Critikon PRO 1000V3: troubleshooting and error codes
The ServiceForge knowledge base holds 33 documented symptoms for the Critikon PRO 1000V3, with 87 ranked causes, 34 diagnostic checks with test points and expected values, and 158 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit will not power on caution
Equipment fails to power on when AC power applied and power switch pressed
Ranked causes
- 1. No AC power available (common)
- 2. Faulty power supply (common)
- 3. Faulty PSU power board (occasional)
- 4. Faulty main board (occasional)
- 5. Faulty power cable (occasional)
- 6. Faulty power switch (occasional)
Verify AC power availability at outlet
- Expected
- AC voltage present
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not power on
Device fails to turn on or power up
Ranked causes
- 1. No AC power to unit (common)
- 2. Faulty power supply (occasional)
- 3. Faulty PSU PWA (occasional)
- 4. Faulty main board (occasional)
- 5. Faulty power cables (occasional)
- 6. Faulty power On/Off front panel switch (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powers on but no display caution
Equipment powers up with signs of operation but display screen blank or dark
Ranked causes
- 1. Faulty display (common)
- 2. Faulty backlight driver (occasional)
- 3. Faulty main board (occasional)
- 4. Faulty PSU power board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powers on but no display
Device powers up but display shows nothing
Ranked causes
- 1. Faulty display (occasional)
- 2. Faulty backlight driver (occasional)
- 3. Faulty main board (occasional)
- 4. Faulty PSU PWA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform NIBP function patient safety
Non-invasive blood pressure measurement not working
Ranked causes
- 1. Faulty main board (occasional)
- 2. Faulty pneumatics assembly (occasional)
- 3. Faulty pneumatics cable (occasional)
- 4. Faulty front panel switch (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform NIBP function patient safety
Non-invasive blood pressure monitoring function fails to operate
Ranked causes
- 1. Faulty main board (common)
- 2. Faulty pneumatics assembly (common)
- 3. Faulty pneumatics cable (occasional)
- 4. Faulty front panel switch (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not print
Printer function fails to produce output or paper issues present
Ranked causes
- 1. No paper or incorrect paper loading (common)
- 2. Faulty printer (common)
- 3. Faulty main board (occasional)
- 4. Faulty PSU power board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not print
Printer module not printing documents
Ranked causes
- 1. Unit out of paper or paper incorrectly installed (common)
- 2. Faulty printer (occasional)
- 3. Faulty main board (occasional)
- 4. Faulty PSU PWA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not operate on battery caution
Equipment powered by battery backup fails to operate or battery does not charge
Ranked causes
- 1. Battery disconnect fuse open or unplugged (common)
- 2. Faulty battery pack (common)
- 3. Faulty PSU power board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not operate on battery
Device cannot run on battery power
Ranked causes
- 1. Battery disconnect fuse open or not plugged in (common)
- 2. Faulty battery pack (occasional)
- 3. Faulty PSU PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform ECG function patient safety
Electrocardiogram monitoring function fails to operate
Ranked causes
- 1. Faulty ECG PWA assembly (common)
- 2. Faulty ECG cable assembly (common)
- 3. Faulty main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform ECG function patient safety
Electrocardiogram measurement not working
Ranked causes
- 1. Faulty ECG PWA assembly (occasional)
- 2. Faulty ECG cable assembly (occasional)
- 3. Faulty main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform SPO2 function patient safety
Pulse oximetry (SpO2) monitoring function fails to operate
Ranked causes
- 1. Faulty SPO2 PWA (Nellcor) (common)
- 2. Faulty SPO2 PWA (Masimo) (common)
- 3. Faulty main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform Temperature function patient safety
Temperature monitoring via probe fails to operate
Ranked causes
- 1. Faulty temperature probe (common)
- 2. Faulty temperature probe sensor (common)
- 3. Faulty main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform Temperature function patient safety
Temperature measurement not working
Ranked causes
- 1. Faulty temperature probe (occasional)
- 2. Faulty temperature probe sensor (occasional)
- 3. Faulty main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder print quality and speed
Verification that recorder prints waveforms and text at correct speed with even tone, proper pixel density, and legible spacing.
Ranked causes
- 1. Recorder paper misalignment or print head contamination (occasional)
- 2. Worn or defective recorder mechanism (rare)
Verify paper is loaded in recorder
- Test point
- Recorder Parameter paper compartment
- Expected
- Paper properly loaded and ready for printing
- Tools
- Recorder paper
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit Host Comms not functional
Host communications interface fails to operate
Ranked causes
- 1. Faulty main board (common)
- 2. Faulty PSU power board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit host communications not functional
External device communication or networking not working
Ranked causes
- 1. Faulty main board (occasional)
- 2. Faulty PSU PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not generate sound
Speaker or audio output not functioning
Ranked causes
- 1. Faulty speaker assembly (occasional)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not generate sound
Audio alarm or notification sounds fail to play
Ranked causes
- 1. Faulty speaker assembly (common)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not operate on external DC power caution
Equipment fails to operate when powered from external DC source
Ranked causes
- 1. Faulty DC input jack or cable (common)
- 2. Faulty PSU power board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not operate on external DC power
Device cannot run on external DC power input
Ranked causes
- 1. Faulty DC input jack or DC cable (occasional)
- 2. Faulty PSU PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not perform SpO2 function patient safety
Pulse oximetry (SpO2) measurement not working
Ranked causes
- 1. Faulty SpO2 PWA assembly (occasional)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not produce analog ECG waveform output patient safety
Defibrillator interface or ECG output not functional
Ranked causes
- 1. Faulty Defib cable assembly (occasional)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not produce analog ECG waveform output patient safety
Equipment fails to output ECG signal to external defibrillator or devices
Ranked causes
- 1. Faulty Defib cable assembly (common)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not respond to rotary knob
Rotary encoder control not functioning
Ranked causes
- 1. Faulty encoder (occasional)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not respond to rotary knob
Rotary encoder control on unit does not respond to user input
Ranked causes
- 1. Faulty encoder (common)
- 2. Faulty main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG signal detection and waveform display
Verification of ECG simulator signal acquisition, waveform display, paced marker detection, and alarm functions across heart rate range.
Ranked causes
- 1. ECG lead or connector fault (occasional)
Verify ECG waveform display at 80 BPM
- Test point
- LCD display waveform and HR indication
- Expected
- ECG waveform displayed on LCD
- Tools
- ECG simulator set to 80 BPM Paced, 1.0 mV amplitude
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP leak test caution
Pneumatic system pressure loss rate measurement to verify system integrity before NIBP calibration checks.
Ranked causes
- 1. Cuff or hose defective or leaking (common)
Run automated leak test from Service Menu
- Test point
- NIBP Parameter connection
- Expected
- Leak Test Status displays Passed, Leak Test Results value less than 6, Service Error: None
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP overpressure protection caution
Verification that pump automatically shuts down and releases pressure at safe limits for adult and neonatal cuff settings.
Ranked causes
- 1. Overpressure valve out of tolerance (occasional)
Verify adult overpressure threshold
- Test point
- Manometer pressure reading when pump shuts down
- Expected
- Pump shuts down and pressure released at 300-330 mmHg, 300-330 mmHg
- Tools
- Manometer, calibration kit 320-246
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pressure measurement accuracy caution
Verification that PT1 and PT2 pressure transducers read within tolerance at multiple reference pressures to confirm system linearity and accuracy.
Ranked causes
- 1. Pressure transducers out of calibration (occasional)
Verify PT1 and PT2 pressure at 200 mmHg reference
- Test point
- Service Menu PT1 Pressure and PT2 Pressure display
- Expected
- PT1 and PT2 Pressures within 1 mmHg of manometer reading, ±1 mmHg
- Tools
- Manometer, calibration kit 320-246
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Respiration waveform verification
Verification that respiration waveform and rate readings display correctly at simulated respiration rates.
Ranked causes
- 1. Respiration sensor or signal processing fault (rare)
Verify respiration waveform at 20 BrPM
- Test point
- LCD display respiration waveform
- Expected
- RESP waveform displayed on LCD BrPM
- Tools
- Respiration simulator, delta ohms 1.0, Lead II, Baseline 1K
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature probe accuracy
Verification that temperature probe readings display within tolerance at multiple reference temperatures and probe connection/disconnection detection.
Ranked causes
- 1. Temperature probe fault or misalignment (occasional)
Verify temperature reading at 98.6°F
- Test point
- Monitor temperature display
- Expected
- Displayed temperature 98.6°F ±0.2°F, ±0.2°F
- Tools
- IVAC probe simulator TE 1811
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
101-ECG | ECG board data rate error | ECG data acquisition | error |
101 (ECG) | ECG board data rate error | ECG board data transmission rate incorrect | warning |
101 (Recorder) | output (to printer) queue overflow | Printer output queue buffer overflowed | warning |
101-Recorder | Printer output queue overflow | Printer data buffering | error |
101-Respiration | Memory error at wake up, unable to get data space | Respiration module wake-up | error |
101 (Respiration) | This means something was wrong with memory at wake up. Couldn't get data space. | Memory allocation failure at system wake-up | warning |
102 (ECG) | ECG board revision not compatible | ECG board software/hardware revision incompatible with main system | warning |
102-ECG | ECG board revision not compatible | ECG board identification | error |
102-Recorder | Printer output queue overflow | Printer data buffering | error |
102 (Recorder) | output (to printer) queue overflow | Printer output queue overrun | warning |
102 (Respiration) | Algorithm execution couldn't keep up with data acquisition | Respiration algorithm processing rate too slow | warning |
102-Respiration | Respiration algorithm execution behind data acquisition | Respiration data processing | error |
103-ECG | ECG board hardware error | ECG board self-test | error |
103 (ECG) | ECG board hardware error | Hardware failure detected in ECG board | warning |
103 (Recorder) | output (to printer) queue overflow | Printer output buffer overflow condition | warning |
103-Recorder | Printer output queue overflow | Printer data buffering | error |
103-Respiration | Respiration algorithm execution behind data acquisition | Respiration data processing | error |
103 (Respiration) | Algorithm execution couldn't keep up with data acquisition | Respiration algorithm lagging behind acquisition | warning |
104 (Recorder) | input queue (from system) overflow | Printer input buffer from system overflowed | warning |
104-Recorder | Printer input queue overflow from system | Printer input buffering | error |
104-Respiration | Respiration algorithm execution behind data acquisition | Respiration data processing | error |
104 (Respiration) | Algorithm execution couldn't keep up with data acquisition | Respiration algorithm processing too slow | warning |
105 (Recorder) | queue freeze error | Printer queue processing frozen | warning |
105-Recorder | Printer queue freeze error | Printer queue operation | error |
109-ECG | ECG waveform processing too far behind | ECG data processing | error |
109 (ECG) | Processing of ECG waveform too far behind | ECG waveform processing lagging behind data acquisition | warning |
110 | Overpressure circuit failure | Overpressure detection circuit malfunction | critical |
110 | Overpressure circuit failure | NIBP pressure monitoring | error |
110-Recorder | Invalid printer speed setting | Printer configuration | error |
110 (Recorder) | invalid speed setting | Printer speed parameter invalid or out of range | info |
111 | Software error - state machine bad probe number | Temperature monitoring state machine | error |
111-Recorder | Invalid printer number of waves setting | Printer configuration | error |
111 (Recorder) | invalid number of waves setting | Printer wave display count parameter invalid | info |
111 (Temperature) | Software error - state machine bad probe number | Temperature module software detected invalid probe number in state machine | warning |
112 | Overpressure watchdog error | Overpressure watchdog protection circuit failure | critical |
112 | Overpressure watchdog error | Pressure monitoring watchdog | error |
112-Recorder | Invalid printer density setting | Printer configuration | error |
112 (Recorder) | invalid density setting | Printer density parameter invalid or out of range | info |
112-Temp | Temperature sample rate error | Temperature data acquisition | error |
112 (Temperature) | Temp sample rate error | Temperature sampling rate incorrect | warning |
113 (ECG) | Data requested from ECG data manager is not available | ECG data manager does not have requested data | warning |
113-ECG | Data from ECG data manager not available | ECG data request | error |
113-Temp | Temperature lost synchronization with PIC | Temperature PIC communication | error |
113 (Temperature) | Temp lost synchronization with PIC | Temperature module lost sync with peripheral interface controller | warning |
114 (ECG) | Data requested from ECG data manager is not available | ECG data not available in data manager | warning |
114-ECG | Data from ECG data manager not available | ECG data request | error |
114-Recorder | Bad command to printer | Printer command transmission | error |
114 (Recorder) | bad command | Invalid command sent to printer | warning |
114-Temp | Unable to synchronize with temperature PIC | Temperature PIC synchronization | error |
114 (Temperature) | Unable to synchronize with the Temp PIC | Failed to establish synchronization with temperature peripheral controller | warning |
115-Recorder | Bad command to printer | Printer command transmission | error |
115 (Recorder) | bad command | Invalid printer command detected | warning |
120-Recorder | Printer queue not initialized | Printer initialization | error |
120 (Recorder) | queue not initialized | Printer queue data structure not properly initialized | warning |
121 (Recorder) | annotation queue overflow | Printer annotation buffer overflowed | warning |
121-Recorder | Printer annotation queue overflow | Printer annotation buffering | error |
122-Recorder | Invalid printer queue location | Printer queue management | error |
122 (Recorder) | invalid location | Printer output location parameter invalid | info |
123 (Recorder) | not enough room | Printer buffer insufficient space for requested output | warning |
123-Recorder | Printer queue insufficient room | Printer buffer management | error |
125-SPO2 | Too many reset requests from SPO2 | SPO2 board operation | warning |
125 (SPO2) | too many reset requests | SPO2 module reset requested excessively | warning |
126 (SPO2) | Nellcor has posted a 'serious' FE error | Nellcor SPO2 sensor reports serious functional error | warning |
126-SPO2 | Nellcor posted a serious front-end error | SPO2 Nellcor interface operation | error |
128-132-ECG | Error generating analog output | ECG analog output generation | error |
128-132 (ECG) | Errors returned while generating analog O/P (range) | Multiple analog output generation errors detected | warning |
128 (ECG) | Errors returned while generating analog O/P | Analog output generation failure during ECG processing | warning |
128-ECG | Error generating analog output | ECG analog output generation | error |
129-SPO2 | SPO2 front-end data OK but processing stalled | SPO2 data processing | error |
129 (SPO2) | FE data OK - processing stalled | SPO2 functional error data valid but processing halted | warning |
130 | EEProm read failure | Unable to read data from non-volatile EEPROM memory | critical |
130 | EEPROM read failure | EEPROM access during startup | error |
130-SPO2 | Masimo board or diagnostic failure | SPO2 Masimo interface operation | error |
130 (SPO2) | MASIMO has posted either board or diagnostic failure | MASIMO SPO2 sensor reports board or diagnostic error | warning |
131 | EEProm write failure | Unable to write data to non-volatile EEPROM memory | critical |
131 | EEPROM write failure | EEPROM write operation | error |
131-SPO2 | SPO2 message out of sequence | SPO2 data reception | error |
131 (SPO2) | msg looks out of sequence | SPO2 board message sequence error detected | warning |
132 (SPO2) | missing characters inside a packet | SPO2 data packet incomplete or corrupted | warning |
132-SPO2 | Missing characters in SPO2 packet | SPO2 packet reception | error |
133-SPO2 | Unable to correctly set SPO2 parameter | SPO2 parameter configuration | error |
133 (SPO2) | not able to correctly set parameter | Failed to set SPO2 module parameter | warning |
134 (SPO2) | NELL_SendCommand() called before previous call completed | SPO2 command sent before previous command processing complete | warning |
134-SPO2 | SPO2 SendCommand called before previous call completed | SPO2 command execution | error |
135-SPO2 | Transmit FIFO full for SPO2 communication | SPO2 data transmission | error |
135 (SPO2) | Not enough room left in transmit FIFO to send data | SPO2 transmit buffer full | warning |
136 (SPO2) | Receive FIFO full, probably lost data | SPO2 receive buffer overrun detected | warning |
136-SPO2 | SPO2 receive FIFO full, data lost | SPO2 data reception | error |
137-SPO2 | SPO2 output queue to OEM board full | SPO2 data queuing | error |
137 (SPO2) | queue out to OEM board is full | SPO2 output queue to OEM board at capacity | warning |
138 (SPO2) | Nellcor has posted too many auto resets | Nellcor SPO2 sensor auto-reset exceeded acceptable count | warning |
138-SPO2 | Nellcor posted too many auto resets | SPO2 Nellcor operation | error |
139-SPO2 | Nellcor software error | SPO2 Nellcor module operation | error |
139 (SPO2) | Nellcor software error | Nellcor SPO2 software failure detected | warning |
140 | Transducer initialization failure | Pressure transducer initialization failed during startup | critical |
140 | Transducer initialization failure | Transducer power-up | error |
140-Recorder | Bad command for printer mode | Printer mode-specific command | error |
140 (Recorder) | bad command for this mode | Printer command invalid for current operating mode | warning |
140 (SPO2) | Nellcor has posted too many communication errors | Nellcor SPO2 communication error count exceeded limit | warning |
140-SPO2 | Nellcor posted too many communication errors | SPO2 Nellcor communication | error |
141 | Transducer zero calibration failed | Transducer auto-zero procedure | error |
141 | Calibration of a transducer channel's zero failed | Pressure transducer zero-point calibration failed | critical |
141 (Recorder) | bad command for this mode | Unsupported printer command in current mode | warning |
141-Recorder | Bad command for printer mode | Printer mode-specific command | error |
141-SPO2 | No communication with SPO2 board | SPO2 board initialization | error |
141 (SPO2) | No communication with SPO2 board. Unit may be configured incorrectly | SPO2 board communication failure or non-responsive | warning |
142 | Calibration of a transducer channel's span failed | Pressure transducer span calibration failed | critical |
142 | Transducer span calibration failed | Transducer span calibration | error |
150 | Auto zero failure | Auto-zero calibration procedure | error |
150 | Auto zero failure | Automatic pressure zero calibration failed | critical |
151 | Auto zero verify failed | Verification of auto zero calibration failed | critical |
151 | Auto zero verification failed | Auto-zero verification after calibration | error |
170 | Pump current failure | Pneumatic pump current monitoring | error |
170 | Pump current failure | Pump current measurement out of acceptable range | critical |
171 | Pump current value out of NI_RANGE | Pump current reading outside specified nominal range | critical |
171 | Pump current value out of range | Pump current measurement | error |
180 | Excessive pneumatic leakage detected | Pressure decay monitoring | error |
180 | Excessive leakage | Pressure not maintained in pneumatic system | warning |
190 | Commands out of sequence | NIBP commands received in unexpected order | warning |
190 | Commands received out of sequence | Command sequence validation | error |
200 | Overpressure setpoint not found | Overpressure setpoint determination | error |
200 | Ovp setpoint not found | Overpressure setpoint value not located in memory | critical |
201 (ECG) | ECG board command queue overrun | ECG board command queue exceeded capacity | warning |
201-ECG | ECG board command queue overrun | Command queue processing | error |
210 | Pump stuck on during idle | Idle state pump monitoring | error |
210 | Pump stuck on during idle | Pump continues running when it should be idle | critical |
220 | Valve in illegal state | Pneumatic valve position invalid or unknown | critical |
220 | Valve in illegal state | Valve state monitoring | error |
221 | Pressure too high for too long | Pressure threshold and duration monitoring | error |
221 | Pressure too high for too long | System pressure exceeds setpoint for extended duration | critical |
26631 | Operating system 300 Hertz timer re-entry error | OS timer interrupt re-entry detected | critical |
26631 | Operating system 300 Hz timer re-entry error | OS timer interrupt handling | error |
27268 | Unexpected error condition | Software exception handler | error |
27268 | Unexpected error condition | Unexpected error detected by system | critical |
8193 | HW, Time Base Failure | Hardware time base clock failure detected | critical |
8193 | Time base failure | Hardware initialization | error |
8202 | Power supply system failure | Power supply monitoring | error |
8202 | HW, Power Supply System Failure | Power supply or system power failure detected | critical |
8222 | HW, RAM Test Failure | RAM memory test failure during self-test | critical |
8222 | RAM test failure | RAM memory self-test | error |
8232 | ROM checksum failure | ROM integrity check | error |
8232 | HW, ROM Checksum failure | ROM memory checksum verification failed | critical |
8252 | HW, Secondary processor not compatible | Secondary processor compatibility check failed | critical |
8252 | Secondary processor incompatible | Secondary processor detection | error |
8253 | Secondary communications failure | Inter-processor communication | error |
8253 | HW, Secondary communications failure | Communication with secondary processor failed | critical |
SY-16 | Power fail signal true time is too long | Power failure duration exceeds acceptable limit | critical |
SY-16 | Power fail signal duration exceeds threshold | Power supply monitoring | error |
SY-19 | Power supply output out of limits | Software power supply voltage monitoring | error |
SY-19 | Software detected power supply out of limits failure | Power supply voltage measured outside acceptable range | critical |
SY-20 | Checksum of code in Flash Memory is not valid | Flash memory program code checksum verification failed | critical |
SY-20 | Code checksum in flash memory invalid | Flash memory integrity check | error |
SY-40 | Unexpected interrupt occurred | Interrupt handler execution | error |
SY-40 | Unexpected interrupt | Software detected unexpected interrupt signal | critical |
SY-43 | Real time clock (DS1284) running too slow | Real-time clock oscillator frequency too low | warning |
SY-43 | Real time clock running too slow | Real time clock monitoring | warning |
SY-44 | Real time clock running too fast | Real time clock monitoring | warning |
SY-44 | Real time clock (DS1284) running too fast | Real-time clock oscillator frequency too high | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI