Dräger Narkomed 2B: troubleshooting and error codes
The ServiceForge knowledge base holds 47 documented symptoms for the Dräger Narkomed 2B, with 77 ranked causes, 193 diagnostic checks with test points and expected values, and 24 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Periodic Manufacturer's Certification (PMC) patient safety
Complete PMC inspection procedure for Narkomed 2B anesthesia machine including ventilator valve lubrication, relief valve diaphragm replacement, vaporizer inspection and concentration testing, and certification level determination.
Ranked causes
- 1. Ventilator MJV-2 square Clippard valve lubrication requirement (common)
- 2. Ventilator relief valve diaphragm replacement (common)
- 3. Halothane vaporizer soil condition inspection (occasional)
- 4. Dräger vapor vaporizer concentration verification (common)
- 5. Desflurane vaporizer concentration verification (common)
- 6. Certification level determination per DMI Recommendations (common)
Verify ventilator valve type and lubrication schedule
- Test point
- Ventilator assembly
- Expected
- MJV-2 Clippard valve with lubrication due date documented
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display screens blank upon system power-up
CRT display screens remain blank when system is powered on
Ranked causes
- 1. CRT assembly disconnected or not powered (unknown)
- 2. Power supply voltage failure (unknown)
- 3. Processor assembly failure (unknown)
- 4. CRT assembly failure (unknown)
Verify cables connected to J3 and J4 on processor board
- Test point
- J3 and J4 connectors
- Expected
- Cables are connected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMC certification labeling and documentation patient safety
Proper labeling and record keeping procedures to complete Periodic Manufacturer's Certification process including label selection based on certification level and customer notification.
Ranked causes
- 1. Certification level A (Certified) (unknown)
- 2. Certification level B (Certified with Recommendations) (unknown)
- 3. Certification level BC (Conditionally Certified with Recommendations) (unknown)
- 4. Certification level D (No Certification) (unknown)
Complete header and certification information on PMC form
- Test point
- PM Certification form (P/N S010211)
- Expected
- All header fields completed with machine details and service technician information
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of O2 monitor patient safety
Oxygen monitor display or function is absent or non-functional
Ranked causes
- 1. Oxygen analyzer not functional (unknown)
- 2. Processor assembly failure (unknown)
- 3. Cable disconnected from J10 on processor board (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of respiratory volume monitor patient safety
Respiratory volume monitor display or function is absent or non-functional
Ranked causes
- 1. Volume sensor not functional (unknown)
- 2. Processor assembly failure (unknown)
- 3. Cable disconnected from J9 on processor board (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No keypanel response
Display panel keys are not responding to user input
Ranked causes
- 1. Keypad cables not connected to J1 and J2 (unknown)
- 2. Keypad assembly failure (unknown)
- 3. Processor assembly failure (unknown)
Verify cables connected to J1 and J2 on processor board
- Test point
- J1 and J2 connectors
- Expected
- Cables are connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen ratio monitor alarms patient safety
O2/N2O ratio alarms are not functional or responding
Ranked causes
- 1. ORMC not functional (unknown)
- 2. Alarm channel assembly failure (unknown)
- 3. Processor assembly failure (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen supply pressure alarms patient safety
Oxygen supply pressure alarms are not functional or responding
Ranked causes
- 1. Oxygen supply alarm switch not functional (unknown)
- 2. Alarm channel assembly failure (unknown)
- 3. Processor assembly failure (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vaporizer handwheel does not lock when other vaporizer turned on
One or more vaporizer handwheels fail to lock in zero position when another vaporizer is activated, indicating interlock mechanism misalignment.
