ConMed 7550: troubleshooting and error codes

The ServiceForge knowledge base holds 87 documented symptoms for the ConMed 7550, with 149 ranked causes, 67 diagnostic checks with test points and expected values, and 56 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Err 306 displayed

PW_MEA out of calibration - Power measurement is out of tolerance.

Ranked causes

  1. 1. U7 on Power Control Failed on A4 (occasional)
  2. 2. Ramp transistors Q1 or Q2 failed (occasional)
  3. 3. U14B on A4 Failed (occasional)
  4. 4. U21 - Port A wrong data on A4 and A5 (occasional)
  5. 5. D/A Failed - PCON on A5 (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 302 displayed

IMEA Error - Output current measurement is out of tolerance or power regulation has failed.

Ranked causes

  1. 1. Output Current Error too high on A4 (common)
  2. 2. Too much power into a heavy load (occasional)
  3. 3. Power Regulation Loop failed on A4 (occasional)
  4. 4. D/A Failed - PCON on A5 (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 316 displayed high voltage

HV Idle Test - High voltage monitor or supply circuits not properly in idle state.

Ranked causes

  1. 1. HVMON circuit failed on A1 (occasional)
  2. 2. 10V circuit on A7 Failed (occasional)
  3. 3. J3 disconnected on A7 (occasional)
  4. 4. J1 disconnect on A1 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ABC mode will not initiate or activate

ABC (automatic blend control) mode fails to start or respond to activation request

Ranked causes

  1. 1. Over-voltage circuit active or over-voltage threshold exceeded (common)
  2. 2. U7 failed or pulse width/duty cycle error (occasional)
  3. 3. U4 control signal failure (rare)
First check

Recalibrate ABC over-voltage

Test point
ABC mode control circuit
Expected
Over-voltage threshold reset; ABC mode initiates normally

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dual Pad monitor set will not lock patient safety

Dual Pad electrode monitoring mode fails to lock or maintain connection with monitor set

Ranked causes

  1. 1. PP receptacle resistance too high (>10 ohms) (common)
  2. 2. Monitor set switch on overlay failed (occasional)
  3. 3. A.R.M. (Automatic Return Monitor) calibration error (rare)
First check

Measure PP receptacle resistance

Test point
PP receptacle connector
Expected
>12 ohms, > 12 ohms

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 1 displayed

Front panel switch stuck or defective during power-up; Front Panel overlay failed or U5 microcontroller defective.

Ranked causes

  1. 1. User pressing switch during power-up (common)
  2. 2. Front Panel overlay failed (occasional)
  3. 3. U5 (A5) microcontroller failed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 111 displayed

FMON (Flow Monitor) out of tolerance - Flow control circuit or measurement is out of calibration.

Ranked causes

  1. 1. FMON out of calibration on A1 (common)
  2. 2. Flow Control Valve Driver shorted on A1 (occasional)
  3. 3. Argon tubes to X1 reversed on A1 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 112 displayed

Flow Error - Flow measurement or control is out of calibration or circuit has failed.

Ranked causes

  1. 1. FMON out of Calibration on A1 (common)
  2. 2. Handpiece occluded (common)
  3. 3. Flow control circuit has failed (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 2 displayed

Hand or foot activation switch stuck or failed accessory (handcontrol or footcontrol) preventing normal operation.

Ranked causes

  1. 1. User pressing switch during power-up (common)
  2. 2. Failed Handcontrol or Footcontrol (common)
  3. 3. Failed component on A5 or A11 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 233 displayed

Data transfer to Monitor failed - Communication between Controller and Monitor microcontrollers has failed.

Ranked causes

  1. 1. Mailbox Failed on A5 (occasional)
  2. 2. Data Bus bad on A5 (occasional)
  3. 3. Address Bus bad on A5 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 301 displayed

Flow Error - Flow measurement is too high or low, or D/A converter failure affecting gas flow control.

Ranked causes

  1. 1. Flow too high or low (common)
  2. 2. Flow Rate too high for accessory (common)
  3. 3. D/A Failed - VGAS on A5 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 307 displayed

FB Amp Drive Error - Feedback amplifier drive circuit failed on Power Control PCB.

Ranked causes

  1. 1. U7 on A4 Failed (occasional)
  2. 2. U14A on A4 Failed (occasional)
  3. 3. U21 - Port A wrong data on A4 and A5 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 308 displayed high voltage

HVDC too high - High voltage control circuit or monitoring circuit has failed.

Ranked causes

  1. 1. HV Control Loop failed on A1 (occasional)
  2. 2. HV Monitor Failed on A1 (occasional)
  3. 3. D/A Failed - VCON on A5 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 315 displayed

Flow Idle Error - Flow control circuit not properly inhibited during idle state.

Ranked causes

  1. 1. Flow Control Loop Failed on A1 (occasional)
  2. 2. Driver Q2 Failed on A1 (occasional)
  3. 3. U15B Failed on A1 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 323 displayed

FB Amp Drive Error - Feedback amplifier drive circuit failure on Power Control PCB.

Ranked causes

  1. 1. U7 on A4 Failed (occasional)
  2. 2. U14A on A4 Failed (occasional)
  3. 3. U21 - Port A wrong data on A4 and A5 (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow rate too high caution

Argon gas flow rate exceeds expected rate or control setting on display

Ranked causes

  1. 1. Driver Q2 shorted on A1 board (occasional)
  2. 2. Argon pressure transducer hoses reversed (occasional)
  3. 3. Control loop failure on A1 board (rare)
First check

Measure Q2 emitter voltage during flow operation

Test point
Q2 emitter pin (A1 board)
Expected
Voltage increases progressively with increasing flow rate

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Handsense control not responding or intermittent patient safety

Handcontrol switch activation does not register or registers intermittently; control signals not reaching RF circuits

Ranked causes

  1. 1. Continuity transformer T3 primary drive signal loss (occasional)
  2. 2. Opto-isolator LED failure in handswitch detection circuit (occasional)
  3. 3. Handswitch mechanical contact failure or corrosion (common)
First check

Measure NE555 timer U3 drive signal on T3 primary

Test point
U3 output pin; T3 primary winding
Expected
100 KHz square wave, 20% duty cycle
Tools
Oscilloscope

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low or zero RF output power high voltage

RF Output Assembly fails to produce expected power across Cut, Blend, Pinpoint, Bipolar or Spray modes

Ranked causes

  1. 1. Faulty relay in RF output path (common)
  2. 2. Relay driver failure on Display Assembly (occasional)
  3. 3. Open circuit or disconnection in output relay switching path (occasional)
First check

Verify relay driver is providing drive signal

Test point
Relay driver output on Display Assembly
Expected
Drive signal present for activated mode

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Activation or Tone

Unit will not activate or produce tone when handcontrol or footcontrol is pressed.

Ranked causes

  1. 1. Activation agreement error between Controller and Monitor (common)
  2. 2. Handsense Circuit failed (common)
  3. 3. RF Jack worn out (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pneumatic flow control or solenoid valve not responding caution

Argon gas flow does not modulate or solenoid valve fails to open/close; pneumatic manifold not functioning

Ranked causes

  1. 1. Solenoid valve coil open or shorted (occasional)
  2. 2. Solenoid valve driver transistor failure or open base/collector (occasional)
  3. 3. Proportioning valve controller current not increasing with flow setting (occasional)
First check

Measure solenoid valve coil resistance

Test point
Digital (solenoid) valve coil terminals
Expected
72 ohms
Tools
DMM ohmmeter

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Regulator thermal shutdown activated on 15 V or 5 V supply caution

System loses 15 V or 5 V power due to regulator overheating; load current exceeds safe operating range

Ranked causes

  1. 1. Shorted component on downstream assembly drawing excessive current (common)
  2. 2. Regulator Q1, Q2, or Q5 internal short or degradation (occasional)
  3. 3. 5 V switcher LC filter capacitor failure or degradation (occasional)
First check

Measure regulator output current with all loads connected

Test point
Regulator output line or dedicated current sense resistor
Expected
15 V regulator: <500 mA typical; 5 V switcher: <2 A typical (per schematic)
Tools
DMM or power supply with current monitoring

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

RF leakage detected or excessive RF leakage during operation caution

System detects unbalanced RF current returning via alternate paths; leakage current exceeds balanced condition threshold

Ranked causes

  1. 1. Unbalanced condition in Balun transformer T4 (common)
  2. 2. Damaged or faulty Balun sensing winding (occasional)
  3. 3. Patient return isolation relay failure (K8, K9, K11) (occasional)
First check

Measure balun sense winding voltage

Test point
Balun sense winding
Expected
0 V under balanced condition; voltage increases proportional to leakage current

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 120 displayed

Controller and Monitor microcontrollers disagree on power level, indicating data communication or encoder logic failure.

Ranked causes

  1. 1. Bad data transfer between Controller and Monitor on A5 (occasional)
  2. 2. PIA with encoder logic failed on A5 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 20 displayed caution

Illegal Default Switch Setting - unit is in TEST MODE instead of RUN MODE, or default switch position 1 or 6 is ON.

Ranked causes

  1. 1. Default switch position 1 or 6 is ON (common)
  2. 2. System in TEST MODE (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 203 displayed

Mailbox error - Bad data transfer or mailbox communication failure between Controller and Monitor.

Ranked causes

  1. 1. Bad data transfer on A5 (occasional)
  2. 2. Mailbox failed on A5 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 210 displayed

Bad Calibration Data - Calibration data stored in EEPROM is invalid or corrupted.

Ranked causes

  1. 1. Calibration Data bad on A5 (common)
  2. 2. EEPROM failed on A5 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 252 displayed

EEPROM error - CRC validation of EEPROM data failed or data is corrupted.

Ranked causes

  1. 1. CRC error on A5 (common)
  2. 2. Bad data stored in EEPROM on A5 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 304 displayed

Output Power - Power control circuit or D/A converter failure preventing proper power regulation.

Ranked causes

  1. 1. Power Control Circuit Failed on A4 (occasional)
  2. 2. D/A Failed - PCON on A5 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 313 displayed

IMEA idle - Current measurement idle error indicating improper idle state voltage.

Ranked causes

  1. 1. U14D Failed on A4 (occasional)
  2. 2. U12 or U8 Failed on A5 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 331 displayed high voltage

HV out of range - High voltage DC supply exceeds tolerance during power-on self-test.

Ranked causes

  1. 1. HVDC too high during POST (occasional)
  2. 2. Triac Failure on A7 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 332 displayed high voltage

HV not inhibited by monitor - High voltage control failed to inhibit during power-on self-test.

Ranked causes

  1. 1. HVINH failed to inhibit HVDC during POST (occasional)
  2. 2. Fuse on HV Power Supply open or missing (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Half Power in Cut, Blend, Pinpoint, Bipolar

Unit outputs half power in Cut, Blend, Pinpoint, or Bipolar modes instead of full power.

Ranked causes

  1. 1. A phase of the Full Bridge Amp failed (common)
  2. 2. Open power FET or no RF Drive to power FET (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High voltage supply exceeds 205 VDC high voltage

DC voltage supply reads above 205 VDC in one or more RF modes

Ranked causes

  1. 1. Triac (MT1 or MT2) failed in one mode (common)
  2. 2. Control loop failure (occasional)
First check

Measure DC voltage in Cut mode

Test point
HVDC supply line
Expected
≤205 VDC

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard switches not responding

Front panel switches do not change indicators or modes when pressed; no response to user input

Ranked causes

  1. 1. Front panel overlay switch failure (common)
  2. 2. Switch scanner circuit failure (occasional)
First check

Disconnect J1 harness and apply jumper test

Test point
J1 connector pins 1-4 shorted to pins 5-8
Expected
Front panel indicators change state in response to jumper

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power in Spray or ABC

Unit outputs no power in Spray or ABC modes despite activation.

Ranked causes

  1. 1. Open power FET on SE Amp (common)
  2. 2. No RF Drive on SE Amp (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No RF output in any mode high voltage

System does not produce RF power output from any jack in Cut, Spray, or other modes

Ranked causes

  1. 1. Open relay in output circuit (common)
  2. 2. Relay driver circuit failed (common)
First check

Test RF output from all jacks in Cut mode

Test point
Cut mode RF output jacks
Expected
RF signal present at output jacks

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breaker trips when system is activated high voltage

Main power breaker trips immediately upon system activation or power-on

Ranked causes

  1. 1. Shorted power FET on RF amplifier (common)
First check

Identify affected amplifier

Test point
System activation mode
Expected
FB amp (Cut/Pinpoint) or SE amp (Spray/ABC) identified

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: A.R.M. Max Voltage loading high voltage

Load A.R.M. maximum voltage value into EEPROM for Cut window second number

Ranked causes

  1. 1. A.R.M. Max Voltage value needs loading into EEPROM (common)
First check

Verify S1 Position 1 switch setting and remove Patient Return Adaptor

Test point
Switch 1 (S1) Position 1 and Patient Return Connector
Expected
S1 Position 1 ON (up); Patient Return Adaptor removed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC mode 150W high voltage

ABC mode 150W output power needs calibration adjustment

Ranked causes

  1. 1. ABC 150W output power out of calibration (common)
First check

Measure ABC RF output power at 150W setting

Test point
Patient Return and ABC Output Connector
Expected
150W ±2W, ±2W
Tools
ESU analyzer, 500 ohm load, ABC adaptor

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC mode 15W high voltage

ABC mode 15W output power needs calibration adjustment

Ranked causes

  1. 1. ABC 15W output power out of calibration (common)
First check

Measure ABC RF output power at 15W setting

Test point
Patient Return and ABC Output Connector
Expected
15W ±2W, ±2W
Tools
ESU analyzer, 500 ohm load, ABC adaptor

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC mode Active pulse width high voltage

Verify ABC Active mode pulse width modulation for power control

Ranked causes

  1. 1. ABC Active mode power output out of specification (unknown)
First check

Measure ABC Active mode minimum pulse width

Test point
SE signal in Active mode
Expected
200nS minimum µs
Tools
Oscilloscope

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC mode Target/Booster pulse characteristics high voltage

Verify ABC Target and Booster mode pulse sequences and timing without load

Ranked causes

  1. 1. ABC Target/Booster mode RF leakage elevated (unknown)
First check

Measure ABC Target mode pulse width

Test point
SE signal with no handpiece load
Expected
400nS µs
Tools
Oscilloscope

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC Over Voltage protection high voltage

ABC over-voltage limit setting to prevent output exceeding 3500Vpk

Ranked causes

  1. 1. ABC over-voltage clipping limit out of calibration (common)
First check

Connect oscilloscope probe differentially to ABC and Patient Return outputs

Test point
ABC and Patient Return outputs
Expected
Oscilloscope configured for differential measurement with both channels at maximum amplitude

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC PW_MEA EEPROM loading at 11W high voltage

Load ABC pulse width measurement values at 11W into EEPROM for Coag window display

Ranked causes

  1. 1. ABC 11W PW_MEA value needs loading into EEPROM (common)
First check

Verify S1 Position 1 switch setting

Test point
Back side of display panel, Switch 1 (S1), Position 1
Expected
ON (up)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ABC PW_MEA EEPROM loading at 149W high voltage

Load ABC pulse width measurement values at 149W into EEPROM for Cut window display

Ranked causes

  1. 1. ABC 149W PW_MEA value needs loading into EEPROM (common)
First check

Set and activate ABC at 149W

Test point
ABC power control
Expected
149W

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bipolar RF output high voltage

Bipolar mode output power needs adjustment to meet 50W specification

Ranked causes

  1. 1. Bipolar RF output power out of calibration (common)
First check

Measure Bipolar RF output power

Test point
Bipolar RF Output Jacks
Expected
50W ±2W, ±2W
Tools
ESU analyzer, 100-125 ohm load

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: dP (delta pressure) signal in standby

Verify differential pressure output is zero when unit is in standby mode

Ranked causes

  1. 1. dP signal not zero in standby (unknown)
First check

Measure delta pressure output TP3 in standby mode

Test point
TP3
Expected
0V

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: FMEA signal voltage

Verify argon back pressure monitoring signal FMEA is within operating range

Ranked causes

  1. 1. FMEA signal out of specification (unknown)
First check

Measure FMEA voltage signal

Test point
display assembly input
Expected
2V to 3.5V, +/-0.75 V

Calibration: FMON signal voltage at 4 lpm

Verify proportioning valve current monitor signal FMON is calibrated correctly

Ranked causes

  1. 1. FMON signal out of specification (unknown)
First check

Measure FMON voltage at 4 lpm flow setting

Test point
display assembly input
Expected
2V
Tools
DMM

Calibration: Full Bridge Amplifier drive waveforms high voltage

Verify H-bridge power switching operation and output waveform characteristics

Ranked causes

  1. 1. Output waveform not meeting frequency or timing specifications (unknown)
First check

Measure drive frequency on 01 and 02 signals

Test point
U1 output to Q2 and U2 output
Expected
461 KHz kHz
Tools
Oscilloscope

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Single-Ended Amplifier Spray mode high voltage

Verify Spray mode drive cycle timing and power control voltage

Ranked causes

  1. 1. Spray mode output power out of specification (unknown)
First check

Measure Spray mode drive pulse repetition rate

Test point
SE signal from Power Control Assembly
Expected
35µS repetition rate µs
Tools
Oscilloscope

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Spray Coag PW_MEA EEPROM loading high voltage

Load Spray Coag pulse width measurement values into EEPROM for RF Drive monitoring

Ranked causes

  1. 1. Spray Coag PW_MEA value needs loading into EEPROM (common)
First check

Verify S1 Position 1 switch setting

Test point
Back side of display panel, Switch 1 (S1), Position 1
Expected
ON (up)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 200 displayed

Interrupt error - Unused interrupt was triggered unexpectedly during operation.

Ranked causes

  1. 1. Unused Interrupt tripped (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 201 displayed

Mailbox error - Monitor did not read message from Controller, indicating communication failure.

Ranked causes

  1. 1. Monitor did not read message on A5 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 202 displayed

Mailbox error - Monitor did not respond to Controller message, indicating communication failure.

Ranked causes

  1. 1. Monitor did not respond on A5 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 216 displayed

RAM Failure - External RAM memory on Display PCB has failed.

Ranked causes

  1. 1. External RAM U2 failed on A5 (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 218 displayed

Bad EPROM CRC - EPROM checksum validation failed indicating memory corruption or defective chip.

Ranked causes

  1. 1. EPROM U1 failed on A5 (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 219 displayed

CPU RAM Failed - Microcontroller internal RAM has failed preventing normal operation.

Ranked causes

  1. 1. Micro Internal RAM failed in U15 on A5 (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 220 displayed

CPU Failed - Main microprocessor on Display PCB has failed.

Ranked causes

  1. 1. Microprocessor U15 failed on A5 (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Err 251 displayed

EEPROM error - EEPROM read/write failure on Display PCB.

Ranked causes

  1. 1. Bad EEPROM or U27 on A5 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Err 26 displayed
  • Err 303 displayed
  • Err 305 displayed
  • Err 309 displayed
  • Err 310 displayed
  • Err 311 displayed
  • Err 312 displayed
  • Err 314 displayed
  • Err 318 displayed
  • Err 319 displayed
  • Err 320 displayed
  • Err 321 displayed
  • Err 324 displayed
  • Err 325 displayed
  • Err 326 displayed
  • Err 327 displayed
  • Err 328 displayed
  • Err 329 displayed
  • Err 330 displayed
  • Err 41 displayed
  • Err 42 displayed
  • Err 43 displayed
  • Err 44 displayed
  • Err 55 displayed
  • Fuses or R19 resistor blowing repeatedly
  • Intermittent activation with handcontrol
  • Spray mode will not deliver power

27 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
301Mode or control error with status flagsDuring ABC mode activation with low gas supply and other mode/status conditions activewarning
386Tone monitor signal error or failureT_MON signal from A5 board fails or goes invaliderror
Err 1Stuck Front Panel Switch or Front Panel overlay failed or U5 microcontroller defectiveFront panel switch stuck during power-up or overlay failureerror
Err 111FMON out of tolerance - Flow Monitor out of calibrationFlow measurement outside acceptable range during operationwarning
Err 112Flow Error - Flow measurement out of calibration or control failedFlow rate abnormal or flow control circuit failure detectedwarning
Err 120Controller and Monitor disagree on power levelData mismatch between Controller and Monitor microcontrollers on power settingerror
Err 2Stuck Hand or Foot Activation RequestActivation switch stuck during power-up or defective handcontrol/footcontrolerror
Err 20Illegal Default Switch Setting - position 1 or 6 is ONUnit powered up with incorrect default switch configurationcritical
Err 200Interrupt - Unused Interrupt trippedUnexpected interrupt signal detected during operationerror
Err 201Mailbox - Monitor did not read messageCommunication timeout between Controller and Monitorwarning
Err 202Mailbox - Monitor did not respondCommunication timeout from Monitor to Controllerwarning
Err 203Mailbox error - Bad data transfer or mailbox failedMailbox communication protocol error detectedwarning
Err 210Bad Cal Data - Calibration data invalid or corruptedStored calibration data fails validation during startupwarning
Err 216RAM Failure - External RAM has failedMemory test failure detected during power-on or operationcritical
Err 218Bad EPROM CRC - EPROM checksum validation failedEPROM data integrity check failed during startupcritical
Err 219CPU RAM Failed - Microcontroller internal RAM failedCPU RAM integrity check failed during startup or operationcritical
Err 220CPU Failed - Microprocessor failedCPU test or operation failure detectedcritical
Err 233Data transfer to Monitor failedCommunication error between Controller and Monitor microcontrollerserror
Err 251EEPROM error - Read/Write failureEEPROM read or write operation failederror
Err 252EEPROM error - CRC validation failedEEPROM data integrity check failedwarning
Err 26Stuck encoder bits or Logic device failedEncoder input stuck or logic device failure detectederror
Err 301Flow Error - Flow measurement out of toleranceFlow measurement too high or low during operationwarning
Err 302IMEA Error - Output Current measurement error too highOutput current exceeds tolerance during operationwarning
Err 303Activation Error - Controller and Monitor disagree on activationMismatch in activation request or mode between microcontrollerserror
Err 304Output Power - Power control circuit failedPower control regulation failure detectederror
Err 305Activation Error - Monitor activation request failedMonitor cannot process activation request from Controllererror
Err 306PW_MEA out of calibration - Power measurement out of tolerancePower measurement fails calibration tolerance checkwarning
Err 307FB Amp Drive Error - Feedback amplifier drive circuit failedFeedback amplifier drive signal abnormal or absenterror
Err 308HVDC too high - High voltage DC supply out of tolerance (high)High voltage DC supply exceeds 200V limitcritical
Err 309FBMON idle limit too high - Feedback monitor idle voltage excessiveFeedback monitor idle voltage exceeds 155mV thresholdwarning
Err 31010V Reference - 10V reference voltage out of tolerance10V reference supply fails tolerance check on Display PCBerror
Err 31115V supply - 15V power supply out of tolerance15V power supply fails tolerance checkerror
Err 312PW_MEA idle - Power measurement idle errorPower measurement fails idle state validationwarning
Err 313IMEA idle - Current measurement idle error too highCurrent measurement fails idle state validationwarning
Err 314Stuck Activation Request - Activation signal remains assertedActivation request signal stuck in active stateerror
Err 315Flow Idle Error - Flow control not properly inhibited at idleFlow measurement exceeds 150mV during idle statewarning
Err 316HV Idle Test - High voltage not properly inhibited at idleHigh voltage monitor or supply circuit failure at idle statewarning
Err 318Micro CRC - Monitor microcontroller checksum validation failedMonitor microcontroller firmware integrity check failedcritical
Err 319Micro RAM - Monitor microcontroller internal RAM failedMonitor microcontroller RAM integrity check failedcritical
Err 320Micro Failed - Monitor microcontroller failedMonitor microcontroller failure detectedcritical
Err 321Test Switch - Unit in TEST MODE instead of RUN MODEDefault switch configuration indicates TEST MODE during operationwarning
Err 323FB Amp Drive Error - Feedback amplifier drive circuit failedFeedback amplifier drive signal abnormal or absenterror
Err 324FB Amp not inhibited - Feedback amplifier not disabled during POSTRF feedback amplifier failed to inhibit during power-on self-testerror
Err 325Gas Flow not inhibited - Solenoid valve failed to turn off during POSTGas solenoid valve remains active during power-on self-testerror
Err 326Gas Flow not inhibited by Controller - Flow detected when solenoid offMonitor detected FMON > 150mV when solenoid should be off during POSTerror
Err 327SE Amp not in range - Single-ended amp output power out of tolerance at POSTSingle-ended RF amplifier power measurement fails tolerance during power-on self-testerror
Err 328SE Amp not inhibited by Monitor - RF drive failed to terminate during POSTSingle-ended RF amplifier failed to shut down during power-on self-testerror
Err 329ABC Over Voltage Test failed - ABC protection circuit failed during POSTABC over-voltage protection circuit test failure during power-on self-testerror
Err 330ABC Over Voltage Test failed - Protection or RF drive activation failed at POSTABC over-voltage protection circuit test or RF drive activation failed during power-on self-testerror
Err 331HV out of range - High voltage DC supply exceeds tolerance at POSTHigh voltage DC supply too high during power-on self-testcritical
Err 332HV not inhibited by monitor - High voltage failed to inhibit during POSTHigh voltage DC supply failed to inhibit during power-on self-testcritical
Err 41Calibration Tolerance Error - ARMCAL_10 too lowARMCAL_10 measurement fails calibration tolerance check (too low)warning
Err 42Calibration Tolerance Error - ARMCAL_10 too highARMCAL_10 measurement fails calibration tolerance check (too high)warning
Err 43Calibration Tolerance Error - ARMCAL_150 too lowARMCAL_150 measurement fails calibration tolerance check (too low)warning
Err 44Calibration Tolerance Error - armcal_150 too higharmcal_150 measurement fails calibration tolerance check (too high)warning
Err 55Watchdog Reset - Timing Error in U15Watchdog timer detected timing error in microcontrollererror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.