Dräger Capno: troubleshooting and error codes

The ServiceForge knowledge base holds 9 documented symptoms for the Dräger Capno, with 13 ranked causes, 10 diagnostic checks with test points and expected values, and 9 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

CO2 sensor not detecting patient safety

CO2 sensor unplugged or not positioned correctly on cuvette. Device displays CO2-Sensor? alarm indicating sensor communication failure.

Ranked causes

  1. 1. Sensor physically disconnected (common)
  2. 2. Sensor not positioned on cuvette (occasional)
  3. 3. Sensor defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 cuvette window soiled patient safety

Cuvette window is soiled or CO2 sensor window is dirty, preventing accurate measurement. Device displays Clean CO2 cuvette alarm.

Ranked causes

  1. 1. Cuvette lens contaminated (common)
  2. 2. Sensor window contaminated (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 measurement fault patient safety

CO2 sensor or measurement circuit has detected a fault. Device displays CO2 measurement faulty alarm with high priority. Readings cannot be trusted.

Ranked causes

  1. 1. CO2 sensor defective (common)
  2. 2. Measurement circuit malfunction (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 sensor

Routine calibration procedure for CO2 sensor before measurement. Apparatus may prompt calibration if curve does not return to zero after expiration or when specifically demanded.

Ranked causes

  1. 1. Routine calibration required before CO2 measurement (common)
First check

Install CO2 cuvette into Y-piece patient socket

Test point
Patient socket of Y-piece
Expected
Cuvette windows face sideways

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 calibration failed

Apparatus displays 'CO2 calibration not ok' message instead of success message after calibration cycle.

Ranked causes

  1. 1. Calibration procedure error or sensor fault (unknown)
First check

Observe calibration failure message

Test point
Display
Expected
'CO2 calibration not ok' message

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 calibration out of tolerance patient safety

Zero point calibration of CO2 sensor has drifted out of tolerance. Device displays CO2 calibration? alarm with top priority.

Ranked causes

  1. 1. Calibration drift (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 monitoring disabled patient safety

CO2 monitoring feature has been deactivated. Low priority advisory alarm indicates monitoring is off and patient may not be receiving CO2 data.

Ranked causes

  1. 1. Monitoring was turned off by operator (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

End-tidal CO2 concentration too high patient safety

Measured end-tidal CO2 concentration has exceeded upper alarm limit. High priority alarm indicates patient hypercarbia or ventilation problem.

Ranked causes

  1. 1. Ventilation inadequate or patient hypercarbia (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

End-tidal CO2 concentration too low patient safety

Measured end-tidal CO2 concentration has fallen below lower alarm limit. High priority alarm indicates patient hypocapnia or inadequate ventilation.

Ranked causes

  1. 1. Ventilation excessive or patient hypocapnia (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Clean CO2 cuvetteCuvette or sensor window contamination preventing measurementOptical lens soiled with debris or condensationcritical
CO2 calibration?Zero point calibration drift out of toleranceCalibration lost or drifted; unsuccessful calibration attemptcritical
CO2 calibration interruptedCO2 sensor calibration process halted before completionError or interruption during test gas calibration procedureerror
CO2 calibration not okCO2 sensor calibration procedure completed but calibration unsuccessfulMeasured values outside acceptable tolerance during test gas calibrationerror
CO2 measurement faultyCO2 sensor or measurement electronics malfunctionSensor circuit or sensor element failure; measurement data unreliablecritical
CO2 monitoring offCO2 monitoring feature has been deactivated by operatorOperator disabled monitoring featurewarning
CO2-Sensor?CO2 sensor communication or connection failureSensor unplugged, not positioned on cuvette, or defectivecritical
etCO2 too highEnd-tidal CO2 concentration exceeds upper alarm thresholdMeasured CO2 above alarm limitcritical
etCO2 too lowEnd-tidal CO2 concentration below lower alarm thresholdMeasured CO2 below alarm limitcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.