Dilon Navigator 2.0: troubleshooting and error codes
The ServiceForge knowledge base holds 8 documented symptoms for the Dilon Navigator 2.0, with 21 ranked causes, 17 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Display is dark
No visual indication on control unit display, unit appears non-functional.
Ranked causes
- 1. Power switch is off (common)
- 2. Power switch is broken (occasional)
- 3. Battery is completely discharged (common)
- 4. Fuse is blown or missing (occasional)
- 5. Damaged PCB in control unit (rare)
Verify power switch position
- Test point
- Power switch on rear of control unit
- Expected
- Switch in ON position
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No signal with cabled probe radiation
Display shows zero counts when radioactive source present; cabled probe connection not functioning.
Ranked causes
- 1. No connection between probe, cable, and control unit (common)
- 2. Isotope control set to incorrect isotope (common)
- 3. Open circuit in probe cable (occasional)
- 4. Probe is damaged or detector compromised (occasional)
- 5. Circuit inside control unit damaged (rare)
Verify cable and connector integrity
- Test point
- Probe connector, cable entry at control unit
- Expected
- All connections fully seated, no loose pins or bent connectors
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No wireless signal from probe radiation
Display shows zero counts when radioactive source present; no communication between wireless probe and control unit.
Ranked causes
- 1. No wireless connection between probe and control unit (common)
- 2. Isotope control set to incorrect isotope (common)
- 3. Probe LED illuminated but not transmitting (occasional)
- 4. Circuit inside control unit damaged (rare)
Verify probe battery status
- Test point
- Probe battery compartment
- Expected
- Battery present, '+' terminal toward base of probe, battery not depleted
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe LED does not illuminate
Indicator light on wireless pilot probe does not light up when powered on.
Ranked causes
- 1. Probe battery is dead or installed incorrectly (common)
- 2. Battery was not installed (occasional)
- 3. LED on probe is damaged (rare)
Inspect probe battery compartment
- Test point
- Battery holder in probe base
- Expected
- Battery present with '+' terminal facing base, battery not depleted
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incomplete digits in display
Display shows partial or garbled characters; some segments missing or non-functional.
Ranked causes
- 1. Display or display driver is damaged (rare)
Inspect display for physical damage
- Test point
- Display surface and connectors
- Expected
- No visible cracks, discoloration, or loose connectors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spurious counts with cabled probe away from source radiation
Display shows erroneous count readings when cabled probe is not adjacent to radiopharmaceutical source.
Ranked causes
- 1. Radiopharmaceutical residue on probe (common)
Inspect probe for contamination
- Test point
- Probe surface, cap, and cable entry
- Expected
- No visible residue or staining on probe or connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spurious counts with wireless probe radiation
Display shows erroneous count readings when probe is not adjacent to radiopharmaceutical source.
Ranked causes
- 1. Radiopharmaceutical residue on probe (common)
Inspect probe for contamination
- Test point
- Probe surface and cap
- Expected
- No visible residue or staining on probe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spurious high counts with cabled probe radiation
Display shows very high count readings (e.g., 80,000 counts/second) when probe held in air away from source.
Ranked causes
- 1. Intermittent short in cable (occasional)
Inspect cable for damage
- Test point
- Cable jacket, connectors, strain relief
- Expected
- No visible cuts, kinks, or separated connections
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI