Datascope Passport 2: troubleshooting and error codes
The ServiceForge knowledge base holds 2 documented symptoms for the Datascope Passport 2, with 2 ranked causes, 4 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Overpressure transducer zero point
Overpressure transducer zero point verification and adjustment to ensure proper NIBP measurement accuracy
Ranked causes
- 1. Overpressure transducer zero point out of specification (unknown)
Verify overpressure zero point on screen
- Test point
- Diagnostics menu, Overpressure Voltage display
- Expected
- 0.100 volts ± 0.010 V, ±0.010 V
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Static pressure transducer
Static pressure calibration verification and adjustment to ensure pressure transducer sensitivity and accuracy for NIBP measurements
Ranked causes
- 1. Static pressure reading does not match reference manometer (unknown)
Connect test chamber and enter static pressure calibration test
- Test point
- Side panel fitting, Diagnostics menu
- Expected
- Static pressure calibration menu selected and ready
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI