Covidien LigaSure: troubleshooting and error codes
The ServiceForge knowledge base holds 62 documented symptoms for the Covidien LigaSure, with 122 ranked causes, 123 diagnostic checks with test points and expected values, and 126 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Generator does not respond when turned on high voltage
Unit does not power up or respond to power switch activation
Ranked causes
- 1. Disconnected power cord, faulty wall receptacle, or faulty power cord (common)
- 2. Loose or disconnected internal cables (common)
- 3. Fuse drawer open or fuses blown (common)
- 4. Faulty power entry module or connections (occasional)
- 5. Faulty low voltage power supply (occasional)
- 6. Damaged control board connectors or malfunctioning control board (occasional)
- 7. Faulty power switch (occasional)
- 8. Malfunctioning front panel components (occasional)
- 9. Shorts or disconnects on HV power supply (rare)
- 10. Shorts or disconnects on RF board (rare)
Verify power cord connections
- Test point
- Generator power inlet and wall receptacle
- Expected
- Power cord firmly connected at both ends
3 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator is on but did not complete self-test
Unit powers on but fails to complete self-test sequence during startup
Ranked causes
- 1. An alarm condition exists (common)
- 2. Software malfunction (occasional)
- 3. Loose or disconnected internal cables (occasional)
- 4. Faulty low voltage power supply (occasional)
- 5. Damaged control board connectors or malfunctioning control board (occasional)
- 6. Faulty power switch (rare)
- 7. Malfunctioning front panel components (rare)
- 8. Shorts or disconnects on HV power supply (rare)
- 9. Shorts or disconnects on RF board (rare)
Observe display for alarm number
- Test point
- Display screen
- Expected
- Alarm number displayed or none visible
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator on and activated but no output delivered
Generator powers on and accessory activates but no RF output is produced
Ranked causes
- 1. Malfunctioning footswitching or handswitching instrument (common)
- 2. Footswitch connected to wrong receptacle (common)
- 3. Handset connected to wrong receptacle (common)
- 4. Power set too low (common)
- 5. Alarm condition exists (occasional)
- 6. Blown fuse on power supply (occasional)
- 7. Control board malfunction (occasional)
- 8. RF output stage malfunction (occasional)
13 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buttons do not function correctly when pressed
Front panel buttons do not respond to button presses
Ranked causes
- 1. Faulty ribbon cable between control board and display board (common)
- 2. Incorrect information communicated through control board (occasional)
- 3. Faulty ribbon cable between front panel and display board (occasional)
- 4. Incorrect information from front panel (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch will not activate output
Footswitch does not trigger RF output generation
Ranked causes
- 1. Malfunctioning or damaged footswitch receptacle (common)
- 2. Footswitch activation signal lost on control board (occasional)
- 3. Wrong footswitch type connected (common)
- 4. Invalid smart code (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel button pressed during powerup test
One or more front panel buttons (MEMORY, SEAL_UP, SEAL_DOWN, MACRO_UP, MACRO_DOWN, COAG_UP, COAG_DOWN) appeared pressed during powerup key test.
Ranked causes
- 1. Front panel button physically stuck or pressed during powerup (common)
- 2. Front panel ribbon cable loose or faulty (occasional)
- 3. Display board ribbon cable loose or faulty (occasional)
- 4. Control board fault (occasional)
Power cycle without button presses
- Test point
- Power switch
- Expected
- No front panel button errors 126-132 on powerup
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank or confusing LED display
Display shows no information or incorrect data
Ranked causes
- 1. Faulty ribbon cable between control board and display board (common)
- 2. Incorrect display information communicated through control board (occasional)
- 3. Display board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous monitor interference caution
Monitor displays interference when generator or devices are in operation
Ranked causes
- 1. Faulty chassis-to-ground connections (common)
- 2. Electrical equipment grounding connected to different objects (occasional)
- 3. Malfunctioning monitor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback DOSERR circuit test failed
Powerup test of DOSERR circuitry failed. Failure detected by feedback processor.
Ranked causes
- 1. Control board DOSERR circuit fault (occasional)
- 2. High voltage power supply failure (occasional)
- 3. RF output FET Q5 failure on RF board (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- DOSERR test passes without error 145
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch key down during powerup test
One or more footswitch keys (VS_CUT, VS_COAG, BIP_CUT, BIP_COAG, VS_SEAL, VS_CUT_HAND) activated during powerup key test.
Ranked causes
- 1. Footswitch or accessory pressed during powerup or stuck (common)
- 2. Footswitch board ribbon cable loose or damaged (occasional)
- 3. Ribbon cable integrity failure or control board fault (occasional)
Power cycle without button presses
- Test point
- Power switch
- Expected
- No footswitch errors 120-125 on powerup
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch receptacle connection loose or damaged caution
Footswitch does not seat securely in rear panel receptacle or has visible damage
Ranked causes
- 1. Loose footswitch connector/receptacle mating (common)
- 2. Defective footswitch (occasional)
- 3. Damaged footswitch board assembly (occasional)
Inspect rear panel footswitch receptacles
- Test point
- bipolar footswitch receptacle (blue band) and vessel sealing footswitch receptacle (purple band)
- Expected
- No obstructions or visible damage
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handset receptacle connection loose or damaged caution
Bipolar handset does not seat securely in front panel receptacle or has visible damage; receptacle indicator does not illuminate green
Ranked causes
- 1. Loose handset connector/receptacle mating (common)
- 2. Defective handset (occasional)
- 3. Damaged front panel (occasional)
Inspect front panel handset receptacle
- Test point
- bipolar handset receptacle (blue) or vessel sealing handset receptacle (purple)
- Expected
- No obstructions or visible damage
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interference with other devices during generator activation caution
Other equipment experiences interference only when generator is activated
Ranked causes
- 1. Metal-to-metal sparking (common)
- 2. Electrically inconsistent ground wires in operating room (occasional)
- 3. Monitor responding to radiated frequencies (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Master microprocessor DOSERR test failure
Master microprocessor reported a powerup DOSERR test failure.
Ranked causes
- 1. Control board fault (occasional)
- 2. High voltage power supply failure (occasional)
- 3. RF output FET Q5 failure on RF board (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- DOSERR test passes without error 115
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No activation or alarm tone; speaker malfunctioning
Speaker fails to produce activation tone or alarm sound during operation
Ranked causes
- 1. Poor connection or damaged footswitch board ribbon cable (common)
- 2. Faulty connections or speaker on footswitch board (common)
- 3. Audio signal malfunction on control board (occasional)
Inspect footswitch board ribbon cable connection
- Test point
- Footswitch board ribbon cable connector
- Expected
- Cable firmly seated with no visible damage
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power calculation overflow in feedback
Feedback calculation of power overflowed.
Ranked causes
- 1. Sensor signal too high or control board fault (occasional)
- 2. Control board fault (occasional)
- 3. RF board fault (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- Power calculation without overflow, error 164
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Smart card reads invalid handset
System detects invalid handset when Smart Connector card is read
Ranked causes
- 1. Faulty handset (common)
- 2. Handset not valid for receptacle (common)
- 3. Faulty smart card (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal neuromuscular stimulation patient safety
Patient experiences abnormal neuromuscular stimulation; surgery should stop immediately
Ranked causes
- 1. Metal-to-metal sparking (common)
- 2. Abnormal 50-60 Hz leakage currents (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback A/D conversion error during powerup
Feedback processor detected an A/D error during power up.
Ranked causes
- 1. Control board A/D conversion fault (occasional)
- 2. Low voltage power supply failure (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- A/D test passes without error 146
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Master A/D conversion error during powerup
Master processor detected an A/D error during powerup.
Ranked causes
- 1. Control board A/D conversion fault (occasional)
- 2. Low voltage power supply failure (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- A/D test passes without error 134
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overcurrent during voltage sensor calibration caution
Overcurrent condition discovered during voltage sensor (Vsns) calibration.
Ranked causes
- 1. Load impedance higher than specified during calibration (common)
- 2. RF board malfunction (occasional)
Verify load impedance during calibration
- Test point
- Calibration load
- Expected
- Load impedance as specified in calibration instructions
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage during current sensor calibration caution
Overvoltage condition discovered during current sensor (Isns) calibration indicating load impedance issue.
Ranked causes
- 1. Load impedance higher than specified during calibration (common)
- 2. Control board malfunction (occasional)
Verify load impedance during calibration
- Test point
- Calibration load
- Expected
- Load impedance as specified in calibration instructions
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backed RAM CRC corrected
Battery backed RAM CRC test failed during initialization but data was restored from RTC data.
Ranked causes
- 1. Bad battery or failed memory chip (U3) (occasional)
Monitor for error recurrence
- Test point
- Unit operation
- Expected
- No recurrence of error 106
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backed RAM CRC failed at initialization
Battery backed RAM CRC test failed during initialization and correction attempt failed.
Ranked causes
- 1. Bad battery or failed memory chip (U3) (common)
Inspect and test battery
- Test point
- Battery on control board
- Expected
- Battery functional and holding charge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bipolar key press resulted in complete failure patient safety
A bipolar MACRO or bipolar COAG key press resulted in a COMPLETE_FAIL outcome from the feedback processor.
Ranked causes
- 1. Control board or feedback processor fault during RF output (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- Bipolar key press succeeds without error 170
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration data invalid at powerup
Calibration data invalid at powerup and could not be recovered.
Ranked causes
- 1. Corrupted or missing calibration data (common)
Run calibration procedure
- Test point
- Unit calibration system
- Expected
- Calibration completes successfully
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DOSERR ECON and HV_SEN mismatch during keying
Sys_Econ and HV_SNS do not compare within limits during keying.
Ranked causes
- 1. Control board or high voltage power supply fault (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- ECON and HV_SEN within limits during keying
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DOSERR test memory corrupted
Doserror test state variable corrupted during powerup dose error test.
Ranked causes
- 1. Control board memory fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- DOSERR test memory passes without error 116
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DPRAM communication failure at powerup
Master and feedback processors fail to communicate on power up. Timing or content of communication invalid.
Ranked causes
- 1. Control board communication or processor fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- DPRAM communication succeeds without error 118
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dual port RAM failed initialization
DPRAM failed test during initialization.
Ranked causes
- 1. DPRAM chip failure or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- DPRAM passes test without error 103
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error counts CRC invalid at powerup
Error counts kept in battery backed RAM have invalid CRC at powerup.
Ranked causes
- 1. Battery failure or control board fault (common)
Monitor system status
- Test point
- Battery backed RAM
- Expected
- No CRC errors in error counts
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error disable variables corrupted in master
Master error disable variables are equal when they should be inverse: mm_errors_enabled == curr_settings.errs_disabled.
Ranked causes
- 1. Memory corruption in control board (rare)
Power cycle the unit
- Test point
- Power switch
- Expected
- Error variables correctly inverse without error 163
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback calibration data corrupted
Feedback processor detected corrupted calibration data during powerup.
Ranked causes
- 1. Calibration data corruption or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- Calibration data valid without error 143
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback DOSERR test memory corrupted
Feedback processor detected corrupted memory during dose error circuit test at powerup.
Ranked causes
- 1. Memory corruption or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- DOSERR memory test passes without error 144
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback dual port RAM error during powerup
Feedback processor detected an error in dual port RAM during powerup.
Ranked causes
- 1. DPRAM failure or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- DPRAM test passes without error 141
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback ECON calculation overflow
Feedback calculation of starting ECON overflowed.
Ranked causes
- 1. Control board calculation or processor fault (rare)
Power cycle the unit
- Test point
- Power switch
- Expected
- No ECON overflow error 162
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback error disable variables corrupted
Feedback error disable variables are equal when they should be inverse: fb_errors_enabled == local_curr_settings.errs_disabled.
Ranked causes
- 1. Memory corruption in feedback processor (rare)
Power cycle the unit
- Test point
- Power switch
- Expected
- Feedback error variables correctly inverse without error 169
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback external memory error during powerup
Feedback processor detected an error in external memory during powerup.
Ranked causes
- 1. External RAM failure or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- External memory test passes without error 140
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feedback ROM CRC error during powerup
Feedback processor detected an error in program memory during powerup.
Ranked causes
- 1. Flash memory corruption or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- ROM CRC test passes without error 142
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flash memory CRC test failed
CRC test of ST Microelectronics Flash Memory failed during initialization.
Ranked causes
- 1. Flash memory corruption or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- Flash CRC passes test without error 104
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not detect instrument or adapter
Generator fails to recognize vessel sealing instrument, bipolar instrument, or Smart Connector Adapter
Ranked causes
- 1. Instrument not properly inserted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid error code from powerup self test
Bad error code supplied by powerup self test routine on the master side during initialization.
Ranked causes
- 1. Control board self-test failure (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- Unit initializes without error 100
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid output request during feedback
Feedback enable_output_relays detected illegal combination of generator mode and RF output side for output relay configuration.
Ranked causes
- 1. Control board firmware or hardware fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- Unit restarts without error 025
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Last power settings CRC invalid at powerup
Last power settings structure in battery backed RAM has invalid CRC at powerup.
Ranked causes
- 1. Battery failure or control board fault (common)
Monitor system status
- Test point
- Battery backed RAM
- Expected
- No CRC errors in power settings
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Master unable to reset feedback microprocessor
Master unable to reset feedback microprocessor at initialization.
Ranked causes
- 1. Feedback microprocessor or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- Microprocessor resets successfully without error 101
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microprocessor compatibility error
Incompatible software versions on feedback and master processors or master side software incompatible with digital board type.
Ranked causes
- 1. Software version mismatch between processors or boards (occasional)
Power cycle the unit
- Test point
- Power switch
- Expected
- Software versions compatible without error 117
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Minutes on data CRC invalid at powerup
Minutes on data kept in battery backed RAM has invalid CRC at powerup.
Ranked causes
- 1. Battery failure or control board fault (common)
Monitor system status
- Test point
- Battery backed RAM
- Expected
- No CRC errors in minutes on data
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker interference patient safety
Pacemaker experiences interference from generator operation
Ranked causes
- 1. Intermittent connections or metal-to-metal sparking (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power calculation overflow in master
Master calculation of power overflowed.
Ranked causes
- 1. Control board processor or calculation fault (rare)
Power cycle the unit
- Test point
- Power switch
- Expected
- Power calculation without overflow, error 165
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock CRC corrected
Real time clock RAM CRC test failed during initialization but data was restored from battery backed RAM.
Ranked causes
- 1. Bad battery or failed RTC chip (U1) (occasional)
Monitor for error recurrence
- Test point
- Unit operation
- Expected
- No recurrence of error 108
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock CRC failed at initialization
Real time clock RAM CRC test failed during initialization and correction attempt failed.
Ranked causes
- 1. Bad battery or control board fault (common)
Verify battery condition
- Test point
- Battery on control board
- Expected
- Battery functional
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock valid RAM bit erroneous
Real time clock reported an erroneous Valid RAM and time bit during initialization.
Ranked causes
- 1. Battery failure or RTC chip malfunction (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- RTC valid bit correct without error 119
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recent error data CRC invalid at powerup
Recent error data kept in battery backed RAM has invalid CRC at powerup.
Ranked causes
- 1. Battery failure or control board fault (common)
Monitor system status
- Test point
- Battery backed RAM
- Expected
- No CRC errors in error data
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remove generator cover caution
Cover removal requires turning off and unplugging generator, removing three M4×.7×8 screws, sliding cover toward rear, lifting slightly at catch point, then continuing to rear.
Ranked causes
- 1. Access to internal components required (unknown)
Turn off generator
- Test point
- Power switch
- Expected
- Generator powered off
12 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace battery (battery-backed RAM)
Battery replacement involves removing control board, then removing spent battery and installing new 3 V button cell with positive side visible, then reinstalling board and recalibrating.
Ranked causes
- 1. Battery depleted or failed; calibration data lost (unknown)
Remove cover and control board per control board removal procedure
- Test point
- Control board removed from chassis
- Expected
- Control board removed and accessible
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace control board high voltage
Control board requires replacement; procedure involves removing cover, disconnecting display board ribbon cable, removing three screws and brackets, then sliding board up to disconnect.
Ranked causes
- 1. Control board failure or damage (unknown)
Turn off generator and disconnect power cord
- Test point
- Power source
- Expected
- Generator powered off and disconnected
19 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace HV power supply board high voltage
HV power supply board requires replacement; procedure involves removing cover, heat sink bracket, four cables, three screws, then lifting board out of chassis.
Ranked causes
- 1. HV power supply board failure or damage (unknown)
Turn off generator and disconnect power cord from wall receptacle
- Test point
- Power source
- Expected
- Generator powered off and disconnected
20 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Seal parameters invalid on powerup
SEAL parameters in both BATRAM and RTC are bad at power up. Parameters loaded from defaults.
Ranked causes
- 1. Battery failure or control board fault (common)
Verify SEAL parameters loaded from defaults
- Test point
- Unit SEAL parameters
- Expected
- SEAL parameters at default values without error 133
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ST Microelectronics RAM failed initialization
ST Microelectronics external RAM failed test during initialization.
Ranked causes
- 1. RAM chip failure or control board fault (common)
Power cycle the unit
- Test point
- Power switch
- Expected
- RAM passes test without error 102
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Statistical data CRC invalid at powerup
Statistical data kept in battery backed RAM has invalid CRC at powerup.
Ranked causes
- 1. Battery failure or control board fault (common)
Monitor system status
- Test point
- Battery backed RAM
- Expected
- No CRC errors in statistical data
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Tone average off reading deviation
- Tone average reading deviation at calibration
2 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
001 | FB_UNREPORTABLE_ERROR | Master processor detects error and directs feedback to system error state | critical |
003 | MASTER_DPRAM_TEST_TIMEOUT | Master failed to complete DPRAM test in allotted time | critical |
004 | MASTER_CAL_DATA_TIMEOUT | Master failed to place calibration data in DPRAM in allotted time | critical |
005 | FB_CAL_CHECK_TIMEOUT | Feedback failed to check calibration data in DPRAM in allotted time | critical |
006 | AD_CONVERSION_TIMEOUT | Master side A to D conversion failed to complete in allotted time | critical |
007 | FB_AD_CONVERSION_TIMEOUT | Feedback side A to D conversion failed to complete in allotted time | critical |
008 | MASTER_SETTINGS_TIMEOUT | Master failed to write settings into DPRAM within allotted time for feedback microprocessor | critical |
009 | POST_COMPLETE_TIMEOUT | Master failed to signal initialization complete within allotted time | critical |
010 | INVALID_ERROR_CODE_ALARM | Invalid error code sent to error handler | warning |
011 | FB_POST_ERRORS | Errors detected in feedback microprocessor during power on self test (POST) | critical |
012 | FB_INTERNAL_MEMORY_FAULT | Main state variable for feedback microprocessor corrupted to impossible value | critical |
013 | FB_CORRUPTED_STATE_VAR | Main feedback state variable changed to sys_err incorrectly | critical |
014 | INVALID_BIP_OPS | Invalid value in available operations variable read from smart connector table | error |
015 | INVALID_VS_OPS | Invalid value in available operations variable read from smart connector table | error |
016 | FB_ILLEGAL_STATE_TRANSITION | Combination of new state requested by master and current feedback microprocessor state represents illegal transition | error |
017 | FB_ILLEGAL_STATE_VALUE | Illegal state value detected in feedback state transition checking function | error |
018 | FB_ILLEGAL_POWER_DEACTIVATION_STATE_VALUE | Illegal state value detected in feedback power deactivation state machine state variable | error |
019 | FB_INVALID_MODE_CHANGE | Attempt made to transition directly from one RF output mode to another without first disabling RF from master | error |
020 | FB_INVALID_POWER_CHANGE | Attempt made to change power output while SEAL output was activated | error |
021 | FB_INVALID_MODE_COMMAND | RF output mode (function) sent from master to feedback was invalid | error |
022 | FB_INVALID_CAL_DATA | Calibration data copied from DPRAM to local RAM has invalid CRC and is unusable | error |
023 | FB_INVALID_FEEDBACK_MODE | Invalid feedback mode state variable value | error |
024 | MASTER_FPANEL_TEST_TIMEOUT | Master failed to signal feedback that front panel testing completed in time | critical |
025 | Feedback enable_output_relays detected illegal combination of generator mode and RF output side for output relay configuration. | Invalid output relay configuration during initialization. | error |
026 | Overvoltage condition discovered during current sensor (Isns) calibration. | Excessive voltage detected on Isns line during calibration. | error |
027 | Overcurrent condition discovered during voltage sensor (Vsns) calibration. | Excessive current detected during Vsns calibration. | error |
100 | Bad error code supplied by powerup self test routine on the master side. | Self-test routine failure during initialization. | error |
101 | Master unable to reset feedback microprocessor at initialization. | Feedback microprocessor reset command failed. | error |
102 | ST Microelectronics external RAM failed test during initialization. | RAM memory test failure during powerup. | error |
103 | DPRAM failed test during initialization. | Dual port RAM memory test failure. | error |
104 | CRC test of ST Microelectronics Flash Memory failed during initialization. | Flash memory corruption detected. | error |
105 | Calibration data invalid at powerup and could not be recovered. | Missing or corrupted calibration parameters at startup. | error |
105 | Battery has been replaced | Appears after CPU battery replacement when generator is powered on | info |
106 | Battery backed RAM CRC test failed but data restored from RTC data. | CRC error in battery-backed memory during powerup. | warning |
107 | Battery backed RAM CRC test failed during initialization and correction attempt failed. | CRC error in battery-backed memory that could not be corrected. | error |
108 | Real time clock RAM CRC test failed but data restored from battery backed RAM. | CRC error in RTC memory during powerup. | warning |
109 | Real time clock RAM CRC test failed and correction attempt failed. | CRC error in RTC memory that could not be corrected. | error |
110 | Recent error data in battery backed RAM has invalid CRC at powerup. | CRC error in error history memory. | warning |
111 | Error counts in battery backed RAM have invalid CRC at powerup. | CRC error in error count memory. | warning |
112 | Last power settings in battery backed RAM have invalid CRC at powerup. | CRC error in power settings memory. | warning |
113 | Statistical data in battery backed RAM has invalid CRC at powerup. | CRC error in statistical data memory. | warning |
114 | Minutes on data in battery backed RAM has invalid CRC at powerup. | CRC error in usage time data memory. | warning |
115 | Master microprocessor reported a powerup DOSERR test failure. | DOSERR circuit test failure during initialization. | error |
116 | Doserror test state variable corrupted during powerup dose error test. | Memory corruption during DOSERR test. | error |
117 | Incompatible software versions between processors or master software incompatible with digital board type. | Software version mismatch detected at powerup. | error |
118 | Master and feedback processors fail to communicate on power up. | DPRAM communication failure between processors. | error |
119 | Real time clock reported erroneous Valid RAM and time bit during initialization. | RTC VRT bit indicates invalid RAM or time. | error |
120 | VS footswitch MACRO (cut) keying signal activated during powerup key test. | Footswitch pressed or stuck during initialization. | warning |
121 | VS footswitch COAG keying signal activated during powerup key test. | Footswitch pressed or stuck during initialization. | warning |
122 | BIP footswitch MACRO keying signal activated during powerup key test. | Footswitch pressed or stuck during initialization. | warning |
123 | BIP footswitch COAG keying signal activated during powerup key test. | Footswitch pressed or stuck during initialization. | warning |
124 | VS footswitch SEAL keying signal activated during powerup key test. | Footswitch pressed or stuck during initialization. | warning |
125 | VS handswitch MACRO/SEAL keying signal activated during powerup key test. | Handswitch pressed or stuck during initialization. | warning |
126 | Memory button appeared pressed during powerup key test. | Memory button activation detected during initialization. | warning |
127 | Seal power increment button appeared pressed during powerup. | Button activation detected during initialization. | warning |
128 | Seal power decrement button appeared pressed during powerup. | Button activation detected during initialization. | warning |
129 | Macrobipolar power increment button appeared pressed during powerup. | Button activation detected during initialization. | warning |
130 | Macrobipolar power decrement button appeared pressed during powerup. | Button activation detected during initialization. | warning |
131 | Bipolar power increment button appeared pressed during powerup. | Button activation detected during initialization. | warning |
132 | Bipolar power decrement button appeared pressed during powerup. | Button activation detected during initialization. | warning |
133 | SEAL parameters in both BATRAM and RTC are bad at power up. | Missing or corrupted SEAL device parameters. | warning |
134 | Master processor detected an A/D conversion error during powerup. | Analog-to-digital converter test failure. | error |
140 | Feedback processor detected an error in external memory during powerup. | External RAM test failure. | error |
141 | Feedback processor detected an error in dual port RAM during powerup. | DPRAM test failure. | error |
142 | Feedback processor detected an error in program memory during powerup. | Flash memory ROM CRC failure. | error |
143 | Feedback processor detected corrupted calibration data during powerup. | Calibration data CRC failure. | error |
144 | Feedback processor detected corrupted memory during dose error circuit test at powerup. | Memory corruption detected by DOSERR test. | error |
145 | Powerup test of DOSERR circuitry failed. Failure detected by feedback processor. | DOSERR circuit test failure. | error |
146 | Feedback processor detected an A/D conversion error during power up. | Analog-to-digital converter test failure. | error |
162 | Feedback calculation of starting ECON overflowed. | ECON parameter calculation overflow. | error |
163 | Master error disable variables are equal when they should be inverse. | Memory corruption of error control variables. | error |
164 | Feedback calculation of power overflowed. | Power calculation overflow during operation. | error |
165 | Master calculation of power overflowed. | Power calculation overflow in master processor. | error |
166 | Sys_Econ and HV_SNS do not compare within limits during keying. | ECON and HV sensor mismatch during RF output. | error |
167 | ton_avg reading is too different from calibrated ton_avg value. | Tone average deviation from calibration baseline. | error |
168 | ton_avg reading when no RF output is too different from calibrated off value. | Tone average off-state deviation from baseline. | error |
169 | Feedback error disable variables are equal when they should be inverse. | Memory corruption of feedback error control variables. | error |
170 | Bipolar MACRO or COAG key press resulted in COMPLETE_FAIL outcome from feedback processor. | RF output key press resulted in complete failure. | critical |
172 | 5 volt low voltage power supply analog input exceeded 2.5 volts | EXCESSIVE_SYSTEM_VOLTAGE detected | error |
174 | Invalid mode value in last_settings.curr_mode variable | INVALID_LAST_MODE_VALUE detected by record_keying_event function | error |
175 | Data mismatch in feedback microprocessor communication | FB_RENDEZVOUS_INVALID_ECHO - echoed data does not match original | error |
176 | Feedback microprocessor failed to respond in time | FB_RENDEZVOUS_TIMEOUT - settings rendezvous flag not answered | error |
177 | Generator state machine in illegal RECOVERABLE_ERROR state | MNS_ILLEGAL_STATE_VALUE detected in manage_normal_state | error |
178 | Generator state machine has undefined state value | MNS_INVALID_STATE_VALUE detected in manage_normal_state routine | error |
179 | Generator state machine in illegal state for current context | MRE_INVALID_STATE_VALUE - next_generator_state is illegal | error |
180 | ISO hardware isolation test failure, key input detected during test | ISO_TEST_FAILURE - key read as down during ISO test in progress | error |
181 | Invalid debounce state value in key input processing | INVALID_KEY_STATE_VALUE detected by process_key_input function | error |
182 | Invalid overall key state value in input processing | INVALID_OVERALL_KEY_STATE detected by process_key_inputs | error |
183 | Invalid button debounce state in simple button processing | PSB_INVALID_BUTTON_STATE detected by process_simple_button | error |
184 | Invalid button debounce state in auto-repeat button processing | PARB_INVALID_BUTTON_STATE detected by process_auto_repeat_button | error |
185 | Invalid button debounce state in overall button input processing | PBI_INVALID_BUTTON_STATE detected by process_button_inputs | error |
186 | Memory corruption detected in button input state variables | PBI_INTERNAL_MEMORY_FAULT - inconsistent button state values | critical |
187 | Invalid value in key_down variable during mode matching | VKI_INVALID_MODE_DOWN detected by key_capability_match function | error |
188 | Invalid key state value in key input validation | VKI_INVALID_KEY_STATE_VALUE detected by validate_key_inputs | error |
189 | Invalid or corrupted key value in key_down variable | VKI_INVALID_KEY_DOWN detected by validate_key_inputs function | error |
190 | Invalid or corrupted button identifier in button results processing | MBR_INVALID_BUTTON_DOWN detected by manage_button_results | error |
191 | Power settings corrupted or macro/bipolar power exceeds 95 watts | MBR_POWER_SETTINGS_CORRUPTED detected by manage_button_results | error |
192 | Invalid audio state value in audio management | MA_INVALID_AUDIO_STATE detected by manage_audio function | error |
193 | Invalid mode value in audio management | MA_INVALID_MODE_VALUE in curr_settings.curr_mode detected | error |
194 | Invalid mode value in audio tone selection | SAT_INVALID_MODE_VALUE in curr_settings.curr_mode detected | error |
195 | Invalid alert type value in audio alert transition | MA_AT_INVALID_ALERT_TYPE detected in alert type variable | error |
196 | Calibration data corrupted in dual port RAM | CCD_CAL_DATA_CORRUPTED - invalid CRC in dpram cal store | error |
197 | Invalid generator state value in main function | MAIN_INVALID_STATE_VALUE detected in current_generator_state | error |
198 | Invalid serial message type in result execution | SNESR_INVALID_MESSAGE_TYPE in pending_serial_cmd.type field | error |
199 | Invalid LED mask parameter passed to LED update function | USL_INVALID_LED_MASK detected by update_smart_leds | error |
214 | Feedback failed to complete DPRAM test within allotted time | FB_INIT_ERRORS_TIMEOUT - initialization test timeout | error |
215 | Master main loop delay timer failed to expire | MM_MAIN_LOOP_DELAY_TIMEOUT - timer expiration failure | error |
216 | Feedback main loop delay timer failed to expire | FB_MAIN_LOOP_DELAY_TIMEOUT - feedback timer expiration failure | error |
217 | Invalid seal power watts value requested for display | MM_DSP_INVALID_SEAL_POWER - invalid watts value parameter | error |
218 | Master timed out waiting for feedback model data signal | FB_MODEL_DATA_TIMEOUT - DPRAM signal from feedback timeout | error |
219 | Front panel test sequence did not complete as scheduled | FPANEL_TEST_TIMEOUT - lamps, LEDs, displays, tones test timeout | error |
220 | Generator state machine in illegal RECOVERABLE_ERROR state | MCS_ILLEGAL_STATE_VALUE - next_generator_state is illegal | error |
221 | Invalid undefined state value in calibration state machine | MCS_INVALID_STATE_VALUE detected in manage_calibration_state | error |
222 | Invalid or undefined calibration step state value | MCS_ILLEGAL_CAL_STATE_VALUE in manage_calibration_state routine | error |
223 | Calibration state machine in illegal state | CAL_ILLEGAL_STATE_VALUE - undefined calibration state detected | error |
224 | Invalid key_down variable value in key capability matching | KCM_INVALID_KEY_DOWN - key_capability_match called with invalid value | error |
225 | Excessive idle reading on current or voltage sensor with no RF active | DOSRR_EXCESS_IDLE_READING - more than half scale with no keying | error |
226 | System economy over 90% full scale with low voltage/current readings | DOSRR_ECON_RISING - sys_econ > 90%, V_sns and I_sns < 10% | error |
227 | Power difference between master and feedback sensors exceeds 50 watts | DOSRR_EXCESS_DIFFERENCE - power computation mismatch | error |
229 | Master side current sensor is saturated | DOSRR_CURRENT_SATURATION - current measurement at saturation | error |
230 | Master side voltage sensor is saturated | DOSRR_VOLTAGE_SATURATION - voltage measurement at saturation | error |
231 | Master and feedback computed desired values differ | DOSRR_DESIRED_VAL - value computation mismatch between processors | error |
232 | Power on master side exceeds IEC 60601-2-2 excess power limit | DOSRR_POWER_EXCEEDED - computed power > limit specified by standard | error |
233 | Seal deviation calibration data is corrupted or invalid | CCD_SEAL_DEV_DATA_CORRUPTED - invalid data in curr_seal_dev in dpram | error |
234 | Feedback detected illegal power activation state value | FB_INVALID_POWER_ACTIVATION_STATE_VALUE - illegal state in variable | error |
451 | Temperature has risen past the allowable range | Over Temperature Error - thermal limit exceeded | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.