Covidien INVOS 5100C: troubleshooting and error codes
The ServiceForge knowledge base holds 21 documented symptoms for the Covidien INVOS 5100C, with 51 ranked causes, 25 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No rSO2 reading displayed patient safety
System does not display oxygen saturation values; sensor or preamplifier signal not received
Ranked causes
- 1. Error messages displayed; follow on-screen instructions (common)
- 2. Preamplifier not connected to monitor, sensor cable or sensor (common)
- 3. Sensor not applied to patient (common)
- 4. System locked up (occasional)
- 5. Defective CPU (occasional)
- 6. Defective preamplifier cable or connector (occasional)
- 7. Blown fuse on DIB (occasional)
- 8. Defective preamplifier (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up high voltage
System fails to turn on completely; no display or lights
Ranked causes
- 1. No AC power to outlet (common)
- 2. Blown fuse on line filter (common)
- 3. Blown fuse on power supply (occasional)
- 4. Secondary power cable disconnected or keypad connector not seated (occasional)
- 5. Blown fuse on DIB (occasional)
- 6. Defective CPU board (rare)
Verify AC power is available at outlet
- Test point
- Hospital-grade electrical outlet
- Expected
- AC power present
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad does not function
Buttons do not respond to key presses; no input accepted from control panel
Ranked causes
- 1. Defective keypad (common)
- 2. System locked up (common)
- 3. Internal cable not connected (occasional)
- 4. Defective DIB (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No battery power available
System cannot operate on battery backup; battery does not provide power
Ranked causes
- 1. Defective battery (common)
- 2. Defective DIB (occasional)
- 3. Internal cable not connected or keypad connector not seated (occasional)
- 4. Blown fuse on DIB (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal Strength Indicator (SSI): 1-4 bars fluctuating and unstable patient safety
SSI display shows weak, unstable signal with rapid fluctuations; rSO2 reading reliability degraded
Ranked causes
- 1. Signal interference from external sources (common)
- 2. Sensor applied over anatomical anomalies or contaminants (common)
- 3. Defective SomaSensor (occasional)
- 4. Moisture in sensor connection (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: INVOS 5100C System high voltage
Factory-calibrated system requires annual inspection and functional verification. System calibration is set during manufacturing and stored in SomaSensor chip. Includes mechanical and electrical inspection points.
Ranked causes
- 1. Annual inspection required per maintenance schedule (common)
- 2. Functional test with adjustable field test device (common)
- 3. Battery backup replacement after 5 years (occasional)
Inspect chassis and cables for physical damage
- Test point
- Power cord, sensor cables, preamplifiers, battery status, chassis screws, key panel, fan, internal components
- Expected
- No loose screws, cracks, dirt, loose connectors or damaged cords
13 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is not visible
LCD display is dark or blank; no visible text or images
Ranked causes
- 1. Internal cables not properly seated (common)
- 2. Defective LCD (occasional)
- 3. Defective backlight lamp on LCD (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display visible
Screen is completely dark; no LCD output at all
Ranked causes
- 1. Defective CPU (occasional)
- 2. Defective display (occasional)
- 3. Defective inverter board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit locks up or operates erratically
System freezes, intermittently crashes, or behaves unpredictably during operation
Ranked causes
- 1. Erratic AC power input (common)
- 2. Defective CPU board (occasional)
- 3. Defective DIB (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Digital output: COM port unavailable message
Error message displayed when selecting CASE HISTORY in DIGITAL OUTPUT menu; no PC connection established
Ranked causes
- 1. Digital output port not connected to PC with null modem cable (common)
- 2. Wrong COM port specified in Hyper Terminal (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not show expected rSO2 values
When testing with Adjustable Field Test Device set to known values (90, 70, 50, 30), the display either does not show the value or shows out-of-range data on connected channels
Ranked causes
- 1. Monitor requires alignment and testing with Field Test Device (unknown)
Verify rSO2 display matches Field Test Device setting
- Test point
- Main Screen display
- Expected
- rSO2 value of 90, 70, 50, or 30 ±1 or poor sensor signal indication
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is on but green light is not on
LCD display functions but power indicator light not illuminated
Ranked causes
- 1. Defective keypad (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External monitor display mismatch
External monitor screen does not match INVOS 5100C Monitor screen during final functional test
Ranked causes
- 1. Serial or USB connection failure (unknown)
Verify external monitor displays same screen as INVOS 5100C Monitor
- Test point
- External Monitor connected via serial interface
- Expected
- External monitor screen identical to INVOS 5100C Monitor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line leakage current test failure high voltage
Red FAIL light turns on and alarm sounds during line leakage testing; failure mode displayed on tester front panel
Ranked causes
- 1. Safety ground or neutral/reverse fault condition exceeded (unknown)
Monitor line leakage current through eight test configurations
- Test point
- INVOS 5100C Monitor with Line Leakage Tester connected
- Expected
- All eight SFC and NC configurations pass with current within limits
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lithium battery replacement on CPU
Coin-type lithium battery supplies real-time clock circuit. Nominal 3-3.3VDC, service life approximately 5-6 years at 20°C ambient.
Ranked causes
- 1. Lithium battery end-of-life after 5-6 years service (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity with Philips module
System cannot establish or maintain connection with Philips equipment; digital output not working
Ranked causes
- 1. Output mode switched from Printer/PC Link to Philips without restart (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No data received or transmission scrambled
Digital output data missing or corrupted; cannot receive or view transmitted data
Ranked causes
- 1. Hyper Terminal settings not correct (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on LEDs not illuminated high voltage
PWR ON (green) LED or AC Power (blue) LED not lit after pressing On/Off button
Ranked causes
- 1. AC power not connected or power supply fault (common)
Verify AC power cable connected and both indicator LEDs lit
- Test point
- INVOS 5100C Monitor back panel AC input and front panel LEDs
- Expected
- PWR ON (green) LED and AC Power (blue) LED both illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preamplifier AC Withstand Voltage (Hipot) test failure high voltage
Alarm sounds and red RESET light turns on during Hipot testing, indicating preamplifier failed electrical safety test
Ranked causes
- 1. Preamplifier isolation or insulation breakdown at 1800V (unknown)
Verify AC Withstand Voltage Tester display parameters
- Test point
- Tester front panel display
- Expected
- SET M2 1.0s 1.8KVAC 1.00mA
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Saturation value out of range after power-up patient safety
Oximeter saturation reading is not constant or falls outside 47% to 53% range on display during initial verification
Ranked causes
- 1. Hardware defect or sensor simulator fixture malfunction (unknown)
Verify saturation value is constant and within range
- Test point
- INVOS 5100C Monitor display with Sensor Simulator Fixture connected to Preamplifier
- Expected
- Saturation value constant between 47% and 53%
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal Strength Indicator (SSI): 0 bar patient safety
SSI shows no bars; signal is completely lost; rSO2 value not displayed
Ranked causes
- 1. Error condition; see error messages on display (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.