltage at connector J15 pin 14 on NARKOMED 2C processor board

Ranked causes

  1. 1. Power supply failure if voltage out of range (unknown)
First check

Measure +5VDC at J15-14 using DMM

Test point
J15 connector pin 14 (Purple wire) on processor board
Expected
4.85 to 5.15 VDC, +/-0.15 VDC
Tools
Digital multimeter (DMM)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement: +8VDC on alarm channel high voltage

Measurement of +8V power supply voltage at connector J3 pin 9 on NARKOMED 2C alarm channel

Ranked causes

  1. 1. Power supply failure if voltage out of range (unknown)
First check

Measure +8VDC at J3-9 using DMM

Test point
J3 connector pin 9 (White wire) on alarm channel
Expected
7.76 to 8.24 VDC, +/-0.24 VDC
Tools
Digital multimeter (DMM)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
C001O2 zero calibration A storedOxygen sensor zero reference calibration point A recordedinfo
C002O2 zero calibration B storedOxygen sensor zero reference calibration point B recordedinfo
C011Pressure calibration span storedPressure sensor span calibration recordedinfo
C012Pressure calibration zero storedPressure sensor zero calibration recordedinfo
E000Task create errorReal-time OS fails to create a new task threadcritical
E002Task ident errorReal-time OS fails to identify taskcritical
E003Task start errorReal-time OS cannot start a created taskcritical
E045Fatal errorUnrecoverable system exceptioncritical
E049Device initialize errorPeripheral device fails to initializecritical
E050Device open errorDevice driver cannot open peripheral devicecritical
E052Device read errorData read from peripheral device failscritical
E053Device write errorWrite operation to device failscritical
E100PDA subsystem errorPatient data acquisition subsystem failscritical
E200CCM subsystem errorCentral control module subsystem failscritical
E300DSV subsystem errorDual scavenging valve subsystem failscritical
E400UDI subsystem errorUser display interface subsystem failscritical
E500EDC subsystem errorEsophageal detector circuit subsystem failscritical
E600SYS subsystem errorGeneral system subsystem failscritical
E700IOD subsystem errorInput/output data interface failscritical
E800SUP subsystem errorSupply/power subsystem failscritical
F000SRAM test failedStatic RAM memory test fails during self-diagnosiscritical
F010Timer test failedTimer circuit fails self-diagnostic during startupcritical
F020A/D converter test failedAnalog-to-digital converter self-test failscritical
F030Primary audio test failedSpeaker or audio output circuit fails self-testunknown
F040Backup audio test failedBackup audio circuit fails self-diagnosticunknown
F050Serial I/O test failedSerial communication port self-test failscritical
F060Real-time clock test failedRTC circuit fails self-diagnostic during bootcritical
F070NVRAM test failedNon-volatile RAM memory test fails during startupcritical
HP01MECIF communication errorMedical equipment common interface frame/parity errorcritical
HP02MECIF communication lostMECIF connection brokencritical
HP03MECIF communication regainedMECIF link restoredinfo
O001ORLK communication errorOperating room link serial data error occurscritical
O002ORLK communication lostOperating room link connection droppedcritical
O003ORLK communication regainedOperating room link restored after lossinfo
S000View mode accessedService log view function activatedinfo
S001Installation mode accessedInitial system installation/setup procedure beguninfo
S002Preventive maintenance service accessedPreventive maintenance screen enteredinfo
S003Service mode accessedGeneral service screen enteredinfo
S004Secondary service screen entry settingsSecondary service menu accessedinfo
S005Service date resetService date/time reset by technicianinfo
S006Service log clearedService event log history erasedinfo
S010Serial number setEquipment serial number programmedinfo
S011Vaporizer settings configuredAnesthesia vaporizer parameters configuredinfo
S012Vaporizer defaults restoredVaporizer parameters reset to factory defaultsinfo
S013Enable protocols configuredSafety protocol settings modifiedinfo
S014Low O2 supply whistle setting changedLow oxygen alarm parameters configuredinfo
S015Serial ports updatedCommunication port configuration modifiedinfo
V001VTLK communication errorVaporizer telemetry link frame, overrun, or parity errorcritical
V002VTLK communication lostVaporizer telemetry link connection lostcritical
V003VTLK communication regainedVaporizer telemetry link restored after lossinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI