DeVilbiss 505: troubleshooting and error codes
The ServiceForge knowledge base holds 23 documented symptoms for the DeVilbiss 505, with 85 ranked causes, 23 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Fluctuating oxygen flow patient safety
Oxygen output varies during operation
Ranked causes
- 1. Occluded humidifier (common)
- 2. Improper humidifier type (occasional)
- 3. Occluded filters (occasional)
- 4. Occluded or defective patient cannula and tubing (occasional)
- 5. Defective flow meter (occasional)
- 6. System leak (occasional)
- 7. Defective compressor reed valve (occasional)
- 8. Defective check valve (occasional)
- 9. Pressure regulator malfunction (occasional)
Check humidifier condition
- Test point
- Humidifier inlet and outlet ports
- Expected
- Clear flow path with no blockage
1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Little or no oxygen flow patient safety
Minimal or zero oxygen delivery at patient outlet
Ranked causes
- 1. Flow meter not adjusted or disconnected (common)
- 2. Oxygen delivery tubing kinked or blocked (common)
- 3. Occluded humidifier (common)
- 4. Excessive tubing length (occasional)
- 5. Defective flow meter (occasional)
- 6. System leak (occasional)
- 7. Defective compressor or reed valve (occasional)
- 8. Defective check valve (occasional)
- 9. Pressure regulator malfunction (occasional)
Verify flow meter adjustment
- Test point
- Flow meter dial
- Expected
- Set to desired flow rate (typically 5 LPM)
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen flow patient safety
Fan off or fan operating with no indicator lights on PC board
Ranked causes
- 1. Line cord not properly installed or defective (common)
- 2. No power at receptacle (common)
- 3. Oxygen concentrator circuit breaker activated (occasional)
- 4. Line cord quick-connect terminal disconnected (occasional)
- 5. Defective power switch (occasional)
- 6. Defective circuit breaker (occasional)
- 7. Blown fuse on PC board (occasional)
- 8. Defective PC board (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen concentration with long cycle time patient safety
Unit cycles slower than normal and oxygen concentration is low
Ranked causes
- 1. System leak (common)
- 2. Defective sieve bed check valve (occasional)
- 3. Defective compressor reed valve (occasional)
- 4. Worn compressor (occasional)
- 5. Occluded filters (occasional)
Check for system leaks
- Test point
- All hoses and fittings
- Expected
- No bubbles when leak detection solution applied
- Tools
- Leak detection solution
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit shuts off during operation
Compressor stops, audible alarm sounds, 'Service Required' light illuminates, cooling fan continues running
Ranked causes
- 1. High pressure condition (common)
- 2. Long cycle time (worn compressor or leak) (common)
- 3. Short cycle time (sieve bed or four-way valve issue) (common)
- 4. PC board failure (occasional)
- 5. Uneven cycle time during operation (occasional)
Measure pressure in sieve beds
- Test point
- Sieve bed test points at top of beds
- Expected
- Left bed ~26 PSI (180 kPa) pressurized, right bed ~2 PSI (15 kPa) depressurized, ±2 PSI
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit warm to touch, cannot restart for several minutes caution
Compressor ran then shut off, fan operating, D17 or D17 & D18 illuminated on PC board
Ranked causes
- 1. Compressor overheated due to occluded filters (common)
- 2. Restricted input or output air passage (common)
- 3. Low or high line voltage (occasional)
- 4. Defective cooling fan (occasional)
- 5. Defective compressor (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visible alarm on, audible alarm on, compressor off, power light off patient safety
Blinking visible alarm, pulsing audible alarm with compressor off and power light off
Ranked causes
- 1. Wiring harness disconnected or defective (occasional)
- 2. Loose compressor wire (occasional)
- 3. Defective capacitor (rare)
- 4. Defective PC board (rare)
- 5. Defective compressor (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound when unit turned on or during power failure patient safety
No alarm tone when unit starts or loses power
Ranked causes
- 1. Unit unused for extended period (common)
- 2. Defective PC board (occasional)
- 3. Defective power switch (occasional)
- 4. Defective wire harness (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pilot pressure
Pilot pressure system below normal operating range
Ranked causes
- 1. Leak in pilot pressure system (common)
- 2. Defective check valve in pilot accumulator (occasional)
- 3. Defective compressor reed valve (occasional)
- 4. Defective compressor (occasional)
Measure pilot pressure
- Test point
- Pilot accumulator or pilot valve test point
- Expected
- 22-28 PSI (150-200 kPa) during normal operation; minimum 19 PSI (130 kPa), ≥19 PSI
- Tools
- Pressure gauge
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maximum cycle time exceeded patient safety
Compressor ran then shut off, fan operating, D17 or D17 & D18 illuminated on PC board
Ranked causes
- 1. Leak in system (common)
- 2. Defective compressor reed valve (occasional)
- 3. Defective compressor (occasional)
- 4. Four-way valve not fully shifted (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor will not start high voltage
Power switch in 'On' position, power indicator light on, cooling fan running, but compressor does not operate
Ranked causes
- 1. Defective capacitor (common)
- 2. Defective compressor (occasional)
- 3. Defective wire harness or PC board (occasional)
Measure compressor supply voltage
- Test point
- Compressor motor terminals
- Expected
- 115 VAC, ±10%
- Tools
- DMM set to AC voltage
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive noise during operation caution
Unit produces abnormally loud or unusual noise levels
Ranked causes
- 1. Loose or defective motor mounts (common)
- 2. Defective compressor (occasional)
- 3. Defective cooling fan (occasional)
Inspect motor mount fasteners
- Test point
- Motor mounting bolts
- Expected
- All bolts tight with no visible looseness
- Tools
- Wrench
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluctuating or no flow patient safety
System pressure below 9 PSI or 62 kPa, fan operating, D16 or D16 & D18 illuminated on PC board
Ranked causes
- 1. Leak in system (common)
- 2. Defective compressor (occasional)
- 3. Four-way valve not fully shifted (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Severe flow restriction patient safety
Fan operating, no indicator lights illuminated on PC board
Ranked causes
- 1. Crimped tubing (common)
- 2. Occluded humidifier (common)
- 3. Flow meter turned off or set too low (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Four-way valve failed to shift; contaminated sieve beds patient safety
Fan operating, D17 & D16 or all three indicator lights illuminated on PC board
Ranked causes
- 1. Four-way valve failed to shift properly (occasional)
- 2. Contaminated sieve beds (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen concentration with normal cycle times patient safety
Unit cycles normally but oxygen output is below specification
Ranked causes
- 1. Oxygen leak from system (common)
- 2. Purge valve malfunction (occasional)
Inspect for oxygen leaks
- Test point
- Sieve bed tops, accumulator, pressure regulator, flow meter, bacteria filter, outlet port
- Expected
- No visible damage, no hissing sound, no bubbles when leak detection solution applied
- Tools
- Leak detection solution, visual inspection
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OSD no lights illuminated patient safety
Oxygen Sensor Display has no illuminated indicator lights
Ranked causes
- 1. Connector off between OSD and PC board (common)
- 2. Defective OSD (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulsating air noise
Rhythmic or pulsing sound from unit
Ranked causes
- 1. Intake filter not in place or defective (common)
- 2. Compressor intake hose disconnected (common)
Verify intake filter installed
- Test point
- Filter housing inlet
- Expected
- Filter element in place and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Purge valve stuck; all diagnostic lights illuminated patient safety
Fan operating, D17 & D16 or all three indicator lights illuminated on PC board
Ranked causes
- 1. Purge valve stuck in activated position (occasional)
- 2. Defective PC board (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service Required light does not illuminate when unit is turned on
Visible alarm indicator on front panel does not light
Ranked causes
- 1. Defective PC board (occasional)
- 2. Defective light panel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen concentration with short cycle time patient safety
Unit cycles faster than normal and oxygen concentration is low
Ranked causes
- 1. Contaminated sieve beds (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OSD both lights illuminated patient safety
Both OSD indicator lights are illuminated
Ranked causes
- 1. Defective OSD (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OSD yellow low oxygen light illuminated patient safety
Yellow Low Oxygen light is illuminated with or without intermittent audible alarm every five seconds
Ranked causes
- 1. Oxygen level is low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI