Datascope Passport 2: troubleshooting and error codes

The ServiceForge knowledge base holds 2 documented symptoms for the Datascope Passport 2, with 2 ranked causes, 4 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Overpressure transducer zero point

Overpressure transducer zero point verification and adjustment to ensure proper NIBP measurement accuracy

Ranked causes

  1. 1. Overpressure transducer zero point out of specification (unknown)
First check

Verify overpressure zero point on screen

Test point
Diagnostics menu, Overpressure Voltage display
Expected
0.100 volts ± 0.010 V, ±0.010 V

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Static pressure transducer

Static pressure calibration verification and adjustment to ensure pressure transducer sensitivity and accuracy for NIBP measurements

Ranked causes

  1. 1. Static pressure reading does not match reference manometer (unknown)
First check

Connect test chamber and enter static pressure calibration test

Test point
Side panel fitting, Diagnostics menu
Expected
Static pressure calibration menu selected and ready

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI