Covidien Kangaroo ePump: troubleshooting and error codes
The ServiceForge knowledge base holds 87 documented symptoms for the Covidien Kangaroo ePump, with 132 ranked causes, 95 diagnostic checks with test points and expected values, and 79 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
System error displayed patient safety
Device display shows error number with fixed audible alarm; requires power-down to clear
Ranked causes
- 1. Internal timer stopped (rare)
- 2. Flash chip missing or version incompatible (rare)
- 3. Too many graphic screens in memory (rare)
- 4. Motor current ADC interrupt service routine never triggered (rare)
- 5. Battery level ADC interrupt never triggered (rare)
- 6. Battery temperature ADC interrupt never triggered (rare)
- 7. Ultrasonic occlusion/bag-empty ADC interrupt never triggered (rare)
- 8. Ultrasonic downstream occlusion ADC interrupt never triggered (rare)
- 9. RS232 communications transmission error (occasional)
- 10. Flash write error (rare)
- 11. Flash read error (rare)
- 12. General stopcock timeout (occasional)
- 13. Buzzer test timeout (rare)
- 14. Flash screen read error (rare)
- 15. Flash languages read error (rare)
- 16. Cannot find stopcock flush position during prime (occasional)
- 17. Cannot find stopcock feed position during prime (occasional)
- 18. Motor error during tubing autoload (occasional)
- 19. Motor failsafe timeout; possible damaged gearbox encoder or flex cable (occasional)
- 20. Display lockup timeout (rare)
- 21. Motor control failure (occasional)
21 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error with numeric code patient safety
Display shows system error screen with error number; pump does not operate; requires power-down to exit
Ranked causes
- 1. Internal timer stopped (Error 0) (unknown)
- 2. Flash chip missing or incompatible (Error 1) (unknown)
- 3. ADC interrupt service routine failure (Errors 4-8) (unknown)
- 4. Communication or hardware timeout (Errors 9-14, 20-23, 33, 50-52) (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feed error - formula delivery failure patient safety
Feed Error screen appears; enteral formula no longer being delivered due to empty bag or clog between pump and bag
Ranked causes
- 1. Feed or Flush containers are empty (common)
- 2. Feed or Flush tube occluded between bag and pump (common)
- 3. Tube occluded between pump and patient (occasional)
Check feed container status
- Test point
- Feed bag/container
- Expected
- Container has adequate formula level
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feed error - formula not being delivered patient safety
FEED ERROR screen; enteral formula no longer flowing because bag is empty or clog between pump and bag
Ranked causes
- 1. Feed bag is empty (common)
- 2. Occlusion/clog between pump and feed bag (common)
- 3. Feed tube kinked or damaged (occasional)
Inspect feed bag for formula level
- Test point
- Feed bag
- Expected
- Bag contains sufficient formula
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow error - solution not reaching patient patient safety
FLOW ERROR screen; feeding or flushing solution no longer flowing to patient due to occlusion between pump and patient or valve contamination
Ranked causes
- 1. Occlusion between pump and patient (common)
- 2. Valve pocket contamination or moisture (occasional)
- 3. Pump set damaged or improper loading (occasional)
Visually inspect patient line for blockage
- Test point
- Patient line between pump outlet and patient
- Expected
- Line clear with no visible obstruction
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush error - flushing solution not being delivered
FLUSH ERROR screen; flushing solution no longer flowing because flush bag is empty or clog between pump and flush bag
Ranked causes
- 1. Flush bag is empty (common)
- 2. Occlusion/clog between pump and flush bag (common)
- 3. Flush tube kinked or damaged (occasional)
Inspect flush bag for solution level
- Test point
- Flush bag
- Expected
- Bag contains sufficient flush solution
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not hold charge adequately
Backup battery depletes in less than 15 hours at normal feed rates, or fails to recharge when plugged in
Ranked causes
- 1. Battery degraded or defective (rare)
- 2. Battery not fully charged before use (common)
Verify adequate charging time before battery-only operation
- Test point
- Battery charge indicator on display
- Expected
- Full battery symbol shown after 8 hours at A/C outlet
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feed error - no formula delivery patient safety
Feed Error screen appears; enteral formula no longer being delivered to patient
Ranked causes
- 1. Feed bag empty (common)
- 2. Occlusion between pump and feed bag (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feed error during formula delivery patient safety
FEED ERROR screen appears when enteral formula is no longer being delivered, indicating empty bag or clog between pump and bag.
Ranked causes
- 1. Feed bag empty (common)
- 2. Clog between pump and feed bag (common)
Verify feed bag status
- Test point
- feed bag
- Expected
- Bag contains formula
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow error - no solution delivery to patient patient safety
Flow Error screen appears; feeding or flushing solution not reaching patient due to downstream occlusion
Ranked causes
- 1. Occlusion between pump and patient (common)
- 2. Dampness or dirt in valve pocket causing false error (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow Error appears on screen patient safety
FLOW ERROR screen displayed; feeding or flushing solution no longer being delivered
Ranked causes
- 1. Clog between pump and patient (common)
- 2. Dampness or dirt in valve pocket (occasional)
Inspect line for occlusion
- Test point
- Feeding/flushing line from pump to patient
- Expected
- No blockage present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow error during solution delivery patient safety
FLOW ERROR screen appears when feeding or flushing solution is no longer being delivered due to clog between pump and patient.
Ranked causes
- 1. Clog between pump and patient (common)
- 2. Dampness or dirt in valve pocket (occasional)
Inspect patient delivery line for blockage
- Test point
- line from pump to patient
- Expected
- Clear, patent
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush error - no flush solution delivery
Flush Error screen appears; flush solution no longer being delivered
Ranked causes
- 1. Flush bag empty (common)
- 2. Occlusion between pump and flush bag (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush Error appears on screen patient safety
FLUSH ERROR screen displayed; flushing solution no longer being delivered
Ranked causes
- 1. Flush bag empty (common)
- 2. Clog between pump and flush bag (common)
Verify flush bag has solution
- Test point
- Flush bag
- Expected
- Bag contains flushing solution
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush error during flush delivery patient safety
FLUSH ERROR screen appears when flushing solution is no longer being delivered due to empty flush bag or clog between pump and flush bag.
Ranked causes
- 1. Flush bag empty (common)
- 2. Clog between pump and flush bag (common)
Verify flush bag status
- Test point
- flush bag
- Expected
- Contains flushing solution
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lines not primed before feeding patient safety
Air in tubing may prevent proper fluid delivery to patient
Ranked causes
- 1. Auto prime not selected or incomplete (common)
- 2. Feeding attempted after bag emptied without re-priming (occasional)
Verify AUTO PRIME completed successfully
- Test point
- Pump status display
- Expected
- AUTO PRIME COMPLETE message displayed
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not power on
Device fails to respond when POWER button is pressed on lower right of front panel
Ranked causes
- 1. Battery not connected or depleted (common)
- 2. A/C power not connected or faulty (common)
Verify A/C power cord is properly connected
- Test point
- A/C Adapter Port on back of pump
- Expected
- Plug fully inserted with proper pin/tab orientation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump flow rate inaccurate
Water collected during 30-minute test falls outside expected range for programmed hourly rate
Ranked causes
- 1. Pump Set performance degradation or improper calibration (unknown)
- 2. Rotor assembly fouling or misalignment (occasional)
Verify test conditions and pump setup
- Test point
- Room temperature, pump power source, bag height
- Expected
- Room temperature 22°C ± 2°C (72°F ± 3°F), A/C power connected (not battery), bag bottom 18 inches above pump, ±2°C
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump priming not working
Auto prime feature does not progress or manual prime button does not deliver fluid to valve and tubing
Ranked causes
- 1. Pump Set not properly loaded (common)
- 2. Incorrect priming procedure for set type (occasional)
Verify Pump Set is properly loaded
- Test point
- Blue door, valve finger tab, black ring retainer, rotor
- Expected
- SET LOADED displayed on status line
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump stops running unexpectedly patient safety
Pump halts operation during feeding without operator pressing HOLD or POWER button
Ranked causes
- 1. A/C power interrupted and battery depleted (common)
- 2. Error condition triggered (occasional)
Check A/C power status and battery charge indicator
- Test point
- Power cord connection and battery symbol on display
- Expected
- A/C power connected or battery charge showing 2/3 or greater
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery discharge during long-term storage
Battery may discharge and deactivate during non-usage or storage periods exceeding 9 months.
Ranked causes
- 1. Prolonged storage without battery management (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low
BATTERY LOW screen appears with continuous beeping alarm when battery requires recharging. Approximately 15 minutes of battery life remaining.
Ranked causes
- 1. Battery charge depleted (common)
Plug pump into AC power
- Test point
- AC adapter connection
- Expected
- Power indicator shows charging
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Low alarm activates
BATTERY LOW screen appears and alarm beeps continuously; approximately 15 minutes of battery life remaining
Ranked causes
- 1. Battery charge depleted (common)
Connect pump to AC power
- Test point
- AC adapter to wall outlet
- Expected
- Pump plugged in and charging (pump continues operation or returns to prior screen)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low warning
BATTERY LOW screen with continuous beep alarm; approximately 15 minutes of battery remaining before automatic shutdown
Ranked causes
- 1. Battery charge depleted (common)
Connect power cord to wall outlet
- Test point
- Power connection
- Expected
- AC power available at outlet
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bolus interval out of specifications patient safety
Programmed time between bolus deliveries does not match clinical protocol in intermittent mode
Ranked causes
- 1. Bolus interval incorrectly programmed (common)
Verify bolus interval setting
- Test point
- ADJUST BOLUS screen BOLUS INTERVAL field
- Expected
- Interval programmed between 1 and 24 hours as clinically ordered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bolus rate out of specifications patient safety
Programmed bolus delivery rate does not match clinical order in intermittent mode
Ranked causes
- 1. Bolus rate incorrectly programmed (common)
Verify bolus rate setting
- Test point
- ADJUST BOLUS screen BOLUS RATE field
- Expected
- Rate programmed between 1 and 400 ml/hr as ordered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AccuSystem feeding set compatibility patient safety
Ensure KANGAROO ePump is calibrated and tested with genuine AccuSystem√ feeding sets to maintain accurate delivery. Verify feeding set has AccuSystem√ logo.
Ranked causes
- 1. Use of unauthorized or generic feeding sets (common)
Inspect feeding set packaging and labeling for AccuSystem√ logo
- Test point
- Feeding set exterior
- Expected
- AccuSystem√ logo present on genuine product
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuitry damage from unauthorized rework patient safety
Electronic assembly rework by non-authorized technicians may affect pump accuracy and operation.
Ranked causes
- 1. Non-authorized technician service or rework (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feed rate out of specifications patient safety
Programmed feed rate does not match intended delivery rate in continuous mode
Ranked causes
- 1. Feed rate incorrectly programmed (common)
Verify feed rate setting
- Test point
- ADJUST FEED screen FEED RATE field
- Expected
- Rate programmed between 1 and 400 ml/hr as ordered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feed volume to be delivered incorrectly set patient safety
Programmed volume to be delivered does not match clinical order in continuous mode
Ranked causes
- 1. Volume to be delivered incorrectly programmed (common)
Verify feed volume setting
- Test point
- ADJUST FEED screen FEED VTBD field and remaining volume display
- Expected
- Volume programmed as ordered; remaining VTBD shown next to v icon on lower right
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Feeding complete notification
FEEDING COMPLETE information screen after programmed feeding schedule finishes; display notification only, no error condition
Ranked causes
- 1. Programmed feeding cycle completed normally (common)
Verify feeding schedule was fully executed
- Test point
- Program completion timer
- Expected
- All programmed volume delivered or time elapsed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush cycle interrupts bolus delivery patient safety
During intermittent feeding with flushing enabled, flush cycles interrupt active bolus feeding cycles, reducing net feeding volume over the intended time period.
Ranked causes
- 1. Flushing precedence assigned over feeding process in Feed & Flush pump sets (common)
Verify flushing parameters are programmed and active
- Test point
- RUNNING screen
- Expected
- Flushing data displayed if Feed & Flush Pump Set is used
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush interval out of specifications patient safety
Programmed time between flush cycles does not match clinical protocol in continuous mode
Ranked causes
- 1. Flush interval incorrectly programmed (common)
Verify flush interval setting
- Test point
- ADJUST FLUSH screen FLUSH INTERVAL field
- Expected
- Interval programmed between 1 and 24 hours as clinically ordered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush Now feature requires intentional volume entry to stop patient safety
When Flush Now is activated, user must explicitly enter zero volume and press ENTER to stop the flush; accidental activation cannot be easily cancelled.
Ranked causes
- 1. Flush Now requires deliberate action to activate and stop (occasional)
Verify Flush Now settings before confirmation
- Test point
- Flush Now menu (Figure 12B or 14B appearance)
- Expected
- Default shows previously programmed flush volume; user may adjust before pressing ENTER
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush rate is high and may not match patient needs patient safety
Kangaroo ePump flushes at a fixed high rate that may not be appropriate for all patient conditions or feeding set configurations.
Ranked causes
- 1. Fixed flush rate of 1960 mL/hr may exceed patient tolerance or clinical requirements (occasional)
Review programmed flush volume and interval
- Test point
- FLUSH VOLUME and FLUSH INTERVAL menu (Figures 14A–14C)
- Expected
- Flush volume 10–500 mL; interval 1–24 hours, both selected to match patient need
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flush volume out of specifications caution
Programmed flush volume does not match clinical requirements in continuous mode with Feed & Flush set
Ranked causes
- 1. Flush volume incorrectly programmed (common)
Verify flush volume matches patient needs
- Test point
- ADJUST FLUSH screen FLUSH VOLUME field
- Expected
- Volume programmed between 10 and 500 ml as clinically appropriate
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
History data limited to 72 hours of past records
Kangaroo ePump can retrieve maximum 72 hours of feeding and flushing history, excluding the current hour, which may be insufficient for longer-term trend analysis.
Ranked causes
- 1. History storage capacity limited to 72 hours of past data (common)
Verify available history range in MORE OPTIONS menu
- Test point
- HISTORY screen (Figure 20)
- Expected
- User can set time range up to 72 hours back from current time (excluding current hour)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hold error - pump inactive without input
Hold Error screen appears; pump has been in HOLD mode without input for more than 10 minutes; audible alarm sounds at adjustable volume
Ranked causes
- 1. Pump left in HOLD mode longer than 10 minutes without operator input (common)
Confirm pump has been idle in HOLD mode for >10 minutes
- Test point
- Display screen status
- Expected
- HOLD ERROR screen showing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hold mode error triggered after 10 minutes
HOLD ERROR screen with adjustable alarm beep; pump inactive without operator input exceeds timeout
Ranked causes
- 1. Operator did not provide input within 10 minutes (common)
Pump has been in HOLD mode for more than 10 minutes
- Test point
- HOLD mode duration
- Expected
- Timeout at 10 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keep Tube Open (KTO) feature unavailable in EZMODE patient safety
KTO function cannot be used when pump is operating in EZMODE, limiting ability to prevent tube blockage during periods of minimal feeding.
Ranked causes
- 1. EZMODE restricts advanced features including KTO (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
KTO activation increments in variable time steps
KTO time increments change at higher durations (from 5-minute to larger steps above 30 minutes), which may make setting precise durations difficult.
Ranked causes
- 1. User interface design uses non-linear increment steps for KTO duration (common)
Set KTO duration by pressing input button to left of KTO icon
- Test point
- RUNNING screen, KTO icon input button (Figure 15A)
- Expected
- KTO duration cycles through 5, 10, 15, 20, 25, 30, then larger steps up to 240 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MISTIC connector removed during autopriming caution
MISTIC connector removed from pump during autopriming or running operation
Ranked causes
- 1. MISTIC connector dislodged during operation (rare)
Number of boluses out of specifications patient safety
Programmed bolus count does not match clinical order in intermittent mode
Ranked causes
- 1. Number of boluses incorrectly programmed (common)
Verify number of boluses setting
- Test point
- ADJUST BOLUS screen NUMBER OF BOLUSES field
- Expected
- Count programmed as ordered, or zero if running until supply exhausted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord wet or formula spilled on exterior high voltage
Water or formula contamination of pump exterior or power cord creating electrical hazard or equipment damage risk
Ranked causes
- 1. Improper cleaning or accidental liquid spillage during use (occasional)
Visual inspection for liquid contamination
- Test point
- Pump exterior, power cord, connectors
- Expected
- No visible moisture, formula residue, or wet areas on pump or cord
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump alarm sounds if left in HOLDING mode for 10 minutes patient safety
Kangaroo ePump generates audible alarm if paused (HOLDING) for 10 minutes without user interaction, which may indicate forgotten or unattended pump.
Ranked causes
- 1. Safety feature to alert clinician of extended pause period (common)
Monitor HOLDING time limit
- Test point
- RUNNING screen status line showing HOLDING; yellow LED lit
- Expected
- User interaction (RUN, HOLD, or menu selection) within 10 minutes of entering HOLD
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump continues automatic flush after formula delivery
Pump automatically performs an additional 25 mL water flush after completing programmed formula delivery, which may not be anticipated.
Ranked causes
- 1. Automatic post-delivery flush feature enabled when flushing occurred during formula delivery (common)
Observe total flush volume delivered on RUNNING screen
- Test point
- mL FLUSH total on RUNNING screen and HISTORY screen
- Expected
- Actual flush volume equals programmed flush volume per cycle plus 25 mL after final delivery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump in HOLD mode longer than 10 minutes without input patient safety
Pump automatically triggers HOLD ERROR alarm after 10 minutes of inactivity in HOLD mode without operator input
Ranked causes
- 1. Operator inactivity in HOLD mode exceeding 10 minute timeout (common)
Monitor time in HOLD mode
- Test point
- Pump operational state and timer
- Expected
- Operator input received within 10 minutes of entering HOLD mode
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump locked unintentionally
Screen locked preventing operator access to RUN, HOLD, or parameter controls
Ranked causes
- 1. LOCK SCREEN feature activated inadvertently (occasional)
Verify padlock icon on display
- Test point
- RUNNING screen display
- Expected
- Padlock icon visible indicating LOCK SCREEN mode active
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump requires re-certification patient safety
Periodic system performance re-certification needed to maintain safe operation and verify proper functioning
Ranked causes
- 1. Annual maintenance interval due or pump suspected of improper performance (common)
Verify pump is due for re-certification
- Test point
- Service records and last certification date
- Expected
- Last re-certification more than 12 months ago or pump malfunction suspected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump set dislodged error
PUMP SET DISLODGED screen; black ring retainer (MISTIC) is not properly loaded in MISTIC pocket in pump set loading area
Ranked causes
- 1. MISTIC retainer not properly positioned in pocket (common)
Verify MISTIC retainer position in loading pocket
- Test point
- MISTIC pocket in pump set loading area
- Expected
- Black ring retainer fully seated and locked in pocket
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Set dislodged error patient safety
PUMP SET DISLODGED screen appears when black ring retainer (MISTIC) is not properly loaded in MISTIC pocket in Pump Set loading area.
Ranked causes
- 1. Black ring retainer (MISTIC) improperly positioned (common)
Verify MISTIC retainer positioning
- Test point
- black ring retainer in MISTIC pocket
- Expected
- Properly seated in right pocket
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Set Dislodged Error appears on screen patient safety
PUMP SET DISLODGED screen displayed; black ring retainer not properly positioned
Ranked causes
- 1. MISTIC retainer improperly positioned (common)
Verify MISTIC retainer positioning
- Test point
- MISTIC pocket in Pump Set loading area
- Expected
- Black ring retainer properly seated in MISTIC pocket
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Set in use over 24 hours
Usage warning indicator blinking on RUNNING screen; Pump Set has been used 24 or more hours (running time)
Ranked causes
- 1. Normal Pump Set wear limit reached (common)
Verify pump set running time
- Test point
- RUNNING screen indicator
- Expected
- Blinking warning icon in upper left if >= 24 hours of use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Set not installed patient safety
Screen displays 'LOAD A SET' status message and blinking Pump Set icon appears in upper left corner
Ranked causes
- 1. Pump Set not loaded into pump (common)
Verify Pump Set is properly seated in loading slots
- Test point
- Left and right slots on pump loading mechanism
- Expected
- Valve seated in left slot, MISTIC retainer seated in right slot
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Set reused beyond 24 hours patient safety
Pump Set used beyond recommended single-use timeframe may cause performance degradation or contamination
Ranked causes
- 1. Pump Set not replaced after 24 hours (common)
Verify Pump Set age
- Test point
- Pump Set installation time record
- Expected
- Less than 24 hours since initial load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump set used >24 hours warning
Informational indicator blinks on RUNNING screen; pump set has been in use 24 or more hours (actual running time)
Ranked causes
- 1. Pump set cumulative runtime exceeds 24 hours (common)
Observe runtime counter on pump display
- Test point
- Pump set usage timer
- Expected
- Counter displays >= 24 hours when warning appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Set used more than 24 hours
Pump Set usage warning indicator blinks on Running screen after 24+ hours of set use
Ranked causes
- 1. Pump Set in use for 24 or more hours (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduced pump accuracy from unauthorized feeding sets patient safety
Use of generic or non-Tyco Healthcare feeding sets may jeopardize accuracy of the pump.
Ranked causes
- 1. Non-AccuSystem feeding set installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotor assembly misalignment after replacement
Rotor not properly seated on shaft after installation; may cause improper pump operation or flow rate errors
Ranked causes
- 1. Set screw not aligned with output shaft flat during rotor installation (occasional)
Verify rotor alignment before tightening set screw
- Test point
- Rotor hub set screw position relative to output shaft
- Expected
- Set screw aligned with flattened section of output shaft
- Tools
- 5/64" (2mm) allen wrench, visual inspection
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotor error - unusual rotor operating condition
Rotor Error screen appears during RUNNING or PRIMING; pump has detected unusual rotor operating condition, typically from improper pump set tubing loading around rotor
Ranked causes
- 1. Pump set tubing improperly loaded around rotor or pump set damaged (common)
Inspect pump set tubing loading around rotor
- Test point
- Rotor assembly
- Expected
- Tubing properly seated around rotor, no damage visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Rotor error displayed during operation
- Rotor error during operation
- Rotor error during priming or running
- Screen lock can prevent unintended operation changes
- System error - ADC ISR battery level failure
- System error - ADC ISR battery temperature failure
- System error - ADC ISR downstream occlusion detection failure
- System error - ADC ISR motor current failure
- System error - ADC ISR occlusion detection failure
- System error - buzzer test timeout
- System error - display lockup timeout
- System error - excessive graphic screens in memory
- System error - flash chip missing or version incompatible
- System error - flash languages read error
- System error - flash read error
- System error - flash screen read error
- System error - flash write error
- System error - internal timer stopped
- System error - motor control failure
- System error - motor error during tubing autoload
- System error - motor failsafe timeout
- System error - RS232 communications transmission error
- System error - stopcock feed position not found
- System error - stopcock flush position not found
- System error - stopcock timeout
- Unit remains in hold mode without response
- Volume per bolus out of specifications
27 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0 | Internal timer stopped | Timer malfunction detected | critical |
1 | Flash chip missing or version incompatible | System startup or firmware verification | critical |
10 | Flash write error | Firmware or data write to flash memory | critical |
11 | Flash read error | Reading data from flash memory | critical |
12 | General stopcock timeout | Stopcock stepper motor positioning | error |
13 | Timeout during buzzer test | Buzzer self-test during startup | warning |
14 | Downstream occlusion check timeout; rotor may blow fuse | Downstream occlusion detection | critical |
16 | Flash screen read error | Display screen data read from flash | error |
19 | Flash language pack read error | Language data loading from flash | warning |
2 | Too many graphic screens in memory | Screen display buffer management | error |
20 | Could not find stopcock flush position during prime | Stopcock positioning during priming sequence | error |
21 | Could not find stopcock feed position during prime | Stopcock positioning during priming sequence | error |
22 | Motor error during tubing autoload | Rotor motor operation during pump set autoload | error |
23 | Motor failsafe time out error; possible damaged gearbox encoder | Motor failsafe monitoring | critical |
25 | Display lockup timeout | Display system responsiveness check | error |
33 | Motor control failure | Motor control circuit verification | critical |
4 | Never got ADC interrupt for motor current | Motor control ADC initialization | critical |
5 | Never got ADC interrupt for battery level | Battery sensing circuit initialization | error |
50 | Graphic display error - timeout during GDStatusCheck | Display status check function | error |
51 | Graphic display error - timeout during GDAutoModeDataWriteCheck | Display data write verification | error |
52 | Graphic display error - timeout during GDWriteBytes | Display byte write operation | error |
6 | Never got ADC interrupt for battery temperature | Battery temperature sensing initialization | error |
7 | Never got ADC interrupt for ultrasonic occlusion/bag-empty | Ultrasonic sensor ADC initialization | error |
8 | Never got ADC interrupt for downstream occlusion | Downstream occlusion sensor ADC init | error |
9 | RS232 communications transmission error | Serial communication with control interface | error |
BATTERY LOW | Battery charge critically low | Battery charge depleted to critical level | warning |
Error #0 | Internal timer stopped | Timer hardware or firmware failure during operation | critical |
Error #1 | Flash Chip missing or version incompatible | Flash memory module not detected or version mismatch on startup | critical |
Error #10 | Flash Write Error | Flash memory write operation failed | error |
Error #11 | Flash Read Error | Flash memory read operation failed | error |
Error #12 | General Stopcock Time Out | Stopcock mechanism failed to respond within expected timeout period | error |
Error #13 | Time Out for Buzzer Test | Buzzer self-test failed to complete within timeout period | warning |
Error #16 | Flash Screen Read Error | Flash memory read failed for screen display data | error |
Error #19 | Flash Languages Read Error | Flash memory read failed for language data | error |
Error #2 | Too many graphic screens in memory | Memory overflow with graphic screen data during operation | error |
Error #20 | Could not find Stop Cock Flush Position during prime routine | Stopcock failed to reach flush position during priming sequence | error |
Error #21 | Could not find Stop Cock Feed Position during prime routine | Stopcock failed to reach feed position during priming sequence | error |
Error #22 | Motor Error during Tubing AutoLoad | Motor control failure detected during automatic tubing load procedure | error |
Error #23 | Motor Failsafe Time Out Error | Motor failsafe timeout occurred, possible gearbox encoder or flex cable damage | error |
Error #25 | Display Lockup Time Out Error | Display controller failed to respond within timeout period | error |
Error #33 | Motor Control Failure | Motor control system experienced general failure | error |
Error #4 | Never got a ADC ISR for motor current | Analog-to-Digital Converter failed to trigger motor current interrupt | error |
Error #5 | Never got a ADC ISR for battery level | Analog-to-Digital Converter failed to trigger battery level interrupt | error |
Error #6 | Never got a ADC ISR for battery temperature | Analog-to-Digital Converter failed to trigger battery temperature interrupt | error |
Error #7 | Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag-Empty checks | ADC interrupt failed for upstream ultrasonic occlusion or empty bag detection | error |
Error #8 | Never got a ADC ISR for Ultrasonic during Downstream Occlusion | ADC interrupt failed for downstream ultrasonic occlusion detection | error |
Error #9 | RS232 Communications Transmission Error | Serial communications failure during data transmission | error |
FEED ERROR | Feeding solution delivery interrupted | Feeding solution no longer being delivered | warning |
FEED ERROR | Enteral formula no longer being delivered to patient | Feed bag empty or clog between pump and feed bag | error |
FLOW ERROR | Feeding or flushing solution no longer being delivered | Clog between pump and patient, or dampness/dirt in valve pocket | error |
FLUSH ERROR | Flushing solution no longer being delivered | Flush bag empty or clog between pump and flush bag | error |
Hold Error | Pump inactive in HOLD mode without operator input for more than 10 minutes | Pump left in HOLD mode without input for 10+ minutes; operator timeout timer expired | warning |
HOLD ERROR | Pump in hold or pause state | Pump operation paused or held | warning |
PUMP SET DISLODGED | Black ring retainer (MISTIC) not properly positioned in loading area | MISTIC retainer improperly loaded in MISTIC pocket | critical |
ROTOR ERROR | Unusual operating condition detected for the rotor mechanism | Pump Set tubing damaged or improperly loaded around rotor | critical |
ROTOR ERROR | Pump rotor mechanism failure | Rotor unable to rotate or mechanical fault detected | critical |
SYSTEM ERROR | Critical internal pump system failure | Unspecified system-level fault detected | critical |
System Error #0 | Internal timer stopped | System timer failure during pump operation | critical |
System Error #1 | Flash Chip missing OR version incompatible | During system startup or firmware check | critical |
System Error #10 | Flash Write Error | Attempt to write data to flash memory failed | error |
System Error #11 | Flash Read Error | Attempt to read from flash memory failed | critical |
System Error #12 | Stopcock Time Out | Stopcock mechanism failed to respond or move within timeout period | error |
System Error #13 | Time Out for Buzzer Test | Buzzer speaker failed to sound during diagnostic test | warning |
System Error #14 | Time Out during downstream occlusion check, never got RotorOn3=0. May blow fuse. | Rotor failed to reach expected state during downstream occlusion check | critical |
System Error #2 | Too many graphic screens in memory | Display memory overflow during graphic loading or transition | error |
System Error #20 | Could not find Stop Cock Flush Position during prime routine | Stopcock failed to locate flush position during auto-prime sequence | error |
System Error #21 | Could not find Stop Cock Feed Position during prime routine | Stopcock failed to locate feed position during auto-prime sequence | error |
System Error #22 | Motor Error during Tubing AutoLoad | Motor failed to complete rotor positioning during automatic pump set loading | error |
System Error #23 | Motor Failsafe Time Out Error. Possible damaged gearbox encoder or flex cable. | Motor failsafe timeout triggered; rotor position not confirmed within timeout period | critical |
System Error #33 | Motor Control Failure | General motor control circuit or software failure | critical |
System Error #4 | Never got a ADC ISR for motor current | Motor current sensing subsystem failure during operation | error |
System Error #5 | Never got a ADC ISR for battery level | Battery voltage sensing failure during pump operation | error |
System Error #50 | Graphic Display Error - Timeout during GDStatusCheck function | Display status check routine exceeded timeout during execution | error |
System Error #51 | Graphic Display Error - Timeout during GDAutoModeDataWriteCheck function | Display auto-mode data write routine exceeded timeout | error |
System Error #52 | Graphic Display Error - Timeout during GDWriteBytes function | Display byte write function exceeded timeout during data transfer | error |
System Error #6 | Never got a ADC ISR for battery temperature | Battery temperature sensor failure during operation | error |
System Error #7 | Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag-Empty checks | Ultrasonic sensor subsystem failure during priming or operation | error |
System Error #8 | Never got a ADC ISR for Ultrasonic during Downstream Occlusion | Downstream occlusion detection sensor failure during operation | error |
System Error #9 | RS232 Communications Transmit | Serial communication transmission failure | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.