X-ray generators · CM Series family

Del Medical CM: troubleshooting and error codes

The ServiceForge knowledge base holds 21 documented symptoms for the Del Medical CM, with 33 ranked causes, 17 diagnostic checks with test points and expected values, and 77 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: AEC 3D system setup and calibration radiation

Prepare and calibrate automatic exposure control (AEC) board assemblies for CMP 200 X-ray generators across multiple film/screen combinations and kV ranges

Ranked causes

  1. 1. Initial AEC system configuration for multiple film/screen combinations (common)
  2. 2. Configure film/screen response compensation across kV breakpoints (common)
  3. 3. Establish AEC calibration range within 0-10V ramp limits (common)
First check

Verify generator compatibility with specified AEC chambers

Test point
Compatibility statement in Chapter 1
Expected
Factory configured AEC chambers match installed system

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fault in kV or mA output during exposure high voltage

E008 KV/MA FAULT displayed; exposure terminated due to arcing or HT tank failure

Ranked causes

  1. 1. Arcing in X-ray tube (common)
  2. 2. Arcing of HV cables (occasional)
  3. 3. HT tank failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament current too low

E007 FILAMENT FAULT displayed; filament current detected below 2 amps

Ranked causes

  1. 1. Open filament in X-ray tube (common)
  2. 2. Poor connections in cathode cable (occasional)
  3. 3. Open fuses on filament board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC back-up timer exceeded

E014 AEC BUT ERROR; AEC exposure exceeded allowed backup time

Ranked causes

  1. 1. AEC technique settings incorrect (common)
  2. 2. Wrong AEC chamber energized (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC mAs limit exceeded

E015 AEC BU MAS ERR; AEC exposure exceeded allowed backup mAs

Ranked causes

  1. 1. Exposure technique settings incorrect (common)
  2. 2. Wrong AEC chamber selected (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exposure manually terminated

E013 MANUAL TERMIN displayed; operator released exposure switch during active exposure

Ranked causes

  1. 1. Operator action during exposure (unknown)
  2. 2. Faulty switch contacts or wiring (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Precalibration system checks before AEC calibration radiation

Verify AEC chamber installation, board assembly configuration, and system prerequisites before calibration procedures

Ranked causes

  1. 1. Verify AEC chamber is properly installed and compatible with board type (common)
  2. 2. Verify film processing and cassette consistency (common)
First check

Verify AEC chamber installation per factory compatibility requirements

Test point
AEC chamber assembly and electrical connections
Expected
Chamber type (ion or solid state) matches generator configuration; secure connections

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotor fails to start

E006 ROTOR FAULT displayed; rotor starter detects stator current fault or rotor not ready

Ranked causes

  1. 1. Rotor starter detected stator current fault (occasional)
  2. 2. Generator was not ready to start rotor (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tomography back-up timer exceeded

E016 TOMO BUT ERROR; Tomo exposure exceeded backup time limit

Ranked causes

  1. 1. Tomo technique settings incorrect (common)
  2. 2. Backup time limit too short (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anode heat limit exceeded

E019 ANODE HEAT LIMIT; Selected parameters cause tube to exceed programmed heat limit

Ranked causes

  1. 1. Insufficient anode cooling or excessive parameters (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC board gain adjustment radiation

Configure AEC gain potentiometers (R1-R4 or R11-R14 depending on board assembly) for each of 4 channels (table bucky, wall bucky, spot film, auxiliary)

Ranked causes

  1. 1. AEC board requires gain calibration per ion chamber manufacturer specifications (common)
First check

Verify central ray centered relative to image receptor

Test point
Image receptor center
Expected
Central ray aligned with image receptor center

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC high voltage bias supply output adjustment high voltage

Adjust R79 potentiometer to set +300/+500 VDC and +45 VDC outputs for ion chambers requiring separate high voltage bias supply

Ranked causes

  1. 1. High voltage bias supply output requires calibration per ion chamber manufacturer specifications (occasional)
First check

Verify board version includes high voltage bias supply

Test point
AEC board assembly inspection
Expected
Board fitted with R79 and high voltage output connectors

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC short exposure time compensation adjustment radiation

Configure short exposure time compensation potentiometers (R11-R14 or equivalent) for each of 4 AEC channels

Ranked causes

  1. 1. AEC board requires short exposure time compensation adjustment (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Film density verification at baseline kV radiation

Establish baseline film density at knee kV (75 kV) and verify calibration range accommodates all film/screen combinations

Ranked causes

  1. 1. Ensure film density remains constant across multiple film/screen speeds (common)
First check

Measure film optical density at baseline 75 kV with medium speed film/screen

Test point
Processed radiographic film
Expected
Nominal optical density per setup worksheet (e.g., 1.1-1.4)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Erratic console faults in electrically noisy environments

Random fault messages, RAM data errors, intermittent loss of communication displayed on console

Ranked causes

  1. 1. Inadequate console grounding in electrically noisy environment (common)
First check

Verify console ground connection exists and is adequate

Test point
Ground stud on rear of console, main input fuse block
Expected
Ground wire #14 AWG (2.3 mm2) or larger connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator in prep state too long

E018 PREP TIMEOUT; Generator has been in prep state exceeding time limit

Ranked causes

  1. 1. Extended prep time (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator not ready to make exposure

System displays E009 PS NOT READY message, preventing exposure

Ranked causes

  1. 1. Power supply not ready (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

mA during exposure too high

E011 HIGH MA FAULT displayed; CPU detected mA exceeding allowed tolerance

Ranked causes

  1. 1. mA out of tolerance (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

mA during exposure too low

E012 LOW MA FAULT displayed; CPU detected mA below allowed tolerance

Ranked causes

  1. 1. mA out of tolerance (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Selected mA not calibrated

E017 NOT CALIBRATED; Selected mA not calibrated for selected kV

Ranked causes

  1. 1. Missing calibration for selected mA/kV combination (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray tube thermal switch opened caution

E020 THERMAL INT #1; X-ray tube #1 exceeded temperature limit and thermal switch opened

Ranked causes

  1. 1. X-ray tube #1 overheated (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
AEC DEVICE ERRORAEC system fault with no feedback signal or ramp voltage failureRamp voltage fails to reach 4% of expected value when exposure time reaches 20% of selected backup time, or improper technique selected, or AEC fault occurs causing no feedback signalerror
APR MEMORY ERRORAPR data has been corruptedStartup or during operationerror
E001Generator CPU EPROM checksum errorAt startupcritical
E003Generator CPU NVRAM errorAt startuperror
E004Generator CPU Real Time Clock errorAt startupwarning
E005Main Contactor ErrorDuring startup or power-oncritical
E006Rotor FaultDuring rotor startuperror
E007Filament FaultDuring operation or tube checkerror
E008kV / mA FaultDuring exposurecritical
E009Power Supply Not ReadyWhen attempting exposureerror
E011mA During Exposure Too HighDuring exposureerror
E012mA During Exposure Too LowDuring exposureerror
E013Manually Terminated ExposureDuring exposureinfo
E014AEC Back-up Timer - Exposure TerminatedDuring AEC exposureerror
E015AEC mAs Exceeded - Exposure TerminatedDuring AEC exposureerror
E016Tomo Back-up Timer - Exposure TerminatedDuring tomography exposureerror
E017Uncalibrated Exposure ParameterWhen selecting exposure parameterserror
E018Preparation TimeoutAfter extended prep timewarning
E019Anode Heat LimitWhen anode temperature limit exceededwarning
E020Thermal Switch Interlock #1 ErrorWhen X-ray tube #1 overheatserror
E022Door interlock openDoor is openwarning
E028Prep input active during initializationPrep switch or input short circuit during power onwarning
E029X-ray input active during initializationX-ray switch or input short circuit during power onwarning
E032Console communication errorGenerator detects error in communication to consoleerror
E033Lithium battery voltage lowGenerator detects lithium battery voltage is lowwarning
E034+12VDC rail out of tolerance+12VDC rail is out of toleranceerror
E035-12VDC rail out of tolerance-12VDC rail is out of toleranceerror
E038Calibration data corruptGenerator detects corrupt calibration dataerror
E039AEC data corruptGenerator detects corrupt AEC dataerror
E041Receptor data corruptGenerator detects corrupt receptor dataerror
E042Tube data corruptGenerator detects corrupt tube dataerror
E043High voltage detected in non X-ray statekV detected when X-ray switch is not activecritical
E044Invalid communication messageReceived communication message not validinfo
E045Communication message not supportedReceived message is valid but not supported by systeminfo
E046Mode inhibitedReceived message valid but not allowed in present statewarning
E049Function not enabledRequested function not programmed to be enabledwarning
E050Generator limit data corruptGenerator detects corrupt generator limit dataerror
E051AEC feedback errorNo or insufficient feedback signal detected from AEC deviceerror
E052High small focus filament current in standbySmall focus filament current greater than limits in standby modewarning
E053High large focus filament current in standbyLarge focus filament current greater than limits in standby modewarning
E054AEC reference out of rangeAEC reference reached maximum or minimum limitwarning
E055No AEC fields selectedAEC enabled but no fields are selectedwarning
E056Receptor has no X-ray tube programmedAll Receptors have no X-ray tube programmedwarning
E057AEC stop signal in wrong stateAEC stop signal (P.T. stop) active low indicating exposure finished during preperror
E058Console backup timer exceededConsole detects exposure exceeded backup time and terminated exposurecritical
E060High kV errorkV exceeds high kV tolerance levelerror
E061Low kV errorkV exceeds low kV tolerance levelerror
E063Factory defaults mode setJW2 on generator control board set to LOAD DEFAULTS positioninfo
E065Tomo device errorTomo start signal not received within 30 seconds of X-ray switch presserror
E067Power supply duty cycle limit exceededPower supply duty cycle limit exceededcritical
E071DAP dose overflowAccumulated DAP value exceeds display limitinfo
E072DAP device errorDAP device is not functionalerror
E073DAP configuration data corruptDAP configuration data is corruptederror
E074Inverter 1 power circuit failurePower circuit failure in inverter 1critical
E075Inverter 2 power circuit failurePower circuit failure in inverter 2critical
E077HT primary overcurrent detectedResonant circuit detects HT primary overcurrentcritical
E078Bucky 1 interlock errorBucky 1 is not readywarning
E079Bucky 2 interlock errorBucky 2 is not readywarning
E080Interlock 1 openInterlock 1 is openwarning
E081Interlock 2 openInterlock 2 is openwarning
E082kV overvoltageOutput kV exceeded maximum allowed kV limitcritical
E083Anode current exceeded maximum limitAnode current exceeded maximum allowed limitcritical
E084Cathode current exceeded maximum limitCathode current exceeded maximum allowed limitcritical
E085Insufficient current in stator main windingInsufficient current sensed in rotor stator main windingerror
E086Insufficient current in stator shift windingInsufficient current sensed in rotor stator shift windingerror
E100Maximum mA exceeded during calibrationMaximum mA has been exceeded during auto calibrationerror
E101Calibration data table exceededAuto calibration exceeded data table length due to excessive exposureserror
E102Maximum filament current exceededMaximum filament current for selected focus has been reachederror
E103Calibration manually terminatedOperator released exposure button during auto calibrationinfo
E104No mA feedback during calibrationNo mA feedback detected during auto calibrationcritical
E105Minimum mA not calibratedMinimum generator mA was exceeded at start of calibrationerror
E240Printer out of labelsPrinter is out of labelsinfo
E241Labels jammed in printerLabels are jammedwarning
E242Printer self-diagnostics detected problemPrinter self-diagnostics have reported a problemwarning
E243Printer communication errorPrinter has reported a communication errorwarning
E244Paper platen not positioned properlyThe paper platen is not positioned properlywarning
E245Printer is off-linePrinter is off-lineinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI