Critikon PRO 1000V3: troubleshooting and error codes

The ServiceForge knowledge base holds 33 documented symptoms for the Critikon PRO 1000V3, with 87 ranked causes, 34 diagnostic checks with test points and expected values, and 158 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit will not power on caution

Equipment fails to power on when AC power applied and power switch pressed

Ranked causes

  1. 1. No AC power available (common)
  2. 2. Faulty power supply (common)
  3. 3. Faulty PSU power board (occasional)
  4. 4. Faulty main board (occasional)
  5. 5. Faulty power cable (occasional)
  6. 6. Faulty power switch (occasional)
First check

Verify AC power availability at outlet

Expected
AC voltage present

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not power on

Device fails to turn on or power up

Ranked causes

  1. 1. No AC power to unit (common)
  2. 2. Faulty power supply (occasional)
  3. 3. Faulty PSU PWA (occasional)
  4. 4. Faulty main board (occasional)
  5. 5. Faulty power cables (occasional)
  6. 6. Faulty power On/Off front panel switch (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit powers on but no display caution

Equipment powers up with signs of operation but display screen blank or dark

Ranked causes

  1. 1. Faulty display (common)
  2. 2. Faulty backlight driver (occasional)
  3. 3. Faulty main board (occasional)
  4. 4. Faulty PSU power board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit powers on but no display

Device powers up but display shows nothing

Ranked causes

  1. 1. Faulty display (occasional)
  2. 2. Faulty backlight driver (occasional)
  3. 3. Faulty main board (occasional)
  4. 4. Faulty PSU PWA (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform NIBP function patient safety

Non-invasive blood pressure measurement not working

Ranked causes

  1. 1. Faulty main board (occasional)
  2. 2. Faulty pneumatics assembly (occasional)
  3. 3. Faulty pneumatics cable (occasional)
  4. 4. Faulty front panel switch (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform NIBP function patient safety

Non-invasive blood pressure monitoring function fails to operate

Ranked causes

  1. 1. Faulty main board (common)
  2. 2. Faulty pneumatics assembly (common)
  3. 3. Faulty pneumatics cable (occasional)
  4. 4. Faulty front panel switch (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not print

Printer function fails to produce output or paper issues present

Ranked causes

  1. 1. No paper or incorrect paper loading (common)
  2. 2. Faulty printer (common)
  3. 3. Faulty main board (occasional)
  4. 4. Faulty PSU power board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not print

Printer module not printing documents

Ranked causes

  1. 1. Unit out of paper or paper incorrectly installed (common)
  2. 2. Faulty printer (occasional)
  3. 3. Faulty main board (occasional)
  4. 4. Faulty PSU PWA (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not operate on battery caution

Equipment powered by battery backup fails to operate or battery does not charge

Ranked causes

  1. 1. Battery disconnect fuse open or unplugged (common)
  2. 2. Faulty battery pack (common)
  3. 3. Faulty PSU power board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not operate on battery

Device cannot run on battery power

Ranked causes

  1. 1. Battery disconnect fuse open or not plugged in (common)
  2. 2. Faulty battery pack (occasional)
  3. 3. Faulty PSU PWA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform ECG function patient safety

Electrocardiogram monitoring function fails to operate

Ranked causes

  1. 1. Faulty ECG PWA assembly (common)
  2. 2. Faulty ECG cable assembly (common)
  3. 3. Faulty main board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform ECG function patient safety

Electrocardiogram measurement not working

Ranked causes

  1. 1. Faulty ECG PWA assembly (occasional)
  2. 2. Faulty ECG cable assembly (occasional)
  3. 3. Faulty main board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform SPO2 function patient safety

Pulse oximetry (SpO2) monitoring function fails to operate

Ranked causes

  1. 1. Faulty SPO2 PWA (Nellcor) (common)
  2. 2. Faulty SPO2 PWA (Masimo) (common)
  3. 3. Faulty main board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform Temperature function patient safety

Temperature monitoring via probe fails to operate

Ranked causes

  1. 1. Faulty temperature probe (common)
  2. 2. Faulty temperature probe sensor (common)
  3. 3. Faulty main board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform Temperature function patient safety

Temperature measurement not working

Ranked causes

  1. 1. Faulty temperature probe (occasional)
  2. 2. Faulty temperature probe sensor (occasional)
  3. 3. Faulty main board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Recorder print quality and speed

Verification that recorder prints waveforms and text at correct speed with even tone, proper pixel density, and legible spacing.

Ranked causes

  1. 1. Recorder paper misalignment or print head contamination (occasional)
  2. 2. Worn or defective recorder mechanism (rare)
First check

Verify paper is loaded in recorder

Test point
Recorder Parameter paper compartment
Expected
Paper properly loaded and ready for printing
Tools
Recorder paper

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit Host Comms not functional

Host communications interface fails to operate

Ranked causes

  1. 1. Faulty main board (common)
  2. 2. Faulty PSU power board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit host communications not functional

External device communication or networking not working

Ranked causes

  1. 1. Faulty main board (occasional)
  2. 2. Faulty PSU PWA (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not generate sound

Speaker or audio output not functioning

Ranked causes

  1. 1. Faulty speaker assembly (occasional)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not generate sound

Audio alarm or notification sounds fail to play

Ranked causes

  1. 1. Faulty speaker assembly (common)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not operate on external DC power caution

Equipment fails to operate when powered from external DC source

Ranked causes

  1. 1. Faulty DC input jack or cable (common)
  2. 2. Faulty PSU power board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not operate on external DC power

Device cannot run on external DC power input

Ranked causes

  1. 1. Faulty DC input jack or DC cable (occasional)
  2. 2. Faulty PSU PWA (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not perform SpO2 function patient safety

Pulse oximetry (SpO2) measurement not working

Ranked causes

  1. 1. Faulty SpO2 PWA assembly (occasional)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not produce analog ECG waveform output patient safety

Defibrillator interface or ECG output not functional

Ranked causes

  1. 1. Faulty Defib cable assembly (occasional)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not produce analog ECG waveform output patient safety

Equipment fails to output ECG signal to external defibrillator or devices

Ranked causes

  1. 1. Faulty Defib cable assembly (common)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not respond to rotary knob

Rotary encoder control not functioning

Ranked causes

  1. 1. Faulty encoder (occasional)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not respond to rotary knob

Rotary encoder control on unit does not respond to user input

Ranked causes

  1. 1. Faulty encoder (common)
  2. 2. Faulty main board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ECG signal detection and waveform display

Verification of ECG simulator signal acquisition, waveform display, paced marker detection, and alarm functions across heart rate range.

Ranked causes

  1. 1. ECG lead or connector fault (occasional)
First check

Verify ECG waveform display at 80 BPM

Test point
LCD display waveform and HR indication
Expected
ECG waveform displayed on LCD
Tools
ECG simulator set to 80 BPM Paced, 1.0 mV amplitude

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP leak test caution

Pneumatic system pressure loss rate measurement to verify system integrity before NIBP calibration checks.

Ranked causes

  1. 1. Cuff or hose defective or leaking (common)
First check

Run automated leak test from Service Menu

Test point
NIBP Parameter connection
Expected
Leak Test Status displays Passed, Leak Test Results value less than 6, Service Error: None

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP overpressure protection caution

Verification that pump automatically shuts down and releases pressure at safe limits for adult and neonatal cuff settings.

Ranked causes

  1. 1. Overpressure valve out of tolerance (occasional)
First check

Verify adult overpressure threshold

Test point
Manometer pressure reading when pump shuts down
Expected
Pump shuts down and pressure released at 300-330 mmHg, 300-330 mmHg
Tools
Manometer, calibration kit 320-246

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP pressure measurement accuracy caution

Verification that PT1 and PT2 pressure transducers read within tolerance at multiple reference pressures to confirm system linearity and accuracy.

Ranked causes

  1. 1. Pressure transducers out of calibration (occasional)
First check

Verify PT1 and PT2 pressure at 200 mmHg reference

Test point
Service Menu PT1 Pressure and PT2 Pressure display
Expected
PT1 and PT2 Pressures within 1 mmHg of manometer reading, ±1 mmHg
Tools
Manometer, calibration kit 320-246

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Respiration waveform verification

Verification that respiration waveform and rate readings display correctly at simulated respiration rates.

Ranked causes

  1. 1. Respiration sensor or signal processing fault (rare)
First check

Verify respiration waveform at 20 BrPM

Test point
LCD display respiration waveform
Expected
RESP waveform displayed on LCD BrPM
Tools
Respiration simulator, delta ohms 1.0, Lead II, Baseline 1K

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature probe accuracy

Verification that temperature probe readings display within tolerance at multiple reference temperatures and probe connection/disconnection detection.

Ranked causes

  1. 1. Temperature probe fault or misalignment (occasional)
First check

Verify temperature reading at 98.6°F

Test point
Monitor temperature display
Expected
Displayed temperature 98.6°F ±0.2°F, ±0.2°F
Tools
IVAC probe simulator TE 1811

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
101-ECGECG board data rate errorECG data acquisitionerror
101 (ECG)ECG board data rate errorECG board data transmission rate incorrectwarning
101 (Recorder)output (to printer) queue overflowPrinter output queue buffer overflowedwarning
101-RecorderPrinter output queue overflowPrinter data bufferingerror
101-RespirationMemory error at wake up, unable to get data spaceRespiration module wake-uperror
101 (Respiration)This means something was wrong with memory at wake up. Couldn't get data space.Memory allocation failure at system wake-upwarning
102 (ECG)ECG board revision not compatibleECG board software/hardware revision incompatible with main systemwarning
102-ECGECG board revision not compatibleECG board identificationerror
102-RecorderPrinter output queue overflowPrinter data bufferingerror
102 (Recorder)output (to printer) queue overflowPrinter output queue overrunwarning
102 (Respiration)Algorithm execution couldn't keep up with data acquisitionRespiration algorithm processing rate too slowwarning
102-RespirationRespiration algorithm execution behind data acquisitionRespiration data processingerror
103-ECGECG board hardware errorECG board self-testerror
103 (ECG)ECG board hardware errorHardware failure detected in ECG boardwarning
103 (Recorder)output (to printer) queue overflowPrinter output buffer overflow conditionwarning
103-RecorderPrinter output queue overflowPrinter data bufferingerror
103-RespirationRespiration algorithm execution behind data acquisitionRespiration data processingerror
103 (Respiration)Algorithm execution couldn't keep up with data acquisitionRespiration algorithm lagging behind acquisitionwarning
104 (Recorder)input queue (from system) overflowPrinter input buffer from system overflowedwarning
104-RecorderPrinter input queue overflow from systemPrinter input bufferingerror
104-RespirationRespiration algorithm execution behind data acquisitionRespiration data processingerror
104 (Respiration)Algorithm execution couldn't keep up with data acquisitionRespiration algorithm processing too slowwarning
105 (Recorder)queue freeze errorPrinter queue processing frozenwarning
105-RecorderPrinter queue freeze errorPrinter queue operationerror
109-ECGECG waveform processing too far behindECG data processingerror
109 (ECG)Processing of ECG waveform too far behindECG waveform processing lagging behind data acquisitionwarning
110Overpressure circuit failureOverpressure detection circuit malfunctioncritical
110Overpressure circuit failureNIBP pressure monitoringerror
110-RecorderInvalid printer speed settingPrinter configurationerror
110 (Recorder)invalid speed settingPrinter speed parameter invalid or out of rangeinfo
111Software error - state machine bad probe numberTemperature monitoring state machineerror
111-RecorderInvalid printer number of waves settingPrinter configurationerror
111 (Recorder)invalid number of waves settingPrinter wave display count parameter invalidinfo
111 (Temperature)Software error - state machine bad probe numberTemperature module software detected invalid probe number in state machinewarning
112Overpressure watchdog errorOverpressure watchdog protection circuit failurecritical
112Overpressure watchdog errorPressure monitoring watchdogerror
112-RecorderInvalid printer density settingPrinter configurationerror
112 (Recorder)invalid density settingPrinter density parameter invalid or out of rangeinfo
112-TempTemperature sample rate errorTemperature data acquisitionerror
112 (Temperature)Temp sample rate errorTemperature sampling rate incorrectwarning
113 (ECG)Data requested from ECG data manager is not availableECG data manager does not have requested datawarning
113-ECGData from ECG data manager not availableECG data requesterror
113-TempTemperature lost synchronization with PICTemperature PIC communicationerror
113 (Temperature)Temp lost synchronization with PICTemperature module lost sync with peripheral interface controllerwarning
114 (ECG)Data requested from ECG data manager is not availableECG data not available in data managerwarning
114-ECGData from ECG data manager not availableECG data requesterror
114-RecorderBad command to printerPrinter command transmissionerror
114 (Recorder)bad commandInvalid command sent to printerwarning
114-TempUnable to synchronize with temperature PICTemperature PIC synchronizationerror
114 (Temperature)Unable to synchronize with the Temp PICFailed to establish synchronization with temperature peripheral controllerwarning
115-RecorderBad command to printerPrinter command transmissionerror
115 (Recorder)bad commandInvalid printer command detectedwarning
120-RecorderPrinter queue not initializedPrinter initializationerror
120 (Recorder)queue not initializedPrinter queue data structure not properly initializedwarning
121 (Recorder)annotation queue overflowPrinter annotation buffer overflowedwarning
121-RecorderPrinter annotation queue overflowPrinter annotation bufferingerror
122-RecorderInvalid printer queue locationPrinter queue managementerror
122 (Recorder)invalid locationPrinter output location parameter invalidinfo
123 (Recorder)not enough roomPrinter buffer insufficient space for requested outputwarning
123-RecorderPrinter queue insufficient roomPrinter buffer managementerror
125-SPO2Too many reset requests from SPO2SPO2 board operationwarning
125 (SPO2)too many reset requestsSPO2 module reset requested excessivelywarning
126 (SPO2)Nellcor has posted a 'serious' FE errorNellcor SPO2 sensor reports serious functional errorwarning
126-SPO2Nellcor posted a serious front-end errorSPO2 Nellcor interface operationerror
128-132-ECGError generating analog outputECG analog output generationerror
128-132 (ECG)Errors returned while generating analog O/P (range)Multiple analog output generation errors detectedwarning
128 (ECG)Errors returned while generating analog O/PAnalog output generation failure during ECG processingwarning
128-ECGError generating analog outputECG analog output generationerror
129-SPO2SPO2 front-end data OK but processing stalledSPO2 data processingerror
129 (SPO2)FE data OK - processing stalledSPO2 functional error data valid but processing haltedwarning
130EEProm read failureUnable to read data from non-volatile EEPROM memorycritical
130EEPROM read failureEEPROM access during startuperror
130-SPO2Masimo board or diagnostic failureSPO2 Masimo interface operationerror
130 (SPO2)MASIMO has posted either board or diagnostic failureMASIMO SPO2 sensor reports board or diagnostic errorwarning
131EEProm write failureUnable to write data to non-volatile EEPROM memorycritical
131EEPROM write failureEEPROM write operationerror
131-SPO2SPO2 message out of sequenceSPO2 data receptionerror
131 (SPO2)msg looks out of sequenceSPO2 board message sequence error detectedwarning
132 (SPO2)missing characters inside a packetSPO2 data packet incomplete or corruptedwarning
132-SPO2Missing characters in SPO2 packetSPO2 packet receptionerror
133-SPO2Unable to correctly set SPO2 parameterSPO2 parameter configurationerror
133 (SPO2)not able to correctly set parameterFailed to set SPO2 module parameterwarning
134 (SPO2)NELL_SendCommand() called before previous call completedSPO2 command sent before previous command processing completewarning
134-SPO2SPO2 SendCommand called before previous call completedSPO2 command executionerror
135-SPO2Transmit FIFO full for SPO2 communicationSPO2 data transmissionerror
135 (SPO2)Not enough room left in transmit FIFO to send dataSPO2 transmit buffer fullwarning
136 (SPO2)Receive FIFO full, probably lost dataSPO2 receive buffer overrun detectedwarning
136-SPO2SPO2 receive FIFO full, data lostSPO2 data receptionerror
137-SPO2SPO2 output queue to OEM board fullSPO2 data queuingerror
137 (SPO2)queue out to OEM board is fullSPO2 output queue to OEM board at capacitywarning
138 (SPO2)Nellcor has posted too many auto resetsNellcor SPO2 sensor auto-reset exceeded acceptable countwarning
138-SPO2Nellcor posted too many auto resetsSPO2 Nellcor operationerror
139-SPO2Nellcor software errorSPO2 Nellcor module operationerror
139 (SPO2)Nellcor software errorNellcor SPO2 software failure detectedwarning
140Transducer initialization failurePressure transducer initialization failed during startupcritical
140Transducer initialization failureTransducer power-uperror
140-RecorderBad command for printer modePrinter mode-specific commanderror
140 (Recorder)bad command for this modePrinter command invalid for current operating modewarning
140 (SPO2)Nellcor has posted too many communication errorsNellcor SPO2 communication error count exceeded limitwarning
140-SPO2Nellcor posted too many communication errorsSPO2 Nellcor communicationerror
141Transducer zero calibration failedTransducer auto-zero procedureerror
141Calibration of a transducer channel's zero failedPressure transducer zero-point calibration failedcritical
141 (Recorder)bad command for this modeUnsupported printer command in current modewarning
141-RecorderBad command for printer modePrinter mode-specific commanderror
141-SPO2No communication with SPO2 boardSPO2 board initializationerror
141 (SPO2)No communication with SPO2 board. Unit may be configured incorrectlySPO2 board communication failure or non-responsivewarning
142Calibration of a transducer channel's span failedPressure transducer span calibration failedcritical
142Transducer span calibration failedTransducer span calibrationerror
150Auto zero failureAuto-zero calibration procedureerror
150Auto zero failureAutomatic pressure zero calibration failedcritical
151Auto zero verify failedVerification of auto zero calibration failedcritical
151Auto zero verification failedAuto-zero verification after calibrationerror
170Pump current failurePneumatic pump current monitoringerror
170Pump current failurePump current measurement out of acceptable rangecritical
171Pump current value out of NI_RANGEPump current reading outside specified nominal rangecritical
171Pump current value out of rangePump current measurementerror
180Excessive pneumatic leakage detectedPressure decay monitoringerror
180Excessive leakagePressure not maintained in pneumatic systemwarning
190Commands out of sequenceNIBP commands received in unexpected orderwarning
190Commands received out of sequenceCommand sequence validationerror
200Overpressure setpoint not foundOverpressure setpoint determinationerror
200Ovp setpoint not foundOverpressure setpoint value not located in memorycritical
201 (ECG)ECG board command queue overrunECG board command queue exceeded capacitywarning
201-ECGECG board command queue overrunCommand queue processingerror
210Pump stuck on during idleIdle state pump monitoringerror
210Pump stuck on during idlePump continues running when it should be idlecritical
220Valve in illegal statePneumatic valve position invalid or unknowncritical
220Valve in illegal stateValve state monitoringerror
221Pressure too high for too longPressure threshold and duration monitoringerror
221Pressure too high for too longSystem pressure exceeds setpoint for extended durationcritical
26631Operating system 300 Hertz timer re-entry errorOS timer interrupt re-entry detectedcritical
26631Operating system 300 Hz timer re-entry errorOS timer interrupt handlingerror
27268Unexpected error conditionSoftware exception handlererror
27268Unexpected error conditionUnexpected error detected by systemcritical
8193HW, Time Base FailureHardware time base clock failure detectedcritical
8193Time base failureHardware initializationerror
8202Power supply system failurePower supply monitoringerror
8202HW, Power Supply System FailurePower supply or system power failure detectedcritical
8222HW, RAM Test FailureRAM memory test failure during self-testcritical
8222RAM test failureRAM memory self-testerror
8232ROM checksum failureROM integrity checkerror
8232HW, ROM Checksum failureROM memory checksum verification failedcritical
8252HW, Secondary processor not compatibleSecondary processor compatibility check failedcritical
8252Secondary processor incompatibleSecondary processor detectionerror
8253Secondary communications failureInter-processor communicationerror
8253HW, Secondary communications failureCommunication with secondary processor failedcritical
SY-16Power fail signal true time is too longPower failure duration exceeds acceptable limitcritical
SY-16Power fail signal duration exceeds thresholdPower supply monitoringerror
SY-19Power supply output out of limitsSoftware power supply voltage monitoringerror
SY-19Software detected power supply out of limits failurePower supply voltage measured outside acceptable rangecritical
SY-20Checksum of code in Flash Memory is not validFlash memory program code checksum verification failedcritical
SY-20Code checksum in flash memory invalidFlash memory integrity checkerror
SY-40Unexpected interrupt occurredInterrupt handler executionerror
SY-40Unexpected interruptSoftware detected unexpected interrupt signalcritical
SY-43Real time clock (DS1284) running too slowReal-time clock oscillator frequency too lowwarning
SY-43Real time clock running too slowReal time clock monitoringwarning
SY-44Real time clock running too fastReal time clock monitoringwarning
SY-44Real time clock (DS1284) running too fastReal-time clock oscillator frequency too highwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI