Anesthesia machines · Fabius family

Dräger Fabius GS: troubleshooting and error codes

The ServiceForge knowledge base holds 26 documented symptoms for the Dräger Fabius GS, with 76 ranked causes, 130 diagnostic checks with test points and expected values, and 108 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Fresh gas display blank or not working properly patient safety

Fresh gas flow meter display is blank, not responding, or showing incorrect readings

Ranked causes

  1. 1. Display configuration setting incorrect (2-gas vs 3-gas mismatch) (common)
  2. 2. Fresh gas flow sensors need recalibration (common)
  3. 3. Cable connections at flow sensors loose or damaged (occasional)
  4. 4. Cable from flow sensors to J16 on controller PCB damaged or disconnected (occasional)
  5. 5. Individual fresh gas flow sensor defective (occasional)
  6. 6. Cable from display PCB to controller PCB disconnected or faulty (occasional)
  7. 7. Fresh gas display PCB assembly defective (occasional)
  8. 8. Controller PCB assembly faulty (rare)
First check

Verify fresh gas displays are not working properly

Test point
Display output and reading accuracy
Expected
Blank, unresponsive, or incorrect readings

13 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator inoperative patient safety

Ventilator assembly not functioning or delivering gas to breathing system

Ranked causes

  1. 1. Gasket, ventilator cover, hose connections or breathing system not secure (common)
  2. 2. Breathing bag or internal tubing damaged or improperly installed (common)
  3. 3. Roller diaphragm defective or vacuum system leak (occasional)
  4. 4. Pneumatic connections, vacuum supply, or cable connections faulty (occasional)
  5. 5. Ventilator diaphragms or motor defective (occasional)
  6. 6. Controller PCB assembly faulty (rare)
First check

Verify ventilator is not functioning properly

Test point
Ventilator output and bag operation
Expected
No ventilation or gas delivery

11 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect or out-of-range voltage at controller PCB high voltage

Voltages measured at PCB connectors do not fall within acceptable ranges specified in Table 3-1

Ranked causes

  1. 1. External mains fuse open or faulty (common)
  2. 2. Battery fuse F4 open or faulty (common)
  3. 3. Battery fuse F1 open or faulty (common)
  4. 4. Distribution fuse F2 open or faulty (common)
  5. 5. Distribution fuse F3 open or faulty (common)
First check

Measure +28 VDC at power supply input connector

Test point
J6-1 (+28 VDC), J6-4 (Common)
Expected
26.6 VDC to 29.4 VDC, 26.6 to 29.4 VDC
Tools
DMM

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of O2 monitor patient safety

Oxygen concentration display not functioning or reading

Ranked causes

  1. 1. Sensor cord not connected to interface panel (common)
  2. 2. Sensor element not installed correctly or expired (common)
  3. 3. Sensor housing defective (occasional)
  4. 4. Internal cable from interface to controller PCB open or disconnected (occasional)
  5. 5. Controller PCB assembly faulty (rare)
First check

Verify O2 sensor cord is connected to interface panel

Test point
Interface panel connector
Expected
Cord seated and connected

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of respiratory volume monitor patient safety

Respiratory volume measurement display not functioning or reading

Ranked causes

  1. 1. Sensor cord not connected to flow sensor (common)
  2. 2. Spirolog volume sensor not installed properly or connections not secure (common)
  3. 3. Spirolog volume sensor defective or worn (occasional)
  4. 4. Internal cable from interface to controller PCB open or disconnected (occasional)
  5. 5. Controller PCB assembly faulty (rare)
First check

Verify respiratory volume monitor is functioning

Test point
Monitor display output
Expected
Respiratory volume reading displayed normally

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main display blank upon system power-up patient safety

Display screen shows no output or activity when system is powered on

Ranked causes

  1. 1. Software download procedure performed immediately prior to failure (common)
  2. 2. Power supply voltage out of tolerance (occasional)
  3. 3. Display panel cable disconnected or faulty (occasional)
  4. 4. Display panel defective (occasional)
  5. 5. Controller PCB assembly faulty (rare)
First check

Verify display is blank upon system power-up

Test point
Main display screen
Expected
No display activity or output

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad inoperative patient safety

Keyboard or keypad buttons not responding to user input

Ranked causes

  1. 1. Keypad buttons have no tactile feedback or stuck (common)
  2. 2. Cable disconnected between keypad interface and controller PCB (occasional)
  3. 3. Monitor bezel assembly defective (occasional)
  4. 4. Controller PCB assembly faulty (rare)
First check

Verify keypad is not responding to input

Test point
Keypad buttons and system response
Expected
No response to key presses

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of breathing pressure monitor patient safety

Breathing pressure measurement display not functioning

Ranked causes

  1. 1. Pilot line and breathing gauge connections not secure (common)
  2. 2. Pressure monitor calibration out of specification (occasional)
  3. 3. Internal tubing from interface to controller PCB damaged or disconnected (occasional)
  4. 4. Controller PCB assembly faulty (rare)
First check

Verify breathing pressure monitor is functioning properly

Test point
Monitor display output
Expected
Pressure reading displayed normally

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PMC certification level determination patient safety

Assessment and certification of Fabius GS machine based on equipment condition, NAD recommendations, and failure codes. Certification levels are A (Certified), B (Certified with Recommendations), C (Conditionally Certified), or D (No Certification), with additional modifiers I or II for operational constraints.

Ranked causes

  1. 1. Certification level A: No recommendations (unknown)
  2. 2. Certification level B: Certified with Recommendations (unknown)
  3. 3. Certification level C: Conditionally Certified (unknown)
  4. 4. Certification level D: No Certification (unknown)
First check

Identify applicable NAD recommendations and codes for machine

Test point
NAD Recommendation Guidelines Index Anesthesia Systems Part Number S010250
Expected
All applicable recommendations and their certification codes (A, B, C, D, I, II) identified

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial port communication failure

Host unit cannot communicate with peripheral device through serial port

Ranked causes

  1. 1. Peripheral cable not connected to communication port (common)
  2. 2. Serial port configuration mismatch (common)
  3. 3. Internal communication port cable open or disconnected (occasional)
  4. 4. Controller PCB assembly faulty (rare)
First check

Verify host and peripheral communication status

Test point
Serial communication link status indicator
Expected
Host communicates successfully with peripheral device

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No audio alarms patient safety

Audible alarm function not working; no alarm sounds are produced

Ranked causes

  1. 1. Cable disconnected between front panel keypad and controller PCB (common)
  2. 2. Front panel monitor bezel defective (occasional)
  3. 3. Controller PCB assembly faulty (rare)
First check

Verify audio alarm function is not working

Test point
Speaker/alarm output
Expected
No audio output during alarm condition

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No oxygen supply pressure alarms patient safety

Low oxygen supply pressure alarm not triggering when oxygen pressure is insufficient

Ranked causes

  1. 1. Low O2 supply pressure switch, wire harness, or display PCB connections faulty (common)
  2. 2. Fresh gas display PCB defective (occasional)
  3. 3. Controller PCB assembly faulty (rare)
First check

Verify no O2 in system and check alarm function

Test point
Oxygen supply pressure sensor and alarm output
Expected
Low O2 supply pressure alarm triggered with no O2 pressure

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PEEP valve calibration fails caution

PEEP calibration process displays Fail message on completion instead of Pass

Ranked causes

  1. 1. Invalid or incomplete pressure calibration (common)
  2. 2. Test hose improper connection (occasional)
  3. 3. System leak during calibration (occasional)
First check

Verify pressure calibration status

Test point
Service menu calibration history
Expected
Valid pressure calibration present, no 'Inv pres. cal' message

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PMS parts replacement schedule

Scheduled replacement of consumable and wear parts for Fabius GS during periodic manufacturer service. Includes flow sensors, O2 sensors, O-rings, valve discs, filters, diaphragms, batteries, and sealed assemblies.

Ranked causes

  1. 1. Annual maintenance parts replacement required (common)
  2. 2. 3-year maintenance parts replacement required (common)
  3. 3. Proper disposal of batteries and O2 sensors required (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fresh Gas Flow Calibration result Fail

Calibration procedure resulted in Fail status at Store Zero step

Ranked causes

  1. 1. Residual gas flow when valves closed (common)
  2. 2. Sensor malfunction or drift (occasional)
First check

Inspect all flow control valves for proper closure

Test point
Flow control valves
Expected
All valves fully seated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on high voltage

Machine fails to start when power is applied; no lights or display indication

Ranked causes

  1. 1. External mains fuse blown (common)
  2. 2. Battery fuse F4 open circuit (common)
First check

Verify mains power at unit input

Test point
Power entry connector
Expected
Mains voltage present
Tools
DMM or power indicator

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

'Inv pres. cal' message appears during PEEP calibration setup caution

Invalid pressure calibration message displayed when attempting PEEP valve calibration

Ranked causes

  1. 1. Pressure calibration not completed (common)
First check

Access calibration menu and select pressure calibration

Test point
Service menu Calibration option
Expected
Pressure calibration option available

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack internal fault or disconnection caution

Series-connected 12 V battery cells not functioning or disconnected internally

Ranked causes

  1. 1. One or both 12 V battery cells failed or disconnected (rare)
First check

Verify battery cell series connection integrity

Test point
J5 connector on main controller PCB assembly
Expected
Two series 12 V cells = ~24 V nominal minimum, 27–29.6 V at full charge
Tools
Voltmeter/DMM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage out of specification caution

Backup battery voltage measurement not within acceptable range during AC operation or battery operation

Ranked causes

  1. 1. Battery cells degraded or charge insufficient (common)
First check

Measure total battery voltage during AC operation at full charge

Test point
J5 connector on main controller PCB assembly
Expected
27.0 to 29.6 V, ±2.6 V
Tools
Voltmeter/DMM

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cylinder Pressure Regulator caution

Adjust oxygen, nitrous oxide, or air cylinder pressure regulator output to specified pressure range

Ranked causes

  1. 1. Regulator output pressure requires calibration (common)
First check

Set System Power switch to OFF

Test point
System Power switch
Expected
OFF

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Fabius GS periodic manufacturer service patient safety

Complete periodic manufacturer's service (PMS) calibration and certification procedure for Fabius GS anesthesia machine. Includes electrical safety, system diagnostics, battery circuit, configuration, service data, calibrations, site configurations, scavenger, breathing system, vapor interlock, yokes and gauges, gas inlet regulator, cylinder regulator, high pressure leak, oxygen supply failure protection, flowmeters, oxygen monitor, oxygen concentrations, SORC, pressure monitor, ventilator, volume alarms, audio silence, oxygen flush valve, and final tests.

Ranked causes

  1. 1. PMS calibration and certification required (common)
First check

Verify calibration of all test equipment

Test point
Calibration labels on multimeter, electrical safety analyzer, pressure gauge, flowmeter test stand, digital pressure man
Expected
All labels show current calibration dates

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PEEP

PEEP valve calibration procedure. A valid pressure calibration must be completed first. Verify PEEP calibration passes.

Ranked causes

  1. 1. Valid pressure calibration not performed (common)
First check

Verify valid pressure calibration is stored

Test point
System memory/calibration history
Expected
Valid pressure calibration present (no 'Inv pres. cal' message)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Configuration: Model Type (Software Version 2.X)

Verify Model Type listed on system screen is consistent with actual unit. Do not change unless required by software download or PCB replacement.

Ranked causes

  1. 1. Model Type changed incorrectly (rare)
First check

Verify software version is 2.X or later

Test point
System Settings screen
Expected
Software Version 2.X displayed (if applicable)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 Supply Alarm Activates Outside Acceptable Range patient safety

Oxygen supply alarm triggers at pressure below 16 psi or above 24 psi instead of within the correct range

Ranked causes

  1. 1. Alarm set screw adjustment out of specification (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply replacement high voltage

Procedure to replace the power supply unit mounted on the controller assembly back panel

Ranked causes

  1. 1. Power supply failure or malfunction requiring replacement (unknown)
First check

Set main power switch to OFF

Test point
Main power switch
Expected
OFF position

12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vapor interlock system malfunction patient safety

Multiple vaporizers can be adjusted simultaneously instead of being locked to prevent concurrent use

Ranked causes

  1. 1. Vapor interlock mechanism failure (unknown)
First check

Left vapor interlock function

Test point
Left vaporizer handwheel
Expected
Right vaporizer handwheel cannot be adjusted when left is set above zero

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
APNEA FLOW !!!No airflow detected; high-level apnea alarmNo airflow detected for 26-34 seconds after breathing circuit disconnectioncritical
APNEA FLOW !!No airflow detected; mid-level apnea alarmNo airflow detected for 13-17 seconds after breathing circuit disconnectionerror
APNEA PRESSURE !!!No airway pressure detected; high-level apnea alarmNo airway pressure detected after breathing circuit disconnectioncritical
C001Handler BusyCommunication error during device handler operationwarning
C002Incoming OverflowIncoming communication buffer overflowwarning
C003Outgoing queue overflowOutgoing communication queue overflowedwarning
C004Message not registeredCommunication message failed to registerwarning
C005Character not transmittedCharacter transmission failedwarning
C006Interrupt problemHardware interrupt handling failureerror
C007Continuous interruptsContinuous interrupt generationerror
C008Excessive BreaksExcessive break signals on communication linewarning
C009Excessive FramingExcessive framing errors in communicationwarning
C010Excessive ParityExcessive parity errors detectedwarning
C011Excessive OverrunExcessive overrun errors in communicationwarning
C012Non-existent UARTUART port not detectedcritical
CONDITIONALLY FUNCTIONALMinor problem has been detected during diagnosticsPower-up self-test completion with minor issuewarning
F001Diag WatchdogWatchdog timer expired during diagnosticscritical
F002Diag System RamSystem RAM fails on power upcritical
F003Diag Program MemoryFlash memory error detected on startupcritical
F004Diag Video TestVideo memory or display failederror
F005Diag InterruptsInterrupt handling test failederror
F006Diag A/D ConverterAnalog-to-digital converter test failederror
F007Diag NM RamNon-volatile memory corruption detected on startuperror
F008Diag Serial PortSerial port loopback test failed on startuperror
F009Diag ClockSystem clock test failedcritical
F010Diag SpeakerSpeaker not detected on power upwarning
F011Diag Main PowerMains power not detected on power uperror
F012Diag BatteryBattery voltage below 20% on power upwarning
FG01N2O CAL ERRN2O sensor value out of range during calibrationerror
FG02AIR CAL ERRAIR sensor value out of range during calibrationerror
FG03O2 CAL ERRO2 sensor value out of range during calibrationerror
FLOW SENSOR FAIL !Respiratory volume sensor or flow sensor malfunctionRespiratory volume sensor cord disconnected from flow sensorerror
FRESH GAS LOW !!Fresh gas flow insufficientFresh gas flow below minimum threshold for 30 seconds or when inspiratory port disconnectedwarning
FUNCTIONALFabius GS has passed all power-up tests and is fully functionalPower-up self-test completioninfo
I001CPU Diag RecoveryCPU recovered from bad initializationwarning
I011L2 SW WatchdogLevel 2 software watchdog timer expiredcritical
I012L2 Recovered, wLevelTwoErrorCountLevel 2 software watchdog timeout with recoverywarning
I021L3 SW WatchdogLevel 3 software watchdog timer expiredcritical
I022L3 Recovered, wLevelThreeErrorCountLevel 3 software watchdog timeout with recoverywarning
I031L4 SW WatchdogLevel 4 software watchdog timer expiredcritical
I032L4 Recovered, wLevelFourErrorCountLevel 4 software watchdog timeout with recoverywarning
Inv pres. calInvalid pressure calibration detectedAttempt to perform PEEP valve calibration without valid prior pressure calibrationerror
M001O2 <10%Oxygen concentration drops below 10%critical
M002O2 LOW INOPO2 sensor inoperative or disconnectedcritical
M003O2 DIFF HIGHOxygen difference between sensors exceeds thresholderror
M004BATTERY <10%Battery capacity drops below 10% during battery operationcritical
M005BATTERY <20%Battery capacity drops below 20% during battery operationwarning
MINUTE VOLUME HIGH !!Minute volume exceeds high alarm thresholdMinute volume exceeds high alarm limit during ventilationwarning
MINUTE VOLUME LOW !!Minute volume below low alarm thresholdMinute volume falls below low alarm limit during ventilationwarning
No Fresh Gas !!!No fresh gas flow detected (software version 1.3n only)No fresh gas flow detected when No Fresh Gas mode is enabled (version 1.3n)critical
NON-FUNCTIONALSerious problem has been detected; machine will not proceed to Monitor screenPower-up self-test completion with critical failurecritical
NVC0Non-Matching CRCSNVRAM CRCs for primary and secondary do not matcherror
NVF0Pri and Sec NV Files InvBoth primary and secondary NVRAM files invalidcritical
NVF1Pri NV File InvPrimary NVRAM file invaliderror
NVF2Sec NV File InvSecondary NVRAM file invaliderror
NVP1Pat Check Pri FailNVRAM primary pattern write test failederror
NVP2Pat Check Sec FailNVRAM secondary pattern write test failederror
O2 Sensor Calibration FailedOxygen sensor calibration procedure did not complete successfullyoccurs during 21% O2 calibration steperror
PS01Charger ErrorBattery charger timeout or failurewarning
PS02-2.5V Supply FailureNegative 2.5V supply voltage out of rangeerror
PS03+12V Supply FailurePositive 12V supply voltage out of rangeerror
PS04-5V Supply FailureNegative 5V supply voltage out of rangeerror
PS05Battery Less Than 18VBattery voltage drops below 18Vwarning
PS06+28V Supply FailurePositive 28V supply voltage out of rangeerror
PS07Temp Sensor FailureTemperature sensor voltage out of rangeerror
PS082.5V Supply FailurePositive 2.5V supply voltage out of rangeerror
PS093.3V Supply FailurePositive 3.3V supply voltage out of rangeerror
PS10+10V Supply FailurePositive 10V supply voltage out of rangeerror
PS11+16V Supply FailurePositive 16V supply voltage out of rangeerror
S006Service Log ClearService log cleared by technician or software upgradeinfo
SS90test messageFactory test during manufacturinginfo
U001New Device ID CreatedDevice ID created due to NVRAM corruption or missing IDwarning
U002Option EnabledSecure option was enabledinfo
U003Option DisabledSecure option was disabledinfo
U007Model Changed to FGSModel type changed to FGS configurationinfo
U008Model Changed to TiroModel type changed to Tiro configurationinfo
U009Monitoring DisabledMonitoring mode was disabledwarning
U010Monitoring EnabledMonitoring mode was enabledinfo
U012SOS RecoveredSecure Option Setting recovered after partial NVRAM corruptionwarning
U013FGS SOS InvalidFGS Secure Options Setting NVRAM file corrupt, no recoveryerror
U014Tiro SOS InvalidTiro Secure Options Setting NVRAM file corrupt, no recoveryerror
V001VENT PEEP NOT CALAuto ventilation attempted without valid PEEP calibrationerror
V002ZERO POS INVALIDLight barrier crossing at bottom vent not detectederror
V037EX VALVE LEAKExcessive leak detected in exhalation valvewarning
V042Motor Voltage ErrorMotor voltage drops below minimum during auto ventilationerror
V043MOTOR LIM STOPExcessive motor current during auto ventilationerror
V044MEMBRANE PRES LOWMembrane vacuum below normal operating rangewarning
V045MEMBRANE PRES HIGHMembrane vacuum above normal operating rangewarning
V049MOTOR OVER CURRENTPersistent excessive motor current causing shutdowncritical
V050LT Pres Failed>350Leak test pressure exceeded 350 mbarerror
V051LT Pres Rose>250Leak test pressure rose more than 250 mbarerror
V052LT SYS FAILLeak test system failureerror
V053LT PC FAILLeak test pressure control failureerror
V060Comp Unexp StateCompliance test entered unexpected stateerror
V061Comp Bag Pres Too HighCompliance test start pressure exceeds limiterror
V062Comp Cant Fully PresCompliance test unable to achieve full pressurizationerror
V063Comp Cant Hold PresCompliance test unable to maintain plateau pressureerror
V064COMPCant DePresPressure not decreasing during compliance testerror
V065COMPInv PresPressure changing in wrong direction during compliance testerror
V066COMPInv PresPressure changing in wrong direction during compliance testerror
V067COMPSys < VentHoseIrregular compliance test resultswarning
V069Comp Unexpected ErrorCompliance test entered unexpected error stateerror
V070COMPSys Fail in RangeCompliance value unusable within rangeerror
V071COMPSys Fail out of RangeCompliance value unusable outside rangeerror
V072COMPVHFail in RangeVent hose compliance value unusable within rangeerror
V073COMPVHFail out of RangeVent hose compliance value unusable outside rangeerror
value not storedO2 sensor calibration failed to store zero valuesO2 zero calibration procedure failederror
WD00Watch dog time expWatchdog timer firedcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.