Critikon Dinamap Compact: troubleshooting and error codes
The ServiceForge knowledge base holds 68 documented symptoms for the Critikon Dinamap Compact, with 150 ranked causes, 117 diagnostic checks with test points and expected values, and 45 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Incorrect or no LCD output
LCD display shows incorrect information or is blank
Ranked causes
- 1. LCD board fault (common)
- 2. Addressing fault on switch board or main board (occasional)
- 3. LCD negative supply out of specification (occasional)
Test LCD board operation
- Test point
- LCD Board
- Expected
- LCD displays correctly
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect or no LCD output
LCD display shows no information or incorrect display content
Ranked causes
- 1. LCD Board fault (common)
- 2. Addressing fault on Switch Board or Main Board (occasional)
- 3. Negative supply to LCD module out of specification (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect or no LED display output
LED display shows incorrect information, is blank, or partially dark
Ranked causes
- 1. LED display board fault (common)
- 2. Addressing fault on switch board or main board (occasional)
- 3. 4.5V VDISP supply absent or low (occasional)
Measure +4.5V VDISP supply voltage
- Test point
- Display Board power input
- Expected
- +4.5V present and stable
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect or no LED display output
LED display shows no information, is dim, or displays incorrect data
Ranked causes
- 1. LED Display Board fault (common)
- 2. Addressing fault on Switch Board or Main Board (occasional)
- 3. No or low +4.5V VDISP supply (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement pump laboring with low frequency sound (N33 or N00 message) patient safety
During NIBP determination pump operates at lower than normal frequency or intensity; N33 timeout inflation or N00 overpressure warning may display
Ranked causes
- 1. Kinked hose to cuff or faulty cuff (common)
- 2. Blocked or kinked hose from pump output to front panel (occasional)
- 3. Pneumatic assembly fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement timeout - N44, N55, or N99 error patient safety
NIBP measurement fails with N44 (timeout total time), N55 (timeout pressure), or N99 (no determination) error
Ranked causes
- 1. Excessive patient movement (motion artifact) (common)
- 2. Main board NIBP circuitry fault or pressure measurement problem (occasional)
- 3. Pneumatic problem with cuff deflation or pressure measurement (occasional)
Eliminate motion artifact conditions
- Test point
- Patient and monitor position
- Expected
- Unit stable and patient movement minimized
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP overpressure warning displayed when pump starts with cuff attached patient safety
LCD displays NIBP overpressure alert when pump activates with cuff properly connected
Ranked causes
- 1. Kinked hose to cuff or faulty cuff (common)
- 2. Blocked or kinked hose from front panel to main board pressure sensor ML1 (occasional)
- 3. Main board fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP overpressure warning with cuff attached patient safety
NIBP overpressure warning displayed when pump starts with cuff properly attached
Ranked causes
- 1. Kinked hose to cuff or faulty cuff (common)
- 2. Blocked or kinked hose from front panel to main board pressure sensor (common)
- 3. Main board fault (ML1, IC5 or IC7) (occasional)
Test unit with known good cuff
- Test point
- NIBP cuff connector
- Expected
- Unit operates normally with known good cuff
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP pump laboring (low frequency sound) patient safety
When NIBP determination starts pump sounds labored with lower frequency than normal; may include N33 or N00 warnings
Ranked causes
- 1. Kinked hose to cuff or faulty cuff (common)
- 2. Blocked or kinked hose from pump output to front panel (common)
- 3. Pneumatic assembly fault (occasional)
Test unit with known good cuff
- Test point
- NIBP cuff connector
- Expected
- Pump operates normally with known good cuff
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No LCD backlight
LCD display is dark with no backlight illumination
Ranked causes
- 1. LCD Board fault (common)
- 2. Display Board fault (occasional)
- 3. Addressing fault on Switch Board or Main Board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No LCD backlight
LCD display is present but backlight does not illuminate
Ranked causes
- 1. LCD board fault (common)
- 2. Display board fault (common)
- 3. Addressing fault on switch board or main board (occasional)
Check backlight switch TR1 and backlight supply
- Test point
- PL31 Pin 2 with respect to Pin 1
- Expected
- +Vbacklight present
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No rotor response
Unit does not respond when rotor is turned
Ranked causes
- 1. Rotor failure (common)
- 2. Switch board failure (occasional)
- 3. Main board failure (occasional)
Test rotor operation
- Test point
- Rotor assembly
- Expected
- Rotor responds to manual rotation
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No rotor response
Unit does not respond when rotor is manipulated
Ranked causes
- 1. Rotor failure (common)
- 2. Switch Board failure (occasional)
- 3. Main Board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump over-current alarm
Unit alarms with pump over-current message on LCD indicating pump current limit exceeded
Ranked causes
- 1. Pump failure or mechanical jam (common)
- 2. PSU board pump current sense circuitry fault (occasional)
- 3. Main board fault (occasional)
Test pump assembly operation
- Test point
- Pump motor
- Expected
- Pump operates without excessive current draw or jamming
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotor menu selection skips items
When rotor is turned, menus skip over items instead of advancing normally
Ranked causes
- 1. One digit of grey code not functioning (common)
- 2. Rotor failure (common)
- 3. Switch board failure (occasional)
Test rotor grey code output
- Test point
- Rotor grey code outputs
- Expected
- All grey code bits functioning
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 no signal error (P55 error)
LCD displays P55 error message indicating no SpO2 signal detected
Ranked causes
- 1. Faulty finger probe (common)
- 2. Faulty SpO2 module to front panel cable (occasional)
- 3. SpO2 module fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 no signal error (P55 Error)
LCD displays P55 Error indicating SpO2 module is connected but receiving no valid signal
Ranked causes
- 1. Faulty finger probe sensor (common)
- 2. Faulty SpO2 module to front panel cable (common)
- 3. SpO2 module fault (occasional)
Test with known good SpO2 sensor
- Test point
- SpO2 probe connector
- Expected
- Unit detects signal from known good probe
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 probe disconnected error (P00 Error)
LCD displays P00 Error indicating SpO2 probe is disconnected or not detected
Ranked causes
- 1. Faulty finger probe sensor (common)
- 2. Faulty SpO2 module to front panel cable (common)
- 3. SpO2 module fault (occasional)
Test with known good SpO2 sensor
- Test point
- SpO2 probe connector
- Expected
- Unit recognizes known good probe
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 probe disconnected error reported on LCD (P00 error)
LCD displays P00 error message indicating SpO2 probe disconnection
Ranked causes
- 1. Faulty finger probe (common)
- 2. Faulty SpO2 module to front panel cable (occasional)
- 3. SpO2 module fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error P105, P110, or P115 alarm and message on LCD after power-up patient safety
Alarm sounds with system error message displayed on LCD display after unit completes power-up sequence
Ranked causes
- 1. SpO2 module fault (common)
- 2. Isolated power supply module failure on Main Board (occasional)
- 3. Main Board processor SpO2 RS232 communications failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error P105, P110, or P115 at power-up
Unit alarms at power-up with system error message P105, P110, or P115 on LCD
Ranked causes
- 1. SpO2 module fault (common)
- 2. Isolated power supply module failure on main board (occasional)
- 3. Main board processor SpO2 RS232 communications failure (occasional)
Swap and test SpO2 module
- Test point
- SpO2 module
- Expected
- Unit powers up without P105/P110/P115 error
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit alarms 'PIC: +/-12V Out of Spec'
LCD displays '+/-12V Out of Spec' alarm with code line failure indication
Ranked causes
- 1. PSU Board or Main Board fault (common)
- 2. Low input voltage to IC18 or IC19 causing regulator dropout (occasional)
- 3. Excessive current on +12VV supply or erroneous TR10 operation (occasional)
Measure +12VR and -12VR rail voltages
- Test point
- PSU Board +12VR and -12VR outputs
- Expected
- Within specification as documented
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit alarms 'Pump Over Current' patient safety
LCD displays 'Pump Over Current 0,(code line failure),0,0' alarm message
Ranked causes
- 1. Pump failure (common)
- 2. PSU Board failure of pump current sense circuitry (occasional)
- 3. Main Board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on from external DC source
Unit fails to switch on when connected to external DC adapter
Ranked causes
- 1. External DC adapter output out of specification (common)
- 2. External DC fuse blown (common)
- 3. Power supply board fault (occasional)
Measure external DC adapter output voltage
- Test point
- EXT DC adapter output
- Expected
- Within specification
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not switch on from either battery or external DC source caution
Unit completely unresponsive to power input from all sources
Ranked causes
- 1. On/Off switch faulty (common)
- 2. +5VD supply low or out of specification (occasional)
- 3. PIC Processor not running (occasional)
Test unit power on using rotor switch
- Test point
- Rotor switch
- Expected
- Unit powers on
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not switch on from external DC source
Unit does not power on when connected to external DC adapter
Ranked causes
- 1. External DC adapter output out of specification (common)
- 2. External DC fuse blown (common)
- 3. Power supply unit board fault (occasional)
Measure external DC adapter output voltage
- Test point
- EXT DC adapter output terminals
- Expected
- Within specification
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pop-off pressure adjustment caution
Set the neonatal overpressure threshold; adult threshold is automatically double. Requires Main Board disassembly and jumper LK1.
Ranked causes
- 1. Pop-off pressure adjustment required (occasional)
- 2. Overpressure during calibration (rare)
Verify zero pressure baseline before calibration
- Test point
- calibration pressure measurement system
- Expected
- 0 mmHg (0 mBar)
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pop-off pressure check (Adult mode)
Verify the adult mode overpressure detection system prevents excessive cuff inflation by checking pop-off pressure.
Ranked causes
- 1. Periodic pop-off pressure verification required for adult mode (common)
- 2. Adult pop-off pressure out of range (rare)
Monitor inflation and pop-off in adult mode
- Test point
- calibration pressure measurement system
- Expected
- System inflates then immediately deflates; pump stops; Overpressure alarm reported
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pop-off pressure check (Neonate mode) caution
Verify the overpressure detection system prevents excessive cuff inflation by checking pop-off pressure is within specifications.
Ranked causes
- 1. Periodic pop-off pressure verification required (common)
- 2. Pop-off pressure out of range (rare)
Verify calibration kit connection
- Test point
- calibration kit to DINAMAP™ Compact connector
- Expected
- Connected as shown on page 41
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pressure
Monitor requires pressure calibration using external measurement system to establish zero and 200 mmHg reference points for accurate blood pressure readings.
Ranked causes
- 1. Calibration procedure required (common)
- 2. Calibration failed (occasional)
Verify zero pressure reading
- Test point
- calibration pressure measurement system display
- Expected
- 0 mmHg (0 mBar)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Predictive temperature
Adjustment of temperature calibration circuitry via potentiometers RV2 and RV3 on Main Board. Procedure varies by software version.
Ranked causes
- 1. Temperature calibration requires potentiometer adjustment per software version (common)
- 2. Record calibration timestamp (common)
Verify temperature calibration plug resistance
- Test point
- Pins 1 and 2 of temperature calibration plug (320-635)
- Expected
- 6090Ω, ±7Ω
7 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature sensor adjustment caution
Calibrate predictive temperature circuitry by adjusting potentiometers RV2 and RV3 on Main Board. Requires disassembly and precise edge count measurements.
Ranked causes
- 1. Temperature calibration adjustment required (occasional)
- 2. Temperature reading not 98.6°F after adjustment (rare)
Adjust RV2 potentiometer
- Test point
- RV2 potentiometer on Main Board; edge counter display
- Expected
- Edge count for Calibration PLUG = 957
- Tools
- edge counter display on UUT LCD
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature sensor check caution
Verify temperature detection circuit calibration using precision calibration plug (320-635) every 12 months or when readings are doubtful.
Ranked causes
- 1. Periodic temperature calibration check required (common)
- 2. Temperature reading out of calibration (occasional)
Insert calibration plug without twisting
- Test point
- temperature probe connector on front of monitor
- Expected
- Plug seated firmly without rotation
- Tools
- calibration plug part 320-635
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
N12 error displayed with alarm and blank LCD patient safety
N12 error code shown on systolic display, alarm sounds, and LCD display is blank
Ranked causes
- 1. Pneumatic system state does not match software expectation (common)
- 2. Valve control circuitry fault preventing cuff deflation (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
N12 error with blank LCD and alarm patient safety
N12 error displayed on systolic display with blank LCD and alarm sounding; indicates pneumatic system state mismatch
Ranked causes
- 1. Valve control circuitry fault preventing cuff deflation (common)
- 2. Main board fault (IC3) (occasional)
Determine valve drive status
- Test point
- Pneumatics Control circuitry
- Expected
- Valve drive circuit functions correctly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
N13 error displayed with alarm and blank LCD patient safety
N13 error code shown on systolic display, alarm sounds, and LCD display is blank
Ranked causes
- 1. Pump runs continuously on power-up or when it should be off (common)
- 2. Main board failure with IC3 defect (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
N13 error with blank LCD and alarm patient safety
N13 error displayed on systolic display with blank LCD and alarm sounding; indicates pump continues running when it should be off
Ranked causes
- 1. Pump continuous operation or failure to turn off (Pneumatics Control IC1 or defective pump current measurement) (common)
- 2. Main board fault (IC3) (occasional)
Check pump operating status and current measurement
- Test point
- Pump motor current sense circuitry on PSU Board
- Expected
- Pump turns off when commanded; current measurement is accurate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement no determination error (N99) patient safety
LCD displays 'N99-No determination' message when NIBP measurement is attempted
Ranked causes
- 1. Patient motion artifact or main board NIBP circuitry fault or PSU board pneumatic control fault or pneumatic assembly fault (common)
- 2. NIBP pulse determination circuitry fault on main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement timeout pressure error (N55) patient safety
LCD displays 'N55-NIBP: Timeout Pressure' message during NIBP measurement attempt
Ranked causes
- 1. Patient motion artifact (common)
- 2. Main board NIBP circuitry fault or PSU board pneumatic control fault or pneumatic assembly fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement timeout total time error (N44) patient safety
LCD displays 'N44-NIBP: Timeout Total Time' message during NIBP measurement attempt
Ranked causes
- 1. Patient motion artifact (common)
- 2. Main board NIBP circuitry fault or PSU board pneumatic control fault or pneumatic assembly fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP overpressure warning displayed when pump starts (no cuff attached) patient safety
LCD displays NIBP overpressure alert when pump activates even without cuff connected to unit
Ranked causes
- 1. Blocked or kinked hose from pump output to front panel (common)
- 2. Pneumatic assembly fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP overpressure warning without cuff attached patient safety
NIBP overpressure warning displayed on LCD when pump starts, even without cuff attached
Ranked causes
- 1. Blocked or kinked hose from pump output to front panel (common)
- 2. Pneumatic assembly fault (occasional)
Inspect pneumatic hose from pump to front panel
- Test point
- Pump output hose
- Expected
- Hose clear and not kinked
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No keypad response
Unit does not respond when keys are pressed
Ranked causes
- 1. Switch board failure (common)
- 2. Main board failure (occasional)
Test switch board operation
- Test point
- Switch Board
- Expected
- Keypad signals present and correct
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No keyswitch response
Unit does not respond when keys are pressed
Ranked causes
- 1. Switch Board failure (common)
- 2. Main Board failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No response to host communications
Unit does not communicate with external host system
Ranked causes
- 1. Power supply board fault (common)
- 2. Main board fault (occasional)
Check PSU board to host communications port cable
- Test point
- Rear panel host comms cable
- Expected
- Cable continuity confirmed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No response to Host Communications
Unit does not respond to host communications requests
Ranked causes
- 1. PSU Board fault (common)
- 2. Main Board fault (common)
Check cable continuity
- Test point
- PSU Board to Host Comms Port cable on rear panel
- Expected
- Continuity present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressing rotor switch fails to power on unit but menu selection is possible
Rotor selection works but unit cannot be powered on via rotor; unit appears to have power for other functions
Ranked causes
- 1. PSU Board fault (probable D3 short circuit) (common)
- 2. Switch Board failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote alarm output not operating
Remote alarm output does not activate when alarm condition is triggered
Ranked causes
- 1. Power supply board or main board fault (common)
- 2. Remote alarm cable continuity fault (occasional)
Stimulate alarm and verify remote alarm output
- Test point
- Remote alarm output socket
- Expected
- Remote alarm output activates when alarm is triggered
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotor skips menus when revolved
Menu selections skip steps when rotor is turned
Ranked causes
- 1. Rotor failure (common)
- 2. Switch Board failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotor switch press fails to power on unit
Pressing rotor switch does not turn unit on, but menu selection is possible
Ranked causes
- 1. Power supply board fault (common)
- 2. Switch board fault (occasional)
Test PSU board on/off control
- Test point
- PSU Board
- Expected
- PSU board responds to power-on signal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature calibration error exceeds compensation (TSH: SE_TACF)
Unit alarms during temperature calibration with error greater than compensation capability
Ranked causes
- 1. VREFI supply out of specification (common)
- 2. Main board failure (temperature reference or measurement circuit) (occasional)
Measure VREFI reference supply voltage
- Test point
- Main Board VREFI output
- Expected
- VREFI within specification
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature calibration failed at 54°F point (TSH: HE_54CP)
Unit alarms during temperature calibration with failure at 54°F point
Ranked causes
- 1. Main board isolated supply voltage out of specification (common)
- 2. Main board failure (reference or temperature measurement circuit) (occasional)
Measure isolated supply voltages on main board
- Test point
- Main Board isolated supply outputs (+5VI, +/-15VI, VREFI)
- Expected
- +5VI, +/-15VI, VREFI within specification
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature calibration failed at 98°F point (TSH: HE_98CP)
Unit alarms during temperature calibration with failure at 98°F point
Ranked causes
- 1. Main board isolated supply voltage out of specification (common)
- 2. Main board failure (reference or temperature measurement circuit) (occasional)
Measure isolated supply voltages on main board
- Test point
- Main Board isolated supply outputs (+5VI, +/-15VI, VREFI)
- Expected
- +5VI, +/-15VI, VREFI within specification
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on from battery or external DC source
Unit completely fails to switch on from either power source
Ranked causes
- 1. On/Off switch faulty (common)
- 2. 5V DC supply low (occasional)
Test rotor switch response
- Test point
- Rotor switch
- Expected
- Unit powers on when rotor switch pressed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on from battery source
Unit fails to switch on when powered by internal or external battery
Ranked causes
- 1. Battery discharged (common)
- 2. Power supply board fault (occasional)
Measure battery terminal voltage
- Test point
- Battery terminals
- Expected
- Battery voltage within specification
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not switch on from battery
Unit does not power on when powered by battery source
Ranked causes
- 1. Battery discharged (common)
- 2. Power supply unit board fault (occasional)
Measure battery terminal voltage
- Test point
- Battery terminals
- Expected
- Within normal operating range
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration timestamp recording
Recording the date and time when temperature calibration was completed in the unit's memory.
Ranked causes
- 1. Timestamp must be recorded after each temperature calibration (common)
Record timestamp after temperature calibration
- Test point
- TEMP > CAL menu after calibration steps complete
- Expected
- Current date and time recorded
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP transducers
Verify NIBP pressure transducers (PT1 main, PT2 pneumatic assembly) read correctly; used for checking and recalibrating pressure measurement circuitry
Ranked causes
- 1. NIBP transducer calibration drift or error (unknown)
Verify NIBP transducer baseline reading with no hose attached
- Test point
- PT1 (main transducer) and PT2 (pneumatic assembly) displays
- Expected
- 000 for both systolic and diastolic
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pop Off (overpressure detection) patient safety
Verify and calibrate the overpressure detection circuitry to ensure proper pop-off response during NIBP measurement cycles
Ranked causes
- 1. Pop Off calibration drift (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Record temperature calibration timestamp
Enter current date and time as the calibration timestamp for temperature sensor calibration record.
Ranked causes
- 1. Temperature calibration timestamp recording required (common)
Record temperature calibration timestamp
- Test point
- monitor service menu CAL function
- Expected
- Current date and time recorded as calibration timestamp
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Temperature sensor and algorithm
- NIBP measurement timeout - inflation (N33 error)
- NIBP measurement timeout inflation error (N33)
- Remote Alarm Output not operating
- Unit alarms 'PIC: Incompatible SW Version'
- Unit alarms 'TSH: HE_54CP'
- Unit alarms 'TSH: HE_98CP'
- Unit alarms 'TSH: SE_TACF'
8 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E00 | Temperature: fail (determination time exceeded) | Removal of probe before temperature determination complete or exceeds 60 sec | error |
E11 | Temperature: fail (tissue contact loss) | Predictive temperature loss of tissue contact > 30 sec | error |
E33 | Temperature: fail | Temperature probe not connected or inoperable | error |
Low Battery | Low battery condition | Battery voltage drops below threshold | warning |
N00 | NIBP overpressure warning during inflation | Cuff pressure exceeds normal limits during measurement | warning |
N00 | Overpressure detected | System pressure whilst idle 5 or 15 mmHg above ambient for 30 seconds | error |
N00 | NIBP overpressure | Pressure in NIBP system exceeds safe limits or cuff fails to deflate | error |
N12 | NIBP system state error - cuff fails to deflate | Valve control circuitry prevents cuff deflation or system state does not match software expectation | critical |
N12 | System pressure error (idle) | System pressure whilst idle 5 or 15 mmHg above ambient for 30 sec (neonate or adult cuff) | critical |
N12 | Pneumatic system state mismatch with valve control fault preventing deflation | Pressure in system exceeds ambient despite being set open or valve circuit failure detected | critical |
N13 | Pump running continuously or at improper times | Pump remains on when it should be off or runs continuously on power-up | critical |
N13 | Pump status mismatch | Pump on/off status does not concur with Main board software expected status | critical |
N13 | NIBP pump control error - pump does not turn off | Pump runs continuously on power up or remains on when it should be off | critical |
N33 | NIBP timeout - inflation phase | NIBP measurement fails to reach target pressure within timeout period | error |
N33 | NIBP timeout during inflation | Pressure fails to reach target during NIBP inflation within timeout period | error |
N44 | NIBP timeout total time | NIBP measurement exceeds total allowable time limit | warning |
N44 | NIBP timeout - total measurement time | NIBP measurement exceeds total allowable time | warning |
N55 | NIBP timeout - pressure measurement | NIBP pressure measurement fails within timeout period | warning |
N55 | NIBP timeout pressure | Pressure measurement channel fails or cuff deflation takes too long | warning |
N99 | NIBP no determination obtained | NIBP measurement cannot be completed or pulse determination fails | error |
N99 | NIBP failed | Unable to make an NIBP determination | error |
N99 | NIBP no determination | NIBP measurement algorithm unable to determine pressure values | error |
NIBP Range error | NIBP algorithm value out of range | NIBP algorithm returned value outside specified accuracy range | error |
N, P, E, I, S | Internal system fault | Other error codes not listed | critical |
P00 | SpO2: no sensor | SpO2 sensor not connected or no sensor code detected | error |
P00 | SpO2 probe disconnected | SpO2 probe signal missing or not detected | warning |
P00 | SpO2 probe disconnected | SpO2 probe not detected or connection lost | error |
P105 | System error - isolated supply or SpO2 comms failure | Power-up sequence detects isolated power supply problem or SpO2 RS232 communications failure | critical |
P105 | System error related to SpO2 isolated power supply | Power-up sequence complete with isolated supply fault | error |
P110 | System error related to SpO2 isolated power supply | Power-up sequence complete with isolated supply fault | error |
P110 | System error - isolated supply or SpO2 comms failure | Power-up sequence detects isolated power supply problem or SpO2 RS232 communications failure | critical |
P115 | System error - isolated supply or SpO2 comms failure | Power-up sequence detects isolated power supply problem or SpO2 RS232 communications failure | critical |
P115 | System error related to SpO2 isolated power supply | Power-up sequence complete with isolated supply fault | error |
P55 | SpO2 no signal | SpO2 measurement signal not detected from probe | warning |
P55 | SpO2 no signal | SpO2 probe connected but no valid signal detected | error |
PIC: +/-12V Out of Spec | Supply rails +12VR or -12VR operating outside specification | PSU Board PIC software detects voltage deviation during monitoring | error |
PIC: Incompatible SW Version | PIC software version mismatch between boards detected | Main Board and PSU Board software versions incompatible during initialization | error |
Printer - No Paper | Printer out of paper | Paper expired or printer door open | warning |
Pump Over Current | Pump over current condition detected by Main Board processor | Pump current exceeds safe limits or sense circuit failure | error |
TSH: HE_54CP | Unit failed temperature calibration for 54°F point | Temperature sensor calibration error during startup or maintenance | error |
TSH: HE_54CP | Temperature calibration failed at 54°F point | During temperature calibration, 54°F reference point measurement fails | error |
TSH: HE_98CP | Temperature calibration failed at 98°F point | During temperature calibration, 98°F reference point measurement fails | error |
TSH: HE_98CP | Unit failed temperature calibration for 98°F point | Temperature sensor calibration error during startup or maintenance | error |
TSH: SE_TACF | Temperature calibration error exceeding compensation capability | Temperature error greater than compensatable range detected | error |
TSH: SE_TACF | Temperature calibration error exceeds compensation range | Temperature calibration detects error greater than circuit can compensate | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.