Dräger Apollo: troubleshooting and error codes
The ServiceForge knowledge base holds 5 documented symptoms for the Dräger Apollo, with 6 ranked causes, 14 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Gas measurement module assembly and installation caution
Replacing tube stump at T piece, connecting flow restrictor to module housing connector 3, aligning flow restrictor adhesive join toward T piece, ensuring tube is not kinked, closing housing without trapping tubes, reinserting module, affixing FLOW sticker, and performing test
Ranked causes
- 1. Initial module assembly or tube replacement required (unknown)
- 2. Post-installation testing required (common)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device startup in normal measurement mode and handover
Final startup and validation that device has reached ISO accuracy level for gas measurement
Ranked causes
- 1. Final validation and handover (common)
Restart device and initiate self-test
- Test point
- Device display
- Expected
- Device powered on; self-test initiated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak tightness test of pneumatics failing
Pressure rise exceeds tolerance during leak tightness test; indicates leak in gas measurement module pneumatic system
Ranked causes
- 1. Pneumatic leak in gas measurement module (common)
Connect service laptop via RS 232 extension cable and switch on equipment
- Test point
- Service laptop and device
- Expected
- Service laptop and device powered; service mode activated if applicable; VISIA software started and connected
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Sample gas flow rate out of specification
Sample gas flow rate does not fall within specified tolerance; indicates flow restrictor malfunction or tubing obstruction
Ranked causes
- 1. Sample gas flow rate outside tolerance (common)
Set up sample gas flow rate test equipment as per Fig. 5
- Test point
- Test setup
- Expected
- Flowmeter 300 mL, tube, and WaterLock water trap connector assembled
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Suction rate too high patient safety
Flow restrictor in anaesthetic gas measurement module allows excessive suction rate; occurs in machines manufactured June 2007 to November 2008
Ranked causes
- 1. Flow restrictor requires replacement or bypass modification (common)
Verify correct modification kit selected
- Test point
- Pump regulator size
- Expected
- 150ml pump uses kit 6872185; 200ml pump uses kit 6872195
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI