Datex-Ohmeda Panda: troubleshooting and error codes

The ServiceForge knowledge base holds 318 documented symptoms for the Datex-Ohmeda Panda, with 622 ranked causes, 562 diagnostic checks with test points and expected values, and 110 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Procedure light does not illuminate

Procedure light assembly does not produce light when switched on

Ranked causes

  1. 1. Burned-out or defective halogen bulb (common)
  2. 2. 12V DC power source or control circuitry failure (occasional)
  3. 3. No mains power to 12V DC power supply (occasional)
  4. 4. AC short causing blown fuse on Power PCA (occasional)
  5. 5. AC short on Power PCA (occasional)
  6. 6. AC short in wiring to 12V DC power supply (rare)
  7. 7. AC short in 12V DC power supply (rare)
  8. 8. DC power supply output failure (occasional)
  9. 9. Disconnected or faulty harness (occasional)
  10. 10. Faulty Procedure Light harness (occasional)
  11. 11. Faulty Alarm/Light PCA ribbon cable or defective PCA (rare)
First check

Test halogen bulb for continuity

Test point
Halogen bulb
Expected
<1 Ohm
Tools
Ohm Meter

13 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Procedure light does not illuminate

No light output from procedure light assembly when switched on

Ranked causes

  1. 2. No 12V DC power to bulb socket (common)
  2. 4. Blown AC fuse on Power PCA (occasional)
  3. 5. AC short on Power PCA (rare)
  4. 6. AC short in wiring to 12V DC Power Supply (occasional)
  5. 7. AC short in 12V DC Power Supply (rare)
  6. 8. PF choke failure (rare)
  7. 9. Defective or disconnected 12V DC Power Supply (occasional)
  8. 10. Open in harness between power supply and bulb (occasional)
  9. 11. Faulty procedure light assembly harness (occasional)
  10. 12. No 12V DC to procedure light assembly connector (occasional)
  11. 13. Faulty Alarm/Light PCA ribbon cable (rare)
First check

Test halogen bulb continuity

Test point
Halogen bulb terminals
Expected
<1 Ohm
Tools
Ohm Meter

12 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 reading inaccurate or absent patient safety

Pulse oximetry measurements affected by patient conditions, environmental factors, or sensor application errors

Ranked causes

  1. 1. Excessive patient movement (common)
  2. 2. Sensor applied too tightly (common)
  3. 3. Ambient light interference (common)
  4. 4. Venous pulsations (occasional)
  5. 5. Intravascular dyes or external coloring (occasional)
  6. 6. Sensor on extremity with blood pressure cuff or catheter (occasional)
  7. 7. Poor peripheral perfusion (occasional)
  8. 8. Arterial occlusion proximal to sensor (rare)
  9. 9. Dysfunctional hemoglobins (carboxyhemoglobin, methemoglobin, sulphemoglobin) (rare)
  10. 10. Anemia with hemoglobin below 5 gm/dl (rare)
First check

Verify sensor is properly and securely applied

Test point
Sensor site on patient extremity
Expected
Sensor firmly positioned, not too tight, in contact with skin

3 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate pulse oximetry readings patient safety

SpO2 measurements do not reflect actual patient oxygenation; readings inconsistent with clinical assessment

Ranked causes

  1. 1. Incorrect sensor application or use (common)
  2. 2. Significant dysfunctional hemoglobin levels (occasional)
  3. 3. Intravascular dyes present (rare)
  4. 4. Excessive ambient light exposure (occasional)
  5. 5. Excessive patient movement (common)
  6. 6. Venous pulsations detected (occasional)
  7. 7. Blood pressure cuff or catheter on same extremity (common)
  8. 8. Non-specified accessories or cables used (occasional)
First check

Verify sensor application per directions for use

Test point
Sensor site
Expected
Sensor applied correctly per manufacturer instructions

3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base no motion caution

Elevating base (E-Base) does not move up or down

Ranked causes

  1. 1. Defective elevating base unit (occasional)
  2. 2. No power to elevating base (occasional)
  3. 3. Blown fuse on power distribution (occasional)
  4. 4. Disconnected E-Base harness (occasional)
  5. 5. Defective E-Base harness (occasional)
  6. 6. Defective Power PCA (occasional)
  7. 7. Defective Control PCA (occasional)
First check

Test E-Base movement and power indicator

Expected
Green light ON; E-Base responds to foot switch pedals

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate SpO2 measurements patient safety

Pulse oximetry readings appear unreasonable or do not correlate with patient vital signs

Ranked causes

  1. 1. Incorrect sensor application or placement (common)
  2. 2. Dysfunctional hemoglobin present (carboxyhemoglobin or methemoglobin) (occasional)
  3. 3. Intravascular dyes present (occasional)
  4. 4. Excessive illumination interfering with sensor (common)
  5. 5. Excessive patient movement or artifact (occasional)
  6. 6. Venous pulsations present (rare)
  7. 7. Blood pressure cuff or vascular access on same extremity (common)
First check

Cross-check SpO2 reading with alternate measurement method and patient vital signs

Test point
Patient examination and alternate pulse oximetry or blood gas analysis
Expected
Pulse oximeter reading correlates reasonably with patient condition and alternate measurements

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No light from procedure light (halogen)

Procedure light halogen bulb does not illuminate when switch is turned on; heater may be powered on normally

Ranked causes

  1. 1. Burned-out or defective halogen bulb (common)
  2. 2. No mains AC power to 12V DC power supply or supply defect (occasional)
  3. 3. No mains AC power reaching power PCA J28 connector (occasional)
  4. 4. AC short causing blown fuse in power distribution (occasional)
  5. 5. AC short circuit on power PCA PCB (rare)
  6. 6. AC short in wiring between power PCA and 12V DC supply (rare)
  7. 7. AC short or failure in 12V DC power supply unit (rare)
First check

Test halogen bulb continuity

Test point
Halogen bulb
Expected
<1 Ohm (continuous)
Tools
Ohmmeter

8 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No motion from elevating base caution

Elevating base does not respond to footswitch controls or shows no movement

Ranked causes

  1. 1. Defective elevating base unit (common)
  2. 2. No power to E-Base (common)
  3. 3. Blown fuse (occasional)
  4. 4. Disconnected AC harness (occasional)
  5. 5. Defective E-Base harness (occasional)
  6. 6. Defective Power PCA (occasional)
  7. 7. Defective Control PCA (occasional)

18 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 probe not detected patient safety

System fails to recognize SpO2 probe during startup or operation; probe icon does not appear on screen

Ranked causes

  1. 1. Dislodged SpO2 connector from probe plate (occasional)
  2. 2. Poor probe connector seating at probe panel (occasional)
  3. 3. Reverse probe connection (occasional)
  4. 4. Defective SpO2 flex circuit connection (occasional)
  5. 5. Defective SpO2 probe (occasional)
  6. 6. No communication between SpO2 I/F PCA and Control PCA (common)
  7. 7. Defective SpO2 I/F PCA or Control PCA (occasional)
First check

Verify probe connector seating at probe panel

Expected
Connector fully seated at jack plate

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Both bulbs do not light (Halogen) high voltage

Neither of the two halogen observation bulbs illuminates

Ranked causes

  1. 1. Loose connector and wiring (common)
  2. 2. No mains to transformer (common)
  3. 3. Bad dimmer pot switch (occasional)
  4. 4. Bad fuse (occasional)
  5. 5. Bad relay on power PCA (occasional)
  6. 6. Bad transformer (rare)
First check

Check lamp wiring

Test point
Harness at J26
Expected
Continuity and no shorts

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Both observation light bulbs do not illuminate

Neither of the two observation light bulbs produce light output.

Ranked causes

  1. 1. Bad lamp wiring or connection (common)
  2. 2. No mains voltage to transformer (common)
  3. 3. Bad dimmer potentiometer switch (occasional)
  4. 4. Bad fuse F11 or F12 on power PCA (occasional)
  5. 5. Bad relay on power PCA (rare)
  6. 6. Bad transformer assembly (occasional)
First check

Test mains voltage at transformer input

Test point
J24 pins 3 and 4 on power PCA
Expected
Mains voltage present with dimmer ON
Tools
AC Volt Meter

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hands Free Alarm Silence does not work caution

Hands free silence feature cannot silence alarm when hand waved in front of panel

Ranked causes

  1. 1. Too much ambient sunlight (occasional)
  2. 2. Alarm/Light PCA not programmed or defective (occasional)
  3. 3. Hands free silence disabled in software (occasional)
  4. 4. Open ribbon cable connection (occasional)
  5. 5. Defective emitter/receiver circuit (rare)
  6. 6. Defective Control PCA (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of pulse signal patient safety

Pulse waveform no longer detected by pulse oximeter despite sensor connection

Ranked causes

  1. 1. Sensor applied too tightly (common)
  2. 2. Excessive illumination at sensor site (common)
  3. 3. Blood pressure cuff inflated on sensor extremity (common)
  4. 4. Patient hypotension, severe vasoconstriction, severe anemia, or hypothermia (occasional)
  5. 5. Arterial occlusion proximal to sensor (rare)
  6. 6. Patient in cardiac arrest or shock (rare)
First check

Assess sensor tightness on extremity

Test point
Sensor application site
Expected
Sensor snug but not tight; skin blanching minimal if any

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of up/down control from one footswitch caution

One footswitch pedal does not raise or lower base; other pedal may work normally

Ranked causes

  1. 2. Blown fuse in power circuit (occasional)
  2. 3. Disconnected AC harness to Power PCA (occasional)
  3. 3. Low AC voltage to E-Base harness (occasional)
  4. 4. Defective E-Base harness (occasional)
  5. 5. Defective Power PCA (rare)
  6. 6. Defective Control PCA (rare)
First check

Test fuses F1, F4 on Power PCA for continuity

Test point
Fuses F1, F4 on Power PCA after removing observation light transformer bracket
Expected
Fuse intact (ohm meter shows continuity); not blown
Tools
5.5 mm Hex Wrench, Needle nose pliers, Ohm Meter

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No SpO2 signal or zero saturation displayed patient safety

Module reports zero saturation with loss-of-pulse condition and audible alarm; signal search icon appears.

Ranked causes

  1. 1. Sensor too tight on patient extremity (common)
  2. 2. Excessive ambient illumination interfering with measurement (common)
  3. 3. Blood pressure cuff inflated on same extremity as sensor (occasional)
  4. 4. Patient hypotension, vasoconstriction, anemia, or hypothermia (occasional)
  5. 5. Arterial occlusion proximal to sensor site (rare)
  6. 6. Patient in cardiac arrest or shock (rare)
First check

Verify signal strength indicator

Test point
Signal strength indicator on display
Expected
Strong signal

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One bulb does not light (LED) caution

Only one of two observation LEDs illuminates when switched on

Ranked causes

  1. 1. Defective bulb (common)
  2. 2. Loose contact at LED socket (common)
  3. 3. Loose contact of mating connector at Power PCA (occasional)
  4. 4. Faulty lamp wiring (occasional)
  5. 5. Bad socket (occasional)
  6. 6. Bad Power PCA (rare)
First check

Reseat LED in socket

Test point
LED socket
Expected
LED glows when knob rotated

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm message: Confirm Probe Jack Connection patient safety

Display prompts user to confirm or check probe jack connection

Ranked causes

  1. 1. Defective or improperly seated temperature probe (common)
  2. 2. Cold probe (less than 24°C) (common)
  3. 3. Defective rail harness (occasional)
  4. 4. Defective Probe Panel Interface Board (occasional)
  5. 5. Defective Control PCA (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Both observation bulbs do not light

Both observation light bulbs fail to illuminate

Ranked causes

  1. 1. No mains power to transformer (common)
  2. 2. Bad dimmer pot switch (occasional)
  3. 3. Bad fuse on Power PCA (occasional)
  4. 4. Bad relay on Power PCA (occasional)
  5. 5. Bad transformer (occasional)
First check

Check lamp wiring continuity

Test point
Harness at J26
Expected
Continuity at pins 1-3 and 2-4

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Both observation bulbs do not light

Neither bulb in observation light assembly illuminates

Ranked causes

  1. 1. No AC mains to observation light transformer (common)
  2. 2. Bad dimmer pot switch (occasional)
  3. 3. Bad fuse on Power PCA (occasional)
  4. 4. Bad relay on Power PCA (occasional)
  5. 5. Bad observation light transformer (occasional)
First check

Check lamp wiring continuity

Test point
Harness at J26: pins 1-3 and 2-4 with bulbs installed
Expected
Continuity present
Tools
Ohm Meter

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Confirm Probe Jack Connection alarm patient safety

Probe temperature below 24C or probe connection not properly detected

Ranked causes

  1. 1. Defective probe (common)
  2. 2. Cold probe below 24C (common)
  3. 3. Damaged rail harness (occasional)
  4. 4. Defective Probe Panel Interface Board (occasional)
  5. 5. Defective Control PCA (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of up/down control from one foot switch caution

One foot switch pedal does not move the elevating base up or down

Ranked causes

  1. 2. Disconnected harness at Power PCA J28 (occasional)
  2. 3. Defective E-Base harness (occasional)
  3. 4. Defective Power PCA (occasional)
  4. 5. Defective Control PCA (occasional)
  5. 6. Blown fuse F1 or F4 (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No light from procedure light (LED)

Procedure light LED does not illuminate when switch is turned on; heater may be powered on normally

Ranked causes

  1. 1. Burned-out or defective LED bulb (common)
  2. 2. 12V DC power supply or control circuitry failure (occasional)
  3. 3. Loose LED socket contact (occasional)
  4. 4. Loose mating connector at Power PCA J9 (occasional)
  5. 5. Faulty procedure light harness (occasional)
First check

Measure LED bulb impedance

Test point
LED bulb
Expected
>10 Kilo-Ohm for good bulb
Tools
Multimeter

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One bulb does not light (Halogen) high voltage

Only one of two halogen observation bulbs illuminates

Ranked causes

  1. 1. Bad bulb (common)
  2. 2. Bad harness or bad socket (occasional)
  3. 3. Short in lamp wiring (occasional)
  4. 4. Bad socket (rare)
  5. 5. Bad harness connection (rare)
First check

Test bulb continuity

Test point
Bulb terminals
Expected
< 1 Ohm

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One observation bulb does not light

Single bulb in observation light assembly does not illuminate while the other may work

Ranked causes

  1. 1. Bad bulb (common)
  2. 2. Bad harness or bad socket (occasional)
  3. 3. Open in lamp wiring (occasional)
  4. 4. Short in lamp wiring (rare)
  5. 5. Bad bulb socket (occasional)
First check

Test bulb continuity

Test point
Bulb terminals
Expected
<1 Ohm
Tools
Ohm Meter

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm message: Temp Probe Failure patient safety

Display shows alarm message indicating temperature probe not functioning

Ranked causes

  1. 1. Defective temperature probe (common)
  2. 2. Defective rail harness (occasional)
  3. 3. Defective Probe Panel Interface Board (occasional)
  4. 4. Defective Control PCA (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blank black screen with no alarms

Display shows no image and no alarm messages are generated.

Ranked causes

  1. 1. Loss of 12V to Control PCA (common)
  2. 2. LCD backlight not connected to LED driver board (common)
  3. 3. Protective film on LCD (occasional)
  4. 4. Bad LCD cable connection (common)
First check

Verify LED7 indicator is lit

Test point
LED7 on Control PCA
Expected
LED lit (red indicates power to LCD module)

5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flashing alarm light; warmer will not initialize caution

Alarm light flashes continuously and unit fails to initialize

Ranked causes

  1. 1. DC supply voltage low (common)
  2. 2. Defective power supply (occasional)
  3. 3. Defective warmer head harness (occasional)
  4. 4. Defective Power PCA (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hands free alarm silence not working

Hands free alarm silence feature does not function when hand waved

Ranked causes

  1. 1. Too much ambient sunlight (occasional)
  2. 2. Alarm/Light PCA not programmed (occasional)
  3. 3. Function disabled in software (occasional)
  4. 4. Open connection or defective emitter/receiver (occasional)
First check

Ambient light conditions

Expected
Minimal direct sunlight, procedure light not maximum

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Light output flickers at lowest LED observation light setting

Observation light (LED) flickers or flashes at lowest brightness setting; higher settings may work normally

Ranked causes

  1. 1. Defective LED bulb (common)
  2. 2. Loose LED contact in socket (occasional)
  3. 3. Loose J9 mating connector on Power PCA (occasional)
  4. 4. Faulty Power PCA board (occasional)
First check

Measure observation LED bulb impedance

Test point
LED bulb
Expected
>10 kilo Ohm for good bulb
Tools
Multimeter

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No audio indication of alarms patient safety

No audible tone when alarm conditions occur

Ranked causes

  1. 2. Door switch harness connection loose (occasional)
  2. 3. Solder connection failure at speaker (occasional)
  3. 4. Defective speaker (occasional)
  4. 5. Control PCA defective (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No SpO2 probe detected patient safety

SpO2 sensor not recognized when connected

Ranked causes

  1. 1. Improper probe mating with connector (common)
  2. 2. Dislodged SpO2 connector (common)
  3. 3. Reverse probe connection (occasional)
  4. 4. Defective SpO2 Flex Circuit (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

One observation light bulb does not illuminate

One of the two observation light bulbs fails to produce light.

Ranked causes

  1. 1. Bad or burned-out bulb (common)
  2. 2. Bad harness or bad socket (occasional)
  3. 3. Faulty bulb socket (occasional)
  4. 4. Bad harness connection at power PCA (rare)
First check

Test bulb continuity

Test point
Bulb terminals
Expected
<1 Ohm
Tools
Ohm Meter, Bladed screwdriver

3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Red screen with flashing alarm lights and HP alarm

Display shows red screen; alarm lights flash; HP alarm sounds

Ranked causes

  1. 1. NVRAM failure (common)
  2. 2. Defective or disconnected ribbon cable (occasional)
  3. 3. Control PCA failure (occasional)
  4. 4. Real Time Clock / NVRAM hardware failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale does not return a weight caution

Scale fails to display weight readings or produces consistently inaccurate weights

Ranked causes

  1. 1. Hardware issue or miscalibration (common)
  2. 2. Software malfunction (occasional)
  3. 3. Outdated scale software version (occasional)
  4. 4. Scale slow to return weight (occasional)
First check

Perform scale hardware check

Test point
Scale hardware sensors
Expected
All hardware components functional per procedure 3.2.5.1

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sys Fail 11 [AA AA AA AA] patient safety

Heat Control Processor communication failed due to Power-On Self Test failure

Ranked causes

  1. 1. Wrong mains voltage (occasional)
  2. 2. Open or loose heater AC fuse connection (common)
  3. 3. Open T-Stat at Power-on (occasional)
  4. 4. Defective/Disconnected cable (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sys Fail 11 [xx 00 00 00] or [FF FF FF FF] patient safety

Heat Control Processor stopped communicating with Main Application Processor

Ranked causes

  1. 1. Defective/Disconnected ribbon cable (common)
  2. 2. No +5V DC at TP20 pin 2 (common)
  3. 3. Invalid /Reset signal (common)
  4. 4. Open or faulty fuses F9 & F10 (common)
First check

Inspect ribbon cable seating

Test point
20 pin ribbon cable between Power and Control PCA
Expected
Fully seated
Tools
Visual inspection

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 11 (Sys Fail 11 [xx 00 00 00] or [FF FF FF FF]) caution

Unit displays Sys Fail 11; heat control processor stopped communicating with main application processor.

Ranked causes

  1. 1. Defective or disconnected cable (common)
  2. 2. Power PCA voltage failure (common)
  3. 3. Reset signal failure (common)
  4. 4. Open or loose heater AC fuse connection (occasional)
First check

Verify ribbon cable seating

Test point
20 pin ribbon between Power and Control PCA
Expected
Cable fully seated at both connectors

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temp Probe Failure alarm message patient safety

Temperature probe sensor not detected or communication failure with probe

Ranked causes

  1. 1. Incorrect probe connections (common)
  2. 2. Damaged probe rail harness (occasional)
  3. 3. Defective Probe Panel Interface Board (occasional)
  4. 4. Defective Control PCA (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature ADC channels out of specification

Temperature sensor readings or reference voltages exceed expected tolerances during diagnostics

Ranked causes

  1. 1. ADC channel 1 (Vthref) voltage out of range (unknown)
  2. 2. Patient probe thermistor mismatch (unknown)
  3. 3. Reference temperature (37°C) voltage out of range (unknown)
  4. 4. Reference 1P235 voltage out of range (unknown)
First check

Measure ADC channel 1 Vthref

Test point
ADC Channel 1
Expected
Raw Value 2048 (±4), Corrected Value 1500 (±3) mV raw, ±4

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not turn on (power fail alarm) high voltage

Unit does not power up; power fail alarm displayed or triggered; no response to power switch

Ranked causes

  1. 1. No 12V DC power supply output (common)
  2. 2. Blown fuse on Power PCA (F7 or F8) (occasional)
  3. 3. No AC voltage to Power PCA (front switch, harness, line filter) (common)
  4. 4. No signal from Heat Control Processor (rare)
First check

Measure 12V DC from chassis to top of fuse F7 on Power PCA

Test point
Power PCA fuse F7 (top side)
Expected
12V DC, nominal
Tools
DC Volt Meter

10 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backlight not working

Display backlight does not illuminate or is very dim

Ranked causes

  1. 1. Fuse F7 open on Power PCA (common)
  2. 2. Bad backlight connections (common)
  3. 3. Defective Control PCA or LCD module (occasional)
First check

Check fuse F7 resistance

Test point
Fuse F7 on Power PCA
Expected
< 1 Ohm if intact, open if blown
Tools
Ohm Meter, Needle nose pliers

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blank black screen with no alarms

Display is completely black with no alarm indications

Ranked causes

  1. 1. Loss of 12V to Control PCA (common)
  2. 2. LCD backlights not connected (common)
  3. 3. Fuse F7 failure on Power PCA (occasional)
First check

Check LED7 light status on Control PCA

Test point
LED7 on Control PCA
Expected
Red light indicates LCD receiving power

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Both bulbs do not light (LED) caution

Neither of the two observation LEDs illuminates when switched on

Ranked causes

  1. 1. Loose connector and wiring (common)
  2. 2. Bad dimmer pot switch (occasional)
  3. 3. Bad Power PCA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display not visible or unresponsive

LCD screen shows nothing or cannot be interacted with; backlight may not be illuminated.

Ranked causes

  1. 1. Protective film on LCD (common)
  2. 2. Bad LCD cable connection (common)
  3. 3. Backlight connection failure (occasional)
First check

Verify protective film removal

Test point
LCD surface
Expected
No film present

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

E-Base not moving caution

Elevating base does not respond to foot switch pedals; no vertical bed movement

Ranked causes

  1. 1. Defective E-Base unit (common)
  2. 2. No power to E-Base (common)
  3. 3. Blown fuse on Power PCA (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base no motion caution

Footswitch pedals do not raise or lower base; green light may or may not be present

Ranked causes

  1. 1. Defective E-Base unit (occasional)
  2. 2. No power to E-Base (common)
  3. 3. Loose footswitch connectors (common)
First check

Plug spare line cord directly into E-Base AC inlet and observe indicator light

Test point
E-Base AC inlet; green light indicator on E-Base column
Expected
Green light ON

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flashing alarm light; warmer will not initialize

Alarm light flashing and warmer fails to initialize properly

Ranked causes

  1. 1. DC Supply voltage issue (common)
  2. 3. Defective harness (occasional)
  3. 4. Defective Power PCA (occasional)
First check

DC supply voltage

Test point
Power PCA TP20 pins 3 and 22-3
Expected
at least 11.5V, >= 11.5

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hands Free Alarm Silence does not work

Hands free alarm silence feature is not responding to hand gestures in front of alarm panel

Ranked causes

  1. 2. Alarm/light PCA not programmed (occasional)
  2. 3. Function disabled or not enabled in software (occasional)
  3. 4. Open connection in ribbon cable (occasional)
First check

Verify device location and lighting

Test point
Ambient light conditions and light settings
Expected
Device not in direct sunlight; light settings not at maximum

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD backlight dim or non-uniform lighting

LCD backlight is dim or illumination is uneven across display

Ranked causes

  1. 1. Bad LCD backlight component (common)
  2. 2. Loss of 12V power to Control PCA (occasional)
  3. 3. LCD backlights not connected to Control PCA (occasional)
First check

Check backlight connector seating at Control PCA

Test point
J4, J8 on Control PCA
Expected
Connectors mated and pinned correctly, no bent pins

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No audible Power Fail Alarm high voltage

No audible alarm tone when unit loses AC power

Ranked causes

  1. 2. Ribbon cable disconnected or improperly seated (occasional)
  2. 3. Front panel switch connection loose (occasional)
  3. 4. Defective or depleted battery (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No scale icon displayed caution

Scale icon does not appear on display screen for bedded scale unit

Ranked causes

  1. 2. Defective cable (common)
  2. 3. Defective printed circuit board (occasional)
  3. 4. Defective scale receptacle on unit (occasional)
First check

Verify scale connection

Test point
Scale connector at probe panel
Expected
Scale properly connected and seated

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Potential patient harm from misconfigured alarms patient safety

Patient safety compromised if alarm limits not verified or alarms silenced inappropriately

Ranked causes

  1. 1. Alarm limits not verified for current patient (common)
  2. 2. Audible alarm function silenced or volume decreased inappropriately (occasional)
  3. 3. Alarm speaker blocked or obstructed (occasional)
First check

Verify alarm limits are set appropriately for patient

Test point
Pulse oximeter alarm limit settings
Expected
Low and high SpO2 alarm thresholds appropriate for patient age and clinical condition

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale icon disappears/reappears repeatedly caution

Scale icon on control panel intermittently disappears and reappears; scale makes repeated clicking noise

Ranked causes

  1. 1. Defective or disconnected scale cable (common)
  2. 2. Defective printed circuit board (occasional)
  3. 3. Defective scale receptacle on unit (occasional)
First check

Verify connector insertion

Test point
Scale connector at receptacle
Expected
Connector fully inserted into receptacle

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor disconnect error or alarm patient safety

Sensor disconnect error message displayed and associated alarm triggered

Ranked causes

  1. 1. Sensor disconnected from patient or control unit (common)
  2. 2. Faulty wiring or loose connector (occasional)
  3. 3. Damaged sensor (occasional)
First check

Verify sensor physical connection to control unit

Test point
Sensor connector port on control panel
Expected
Connector fully seated and secure

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor disconnect or no signal patient safety

Pulse oximeter displays sensor disconnect error message and associated alarm; no reading obtained

Ranked causes

  1. 1. Sensor disconnected from module (common)
  2. 2. Faulty wiring or cable damage (occasional)
  3. 3. Defective sensor (occasional)
First check

Verify sensor connector is fully seated

Test point
Sensor port on Panda iRes or Giraffe control panel
Expected
Connector fully inserted; audible click heard

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 readings affected by environmental or application conditions patient safety

Pulse oximetry readings may be inaccurate due to sensor application errors, ambient environmental conditions, or patient conditions affecting measurement.

Ranked causes

  1. 1. Sensor applied incorrectly or placed on unsuitable site (common)
  2. 2. Ambient environmental conditions interfering with optical measurement (common)
  3. 3. Patient motion or unstable perfusion (occasional)
First check

Ensure sensor is applied to pulsating arteriolar vascular bed

Test point
Sensor placement site
Expected
Finger, toe, or other appropriate pulsating arteriolar site

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sys Fail 03

Memory error occurred in NVRAM or control PCA

Ranked causes

  1. 1. NVRAM chip cover not properly seated (common)
  2. 2. Corrupted application flash (occasional)
  3. 3. Control PCA failure (occasional)
First check

Inspect NVRAM chip cover

Test point
NVRAM chip on control PCA
Expected
Cover fully and properly seated; battery retained

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sys Fail 06 [0B XX 00 00] patient safety

Over temperature thermostat is open. Added in MAP Version 1.1.0 and later.

Ranked causes

  1. 1. Open over-temperature thermostat (common)
  2. 2. Defective thermostat or harness (common)
  3. 3. Defective power PCA (occasional)
First check

Inspect heater vents for obstruction

Test point
Heater head cooling vents and heater duct
Expected
No obstructions present

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sys Fail 06 [0B XX 00 00] patient safety

Over temp T-Stat open condition detected

Ranked causes

  1. 1. Open overtemp T-Stat (common)
  2. 2. Defective T-Stat or Harness (common)
  3. 3. Defective Power PCA (occasional)
First check

Inspect for heater blockage

Test point
Heater duct and heater head vents
Expected
No obstructions visible

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sys Fail 06 overtemp T-Stat open caution

Sys Fail 06 error, over temperature T-Stat open circuit detected

Ranked causes

  1. 1. Open overtemp T-Stat (common)
  2. 2. Defective T-Stat or harness (occasional)
  3. 3. Defective Power PCA (occasional)
First check

Check heater duct and vents for blockage

Test point
Heater duct, heater head vents
Expected
Clear passage with no obstruction

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Sys Fail 09 [XX 80 XX XX] / [XX 10 XX XX] / [XX 50 XX XX] / [0A XX 00 00]
  • Sys Fail 11 [AA AA AA AA]
  • System Failure 02 (Sys Fail 02)
  • System Failure 06 (Sys Fail 06 [0B XX 00 00])
  • System Failure 08 (Sys Fail 08 [xx xx 00 00])
  • Unit will not turn on
  • Alarm message: Confirm Probe Jack Connection
  • Alarm sounds inaudible or suppressed
  • Bed does not tilt
  • Bed does not tilt
  • Button not working
  • Button not working
  • Calibration: Heater head alignment
  • Cannot enter Service Mode
  • Check SpO2 probe quality
  • Damaged sensor or cable detected
  • Device exposure to moisture or extreme conditions
  • Dim or non-uniform light
  • HR missing and scale missing since startup
  • Inadequate patient cable or sensor integrity
  • LCD display blank black screen with no alarms
  • LCD display blank black screen with no alarms
  • LCD display not visible or has arrow after replacement
  • Light output dim at highest LED observation light setting
  • Light output dim at highest setting (Halogen)
  • Light output flickers at lowest setting (Halogen)
  • Lights are not dimmable (Halogen)
  • Lights are not dimmable (LED)
  • Lights stay dim with switch off (Halogen)
  • Lights stay ON with observation knob switched off (LED)
  • Lithium-ion battery installation error
  • No POST pass tone heard at startup
  • No SpO2 probe detected
  • Observation light dim at highest brightness setting
  • Observation light dim at highest setting
  • Observation light flickers at lowest setting
  • Observation light flickers at lowest setting
  • Observation light output dim at highest setting
  • Observation light output flickers at lowest setting
  • Observation lights dim at highest setting
  • Observation lights flicker at lowest setting
  • Observation lights not dimmable
  • Observation lights not dimmable
  • Observation lights stay dim with switch off
  • Observation lights stay dim with switch off
  • One observation bulb does not light
  • Oximetry system failure
  • Sensor tissue damage from incorrect application or prolonged use
  • SpO2 reading unusually high or low despite normal clinical appearance
  • Sys Fail 01 [FF xx xx xx]
  • Sys Fail 01 [FF xx xx xx]
  • Sys Fail 01 with flash memory error
  • Sys Fail 02
  • Sys Fail 02 power mismatch
  • Sys Fail 03 memory error
  • Sys Fail 04 memory redundancy error
  • Sys Fail 05 ADC configuration error
  • Sys Fail 08 [xx xx 00 00]
  • Sys Fail 08 [xx xx 00 00]
  • Sys Fail 08 with button shorted
  • Sys Fail 09 [0A XX 00 00]
  • Sys Fail 09 heater safety relay open
  • Sys Fail 11 [AA AA AA AA] alarm
  • Sys Fail 11 [xx 00 00 00] or [FF FF FF FF]
  • System Fail 11 - Heat Control Processor communication failure
  • System Failure 01 (Sys Fail 01 [FF xx xx xx])
  • System Failure 03 (Sys Fail 03)
  • System Failure 04 (Sys Fail 04)
  • System Failure 09 (Sys Fail 09 [0A XX 00 00])
  • System Failure 11 (Sys Fail 11 [AA AA AA AA])
  • System failure detected in event logs
  • Temp probe failure alarm
  • Touch panel button not responding
  • Touch panel button not working
  • Touch panel buttons or switches not responding
  • Touch panel or LCD assembly removal
  • White display, striped display, or color shift
  • 25% Output Check reading incorrect
  • 25% output test operation
  • 25% output test shows incorrect reading
  • Access service mode from powered-on device
  • Alarm limits inappropriate for patient
  • Alarm message: Temp Probe Failure
  • Baby Cold - Check Temp Probe
  • Baby Cold - Check Temp Probe (1.0°C threshold)
  • Baby Hot - Check Temp Probe
  • Baby Hot - Check Temp Probe (1.0°C threshold)
  • Base and rail mounting block replacement (freestanding)
  • Bed height adjustment on crank handle system
  • Bed height adjustment on manual pin system
  • Black background with red, green, or blue tint
  • Blank screen with alarm and indicator lights on
  • Blank screen with alarm lights and HP alarm
  • Blank screen with alarm lights on
  • Blank screen with alarm lights on and HP alarm
  • Blank Screen, Alarm Lights On, HP Alarm
  • Calibration: Date and time not synchronized
  • Calibration: Heater head alignment
  • Calibration: Scale
  • Calibration: Scale calibration required
  • Calibration: Scale weight
  • Calibration: Scale weight accuracy
  • Calibration: Scale weight measurement
  • Calibration: Scale weight measurement
  • Callback Timer Complete
  • Cannot enter service mode
  • Caster replacement needed
  • Check Baby - Heat Off (CPAD limit exceeded)
  • Check Baby - Heat Off (radiant power 100%)
  • Check Baby (CPAD limit exceeded)
  • Check Baby (radiant power 100%)
  • Check SpO2 - defective probe
  • Check SpO2 - defective probe indicated
  • Check SpO2 - excessive ambient light
  • Check SpO2 - interference detected
  • Control board removal
  • Control board replacement needed
  • Damaged equipment from sterilization method
  • Defective scale
  • Defective scale
  • Defective scale (Bedded Models Only)
  • Defective SpO2 probe
  • Defibrillation or electrosurgery interference
  • Device will not boot up
  • Dim or non-uniform LCD backlight
  • Disconnected temperature probe
  • Display arrow present after installing new LCD
  • Display arrow present after installing new LCD
  • Display has pixel defects or color errors
  • Display shows abnormal colors or incomplete test pattern
  • Display test pattern abnormal
  • Display test pattern shows dropout or color errors
  • Distorted audio from speaker
  • Distorted audio from speaker
  • Distorted audio speaker
  • Distorted display: test or random pattern
  • Dovetail rail suction tubing replacement
  • Electronic enclosure access restricted or damaged
  • EU compliance verification required before operation
  • F7 or F8 fuse failure or loose connection
  • F7 or F8 fuse replacement needed
  • Firmware update execution
  • Fuse F7 or F8 damaged or loose connection
  • Hands-free alarm silence not responding
  • Heat engine diagnostic fails or shows abnormal state
  • Heat engine diagnostic parameters query
  • Heat engine diagnostics unavailable
  • Heat engine health check perform
  • Heat engine output not at specified level
  • Heat engine output out of range
  • Heater resistance measurement
  • High Pulse Rate
  • High SpO2
  • High SpO2 reading
  • HR active and scale missing
  • HR subsystem failure and scale active
  • HR subsystem failure and scale missing
  • Inadequate pulse oximetry as apnea monitor
  • Incorrect cable connected to sensor port
  • LCD display color test pattern abnormal
  • LCD display dim or non-uniform lighting
  • LCD display distorted with test or random pattern
  • LCD display distorted with test pattern or random pixels
  • LCD display shows distorted test or random pattern
  • LCD display shows white, striped, or color shifted image
  • LCD display white, striped, or color shift
  • LCD display white, striped, or color shifted
  • LCD display white, striped, or color shifted
  • Lead off detected
  • Lights stay ON with observation light knob switched off
  • Lithium-ion battery replacement
  • Lithium-ion battery replacement needed
  • Log dump fails or times out
  • Log dump transfer timeout
  • Low Perfusion (SpO2 Alert)
  • Low perfusion detected at probe site
  • Low pulse rate
  • Low Signal IQ (SpO2 Alert)
  • Low Signal IQ alert
  • Low SpO2
  • Low SpO2 reading
  • Maneuvering handle replacement
  • MRI procedure with pulse oximeter attached
  • No audible power fail alarm
  • No audio indication of alarms
  • No scale icon displayed
  • No SpO2 Board
  • No SpO2 option on screen
  • No SpO2 option on screen
  • No SpO2 probe connected
  • Observation light dimmer control failure
  • Observation light switch or dimmer control replacement
  • Outlet panel component failure
  • Outlet panel component replacement
  • Patient cable entanglement or strangulation hazard
  • PC command line interface access failure
  • Potential patient harm from carboxyhemoglobin interference
  • Potential skin damage or burns from MRI use
  • Power board removal
  • Power board replacement needed
  • Procedure light bulb burned out or dim
  • Procedure light bulb replacement
  • Pulse oximeter not functioning after power-on self-test
  • Readings potentially inaccurate during defibrillation
  • Red screen with flashing alarm and indicator lights
  • Red screen with flashing alarm lights
  • Red screen, alarm lights flashing
  • Red screen, alarm lights flashing
  • Scale calibration failed
  • Scale control board replacement
  • Scale gives consistently inaccurate weights
  • Scale gives erratically different weights
  • Scale load cell replacement
  • Scale stuck on weight display screen
  • Sensor or cable damage from water or chemical immersion
  • SpO2 display shows no signal
  • SpO2 interference detected
  • SpO2 lost communications
  • SpO2 Probe Off Baby
  • SpO2 probe off patient
  • SpO2 system fails to calibrate
  • Standby switch failure
  • Standby switch replacement
  • Sys Fail 02
  • Sys Fail 03
  • Sys Fail 04
  • Sys Fail 05
  • Sys Fail 05
  • Sys Fail 09 [0A XX 00 00]
  • Sys Fail 09 [0B 00 00 00]
  • Sys Fail 09 [0B 00 00 00]
  • Sys Fail 09 [0E 72 00 00]
  • Sys Fail 09 [XX 80 XX XX]
  • Sys Fail 09 [XX 80 XX XX]
  • Sys Fail 09 heat control processor failure during runtime
  • Sys Fail 10 [XX XX XX XX]
  • Sys Fail 11 [FF FF FF FF] alarm
  • Sys Fail 12 [xx 00 00 00]
  • Sys Fail 12 [xx 00 00 00] alarm
  • System Fail 12 - Audio Processor communication failure
  • System Failure 05 (Sys Fail 05)
  • System Failure 09 (Sys Fail 09 [0B 00 00 00])
  • System Failure 09 (Sys Fail 09 [XX 80 XX XX])
  • System Failure 10 (Sys Fail 10 [XX XX XX XX])
  • System Failure 11 (Sys Fail 11 [FF FF FF FF])
  • System Failure 12 (Sys Fail 12 [xx 00 00 00])
  • System log retrieval and analysis
  • Temperature ADC channel readings out of range
  • Temperature ADC readings out of range
  • Temperature probe failure
  • Too much ambient light (SpO2)
  • Touch panel or LCD display malfunction
  • Touch panel or LCD display removal difficulty
  • Unit will not boot up
  • Unit will not boot up
  • Unrecognized SpO2 probe
  • Unrecognized SpO2 probe attached
  • VGA display test pattern defects

258 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Baby Check SpO2 (Ambient Light)Too much ambient light detected by SpO2 boardMasimo or Nellcor board indicates too much ambient lighterror
Baby Cold - Check Temp ProbeBaby temperature more than 2.0°C below control setpoint>2.0° from Control Temp (below)warning
Baby Cold - Check Temp ProbeBaby temperature more than 2.0°C below control temperature<2.0° from Control Tempwarning
Baby Cold - Check Temp Probe (Manual Mode)Baby temperature more than 1.0°C below control temperature in Manual mode< 1.0°C (<0.5°C) from control tempwarning
Baby Hot - Check Temp ProbeBaby temperature more than 2.0°C above control setpoint>2.0° from Control Tempwarning
Baby Hot - Check Temp ProbeBaby temperature more than 2.0°C above control temperature>2.0° from Control Tempwarning
Baby Hot - Check Temp Probe (Manual Mode)Baby temperature more than 1.0°C above control temperature in Manual mode> 1.0°C (>0.5°C) from control tempwarning
Callback Timer CompleteCallback timer has reached target timeCurrent callback timer has reached target timeinfo
Callback Timer CompleteScheduled callback timer has expiredCurrent callback timer has reached target timeinfo
Check BabyRadiant heater power above CPAD limit for extended durationRadiant Heater % above 'Check Baby Alarm Disabled' (CPAD) limit for > 12 minwarning
Check BabyRadiant heater at 100% power for more than 12 minutesRadiant % power @ 100% for > 12 minwarning
Check Baby (CPAD Limit)Radiant heater above Check Baby Alarm Disabled limit for extended timeRadiant Heater % above CPAD limit for > 12 minwarning
Check Baby - Heat OffRadiant heater at excessive power for extended timeRadiant % power @ 100% for >15 min and alarm not silencedwarning
Check Baby - Heat OffRadiant heater at 100% power for extended time without alarm silencingRadiant % power @ 100% for >15 min and alarm not silencederror
Check Baby - Heat Off (CPAD)Radiant heater above Check Baby Alarm Disabled limit for extended timeRadiant heater % above CPAD limit for >15 min and alarm not silencederror
Check Baby (Radiant Power)Radiant heater at 100% power for extended time in temperature controlRadiant % power @ 100% for > 12 minwarning
Check SpO2Defective SpO2 probe detectedMasimo or Nellcor board indicates defective probewarning
Check SpO2Defective SpO2 probe, interference, or too much ambient light detectedMasimo or Nellcor board indicates defect, interference, or light issuewarning
Check SpO2SpO2 signal quality is poor or probe is defectiveSystem detects defective probe, electromagnetic interference, or excessive ambient lighterror
Check SpO2SpO2 measurement interference detectedMasimo or Nellcor board indicates interference detectedwarning
Check SpO2Probe defect detected by Masimo boardMasimo board diagnostic indicates defective probe after 2 minuteserror
Check SpO2 (Defective Probe)SpO2 probe detected as defective by boardMasimo or Nellcor board indicates defective probeerror
Check SpO2 (Interference)Interference detected by SpO2 boardMasimo or Nellcor board indicates interference detectederror
Confirm Probe Jack ConnectionDisconnected temperature probeNo longer getting a reading from the temperature probeerror
Confirm Probe Jack ConnectionTemperature probe disconnected or no reading receivedNo longer getting a reading from the temperature probeerror
Confirm Probe Jack ConnectionNo reading from temperature probe detectedNo longer getting a reading from the temperature probeerror
High Pulse RatePatient pulse rate exceeds alarm limitPulse rate exceeds set high alarm threshold (default 200 bpm, user adjustable 111-240 bpm)warning
High Pulse RatePulse rate above thresholdPR >200 (floats with user adjustable limits)warning
High SpO2SpO2 measurement exceeds upper alarm limit (default >100%)Measured SpO2 exceeds user-adjustable upper threshold for 2 minuteswarning
High SpO2SpO2 > 100 (floats with user adjustable limits)SpO2 > 100 (floats with user adjustable limits)warning
High SpO2Oxygen saturation above thresholdSpO2 >100 (floats with user adjustable limits)critical
High SpO2Patient blood oxygen saturation exceeds alarm limitSpO2 percentage exceeds set high alarm threshold (default 100%, user adjustable 86-100%)info
Low PerfusionMasimo board detects low blood perfusion at measurement site (SpO2 alert)Masimo board indicates low perfusionwarning
Low PerfusionLow perfusion detected by SpO2 boardMasimo board indicates low perfusionwarning
Low Pulse RatePulse rate below thresholdPR < 100 (user adjustable limits)error
Low Pulse RatePulse rate below 100 BPM (user adjustable)PR < 100 (floats with user adjustable limits)warning
Low Pulse RatePatient pulse rate falls below alarm limitPulse rate drops below set low alarm threshold (default 100 bpm, user adjustable 25-110 bpm)warning
Low Signal IQMasimo board detects poor SpO2 signal quality (SpO2 alert)Masimo board indicates low signal IQwarning
Low Signal IQSystem confidence in detected arterial pulse signal is very lowSystem cannot reliably identify arterial pulse due to motion or low perfusionwarning
Low Signal IQLow signal quality detected by SpO2 boardMasimo board indicates low signal IQwarning
Low SpO2Patient blood oxygen saturation falls below alarm limitSpO2 percentage drops below set low alarm threshold (default 85%, user adjustable 20-99%)critical
Low SpO2SpO2 < 85 (floats with user adjustable limits)SpO2 < 85 (floats with user adjustable limits)warning
Low SpO2Oxygen saturation below thresholdSpO2 <85 (user adjustable limits)error
Low SpO2SpO2 measurement falls below lower alarm limit (default <85%)Measured SpO2 below user-adjustable lower threshold for 2 minutescritical
No SpO2 BoardMasimo or Nellcor board not detected; only available if SpO2 runningNo board connectederror
No SpO2 BoardNo SpO2 probe or module connectedMasimo or Nellcor board indicates no board connectedwarning
No SpO2 BoardSpO2 board not detected as connectedMasimo or Nellcor board indicates no board connectederror
No SpO2 ProbeNellcor board indicates no probe connectionNellcor board indicates no probe connectionerror
No SpO2 ProbeSpO2 probe is not detected as connectedProbe jack detects no connected probe at startup or during monitoringwarning
Oximetry System FailureOximetry system has failed, such as failure to calibrate successfullySystem failure event during operation or startuperror
Oximetry System FailureSpO2 board lost communication with main processor for 2 seconds or more; SpO2 module automatically turns offSpO2 board has not communicated with the main processor for at least 2 secondscritical
Oximetry System FailureSpO2 module has failed or lost communication with warmer systemSystem startup or during operation when calibration fails or communications lostcritical
Oximetry System FailureSpO2 board has not communicated with the main processor for at least 2 secondsLoss of communication between SpO2 board and main processor for ≥2 secondscritical
Processor update failed exit statusFirmware update to audio or heat control processor did not complete successfullyOccurs when 'update' command fails after processor re-programming has commencedcritical
SpO2 lost communicationsSpO2 board has not communicated with the main processor for at least 2 secondsSpO2 board communication failurecritical
SpO2 Probe OffSpO2 probe indicated as off babyProbe off baby is indicated from the Masimo or Nellcor boarderror
SpO2 Probe Off BabyProbe is not in contact with patientProbe loses contact with patient skin or is not properly attachedwarning
SpO2 Probe Off BabyProbe has been removed from babyProbe off baby is indicated from the Masimo or Nellcor boardwarning
SpO2 Probe Off BabySpO2 probe detected as displaced from patientProbe off baby is indicated from the Masimo or Nellcor boardwarning
SpO2 Probe Off BabySpO2 probe no longer in contact with patientProbe off baby is indicated from the Masimo or Nellcor boardwarning
Sys Fail 01 [FF xx xx xx]FLASH Memory File System ErrorFlash memory file system corrupted or inaccessiblecritical
Sys Fail 01 [FF xx xx xx]FLASH memory file system errorMemory check during power-on detects flash corruption or invalid file systemerror
Sys Fail 02Reported power from power PCA does not match commanded power from control PCAPower feedback mismatch during operationerror
Sys Fail 02Power mismatch between Power PCA and Control PCAReported power from Power PCA does not match commanded power from Control PCAerror
Sys Fail 03Memory error occurredSystem detected memory error during operationerror
Sys Fail 03Memory error occurredNVRAM or memory validation check fails during operationerror
Sys Fail 04Redundant stored variables do not matchMemory redundancy check detects mismatch in stored dataerror
Sys Fail 04Redundant stored variables do not match (Memory Redundancy Error)System detected mismatch in redundant memory variableserror
Sys Fail 05ADC Configuration Error or ADC Failure; Reference channel outside range (1597-1775)ADC reference channel reading outside specified range 1597-1775error
Sys Fail 05ADC configuration error or ADC failure; reference channel outside of range (1597-1775)ADC reference check at power-on detects out-of-range valueerror
Sys Fail 06 [0B XX 00 00]Over temperature thermostat is openT-Stat contact does not close when heated or temperature exceeds safe limitscritical
Sys Fail 06 [0B XX 00 00]System Failure 06 with specified auxiliary codeProcessor communication or hardware fault detectederror
Sys Fail 06 [0B XX 00 00]Overtemp T-Stat is openOvertemperature thermostat detected open circuit during operationcritical
Sys Fail 08 [xx xx 00 00]Touch panel button shorted or single button pressed during power onShorted button detected on touch panel during startup or operationerror
Sys Fail 08 [xx xx 00 00]Touch panel button shortedContinuous button input detected during operationwarning
Sys Fail 08 [xx xx 00 00]System Failure 08 with variable auxiliary codesSensor or board communication errorerror
Sys Fail 09 [0A XX 00 00]System Failure 09 with first auxiliary byte 0x0ASpecific processor or peripheral faulterror
Sys Fail 09 [0A XX 00 00]Heat control processor failure during run time; Heater Safety Relay is openSafety relay in heater control circuit detected opencritical
Sys Fail 09 [0B 00 00 00]Heat control processor failure during run time; Open T-StatThermostat detected open circuit during heat control operationcritical
Sys Fail 09 [0B 00 00 00]Heat control processor failure during runtime; open T-StatOver-temperature thermostat remains open during heat operationcritical
Sys Fail 09 [0B 00 00 00]System Failure 09 with first auxiliary byte 0x0BHeat control processor communication failureerror
Sys Fail 09 [0E 72 00 00]Heat control processor failure during startup; disconnected heater cableHeater cable not connected or loose at power-oncritical
Sys Fail 09 [XX 80 XX XX]System Failure 09 with first auxiliary byte 0x80Power or alarm board faulterror
Sys Fail 09 [XX 80 XX XX]Heat control processor failure during run time; Heater Control FailureHeat control processor detected failure during normal operationcritical
Sys Fail 09 [XX 80 XX XX] / [XX 10 XX XX] / [XX 50 XX XX] /Heat control processor failure during runtime; heater control failureHeat control processor communication or operation failure during heating cyclecritical
Sys Fail 10 [XX XX XX XX]Self-calibration failure occurredSelf-calibration routine fails during power-up or initializationerror
Sys Fail 10 [XX XX XX XX]System Failure 10 with variable auxiliary codesMultiple processor or board faultserror
Sys Fail 10 [XX XX XX XX]Self-calibration failure occurredUnit failed to complete self-calibration during startup or diagnosticserror
Sys Fail 11 [AA AA AA AA]Heat Control Processor stopped communicating due to POST (Power-On Self Test) failurePower PCA POST failed during power-on sequencecritical
Sys Fail 11 [AA AA AA AA]System Failure 11 with alternating pattern auxiliary codeHeat control processor communication with specific fault patterncritical
Sys Fail 11 [AA AA AA AA]No communication of heat control processor with main processor since power-onHeat control processor not responding since device powered onerror
Sys Fail 11 [AA AA AA AA]Heat Control Processor communication failure due to POST failurePower-on self-test failureerror
Sys Fail 11 [FF FF FF FF]Heat control processor not responding; communication completely lostHeat control processor power loss or complete failurecritical
Sys Fail 11 [FF FF FF FF]Communication message failedInter-processor communication errorerror
Sys Fail 11 [FF FF FF FF]Heat Control Processor communication message failedCommunication message error between processors detectederror
Sys Fail 11 [xx 00 00 00]Heat control processor lost communication with main application processorHeat control processor stopped responding or communication link failedcritical
Sys Fail 11 [xx 00 00 00] -or- [FF FF FF FF]Heat control processor stopped communicating with main processor; communication lossCommunication timeout between heat control and main processor during operationerror
Sys Fail 11 [xx 00 00 00] or [FF FF FF FF]Heat Control Processor stopped communicating with Main Application Processor (xx = seconds since last communication)Communication loss between heat control and main processor during operationcritical
Sys Fail 12 [xx 00 00 00]Audio Processor stopped communicating with Main Application Processor (xx = seconds since last communication)Communication loss between audio processor and main processorwarning
Sys Fail 12 [xx 00 00 00]Audio Processor stopped communicating with Main Application ProcessorAudio processor communication timeouterror
Sys Fail 12 [xx 00 00 00]System Failure 12 with specified auxiliary codePower or control board processor faulterror
Temp. Probe FailureTwo thermistors in probe differ by 0.5°C or moreTwo Thermistors in a probe differ by 0.5° C or moreerror
Temp. Probe FailureTwo thermistors in probe differ by 0.5°C or moreTwo Thermistors in a probed differ by 0.5° C or moreerror
Unrecognized SpO2 ProbeNellcor board indicates unrecognized probeNellcor board indicates unrecognized probeerror
Unrecognized SpO2 ProbeProbe model not recognized by Masimo boardMasimo board fails to identify probe model after 2 minutes; audio progresses 2 to 1 after 10 secserror
Unrecognized SpO2 ProbeUnrecognized SpO2 probe type connectedUnrecognized probe indicated from Masimo or Nellcor boardwarning
Unrecognized SpO2 ProbeAttached probe is not compatible with this warmer systemSystem detects probe with incompatible connector or incorrect modelerror
Unrecognized SpO2 ProbeSpO2 probe not recognized by boardUnrecognized probe indicated from Masimo or Nellcor boarderror
xmodem error: -69Xmodem file transfer protocol error during log dumpOccurs when log dump command times out or communication is interrupted during xmodem transfererror
xmodem error: -69Log dump xmodem transfer failedDuring logs dump file transfer via xmodemerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI