Datex-Ohmeda Tec 6 Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Datex-Ohmeda Tec 6 Plus, with 21 ranked causes, 23 diagnostic checks with test points and expected values, and 10 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No output alarm patient safety
Vaporizer has shut down and is no longer delivering vapor. Amber light illuminates.
Ranked causes
- 1. Agent level below 20 ml (common)
- 2. Vaporizer tilted for more than 10 seconds (occasional)
- 3. Power failure longer than 10 seconds (occasional)
- 4. Internal malfunction detected (rare)
Verify agent level
- Test point
- Agent reservoir
- Expected
- Above 20 ml
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm and display test incomplete or non-functional patient safety
During power-on self-test, one or more of the five indicator lights, LCD agent level bars, or the auditory alarm does not flash or sound for approximately four seconds
Ranked causes
- 1. Defective indicator light or LED (unknown)
- 2. Defective LCD agent level indicator bars (unknown)
- 3. Defective auditory alarm (unknown)
Verify all five indicator lights flash
- Test point
- Front display panel lights (DESFLURANE, NO OUTPUT, LOW AGENT, ALARM, BATTERY LOW, WARM-UP, OPERATIONAL)
- Expected
- All five lights flash simultaneously for approximately four seconds
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low agent level in sump patient safety
LOW AGENT alarm occurs. Agent level in sump reservoir below 60 ml. If below 20 ml, NO OUTPUT alarm activated.
Ranked causes
- 1. Agent level below 60 ml in sump (common)
- 2. Critical agent depletion below 20 ml (occasional)
Observe agent level in sump reservoir
- Test point
- sump viewing window
- Expected
- above 60 ml minimum; ideally full
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power failure or internal malfunction patient safety
Vaporizer goes to red alarm or does not reach operational status. Power failure detected or internal system malfunction.
Ranked causes
- 1. External power loss (common)
- 2. Internal power failure or malfunction (rare)
Check for power supply status and alarm tone
- Test point
- power indicator and audio alarm
- Expected
- normal operational tone (not higher pitched failure tone)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vaporizer dial cannot turn despite clicking mechanism patient safety
Dial release depresses but dial will not rotate. Clicking heard on mechanism when pressure applied. Dial remains locked.
Ranked causes
- 1. Vaporizer in warm-up mode, interlock prevents use (common)
- 2. Another vaporizer interlock mechanism is active (occasional)
Verify green OPERATIONAL light status
- Test point
- front panel indicator light
- Expected
- green OPERATIONAL light illuminated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm battery low
ALARM BATTERY LOW indicator activates; green OPERATIONAL light remains illuminated during normal operation
Ranked causes
- 1. Alarm battery voltage low (common)
Verify alarm battery voltage is low
- Test point
- Alarm battery
- Expected
- Low voltage condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low alarm indication
BATTERY LOW alarm appears on display. Battery voltage insufficient for backup power.
Ranked causes
- 1. Battery voltage depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tec 6 Plus vaporizer output concentrations patient safety
Verify vaporizer delivers correct anesthetic agent concentrations across calibrated range
Ranked causes
- 1. Factory calibration verification (common)
Verify calibration conditions
- Test point
- Atmospheric pressure
- Expected
- 760 torr
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical supply verification high voltage
Confirm AC power supply meets vaporizer requirements
Ranked causes
- 1. Incorrect AC supply voltage or frequency (rare)
Measure AC supply voltage
- Test point
- Facility wall outlet or disconnect
- Expected
- 100–120V AC or 220–240V AC, 100–120 or 220–240
- Tools
- Multimeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LOW AGENT light remains illuminated after test
The LOW AGENT indicator light remains illuminated after the alarm and display test is completed
Ranked causes
- 1. Agent (desflurane) level low or empty (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NO OUTPUT light flashes during operation patient safety
Vaporizer tilted for 10 seconds while dial is on in operational mode; flashing NO OUTPUT light and auditory alarm indicate vapor delivery stopped
Ranked causes
- 1. Vaporizer tilted during operation (common)
Verify vaporizer position
- Test point
- Vaporizer body
- Expected
- Level, not tilted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operational light does not illuminate within 10 minutes
After warm-up period of up to 10 minutes, the amber WARM-UP light does not extinguish and the green OPERATIONAL light does not illuminate
Ranked causes
- 1. Warm-up cycle failure (unknown)
Verify sufficient agent in sump
- Test point
- Sump level
- Expected
- Sufficient agent present
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Recalibration authorization requirement patient safety
Technician must be aware that only authorized service centers perform recalibration
Ranked causes
- 1. Recalibration required after service or suspected drift (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ALARM BATTERY LOW | Alarm battery voltage is insufficient | Alarm battery voltage drops below threshold | warning |
BATTERY LOW | Backup battery voltage insufficient | Battery voltage drops below acceptable threshold during operation | warning |
Internal malfunction | Internal system fault detected, vaporizer not safe for use | WARM-UP light remains illuminated beyond 10 minutes after power restoration | critical |
LOW AGENT | Anesthetic agent level below 60 ml threshold | Sump reservoir level drops below 60 ml during operation | warning |
NO OUTPUT | Anesthetic agent delivery halted, sump nearly empty | Agent level drops below 20 ml in sump reservoir | critical |
NO OUTPUT | Vapor delivery stopped | Vaporizer tilted for 10 seconds while dial is on in operational mode | error |
NO OUTPUT | Vaporizer has shut down and is no longer delivering vapor | Dial in ON position with agent below 20 ml, vaporizer tilted >10 seconds, power failure >10 seconds, or internal malfunction detected | critical |
OPERATIONAL | Vaporizer ready for use | Warm-up complete after power application | info |
Power failure alarm | External AC power loss or restoration detected | Power supply interruption or recovery | error |
WARM-UP | Vaporizer in warm-up mode, not yet operational | Power applied to vaporizer or system reset | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.