Del Medical EV200: troubleshooting and error codes

The ServiceForge knowledge base holds 16 documented symptoms for the Del Medical EV200, with 17 ranked causes, 34 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Collision sensor switch adjustment high voltage

Adjust the height of the collision sensor switch in relation to the rail and upper table top support bearing to set the trip point. Preliminary setting obtained using adjustment block part #242-5037P1.

Ranked causes

  1. 1. Collision sensor switch requires trip point calibration (common)
  2. 2. Activation lever requires proper tension against table guide rail (common)
First check

Turn power to table OFF before adjustment

Test point
Main power switch
Expected
Table powered down

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Adjustable bearing assembly not functioning properly caution

Adjustable tension bearing not riding lightly on lower surface of table top guide rail during initial power-on check.

Ranked causes

  1. 1. Bearing assembly assembly misalignment or improper adjustment (common)
First check

Inspect bearing assembly operation

Test point
Adjustable tension bearing on guide rail
Expected
Bearing riding lightly on lower surface of table top guide rail

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 40 inch S.I.D. limit switch high voltage

Adjustment of X-ray source to film plane distance limit switch to ensure proper positioning at 40 inches when switch actuates.

Ranked causes

  1. 1. Initial limit switch installation requires calibration (common)
First check

Position film plane at correct distance

Test point
Film plane of Bucky tray relative to X-ray source
Expected
40 inches from X-ray source when switch is depressed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Adjustable tension bearing assembly height patient safety

Table top guide rail collision sensor switch not held properly closed during normal operation; bearing assembly height requires adjustment

Ranked causes

  1. 1. Bearing assembly set below lower surface of guide rail (unknown)
First check

Slide table top over each adjustable bearing assembly

Test point
adjustable bearing assembly
Expected
bearing not set below lower surface of guide rail

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Collision sensor switch operation patient safety

Verify collision sensor switch inhibits DOWN elevator command when table top is in raised position; switch must release when table top is lowered

Ranked causes

  1. 1. Collision sensor switch not properly adjusted or collision bearing assembly misaligned (unknown)
First check

Table top installed and power turned ON

Test point
EV200 power
Expected
power ON

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SID switch positioning radiation

Set the 40 inch S.I.D. limit switch so film plane is 40 inches from x-ray source when depressed

Ranked causes

  1. 1. SID switch requires positioning adjustment (common)
First check

Measure distance from x-ray source to film plane when SID switch depressed

Test point
Bucky tray film plane to x-ray source
Expected
40 inches

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collision sensor switch malfunction during DOWN command patient safety

DOWN elevator command not inhibited when table top is in collision position (lifted at corner), or DOWN elevator will not activate when table top is released

Ranked causes

  1. 1. Collision sensor switch not properly adjusted or bearing assembly not maintaining tension (common)
First check

Verify collision sensor switch inhibits DOWN when table top corner is lifted

Test point
collision sensor switch; DOWN foot pedal
Expected
DOWN elevator command inhibited (no movement)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collision sensor switches not inhibiting DOWN command patient safety

When lifting table top corner and depressing DOWN foot pedal, switch fails to prevent downward motion, indicating improper adjustment.

Ranked causes

  1. 1. Collision sensing switch improperly adjusted (common)
First check

Verify collision sensor switch function

Test point
Each collision sensing switch
Expected
DOWN command inhibited when table top corner is lifted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collision switch does not inhibit downward motion patient safety

Table top continues to move down when pulled up at corner, collision sensor fails to stop elevator

Ranked causes

  1. 1. Adjustable bearing assembly set below guide rail lower surface (common)
First check

Slide table top over adjustable bearing assembly

Test point
Adjustable bearing assembly
Expected
Bearing contacts lower surface of guide rail

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient or operator contact with tubestand during height adjustment patient safety

Risk of collision or crush injury during tabletop height changes

Ranked causes

  1. 1. Inadequate clearance verification before height adjustment (common)
First check

Confirm clearance around table

Test point
Patient position, operator position, tubestand proximity, Bucky area
Expected
Patient and operator physically clear of all equipment

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Risk of pinching or striking during table height adjustment patient safety

Patient or operator reports potential for contact with tubestand equipment or Bucky assembly during tabletop vertical movement

Ranked causes

  1. 1. Inadequate clearance awareness during height adjustment (common)
First check

Confirm clearance of patient and operator from tubestand and Bucky

Test point
Visual inspection around table perimeter, tubestand, and Bucky assembly
Expected
No personnel or patient limbs in collision path

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table does not lock without power patient safety

Patient or operator observes that table locking mechanism fails to engage when equipment is unpowered

Ranked causes

  1. 1. Table power is OFF (common)
First check

Verify table power switch status

Test point
Table power control
Expected
Power ON

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table locks inoperative patient safety

Table locking mechanism fails to engage or disengage; patient or operator cannot safely secure table position

Ranked causes

  1. 1. Table power is OFF (common)
First check

Verify table power status

Test point
Main power switch
Expected
Switch in ON position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table top collision switch inhibits motion incorrectly patient safety

DOWN elevator cannot activate even when table top is not lifted, collision sensor always blocking

Ranked causes

  1. 1. Adjustable bearing assembly set too high, over-pressing sensor (occasional)
First check

Center table top and release from any lifted position

Test point
Table top and collision sensor
Expected
Table top fully seated on guide rail

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table top installation in small rooms caution

Alternate mounting procedure when standard installation clearance is unavailable due to room size constraints.

Ranked causes

  1. 1. Insufficient room clearance for standard table top installation (occasional)
First check

Remove bumper stops and guide components

Test point
Bumper stops, adjustable tension bearing assemblies, plastic guide rollers
Expected
Components removed from front rail of inner frame

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verify collision sensor operation patient safety

Test that collision sensor switch interrupts DOWN command when table encounters resistance

Ranked causes

  1. 1. Collision sensor needs functional verification (common)
First check

Activate elevator in DOWN direction

Test point
Elevator motor control
Expected
Table descends smoothly

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI