Electrosurgical generators

ConMed System 5000: troubleshooting and error codes

The ServiceForge knowledge base holds 25 documented symptoms for the ConMed System 5000, with 45 ranked causes, 65 diagnostic checks with test points and expected values, and 88 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No output power high voltage

System does not produce RF output power when activated

Ranked causes

  1. 1. Internal connections bad (common)
  2. 2. Faulty accessory with blade not fully seated (occasional)
  3. 3. Output board jacks not making contact (occasional)
  4. 4. Low voltage power supply bad (occasional)
  5. 5. Shorted RF Amplifier MOSFET (occasional)
  6. 6. HVPS output too low (occasional)
  7. 7. Improper gate drive (occasional)
First check

Verify internal connector seating

Test point
A6J1, A6J2, A6J3, A6J4, A7J1, A7J2 connectors
Expected
All connectors fully seated

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display panel blank when powered on

No display visible when system is turned on

Ranked causes

  1. 1. Power cord not fully seated (common)
  2. 2. Fuses blown or improperly installed (common)
  3. 3. Controller board or harness not seated (occasional)
  4. 4. Low voltage power supply failure (occasional)
First check

Verify power cord connection

Test point
Power receptacle on back panel
Expected
Fully seated

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dispersive electrode status indicators flashing alternating red patient safety

Single or dual dispersive electrode alarm LEDs flashing red alternately

Ranked causes

  1. 1. Dispersive electrode not connected (common)
  2. 2. Dual dispersive electrode not properly applied to patient (common)
  3. 3. Dispersive electrode connector not fully seated (occasional)
First check

Verify dispersive electrode connection

Test point
Dispersive Electrode Receptacle
Expected
Electrode connected and seated

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows A[[ [P xx (control panel error)

Control panel error code displayed on screen

Ranked causes

  1. 1. Faulty push button on front control panel (occasional)
  2. 2. Push button pressed during power-on (occasional)
  3. 3. Faulty cable connection between front panel and display board (occasional)
First check

Verify no buttons pressed during power-on

Test point
Control panel buttons
Expected
No buttons actuated

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows A[[ xxx (accessory error)

Faulty accessory error code displayed on screen

Ranked causes

  1. 1. Faulty accessory connected (common)
  2. 2. Hand-controlled accessory erroneously actuated (occasional)
  3. 3. Footswitch erroneously actuated (occasional)
First check

Verify no footswitches are stacked

Test point
Monopolar and Bipolar footswitch connectors
Expected
Footswitches not stacked

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect bipolar output

Bipolar output power does not match set value or expected output

Ranked causes

  1. 1. Bad calibration (common)
  2. 2. Incorrect load resistor when checking output (occasional)
  3. 3. Bad voltage or current feedback (occasional)
First check

Perform bipolar output calibration

Test point
Bipolar output with calibrated load resistor
Expected
Output matches calibration targets per procedure

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect monopolar output

Monopolar output power does not match set value or expected output

Ranked causes

  1. 1. Bad calibration (common)
  2. 2. Incorrect load resistor when checking output (occasional)
  3. 3. Bad voltage or current feedback (occasional)
First check

Perform monopolar output calibration

Test point
Monopolar output with calibrated load resistor
Expected
Output matches calibration targets per procedure

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Real Time Clock battery voltage low or not charged

Date defaults to 01 01 01 or time defaults to 12P 0 1, indicating Real Time Clock battery insufficient

Ranked causes

  1. 1. Real Time Clock battery voltage below 2V (common)
  2. 2. Real Time Clock battery not charged from previous service (occasional)
First check

Verify date and time display

Test point
Display panel (Last Fault Code Mode)
Expected
Date shows actual month-day-year; time shows actual hours-minutes-seconds

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

A.R.M. dual electrode lower limit out of range patient safety

A.R.M. does not show green flashing with eight bargraph bars at lower resistance limit

Ranked causes

  1. 1. A.R.M. circuit calibration error or component fault (unknown)
First check

Verify A.R.M. dual electrode lower limit indicator and bargraph response

Test point
Dual Dispersive Electrode Status/Alarm Indicator and Bargraph
Expected
Flashing green with eight bars illuminated Ohms
Tools
Decade Resistance Box

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

A.R.M. dual electrode upper limit out of range patient safety

A.R.M. does not show green flashing with single bargraph bar at upper resistance limit

Ranked causes

  1. 1. A.R.M. circuit calibration error or component fault (unknown)
First check

Verify A.R.M. dual electrode upper limit indicator and bargraph response

Test point
Dual Dispersive Electrode Status/Alarm Indicator and Bargraph
Expected
Flashing green with one bar illuminated Ohms
Tools
Decade Resistance Box

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

A.R.M. single electrode upper limit out of range patient safety

A.R.M. does not show steady green indicator without flashing at upper resistance limit

Ranked causes

  1. 1. A.R.M. circuit calibration error or component fault (unknown)
First check

Verify A.R.M. single electrode upper limit indicator response

Test point
Single Dispersive Electrode Status/Alarm Indicator
Expected
Green and not flashing Ohms
Tools
Decade Resistance Box

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Automatic Return Monitor (A.R.M.) dual electrode alarm limit out of range patient safety

A.R.M. does not flash red in alternating fashion when dispersive electrode resistance is at alarm threshold

Ranked causes

  1. 1. A.R.M. circuit calibration drift or component failure (unknown)
First check

Verify A.R.M. dual electrode alarm limit indicator response

Test point
Single and Dual Dispersive Electrode Status/Alarm Indicators
Expected
Red alternating flash Ohms
Tools
Decade Resistance Box, dispersive electrode cable adapter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: A.R.M.™ Dispersive Electrode connection high voltage

Calibrate Automatic Return Electrode Monitoring against two known precision resistances

Ranked causes

  1. 1. A.R.M.™ requires calibration with paired resistances (common)
First check

Initiate A.R.M.™ calibration

Test point
Remote Power Control Key
Expected
Resistance value appears in Monopolar Coag Power Digital Display

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bipolar modes high voltage

Calibrate Bipolar Macro mode using measured current or power against a precision resistor

Ranked causes

  1. 1. Bipolar mode requires power calibration adjustment (common)
First check

Select calibration units (current or power)

Test point
Bipolar Power Adjustment Keys
Expected
A for current or P for power

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Complete and exit calibration mode caution

Exit calibration mode and prepare ESU for normal operation

Ranked causes

  1. 1. Calibration must be completed before normal operation (common)
First check

Turn off main power

Test point
Main power switch
Expected
Power indicator extinguished

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monopolar modes high voltage

Adjust Monopolar Cut, Blend, Pinpoint Coag, Standard Coag, and Spray Coag modes using measured current or power against a precision resistor

Ranked causes

  1. 1. Mode requires power calibration adjustment (common)
First check

Select calibration units (current or power)

Test point
Bipolar Power Adjustment Keys
Expected
A for current or P for power

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: System 5000 monopolar and bipolar output power

Verify and adjust output power calibration for Cut, Coag, Blend modes and Bipolar to match target values using load resistance and meter

Ranked causes

  1. 1. Calibration drift after component replacement or repair (common)
First check

Set calibration system configuration DIP switch

Test point
Controller board A4SW2.2 and A4SW2.1
Expected
A4SW2.2 ON (UP), A4SW2.1 OFF (DOWN)

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows [al (calibration mode)

Calibration mode is displayed on screen

Ranked causes

  1. 1. DIP switch 2 set to ON and Remote Power Control Key pressed during power-on (common)
First check

Verify DIP switch 2 position

Test point
Controller assembly DIP switch position 2
Expected
OFF position for normal operation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows Err xxx (error code)

Error code is displayed on screen indicating a system fault

Ranked causes

  1. 1. System fault detected during POST or operation (unknown)
First check

Attempt system reset via power cycle

Test point
Main power switch
Expected
Error clears or POST identifies fault

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows LF[ xxx (last fault code mode)

Last fault code mode is displayed instead of normal operation

Ranked causes

  1. 1. Remote Power Control Key pressed during power initialization (common)
First check

Power cycle system without pressing buttons

Test point
Main power switch
Expected
Normal operation display

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 140: Calibration memory error

Stored calibration values are corrupted or inaccessible, affecting output power accuracy

Ranked causes

  1. 1. Calibration memory corruption or Controller board failure (rare)
First check

Verify Err 140 display

Test point
Monopolar Cut Power Digital Display
Expected
Err 140

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 143: Mode calibration required

One or more operating modes require calibration after board replacement or calibration drift detected

Ranked causes

  1. 1. Board replacement (Controller, Transformer, or Output) or calibration drift (common)
First check

Verify Err 143 display and affected modes

Test point
Monopolar Cut Power Digital Display and mode indicators
Expected
nEr with flashing minor mode indicators for modes requiring calibration

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flashing decimal points on display

Decimal point indicators on display are flashing

Ranked causes

  1. 1. Test Mode is active (common)
First check

Verify DIP switch 1 position

Test point
Controller assembly DIP switch position 1
Expected
OFF position for normal operation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leakage current out of specification high voltage

Active RF output to neutral exceeds allowable limit during monopolar operation with footswitch or hand control activation

Ranked causes

  1. 1. Faulty monopolar active output relay or circuit (unknown)
First check

Measure leakage current from footswitched monopolar active output to neutral with normal line/GND

Test point
Active output terminal to neutral
Expected
≤15 μA, max
Tools
Leakage current meter, footswitch

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Output coupling capacitor out of specification caution

Capacitance measured at output coupling does not meet expected values for monopolar or bipolar modes

Ranked causes

  1. 1. Faulty output coupling capacitor or circuit component (unknown)
First check

Measure baseline output coupling capacitance without activation

Test point
Shorting plug adapter to footswitched ReadiPlug Universal Accessory Receptacle
Expected
<0.5 nF, max
Tools
Capacitance meter (≤1 kHz), shorting plug adapter, ≤6" test leads

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100Test or CAL mode dip switch with no remote power control key pressRemote Power Control Key not depressed during power initialization or not released before last tone soundserror
101Controller/Monitor DIP switch mismatchSystem Controller and Monitor detect different Configuration Dipswitch settingserror
102Corrupted communications between System Controller and MonitorFaulty data transmission from System Controller to Monitor through HPI porterror
103Corrupted communications between System Controller and RF ControllerFaulty data transmission from System Controller to RF Controller through HPI porterror
104Monitor TOKEN not incremented, monitor program scrambledFaulty data transmitted from System Controller to Monitor through HPI porterror
105Controller failed RAM testMemory errors detected during POSTerror
106RF controller bootload program does not match controller ROM tableFaulty program transmission from System Controller to RF Controller at startuperror
107Monitor bootload program does not match controller ROM tableFaulty program transmission from System Controller to Monitor at startuperror
108RF TOKEN not incremented, RF controller program scrambledFaulty or corrupted program transmission from System Controller to RF Controller at startuperror
109Communications lost during POST testsFaulty communications through HPI porterror
112Pre-activation handshake, monitor does not read an activation signalActivation signals detected by System Controller and Monitor do not matcherror
113Pre-activation handshake, monitor does not read a tone signalActivation tone not present when Monitor checks prior to enabling electrosurgical outputerror
114Display or keyboard SPI communication failureCommunication failure between Control assembly and Display assembly over SPI buserror
1175.0 volt too low5-volt power supply voltage drops below specificationerror
1185.0 volt too high5-volt power supply voltage exceeds specificationerror
122System Controller flash program CRC does not match calculated CRCSelf-check CRC for System Controller not internally consistenterror
135A.R.M. calibration (10/150 Ohm) EEPROM CRC failedCRC for calibration coefficient memory not self consistenterror
136A.R.M. circuit dropped below 0.4V (circuit failed)A.R.M. circuit voltage falls below 0.4Verror
137RF controller calibration EEPROM CRC failedCRC for RF controller calibration coefficient memory not self consistenterror
138Monitor calibration EEPROM CRC failedCRC for Monitor calibration coefficient memory not self consistenterror
139During calibration, not all points were entered for the modeCalibration not completed in correct mannerwarning
140A calibration EEPROM CRC failed (POST ERROR)CRC for calibration coefficient memory not self consistent during POSTerror
141Test and CAL dip switches are both turned onCalibration and Test Configuration Dipswitches both in ON positionerror
142RF PWM (RF_PROT) gate protection circuit shutdownPulse width commanded by RF Controller exceeds maximum allowed for selected modeerror
143Unit or a mode was never calibratedSystem or mode never calibrated or Controller replaced without calibrationwarning
144RF controller and Monitor power levels do not matchMonitor and RF Controller sense different electrosurgical output power during activationerror
170Controller waveform inhibit failed (will not disable RF) (POST ERROR)Monitor detects that System Controller cannot disable RF Drive signal during POSTcritical
200Modality function timed out while waiting for ADC DMAs to finishComponent failure preventing ADC DMA operations from completingerror
201RF Controller emulation restartProgramming function restartinfo
202Inter-processor communications errorController Assembly Host Port communication failureerror
204Processor system clock change failedMicroprocessor clock frequencies do not change to correct frequency after initializationerror
205Halt mode active beyond limitHalt mode operation exceeds time limiterror
206RF Controller discovered runtime bad CRCErrors found in software downloaded from System Controller to RF Controller during initializationerror
207Unexpected mode change while activeSelected output mode changes while system is activatederror
236Bad CRC on modality coefficientsModality coefficient CRC check failureerror
237POST memory failureErrors found in microprocessor memories during power initializationerror
239RF Controller rejected a calibration valueCalibration failed for one or more modeserror
311POST Monitor memory failureErrors found in Monitor microprocessor memory during power initializationerror
314POST Monitor or RF controller clock failureRF Controller or Monitor clock oscillator failure detected during POSTerror
315POST Monitor activation circuitry failureActivation request signal states do not match between Monitor and System Controllererror
320Monitor failed program CRC testErrors found in software downloaded from System Controller to Monitor during initializationerror
321Monitor failed calibration data CRC testMonitor calibration data CRC check failureerror
322Monitor or RF controller did not return to IDLE stateActivation continues after activation command ceaseserror
330POST RF AC voltage cut feedback too low (A/D channel 0)Monitor senses electrosurgical voltage output too low during POST testerror
331POST RF AC voltage cut feedback too high (A/D channel 0)Monitor senses electrosurgical voltage output too high during POST testerror
332POST RF AC voltage coag feedback too low (A/D channel 2)Monitor senses electrosurgical voltage output too low during POST testerror
333POST RF AC voltage coag feedback too high (A/D channel 2)Monitor senses electrosurgical voltage output too high during POST testerror
334POST RF AC voltage coag feedback too low (A/D channel 1)Monitor senses electrosurgical voltage output too low during POST testerror
335POST RF AC voltage coag feedback too high (A/D channel 1)Monitor senses electrosurgical voltage output too high during POST testerror
336Calibration RF AC voltage cut feedback too highMonitor senses electrosurgical voltage output too high during POST testerror
337Calibration RF AC voltage cut feedback too lowMonitor senses electrosurgical voltage output too low during POST testerror
338Calibration RF AC voltage coag feedback too highMonitor senses electrosurgical voltage output too high during POST testerror
339Calibration RF AC voltage coag feedback too lowMonitor senses electrosurgical voltage output too low during POST testerror
340Monitor Timer0 interrupt failedInternal timer failure during operationcritical
341Monitor Timer0 interrupt failed POSTInternal timer failure detected during POSTcritical
342Monitor Timer1 interrupt failed POSTInternal timer failure detected during POSTcritical
343Monitor has lost communications - controller TOKENS not incrementedLoss of communication between Monitor and Controllercritical
344Monitor Isense DMA or SPI hungMonitor DMA or SPI interface failurecritical
345Monitor Power DMA or SPI hungMonitor DMA or SPI interface failurecritical
350Monitor output channel current sensor failureCurrent mismatch between output channel and power monitoring sensorserror
351Monitor sensed over power condition for 300mS running averageOver power at electrosurgical output persists for more than 300 mserror
352Monitor sensed 4x over power condition for short periodExcessive power at electrosurgical output detected brieflyerror
357Monitor sensed power @ 0W dial power settingMonitor detects electrosurgical output power during activation at zero power dial settingerror
358Monitor sensed gate drive in idle @ 0W dial power settingDrive signal detected while system is not activatedwarning
359Monitor sensed peak voltages above the LAP LimitVoltage peaks exceed limits in Lap Advanced Specialty modeerror
360Monitor program counter landed on an unused interrupt vectorMonitor processor anomalycritical
371Monitor will not disable RF waveform on the Controller assembly during POSTMonitor cannot disable RF waveform during POSTcritical
372Monitor will not disable high voltage during POSTMonitor cannot disable HVPS during POSTcritical
373Damping circuit failed POST – damping not detectedDamping drive signal failure detected during POSTerror
374Damping circuit failed POST – damping stuck onDamping drive signal stuck on during POST, no RF outputcritical
379RF calibration failed POSTSystem calibration failure during POSTerror
381No tone feedback during POSTMonitor did not detect tones during POSTwarning
382Activation without a toneMonitor did not detect a tone during activationwarning
383RF Current sensed at an unselected outputCurrent flowing at an output that has not been selectederror
384POST HVDC to 30V failed - supply too lowHVPS output is too low during Power-On Self Testerror
385POST HVDC to 30V failed - supply too highHVPS output is too high during Power-On Self Testerror
386POST HVDC to 140V failed - supply too lowHVPS output is too low at 140V setting during POSTerror
387POST HVDC to 140V failed - supply too highHVPS output is too high at 140V setting during POSTerror
388POST monitor timed out - did not respond to POST commandMonitor did not respond to Power-On Self Test command within timeout perioderror
389Dial setting over maximum limitFront panel power setting exceeds limit for selected modewarning
391Monitor detected the wrong RF waveformMonitor detected improper RF Amplifier drive waveform or frequency for selected modeerror
392Monitor detected a pulse mode on-time that was too longActive pulse mode pulses exceed nominal duration limitswarning
393Vbias signal too highReference voltage on one or more ADC channels on Controller exceeds nominalerror
394Vbias signal too lowReference voltage on one or more ADC channels on Controller below nominalerror
A[[ xxxAccessory error codeFaulty accessory connected, hand-controlled accessory erroneously actuated, or footswitch erroneously actuatederror
Err 100Error 100 fault codeDisplayed when DIP switch 1 or 2 is ON and Remote Power Control Key is not pressed until Test Tone beginserror
Err 140Calibration memory error; stored calibration values detected as invalidUnit detects corruption or loss of calibration data in non-volatile memoryerror
Err 143One or more modes require calibration; specific modes lack valid calibration dataAfter board replacement (Controller P/N 61-6431, Transformer P/N 61-6445, Output P/N 61-6461) or calibration drift detected during preventative maintenanceerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.