Del Medical EV200: troubleshooting and error codes
The ServiceForge knowledge base holds 16 documented symptoms for the Del Medical EV200, with 17 ranked causes, 34 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Collision sensor switch adjustment high voltage
Adjust the height of the collision sensor switch in relation to the rail and upper table top support bearing to set the trip point. Preliminary setting obtained using adjustment block part #242-5037P1.
Ranked causes
- 1. Collision sensor switch requires trip point calibration (common)
- 2. Activation lever requires proper tension against table guide rail (common)
Turn power to table OFF before adjustment
- Test point
- Main power switch
- Expected
- Table powered down
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjustable bearing assembly not functioning properly caution
Adjustable tension bearing not riding lightly on lower surface of table top guide rail during initial power-on check.
Ranked causes
- 1. Bearing assembly assembly misalignment or improper adjustment (common)
Inspect bearing assembly operation
- Test point
- Adjustable tension bearing on guide rail
- Expected
- Bearing riding lightly on lower surface of table top guide rail
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 40 inch S.I.D. limit switch high voltage
Adjustment of X-ray source to film plane distance limit switch to ensure proper positioning at 40 inches when switch actuates.
Ranked causes
- 1. Initial limit switch installation requires calibration (common)
Position film plane at correct distance
- Test point
- Film plane of Bucky tray relative to X-ray source
- Expected
- 40 inches from X-ray source when switch is depressed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Adjustable tension bearing assembly height patient safety
Table top guide rail collision sensor switch not held properly closed during normal operation; bearing assembly height requires adjustment
Ranked causes
- 1. Bearing assembly set below lower surface of guide rail (unknown)
Slide table top over each adjustable bearing assembly
- Test point
- adjustable bearing assembly
- Expected
- bearing not set below lower surface of guide rail
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Collision sensor switch operation patient safety
Verify collision sensor switch inhibits DOWN elevator command when table top is in raised position; switch must release when table top is lowered
Ranked causes
- 1. Collision sensor switch not properly adjusted or collision bearing assembly misaligned (unknown)
Table top installed and power turned ON
- Test point
- EV200 power
- Expected
- power ON
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SID switch positioning radiation
Set the 40 inch S.I.D. limit switch so film plane is 40 inches from x-ray source when depressed
Ranked causes
- 1. SID switch requires positioning adjustment (common)
Measure distance from x-ray source to film plane when SID switch depressed
- Test point
- Bucky tray film plane to x-ray source
- Expected
- 40 inches
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collision sensor switch malfunction during DOWN command patient safety
DOWN elevator command not inhibited when table top is in collision position (lifted at corner), or DOWN elevator will not activate when table top is released
Ranked causes
- 1. Collision sensor switch not properly adjusted or bearing assembly not maintaining tension (common)
Verify collision sensor switch inhibits DOWN when table top corner is lifted
- Test point
- collision sensor switch; DOWN foot pedal
- Expected
- DOWN elevator command inhibited (no movement)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collision sensor switches not inhibiting DOWN command patient safety
When lifting table top corner and depressing DOWN foot pedal, switch fails to prevent downward motion, indicating improper adjustment.
Ranked causes
- 1. Collision sensing switch improperly adjusted (common)
Verify collision sensor switch function
- Test point
- Each collision sensing switch
- Expected
- DOWN command inhibited when table top corner is lifted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collision switch does not inhibit downward motion patient safety
Table top continues to move down when pulled up at corner, collision sensor fails to stop elevator
Ranked causes
- 1. Adjustable bearing assembly set below guide rail lower surface (common)
Slide table top over adjustable bearing assembly
- Test point
- Adjustable bearing assembly
- Expected
- Bearing contacts lower surface of guide rail
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient or operator contact with tubestand during height adjustment patient safety
Risk of collision or crush injury during tabletop height changes
Ranked causes
- 1. Inadequate clearance verification before height adjustment (common)
Confirm clearance around table
- Test point
- Patient position, operator position, tubestand proximity, Bucky area
- Expected
- Patient and operator physically clear of all equipment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Risk of pinching or striking during table height adjustment patient safety
Patient or operator reports potential for contact with tubestand equipment or Bucky assembly during tabletop vertical movement
Ranked causes
- 1. Inadequate clearance awareness during height adjustment (common)
Confirm clearance of patient and operator from tubestand and Bucky
- Test point
- Visual inspection around table perimeter, tubestand, and Bucky assembly
- Expected
- No personnel or patient limbs in collision path
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table does not lock without power patient safety
Patient or operator observes that table locking mechanism fails to engage when equipment is unpowered
Ranked causes
- 1. Table power is OFF (common)
Verify table power switch status
- Test point
- Table power control
- Expected
- Power ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table locks inoperative patient safety
Table locking mechanism fails to engage or disengage; patient or operator cannot safely secure table position
Ranked causes
- 1. Table power is OFF (common)
Verify table power status
- Test point
- Main power switch
- Expected
- Switch in ON position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table top collision switch inhibits motion incorrectly patient safety
DOWN elevator cannot activate even when table top is not lifted, collision sensor always blocking
Ranked causes
- 1. Adjustable bearing assembly set too high, over-pressing sensor (occasional)
Center table top and release from any lifted position
- Test point
- Table top and collision sensor
- Expected
- Table top fully seated on guide rail
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table top installation in small rooms caution
Alternate mounting procedure when standard installation clearance is unavailable due to room size constraints.
Ranked causes
- 1. Insufficient room clearance for standard table top installation (occasional)
Remove bumper stops and guide components
- Test point
- Bumper stops, adjustable tension bearing assemblies, plastic guide rollers
- Expected
- Components removed from front rail of inner frame
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify collision sensor operation patient safety
Test that collision sensor switch interrupts DOWN command when table encounters resistance
Ranked causes
- 1. Collision sensor needs functional verification (common)
Activate elevator in DOWN direction
- Test point
- Elevator motor control
- Expected
- Table descends smoothly
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI