ConMed 7550: troubleshooting and error codes
The ServiceForge knowledge base holds 87 documented symptoms for the ConMed 7550, with 149 ranked causes, 67 diagnostic checks with test points and expected values, and 56 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Err 306 displayed
PW_MEA out of calibration - Power measurement is out of tolerance.
Ranked causes
- 1. U7 on Power Control Failed on A4 (occasional)
- 2. Ramp transistors Q1 or Q2 failed (occasional)
- 3. U14B on A4 Failed (occasional)
- 4. U21 - Port A wrong data on A4 and A5 (occasional)
- 5. D/A Failed - PCON on A5 (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 302 displayed
IMEA Error - Output current measurement is out of tolerance or power regulation has failed.
Ranked causes
- 1. Output Current Error too high on A4 (common)
- 2. Too much power into a heavy load (occasional)
- 3. Power Regulation Loop failed on A4 (occasional)
- 4. D/A Failed - PCON on A5 (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 316 displayed high voltage
HV Idle Test - High voltage monitor or supply circuits not properly in idle state.
Ranked causes
- 1. HVMON circuit failed on A1 (occasional)
- 2. 10V circuit on A7 Failed (occasional)
- 3. J3 disconnected on A7 (occasional)
- 4. J1 disconnect on A1 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ABC mode will not initiate or activate
ABC (automatic blend control) mode fails to start or respond to activation request
Ranked causes
- 1. Over-voltage circuit active or over-voltage threshold exceeded (common)
- 2. U7 failed or pulse width/duty cycle error (occasional)
- 3. U4 control signal failure (rare)
Recalibrate ABC over-voltage
- Test point
- ABC mode control circuit
- Expected
- Over-voltage threshold reset; ABC mode initiates normally
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dual Pad monitor set will not lock patient safety
Dual Pad electrode monitoring mode fails to lock or maintain connection with monitor set
Ranked causes
- 1. PP receptacle resistance too high (>10 ohms) (common)
- 2. Monitor set switch on overlay failed (occasional)
- 3. A.R.M. (Automatic Return Monitor) calibration error (rare)
Measure PP receptacle resistance
- Test point
- PP receptacle connector
- Expected
- >12 ohms, > 12 ohms
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 1 displayed
Front panel switch stuck or defective during power-up; Front Panel overlay failed or U5 microcontroller defective.
Ranked causes
- 1. User pressing switch during power-up (common)
- 2. Front Panel overlay failed (occasional)
- 3. U5 (A5) microcontroller failed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 111 displayed
FMON (Flow Monitor) out of tolerance - Flow control circuit or measurement is out of calibration.
Ranked causes
- 1. FMON out of calibration on A1 (common)
- 2. Flow Control Valve Driver shorted on A1 (occasional)
- 3. Argon tubes to X1 reversed on A1 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 112 displayed
Flow Error - Flow measurement or control is out of calibration or circuit has failed.
Ranked causes
- 1. FMON out of Calibration on A1 (common)
- 2. Handpiece occluded (common)
- 3. Flow control circuit has failed (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 2 displayed
Hand or foot activation switch stuck or failed accessory (handcontrol or footcontrol) preventing normal operation.
Ranked causes
- 1. User pressing switch during power-up (common)
- 2. Failed Handcontrol or Footcontrol (common)
- 3. Failed component on A5 or A11 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 233 displayed
Data transfer to Monitor failed - Communication between Controller and Monitor microcontrollers has failed.
Ranked causes
- 1. Mailbox Failed on A5 (occasional)
- 2. Data Bus bad on A5 (occasional)
- 3. Address Bus bad on A5 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 301 displayed
Flow Error - Flow measurement is too high or low, or D/A converter failure affecting gas flow control.
Ranked causes
- 1. Flow too high or low (common)
- 2. Flow Rate too high for accessory (common)
- 3. D/A Failed - VGAS on A5 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 307 displayed
FB Amp Drive Error - Feedback amplifier drive circuit failed on Power Control PCB.
Ranked causes
- 1. U7 on A4 Failed (occasional)
- 2. U14A on A4 Failed (occasional)
- 3. U21 - Port A wrong data on A4 and A5 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 308 displayed high voltage
HVDC too high - High voltage control circuit or monitoring circuit has failed.
Ranked causes
- 1. HV Control Loop failed on A1 (occasional)
- 2. HV Monitor Failed on A1 (occasional)
- 3. D/A Failed - VCON on A5 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 315 displayed
Flow Idle Error - Flow control circuit not properly inhibited during idle state.
Ranked causes
- 1. Flow Control Loop Failed on A1 (occasional)
- 2. Driver Q2 Failed on A1 (occasional)
- 3. U15B Failed on A1 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 323 displayed
FB Amp Drive Error - Feedback amplifier drive circuit failure on Power Control PCB.
Ranked causes
- 1. U7 on A4 Failed (occasional)
- 2. U14A on A4 Failed (occasional)
- 3. U21 - Port A wrong data on A4 and A5 (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate too high caution
Argon gas flow rate exceeds expected rate or control setting on display
Ranked causes
- 1. Driver Q2 shorted on A1 board (occasional)
- 2. Argon pressure transducer hoses reversed (occasional)
- 3. Control loop failure on A1 board (rare)
Measure Q2 emitter voltage during flow operation
- Test point
- Q2 emitter pin (A1 board)
- Expected
- Voltage increases progressively with increasing flow rate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handsense control not responding or intermittent patient safety
Handcontrol switch activation does not register or registers intermittently; control signals not reaching RF circuits
Ranked causes
- 1. Continuity transformer T3 primary drive signal loss (occasional)
- 2. Opto-isolator LED failure in handswitch detection circuit (occasional)
- 3. Handswitch mechanical contact failure or corrosion (common)
Measure NE555 timer U3 drive signal on T3 primary
- Test point
- U3 output pin; T3 primary winding
- Expected
- 100 KHz square wave, 20% duty cycle
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or zero RF output power high voltage
RF Output Assembly fails to produce expected power across Cut, Blend, Pinpoint, Bipolar or Spray modes
Ranked causes
- 1. Faulty relay in RF output path (common)
- 2. Relay driver failure on Display Assembly (occasional)
- 3. Open circuit or disconnection in output relay switching path (occasional)
Verify relay driver is providing drive signal
- Test point
- Relay driver output on Display Assembly
- Expected
- Drive signal present for activated mode
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Activation or Tone
Unit will not activate or produce tone when handcontrol or footcontrol is pressed.
Ranked causes
- 1. Activation agreement error between Controller and Monitor (common)
- 2. Handsense Circuit failed (common)
- 3. RF Jack worn out (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic flow control or solenoid valve not responding caution
Argon gas flow does not modulate or solenoid valve fails to open/close; pneumatic manifold not functioning
Ranked causes
- 1. Solenoid valve coil open or shorted (occasional)
- 2. Solenoid valve driver transistor failure or open base/collector (occasional)
- 3. Proportioning valve controller current not increasing with flow setting (occasional)
Measure solenoid valve coil resistance
- Test point
- Digital (solenoid) valve coil terminals
- Expected
- 72 ohms
- Tools
- DMM ohmmeter
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator thermal shutdown activated on 15 V or 5 V supply caution
System loses 15 V or 5 V power due to regulator overheating; load current exceeds safe operating range
Ranked causes
- 1. Shorted component on downstream assembly drawing excessive current (common)
- 2. Regulator Q1, Q2, or Q5 internal short or degradation (occasional)
- 3. 5 V switcher LC filter capacitor failure or degradation (occasional)
Measure regulator output current with all loads connected
- Test point
- Regulator output line or dedicated current sense resistor
- Expected
- 15 V regulator: <500 mA typical; 5 V switcher: <2 A typical (per schematic)
- Tools
- DMM or power supply with current monitoring
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF leakage detected or excessive RF leakage during operation caution
System detects unbalanced RF current returning via alternate paths; leakage current exceeds balanced condition threshold
Ranked causes
- 1. Unbalanced condition in Balun transformer T4 (common)
- 2. Damaged or faulty Balun sensing winding (occasional)
- 3. Patient return isolation relay failure (K8, K9, K11) (occasional)
Measure balun sense winding voltage
- Test point
- Balun sense winding
- Expected
- 0 V under balanced condition; voltage increases proportional to leakage current
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 120 displayed
Controller and Monitor microcontrollers disagree on power level, indicating data communication or encoder logic failure.
Ranked causes
- 1. Bad data transfer between Controller and Monitor on A5 (occasional)
- 2. PIA with encoder logic failed on A5 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 20 displayed caution
Illegal Default Switch Setting - unit is in TEST MODE instead of RUN MODE, or default switch position 1 or 6 is ON.
Ranked causes
- 1. Default switch position 1 or 6 is ON (common)
- 2. System in TEST MODE (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 203 displayed
Mailbox error - Bad data transfer or mailbox communication failure between Controller and Monitor.
Ranked causes
- 1. Bad data transfer on A5 (occasional)
- 2. Mailbox failed on A5 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 210 displayed
Bad Calibration Data - Calibration data stored in EEPROM is invalid or corrupted.
Ranked causes
- 1. Calibration Data bad on A5 (common)
- 2. EEPROM failed on A5 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 252 displayed
EEPROM error - CRC validation of EEPROM data failed or data is corrupted.
Ranked causes
- 1. CRC error on A5 (common)
- 2. Bad data stored in EEPROM on A5 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 304 displayed
Output Power - Power control circuit or D/A converter failure preventing proper power regulation.
Ranked causes
- 1. Power Control Circuit Failed on A4 (occasional)
- 2. D/A Failed - PCON on A5 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 313 displayed
IMEA idle - Current measurement idle error indicating improper idle state voltage.
Ranked causes
- 1. U14D Failed on A4 (occasional)
- 2. U12 or U8 Failed on A5 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 331 displayed high voltage
HV out of range - High voltage DC supply exceeds tolerance during power-on self-test.
Ranked causes
- 1. HVDC too high during POST (occasional)
- 2. Triac Failure on A7 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 332 displayed high voltage
HV not inhibited by monitor - High voltage control failed to inhibit during power-on self-test.
Ranked causes
- 1. HVINH failed to inhibit HVDC during POST (occasional)
- 2. Fuse on HV Power Supply open or missing (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Half Power in Cut, Blend, Pinpoint, Bipolar
Unit outputs half power in Cut, Blend, Pinpoint, or Bipolar modes instead of full power.
Ranked causes
- 1. A phase of the Full Bridge Amp failed (common)
- 2. Open power FET or no RF Drive to power FET (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High voltage supply exceeds 205 VDC high voltage
DC voltage supply reads above 205 VDC in one or more RF modes
Ranked causes
- 1. Triac (MT1 or MT2) failed in one mode (common)
- 2. Control loop failure (occasional)
Measure DC voltage in Cut mode
- Test point
- HVDC supply line
- Expected
- ≤205 VDC
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard switches not responding
Front panel switches do not change indicators or modes when pressed; no response to user input
Ranked causes
- 1. Front panel overlay switch failure (common)
- 2. Switch scanner circuit failure (occasional)
Disconnect J1 harness and apply jumper test
- Test point
- J1 connector pins 1-4 shorted to pins 5-8
- Expected
- Front panel indicators change state in response to jumper
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power in Spray or ABC
Unit outputs no power in Spray or ABC modes despite activation.
Ranked causes
- 1. Open power FET on SE Amp (common)
- 2. No RF Drive on SE Amp (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No RF output in any mode high voltage
System does not produce RF power output from any jack in Cut, Spray, or other modes
Ranked causes
- 1. Open relay in output circuit (common)
- 2. Relay driver circuit failed (common)
Test RF output from all jacks in Cut mode
- Test point
- Cut mode RF output jacks
- Expected
- RF signal present at output jacks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breaker trips when system is activated high voltage
Main power breaker trips immediately upon system activation or power-on
Ranked causes
- 1. Shorted power FET on RF amplifier (common)
Identify affected amplifier
- Test point
- System activation mode
- Expected
- FB amp (Cut/Pinpoint) or SE amp (Spray/ABC) identified
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: A.R.M. Max Voltage loading high voltage
Load A.R.M. maximum voltage value into EEPROM for Cut window second number
Ranked causes
- 1. A.R.M. Max Voltage value needs loading into EEPROM (common)
Verify S1 Position 1 switch setting and remove Patient Return Adaptor
- Test point
- Switch 1 (S1) Position 1 and Patient Return Connector
- Expected
- S1 Position 1 ON (up); Patient Return Adaptor removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC mode 150W high voltage
ABC mode 150W output power needs calibration adjustment
Ranked causes
- 1. ABC 150W output power out of calibration (common)
Measure ABC RF output power at 150W setting
- Test point
- Patient Return and ABC Output Connector
- Expected
- 150W ±2W, ±2W
- Tools
- ESU analyzer, 500 ohm load, ABC adaptor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC mode 15W high voltage
ABC mode 15W output power needs calibration adjustment
Ranked causes
- 1. ABC 15W output power out of calibration (common)
Measure ABC RF output power at 15W setting
- Test point
- Patient Return and ABC Output Connector
- Expected
- 15W ±2W, ±2W
- Tools
- ESU analyzer, 500 ohm load, ABC adaptor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC mode Active pulse width high voltage
Verify ABC Active mode pulse width modulation for power control
Ranked causes
- 1. ABC Active mode power output out of specification (unknown)
Measure ABC Active mode minimum pulse width
- Test point
- SE signal in Active mode
- Expected
- 200nS minimum µs
- Tools
- Oscilloscope
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC mode Target/Booster pulse characteristics high voltage
Verify ABC Target and Booster mode pulse sequences and timing without load
Ranked causes
- 1. ABC Target/Booster mode RF leakage elevated (unknown)
Measure ABC Target mode pulse width
- Test point
- SE signal with no handpiece load
- Expected
- 400nS µs
- Tools
- Oscilloscope
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC Over Voltage protection high voltage
ABC over-voltage limit setting to prevent output exceeding 3500Vpk
Ranked causes
- 1. ABC over-voltage clipping limit out of calibration (common)
Connect oscilloscope probe differentially to ABC and Patient Return outputs
- Test point
- ABC and Patient Return outputs
- Expected
- Oscilloscope configured for differential measurement with both channels at maximum amplitude
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC PW_MEA EEPROM loading at 11W high voltage
Load ABC pulse width measurement values at 11W into EEPROM for Coag window display
Ranked causes
- 1. ABC 11W PW_MEA value needs loading into EEPROM (common)
Verify S1 Position 1 switch setting
- Test point
- Back side of display panel, Switch 1 (S1), Position 1
- Expected
- ON (up)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABC PW_MEA EEPROM loading at 149W high voltage
Load ABC pulse width measurement values at 149W into EEPROM for Cut window display
Ranked causes
- 1. ABC 149W PW_MEA value needs loading into EEPROM (common)
Set and activate ABC at 149W
- Test point
- ABC power control
- Expected
- 149W
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bipolar RF output high voltage
Bipolar mode output power needs adjustment to meet 50W specification
Ranked causes
- 1. Bipolar RF output power out of calibration (common)
Measure Bipolar RF output power
- Test point
- Bipolar RF Output Jacks
- Expected
- 50W ±2W, ±2W
- Tools
- ESU analyzer, 100-125 ohm load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: dP (delta pressure) signal in standby
Verify differential pressure output is zero when unit is in standby mode
Ranked causes
- 1. dP signal not zero in standby (unknown)
Measure delta pressure output TP3 in standby mode
- Test point
- TP3
- Expected
- 0V
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: FMEA signal voltage
Verify argon back pressure monitoring signal FMEA is within operating range
Ranked causes
- 1. FMEA signal out of specification (unknown)
Measure FMEA voltage signal
- Test point
- display assembly input
- Expected
- 2V to 3.5V, +/-0.75 V
Calibration: FMON signal voltage at 4 lpm
Verify proportioning valve current monitor signal FMON is calibrated correctly
Ranked causes
- 1. FMON signal out of specification (unknown)
Measure FMON voltage at 4 lpm flow setting
- Test point
- display assembly input
- Expected
- 2V
- Tools
- DMM
Calibration: Full Bridge Amplifier drive waveforms high voltage
Verify H-bridge power switching operation and output waveform characteristics
Ranked causes
- 1. Output waveform not meeting frequency or timing specifications (unknown)
Measure drive frequency on 01 and 02 signals
- Test point
- U1 output to Q2 and U2 output
- Expected
- 461 KHz kHz
- Tools
- Oscilloscope
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Single-Ended Amplifier Spray mode high voltage
Verify Spray mode drive cycle timing and power control voltage
Ranked causes
- 1. Spray mode output power out of specification (unknown)
Measure Spray mode drive pulse repetition rate
- Test point
- SE signal from Power Control Assembly
- Expected
- 35µS repetition rate µs
- Tools
- Oscilloscope
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Spray Coag PW_MEA EEPROM loading high voltage
Load Spray Coag pulse width measurement values into EEPROM for RF Drive monitoring
Ranked causes
- 1. Spray Coag PW_MEA value needs loading into EEPROM (common)
Verify S1 Position 1 switch setting
- Test point
- Back side of display panel, Switch 1 (S1), Position 1
- Expected
- ON (up)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 200 displayed
Interrupt error - Unused interrupt was triggered unexpectedly during operation.
Ranked causes
- 1. Unused Interrupt tripped (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 201 displayed
Mailbox error - Monitor did not read message from Controller, indicating communication failure.
Ranked causes
- 1. Monitor did not read message on A5 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 202 displayed
Mailbox error - Monitor did not respond to Controller message, indicating communication failure.
Ranked causes
- 1. Monitor did not respond on A5 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 216 displayed
RAM Failure - External RAM memory on Display PCB has failed.
Ranked causes
- 1. External RAM U2 failed on A5 (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 218 displayed
Bad EPROM CRC - EPROM checksum validation failed indicating memory corruption or defective chip.
Ranked causes
- 1. EPROM U1 failed on A5 (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 219 displayed
CPU RAM Failed - Microcontroller internal RAM has failed preventing normal operation.
Ranked causes
- 1. Micro Internal RAM failed in U15 on A5 (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 220 displayed
CPU Failed - Main microprocessor on Display PCB has failed.
Ranked causes
- 1. Microprocessor U15 failed on A5 (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Err 251 displayed
EEPROM error - EEPROM read/write failure on Display PCB.
Ranked causes
- 1. Bad EEPROM or U27 on A5 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Err 26 displayed
- Err 303 displayed
- Err 305 displayed
- Err 309 displayed
- Err 310 displayed
- Err 311 displayed
- Err 312 displayed
- Err 314 displayed
- Err 318 displayed
- Err 319 displayed
- Err 320 displayed
- Err 321 displayed
- Err 324 displayed
- Err 325 displayed
- Err 326 displayed
- Err 327 displayed
- Err 328 displayed
- Err 329 displayed
- Err 330 displayed
- Err 41 displayed
- Err 42 displayed
- Err 43 displayed
- Err 44 displayed
- Err 55 displayed
- Fuses or R19 resistor blowing repeatedly
- Intermittent activation with handcontrol
- Spray mode will not deliver power
27 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
301 | Mode or control error with status flags | During ABC mode activation with low gas supply and other mode/status conditions active | warning |
386 | Tone monitor signal error or failure | T_MON signal from A5 board fails or goes invalid | error |
Err 1 | Stuck Front Panel Switch or Front Panel overlay failed or U5 microcontroller defective | Front panel switch stuck during power-up or overlay failure | error |
Err 111 | FMON out of tolerance - Flow Monitor out of calibration | Flow measurement outside acceptable range during operation | warning |
Err 112 | Flow Error - Flow measurement out of calibration or control failed | Flow rate abnormal or flow control circuit failure detected | warning |
Err 120 | Controller and Monitor disagree on power level | Data mismatch between Controller and Monitor microcontrollers on power setting | error |
Err 2 | Stuck Hand or Foot Activation Request | Activation switch stuck during power-up or defective handcontrol/footcontrol | error |
Err 20 | Illegal Default Switch Setting - position 1 or 6 is ON | Unit powered up with incorrect default switch configuration | critical |
Err 200 | Interrupt - Unused Interrupt tripped | Unexpected interrupt signal detected during operation | error |
Err 201 | Mailbox - Monitor did not read message | Communication timeout between Controller and Monitor | warning |
Err 202 | Mailbox - Monitor did not respond | Communication timeout from Monitor to Controller | warning |
Err 203 | Mailbox error - Bad data transfer or mailbox failed | Mailbox communication protocol error detected | warning |
Err 210 | Bad Cal Data - Calibration data invalid or corrupted | Stored calibration data fails validation during startup | warning |
Err 216 | RAM Failure - External RAM has failed | Memory test failure detected during power-on or operation | critical |
Err 218 | Bad EPROM CRC - EPROM checksum validation failed | EPROM data integrity check failed during startup | critical |
Err 219 | CPU RAM Failed - Microcontroller internal RAM failed | CPU RAM integrity check failed during startup or operation | critical |
Err 220 | CPU Failed - Microprocessor failed | CPU test or operation failure detected | critical |
Err 233 | Data transfer to Monitor failed | Communication error between Controller and Monitor microcontrollers | error |
Err 251 | EEPROM error - Read/Write failure | EEPROM read or write operation failed | error |
Err 252 | EEPROM error - CRC validation failed | EEPROM data integrity check failed | warning |
Err 26 | Stuck encoder bits or Logic device failed | Encoder input stuck or logic device failure detected | error |
Err 301 | Flow Error - Flow measurement out of tolerance | Flow measurement too high or low during operation | warning |
Err 302 | IMEA Error - Output Current measurement error too high | Output current exceeds tolerance during operation | warning |
Err 303 | Activation Error - Controller and Monitor disagree on activation | Mismatch in activation request or mode between microcontrollers | error |
Err 304 | Output Power - Power control circuit failed | Power control regulation failure detected | error |
Err 305 | Activation Error - Monitor activation request failed | Monitor cannot process activation request from Controller | error |
Err 306 | PW_MEA out of calibration - Power measurement out of tolerance | Power measurement fails calibration tolerance check | warning |
Err 307 | FB Amp Drive Error - Feedback amplifier drive circuit failed | Feedback amplifier drive signal abnormal or absent | error |
Err 308 | HVDC too high - High voltage DC supply out of tolerance (high) | High voltage DC supply exceeds 200V limit | critical |
Err 309 | FBMON idle limit too high - Feedback monitor idle voltage excessive | Feedback monitor idle voltage exceeds 155mV threshold | warning |
Err 310 | 10V Reference - 10V reference voltage out of tolerance | 10V reference supply fails tolerance check on Display PCB | error |
Err 311 | 15V supply - 15V power supply out of tolerance | 15V power supply fails tolerance check | error |
Err 312 | PW_MEA idle - Power measurement idle error | Power measurement fails idle state validation | warning |
Err 313 | IMEA idle - Current measurement idle error too high | Current measurement fails idle state validation | warning |
Err 314 | Stuck Activation Request - Activation signal remains asserted | Activation request signal stuck in active state | error |
Err 315 | Flow Idle Error - Flow control not properly inhibited at idle | Flow measurement exceeds 150mV during idle state | warning |
Err 316 | HV Idle Test - High voltage not properly inhibited at idle | High voltage monitor or supply circuit failure at idle state | warning |
Err 318 | Micro CRC - Monitor microcontroller checksum validation failed | Monitor microcontroller firmware integrity check failed | critical |
Err 319 | Micro RAM - Monitor microcontroller internal RAM failed | Monitor microcontroller RAM integrity check failed | critical |
Err 320 | Micro Failed - Monitor microcontroller failed | Monitor microcontroller failure detected | critical |
Err 321 | Test Switch - Unit in TEST MODE instead of RUN MODE | Default switch configuration indicates TEST MODE during operation | warning |
Err 323 | FB Amp Drive Error - Feedback amplifier drive circuit failed | Feedback amplifier drive signal abnormal or absent | error |
Err 324 | FB Amp not inhibited - Feedback amplifier not disabled during POST | RF feedback amplifier failed to inhibit during power-on self-test | error |
Err 325 | Gas Flow not inhibited - Solenoid valve failed to turn off during POST | Gas solenoid valve remains active during power-on self-test | error |
Err 326 | Gas Flow not inhibited by Controller - Flow detected when solenoid off | Monitor detected FMON > 150mV when solenoid should be off during POST | error |
Err 327 | SE Amp not in range - Single-ended amp output power out of tolerance at POST | Single-ended RF amplifier power measurement fails tolerance during power-on self-test | error |
Err 328 | SE Amp not inhibited by Monitor - RF drive failed to terminate during POST | Single-ended RF amplifier failed to shut down during power-on self-test | error |
Err 329 | ABC Over Voltage Test failed - ABC protection circuit failed during POST | ABC over-voltage protection circuit test failure during power-on self-test | error |
Err 330 | ABC Over Voltage Test failed - Protection or RF drive activation failed at POST | ABC over-voltage protection circuit test or RF drive activation failed during power-on self-test | error |
Err 331 | HV out of range - High voltage DC supply exceeds tolerance at POST | High voltage DC supply too high during power-on self-test | critical |
Err 332 | HV not inhibited by monitor - High voltage failed to inhibit during POST | High voltage DC supply failed to inhibit during power-on self-test | critical |
Err 41 | Calibration Tolerance Error - ARMCAL_10 too low | ARMCAL_10 measurement fails calibration tolerance check (too low) | warning |
Err 42 | Calibration Tolerance Error - ARMCAL_10 too high | ARMCAL_10 measurement fails calibration tolerance check (too high) | warning |
Err 43 | Calibration Tolerance Error - ARMCAL_150 too low | ARMCAL_150 measurement fails calibration tolerance check (too low) | warning |
Err 44 | Calibration Tolerance Error - armcal_150 too high | armcal_150 measurement fails calibration tolerance check (too high) | warning |
Err 55 | Watchdog Reset - Timing Error in U15 | Watchdog timer detected timing error in microcontroller | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.