Dräger Fabius GS: troubleshooting and error codes
The ServiceForge knowledge base holds 26 documented symptoms for the Dräger Fabius GS, with 76 ranked causes, 130 diagnostic checks with test points and expected values, and 108 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Fresh gas display blank or not working properly patient safety
Fresh gas flow meter display is blank, not responding, or showing incorrect readings
Ranked causes
- 1. Display configuration setting incorrect (2-gas vs 3-gas mismatch) (common)
- 2. Fresh gas flow sensors need recalibration (common)
- 3. Cable connections at flow sensors loose or damaged (occasional)
- 4. Cable from flow sensors to J16 on controller PCB damaged or disconnected (occasional)
- 5. Individual fresh gas flow sensor defective (occasional)
- 6. Cable from display PCB to controller PCB disconnected or faulty (occasional)
- 7. Fresh gas display PCB assembly defective (occasional)
- 8. Controller PCB assembly faulty (rare)
Verify fresh gas displays are not working properly
- Test point
- Display output and reading accuracy
- Expected
- Blank, unresponsive, or incorrect readings
13 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator inoperative patient safety
Ventilator assembly not functioning or delivering gas to breathing system
Ranked causes
- 1. Gasket, ventilator cover, hose connections or breathing system not secure (common)
- 2. Breathing bag or internal tubing damaged or improperly installed (common)
- 3. Roller diaphragm defective or vacuum system leak (occasional)
- 4. Pneumatic connections, vacuum supply, or cable connections faulty (occasional)
- 5. Ventilator diaphragms or motor defective (occasional)
- 6. Controller PCB assembly faulty (rare)
Verify ventilator is not functioning properly
- Test point
- Ventilator output and bag operation
- Expected
- No ventilation or gas delivery
11 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect or out-of-range voltage at controller PCB high voltage
Voltages measured at PCB connectors do not fall within acceptable ranges specified in Table 3-1
Ranked causes
- 1. External mains fuse open or faulty (common)
- 2. Battery fuse F4 open or faulty (common)
- 3. Battery fuse F1 open or faulty (common)
- 4. Distribution fuse F2 open or faulty (common)
- 5. Distribution fuse F3 open or faulty (common)
Measure +28 VDC at power supply input connector
- Test point
- J6-1 (+28 VDC), J6-4 (Common)
- Expected
- 26.6 VDC to 29.4 VDC, 26.6 to 29.4 VDC
- Tools
- DMM
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of O2 monitor patient safety
Oxygen concentration display not functioning or reading
Ranked causes
- 1. Sensor cord not connected to interface panel (common)
- 2. Sensor element not installed correctly or expired (common)
- 3. Sensor housing defective (occasional)
- 4. Internal cable from interface to controller PCB open or disconnected (occasional)
- 5. Controller PCB assembly faulty (rare)
Verify O2 sensor cord is connected to interface panel
- Test point
- Interface panel connector
- Expected
- Cord seated and connected
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of respiratory volume monitor patient safety
Respiratory volume measurement display not functioning or reading
Ranked causes
- 1. Sensor cord not connected to flow sensor (common)
- 2. Spirolog volume sensor not installed properly or connections not secure (common)
- 3. Spirolog volume sensor defective or worn (occasional)
- 4. Internal cable from interface to controller PCB open or disconnected (occasional)
- 5. Controller PCB assembly faulty (rare)
Verify respiratory volume monitor is functioning
- Test point
- Monitor display output
- Expected
- Respiratory volume reading displayed normally
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main display blank upon system power-up patient safety
Display screen shows no output or activity when system is powered on
Ranked causes
- 1. Software download procedure performed immediately prior to failure (common)
- 2. Power supply voltage out of tolerance (occasional)
- 3. Display panel cable disconnected or faulty (occasional)
- 4. Display panel defective (occasional)
- 5. Controller PCB assembly faulty (rare)
Verify display is blank upon system power-up
- Test point
- Main display screen
- Expected
- No display activity or output
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad inoperative patient safety
Keyboard or keypad buttons not responding to user input
Ranked causes
- 1. Keypad buttons have no tactile feedback or stuck (common)
- 2. Cable disconnected between keypad interface and controller PCB (occasional)
- 3. Monitor bezel assembly defective (occasional)
- 4. Controller PCB assembly faulty (rare)
Verify keypad is not responding to input
- Test point
- Keypad buttons and system response
- Expected
- No response to key presses
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of breathing pressure monitor patient safety
Breathing pressure measurement display not functioning
Ranked causes
- 1. Pilot line and breathing gauge connections not secure (common)
- 2. Pressure monitor calibration out of specification (occasional)
- 3. Internal tubing from interface to controller PCB damaged or disconnected (occasional)
- 4. Controller PCB assembly faulty (rare)
Verify breathing pressure monitor is functioning properly
- Test point
- Monitor display output
- Expected
- Pressure reading displayed normally
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMC certification level determination patient safety
Assessment and certification of Fabius GS machine based on equipment condition, NAD recommendations, and failure codes. Certification levels are A (Certified), B (Certified with Recommendations), C (Conditionally Certified), or D (No Certification), with additional modifiers I or II for operational constraints.
Ranked causes
- 1. Certification level A: No recommendations (unknown)
- 2. Certification level B: Certified with Recommendations (unknown)
- 3. Certification level C: Conditionally Certified (unknown)
- 4. Certification level D: No Certification (unknown)
Identify applicable NAD recommendations and codes for machine
- Test point
- NAD Recommendation Guidelines Index Anesthesia Systems Part Number S010250
- Expected
- All applicable recommendations and their certification codes (A, B, C, D, I, II) identified
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial port communication failure
Host unit cannot communicate with peripheral device through serial port
Ranked causes
- 1. Peripheral cable not connected to communication port (common)
- 2. Serial port configuration mismatch (common)
- 3. Internal communication port cable open or disconnected (occasional)
- 4. Controller PCB assembly faulty (rare)
Verify host and peripheral communication status
- Test point
- Serial communication link status indicator
- Expected
- Host communicates successfully with peripheral device
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio alarms patient safety
Audible alarm function not working; no alarm sounds are produced
Ranked causes
- 1. Cable disconnected between front panel keypad and controller PCB (common)
- 2. Front panel monitor bezel defective (occasional)
- 3. Controller PCB assembly faulty (rare)
Verify audio alarm function is not working
- Test point
- Speaker/alarm output
- Expected
- No audio output during alarm condition
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen supply pressure alarms patient safety
Low oxygen supply pressure alarm not triggering when oxygen pressure is insufficient
Ranked causes
- 1. Low O2 supply pressure switch, wire harness, or display PCB connections faulty (common)
- 2. Fresh gas display PCB defective (occasional)
- 3. Controller PCB assembly faulty (rare)
Verify no O2 in system and check alarm function
- Test point
- Oxygen supply pressure sensor and alarm output
- Expected
- Low O2 supply pressure alarm triggered with no O2 pressure
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PEEP valve calibration fails caution
PEEP calibration process displays Fail message on completion instead of Pass
Ranked causes
- 1. Invalid or incomplete pressure calibration (common)
- 2. Test hose improper connection (occasional)
- 3. System leak during calibration (occasional)
Verify pressure calibration status
- Test point
- Service menu calibration history
- Expected
- Valid pressure calibration present, no 'Inv pres. cal' message
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMS parts replacement schedule
Scheduled replacement of consumable and wear parts for Fabius GS during periodic manufacturer service. Includes flow sensors, O2 sensors, O-rings, valve discs, filters, diaphragms, batteries, and sealed assemblies.
Ranked causes
- 1. Annual maintenance parts replacement required (common)
- 2. 3-year maintenance parts replacement required (common)
- 3. Proper disposal of batteries and O2 sensors required (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fresh Gas Flow Calibration result Fail
Calibration procedure resulted in Fail status at Store Zero step
Ranked causes
- 1. Residual gas flow when valves closed (common)
- 2. Sensor malfunction or drift (occasional)
Inspect all flow control valves for proper closure
- Test point
- Flow control valves
- Expected
- All valves fully seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Machine fails to start when power is applied; no lights or display indication
Ranked causes
- 1. External mains fuse blown (common)
- 2. Battery fuse F4 open circuit (common)
Verify mains power at unit input
- Test point
- Power entry connector
- Expected
- Mains voltage present
- Tools
- DMM or power indicator
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
'Inv pres. cal' message appears during PEEP calibration setup caution
Invalid pressure calibration message displayed when attempting PEEP valve calibration
Ranked causes
- 1. Pressure calibration not completed (common)
Access calibration menu and select pressure calibration
- Test point
- Service menu Calibration option
- Expected
- Pressure calibration option available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery pack internal fault or disconnection caution
Series-connected 12 V battery cells not functioning or disconnected internally
Ranked causes
- 1. One or both 12 V battery cells failed or disconnected (rare)
Verify battery cell series connection integrity
- Test point
- J5 connector on main controller PCB assembly
- Expected
- Two series 12 V cells = ~24 V nominal minimum, 27–29.6 V at full charge
- Tools
- Voltmeter/DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of specification caution
Backup battery voltage measurement not within acceptable range during AC operation or battery operation
Ranked causes
- 1. Battery cells degraded or charge insufficient (common)
Measure total battery voltage during AC operation at full charge
- Test point
- J5 connector on main controller PCB assembly
- Expected
- 27.0 to 29.6 V, ±2.6 V
- Tools
- Voltmeter/DMM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cylinder Pressure Regulator caution
Adjust oxygen, nitrous oxide, or air cylinder pressure regulator output to specified pressure range
Ranked causes
- 1. Regulator output pressure requires calibration (common)
Set System Power switch to OFF
- Test point
- System Power switch
- Expected
- OFF
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fabius GS periodic manufacturer service patient safety
Complete periodic manufacturer's service (PMS) calibration and certification procedure for Fabius GS anesthesia machine. Includes electrical safety, system diagnostics, battery circuit, configuration, service data, calibrations, site configurations, scavenger, breathing system, vapor interlock, yokes and gauges, gas inlet regulator, cylinder regulator, high pressure leak, oxygen supply failure protection, flowmeters, oxygen monitor, oxygen concentrations, SORC, pressure monitor, ventilator, volume alarms, audio silence, oxygen flush valve, and final tests.
Ranked causes
- 1. PMS calibration and certification required (common)
Verify calibration of all test equipment
- Test point
- Calibration labels on multimeter, electrical safety analyzer, pressure gauge, flowmeter test stand, digital pressure man
- Expected
- All labels show current calibration dates
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PEEP
PEEP valve calibration procedure. A valid pressure calibration must be completed first. Verify PEEP calibration passes.
Ranked causes
- 1. Valid pressure calibration not performed (common)
Verify valid pressure calibration is stored
- Test point
- System memory/calibration history
- Expected
- Valid pressure calibration present (no 'Inv pres. cal' message)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration: Model Type (Software Version 2.X)
Verify Model Type listed on system screen is consistent with actual unit. Do not change unless required by software download or PCB replacement.
Ranked causes
- 1. Model Type changed incorrectly (rare)
Verify software version is 2.X or later
- Test point
- System Settings screen
- Expected
- Software Version 2.X displayed (if applicable)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 Supply Alarm Activates Outside Acceptable Range patient safety
Oxygen supply alarm triggers at pressure below 16 psi or above 24 psi instead of within the correct range
Ranked causes
- 1. Alarm set screw adjustment out of specification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply replacement high voltage
Procedure to replace the power supply unit mounted on the controller assembly back panel
Ranked causes
- 1. Power supply failure or malfunction requiring replacement (unknown)
Set main power switch to OFF
- Test point
- Main power switch
- Expected
- OFF position
12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vapor interlock system malfunction patient safety
Multiple vaporizers can be adjusted simultaneously instead of being locked to prevent concurrent use
Ranked causes
- 1. Vapor interlock mechanism failure (unknown)
Left vapor interlock function
- Test point
- Left vaporizer handwheel
- Expected
- Right vaporizer handwheel cannot be adjusted when left is set above zero
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
APNEA FLOW !!! | No airflow detected; high-level apnea alarm | No airflow detected for 26-34 seconds after breathing circuit disconnection | critical |
APNEA FLOW !! | No airflow detected; mid-level apnea alarm | No airflow detected for 13-17 seconds after breathing circuit disconnection | error |
APNEA PRESSURE !!! | No airway pressure detected; high-level apnea alarm | No airway pressure detected after breathing circuit disconnection | critical |
C001 | Handler Busy | Communication error during device handler operation | warning |
C002 | Incoming Overflow | Incoming communication buffer overflow | warning |
C003 | Outgoing queue overflow | Outgoing communication queue overflowed | warning |
C004 | Message not registered | Communication message failed to register | warning |
C005 | Character not transmitted | Character transmission failed | warning |
C006 | Interrupt problem | Hardware interrupt handling failure | error |
C007 | Continuous interrupts | Continuous interrupt generation | error |
C008 | Excessive Breaks | Excessive break signals on communication line | warning |
C009 | Excessive Framing | Excessive framing errors in communication | warning |
C010 | Excessive Parity | Excessive parity errors detected | warning |
C011 | Excessive Overrun | Excessive overrun errors in communication | warning |
C012 | Non-existent UART | UART port not detected | critical |
CONDITIONALLY FUNCTIONAL | Minor problem has been detected during diagnostics | Power-up self-test completion with minor issue | warning |
F001 | Diag Watchdog | Watchdog timer expired during diagnostics | critical |
F002 | Diag System Ram | System RAM fails on power up | critical |
F003 | Diag Program Memory | Flash memory error detected on startup | critical |
F004 | Diag Video Test | Video memory or display failed | error |
F005 | Diag Interrupts | Interrupt handling test failed | error |
F006 | Diag A/D Converter | Analog-to-digital converter test failed | error |
F007 | Diag NM Ram | Non-volatile memory corruption detected on startup | error |
F008 | Diag Serial Port | Serial port loopback test failed on startup | error |
F009 | Diag Clock | System clock test failed | critical |
F010 | Diag Speaker | Speaker not detected on power up | warning |
F011 | Diag Main Power | Mains power not detected on power up | error |
F012 | Diag Battery | Battery voltage below 20% on power up | warning |
FG01 | N2O CAL ERR | N2O sensor value out of range during calibration | error |
FG02 | AIR CAL ERR | AIR sensor value out of range during calibration | error |
FG03 | O2 CAL ERR | O2 sensor value out of range during calibration | error |
FLOW SENSOR FAIL ! | Respiratory volume sensor or flow sensor malfunction | Respiratory volume sensor cord disconnected from flow sensor | error |
FRESH GAS LOW !! | Fresh gas flow insufficient | Fresh gas flow below minimum threshold for 30 seconds or when inspiratory port disconnected | warning |
FUNCTIONAL | Fabius GS has passed all power-up tests and is fully functional | Power-up self-test completion | info |
I001 | CPU Diag Recovery | CPU recovered from bad initialization | warning |
I011 | L2 SW Watchdog | Level 2 software watchdog timer expired | critical |
I012 | L2 Recovered, wLevelTwoErrorCount | Level 2 software watchdog timeout with recovery | warning |
I021 | L3 SW Watchdog | Level 3 software watchdog timer expired | critical |
I022 | L3 Recovered, wLevelThreeErrorCount | Level 3 software watchdog timeout with recovery | warning |
I031 | L4 SW Watchdog | Level 4 software watchdog timer expired | critical |
I032 | L4 Recovered, wLevelFourErrorCount | Level 4 software watchdog timeout with recovery | warning |
Inv pres. cal | Invalid pressure calibration detected | Attempt to perform PEEP valve calibration without valid prior pressure calibration | error |
M001 | O2 <10% | Oxygen concentration drops below 10% | critical |
M002 | O2 LOW INOP | O2 sensor inoperative or disconnected | critical |
M003 | O2 DIFF HIGH | Oxygen difference between sensors exceeds threshold | error |
M004 | BATTERY <10% | Battery capacity drops below 10% during battery operation | critical |
M005 | BATTERY <20% | Battery capacity drops below 20% during battery operation | warning |
MINUTE VOLUME HIGH !! | Minute volume exceeds high alarm threshold | Minute volume exceeds high alarm limit during ventilation | warning |
MINUTE VOLUME LOW !! | Minute volume below low alarm threshold | Minute volume falls below low alarm limit during ventilation | warning |
No Fresh Gas !!! | No fresh gas flow detected (software version 1.3n only) | No fresh gas flow detected when No Fresh Gas mode is enabled (version 1.3n) | critical |
NON-FUNCTIONAL | Serious problem has been detected; machine will not proceed to Monitor screen | Power-up self-test completion with critical failure | critical |
NVC0 | Non-Matching CRCS | NVRAM CRCs for primary and secondary do not match | error |
NVF0 | Pri and Sec NV Files Inv | Both primary and secondary NVRAM files invalid | critical |
NVF1 | Pri NV File Inv | Primary NVRAM file invalid | error |
NVF2 | Sec NV File Inv | Secondary NVRAM file invalid | error |
NVP1 | Pat Check Pri Fail | NVRAM primary pattern write test failed | error |
NVP2 | Pat Check Sec Fail | NVRAM secondary pattern write test failed | error |
O2 Sensor Calibration Failed | Oxygen sensor calibration procedure did not complete successfully | occurs during 21% O2 calibration step | error |
PS01 | Charger Error | Battery charger timeout or failure | warning |
PS02 | -2.5V Supply Failure | Negative 2.5V supply voltage out of range | error |
PS03 | +12V Supply Failure | Positive 12V supply voltage out of range | error |
PS04 | -5V Supply Failure | Negative 5V supply voltage out of range | error |
PS05 | Battery Less Than 18V | Battery voltage drops below 18V | warning |
PS06 | +28V Supply Failure | Positive 28V supply voltage out of range | error |
PS07 | Temp Sensor Failure | Temperature sensor voltage out of range | error |
PS08 | 2.5V Supply Failure | Positive 2.5V supply voltage out of range | error |
PS09 | 3.3V Supply Failure | Positive 3.3V supply voltage out of range | error |
PS10 | +10V Supply Failure | Positive 10V supply voltage out of range | error |
PS11 | +16V Supply Failure | Positive 16V supply voltage out of range | error |
S006 | Service Log Clear | Service log cleared by technician or software upgrade | info |
SS90 | test message | Factory test during manufacturing | info |
U001 | New Device ID Created | Device ID created due to NVRAM corruption or missing ID | warning |
U002 | Option Enabled | Secure option was enabled | info |
U003 | Option Disabled | Secure option was disabled | info |
U007 | Model Changed to FGS | Model type changed to FGS configuration | info |
U008 | Model Changed to Tiro | Model type changed to Tiro configuration | info |
U009 | Monitoring Disabled | Monitoring mode was disabled | warning |
U010 | Monitoring Enabled | Monitoring mode was enabled | info |
U012 | SOS Recovered | Secure Option Setting recovered after partial NVRAM corruption | warning |
U013 | FGS SOS Invalid | FGS Secure Options Setting NVRAM file corrupt, no recovery | error |
U014 | Tiro SOS Invalid | Tiro Secure Options Setting NVRAM file corrupt, no recovery | error |
V001 | VENT PEEP NOT CAL | Auto ventilation attempted without valid PEEP calibration | error |
V002 | ZERO POS INVALID | Light barrier crossing at bottom vent not detected | error |
V037 | EX VALVE LEAK | Excessive leak detected in exhalation valve | warning |
V042 | Motor Voltage Error | Motor voltage drops below minimum during auto ventilation | error |
V043 | MOTOR LIM STOP | Excessive motor current during auto ventilation | error |
V044 | MEMBRANE PRES LOW | Membrane vacuum below normal operating range | warning |
V045 | MEMBRANE PRES HIGH | Membrane vacuum above normal operating range | warning |
V049 | MOTOR OVER CURRENT | Persistent excessive motor current causing shutdown | critical |
V050 | LT Pres Failed>350 | Leak test pressure exceeded 350 mbar | error |
V051 | LT Pres Rose>250 | Leak test pressure rose more than 250 mbar | error |
V052 | LT SYS FAIL | Leak test system failure | error |
V053 | LT PC FAIL | Leak test pressure control failure | error |
V060 | Comp Unexp State | Compliance test entered unexpected state | error |
V061 | Comp Bag Pres Too High | Compliance test start pressure exceeds limit | error |
V062 | Comp Cant Fully Pres | Compliance test unable to achieve full pressurization | error |
V063 | Comp Cant Hold Pres | Compliance test unable to maintain plateau pressure | error |
V064 | COMPCant DePres | Pressure not decreasing during compliance test | error |
V065 | COMPInv Pres | Pressure changing in wrong direction during compliance test | error |
V066 | COMPInv Pres | Pressure changing in wrong direction during compliance test | error |
V067 | COMPSys < VentHose | Irregular compliance test results | warning |
V069 | Comp Unexpected Error | Compliance test entered unexpected error state | error |
V070 | COMPSys Fail in Range | Compliance value unusable within range | error |
V071 | COMPSys Fail out of Range | Compliance value unusable outside range | error |
V072 | COMPVHFail in Range | Vent hose compliance value unusable within range | error |
V073 | COMPVHFail out of Range | Vent hose compliance value unusable outside range | error |
value not stored | O2 sensor calibration failed to store zero values | O2 zero calibration procedure failed | error |
WD00 | Watch dog time exp | Watchdog timer fired | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI