Datex-Ohmeda Panda: troubleshooting and error codes
The ServiceForge knowledge base holds 318 documented symptoms for the Datex-Ohmeda Panda, with 622 ranked causes, 562 diagnostic checks with test points and expected values, and 110 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Procedure light does not illuminate
Procedure light assembly does not produce light when switched on
Ranked causes
- 1. Burned-out or defective halogen bulb (common)
- 2. 12V DC power source or control circuitry failure (occasional)
- 3. No mains power to 12V DC power supply (occasional)
- 4. AC short causing blown fuse on Power PCA (occasional)
- 5. AC short on Power PCA (occasional)
- 6. AC short in wiring to 12V DC power supply (rare)
- 7. AC short in 12V DC power supply (rare)
- 8. DC power supply output failure (occasional)
- 9. Disconnected or faulty harness (occasional)
- 10. Faulty Procedure Light harness (occasional)
- 11. Faulty Alarm/Light PCA ribbon cable or defective PCA (rare)
Test halogen bulb for continuity
- Test point
- Halogen bulb
- Expected
- <1 Ohm
- Tools
- Ohm Meter
13 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Procedure light does not illuminate
No light output from procedure light assembly when switched on
Ranked causes
- 2. No 12V DC power to bulb socket (common)
- 4. Blown AC fuse on Power PCA (occasional)
- 5. AC short on Power PCA (rare)
- 6. AC short in wiring to 12V DC Power Supply (occasional)
- 7. AC short in 12V DC Power Supply (rare)
- 8. PF choke failure (rare)
- 9. Defective or disconnected 12V DC Power Supply (occasional)
- 10. Open in harness between power supply and bulb (occasional)
- 11. Faulty procedure light assembly harness (occasional)
- 12. No 12V DC to procedure light assembly connector (occasional)
- 13. Faulty Alarm/Light PCA ribbon cable (rare)
Test halogen bulb continuity
- Test point
- Halogen bulb terminals
- Expected
- <1 Ohm
- Tools
- Ohm Meter
12 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 reading inaccurate or absent patient safety
Pulse oximetry measurements affected by patient conditions, environmental factors, or sensor application errors
Ranked causes
- 1. Excessive patient movement (common)
- 2. Sensor applied too tightly (common)
- 3. Ambient light interference (common)
- 4. Venous pulsations (occasional)
- 5. Intravascular dyes or external coloring (occasional)
- 6. Sensor on extremity with blood pressure cuff or catheter (occasional)
- 7. Poor peripheral perfusion (occasional)
- 8. Arterial occlusion proximal to sensor (rare)
- 9. Dysfunctional hemoglobins (carboxyhemoglobin, methemoglobin, sulphemoglobin) (rare)
- 10. Anemia with hemoglobin below 5 gm/dl (rare)
Verify sensor is properly and securely applied
- Test point
- Sensor site on patient extremity
- Expected
- Sensor firmly positioned, not too tight, in contact with skin
3 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate pulse oximetry readings patient safety
SpO2 measurements do not reflect actual patient oxygenation; readings inconsistent with clinical assessment
Ranked causes
- 1. Incorrect sensor application or use (common)
- 2. Significant dysfunctional hemoglobin levels (occasional)
- 3. Intravascular dyes present (rare)
- 4. Excessive ambient light exposure (occasional)
- 5. Excessive patient movement (common)
- 6. Venous pulsations detected (occasional)
- 7. Blood pressure cuff or catheter on same extremity (common)
- 8. Non-specified accessories or cables used (occasional)
Verify sensor application per directions for use
- Test point
- Sensor site
- Expected
- Sensor applied correctly per manufacturer instructions
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base no motion caution
Elevating base (E-Base) does not move up or down
Ranked causes
- 1. Defective elevating base unit (occasional)
- 2. No power to elevating base (occasional)
- 3. Blown fuse on power distribution (occasional)
- 4. Disconnected E-Base harness (occasional)
- 5. Defective E-Base harness (occasional)
- 6. Defective Power PCA (occasional)
- 7. Defective Control PCA (occasional)
Test E-Base movement and power indicator
- Expected
- Green light ON; E-Base responds to foot switch pedals
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate SpO2 measurements patient safety
Pulse oximetry readings appear unreasonable or do not correlate with patient vital signs
Ranked causes
- 1. Incorrect sensor application or placement (common)
- 2. Dysfunctional hemoglobin present (carboxyhemoglobin or methemoglobin) (occasional)
- 3. Intravascular dyes present (occasional)
- 4. Excessive illumination interfering with sensor (common)
- 5. Excessive patient movement or artifact (occasional)
- 6. Venous pulsations present (rare)
- 7. Blood pressure cuff or vascular access on same extremity (common)
Cross-check SpO2 reading with alternate measurement method and patient vital signs
- Test point
- Patient examination and alternate pulse oximetry or blood gas analysis
- Expected
- Pulse oximeter reading correlates reasonably with patient condition and alternate measurements
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No light from procedure light (halogen)
Procedure light halogen bulb does not illuminate when switch is turned on; heater may be powered on normally
Ranked causes
- 1. Burned-out or defective halogen bulb (common)
- 2. No mains AC power to 12V DC power supply or supply defect (occasional)
- 3. No mains AC power reaching power PCA J28 connector (occasional)
- 4. AC short causing blown fuse in power distribution (occasional)
- 5. AC short circuit on power PCA PCB (rare)
- 6. AC short in wiring between power PCA and 12V DC supply (rare)
- 7. AC short or failure in 12V DC power supply unit (rare)
Test halogen bulb continuity
- Test point
- Halogen bulb
- Expected
- <1 Ohm (continuous)
- Tools
- Ohmmeter
8 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No motion from elevating base caution
Elevating base does not respond to footswitch controls or shows no movement
Ranked causes
- 1. Defective elevating base unit (common)
- 2. No power to E-Base (common)
- 3. Blown fuse (occasional)
- 4. Disconnected AC harness (occasional)
- 5. Defective E-Base harness (occasional)
- 6. Defective Power PCA (occasional)
- 7. Defective Control PCA (occasional)
18 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 probe not detected patient safety
System fails to recognize SpO2 probe during startup or operation; probe icon does not appear on screen
Ranked causes
- 1. Dislodged SpO2 connector from probe plate (occasional)
- 2. Poor probe connector seating at probe panel (occasional)
- 3. Reverse probe connection (occasional)
- 4. Defective SpO2 flex circuit connection (occasional)
- 5. Defective SpO2 probe (occasional)
- 6. No communication between SpO2 I/F PCA and Control PCA (common)
- 7. Defective SpO2 I/F PCA or Control PCA (occasional)
Verify probe connector seating at probe panel
- Expected
- Connector fully seated at jack plate
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both bulbs do not light (Halogen) high voltage
Neither of the two halogen observation bulbs illuminates
Ranked causes
- 1. Loose connector and wiring (common)
- 2. No mains to transformer (common)
- 3. Bad dimmer pot switch (occasional)
- 4. Bad fuse (occasional)
- 5. Bad relay on power PCA (occasional)
- 6. Bad transformer (rare)
Check lamp wiring
- Test point
- Harness at J26
- Expected
- Continuity and no shorts
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both observation light bulbs do not illuminate
Neither of the two observation light bulbs produce light output.
Ranked causes
- 1. Bad lamp wiring or connection (common)
- 2. No mains voltage to transformer (common)
- 3. Bad dimmer potentiometer switch (occasional)
- 4. Bad fuse F11 or F12 on power PCA (occasional)
- 5. Bad relay on power PCA (rare)
- 6. Bad transformer assembly (occasional)
Test mains voltage at transformer input
- Test point
- J24 pins 3 and 4 on power PCA
- Expected
- Mains voltage present with dimmer ON
- Tools
- AC Volt Meter
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hands Free Alarm Silence does not work caution
Hands free silence feature cannot silence alarm when hand waved in front of panel
Ranked causes
- 1. Too much ambient sunlight (occasional)
- 2. Alarm/Light PCA not programmed or defective (occasional)
- 3. Hands free silence disabled in software (occasional)
- 4. Open ribbon cable connection (occasional)
- 5. Defective emitter/receiver circuit (rare)
- 6. Defective Control PCA (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of pulse signal patient safety
Pulse waveform no longer detected by pulse oximeter despite sensor connection
Ranked causes
- 1. Sensor applied too tightly (common)
- 2. Excessive illumination at sensor site (common)
- 3. Blood pressure cuff inflated on sensor extremity (common)
- 4. Patient hypotension, severe vasoconstriction, severe anemia, or hypothermia (occasional)
- 5. Arterial occlusion proximal to sensor (rare)
- 6. Patient in cardiac arrest or shock (rare)
Assess sensor tightness on extremity
- Test point
- Sensor application site
- Expected
- Sensor snug but not tight; skin blanching minimal if any
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of up/down control from one footswitch caution
One footswitch pedal does not raise or lower base; other pedal may work normally
Ranked causes
- 2. Blown fuse in power circuit (occasional)
- 3. Disconnected AC harness to Power PCA (occasional)
- 3. Low AC voltage to E-Base harness (occasional)
- 4. Defective E-Base harness (occasional)
- 5. Defective Power PCA (rare)
- 6. Defective Control PCA (rare)
Test fuses F1, F4 on Power PCA for continuity
- Test point
- Fuses F1, F4 on Power PCA after removing observation light transformer bracket
- Expected
- Fuse intact (ohm meter shows continuity); not blown
- Tools
- 5.5 mm Hex Wrench, Needle nose pliers, Ohm Meter
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 signal or zero saturation displayed patient safety
Module reports zero saturation with loss-of-pulse condition and audible alarm; signal search icon appears.
Ranked causes
- 1. Sensor too tight on patient extremity (common)
- 2. Excessive ambient illumination interfering with measurement (common)
- 3. Blood pressure cuff inflated on same extremity as sensor (occasional)
- 4. Patient hypotension, vasoconstriction, anemia, or hypothermia (occasional)
- 5. Arterial occlusion proximal to sensor site (rare)
- 6. Patient in cardiac arrest or shock (rare)
Verify signal strength indicator
- Test point
- Signal strength indicator on display
- Expected
- Strong signal
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One bulb does not light (LED) caution
Only one of two observation LEDs illuminates when switched on
Ranked causes
- 1. Defective bulb (common)
- 2. Loose contact at LED socket (common)
- 3. Loose contact of mating connector at Power PCA (occasional)
- 4. Faulty lamp wiring (occasional)
- 5. Bad socket (occasional)
- 6. Bad Power PCA (rare)
Reseat LED in socket
- Test point
- LED socket
- Expected
- LED glows when knob rotated
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm message: Confirm Probe Jack Connection patient safety
Display prompts user to confirm or check probe jack connection
Ranked causes
- 1. Defective or improperly seated temperature probe (common)
- 2. Cold probe (less than 24°C) (common)
- 3. Defective rail harness (occasional)
- 4. Defective Probe Panel Interface Board (occasional)
- 5. Defective Control PCA (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both observation bulbs do not light
Both observation light bulbs fail to illuminate
Ranked causes
- 1. No mains power to transformer (common)
- 2. Bad dimmer pot switch (occasional)
- 3. Bad fuse on Power PCA (occasional)
- 4. Bad relay on Power PCA (occasional)
- 5. Bad transformer (occasional)
Check lamp wiring continuity
- Test point
- Harness at J26
- Expected
- Continuity at pins 1-3 and 2-4
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both observation bulbs do not light
Neither bulb in observation light assembly illuminates
Ranked causes
- 1. No AC mains to observation light transformer (common)
- 2. Bad dimmer pot switch (occasional)
- 3. Bad fuse on Power PCA (occasional)
- 4. Bad relay on Power PCA (occasional)
- 5. Bad observation light transformer (occasional)
Check lamp wiring continuity
- Test point
- Harness at J26: pins 1-3 and 2-4 with bulbs installed
- Expected
- Continuity present
- Tools
- Ohm Meter
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Confirm Probe Jack Connection alarm patient safety
Probe temperature below 24C or probe connection not properly detected
Ranked causes
- 1. Defective probe (common)
- 2. Cold probe below 24C (common)
- 3. Damaged rail harness (occasional)
- 4. Defective Probe Panel Interface Board (occasional)
- 5. Defective Control PCA (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of up/down control from one foot switch caution
One foot switch pedal does not move the elevating base up or down
Ranked causes
- 2. Disconnected harness at Power PCA J28 (occasional)
- 3. Defective E-Base harness (occasional)
- 4. Defective Power PCA (occasional)
- 5. Defective Control PCA (occasional)
- 6. Blown fuse F1 or F4 (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No light from procedure light (LED)
Procedure light LED does not illuminate when switch is turned on; heater may be powered on normally
Ranked causes
- 1. Burned-out or defective LED bulb (common)
- 2. 12V DC power supply or control circuitry failure (occasional)
- 3. Loose LED socket contact (occasional)
- 4. Loose mating connector at Power PCA J9 (occasional)
- 5. Faulty procedure light harness (occasional)
Measure LED bulb impedance
- Test point
- LED bulb
- Expected
- >10 Kilo-Ohm for good bulb
- Tools
- Multimeter
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One bulb does not light (Halogen) high voltage
Only one of two halogen observation bulbs illuminates
Ranked causes
- 1. Bad bulb (common)
- 2. Bad harness or bad socket (occasional)
- 3. Short in lamp wiring (occasional)
- 4. Bad socket (rare)
- 5. Bad harness connection (rare)
Test bulb continuity
- Test point
- Bulb terminals
- Expected
- < 1 Ohm
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One observation bulb does not light
Single bulb in observation light assembly does not illuminate while the other may work
Ranked causes
- 1. Bad bulb (common)
- 2. Bad harness or bad socket (occasional)
- 3. Open in lamp wiring (occasional)
- 4. Short in lamp wiring (rare)
- 5. Bad bulb socket (occasional)
Test bulb continuity
- Test point
- Bulb terminals
- Expected
- <1 Ohm
- Tools
- Ohm Meter
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm message: Temp Probe Failure patient safety
Display shows alarm message indicating temperature probe not functioning
Ranked causes
- 1. Defective temperature probe (common)
- 2. Defective rail harness (occasional)
- 3. Defective Probe Panel Interface Board (occasional)
- 4. Defective Control PCA (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank black screen with no alarms
Display shows no image and no alarm messages are generated.
Ranked causes
- 1. Loss of 12V to Control PCA (common)
- 2. LCD backlight not connected to LED driver board (common)
- 3. Protective film on LCD (occasional)
- 4. Bad LCD cable connection (common)
Verify LED7 indicator is lit
- Test point
- LED7 on Control PCA
- Expected
- LED lit (red indicates power to LCD module)
5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flashing alarm light; warmer will not initialize caution
Alarm light flashes continuously and unit fails to initialize
Ranked causes
- 1. DC supply voltage low (common)
- 2. Defective power supply (occasional)
- 3. Defective warmer head harness (occasional)
- 4. Defective Power PCA (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hands free alarm silence not working
Hands free alarm silence feature does not function when hand waved
Ranked causes
- 1. Too much ambient sunlight (occasional)
- 2. Alarm/Light PCA not programmed (occasional)
- 3. Function disabled in software (occasional)
- 4. Open connection or defective emitter/receiver (occasional)
Ambient light conditions
- Expected
- Minimal direct sunlight, procedure light not maximum
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light output flickers at lowest LED observation light setting
Observation light (LED) flickers or flashes at lowest brightness setting; higher settings may work normally
Ranked causes
- 1. Defective LED bulb (common)
- 2. Loose LED contact in socket (occasional)
- 3. Loose J9 mating connector on Power PCA (occasional)
- 4. Faulty Power PCA board (occasional)
Measure observation LED bulb impedance
- Test point
- LED bulb
- Expected
- >10 kilo Ohm for good bulb
- Tools
- Multimeter
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio indication of alarms patient safety
No audible tone when alarm conditions occur
Ranked causes
- 2. Door switch harness connection loose (occasional)
- 3. Solder connection failure at speaker (occasional)
- 4. Defective speaker (occasional)
- 5. Control PCA defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 probe detected patient safety
SpO2 sensor not recognized when connected
Ranked causes
- 1. Improper probe mating with connector (common)
- 2. Dislodged SpO2 connector (common)
- 3. Reverse probe connection (occasional)
- 4. Defective SpO2 Flex Circuit (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
One observation light bulb does not illuminate
One of the two observation light bulbs fails to produce light.
Ranked causes
- 1. Bad or burned-out bulb (common)
- 2. Bad harness or bad socket (occasional)
- 3. Faulty bulb socket (occasional)
- 4. Bad harness connection at power PCA (rare)
Test bulb continuity
- Test point
- Bulb terminals
- Expected
- <1 Ohm
- Tools
- Ohm Meter, Bladed screwdriver
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red screen with flashing alarm lights and HP alarm
Display shows red screen; alarm lights flash; HP alarm sounds
Ranked causes
- 1. NVRAM failure (common)
- 2. Defective or disconnected ribbon cable (occasional)
- 3. Control PCA failure (occasional)
- 4. Real Time Clock / NVRAM hardware failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale does not return a weight caution
Scale fails to display weight readings or produces consistently inaccurate weights
Ranked causes
- 1. Hardware issue or miscalibration (common)
- 2. Software malfunction (occasional)
- 3. Outdated scale software version (occasional)
- 4. Scale slow to return weight (occasional)
Perform scale hardware check
- Test point
- Scale hardware sensors
- Expected
- All hardware components functional per procedure 3.2.5.1
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sys Fail 11 [AA AA AA AA] patient safety
Heat Control Processor communication failed due to Power-On Self Test failure
Ranked causes
- 1. Wrong mains voltage (occasional)
- 2. Open or loose heater AC fuse connection (common)
- 3. Open T-Stat at Power-on (occasional)
- 4. Defective/Disconnected cable (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sys Fail 11 [xx 00 00 00] or [FF FF FF FF] patient safety
Heat Control Processor stopped communicating with Main Application Processor
Ranked causes
- 1. Defective/Disconnected ribbon cable (common)
- 2. No +5V DC at TP20 pin 2 (common)
- 3. Invalid /Reset signal (common)
- 4. Open or faulty fuses F9 & F10 (common)
Inspect ribbon cable seating
- Test point
- 20 pin ribbon cable between Power and Control PCA
- Expected
- Fully seated
- Tools
- Visual inspection
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 11 (Sys Fail 11 [xx 00 00 00] or [FF FF FF FF]) caution
Unit displays Sys Fail 11; heat control processor stopped communicating with main application processor.
Ranked causes
- 1. Defective or disconnected cable (common)
- 2. Power PCA voltage failure (common)
- 3. Reset signal failure (common)
- 4. Open or loose heater AC fuse connection (occasional)
Verify ribbon cable seating
- Test point
- 20 pin ribbon between Power and Control PCA
- Expected
- Cable fully seated at both connectors
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temp Probe Failure alarm message patient safety
Temperature probe sensor not detected or communication failure with probe
Ranked causes
- 1. Incorrect probe connections (common)
- 2. Damaged probe rail harness (occasional)
- 3. Defective Probe Panel Interface Board (occasional)
- 4. Defective Control PCA (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature ADC channels out of specification
Temperature sensor readings or reference voltages exceed expected tolerances during diagnostics
Ranked causes
- 1. ADC channel 1 (Vthref) voltage out of range (unknown)
- 2. Patient probe thermistor mismatch (unknown)
- 3. Reference temperature (37°C) voltage out of range (unknown)
- 4. Reference 1P235 voltage out of range (unknown)
Measure ADC channel 1 Vthref
- Test point
- ADC Channel 1
- Expected
- Raw Value 2048 (±4), Corrected Value 1500 (±3) mV raw, ±4
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not turn on (power fail alarm) high voltage
Unit does not power up; power fail alarm displayed or triggered; no response to power switch
Ranked causes
- 1. No 12V DC power supply output (common)
- 2. Blown fuse on Power PCA (F7 or F8) (occasional)
- 3. No AC voltage to Power PCA (front switch, harness, line filter) (common)
- 4. No signal from Heat Control Processor (rare)
Measure 12V DC from chassis to top of fuse F7 on Power PCA
- Test point
- Power PCA fuse F7 (top side)
- Expected
- 12V DC, nominal
- Tools
- DC Volt Meter
10 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight not working
Display backlight does not illuminate or is very dim
Ranked causes
- 1. Fuse F7 open on Power PCA (common)
- 2. Bad backlight connections (common)
- 3. Defective Control PCA or LCD module (occasional)
Check fuse F7 resistance
- Test point
- Fuse F7 on Power PCA
- Expected
- < 1 Ohm if intact, open if blown
- Tools
- Ohm Meter, Needle nose pliers
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank black screen with no alarms
Display is completely black with no alarm indications
Ranked causes
- 1. Loss of 12V to Control PCA (common)
- 2. LCD backlights not connected (common)
- 3. Fuse F7 failure on Power PCA (occasional)
Check LED7 light status on Control PCA
- Test point
- LED7 on Control PCA
- Expected
- Red light indicates LCD receiving power
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both bulbs do not light (LED) caution
Neither of the two observation LEDs illuminates when switched on
Ranked causes
- 1. Loose connector and wiring (common)
- 2. Bad dimmer pot switch (occasional)
- 3. Bad Power PCA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display not visible or unresponsive
LCD screen shows nothing or cannot be interacted with; backlight may not be illuminated.
Ranked causes
- 1. Protective film on LCD (common)
- 2. Bad LCD cable connection (common)
- 3. Backlight connection failure (occasional)
Verify protective film removal
- Test point
- LCD surface
- Expected
- No film present
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
E-Base not moving caution
Elevating base does not respond to foot switch pedals; no vertical bed movement
Ranked causes
- 1. Defective E-Base unit (common)
- 2. No power to E-Base (common)
- 3. Blown fuse on Power PCA (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base no motion caution
Footswitch pedals do not raise or lower base; green light may or may not be present
Ranked causes
- 1. Defective E-Base unit (occasional)
- 2. No power to E-Base (common)
- 3. Loose footswitch connectors (common)
Plug spare line cord directly into E-Base AC inlet and observe indicator light
- Test point
- E-Base AC inlet; green light indicator on E-Base column
- Expected
- Green light ON
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flashing alarm light; warmer will not initialize
Alarm light flashing and warmer fails to initialize properly
Ranked causes
- 1. DC Supply voltage issue (common)
- 3. Defective harness (occasional)
- 4. Defective Power PCA (occasional)
DC supply voltage
- Test point
- Power PCA TP20 pins 3 and 22-3
- Expected
- at least 11.5V, >= 11.5
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hands Free Alarm Silence does not work
Hands free alarm silence feature is not responding to hand gestures in front of alarm panel
Ranked causes
- 2. Alarm/light PCA not programmed (occasional)
- 3. Function disabled or not enabled in software (occasional)
- 4. Open connection in ribbon cable (occasional)
Verify device location and lighting
- Test point
- Ambient light conditions and light settings
- Expected
- Device not in direct sunlight; light settings not at maximum
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD backlight dim or non-uniform lighting
LCD backlight is dim or illumination is uneven across display
Ranked causes
- 1. Bad LCD backlight component (common)
- 2. Loss of 12V power to Control PCA (occasional)
- 3. LCD backlights not connected to Control PCA (occasional)
Check backlight connector seating at Control PCA
- Test point
- J4, J8 on Control PCA
- Expected
- Connectors mated and pinned correctly, no bent pins
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audible Power Fail Alarm high voltage
No audible alarm tone when unit loses AC power
Ranked causes
- 2. Ribbon cable disconnected or improperly seated (occasional)
- 3. Front panel switch connection loose (occasional)
- 4. Defective or depleted battery (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No scale icon displayed caution
Scale icon does not appear on display screen for bedded scale unit
Ranked causes
- 2. Defective cable (common)
- 3. Defective printed circuit board (occasional)
- 4. Defective scale receptacle on unit (occasional)
Verify scale connection
- Test point
- Scale connector at probe panel
- Expected
- Scale properly connected and seated
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Potential patient harm from misconfigured alarms patient safety
Patient safety compromised if alarm limits not verified or alarms silenced inappropriately
Ranked causes
- 1. Alarm limits not verified for current patient (common)
- 2. Audible alarm function silenced or volume decreased inappropriately (occasional)
- 3. Alarm speaker blocked or obstructed (occasional)
Verify alarm limits are set appropriately for patient
- Test point
- Pulse oximeter alarm limit settings
- Expected
- Low and high SpO2 alarm thresholds appropriate for patient age and clinical condition
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale icon disappears/reappears repeatedly caution
Scale icon on control panel intermittently disappears and reappears; scale makes repeated clicking noise
Ranked causes
- 1. Defective or disconnected scale cable (common)
- 2. Defective printed circuit board (occasional)
- 3. Defective scale receptacle on unit (occasional)
Verify connector insertion
- Test point
- Scale connector at receptacle
- Expected
- Connector fully inserted into receptacle
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor disconnect error or alarm patient safety
Sensor disconnect error message displayed and associated alarm triggered
Ranked causes
- 1. Sensor disconnected from patient or control unit (common)
- 2. Faulty wiring or loose connector (occasional)
- 3. Damaged sensor (occasional)
Verify sensor physical connection to control unit
- Test point
- Sensor connector port on control panel
- Expected
- Connector fully seated and secure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor disconnect or no signal patient safety
Pulse oximeter displays sensor disconnect error message and associated alarm; no reading obtained
Ranked causes
- 1. Sensor disconnected from module (common)
- 2. Faulty wiring or cable damage (occasional)
- 3. Defective sensor (occasional)
Verify sensor connector is fully seated
- Test point
- Sensor port on Panda iRes or Giraffe control panel
- Expected
- Connector fully inserted; audible click heard
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 readings affected by environmental or application conditions patient safety
Pulse oximetry readings may be inaccurate due to sensor application errors, ambient environmental conditions, or patient conditions affecting measurement.
Ranked causes
- 1. Sensor applied incorrectly or placed on unsuitable site (common)
- 2. Ambient environmental conditions interfering with optical measurement (common)
- 3. Patient motion or unstable perfusion (occasional)
Ensure sensor is applied to pulsating arteriolar vascular bed
- Test point
- Sensor placement site
- Expected
- Finger, toe, or other appropriate pulsating arteriolar site
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sys Fail 03
Memory error occurred in NVRAM or control PCA
Ranked causes
- 1. NVRAM chip cover not properly seated (common)
- 2. Corrupted application flash (occasional)
- 3. Control PCA failure (occasional)
Inspect NVRAM chip cover
- Test point
- NVRAM chip on control PCA
- Expected
- Cover fully and properly seated; battery retained
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sys Fail 06 [0B XX 00 00] patient safety
Over temperature thermostat is open. Added in MAP Version 1.1.0 and later.
Ranked causes
- 1. Open over-temperature thermostat (common)
- 2. Defective thermostat or harness (common)
- 3. Defective power PCA (occasional)
Inspect heater vents for obstruction
- Test point
- Heater head cooling vents and heater duct
- Expected
- No obstructions present
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sys Fail 06 [0B XX 00 00] patient safety
Over temp T-Stat open condition detected
Ranked causes
- 1. Open overtemp T-Stat (common)
- 2. Defective T-Stat or Harness (common)
- 3. Defective Power PCA (occasional)
Inspect for heater blockage
- Test point
- Heater duct and heater head vents
- Expected
- No obstructions visible
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sys Fail 06 overtemp T-Stat open caution
Sys Fail 06 error, over temperature T-Stat open circuit detected
Ranked causes
- 1. Open overtemp T-Stat (common)
- 2. Defective T-Stat or harness (occasional)
- 3. Defective Power PCA (occasional)
Check heater duct and vents for blockage
- Test point
- Heater duct, heater head vents
- Expected
- Clear passage with no obstruction
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Sys Fail 09 [XX 80 XX XX] / [XX 10 XX XX] / [XX 50 XX XX] / [0A XX 00 00]
- Sys Fail 11 [AA AA AA AA]
- System Failure 02 (Sys Fail 02)
- System Failure 06 (Sys Fail 06 [0B XX 00 00])
- System Failure 08 (Sys Fail 08 [xx xx 00 00])
- Unit will not turn on
- Alarm message: Confirm Probe Jack Connection
- Alarm sounds inaudible or suppressed
- Bed does not tilt
- Bed does not tilt
- Button not working
- Button not working
- Calibration: Heater head alignment
- Cannot enter Service Mode
- Check SpO2 probe quality
- Damaged sensor or cable detected
- Device exposure to moisture or extreme conditions
- Dim or non-uniform light
- HR missing and scale missing since startup
- Inadequate patient cable or sensor integrity
- LCD display blank black screen with no alarms
- LCD display blank black screen with no alarms
- LCD display not visible or has arrow after replacement
- Light output dim at highest LED observation light setting
- Light output dim at highest setting (Halogen)
- Light output flickers at lowest setting (Halogen)
- Lights are not dimmable (Halogen)
- Lights are not dimmable (LED)
- Lights stay dim with switch off (Halogen)
- Lights stay ON with observation knob switched off (LED)
- Lithium-ion battery installation error
- No POST pass tone heard at startup
- No SpO2 probe detected
- Observation light dim at highest brightness setting
- Observation light dim at highest setting
- Observation light flickers at lowest setting
- Observation light flickers at lowest setting
- Observation light output dim at highest setting
- Observation light output flickers at lowest setting
- Observation lights dim at highest setting
- Observation lights flicker at lowest setting
- Observation lights not dimmable
- Observation lights not dimmable
- Observation lights stay dim with switch off
- Observation lights stay dim with switch off
- One observation bulb does not light
- Oximetry system failure
- Sensor tissue damage from incorrect application or prolonged use
- SpO2 reading unusually high or low despite normal clinical appearance
- Sys Fail 01 [FF xx xx xx]
- Sys Fail 01 [FF xx xx xx]
- Sys Fail 01 with flash memory error
- Sys Fail 02
- Sys Fail 02 power mismatch
- Sys Fail 03 memory error
- Sys Fail 04 memory redundancy error
- Sys Fail 05 ADC configuration error
- Sys Fail 08 [xx xx 00 00]
- Sys Fail 08 [xx xx 00 00]
- Sys Fail 08 with button shorted
- Sys Fail 09 [0A XX 00 00]
- Sys Fail 09 heater safety relay open
- Sys Fail 11 [AA AA AA AA] alarm
- Sys Fail 11 [xx 00 00 00] or [FF FF FF FF]
- System Fail 11 - Heat Control Processor communication failure
- System Failure 01 (Sys Fail 01 [FF xx xx xx])
- System Failure 03 (Sys Fail 03)
- System Failure 04 (Sys Fail 04)
- System Failure 09 (Sys Fail 09 [0A XX 00 00])
- System Failure 11 (Sys Fail 11 [AA AA AA AA])
- System failure detected in event logs
- Temp probe failure alarm
- Touch panel button not responding
- Touch panel button not working
- Touch panel buttons or switches not responding
- Touch panel or LCD assembly removal
- White display, striped display, or color shift
- 25% Output Check reading incorrect
- 25% output test operation
- 25% output test shows incorrect reading
- Access service mode from powered-on device
- Alarm limits inappropriate for patient
- Alarm message: Temp Probe Failure
- Baby Cold - Check Temp Probe
- Baby Cold - Check Temp Probe (1.0°C threshold)
- Baby Hot - Check Temp Probe
- Baby Hot - Check Temp Probe (1.0°C threshold)
- Base and rail mounting block replacement (freestanding)
- Bed height adjustment on crank handle system
- Bed height adjustment on manual pin system
- Black background with red, green, or blue tint
- Blank screen with alarm and indicator lights on
- Blank screen with alarm lights and HP alarm
- Blank screen with alarm lights on
- Blank screen with alarm lights on and HP alarm
- Blank Screen, Alarm Lights On, HP Alarm
- Calibration: Date and time not synchronized
- Calibration: Heater head alignment
- Calibration: Scale
- Calibration: Scale calibration required
- Calibration: Scale weight
- Calibration: Scale weight accuracy
- Calibration: Scale weight measurement
- Calibration: Scale weight measurement
- Callback Timer Complete
- Cannot enter service mode
- Caster replacement needed
- Check Baby - Heat Off (CPAD limit exceeded)
- Check Baby - Heat Off (radiant power 100%)
- Check Baby (CPAD limit exceeded)
- Check Baby (radiant power 100%)
- Check SpO2 - defective probe
- Check SpO2 - defective probe indicated
- Check SpO2 - excessive ambient light
- Check SpO2 - interference detected
- Control board removal
- Control board replacement needed
- Damaged equipment from sterilization method
- Defective scale
- Defective scale
- Defective scale (Bedded Models Only)
- Defective SpO2 probe
- Defibrillation or electrosurgery interference
- Device will not boot up
- Dim or non-uniform LCD backlight
- Disconnected temperature probe
- Display arrow present after installing new LCD
- Display arrow present after installing new LCD
- Display has pixel defects or color errors
- Display shows abnormal colors or incomplete test pattern
- Display test pattern abnormal
- Display test pattern shows dropout or color errors
- Distorted audio from speaker
- Distorted audio from speaker
- Distorted audio speaker
- Distorted display: test or random pattern
- Dovetail rail suction tubing replacement
- Electronic enclosure access restricted or damaged
- EU compliance verification required before operation
- F7 or F8 fuse failure or loose connection
- F7 or F8 fuse replacement needed
- Firmware update execution
- Fuse F7 or F8 damaged or loose connection
- Hands-free alarm silence not responding
- Heat engine diagnostic fails or shows abnormal state
- Heat engine diagnostic parameters query
- Heat engine diagnostics unavailable
- Heat engine health check perform
- Heat engine output not at specified level
- Heat engine output out of range
- Heater resistance measurement
- High Pulse Rate
- High SpO2
- High SpO2 reading
- HR active and scale missing
- HR subsystem failure and scale active
- HR subsystem failure and scale missing
- Inadequate pulse oximetry as apnea monitor
- Incorrect cable connected to sensor port
- LCD display color test pattern abnormal
- LCD display dim or non-uniform lighting
- LCD display distorted with test or random pattern
- LCD display distorted with test pattern or random pixels
- LCD display shows distorted test or random pattern
- LCD display shows white, striped, or color shifted image
- LCD display white, striped, or color shift
- LCD display white, striped, or color shifted
- LCD display white, striped, or color shifted
- Lead off detected
- Lights stay ON with observation light knob switched off
- Lithium-ion battery replacement
- Lithium-ion battery replacement needed
- Log dump fails or times out
- Log dump transfer timeout
- Low Perfusion (SpO2 Alert)
- Low perfusion detected at probe site
- Low pulse rate
- Low Signal IQ (SpO2 Alert)
- Low Signal IQ alert
- Low SpO2
- Low SpO2 reading
- Maneuvering handle replacement
- MRI procedure with pulse oximeter attached
- No audible power fail alarm
- No audio indication of alarms
- No scale icon displayed
- No SpO2 Board
- No SpO2 option on screen
- No SpO2 option on screen
- No SpO2 probe connected
- Observation light dimmer control failure
- Observation light switch or dimmer control replacement
- Outlet panel component failure
- Outlet panel component replacement
- Patient cable entanglement or strangulation hazard
- PC command line interface access failure
- Potential patient harm from carboxyhemoglobin interference
- Potential skin damage or burns from MRI use
- Power board removal
- Power board replacement needed
- Procedure light bulb burned out or dim
- Procedure light bulb replacement
- Pulse oximeter not functioning after power-on self-test
- Readings potentially inaccurate during defibrillation
- Red screen with flashing alarm and indicator lights
- Red screen with flashing alarm lights
- Red screen, alarm lights flashing
- Red screen, alarm lights flashing
- Scale calibration failed
- Scale control board replacement
- Scale gives consistently inaccurate weights
- Scale gives erratically different weights
- Scale load cell replacement
- Scale stuck on weight display screen
- Sensor or cable damage from water or chemical immersion
- SpO2 display shows no signal
- SpO2 interference detected
- SpO2 lost communications
- SpO2 Probe Off Baby
- SpO2 probe off patient
- SpO2 system fails to calibrate
- Standby switch failure
- Standby switch replacement
- Sys Fail 02
- Sys Fail 03
- Sys Fail 04
- Sys Fail 05
- Sys Fail 05
- Sys Fail 09 [0A XX 00 00]
- Sys Fail 09 [0B 00 00 00]
- Sys Fail 09 [0B 00 00 00]
- Sys Fail 09 [0E 72 00 00]
- Sys Fail 09 [XX 80 XX XX]
- Sys Fail 09 [XX 80 XX XX]
- Sys Fail 09 heat control processor failure during runtime
- Sys Fail 10 [XX XX XX XX]
- Sys Fail 11 [FF FF FF FF] alarm
- Sys Fail 12 [xx 00 00 00]
- Sys Fail 12 [xx 00 00 00] alarm
- System Fail 12 - Audio Processor communication failure
- System Failure 05 (Sys Fail 05)
- System Failure 09 (Sys Fail 09 [0B 00 00 00])
- System Failure 09 (Sys Fail 09 [XX 80 XX XX])
- System Failure 10 (Sys Fail 10 [XX XX XX XX])
- System Failure 11 (Sys Fail 11 [FF FF FF FF])
- System Failure 12 (Sys Fail 12 [xx 00 00 00])
- System log retrieval and analysis
- Temperature ADC channel readings out of range
- Temperature ADC readings out of range
- Temperature probe failure
- Too much ambient light (SpO2)
- Touch panel or LCD display malfunction
- Touch panel or LCD display removal difficulty
- Unit will not boot up
- Unit will not boot up
- Unrecognized SpO2 probe
- Unrecognized SpO2 probe attached
- VGA display test pattern defects
258 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Baby Check SpO2 (Ambient Light) | Too much ambient light detected by SpO2 board | Masimo or Nellcor board indicates too much ambient light | error |
Baby Cold - Check Temp Probe | Baby temperature more than 2.0°C below control setpoint | >2.0° from Control Temp (below) | warning |
Baby Cold - Check Temp Probe | Baby temperature more than 2.0°C below control temperature | <2.0° from Control Temp | warning |
Baby Cold - Check Temp Probe (Manual Mode) | Baby temperature more than 1.0°C below control temperature in Manual mode | < 1.0°C (<0.5°C) from control temp | warning |
Baby Hot - Check Temp Probe | Baby temperature more than 2.0°C above control setpoint | >2.0° from Control Temp | warning |
Baby Hot - Check Temp Probe | Baby temperature more than 2.0°C above control temperature | >2.0° from Control Temp | warning |
Baby Hot - Check Temp Probe (Manual Mode) | Baby temperature more than 1.0°C above control temperature in Manual mode | > 1.0°C (>0.5°C) from control temp | warning |
Callback Timer Complete | Callback timer has reached target time | Current callback timer has reached target time | info |
Callback Timer Complete | Scheduled callback timer has expired | Current callback timer has reached target time | info |
Check Baby | Radiant heater power above CPAD limit for extended duration | Radiant Heater % above 'Check Baby Alarm Disabled' (CPAD) limit for > 12 min | warning |
Check Baby | Radiant heater at 100% power for more than 12 minutes | Radiant % power @ 100% for > 12 min | warning |
Check Baby (CPAD Limit) | Radiant heater above Check Baby Alarm Disabled limit for extended time | Radiant Heater % above CPAD limit for > 12 min | warning |
Check Baby - Heat Off | Radiant heater at excessive power for extended time | Radiant % power @ 100% for >15 min and alarm not silenced | warning |
Check Baby - Heat Off | Radiant heater at 100% power for extended time without alarm silencing | Radiant % power @ 100% for >15 min and alarm not silenced | error |
Check Baby - Heat Off (CPAD) | Radiant heater above Check Baby Alarm Disabled limit for extended time | Radiant heater % above CPAD limit for >15 min and alarm not silenced | error |
Check Baby (Radiant Power) | Radiant heater at 100% power for extended time in temperature control | Radiant % power @ 100% for > 12 min | warning |
Check SpO2 | Defective SpO2 probe detected | Masimo or Nellcor board indicates defective probe | warning |
Check SpO2 | Defective SpO2 probe, interference, or too much ambient light detected | Masimo or Nellcor board indicates defect, interference, or light issue | warning |
Check SpO2 | SpO2 signal quality is poor or probe is defective | System detects defective probe, electromagnetic interference, or excessive ambient light | error |
Check SpO2 | SpO2 measurement interference detected | Masimo or Nellcor board indicates interference detected | warning |
Check SpO2 | Probe defect detected by Masimo board | Masimo board diagnostic indicates defective probe after 2 minutes | error |
Check SpO2 (Defective Probe) | SpO2 probe detected as defective by board | Masimo or Nellcor board indicates defective probe | error |
Check SpO2 (Interference) | Interference detected by SpO2 board | Masimo or Nellcor board indicates interference detected | error |
Confirm Probe Jack Connection | Disconnected temperature probe | No longer getting a reading from the temperature probe | error |
Confirm Probe Jack Connection | Temperature probe disconnected or no reading received | No longer getting a reading from the temperature probe | error |
Confirm Probe Jack Connection | No reading from temperature probe detected | No longer getting a reading from the temperature probe | error |
High Pulse Rate | Patient pulse rate exceeds alarm limit | Pulse rate exceeds set high alarm threshold (default 200 bpm, user adjustable 111-240 bpm) | warning |
High Pulse Rate | Pulse rate above threshold | PR >200 (floats with user adjustable limits) | warning |
High SpO2 | SpO2 measurement exceeds upper alarm limit (default >100%) | Measured SpO2 exceeds user-adjustable upper threshold for 2 minutes | warning |
High SpO2 | SpO2 > 100 (floats with user adjustable limits) | SpO2 > 100 (floats with user adjustable limits) | warning |
High SpO2 | Oxygen saturation above threshold | SpO2 >100 (floats with user adjustable limits) | critical |
High SpO2 | Patient blood oxygen saturation exceeds alarm limit | SpO2 percentage exceeds set high alarm threshold (default 100%, user adjustable 86-100%) | info |
Low Perfusion | Masimo board detects low blood perfusion at measurement site (SpO2 alert) | Masimo board indicates low perfusion | warning |
Low Perfusion | Low perfusion detected by SpO2 board | Masimo board indicates low perfusion | warning |
Low Pulse Rate | Pulse rate below threshold | PR < 100 (user adjustable limits) | error |
Low Pulse Rate | Pulse rate below 100 BPM (user adjustable) | PR < 100 (floats with user adjustable limits) | warning |
Low Pulse Rate | Patient pulse rate falls below alarm limit | Pulse rate drops below set low alarm threshold (default 100 bpm, user adjustable 25-110 bpm) | warning |
Low Signal IQ | Masimo board detects poor SpO2 signal quality (SpO2 alert) | Masimo board indicates low signal IQ | warning |
Low Signal IQ | System confidence in detected arterial pulse signal is very low | System cannot reliably identify arterial pulse due to motion or low perfusion | warning |
Low Signal IQ | Low signal quality detected by SpO2 board | Masimo board indicates low signal IQ | warning |
Low SpO2 | Patient blood oxygen saturation falls below alarm limit | SpO2 percentage drops below set low alarm threshold (default 85%, user adjustable 20-99%) | critical |
Low SpO2 | SpO2 < 85 (floats with user adjustable limits) | SpO2 < 85 (floats with user adjustable limits) | warning |
Low SpO2 | Oxygen saturation below threshold | SpO2 <85 (user adjustable limits) | error |
Low SpO2 | SpO2 measurement falls below lower alarm limit (default <85%) | Measured SpO2 below user-adjustable lower threshold for 2 minutes | critical |
No SpO2 Board | Masimo or Nellcor board not detected; only available if SpO2 running | No board connected | error |
No SpO2 Board | No SpO2 probe or module connected | Masimo or Nellcor board indicates no board connected | warning |
No SpO2 Board | SpO2 board not detected as connected | Masimo or Nellcor board indicates no board connected | error |
No SpO2 Probe | Nellcor board indicates no probe connection | Nellcor board indicates no probe connection | error |
No SpO2 Probe | SpO2 probe is not detected as connected | Probe jack detects no connected probe at startup or during monitoring | warning |
Oximetry System Failure | Oximetry system has failed, such as failure to calibrate successfully | System failure event during operation or startup | error |
Oximetry System Failure | SpO2 board lost communication with main processor for 2 seconds or more; SpO2 module automatically turns off | SpO2 board has not communicated with the main processor for at least 2 seconds | critical |
Oximetry System Failure | SpO2 module has failed or lost communication with warmer system | System startup or during operation when calibration fails or communications lost | critical |
Oximetry System Failure | SpO2 board has not communicated with the main processor for at least 2 seconds | Loss of communication between SpO2 board and main processor for ≥2 seconds | critical |
Processor update failed exit status | Firmware update to audio or heat control processor did not complete successfully | Occurs when 'update' command fails after processor re-programming has commenced | critical |
SpO2 lost communications | SpO2 board has not communicated with the main processor for at least 2 seconds | SpO2 board communication failure | critical |
SpO2 Probe Off | SpO2 probe indicated as off baby | Probe off baby is indicated from the Masimo or Nellcor board | error |
SpO2 Probe Off Baby | Probe is not in contact with patient | Probe loses contact with patient skin or is not properly attached | warning |
SpO2 Probe Off Baby | Probe has been removed from baby | Probe off baby is indicated from the Masimo or Nellcor board | warning |
SpO2 Probe Off Baby | SpO2 probe detected as displaced from patient | Probe off baby is indicated from the Masimo or Nellcor board | warning |
SpO2 Probe Off Baby | SpO2 probe no longer in contact with patient | Probe off baby is indicated from the Masimo or Nellcor board | warning |
Sys Fail 01 [FF xx xx xx] | FLASH Memory File System Error | Flash memory file system corrupted or inaccessible | critical |
Sys Fail 01 [FF xx xx xx] | FLASH memory file system error | Memory check during power-on detects flash corruption or invalid file system | error |
Sys Fail 02 | Reported power from power PCA does not match commanded power from control PCA | Power feedback mismatch during operation | error |
Sys Fail 02 | Power mismatch between Power PCA and Control PCA | Reported power from Power PCA does not match commanded power from Control PCA | error |
Sys Fail 03 | Memory error occurred | System detected memory error during operation | error |
Sys Fail 03 | Memory error occurred | NVRAM or memory validation check fails during operation | error |
Sys Fail 04 | Redundant stored variables do not match | Memory redundancy check detects mismatch in stored data | error |
Sys Fail 04 | Redundant stored variables do not match (Memory Redundancy Error) | System detected mismatch in redundant memory variables | error |
Sys Fail 05 | ADC Configuration Error or ADC Failure; Reference channel outside range (1597-1775) | ADC reference channel reading outside specified range 1597-1775 | error |
Sys Fail 05 | ADC configuration error or ADC failure; reference channel outside of range (1597-1775) | ADC reference check at power-on detects out-of-range value | error |
Sys Fail 06 [0B XX 00 00] | Over temperature thermostat is open | T-Stat contact does not close when heated or temperature exceeds safe limits | critical |
Sys Fail 06 [0B XX 00 00] | System Failure 06 with specified auxiliary code | Processor communication or hardware fault detected | error |
Sys Fail 06 [0B XX 00 00] | Overtemp T-Stat is open | Overtemperature thermostat detected open circuit during operation | critical |
Sys Fail 08 [xx xx 00 00] | Touch panel button shorted or single button pressed during power on | Shorted button detected on touch panel during startup or operation | error |
Sys Fail 08 [xx xx 00 00] | Touch panel button shorted | Continuous button input detected during operation | warning |
Sys Fail 08 [xx xx 00 00] | System Failure 08 with variable auxiliary codes | Sensor or board communication error | error |
Sys Fail 09 [0A XX 00 00] | System Failure 09 with first auxiliary byte 0x0A | Specific processor or peripheral fault | error |
Sys Fail 09 [0A XX 00 00] | Heat control processor failure during run time; Heater Safety Relay is open | Safety relay in heater control circuit detected open | critical |
Sys Fail 09 [0B 00 00 00] | Heat control processor failure during run time; Open T-Stat | Thermostat detected open circuit during heat control operation | critical |
Sys Fail 09 [0B 00 00 00] | Heat control processor failure during runtime; open T-Stat | Over-temperature thermostat remains open during heat operation | critical |
Sys Fail 09 [0B 00 00 00] | System Failure 09 with first auxiliary byte 0x0B | Heat control processor communication failure | error |
Sys Fail 09 [0E 72 00 00] | Heat control processor failure during startup; disconnected heater cable | Heater cable not connected or loose at power-on | critical |
Sys Fail 09 [XX 80 XX XX] | System Failure 09 with first auxiliary byte 0x80 | Power or alarm board fault | error |
Sys Fail 09 [XX 80 XX XX] | Heat control processor failure during run time; Heater Control Failure | Heat control processor detected failure during normal operation | critical |
Sys Fail 09 [XX 80 XX XX] / [XX 10 XX XX] / [XX 50 XX XX] / | Heat control processor failure during runtime; heater control failure | Heat control processor communication or operation failure during heating cycle | critical |
Sys Fail 10 [XX XX XX XX] | Self-calibration failure occurred | Self-calibration routine fails during power-up or initialization | error |
Sys Fail 10 [XX XX XX XX] | System Failure 10 with variable auxiliary codes | Multiple processor or board faults | error |
Sys Fail 10 [XX XX XX XX] | Self-calibration failure occurred | Unit failed to complete self-calibration during startup or diagnostics | error |
Sys Fail 11 [AA AA AA AA] | Heat Control Processor stopped communicating due to POST (Power-On Self Test) failure | Power PCA POST failed during power-on sequence | critical |
Sys Fail 11 [AA AA AA AA] | System Failure 11 with alternating pattern auxiliary code | Heat control processor communication with specific fault pattern | critical |
Sys Fail 11 [AA AA AA AA] | No communication of heat control processor with main processor since power-on | Heat control processor not responding since device powered on | error |
Sys Fail 11 [AA AA AA AA] | Heat Control Processor communication failure due to POST failure | Power-on self-test failure | error |
Sys Fail 11 [FF FF FF FF] | Heat control processor not responding; communication completely lost | Heat control processor power loss or complete failure | critical |
Sys Fail 11 [FF FF FF FF] | Communication message failed | Inter-processor communication error | error |
Sys Fail 11 [FF FF FF FF] | Heat Control Processor communication message failed | Communication message error between processors detected | error |
Sys Fail 11 [xx 00 00 00] | Heat control processor lost communication with main application processor | Heat control processor stopped responding or communication link failed | critical |
Sys Fail 11 [xx 00 00 00] -or- [FF FF FF FF] | Heat control processor stopped communicating with main processor; communication loss | Communication timeout between heat control and main processor during operation | error |
Sys Fail 11 [xx 00 00 00] or [FF FF FF FF] | Heat Control Processor stopped communicating with Main Application Processor (xx = seconds since last communication) | Communication loss between heat control and main processor during operation | critical |
Sys Fail 12 [xx 00 00 00] | Audio Processor stopped communicating with Main Application Processor (xx = seconds since last communication) | Communication loss between audio processor and main processor | warning |
Sys Fail 12 [xx 00 00 00] | Audio Processor stopped communicating with Main Application Processor | Audio processor communication timeout | error |
Sys Fail 12 [xx 00 00 00] | System Failure 12 with specified auxiliary code | Power or control board processor fault | error |
Temp. Probe Failure | Two thermistors in probe differ by 0.5°C or more | Two Thermistors in a probe differ by 0.5° C or more | error |
Temp. Probe Failure | Two thermistors in probe differ by 0.5°C or more | Two Thermistors in a probed differ by 0.5° C or more | error |
Unrecognized SpO2 Probe | Nellcor board indicates unrecognized probe | Nellcor board indicates unrecognized probe | error |
Unrecognized SpO2 Probe | Probe model not recognized by Masimo board | Masimo board fails to identify probe model after 2 minutes; audio progresses 2 to 1 after 10 secs | error |
Unrecognized SpO2 Probe | Unrecognized SpO2 probe type connected | Unrecognized probe indicated from Masimo or Nellcor board | warning |
Unrecognized SpO2 Probe | Attached probe is not compatible with this warmer system | System detects probe with incompatible connector or incorrect model | error |
Unrecognized SpO2 Probe | SpO2 probe not recognized by board | Unrecognized probe indicated from Masimo or Nellcor board | error |
xmodem error: -69 | Xmodem file transfer protocol error during log dump | Occurs when log dump command times out or communication is interrupted during xmodem transfer | error |
xmodem error: -69 | Log dump xmodem transfer failed | During logs dump file transfer via xmodem | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.