Ranked causes
- 1. Center set screw insufficiently tight (when center vaporizer turned on) (common)
- 2. Right set screw insufficiently adjusted (when left vaporizer turned on) (common)
- 3. Left set screw insufficiently adjusted (when right vaporizer turned on) (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vitalink failure
Host fails to communicate with peripheral device via Vitalink serial interface
Ranked causes
- 1. Peripheral cable not connected to serial port (unknown)
- 2. Incorrect serial port configuration settings (unknown)
- 3. Processor assembly failure (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration reading out of range
Displayed PRESSURE SPAN reading at 60 cm H2O is outside tolerance window of 465–519
Ranked causes
- 1. Pressure sensor calibration error (common)
- 2. Test fixture or gauge inaccuracy (occasional)
Measure circuit pressure on external gauge
- Test point
- External test gauge on TEE connector
- Expected
- 60 cm H2O
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of breathing pressure monitor patient safety
Breathing pressure monitor display or function is absent or non-functional
Ranked causes
- 1. Breathing pressure tubing disconnected or obstructed (unknown)
- 2. Processor assembly failure (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor housing not seating properly patient safety
Sensor housing fails to fully seat in inspiratory valve dome or copper rings not contacting electrical contacts
Ranked causes
- 1. Copper rings misaligned with electrical contacts (occasional)
- 2. Sensor housing mechanically misaligned or damaged (rare)
Verify copper ring alignment with electrical contacts
- Test point
- Sensor capsule copper rings and housing electrical contacts
- Expected
- Rings fully aligned and visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace ventilator bellows valve and guide assembly high voltage
Replacement of two-piece or one-piece tidal volume adjustment and valve case assembly in bellows box on left side of machine
Ranked causes
- 1. Valve case assembly requires replacement (unknown)
- 2. One-piece valve case and guide assembly requires replacement (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Roman Numeral conditions I and II requirements patient safety
Additional operational requirements that do not affect base certification level but mandate specific monitoring equipment and operator procedures.
Ranked causes
- 1. Roman Numeral I: CO2 monitor with apnea warning required (occasional)
- 2. Roman Numeral II: Oxygen analyzer with low oxygen warning required (occasional)
Verify CO2 monitor with apnea warning if Condition I applies
- Test point
- CO2 monitoring system
- Expected
- Functional CO2 monitor with apnea alarm capability present on system
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm lamp does not light at initial ORMC setting patient safety
Yellow O2/N2O FLOW RATIO lamp on alarm panel is not illuminated when O2/N2O flowmeter reaches 10 L/min at 25% O2 concentration.
Ranked causes
- 1. Switch adjustment cam and switch mount not properly positioned (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm lamp does not turn off at low N2O flow patient safety
Yellow O2/N2O FLOW RATIO lamp remains on when N2O flow drops to between 100 and 200 ml per minute, outside expected turn-off range.
Ranked causes
- 1. MPL switch setpoint misaligned (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of specification patient safety
Battery voltage measured at J7-3/J7-4 outside normal ranges; affects system backup operation
Ranked causes
- 1. Battery requires charging or replacement (common)
Measure battery voltage while operating from AC line at full charge
- Test point
- J7-3 (positive) to J7-4 (negative)
- Expected
- 12.83 to 14.18 VDC, 12.83-14.18 VDC
- Tools
- Voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing pressure hose disconnect failed
Breathing pressure hose cannot be disconnected from absorber during calibration setup
Ranked causes
- 1. Hose stuck or over-tightened connection (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air flowmeter (if applicable)
Verify air flow can be adjusted over full range and O2 concentration correct at 2 l/min with proper knob and label
Ranked causes
- 1. O2 concentration out of spec with air at 2 l/min (occasional)
Connect air pipeline hose if no cylinder yoke configured
- Test point
- Air inlet
- Expected
- Pipeline hose securely connected
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Auxiliary oxygen flowmeter (if applicable)
Verify auxiliary oxygen outlet maintains pressure stability at 50 cm H2O within 40–60 cm H2O tolerance after 10 seconds
Ranked causes
- 1. Auxiliary oxygen regulator or supply leak causing pressure drift (occasional)
Connect test pressure monitor to auxiliary O2 outlet
- Test point
- O2 outlet port
- Expected
- Monitor securely connected via PDM/Suction adapter (P/N 4115038)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Carbon dioxide flowmeter (if applicable)
Verify CO2 flow over 550 ml/min range and O2 concentration correct at CO2 500 ml/min + O2 1000 ml/min with proper knob and label
Ranked causes
- 1. O2 concentration out of spec with CO2 at 500 ml/min (occasional)
Set gas selector to ALL GAS
- Test point
- Gas selector switch
- Expected
- Switch in ALL GAS position
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 cylinder pressure regulator
Adjustment of CO2 regulator to ensure correct flowmeter reading at full flow
Ranked causes
- 1. CO2 regulator requires calibration adjustment (common)
Open CO2 cylinder valve
- Test point
- CO2 cylinder valve
- Expected
- Open
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cylinder pressure regulator adjustment high voltage
Adjust cylinder pressure regulator output to specified pressures based on cylinder supply pressure, with compensated settings for O2, Air, He, He/O2, N2 and direct setting for N2O
Ranked causes
- 1. Regulator output pressure requires adjustment to compensate for cylinder supply variations (common)
Verify minimum cylinder pressures before adjustment
- Test point
- Cylinder pressure gauge
- Expected
- N2O & CO2: 600 psi minimum; O2, Air, He, He/O2, N2: 1000 psi minimum
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Helium flowmeter (if applicable)
Verify helium flow can be adjusted over full range and O2 concentration is correct at 2 l/min with proper knob and label
Ranked causes
- 1. O2 concentration out of spec with helium at 2 l/min (occasional)
Set gas selector to ALL GAS
- Test point
- Gas selector switch
- Expected
- Switch in ALL GAS position
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nitrogen flowmeter (if applicable)
Verify nitrogen flow can be adjusted over full range and O2 concentration is correct at 2 l/min with proper knob and label
Ranked causes
- 1. O2 concentration out of spec with nitrogen at 2 l/min (occasional)
Set gas selector to ALL GAS
- Test point
- Gas selector switch
- Expected
- Switch in ALL GAS position
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nitrous oxide flowmeter
Verify nitrous oxide flow and O2 concentration response at 2 l/min with proper knob and label
Ranked causes
- 1. O2 concentration out of spec with N2O at 2 l/min (occasional)
Set nitrous oxide flow to 2 l/min
- Test point
- N2O flowmeter display
- Expected
- 2 l/min
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen flowmeter range adjustment
Verify oxygen flowmeter can be adjusted over full operating range and correct control knob and label are attached
Ranked causes
- 1. Oxygen flowmeter valve stuck or restricted (rare)
Adjust oxygen flow through full range
- Test point
- Oxygen flowmeter control knob
- Expected
- Smooth adjustment possible from minimum to maximum
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen ratio control (if applicable) patient safety
Verify O2 concentration remains in 21–29% range at varying oxygen flows with N2O at maximum, and N2O flow remains ≥600 ml/min at minimum O2
Ranked causes
- 1. O2/N2O ratio out of safe range or safety lock failure (occasional)
Open N2O flow control valve to stop position
- Test point
- N2O flowmeter control
- Expected
- Valve fully open
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen ratio monitor controller (if applicable, non-2-gas) patient safety
Verify N2O flow begins at O2 200–400 ml/min, O2/N2O ratio alarm functions correctly, and O2 concentration stays 21–29% across flow adjustments
Ranked causes
- 1. Oxygen ratio monitor threshold or alarm circuit malfunction (occasional)
Close oxygen flow control valve completely
- Test point
- Oxygen flowmeter control
- Expected
- Valve fully closed
19 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Ratio Monitor/Controller (ORMC) patient safety
Adjustment of ORMC to ensure correct O2/N2O ratio control and alarm lamp function on Narkomed 2B anesthesia machine.
Ranked causes
- 1. ORMC requires initial setup and calibration per procedure (common)
Initial N2O flow with closed O2 valve
- Test point
- N2O flow control valve
- Expected
- No nitrous oxide flow
11 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen sensor patient safety
Replace oxygen sensor capsule and perform 21% calibration after 15-minute stabilization
Ranked causes
- 1. Oxygen sensor requires replacement (unknown)
Set System Power switch to STANDBY
- Test point
- System Power switch
- Expected
- STANDBY position
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen supply pressure alarm switch patient safety
Oxygen supply low-pressure alarm activation setpoint adjustment for Narkomed 2B anesthesia machines with alarm switch on alarm channel
Ranked causes
- 1. Alarm setpoint requires adjustment to specification (unknown)
Disconnect all pipeline hoses
- Test point
- pipeline connections
- Expected
- all hoses removed
21 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen-Helium flowmeter (if applicable)
Verify Oxygen-Helium mixture can be adjusted over full range and concentration is correct at 2 l/min with proper knob and label
Ranked causes
- 1. O2-He concentration out of specification at baseline flow (occasional)
Set gas selector to ALL GAS
- Test point
- Gas selector switch
- Expected
- Switch in ALL GAS position
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment certification level precedence rules patient safety
Guidelines for determining final certification level when multiple DMI Recommendations or Failure Codes apply using lowest common denominator and precedence hierarchy.
Ranked causes
- 1. Multiple conditions apply with precedence hierarchy (occasional)
Identify all applicable DMI Recommendations or Failure Codes
- Test point
- Equipment condition assessment
- Expected
- Complete list of all codes and recommendations that apply
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect display
Wrong monitor screen is displayed on the system
Ranked causes
- 1. Processor assembly failure (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio alarms patient safety
Audible warnings and cautions are not being announced by the system
Ranked causes
- 1. Processor assembly failure (unknown)
Perform complete PMC on unit
- Test point
- Unit
- Expected
- Unit passes PMC
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 supply alarm does not activate at correct pressure patient safety
Oxygen supply alarm switch is not activating at the specified setpoint of 40 to 34 psi
Ranked causes
- 1. Alarm switch setpoint misaligned (common)
Measure O2 pressure at alarm activation
- Test point
- test gauge connected to TEE fitting in O2 regulator output
- Expected
- alarm activates between 40 and 34 psi, 40 to 34 psi
- Tools
- dedicated O2 test gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ORC adjustment screw not responsive
Adjusting screw fails to change N2O flow rate during calibration when turned
Ranked causes
- 1. Adjusting screw locknut too tight or screw stuck (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen analyzer calibration verification patient safety
Verify O2 sensor calibration and response after O2 CAL key activation; sensor must stabilize in ambient air before calibration
Ranked causes
- 1. O2 sensor not calibrated or unstable (common)
Verify O2 sensor stabilization in ambient air
- Test point
- O2 analyzer display
- Expected
- Reading stabilizes
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen analyzer low alarm generation patient safety
Verify low oxygen alarm message and audio alarm appear on central display and trigger continuous alarm sound
Ranked causes
- 1. O2 concentration below alarm threshold after calibration (common)
Verify % OXYGEN LOW message displays
- Test point
- Central alarm display
- Expected
- % OXYGEN LOW message
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen analyzer response at steady O2 flow patient safety
Verify O2 sensor reads correctly when oxygen is flowing at steady state 10 l/min
Ranked causes
- 1. O2 analyzer sensor drift or calibration error at operating flow (occasional)
Set oxygen flow to 10 l/min
- Test point
- Oxygen flowmeter valve
- Expected
- Flow reads 10 l/min on meter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen analyzer response to oxygen flush patient safety
Verify O2 sensor correctly reads high oxygen concentration during 10-second O2 flush at inspiratory valve
Ranked causes
- 1. O2 sensor not responding to flush or slow response (occasional)
Position O2 sensor in inspiratory valve dome adapter
- Test point
- Inspiratory valve dome adapter
- Expected
- Sensor seated and secure
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen flush cessation response patient safety
Verify O2 flow stops immediately when O2 Flush control is released
Ranked causes
- 1. O2 Flush valve sticking or leaking (occasional)
Release O2 Flush control
- Test point
- O2 Flush button
- Expected
- Button returns to unpressed state
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen supply alarm activates outside setpoint range patient safety
O2 supply pressure alarm activates below 34 psi or above 40 psi instead of within specified range
Ranked causes
- 1. Alarm setpoint incorrectly calibrated (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor board voltage measurement caution
Reference test points and acceptable voltage ranges for troubleshooting power supply issues in the NARKOMED 2B.
Ranked causes
- 1. Power supply voltages out of range (unknown)
Measure +5 VDC at capacitor C23 (+) end
- Test point
- C23 (+) end on processor board
- Expected
- 4.85 to 5.15 VDC, 4.85 to 5.15 VDC
- Tools
- Voltmeter, DMM
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator output requires compensation for cylinder pressure caution
O2 and other gas regulators output decreases 0.5 psi per 100 psi cylinder pressure increase above 1000 psi; N2O compensation not required
Ranked causes
- 1. Cylinder pressure above 1000 psi reduces regulator output (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Group 1 Left: 1 | System memory fail (checkerboard test) | Diagnostics menu self-test | error |
Group 1 Left: 2 | System memory fail (ROM compare test) | Diagnostics menu self-test | error |
Group 1 Left: 4 | CTC fail (timer function test) | Diagnostics menu self-test | error |
Group 1 Left: 8 | CTC fail (interrupt logic test) | Diagnostics menu self-test | error |
Group 1 Right: 1 | CRT controller fail (readback test) | Diagnostics menu self-test | error |
Group 1 Right: 2 | CRT memory fail (checkerboard test) | Diagnostics menu self-test | error |
Group 1 Right: 4 | CRT memory fail (ROM compare test) | Diagnostics menu self-test | error |
Group 1 Right: 8 | ROM fail (firmware CRT test) | Diagnostics menu self-test | error |
Group 2 Left: 1 | Backup audio fail (sound generation test) | Diagnostics menu self-test | warning |
Group 2 Left: 2 | Serial port B fail (channel A on UART) | Diagnostics menu self-test | error |
Group 2 Left: 4 | Serial port A fail (channel B on UART) | Diagnostics menu self-test | error |
Group 2 Left: 8 | Clock fail (cannot read) | Diagnostics menu self-test | error |
Group 2 Right: 1 | A/D conv. fail (analog V out of range) | Diagnostics menu self-test | error |
Group 2 Right: 2 | Primary audio fail (speaker test) | Diagnostics menu self-test | warning |
Group 2 Right: 4 | Primary audio fail (sound gen. test) | Diagnostics menu self-test | warning |
Group 2 Right: 8 | Backup audio fail (speaker test) | Diagnostics menu self-test | warning |
Group 3 Left: 1 | Backup memory fail (ROM compare test) | Diagnostics menu self-test | error |
Group 3 Left: 2 | Backup memory fail (array recall test) | Diagnostics menu self-test | error |
Group 3 Left: 4 | AC power fail | Diagnostics menu self-test | critical |
Group 3 Left: 8 | Reserve power fail (batt. low/disconnected) | Diagnostics menu self-test | critical |
Group 3 Right: 1 | Clock fail (write/readback test) | Diagnostics menu self-test | error |
Group 3 Right: 2 | Clock fail (original time test) | Diagnostics menu self-test | error |
Group 3 Right: 4 | Clock fail (seconds not incrementing) | Diagnostics menu self-test | error |
Group 3 Right: 8 | Backup memory fail (checkerboard test) | Diagnostics menu self-test | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI